Radial Hadamard-encoded 19F-MRI

Kian Tadjalli Mehr 1,, Johannes Fischer 1, Felix Spreter 1, Simon Reiss 1, David Boll 2, Ali Caglar Özen 1, Deepa Gunashekar 1, Constantin von zur Mühlen 2, Alexander Maier 2, Michael Bock 1
PMCID: PMC12638425  PMID: 40445260

Abstract

Objectives

Developing a 19F imaging method to acquire images of the molecular inflammation tracer perfluorooctyl bromide (PFOB) without chemical shift artifacts.

Materials and methods

PFOB is a molecular tracer that can be used to track the response of myeloid cells. However, imaging of PFOB with 19F-MRI is challenging due to its complex spectrum which leads to unwanted chemical shift artifacts. Spectral HE allows for separate reconstructions of each peak of the PFOB spectrum, which was combined into a single image after resonance shift correction. In this work, a Hadamard-encoded (HE) radial 3D UTE sequence was tested in phantoms and in vivo in a pig, measuring the 19F signal in the spleen at different times after injection.

Results

Chemical shift artifacts were effectively suppressed with HE, and an SNR > 100 was observed for the 19F signal in the spleen 2 days after injection. The signal decreased over time, and 7 days after injection it was reduced by 30%.

Discussion

Chemical shift artifact correction using HE allowed for in vivo 19F PFOB imaging of labeled monocytes with a high SNR. Compared to spectrally selective excitation, HE increased the PFOB 19F-MRI signal by 10%, and the simple HE-algorithm could be directly integrated into the image reconstruction of the MRI system.

Keywords: Magnetic resonance imaging, Perfluorocarbons, Fluorine-19 magnetic resonance imaging

Introduction

Novel therapeutic approaches to treat inflammatory diseases make molecular inflammation imaging an attractive therapeutic tool [1]. Myeloid cells are involved in inflammatory healing processes. They accumulate in the spleen, where they remain silent until activation in response to inflammatory events such as myocardial infarction (MI), stroke, or others. These cells can be labeled using biologically inert perfluorocarbons (PFCs) that are injected into the blood stream [2]. The distribution of the PFC-labeled cells can be imaged with 19F-MRI [3], which makes PFCs an excellent imaging marker for inflammation due to the absence of other 19F-MRI background signals [4, 5]. Fluorine-19 has a natural abundance of nearly 100% and a gyromagnetic ratio of 40.05 MHz/T leading to a high MR sensitivity of 83% of that of protons [6]; however, 19F-MRI of PFCs is limited by the low millimolar concentrations of 19F in an in vivo application.

Due to the background-free 19F-MRI signal, the concentration of inflammatory cells and, in turn, the inflammatory reaction can be directly quantified from the 19F-MRI signal of the PFCs [7]. Several different PFCs have been proposed for this application [810]. In this work, we use perfluorooctyl bromide (PFOB), which has been used in clinical trials before and has a relatively low biological half-life of about 12 days [8]. Unlike some other PFCs, PFOB is known to exhibit significant chemical shift artifacts in 19F-MRI due to its complex 19F NMR-spectrum. One approach to mitigate chemical shift artifacts is the selective excitation of single resonances of the spectrum at the cost of an overall reduced signal intensity. Previous works have performed cellular 19F imaging of PFOB in this way [11, 12]. Alternatively, all resonances can be excited and the chemical shift artifacts can be removed during data post-processing using a deconvolution approach [1315], using phase information at different echo times [16, 17] or 19F-MRI data can be acquired with two different directions of the readout direction leading to a characteristic change of the chemical shift artifact patterns which can be exploited to suppress them [18].

In this work, we use a different concept for chemical shift correction which is based on spectral Hadamard-encoding (HE) [19, 20]. HE is combined with radial data acquisition based on an ultra-short TE (UTE) sequence, and the method is tested in vivo in pigs with 19F-MRI of the spleen.

Materials and methods

Perfluorooctyl bromide (PFOB)

Perflubron or PFOB is a perfluorinated derivative of the hydrocarbon octane, where all 18 hydrogen atoms have been replaced with 17 fluorine (19F) and 1 bromine (Br) atoms leading to 1 terminal CF3 group, 1 terminal CF2Br group, and 6 CF2 groups (Fig. 1). In the 19F MR spectrum, these groups have different chemical shifts of δCF2Br=+18.1ppm and δCF2=-40.1ppm relative to the central CF3 resonance (here set to δCF3=0ppm) [8, 21]. The terminal Br atom breaks the structural symmetry of the molecule and gives rise to small variations in chemical shift of the CF2 resonances ranging from -35.7ppm for the closest CF2 group, β, to the most distant group η at -44.8ppm. In this work, these resonances cannot be resolved so that we consider only three distinct resonances.

Fig. 1.

Fig. 1

Chemical structure and chemical shift spectrum of PFOB at 3 T

Hadamard-encoding scheme

In MRI with conventional Cartesian encoding, chemical shift artifacts manifest as ghost images of the different resonances that appear spatially shifted along the readout direction by

Δx=δ·γB0BW, 1

where BW is the receiver bandwidth and δ is the chemical shift of the resonance. The most common occurrence of chemical shift correction is the fat–water shift in 1H MRI. Many methods were developed to remove the chemical shift artifacts by suppressing the fat signal [22]. However, in 19F-MRI, the tracer concentration is low, and signal from all resonances is desired to improve SNR. Thus, rather than using suppression techniques to avoid chemical shift artifacts, we propose a spectral encoding/decoding scheme that allows for the simultaneous acquisition of signal from all resonances.

HE has been previously applied in MRI, for example, to counteract aliasing artifacts [23], or in simultaneous multislice acquisitions [24]. The general concept is to encode a vector of information by multiplying it with a matrix of orthogonal column vectors, which can later be decoded by multiplication with the transposed matrix. Here, we use a Hadamard matrix H, to encode the signal from each separate resonance, by acquiring the same k-space data four times with four different RF excitation pulses B1.

B1,1tB1,2tB1,3tB1,4t=HA=1111-1-111-11-11·B1,μt·exp-iδω1tB1,νtB1,ξt·exp-iδω2t, 2

where B1,μ is the RF-envelope of the CF2Br-Peak, B1,ν of the CF3-Peak, and B1,ξ of the CF2-Peaks and δω1=γB0δCF2Br and δω2=γB0δCF2 are the off-resonances of the respective resonances relative to the CF3 group. Note that the Hadamard matrix is reduced to 4 × 3, because we only need to encode three resonances and the fourth encoding combination is not needed. The frequency of the MRI system is set to the resonance frequency of the CF3 resonance. Negative signs in the encoding matrix are effectively 180° phase shifts in the excitation, which means that received signals of the same resonance excited with a negative and a positive sign cancel out after addition of the raw data. While it would be possible to separate the individual CF2 resonances as well using a larger Hadamard matrix, this would require longer pulses causing signal loss due to the longer echo time, while only gaining a minimal amount of signal.

During post-processing, the signal from the three resonances can, therefore, be Hadamard-decoded using the following combinations of signals:

SμSνSξ=HTS=11111-11-11-1-11S1S2S3S4, 3

where S1,,S4 are the signals received after each HE pulse, B1,i. In Eq. (3), the signals from the resonances Sμ,Sν, or Sξ add up constructively, while signals from the other resonances cancel, such that after the four pulses each resonance has been acquired without time loss compared to selective excitation with four averages. After reversing the chemical shift for each of the resonances individually, their signals can be added using a sum of squares combination (Fig. 2). In the ideal case, where all resonances contribute equally, the 3 fluorine atoms from the CF3 resonance, the 2 fluorine atoms from the CF2Br resonance and the 12 fluorine atoms of the CF2 resonances would increase the signal by a factor of 122+32+2212=1.04, over an acquisition with the CF2 resonances alone—thus, only a 4% gain would be measurable. In reality, however, the CF2-signals dephase quickly and blur due to the uncorrected chemical shifts, which reduces their contribution to the total signal, making it more worthwhile to add the contribution from the other resonances.

Fig. 2.

Fig. 2

Different steps during the reconstruction process shown with the same windowing. Left: Chemical shift artifacts in a radial sequence, Middle: Reconstructed signals for each peak after Hadamard-decoding. Due to chemical shifts within the CF2 multi resonance, this image appears blurred. Right: Combination of all signals via sum of squares

Chemical shift correction

In radial imaging, the readout direction is different for each k-space line (spoke). Thus, artifacts from chemical shifts do not result in a constant spatial offset as in Cartesian imaging, but rather in a blurring of the reconstructed image [15]. Therefore, a chemical shift correction must be performed prior to regridding of the radial data to a Cartesian grid. The evolution of the MR signal phase during data acquisition of an isochromate with chemical shift δ is given by:

φr,t=γ1+δB0+Gtrdt=1+δγB0t+γGtrdt+δγGtrdt. 4

The first term describes a linearly increasing phase over time, independent of the spatial location, compared to the resonant signal. The second term is the desired on-resonant phase evolution due to spatial encoding in a magnetic field gradient G. The last term is also location dependent, but can be neglected as B0G·r. The phase evolution of the chemical shift is, therefore, governed by δγB0t, i.e. a linear increase of the phase with time. Thus, after signals of the different resonances are separated using HE, the chemical shift can be corrected by multiplication of each acquired line in k-space with a linearly increasing phase factor exp+iδγB0κτ, where κ is the index of the pixel within that line and τ is the dwell time.

RF pulse shapes

To excite the individual resonances, both Gaussian and sinc shapes were tested as RF-pulses. The Gaussian pulses were defined as

B1Gausst=B1Gauss·exp-π2δωGauss2t22. 5

Compared to Gauss pulses, sinc pulses can excite a broader range of the spectrum homogeneously. Here, a Hamming-windowed pulse [25] was applied:

B1sinct=B1sinc·0.54+0.46·cosπ·δωsinc·t6·sinπ·δωsinc·tπ·δωsinc·t, 6

where δωGauss=541Hz and δωSinc=4000Hz describe the width of the pulse in the frequency domain. The respective bandwidths were chosen for a field strength of 3 T, which is the clinical field strength which will later be applied (see below). Phantom measurements with a 10% PFOB fat–water emulsion were performed with both pulse shapes for all resonances using a single-channel Tx/Rx 19F solenoid coil. The signal intensities and chemical shift artifacts of the multi-resonance were compared.

UTE sequence with Hadamard-encoding

K-space data were acquired using a center-out radial trajectory with a spiral phyllotaxis pattern [26]. Here, all acquired spokes in k-space intersect a spherical spiral, where the azimuthal angle is rotated by the golden angle for consecutive points (Fig. 3). The spokes are acquired in an interleaved fashion starting with every n-th point, then rotating the starting point by a golden angle. For n being a Fibonacci number, consecutively acquired spokes are close to each other, reducing eddy current effects and improving gradient spoiling [27]. This approach was taken for two reasons: First, from FID measurements, we estimated T2* values for all resonances to be below 10 ms, so that an ultra-short echo time (UTE) sequence was chosen to maximize the signal. Second, to acquire images in the abdomen, the sequence needs to be robust to motion. While this is the case for radial imaging, the phyllotaxis trajectory ensures homogeneous sampling of k-space over small time scales compared to the total imaging time.

Fig. 3.

Fig. 3

Phyllotaxis pattern of consecutively acquired spokes intersecting the unit sphere. The end-points of consecutive k-space spokes follow a spiral on the surface of a sphere with radius kMax. A spiral where the kz-component decreases with every spoke (red) is followed by a spiral with increasing kz-component (orange)

PFOB nanoemulsion

A nanoemulsion consisting of 35% PFOB was used for the in vivo experiments. The PFOB nanoemulsion (PFOB-NE) was produced similar as described previously [12]: a buffer of sodium dihydrogen phosphate (0.24 g), disodium hydrogen phosphate (2.46 g), sodium chloride (1.15 g) in sterile water for injection (284.5 g) was mixed with 18.4 g Lipoid E 80SN (Lipoid GmbH, Ludwigshafen, Germany), 322 g perfluorooctyl bromide (PFOB) and 8.62 g 1-perfluoro-n-hexy-decane (both abcr GmbH, Karlsruhe, Germany). This mixture was pre-emulsified with a tip sonicator (Branson Ultrasonics Corporation) for 5 min in portions of 50 ml and then transferred to a CF1 homogenizer (Constant Systems Ltd., Daventry, UK) for 10 cycles of 1000 bar. Then, after renewed emulsification with the tip sonicator, the resulting emulsion was autoclaved for 15 min at 121 °C. The nanoparticles of the final emulsion had a typical mean of the number average size distribution of 170–190 nm measured by dynamic light scattering (Malvern Panalytical). The PFOB-NE was injected 3–4 days after production and was stored at 4 °C.

T1 measurements

To optimize the flip angle for each resonance individually, T1 was determined in a 10% PFOB fat–water emulsion. To estimate T1, the flip angle of an FID MRS sequence (TE > 1.5 ms, dwell time = 50 µs) was varied (α = 0°–45°) in steps of 1° at a fixed TR = 30 ms, again using the single-channel Tx/Rx 19F solenoid coil. For RF excitation, a sinc pulse (duration: 3 ms, spectral width: 100 ppm) was used to ensure homogeneous excitation over the entire PFOB spectrum. Spectral boundaries for five resonances (CF2Br, CF3 and the three most prominent peaks (β,δ+ϵ,η) from the CF2 multi-peak spectrum) were chosen, over which the signal strength was integrated for each flip angle. T1 was then estimated for each of these resonances using the FLASH signal equation [25].

Phantom experiments

To test the validity of the method, phantom experiments were performed using the coil setup for the in vivo experiments, which consists of an eight-channel Tx loop coil array and an eight-channel Rx coil array containing two coils posterior and a six-channel anterior flexible array of coils [28]. The measurement was performed using the HE UTE sequence with TE = 1.7 ms, TR = 10 ms, readout bandwidth = 390 Hz/px, with 377 spirals and 26 spokes per spiral. The FWHM of the point-spread function was 15.7 mm in every direction, considering both regridding artifacts and T2*-decay, as well as the CF2 off-resonances. Regridding was performed using a voxel size of (5 mm)3. 8 Phantoms containing a 5% PFOB fat–water emulsion were measured ten times consecutively, resulting in a total acquisition time of 65:20 min:s. Due to the uneven distribution of noise in radial images, this allowed for a better noise estimation by creation of a standard deviation map [29]. SNR values of the reconstructed images for each peak and of the combined image were compared.

Comparison with iterative deconvolution

We compared the HE UTE reconstruction with an iterative deconvolution as it is the most commonly used method for chemical shift artifact correction of PFOB, and can be applied to the same data set, allowing for a direct comparison. For this, we acquired an image of the same phantoms as before using the HE UTE sequence (TE = 1.7 ms, TR = 10 ms, readout bandwidth = 390 Hz/px, with 1597 spirals and 26 spokes per spiral) and measured an FID with the same bandwidth and TR with 500 averages. Then we applied an iterative sparse deconvolution algorithm to the signal for which each peak was excited with a positive sign, minimizing the function

x=argminx~Cx~-y22+λx~1, 7

where C is the convolution matrix calculated from the FID signal, y the acquired data and λ an arbitrary factor for LASSO (least absolute shrinkage and selection operator) regularization [18, 30]. λ was chosen by trial and error, such that the final image yielded the maximum SNR without showing major artifacts. Reconstruction was done with Matlab (R2023a, The MathWorks Inc., Natick, USA) and took 2–4 h for each value of λ using a GPU (NVIDIA GeForce RTX 2080 Super). SNR was measured for both methods, adjusted by a factor of 2 for the deconvolution method, as only ¼ of the data was used, and compared.

In vivo experiments

One healthy juvenile pig (German landrace, 63 kg) was injected with a PFOB nanoemulsion (5 ml per kilogram body weight), and 19F-MRI scans were performed with the proposed HE UTE sequence 2, 4 and 7 days after the injection. To estimate the evolution of the 19F signal over time, the mean values of the acquired signals were compared in regions of the spleen that were equidistant from the coils in all measurements. Anesthesia was introduced and maintained as described previously [31]. The MRI protocol consisted of a T1 VIBE-Dixon sequence for 1H images, with TE = 1.29 ms, TR = 3.97 ms, readout bandwidth 1042 Hz/px and a matrix size of 260 × 320 per slice, 52 slices, voxel size 1.19 mm × 1.19 mm × 5 mm acquired in one breath hold through a brief interruption of mechanical ventilation. The HE UTE sequence to measure the 19F signal had the following acquisition parameters: TE = 1.7 ms, TR = 10 ms, readout bandwidth = 390 Hz/Px, with 1597 spirals and 26 spokes per spiral, resulting in a total acquisition time of 27:41 min:s. All measurements were performed on a clinical 3 T system (Prisma, Siemens Healthineers, Erlangen, Germany) and the 19F images were reconstructed using Kaiser–Bessel regridding in Matlab (R2023a, The MathWorks Inc., Natick, USA).

Results

T1 measurements

Figure 4a shows the PFOB spectrum together with the FLASH signal curves of the CF3 resonance, the CF2Br resonance and the most prominent resonances (β, δ + ε, η) of the CF2 multi-peak (Fig. 4b). T1 values of the different PFOB resonances were between 870 and 970 ms for all resonances (Table 1). Thus, at a TR of 10 ms as used in the imaging sequence, the Ernst angles differ by less than 0.5°. A variation of the flip angle by that amount changes the signal by only 0.3%. Therefore, all resonances were excited with the same flip angle (α = 8°) in future measurements.

Fig. 4.

Fig. 4

a Spectrum of perfluorooctyl bromide (PFOB), b measured FLASH signal curves of different resonances at TR = 10 ms

Table 1.

T1 values for different PFOB resonances together with their frequency range relative to the central resonance

Resonance T1 [ms] Frequency range [ppm]
CF2Br 966 ± 8 [12.6, 24.1]
CF3 949 ± 8 [− 4.0, 2.6]
CF2 (β) 874 ± 15 [− 37.0, − 34.3]
CF2 (δ, ε) 952 ± 12 [− 41.0, − 39.7]
CF2 (η) 940 ± 12 [− 53.6, − 43.6]

RF pulse envelopes

A comparison between signals of Gaussian and sinc pulse envelopes is shown in Fig. 5. As expected, the SNR of the CF2 multi-peak was increased with the sinc pulse due to the more homogeneous excitation of its full spectrum leading to an SNR gain of 50%. Due to a broader excitation of the spectrum, a sinc envelope also increases chemical shift artifacts from these resonances. No significant difference was observed for the other resonances. Thus, for all further measurements, a Gaussian pulse was used for the CF2Br and CF3 resonances (δωGauss=541Hz) and a sinc pulse was used for the CF2 multi-peak (δωSinc=4000Hz).

Fig. 5.

Fig. 5

Comparison between excitation with Gaussian and sinc-shaped RF pulses for all PFOB resonances

Phantom experiments

Images and noise maps are seen in Fig. 6. SNR values for the single resonance images were SNR(CF2BR) = 55, SNR(CF3) = 65 and SNR(CF2) = 120 and for the combined image SNR(total) = 139. Thus, the SNR gain compared to the CF2 image was 15%.

Fig. 6.

Fig. 6

19F Phantom images acquired with the HE UTE sequence. Top: Comparison of individual reconstructions of each resonance with the combined image (mean of 10 averages). Bottom: Comparison of noise between each resonance and the combined image (Standard deviation of 10 averages)

Comparison with iterative deconvolution

Reconstruction of the data with both Hadamard-decoding and iterative deconvolution are shown in Fig. 7. The best reconstruction was found for λ=3·10-8. We measured an SNR of 297 for Hadamard-decoding and 133 for iterative deconvolution. The SNR ratio was SNRHD2SNR(ID)=1.11 between both methods.

Fig. 7.

Fig. 7

Image reconstructed with Hadamard-decoding (left) and iterative deconvolution (right)

In vivo results

Figure 8 shows the 19F signal in the spleen of a pig 2 days after injection of 5 ml/kg of PFOB-NE using the HE UTE 19F technique. A 19F signal with SNR over 100 was observed in the spleen, confirming the viability of the method of HE for this application. To quantify the temporal change of the PFOB concentration, the spleen was segmented on the 1H images. Figure 9 shows the average 19F signal intensity inside the spleen at 2, 4 and 7 days after injection on a logarithmic scale. By fitting the signal, the half-life of PFOB in the spleen is roughly estimated to be 9 ± 5 days. The adjusted R2 of the fit was 0.97 with a RMSE of 8% of the day 2 signal. Further PFOB signals were found in liver, bone marrow and subcutaneous fat in all measurements.

Fig. 8.

Fig. 8

a Combined 1H (gray) and 19F (green) image of a porcine abdomen in vivo 2 days after PFOB-NE injection. b Outlines of a segmentation of the spleen together with the fluorine signal. The three green spots outside of the body of the pig are 19F marker vials. Due to blurring from the CF2 off-resonances, the edges appear less sharp in the 19F-signal

Fig. 9.

Fig. 9

19F signal in the spleen at different times after injection on a logarithmic scale

Discussion and conclusion

In this work, a HE 3D UTE sequence was successfully implemented to image 19F PFOB in vivo. In phantom experiments, the T1 values of the PFOB resonances varied by only 3% so that a single flip angle could be used for all resonances. Exciting the CF2 resonances with a sinc pulse instead of a Gaussian pulse increased the SNR substantially, as the sinc pulse covers a larger spectral range and includes all CF2 resonances. With these optimizations, an SNR > 100 could be achieved in vivo which enabled the observation of the decreasing 19F signal in the spleen over a time-course of 7 days caused by the natural decay of monocytes.

Many 19F-MRI studies have been conducted so far in mice to image the spleen. For this, a range of different PFCs, such as PFOB [17, 21, 30], perfluorocrownether (PFCE) [16], FC-43 [14] or perfluoro(t-butylcyclohexane) (ABL-101) [32] were used and chemical shift artifact correction was performed with a variety of methods, for example iterative deconvolution [30] or spectroscopic imaging [21]. In this study a large animal model was used, and the chemical shift correction of the multi-peak 19F signal was performed by HE. The large animal model was chosen due to its geometric similarity to humans [33], as we want to use this method in studies of coronary artery interventions after myocardial infarction in the future [31, 34]. In general, a direct comparison of the SNR in this study with previous small animal studies is difficult due to the use of different RF coils, field strengths and voxel dimensions. Only few studies on pigs at a clinical 3 T system were performed which focused on imaging macrophages and other myeloid cells in the infarcted myocardium [11, 12]. So far, these studies used selective excitation of a single peak of the PFOB spectrum. Here, we have shown the viability of HE to image the whole PFOB spectrum at 3 T: HE enables a combination of the individual resonance signals to remove the chemical shift artifact, and it even allows to study the different resonances separately. With HE an increase in SNR by 15% was achieved over the excitation of a single resonance (CF2) and by 11% over using a time intensive iterative deconvolution approach, but the use of the composite HE RF pulses led to an increased TE, and thus to a slightly reduced SNR. Here, a TE = 1.7 ms was used, whereas a TE = 0.1 ms was possible for a single resonance. To further shorten the echo time in HE, asymmetric pulses could be applied which would lead to an estimated value of TE of about 0.2 ms which would further increase the 19F-MRI signal in a HE acquisition.

The HE signal shows a high 19F concentration in the spleen as seen in Fig. 7. 19F signal is also visible outside of the spleen, which is a consequence of the remaining uncorrected chemical shifts of the CF2 multi-peaks. This effect could be reduced either by exciting a narrower part of the CF2 spectrum, which in turn would lower the SNR, or using a deconvolution approach on top. A perfect deconvolution would remove all chemical shift artifacts; however, in practice deconvolution is a mathematically ill-posed problem that cannot be solved analytically and requires a computationally intensive reconstruction process. With the current sequence parameter settings, the blurring of the multi-peak CF2 signals is acceptable, as it extends about two pixels only. This could be reduced further by an increased pixel bandwidth. The HE method, however, is very simple as compared to a deconvolution approach, and it can be integrated directly into the image reconstruction so that 19F images are directly available after the image acquisition.

Over the course of the 30 min-long measurement breathing is expected to cause motion artifacts, which are not seen in underlying 1H images which were acquired in a breath hold. As 19F data are acquired with radial sampling, which is largely insensitive to periodic motion, these artifacts are expected to be small. Furthermore, they could be corrected using methods based on the low-frequency k-space data such as self-gating [35], which allows for sorting of the data into different phases of the breathing cycle at the cost of SNR.

A limitation of the longitudinal animal experiment is the comparison of the 19F signals acquired on subsequent days: the spleen’s position varied between measurement days, which can cause variations in its positioning within the Rx sensitivity profile. Nevertheless, the acquired data provide a good estimate of the temporal behavior of the injected PFOB, as they are integrated over the full volume of the spleen. Furthermore, the positioning of the animal was performed carefully to minimize re-positioning errors. All acquisitions show an SNR > 100 within the spleen, and, therefore, produce a relevant 19F signal.

In summary, using optimized flip angles and excitation pulses, HE 19F-MRI can be successfully applied to detect PFOB with high SNR in large animals. With this method, longitudinal studies might become feasible, for example to study the inflammatory reaction after acute cardiovascular events.

Acknowledgements

The authors would like to thank Dr. Heidi Cristina Ramona Schmitz for assisting with the animal experiments, Phillip Groß for his contribution to NMR measurements, Prof. Dominik von Elverfeldt for providing materials for phantoms, and Dr. Jochen Leupold and Dr. Wilfried Reichert for additional fluorine measurements at high fields. The authors thank the Core Facility AMIRCF (DFG-RIsources N° RI_00052) for support in these measurements.

Funding

Open Access funding enabled and organized by Projekt DEAL. This work was supported by the German Science Foundation (DFG) under grant number 492563001 to Alexander Maier and Michael Bock and through the SFB 1425 (project number 422681845). Alexander Maier received support from the Berta Ottenstein Program for Advanced Clinician Scientists, Faculty of Medicine, University of Freiburg. David Boll was supported by a grant of the German Heart Foundation.

Data availability

Data will be provided upon reasonable request by the corresponding author.

Declarations

Conflict of interest

All authors declare no conflict of interest.

Ethical approval

The animal experiment was conducted in accordance with FELASA and GV-SOLAS standards for animal welfare. Experiments were approved by the local ethics committee of Freiburg University and the regional council of Freiburg, Baden-Wuerttemberg, Germany (licence number 35-9185.81/G-23/012).

Footnotes

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Maier A, Teunissen AJP, Nauta SA, Lutgens E, Fayad ZA, van Leent MMT (2024) Uncovering atherosclerotic cardiovascular disease by PET imaging. Nat Rev Cardiol. 10.1038/s41569-024-01009-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Senders ML, Meerwaldt AE, van Leent MMT, Sanchez-Gaytan BL, van de Voort JC, Toner YC, Maier A, Klein ED, Sullivan NAT, Sofias AM, Groenen H, Faries C, Oosterwijk RS, van Leeuwen EM, Fay F, Chepurko E, Reiner T, Duivenvoorden R, Zangi L, Dijkhuizen RM, Hak S, Swirski FK, Nahrendorf M, Pérez-Medina C, Teunissen AJP, Fayad ZA, Calcagno C, Strijkers GJ, Mulder WJM (2020) Probing myeloid cell dynamics in ischemic heart disease by nanotracer hot spot imaging. Nat Nanotechnol 15:398–405 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.van Heeswijk RB, Bauer WR, Bönner F, Janjic JM, Mulder WJM, Schreiber LM, Schwitter J, Flögel U (2023) Cardiovascular molecular imaging with fluorine-19 MRI: the road to the clinic. Circ Cardiovasc Imaging 16:e014742 [DOI] [PubMed] [Google Scholar]
  • 4.Flögel U, Ding Z, Hardung H, Jander S, Reichmann G, Jacoby C, Schubert R, Schrader J (2008) In vivo monitoring of inflammation after cardiac and cerebral ischemia by fluorine magnetic resonance imaging. Circulation 118:140–148 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Maier A, Toner YC, Munitz J, Sullivan NAT, Sakurai K, Meerwaldt AE, Brechbühl EES, Prévot G, van Elsas Y, Maas RJF, Ranzenigo A, Soultanidis G, Rashidian M, Pérez-Medina C, Heo GS, Gropler RJ, Liu Y, Reiner T, Nahrendorf M, Swirski FK, Strijkers GJ, Teunissen AJP, Calcagno C, Fayad ZA, Mulder WJM, van Leent MMT (2023) Multiparametric immunoimaging maps inflammatory signatures in murine myocardial infarction models. JACC Basic Transl Sci 8:801–816 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Maxouri O, Bodalal Z, Daal M, Rostami S, Rodriguez I, Akkari L, Srinivas M, Bernards R, Beets-Tan R (2023) How to 19F MRI: applications, technique, and getting started. BJR|Open 5:20230019 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Srinivas M, Heerschap A, Ahrens ET, Figdor CG, de Vries IJM (2010) 19F MRI for quantitative in vivo cell tracking. Trends Biotechnol 28:363–370 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Jacoby C, Temme S, Mayenfels F, Benoit N, Krafft MP, Schubert R, Schrader J, Flögel U (2014) Probing different perfluorocarbons for in vivo inflammation imaging by 19F MRI: image reconstruction, biological half-lives and sensitivity. NMR Biomed 27:261–271 [DOI] [PubMed] [Google Scholar]
  • 9.Shukla HP, Mason RP, Woessner DE, Antich PP (1995) A comparison of three commercial perfluorocarbon emulsions as high-field 19F NMR probes of oxygen tension and temperature. J Magn Reson B 106:131–141 [Google Scholar]
  • 10.Colotti R, Bastiaansen JAM, Wilson A, Flögel U, Gonzales C, Schwitter J, Stuber M, van Heeswijk RB (2017) Characterization of perfluorocarbon relaxation times and their influence on the optimization of fluorine-19 MRI at 3 tesla. Magn Reson Med 77:2263–2271 [DOI] [PubMed] [Google Scholar]
  • 11.Rothe M, Jahn A, Weiss K, Hwang J-H, Szendroedi J, Kelm M, Schrader J, Roden M, Flögel U, Bönner F (2019) In vivo 19F MR inflammation imaging after myocardial infarction in a large animal model at 3 T. Magma NY N 32:5–13 [DOI] [PubMed] [Google Scholar]
  • 12.Bönner F, Gastl M, Nienhaus F, Rothe M, Jahn A, Pfeiler S, Gross U, Schultheiss H-P, Ibanez B, Kozerke S, Szendroedi J, Roden M, Westenfeld R, Schrader J, Flögel U, Heusch G, Kelm M (2022) Regional analysis of inflammation and contractile function in reperfused acute myocardial infarction by in vivo 19F cardiovascular magnetic resonance in pigs. Basic Res Cardiol 117:21 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Lee HK, Nalcioglu O (1992) Reblurred deconvolution method for chemical shift removal in F-19 (PFOB) MR imaging. J Magn Reson Imaging JMRI 2:53–61 [DOI] [PubMed] [Google Scholar]
  • 14.Busse LJ, Pratt RG, Thomas SR (1988) Deconvolution of chemical shift spectra in two- or three-dimensional [19F] MR imaging. J Comput Assist Tomogr 12:824–835 [DOI] [PubMed] [Google Scholar]
  • 15.Chen J, Ahrens ET, Pal P (2023) Physics-driven learned deconvolution of multi-spectral cellular MRI with radial sampling. In: 2023 57th asilomar conf. signals syst. comput., pp 1453–1457 [DOI] [PMC free article] [PubMed]
  • 16.Ludwig KD, Hernando D, Roberts NT, van Heeswijk RB, Fain SB (2018) A chemical shift encoding (CSE) approach for spectral selection in fluorine-19 MRI. Magn Reson Med 79:2183–2189 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.van Heeswijk RB, Colotti R, Darçot E, Delacoste J, Pellegrin M, Piccini D, Hernando D (2018) Chemical shift encoding (CSE) for sensitive fluorine-19 MRI of perfluorocarbons with complex spectra. Magn Reson Med 79:2724–2730 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Schoormans J, Calcagno C, Daal MRR, Wüst RCI, Faries C, Maier A, Teunissen AJP, Naidu S, Sanchez-Gaytan BL, Nederveen AJ, Fayad ZA, Mulder WJM, Coolen BF, Strijkers GJ (2020) An iterative sparse deconvolution method for simultaneous multicolor 19 F-MRI of multiple contrast agents. Magn Reson Med 83:228–239 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Bock M, Ilbey S, Maier A, Özen AC (2022) Chemical shift artifact correction for 19F MRI using spectral Hadamard encoding. Proc Intl Soc Mag Reson Med. 10.58530/2022/1008 [Google Scholar]
  • 20.Goelman G (1994) Fast Hadamard Spectroscopic Imaging Techniques. J Magn Reson B 104:212–218 [DOI] [PubMed] [Google Scholar]
  • 21.Yildirim M, Kovalyk X, Scholtz P, Schütz M, Lindemeyer J, Lamerichs R, Grüll H, Isik EO (2024) Fast 19 F spectroscopic imaging with pseudo-spiral k-space sampling. NMR Biomed 37:e5086 [DOI] [PubMed] [Google Scholar]
  • 22.Bley TA, Wieben O, François CJ, Brittain JH, Reeder SB (2010) Fat and water magnetic resonance imaging. J Magn Reson Imaging 31:4–18 [DOI] [PubMed] [Google Scholar]
  • 23.Fletcher DW, Haselgrove JC, Bolinger L (1999) High-resolution imaging using Hadamard encoding. Magn Reson Imaging 17:1457–1468 [DOI] [PubMed] [Google Scholar]
  • 24.Souza SP, Szumowski J, Dumoulin CL, Plewes DP, Glover G (1988) SIMA: simultaneous multislice acquisition of MR images by Hadamard-encoded excitation. J Comput Assist Tomogr 12:1026–1030 [PubMed] [Google Scholar]
  • 25.Bernstein MA, King KF, Zhou XJ (2004) Handbook of MRI Pulse Sequences. Academic Press, Amsterdam, Boston [Google Scholar]
  • 26.Piccini D, Littmann A, Nielles-Vallespin S, Zenge MO (2011) Spiral phyllotaxis: the natural way to construct a 3D radial trajectory in MRI. Magn Reson Med 66:1049–1056 [DOI] [PubMed] [Google Scholar]
  • 27.Zhu X, Tan F, Johnson K, Larson P (2021) Optimizing trajectory ordering for fast radial ultra-short TE (UTE) acquisitions. J Magn Reson 327:106977 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Özen AC, Spreter F, Schimpf W, Fischer J, Ilbey S, Reiss S, Maier A, von Elverfeldt D, Heidt T, von Zur MC, Bock M (2023) Scalable and modular 8-channel transmit and 8-channel flexible receive coil array for 19 F MRI of large animals. Magn Reson Med 89:1237–1250 [DOI] [PubMed] [Google Scholar]
  • 29.Dietrich O, Raya JG, Reeder SB, Reiser MF, Schoenberg SO (2007) Measurement of signal-to-noise ratios in MR images: influence of multichannel coils, parallel imaging, and reconstruction filters. J Magn Reson Imaging 26:375–385 [DOI] [PubMed] [Google Scholar]
  • 30.Meissner M, Reisert M, Hugger T, Hennig J, von Elverfeldt D, Leupold J (2015) Revealing signal from noisy (19) F MR images by chemical shift artifact correction. Magn Reson Med 73:2225–2233 [DOI] [PubMed] [Google Scholar]
  • 31.Heidt T, Reiss S, Krafft AJ, Özen AC, Lottner T, Hehrlein C, Galmbacher R, Kayser G, Hilgendorf I, Stachon P, Wolf D, Zirlik A, Düring K, Zehender M, Meckel S, von Elverfeldt D, Bode C, Bock M, von zur Mühlen C (2019) Real-time magnetic resonance imaging—guided coronary intervention in a porcine model. Sci Rep 9:8663 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Darçot E, Colotti R, Brennan D, Deuchar GA, Santosh C, van Heeswijk RB (2020) A characterization of ABL-101 as a potential tracer for clinical fluorine-19 MRI. NMR Biomed 33:e4212 [DOI] [PubMed] [Google Scholar]
  • 33.Meissner F, Galbas MC, Szvetics S, von zur Mühlen C, Heidt T, Maier A, Bock M, Czerny M, Bothe W, Reiss S (2024) Cardioaortic dimensions in German landrace pigs derived from cardiac magnetic resonance imaging. Sci Rep 14:1869 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Özen AC, Lottner T, Reiss S, Heidt T, Zur Mühlen CV, Bock M (2022) Real-time control of active catheter signals for better visual profiling during cardiovascular interventions under MRI guidance. IEEE Access 10:20581–20589 [Google Scholar]
  • 35.Larson AC, White RD, Laub G, McVeigh ER, Li D, Simonetti OP (2004) Self-gated cardiac cine MRI. Magn Reson Med 51:93–102 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Data will be provided upon reasonable request by the corresponding author.


Articles from Magma (New York, N.y.) are provided here courtesy of Springer