Skip to main content
Pathogens and Global Health logoLink to Pathogens and Global Health
. 2025 Sep 11;119(8):311–320. doi: 10.1080/20477724.2025.2551508

Diagnostic performance of hemozoin-based magneto-optical detection assay and RDT: a prospective observational study

Vishnu Teja Nallapati a, Manoj A R a, Sushma Belurkar b, Naveenchandra Kulal c, Prashanth Bhat d, Shama Prasada K e, Nitin Gupta a, Manjunath H Hande f, Priyaleela Thota g, David Bell g, Kavitha Saravu a,
PMCID: PMC12642893  PMID: 40934287

ABSTRACT

Early detection and effective management of malaria are crucial for elimination efforts. Microscopy and rapid diagnostic tests (RDTs) have been the main diagnostic methods for over fifteen years, but they have limitations, especially in cases of low parasite density or deletions of target markers (HRP2/3). This study compares the diagnostic performance of a novel hemozoin-based diagnostic assay (Hz-MOD) with RDTs for detecting malaria in febrile patients in southwestern India. A prospective observational study involved 480 patients screened with Hz-MOD, RDT, microscopy, and nested PCR. Among the samples, 121 were positive by both microscopy and PCR. The sensitivity of Hz-MOD was 94.21% compared to microscopy and 91.74% compared to PCR. For RDTs, sensitivity was 90.91% compared to microscopy and 87.60% compared to PCR. In terms of specificity, Hz-MOD showed 98.61% compared to microscopy and 97.77% compared to PCR, while RDTs had 100% specificity against microscopy and 98.89% against PCR. These results suggest that the hemozoin-based test demonstrates similar sensitivity to RDTs and could serve as an effective screening tool for malaria detection.

KEYWORDS: Hemozoin, magneto-optical detection, rapid diagnostic test, malaria diagnosis, diagnostic performance

1. Introduction

Malaria is a vector-borne disease transmitted by infected female Anopheles mosquitoes and caused by the genus Plasmodium. Five parasite species cause malaria in humans, and two of these species – Plasmodium falciparum and Plasmodium vivax – pose the greatest threat. The infection is endemic in 83 countries, with approximately 263 million cases and 597,000 deaths reported globally in 2023 [1]. Regardless, proper malaria diagnosis and the species differentiation of Plasmodium parasites are still challenging. In 2015, the Global Technical Strategy for Malaria 2016–2030 was adopted by the World Health Organization (WHO) to reduce global malaria incidence and mortality rates by at least 90% by 2030 [2]. This can only be achieved by accurate diagnosis and the appropriate treatment of malaria cases. Even though the available control strategies are effective, they are limited by the ineffectiveness of early diagnostic tools for detection, especially at the low parasite densities required for surveillance in low-transmission settings [3].

As per WHO recommendations, malaria is primarily diagnosed using either microscopy or Rapid Diagnostic Tests (RDT). Even though malaria microscopy is the gold standard technique for diagnosis, it is labor-intensive and highly dependent on the laboratory staff’s degree of expertise. The RDT is an immunochromatographic method in which antibodies are used to capture one or several parasite-specific antigens, such as Plasmodium falciparum histidine-rich protein 2 (HRP-2) and Plasmodium lactate dehydrogenase (pLDH) or aldolase. More than 200 different RDTs are commercially available and are rapid, easy to perform, and relatively inexpensive. Drawbacks include their inability to quantify parasite density, false positive results due to the persistence of HRP2 in the blood sample for several days after clearance of infection, and false negative results due to the deletion of the target markers (pfhrp-2/3 genes) [4,5]. In the past decade, alternative and advanced methods such as Polymerase Chain Reaction (PCR), quantitative PCR (qPCR), real-time PCR (RT PCR), and Loop-mediated isothermal amplification (LAMP) methods have been developed and successfully tested. Although these methods have higher sensitivity and specificity, they are either expensive or require expert personnel. Due to the need for advanced laboratory facilities, highly trained staff, and high costs, these tests cannot be utilized in resource-limited settings [6]. However, the surveillance of low-density parasitemia is crucial for malaria control and elimination, which requires a highly sensitive and field-deployable diagnostic test.

Recently, multiple novel Hz-MOD assays have been developed to detect malaria parasites. These assays are rapid, robust, and sensitive devices that detect hemozoin particles in the blood of malaria-infected patients. Gazelle is one such device, which is a portable, battery-operated point-of-care hemozoin-based malaria diagnostic device with the principle of magneto-optical detection [7–9]. Hence, this study aimed to compare the diagnostic performance of Hz-MOD assays and RDT with microscopy and PCR as reference standards.

2. Materials and methods

2.1. Study site

A multi-centric prospective observational study was conducted in Kasturba Medical College, Manipal, Malaria Clinic, District Hospital, Udupi, and Malaria Clinic, Govt. Wenlock Hospital Mangalore, between July 2022 and September 2024.

2.2. Study settings

Both Udupi (13.3330, 74.7426) and Mangalore (12.8678, 74.8430) are located in the coastal Karnataka region of India, a tropical area characterized by a malaria annual parasite index of 0.01 in Udupi and 0.03 in Mangalore for 2023, with seasonal peaks coinciding with the high rainfall during the monsoon months [10]. The region’s unique combination of urban and rural landscapes, including wetland habitats, provide ideal conditions for mosquito breeding and disease transmission.

2.3. Sample size

Assuming sensitivity and specificity to be 90% with 6% of the absolute precision sample size required is 96 malaria-positive and 96 malaria-negative samples. A total of 480 samples were screened to achieve this.

2.4. Study procedure

The study was approved by the institutional ethics committee of Kasturba Medical College and Kasturba Hospital (IEC-352/2021) as well as by the technical advisory committee, Directorate of Health & Family Welfare Services, Bengaluru, Karnataka (NHM/SPM/4-Part file/2020–21). The confirmed positive and negative malaria cases (febrile patients) by Microscopy were enrolled after obtaining written informed consent. After recruitment, 2 ml of venous blood samples were collected from all the patients before the administration of antimalarials. Then all the samples were transported to the laboratory at Kasturba Medical College for further analysis.

2.5. Laboratory investigations

2.5.1. Malaria microscopy

Thin and thick blood smears were prepared and stained with Leishman’s stain, and two experienced microscopists examined them. A minimum of 200 oil immersion fields were assessed to confirm negative results. The parasite density (parasites/µL) was calculated by dividing the number of asexual parasites counted by the number of white blood cells (WBCs) counted, and then multiplying by an assumed WBC density of 8,000 per µL [11].

2.5.2. RDT

All the samples were tested using the Malaria Ag P.f/P.v test (Abbott Bioline™) and interpreted according to the manufacturer’s instructions. This malaria RDT targets detecting histidine-rich protein II (HRP-II) antigen of P. falciparum and lactate dehydrogenase (pLDH) of P. vivax species in human whole blood.

2.5.3. Hemozoin-based magneto-optical detection assay

A hemozoin-based magneto-optical device, known as Gazelle, developed by Hemex Health, U.S.A., was used for malaria detection. It is a small tabletop reader along with single-use cartridges. Prior to the sample processing, positive and negative controls, along with a blank (empty cartridge), were analyzed for quality control (QC). All the samples were processed as per the manufacturer’s instructions. Briefly, 65 µL of the assay buffer was added to the test cartridge (a disposable single-use item), followed by the addition of 30 µL of whole blood, and then the cartridge was loaded into the device for interpretation. In three minutes, the device displayed the results, and those were recorded.

2.5.4. Molecular diagnosis

DNA extraction was done using the QIAamp DNA Blood Mini Kit (Qiagen, Germany) and eluted with 100 μL of elution buffer. The DNA was amplified by nested PCR to detect malaria parasites using genus/species-specific primers targeting the 18S rRNA gene following the protocol of Snounou et al. [12]. All reactions were carried out in a final volume of 25 µL containing 2 mM MgCl2, 50 mM KCl, 10 mM Tris – HCl pH 8.3, 125 µM of each dNTPs, 250 nM of each oligonucleotide primer, 0.5 units of Taq DNA polymerase, and 2 µL of template DNA. The amplification of amplicons were visualized on a 1.5% agarose gel with EtBr staining.

2.6. Statistical analysis

Two-by-two contingency tables were constructed to include True Positives (TP), True Negatives (TN), False Positives (FP), and False Negatives (FN) for each index test, as well as various combinations of the components of the combination tests, compared to the reference tests. We analyzed the results for sensitivity, specificity, positive and negative likelihood ratios, and diagnostic odds ratio (DOR). Likelihood ratios for positive (LR+) and negative (LR−) test results were considered good when LR+ was > 10, and LR− < 0.1 [13]. The higher DOR indicates better test performance, with values >1 suggesting the test discriminates well between diseased and non-diseased individuals. For example, a DOR of 10 means the odds of a positive test are 10 times higher in those with the condition [14]. Diagnostic accuracy was assessed through receiver-operating characteristics (ROC) curves, and the area under the curve (AUC) was interpreted as follows: 0.9–1.0, excellent; 0.8–0.9, very good; 0.7–0.8, good; 0.6–0.7, sufficient; 0.5–0.6, bad; <0.5, test not useful [15]. The data analysis was performed by using SPSS version 20 and MedCalc’s software [16].

3. Results

A total of 480 febrile patients were enrolled, with 121 confirmed positive by microscopy and 359 confirmed negative for malaria (Figure 1). However, we observed discrepancies between the microscopy and PCR results. Of the total samples, 68.5% were males and 31.5% were females; their mean age was 38.07 ± 16.85 years. Among the malaria-positive cases, 105 (86.77%) were P. vivax, 11 (9.09%) were P. falciparum, and 5 (4.13%) were mixed infections. The median parasite density of the samples was 3388 parasites/μL (interquartile range [IQR]: 988 parasites/μL, 8827 parasites/μL). In P. vivax samples, the median parasite density was 2456 parasites/μL (IQR: 917 parasites/μL, 7484 parasites/μL), whereas it was 8294 parasites/μL (IQR: 1751 parasites/μL 12,088 parasites/μL) in P. falciparum samples. When evaluating diagnostic performance, out of 121 microscopy-positive malaria samples, 114 tested positive by Hz-MOD, 110 by RDT, and 118 by PCR (Figure 1). Conversely, among the 121 PCR-positive malaria samples, 111 tested positive by Hz-MOD, 106 by RDT, and 118 by microscopy (Figure 1).

Figure 1.

Figure 1.

Flow chart illustrating the diagnostic workflow and test results for the 480 patients recruited in the study.

3.1. Diagnostic accuracy of index tests as compared to microscopy

The sensitivity and specificity of the Hz-MOD assay were 94.21% (95% CI: 88.44–97.64) and 98.61% (95% CI: 96.78–99.55), while the RDT sensitivity and specificity were 90.91% (95% CI: 84.32–95.37) and 100% (95% CI: 98.98–100), respectively (Table 1). The pooled estimates for the LR+, LR−, and DOR for hemozoin-based diagnostic assay were 67.65 (95% CI: 28.30–161.71), 0.06 (95% CI: 0.03–0.12), and 1153.03 (95% CI: 358.97–3703.57), respectively, and the overall LR− and DOR for RDT were 0.09 (95% CI: 0.05–0.16), and 6908.65 (95% CI: 403.85–118186.58), respectively (Table 1). However, LR+ for RDTs was not available. The AUC for Hz-MOD and RDT were 0.964 (95% CI: 0.939–0.989) and 0.955 (95% CI: 0.924–0.985), respectively (Figure 2).

Table 1.

Diagnostic performance of Hz-MOD and RDT for diagnosis of malaria.

Reference standard Index tests Sensitivity % (95%CI) Specificity % (95%CI) Positive Likelihood Ratio (95%CI) Negative Likelihood Ratio (95%CI) Diagnostic Accuracy
(95%CI)
Diagnostic Odds Ratio (95%CI) P value
Microscopy as the reference standard Hz-based diagnostic assay 94.21%
(88.44–97.64)
98.61
(96.78–99.55)
67.65
(28.30–161.71)
0.06
(0.03–0.12)
97.50
(95.67–98.70)
1153.02
(358.97–3703.57)
 < 0.0001
RDT 90.91
(84.32–95.37)
100 (98.98–100) 0.09
(0.05–0.16)
97.71
(96.94–98.85)
6908.65
(403.84–118186.58)
 < 0.0001
PCR as the reference standard Hz-based diagnostic assay 91.74
(85.33–95.97)
97.77
(95.66–99.03)
41.17
(20.70–81.85)
0.08
(0.05–0.15)
96.25
(94.14–97.76)
487.01
(187.61–1264.18)
 < 0.0001
RDT 87.60
(80.38–92.89)
98.89
(97.17–99.70)
78.62
(29.60–208.83)
0.13
(0.08–0.20)
96.04
(93.89–97.60)
627.16
(203.79–1930.04)
 < 0.0001

Abbreviations: TP-true positive; FP-false positive; FN-false negative; TN-true negative; PCR – polymerase chain reaction; RDT – rapid diagnostic tests; Hz-MOD – hemozoin-based magneto-optical detection assay.

Figure 2.

Figure 2.

ROC curves for Hz-MOD and RDT compared to (a) microscopy and (b) PCR.

In P. vivax samples, the sensitivity and specificity of the Hz-MOD assay were 94.29% (95% CI: 87.98–97.87) and 98.61% (95% CI: 96.78–99.55), respectively, while the RDT sensitivity and specificity were 90.48% (95% CI: 83.18–95.34) and 100% (95% CI: 98.98–100). In contrast, in P. falciparum samples, the sensitivity and specificity of the Hz-MOD assay were 90.91% (95% CI: 58.72–99.77) and 98.61% (95% CI: 96.78–99.55), while the RDT sensitivity and specificity were both 100% (95% CI: 71.51–100.00) and 100% (95% CI: 98.98–100), respectively (Table 2).

Table 2.

Comparison of sensitivity and specificity of Hz-MOD and RDT for the detection of P. vivax and P. falciparum.

    Microscopy as a reference test
PCR as a reference test
    Sensitivity
% (95% CI)
Specificity
% (95% CI)
Sensitivity
% (95% CI)
Specificity
% (95% CI)
Hz-MOD In P. vivax samples 94.29
(87.98–97.87)
98.61
(96.78–99.55)
91.51
(84.49–96.04)
97.77
(95.66–99.03)
In P. falciparum samples 90.91
(58.72–99.77)
98.61
(96.78–99.55)
91.67
(61.52–99.79)
97.77
(95.66–99.03)
RDT In P. vivax samples 90.48
(83.18–95.34)
100
(98.98–100)
95.88
(89.78–98.87)
100
(98.98–100)
In P. falciparum samples 100
(71.51–100.00)
100
(98.98–100)
100
(73.54–100.00)
100
(98.98–100)

Abbreviations: PCR – polymerase chain reaction; RDT – rapid diagnostic tests; Hz-MOD – hemozoin-based magneto-optical detection assay.

For samples with high parasite density (> 200 parasites/μL), Hz-MOD showed a sensitivity of 95.54% (95% CI: 89.89–98.53), surpassing RDT, which had a sensitivity of 91.07% (95% CI: 84.19–95.64). For samples with low parasite density (< 200 parasites/μL), Hz-MOD demonstrated a sensitivity of 100% (95% CI: 66.37–100), while RDT showed a sensitivity of 88.89% (95% CI: 51.75–99.72%).

3.2. Diagnostic accuracy of index tests as compared to PCR

The sensitivity and specificity of the Hz-MOD assay were 91.74% (95% CI: 85.33–95.97) and 97.77% (95% CI: 95.66–99.03), while the RDT sensitivity and specificity were 87.60% (95% CI: 80.38–92.89) and 98.89% (95% CI: 97.17–99.70), respectively (Table 1). The pooled estimates for the LR+, LR−, and DOR for hemozoin-based diagnostic assay were 41.17 (95% CI: 20.70–81.85), 0.08 (95% CI: 0.05–0.15), and 487.01 (95% CI: 187.62–1264.19), respectively, and the overall LR+, LR−, and DOR for RDT were 78.62 (95% CI: 29.60–208.83), 0.13 (95% CI: 0.08–0.20), and 627.17 (95% CI: 203.79–1930.04), respectively (Table 1). The AUC for Hz-MOD and RDT were 0.948 (95% CI: 0.918–0.977) and 0.932 (95% CI: 0.897–0.968), respectively (Figure 2).

In P. vivax samples, the sensitivity and specificity of the Hz-MOD assay were 91.51% (95% CI: 84.49–96.04) and 97.77% (95% CI: 95.66–99.03), respectively, while the RDT sensitivity and specificity were 95.88% (95% CI: 89.78–98.87) and 100% (95% CI: 98.98–100). In contrast, in P. falciparum samples, the sensitivity and specificity of the Hz-MOD assay were 91.67% (95% CI: 61.52–99.79) and 97.77% (95% CI: 95.66–99.03), while the RDT sensitivity and specificity were both 100% (95% CI: 73.54–100.00) and 100% (95% CI: 98.98–100), respectively (Table 2).

4. Discussion

The Global Technical Strategy for Malaria 2016–2030 aims to reduce malaria incidence and mortality at least by 90% by 2030 [2]. Early detection and management of malaria is crucial; however, detecting low-density parasitemia is challenging. Furthermore, many countries are progressing from malaria control to elimination. Therefore, accurate, reliable, and field-deployable rapid diagnostic approaches are essential to achieve this goal. Thus, in this study, we compared the diagnostic performance of one such Hz-MOD assay called Gazelle by Hemex Health, U.S.A. [17,18], with the malaria RDT (PfHRP2/pLDH).

In our analysis, we observed that the overall diagnostic accuracy of Hz-MOD and RDT compared to Microscopy were 97.5% (95% CI: 95.6–98.7) and 97.7% (95% CI: 95% CI: 96.9–98.8). Compared to PCR, the diagnostic accuracy of Hz-MOD and RDT were 96.25% (95% CI: 94.1–97.7) and 96.0% (95% CI: 93.8–97.6). These results were similar to the study conducted by Kumar et al., where the diagnostic accuracy of Hz-MOD and RDT were 96.9% (95% CI: 94.1–98.7) and 93.5% (95% CI: 89.8–96.2) compared to microscopy, whereas it was 95.4% (95% CI: 92.1–97.6) and 94.3% (95% CI: 90.7–96.8) compared to PCR [8].

In our study, we assessed the sensitivity of diagnostic assays by parasite density. The Hz-MOD assay successfully detected 69 parasites/µL in clinical samples of P. vivax. Notably, these results suggest that Hz-MOD may offer superior performance, particularly in detecting low-density infections where conventional RDTs are less reliable. However, the sample size for the low parasite density group was limited (n = 9), resulting in wide confidence intervals and reduced statistical power. This limitation may affect the precision and generalizability of our sensitivity estimates, especially for low-density parasitemia cases. Therefore, larger studies are needed to assess diagnostic performance in low-transmission or pre-elimination settings, where low-density infections are more prevalent and present a significant challenge for malaria elimination efforts.

Our findings are consistent with reports from other endemic regions, such as Latin America, where the diagnostic sensitivity of both Hz-MOD and RDTs tends to be lower in low-transmission settings characterized by a higher prevalence of low-density infections. This pattern is evident in studies from the Brazilian Amazon, where the diagnostic accuracy of Hz-MOD and RDT for P. vivax malaria was high when compared to microscopy (98.2% and 92.4%, respectively), but substantially lower when compared to PCR (82.3% and 76.5%, respectively) [9]. Similar trends were observed in another study from the same region, with Hz-MOD and RDT reporting the diagnostic accuracy of Hz-MOD and RDT as 96.7% and 94.0% compared to microscopy, while it was 92.9% and 90.2% compared to PCR [19].

Furthermore, the Hz-MOD demonstrated a sensitivity of 84.6% in P. knowlesi cases (n = 26) with ≤200 parasites/µL, with the lowest parasitemia detected being 18/µL [20]. Moreover, previous studies have reported the limit of detection as low as 50 parasites/µL in the cultured P. falciparum isolates with an accuracy of 95%. In contrast, in clinical samples of patients infected with P. vivax malaria, the limit of detection was found to be 35 parasites/µL, with an accuracy of 100% [8]. Various studies that evaluated the diagnostic performance of the Hz-MOD (Gazelle) assay and RDT have reported diagnostic accuracy ranges of 95.1%−99.6% for the Hz-MOD and 92.4%−100% for RDT when compared to microscopy. In contrast, the diagnostic accuracies of Hz-MOD and RDT ranges were observed to be 85.1%–99.6% and 76.4%–100% compared to PCR (Table 3) [8,9,19–22].

Table 3.

Comparison of diagnostic performance between Hz-MOD and RDT.

Author with reference Year of the study Study population Location of the study Microscopy as a reference test
PCR as a reference test
Hz-MOD (Gazelle)
Malaria RDT
Hz-MOD (Gazelle)
Malaria RDT
Sensitivity % (95%CI) Specificity % (95%CI) Sensitivity % (95%CI) Specificity % (95%CI) Sensitivity % (95%CI) Specificity % (95%CI) Sensitivity % (95%CI) Specificity % (95%CI)
Kumar et al. [8], 2020 Malaria infections ICMR-NIRTH, Jagdalpur, Chhattisgarh, India 97.6
(87.4–99.9)
96.8
(93.6–98.7)
100.0
(91.6–100.0)
92.3
(87.9–95.4)
82.1
(69.6–91.1)
99.0
(96.5–99.9)
89.3
(78.1–96.0)
95.6
(91.9–98.0)
De Melo, G.C., et al, [9] 2021 P. vivax samples Western Brazilian Amazon 96.1
(91.3–98.3)
100
(97.4–100)
83.8
(76.3–89.7)
100
(97.51–100)
72.1
(65.0–78.3)
99.0
(94.8–99.9)
62.79
(55.11–70.03)
99.04
(94.7–99.9)
Valdivia HO et al. [19], 2021 P. vivax samples Peruvian Amazon basin 96.15
(89.1–99.2)
98.04
(93.10–99.7)
89.74
(80.7–95.4)
100
(96.45–100)
88.2
(79.4–94.2)
97.9
(92.6–99.7)
82.4
(72.6–89.8)
100
(96.2–100.0)
Fernando, D et al. [21], 2022 Malaria infections AMC Headquarters, Colombo, Sri Lanka. 77.8
(40.0–97.2)
100.0
(99.2–100.0)
100.0
(66.3–100.0)
100.0
(99.1–100.0)
77.8
(40.0–97.2)
100.0
(99.2–100.0)
100.0
(66.3–100.0)
100.00
(99.1–100.0)
Fontecha, G. et al. [22], 2022 Malaria infections Department of Gracias a Dios in the Moskitia region, Eastern Honduras 92.00
(80.7–97.7)
98.82
(95.8–99.8)
59.7
(57.9–61.6)
98.6
(98.2–99.0)
Tan, A.F. et al. [20], 2023 P. knowlesi samples Ranau District Hospital in Sabah, Malaysia
Menzies School of Health Research, Australia
94.2
(90.5–96.8)
100.0
(91.9–100.0)
92.71
(88.7–95.6)
100.00
(91.1–100.0)

The variation in diagnostic accuracy across different epidemiological contexts is likely multifactorial. One key factor is the higher proportion of asymptomatic or sub-patent infections in low-transmission settings, which often present with parasite densities below the detection thresholds of both Hz-MOD and RDTs [23,24]. Additionally, differences in host immunity may allow infections to persist at lower densities for longer periods, further complicating detection [25]. Variations in parasite biology, such as differences in hemozoin production or antigen expression, may also influence assay performance [26,27]. Finally, methodological factors, including the choice of reference standard and sample handling, can impact measured diagnostic accuracy [28]. Collectively, these considerations highlight the need for context-specific validation and optimization of malaria diagnostic tools, particularly as control programs shift focus toward elimination and the detection of low-density infections becomes increasingly critical [29].

In this study, we found that Hz-MOD (Gazelle) demonstrated promising sensitivity, particularly when compared to the RDT, excelling in the detection of symptomatic malaria infections. This suggests that the Hz-MOD assay may address some of the inherent limitations associated with RDTs. Notably, the false positivity rate of Hz-MOD was 1.39%, which is well within the current WHO selection criteria for RDT procurement (< 10%) [30]. One of the most significant advantages of Hz-MOD (Gazelle) is its remarkably quick turnaround time with a processing time of just three minutes, it far outperforms traditional microscopy (30 minutes), PCR (7–8 hours), and even RDT (15–20 min) making it a valuable diagnostic tool for fieldwork, where rapid and accurate diagnosis is essential (Table 4). Such rapid detection could potentially improve malaria management in resource-limited settings, where speed and accessibility are crucial.

Table 4.

Comparative cost analysis of Hz-MOD and Conventional malaria diagnostic assays.

Diagnostic Method Cost per Test (USD) Cost Components Time per Test Advantages Disadvantages Reference
Hz-MOD (Gazelle) $2.18–$2.36 Device amortization, materials, and minimal labor ~3 min Sensitive; Rapid; Cost effective Cannot differentiate species [31]
RDT (Rapid Diagnostic Test) $1.00–$2.19 Materials ($1.00–$1.51), labor ($0.68) ~15–20 min Rapid, easy to perform, and inexpensive Unable to quantify parasitemia; Absence of HRP 2 leads to false negative results [30,32]
Microscopy $6.98 Materials ($0.18), labor ($6.80) ~30–60 min Gold standard test and high specificity Labor-intensive and slow, and limited sensitivity at low parasitemia [32]
PCR $20–$60 Reagents, skilled labor, and equipment 7–8 hours Highest sensitivity/specificity Costly, slow, and requires lab infrastructure; not suitable for field use [33,34]

The estimated cost covers reagents and supplies for DNA extraction and slide preparation, including equipment and labor costs.

However, despite its promising performance, the Hz-MOD (Gazelle) assay has limitations. It cannot differentiate between malaria species, which is a critical shortcoming in areas where both P. vivax and P. falciparum co-exist. Consequently, any positive result obtained from Hz-MOD would require further species-specific testing via microscopy, which increases both time and cost. Additionally, hemozoin-based assays like Hz-MOD (Gazelle) are prone to false negatives in early or low-density infections due to insufficient hemozoin production and false positives from residual hemozoin from prior infections or technical artifacts [22,35,36]. These factors can reduce diagnostic accuracy. Although only minimal training is required to operate the device, the initial cost of the Hz-MOD (Gazelle) remains a significant financial consideration. Given these factors, there is a clear need for an updated version of the Hz-MOD (Gazelle) that can distinguish between different malaria species while also enhancing sensitivity and specificity. Such improvements could make Hz-MOD an even more effective diagnostic tool, potentially transforming malaria detection across diverse settings.

5. Limitations

The study was conducted in a region with a high prevalence of P. vivax malaria and relatively fewer cases of P. falciparum, and analysis would benefit from a larger sample size.

6. Conclusion

Hz-MOD (Gazelle) assay has shown better sensitivity than RDT in detecting Plasmodium infections with similar diagnostic accuracy. This suggests that the Hz-MOD assay could be an alternative for malaria screening. However, the Hz-MOD assay cannot distinguish between malaria species. Hence, to fully understand the true potential of the Hz-MOD assay, it needs an evaluation on a large number of febrile patients with a modified version of the assay which can differentiate among the species across various malaria settings.

Acknowledgements

The authors would like to acknowledge the contributions of the following individuals in the enrollment of patients: Mrs. Saroja, Laboratory Technician at Kasturba Medical College, Manipal; Mr. Ibrahim Khaleel, Laboratory Technical Officer at the Malaria Clinic, Government Wenlock Hospital, Mangalore; Mrs. Sunitha, Laboratory Technical Officer at the Malaria Clinic, Government Wenlock Hospital, Mangalore; and Mrs. Shubha Sree, Laboratory Technician at the Malaria Clinic, District Hospital, Udupi. Additionally, the authors would like to express their sincere gratitude to the study participants for their valuable contributions.

Correction Statement

This article has been corrected with minor changes. These changes do not impact the academic content of the article.

Funding Statement

This study was partially sponsored by HemexDX Private Limited, India, which provided the Hz-MOD device and test cartridges to the corresponding author. Additionally, the first author has received the Dr. T M A Pai PhD Scholarship from Manipal Academy of Higher Education, Manipal, Karnataka, India.

Authors contributions

Vishnu Teja Nallapati: Conceptualization, Methodology, Validation, Investigation, Data Curation, Formal analysis, Writing – Original Draft, Writing – Review & Editing, Visualization.

Manoj A.R.: Methodology, Investigation, Data Curation, Formal analysis, Writing – Review & Editing.

Sushma Belurkar: Methodology, Writing – Review & Editing.

Naveenchandra Kulal: Methodology, Writing – Review & Editing.

Prashanth Bhat: Methodology, Writing – Review & Editing.

Shama Prasada K: Methodology, Writing – Review & Editing.

Nitin Gupta: Methodology, Writing – Review & Editing.

Manjunatha Hande H: Methodology, Writing – Review & Editing.

Priyaleela Thota: Writing – Review & Editing.

David Bell: Writing – Review & Editing.

Kavitha Saravu: Conceptualization, Methodology, Validation, Investigation, Formal analysis, Software, Resources, Writing – Review & Editing, Visualization, Supervision, Project administration, Funding acquisition. All authors have read and approved the final version of the manuscript.

Disclosure statement

No potential conflict of interest was reported by the author(s).

Declaration of interest

The first and corresponding authors have received the Hz-MOD device and test cartridges from HemexDX Private Limited, India. Priya Thota and David Bell are affiliated with Hemex Health, U.S.A. The sponsor had no role in data collection, analysis, or interpretation. The remaining authors declare no competing interests.

References

  • [1].World . 2024. malaria report [Internet]. [cited 2025 Jan 22]. Available from: https://www.who.int/teams/global-malaria-programme/reports/world-malaria-report-2024
  • [2].Global technical strategy for malaria 2016–2030, 2021 update [Internet]. [cited 2023 Dec 28]. Available from: https://www.who.int/publications-detail-redirect/9789240031357
  • [3].Krampa FD, Aniweh Y, Kanyong P, et al. Recent advances in the development of biosensors for malaria diagnosis. Sensors (Basel). 2020. Feb 1;20(3):799. doi: 10.3390/s20030799 PMID: 32024098; PMCID: PMC7038750. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [4].Berzosa P, de Lucio A, Romay-Barja M, et al. Comparison of three diagnostic methods (microscopy, RDT, and PCR) for the detection of malaria parasites in representative samples from Equatorial Guinea. Malar J. 2018. Sep 17;17(1):333. doi: 10.1186/s12936-018-2481-4 PMID: 30223852; PMCID: PMC6142353. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [5].Nderu D, Kimani F, Thiong’o K, et al. Plasmodium falciparum histidine-rich protein (PfHRP2 and 3) diversity in Western and Coastal Kenya. Sci Rep. 2019;9(1):1709. doi: 10.1038/s41598-018-38175-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [6].Oth JM, Korevaar DA, Leeflang MM, et al. Molecular malaria diagnostics: a systematic review and meta-analysis. Crit Rev Clin Lab Sci. 2016;53(2):87–105. doi: 10.3109/10408363.2015.1084991 Epub 2015 Sep 17. PMID: 26376713. [DOI] [PubMed] [Google Scholar]
  • [7].Gazelle Diagnostic Device [Internet]. Engineering For Change . [cited 2025 May 20]. Available from: https://www.engineeringforchange.org/solutions/product/gazelle-diagnostic-device/
  • [8].Kumar R, Verma AK, Shrivas S, et al. First successful field evaluation of new, one-minute haemozoin-based malaria diagnostic device. EClinicalMedicine. 2020. May 22;22:100347. doi: 10.1016/j.eclinm.2020.100347 PMID: 32490369; PMCID: PMC7256309. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [9].de Melo GC, Netto RLA, Mwangi VI, et al. Performance of a sensitive haemozoin-based malaria diagnostic test validated for vivax malaria diagnosis in Brazilian Amazon. Malar J. 2021. Mar 12;20(1):146. doi: 10.1186/s12936-021-03688-0 PMID: 33712019; PMCID: PMC7953757. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [10].Malaria-AnnualReport-2023.pdf [Internet]. [cited 2025 Feb 5]. Available from: https://ncvbdc.mohfw.gov.in/Doc/Malaria-AnnualReport-2023.pdf
  • [11].Organization WH. Diseases UBSP for R and T in T. Microscopy for the detection, identification and quantification of malaria parasites on stained thick and thin blood films in research settings (version 1.0): procedure: methods manual [Internet]. World Health Organization. 2015. [cited 2025 May 20]. Available from: https://iris.who.int/handle/10665/163782
  • [12].Snounou G, Viriyakosol S, Zhu XP, et al. High sensitivity of detection of human malaria parasites by the use of nested polymerase chain reaction. Mol Biochem Parasitol. 1993. Oct;61(2):315–320. doi: 10.1016/0166-6851(93)90077-b PMID: 8264734. [DOI] [PubMed] [Google Scholar]
  • [13].Florkowski CM. Sensitivity, specificity, receiver-operating characteristic (ROC) curves and likelihood ratios: communicating the performance of diagnostic tests. Clin Biochem Rev. 2008. Aug;29(Suppl 1):S83–7. PMID: 18852864; PMCID: PMC2556590. [PMC free article] [PubMed] [Google Scholar]
  • [14].Glas AS, Lijmer JG, Prins MH, et al. The diagnostic odds ratio: a single indicator of test performance. J Clin Epidemiol. 2003. Nov;56(11):1129–1135. doi: 10.1016/s0895-4356(03)00177-x PMID: 14615004. [DOI] [PubMed] [Google Scholar]
  • [15].Cappellini MD, Motta I. Anemia in clinical practice-definition and classification: does hemoglobin change with aging? Semin Hematol. 2015. Oct;52(4):261–269. doi: 10.1053/j.seminhematol.2015.07.006 Epub 2015 Jul 17. PMID: 26404438. [DOI] [PubMed] [Google Scholar]
  • [16].MedCalc SF. MedCalc’s diagnostic test evaluation calculator. 2025. Jan 22. Available from: https://www.medcalc.org/calc/diagnostic_test.php
  • [17].Mbanefo A, Kumar N. Evaluation of malaria diagnostic methods as a key for successful control and elimination programs. Trop Med Infect Dis. 2020. Jun 19;5(2):102. doi: 10.3390/tropicalmed5020102 PMID: 32575405; PMCID: PMC7344938. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [18].Zainabadi K, Kraft CS. Ultrasensitive diagnostics for low-density asymptomatic Plasmodium falciparum infections in low-transmission settings. J Clin Microbiol. 2021. Mar 19;59(4):e01508–20. doi: 10.1128/JCM.01508-20 PMID: 33148707; PMCID: PMC8092727. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [19].Valdivia HO, Thota P, Braga G, et al. Field validation of a magneto-optical detection device (Gazelle) for portable point-of-care Plasmodium vivax diagnosis. PLOS ONE. 2021. Jun 22;16(6):e0253232. doi: 10.1371/journal.pone.0253232 PMID: 34157032; PMCID: PMC8219132. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [20].Tan AF, Thota P, Sakam SSB, et al. Evaluation of a point-of-care haemozoin assay (Gazelle device) for rapid detection of Plasmodium knowlesi malaria. Sci Rep. 2023. Mar 23;13(1):4760. doi: 10.1038/s41598-023-31839-7 PMID: 36959462; PMCID: PMC10036474. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [21].Fernando D, Thota P, Semege S, et al. Evaluation of a haemozoin-based rapid diagnostic test for diagnosis of imported malaria during the phase of prevention of reestablishment in Sri Lanka. Malar J. 2022. Sep 10;21(1):263. doi: 10.1186/s12936-022-04283-7 PMID: 36088431; PMCID: PMC9464370. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [22].Fontecha G, Escobar D, Ortiz B, et al. Field evaluation of a hemozoin-based malaria diagnostic device in Puerto Lempira, Honduras. Diagnostics (Basel). 2022. May 11;12(5):1206. doi: 10.3390/diagnostics12051206 PMID: 35626361; PMCID: PMC9140950. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [23].Okell LC, Ghani AC, Lyons E, et al. Submicroscopic infection in Plasmodium falciparum-endemic populations: a systematic review and meta-analysis. J Infect Dis. 2009. Nov 15;200(10):1509–1517. doi: 10.1086/644781 PMID: 19848588. [DOI] [PubMed] [Google Scholar]
  • [24].Bousema T, Okell L, Felger I, et al. Asymptomatic malaria infections: detectability, transmissibility and public health relevance. Nat Rev Microbiol. 2014. Dec;12(12):833–840. doi: 10.1038/nrmicro3364 Epub 2014 Oct 20. PMID: 25329408. [DOI] [PubMed] [Google Scholar]
  • [25].Chen K, Perlaki C, Xiong A, et al. Review of surface enhanced Raman spectroscopy for malaria diagnosis and a new approach for the detection of single parasites in the ring stage. IEEE J Select Top Quantum Electron. 2016. Jan 18;22(4):179–187. doi: 10.1109/JSTQE.2016.2518959 [DOI] [Google Scholar]
  • [26].Hanssen E, Knoechel C, Dearnley M, et al. Soft X-ray microscopy analysis of cell volume and hemoglobin content in erythrocytes infected with asexual and sexual stages of Plasmodium falciparum. J Struct Biol. 2012. Feb;177(2):224–232. doi: 10.1016/j.jsb.2011.09.003 Epub 2011 Sep 16. PMID: 21945653; PMCID: PMC3349340. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [27].Wongsrichanalai C, Barcus MJ, Muth S, et al. A review of malaria diagnostic tools: microscopy and rapid diagnostic test (RDT). Am J Trop Med Hyg. 2007. Dec;77(6 Suppl):119–127. PMID: 18165483. doi: 10.4269/ajtmh.2007.77.119 [DOI] [PubMed] [Google Scholar]
  • [28].Bossuyt PM, Reitsma JB, Bruns DE, et al. Stard 2015: an updated list of essential items for reporting diagnostic accuracy studies. BMJ. 2015. Oct 28;351:h5527. doi: 10.1136/bmj.h5527 PMID: 26511519; PMCID: PMC4623764. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [29].World Health Organization . Malaria microscopy quality assurance manual [Internet]. version 2. Geneva: World Health Organization; 2016. [cited 2025 May 20]. p. 121 p. Available from: https://iris.who.int/handle/10665/204266 [Google Scholar]
  • [30].Malaria rapid diagnostic test performance. Results of WHO product testing of malaria RDTs: round. 2016–2018) [Internet]. [cited 2023 Dec 29];8. Available from: https://www.who.int/publications-detail-redirect/9789241514965 [Google Scholar]
  • [31].Rathi A, Chowdhry Z, Patel A, et al. Hemozoin in malaria eradication—from material science, technology to field test. NPG Asia Mater. 2023;15(70). doi: 10.1038/s41427-023-00516-6 [DOI] [Google Scholar]
  • [32].Du YQ, Ling JJ XX, Zhou HY, et al. Cost-effectiveness analysis of malaria rapid diagnostic test in the elimination setting. Infect Dis Poverty. 2020. Sep 29;9(1):135. doi: 10.1186/s40249-020-00745-9 PMID: 32993762; PMCID: PMC7523355. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [33].Panwar V, Bansal S, Chauhan C, et al. Cost analyses for malaria molecular diagnosis for research planners in India and beyond. Expert Rev Mol Diagn. 2024. Jun;24(6):549–559. doi: 10.1080/14737159.2024.2356172 Epub 2024 May 20. PMID: 38768107. [DOI] [PubMed] [Google Scholar]
  • [34].Mahjabeen SS, Mousa R, Salah R, et al. Comparative analysis of the diagnostic accuracy of malaria parasite by microscopy, RDTs (PfHRP-2 & PLDH) and PCR. Indian J Pathol Oncol. 2022;9(2):112–115. doi: 10.18231/j.ijpo.2022.028 [DOI] [Google Scholar]
  • [35].Grobusch MP, Hänscheid T, Krämer B, et al. Sensitivity of hemozoin detection by automated flow cytometry in non- and semi-immune malaria patients. Cytometry B Clin Cytom. 2003. Sep;55(1):46–51. doi: 10.1002/cyto.b.10039 PMID: 12949959. [DOI] [PubMed] [Google Scholar]
  • [36].Delahunt C, Horning MP, Wilson BK, et al. Limitations of haemozoin-based diagnosis of Plasmodium falciparum using dark-field microscopy. Malar J. 2014;13(1):147. doi: 10.1186/1475-2875-13-147 [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Pathogens and Global Health are provided here courtesy of Taylor & Francis

RESOURCES