Skip to main content
Wiley Open Access Collection logoLink to Wiley Open Access Collection
. 2025 Sep 4;95(1):628–636. doi: 10.1002/mrm.70062

A theoretical interpretation of diffusion weighted and intravoxel incoherent motion imaging for cerebrospinal fluid flow

Tomohiro Otani 1,, Yoshitaka Bito 2, Shigeki Yamada 3,4, Yoshiyuki Watanabe 5, Shigeo Wada 1
PMCID: PMC12620170  PMID: 40906884

Abstract

Purpose

Diffusion‐weighted imaging (DWI) and intravoxel incoherent motion (IVIM) imaging are well‐established approaches for evaluating cerebrospinal fluid (CSF) flow in subarachnoid and perivascular spaces, and have recently been applied to study ventricular CSF flow. However, DWI does not directly measure flow velocity, and the physical implications of DWI measurements are unclear. This study aimed to provide a theoretical interpretation of the DWI and IVIM imaging of CSF flow velocity fields.

Theory

The general semi‐analytical form of the signal attenuations caused by fluid flow and the resultant apparent diffusion coefficient were derived from the Bloch–Torrey equation for arbitrary b values.

Methods

The fundamental properties of the signal attenuation in laminar flow velocity fields were investigated. A Monte Carlo simulation of the IVIM parameter estimation was performed based on these signal attenuations, taking background noise into consideration.

Results

The developed theoretical framework indicates that signal attenuations in DWI detect intravoxel flow velocity standard deviations ranging from approximately 0.1 to 10 mm/s within the range of practical scan parameter settings. The lower bounds of the DWI flow profiles appeared where the flow effect was an order of magnitude lower than the molecular diffusion effects, even when b increased. The IVIM fitting parameters reflected the flow effects of the signal attenuations despite an inconsistency with the original IVIM model assumptions.

Conclusion

The physical implications of signal attenuation in DWI have been theoretically clarified. This framework provides a useful basis for understanding CSF flow dynamics and considering appropriate imaging settings.

Keywords: cerebrospinal fluid, diffusion, intravoxel incoherent motion, intravoxel standard deviation, neurofluids

1. INTRODUCTION

Intracranial cerebrospinal fluid (CSF) plays essential roles in the transport of molecular signals and the clearance of waste products 1 , and its flow characteristics have gained much attention. Magnetic resonance imaging (MRI) provides a unique tool for the non‐invasive evaluation of subject‐specific CSF characteristics, and various approaches to CSF flow imaging have been established, 2 such as phase‐contrast (PC) imaging, 3 , 4 diffusion‐weighted imaging (DWI), 5 , 6 , 7 spin‐labeling 8 , 9 , 10 and functional MRI 11 .

Because of the slow flow properties of intracranial CSF flow, its imaging presents severe technical difficulties. In this context, DWI and the apparent diffusion coefficient (ADC) provide a promising approach. 6 , 12 , 13 In particular, low‐b DWI is thought to be able to capture the signal attenuation originating from CSF flow while avoiding molecular diffusion effects. 14 , 15 , 16 , 17 Furthermore, several studies have applied DWI model‐based analysis using the intravoxel incoherent motion (IVIM) model 18 to CSF flow imaging in the brain parenchyma and subarachnoid spaces. 19 , 20 , 21 , 22 , 23 This method assumes two compartments within a voxel (perfusion and diffusion) and estimates the perfusion fraction. 18 , 24 , 25 In recent studies, the IVIM model and its parameters have been extended to ventricular CSF flow analyses. 26 , 27 , 28 , 29

A key limitation of DWI in the context of fluid flow imaging is that it does not directly measure flow velocity, 6 and both ADC and IVIM models rely on assumptions that may not hold in single‐component flows. Nevertheless, Jang et al. 30 reported strong correlations between ADC and peak flow velocity according to computational simulations, and thus DWI may reflect certain properties of intravoxel fluid flow. Bito et al. 14 investigated the theoretical properties of the ADC and showed that the limit of the ADC with sufficiently low b is a function of the variance of intravoxel flow velocity. Equivalent formulations were developed in both DWI modeling 31 and PC imaging, 32 and have been widely used for estimating turbulent kinetic energy, 33 particularly in PC imaging, where the intravoxel flow distribution is typically assumed to follow a normal distribution. Combining concepts from both DWI‐ and PC‐based formulations could provide a general theoretical form for the ADC for arbitrary b, and could establish a theoretical framework for interpreting the DWI and IVIM imaging of CSF flow fields.

The study aimed to theoretically interpret DWI and IVIM signals for slow flows, such as intracranial CSF flow. A general semi‐analytical form of the signal attenuations by the diffusion gradient pulses characterized by b was derived from the Bloch–Torrey equation. The fundamental properties of the signal attenuations and detectable ranges of intravoxel flow velocity distributions were estimated considering clinically practical conditions, and the physical meaning of the IVIM model fitting for single CSF flow was demonstrated.

2. THEORY

2.1. Signal attenuation of magnetization

A three‐dimensional domain is defined in a Cartesian coordinate system (oxyz) with the static magnetic field directed along the z‐axis. Considering a magnetization vector m=(mx,my,mz) in the target fluids filling in this domain, the Bloch–Torrey equation for the transverse component of m (m=mx+imy) in a rotating frame is given by,

mt=iγ(G·x)mmT2+D2m, (1)

where γ is the gyromagnetic ratio, G(t) is the spatial gradient magnetic field along the z‐axis, x(t) is the position vector of m, T2 is the transverse relaxation time, and D is the diffusion coefficient of the target fluids. The position vector x(t) is defined as

x(t)=x(0)+0tv(x;τ)dτ (2)

where v(x;t) is the flow velocity vector expressed as a smooth and continuous spatiotemporal function. The semi‐analytical solution of m can be derived as 34 , 35

m(x;t)=CexptT2relaxation·expγ2D0t0tG(τ)dτ2dtdiffusion·expiγ0tx(τ)·G(τ)dτprecession, (3)

where C is a constant.

From here, G(t) and v(x;t) along the same direction are considered, and these values are denoted as scalar values G(t) and v(x;t), respectively. Here, the scalar v is defined as the projection of the flow velocity vector field along the direction of the diffusion gradient. Furthermore, v(x;t) is assumed to be constant during the diffusion gradient pulse T. Using these assumptions, the signal attenuation is expressed as the change ratio of m with and without the gradient pulse, given by 34

mm0=exp(bD)·exp(ikv), (4)

where

b=γ20T0tG(τ)dτ2dt,k=γ0TtGdt, (5)

and m0 is the transverse magnetization without the gradient pulse (G(t)=0). If the Stejskal–Tanner pulse is applied, these values are uniquely determined as b=γ2G2δ2τd and k=γGδΔ, where δ and Δ are the duration and separation of the gradient pulse, respectively, and τd=Δδ3. Here, the relationship between k and b is given by

k2=bΔ2τd. (6)

From here, the Stejskal–Tanner pulse is assumed as the gradient pulse unless otherwise noted.

2.2. Intravoxel signal attenuation

To extend Equation 4 to the total signal attenuation in a voxel, an integral form with respect to flow velocity was derived in PC 32 and DW 14 , 31 imaging studies, as follows.

SS0=exp(bD)f(v)exp(ikv)dv, (7)

where signals S and S0 are summations of m and m0 distributed in a voxel, respectively, and f(v) is the probability density function of v. Since the above integral is equivalent to the Fourier transform of f(v) 32 (i.e., the characteristic function 36 ), Equation 7 can be rewritten as

SS0=exp(bD)F^(k)signal attenuation·exp(ikvm)phase shift, (8)

where F^(k) is the characteristic function of f^(v)=f(vvm) by the shift theorem, and vm is the mean of f(v) extracted in PC imaging. From Equation 8, the ADC is expressed as

ADC:=1blnSS0=D1bln|F^(k)|,SS0=exp(bD)F^(k). (9)

Thus, the ADC can be understood as a summation of the diffusion coefficient D and the intravoxel flow distributions independent of vm. Since |F^(k)|1 36 and b0, the ADC is positive.

Remark

Bito et al. 14 pointed out that the limit of the ADC as b goes to zero is uniquely determined in arbitrary f^(v), as follows.

limb0ADC=D+σv2Δ22τd, (10)

where σv2 is the variance of the intravoxel flow velocity. From Equation 8, this limit operation can be understood from the limit of F^(k) as k goes to zero. Consider F^(k) with small k,

F^(k)=f^(v)1ikvk2v22+O(k3)dvf^(v)dvikf^(v)vdvk22f^(v)v2dv=1k2σv22, (11)

which is equivalent to that of a normal distribution with zero mean and variance of σv2, such that

F^(k)=expk2σv22,=1k2σv22+𝒪(k3). (12)

Thus, substituting Equation 12 into Equation 9 leads to Equation 10 without a limit operator, such that

ADC=D1blnexpk2σv22=D+σv2Δ22τd. (13)

The correspondence of Equations 11 and 12 is commonly used in the proof of the central limit theorem (e.g., Reference [37]), where high‐order terms of the characteristic function are omitted as the number of trials increases.

2.3. f^(v) and F^(k) in laminar flow

The flow velocity field is assumed to be laminar and sufficiently slow such that it can be expressed as a low‐order polynomial function within a voxel. Furthermore, the flow field is assumed to be temporally constant, and the flow variance discussed in the following sections reflects only spatial variation.

Let us consider a local coordinate system defined within a voxel (right‐handed), where the origin is placed at the voxel center and the orthogonal unit vectors are aligned with the voxel edges. Assuming that v varies linearly in space, it is given by

v(x,y,z)=v0+vxx+vyy+vzz=v0+a1x+a2y+a3z, (14)

where v0 is the velocity at the voxel center, which coincides with the spatial average of v within the voxel, and ai (i=1,2,3) is the spatial velocity gradients along the respective coordinate directions. Here, the voxel size is given by [L1,L1]×[L2,L2]×[L3,L3] and the range of v is normalized to [vd,vd].

Following the above definition, f^(v) is defined based on a cross‐sectional area of the v iso‐surface in the voxel (Figure 1 (left)). First, if two components are negligible (e.g., a2=a3=0), the f^(v) is a uniform distribution denoted as f^1D(a1,L1), given by

f^1D(a1,L1)=12a1L1,vd=a1L1. (15)

Thus, the corresponding F^(k) is a sinc function, such that

F^(k)=sinc(ka1L1). (16)

In general cases (a1,a2,a30), f^(v) is expressed as a convolution integral 37 of the uniform distribution f^1D assigned to each axis of interest, such that

f^(v)=f^1D(a1,L1)f^1D(a2,L2)f^1D(a3,L3),vd=i=13aiLi, (17)

where denotes the convolution integral. The variance of flow velocity σv2 is also generalized as

σv2=(a1L1)2+(a2L2)2+(a3L3)23. (18)

Finally, the corresponding F^(k) is given by

F^(k)=i=13sinc(kaiLi). (19)

FIGURE 1.

FIGURE 1

Representative cases of v iso‐surfaces in an isotropic voxel (left), corresponding probability density functions f^(v) of the flow velocity v[vd,vd] (center), and the characteristic function F^(k) with σv=1,0.1, and 0.01 (right) in cases that (i) the velocity gradient is negligible in two‐directions, (ii) the velocity gradient is negligible in one direction, and (iii) the velocity gradient is comparable in all directions.

Figure 1 (center and right) shows representative examples of f^(v) curves with constant vd and F^(k) curves with σv=0.01,0.1, and 1. When f^(v) follows a uniform distribution, F^(k) becomes a sinc function and exhibits oscillatory behavior around zero as k increases. As velocity gradients become more multidirectional and comparable in magnitude, f^(v) approaches a triangular‐like distribution, and the oscillations in F^(k) become weak. Curves of F^(k) between original and corresponding normal distributions with the same σv are closed when F^(k) is higher than 0.5, regardless of flow distributions. Although F^(k) in normal distributions mildly decay compared to those of the original F^(k), these differences are at most 4% in the case of a1=a2=a3. This excellent agreement is reasonable because the multiple convolution integrals of the probability density function, which is independent and identically distributed, approach the normal distribution based on the central limit theorem. 37

3. METHODS

The fundamental properties of the intravoxel signal attenuation and the physical interpretation of the DWI and IVIM imaging of the laminar flow were investigated with consideration of the CSF flow imaging. We set D of the CSF to 3.0×103 mmInline graphic/s as the diffusion coefficient of pure water at 37°C. 38 For modeling F^(k) in Equation 19, we considered a simplified case with isotropic voxel (L1=L2=L3) and velocity gradients (a1=a2=a3), which represents a fully three‐dimensional flow distribution in a representative manner.

The DWI signal attenuations were evaluated under ideal and noise‐affected conditions. In the latter case, the background noise of |S/S0| was modeled using a Rician distribution, 39 with the signal‐to‐noise ratio (SNR) set to 20 as a representative value. Assuming that the true MRI signal is zero, the mean value of the background noise ϵ is given by 39

ϵ=σπ/2 (20)

where σ=1/SNR, assuming that the S0 is normalized to 1. This value ϵ was used to evaluate the non‐zero baseline in the low‐signal regime (i.e., noise floor 40 ).

First, the ranges of σv detectable using DWI were investigated. To consider the relative extents of the intravoxel flow velocity distributions with respect to the diffusion effect, we introduced the scaling factor α0 from Equation 10 and rewrote the ADC (Equation 13) in the sufficiently low‐b case (b0), as follows.

σv2Δ22τd=αD, (21)
limb0ADC=(1+α)D. (22)

For the evaluation, we set α=0 (pure diffusion), 0.1,1,10,100, and 1000. Furthermore, the sensitivities of δ and Δ were evaluated in the case of σv=0.4 mm/s and b=100 s/mmInline graphic. Both δ and Δ were set from 0 ms to 60 ms, and Δδ were set according to the definitions.

Next, the IVIM model fitting was applied to the signal attenuations of the representative three cases of α = 0.1, 1, and 10. The IVIM model is defined as the following bi‐exponential function 18 :

SS0IVIM:=fVOFexp(bDp)perfusion+(1fVOF)exp(bDd)diffusion, (23)

where Dp is the pseudo diffusion coefficient originating from the perfusion effect, Dd is the molecular diffusion coefficient estimated in this model, and fVOF is the volume fraction of the perfusion components (fVOF[0,1]). To consider practical conditions, we extracted signal attenuations at b=0,50,100,250,500, and 1000 s/mmInline graphic based on Reference [28]. Based on the Rician noise characteristics described above, Monte Carlo simulations of the IVIM model fitting were performed with a number of trials of 104 in each case of α. These fits were computed using the curve‐fit algorithm implemented in SciPy 41 with the constraints of Dp>Dd5×105 mmInline graphic/s for stability.

4. RESULTS

Figure 2 shows signal attenuation curves with increasing b in representative cases of α. The signal of α = 0.1 was almost equal to that of α=0 (pure diffusion) at b<10 s/mmInline graphic, was slightly lower at b>10 s/mmInline graphic, and then approached zero at b of around 1000 s/mmInline graphic. These signal curves shifted to lower b with increasing α, and that of α = 1000 decayed at low b from 0.01 to 1 s/mmInline graphic and approached zero at b1 s/mmInline graphic.

FIGURE 2.

FIGURE 2

Degree of signal attenuation with increasing b in representative cases with intravoxel velocity disturbances of α = 0 (diffusion only), 0.1, 1, 10, 100, and 1000.

For practical interpretation of the detectable velocity range, the signal attenuation curves with respect to σv are summarized for different b at (δ,Δ) = (20, 40) ms and (20, 100) ms, as shown in Figure 3. In the case of (δ,Δ) = (20, 40) ms, the signal decreased in σv of 𝒪(1) mm/s at b=1 s/mmInline graphic. The curves were shifted to lower σv with increasing b, while the signal baselines (e.g., those at σv=0 mm/s) also became lower and reached values comparable to the noise floor (ϵ0.06) at b=1000 s/mmInline graphic. In addition, the signal attenuation curves were shifted to relatively low σv as Δ increased, as shown in the case of (δ,Δ) = (20,100) ms.

FIGURE 3.

FIGURE 3

Sensitivities of σv on signal attenuation in the ranges of b = 1, 10, 50, 100, 200, 500, and 1000 at (δ, Δ) = (20, 40) ms (20, 100) ms, as representatives. Dashed lines in red in horizontal directions show the mean of background noises (noise floor) estimated as a Rician distribution (SNR = 20).

The effects of δ and Δ on the extent of signal attenuation are summarized in Figure 4 in the representative case of σv = 0.4 mm/s and b = 100 s/mmInline graphic. The signal decreased monotonically, while the degree of attenuation became mild with increasing δ and Δ, and ranged from approximately 0.7 to 0.3.

FIGURE 4.

FIGURE 4

Sensitivities of δ and Δ on signal attenuations in the ranges of δ[0,60] and Δ[0,60] at b = 100 s/mmInline graphic and σv = 0.4 mm/s.

Finally, the IVIM model was fitted to the signal attenuation curves obtained at multiple b‐values. Figure 5 shows IVIM curves fitted to the mean of the signals with Rician noise (solid lines) and the original noise‐free signal attenuations (dashed lines) in three representative cases of α= 0.1, 1, and 10. As the b increased, the signals approached the noise floor, and thus resulted in bi‐exponential‐like curves whose characteristics depend on the flow effects α. In this setting, the IVIM parameters (particularly fVOF and Dp) reflected the slope of flow‐induced signal attenuation in the low‐b range. The corresponding values and their variability are provided in Supplementary 1.

FIGURE 5.

FIGURE 5

Signal attenuations at b = 0, 50, 100, 250, 500, and 1000 in cases of α = 0.1, 1, and 10. Shaded domains show the range of noise of SNR = 20, and the dotted line shows the mean of the noise (noise floor). Solid and dashed lines show the fitting curves of the IVIM model (Equation 23) on the mean signal values and original signal attenuation curves without noise, respectively. The kink observed in the dashed line for α=10 near |S/S0|=102 is attributed to the oscillatory behavior of the F^(k) in the signal attenuation curve.

5. DISCUSSION

This study aimed to extend the existing theory of the DWI of fluid flow with sufficiently low b 14 to the generalized theoretical framework in arbitrary b by combining knowledge on both DWI 14 and PC 32 imaging. The semi‐analytical ADC expression (Equation 9) consists of the molecular diffusion term and a flow term, determined solely by the velocity distribution and independent of mean velocity. From this theory, the ADC limit with sufficiently low b 14 can be understood from the property of the characteristic function F^(k).

Using the theoretically derived signal attenuation, we investigated the range of CSF flow velocities detectable under clinically relevant scan settings. The framework demonstrated that DWI is capable of detecting signal attenuation originating from flow effects when the velocity standard deviation σv falls within approximately 0.1–10 mm/s (Figure 3). If the molecular diffusion coefficient D is known a priori (e.g., the imaging target is pure fluids such as CSF), the flow contribution can be extracted by subtracting the molecular diffusion effects from the signal attenuations. Moreover, the signal intensity is influenced not only by b, but also by δ and Δ 15 (Figure 4). Thus, to estimate flow‐related contributions more robustly, one possible strategy is to vary δ and Δ while keeping b constant. This allows the diffusion‐related component of the signal to remain fixed, while the flow‐related component varies according to changes in the gradient waveform. Therefore, signal differences across multiple (δ,Δ) combinations at constant b could help extract intravoxel flow variance. Although further validation with actual MRI sequences and implementation would be required, this theoretical framework provides a promising basis for future development of DWI‐based flow quantification.

Furthermore, we explored how IVIM model fitting behaves when applied to single‐component flow fields, despite the original IVIM model being designed for two‐compartment systems. Under noise‐affected conditions, the signal attenuation curves approached a bi‐exponential‐like shape due to the noise floor (Figure 5), and the fitted IVIM parameters (particularly fVOF and Dp) reflected the slope of the flow‐induced signal attenuation in the low‐b region (Supplementary 1). Although the fitted parameters are not consistent with their original physical meaning, these findings may help in understanding recent IVIM studies on ventricular CSF flow, 26 , 27 , 28 , 29 where a single dominant flow component is likely present.

This study has three main limitations and perspectives for future work. The first is the assumption of isotropic velocity gradients (a1=a2=a3), which was adopted to simplify the modeling and enable tractable analysis of signal attenuation behavior. While the signal attenuation curves are consistent with same σv when F^(k)>0.5 regardless of the flow distribution (Figure 1), anisotropic flow profiles can induce oscillatory features in F^(k), particularly in low‐signal regimes (Figure 1). As these regimes are often dominated by low SNR (Figure 2), further investigation would be needed to fully capture the impact of anisotropy under realistic acquisition conditions. The second is that we assumed the flow velocity to be sufficiently slow such that motion artifacts due to magnetization advection 42 could be neglected. Since the repetition time of general DWI using echo‐planar imaging is on the order of 𝒪(1) seconds, this assumption may be critical for relatively high‐velocity flow fields. 43 In such cases, the transport of magnetization over the duration of the pulse sequence may introduce significant artifacts, and theoretical modeling of these conditions becomes more challenging. The third is the assumption of steady flow. For ventricular CSF flow, the velocity is typically slow and synchronized with the cardiac cycle, and thus, the unsteady effects can be reduced by ECG‐gated DWI acquisition. However, in more general cases involving arbitrary time‐varying flows, unsteady effects may lead to non‐negligible artifacts. These effects arise from the steady‐flow assumption used in the derivation of b (Equation 4) and are similar to temporal misregistration artifacts observed in phase‐contrast MRI under similar assumptions. 44 To address the above limitations, computational simulation of flow MRI under unsteady and anisotropic flow conditions 45 , 46 would be useful in future work. By modeling the spatiotemporal evolution of magnetization in arbitrary flow velocity fields, such simulations may help interpret the resulting DWI signals and the associated artifacts.

6. CONCLUSIONS

This study developed a general theoretical framework to understand the physical implications of DWI and IVIM imaging with arbitrary b. According to this theory, DWI can detect intravoxel flow velocity standard deviations ranging from 0.1 to 10 mm/s under practical conditions. Furthermore, the IVIM parameter fits of the single flow domain provide the effects of the intravoxel flow velocity distribution, although the original meaning of the IVIM model is inconsistent in this situation. These examples successfully highlight the usefulness of the developed theoretical framework, and therefore, we expect that this framework can provide attractive insights for understanding the DWI of fluid flow, help parameter tuning to detect the preferred flow velocity range, and assist in the development of further advanced imaging protocols.

CONFLICT OF INTEREST STATEMENT

The authors declare no potential conflict of interest.

FUNDING INFORMATION

This work was partially supported by Japan Society for the Promotion of Science Grants‐in‐Aid for Scientific Research (Grant Nos. 23K11830, 24K02408, 24K02557, 25K15857, and 25K03452); Program for Promoting Research on the Supercomputer Fugaku (Development of human digital twins for cerebral circulation using Fugaku) funded by the Ministry of Education, Culture, Sports, Science and Technology (Grant No. JPMXP1020230118); and the Nakatani Foundation.

Supporting information

Table S1. IVIM model parameters (N = 10 000).

MRM-95-628-s001.pdf (147.5KB, pdf)

ACKNOWLEDGMENTS

We thank Satoshi Ii (Institute of Science Tokyo), Marie Oshima (The University of Tokyo), Tetsuro Sekine (Nihon Medical School), and Kota Kitamura (The University of Osaka) for fruitful discussions. We thank Edanz (https://jp.edanz.com/ac) for editing a draft of this manuscript.

Otani T., Bito Y., Yamada S., Watanabe Y., and Wada S., “A theoretical interpretation of diffusion weighted and intravoxel incoherent motion imaging for cerebrospinal fluid flow,” Magnetic Resonance in Medicine 95, no. 1 (2026): 628–636, 10.1002/mrm.70062.

DATA AVAILABILITY STATEMENT

The data that support the findings of this study are available from the corresponding author upon reasonable request.

REFERENCES

  • 1. Bohr T, Hjorth PG, Holst SC, et al. The glymphatic system: Current understanding and modeling. iScience. 2022;25:104987. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Agarwal N, Lewis LD, Hirschler L, et al. Current understanding of the anatomy, physiology, and magnetic resonance imaging of neurofluids: Update from the 2022 ISMRM imaging neurofluids study group workshop in Rome. J Magnet Reson Imaging. 2024;59:431‐449. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Yamada S, Ishikawa M, Ito H, et al. Cerebrospinal fluid dynamics in idiopathic normal pressure hydrocephalus on four‐dimensional flow imaging. Eur Radiol. 2020;30:4454‐4465. [DOI] [PubMed] [Google Scholar]
  • 4. Vikner T, Johnson KM, Cadman RV, et al. CSF dynamics throughout the ventricular system using 4D flow MRI: Associations to arterial pulsatility, ventricular volumes, and age. Fluids Barriers CNS. 2024;21:68. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Taoka T, Masutani Y, Kawai H, et al. Evaluation of glymphatic system activity with the diffusion MR technique: Diffusion tensor image analysis along the perivascular space (DTI‐ALPS) in Alzheimer's disease cases. Jpn J Radiol. 2017;35:172‐178. [DOI] [PubMed] [Google Scholar]
  • 6. Wright AM, Wu Y‐C, Feng L, Wen Q. Diffusion magnetic resonance imaging of cerebrospinal fluid dynamics: Current techniques and future advancements. NMR Biomed. 2024;37:e5162. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Ishida S, Yamada S, Oki T, et al. Evaluation of interstitial fluid volume and diffusivity in patients with idiopathic normal pressure hydrocephalus using spectral diffusion analysis. J Magnet Reson Imaging. 2025. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Yamada S, Miyazaki M, Kanazawa H, et al. Visualization of cerebrospinal fluid movement with spin labeling at MR imaging: Preliminary results in normal and pathophysiologic conditions. Radiology. 2008;249:644‐652. [DOI] [PubMed] [Google Scholar]
  • 9. Petitclerc L, Hirschler L, Örzsik B, Asllani I, Osch MJP. Arterial spin labeling signal in the CSF: Implications for partial volume correction and blood‐CSF barrier characterization. NMR Biomed. 2023;36:e4852. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Wu Y, Xu F, Zhu D, et al. Cerebrospinal fluid flow within ventricles and subarachnoid space evaluated by velocity selective spin labeling MRI. NeuroImage. 2025;309:121095. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Diorio TC, Nair VV, Patel NM, Hedges LE, Rayz VL, Tong Y. Real‐time quantification of in vivo cerebrospinal fluid velocity using the functional magnetic resonance imaging inflow effect. NMR Biomed. 2024;37:e5200. [DOI] [PubMed] [Google Scholar]
  • 12. Harrison IF, Siow B, Akilo AB, et al. Non‐invasive imaging of CSF‐mediated brain clearance pathways via assessment of perivascular fluid movement with diffusion tensor MRI. eLife. 2018;7:e34028. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Evans PG, Sajic M, Yu Y, et al. Changes in cardiac‐driven perivascular fluid movement around the MCA in a pharmacological model of acute hypertension detected with non‐invasive MRI. J Cereb Blood Flow Metabol. 2024;44:508‐515. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Bito Y, Harada K, Ochi H, Kudo K. Low b‐value diffusion tensor imaging for measuring pseudorandom flow of cerebrospinal fluid. Magn Reson Med. 2021;86:1369‐1382. [DOI] [PubMed] [Google Scholar]
  • 15. Bito Y, Ochi H, Shirase R, Yokohama W, Harada K, Kudo K. Low b‐value diffusion tensor imaging to analyze the dynamics of cerebrospinal fluid: Resolving intravoxel pseudorandom motion into ordered and disordered motions. Magn Reson Med Sci. 2025;24(1):46‐57. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. Chen Y, Hong H, Nazeri A, Markus HS, Luo X. Cerebrospinal fluid‐based spatial statistics: Towards quantitative analysis of cerebrospinal fluid pseudo diffusivity. Fluids Barriers CNS. 2024;21:59. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Nazeri A, Hosseini H, Dehkharghanian T, et al. Characterizing the spatial patterns and determinants of cerebrospinal fluid pseudorandom flow in the human brain with low B‐value diffusion MRI. Imaging Neurosci. 2025;3:imag_a_00473. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Le Bihan D, Breton E, Lallemand D, Aubin ML, Vignaud J, Laval‐Jeantet M. Separation of diffusion and perfusion in intravoxel incoherent motion MR imaging. Radiology. 1988;168:497‐505. [DOI] [PubMed] [Google Scholar]
  • 19. Federau C, Maeder P, O'Brien K, Browaeys P, Meuli R, Hagmann P. Quantitative measurement of brain perfusion with intravoxel incoherent motion MR imaging. Radiology. 2012;265:874‐881. [DOI] [PubMed] [Google Scholar]
  • 20. Federau C. Intravoxel incoherent motion MRI as a means to measure in vivo perfusion: A review of the evidence. NMR Biomed. 2017;30:e3780. [DOI] [PubMed] [Google Scholar]
  • 21. Dietrich O, Cai M, Tuladhar AM, et al. Integrated intravoxel incoherent motion tensor and diffusion tensor brain MRI in a single fast acquisition. NMR Biomed. 2023;36:e4905. [DOI] [PubMed] [Google Scholar]
  • 22. Yamada S, Hiratsuka S, Otani T, et al. Usefulness of intravoxel incoherent motion MRI for visualizing slow cerebrospinal fluid motion. Fluids Barriers CNS. 2023;20:16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Levendovszky Swati R, Flores J, Peskind ER, Václavů L, Osch MJ, Iliff J. Preliminary investigations into human neurofluid transport using multiple novel non‐contrast MRI methods. J Cereb Blood Flow Metabol. 2024;44:1580‐1592. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Iima M, Le Bihan D. Clinical intravoxel incoherent motion and diffusion MR imaging: Past, present, and future. Radiology. 2016;278:13‐32. [DOI] [PubMed] [Google Scholar]
  • 25. Le Bihan D. What can we see with IVIM MRI? NeuroImage. 2019;187:56‐67. [DOI] [PubMed] [Google Scholar]
  • 26. Becker AS, Boss A, Klarhoefer M, Finkenstaedt T, Wurnig MC, Rossi C. Investigation of the pulsatility of cerebrospinal fluid using cardiac‐gated Intravoxel Incoherent Motion imaging. NeuroImage. 2018;169:126‐133. [DOI] [PubMed] [Google Scholar]
  • 27. Surer E, Rossi C, Becker AS, et al. Cardiac‐gated intravoxel incoherent motion diffusion‐weighted magnetic resonance imaging for the investigation of intracranial cerebrospinal fluid dynamics in the lateral ventricle: A feasibility study. Neuroradiology. 2018;60:413‐419. [DOI] [PubMed] [Google Scholar]
  • 28. Yamada S, Otani T, Ii S, et al. Modeling cerebrospinal fluid dynamics across the entire intracranial space through integration of four‐dimensional flow and intravoxel incoherent motion magnetic resonance imaging. Fluids Barriers CNS. 2024;21:47. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Voorter PHM, Jansen JFA, Thiel MM, et al. Diffusion‐derived intravoxel‐incoherent motion anisotropy relates to CSF and blood flow. Magn Reson Med. 2025;93:930‐941. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Jang M, Han S, Cho H. D* from diffusion MRI reveals a correspondence between ventricular cerebrospinal fluid volume and flow in the ischemic rodent model. J Cereb Blood Flow Metab. 2022;42:572‐583. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Ahlgren A, Knutsson L, Wirestam R, et al. Quantification of microcirculatory parameters by joint analysis of flow‐compensated and non‐flow‐compensated intravoxel incoherent motion (IVIM) data: Joint Analysis of Flow‐Compensated and Non‐Flow‐Compensated IVIM Data. NMR Biomed. 2016;29:640‐649. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Dyverfeldt P, Sigfridsson A, Kvitting Escobar J‐P, Ebbers T. Quantification of intravoxel velocity standard deviation and turbulence intensity by generalizing phase‐contrast MRI. Magn Reson Med. 2006;56:850‐858. [DOI] [PubMed] [Google Scholar]
  • 33. Petter D, Escobar KJ‐P, Andreas S, Jan E, Bolger Ann F, Tino E. Assessment of fluctuating velocities in disturbed cardiovascular blood flow: In vivo feasibility of generalized phase‐contrast MRI. J Magnet Reson Imaging. 2008;28:655‐663. [DOI] [PubMed] [Google Scholar]
  • 34. Stejskal EO. Use of spin echoes in a pulsed magnetic‐field gradient to study anisotropic, restricted diffusion and flow. J Chem Phys. 1965;43:3597‐3603. [Google Scholar]
  • 35. Price WS. Pulsed‐field gradient nuclear magnetic resonance as a tool for studying translational diffusion: Part 1. Basic theory. Concepts Magn Reson. 1997;9:299‐336. [Google Scholar]
  • 36. Lukacs E. Characteristic Functions. Charles Griffin & Company; 1970. [Google Scholar]
  • 37. Rohatgi Vijay K, Ehsanes Saleh AK. An Introduction to Probability and Statistics. Wiley Series in Probability and Statistics. 2nd ed. John Wiley & Sons; 2000. [Google Scholar]
  • 38. Mills R. Self‐diffusion in normal and heavy water in the range 1‐45.deg. J Phys Chem. 1973;77:685‐688. [Google Scholar]
  • 39. Gudbjartsson H, Patz S. The Rician distribution of noisy MRI data. Magnet Reson Med. 1995;34:910‐914. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Iima M, Yano K, Kataoka M, et al. Quantitative non‐Gaussian diffusion and intravoxel incoherent motion magnetic resonance imaging: Differentiation of malignant and benign breast lesions. Investig Radiol. 2015;50:205‐211. [DOI] [PubMed] [Google Scholar]
  • 41. Virtanen P, Gommers R, Oliphant TE, et al. SciPy 1.0: Fundamental algorithms for scientific computing in python. Nature Methods. 2020;17:261‐272. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42. Kouwenhoven M, Hofman MB, Sprenger M. Motion induced phase shifts in MR: Acceleration effects in quantitative flow measurements–a reconsideration. Magnet Reson Med. 1995;33:766‐777. [DOI] [PubMed] [Google Scholar]
  • 43. Dillinger H, Walheim J, Kozerke S. On the limitations of echo planar 4D flow MRI. Magnet Reson Med. 2020;84:1806‐1816. [DOI] [PubMed] [Google Scholar]
  • 44. Doyle M, Kortright E, Anayiotos AS, et al. Correction of temporal misregistration artifacts in jet flow by conventional phase‐contrast MRI. J Magnet Reson Imaging. 2007;25:1256‐1262. [DOI] [PubMed] [Google Scholar]
  • 45. Puiseux T, Sewonu A, Moreno R, Mendez S, Nicoud F. Numerical simulation of time‐resolved 3D phase‐contrast magnetic resonance imaging. PLoS One. 2021;16:e0248816. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46. Otani T, Sekine T, Sato Y, Cavalcante Alves E, Wada S. An Eulerian formulation for the computational modeling of phase‐contrast MRI. Magnet Reson Med. 2025;93:828‐841. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Table S1. IVIM model parameters (N = 10 000).

MRM-95-628-s001.pdf (147.5KB, pdf)

Data Availability Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request.


Articles from Magnetic Resonance in Medicine are provided here courtesy of Wiley

RESOURCES