Abstract
Ambulatory electrocardiogram (AECG) monitors are frequently used to evaluate arrhythmias and assess symptom-rhythm correlation. The extended use of AECG monitors is limited by complications, such as allergic contact dermatitis, secondary to the adhesive. We present the case of a 70-year-old woman with newly diagnosed paroxysmal atrial fibrillation and a known adhesive allergy who developed an allergic contact dermatitis shortly after the application of an AECG monitor. After her symptoms were managed with a topical corticosteroid, fluticasone propionate nasal spray was used as a novel, off-label prophylaxis before the second application of the AECG monitor. The patient then tolerated the full 14-day study without issues, highlighting the potential of aerosolized corticosteroids as a prophylaxis for managing adhesive intolerance in patients requiring long-term AECG monitoring.
Keywords: ambulatory electrocardiography devices, atrial fibrillation (af), contact allergic dermatitis, fluticasone, remote telemetry
Introduction
Ambulatory electrocardiogram (AECG) monitors are a commonly used non-invasive modality for the evaluation of cardiac symptoms and are used to identify the presence of arrhythmias, quantify disease burden, and establish the presence of symptom-rhythm correlation [1,2]. To fully elucidate the burden of disease, patients are often prescribed these devices for weeks at a time [3-6]. Adhesive intolerance, most commonly in the form of allergic contact dermatitis (ACD), is a significant complication for patients requiring long-term AECG monitoring. Despite attempts to reduce the risk of allergic reactions and improve patient adherence [7], a recent screening study for atrial fibrillation using AECG revealed that 37.9% of users reported limited adverse skin reactions, with 1.2% of patients discontinuing the study due to intolerable side effects [8]. The findings reflect a broader issue inherent to medical adhesives overall [9-11]. While the management strategy of ACD involves allergen avoidance, skin protection, and the use of corticosteroids, there are limited studies that use corticosteroids as a method of prevention for specific medical adhesives. Paret et al. led a study using intranasal fluticasone propionate for the prevention of hypersensitivity to the use of a continuous glucometer [12]; thus, this method was applied for our patient.
We present a case of a patient with a vague history of adhesive allergy that initially prohibited the use of a current-generation AECG monitor. The patient was subsequently pretreated with aerosolized corticosteroid spray at the site of the intended AECG application, which allowed the patient to successfully complete the study.
Case presentation
Our patient is a 70-year-old woman who presented to the clinic following a recent emergency department visit where she was first diagnosed with atrial fibrillation.
On the day of diagnosis, she was minimally symptomatic, reporting only a vague sense of unease that prompted her evaluation. While on telemetry in the emergency department, she was found to be in atrial fibrillation and spontaneously converted to normal sinus rhythm without intervention. She was discharged with systemic anticoagulation (apixaban 5 mg twice daily), a rate-controlling medication (metoprolol succinate 50 mg daily), and instructions to follow up with her cardiologist.
Following discharge, the patient reported no further symptoms. To better assess her atrial fibrillation burden and identify any subtle symptom-rhythm correlation, she was discharged from the clinic with a 14-day Zio AECG monitor (Figure 1). Within 24 hours, the patient developed intolerable pruritus, erythema, and swelling at the site of AECG monitor placement on the left chest. The device and adhesive were removed, and she was treated with topical hydrocortisone 0.5% cream daily for one week, resulting in complete resolution of her skin reaction. She was not referred to Dermatology for patch testing.
Figure 1. Zio ambulatory electrocardiogram (AECG) monitor overlying affixed to the patient's skin.
At her follow-up visit, before reapplication of the AECG monitor, two pumps of intranasal fluticasone propionate 50 mcg spray were applied directly to the uncompromised, fully healed skin of intended AECG placement and allowed to dry. The AECG monitor was then reapplied, and the patient completed a full 14-day monitoring period without recurrence of irritation. The patient's skin was self-monitored during the study (Figure 2).
Figure 2. Skin demonstrating allergic contact dermatitis (ACD) following removal of the Zio ambulatory electrocardiogram (AECG) monitor.
Discussion
AECG monitors are useful not only in diagnosing arrhythmias but also in establishing diagnosis, arrhythmia burden, and symptom-rhythm correlation. Guidelines recommend AECG monitoring as a useful diagnostic modality for patients with paroxysmal atrial fibrillation, particularly if episodes or symptoms are infrequent or not captured in the clinic or hospital setting on a 12-lead ECG [1,2]. Extending the wear time of AECG monitors increases the probability of capturing evidence of atrial fibrillation; as such, patch-based AECG monitors are designed to provide continuous telemetry monitoring for up to two weeks [3-6]. Current generation models are designed to reduce the risk of allergic reactions and improve patient adherence [7].
ACD is a type IV hypersensitivity reaction, which is a cell-mediated reaction from the interaction between sensitized T lymphocytes and specific antigens, which leads to an inflammatory reaction from the release of cytokines from CD4+ and CD8+ cells 24 to 72 hours after antigen exposure, ultimately presenting as a delayed-type reaction. The prevalence of ACD has been estimated at around 20% of the general population, according to a systematic review and meta-analysis of studies published between 2007 and 2017 [13]. ACD has been increasingly reported in patients using various types of medical adhesives, including, but not limited to, tapes, dressings, skin adhesives/suture glue, and adhesive devices [9-11]. Although many ingredients, such as acrylates and colophony, have been cited as common allergens in medical adhesive devices causing ACD, identifying the exact culprit antigen may not always be feasible due to a variety of supplemental antigens and materials used in allergens, possible selection bias in patient referral population for patch testing, and user-dependent interpretation of patch testing [10,11,14].
The initial management strategy of ACD involves allergen avoidance and skin protection. Further management is aimed at reducing inflammation. Corticosteroids are commonly utilized as a first-line pharmacologic agent to reduce inflammation. The form (topical vs. systemic) and potency of corticosteroids depend on the location of the lesions and the total body surface area involved [15]. While corticosteroids are predominantly used for the treatment of ACD, some studies have demonstrated reduced reactions to patch tests when pre-treated with corticosteroids, highlighting their potential effectiveness when used as a prophylaxis [16-18]. Additionally, pretreatment of a localized skin area with a topical corticosteroid has shown potential to mitigate the inflammatory response from an antigen [19]. Lastly, the use of aerosolized corticosteroids has also shown potential in preventing dermatitis [20].
This study highlights the potential use of aerosolized corticosteroids as a preventive measure for ACD. The low concentration of corticosteroids in aerosolized products may help reduce the side effects typically associated with traditional topical steroids, including skin thinning, striae, and discoloration. Although this approach has previously been studied for the mitigation of reactions to other medical devices (such as continuous glucose monitors) [12], further research is needed to validate this study on AECG, given that this is a single case. In addition, the lack of a direct comparison of the amount of steroids used between topical and aerosolized forms, standardized skin monitoring, and grading of skin reactions during the study, as well as longitudinal data on the safety profile of the off-label use of intranasal corticosteroids directly on the outer skin, warrants caution prior to broader application.
Conclusions
This case highlights the increasing challenges of ACD from medical adhesive devices. The successful use of aerosolized corticosteroid in mitigating ACD from AECG underscores the possibility of its use as a prophylaxis. However, further investigation is needed to evaluate the broader applicability of this approach, given that this is a single case of an off-label use of an intranasal corticosteroid lacking longitudinal safety data.
Acknowledgments
We would like to acknowledge the contributions of Dr. Christine Persaud of the Brooke Army Medical Center Division of Endocrinology and Mr. Jerome Abella of the Brooke Army Medical Center Division of Cardiology to this work.
Disclosures
Human subjects: Informed consent for treatment and open access publication was obtained or waived by all participants in this study.
Conflicts of interest: In compliance with the ICMJE uniform disclosure form, all authors declare the following:
Payment/services info: All authors have declared that no financial support was received from any organization for the submitted work.
Financial relationships: All authors have declared that they have no financial relationships at present or within the previous three years with any organizations that might have an interest in the submitted work.
Other relationships: All authors have declared that there are no other relationships or activities that could appear to have influenced the submitted work.
Author Contributions
Concept and design: Loreto L. Calaquian, Noah Y. Jo, Stephanie A. Howes, John-Henry L. Dean, Serafim Perdikis
Acquisition, analysis, or interpretation of data: Loreto L. Calaquian, Noah Y. Jo, Stephanie A. Howes, John-Henry L. Dean, Serafim Perdikis
Drafting of the manuscript: Loreto L. Calaquian, Noah Y. Jo, Stephanie A. Howes, John-Henry L. Dean, Serafim Perdikis
Critical review of the manuscript for important intellectual content: Loreto L. Calaquian, Noah Y. Jo, Stephanie A. Howes, John-Henry L. Dean, Serafim Perdikis
Supervision: Serafim Perdikis
References
- 1.2023 ACC/AHA/ACCP/HRS guideline for the diagnosis and management of atrial fibrillation: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Joglar JA, Chung MK, Armbruster AL, et al. J Am Coll Cardiol. 2024;83:109–279. doi: 10.1016/j.jacc.2023.08.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.2017 ISHNE-HRS expert consensus statement on ambulatory ECG and external cardiac monitoring/telemetry. Steinberg JS, Varma N, Cygankiewicz I, et al. Ann Noninvasive Electrocardiol. 2017;22 doi: 10.1111/anec.12447. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.European Heart Rhythm Association (EHRA)/Heart Rhythm Society (HRS)/Asia Pacific Heart Rhythm Society (APHRS)/Latin American Heart Rhythm Society (LAHRS) expert consensus on risk assessment in cardiac arrhythmias: use the right tool for the right outcome, in the right population. Nielsen JC, Lin YJ, de Oliveira Figueiredo MJ, et al. J Arrhythm. 2020;36:553–607. doi: 10.1002/joa3.12338. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Ambulatory external electrocardiographic monitoring: focus on atrial fibrillation. Mittal S, Movsowitz C, Steinberg JS. J Am Coll Cardiol. 2011;58:1741–1749. doi: 10.1016/j.jacc.2011.07.026. [DOI] [PubMed] [Google Scholar]
- 5.New directions for ambulatory monitoring following 2017 HRS-ISHNE expert consensus. Locati ET. J Electrocardiol. 2017;50:828–832. doi: 10.1016/j.jelectrocard.2017.08.009. [DOI] [PubMed] [Google Scholar]
- 6.Diagnostic yield, outcomes, and resource utilization with different ambulatory electrocardiographic monitoring strategies. Gupta N, Yang J, Reynolds K, et al. Am J Cardiol. 2022;166:38–44. doi: 10.1016/j.amjcard.2021.11.027. [DOI] [PubMed] [Google Scholar]
- 7.Diagnostic utility of a novel leadless arrhythmia monitoring device. Turakhia MP, Hoang DD, Zimetbaum P, et al. Am J Cardiol. 2013;112:520–524. doi: 10.1016/j.amjcard.2013.04.017. [DOI] [PubMed] [Google Scholar]
- 8.Screening for atrial fibrillation in the older population: a randomized clinical trial. Gladstone DJ, Wachter R, Schmalstieg-Bahr K, et al. JAMA Cardiol. 2021;6:558–567. doi: 10.1001/jamacardio.2021.0038. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Medical adhesive-related skin injury prevalence at the peripherally inserted central catheter insertion site: a cross-sectional, multiple-center study. Zhao H, He Y, Wei Q, Ying Y. J Wound Ostomy Continence Nurs. 2018;45:22–25. doi: 10.1097/WON.0000000000000394. [DOI] [PubMed] [Google Scholar]
- 10.Incidence and characteristics of medical adhesive-related skin injuries in patients following spinal surgery: a prospective observational study. Yang TY, Park JE, Kim JB, Lee NH. Int Wound J. 2025;22:0. doi: 10.1111/iwj.70457. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Postoperative contact dermatitis caused by skin adhesives used in orthopedic surgery: incidence, characteristics, and difference from surgical site infection. So SP, Yoon JY, Kim JW. Medicine (Baltimore) 2021;100:0. doi: 10.1097/MD.0000000000026053. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12."Out of the box" solution for skin problems due to glucose-monitoring technology in youth with type 1 diabetes: real-life experience with fluticasone spray. Paret M, Barash G, Rachmiel M. Acta Diabetol. 2020;57:419–424. doi: 10.1007/s00592-019-01446-y. [DOI] [PubMed] [Google Scholar]
- 13.Prevalence of contact allergy in the general population: a systematic review and meta-analysis. Alinaghi F, Bennike NH, Egeberg A, Thyssen JP, Johansen JD. Contact Dermatitis. 2019;80:77–85. doi: 10.1111/cod.13119. [DOI] [PubMed] [Google Scholar]
- 14.Medical adhesive allergens: retrospective analysis of cross-sectional data from the North American Contact Dermatitis Group, 2001-2018. Atwater AR, Bembry R, Liu B, et al. J Am Acad Dermatol. 2022;87:1024–1032. doi: 10.1016/j.jaad.2021.11.055. [DOI] [PubMed] [Google Scholar]
- 15.Diagnosis and management of contact dermatitis. Usatine RP, Riojas M. https://www.aafp.org/pubs/afp/issues/2010/0801/p249.html. Am Fam Physician. 2010;82(3):249–255. [PubMed] [Google Scholar]
- 16.The effect of topically applied corticosteroid on irritant and allergic patch test reactions. Green C. Contact Dermatitis. 1996;35:331–333. doi: 10.1111/j.1600-0536.1996.tb02412.x. [DOI] [PubMed] [Google Scholar]
- 17.Efficacy of topical corticosteroids in nickel-induced contact allergy. Hachem JP, De Paepe K, Vanpée E, Bogaerts M, Kaufman L, Rogiers V, Roseeuw D. Clin Exp Dermatol. 2002;27:47–50. doi: 10.1046/j.0307-6938.2001.00963.x. [DOI] [PubMed] [Google Scholar]
- 18.Pre-emptive avoidance strategy (P.E.A.S.) - addressing allergic contact dermatitis in pediatric populations. Hill H, Goldenberg A, Golkar L, Beck K, Williams J, Jacob SE. Expert Rev Clin Immunol. 2016;12:551–561. doi: 10.1586/1744666X.2016.1142373. [DOI] [PubMed] [Google Scholar]
- 19.Mitigation or prevention of mild hair dye contact dermatitis after pretreatment with clobetasol foam. Rickstrew J, Wang T, Liu D, Rajpara A, Aires DJ. J Am Acad Dermatol. 2017;77:0. doi: 10.1016/j.jaad.2017.07.021. [DOI] [PubMed] [Google Scholar]
- 20.Steroid aerosol spray in contact dermatitis. Fisher AA. Arch Dermatol. 1964;89:841–843. doi: 10.1001/archderm.1964.01590300069020. [DOI] [PubMed] [Google Scholar]