Low-tube-potential ultra-high-resolution coronary CTA with photon-counting detector CT for stent evaluation: a comparative feasibility study

Suguru Araki 1, Satoshi Nakamura 2,✉, Shiko Okabe 1, Masafumi Takafuji 1, Yasutaka Ichikawa 1, Hajime Sakuma 1, Kakuya Kitagawa 2,3
PMCID: PMC12647213  PMID: 40782242

Abstract

Objectives

This study aimed to evaluate the feasibility of low-tube-potential ultra-high-resolution (UHR) coronary CT angiography (CCTA) using photon-counting detector CT (PCD-CT) in comparison to low-tube-potential CCTA using energy-integrating detector CT (EID-CT), with a specific focus on the evaluation of coronary stents.

Materials and methods

This retrospective study included 54 patients (88 stents) who underwent CCTA in UHR mode on PCD-CT at low tube potential (PCDUHR-low; 27 patients [45 stents]) or CCTA with EID-CT at low tube potential (EID-low; 27 patients [43 stents]). The EID-low cohort was selected by propensity score matching to the PCDUHR-low cohort. Image quality of in-stent lumen was assessed using a 4-point Likert scale, with a score of 4 indicating “excellent.” Quantitative assessment of stents included stent-induced blooming, edge sharpness, stent full width at half maximum (FWHM) and ΔHUin-stent which quantifies the increase in CT attenuation within the stent lumen. Radiation dose was evaluated using CT dose index volume (CTDIvol) and dose–length product (DLP).

Results

PCDUHR-low had higher image quality scores than EID-low (scores of 3 and 4, 82.2% vs. 53.5%; p = 0.02). PCDUHR-low showed reduced stent-induced blooming, superior edge sharpness, smaller stent FWHM and smaller ΔHUin-stent than EID-low (all, p < 0.01). CTDIvol and DLP in PCDUHR-low (9.3 ± 4.4 mGy and 107.1 ± 50.4 mGy・cm) were comparable to those of EID-low (10.5 ± 5.9 mGy and 118.2 ± 74.9 mGy・cm), respectively. (p = 0.26 and 0.47).

Conclusion

CCTA using UHR mode in PCD-CT at low tube potential achieved superior image quality for implanted stent visualization as compared to EID-CT at low tube potential, while maintaining comparable radiation doses.

Supplementary Information

The online version contains supplementary material available at 10.1007/s11604-025-01846-x.

Keywords: Coronary artery disease, Computed tomography, Stents, Radiation dosage

Introduction

Cardiac computed tomography (CT) is widely used as a noninvasive imaging modality in clinical practice [1, 2]. Coronary CT angiography (CCTA) can rule out coronary artery disease (CAD) with a high negative predictive value [3], and recent guidelines recommend CCTA as the modality of choice for evaluating coronary arteries in patients with stable chest pain and low to intermediate risk of coronary artery disease [4]. Despite its widespread utilization, the evaluation of coronary stenosis in the presence of coronary stents has been limited when using conventional energy-integrating detector CT (EID-CT) scanners. This limitation is mainly due to stent-induced blooming and beam-hardening artifacts, which compromise the diagnostic value of CCTA [5–8]. Stent-induced blooming has been primarily attributed to partial volume averaging, which is influenced by spatial resolution and detector cell size [9, 10], leading to inaccurate evaluation of stenosis. The structural limitations of EID-CT detector elements restrict spatial resolution and accurate delineation of stented segments, indicating the need for novel detector technologies.

Recently, photon-counting detector CT (PCD-CT) scanners equipped with dual-source CT technology have been introduced into clinical practice. A key advantage of PCD-CT is its improved photon efficacy, largely attributable to its spectral capabilities and the ability to exclude electronic noise. By applying appropriate energy thresholds, PCD-CT can effectively remove electronic noise from the signal, thereby improving the signal-to-noise ratio (SNR) and potentially reducing radiation dose [11, 12]. In addition, PCD-CT provides higher spatial resolution than EID-CT, particularly in its ultra-high-resolution (UHR) mode. The absence of septa between detector pixels in PCD-CT enables a slice thickness of 0.2 mm in UHR mode, reducing partial volume effects and improving stent-induced blooming artifacts [13–15]. Prior investigations have optimized UHR imaging parameters such as the choice of reconstruction kernels and the use of quantum iterative reconstruction (QIR) algorithms [16, 17].

However, while the previous studies have addressed several aspects of UHR imaging, the feasibility and utility of low-tube-potential protocols in UHR mode remain insufficiently explored. Tube potential affects contrast-to-noise ratio (CNR), the attenuation profiles of plaques and stents, and overall image quality. On the radiation dose side, the narrow X-ray beam width required by UHR mode [18] may lead to higher radiation exposure as compared to normal resolution mode. Considering these, it is desirable to achieve UHR-mode imaging at as low a dose as possible without compromising image quality.

Therefore, the aim of this study was to evaluate the image quality and radiation dose of low tube potential CCTA using PCD-CT UHR mode (PCDUHR-low) in the presence of coronary stents, and to compare these results with those obtained using EID-CT at low tube potential (EID-low), with the EID-low cohort selected by propensity score matching to the PCDUHR-low cohort. Through this comparative assessment, we seek to clarify the potential benefits of low tube potential and thereby guide more effective clinical use of UHR-mode PCD-CT for stent assessment.

Materials and methods

Study design

This retrospective, single-center study focused on the assessment of UHR-CCTA with a low tube potential of 70 or 90 kVp using a PCD-CT scanner. The institutional review board of our hospital approved the study protocol and waived the need for individual consent due to its retrospective design.

Between October 2023 and September 2024, 92 patients underwent CCTA for coronary artery stent evaluation using a PCD-CT scanner (NAEOTOM Alpha, Siemens Healthineers, Forchheim, Germany). Thirteen patients with a history of coronary artery bypass grafting (CABG) were excluded. Of the remaining 79 patients, 43 were further excluded because their scans were performed with normal resolution mode or with ECG-gated retrospective helical scans in UHR mode, which were used for patients with nonsinus rhythm or preoperative evaluation for valvular disease. Among these, 27 underwent CCTA scans performed with ECG-triggered prospective axial acquisition in UHR mode at low tube potentials (70 or 90 kVp). These 27 patients were classified as the PCDUHR-low group. Patients were not excluded based on heart rate variability, body weight, or body mass index (BMI). Among these 27 patients, 16 underwent scans at 90 kVp, whereas 11 underwent scans at 70 kVp.

For comparison, 68 patients underwent CCTA at a low tube potential of 70 or 80 kVp using a third-generation dual-source EID-CT scanner (Somatom Force, Siemens Healthineers) at our hospital between October 2023 and May 2024. Two patients with a history of CABG and two patients with retrospective helical scans were excluded, leaving 64 eligible patients. Among these, 27 patients (EID-low) who matched the 27 patients included in the PCDUHR-low group were selected using propensity score matching.

The final study population consisted of 27 patients in the PCDUHR-low group, and 27 propensity score-matched patients in the EID-low group. The patient selection process is illustrated in Fig. 1.

Fig. 1.

Fig. 1

Study flow chart. CCTA, coronary computed tomography angiography; PCD, photon-counting detector; EID, energy-integrating detector; CABG, coronary artery bypass graft; UHR, ultra-high-resolution; PCDUHR-low, CCTA scans with ECG-triggered prospective axial acquisition, UHR scan mode, and a low tube potential of 70 or 90 kVp using a PCD-CT scanner; EID-low, CCTA scans with ECG-triggered prospective axial acquisition, and a low tube potential of 70 or 90 kVp using EID-CT

In addition, 9 patients who underwent scans with UHR mode at standard tube potentials of 120 or 140 kVp between October 2023 and May 2024 were included in the PCDUHR-std group. Among these 9 patients, 2 underwent scans at 140 kVp, whereas 7 underwent scans at 120 kVp. These patients were not included in the propensity score matching but were analyzed separately to evaluate the effects of standard tube potential settings.

Image acquisition

All scans in PCDUHR-low and PCDUHR-std were performed using a first-generation dual-source PCD-CT scanner (NAEOTOM Alpha, Siemens Healthineers, Forchheim, Germany), while scans in the EID-low were performed using a third-generation dual-source EID-CT scanner (Somatom Force, Siemens Healthineers). In PCD-CT, CCTA with ECG-triggered prospective axial UHR scan mode was performed with a phase of 65–85% R-R interval by bolus injection of 26 mgI/kg/s of iopamidol (Iopamilon-370, Bayer AG, Leverkusen, Germany) over 12 s, followed by a 20 mL saline flush, with the coronary arteries dilated with a sublingual nitrate (Myocor sprays, Toa Eiyo Co., Ltd., Tokyo, Japan). The acquisition parameters of CCTA in PCD-CT were as follows: collimation of 144 mm × 0.2 mm, gantry rotation time of 0.25 s, and tube voltage of 70 or 90 kVp in PCDUHR-low and 120 or 140 kVp in PCDUHR-std. Automatic exposure control (CAREkeV, Siemens Healthineers, Forchheim, Germany) was used to determine tube current for each patient. Axial images of CCTA were reconstructed with a slice thickness and increment of 0.2/0.2 mm, quantum iterative reconstruction level of 4, vascular convolution kernel (Bv64), image matrix of 1024 × 1024, and a field-of-view restricted to the heart. The in-plane spatial resolution of PCD-CT in UHR mode was approximately 0.11 mm. Since the PCD-CT system did not allow for the collection of spectral information with UHR mode, polychromatic images called T3D were reconstructed for CCTA with UHR mode.

In EID-CT, ECG-triggered prospective axial CCTA was performed. The contrast medium was injected the same as the protocol of PCD-CT. The acquisition parameters of CCTA in EID-CT were as follows: collimation of 192 mm × 0.6 mm, gantry rotation time of 0.25 s, and tube voltage of 70 or 80 kVp. Automatic exposure control (CAREkVp, Siemens Healthineers, Forchheim, Germany) was used to determine tube current for each patient. Axial images of CCTA were reconstructed with a slice thickness and increment of 0.6/0.6 mm, vascular convolution kernel (Bv49), image matrix of 512 × 512, and a field-of-view restricted to the heart. The in-plane spatial resolution of EID-CT under these reconstruction conditions was approximately 0.24 mm.

Heart rate was controlled before CCTA with intravenous injection of landiolol hydrochloride (Corebeta, Ono Pharmaceutical Co., Ltd., Osaka, Japan), if necessary.

CCTA image quality analysis

Although the details of the CCTA image quality assessment are described in the Supplementary Method, briefly, two radiologists, with 4 (S.A.) and 10 (S.N.) years of experience in cardiovascular imaging, performed a consensus visual evaluation of CCTA image quality on a four-point scale (1, nondiagnostic; 2, fair; 3, good; 4, excellent) (Fig. 2), taking into account motion artifacts, sharpness, noise, contrast enhancement, and beam hardening. Objective image quality parameters (signal intensity, image noise, contrast to noise ratio [CNR], and signal to noise ratio [SNR]) were calculated using CT attenuation values from coronary artery lumens and the left ventricular lateral wall.

Fig. 2.

Fig. 2

Illustration of image quality scores of CCTA images. The figures show a four-point scale used to assess the image quality of the CCTA images: 1, nondiagnostic; 2, fair; 3, good; and 4, excellent

Implanted stent image quality analysis

Two radiologists, with 4 (S.A.) and 10 (S.N.) years of experience in cardiovascular imaging, performed a consensus visual assessment of in-stent lumen image quality for each stent using a 4-point Likert scale (1 = nondiagnostic, 2 = fair, 3 = good, 4 = excellent). The evaluation included four specific image quality attributes critical for assessing in-stent morphology and plaque characteristics: motion artifacts, vessel contrast enhancement, sharpness of the stent, and beam hardening artifacts. Each attribute was assessed separately to allow for a more objective comparison between imaging techniques. In addition to these individual parameters, overall image quality was also graded to reflect the general diagnostic acceptability of the in-stent lumen. A score of 1 (“nondiagnostic”) was assigned when severe artifacts prevented the assessment of stent patency. A score of 2 (“fair”) indicated that although artifacts were present, interpretation remained feasible. A score of 3 (“good”) represented minimal interference from artifacts and clear visualization of the stent lumen. A score of 4 (“excellent”) was given when the lumen appeared sharply defined and free of any limiting artifacts (Fig. 3) [19].

Fig. 3.

Fig. 3

Illustration of image quality scores of in-stent lumens on CCTA images. The figures show a four-point scale used to assess the image quality of the CCTA images: 1, nondiagnostic; 2, fair; 3, good; and 4, excellent

To assess the stent-induced blooming, internal and external stent diameters on multiplanar reformatted CT images, focusing on the slice that best represents the stent geometry for each case. When two stents were connected, each stent was measured and analyzed independently. There was no limitation on the maximum number of stents that could be assessed per vessel. The estimated stent-induced blooming was calculated using the following formula [19, 20].

Stent - induced blooming%=stent diameterexternalmm-stent diameterinternalmm/stent diameterexternalmm×100

The effect on in-stent lumen attenuation was also evaluated. Three manually drawn regions of interest (ROIs) were utilized: The first, designated as ROIin-stent, was drawn as large as possible to encompass the in-stent lumen while avoiding the vessel walls and stent struts, with a minimum area of 4 mm2. Two additional ROIs were placed in the coronary artery lumen—one proximal to the stent (ROIprox) and the other distal to the stent (ROIdist), both within a maximum distance of 10 mm from the stent edges. From these ROIs, the average Hounsfield unit (HU) values and standard deviation (SD) were measured. Stent struts, vessel walls, and calcified or noncalcified plaques were carefully avoided from the ROIs. With reference to the methods described by Hagar MT, et al. and Boccalini S, et al. [19, 20], ΔHUin-stent were calculated to evaluate the effects of the stent on in-stent lumen attenuation using the three ROIs. ΔHUin-stent quantifies the increase in CT attenuation within the stent lumen. This parameter was derived using the following formulas:

ROIex - stent=ROIproxHU+ROIdistHU/2
ΔHUin - stent=ROIin - stentHU-ROIex - stentHU∗100/ROIex - stentHU%

This metric quantitatively captures the stent’s impact on lumen attenuation. Edge sharpness and full width at half maximum (FWHM) of coronary stent were obtained from the vessel profile using the definitions shown in Fig. 4 by a radiologist (S.A.) setting a linear region of interest on a cross-sectional image using the commercially available software (Ziostation2, Ziosoft Inc., Tokyo, Japan) [21].

Fig. 4.

Fig. 4

Profile curve shows definition of edge sharpness and full width at half maximum (FWHM). Edge sharpness of stent strut is defined as the maximum slope of the profile curve. FWHMs of lumen (FWHM-lumen) and stent strut (FWHM-stent) were measured

Radiation dose estimation

The CT dose index volume (CTDIvol) and dose-length product (DLP) were extracted from the participant protocols.

Statistical analysis

Descriptive statistics (mean, standard deviation) were used to summarize the data. Student’s t test was used to assess differences in continuous variables between two groups after examining the normality of the data. The chi-squared test was used to assess differences in proportion between categorical variables. Values of p < 0.05 were considered indicative of statistical significance.

Propensity-score matching was performed to select control patients who underwent CCTA with EID-CT and a propensity score was calculated based on age, sex, body height, body weight, and heart rate. Propensity-score matching was performed using the SPSS statistical package (version 23.0, IBM Corp., Armonk, NY, USA). All the other statistical analyses were performed using JMP version 10 software (SAS Institute Inc., Cary, NC, USA).

Results

Patient characteristics and radiation dose metrics

Patient characteristics and CCTA imaging parameters are summarized in Table 1. Of the 54 patients included, 38 (70.4%) were male, and the mean age was 70 ± 10 years. The mean body weight was 63.4 ± 14.1 kg, and the mean BMI was 23.8 ± 3.9 kg/m2. The mean heart rate during CCTA was 64 ± 8 bpm. There were no significant differences between the PCDUHR-low and EID-low groups in terms of patient characteristics, including sex, age, height, weight, and BMI. Regarding radiation exposure, the mean CTDIvol was 9.3 ± 4.4 mGy in the PCDUHR-low group and 10.5 ± 5.9 mGy in the EID-low group (p = 0.26). Similarly, the mean DLP was 107.1 ± 50.4 mGy·cm and 118.2 ± 74.9 mGy·cm, respectively (p = 0.47).

Table 1.

Patient characteristics, CCTA parameters and image quality of CCTA images

All participants
(n = 54)
PCDUHR-low
(n = 27)
EID-low
(n = 27)
p value
Patient characteristics
Male 38 (70.4) 19 (70.4) 19 (70.4) N.A
Age [yrs] 70 ± 10 71 ± 2 69 ± 2 0.51
Body height [cm] 162.5 ± 9.8 162.1 ± 8.8 162.7 ± 10.9 0.82
Body weight [kg] 63.4 ± 14.1 62.0 ± 2.8 64.8 ± 2.8 0.48
BMI [kg/m2] 23.8 ± 3.9 23.4 ± 0.8 24.2 ± 0.8 0.47
CCTA parameters
Heart rate [bpm] 64 ± 8 64 ± 2 64 ± 2 0.86
Amount of contrast [ml] 52.5 ± 8.1 54.3 ± 6.8 50.7 ± 9.0 0.11
Flow rate [ml/s] 4.3 ± 0.7 4.4 ± 0.5 4.3 ± 0.8 0.38
CTDIvol [mGy] 9.9 ± 5.2 9.3 ± 4.4 10.5 ± 5.9 0.26
DLP [mGy・cm] 112.7 ± 63.5 107.1 ± 50.4 118.2 ± 74.9 0.47

Data are presented as the mean ± standard deviation or number of patients (%). p value indicates comparison of parameters between the groups. BMI body mass index, CCTA coronary computed tomography angiography, CTDIvol computed tomography dose index volume, DLP dose-length product

Patient characteristics and CCTA imaging parameters of the PCDUHR-std group are summarized in Supplementary Table 1 and 2, accompanied by those of the PCDUHR-low group for reference. Regarding radiation dose estimates, the mean CTDIvol and DLP in the PCDUHR-std group were 22.5 ± 7.7 mGy and 256.8 ± 83.2 mGy∙cm, respectively, which were significantly higher than those in the PCDUHR-low group (p < 0.01, both).

Table 2.

Image quality of CCTA images

All participants
(n = 54)
PCDUHR-low (n = 27) EID-low
(n = 27)
p value
Subjective image quality  < 0.01
1. Nondiagnostic 7 (1.0) 4 (1.1) 3 (0.8)
2. Fair 70 (9.6) 10 (2.8) 60 (16.3)
3. Good 388 (53.3) 140 (38.8) 248 (67.6)
4. Excellent 263 (36.1) 207 (57.3) 56 (15.3)
Objective Image quality
Signal intensity [HU] 794.8 ± 177.1 782.9 ± 212.9 806.7 ± 135.3 0.11
Image noise [HU] 97.6 ± 17.2 106.7 ± 15.3 88.4 ± 14.0  < 0.01
Contrast-to-noise ratio 8.4 ± 2.2 6.2 ± 2.0 7.3 ± 1.4 0.01
Signal-to-noise ratio 6.7 ± 1.8 7.5 ± 2.2 9.3 ± 1.8  < 0.01

Data are presented as the mean ± standard deviation or number of patients (%). p value indicates comparison of parameters between the groups. HU Hounsfield unit

CCTA image quality

Table 2 summarizes the subjective and objective image quality of CCTA. Regarding subjective image quality, among all participants, 36.1% of segments were rated as excellent, 53.3% as good, 9.6% as fair, and 1.0% as nondiagnostic. The PCDUHR-low group showed a higher proportion of excellent scores (57.3% vs. 15.3%) and a lower proportion of fair or nondiagnostic scores (3.9% vs. 17.1%) as compared to the EID-low group (p < 0.01). Objectively, the PCDUHR-low group showed higher image noise (106.7 ± 15.3 HU) than EID-low group, while its signal intensity (782.9 ± 212.9 HU) was comparable to EID-low (p = 0.11). The PCDUHR-low group had lower CNR and SNR than the EID-low group (both p < 0.01). Subjective and objective image quality of CCTA in PCDUHR-std are summarized in Supplementary Table 3. Subjective image quality in the PCDUHR-std group was comparable to that in the PCDUHR-low group, with approximately 95% of segments rated as good or excellent in both groups. SNR and CNR were also comparable between the two groups.

Image quality of implanted stents

Table 3 summarizes the location and caliber of all the implanted stents among the three subgroups. There were no significant differences in stent location (p = 0.59) or stent caliber (p = 0.85) among the groups. Characteristics of implanted stents in PCDUHR-std are summarized in Supplementary Table 4.

Table 3.

Implanted stent characteristics

All stents
(n = 88)
PCDUHR-low
(n = 45)
EID-low
(n = 43)
p value
Stent location 0.59
LMT 4 (4.5) 2 (4.4) 2 (4.7)
LAD 41 (46.6) 21 (46.7) 20 (46.5)
LCX 21 (23.9) 13 (28.9) 8 (18.6)
RCA 22 (25.0) 9 (20.0) 13 (30.2)
Stent caliber 0.85
 ≥ 3.5 25 (28.4) 11 (24.5) 14 (32.6)
3.5 >, ≥ 3.0 27 (30.7) 15 (33.3) 12 (27.9)
3.0 >, ≥ 2.5 26 (29.5) 14 (31.1) 12 (27.9)
 < 2.5 10 (11.4) 5 (11.1) 5 (11.6)

Data are presented as the number of patients (%). p value indicates comparison of parameters between the groups. LMT left main coronary artery, LAD left anterior descending artery, LCX left circumflex artery, RCA right coronary artery

Table 4 summarizes the subjective image quality of implanted stents. No significant differences were observed between the groups in terms of motion artifacts and vessel contrast enhancement (p = 0.28 and 0.17, respectively). For motion, only one segment was rated as nondiagnostic in the PCDUHR-low group. For contrast, more than 95% of segments were rated as good or excellent in both groups. In contrast, differences were observed in sharpness and beam hardening artifacts. In the PCDUHR-low group, 100% of segments were rated as excellent or good for both parameters, whereas in the EID-low group, 60.4% were rated as excellent or good for sharpness, and 90.7% for beam hardening. Subjective image quality of implanted stents in PCDUHR-std are summarized in Supplementary Table 5. Subjective image quality of implanted stents was comparable between the PCDUHR-low and PCDUHR-std groups across all evaluated parameters, including motion, contrast, sharpness, beam hardening, and overall quality.

Table 4.

Subjective image quality of implanted stents

All participants
(n = 88)
PCDUHR-low (n = 45) EID-low
(n = 43)
p value
Subjective image quality
motion 0.28
1. Nondiagnostic 1 (1.1) 1 (2.2) 0 (0)
2. Fair 8 (9.1) 6 (13.3) 2 (4.7)
3. Good 5 (5.7) 3 (6.7) 2 (4.7)
4. Excellent 74 (84.1) 35 (77.8) 39 (90.6)
Contrast 0.17
1. Nondiagnostic 0 (0) 0 (0) 0 (0)
2. Fair 2 (2.3) 0 (0) 2 (4.7)
3. Good 12 (13.6) 5 (11.1) 7 (16.3)
4. Excellent 74 (84.1) 40 (88.9) 34 (79.0)
Sharpness  < 0.01
1. Non-diagnostic 3 (3.4) 0 (0) 3 (7.0)
2. Fair 14 (15.9) 0 (0) 14 (32.6)
3. Good 17 (19.3) 0 (0) 17 (39.5)
4. Excellent 54 (61.4) 45 (100) 9 (20.9)
Beam hardening  < 0.01
1. Nondiagnostic 0 (0) 0 (0) 0 (0)
2. Fair 4 (4.6) 0 (0) 4 (9.3)
3. Good 42 (47.7) 18 (40.0) 24 (55.8)
4. Excellent 42 (47.7) 27 (60.0) 15 (34.9)

Data are presented as the mean ± standard deviation or number of patients (%). p value indicates comparison of parameters between the groups

Figure 5 illustrates the distribution of subjective overall image quality scores of implanted stents. For all stents (n = 88), the proportions of segments rated as “excellent” (score 4), “good” (score 3), “fair” (score 2), and “nondiagnostic” (score 1) were 35.2% (n = 31), 33.0% (n = 29), 25.0% (n = 22), and 6.8% (n = 6), respectively. When stratified by group, in the PCDUHR-low group (n = 45), the proportions were 48.9% (score 4, n = 22), 33.3% (score 3, n = 15), 15.6% (score 2, n = 7), and 2.2% (score 1, n = 1). In the EID-low group (n = 43), the proportions were 20.9% (score 4, n = 9), 32.6% (score 3, n = 14), 34.9% (score 2, n = 15), and 11.6% (score 1, n = 5). The PCDUHR-low group demonstrated a significantly better image quality distribution compared to the EID-low group (p = 0.02). When stents were stratified by diameter, the differences in image quality distribution became more apparent. For stents with a diameter of ≥ 3 mm (n = 52), no significant differences were observed between the PCD UHR-low and EID-low groups (p = 0.07). However, for stents with a diameter of < 3 mm (n = 36), the PCDUHR-low group demonstrated a significantly better image quality distribution compared to the EID-low group (p = 0.01).

Fig. 5.

Fig. 5

Image quality score of stents. The figure shows distribution of subjective image quality scores based on the Likert scale for (A) all stents (n = 110), B stents with a diameter of ≥ 3 mm (n = 63), and C stents with a diameter of < 3 mm (n = 47). The image quality scores are categorized as 1: non-diagnostic (gray), 2: fair (orange), 3: good (light blue), and 4: excellent (dark blue). Comparisons among groups are shown for PCDUHR-low (a), and EID-low (b). There were significant differences for all stents (p = 0.02 for PCDUHR-low vs. EID-low) and for stents < 3 mm (p = 0.01 for PCDUHR-low vs. EID-low), whereas no significant differences were found between groups for stents ≥ 3 mm. PCDUHR-low, CCTA scans with ECG-triggered prospective axial acquisition, UHR scan mode, and a low tube potential of 70 or 90 kVp using a PCD-CT scanner; EID-low, CCTA scans with ECG-triggered prospective axial acquisition, and a low tube potential of 70 or 90 kVp using EID-CT

The objective image quality metrics specific to implanted stents are detailed in Table 5, Supplementary Table 6 and Supplementary Table 7. For all stents, the mean stent-induced blooming was 37.4 ± 14.6%. ΔHUin-stent was 22.1 ± 22.9%. The PCDUHR-low group demonstrated significantly lower stent-induced blooming (28.1 ± 9.8%) as compared to the EID-low group (47.1 ± 12.2%, p < 0.01), as well as a lower ΔHUin-stent (15.1 ± 13.4%) as compared to the EID-low group (29.4 ± 28.1%, p < 0.01). The objective image quality metrics in PCDUHR-std are summarized in Supplementary Table 8. In the PCDUHR-std group, ΔHUin-stent was significantly lower than in the PCDUHR-low group, while other objective image quality metrics were comparable between the two groups.

Table 5.

Objective image quality of implanted stent (all)

All stents
(n = 88)
PCDUHR-low (n = 45) EID-low
(n = 43)
p value
Stent-induced blooming [%] 37.4 ± 14.6 28.1 ± 9.8 47.1 ± 12.2  < 0.01
ΔHUin_stent [%] 22.1 ± 22.9 15.1 ± 13.4 29.4 ± 28.1  < 0.01
Edge sharpness [HU/mm] 4792.0 ± 3068.3 7178.2 ± 2550.5 2294.8 ± 448.8  < 0.01
FWHM-stent [mm] 0.61 ± 0.17 0.50 ± 0.11 0.71 ± 0.15  < 0.01
FWHM-lumen [mm] 2.19 ± 0.62 2.36 ± 0.56 2.02 ± 0.64  < 0.01

Data are presented as the mean ± standard deviation. p value indicates comparison of parameters between the groups. FWHM full width at half maximum

Figure 6 illustrates representative cases of group-wise comparisons stratified by stent diameter. The upper row demonstrates the appearance of stents in the two groups, highlighting the visual differences in stent depiction. The lower row shows attenuation profile curves, which reveal that stents in the PCDUHR-low group are depicted with thinner stent struts (as indicated by smaller FWHM-stent values) and larger lumen diameters (as indicated by larger FWHM-lumen values) as compared to the EID-low group.

Fig. 6.

Fig. 6

Representative cases of stents with different diameters. The figures show representative images and attenuation profile curves for different stent diameters (A: 4 mm, B: 3 mm, C: 2.5 mm, D: 2.25 mm) between PCDUHR-low and EID-low. The top row displays cross-sectional and longitudinal images of the stents between the groups. The middle row shows attenuation profile curves, indicating thinner stent struts (narrower FWHM-stent) and larger lumen diameters (wider FWHM-lumen) in the PCDUHR-low group as compared to the EID-low groups. PCDUHR-low, CCTA scans with ECG-triggered prospective axial acquisition, UHR scan mode, and a low tube potential of 70 or 90 kVp using a PCD-CT scanner; EID-low, CCTA scans with ECG-triggered prospective axial acquisition, and a low tube potential of 70 or 90 kVp using EID-CT; FWHM, full width at half maximum

Discussion

For the primary analysis of implanted stents, the PCDUHR-low group had significantly better overall subjective image quality with superior ratings for sharpness and beam hardening artifacts than the EID-low group; this benefit was particularly pronounced in stents with a diameter of less than 3 mm. Objectively, the PCDUHR-low protocol resulted in significantly less stent-induced blooming and superior edge sharpness as compared to the EID-low protocol. When compared to the PCDUHR-std group, the PCDUHR-low protocol provided comparable image quality but at a significantly lower radiation dose.

EID-CT has demonstrated efficacy in evaluating coronary artery stents, particularly with advancements in multidetector row technology [22]. Studies using 320-row EID-CT have shown acceptable diagnostic accuracy for in-stent restenosis, suggesting a potential non-invasive alternative to invasive coronary angiography [9]. However, EID-CT has limitations in stent evaluation due to artifacts and limited spatial resolution, especially in smaller stents or cases with severe blooming artifacts from metal stent struts, which can reduce diagnostic accuracy [11]. Previous studies have reported that 8–28% of stent lumens cannot be assessed due to stent-induced artifacts [5, 9, 23–25], primarily caused by partial volume averaging and beam hardening. Consequently, the effectiveness of CCTA using EID-CT in patients with previous stent implantation, particularly those with smaller diameter stents, remains controversial due to the rate of unassessable stents.

To date, no consensus has been reached regarding the optimal tube voltage for coronary stent imaging with EID-CT, although 120 kVp is typically used; higher tube voltages, such as 140 kVp, have been employed in some cases due to their ability to penetrate dense stent struts and reduce blooming artifacts [10], but this approach can increase radiation exposure, which is a significant concern for patients undergoing frequent imaging [4]. However, some reports have shown that lowering the tube voltage can improve the CNR, thereby enhancing stent lumen visibility; studies by Eisentopf et al. [24] and Lee et al. [25] suggested that lower tube voltages (80 to 100 kVp) can provide comparable image quality to that produced by higher tube voltages for stent assessment while reducing radiation doses, offering a viable alternative for routine clinical practice.

PCD-CT have represented a significant advancement in cardiac CT, offering substantial improvements in image quality and noise reduction. This technology, which directly converts X-ray photons into electronic signals, employs smaller detector elements that enable UHR imaging and achieves superior SNR by effectively eliminating electronic noise [12]. Studies such as Geering et al. [26] and von Spiczak et al. [16] have demonstrated that UHR mode offered superior spatial resolution, potentially enabling clearer visualization of stent struts and more accurate assessment of in-stent restenosis compared to EID-CT. These benefits are attributed not only to the increased spatial resolution provided by smaller detector elements, but also to the use of dedicated sharp convolution kernels, which may help reduce blooming artifacts and enhance stent lumen delineation. In the present study, the use of PCD-CT, as compared to EID-CT, also demonstrated a significant reduction in stent-induced artifacts.

In this study, low-tube-potential UHR CCTA using PCD-CT (PCDUHR-low) demonstrated clear superiority over low-tube-potential EID-CT (EID-low) in coronary stent evaluation. Notably, this superiority was evident in overall image quality, including subjective assessments and several objective measures, despite the comparable radiation dose between the two protocols. Importantly, EID-low has been validated in recent studies as a clinically feasible approach for stent evaluation, owing to advancements in image reconstruction techniques such as deep learning and sharp kernel methods [27, 28]. Against this validated reference, PCDUHR-low provided further improvements in image quality. Specifically, it effectively reduced blooming artifacts and improved the visualization of the stent lumen, likely owing to the intrinsic advantages of PCD technology, such as smaller detector elements and the elimination of electronic noise. The PCD-CT group exhibited higher image noise (SD) and lower SNR and CNR compared with the EID-CT group; however, these objective differences can be largely attributed to the employment of a sharper reconstruction kernel (Bv64 vs. Bv49), which enhances spatial resolution at the cost of increased noise. Importantly, overall image quality must integrate both quantitative and qualitative aspects. In our combined evaluation, PCD-CT’s superior spatial resolution and visually sharp stent delineation resulted in higher subjective scores, compensating for its increased noise. This finding indicates the necessity of balancing objective noise measures against subjective image‐quality characteristics when optimizing UHR CCTA protocols. By providing higher-quality images with fewer artifacts while maintaining a low radiation dose, low-tube-potential UHR CCTA using PCD-CT may enable more accurate and reliable detection of in-stent restenosis, potentially contributing to more appropriate and efficient selection of patients for invasive coronary angiography. On the other hand, low-tube-potential UHR CCTA provided comparable image quality for stent evaluation while achieving a significantly lower radiation dose. In our cohort, the mean DLP of the low-tube-potential UHR CCTA was 107.1 ± 50.4 mGy∙cm, significantly lower than that of the standard UHR protocol (256.8 ± 83.2 mGy∙cm). This reduction is also notable when compared with previous reports of UHR CCTA using standard tube potentials, with previously reported DLP values ranging from approximately 240 to 936 mGy∙cm [29–32].

This study has several limitations. First, it was a single-center retrospective investigation, which may limit the generalizability of the findings. Second, the imaging modes were not compared within the same patients, potentially introducing selection bias and limiting the robustness of direct comparisons. Third, different reconstruction kernels were used for PCD-CT and EID-CT, which might have influenced stent evaluation, including blooming artifacts and lumen visibility. In this study, we prioritized selecting the reconstruction kernels that were considered most appropriate for stent evaluation with each scanner [13, 24, 33]. Forth, different stent materials may result in varying levels of x-ray attenuation and thus could have influenced image quality. However, because we lacked complete data on the stent materials used in some patients, we neither classified stents according to their propensity to produce artifacts nor performed statistical analyses comparing image quality by stent material. Moreover, even if we restricted our analysis to those stents with known materials, the wide variety of stent types would necessitate dividing them into multiple subgroups, which would likely result in insufficient statistical power. The fifth limitation is the lack of standardization in tube voltage selection between the two CT systems. This reflects the retrospective nature of the study, where tube voltage was determined according to routine clinical protocols, and available voltage options differed between the systems. Such differences may have influenced image quality metrics, particularly noise and contrast. Sixth, although our preliminary retrospective analysis suggests that stent visualization in PCDUHR-low may be comparable to that at PCDUHR-std, the small sample size of the PCDUHR-std group and the lack of matched baseline characteristics between the two groups limits the strength and generalizability of these observations. Future work should therefore include prospective imaging of the same patients across a spectrum of tube voltages under standardized conditions or, alternatively, validation using a phantom model. Such studies will be critical to establish the optimal tube voltage in UHR CCTA. Finally, the absence of a reference standard such as invasive angiography leaves uncertainty regarding the true accuracy of the imaging findings and warrants further validation by standardized reference assessments.

Conclusion

This study demonstrated that the low-tube-potential UHR mode in PCD-CT offers a favorable balance between dose reduction and image quality improvement in coronary stent imaging. Compared with low-tube-potential imaging with EID-CT, this approach achieved similar radiation doses while yielding superior subjective and objective image quality, including fewer stent-related artifacts. These findings suggested the importance of optimizing tube voltage in UHR mode to achieve safer, more accurate coronary stent assessments.

Supplementary Information

Below is the link to the electronic supplementary material.

Funding

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

Declarations

Conflict of interest

Department of Advanced Diagnostic Imaging, where Dr. Nakamura and Dr. Kitagawa currently belong to, is an endowment department supported with an unrestricted grant from Siemens Healthcare K.K. Hajime Sakuma receives departmental research grant support from Eisai Co., Ltd., Guerbet Japan KK, FUJIFILM Toyama Chemical Co., Ltd., Nihon Medi-Physics Co., Ltd., PDRadiopharma Inc. and GE Healthcare Japan. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose.

Ethics approval

This study was approved by the Clinical Research Ethics Review Committee of Mie University Hospital (approval number: H2019-207).

Informed consent

The institutional review board of our hospital approved the study protocol and waived the need for individual consent due to its retrospective design.

Footnotes

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Knuuti J, Wijns W, Saraste A, Capodanno D, Barbato E, Funck-Brentano C, et al. 2019 ESC guidelines for the diagnosis and management of chronic coronary syndromes. Eur Heart J. 2020;41:407–77. [DOI] [PubMed] [Google Scholar]
  • 2.Wolk MJ, Bailey SR, Doherty JU, Douglas PS, Hendel RC, Kramer CM, et al. ACCF/AHA/ASE/ASNC/HFSA/HRS/SCAI/SCCT/SCMR/STS 2013 multimodality appropriate use criteria for the detection and risk assessment of stable ischemic heart disease: a report of the American College of Cardiology Foundation appropriate use criteria task force. J Am Coll Cardiol. 2014;63:380–406. [DOI] [PubMed] [Google Scholar]
  • 3.Budoff MJ, Dowe D, Jollis JG, Gitter M, Sutherland J, Halamert E, et al. Diagnostic performance of 64-multidetector row coronary computed tomographic angiography for evaluation of coronary artery stenosis in individuals without known coronary artery disease: results from the prospective multicenter ACCURACY trial. J Am Coll Cardiol. 2008;52:1724–32. [DOI] [PubMed] [Google Scholar]
  • 4.Gulati M, Levy PD, Mukherjee D, Amsterdam E, Bhatt DL, Birtcher KK, et al. 2021 AHA/ACC/ASE/CHEST/SAEM/SCCT/SCMR guideline for the evaluation and diagnosis of chest pain: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. J Cardiovasc Comput Tomogr. 2022;16:54–122. [DOI] [PubMed] [Google Scholar]
  • 5.Eckert J, Renczes-Janetzko P, Schmidt M, Rolf A, Ropers D, Ropers U, et al. Coronary CT angiography (CCTA) using third-generation dual-source CT for ruling out in-stent restenosis. Clin Res Cardiol. 2019;108:402–10. [DOI] [PubMed] [Google Scholar]
  • 6.Maintz D, Juergens KU, Wichter T, Grude M, Heindel W, Fischbach R. Imaging of coronary artery stents using multislice computed tomography: in vitro evaluation. Eur Radiol. 2003;13:830–5. [DOI] [PubMed] [Google Scholar]
  • 7.Gassenmaier T, Petri N, Allmendinger T, Schoenberg SO, Schoepf UJ, Nikolaou K, et al. Next generation coronary CT angiography: in vitro evaluation of 27 coronary stents. Eur Radiol. 2014;24:2953–61. [DOI] [PubMed] [Google Scholar]
  • 8.de Graaf FR, Schuijf JD, van Velzen JE, Kroft LJ, de Roos A, Doevendans PA, et al. Diagnostic accuracy of 320-row multidetector computed tomography coronary angiography to noninvasively assess in-stent restenosis. Invest Radiol. 2010;45:331–7. [DOI] [PubMed] [Google Scholar]
  • 9.Ghekiere O, Salgado R, Buls N, De Bondt P, Mancini I, Van den Broeck R, et al. Image quality in coronary CT angiography: challenges and technical solutions. Br J Radiol. 2017;90:20160567. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Kalisz K, Buethe J, Saboo SS, Abbara S, Halliburton SS, Rajiah P. Artifacts at cardiac CT: physics and solutions. Radiographics. 2016;36:2064–83. [DOI] [PubMed] [Google Scholar]
  • 11.Sandfort V, Persson M, Pourmorteza A, Noël PB, Fleischmann D, Marin D. Spectral photon-counting CT in cardiovascular imaging. J Cardiovasc Comput Tomogr. 2021;15:218–25. [DOI] [PubMed] [Google Scholar]
  • 12.Leng S, Bruesewitz M, Tao S, Rajendran K, Halaweish AF, Campeau NG, et al. Photon-counting detector CT: system design and clinical applications of an emerging technology. Radiographics. 2019;39:729–43. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Mergen V, Sartoretti T, Baer-Beck M, Morsbach F, Euler A, Alkadhi H, et al. Ultra-high-resolution coronary CT angiography with photon-counting detector CT: feasibility and image characterization. Invest Radiol. 2022;57:780–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Greffier J, Si-Mohamed SA, Lacombe H, Dabli D, Boussel L, Douek P, et al. Virtual monochromatic images for coronary artery imaging with a spectral photon-counting CT in comparison to dual-layer CT systems: a phantom and a preliminary human study. Eur Radiol. 2023;33:5476–88. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.von Spiczak J, Mannil M, Peters B, Baer M, Morsbach F, Alkadhi H, et al. Photon-counting computed tomography with dedicated sharp convolution kernels: tapping the potential of a new technology for stent imaging. Invest Radiol. 2018;53:486–94. [DOI] [PubMed] [Google Scholar]
  • 16.Vattay B, Boussoussou M, Vecsey-Nagy M, Varga-Szemes A, Schoepf UJ, De Cecco CN, et al. Qualitative and quantitative image quality of coronary CT angiography using photon-counting computed tomography: standard and ultra-high-resolution protocols. Eur J Radiol. 2024;175: 111426. [DOI] [PubMed] [Google Scholar]
  • 17.Decker JA, O’Doherty J, Schoepf UJ, Varga-Szemes A, Emrich T, De Cecco CN, et al. Stent imaging on a clinical dual-source photon-counting detector CT system-impact of luminal attenuation and sharp kernels on lumen visibility. Eur Radiol. 2023;33:2469–77. [DOI] [PubMed] [Google Scholar]
  • 18.Kwan AC, Pourmorteza A, Stutman D, McCollough CH. Next-generation hardware advances in CT: cardiac applications. Radiology. 2021;298:3–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Hagar MT, Soschynski M, Saffar R, Varga-Szemes A, De Cecco CN, Schoepf UJ, et al. Ultra-high-resolution photon-counting detector CT in evaluating coronary stent patency: a comparison to invasive coronary angiography. Eur Radiol. 2024;34:4273–83. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Boccalini S, den Harder AM, Witsenburg M, Roos-Hesselink JW, van der Lugt A, de Roos A, et al. Computed tomography image quality of aortic stents in patients with aortic coarctation: a multicentre evaluation. Eur Radiol Exp. 2018;2:17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Takafuji M, Kitagawa K, Mizutani S, Kuroda K, Tanimoto T, Tanaka R, et al. Super-resolution deep learning reconstruction for improved image quality of coronary CT angiography. Radiol Cardiothorac Imaging. 2023;5: e230085. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Das KM, El-Menyar AA, Salam AM, Al Suwaidi J, Al Binali HA. Contrast-enhanced 64-section coronary multidetector CT angiography versus conventional coronary angiography for stent assessment. Radiology. 2007;245:424–32. [DOI] [PubMed] [Google Scholar]
  • 23.Carbone I, Francone M, Algeri E, Agati L, Catalano C, Passariello R. Non-invasive evaluation of coronary artery stent patency with retrospectively ECG-gated 64-slice CT angiography. Eur Radiol. 2008;18:234–43. [DOI] [PubMed] [Google Scholar]
  • 24.Eisentopf J, Achenbach S, Ulzheimer S, Pflederer T, Ropers D, Daniel WG, et al. Low-dose dual-source CT angiography with iterative reconstruction for coronary artery stent evaluation. JACC Cardiovasc Imaging. 2013;6:458–65. [DOI] [PubMed] [Google Scholar]
  • 25.Lee JW, Kim CW, Lee HC, Kim YJ, Kim TH, Kim JK, et al. High-definition computed tomography for coronary artery stents: image quality and radiation doses for low voltage (100 kVp) and standard voltage (120 kVp) ECG-triggered scanning. Int J Cardiovasc Imaging. 2015;31(Suppl 1):39–49. [DOI] [PubMed] [Google Scholar]
  • 26.Geering L, Sartoretti T, Mergen V, Baer-Beck M, Euler A, Alkadhi H, et al. First in-vivo coronary stent imaging with clinical ultra-high-resolution photon-counting CT. J Cardiovasc Comput Tomogr. 2023. 10.1016/j.jcct.2023.03.001. [DOI] [PubMed] [Google Scholar]
  • 27.Wu W, Zhan H, Wang Y, Zhang Y, Liu X, Li Y, et al. Improving image quality and in-stent restenosis diagnosis with high-resolution “double-low” coronary CT angiography in patients after percutaneous coronary intervention. Front Cardiovasc Med. 2024;11:1330824. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Feng R, Tong J, Liu X, Zhao Y, Zhang L. High-pitch coronary CT angiography at 70 kVp adopting a protocol of low injection speed and low volume of contrast medium. Korean J Radiol. 2017;18(5):763–72. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Hoe J, Toh KH. Initial clinical experience with coronary CT angiography performed on dual-source photon counting CT using different cardiac scan modes-analysis of image quality and radiation dose. Cardiovasc Imaging Asia. 2023;7:39–48. [Google Scholar]
  • 30.Hagar MT, Soschynski M, Saffar R, Varga-Szemes A, De Cecco CN, Schoepf UJ, et al. Accuracy of ultra-high-resolution photon-counting CT for detecting coronary artery disease in a high-risk population. Radiology. 2023;307: e223305. [DOI] [PubMed] [Google Scholar]
  • 31.Sharma SP, Verhemel S, Hirsch A, Varga-Szemes A, De Cecco CN, Schoepf UJ, et al. Diagnostic performance of high and ultra-high-resolution photon-counting CT for detection of coronary artery disease in patients evaluated for transcatheter aortic valve implantation. Int J Cardiovasc Imaging. 2024. 10.1007/s10554-024-02827-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Vecsey-Nagy M, Varga-Szemes A, Schoepf UJ, De Cecco CN, Emrich T, Decker JA, et al. Ultra-high-resolution coronary CT angiography on photon-counting detector CT: bi-centre study on the impact of quantum iterative reconstruction on image quality and accuracy of stenosis measurements. Eur J Radiol. 2024;176: 111517. [DOI] [PubMed] [Google Scholar]
  • 33.McCollough CH, Rajiah P, Bois JP, Winfree TN, Carter RE, Rajendran K, et al. Comparison of photon-counting detector and energy-integrating detector CT for visual estimation of coronary percent luminal stenosis. Radiology. 2023;309(3): e230853. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials


Articles from Japanese Journal of Radiology are provided here courtesy of Springer

RESOURCES