Abstract
Purpose
‐separation is an advanced quantitative susceptibility mapping (QSM) method that is designed to generate paramagnetic () and diamagnetic () susceptibility maps, reflecting the distribution of iron and myelin in the brain. However, vessels have shown artifacts, interfering with the accurate quantification of iron and myelin in applications. To address this challenge, a new vessel segmentation method for ‐separation is developed.
Methods
The method comprises three steps: (1) seed generation from and the product of and maps; (2) region growing, guided by vessel geometry, creating a vessel mask; (3) refinement of the vessel mask by excluding non‐vessel structures. The performance of the method was compared to other vessel segmentation methods both qualitatively and quantitatively. To demonstrate the utility of the method, it was tested in two applications: quantitative evaluation of a neural network‐based ‐separation reconstruction method (‐sepnet‐) and population‐averaged region of interest (ROI) analysis.
Results
The proposed method demonstrates superior performance to other vessel segmentation methods, effectively excluding the non‐vessel structures, achieving the highest Dice score coefficient against manually segmented vessel masks (3 T: 76.7% for and 68.7% for , 7 T: 76.9% for and 72.6% for ). For the applications, applying vessel masks report notable improvements for the quantitative evaluation of ‐sepnet‐ and statistically significant differences in population‐averaged ROI analysis. These applications suggest excluding vessels when analyzing the ‐separation maps provide more accurate evaluations.
Conclusion
The proposed method has the potential to facilitate various applications, offering reliable analysis through the generation of a high‐quality vessel mask.
Keywords: ‐separation, image analysis, vessel segmentation, x‐separation
1. INTRODUCTION
Iron and myelin have critical roles in normal brain functions, 1 , 2 and alterations in their levels are often associated with neurodegenerative diseases such as Parkinson's disease (PD), Alzheimer's disease (AD), and multiple sclerosis (MS). 3 , 4 , 5 , 6 , 7 These changes highlight the potential of iron and myelin as biomarkers for the diseases, suggesting the need for imaging techniques that can accurately quantify iron and myelin.
‐separation (chi‐separation or x‐separation) is an advanced QSM method that generates paramagnetic and diamagnetic susceptibility maps, potentially reflecting the distribution of iron and myelin in the brain. 8 , 9 This method combines local field information with reversible transverse relaxation rates () to separate the paramagnetic and diamagnetic susceptibility sources within a voxel. The method has been applied to various studies in evaluating neurological disorders. 10 , 11 , 12 , 13 , 14 , 15 , 16 Additionally, similar susceptibility source separation techniques have been developed, expanding the scope of potential applications. 17 , 18 , 19 , 20
In ‐separation, vessels have shown to create erroneous artifacts, hampering its applications. 9 , 21 Large susceptibility differences around veins can induce mesoscopic field inhomogeneities, 22 , 23 creating non‐local (or ) effects. These effects contradict with the assumption that (or ) effects are fully localized in each voxel, 9 resulting in artifacts near the veins. Furthermore, flow inside the vessels causes spatial displacement, 24 , 25 leading to inconsistent signal decay across echo times within and near the vessels. This inconsistency results in inaccurate values, propagating errors in ‐separation maps. 21 , 26 These vessel artifacts can interfere with the accurate quantification of iron and myelin in some applications. 27 , 28 For reliable analysis, the exclusion of vessels via vessel segmentation is an option.
Until now, several vessel segmentation methods have been developed and applied to susceptibility imaging such as SWI and QSM for disease assessment or oxygen extraction fraction (OEF). 29 , 30 , 31 , 32 , 33 , 34 , 35 , 36 , 37 , 38 , 39 , 40 Among them, Hessian‐based vessel enhancement filters, such as Frangi filter, 41 stand out because of their effectiveness in highlighting tubular structures. The eigenvalues of the second order derivatives matrix (Hessian matrix) of the image intensity have been used to calculate the “vesselness,” which indicates the likelihood of each voxel belonging to a vessel. 41 Based on a similar idea, a few important improvements have been made. 42 , 43 , 44 , 45 However, applying only Hessian‐based vessel enhancement filters to susceptibility images poses challenges in differentiating vessels from regions with high susceptibility values such as deep brain nuclei. To address this challenge, Monti et al.46 and Straub et al.47 have been proposed to apply the Hessian‐based vessel enhancement filter with additional steps to reduce false positive voxels. These approaches improve vesselness by suppressing contrast for non‐vessel structures using a high‐pass filter, such as an inverse Hamming filter, before applying the vessel enhancement filter 47 , 48 and by incorporating a regularization function that penalizes regions with large field inhomogneities. 46 , 47 Additionally, Monti et al. and Straub et al. proposed to exclude falsely included voxels by removing small size clusters. 46 , 47
This study proposes a new vessel segmentation method for ‐separation. The method incorporates the physics of ‐separation and the geometry characteristics of vessels such that it creates a high‐quality vessel mask without non‐vessel structures. The method is expected to facilitate various applications, providing reliable analysis.
2. METHODS
2.1. Proposed vessel segmentation method
The overview of the proposed method is illustrated in Figure 1, and a more detailed depiction is provided in Figure S1. The method comprised three steps. In the first step, seeds for vessels were obtained. Next, region growing, guided by vessel geometry characteristics, was applied to generate a vessel mask. In the final step, this mask was refined by removing non‐vessel structures. This method used four inputs: , paramagnetic susceptibility () and diamagnetic susceptibility () maps, and a brain mask, all of which were generated during the ‐separation processing.
FIGURE 1.

Overview of the proposed pipeline for vessel segmentation. The pipeline has three steps: step 1 for seed generation, step 2 for an initial vessel mask created by region growing guided by the characteristics of vessel geometry, and step 3 for non‐vessel structures removal.
2.1.1. Step 1: Seed generation
In this step, a seed map for vessels was generated. Seeds for large and small vessels were obtained separately and then combined to form the final seed map. This approach improved identification of small vessels. The , , and maps, which exhibited high signal intensity in vessels (see Figure S2), were used as the inputs for the seed generation. The map was exploited for large vessels, whereas the product of the two susceptibility maps ) was applied to identify small vessels.
For the large vessel seed map, the approach proposed by Straub et al. 47 was adopted. A high pass filter (inverse Hamming filter) (see Appendix A.1) was first applied to the map, suppressing large non‐vessel structures of high (e.g., basal ganglia). 47 , 48 To remove high frequency residuals at the boundary of the brain, the map was inpainted outside the brain mask using coherent voxel values (inpaintCoherent, MATLAB) before the high pass filtering. 49 From this non‐vessel suppressed map, a “vesselness” () map was calculated by applying a multi‐scale fractional anisotropy tensor (MFAT) filter (see Appendix A.2). 44 , 47 Finally, a threshold with a high cutoff value (= ) was applied to the vesselness map, creating the seed map for large vessels.
For small vessels, maximum intensity projection (MIP) was applied to to enhance the visibility of small vessels, creating . 36 , 50 The process was conducted every 16 mm in axial, with half of the slices overlapping. During the MIP process, voxel positions corresponding to the maximum intensity position of were stored. To retain only small vessels, the large vessel seeds were removed from by negating the projection of large vessel seeds at the stored positions () and then multiplying it to . From this result, a vesselness map was calculated using the MFAT filter on each 2D MIP image, and then the map was binarized with a low threshold (= ). These seeds were back‐projected to their original locations in 3D using the stored voxel positions, creating the small vessel seed map. Finally, the large and small vessel seeds were combined to generate a final seed map.
2.1.2. Step 2: Vessel geometry guided‐region growing
In this step, two vessel masks, one for and the other for , were generated through region growing guided by the geometry of vessels. Both masks were initialized with the final seed map from step 1.
The region growing process began with defining a queue, which stores all the seed voxels of the final seed map. The queue was re‐ordered by the size of the seed cluster, prioritizing the largest seed cluster (bwconncomp, MATLAB) within which the smallest linear index of the voxel was queued first (sub2ind, MATLAB). Using this queue, the region growing algorithm ran as follows: the first element in the queue was selected as the starting seed voxel and removed from the queue. If an adjacent voxel of this seed voxel, which was not included in the vessel mask, met a region growing condition (see below), it was added to the queue and to the mask. The region growing process ended when the queue was empty.
The region growing condition is a criterion to include a voxel into the vessel mask. It has two susceptibility value limits:
| (1) |
| (2) |
where is either or map, is the final seed map, and and are hyperparameters to determine the intensity limits. If an adjacent voxel had an intensity higher than the upper limit, the voxel was added to the vessel mask. If the intensity of an adjacent voxel was lower than the lower limit, the voxel was ignored. If the intensity was between the upper and lower limits, the voxel was incorporated into the mask when it satisfied Eq. (3), which is composed of directionality similarity (Eq. (4)), intensity similarity (Eq. (5)), and anisotropy (Eq. (6)):
| (3) |
where
| (4) |
| (5) |
| (6) |
with is the location of the seed voxel, is the location of an adjacent voxel, is vesselness determined by the MFAT filter from the susceptibility map (either or ), and denotes the intensity of the image. is an eigenvector corresponding to the smallest eigenvalue of the Hessian matrix of the susceptibility map, indicating the direction of the vessel. and are the other two eigenvalues, which characterize variances in the directions perpendicular to the vessel direction. Eq. (3) was modified from Kerkeni et al., 51 which combined vesselness and directional information of vessels (i.e., directionality similarity). The two new criteria, intensity similarity and anisotropy, were designed to remove non‐vessel structures effectively in step 3 (see Figure S3). The anisotropy criterion was formulated in exponential form to express it as a probability value in Eq. (3) as suggested by Frangi et al. 41
This vessel geometry guided‐region growing generated initial vessel masks, one for and the other for .
2.1.3. Step 3: Non‐vessel structure removal
In this step, the initial vessel masks were refined by excluding non‐vessel structures, generating a final vessel mask.
In the initial vessel mask, non‐vessel structures such as globus pallidus could also be included because of their high susceptibility values (see CC2 and CC3 in Figure S1). To remove these structures, the structural characteristic of vessels, which reported a high anisotropy value, was exploited. The initial vessel mask was clustered into connected components (CCs) and then CCs with low anisotropy were removed as follows:
| (7) |
where indicates the number of voxels in a CC, denotes the nth CC, is an anisotropy threshold, and is anisotropy criterion defined in step 2.
The final vessel mask was generated by accumulating the remaining CCs and binarizing it. Two masks, one for and the other for , were produced.
2.2. MRI data acquisition and data processing
In this study, three datasets from previous studies were used: the ‐sepnet dataset, 10 , 52 the ‐separation template dataset, 28 and the high‐resolution ‐separation dataset. 53 The study was approved by the institutional review board.
The ‐sepnet dataset is from 12 subjects and is composed of 3D multi‐echo gradient echo (GRE) data in six head orientations and 2D multi‐echo spin echo (MESE) data in a single head orientation (3 T, Siemens Tim Trio). The ‐separation template dataset has 106 subjects, with 3D multi‐echo GRE images and T1‐weighted images obtained using MPRAGE (3 T, Philips Ingenia CX and Ingenia Elition X). In the high‐resolution ‐separation dataset, 3D multi‐echo GRE data from eight subjects were used (7 T, Siemens Magnetom Terra). The acquisition parameters of the three datasets are summarized in Table S1.
For all three datasets, a common data processing pipeline was applied: a brain mask was generated from the first echo magnitude image of the multi‐echo GRE data, using the brain extraction tool (FSL, FMRIB). 54 Phase processing for a local field map followed the QSM consensus guideline. 55 Briefly, phase images from the multi‐echo GRE data were unwrapped using the rapid opensource minimum spanning tree algorithm (ROMEO). 56 The unwrapped phase images were averaged using a weighted echo sum to produce a combined phase image. 57 Next, background field removal using V‐SHARP was applied, creating a local field map. 58 , 59 From the multi‐echo GRE magnitude images, an map was generated by voxel‐wise fitting of a mono‐exponential decay function using a nonlinear least square solver (lsqnonlin, MATLAB). For mapping, a simulated dictionary of spin‐echo decay, constructed with the StimFit toolbox, 60 , 61 was used to match for the MESE magnitude images.
For the ‐sepnet dataset, the local field and maps from each orientation, as well as the map, were registered to the first head orientation using FSL FLIRT. 62 A registration matrix was computed from the first echo magnitude images of GRE and MESE, and subsequently applied to align the local field, , and maps. The map was then generated by subtracting the registered map from each orientation map, with negative values set to zero. Finally, and maps were generated via four ‐separation algorithms: ‐sep‐COSMOS, 63 ‐sep‐MEDI, 9 ‐sep‐iLSQR, 9 and ‐sepnet‐. 52 For ‐sep‐COSMOS, which is a multi‐orientation ‐separation algorithm, the local field and maps from all orientations were used. For ‐sep‐MEDI and ‐sep‐iLSQR, which are single‐orientation conventional ‐separation algorithms, the local field and maps from the first orientation were used as the input. Last, for ‐sepnet‐, a deep learning‐based ‐separation algorithm designed for single‐orientation GRE data only, the local field and maps from the first orientation were used. For the ‐separation template dataset, and maps were generated using ‐sepnet‐. For the high‐resolution ‐separation dataset, the pipeline proposed by Kim et al. 53 was applied to produce high‐resolution ‐separation maps.
Finally, the proposed vessel segmentation method was applied to all the ‐separation maps. Out of the hyperparameters, and were fixed as 0.5 and −0.5, respectively. Moreover, the anisotropy threshold () was adjusted for each subject with 1.2 10−3 as the starting point for both and maps. In most cases, this starting point value created a high‐quality outcome, however, in some cases, the value was manually increased in the increment of 0.3 10−3 when deep gray matter regions were not properly excluded (Figure S4).
2.3. Comparison with previously proposed vessel segmentation methods
The proposed vessel segmentation method was compared with two previously proposed methods: the Frangi filter 41 and a GRE‐based vessel segmentation method. 47 For the Frangi filter, a (or ) map was used as input. The Frangi filter had six parameters: scale range (), scale ratio (), Frangi vesselness constants (, , and ), and a threshold. 41 Based on previous studies, 35 , 47 the parameters were set as follows: , , , , half of the maximal Hessian norm, and threshold . The Frangi filter was publicly available (https://kr.mathworks.com/matlabcentral/fileexchange/24409‐hessian‐based‐frangi‐vesselness‐filter). The GRE‐based vessel segmentation method was originally developed to extract veins by leveraging QSM, SWI, and . To adapt this method for ‐separation, QSM was replaced by the (or ) map. The GRE‐based vessel segmentation included 10 hyperparameters, which were set as default. 47 This method was available online (https://github.com/SinaStraub/GRE_vessel_seg).
For the evaluation of the performances, three subjects from the ‐sepnet dataset (3 T) and three subjects from the high‐resolution ‐separation dataset (7 T) were manually segmented using ITK‐snap 64 to produce ground truth segmentation results (see Figure S5). This manual segmentation was conducted on central 12 consecutive slices on the axial, sagittal, and coronal planes for each subject (total 36 slices per subject), including deep gray matter regions and large and small cerebral vessels. For quantitative assessment, the Dice similarity coefficient (DSC) was calculated with respect to manually segmented vessel masks. 65 Additionally, processing time and memory usage of each method were evaluated.
All methods were executed on a workstation with an Intel Xeon CPU E5‐2699 v4 @ 2.20 GHz and 396 gigabytes (GB) random access memory (RAM).
2.4. Robustness of the proposed vessel segmentation method
To assess the robustness of the proposed vessel segmentation method, the method was applied to ‐separation maps from the four ‐separation algorithms (‐sep‐COSMOS, ‐sep‐MEDI, ‐sep‐iLSQR, and ‐sepnet‐). The three 3 T subject data with the manual segmentation masks were used to calculate DSC for the vessel mask generated from the ‐separation map of each algorithm.
2.5. Applications of the proposed vessel segmentation method
To demonstrate the utility of vessel masks when analyzing ‐separation maps, vessel masks were applied to two applications: quantitative evaluation of ‐sepnet‐ and population‐averaged region of interest (ROI) analysis.
To evaluate the effects of vessels to the quantitative metrics assessing the reconstruction performance of ‐sepnet‐, the root mean squared error (RMSE), peak SNR (PSNR), and structure similarity index (SSIM) were calculated in three conditions: whole brain including vessels (without applying the vessel mask), whole brain excluding vessels (with the vessel mask), and only within the vessel mask. Each metric was computed with respect to ‐sep‐COSMOS as the reference. The ‐sepnet‐ trained in Kim et al. 52 was used, and the evaluation was conducted using the test data from the ‐sepnet dataset (6 subjects with the 6 head orientations). Statistical significance between the metrics calculated with and without applying the vessel mask was analyzed using paired t‐tests, after verifying the normality of the differences with the Shapiro–Wilk test using IBM SPSS Statistics.
To assess the impact of vessels on the population‐averaged ROI analysis, the proportion of vessels and the population average of the mean susceptibility values with and without vessels were quantified across 27 ROIs defined in the ‐separation atlas. 28 Data from 106 subjects in the ‐separation template dataset were used for this analysis, and the ROIs were transformed into each subject's space using deformation matrices in Min et al. 28 The proportion of vessels in each ROI was calculated as the ratio of the number of voxels included in the vessel mask to the total number of voxels in the ROI. The population average of mean susceptibility values for each ROI was computed with and without vessels. A paired t‐test was conducted to discern statistically significant differences between two measurements, with significance determined at a Bonferroni‐corrected threshold of .
3. RESULTS
Vessel segmentation results of the two previously proposed methods and the proposed method are illustrated in Figure 2 for and Figure 3 for . The proposed method demonstrates superior performance by effectively excluding non‐vessel structures (yellow arrows in Figures 2 and 3: globus pallidus in and optic radiation in ), while maintaining sensitivity to vessels. Furthermore, other methods fail to capture parts of vessels (green arrows in Figures 2 and 3), whereas the proposed method achieves clear masking of vessels. In a few small vessels, however, the GRE‐based vessel segmentation method results in better outcomes than the proposed method (blue arrows in Figures 2 and 3). The superior performance of the proposed method is consistent across the resolutions of 1 1 1 mm3 (3 T) and 0.65 0.65 0.65 mm3 (7 T), which can be confirmed in the MIP images of Figure 4. The quantitative metrics in Table 1 consolidate that the proposed method achieves the best performance, reporting the highest DSCs calculated with respect to the manually segmented vessel masks.
FIGURE 2.

Results of the vessel segmentation methods applied to . The maps (first column) and the three vessel segmentation masks overlaid on (second column: Frangi filter, third column: gradient echo (GRE)‐based method, and fourth column: proposed method) are displayed. Three representative slices that include the globus pallidus (first row), a large vein (second row), and small vessels (third row) reveal that the proposed method effectively excludes non‐vessel structures (yellow arrows), providing a high‐quality vessel mask (green arrows). For small vessels, however, the GRE‐based method shows more sensitivity (blue arrows).
FIGURE 3.

Results of the vessel segmentation methods applied to . The maps (first column) and the three vessel segmentation masks overlaid on (second column: Frangi filter, third column: gradient echo (GRE)‐based method, and fourth column: proposed method) are displayed. Three representative slices that include the optic radiation (first row), cortical vessels (second row), and small vessels (third row) reveal that the proposed method effectively excludes non‐vessel structures (yellow arrows), providing a high‐quality vessel mask (green arrows). For small vessels, however, the GRE‐based method shows more sensitivity (blue arrows).
FIGURE 4.

Maximum intensity projection (MIP) of and maps, and vessel segmentation outcomes from 3 T data (upper rows, 1 1 1 mm3 resolution) and 7 T data (lower rows, 0.65 0.65 0.65 mm3 resolution). The previously proposed methods erroneously segment deep gray matter structures (yellow arrows) or miss large vessels (green arrows), whereas the proposed method delivers more accurate results.
TABLE 1.
Quantitative comparison of the vessel segmentation methods.
|
|
|
||||||||
|---|---|---|---|---|---|---|---|---|---|
| Frangi filter | GRE‐based method | Proposed | Frangi filter | GRE‐based method | Proposed | ||||
| DSC (%) | 1 1 1 mm3 (3 T) | 55.8 4.2 | 60.8 1.8 |
76.7
|
44.0 5.9 | 53.2 4.6 |
68.7
|
||
| 0.65 0.65 0.65 mm3 (7 T) | 65.5 3.4 | 66.4 1.4 |
76.9
|
55.6 5.5 | 63.9 3.8 |
72.6
|
|||
Note: This table reports DSCs obtained between vessel masks from each method and manually segmented vessel masks for 3 T and 7 T data. Best DSCs are highligted in bold.
Abbreviations: DSC, Dice similarity coefficient; GRE, gradient echo.
The proposed method significantly reduces both processing time and memory usage compared to the GRE‐based vessel segmentation method. For the 3 T data (matrix size: 256 224 176), the proposed method requires only 2 GB of RAM and takes 4 min for processing, while the GRE‐based vessel segmentation method demands approximately 60 GB of RAM and 28 min for processing. For the 7 T high‐resolution data (matrix size: 350 284 224), the proposed method needs 4 GB of RAM and 16 min, whereas the GRE‐based vessel segmentation method requires 140 GB of RAM and 80 min. The Frangi filter takes 76 s with 1.2 GB of RAM for processing the 3 T data and 3 min with 2 GB of RAM for processing the 7 T data.
The proposed vessel segmentation method demonstrates robust performance across the four ‐separation algorithms (Figure 5), showing consistent segmentation results for both and . Minor differences are observed because of the characteristics of each algorithm, particularly in cortical areas (yellow arrows in Figure 5; see Discussion). The DSC scores (Table 2) also confirm comparable segmentation performance across the four ‐separation algorithms.
FIGURE 5.

Vessel segmentation results from the four different ‐separation algorithms: ‐sep‐COSMOS, ‐sep‐MEDI, ‐sep‐iLSQR, and ‐sepnet‐. The results for (A) and (B) demonstrate consistent segmentation of vessels across all algorithms, showing the robustness of the proposed method. Minor differences occur in small vessels within cortical regions (yellow arrows).
TABLE 2.
DSCs between vessel masks from the four different ‐separation algorithms and manual segmentation.
| ‐sep‐COSMOS | ‐sep‐MEDI | ‐sep‐iLSQR | ‐sepnet‐ | |||
|---|---|---|---|---|---|---|
| DSC (%) |
|
76.7 4.2 | 72.0 3.8 | 72.0 4.0 | 73.6 4.1 | |
|
|
68.7 7.9 | 63.9 7.4 | 63.7 7.2 | 64.5 7.1 |
Abbreviations: DSC, Dice similarity coefficient.
When the vessel mask was applied to the performance evaluation of ‐sepnet‐ against ‐sep‐COSMOS, the quantitative metrics (RMSE, PSNR, and SSIM) report statistically significant improvements (p < 0.05; Table 3), demonstrating a potential value of masking out vessels if they are not of interest. Within the vessel mask, we observed higher RMSE and lower PSNR and SSIM than the non‐vessel regions (i.e., with vessel mask results).
TABLE 3.
Reconstruction quality of ‐sepnet‐ with respect to ‐sep‐COSMOS when analyzed with, without, and within the vessel mask.
|
|
|
|||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| RMSE (↓) | PSNR (↑) | SSIM (↑) | RMSE (↓) | PSNR (↑) | SSIM (↑) | |||||||
| ‐sepnet‐ (without vessel mask) | 0.0154 0.0015 | 36.3 0.8 | 0.926 0.007 | 0.0145 0.0012 | 36.8 0.7 | 0.920 0.009 | ||||||
| ‐sepnet‐ (with vessel mask) |
0.0121
|
38.4
|
0.927
|
0.0118
|
38.6
|
0.922
|
||||||
| p‐value (between without and with vessel mask) | 2.4 10−5 | 2.6 10−5 | 1.5 10−4 | 6.0 10−5 | 5.8 10−4 | 1.2 10−4 | ||||||
| ‐sepnet‐ (within vessel mask) | 0.0540 0.0061 | 25.4 1.0 | 0.894 0.009 | 0.0512 0.0060 | 25.9 1.0 | 0.840 0.016 | ||||||
Abbreviations: PSNR, peak SNR; RMSE, root mean squared error; SSIM, structure similarity index.
In the evaluation of the population‐averaged ROI values of and , the vessel mask does make statistically significant differences in the susceptibility values in 16 of 27 ROIs (Table 4). For example, of caudate, which reports the highest vessel proportion (3.76% 1.67%) because of the inclusion of the anterior terminal vein (Figure 6), shows a statistically significant difference in the susceptibility value (47.5 7.2 ppb without the mask vs. 44.4 6.8 ppb with the mask). In , statistically significant differences are observed in most ROIs, with the genu of the corpus callosum showing the greatest reduction in the population‐averaged mean susceptibility values (32.2 3.0 ppb without the mask vs. 30.7 2.9 ppb with the mask). The genu ROI is primarily influenced by the septal vein, requiring exclusion of the vessel (Figure 6). These examples suggest the importance of the vessel mask for accurate estimation of the ROI susceptibility values (see Figure S6 for additional ROIs).
TABLE 4.
Application for the population‐averaged ROI analysis.
| Population average of mean [ppb] | |||||||
|---|---|---|---|---|---|---|---|
| Vessel portions [%] | Without vessel mask | With vessel mask | p‐value | ||||
|
|
Subcortical nuclei | Caudate | 3.76 1.67 | 47.5 7.2 | 44.4 6.8 | <0.0001* | |
| Putamen | 0.17 0.26 | 82.2 21.5 | 82.1 21.5 | <0.00011* | |||
| Globus pallidus | 0.02 0.14 | 126.2 16.1 | 126.1 16.2 | 0.3276 | |||
| Nucleus accumbens | 0.01 0.02 | 57.9 15.0 | 57.9 15.0 | 0.1539 | |||
| Substantia nigra | 0.01 0.07 | 111.6 17.2 | 111.5 18.4 | 0.3261 | |||
| Red nucleus | 0.00 0.00 | 104.4 17.2 | 104.4 17.2 | – | |||
| Ventral pallidum | 0.01 0.07 | 134.9 26.5 | 134.7 27.0 | 0.2497 | |||
| Subthalamic nucleus | 0.00 0.00 | 103.4 15.2 | 103.4 15.2 | 0.3197 | |||
| Thalamic nuclei | Medial thalamic nuclei | 0.30 0.19 | 34.8 7.8 | 31.0 7.4 | <0.00011* | ||
| Lateral thalamic nuclei | 0.03 0.06 | 22.6 5.1 | 22.3 5.0 | <0.00011* | |||
| Pulvinar | 0.02 0.14 | 50.8 11.9 | 50.7 12.0 | 0.1329 | |||
|
|
White matter | Genu of corpus callosum | 1.89 0.65 | 32.2 3.0 | 30.7 2.9 | <0.00011* | |
| Body of corpus callosum | 0.66 0.39 | 35.1 2.6 | 34.7 2.6 | <0.00011* | |||
| Splenium of corpus callosum | 0.53 0.36 | 42.4 4.0 | 42.2 4.0 | <0.00011* | |||
| Cerebral peduncle | 0.16 0.25 | 44.5 3.9 | 44.2 4.0 | 0.00021* | |||
| Anterior limb of internal capsule | 0.13 0.14 | 40.5 4.1 | 40.3 4.0 | <0.00011* | |||
| Posterior limb of internal capsule | 0.03 0.06 | 52.1 3.4 | 52.1 3.4 | 0.0096 | |||
| Retrolentocular part of internal capsule | 0.09 0.14 | 39.7 3.5 | 39.6 3.4 | <0.00011* | |||
| Anterior corona radiata | 0.02 0.05 | 27.5 2.8 | 27.5 2.8 | 0.00161* | |||
| Superior corona radiata | 0.03 0.06 | 31.4 2.9 | 31.4 2.9 | <0.00011* | |||
| Posterior corona radiata | 0.02 0.07 | 31.9 2.6 | 31.8 2.6 | 0.0719 | |||
| Posterior thalamic radiation | 0.04 0.06 | 40.8 4.4 | 40.8 4.4 | <0.00011* | |||
| Sagittal stratum | 0.01 0.02 | 36.8 3.9 | 36.8 3.9 | 0.00051* | |||
| Superior longitudinal fasciculus | 0.01 0.04 | 32.2 2.8 | 32.2 2.8 | 0.0650 | |||
| Thalamic nuclei | Medial thalamic nuclei | 0.24 0.20 | 14.5 5.9 | 13.0 5.3 | <0.00011* | ||
| Lateral thalamic nuclei | 0.03 0.06 | 24.0 4.7 | 23.9 4.7 | <0.00011* | |||
| Pulvinar | 0.01 0.09 | 6.3 4.3 | 6.2 4.2 | 0.1217 | |||
Note: This table reports the proportion of vessels, and the mean susceptibility in each ROI analyzed with and without vessel masks.
Abbreviation: ROI, region of interest.
p < 0.05 with Bonferroni‐correction.
FIGURE 6.

Representative regions of interest (ROIs) including vessels. Caudate and corpus callosum show the highest vessel portion for and , respectively. Caudate primarily includes the anterior terminal veins whereas corpus callosum has septal veins (yellow arrows).
4. DISCUSSION
In our study, a new vessel segmentation method is developed to enhance the quantification of iron and myelin content in ‐separation. The overall processing pipeline consists of three steps: seed generation, vessel geometry guided‐region growing, and non‐vessel structure removal. In the first step, we developed a new seed generation that enhances the detection for small vessels by separating seeds for large and small vessels and using MIP for small vessels. In the region growing step, we extended the method from Kerkeni et al., 51 which combined vesselness and directional information, by incorporating intensity similarity and anisotropy criteria to reflect the geometric characteristics of vessels. Finally, non‐vessel structure removal was improved by exploiting the geometric properties of vessel structures, which is in contrast to the previous approach of thresholding for a cluster size. 46 , 47 Our results demonstrated superior performance of the proposed method compared to the previously proposed approaches by effectively excluding non‐vessel structures, achieving the highest DSC when calculated with respect to manually segmented vessel masks.
The proposed method generates two separate vessel masks for and . Vessel‐related artifacts in ‐separation appear in both susceptibility maps with erroneously values, but not all vessels are visible in both maps, depending on the vessel type and size (see Figures S2 and S5). 26 To ensure the inclusion of voxels affected by vessel‐related artifacts, seeds were generated from the product of the two maps. The region growing, however, was applied independently to each map to target vessels visible in each susceptibility map.
Our method has three hyperparameters (, , and anisotropy threshold) that can be adjusted for segmentation quality. Among them, and were fixed in this study, whereas the anisotropy threshold was optimized for each subject. The choice of the anisotropy threshold is a balance between excluding non‐vessel structures and including small vessels. When tested on the ‐separation maps from ‐sep‐COSMOS and ‐sepnet‐, a typical anisotropy threshold value ranged from 0.0012 to 0.0048 for and from 0.0012 to 0.0024 for . It was higher for ‐sep‐MEDI and ‐sep‐iLSQR maps (from 0.0072 to 0.0108 for and from 0.0018 to 0.0048 for ). This increase was likely because of relatively noisy maps from single‐orientation data when using conventional reconstructions (i.e., ‐sep‐MEDI and ‐sep‐iLSQR). It might have led to the exclusion of small vessels with low anisotropy values (yellow arrows in Figure 5) and consequently resulted in slightly reduced DSC values (Table 2).
The proposed method also allows adjusting the parameters as optional hyperparameters to further improve performance. As suggested in Eq. (1), controls the inclusion of voxels with high intensity, affecting sensitivity to vessels. A low value enhances the delineation of vessels, but may lead to the inclusion of non‐vessel structures with high intensity. In our study, the default settings () worked well in most cases. In some cases, however, falsely included deep gray matter regions, because of a low , was connected to nearby large vessels (see Figure S4). Such outcomes tended to occur in datasets from old populations, where increased susceptibility values in deep gray matter regions are observed because of iron accumulation. 66 , 67 By using a higher value, these masks can be improved. Additionally, for maps, meninges, which reveal high absolute susceptibility values, were prone to be included in the vessel masks when was low.
We observed that the previously proposed methods often included non‐vessel structures such as deep gray matter and highly myelinated fibers (see Figures 2, 3, 4). The Frangi filter has adjustable parameters, which was set to the values of previous QSM studies. 35 , 47 When it was adjusted for individual subjects, performance improved but the results still faced a trade‐off between sensitivity to vessels and the inclusion of non‐vessel structures (see Figures S7 and S8). The GRE‐based vessel segmentation method improved the performance by leveraging multiple contrasts (SWI and ) and incorporating additional techniques before and after applying vessel enhancement filter. For example, a regularization function was used to penalize regions with large field inhomogeneity, reducing false positives (see top row in Figure 3). This approach, however, still failed to completely exclude non‐vessels.
Our method has three challenges. First, subject‐wise parameter tuning can be challenging for a large number of subjects. Second, the method struggles to differentiate calcification from vessels because of their hyperintensities as well as their tube‐like structures in some cases (see Figure S5). Last, as previously mentioned, some small vessels were removed for the fixed parameters.
Deep learning has been widely applied to segmentation, 68 , 69 , 70 including brain vessel segmentation. 71 , 72 , 73 These approaches adopt data‐driven manners to generate a vessel mask. By learning structural information, deep learning models have potential to improve segmentation performance. However, generation of a training dataset requires high‐quality vessel masks. Vessel masks obtained using the proposed method, combined with individually tuned hyperparameters and manual refinements, could serve as a high‐quality label for training. This holds the potential to develop a method that is less reliant on hyperparameters. Nonetheless, generalization, such as resolution or contrasts, remains as critical consideration. 71 , 72 , 74
Another progress in deep learning is foundation models for universal segmentation, such as MedSAM. 75 These models aim to generalize across a wide range of segmentation tasks. However, foundation models experience performance degradation on less‐represented contrasts, necessitating fine‐tuning. 75 , 76 Additionally, they have reported difficulties in segmenting vessel‐like branching structures. 75 This limitation suggests that further research is required to adapt these models effectively for vessel segmentation tasks.
As for future work, our proposed method can be generalized for other susceptibility imaging contrasts, SWI and QSM, for instance, are potential. 30 , 31 , 77 By modifying the inputs of the method (i.e., and the product of and ) to the corresponding contrasts (i.e., SWI and QSM) and adjusting the hyperparameters, the proposed method may achieve vessel segmentation for the contrasts.
5. CONCLUSION
This study proposes a vessel segmentation method for ‐separation, leveraging region growing guided by vessel geometry. The vessel mask from this method outperforms the previously proposed methods, effectively excluding the non‐vessel structures such as deep gray matters in and myelinated fibers in . We demonstrate that the method generates robust results across different resolutions and ‐separation algorithms. Finally, the utility of the vessel masks in the analysis of ‐separation maps suggest improvements in the reliability and accuracy of the analysis.
Supporting information
Figure S1. Detailed pipeline for the proposed vessel segmentation algorithm.
Figure S2. Representative slice of , , and showing high signal intensities within vessels across all maps. Veins are typically paramagnetic whereas arteries possess susceptibility close to surrounding tissues. In ‐separation, however, both veins and arteries exhibit inaccurate values within and near vessel regions because of flow artifacts and non‐local effects, creating vessel artifacts. 5 Arteries appear on both susceptibility maps with high susceptibility values (red arrows). Some parts of veins also appear in map, despite vein's paramagnetic properties (blue arrows).
Figure S3. Ablation study of the region growing conditions: (A) Kerkeni et al.'s condition without intensity limits. (B) Proposed condition without intensity limits. (C) Kerkeni et al.'s condition with intensity limits. (D) Proposed condition with intensity limits. When intensity limits were not used, the algorithm showed reduced sensitivity to large vessels (blue arrows). Additionally, the use of the proposed condition demonstrated its effectiveness in excluding non‐vessel structures, such as basal ganglia (yellow arrows).
Figure S4. Effects of hyperparameters on vessel segmentation displayed on maps. (A) A lower value results in clearer vessel delineation (yellow arrow). (B) A higher anisotropy threshold better excludes non‐vessel structures, but may also exclude vessels (orange arrows). (C) The worst‐case scenario, where either non‐vessel structures are not excluded or both non‐vessels and vessels are excluded together, can occur (cyan box).
Figure S5. Manual segmentation results for and displayed in (A, D) axial, (B, E) sagittal, and (C, F) coronal views. Manual segmentation was conducted on 12 consecutive slices in the axial, sagittal, and coronal planes (a total of 36 slices per subject for each susceptibility map) using ITK‐snap. Calcifications (yellow arrows), meninges (orange arrows), and artifacts caused by mis‐registration between and (blue arrows), which exhibit high intensity in both and , were excluded from these masks.
Figure S6. Representative ROIs including vessels from eight subjects. Four ROIs (caudate and medial thalamic nuclei for ; genu of corpus callosum and superior corona radiata for ) are overlaid on the susceptibility maps. When vessels are excluded using the vessel mask, ROIs with a higher portion (first column) show relatively large differences in mean susceptibility values. Even in ROIs with a lower vessel portion (second column), although the differences are smaller, changes in susceptibility values can still be observed.
Figure S7. ROC curves for optimizing the Frangi filter parameters for (A) 3 T , (B) 3 T , (C) 7 T and (D) 7 T . The optimum parameters were selected that achieved the maximum specificity while exceeding the sensitivity of the proposed method.
Figure S8. Comparison of vessel masks generated using the Frangi filter with the default and optimized parameters. Vessel masks generated with the optimized parameters show improved results, excluding more deep gray matter regions in (yellow arrows) and capturing more small vessels in (orange arrows). The DSC values were calculated with respect to manually segmented vessel masks. The DSC values at the bottom of the zoomed‐in images confirm the enhanced results for the optimized parameters. Despite the improvement, the vessel masks still include non‐vessels structures, reporting lower DSC values than the proposed method.
Table S1. Summary of the MRI acquisition parameters for the three datasets.
ACKNOWLEDGMENTS
This work was supported by National Research Foundation of Korea (RS‐2024‐00349509), Korea Health Industry Development Institute (RS‐2024‐00439677), Korea Basic Science Institute (RS‐2024‐00435727), Institute of Information and communications Technology Planning and Evaluation (IITP) (IITP‐2025‐RS‐2023‐00256081), Samsung Electronics Co., Ltd. (IO201216‐08215‐01), and Institute of New Media and Communications, Institute of Engineering Research, and AI‐BIO Research Grant of Seoul National University.
APPENDIX A.
A.1. Inverse Hamming filter
An inverse Hamming filter, which is a high pass filter for suppressing non‐vessel structures and enhancing vascular structures, 47 , 48 is formulated as follows:
| (A1) |
where , , and are k‐space index, and , , and are filter sizes in each axis. In this study, the filter sizes suggested by Straub et al. 47 () were used.
A.2. Multi‐scale fractional anisotropy tensor vesselness ()
Multi‐scale fractional anisotropy tensor vesselness () reports the likelihood of a voxel being vascular. This is an extension of fractional anisotropy tensor vesselness (), effectively identifying vessels of varying sizes. 44
is derived from the eigenvalues (; ) of a Hessian matrix of an input image 44 :
| (A2) |
where is the mean of the eigenvalues (), and and are
| (A3) |
where r is the voxel position, and and are scaling constants between 0 and 1, modifying the eigenvalues to ensure the robustness of vesselness to low magnitudes of and .
To consider various sizes of vessels, is calculated for input images filtered by Gaussian kernels with different standard deviations . Next, is obtained as follows 44 :
| (A4) |
| (A5) |
| (A6) |
where is the current SD, is a previous SD, is the step size, and is defined as:
| (A7) |
In this study, the default parameters in Alhasson et al. 44 were used ( with an increment () of , , , and . The implementation of MFAT is available online (https://github.com/Haifafh/MFAT).
Kim T., Ji S., Min K., et al., “Vessel segmentation for ‐separation in quantitative susceptibility mapping,” Magnetic Resonance in Medicine 95, no. 1 (2026): 382–399, 10.1002/mrm.70054.
DATA AVAILABILITY STATEMENT
The proposed method is implemented in MATLAB and is available as a part of the ‐separation toolbox (https://github.com/SNU‐LIST/chi‐separation). The data used in this study are available on request from the corresponding author and can be shared following Institutional Review Board approval because of privacy or ethical restrictions.
REFERENCES
- 1. MacKay AL, Laule C. Magnetic resonance of myelin water: an in vivo marker for myelin. Brain Plast. 2016;2:71‐91. doi: 10.3233/bpl-160033 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Ward RJ, Zucca FA, Duyn JH, Crichton RR, Zecca L. The role of iron in brain ageing and neurodegenerative disorders. Lancet Neurol. 2014;13:1045‐1060. doi: 10.1016/s1474-4422(14)70117-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Bouhrara M, Reiter DA, Bergeron CM, et al. Evidence of demyelination in mild cognitive impairment and dementia using a direct and specific magnetic resonance imaging measure of myelin content. Alzheimers Dement. 2018;14:998‐1004. doi: 10.1016/j.jalz.2018.03.007 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. van Duijn S, Bulk M, van Duinen SG, Nabuurs RJA, et al. Cortical iron reflects severity of Alzheimer's disease. J Alzheimer's Dis. 2017;60:1‐13. doi: 10.3233/jad-161143 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Möller HE, Bossoni L, Connor JR, et al. Iron, myelin, and the brain: neuroimaging meets neurobiology. Trends Neurosci. 2019;42:384‐401. doi: 10.1016/j.tins.2019.03.009 [DOI] [PubMed] [Google Scholar]
- 6. Stephenson E, Nathoo N, Mahjoub Y, Dunn JF, Yong VW. Iron in multiple sclerosis: roles in neurodegeneration and repair. Nat Rev Neurol. 2014;10:459‐468. doi: 10.1038/nrneurol.2014.118 [DOI] [PubMed] [Google Scholar]
- 7. Zecca L, Youdim MBH, Riederer P, Connor JR, Crichton RR. Iron, brain ageing and neurodegenerative disorders. Nat Rev Neurosci. 2004;5:863‐873. doi: 10.1038/nrn1537 [DOI] [PubMed] [Google Scholar]
- 8. Lee J, Nam Y, Choi JY, Shin HG, Hwang T, Lee J. Separating positive and negative susceptibility sources in QSM. In: Proceedings of the 25th Annual Meeting of ISMRM, Honolulu, HI, USA 2017.
- 9. Shin HG, Lee J, Yun YH, et al. Χ‐Separation: magnetic susceptibility source separation toward iron and myelin mapping in the brain. Neuroimage. 2021;240:118371. doi: 10.1016/j.neuroimage.2021.118371 [DOI] [PubMed] [Google Scholar]
- 10. Ji S, Jang J, Kim M, et al. Comparison between R2′‐based and R2*‐based χ‐separation methods: a clinical evaluation in individuals with multiple sclerosis. NMR Biomed. 2024;37:e5167. doi: 10.1002/nbm.5167 [DOI] [PubMed] [Google Scholar]
- 11. Jang M, Dimov AV, Kapse K, et al. Quantitative susceptibility mapping with source separation in normal brain development of newborns. Am J Neuroradiol. 2024;46:ajnr.A8488. doi: 10.3174/ajnr.a8488 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Müller J, Lu PJ, Cagol A, et al. Quantifying remyelination using χ‐separation in white matter and cortical multiple sclerosis lesions. Neurology. 2024;103:e209604. doi: 10.1212/wnl.0000000000209604 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Nakashima M, Kan H, Kawai T, et al. Quantitative susceptibility mapping analyses of white matter in Parkinson's disease using susceptibility separation technique. Parkinsonism Relat Disord. 2024;128:107135. doi: 10.1016/j.parkreldis.2024.107135 [DOI] [PubMed] [Google Scholar]
- 14. Zhu Z, Naji N, Esfahani JH, et al. MR susceptibility separation for quantifying lesion paramagnetic and diamagnetic evolution in relapsing–remitting multiple sclerosis. J Magn Reson Imaging. 2024; 60:1867‐1879. doi: 10.1002/jmri.29266 [DOI] [PubMed] [Google Scholar]
- 15. Straub S, El‐Sanosy E, Emmerich J, Sandig FL, Ladd ME, Schlemmer H. Quantitative magnetic resonance imaging biomarkers for cortical pathology in multiple sclerosis at 7 T. NMR Biomed. 2023;36:e4847. doi: 10.1002/nbm.4847 [DOI] [PubMed] [Google Scholar]
- 16. Kim W, Shin HG, Lee H, et al. Χ‐Separation imaging for diagnosis of multiple sclerosis versus neuromyelitis Optica Spectrum disorder. Radiology. 2023;307:e220941. doi: 10.1148/radiol.220941 [DOI] [PubMed] [Google Scholar]
- 17. Chen J, Gong NJ, Chaim KT, Otaduy MCG, Liu C. Decompose quantitative susceptibility mapping (QSM) to sub‐voxel diamagnetic and paramagnetic components based on gradient‐echo MRI data. Neuroimage. 2021;242:118477. doi: 10.1016/j.neuroimage.2021.118477 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Emmerich J, Bachert P, Ladd ME, Straub S. On the separation of susceptibility sources in quantitative susceptibility mapping: theory and phantom validation with an in vivo application to multiple sclerosis lesions of different age. J Magn Reson. 2021;330:107033. doi: 10.1016/j.jmr.2021.107033 [DOI] [PubMed] [Google Scholar]
- 19. Li Z, Feng R, Liu Q, et al. APART‐QSM: an improved sub‐voxel quantitative susceptibility mapping for susceptibility source separation using an iterative data fitting method. Neuroimage. 2023;274:120148. doi: 10.1016/j.neuroimage.2023.120148 [DOI] [PubMed] [Google Scholar]
- 20. Kan H, Uchida Y, Kawaguchi S, Kasai H, Hiwatashi A, Ueki Y. Quantitative susceptibility mapping for susceptibility source separation with adaptive relaxometric constant estimation (QSM‐ARCS) from solely gradient‐echo data. Neuroimage. 2024;296:120676. doi: 10.1016/j.neuroimage.2024.120676 [DOI] [PubMed] [Google Scholar]
- 21. Lee J, Ji S, Oh SH. So You want to image myelin using MRI: magnetic susceptibility source separation for myelin imaging. Magn Reson Med Sci. 2024;23:291‐306. doi: 10.2463/mrms.rev.2024-0001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Hernando D, Vigen KK, Shimakawa A, Reeder SB. R mapping in the presence of macroscopic B0 field variations. Magn Reson Med. 2012;68:830‐840. doi: 10.1002/mrm.23306 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Yablonskiy DA, Sukstanskii AL, Luo J, Wang X. Voxel spread function method for correction of magnetic field inhomogeneity effects in quantitative gradient‐echo‐based MRI. Magn Reson Med. 2013;70:1283‐1292. doi: 10.1002/mrm.24585 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Larson TC, Kelly WM, Ehman RL, Wehril FW. Spatial misregistration of vascular flow during MR imaging of the CNS. ASNR. 1990; 11:1041‐1048. [PMC free article] [PubMed] [Google Scholar]
- 25. Nishimura DG, Jackson JI, Pauly JM. On the nature and reduction of the displacement artifact in flow images. Magn Reson Med. 1991;22:481‐492. doi: 10.1002/mrm.1910220255 [DOI] [PubMed] [Google Scholar]
- 26. Kim T, Ji S, Lee J. Exploring vessel artifacts in χ‐separation. In: Proceedings of the 33rd Annual Meeting ISMRM, Honolulu, HI, USA, 2025. [Google Scholar]
- 27. Lee S, Shin HG, Kim M, Lee J. Depth‐wise profiles of iron and myelin in the cortex and white matter using χ‐separation: a preliminary study. Neuroimage. 2023;273:120058. doi: 10.1016/j.neuroimage.2023.120058 [DOI] [PubMed] [Google Scholar]
- 28. Min K, Sohn B, Kim WJ, et al. A human brain atlas of χ‐separation for normative iron and myelin distributions. NMR Biomed. 2024;37:e5226. doi: 10.1002/nbm.5226 [DOI] [PubMed] [Google Scholar]
- 29. Berg RC, Preibisch C, Thomas DL, Shmueli K, Biondetti E. Investigating the effect of flow compensation and quantitative susceptibility mapping method on the accuracy of venous susceptibility measurement. Neuroimage. 2021;240:118399. doi: 10.1016/j.neuroimage.2021.118399 [DOI] [PubMed] [Google Scholar]
- 30. Biondetti E, Rojas‐Villabona A, Sokolska M, et al. Investigating the oxygenation of brain arteriovenous malformations using quantitative susceptibility mapping. Neuroimage. 2019;199:440‐453. doi: 10.1016/j.neuroimage.2019.05.014 [DOI] [PubMed] [Google Scholar]
- 31. Breiding PS, Kellner‐Weldon F, Grunder L, et al. Quantification of cerebral veins in patients with acute migraine with aura: a fully automated quantification algorithm using susceptibility‐weighted imaging. PLoS One. 2020;15:e0233992. doi: 10.1371/journal.pone.0233992 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Ge Y, Zohrabian VM, Osa E, et al. Diminished visibility of cerebral venous vasculature in multiple sclerosis by susceptibility‐weighted imaging at 3.0 tesla. J Magn Reson Imaging. 2009;29:1190‐1194. doi: 10.1002/jmri.21758 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Koopmans PJ, Manniesing R, Niessen WJ, Viergever MA, Barth M. MR venography of the human brain using susceptibility weighted imaging at very high field strength. Magn Reson Mater Phys Biol Med. 2008;21:149‐158. doi: 10.1007/s10334-007-0101-3 [DOI] [PubMed] [Google Scholar]
- 34. Winchell AM, Taylor BA, Song R, et al. Evaluation of SWI in children with sickle cell disease. Am J Neuroradiol. 2014;35:1016‐1021. doi: 10.3174/ajnr.a3794 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Ma Y, Mazerolle EL, Cho J, Sun H, Wang Y, Pike GB. Quantification of brain oxygen extraction fraction using QSM and a hyperoxic challenge. Magn Reson Med. 2020;84:3271‐3285. doi: 10.1002/mrm.28390 [DOI] [PubMed] [Google Scholar]
- 36. Jorge J, Gretsch F, Najdenovska E, et al. Improved susceptibility‐weighted imaging for high contrast and resolution thalamic nuclei mapping at 7T. Magn Reson Med. 2020;84:1218‐1234. doi: 10.1002/mrm.28197 [DOI] [PubMed] [Google Scholar]
- 37. Ward PGD, Ferris NJ, Raniga P, et al. Combining images and anatomical knowledge to improve automated vein segmentation in MRI. Neuroimage. 2018;165:294‐305. doi: 10.1016/j.neuroimage.2017.10.049 [DOI] [PubMed] [Google Scholar]
- 38. Bernier M, Cunnane SC, Whittingstall K. The morphology of the human cerebrovascular system. Hum Brain Mapp. 2018;39:4962‐4975. doi: 10.1002/hbm.24337 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Zhang X, Zhang Y, Hu Q. Deep learning based vein segmentation from susceptibility‐weighted images. Computing. 2019;101:637‐652. doi: 10.1007/s00607-018-0677-7 [DOI] [Google Scholar]
- 40. Huck J, Wanner Y, Fan AP, et al. High resolution atlas of the venous brain vasculature from 7 T quantitative susceptibility maps. Brain Struct Funct. 2019;224:2467‐2485. doi: 10.1007/s00429-019-01919-4 [DOI] [PubMed] [Google Scholar]
- 41. Frangi AF, Niessen WJ, Vincken KL, Viergever MA. Multiscale vessel enhancement filtering. In: International Conference on Medical Image Computing and Computer‐Assisted Intervention. Springer Berlin Heidelberg, 1998:130‐137. doi: 10.1007/bfb0056195 [DOI] [Google Scholar]
- 42. Manniesing R, Viergever MA, Niessen WJ. Vessel enhancing diffusion a scale space representation of vessel structures. Med Image Anal. 2006;10:815‐825. doi: 10.1016/j.media.2006.06.003 [DOI] [PubMed] [Google Scholar]
- 43. Jerman T, Pernuš F, Likar B, Špiclin Ž. Enhancement of vascular structures in 3D and 2D angiographic images. IEEE Trans Med Imaging. 2016;35:2107‐2118. doi: 10.1109/tmi.2016.2550102 [DOI] [PubMed] [Google Scholar]
- 44. Alhasson HF, Alharbi SS, Obara B. 2D and 3D vascular structures enhancement via multiscale fractional anisotropy tensor. arXiv 2019. doi: 10.48550/arxiv.1902.00550 [DOI]
- 45. Cui H, Xia Y, Zhang Y. 2D and 3D vascular structures enhancement via improved Vesselness filter and vessel enhancing diffusion. IEEE Access. 2019;7:123969‐123980. doi: 10.1109/access.2019.2938392 [DOI] [Google Scholar]
- 46. Monti S, Cocozza S, Borrelli P, et al. MAVEN: an algorithm for multi‐parametric automated segmentation of brain veins from gradient Echo acquisitions. IEEE Trans Med Imaging. 2017;36:1054‐1065. doi: 10.1109/tmi.2016.2645286 [DOI] [PubMed] [Google Scholar]
- 47. Straub S, Stiegeler J, El‐Sanosy E, Bendszus M, Ladd ME, Schneider TM. A novel gradient echo data based vein segmentation algorithm and its application for the detection of regional cerebral differences in venous susceptibility. Neuroimage. 2022;250:118931. doi: 10.1016/j.neuroimage.2022.118931 [DOI] [PubMed] [Google Scholar]
- 48. Jin Z, Xia L, Zhang M, Du YP. Background‐suppressed MR venography of the brain using magnitude data: a high‐pass filtering approach. Comput Math Methods Med. 2014;2014:812785. doi: 10.1155/2014/812785 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49. Bornemann F, März T. Fast image inpainting based on coherence transport. J Math Imaging Vis. 2007;28:259‐278. doi: 10.1007/s10851-007-0017-6 [DOI] [Google Scholar]
- 50. Murayama K, Suzuki S, Matsukiyo R, et al. Preliminary study of time maximum intensity projection computed tomography imaging for the detection of early ischemic change in patient with acute ischemic stroke. Medicine. 2018;97:e9906. doi: 10.1097/md.0000000000009906 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51. Kerkeni A, Benabdallah A, Manzanera A, Bedoui MH. A coronary artery segmentation method based on multiscale analysis and region growing. Comput Med Imaging Graph. 2016;48:49‐61. doi: 10.1016/j.compmedimag.2015.12.004 [DOI] [PubMed] [Google Scholar]
- 52. Kim M, Ji S, Kim J, et al. Χ‐Sepnet: deep neural network for magnetic susceptibility source separation. Hum Brain Mapp. 2025;46:e70136. doi: 10.1002/hbm.70136 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53. Kim J, Kim M, Ji S, et al. In‐vivo high‐resolution χ‐separation at 7T. Neuroimage. 2025;308:121060. doi: 10.1016/j.neuroimage.2025.121060 [DOI] [PubMed] [Google Scholar]
- 54. Smith SM. Fast robust automated brain extraction. Hum Brain Mapp. 2002;17:143‐155. doi: 10.1002/hbm.10062 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55. Committee QCO , Bilgic B, Costagli M, et al. Recommended implementation of quantitative susceptibility mapping for clinical research in the brain: a consensus of the ISMRM electro‐magnetic tissue properties study group. Magn Reson Med. 2024;91:1834‐1862. doi: 10.1002/mrm.30006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56. Dymerska B, Eckstein K, Bachrata B, et al. Phase unwrapping with a rapid opensource minimum spanning tree algorithm (ROMEO). Magn Reson Med. 2021;85:2294‐2308. doi: 10.1002/mrm.28563 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57. Wu B, Li W, Avram AV, Gho SM, Liu C. Fast and tissue‐optimized mapping of magnetic susceptibility and T2* with multi‐echo and multi‐shot spirals. Neuroimage. 2012;59:297‐305. doi: 10.1016/j.neuroimage.2011.07.019 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58. Schweser F, Deistung A, Lehr BW, Reichenbach JR. Quantitative imaging of intrinsic magnetic tissue properties using MRI signal phase: an approach to in vivo brain iron metabolism? Neuroimage. 2011;54:2789‐2807. doi: 10.1016/j.neuroimage.2010.10.070 [DOI] [PubMed] [Google Scholar]
- 59. Wu B, Li W, Guidon A, Liu C. Whole brain susceptibility mapping using compressed sensing. Magn Reson Med. 2012;67:137‐147. doi: 10.1002/mrm.23000 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60. Lebel RM, Wilman AH. Transverse relaxometry with stimulated echo compensation. Magn Reson Med. 2010;64:1005‐1014. doi: 10.1002/mrm.22487 [DOI] [PubMed] [Google Scholar]
- 61. Lebel RM. StimFit: A Toolbox for Robust T2 Mapping with Stimulated Echo Compensation. In: Proceedings of the 20th Annual Meeting of ISMRM, Melbourne, Australia. 2012.
- 62. Smith SM, Jenkinson M, Woolrich MW, et al. Advances in functional and structural MR image analysis and implementation as FSL. Neuroimage. 2004;23:S208‐S219. doi: 10.1016/j.neuroimage.2004.07.051 [DOI] [PubMed] [Google Scholar]
- 63. Shin HG, Seo J, Lee Y, et al. chi‐separation using multi‐orientation data in invivo and exvivo brains: Visualization of histology up to the resolution of 350 um. In: Proceedings of the 30th Annual Meeting of ISMRM, London, England, UK. 2022.
- 64. Yushkevich PA, Piven J, Hazlett HC, et al. User‐guided 3D active contour segmentation of anatomical structures: significantly improved efficiency and reliability. Neuroimage. 2006;31:1116‐1128. doi: 10.1016/j.neuroimage.2006.01.015 [DOI] [PubMed] [Google Scholar]
- 65. Dice LR. Measures of the amount of ecologic association between species. Ecology. 1945;26:297‐302. [Google Scholar]
- 66. Bilgic B, Pfefferbaum A, Rohlfing T, Sullivan EV, Adalsteinsson E. MRI estimates of brain iron concentration in normal aging using quantitative susceptibility mapping. Neuroimage. 2012;59:2625‐2635. doi: 10.1016/j.neuroimage.2011.08.077 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67. Ficiarà E, Stura I, Guiot C. Iron deposition in brain: Does aging matter? Int J Mol Sci. 2022;23:10018. doi: 10.3390/ijms231710018 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68. Akkus Z, Galimzianova A, Hoogi A, Rubin DL, Erickson BJ. Deep learning for brain MRI segmentation: state of the art and future directions. J Digit Imaging. 2017;30:449‐459. doi: 10.1007/s10278-017-9983-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69. Minaee S, Boykov Y, Porikli F, Plaza A, Kehtarnavaz N, Terzopoulos D. Image segmentation using deep learning: a survey. IEEE Trans Pattern Anal Mach Intell. 2022;44:3523‐3542. doi: 10.1109/tpami.2021.3059968 [DOI] [PubMed] [Google Scholar]
- 70. Wang R, Lei T, Cui R, Zhang B, Meng H, Nandi AK. Medical image segmentation using deep learning: a survey. IET Image Process. 2022;16:1243‐1267. doi: 10.1049/ipr2.12419 [DOI] [Google Scholar]
- 71. Tetteh G, Efremov V, Forkert ND, et al. DeepVesselNet: vessel segmentation, centerline prediction, and bifurcation detection in 3‐D angiographic volumes. Front Neurosci. 2020;14:592352. doi: 10.3389/fnins.2020.592352 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72. Hilbert A, Madai VI, Akay EM, et al. BRAVE‐NET: fully automated arterial brain vessel segmentation in patients with cerebrovascular disease. Front Artif Intell. 2020;3:552258. doi: 10.3389/frai.2020.552258 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73. Goni MR, Ruhaiyem NIR, Mustapha M, Achuthan A, Nassir CMNCM. Brain vessel segmentation using deep learning—a review. IEEE Access. 2022;10:111322‐111336. doi: 10.1109/access.2022.3214987 [DOI] [Google Scholar]
- 74. Jung W, Bollmann S, Lee J. Overview of quantitative susceptibility mapping using deep learning: current status, challenges and opportunities. NMR Biomed. 2022;35:e4292. doi: 10.1002/nbm.4292 [DOI] [PubMed] [Google Scholar]
- 75. Ma J, He Y, Li F, Han L, You C, Wang B. Segment anything in medical images. Nat Commun. 2024;15:654. doi: 10.1038/s41467-024-44824-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76. Wei X, Cao J, Jin Y, Lu M, Wang G, Zhang S. I‐MedSAM: implicit medical image segmentation with segment anything. arXiv 2023. doi: 10.48550/arxiv.2311.17081 [DOI]
- 77. Bazin PL, Plessis V, Fan AP, Villringer A, Gauthier CJ. Vessel segmentation from quantitative susceptibility maps for local oxygenation venography. In: 2016 IEEE 13th Int Symp Biomed Imaging (ISBI), IEEE;2016:1135‐1138. doi: 10.1109/isbi.2016.7493466 [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Figure S1. Detailed pipeline for the proposed vessel segmentation algorithm.
Figure S2. Representative slice of , , and showing high signal intensities within vessels across all maps. Veins are typically paramagnetic whereas arteries possess susceptibility close to surrounding tissues. In ‐separation, however, both veins and arteries exhibit inaccurate values within and near vessel regions because of flow artifacts and non‐local effects, creating vessel artifacts. 5 Arteries appear on both susceptibility maps with high susceptibility values (red arrows). Some parts of veins also appear in map, despite vein's paramagnetic properties (blue arrows).
Figure S3. Ablation study of the region growing conditions: (A) Kerkeni et al.'s condition without intensity limits. (B) Proposed condition without intensity limits. (C) Kerkeni et al.'s condition with intensity limits. (D) Proposed condition with intensity limits. When intensity limits were not used, the algorithm showed reduced sensitivity to large vessels (blue arrows). Additionally, the use of the proposed condition demonstrated its effectiveness in excluding non‐vessel structures, such as basal ganglia (yellow arrows).
Figure S4. Effects of hyperparameters on vessel segmentation displayed on maps. (A) A lower value results in clearer vessel delineation (yellow arrow). (B) A higher anisotropy threshold better excludes non‐vessel structures, but may also exclude vessels (orange arrows). (C) The worst‐case scenario, where either non‐vessel structures are not excluded or both non‐vessels and vessels are excluded together, can occur (cyan box).
Figure S5. Manual segmentation results for and displayed in (A, D) axial, (B, E) sagittal, and (C, F) coronal views. Manual segmentation was conducted on 12 consecutive slices in the axial, sagittal, and coronal planes (a total of 36 slices per subject for each susceptibility map) using ITK‐snap. Calcifications (yellow arrows), meninges (orange arrows), and artifacts caused by mis‐registration between and (blue arrows), which exhibit high intensity in both and , were excluded from these masks.
Figure S6. Representative ROIs including vessels from eight subjects. Four ROIs (caudate and medial thalamic nuclei for ; genu of corpus callosum and superior corona radiata for ) are overlaid on the susceptibility maps. When vessels are excluded using the vessel mask, ROIs with a higher portion (first column) show relatively large differences in mean susceptibility values. Even in ROIs with a lower vessel portion (second column), although the differences are smaller, changes in susceptibility values can still be observed.
Figure S7. ROC curves for optimizing the Frangi filter parameters for (A) 3 T , (B) 3 T , (C) 7 T and (D) 7 T . The optimum parameters were selected that achieved the maximum specificity while exceeding the sensitivity of the proposed method.
Figure S8. Comparison of vessel masks generated using the Frangi filter with the default and optimized parameters. Vessel masks generated with the optimized parameters show improved results, excluding more deep gray matter regions in (yellow arrows) and capturing more small vessels in (orange arrows). The DSC values were calculated with respect to manually segmented vessel masks. The DSC values at the bottom of the zoomed‐in images confirm the enhanced results for the optimized parameters. Despite the improvement, the vessel masks still include non‐vessels structures, reporting lower DSC values than the proposed method.
Table S1. Summary of the MRI acquisition parameters for the three datasets.
Data Availability Statement
The proposed method is implemented in MATLAB and is available as a part of the ‐separation toolbox (https://github.com/SNU‐LIST/chi‐separation). The data used in this study are available on request from the corresponding author and can be shared following Institutional Review Board approval because of privacy or ethical restrictions.