. 2025 Oct 29;27(11):euaf268. doi: 10.1093/europace/euaf268

Comparison of readmission outcomes and complications between leadless and traditional transvenous pacemakers in older adults: a nationwide readmission analysis of 49852 admission events

Jiaqi He 1, Keting Liang 2, Ruijian Huang 3, Cunhua Su 4, Jiancheng Zhou 5,✉,3,2, Lingli Wang 6,✉,3,2, Jifang Zhou 7,✉,3,2
PMCID: PMC12596647  PMID: 41160713

Abstract

Aims

While transvenous pacemakers (TV-VVI) are standard for bradyarrhythmia, lead- and pocket-related complications remain concerns. Leadless pacemakers (LPMs) may reduce these risks. However, direct comparisons between LPMs and single-chamber TV pacemakers are limited. This study aimed to compare clinically meaningful outcomes between LPM and TV-VVI using real-world data.

Methods and results

Using the National Readmissions Database (NRD), we analysed demographics, readmission rates, and 30-day outcomes of patients aged ≥65 years who underwent LPM or TV-VVI implantation between 2016 and 2022. Admissions were identified via ICD-10 codes. Outcomes were assessed in the propensity score-matched population (10 594 patients per group) through multivariable logistic regression after 1:1 high-dimensional propensity score matching (caliper 0.1 SD) to adjust for confounding. Among 49 852 patients, 44.8% received LPM. Median age was 84 vs. 81 years in TV-VVI and LPM groups; 46.2% were female. TV-VVI patients had significantly higher rates of device-related complications [adjusted OR (aOR): 0.45, 95% CI (0.30–0.65), P < 0.001], device revision or replacement [aOR: 0.20, 95% CI (0.11–0.36), P < 0.001], implant-related complications [aOR: 0.58, 95% CI (0.34–0.97), P = 0.040]. Crude rates of arteriovenous fistula, pseudoaneurysm, and pericardial complications were higher in LPM, but adjusted differences were non-significant. Thirty-day readmission rates were similar between LPM and TV-VVI groups at 15.5% and 15.9%, respectively. Mortality and prolonged length of hospital stay also showed no significant differences.

Conclusion

Nationally representative data indicate that LPM implantation is associated with fewer device-related complications compared to TV-VVI, though further studies are needed to evaluate long-term outcomes.

Keywords: Leadless pacemaker, TV-VVI, Complications, Mortality, Outcomes, Readmission

Graphical Abstract

Graphical Abstract.

Graphical Abstract


What’s new?

  • Leadless pacemaker (LPM) implantation has increased annually, surpassing traditional single-chamber transvenous pacemakers (TV-VVI) in volume since 2020, while TV-VVI implantation has declined overall.

  • LPM implantation was associated with fewer device-related complications, including lower rates of device revision or replacement, implant-related complications and device infection.

  • No significant differences in 30-day readmission rates were observed between the LPM and TV-VVI groups.

  • No significant differences in all-cause in-hospital mortality or prolonged length of stay (≥30 days) were observed between the LPM and TV-VVI groups within 30 days of readmission.

Introduction

With the accelerating process of global population aging, the demand for pacemaker implantation has been steadily increasing. According to the 2021 European Society of Cardiology (ESC) Guidelines for Cardiac Pacing and Resynchronization Therapy, ∼1 million cardiac devices are implanted globally each year, and this trend continues to rise.1 Pacemakers are essential medical devices for the treatment of bradyarrhythmia, such as atrioventricular block and sick sinus syndrome.1–3

Although the traditional single-chamber transvenous pacemakers (TV-VVI) have been widely used and offer significant clinical benefits, the presence of leads and the subcutaneous pulse generator pocket may cause device-related complications.4  ,5

In recent years, leadless pacemakers (LPM) have emerged as a novel technology. LPM avoid the need for leads and the pocket structure typically required by traditional pacemakers. The first single-chamber right ventricular LPM received CE certification in 2015 and FDA approval in 2016. Early studies demonstrated a high implantation success rate of up to 99.6% and a primary complication rate of 1.51%.6 As outlined in the 2021 ESC guidelines, LPM are considered a suitable alternative for high-risk patients, including those with a history of pocket infections, those undergoing haemodialysis, and those with venous access obstruction that precludes the implantation of traditional transvenous pacemakers (TVPs).1

However, randomized controlled trials comparing LPM and TV-VVI are limited, and real-world evidence remains scarce. Registry data in haemodialysis patients suggest that LPM are associated with lower device-related infections and improved survival compared with single-chamber transvenous devices; however, these findings are restricted to a high-risk population and may not generalize to older adults with broader comorbidity profiles.7

Hospital readmission is a widely accepted indicator of healthcare quality and short-term patient outcomes, particularly in cardiovascular care.8 Older adults receiving pacemakers often have multiple comorbidities and are prone to early rehospitalizations.9 Therefore, 30-day readmission is a clinically meaningful outcome to assess short-term differences in outcomes between leadless and transvenous pacing in real-world practice. Therefore, we aim to evaluate clinical outcomes, including 30-day readmission, between LPM and TV-VVI in this large real-world retrospective cohort study based on the National Readmissions Database (NRD). A high-dimensional propensity score (HDPS) matching approach was applied to balance baseline characteristics and reduce confounding.

Methods

Data sources

This study utilized data from the NRD spanning 2016–22. The NRD is part of the Healthcare Cost and Utilization Project (HCUP), supported by the Agency for Healthcare Research and Quality (AHRQ) and contains patient-level linkage variables that enable tracking of readmissions across hospitals within a calendar year. It includes nationally representative in-patient discharge records from multiple U.S. states, covering ∼35 million weighted discharges annually (excluding rehabilitation and long-term acute care facilities). All data are de-identified, and thus, Institutional Review Board approval was not required. Data for the study were systematically obtained using the International Classification of Diseases, Tenth Revision, Clinical Modification and Procedure Coding System (ICD-10-CM/PCS) for the identification of diagnoses and procedures, respectively. The database also includes key clinical variables, such as patient demographics, in-hospital mortality, and length of stay.

Sample extraction

We queried the NRD from 2016 to 2022 to include patients aged ≥65 years with an admission diagnosis of sick sinus syndrome, second-degree atrioventricular block, or third-degree atrioventricular block, who underwent either a LPM or TV-VVI implantation during hospitalization. LPM implantation was identified using the ICD-10-PCS code (02HK3NZ), which specifically denotes an entirely intracardiac device without transvenous leads. TV-VVI was defined by the presence of both single-chamber pacemaker codes (0JH604Z, 0JH605Z) and right ventricular lead insertion codes (02HK3JZ, 02HK4JZ, 02HK3MZ, 02HK4MZ). Internal checks confirmed that no LPM-coded patients had concurrent lead insertion codes, ensuring accurate classification.

Patients were excluded if they had a prior history of cardiovascular implantable electronic devices (CIEDs), were younger than 65 years, or had missing in-hospital mortality information. The hospitalization during which the pacemaker was implanted was defined as the ‘index admission’, and any readmission occurring within a specified readmission period after discharge from the index admission was defined as the corresponding ‘readmission’ (e.g. 30-day readmission). If multiple readmissions occurred within the same readmission window, only the first readmission was included. To ensure the accuracy of readmission rate estimation across different readmission periods, patients with insufficient observation window were excluded, specifically, patients enrolled in December were excluded from the 30-day readmission analysis. We additionally analysed the causes of 30-day readmissions, which were classified based on the primary diagnosis ICD-10-CM code at readmission. Causes were categorized as cardiac or non-cardiac, and further stratified as elective or non-elective.

Statistical analysis

To control for potential confounders, a range of patient-level and hospital-level covariates were included in this study. Patient-level covariates included: age, sex, insurance type, median household income for the patient's zip code, Charlson comorbidity index (CCI) score, aortic aneurysm, atrial fibrillation, cardiac aneurysm, percutaneous coronary intervention (PCI), ischaemic heart disease (IHD), myocardial infarction (MI), coronary artery bypass grafting (CABG), non-ST elevation myocardial infarction (NSTEMI), ST elevation myocardial infarction (STEMI), dementia, ischaemic stroke, haemorrhagic stroke, transient ischaemic attack (TIA), dyslipidaemia, hypertension, diabetes, obesity, coronary artery disease (CAD), peripheral artery disease (PAD), chronic kidney disease (CKD), end-stage renal disease (ESRD), chronic obstructive pulmonary disease (COPD), congestive heart failure (CHF), cardiogenic shock, anaemia, thrombocytopenia, chronic liver disease, tobacco use, alcohol use, substance abuse, history of malignancy, and venous thromboembolism. Hospital-level covariates included hospital bed size and hospital location/teaching status.

Categorical variables were presented as frequencies and percentages, while continuous variables were reported using medians and inter-quartile ranges (IQRs).

To further control for high-dimensional confounders, we employed the HDPS method. HDPS uses large-scale diagnostic and procedural coding information, combining the frequency of variables and their association strength with the study outcome.10 From the patient's diagnostic and procedural codes, the 100 most relevant variables were selected and included in the propensity score model, alongside the pre-specified clinical covariates.11,12

Subsequently, propensity score matching (PSM) was performed using a 1:1 nearest-neighbour matching method without replacement and employing a greedy algorithm. A caliper width of 0.1 standard deviations (SD) was used to construct a well-matched propensity-matched cohort, ensuring independence of matched pairs to improve the validity of statistical inference and reduce potential bias, even though this reduced the sample size. The balance of baseline characteristics was primarily assessed using standardized mean differences (SMD), with an SMD < 0.1 indicating negligible differences. Before matching, crude odds ratios (OR) were estimated using a univariate logistic regression model. After matching, multivariate logistic regression was performed on the matched cohort to estimate the adjusted odds ratios (aOR). The aOR model included covariates that remained imbalanced after matching (SMD > 0.1) as well as covariates with SMD close to 0.1 that were considered clinically important.

To supplement the observation of longer-term clinical outcomes and enhance the robustness of our findings, sensitivity analyses were conducted to assess readmission outcomes at 60, 90, and 180 days.

All analyses were conducted using SAS version 9.4 (SAS Institute Inc., Cary, NC, USA) and R version 4.3.1 (R Foundation for Statistical Computing, Vienna, Austria). Two-sided statistical tests with an α level of <0.05 were considered statistically significant.

Results

Baseline characteristic and readmission rate

Our cohort included 49 852 patients who underwent pacemaker implantation during the study period. Of these, 27 494 (55.2%) received TV-VVI, and 22 358 (44.8%) received LPM (Figure 1). Between 2016 and 2022, the number of LPM implantations increased markedly from 98 to 6388. LPM implantations surpassed TV-VVI in volume beginning in 2020 and continued to exceed them thereafter (Figure 2). Overall, the majority of patients were aged ≥80 years (67.8% in the TV-VVI group and 56.8% in the LPM group), and 46.2% were female. Most implantations were performed at large, urban teaching hospitals. The most prevalent comorbidities were atrial fibrillation (74.7%), hypertension (65.2%), hyperlipidaemia (60.7%), and heart failure (54.4%) (Table 1). Compared with the LPM group, the TV-VVI group had a higher prevalence of atrial fibrillation (85.7% vs. 61.1%), whereas the LPM group had a higher prevalence of ESKD (10.5% vs. 2.7%). In the overall cohort, the 30-day readmission rates were 15.9% in the LPM group and 15.3% in the TV-VVI group, no significant difference was observed [OR: 1.05, 95% CI (0.99–1.10), P = 0.079] (see Supplementary material online, Figure S1 and Supplementary material online, Table S5). Among 30-day readmissions, 40.8% were due to cardiac causes and 59.2% to non-cardiac causes. The majority of admissions were non-elective, accounting for 88.1% (see Supplementary material online, Table S1).

Figure 1.

CONSORT diagram showing the selection of the study cohort from the NRD between 2016 and 2022. Initially, 89,258,036 admissions were screened. Inclusion criteria were insertion of a leadless or right ventricular lead pacemaker with a primary diagnosis of sick sinus syndrome or second-/third-degree AV block. Exclusions included prior CIED implantation, age <65 years, or missing in-hospital mortality data. After applying these criteria, 49,852 patients remained in the index admission cohort, with 27,494 receiving single-lead pacemakers and 22,358 receiving leadless pacemakers. Admissions in December or without readmissions within 30 days were excluded for the 30-day readmission analysis, leaving 7,082 readmissions (3,666 single-lead and 3,216 leadless). RV = right ventricle; CIED = cardiovascular implantable electronic device.

Consort diagram showing the derivation of our study sample. RV, right ventricle; CIED, cardiovascular implantable electronic device.

Figure 2.

Bar chart showing the number of LPM (leadless pacemakers) and TV-VVI (single-chamber transvenous pacemakers) implantations per year from 2016 to 2022. The x-axis represents the year, and the y-axis represents the number of patients. LPM implantations increased steadily each year, surpassing TV-VVI in 2020 and remaining higher thereafter. TV-VVI implantations showed an overall declining trend. The chart clearly illustrates the annual changes in the use of the two types of pacemakers.

Temporal trends of LPM or traditional TVP. LPM, leadless pacemaker; TV-VVI, transvenous single-chamber ventricular pacemaker.

Table 1.

Baseline characteristics of pacemaker implantation patients

Variable Overall Before propensity matching After propensity matching
Treatment |SMD| Treatment |SMD|
TV-VVI Leadless TV-VVI Leadless
Population 49 852 27 494 22 358 10 594 10 594
Age 83 [76, 88] 84 [78, 89] 81 [74, 87] 0.264 82 [76, 88] 82 [76, 88] 0.022
Age group (%)
 65–69 3847 (7.7) 1594 (5.8) 2253 (10.1) 0.251 809 (7.6) 844 (8.0) 0.028
 70–74 6192 (12.4) 2840 (10.3) 3352 (15.0) 1307 (12.3) 1392 (13.1)
 75–79 8485 (17.0) 4426 (16.1) 4059 (18.2) 1873 (17.7) 1849 (17.5)
 80+ 31 328 (62.8) 18 634 (67.8) 12 694 (56.8) 6605 (62.3) 6509 (61.4)
Female sex (%) 23 008 (46.2) 12 695 (46.2) 10 313 (46.1) 0.001 4903 (46.3) 4888 (46.1) 0.003
Insurance type (%)
 Medicare 45 956 (92.2) 25 510 (92.8) 20 446 (91.4) 0.080 9724 (91.8) 9663 (91.2) 0.030
 Medicaid 555 (1.1) 246 (0.9) 309 (1.4) 126 (1.2) 136 (1.3)
 Private Insurance 2132 (4.3) 1185 (4.3) 947 (4.2) 460 (4.3) 486 (4.6)
 Self-pay 136 (0.3) 61 (0.2) 75 (0.3) 31 (0.3) 33 (0.3)
 Other 1034 (2.1) 478 (1.7) 556 (2.5) 244 (2.3) 264 (2.4)
 Missing 39 (0.1) 14 (0.1) 25 (0.1) 9 (0.1) 12 (0.1)
Median household income for the patient zip code, n (%)
 Q1: 0–25th percentile 10 835 (21.7) 5934 (21.6) 4901 (21.9) 0.080 2280 (21.5) 2259 (21.3) 0.012
 Q2: 26th−50th percentile 13 055 (26.2) 7577 (27.6) 5478 (24.5) 2804 (26.5) 2815 (26.6)
 Q3: 51st–75th percentile 12 861 (25.8) 7047 (25.6) 5814 (26.0) 2741 (25.9) 2747 (25.9)
 Q4: 76th–100th percentile 12 564 (25.2) 6622 (24.1) 5942 (26.6) 2667 (25.2) 2659 (25.1)
 Missing 537 (1.1) 314 (1.1) 223 (1.0) 102 (1.0) 114 (1.1)
CCI (%)
 0–5 9494 (19.0) 5110 (18.6) 4384 (19.6) 0.081 2016 (19.0) 2004 (18.9) 0.008
 6–7 17 314 (34.7) 10 021 (36.4) 7293 (32.6) 3607 (34.0) 3575 (33.7)
 8+ 23 044 (46.2) 12 363 (45.0) 10 681 (47.8) 4971 (46.9) 5015 (47.3)
Bed size (%)
 Small 5541 (11.1) 3714 (13.5) 1827 (8.2) 0.276 995 (9.4) 1106 (10.4) 0.050
 Medium 13 450 (27.0) 8354 (30.4) 5096 (22.8) 2836 (26.8) 2640 (24.9)
 Large 30 861 (61.9) 15 426 (56.1) 15 435 (69.0) 6763 (63.8) 6848 (64.6)
Location/teaching status (%)
 Rural 10 607 (21.3) 6702 (24.4) 3905 (17.5) 0.292 2180 (20.6) 1797 (17.0) 0.227
 Urban teaching 37 619 (75.5) 19 434 (70.7) 18 185 (81.3) 7917 (74.7) 8657 (81.7)
 Urban non-teaching 1626 (3.3) 1358 (4.9) 268 (1.2) 497 (4.7) 140 (1.3)
Years (%)
 2016 4612 (9.3) 4514 (16.4) 98 (0.4) 1.084 416 (3.9) 94 (0.9) 0.311
 2017 6320 (12.7) 5390 (19.6) 930 (4.2) 956 (9.0) 808 (7.6)
 2018 6603 (13.2) 4677 (17.0) 1926 (8.6) 1377 (13.0) 1512 (14.3)
 2019 6807 (13.7) 3962 (14.4) 2845 (12.7) 1680 (15.9) 1979 (18.7)
 2020 7746 (15.5) 3359 (12.2) 4387 (19.6) 1816 (17.1) 2348 (22.2)
 2021 8669 (17.4) 2885 (10.5) 5784 (25.9) 2017 (19.0) 2320 (21.9)
 2022 9095 (18.2) 2707 (9.8) 6388 (28.6) 2332 (22.0) 1533 (14.5)
Comorbidities (%)
 Aortic Aneurysm 1385 (2.8) 778 (2.8) 607 (2.7) 0.007 300 (2.8) 306 (2.9) 0.003
 Atrial Fibrillation 37 219 (74.7) 23 560 (85.7) 13 659 (61.1) 0.580 8015 (75.7) 7785 (73.5) 0.050
 Cardiac Aneurysm 49 (0.1) 21 (0.1) 28 (0.1) 0.015 12 (0.1) 9 (0.1) 0.009
 PCI 5220 (10.5) 2841 (10.3) 2379 (10.6) 0.010 1113 (10.5) 1117 (10.5) 0.001
 IHD 22 208 (44.5) 12 239 (44.5) 9969 (44.6) 0.001 4726 (44.6) 4749 (44.8) 0.004
 MI 4471 (9.0) 2458 (8.9) 2013 (9.0) 0.002 957 (9.0) 955 (9.0) 0.001
 CABG 5415 (10.9) 3138 (11.4) 2277 (10.2) 0.040 1147 (10.8) 1105 (10.4) 0.013
 NSTEMI 1836 (3.7) 941 (3.4) 895 (4.0) 0.031 413 (3.9) 424 (4.0) 0.005
 STEMI 391 (0.8) 170 (0.6) 221 (1.0) 0.041 105 (1.0) 104 (1.0) 0.001
 Dementia 8310 (16.7) 4643 (16.9) 3667 (16.4) 0.013 1825 (17.2) 1837 (17.3) 0.003
 Stroke, Ischaemic 2861 (5.7) 1545 (5.6) 1316 (5.9) 0.011 632 (6.0) 598 (5.6) 0.014
 Stroke, Haemorrhagic 365 (0.7) 174 (0.6) 191 (0.9) 0.026 83 (0.8) 93 (0.9) 0.010
 TIA 398 (0.8) 270 (1.0) 128 (0.6) 0.047 66 (0.6) 76 (0.7) 0.012
 Dyslipidaemia 30 270 (60.7) 16 404 (59.7) 13 866 (62.0) 0.048 6466 (61.0) 6471 (61.1) 0.001
 Hypertension 32 502 (65.2) 18 322 (66.6) 14 180 (63.4) 0.068 6785 (64.0) 6735 (63.6) 0.010
 Diabetes 18 415 (36.9) 9504 (34.6) 8911 (39.9) 0.110 3909 (36.9) 3991 (37.7) 0.016
 Obesity 7650 (15.3) 4188 (15.2) 3462 (15.5) 0.007 1622 (15.3) 1654 (15.6) 0.008
 CAD 22 168 (44.5) 12 217 (44.4) 9951 (44.5) 0.001 4719 (44.5) 4744 (44.8) 0.005
 PAD 4581 (9.2) 2648 (9.6) 1933 (8.6) 0.034 972 (9.2) 967 (9.1) 0.002
 CKD 19 210 (38.5) 10 108 (36.8) 9102 (40.7) 0.081 4129 (39.0) 4128 (39.0) <0.001
 ESRD 3108 (6.2) 755 (2.7) 2353 (10.5) 0.316 550 (5.2) 658 (6.2) 0.044
 COPD 11 974 (24.0) 6756 (24.6) 5218 (23.3) 0.029 2561 (24.2) 2553 (24.1) 0.002
 CHF 27 095 (54.4) 15 517 (56.4) 11 578 (51.8) 0.093 5803 (54.8) 5781 (54.6) 0.004
 Cardiogenic Shock 2580 (5.2) 1037 (3.8) 1543 (6.9) 0.140 630 (5.9) 625 (5.9) 0.002
 Anaemia 3425 (6.9) 1787 (6.5) 1638 (7.3) 0.033 774 (7.3) 751 (7.1) 0.008
 Thrombocytopenia 4676 (9.4) 2446 (8.9) 2230 (10.0) 0.037 993 (9.4) 989 (9.3) 0.001
 Chronic Liver Disease 1561 (3.1) 731 (2.7) 830 (3.7) 0.060 368 (3.5) 377 (3.6) 0.005
 Tobacco Use 16 027 (32.1) 8858 (32.2) 7169 (32.1) 0.003 3398 (32.1) 3402 (32.1) 0.001
 Alcohol Use 1035 (2.1) 548 (2.0) 487 (2.2) 0.013 239 (2.3) 221 (2.1) 0.012
 Substance Abuse 2688 (5.4) 1364 (5.0) 1324 (5.9) 0.042 594 (5.6) 611 (5.8) 0.007
 History of Malignancy 8274 (16.6) 4506 (16.4) 3768 (16.9) 0.012 1748 (16.5) 1765 (16.7) 0.004
 Venous Thromboembolism 1967 (3.9) 802 (2.9) 1165 (5.2) 0.116 441 (4.2) 470 (4.4) 0.013

CCI, Charlson comorbidity index; PCI, percutaneous coronary intervention; IHD, ischaemic heart disease; MI, myocardial infarction; CABG, coronary artery bypass grafting; NSTEMI, non-ST elevation myocardial infarction; STEMI, ST elevation myocardial infarction; TIA, transient ischaemic attack; CAD, coronary artery disease; PAD, peripheral artery disease; CKD, chronic kidney disease; ESRD, end-stage renal disease; COPD, chronic obstructive pulmonary disease; CHF, congestive heart failure.

In-patient mortality, length of stay, and complications

After 1:1 HDPS matching, a total of 21 188 hospitalizations were included, with 10 594 patients in each group. All subsequent outcome analyses, including in-hospital mortality, length of stay (LOS), and complications, were conducted in the matched population. In this cohort, all-cause in-hospital mortality during 30-day readmission was 9.3% in the LPM group and 10.3% in the TV-VVI group [aOR: 0.94, 95% CI (0.41–2.85), P = 0.881], with no statistically significant difference. Similarly, prolonged length of stay (≥30 days) occurred in 1.3% of the LPM group and 2.0% of the TV-VVI group [aOR: 0.62, 95% CI (0.34–1.10), P = 0.107], also without a significant difference (Table 2 and Figure 3).

Table 2.

Breakdown of primary outcomes and logistic regression analysis before and after PSM among patients with 30-day rehospitalization

Primary outcome Before propensity matching After propensity matching Crude ORa (95% CI) P-value Adjusted ORa (95% CI) P-value
Overall n = 7082 (%) Leadless n = 3216 (%) TV-VVI n = 3866 (%) Overall n = 3046 (%) Leadless n = 1519 (%) TV-VVI n = 1527 (%)
All-cause in-hospital mortality 643 (9.1) 299 (9.3) 344 (8.9) 300 (9.8) 142 (9.3) 158 (10.3) 1.05 (0.89–1.25) 0.551 0.94 (0.41–2.85) 0.881
Prolonged LOSa (≥30 days) 101 (1.4) 45 (1.4) 56 (1.4) 49 (1.6) 19 (1.3) 30 (2.0) 0.94 (0.61–1.42) 0.761 0.62 (0.34–1.10) 0.107
Complications 1090 (15.4) 539 (16.8) 551 (14.3) 494 (16.2) 238 (15.7) 256 (16.8) 1.19 (1.04–1.36) 0.014 0.92 (0.76–1.11) 0.376
Vascular complications 709 (10.0) 373 (11.6) 336 (8.7) 328 (10.8) 176 (11.6) 152 (10.0) 1.37 (1.16–1.62) <0.001 1.17 (0.93–1.48) 0.169
 Post-operative bleeding/Hematoma 31 (0.4) 19 (0.6) 12 (0.3) 18 (0.6) 11 (0.7) <11b 2.07 (0.99–4.47) 0.055 1.62 (0.63–4.42) 0.324
 Post-operative anaemia 666 (9.4) 346 (10.8) 320 (8.3) 301 (9.9) 160 (10.5) 141 (9.2) 1.32 (1.12–1.57) <0.001 1.15 (0.90–1.46) 0.266
 Accidental puncture <11b <11b <11b <11b <11b <11b 1.28 (0.14–11.41) 0.810 1.09 (0.04–28.10) 0.950
 Arteriovenous fistula <11b <11b <11b <11b <11b 0 12.09 (2.01–232.52) 0.023 * 0.992
 Vascular pseudoaneurysm 35 (0.5) 22 (0.7) 13 (0.3) 24 (0.8) 15 (1.0) <11b 2.19 (1.08–4.58) 0.032 1.69 (0.74–4.34) 0.216
Pericardial complications 192 (2.7) 114 (3.5) 78 (2.0) 89 (2.9) 45 (3.0) 44 (2.9) 1.74 (1.28–2.36) <0.001 1.04 (0.68–1.59) 0.861
 Pericardial effusion 183 (2.6) 108 (3.4) 75 (1.9) 82 (2.7) 40 (2.6) 42 (2.8) 1.70 (1.24–2.33) <0.001 0.96 (0.62–1.50) 0.869
 Haemopericardium <11b <11b <11b <11b <11b 0 3.99 (0.55–80.00) 0.226 * 0.997
 Cardiac tamponade 27 (0.4) 19 (0.6) <11b 14 (0.5) <11b <11b 2.62 (1.14–6.55) 0.028 2.58 (0.86–9.44) 0.111
 Pericardiocentesis 31 (0.4) 21 (0.7) <11b 18 (0.6) <11b <11b 2.45 (1.14–5.56) 0.025 1.25 (0.49–3.29) 0.636
Post-operative shock <11b <11b <11b <11b <11b <11b 1.77(0.44–7.50) 0.414 1.03 (0.04–26.7) 0.981
Pneumothorax 39 (0.6) 14 (0.4) 25 (0.6) 19 (0.6) <11b 11 (0.7) 0.68 (0.33–1.33) 0.267 0.75 (0.29–1.88) 0.544
Device-related complications 276 (3.9) 101 (3.1) 175 (4.5) 124 (4.1) 39 (2.6) 85 (5.6) 0.61 (0.47–0.80) <0.001 0.45 (0.30–0.65) <0.001
 Device-related revision/replacement 96 (1.4) 13 (0.4) 83 (2.1) 48 (1.6) <11b 40 (2.6) 0.18 (0.10–0.33) <0.001 0.20 (0.11–0.36) <0.001
 Implant related complication (Hemorrhage, stenosis, lead breakdown) 141 (2.0) 53 (1.6) 88 (2.3) 62 (2.0) 23 (1.5) 39 (2.6) 0.60 (0.41–0.87) 0.007 0.58 (0.34–0.97) 0.040
 Device thrombus <11b 0 <11b <11b 0 <11b * 0.996 * 0.998
 Device infection/Inflammation 112 (1.6) 48 (1.5) 64 (1.7) 52 (1.7) 17 (1.1) 35 (2.3) 0.85 (0.57–1.26) 0.417 0.48 (0.26–0.85) 0.014

aCrude OR, odds ratio before propensity score matching; Adjusted OR: odds ratio after propensity score matching; CI, confidence interval; PSM, propensity score matching; LOS, length of stay; TV-VVI, transvenous ventricular-inhibited pacing.

bPer HCUP policy, counts <11 are suppressed to protect privacy.

Figure 3.

Summary of primary outcomes and results of logistic regression analyses before and after propensity score matching (PSM) among patients with 30-day rehospitalization. Odds ratios (OR) compare LPM (leadless pacemakers) to TV-VVI (single-chamber transvenous pacemakers), with OR < 1 indicating a lower risk of the outcome in the leadless pacemaker group.

Breakdown of primary outcomes and logistic regression before and after PSM among patients with 30-day rehospitalization. OR compares LPMs to TV-VVI pacemakers; OR < 1 indicates lower risk in the leadless group.

As shown in Table 2, the incidence of arteriovenous fistula was 0.2% in the LPM group and <0.1% in the TV-VVI group; adjusted analysis was not feasible due to the very low number of events. Vascular pseudoaneurysm occurred more frequently in the LPM group (1.0% vs. 0.6%), with a non-significant trend after adjustment [aOR: 1.69; 95% CI (0.74–4.34); P = 0.216]. Pericardial complications occurred in 3.0% of LPM patients and 2.9% of TV-VVI patients after adjustment [aOR: 1.04; 95% CI (0.68–1.59); P = 0.861], representing a non-significant trend. In contrast, device-related complications were significantly lower in the LPM group (2.6% vs. 5.6%) [aOR: 0.45; 95% CI (0.30–0.65); P < 0.001], as were device infections or inflammations (1.1% vs. 2.3%) [aOR: 0.48; 95% CI (0.26–0.85); P = 0.014] (Table 2 and Figure 3).

Sensitivity analyses for longer-term readmission

The sensitivity analyses showed that the readmission rates at 60, 90, and 180 days were similar between groups and the trends for in-hospital mortality and major complications were consistent with the primary 30-day results. Detailed numerical results are provided in the Supplementary Materials.

Discussion

This real-world, population-based study included 49 852 readmitted patients, of whom 27 494 (55.2%) received TV-VVI and 22 358 (44.8%) received LPM. We analysed one of the largest contemporary retrospective cohorts comparing LPM and TV-VVI implantations and identified several key findings: (i) LPM adoption increased steadily over time, surpassing TV-VVI volumes in 2020, while TV-VVI use declined; (ii) Despite this trend, TV-VVI recipients experienced higher rates of device-related complications, including revisions or replacements, implant-site issues, and device-associated infections or inflammations; (iii) In contrast, 30-day readmission rates, all-cause in-hospital mortality, and prolonged length of stay (≥30 days) after readmission were similar between groups; (iv) Although unadjusted data indicated more frequent arteriovenous fistulas, vascular pseudoaneurysms, and pericardial complications among LPM patients during readmission, these represent non-significant trends after multivariable adjustment.

We observed 30-day readmission rates of 15.9% and 15.3% in the LPM and TV-VVI groups, respectively. A previous study based on the NRD reported a 30-day readmission rate of ∼13% following transvenous permanent pacemaker implantation,13 while another study focusing on LPMs demonstrated a higher all-cause readmission rate of 17.9%,14 aligning closely with our LPM findings. Moreover, we found no statistically significant difference in readmission rates between the LPM and TV-VVI groups, consistent with Mararenko et al.15

Regarding short-term mortality among readmitted patients, unadjusted all-cause mortality was initially lower in the TV-VVI group. However, after adjustment for confounders, the TV-VVI group showed a slightly higher mortality rate (10.3% vs. 9.3%), representing a non-significant trend. This suggests that baseline illness severity, rather than device type, may have driven mortality. Prior NRD analyses noted that LPM recipients often have greater comorbidity burdens, including end-stage renal disease and chronic haemodialysis, both of which are recognized as primary indications for LPM use.14 Moreover, a meta-analysis involving 2496 patients found no significant difference in all-cause mortality between the groups [RR: 0.45, 95% CI (0.15–1.35), P = 0.160].16 The current literature on both 30-day and long-term mortality outcomes is generally consistent with our findings.17,18 Differences compared with previous NRD studies reporting higher short-term mortality in LPM recipients are likely due to differences in study period and statistical methods. Our analysis included 2016–22, capturing the period of rapidly increasing LPM use, while other studies focused on earlier years.19 Variations in matching criteria and statistical approaches may also contribute to these discrepancies.

Concerning procedural outcomes, several studies have compared LPMs and transvenous systems.20–24 Breeman et al. reported that although LPM recipients may experience certain acute, procedure-related complications in the early post-operative phase, their long-term complication rates remain low, and pacing performance is maintained over a 5-year follow-up period.22 A Danish cohort study involving 5918 patients found that 9.5% experienced at least one complication during follow-up.25 Other studies have suggested that LPMs are associated with fewer complications compared to conventional transvenous systems,23 which aligns with our results, though adjusted differences were not statistically significant.

For vascular complications, arteriovenous fistula (0.4%) and pseudoaneurysm (1.0%) were more frequent in the LPM group, representing non-significant trends. This may be closely linked to the unique implantation approach and procedural characteristics. LPM are typically delivered via the femoral vein using an introducer sheath up to 23 French in diameter, which is significantly larger than the sheaths used for conventional transvenous systems.26–28 This may increase the risk of vascular injury at the access site and contribute to a higher rate of post-procedural complications. These non-significant trends should be interpreted cautiously and do not imply definitive clinical differences between the groups. Due to limited follow-up duration and small event counts, these results may not reflect potential differences in longer-term follow-up. Roberts et al. reported that within 30 days after LPM implantation, 13 major complications occurred in 12 patients, with an overall incidence of 1.51%, including arteriovenous fistula (0.13%) and vascular pseudoaneurysm (0.13%).6 In a prospective, single-arm clinical study that enrolled 744 patients, 725 patients underwent attempted device implantation, and 719 were successfully implanted. A total of 28 major complications were reported, including 11 cases of cardiac injury and 5 cases of groin puncture site complications.29 The complication rates reported in this study were comparable to those observed in our analysis.

Regarding cardiac injury, our findings showed that prior to adjustment, the LPM group had higher rates of pericardial effusion, cardiac tamponade, and pericardiocentesis; however, after adjusting for confounding factors, there was no statistically significant difference between the groups. This may be attributed to the device’s structure. Although LPM is compact, it is encased in a hard metal shell, with a nitinol anchor hook at the front that directly attaches to the right ventricular wall. If the fixation is too deep or misaligned, it may puncture or tear the myocardium, leading to complications such as pericardial effusion, pericarditis, or even cardiac tamponade. Previous reports on the acute and six-month outcomes of the Micra CED study found a higher incidence of pericardial effusion and/or perforation in patients implanted with leadless VVI pacemakers.30 In a retrospective study, Sattar et al. analysed 204 patients with LPM and found that the risk of pericardial effusion was approximately twice that of traditional pacemakers, although this difference did not reach statistical significance, representing a non-significant trend.31 Some studies have also found a higher incidence of pericarditis and pericardial effusion, which are similar to our results.17,24,32

Regarding device-related complications, our study found that LPM recipients experienced lower rates of device revisions, implant-site issues, and device-related infections, whereas TV-VVI recipients experienced more complications, most of which were lead- or pocket-related. Even if not immediately fatal, these lead- or pocket-related complications substantially increase healthcare resource use, hospitalization costs, and the likelihood of repeat interventions.33,34 Long-term evidence further indicates that device-related infections significantly elevate mortality risk: cardiac device infection after transvenous lead extraction was associated with higher 1-year and 10-year mortality,35 and in a real-world CIED cohort, infection was associated with a 12-month adjusted hazard ratio for death of 2.73 (95% CI 2.10–3.54) compared with uninfected patients.36 Collectively, these findings underscore the clinical importance of preventing TV-VVI complications, both to reduce healthcare burden and to avoid potentially life-threatening outcomes. The lower complication rate in LPM recipients can be attributed to design features that avoid the need for a subcutaneous pocket and transvenous leads.24,30 The LPM is compact, lightweight, and precisely engineered, incorporating high-density batteries, low-power electronics, catheter-based delivery systems, and materials like nitinol.15,37,38 As a single-chamber device, it is implanted directly into the right ventricular wall via the femoral vein, eliminating the need for a subcutaneous pocket or transvenous leads.26,28,39 This design minimizes typical complications associated with traditional pacemakers, such as infection, lead dislodgement, and mechanical failure, thereby enhancing safety, particularly in patients at elevated risk for conventional systems. Cantillon et al. reported that within both the short- to mid-term period (≤1 month) and the mid-term period (>1–18 months), leadless cardiac pacemaker (LCP) recipients did not experience TVP-related complications such as lead failure, pocket complications, or device-related infections.24,26,39 These data support that LCP significantly reduce the risk of device-related complications and infections.

Limitations

Several limitations should be considered when interpreting the findings of this study. First, the analysis was based on a large administrative database in which comorbidities and complications were identified through ICD codes. While these coding algorithms have been widely validated in prior research, their accuracy in routine clinical settings remains variable. Nonetheless, any misclassification is likely to be non-differential and would therefore bias the results towards the null. Second, the retrospective observational design precludes definitive causal inference and may be subject to residual confounding, despite multivariable adjustment. A major constraint of the National Readmissions Database is the absence of granular clinical data such as medication use, laboratory parameters, hemodynamic status, or pacing indications. Consequently, we were unable to determine the precise causes of in-hospital mortality, and deaths occurring outside the hospital were not captured. While causality cannot be firmly established, it is likely that bradycardia-related deaths reflect underlying comorbid conditions rather than device-related failure. Third, comparisons with other studies should be interpreted cautiously due to potential differences in outcome definitions and populations. The lack of longitudinal linkage across states or calendar years also limits the ability to capture readmissions comprehensively, potentially underestimating the true readmission burden. Moreover, the dataset does not include device-specific characteristics such as pacing thresholds, battery longevity, or implantation techniques, limiting our ability to explore potential mechanisms underlying the observed outcome differences. Finally, the present study focused exclusively on single-chamber LPMs; recent reports on dual-chamber–capable LPMs suggest differing complication and survival profiles, highlighting that our findings may not be directly extrapolated to these newer devices.40

Conclusions

Overall, LPM demonstrated significantly lower device-related complications, driven by reduced infections and lead revisions, despite comparable in the short-term readmission rates and in-hospital mortality between LPM and TV-VVI. Large-scale, prospective, randomized controlled trials are needed to further elucidate the long-term risks and clinical benefits of LPMs compared to traditional transvenous systems. These findings pertain only to single-chamber LPMs and may not be generalizable to dual-chamber leadless systems.

Supplementary Material

euaf268_Supplementary_Data

Contributor Information

Jiaqi He, School of International Business, China Pharmaceutical University, No. 639 Longmian Avenue, Jiangning District, Nanjing, Jiangsu 211198, China.

Keting Liang, School of International Business, China Pharmaceutical University, No. 639 Longmian Avenue, Jiangning District, Nanjing, Jiangsu 211198, China.

Ruijian Huang, School of International Business, China Pharmaceutical University, No. 639 Longmian Avenue, Jiangning District, Nanjing, Jiangsu 211198, China.

Cunhua Su, Department of Thoracic and Cardiovascular Surgery, Nanjing First Hospital, Nanjing Medical University, Nanjing, Jiangsu, China.

Jiancheng Zhou, Department of Pharmacy, Research Division of Clinical Pharmacology, First Affiliated Hospital of Nanjing Medical University, No. 300 Guangzhou Road, Nanjing, Jiangsu 210029, China.

Lingli Wang, Department of Cardiovascular Medicine, The Affiliated Aoyang Hospital of Jiangsu University, No. 279 Jinggang Road, Zhangjiagang, Suzhou, Jiangsu 215600, China.

Jifang Zhou, School of International Business, China Pharmaceutical University, No. 639 Longmian Avenue, Jiangning District, Nanjing, Jiangsu 211198, China.

Author contributions

Jiaqi He: Conception and design, acquisition of data, analysis, and interpretation of data, drafting the manuscript, and revising it critically for important intellectual content. Keting Liang: Acquisition of data, analysis, and interpretation of data. Ruijian Huang: Revising the manuscript critically for important intellectual content. Cunhua Su: Revising the manuscript critically for important intellectual content. Jiancheng Zhou: Conception and design, analysis and interpretation of data, revising the manuscript critically for important intellectual content, final approval of the version to be published. Lingli Wang: Revising the manuscript critically for important intellectual content, final approval of the version to be published. Jifang Zhou: Conception and design, analysis and interpretation of data, revising the manuscript critically for important intellectual content, final approval of the version to be published.

Supplementary material

Supplementary material is available at Europace online.

Funding

This study was not supported by any specific grant from public, commercial, or non-profit funding agencies.

Data availability

The data underlying this article are available in the National Readmissions Database at https://www.hcup-us.ahrq.gov/ and can be accessed by purchasing them directly from the HCUP website.

Ethics approval and consent to participate

This study did not involve human or animal subjects or clinical trials. Ethical approval was waived as de-identified administrative data were used, per institutional and national guidelines.

References

  • 1. Glikson  M, Nielsen  JC, Kronborg  MB, Michowitz  Y, Auricchio  A, Barbash  IM  et al.  2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. Eur Heart J  2021;42:3427–520. [DOI] [PubMed] [Google Scholar]
  • 2. Kusumoto  FM, Schoenfeld  MH, Barrett  C, Edgerton  JR, Ellenbogen  KA, Gold  MR  et al.  2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation  2019;140:e382–482. [DOI] [PubMed] [Google Scholar]
  • 3. Epstein  AE, Dimarco  JP, Ellenbogen  KA, Estes  NA, Freedman  RA, Gettes  LS  et al.  ACC/AHA/HRS 2008 Guidelines for device-based therapy of cardiac rhythm abnormalities. Heart Rhythm  2008;5:e1–62. [DOI] [PubMed] [Google Scholar]
  • 4. Alhuarrat  MAD, Kharawala  A, Renjithlal  S, Magdi Eid  M, Varrias  D, Mohammed  M  et al.  Comparison of in-hospital outcomes and complications of leadless pacemaker and traditional transvenous pacemaker implantation. Europace  2023;25:euad269. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Udo  EO, Zuithoff  NP, van Hemel  NM, de Cock  CC, Hendriks  T, Doevendans  PA  et al.  Incidence and predictors of short- and long-term complications in pacemaker therapy: the FOLLOWPACE study. Heart Rhythm  2012;9:728–35. [DOI] [PubMed] [Google Scholar]
  • 6. Roberts  PR, Clementy  N, Al Samadi  F, Garweg  C, Martinez-Sande  JL, Iacopino  S  et al.  A leadless pacemaker in the real-world setting: The Micra Transcatheter Pacing System Post-Approval Registry. Heart Rhythm  2017;14:1375–9. [DOI] [PubMed] [Google Scholar]
  • 7. Panico  A, Flahault  A, Guillemin  F, Varlet  E, Couchoud  C, Bauwens  M  et al.  Improved outcomes with leadless vs. single-chamber transvenous pacemaker in haemodialysis patients. Europace  2024;26:euae257. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Jencks  SF, Williams  MV, Coleman  EA. Rehospitalizations among patients in the Medicare fee-for-service program. N Engl J Med  2009;360:1418–28. [DOI] [PubMed] [Google Scholar]
  • 9. Sagoo  NS, Sagoo  R, Srinivasaraghavan  A, Bose  G, Haider  AS, Charate  R  et al.  Impact of advanced age on in-hospital outcomes following leadless pacemaker implantation. J Cardiovasc Electrophysiol  2025;36:2151–9. [DOI] [PubMed] [Google Scholar]
  • 10. Schneeweiss  S, Rassen  JA, Glynn  RJ, Avorn  J, Mogun  H, Brookhart  MA. High-dimensional propensity score adjustment in studies of treatment effects using health care claims data. Epidemiology  2009;20:512–22. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Gao  Q, Zhang  Y, Sun  H, Wang  T. Evaluation of propensity score methods for causal inference with high-dimensional covariates. Brief Bioinform  2022;23:bbac227. [DOI] [PubMed] [Google Scholar]
  • 12. Loke  YK, Mattishent  K. Propensity score methods in real-world epidemiology: a practical guide for first-time users. Diabetes Obes Metab  2020;22:13–20. [DOI] [PubMed] [Google Scholar]
  • 13. Pasupula  DK, Rajaratnam  A, Rattan  R, Munir  MB, Ahmad  S, Adelstein  E  et al.  Trends in hospital admissions for and readmissions after cardiac implantable electronic device procedures in the United States: an analysis from 2010 to 2014 using the national readmission database. Mayo Clin Proc  2019;94:588–98. [DOI] [PubMed] [Google Scholar]
  • 14. Haddadin  F, Majmundar  M, Jabri  A, Pecha  L, Scott  C, Daher  M  et al.  Clinical outcomes and predictors of complications in patients undergoing leadless pacemaker implantation. Heart Rhythm  2022;19:1289–96. [DOI] [PubMed] [Google Scholar]
  • 15. Mararenko  A, Udongwo  N, Pannu  V, Miller  B, Alshami  A, Ajam  F  et al.  Intracardiac leadless versus transvenous permanent pacemaker implantation: impact on clinical outcomes and healthcare utilization. J Cardiol  2023;82:378–87. [DOI] [PubMed] [Google Scholar]
  • 16. Darlington  D, Brown  P, Carvalho  V, Bourne  H, Mayer  J, Jones  N  et al.  Efficacy and safety of leadless pacemaker: a systematic review, pooled analysis and meta-analysis. Indian Pacing Electrophysiol J  2022;22:77–86. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. El-Chami  MF, Bockstedt  L, Longacre  C, Higuera  L, Stromberg  K, Crossley  G  et al.  Leadless vs. transvenous single-chamber ventricular pacing in the Micra CED study: 2-year follow-up. Eur Heart J  2022;43:1207–15. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Bodin  A, Clementy  N, Bisson  A, Pierre  B, Herbert  J, Babuty  D  et al.  Leadless or conventional transvenous ventricular permanent pacemakers: a nationwide matched control study. J Am Heart Assoc  2022;11:e025339. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Parulkar  A, Sheikh  W, Has  P, Lhungay  TP, Sharma  E, Vogt  B  et al.  Leadless versus single chamber pacemakers: an analysis from the national readmissions database of adverse outcomes. J Cardiovasc Electrophysiol  2025;36:1538–47. [DOI] [PubMed] [Google Scholar]
  • 20. Middour  TG, Chen  JH, El-Chami  MF. Leadless pacemakers: a review of current data and future directions. Prog Cardiovasc Dis  2021;66:61–9. [DOI] [PubMed] [Google Scholar]
  • 21. Shtembari  J, Shrestha  DB, Awal  S, Raut  A, Gyawali  P, Abe  T  et al.  Comparative assessment of safety with leadless pacemakers compared to transvenous pacemakers: a systemic review and meta-analysis. J Interv Card Electrophysiol  2023;66:2165–75. [DOI] [PubMed] [Google Scholar]
  • 22. Breeman  KTN, Oosterwerff  EFJ, de Graaf  MA, Juffer  A, Saleem-Talib  S, Maass  AH  et al.  Five-year safety and efficacy of leadless pacemakers in a Dutch cohort. Heart Rhythm  2023;20:1128–35. [DOI] [PubMed] [Google Scholar]
  • 23. Vincent  L, Grant  J, Peñalver  J, Ebner  B, Maning  J, Olorunfemi  O  et al.  Early trends in leadless pacemaker implantation: evaluating nationwide in-hospital outcomes. Heart Rhythm  2022;19:1334–42. [DOI] [PubMed] [Google Scholar]
  • 24. Cantillon  DJ, Dukkipati  SR, Ip  JH, Exner  DV, Niazi  IK, Banker  RS  et al.  Comparative study of acute and mid-term complications with leadless and transvenous cardiac pacemakers. Heart Rhythm  2018;15:1023–30. [DOI] [PubMed] [Google Scholar]
  • 25. Kirkfeldt  RE, Johansen  JB, Nohr  EA, Jørgensen  OD, Nielsen  JC. Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J  2014;35:1186–94. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Duray  GZ, Ritter  P, El-Chami  M, Narasimhan  C, Omar  R, Tolosana  JM  et al.  Long-term performance of a transcatheter pacing system: 12-month results from the Micra Transcatheter Pacing Study. Heart Rhythm  2017;14:702–9. [DOI] [PubMed] [Google Scholar]
  • 27. Thomas  H, Plummer  C, Wright  IJ, Foley  P, Turley  AJ. Guidelines for the peri-operative management of people with cardiac implantable electronic devices: Guidelines from the British Heart Rhythm Society. Anaesthesia  2022;77:808–17. [DOI] [PubMed] [Google Scholar]
  • 28. Ritter  P, Duray  GZ, Steinwender  C, Soejima  K, Omar  R, Mont  L  et al.  Early performance of a miniaturized leadless cardiac pacemaker: the Micra Transcatheter Pacing Study. Eur Heart J  2015;36:2510–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Reynolds  DW, Ritter  P. A leadless intracardiac transcatheter pacing system. N Engl J Med  2016;374:2604–5. [DOI] [PubMed] [Google Scholar]
  • 30. Piccini  JP, El-Chami  M, Wherry  K, Crossley  GH, Kowal  RC, Stromberg  K  et al.  Contemporaneous comparison of outcomes among patients implanted with a leadless vs transvenous single-chamber ventricular pacemaker. JAMA Cardiol  2021;6:1187–95. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Sattar  Y, Ullah  W, Roomi  S, Rauf  H, Mukhtar  M, Ahmad  A  et al.  Complications of leadless vs conventional (lead) artificial pacemakers—a retrospective review. J Community Hosp Intern Med Perspect  2020;10:328–33. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Boveda  S, Higuera  L, Longacre  C, Wolff  C, Wherry  K, Stromberg  K  et al.  Two-year outcomes of leadless vs. transvenous single-chamber ventricular pacemaker in high-risk subgroups. Europace  2023;25:1041–50. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33. Clémenty  N, Fernandes  J, Carion  PL, de Léotoing  L, Lamarsalle  L, Wilquin-Bequet  F  et al.  Pacemaker complications and costs: a nationwide economic study. J Med Econ  2019;22:1171–8. [DOI] [PubMed] [Google Scholar]
  • 34. Ludwig  S, Theis  C, Wolff  C, Nicolle  E, Witthohn  A, Götte  A. Complications and associated healthcare costs of transvenous cardiac pacemakers in Germany. J Comp Eff Res  2019;8:589–97. [DOI] [PubMed] [Google Scholar]
  • 35. Gomes  S, Cranney  G, Bennett  M, Giles  R. Long-term outcomes following transvenous lead extraction. Pacing Clin Electrophysiol  2016;39:345–51. [DOI] [PubMed] [Google Scholar]
  • 36. Shawon  MSR, Sotade  OT, Hill  M, Strachan  L, Challis  G, Ooi  SY  et al.  Real-world evidence on the association between cardiac implantable electronic device infection and all-cause mortality. Europace  2023;25:euad274. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37. Mela  T, Singh  JP. Leadless pacemakers: leading us into the future?  Eur Heart J  2015;36:2520–2. [DOI] [PubMed] [Google Scholar]
  • 38. El-Chami  MF, Al-Samadi  F, Clementy  N, Garweg  C, Martinez-Sande  JL, Piccini  JP  et al.  Updated performance of the Micra transcatheter pacemaker in the real-world setting: a comparison to the investigational study and a transvenous historical control. Heart Rhythm  2018;15:1800–7. [DOI] [PubMed] [Google Scholar]
  • 39. Ritter  P, Duray  GZ, Zhang  S, Narasimhan  C, Soejima  K, Omar  R  et al.  The rationale and design of the Micra Transcatheter Pacing Study: safety and efficacy of a novel miniaturized pacemaker. Europace  2015;17:807–13. [DOI] [PubMed] [Google Scholar]
  • 40. El-Chami  MF, Higuera  L, Longacre  C, Stromberg  K, Crossley  G, Piccini  JP. Two-year outcomes of Micra AV leadless pacemakers in the Micra AV CED study. Europace  2024;26:euae273. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

euaf268_Supplementary_Data

Data Availability Statement

The data underlying this article are available in the National Readmissions Database at https://www.hcup-us.ahrq.gov/ and can be accessed by purchasing them directly from the HCUP website.


Articles from Europace are provided here courtesy of Oxford University Press