To the Editor,
Malposition of a nasogastric tube (NGT) can occasionally occur [1]. This may potentially lead to serious adverse events [2]. The Tumguide® LED Light Source (Neuroceuticals Inc., Tokyo, Japan; distributed by Otsuka Pharmaceutical Factory, Inc., Tokushima, Japan) emits highly bio-permeable red light (wavelength 660 nm) and may help identify the position of the NGT tip by allowing visual confirmation in the epigastric area from outside the body [3].
In cases of gastrectomy, the inserted NGT should be sufficiently withdrawn during resection to prevent interference with the surgical field or accidental transection. After the anastomosis is completed, the NGT should be carefully re-advanced to the appropriate position. Caution is essential to avoid blind or forceful advancement because this may injure the anastomotic site. There have been reports of adverse events in which NGTs were accidentally cut or sutured during surgery due to improper positioning [4–6].
In laparoscopic surgery, the anesthesiologist and surgeon estimate the NGT’s position based on the insertion length or by subtle movements of the gastric wall in response to tube advancement. However, these usual approaches may not always provide a reliable confirmation of the NGT’s position. Because the tube lies entirely within the stomach and is not directly visible, its position must often be inferred, and unintentional advancement can occur. We are sometimes unable to confirm the tube’s movement, and it may be unintentionally advanced further.
This case involved an 88-year-old man (height 169 cm, weight 62 kg) who underwent laparoscopic total gastrectomy for gastric cancer. In this case, the Tumguide® allowed us to easily confirm the position of the NGT tip and safely advance it to the appropriate location after anastomosis (Fig. 1). Once the Tumguide® light is turned on, it can be switched on and off as needed. However, if the total time it remains off exceeds 15 min after first activation, it will stop functioning and cannot be used again. Therefore, if we use the light to check the tube tip during stomach transection, it will no longer be available for confirming the tube’s position after anastomosis. One limitation of the Tumguide® is that it is intended for temporary, single-use only and should not be reinserted into the nasogastric tube after removal.
Fig. 1.

Laparoscopic view of the esophagojejunostomy site. The arrow (⇧) in each panel indicates the esophagojejunostomy site. a As the nasogastric tube passed through the diaphragm, the Tumguide® light became visible within the esophagus. b The gastric tube was advanced beyond the esophagojejunostomy into the jejunum. The red light was clearly visible through the laparoscopic camera, allowing safe confirmation of the tube’s position just distal to the anastomosis
Tumguide® allows both the surgeon and the anesthesiologist to confirm the position of the NGT intraoperatively, enabling shared recognition of its correct positioning and preventing over-insertion or excessive withdrawal.
Since the Tumguide® is a disposable and relatively expensive device, its routine use in all laparoscopic surgeries requiring NGT confirmation is impractical. From a safety management perspective, it may still be reasonable to consider the Tumguide® as an option for patients in whom NGT placement is particularly difficult or whose tube position is hard to confirm.
In this single case, we were able to successfully identify the Tumguide® during the procedure. However, care must be taken as the laparoscopic light source is very intense, and there is a risk that the Tumguide® may be overlooked.
Acknowledgements
We thank Angela Morben, DVM, ELS, from Edanz (https://jp.edanz.com/ac), for editing a draft of this manuscript.
Abbreviations
- NGT
Nasogastric tube
Authors’ contributions
YR wrote the manuscript. CN was responsible for anesthesia management. SH performed the surgery. CN reviewed and edited the manuscript. SI helped write the manuscript. All authors read and approved the final manuscript.
Funding
Not applicable.
Data availability
Not applicable.
Declarations
Ethics approval and consent to participate
Not applicable.
Consent for publication
Written informed consent was obtained from the patient.
Competing interests
The authors declare no competing interests.
Footnotes
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Bankier AA, Wiesmayr MN, Henk C, Turetschek K, Winkelbauer F, Mallek R, et al. Radiographic detection of intrabronchial malpositions of nasogastric tubes and subsequent complications in intensive care unit patients. Intensive Care Med. 1997;23(4):406–10. 10.1007/s001340050348. [DOI] [PubMed] [Google Scholar]
- 2.Motta APG, Rigobello MCG, Silveira RCCP, Gimenes FRE. Nasogastric/nasoenteric tube-related adverse events: an integrative review. Rev Lat Am Enfermagem. 2021;29:e3400. 10.1590/1518-8345.3355.3400. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Masaki H, Hirano H, Takahashi J, Kamada T, Masaki E. An improved biologically transparent illumination system that increases the accuracy of detecting the correct position of a nasogastric tube in the stomach. PLoS ONE. 2023;18(12):e0295415. 10.1371/journal.pone.0295415. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Chen CN, Lee WJ, Cheng TJ, Chang KJ. Endoscopic removal of nasogastric tube sutured unintentionally to gastrojejunostomy. Surg Laparosc Endosc. 1997;7(4):359–60. [PubMed] [Google Scholar]
- 5.Wilkinson MN, Jayaraman V, Watkins K, Bao P, Buscaglia JM. Newly developed flexible endoscopic scissors for removal of a nasogastric tube trapped within a gastrojejunal anastomotic staple line. Endoscopy. 2011;43 Suppl 2 UCTN:E234–5. 10.1055/s-0030-1256426. [DOI] [PubMed]
- 6.Shaaban H, Armstrong C. Nasogastric tube accidentally stitched to the stomach during laparoscopic antireflux surgery. Endoscopy. 2009;41(Suppl 2):E61. 10.1055/s-0028-1103464. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Not applicable.