. 2005 Oct;18(4):583–607. doi: 10.1128/CMR.18.4.583-607.2005

Current and Developing Technologies for Monitoring Agents of Bioterrorism and Biowarfare

Daniel V Lim 1,*, Joyce M Simpson 1, Elizabeth A Kearns 1, Marianne F Kramer 1
PMCID: PMC1265906  PMID: 16223949

Abstract

Recent events have made public health officials acutely aware of the importance of rapidly and accurately detecting acts of bioterrorism. Because bioterrorism is difficult to predict or prevent, reliable platforms to rapidly detect and identify biothreat agents are important to minimize the spread of these agents and to protect the public health. These platforms must not only be sensitive and specific, but must also be able to accurately detect a variety of pathogens, including modified or previously uncharacterized agents, directly from complex sample matrices. Various commercial tests utilizing biochemical, immunological, nucleic acid, and bioluminescence procedures are currently available to identify biological threat agents. Newer tests have also been developed to identify such agents using aptamers, biochips, evanescent wave biosensors, cantilevers, living cells, and other innovative technologies. This review describes these current and developing technologies and considers challenges to rapid, accurate detection of biothreat agents. Although there is no ideal platform, many of these technologies have proved invaluable for the detection and identification of biothreat agents.

INTRODUCTION

Recent bioterrorist events have emphasized the need to immediately detect and identify biothreat agents. Rapid, accurate identification of such agents is important not only to confirm that a bioterrorism event has occurred, but also to determine whether suitable measures should be implemented to protect public health. Clinical microbiology laboratories have access to the Laboratory Response Network, which is an organized system designed by the Centers for Disease Control and Prevention (CDC) to provide structured guidance for the detection, diagnosis, and reporting of biological threat agents (263; http://www.bt.cdc.gov/lrn/factsheet.asp).

The Laboratory Response Network is structured into four laboratory levels based upon testing capabilities and likelihood of handling biothreat agents. Level A laboratories are standard clinical laboratories which may handle specimens for likely biothreat agents as part of classic diagnostic analysis on hospital cultures. These laboratories are the first line of detection and need to be familiar with likely agents such as Bacillus anthracis, Yersinia pestis, and Francisella tularensis. Level A laboratories conduct diagnostic analyses as dictated by the CDC. Level B laboratories can confirm identification of suspicious isolates and are typically public health laboratories. Level C laboratories use molecular methods and typing procedures to further identify or confirm identification of species and strains of biothreat agents and are either typing laboratories or public health laboratories. Level D laboratories located at the CDC and U.S. Army Medical Research Institute of Infectious Diseases are able to do high-level characterization of biothreat isolates under tightly controlled conditions.

Various tests have been developed to detect and identify biothreat agents. Some of these tests were available before 11 September 2001; other tests have been developed since that time. Although many of these technologies claim to be rapid, accurate, and reliable, few have been extensively evaluated under field conditions. This review describes documented current and developing technologies for detection and identification of biothreat agents and addresses the challenges associated with detection in complex sample matrices.

BRIEF HISTORY

There have been many reviews written regarding the history, theory, and use of bioterrorism, biothreats, biological weapons, and biological warfare from the 14th century to today. The reader is referred to the following for additional information: Atlas (17), Christopher et al. (55), Hawley and Eitzen (115), Heden (117), Klietmann and Ruoff (160), van Courtland Moon (293), and Tucker (283).

One of the most well known early attempts to use biological agents was during the 14th century medieval siege of Kaffa (Feodosiya, Ukraine) (55, 305). The attacking Tartars (Mongols) catapulted dead and dying plague victims into the city in the attempt to spread the disease. There has been speculation that escaping Kaffa victims may have carried the plague (also known as black death, bubonic plague, or black plague) to other parts of Europe, thereby hastening the ensuing pandemic. An attempt was made in 1763 by the British at Fort Pitt in the Ohio River Valley to use blankets to transfer smallpox to Native Americans (55, 305). The fort had been afflicted with smallpox among the troops, and used linens from the infirmary were systematically dispensed to the neighboring Indian populations. There is also anecdotal reference of attempts to spread smallpox via infected British soldiers during the American Revolutionary War (1776 to 1781) and by contaminated clothing during the American Civil War (1860 to 1864) (55, 305).

The era of biological weapons was significantly advanced in the 20th century by modern microbiology and multiple international wars. The biological and chemical horrors inflicted during World War I resulted in the drafting of the 1925 Geneva Protocol for the Prohibition of the Use in War of Asphyxiating, Poisonous or Other Gases and of Bacteriological Methods of Warfare. However, many countries that signed the document did so with contingencies in the event of attack by a nonratifying entity and with the stipulation that the protocol did not prevent investigative research (http://www.state.gov/t/ac/trt/4784.htm).

Subsequently, Germany, Japan, the Union of Soviet Socialist Republics, and the United States initiated research programs to develop and refine biological weapons (55, 293). In 1969, under President Richard M. Nixon, the United States began dismantling its offensive biological weapon programs. Henceforth, all biothreat agent research programs in the United States were of a defensive nature, and the 1972 Geneva Convention on the Prohibition of the Development, Production, and Stockpiling of Bacteriological and Toxin Weapons and on Their Destruction was developed and ratified (http://www.state.gov/t/ac/trt/4718.htm#treaty). However, several countries that signed the convention (notably the Union of Soviet Socialist Republics and Iraq) continued offensive research and production of biological agents as recently as the mid-1990s. Additionally, there have been increasingly more subnational terrorist and radical groups that have independently worked on offensive use of biological weapons since the mid-1980s to today (283, 285).

The Chemical and Biological Weapons Nonproliferation Program maintains an open-source database containing information on all known incidents worldwide involving chemical, biological, or nuclear materials since 1900 (285). This database is searchable, and annual summary reports of activity are available; however, large amounts of biological weapons material remain unaccounted for and, thus, there is a critical need to develop methods for rapid, accurate detection of biothreat agents.

CHALLENGES TO DETECTION

The prospect of biological terrorism presents many challenges for detection platforms. Some challenges are unique to bioterrorism and others are common for all testing situations. Ideally, detection platforms should be capable of rapidly detecting and confirming biothreat agents, including modified or previously uncharacterized agents, directly from complex matrix samples, with no false results. Furthermore, the instrument should be portable, user-friendly, and capable of testing for multiple agents simultaneously. Although several detection platforms exhibit many of the desired characteristics, no one system satisfies all of the criteria.

Detection assays must be sensitive and specific, capable of detecting low concentrations of target agents without interference from background materials. Although many chemical detectors can detect chemical agents at levels that pose a risk to human health, biological detectors can only rarely detect microorganisms directly from samples at or below human risk levels because of their lack of sensitivity.

In general, nucleic acid-based detection systems are more sensitive than antibody-based detection systems. The PCR assay can detect 10 or fewer microorganisms in a short period of time (23, 95, 133). However, PCR requires a clean sample and is unable to detect protein toxins and other non-nucleic acid-containing analytes such as prions. Furthermore, cultures of the target organism are not available for archiving and additional tests after PCR analysis.

Specificity is as important as sensitivity in the detection of biothreat agents. High specificity is important to minimize background signals and false-positive results from samples that are often complex, uncharacterized mixtures of organic and inorganic materials. Specificity can be affected not only by humics and other background particles, but also by high concentrations of competing antigens and DNA. In the case of PCR, its high sensitivity can also be a major weakness because contaminating or carryover DNA can be amplified, resulting in false-positive results.

In addition to sensitivity and specificity, reproducibility is an important requirement for detection platforms. Detection platforms that do not provide reproducible results are unreliable and may exacerbate a terrorist event. Many factors can affect the repeatability of bioassays, including the stability and consistency of reagents and differences in assay conditions. These variations can often be reduced by standardizing assay conditions and procedures.

Detection platforms must be capable of detecting a variety of biothreat agents in samples. This multiplex capability is vital because suspect samples may contain toxins, bacteria, viruses, or other types of analytes. In some instances, known biothreat agents may have been deliberately altered through genetic, antigenic, or chemical modifications or may represent new or uncommon variants of known microorganisms. Such modifications can make detection of a biothreat agent difficult.

Even without modifications, conventional biothreat agents are difficult to detect in complex sample matrices. Samples such as human specimens (for example, blood and stool), powder, food, water, and even air present challenges for detection assays. Anticoagulants, leukocyte DNA, and heme compounds in blood inhibit PCRs (97, 206). Lipids in ground beef and high numbers of background bacteria in stool specimens interfere with immunoassays. For this reason, target analytes usually must be isolated or purified from such samples prior to analysis and identification. These steps can add hours or days to detection protocols and often cannot be performed in the field. A further complication is that some viable microbes may not be culturable or may require specific nutritional requirements for culture.

An important consideration in biodetection is the collection and handling of samples. Issues associated with sampling include the type of material to be tested, the collection procedure, and sample transport. Air-borne and water-borne samples generally must be concentrated from large volumes to detect low levels of target analytes. Air-borne samples must also be extracted to a liquid because most detection platforms process only liquid samples. The efficiency of recovery from concentration and extraction procedures can vary and affect detection limits. Sample size, number, and distribution should be considered, as well as the transport time and method, particularly for fastidious, living microbes that may require specific environmental and nutritional conditions for survival. In some instances, confirmation of microbe viability may be important to ascertain whether the microorganism poses a health threat.

SAMPLE MATRIX PROCESSING

Sample Processing

Conventional culture and staining techniques are currently the gold standard for isolation, detection, and identification of target biothreat agents. These culture and isolation methods are based on the ability of healthy bacterial cells to multiply in nutrient-rich medium containing selective and differential agents that inhibit the growth of nontarget organisms and differentiate target from nontarget organisms. Use of culture enrichment and selection, however, results in lengthy assays, which can take days for preliminary results. Rapid detection methods replace the selective and differential culturing steps with DNA hybridization, nucleic acid amplification, antibody agglutination, and/or enzyme immunoassays. The majority of these rapid detection methods are suitable only when the biothreat agent is present in high numbers and/or in the absence of interfering substances.

In most cases, rapid detection methods require steps to grow or concentrate the target biothreat agent and/or purify the target analyte from the sample matrix prior to rapid detection. For example, PCR and nucleic acid sequence-based amplification (NASBA) enrich a single specific DNA or RNA sequence up to 106-fold in 20 min to a few hours and theoretically have a sensitivity of a single bacterial cell. These methods give rapid, specific detection but are limited by small sample volumes (e.g., 5 μl). In addition, substances such as bile salts, polysaccharides, heme, and humic acids in sample matrices inhibit enzymatic reactions required for nucleic acid amplification (236). Low levels of target analyte in samples require concentration and/or cultural enrichments to provide sufficient target for amplification.

Methods for Sample Processing

Ideally the method to separate, concentrate, and purify the target biothreat agent should be universal, utilizable for all samples for all types of target analytes. In addition, the sample preparation method should be capable of rapidly removing the sample matrix that could inhibit detection capabilities and concentrating the analyte. The majority of current sample preparation methods can do neither of the above. The preparation procedure is usually limited to specific types of samples and is generally time-consuming and sometimes labor intensive. In addition, the concentration/purification method should ideally maintain cell viability so that, upon positive detection, the target organism can be cultured for confirmation of viability, further characterization and archived as evidence in the case of a criminal investigation.

Many sample preparation methods are currently under investigation. These methods include chemical, physical, and biological manipulation of the sample and are described elsewhere (25, 236, 272). Many of these methods have been developed for foods because of the historic need to test food for the presence of pathogens. Therefore, the majority of the methods described in this review will be for food sample preparation, but it would be possible to modify many of these methods for other types of sample matrices. This discussion will focus on rapid and/or automatic sample preparation methods for various solids, liquids, and aerosols and will include methods that are commonly used for detection of biothreat agents in samples such as blood, powder, and air.

Centrifugation.

Centrifugation is a method that has conventionally been used to concentrate and recover microorganisms from liquid samples. Centrifugation times can vary from 30 seconds to 1 hour depending on the type of sample and the number of wash steps. Undesired sample debris may also be concentrated during the process and the centrifugation process cannot be easily automated. Escherichia coli O157:H7 cells have been recovered directly from a ground beef/buffer suspension by a 5-minute differential centrifugation step that separated the suspension into three distinct layers (69, 71). The middle layer containing the majority of the target cells was used for rapid detection of E. coli O157:H7 with an evanescent wave fiber optic biosensor. The efficiency of E. coli O157:H7 cells recovery from the ground beef by this procedure was not determined. Buoyant density gradient centrifugation has been used to separate and concentrate Yersinia enterocolitica in meat fluids from pork (311). This centrifugation procedure removed dead cells, and the concentrated samples contained only viable cells that were then used directly for PCR. Buoyant density centrifugation has also been successfully used to separate and concentrate bacteria from food in a one minute procedure (184, 185).

Filtration.

Filtration can be used to separate microorganisms on the basis of cell size. Although liquid samples can be rapidly forced through filters of different pore sizes, sample debris can clog filters and retain bacteria. Removal of bacteria from filters following filtration can also be difficult. The commercially available Iso-Grid (Neogen Corp., Lansing, MI) is a dual filtration method developed for food products (227). Prefiltered food is passed through a 0.45-μm filter, which is then placed on an agar plate. Bacteria captured on the filter are grown and detected using selective and differential media. Iso-Grid filtration for E. coli O157:H7 has been approved by AOAC International, AOAC Official Method 997.11 (14, 88).

FTA filters (Whatman, Springfield, KY) have been developed for rapid isolation of nucleic acids from environmental, clinical, or food samples. Samples are added directly to the filter. The filter can then be washed and the nucleic acid remains bound to the filter. The filter is then ready for processing or long term storage at room temperature. FTA filters have been used for the detection of Bacillus subtilis, Bacillus cereus, and Bacillus megaterium spores using nested PCR. The reported sensitivity of this method is 53 spores in the first round of PCR, and 5 spores after the second nested PCR round (169). The preparation of template DNA from biothreat agents such as Bacillus anthracis spores using FTA filters should be similar although samples such as powder have not yet been tested.

Dielectrophoresis.

Dielectrophoresis takes advantage of the intrinsic dielectric properties of bioparticles to enable separation of particles in nonuniform electric fields. In dielectrophoresis, a nonuniform electric field consisting of positive and negative dielectrophoretic forces is generated by microelectrodes in a small chamber. Different bacterial cells and other components in the sample can be separated based on each particle's effective conductivity. Particles are released from regions near the electrodes when the dielectrophoretic response of each particle changes from attraction to repulsion from such regions. Different species of bacteria have different cell wall structures and compositions, and these differences give rise to large differences in particle conductivities. Changes in the physiological state of the cell will also affect particle conductivities. Dielectrophoresis has been incorporated into a microfabricated bioelectronic chip and used to separate E. coli from the sample (54, 315). Dielectrophoresis has also been used to clean target cells by removing inhibitors (228) and to isolate parasitized cells (98). Insulator-based dielectrophoresis has been used to isolate and trap E. coli and vegetative cells of Bacillus species into two distinct locations (170).

Immunomagnetic separation.

Antibodies also can be used to purify and concentrate target biothreat agents. Magnetic particles (beads) with antibodies immobilized on their surfaces can bind to target cells, toxins, or other molecules found in samples. The magnetized particles are then collected using a magnetic field. Sample debris and nontarget organisms and molecules are removed by washing. The particles are released when the magnetic field is removed. The remaining solution contains mainly concentrated cells or molecules of interest that have bound to the antibody attached to the magnetic particle. This antibody-based method of cell separation is referred to as immunomagnetic separation. Immunomagnetic separation has been used to isolate and concentrate Listeria monocytogenes, Salmonella spp., and E. coli O157:H7 from stool and food samples (131, 148, 243, 267, 291). Cryptosporidium spp. have been successfully isolated from water samples using immunomagnetic separation (41), and the method is a component of U.S. Environmental Protection Agency Method 1622 and Method 1623 for the recovery and identification of Cryptosporidium spp. and Giardia spp. in environmental waters (289, 290). Dynabeads (Dynal Inc., Oslo, Norway) supplies immunomagnetic separation beads that are functionalized for covalent linkage to antibodies or peptides that target biothreat agents. Beads with protein A or G linkage for noncovalent attachment of antibodies to the beads are also available.

Immunomagnetic separation has been used in a field-deployable automated electomagnetic flow cell fluidics system (Biodetection Enabling Analyte Delivery System [BEADS]) to separate and concentrate pathogenic bacterial cells as well as nucleic acid. The BEADS system engineered by Pacific Northwest National Laboratory (Richland, WA) was designed specifically for processing environmental and clinical samples prior to biodetection. Porous Ni foam was used to enhance the magnetic field gradient within the flow path so that the immunomagnetic separation particles could be immobilized throughout the fluid rather than at the tubing wall (50-52). E. coli O157:H7 was recovered at 32% efficiency directly from poultry carcass rinse without cell growth and enrichment using the BEADS system. Commercially available antibodies targeted against the O and K antigens were used for capturing the E. coli onto beads, and the recovered beads were used for direct PCR amplification and microarray detection (50). As long as antibodies or other affinity-based molecules are available for attachment to the beads, immunomagnetic separation can be a versatile technique potentially available for the purification of target biothreat agents and their products from heterogeneous sample matrices.

Nucleic acid extraction.

The efficiency of extraction and purification of nucleic acid (DNA or RNA) from cells or spores of potential biothreat agents influences the sensitivity, reproducibility, and accuracy of any nucleic acid detection method. Substances in complex sample matrices can inhibit hybridization and enzymatic reactions, degrade the nucleic acid, and reduce the efficiency of cell or spore lysis. In addition, spores must be disrupted (e.g., through sonication) prior to nucleic acid extraction. In large samples, the nucleic acid must also be concentrated, often over 1,000-fold, into a smaller volume appropriate for nucleic acid analysis. There are many kits commercially available for purifying nucleic acid (56). The following are a few examples of rapid nucleic acid purification methods that have been developed to purify and concentrate nucleic acid from complex sample matrices.

The Cepheid GeneXpert and Biothreat agent detection system (Cepheid, Sunnyvale, CA) integrates sample preparation, PCR, and detection into a disposable cartridge. An instrument automatically processes the cartridge, allowing sample preparation in fewer than 5 minutes and detection (four-color real-time PCR) in 25 min. No special skills are required to use the system. Large volumes (100 μl to 5 ml) of raw sample can be handled, and up to four targets can be detected simultaneously per cartridge. The system automatically sonicates and purifies the sample; PCR reagent is then added to the extracted nucleic acid, and the mixture is dispensed into a PCR tube (199). Swab extracts, serum, cerebrospinal fluid, urine, bone marrow, sputum, tissue, and whole blood samples have been tested with this system (21, 22, 199). Bacillus anthracis (Ames strain) spores have also been detected using the system (199).

Extraction of many types of samples for nucleic acid purification has also been automated using the MagNaPure compact or LC instrument (Roche Applied Science, Indianapolis, IN). Eight or 32 samples, respectively, can be purified in 1 to 3 hours. The purification process involves the binding of nucleic acid to surfaces of magnetic glass particles, extensive washing, and then elution of nucleic acid. The MagNaPure instrument has been used successfully to purify nucleic acid from stool (254, 312), serum (104), whole blood (91, 151, 201), urine and swab specimens (63), atherosclerotic tissue (210), sputum (237), cerebrospinal fluid, synovial fluid, and ticks (90), and powders (190).

Sample Matrix Processing

Sample matrices can be of different compositions and, therefore, may require different methods for processing. The following are examples of sample preparation methods and the impediments encountered for specific sample types.

Food.

The vast variety of food types makes it difficult to develop a universal sample preparation method for food. Many current methods rely on the time-consuming enrichment culture for the growth of the target microorganisms in the sample. More rapid methods such as immunomagnetic separation require specialized equipment and/or skilled labor. Obstacles to preparation of food samples for testing using rapid detection protocols include the requirement that food volumes need to be large (25 g or ml) and homogenous. The presence of food components such as lipids, polysaccharides, acid, and salts will inhibit enzymatic reactions required for certain rapid detection approaches such as PCR. Low concentrations of the target pathogen within the large sample will also decrease sensitivity of detection if small volumes are required for rapid detection analysis.

In conventional methods, solid food is diluted in a buffer solution and is blended or homogenized to obtain a homogenous mixture. The homogenous mixture is then further purified or cultured prior to testing. A homogenous mixture must still be prepared even when a rapid detection technique is used. The target microorganism can then be concentrated from the mixture by methods discussed previously. Unfortunately, the most effective technique for obtaining large numbers of target organisms in food is growth of the organisms by culture enrichment. Rapid effective separation and concentration of target pathogens remain a barrier to the rapid detection of microorganisms in foods.

Water.

Microorganisms and their by-products in drinking water and recreational waters are usually too dilute for direct measurements. Water samples generally must be concentrated to obtain the number of target microorganisms necessary for microbiological analysis. Detection of microorganisms or their toxins from water samples is difficult because the recovery efficiency is poor and/or substances that are inhibitory to the detection technique are present. Target microorganisms must be isolated and concentrated by several orders of magnitude from the water samples. Concentration is most often performed using ultrafiltration in which parasites, bacteria, viruses, and potentially some high-molecular-weight biotoxins are retained on the filter (76, 205, 289, 290). The material that has accumulated on the filter is removed, collected, and analyzed. Potential interferents such as chlorine, salts, humic acids, and indigenous bacteria may also accumulate on the filter and inhibit rapid detection. Target microbial pathogens have also been recovered from the water sample using immunomagnetic separation (87, 192).

Human specimens. (i) Blood.

Blood contains heme, erythrocytes, and other cells and biologically active components that can interfere with antibody- and nucleic acid-based assays. Removal of these components and labeling of the target pathogen in such a complex mixture are major hurdles that must be overcome when processing blood samples for rapid detection. Instead of removing the sample matrix, Spectral Diagnostics, Inc. (Canada), took advantage of the components in blood by developing a rapid whole-blood elimination test for gram-negative bacterial infection for patients in the intensive care unit. The U.S. Food and Drug Administration has approved the test to identify patients at risk for developing severe sepsis on admission to intensive care units. The test is performed directly on whole blood; separation of the target bacteria from the blood is not necessary. Endotoxin, if present, reacts with antiendotoxin antibody in the test. The endotoxin-antibody binding reaction primes the patient's neutrophils, resulting in an enhanced respiratory burst in the presence of zymosan. Zymosan is an insoluble preparation of yeast cells and has been shown to activate macrophages via toll-like receptor 2. The release of oxygen radicals results in chemiluminescence in the presence of luminol. The magnitude of the priming influence is proportional to the concentration of the endotoxin/antiendotoxin antibody complex (246).

An integrated miniaturized biochip device that automatically integrates sample preparation with PCR and DNA microarray detection has been reported by Liu et al. (186). This automated device mixes the blood sample with immunomagnetic separation beads in a sample storage chamber using air bubbles in a phenomenon termed microstreaming. The target pathogen bound to the immunomagnetic separation beads is then pumped to a PCR chamber, where the pathogen is concentrated using the magnetic element of the device and separated from whole blood using a washing step. The sample is then ready for direct addition of PCR reagents to the PCR chamber (186).

Blood is slightly acidic, thereby inhibiting antibody binding and potentially other enzymatic reactions. Taq DNA polymerase, the enzyme used to amplify DNA in PCR, is totally inhibited by as little as 0.004% (vol/vol) human blood (3); this inhibition is due to the heme in the blood (4). In addition, in many situations, the target microorganism must first be concentrated from large volumes of blood prior to DNA extraction and analysis. A procedure to rapidly process large-volume (6 ml) blood samples using a fibrin lysis cocktail and the Roche LC DNA isolation kit III (Roche Applied Science, Indianapolis, IN) has been reported (M. S. Ewert, personal communication). DNA preparations from blood processed using this procedure were analyzed by real-time PCR to detect low concentrations (<44 CFU/6 ml) of Yersinia pestis, Brucella suis, and Bacillus anthracis.

Dielectrophoresis can be used to discriminate and separate Plasmodium falciparum-infected cells from blood (98, 99). The plasma membranes of the infected erythrocytes have an increased ionic permeability, so that infected cells lose internal ions when suspended in a low-conductivity medium (98).

(ii) Urine.

It is unlikely that biothreat agents would need to be detected in the course of a urinary tract infection; however, urine contains yet-unidentified substances that are inhibitory to PCR (35, 40, 282, 297), and many components such as bacterial contamination commonly found in urine can interfere with detection results. For example, Western blots of urine samples used to detect the presence of protease-resistant prion proteins could be misinterpreted. Bands with a molecular mass similar to prion proteins can be detected from these Western blots (96).

There are several rapid methods available for processing urine in order to detect the total number of bacteria present. These methods include the FiltraCheck-UTI (Applied Polytchnology, Inc., Houston, TX), the UTIscreen (Los Alamos Diagnostics, Los Alamos, NM), and the Flash Track (Gen-Probe, San Diego, CA). These tests, along with the conventional cytospin Gram stain smear method, are used to process urine when large numbers of bacteria (≥105 CFU/ml) are present (189, 244). Studies that have investigated sample preparation methods for urine PCR have found that centrifugation of the specimen and DNA extraction of the pellet were critical factors to reducing PCR inhibition by urine constituents (26, 35, 297). A urine sample preparation method for PCR has been developed for Borrelia bugdorferi infection using seven different extraction methods (26). Only one of these methods, the DNAzol extraction method (Molecular Research Center, Inc., Cincinnati, OH), yielded positive results with 10 cells per PCR mixture. This method included a preextraction step, in which the sample was centrifuged at 36,000 × g for 30 min and ethanol precipitated after the addition of DNAzol (26).

(iii) Stool specimens.

Stool specimens contain many substances, such as bile salts and complex polysaccharides, that will inhibit PCR and other enzymatic reactions. In addition, there will be large numbers of background bacteria that will interfere with immunoassays and other affinity-based detection techniques. Immunomagnetic separation and the MagNaPure system, previously described in this article, are techniques that can be used to isolate and concentrate bacteria and nucleic acid from stool specimens. The QIAamp DNA stool purification kit (QIAGEN, Inc., Valencia, CA), intended for Taq DNA polymerase inhibitor removal, has been tested for detection of Shiga-toxin-producing E. coli by PCR directly from cattle fecal samples (106). The stx gene, encoding Shiga toxin, a defining characteristic of enterohemorrhagic E. coli strains, was amplified in the PCR, and the results were equivalent to culture, suggesting successful removal of PCR-interfering inhibitors from the feces.

Detection of Helicobacter pylori is usually performed on gastric biopsy samples because noninvasive detection directly from feces has low sensitivity. Immunomagnetic separation was used with magnetic beads coated with anti-Helicobacter pylori antibody to concentrate the bacteria from the feces of patients with Helicobacter pylori-positive gastric biopsies. The sensitivity of PCR detection was improved with immunomagnetic separation compared with direct detection from feces. However, only 61% of the fecal samples tested positive for Helicobacter pylori, suggesting that gastric biopsies are still more effective for this diagnosis (303).

(iv) Nasal and throat swab specimens.

Luna et al. (190) have developed a method to safely extract DNA from powder samples suspected of containing Bacillus anthracis spores (described later in this review). This method has also been used to safely extract DNA from nasal swabs that contain Bacillus anthracis spores.

Vaccinia virus is commonly used as a surrogate for measuring the effectiveness of detection tests for variola, the smallpox virus. Donaldson et al. (81) have described a procedure in which throat swab specimens spiked with vaccinia virus were resuspended in buffer and directly detected using an evanescent wave fiber optic biosensor. Vaccinia virus-specific antibodies attached to the biosensor's fiber optic waveguide were able to separate and capture virus particles from other cells and debris in the specimens.

Powders and soil.

During the most recent threat of Bacillus anthracis spores (47), powders were one of the most common nonclinical specimens submitted to designated laboratories (http://www.bt.cdc.gov/lrn/factsheet.asp). A rapid method was developed by Luna et al. (190) to render powder and other environmental samples harmless for safe extraction and identification of DNA. DNA was safely extracted from ≤10 Bacillus anthracis spores and detected using PCR. Samples were prepared for PCR by germinating spores by heat shock (80°C for 2 minutes), followed by sonication (30 minutes) and autoclaving (121°C, 20 minutes). The DNA was then purified using the MagNaPure instrument as previously described in this review.

An evanescent wave fiber optic biosensor method has also been developed to directly test powders for Bacillus anthracis spores (281). Talc-based baby powder, corn starch-based baby powder, confectioner's sugar, baking soda, and Bacillus thuringiensis-based pesticide spiked with Bacillus anthracis spores were successfully processed by this method. Powders were resuspended in buffer and tested directly by the biosensor. Use of unspiked talc-based baby powder as a reference baseline eliminated 100% of potential false-positive readings from powder samples.

Numerous kits have been developed to expedite extraction of total DNA from natural microbial communities within soils and sediments. Three commercially available kits for DNA purification were compared by extracting target DNA from E. coli DH5α that had been seeded into freshwater lake and river sediments. DNA recovery was higher for the two kits (UltraClean Soil DNA Kit, MoBio Inc., Solana, CA, and FastDNA SPIN Kit, Bio 101, Carlsbad, CA) that used mechanical lysis through bead beating. The third kit (SoilMaster DNA Extraction Kit, EpiCentre, Madison, WI) lysed cells with hot detergent and recovered less but a higher quality of DNA (208).

Aerosols.

Particles of two major size distribution groups are present in infectious aerosols: particles >5 μm in diameter and particles 1 to 5 μm in diameter. Particles 1 to 5 μm in diameter behave like gases (61). Larger particles (>5 μm in diameter) settle from the atmosphere and bind to surfaces (82). The smaller 1- to 5-μm particles can be collected by processing large volumes of air and passing the air through a filter or impinging the particles from the collected air into a liquid or semisolid sample (271).

Impaction samplers collect culturable air-borne bacteria and fungi by depositing air-borne particles onto a semisolid agar surface. The semisolid surface then must be cultured in order to detect the presence of collected microorganisms. Although these samplers are capable of collecting small numbers of bacteria (approximately 20 CFU), their usefulness is limited at higher levels of bacteria (>104 CFU) and by sampling time (43). Longer sampling times (>5 min) dry the semisolid collection surface, thereby reducing collection efficiency. Impingement samplers that collect air-borne cells into a liquid can collect particles over longer periods of time and are not limited by the types of analytical methods that can be used for detection of the collected microbes.

Collected air can also be passed through a porous filter. Filtration may desiccate vegetative cells, but the DNA from the cells would be preserved for PCR analysis. Evidence of only the DNA, however, may not be an ideal way to detect a potential biothreat agent because the presence of intact viable cells cannot be confirmed. Filtration collectors currently are used in more than 30 major cities in the United States in the BioWatch biosurveillance program. Filters from these collectors are periodically removed and processed manually for the presence or absence of biothreat agents by participating laboratories. The Department of Homeland Security and the Centers for Disease Control and Prevention's Laboratory Response Network provide early detection of biothreat agents (37).

The Lawrence Livermore National Laboratory (Livermore, CA) has developed a high air volume to low liquid volume aerosol sampler that concentrates air-borne materials from large volumes (2,300 liters/min) of air into a 4-ml liquid sample for subsequent automatic analysis. This sampler, part of the Advanced Pathogen Detection System, has been field tested for viable biothreat agents Bacillus anthracis and Yersinia pestis in a biosafety level 3 facility (197).

Northrop Grumman (Arlington, VA) has incorporated the existing GeneXpert technology into its high-volume Biohazard Detection System for screening mail at U.S. Postal Service facilities. During mail operations, the Biohazard Detection System collects air samples directly above the cancellation equipment and concentrates air samples for 1 hour by absorbing and concentrating air-borne particles into a sterile water base. The fully automated system is a Cepheid GeneXpert module that identifies the presence of Bacillus anthracis spores from air samples. In the fall of 2003, after extensive testing, Northrop Grumman was awarded a production contract to install and manage these systems at U.S. Postal Service sorting centers nationwide (141).

Testing for potential biothreat agents in air is possible only at facilities that contain approved air chambers, such as the U.S. Army Edgewood Chemical Biological Center (Aberdeen Proving Ground, MD) and U.S. Army Dugway Proving Ground, Utah. The Harry Reid Center at the University of Nevada (Las Vegas, NV) contains aerosol chambers and wind tunnels capable of handling items contaminated with chemical, biological, or radiological materials (or surrogates) of interest. Commercial and academic facilities cannot test select agents unless registered with the Centers for Disease Control.

Surfaces.

The Brentwood Mail Processing and Distribution Center in Washington, D.C., was extensively contaminated with Bacillus anthracis spores in 2001 after two letters containing spores were processed at the facility (47). The level of Bacillus anthracis spores obtained by surface wet or dry-swab, wipe, and HEPA vacuum sock sampling methods was measured at the Brentwood facility. The numbers of spores detected using the wipe and HEPA vacuum methods were similar, but the wet-swab and dry-swab methods did not detect spores >33% and 66% of the time, respectively. No single sampling method was found to completely remove spores from sampled surfaces (250). Spores from the Bacillus anthracis surrogate Bacillus atrophaeus sprayed onto different surfaces were used to study the effectiveness of sampling using a swipe (Speci-songe, Nasco, Fort Atkinson, WI), heavy wipe (Handy Wipes Heavy Wipe, First Brands Corp., Danbury, CO), or swab sample processing kit (ASD, Ft. Lauderdale, FL). The three sampling methods performed equally well, but no method completely recovered all spores (42). Different swab materials were also found to affect the recovery of Bacillus anthracis spores from steel surfaces. Premoistened macrofoam and cotton swabs that had been mixed 2 minutes recovered the greatest percentage (>40%) of spores (247).

The Microbial-Vac (M-Vac, Microbial-Vac Systems, Inc., Jerome, ID) uses vacuum and centrifugation to sample and concentrate microorganisms from fruit and vegetable, carcass, and equipment surfaces. A study compared recovery of Salmonella enterica from cantaloupe rind surfaces using M-Vac and by resuspending rinds in peptone water containing Tween 20 and massaging the rinds in solution by hand or by stomacher (31). M-Vac was the only method that could recover S. enterica (6% recovery) from the rind surface when the surface was seeded with a low quantity of the bacterium (2.0 × 102 CFU/100 cm2).

COMMERCIAL DETECTION AND IDENTIFICATION OF BIOLOGICAL THREAT AGENTS

Numerous methods for detection and identification of human pathogens have been used for more than a century and thus are well established and well understood (209). These methods rely on cultivation and biochemical assays which can take from 24 h to 1 month to perform in the hands of a well-trained clinical technician. While these methods are reliable, they generally cannot be used in the field or are not capable of providing real-time or nearly real-time detection and identification of biothreat agents in a bioterrorism event. With recent bioterrorism incidences, there is a critical need to develop more rapid, accurate methods to detect and identify biothreat agents.

Much research and money have been invested in commercial development of technologies that can rapidly and accurately detect and identify biothreat agents both in the field and in the clinical laboratory. The following sections cover only those technologies that are currently available in an “off-the-shelf” format. Those products still in development or in beta-testing stages are discussed in a later section of this review. While an attempt has been made to broadly and thoroughly cover each area of detection methodology, the following discussion is by no means an exhaustive accounting of products available on the market. Many products are too new to have been tested and reported upon in the scientific literature; others are proprietarily owned or licensed to the U.S. Government for defense purposes and, thus, are excluded from this review. Detection assay types discussed include those based on biochemistry, antibodies, and nucleic acids.

Manual Biochemical Tests

Local clinics and hospital laboratories are the first facilities likely to encounter biothreat agents as infected or exposed individuals seek medical assistance (263). Recognition of unusual symptoms or culture test results is of utmost importance, while at the same time unnecessary alarm at false positives should be minimized. Widely available microbial identification systems used on a routine basis by clinical laboratories are not designed or optimized for detection of biothreat agents. Thus, screening hierarchies have been established using standard microbiology biochemical tests to reduce the potential for false alarms and rule out biothreat agents as quickly as possible at the clinical laboratory level. For example, testing and presumptive agent identification protocols for Bacillus anthracis, Yersinia pestis, Francisella tularensis, and Brucella spp. can be obtained online at http://www.bt.cdc.gov/labissues/index.asp#testing. These protocols, which outline the steps to determine if a suspicious culture can be determined not to be of significance or must be further examined by additional tests, are summarized in Table 1.

TABLE 1.

Examples of biothreat agents and identification tests for Laboratory Response Network level A laboratories

Screening test or medium Suitability of test for bacterium
B. anthracis Y. pestis F. tularensis Brucella spp.
Gram stain + + + +
Catalase test + + + +
Oxidase test + + +
Urease test + + +
India ink +
Wright-Giemsa stain +
Beta-lactamase +
XV or S. aureusa + +
Motility medium +
5% sheep blood agar + + + +
Chocolate agar + + +
MacConkey agar + + + +
Phenyl ethyl alcohol agar +
Tryptic soy broth +
Thioglycollate broth + +
Brain heart infusion agar +
Eosin methylene blue agarb + +
Thayer-Martin agarc + +
a

XV factors or S. aureus for satellite growth.

b

Alternative to MacConkey agar.

c

Alternative to chocolate agar.

Automated Biochemical Tests

Most commercial biochemistry-based clinical microbe identification systems are not developed specifically for identification of biothreat agents. However, some commercial systems that utilize pattern recognition databases or libraries have recently begun to offer biothreat update packages. The patterns generated by these types of systems typically are based either on the bacterium's ability to metabolize specified compounds or on gas-liquid chromatography of cellular components.

Substrate utilization patterns.

Several automated identification systems such as VITEK (bioMérieux, Hazelwood, MO) and MicroLog (BiOLOG, Hayward CA) utilize pattern recognition or “fingerprinting” systems based on metabolized substrates and carbon sources or susceptibility to antimicrobial agents. A turbidometrically controlled aliquot of a pure bacterial suspension is added to analysis or identification cards or plates. These cards or plates contain numerous wells with different types of biochemical substrates (usually proprietary) and can be purchased in many different configurations depending on the expected needs of the laboratory. The inoculated cards or plates are then incubated for a set amount of time and read using an automated system. A similar system offered by bioMérieux is the API Series, which consists of biochemical tests in a strip format that, after inoculation and incubation, are processed and read manually. These systems have been used to identify biothreat agents such as Bacillus anthracis (20), Yersinia spp. (16, 183), Vibrio cholerae (176), and other pathogens (219). BiOLOG has recently introduced a “dangerous pathogen” supplement (DP Database) to its MicroLog system, although no published evaluations of the system are available currently.

Fatty acid profile.

The Microbial Identification System (MIDI Inc., Newark, DE) converts cellular fatty acids from pure cultures of bacteria to fatty acid methyl esters and uses a gas chromatograph for separation and identification. Samples are harvested, saponified, and methylated and then extracted and washed. The resulting organic phase is used for identification. Pattern recognition software is used to identify isolates from the chromatograph reading. This method has been used to identify and differentiate Bacillus spores (266), Burkholderia spp. (136, 268), Francisella spp. (27, 57), and Yersinia spp. (172). MIDI Inc. recently introduced the Sherlock Bioterrorism Library that can be added to its identification system to specifically target biothreat agents and challenge organisms. The MIDI Sherlock system containing the MIDI BIOTER database version 2.0 has been been awarded AOAC Official Methods of Analysis status for confirmatory identification of Bacillus anthracis (13).

Each of these biochemical methods is inherently time-consuming as each requires pure isolates for the tests to be accurate. In addition, well-trained, experienced technicians are required to accurately identify and subsequently handle further testing that may be necessary. While automated systems are initially expensive, once a pure isolate has been obtained, the benefits of such systems include the ability to perform multiple biochemical tests simultaneously, the speed of obtaining results after minimal incubation times, and minimal advanced technician interaction. These systems generally are more convenient (i.e., plug in sample, obtain print out) than manual procedures but typically cannot be used in the field and their usefulness may limited by the extent of their databases, which must be updated on a regular basis.

Immunological Detection Devices

Since its inception, the immunoassay has increasingly been used and developed for detection of infectious diseases, drugs, toxins, and contaminants in the medical, pharmaceutical, and food industries. In addition, immunological detection has been successfully employed for detection of biothreat agents such as bacterial cells, spores, viruses, and toxins based on the concept that any compound capable of triggering an immune response can be targeted as an antigen. The various types of immunoassays used for biothreat detection have been reviewed previously (12, 137, 230, 231).

Andreotti et al. (12) summarized the basic principles as follows. “Immunoassays rely upon four basic components regardless of the application and underlying technology: (i) the antigen to be detected; (ii) the antibody or antiserum used for detection; (iii) the method to separate bound antigen and antibody complexes from unbound reactants; and (iv) the detection method. The efficacy of any given immunoassay is dependent on two major factors: the efficiency of antigen-antibody complex formation and the ability to detect these complexes.”

Typically, immunoassays generally test for only one analyte per assay. This limitation means that multiple simultaneous or sequential assays must be performed to detect more than one analyte in a sample or specimen. Advances in assay design and in matrix format have resulted in development of multiplex assays that can be performed on multiple samples simultaneously by automated systems. However, the specificity of immunoassays is limited by antibody quality, and sensitivity (detection limits ∼105 CFU) is typically lower than with PCR and other DNA-based assays. As improvements are made in antibody quality (e.g., production of antibodies from recombinant libraries) and in assay parameters, it may be possible to increase immunoassay sensitivity and specificity.

Many different immunoassay formats are currently commercially available for a wide variety of detection needs (Table 2). Many formats are similar to, or derived from, the classic sandwich assay based on the enzyme-linked immunosorbent assay (ELISA) design (209).

TABLE 2.

Commercially available biothreat detection systemsa

Type Test format Test name Manufacturer Targetsb Referencesc
Biochemical FAME-GC Sherlock Bioterrorism Library MIDI, Inc. B. anthracis, Y. pestis, Brucella spp., F. tularensis, B. mallei, B. pseudomallei 27, 57, 136, 152, 172, 266, 268
Substrate utilization MicroLog Dangerous Pathogen Database BiOLOG B. anthracis, Y. pestis, Brucella spp., F. tularensis, B. mallei, B. pseudomallei 20, 219
Vitek bioMérieux Y. pestis, V. cholerae, C. botulinum 176, 183, 219
API Series bioMérieux B. anthracis, C. botulinum, Y. pestis, Salmonella spp. 16, 188, 258, 309
Antibody Microarray NanoChip Nanogen Pathogens, SEB, V. cholerae toxin B 89
ELISA xMap Technology Luminex Pathogens 29, 30, 123, 196
BV Technology BioVeris Corp. B. anthracis, E. coli O157:H7, Listeria spp., Salmonella spp., C. botulinum toxins A, B, E, & F, ricin, SEB 100, 154, 260, 318, 319
QuickELISA B. anthracis-PA kit Immunetics Inc. B. anthracis NA
RAPTOR Research International B. anthracis, F. tularensis, 9-11, 44, 180
Y. pestis, Brucella spp., V. cholerae, ricin, SEB, E. coli O157:H7, Salmonella spp.
Bio-Detector Smiths Assorted organisms NA
Smart Tickets Instant Check EY Laboratories Ricin NA
BioThreat Alert Tetracore B. anthracis, F. tularensis, Y. pestis, C. botulinum, Brucella spp., ricin, SEB 157
Redline Alert Tetracore B. anthracis NA
SMART-II New Horizons Diagnostic, Inc. B. anthracis, F. tularensis, Y. pestis, C. botulinum, V. cholerae, ricin, SEB, E. coli O157:H7, Salmonella spp. 157
Biowarfare Agent Detection Device (BADD) Osborne Scientific B. anthracis, C. botulinum, ricin 157
RAMP Response Biomedical Corp. B. anthracis, C. botulinum, Y. pestis, ricin, smallpox NA
HHA & HHMA ANP Tech B. anthracis, C. botulinum, Y. pestis, smallpox NA
TRF DELFIA Perkin-Elmer Pathogens 229, 320
DNA Q-PCR GeneXpert/Smart Cycler Cepheid B. anthracis 126, 134, 167, 193
RAPID Idaho Technology B. anthracis, F. tularensis, Y. pestis, C. botulinum, E. coli O157:H7, Salmonella spp., Brucella spp., Listeria spp. 121, 296
LightCycler detection kit Roche B. anthracis 23, 126, 134, 167, 193
BioSeeq Smiths Detection Pathogens 83
B. anthracis PCR kit Takara Mirus Bio B. anthracis 193
O157:H7; StxI; StxII Applied Biosystems Inc. E. coli O157:H7 64
RealArt PCR kits Artus B. anthracis, Salmonella spp. dengue virus orthopox virus, other viruses NA
PathAlert detection system Invitrogen B. anthracis, Y. pestis, F. tularensis, smallpox NA
Certified Lux primer set Invitrogen B. anthracis, Y. pestis, F. tularensis, smallpox, C. botulinum NA
Other Mass spectrometry TEEMmate JEOL Spores 28
Bioluminescence Profile-1 New Horizon Diagnostic, Inc. Pathogens plus spores 62, 174, 273
a

Abbreviations: FAME, fatty acid methylester; GC, gas chromatography; SEB, staphylococcal enterotoxin B; NA, not available to public; HHA, handheld assay; HHMA, handheld microarray assay.

b

Examples of targets detected, not a complete accounting.

c

Peer-reviewed publications accessible to the public.

Commercial versions of the sandwich method have expanded to include newer innovations of biotechnology, principally in detection capabilities. Assays can be performed with a variety of substrates and labels (fluorescent, chemiluminescent, and electrochemiluminescent), as well as on multiple platform types (biosensors, flow cytometry, microarray, and lateral flow diffusion devices). Many systems available today use different combinations of these advances. However, there are two main categories of immunoassays used with a few adaptations, mainly variations of solid-support models and lateral flow diffusion apparatuses, also known as “smart ticket” technology or hand-held assay devices.

Solid-support platforms. (i) Luminex xMAP.

Luminex xMAP (Luminex Corp, Austin, TX) technology uses the basic “sandwich” assay format, but the capture antibody is coated onto the surface of a polystyrene bead rather than in a microwell plate. These beads are processed through the assay and separated for analysis via flow cytometry (298). The beads (microspheres) are spectrally unique and color coded into different sets that can be differentiated by the Luminex100 analyzer. As each type of bead can be labeled with different antibodies, this labeling enables multiple analyses to be simultaneously performed in the same well. The beads are maintained in solution throughout the assay, thereby permitting liquid phase binding of target and separation spectrally by a dual laser detection system. The laser detects excitation of internal bead dyes and reporter dyes of any captured targets. The signal intensity is proportional to the amount of target analyte in the sample. The xMAP technology has been incorporated into continuous environmental monitoring systems (123, 196) and for detection of exposure to biothreat agents in humans (29, 30).

(ii) BV technology.

The BV M-Series instrument (BioVeris Corp., Gaithersburg, MD) (formerly ORIGIN-Igen Corp.) uses electrochemiluminescence to detect reporter molecules used in the sandwich assay. BV assays use paramagnetic beads as the support structure for the capture antibodies. After the target is captured, the reporter antibodies labeled with BV-TAG, Ru(bpy)32+, are passed through the system and any target analyte captured on the beads is labeled. The beads are passed via a flow cell over a magnet on the surface of an electrode. The beads are captured by the magnet and are thus separated from unbound label and remaining matrix components. An electrical potential is applied to the electrode that excites the BV-TAGs, which then emit light that can be detected and processed. This technology has been mainly used in clinical settings, but is being applied to detection of biothreat agents such as E. coli O157 (260, 318, 319), Bacillus spores (100, 318), Yersinia spp. (318), S. enterica serovar Typhimurium (319), and toxins and toxoids (100). Numerous biothreat agent assays have been developed and many are available as kits (Table 2).

(iii) Bio-Detector.

The Bio-Detector (Smiths Detection, Edgewood, MD) uses ELISA principles on a tape format in a portable rugged housing. Liquid samples are injected and separated for detection of different biothreat agents. During the reaction, each subsample is mixed with biotin-labeled and fluorescein-labeled antibodies as well as with streptavidin, which attaches to the biotin label. After the labeled antibodies have attached to the targets within the subsample, they are filtered and captured by biotin-coated biotape in different locations for each subsample. The streptavidin acts as a bridge between the biotin-tagged targets and the biotin-coated tape. A solution of antifluorescein antibody, conjugated to the enzyme urease, is filtered through the tape and tags the bound target. The tape is then positioned over the sensor, where it is covered in a solution of urea. If the tape has trapped a targeted agent, the urease reacts with the urea and causes a change in pH. The rate of change is proportional to the amount of target present. The sensor processes the signal and determines the presence and quantity of target agent present.

(iv) DELFIA.

The DELFIA (dissociation-enhanced lanthanide fluorescence immunoassay) (Perkin-Elmer Life Sciences, Akron, OH) system is one example of a format based on time-resolved fluorescence. This technique relies on lanthanide chelate labels that have long fluorescence decay times, which allow measurements of fluorescence without interference from background signals. Typically, this chemistry is incorporated into standard ELISA microplate assay format (detection antibodies are labeled with the lanthanide) and, after the reaction has occurred, the label is disassociated from the detection antibodies using a low-pH enhancement solution. The free molecules rapidly form new stable highly fluorescent chelates, which can be read by the system (12, 230). The DELFIA system has been used to detect Francisella tularensis, Clostridium botulinum toxin, staphylococcal enterotoxin B and E. coli O157:H7 (229, 320).

Lateral flow platforms.

Lateral flow devices have been principally developed for rapid field assay formats but are quickly becoming incorporated in clinical laboratory settings. These tests are usually single-use, disposable cartridge tests in the form of either immunochromatographic line assays or enzyme immunoassays generating detectable colored end products which give a yes or no answer. Basically, these assays use antibodies mounted on a paper strip or membrane as the capture antibody and use capillary flow to move colloidal gold or colored microparticle-labeled antigen detection antibody complexes in the fluid phase toward the capture antibody. A positive result is obtained from the capture of the labeled antigen-antibody complex with a second immobilized antispecies antibody (typically an anti-immunoglobulin G for the detection antibody host) and the formation of a line or pattern in the appropriate result window (209). Also incorporated are control reagents which result in either in a contrasting design or pattern appearance in a separate window for a negative result depending on the format. The negative result (control) is also an indicator that the test worked correctly and the sample was passed through the testing field. These devices are commonly referred to as dipstick tests, and the public is most familiar with their use as off-the-shelf pregnancy test kits.

Although lateral flow device assays are easier to perform and more rapid than classic instrument-based immunoassays, they typically are not as sensitive and can have the potential for higher rates of false positives. However, such devices can be useful in rapid preliminary screening of samples for a biothreat agent, and any positives should be followed by confirmatory tests such as PCR.

Lateral flow devices have been developed by many companies for several biothreat agents, including Bacillus anthracis, Francisella tularensis, Yersinia pestis, Clostridium botulinum, and several toxins such as ricin and staphylococcal enterotoxin B (Table 2). Although lateral flow devices have been available for use in the field for some time, there is little information available about their performance other than company literature, promotional items, and news reviews. With the exception of one paper by King et al. (157), no peer-reviewed published data are available on these kits. However, the U.S. Food and Drug Administration recently approved two test kits (Redline Alert, Tetracore Inc., Gaithersburg, MD; and RAMP, Response Biomedical Corp., Burnaby, British Columbia, Canada) for use with pure culture isolates to test for Bacillus anthracis in the clinical setting (Redline: http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRL/listing.cfm?&ID-90045; RAMP: http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRL/listing.cfm?&ID=79402) In addition, the RAMP system was also awarded AOAC Official Methods status (13, 111). These approvals may lead to more thorough and descriptive testing and reporting of these types of systems.

Nucleic Acid Detection via Quantitative PCR

Quantitative real-time PCR (Q-PCR) combines PCR amplification with simultaneous detection of amplified products based on changes in reporter fluorescence proportional to the increase in product (118, 187). There are two formats of Q-PCR, nonspecific detection and specific detection. Nonspecific detection uses DNA-intercalating dyes that fluoresce when bound to DNA (e.g., SYBR green) (Molecular Probes Inc., Eugene, OR). As the DNA is amplified, the dye is intercalated into the product. After amplification, a melting curve is performed and, as the DNA dissociates, the fluorescent signal decreases proportionally. This format is useful for optimizing PCR conditions and checking specificity of primer designs and is cost efficient compared to specific detection formats. For specific detection, the change in fluorescence relies on the use of dual-labeled fluorogenic probes containing both a reporter fluorescent dye and a quencher dye. An increase in fluorescence indicates that the probe has hybridized to the target DNA and the quencher dye is no longer able to mask the signal of the fluorescent dye. Q-PCR is used for a variety of applications from quantitative presence/absence tests and rapid confirmation tests to monitoring gene expression (the most common use).

The main Q-PCR format used for biothreat agents is specific target detection, and a wide variety of primer and probe combinations are available from many companies in a multitude of configurations. Many of these specific target configurations rely on mechanistic variants in the primer/probe construction and combinations which can include TaqMan probes (double-dye probes) (127), locked nucleic acid probes (161, 218), black hole quenchers (146), molecular beacons (287), lux primers (213), and scorpion primers (306). Each type has a slightly different method of separating the fluorophore from the quencher for reporting the amplification process, and all are available commercially and can be customized.

As the fluorescent reporter dyes attached to these primer and probe combinations can have different excitation/emission spectra, Q-PCR can be utilized to detect several targets simultaneously using different reporter dyes for different targets. Some of the more common reporter dyes include 6-carboxyfluorescein, 6-carboxytetramethylrhodamine, Cy5, Cy3, Rox, Texas Red, rhodamine, fluorescein, and Oregon Green. Quenchers have the ability to quench throughout their absorption spectrum, but optimally have their maximum absorption matched closely with the reporter dye emission for the best performance.

Typically the Q-PCR thermocyclers incorporate intricate software to monitor the progression of the reactions. At the point when the product signal is detected above the background (Ct value) and in the exponential phase, the software can quantitate and compare the signal to standards. The more abundant the target, the faster it is detected above background and the lower the Ct value. This Ct value can be either qualitative or quantitative, depending on how the standards are set up and based on the needs of the user. As technology progresses, these instruments are becoming smaller, faster and more sensitive, thus making them highly appealing as rapid detection methods for biothreat agents.

However, accurate characterization or identification of bacteria by Q-PCR is limited by the same bias and variations that are inherent in many nucleic acid techniques. The main concerns are biased nucleic acid extraction (e.g., efficiency of extraction or cell lysis if using whole-cell methods), degradation of nucleic acids by nucleases, probe and primer reactivity (i.e., sensitivity, specificity, accessibility, and quantitation), and inherent PCR bias (e.g., variances in polymerase, buffer, and thermocycler performances). The ability to either extract the DNA or rupture the cells or spores for accessibility significantly influences the sensitivity, reproducibility and accuracy of any PCR biothreat agent detection method. Additionally, the presence of inhibitors such as humic acids or chelating agents can interfere with target sites of the probes and primers, thereby resulting in false negatives. Care must be taken to remove these inhibitory compounds prior to analysis.

In spite of the limitations, PCR-based analysis can be highly specific and sensitive for the target of interest if the number of cells present is at or above the detection limits of the particular assay (typically 10 to 100 cells). Use of Q-PCR to obtain rapid quantitative estimates for biothreat agent presence is an invaluable asset. The new advances in size reduction and speed of thermocycling enable these units to be used both as portable and as laboratory-based platforms.

Q-PCR thermocycler platforms.

While many Q-PCR technologies are available, a few have had more focused development towards pathogen and biothreat agent detection. Most of these systems are sophisticated molecular-based platforms that perform best in a laboratory setting such as hospitals, monitoring offices, and research laboratories. The GeneXpert Q-PCR DNA detection system (Cepheid, Sunnyvale, CA) incorporates the standard Q-PCR features of the Smart Cycler system (Cepheid) with disposable automated testing cartridges that can process “dirty” samples and perform amplifications followed by product detection (22). This automated process removes the human variable and speeds the processing time by eliminating the need to extract DNA from samples prior to analysis. A GeneXpert test cartridge specific for Bacillus anthracis is available, and other targets are in development.

The LightCycler system (Roche Diagnostics GmbH, Mannheim, Germany) also has assays specifically developed for biothreat agents and a LightCycler-Bacillus anthracis detection kit (Roche Applied Science) is commercially available (23, 288). The LightCycler system utilizes a disposable glass capillary system into which extracted DNA is placed and analyzed, instead of a cartridge system (310). Applied Biosystems (ABI, Foster City, CA) offers several 96-well plate format Q-PCR systems, including the 7300 and the 7500 real-time PCR systems and the 7000 and the 7900 ABI Prism sequence detection systems. ABI also sells TaqMan-based detection kits for pathogens (e.g., Taqman E. coli O157:H7 and E. coli Stx1 and Stx2) (64). Numerous Q-PCR assays have been developed and validated for several biothreat agents (e.g., Bacillus anthracis and variola virus), and many have had their conditions optimized for use on these different platform types (126, 134, 139, 167).

While these Q-PCR instruments are useful in a laboratory environment, recent bioterrorism events have underscored the need for fieldable Q-PCR platforms. The Ruggedized Advanced Pathogen Identification Device (RAPID) (Idaho Technology, Salt Lake City, UT) is an automated, portable Q-PCR platform that integrates the LightCycler technology in a compact unit. This instrument was designed for use by military field hospitals and first responders and in other rugged environments that require a portable impact-resistant platform. Many biothreat agent and pathogen detection kits are available in freeze-dried format for the RAPID (121, 296). Another portable instrument is BioSeeq (Smiths Detection, Edgewood, MD), a commercial version of the handheld advanced nucleic acid analyzer by Lawrence Livermore National Laboratory. This unit has a small, rugged design with features that make it possible to operate with the heavy gloves of a hazmat suit and was designed for military field use. BioSeeq uses either a prepared PCR cartridge and preset protocols to analyze samples in the field or individual samples with programmable protocols in a laboratory mode for customized assay designs (83).

PCR reagent kits for biothreat agents.

There are innumerable sets of probes and primers in the literature databases that can be used with any of a number of pathogens or biothreat agents. Recently, however, several companies have begun to offer PCR kits in various formats for detection of biothreat agents (Table 2). These kits eliminate the need for extensive primer/probe design and facilitate rapid detection and monitoring programs. The kits often come with prepared controls and only lack the sample DNA to be tested. The removal of inadvertent investigator influence (e.g., pipeting or calculation errors and cross-contamination issues) on these systems results in simplified processing and reproducible data. One kit, PathAlert (Invitrogen Corp., Frederick, MD) (Table 2), can be coupled with the Agilent 2100 Bioanalyzer (Agilent Technologies, Wilmington, DE) for an automated multiplex assay in a lab-on-a-chip format. PCR products produced with the Q-PCR kit are passed through the Bioanalyzer and identified.

Bioluminescence Detection

Bioluminescence has been used widely in clinical, food, and environmental settings for monitoring incidences of bacterial contamination. The most common example of bioluminescence is the luciferin-luciferase reaction in the presence of ATP. The basic principle of ATP bioluminescence monitoring is that the amount of ATP in a sample correlates proportionally to the biomass. Since ATP is found in all living cells, steps must be taken during sample processing to eliminate ATP contamination from nontarget sources (e.g., plant cells, mammalian cells, and nontarget bacteria). Depending on the selected target, different strengths of detergents are used to lyse nontarget cells (e.g., mold, pollen, somatic cells) in the sample, so that only the intended cells are monitored (62).

The Profile-1 hand-held system (New Horizon Diagnostics Inc., Columbia, MD) uses a microluminometer to read sample bioluminescence and a Filtravette (New Horizon Diagnostics Inc.) sample processing unit to remove nonbacterial ATP sources. This combined system has been used to examine contamination on meat carcasses (62), test biological aerosols (273) and detect Bacillus species spores in powders (174). The main limitation to this method is that it is nonspecific for biothreat agents. ATP monitoring can report the presence and relative abundance of bacteria or germinating spores in a sample but cannot identify the type of bacteria. Processing samples via this method only indicates if ATP-producing microorganisms are present; further processing of the sample by other methods is needed to determine whether or not a serious threat is present. While this type of test may be well suited to the food industry, where the presence of any bacterial contamination may be important, its usefulness for biothreat agent detection is limited.

DEVELOPING TECHNOLOGIES

Developing sensor technologies for bioterrorism defense can be divided into a few broad groups according to the way they recognize and respond to a biological threat agent. For the purposes of this review, these groups are defined as biochemical, immunological, nucleic acid, cell/tissue, and chemical/physical technologies. Each of these recognition systems must be paired with a transducer that can transform the response into an analyzable signal. Detection platforms that combine a biological recognition system and a physical transducer are termed biosensors. There are a variety of transducers that can be paired with a given biological recognition system to produce an analyzable signal, including electrochemical, optical, mass, thermal, and high frequency (2, 67). Hence, this review will focus on only some of the most active areas of research and advanced biosensor development efforts reported in the literature. Additional information on biosensor development and potential applications can be found in other review articles (67, 103, 113, 137, 138).

Biochemical Detection

Biochemical systems for recognition of biological targets include measurements of products and enzymatic activity associated with microbial metabolism. These methods generally are not as specific as antibody- or nucleic acid-based methods because the targeted product or enzyme may also be present in other organisms.

One example of a technology that detects the metabolic products of organisms is electronic nose devices. This technology uses a transducer, such as a cantilever, acoustic wave, or conducting polymer, which has been coated with a chemical that reacts with specific volatile organic compounds or gases to create a sensing element (67). An array of sensors, each specific for different vapors or gases, can be constructed and used to detect multiple analytes. The technology has been used by researchers to detect and identify volatile organic compounds produced by specific bacteria or fungi (1, 128, 149, 150, 191, 224-226, 252, 286). Electronic nose devices have been used for detection of microbes in food (150, 191) and infections in humans (225, 226). This technology requires complex pattern recognition software to interpret the results. It can be rapid and sensitive but is not very specific because the compounds produced by microorganisms can fluctuate with energy and/or carbon source and with environmental conditions. Furthermore, different organisms can produce similar volatile products.

Conducting polymers may also be used to detect biologically produced chemicals other than volatile organic compounds and could be used to detect products of microbial metabolism such as toxins. Conducting polymers are organic polymers that can conduct electricity. These polymers can be doped with enzymes, antibodies, or other biomolecules that allow them to capture specific types of target biological compounds. Much of the work in this area has concentrated on detection of biological compounds, such as glucose, urea, and cholesterol, making them useful as biochemical detectors. A recent review by Gerard et al. (103) provides detailed information on the nature of these polymers and their potential applications in health, food, and environmental monitoring.

Biosensors that detect the presence of a bacterial toxin based on inhibition of enzyme activity have been reported in the literature. A portable sensor for anatoxin-a, a toxin released by certain species of cyanobacteria, detected inhibition of acetylcholinesterase activity using an electrochemical transducer (299). The sensor performed well; however, it is nonspecific since several other compounds, including common organophosphate and carbamate pesticides, also inhibit acetylcholinesterase activity (74, 256). In an attempt to compensate for the lack of specificity, mutations were induced in acetylcholinesterase and mutant enzymes were screened for both increased sensitivity and specificity to anatoxin-a (74).

β-Galactosidase activity with amperometric detection was used to quantify both E. coli and Klebsiella pneumoniae, members of the fecal coliform group of bacteria, in sterile water (202). Specific detection of E. coli was also accomplished by selective capture of the bacteria on an electrode functionalized with antibodies specific for the bacteria. An alternative method increased specificity by using phage specific for E. coli to lyse the bacteria followed by amperometric detection of β-galactosidase activity (214).

Immunological Detection

Antibody-based and similar affinity probes are used in many of the technologies currently under development. Technologies based on this type of detection utilize the specificity of the immune system to target agents of interest. Shape recognition technologies can be used to detect a range of potential threat agents, including viruses, bacteria, toxins, and bioregulators. A substantial amount of research is concentrated on improving antibody sensitivity and specificity through the generation of recombinant antibodies, antibody fragments, and phage probes. Other types of shape recognition probes are also under investigation, including receptors, aptamers, and peptide ligands. A primary advantage to using affinity probes is the potential to recover live organisms for culture and further study (165, 280). Several reviews have been published that discuss the use of antibodies and other affinity probes and their applications (36, 125, 137, 166, 232).

Antibodies and fragments.

Conventional antibodies are the predominant affinity probe used in shape recognition-based technologies. However, other forms of antibodies, including mono- and divalent antibody fragments such as Fab′ and F(ab′)2 respectively, and single-chain variable regions have also been explored to determine any advantages they might offer in sensitivity, specificity, or durability compared to antibodies (84, 163, 177, 232, 238, 323). Such fragments may also be modified using recombinant technology in an attempt to improve the binding kinetics of the antibody fragments, thus making them more favorable as probes (232). Phage display libraries have improved the development of affinity probes by permitting biopanning of libraries containing thousands of possible peptides for the ones that bind best to a specific antigen. Identified antibodies or fragments can then be chemically synthesized or produced in large quantities in a recombinant host (137, 233), or the phage expressing the antibody or fragment may also be used as the probe (233). Using this methodology, probes have been developed that bind to bacteria, viruses, and toxins, e.g., Clostridium difficile toxin B (73), Brucella melitensis (116), vaccinia virus (253), and botulinum toxin (84). The final products may be isolated or remain attached to the phage (73, 116, 233).

Aptamers and peptide ligands.

Aptamers and peptide ligands are other alternatives to antibodies. Aptamers are small DNA or RNA ligands that recognize a target by shape, not by sequence, and that are generated using combinatorial methods. Aptamers have been used to detect the toxin ricin in a bead-based biochip sensor (159) and cholera toxin, staphylococcal enterotoxin B, and Bacillus anthracis spores in an electrochemiluminescence assay (38, 39). Two variations of aptamers include ribozymes, autocatalytic RNA aptamers that can be engineered to generate a signal after target capture (119), and photoaptamers, DNA aptamers that have been modified to bond covalently to a bound target when exposed to UV light (262). A review by Breaker (33) provides more detailed information on the characteristics and potential applications of aptamers and similar nucleic acid affinity probes.

Short recombinant peptide sequences have also been tested as capture and detection elements in biosensors. As with phage display-generated antibodies and fragments, these peptide ligands may be chemically synthesized or remain as a phage probe. Sequences that bind specifically to ricin (153), Bacillus anthracis and other Bacillus species spores (34, 284, 307), protein A of Staphylococcus aureus (195), and staphylococcal enterotoxin B (108) have been developed and tested. In addition, hemin tethered to glass slides has been used to capture and detect various types of bacteria using intrinsic NADH/NADPH fluorescence (156, 195).

Flow cytometry.

Biosensors that exploit shape-based recognition to capture and detect specific biothreat agents are being developed. One example is the flow cytometer that serves as the detector for the Autonomous Pathogen Detection System (APDS) developed at Lawrence Livermore National Laboratory. This flow cytometer is based on proprietary Luminex technology, in which Luminex color-coded beads are conjugated to antibodies that bind to specific target agents (196, 197). Each differently coded bead is labeled with a target-specific antibody, thereby providing extensive multiplex capabilities. The APDS has been used to detect the threat agent simulants Bacillus globigii, Erwinia herbicola, MS2, and ovalbumin singly and in mixtures (196) and to simultaneously detect of Bacillus anthracis and Yersinia pestis in air (197). Researchers have recently reported a successful integration of PCR with the APDS to produce a completely automated system that will automatically perform confirmatory PCR on any sample that produces a positive immunoassay result (124). During aerosol chamber tests with Bacillus anthracis, Yersinia pestis, Bacillus globigii, and botulinum toxoid, all agents were detected using the automated immunoassay procedure, and the presence of the three bacteria was confirmed using the automated PCR process.

Biochip arrays.

Several methods of detecting targets using biochip technology are under investigation. One approach uses dielectrophoresis to concentrate target agents for identification and then detects them using an electric-field-driven immunoassay (89, 132, 315). Bioparticle separation and enrichment using dielectrophoresis was discussed previously under Sample Matrix Processing. Dielectrophoresis and electric fields are used to direct the assay components and targets to the proper position on the array. The resulting fluorescence is visualized microscopically. The method has been tested with E. coli O157:H7 and Bacillus globigii spores (132, 315) and staphylococcal enterotoxin B and cholera toxin B (89, 315).

A second method uses a sensor array constructed on a complementary metal oxide semiconductor integrated circuit to detect targets on a microarray (207). The optical detector can interrogate all spots on the array simultaneously or individually. A system based on this technology has been designed for detection of Bacillus globigii spores in air by integrating a portable air sampler with an on-chip ELISA (274).

A complementary metal oxide semiconductor platform is also used in an immunoassay microarray that self-assembles oligonucleotide-labeled antibodies onto the chip through matching with their complementary oligonucleotides on the array (77). Detector antibodies labeled with horseradish peroxidase provide amplified electrochemical detection. The system has been tested with ricin, M13 phage, and Bacillus globigii spores.

A nanoscale detector comprised of porous silicon has been developed that can rapidly distinguish gram-positive and gram-negative microorganisms (48). Microcavities in the silicon were functionalized with an antibody alternative, a synthetic organic receptor specific for lipid A. Binding of gram-negative bacteria produced a photoluminescence red shift. No shift was observed with gram-positive bacteria.

Microspheres arranged in cavities micromachined into a silicon wafer constitute another version of a biochip. The micropheres are functionalized with antibodies or aptamers directed to specific targets. This “taste chip” technology has been tested for its ability to detect several proteins, including ricin (109, 159).

Surface plasmon resonance-based biosensors.

Biosensors based on surface plasmon resonance are also being studied for use as detectors of biothreat agents. These sensors directly detect target analytes by measuring the refractive index changes that occur when a target binds to the surface of a metal-coated (generally gold or silver) surface (200, 203). Commercial instruments are large, but a small unit for field applications has been developed (Spreeta, Texas Instruments, Attleboro, MA) and is being tested for detection of staphylococcal enterotoxin B (211, 313). The miniature gold-coated sensor uses a gold-binding peptide to functionalize the surface for a specific target and tracks the angle of reflectance at a fixed wavelength to produce a signal. It is temperature controlled and has two channels, with one channel serving as a control channel. Another miniature surface plasmon resonance device uses a side-polished, single-mode optical fiber to monitor spectral changes at a fixed angle of incidence (261). This device was used to directly detect staphylococcal enterotoxin B in milk. Sensitivity was increased when a secondary antibody was used to amplify the signal (129).

Evanescent-wave biosensors.

Evanescent-wave excitation of fluorophore-labeled antibodies attached to targets captured on a waveguide surface is another method of using biosensors for detection (200, 203). Several of these devices are being developed and tested for detection of biothreat agents. The RAPTOR (Research International, Monroe, WA) is a rugged biosensor that was developed by the Naval Research Laboratory (Washington, DC) (9-11, 158, 180, 212). For this format, the capture antibodies are coated onto polystyrene fiber optic waveguides encased in a coupon that can be prepared in advance. Samples are passed over the waveguides, and target analytes are captured by the antibodies. Fluorophore-labeled reporter antibodies are then passed through the coupon and bind to the captured target analytes on the waveguides. Fluorescent reporter molecules within 100 to 1,000 nm of the waveguide surface are excited by the evanescent field of the laser, and a portion of the emission energy recouples into the fiber and is quantified by the photodiode. Increases in fluorescence are proportional in rate and magnitude to the target analyte concentration and are recorded as pAmp signals.

The RAPTOR biosensor and its predecessor, the Analyte 2000 (Research International), have been used to develop biothreat agent assays in many different complex matrices. These include foods (69-71, 102, 164, 180, 181), river and spent sprout irrigation waters containing potentially interfering organic compounds and microbes (164, 212), powders (281), and clinical specimens (44, 81, 278). Furthermore, methods have been developed to recover live bacteria from the waveguides used in RAPTOR/Analyte 2000 assays to confirm viability and perform PCR and other confirmatory tests (165, 280).

The Naval Research Laboratory has also developed an array biosensor and an integrating waveguide biosensor. In the array biosensor, target analytes are captured by antibodies on a patterned glass slide. Fluorescent reporter antibodies are added and, after evanescent wave excitation by a laser, images are recorded by a charge-coupled device camera (68, 179, 276, 277). The array biosensor has been tested with target analytes in different complex matrices, including toxins in clinical fluids, environmental samples and food matrices (179), and bacteria in food (251, 277). The assay process has been automated with development of a small fluidics module that can be coupled to the array detector (79). The integrating waveguide biosensor uses a glass capillary tube as the waveguide and has been used to detect staphylococcal enterotoxin B at pg/ml levels (178).

Cantilever and acoustic wave.

Cantilever technology uses changes in mass to detect target agents trapped on the device surface. Target specificity is achieved by adsorbing or attaching some type of capture molecule to the surface substrate. As target is captured, the added mass results in a frequency shift from the normal oscillation of the cantilever. An additional shift in resonance frequency is observed upon capture of the target (113). A cantilever biosensor was constructed and used to detect E. coli O157:H7 following immersion of the cantilevers in suspensions containing 106 to 109 cells/ml (135). No frequency shifts were observed when buffer alone or buffer containing Salmonella enterica serovar Typhimurium was incubated with the cantilevers.

A magnetoelastic cantilever immunosensor was developed that used a magnetic field to induce oscillation of the sensor (248, 249). The sensor surface was coated with antibodies to permit specific capture of the desired target agent. Following agent capture, alkaline phosphatase-labeled antibodies to the target were added to amplify the signal by increasing the total mass on the sensor. The sensor was tested with both E. coli O157:H7 and staphylococcal enterotoxin B with reported sensitivities of 102 cells/ml and 0.5 ng/ml, respectively.

Impedance analysis of an oscillating quartz crystal was used as the basis of a piezoelectric biosensor that was tested with Salmonella enterica serovar Typhimurium (155). Antibodies coated onto paramagnetic microspheres provided specificity and increased the response of the sensor. The detection limit was reported as 103 cells/ml.

A device based on surface acoustic wave technology is under development and has been tested using spores of Bacillus thuringiensis as a surrogate for Bacillus anthracis spores (32). The use of shear horizontal waves permitted acoustic measurements in a liquid environment and preparation of the surface with a monoclonal antibody provided specific capture of the target. Detection of Bacillus thuringiensis spores at or below Bacillus anthracis spore levels for human infection via inhalation was reported.

Quantum dots and upconverting phosphors.

Quantum dots are inorganic fluorescent nanocrystals that are color-tunable by varying the size and composition of the crystal core. They have a wide absorption spectrum and a narrow emission peak, which makes them ideal for multiplexing. In addition, quantum dots are highly photostable and have a high quantum yield. These novel reporters and their potential use as labels in biological assays are discussed in more detail by Chan et al. (49) and Riegler and Nann (242). Quantum dots have been used as tags to replace more traditional fluorescent dyes such as fluorescein and rhodamine. Assays to detect E. coli O157:H7 (275), Cryptosporidium and Giardia spp. (175, 324), and a multiplexed assay to detect cholera toxin, ricin, shiga toxin 1 (called shiga-like toxin 1 or SLT-1 by the authors), and staphylococcal enterotoxin B (107) have been developed using quantum dot-labeled antibodies.

Upconverting phosphor technology provides another labeling replacement for fluorescent dyes and utilizes unique submicron-sized ceramic particles containing a rare earth element (114, 216). These unusual labels absorb more than one photon of low energy (in the infrared) to achieve a higher energy state that is emitted as phosphorescence when the crystal returns to its ground state. Since this process does not occur naturally, there should be no background when these reporters are used as labels. Hampl et al. (114) attached upconverting phosphors to antibodies and used them in both lateral flow and plate assay formats to detect human chorionic gonadotropin and in a duplex assay to detect both human chorionic gonadotropin and ovalbumin simultaneously. Niedbala et al. (216) used upconverting phosphor technology in a lateral flow sandwich assay format. These researchers were able to detect 103 CFU/ml E. coli O157:H7 spiked into an enrichment medium containing 109 organisms/ml grown from a ground beef inoculum. This enriched medium served as a negative control background for the assays. Upconverting phosphor technology has also been tested in nucleic acid-based assays (60, 294, 325).

Nucleic Acid Detection

Nucleic acid-based detection uses the specificity of base pair matching to detect and identify biothreat agents. Any biothreat agent that contains DNA or RNA can be detected using nucleic acid-based detection methods; however, a limitation of these types of sensors is that they are not able to detect protein threat agents such as toxins and bioregulators. Many of the nucleic acid approaches for detection of biothreat agents are described in recent review articles (137, 139, 182, 269). Some of these development efforts focus on improving the speed, portability and simplicity of PCR while maintaining the sensitivity and specificity. Additional approaches also investigate methods to detect nucleic acids from target agents of interest using isothermal amplification or directly from samples without using an amplification step.

Amplification methods.

A hand-held real-time thermal cycler has been developed that can rapidly analyze up to four prepared samples (122). This platform has been tested successfully with Bacillus anthracis cultured from swab samples taken after a recent bioterrorism incident and E. coli cultured from streamwater samples. SYBR green dye was used as the label and positive results were observed in 13 to 32 minutes. The appearance of background fluorescence was noted as a problem in the E. coli assays. Erwinia herbicola from an overnight culture was detected in approximately 7 minutes using a TaqMan probe and 500 cells as a template.

NASBA relies on the isothermal amplification of single-stranded RNA for detection of target organisms. In this method, a primer binds to the target RNA sequence and a reverse transcriptase produces a cDNA strand. RNase digests the template RNA and a second primer binds to the cDNA, which the reverse transcriptase uses to form a double-stranded cDNA. T7 RNA polymerase is then used to produce RNA transcripts via an amplification process. The uses, advantages, and limitations of NASBA have been discussed in reviews of the technique (58, 66). Assays have been developed and tested for several pathogenic microorganisms, including viruses (46, 120, 144, 145, 168, 204, 321), bacteria (18, 19, 59, 245), fungi (317), and protozoans (255). Successful detection of microorganisms in both environmental (18, 145, 168, 245) and clinical (120, 168, 204, 255, 317) samples has been reported. Data from these investigations indicate that NASBA is a sensitive, specific, and rapid analysis method. The method may also be useful for detecting viable organisms when mRNA is used as the template (58).

Loop-mediated isothermal amplification is a method of isothermally amplifying DNA using a novel strand displacement approach (217). The technique uses DNA polymerase and four specially designed primers that are specific for sequences on the sense and antisense strands of the target DNA. Loop-mediated isothermal amplification-based assays have been used to detect several viruses (222, 234, 235, 279) and bacteria (86, 130, 140, 194, 257, 265) and at least one fungus (85). Successful detection has been reported in tests involving several clinical samples (85, 140, 234, 235, 257, 279).

Microarrays.

A method of identifying bacteria using rRNA was tested using E. coli in a mix containing Bordetella bronchiseptica (101). After single-stranded DNA capture of rRNA from the bacteria on a self-assembled monolayer, the bacteria were tagged with another single-stranded DNA probe labeled with fluorescein. An antifluorescein antibody labeled with peroxidase was then used to amplify the signal, followed by amperometric detection of peroxidase activity. The authors reported detection at approximately 103 E. coli cells with this method.

Many of the biochip technologies discussed under Immunological Detection have also been tested for detection of nucleic acids. The electric-field-driven assay described for immunological detection has also been used to detect bacterial DNA. Bacteria concentrated by dielectrophoresis are lysed and the DNA is denatured in a heated denaturation module. The released DNA is amplified using strand displacement amplification (304, 315) and is analyzed using an on-chip electric-field-driven hybridization assay. This electric-field-driven assay system has been used to amplify and discriminate six gene sequences from six different bacteria (304).

Another antibody-based detection chip technology that has been adapted for detection of DNA uses capillary array electrophoresis and laser-induced fluorescence to detect PCR-amplified DNA (264). DNA from an enterotoxigenic strain of E. coli was used to test chip performance. The results indicated that the method was comparable to gel analysis for identifying specific amplicons and offered the advantages of greater speed and multiplexing capabilities on a microarray platform.

The “taste chip” technology used in antibody-based biochips has also been used to rapidly distinguish a mixture of similar 18-base DNA oligonucleotides (6). Microbeads labeled with DNA probes were placed in micromachined cavities on a silicon chip and samples containing target DNA were recirculated through the cavities. Rapid identification of single nucleotide mismatches and a limit of detection of about 10−13 M using this method were reported.

A nucleic acid-based biochip for detection of cells has been developed and tested with E. coli K-12 cells (186). This biochip incorporates sample capture, preparation, PCR, hybridization, and electrochemical detection on a totally self-contained chip. The authors claimed positive recognition at 103 cells but with a very low signal-to-noise ratio.

Tissue- and Cell-Based Detection

Tissue- or cell-based detection systems use the intrinsic response of a specific cell type to a potentially toxic or infectious foreign substance to identify a biothreat agent. In these devices, the cells constituting the sensor produce an action potential signal that can be measured by an electrode or optical detector (162). The detector cells may originate from a specific unicellular organism or a tissue type, such as nerve or heart cells, and may be either primary or immortalized. Each has its advantages and disadvantages, which are reviewed by Kovacs (162). Potential applications of this approach for the detection of biothreat agents have been described (162, 270). In some instances, whole organisms may be used as the detector, but these are not generally useful in a clinical setting and will not be discussed in this review.

Biosensors based on cells from sources such as neurological and cardiac tissues are being explored for detection of harmful substances including biological threat agents (110, 162, 220, 270). One example of such a cell-based biosensor that has been tested for field use incorporates cardiomyocytes that have been genetically engineered to more selectively respond to specific functional activity (15, 65, 105). An automated biosensor based on neuronal cells has also been tested and was found to respond to the biological toxins tetrodotoxin and tityustoxin (220, 221). Neither of these biosensors has been explored for use with biological agents other than toxins. These cell-based biosensors generally are not as specific as those based on shape or sequence recognition, such as the antibody and nucleic acid sensors described above, but this characteristic can be an advantage when the threat agent is unknown.

Another type of cell-based biosensor, based on B lymphocytes, is also under development (239, 241). B lymphocytes are cells in the human body that display surface antibodies that act as pathogen receptors. Rider et al. engineered the B lymphocytes to express both aequorin, a jellyfish protein that emits light in response to calcium flux, and pathogen-specific surface antibodies (241). Although this type of cell-based biosensor provides increased specificity compared to other cell-based systems, it is subject to the same kind of antibody cross-reactivity problems as other antibody-based technologies. In addition, it faces the same problems with storage and maintenance as other cell-based systems.

Another cell type that has been investigated for detection of biological agents is chromatophores, the brightly colored cells found in the skin of cold-blooded animals (53, 198). Chromatophores are responsible for the pigmentation and camouflage exhibited by these animals and will change color when exposed to many classes of biologically active substances, such as pesticides, neurotransmitters, and bacterial toxins. The color change can be observed with a microscope or spectrophotometer. A biosensor incorporating chromatophores from fish has been developed and tested for its ability to detect harmful chemicals, biological toxins, toxin-producing pathogens, and other biologically active agents (53, 75, 259). One advantage of this biosensor is that the fish chromatophores do not grow and, therefore, the culture medium does not need to be replaced frequently.

Chemical and Physical Detection

Biosensors based on physical and chemical properties respond to specific characteristics intrinsic to the target analyte. Examples of such technologies include mass spectrometry, Raman spectrometry, and intrinsic fluorescence/luminescence. These methods do not require additional biological reagents; however, affinity probes may be used to aid in target capture and increase specificity.

A multiwavelength, UV/visible spectroscopic method of detecting pathogens that uses both the light-scattering and absorbing properties of vegetative cells and spores to generate a spectrum is under development (7, 8). It incorporates a relatively simple model to interpret the resulting spectrum. Data acquired to date indicate that vegetative cells of different species of bacteria may be distinguishable and that vegetative cells can be differentiated from spores. One problem with this method is that it currently requires extensive sample preparation to produce a pure suspension of target in a nonabsorbing medium. Furthermore, the method is nonspecific for most clinical applications.

There have been many attempts to develop a biosensor based on mass spectrometry (93, 171, 292). Mass spectrometry selectively identifies components of a sample by molecular weight analysis. The technique has been used to identify bacterial and viral proteins (72, 301, 316) and intact bacterial cells (28, 142, 147, 300, 302, 308) and to distinguish aerosolized spores of Bacillus thuringiensis and Bacillus atrophaeus (94). Most of these approaches utilize matrix-assisted laser desorption/ionization time-of-flight (MALDI-TOF). These systems are reagentless and rapid, require only a small sample volume, and potentially require little sample preparation. MALDI-TOF/mass spectrometry is theoretically capable of identifying all types of biological agents, including viruses, bacteria, fungi, and spores. Drawbacks to this approach include the requirement for a highly concentrated sample (105 to 107 cells/ml for whole-cell analysis), the need to develop complex spectral fingerprints for every target agent, and the possible lack of specificity in complex matrices or mixtures of targets. Flow field-flow fractionation is being investigated as one way of increasing specificity by separating particles in a complex matrix prior to analysis by MALDI-TOF/mass spectrometry (173, 240).

Electrospray ionization Fourier transform ion cyclotron resonance (EIS-FTICR) mass spectrometry has been used to analyze alleles from variable-number tandem repeats and single-nucleotide polymorphisms simultaneously (295). DNA was amplified and fragment size was kept to less than 200 base pairs to stay within the optimum size range for the mass spectrometry procedure. Analysis of variable-number tandem repeat and single-nucleotide polymorphism PCR products from Bacillus anthracis using EIS-FTICR-mass spectrometry produced results comparable to traditional gel electrophoresis.

Surface-enhanced Raman scattering (SERS) is under study as a means of identifying nucleic acids, pathogens, and toxins. As with mass spectrometry, the technique generally requires generation of fingerprints for each target analyte. Two methods of detecting the dipicolinic acid present in bacterial spores using SERS have been described (24, 322). Bell et al. improved the sensitivity of detection of dipicolinic acid by adding sodium sulfate to aggregate the silver colloid surface material and included thiosulfate as an internal standard to permit quantitation (24). Zhang et al. used a silver film over glass nanospheres to detect dipicolinic acid extracted from Bacillus subtilis spores, obtaining a sensitivity of approximately 103 spores in 20 μl (322). The sensitivity was determined to be approximately 104 spores in 20 μl for a field-portable detector that the authors also tested. SERS technology was incorporated into a biochip platform that used antibodies to specifically capture target analytes (112). The authors presented data indicating detection and identification of Listeria species, Legionella species, Bacillus species spores, and Cryptosporidium parvum and Cryptosporidium meleagridis using their biochip. The authors also reported differentiating between nonviable and viable organisms and identifying specific toxins in a mixture based on their SERS fingerprints.

The optical capture of individual particles using a near-infrared SERS method has been tested (5). Following capture, the authors used near-infrared SERS to discriminate between spores of two strains of Bacillus stearothermophilus. Near-infrared SERS has also been used to differentiate seven strains of Escherichia coli and six clinical isolates from urinary tract infections (143). SERS and a related technology, surface-enhanced resonance Raman scattering (SERRS), have been tested as assay reporters. In this approach, the target was either labeled directly with a SERS active substance (92, 223) or indirectly via gold or silver nanoparticles with attached SERS/SERRS-active compounds (45, 78, 80, 215, 314). SERS-active dyes were used to detect two oligonucleotides using a novel microfluidics platform (223) and spectra and detection limits for eight dye-labeled oligonucleotides were determined using SERRS (92). Using a sandwich hybridization technique on a biochip platform and a nanoparticle probe labeled with a SERS-active dye, Cao et al. (45) were able to differentiate between six different DNA strands and between two different RNA strands. Detection down to the femtomolar range was reported. Detection of rat and goat immunoglobulin G antibodies was reported using a sandwich assay that utilized antibody- and SERS-labeled colloidal gold nanoparticles as reporters (215). Detection of the hepatitis B virus surface antigen was also reported using an antibody- and SERS-labeled probe in a sandwich assay format (314). SERS/SERRS probes are a potential means of increasing the specificity of Raman scattering methods and eliminate the need to generate separate spectra for each target of interest.

CONCLUSION

The intentional release and dissemination of Bacillus anthracis spores by contaminated U.S. Postal Service mail in the fall of 2001 made public health officials acutely aware of the importance of rapidly and accurately detecting such events. Because bioterrorism is difficult to predict or prevent, reliable detection platforms are especially important to minimize dissemination of biothreat agents and to protect the public health.

The United States and other countries have committed large amounts of time, effort, and funds in recent years to develop reliable platforms to rapidly detect and identify biothreat agents. Although many different detection technologies have been introduced, few of these technologies have been extensively evaluated or reviewed under field conditions. Many challenges, including processing of complex sample matrices and detection of multiple types of agents and modified or previously uncharacterized agents in a sample, remain to be resolved.

This review has attempted to provide a survey of commercially available and developing technologies for biothreat agent detection. Only technologies that have been evaluated and published have been included. Many other technologies have not been included because insufficient published data were available to ascertain their accuracy and reliability. While an ideal platform has yet to be developed, many of the systems described in this review have proved invaluable in rapidly and accurately identifying biothreat agents. Although the risk of bioterrorism remains, detection technologies will continue to be improved to meet the challenges of this threat.

Acknowledgments

We thank Eric M. Callahan for critical reading of the manuscript.

REFERENCES

  • 1.Aathithan, S., J. C. Plant, A. N. Chaudry, and G. L. French. 2001. Diagnosis of bacteria by detection of volatile organic compounds in urine using an automated headspace analyzer with multiple conducting polymer sensors. J. Clin. Microbiol. 39:2590-2593. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Aberl, F., and C. Kösslinger. 1998. Biosensor-based methods in clinical diagnosis. In Molecular diagnosis of infectious diseases, vol. 13. Humana Press, Inc., Totowa, N.J. [DOI] [PubMed]
  • 3.Abu Al-Soud, W., and P. Radstrom. 1998. Capacity of nine thermostable DNA polymerases to mediate DNA amplification in the presence of PCR-inhibiting samples. Appl. Environ. Microbiol. 64:3748-3753. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Akane, A., K. Matsubara, H. Nakamura, S. Takahashi, and K. Kimura. 1994. Identification of the heme compound copurified with deoxyribonucleic acid (DNA) from bloodstains, a major inhibitor of polymerase chain reaction (PCR) amplification. J. Forensic Sci. 39:362-372. [PubMed] [Google Scholar]
  • 5.Alexander, T. A., P. M. Pellegrino, and J. B. Gillespie. 2003. Near-infrared surface-enhanced-Raman-scattering-mediated detection of single optically trapped bacterial spores. Appl. Spectrosc. 57:1340-1345. [DOI] [PubMed] [Google Scholar]
  • 6.Ali, M. F., R. Kirby, A. P. Goodey, M. D. Rodriguez, A. D. Ellington, D. P. Neikirk, and J. T. McDevitt. 2003. DNA hybridization and discrimination of single-nucleotide mismatches using chip-based microbead arrays. Anal. Chem. 75:4732-4739. [DOI] [PubMed] [Google Scholar]
  • 7.Alupoaei, C. E., and L. H. García-Rubio. 2004. Growth behavior of microorganisms using UV-Vis spectroscopy: Escherichia coli. Biotechnol. Bioeng. 86:163-167. [DOI] [PubMed] [Google Scholar]
  • 8.Alupoaei, C. E., J. A. Olivares, and L. H. Garcia-Rubio. 2004. Quantitative spectroscopy analysis of prokaryotic cells: vegetative cells and spores. Biosensor. Bioelect. 19:893-903. [DOI] [PubMed] [Google Scholar]
  • 9.Anderson, G. P., J. P. Golden, L. K. Cao, D. Wijesuriya, L. C. Shriver-Lake, and F. S. Ligler. 1994. Development of an evanescent wave fiber optic biosensor. Proc. IEEE June :358-363.
  • 10.Anderson, G. P., M. A. Jacoby, F. S. Ligler, and K. D. King. 1997. Effectiveness of protein A for antibody immobilization for a fiber optic biosensor. Biosensor. Bioelect. 12:329-336. [DOI] [PubMed] [Google Scholar]
  • 11.Anderson, G. P., K. D. King, K. L. Gaffney, and L. H. Johnson. 2000. Multi-analyte interrogation using the fiber optic biosensor. Biosensor. Bioelect. 14:771-777. [DOI] [PubMed] [Google Scholar]
  • 12.Andreotti, P. E., G. V. Ludwig, A. H. Peruski, J. J. Tuite, S. S. Morse, and L. F. Peruski, Jr. 2003. Immunoassay of infectious agents. BioTechniques 35:850-859. [DOI] [PubMed] [Google Scholar]
  • 13.AOAC International. 2004. Initiative yields effective methods for anthrax detection; RAMP and MIDI, Inc., methods approved. Inside Lab. Management 10:3. [Google Scholar]
  • 14.AOAC International. 2003. The Official Methods of Analysis of AOAC International, 17th ed. AOAC International, Gaithersburg, Md.
  • 15.Aravanis, A. M., B. D. DeBusschere, A. J. Chruscinski, K. H. Gilchrist, B. K. Kobilka, and G. T. A. Kovacs. 2001. A genetically engineered cell-based biosensor for functional classification of agents. Biosensor. Bioelect. 16:571-577. [DOI] [PubMed] [Google Scholar]
  • 16.Archer, J. R., R. F. Schell, D. R. Pennell, and P. D. Wick. 1987. Identification of Yersinia spp. with the API 20E system. J. Clin. Microbiol. 25:2398-2399. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Atlas, R. M. 2002. Bioterrorism: from threat to reality. Annu. Rev. Microbiol. 56:167-185. [DOI] [PubMed] [Google Scholar]
  • 18.Baeumner, A. J., R. N. Cohen, V. Miksic, and J. Min. 2003. RNA biosensor for the rapid detection of viable Escherichia coli in drinking water. Biosensor. Bioelect. 18:405-413. [DOI] [PubMed] [Google Scholar]
  • 19.Baeumner, A. J., B. Leonard, J. McElwee, and R. A. Montagna. 2004. A rapid biosensor for viable Bacillus anthracis spores. Anal. Bioanal. Chem. 380:15-23. [DOI] [PubMed] [Google Scholar]
  • 20.Baillie, L. W., M. N. Jones, P. C. Turnbull, and R. J. Manchee. 1995. Evaluation of the BIOLOG® system for the identification of Bacillus anthracis. Lett. Appl. Microbiol. 20:209-211. [DOI] [PubMed] [Google Scholar]
  • 21.Belgrader, P., D. Hansford, G. T. Kovacs, K. Venkateswaran, R. Mariella, Jr., F. P. Milanovich, S. Nasarabadi, M. Okuzumi, F. Pourahmadi, and M. A. Northrup. 1999. A minisonicator to rapidly disrupt bacterial spores for DNA analysis. Anal. Chem. 71:4232-4236. [DOI] [PubMed] [Google Scholar]
  • 22.Belgrader, P., M. Okuzumi, F. Pourahmadi, D. A. Borkholder, and M. A. Northrup. 2000. A microfluidic cartridge to prepare spores for PCR analysis. Biosensor. Bioelect. 14:849-852. [DOI] [PubMed] [Google Scholar]
  • 23.Bell, C. A., J. R. Uhl, T. L. Hadfield, J. C. David, R. F. Meyer, T. F. Smith, and F. R. Cockerill III. 2002. Detection of Bacillus anthracis DNA by LightCycler PCR. J. Clin. Microbiol. 40:2897-2902. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Bell, S. E. J., J. N. Mackle, and N. M. S. Sirimuthu. 2005. Quantitative surface-enhanced Raman spectroscopy of dipicolinic acid-towards rapid anthrax endospore detection. Analyst 130:545-549. [DOI] [PubMed] [Google Scholar]
  • 25.Benoit, P. W., and D. W. Donahue. 2003. Methods for rapid separation and concentration of bacteria in food that bypass time-consuming cultural enrichment. J. Food Prot. 66:1935-1948. [DOI] [PubMed] [Google Scholar]
  • 26.Bergmann, A. R., B. L. Schmidt, A. M. Derler, and E. Aberer. 2002. Importance of sample preparation for molecular diagnosis of Lyme Borreliosis from urine. J. Clin. Microbiol. 40:4581-4584. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Bernard, K., S. Tessier, J. Winstanley, D. Chang, and A. Borczyk. 1994. Early recognition of atypical Francisella tularensis strains lacking a cysteine requirement. J. Clin. Microbiol. 32:551-553. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Beverly, M. B., and K. J. Voorhees. 2000. Electron monochromator mass spectrometry for the analysis of whole bacteria and bacterial spores. Anal. Chem. 72:2428-2432. [DOI] [PubMed] [Google Scholar]
  • 29.Biagini, R. E., D. L. Sammons, J. P. Smith, B. A. MacKenzie, C. A. F. Striley, V. Semenova, E. Steward-Clark, K. Stamey, A. E. Freeman, C. P. Quinn, and J. E. Snawder. 2004. Comparison of a multiplexed fluorescent covalent microsphere immunoassay and an enzyme-linked immunosorbent assay for measurement of human immunoglobulin G antibodies to anthrax toxins. Clin. Diagn. Lab. Immunol. 11:50-55. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Biagini, R. E., D. L. Sammons, J. P. Smith, E. H. Page, J. E. Snawder, C. A. F. Striley, and B. A. MacKenzie. 2004. Determination of serum IgG antibodies to Bacillus anthracis protective antigen in environmental sampling workers using a fluorescent covalent microsphere immunoassay. Occup. Environ. Med. 61:703-708. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Bible, E. A., F. M. Johnson, B. J. Bradley, and D. A. Golden. 2004. Comparison of methods for recovery of Salmonella from cantaloupe rinds. Presented at the International Association for Food Protection, Phoenix, Ariz, 8-11 August.
  • 32.Branch, D. W., and S. M. Brozik. 2004. Low-level detection of a Bacillus anthracis simulant using Love-wave biosensors on 36[deg]YX LiTaO3. Biosensor. Bioelect. 19:849-859. [DOI] [PubMed] [Google Scholar]
  • 33.Breaker, R. R. 2004. Natural and engineered nucleic acids as tools to explore biology. Nature 432:838-845. [DOI] [PubMed] [Google Scholar]
  • 34.Brigati, J., D. D. Williams, I. B. Sorokulova, V. Nanduri, I.-H. Chen, C. L. Turnbough, Jr., and V. A. Petrenko. 2004. Diagnostic probes for Bacillus anthracis spores selected from a landscape phage library. Clin. Chem. 50:1899-1906. [DOI] [PubMed] [Google Scholar]
  • 35.Brinkman, J. A., M. Z. Rahmani, W. E. Jones, A. K. Chaturvedi, and M. E. Hagensee. 2004. Optimization of PCR based detection of human papillomavirus DNA from urine specimens. J. Clin. Virol. 29:230-240. [DOI] [PubMed] [Google Scholar]
  • 36.Brody, E. N., and L. Gold. 2000. Aptamers as therapeutic and diagnostic agents. J. Biotechnol. 4:5-13. [DOI] [PubMed] [Google Scholar]
  • 37.Brown, K. 2004. Up in the air. Science 305:1228-1229. [DOI] [PubMed] [Google Scholar]
  • 38.Bruno, J. G., and J. L. Kiel. 1999. In vitro selection of DNA aptamers to anthrax spores with electrochemiluminescence detection. Biosensor. Bioelect. 14:457-464. [DOI] [PubMed] [Google Scholar]
  • 39.Bruno, J. G., and J. L. Kiel. 2002. Use of magnetic beads in selection and detection of biotoxin aptamers by electrochemiluminescence and enzymatic methods. BioTechniques 32:178-183. [DOI] [PubMed] [Google Scholar]
  • 40.Buffone, G. J., G. J. Demmler, C. M. Schimbor, and J. Greer. 1991. Improved amplification of cytomegalovirus DNA from urine after purification of DNA with glass beads. Clin. Chem. 37:1945-1949. [PubMed] [Google Scholar]
  • 41.Bukhari, Z., R. M. McCuin, C. R. Fricker, and J. L. Clancy. 1998. Immunomagnetic separation of Cryptosporidium parvum from source water samples of various turbidities. Appl. Environ. Microbiol. 64:4495-4499. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Buttner, M. P., P. Cruz, L. D. Stetzenbach, A. K. Klima-Comba, V. L. Stevens, and T. D. Cronin. 2004. Determination of the efficacy of two building decontamination strategies by surface sampling with culture and quantitative PCR analysis. Appl. Environ. Microbiol. 70:4740-4747. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Buttner, M. P., K. Willeke, and S. Grinshpun. 2002. Sampling and analysis of air-borne microorganisms, p. 814-826. In C. J. Hurst, R. L. Crawford, G. Knudsen, M. McInerney, and L. D. Stetzenbach (ed.), Manual of environmental microbiology, 2nd ed. ASM Press, Washington, D.C.
  • 44.Cao, L. K., G. P. Anderson, F. S. Ligler, and J. Ezzell. 1995. Detection of Yersinia pestis fraction 1 antigen with a fiber optic biosensor. J. Clin. Microbiol. 33:336-341. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Cao, Y. C., R. Jin, and C. A. Mirkin. 2002. Nanoparticles with Raman spectroscopic fingerprints for DNA and RNA detection. Science 297:1536-1540. [DOI] [PubMed] [Google Scholar]
  • 46.Casper, E. T., S. S. Patterson, M. C. Smith, and J. H. Paul. 2005. Development and evaluation of a method to detect and quantify enteroviruses using NASBA and internal control RNA (IC-NASBA). J. Virol. Methods 124:149-155. [DOI] [PubMed] [Google Scholar]
  • 47.Centers for Disease Control and Prevention. 2001. Evaluation of Bacillus anthracis contamination inside the Brentwood Mail Processing and Distribution Center B, District of Columbia. Morb. Mortal. Wkly. Rep. 50:1129-1133. [Google Scholar]
  • 48.Chan, S., S. R. Horner, P. M. Fauchet, and B. L. Miller. 2001. Identification of gram negative bacteria using nanoscale silicon microcavities. J. Am. Chem. Soc. 123:11797-11798. [DOI] [PubMed] [Google Scholar]
  • 49.Chan, W. C. W., D. J. Maxwell, X. Gao, R. E. Bailey, M. Han, and S. Nie. 2002. Luminescent quantum dots for multiplexed biological detection and imaging. Curr. Opin. Biotechnol. 13:40-46. [DOI] [PubMed] [Google Scholar]
  • 50.Chandler, D. P., J. Brown, D. R. Call, S. Wunschel, J. W. Grate, D. A. Holman, L. Olson, M. S. Stottlemyre, and C. J. Bruckner-Lea. 2001. Automated immunomagnetic separation and microarray detection of E. coli O157:H7 from poultry carcass rinse. Int. J. Food Microbiol. 70:143-154. [DOI] [PubMed] [Google Scholar]
  • 51.Chandler, D. P., and A. E. Jarrell. 2004. Automated purification and suspension array detection of 16S rRNA from soil and sediment extracts by using tunable surface microparticles. Appl. Environ. Microbiol. 70:2621-2631. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Chandler, D. P., J. R. Stults, S. Cebula, B. L. Schuck, D. W. Weaver, K. K. Anderson, M. Egholm, and F. J. Brockman. 2000. Affinity purification of DNA and RNA from environmental samples with peptide nucleic acid clamps. Appl. Environ. Microbiol. 66:3438-3445. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Chaplen, F. W. R., R. H. Upson, P. N. Mcfadden, and W. Kolodziej. 2002. Fish chromatophores as cytosensors in a microscale device: Detection of environmental toxins and bacterial pathogens. Pigment Cell Res. 15:19-26. [DOI] [PubMed] [Google Scholar]
  • 54.Cheng, J., E. L. Sheldon, L. Wu, A. Uribe, L. O. Gerrue, J. Carrino, M. J. Heller, and J. P. O'Connell. 1998. Preparation and hybridization analysis of DNA/RNA from E. coli on microfabricated bioelectronic chips. Nat. Biotechnol. 16:541-546. [DOI] [PubMed] [Google Scholar]
  • 55.Christopher, G., T. Cieslak, J. Pavlin, and E. Eitzen. 1997. Biological warfare: a historical perspective. JAMA 278:412-417. [PubMed] [Google Scholar]
  • 56.Chua, J. 2004. A buyer's guide to DNA and RNA prep kits. Scientist 18:38-39. [Google Scholar]
  • 57.Clarridge, J., 3rd, T. Raich, A. Sjosted, G. Sandstrom, R. Darouiche, R. Shawar, P. Georghiou, C. Osting, and L. Vo. 1996. Characterization of two unusual clinically significant Francisella strains. J. Clin. Microbiol. 34:1995-2000. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Cook, N. 2003. The use of NASBA for the detection of microbial pathogens in food and environmental samples. J. Microbiol. Methods 53:165-174. [DOI] [PubMed] [Google Scholar]
  • 59.Coombes, B. K., and J. B. Mahony. 2000. Nucleic acid sequence based amplification (NASBA) of Chlamydia pneumoniae major outer membrane protein (ompA) mRNA with bioluminescent detection. Comb. Chem. High Throughput Screen. 3:315-327. [DOI] [PubMed] [Google Scholar]
  • 60.Corstjens, P., M. Zuiderwijk, A. Brink, S. Li, H. Feindt, R. S. Niedbala, and H. Tanke. 2001. Use of up-converting phosphor reporters in lateral-flow assays to detect specific nucleic acid sequences: a rapid, sensitive DNA test to identify human papillomavirus type 16 infection. Clin. Chem. 47:1885-1893. [PubMed] [Google Scholar]
  • 61.Cox, C., and C. Wathes. 1995. Bioaerosols in the environment, p. 11-14. In C. Cox and C. Wathes (ed.), Bioaerosols handbook. Lewis Publishers, Boca Raton, Fla.
  • 62.Cutler, C. N., W. J. Dorsa, and G. R. Sirogusa. 1996. A rapid microbial ATP bioluminescence assay for meat carcasses. Dairy Food Environ. Sanit. 16:726-736. [Google Scholar]
  • 63.Dalesio, N., V. Marsiglia, A. Quinn, T. C. Quinn, and C. A. Gaydos. 2004. Performance of the MagNA Pure LC robot for extraction of Chlamydia trachomatis and Neisseria gonorrhoeae DNA from urine and swab specimens. J. Clin. Microbiol. 42:3300-3302. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Davis, K. C., C. H. Nakatsu, R. Turco, S. D. Weagant, and A. K. Bhunia. 2003. Analysis of environmental Escherichia coli isolates for virulence genes using the TaqMan® PCR system. J. Appl. Microbiol. 95:612-620. [DOI] [PubMed] [Google Scholar]
  • 65.DeBusschere, B. D., and G. T. A. Kovacs. 2001. Portable cell-based biosensor system using integrated CMOS cell-cartridges. Biosensor. Bioelect. 16:543-556. [DOI] [PubMed] [Google Scholar]
  • 66.Deiman, B., P. van Aarle, and P. Sillekens. 2002. Characteristics and applications of nucleic acid sequence-based amplification (NASBA). Mol. Biotechnol. 20:163-180. [DOI] [PubMed] [Google Scholar]
  • 67.Deisingh, A. K., and M. Thompson. 2004. Biosensors for the detection of bacteria. Can. J. Microbiol. 50:69-77. [DOI] [PubMed] [Google Scholar]
  • 68.Delehanty, J. B., and F. S. Ligler. 2002. A microarray immunoassay for simultaneous detection of proteins and bacteria. Anal. Chem. 74:5681-5687. [DOI] [PubMed] [Google Scholar]
  • 69.DeMarco, D. R., and D. V. Lim. 2002. Detection of Escherichia coli O157:H7 in 10- and 25-gram ground beef samples with an evanescent-wave biosensor with silica and polystyrene waveguides. J. Food Prot. 65:596-602. [DOI] [PubMed] [Google Scholar]
  • 70.DeMarco, D. R., and D. V. Lim. 2001. Direct detection of Escherichia coli O157:H7 in unpasteurized apple juice with an evanescent wave biosensor. J. Rapid Methods Automat. Microbiol. 9:241-257. [Google Scholar]
  • 71.DeMarco, D. R., E. W. Saaski, D. A. McCrae, and D. V. Lim. 1999. Rapid detection of Escherichia coli O157:H7 in ground beef using a fiber-optic biosensor. J. Food Prot. 62:711-716. [DOI] [PubMed] [Google Scholar]
  • 72.Demirev, P. A., J. Ramirez, and C. Fenselau. 2001. Tandem mass spectrometry of intact proteins for characterization of biomarkers from Bacillus cereus T spores. Anal. Chem. 73:5725-5731. [DOI] [PubMed] [Google Scholar]
  • 73.Deng, X. K., L. A. Nesbit, and K. J. Morrow, Jr. 2003. Recombinant single-chain variable fragment antibodies drected against Clostridium difficile toxin B produced by use of an optimized phage display system. Clin. Diagn. Lab. Immunol. 10:587-595. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 74.Devic, E., D. Li, A. Dauta, P. Henriksen, G. A. Codd, J.-L. Marty, and D. Fournier. 2002. Detection of anatoxin-a(s) in environmental samples of cyanobacteria by using a biosensor with engineered acetylcholinesterases. Appl. Environ. Microbiol. 68:4102-4106. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 75.Dierksen, K. P., L. Mojovic, B. A. Caldwell, R. R. Preston, R. Upson, J. Lawrence, P. N. McFadden, and J. E. Trempy. 2004. Responses of fish chromatophore-based cytosensor to a broad range of biological agents. J. Appl. Toxicol. 24:363-369. [DOI] [PubMed] [Google Scholar]
  • 76.DiGiorgio, C. L., D. A. Gonzalez, and C. C. Huitt. 2002. Cryptosporidium and Giardia recoveries in natural waters by using Environmental Protection Agency method 1623. Appl. Environ. Microbiol. 68:5952-5955. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 77.Dill, K., D. D. Montgomery, A. L. Ghindilis, K. R. Schwarzkopf, S. R. Ragsdale, and A. V. Oleinikov. 2004. Immunoassays based on electrochemical detection using microelectrode arrays. Biosensor. Bioelect. 20:736-742. [DOI] [PubMed] [Google Scholar]
  • 78.Docherty, F. T., M. Clark, G. McNay, D. Graham, and W. E. Smith. 2003. Multiple labelled nanoparticles for bio detection. Faraday Discuss 126:281-288. [DOI] [PubMed] [Google Scholar]
  • 79.Dodson, J. M., M. J. Feldstein, D. M. Leatzow, L. K. Flack, J. P. Golden, and F. S. Ligler. 2001. Fluidics cube for biosensor miniaturization. Anal. Chem. 73:3776-3780. [DOI] [PubMed] [Google Scholar]
  • 80.Doering, W. E., and S. Nie. 2003. Spectroscopic tags using dye-embedded nanoparticles and surface-enhanced Raman scattering. Anal. Chem. 75:6171-6176. [DOI] [PubMed] [Google Scholar]
  • 81.Donaldson, K. A., M. F. Kramer, and D. V. Lim. 2004. A rapid detection method for Vaccinia virus, the surrogate for smallpox virus. Biosensor. Bioelect. 20:322-327. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 82.Doyle, R. J., R. Nedjat-Haiem, and J. S. Singh. 1984. Hydrophobic characteristics of Bacillus spores. Curr. Microbiol. 10:329-332. [Google Scholar]
  • 83.Emanuel, P. A., R. Bell, J. L. Dang, R. McClanahan, J. C. David, R. J. Burgess, J. Thompson, L. Collins, and T. Hadfield. 2003. Detection of Francisella tularensis within infected mouse tissues by using a hand-held PCR thermocycler. J. Clin. Microbiol. 41:689-693. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 84.Emanuel, P. A., J. Dang, J. S. Gebhardt, J. Aldrich, E. A. E. Garber, H. Kulaga, P. Stopa, J. J. Valdes, and A. Dion-Schultz. 2000. Recombinant antibodies: a new reagent for biological agent detection. Biosensor. Bioelect. 14:751-759. [DOI] [PubMed] [Google Scholar]
  • 85.Endo, S., T. Komori, G. Ricci, A. Sano, K. Yokoyama, A. Ohori, K. Kamei, M. Franco, M. Miyaji, and K. Nishimura. 2004. Detection of gp43 of Paracoccidioides brasiliensis by the loop-mediated isothermal amplification (LAMP) method. FEMS Microbiol. Lett. 234:93-97. [DOI] [PubMed] [Google Scholar]
  • 86.Enosawa, M., S. Kageyama, K. Sawai, K. Watanabe, T. Notomi, S. Onoe, Y. Mori, and Y. Yokomizo. 2003. Use of loop-mediated isothermal amplification of the IS900 sequence for rapid detection of cultured Mycobacterium avium subsp. paratuberculosis. J. Clin. Microbiol. 41:4359-4365. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 87.Enroth, H., and L. Engstrand. 1995. Immunomagnetic separation and PCR for detection of Helicobacter pylori in water and stool specimens. J. Clin. Microbiol. 33:2162-2165. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 88.Entis, P. 1998. Direct 24-hour presumptive enumeration of Escherichia coli O157:H7 in foods using hydrophobic grid memebrane filter followed by serological confirmation: collaborative study. J. AOAC Int. 81:403-418. [PubMed] [Google Scholar]
  • 89.Ewalt, K. L., R. W. Haigis, R. Rooney, D. Ackley, and M. Krihak. 2001. Detection of biological toxins on an active electronic microchip. Anal. Biochem. 289:162-172. [DOI] [PubMed] [Google Scholar]
  • 90.Exner, M. M., and M. A. Lewinski. 2003. Isolation and detection of Borrelia burgdorferi DNA from cerebral spinal fluid, synovial fluid, blood, urine, and ticks using the Roche MagNA Pure system and real-time PCR. Diagn. Microbiol. Infect. Dis. 46:235-240. [DOI] [PubMed] [Google Scholar]
  • 91.Fafi-Kremer, S., K. Brengel-Pesce, G. Bargues, M. J. Bourgeat, O. Genoulaz, J. M. Seigneurin, and P. Morand. 2004. Assessment of automated DNA extraction coupled with real-time PCR for measuring Epstein-Barr virus load in whole blood, peripheral mononuclear cells and plasma. J. Clin. Virol. 30:157-164. [DOI] [PubMed] [Google Scholar]
  • 92.Faulds, K., W. E. Smith, and D. Graham. 2004. Evaluation of surface-enhanced resonance Raman scattering for quantitative DNA analysis. Anal. Chem. 76:412-417. [DOI] [PubMed] [Google Scholar]
  • 93.Fenselau, C., and P. A. Demirev. 2001. Characterization of intact microorganisms by MALDI mass spectrometry. Mass Spectrom Rev. 20:157-171. [DOI] [PubMed] [Google Scholar]
  • 94.Fergenson, D. P., M. E. Pitesky, H. J. Tobias, P. T. Steele, G. A. Czerwieniec, S. C. Russell, C. B. Lebrilla, J. M. Horn, K. R. Coffee, A. Srivastava, S. P. Pillai, M. T. Shih, H. L. Hall, A. J. Ramponi, J. T. Chang, R. G. Langlois, P. L. Estacio, R. T. Hadley, M. Frank, and E. E. Gard. 2004. Reagentless detection and classification of individual bioaerosol particles in seconds. Anal. Chem. 76:373-378. [DOI] [PubMed] [Google Scholar]
  • 95.Fode-Vaughan, K. A., J. S. Maki, J. A. Benson, and M. L. P. Collins. 2003. Direct PCR detection of Escherichia coli O157:H7. Lett. Appl. Microbiol. 37:239-243. [DOI] [PubMed] [Google Scholar]
  • 96.Furukawa, H., K. Doh-ura, R. Okuwaki, S. Shirabe, K. Yamamoto, H. Udono, T. Ito, S. Katamine, and M. Niwa. 2004. A pitfall in diagnosis of human prion diseases using detection of protease-resistant prion protein in urine. J. Biol. Chem. 279:23661-23667. [DOI] [PubMed] [Google Scholar]
  • 97.Garcia, M. E., J. L. Blanco, J. Caballero, and D. Gargallo-Viola. 2002. Anticoagulants interfere with PCR used to diagnose invasive aspergillosis. J. Clin. Microbiol. 40:1567-1568. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 98.Gascoyne, P., C. Mahidol, M. Ruchirawat, J. Satayavivad, P. Watcharasit, and F. F. Becker. 2002. Microsample preparation by dielectrophoresis: isolation of malaria. Lab. Chip 2:70-75. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 99.Gascoyne, P., J. Satayavivad, and M. Ruchirawat. 2004. Microfluidic approaches to malaria detection. Acta Trop. 89:357-369. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 100.Gatto-Menking, D. L., H. Yu, J. G. Bruno, M. T. Goode, M. Miller, and A. W. Zulich. 1995. Sensitive detection of biotoxoids and bacterial spores using an immunomagnetic electrochemiluminescence sensor. Biosensor. Bioelect. 10:501-507. [DOI] [PubMed] [Google Scholar]
  • 101.Gau, J.-J., E. H. Lan, B. Dunn, C.-M. Ho, and J. C. S. Woo. 2001. A MEMS based amperometric detector for E. coli bacteria using self-assembled monolayers. Biosensor. Bioelect. 16:745-755. [DOI] [PubMed] [Google Scholar]
  • 102.Geng, T., M. T. Morgan, and A. K. Bhunia. 2004. Detection of low levels of Listeria monocytogenes cells by using a fiber-optic immunosensor. Appl. Environ. Microbiol. 70:6138-6146. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 103.Gerard, M., A. Chaubey, and B. D. Malhotra. 2002. Application of conducting polymers to biosensors. Biosensor. Bioelect. 17:345-359. [DOI] [PubMed] [Google Scholar]
  • 104.Germer, J. J., M. M. Lins, M. E. Jensen, W. S. Harmsen, D. M. Ilstrup, P. S. Mitchell, F. R. Cockerill, III, and R. Patel. 2003. Evaluation of the MagNaPure LC instrument for extraction of hepatitis C virus RNA for the COBAS AMPLICOR hepatitis C virus test, version 2.0. J. Clin. Microbiol. 41:3503-3508. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 105.Gilchrist, K. H., V. N. Barker, L. E. Fletcher, B. D. DeBusschere, P. Ghanouni, L. Giovangrandi, and G. T. A. Kovacs. 2001. General purpose, field-portable cell-based biosensor platform. Biosensor. Bioelect. 16:557-564. [DOI] [PubMed] [Google Scholar]
  • 106.Gioffre, A., L. Meichtri, M. Zumarraga, R. Rodriguez, and A. Cataldi. 2004. Evaluation of a QIAamp DNA stool purification kit for Shiga-toxigenic Escherichia coli detection in bovine fecal swabs by PCR. Rev. Argent. Microbiol. 36:1-5. [PubMed] [Google Scholar]
  • 107.Goldman, E. R., A. R. Clapp, G. P. Anderson, H. T. Uyeda, J. M. Mauro, I. L. Medintz, and H. Mattoussi. 2004. Multiplexed toxin analysis using four colors of quantum dot fluororeagents. Anal. Chem. 76:684-688. [DOI] [PubMed] [Google Scholar]
  • 108.Goldman, E. R., M. P. Pazirandeh, J. M. Mauro, K. D. King, J. C. Frey, and G. P. Anderson. 2000. Phage-displayed peptides as biosensor reagents. J. Mol. Recog. 13:382-387. [DOI] [PubMed] [Google Scholar]
  • 109.Goodey, A., J. J. Lavigne, S. M. Savoy, M. D. Rodriguez, T. Curey, A. Tsao, G. Simmons, J. Wright, S.-J. Yoo, Y. Sohn, E. V. Anslyn, J. B. Shear, D. P. Neikirk, and J. T. McDevitt. 2001. Development of multianalyte sensor arrays composed of chemically derivatized polymeric microspheres localized in micromachined cavities. J. Am. Chem. Soc. 123:2559-2570. [DOI] [PubMed] [Google Scholar]
  • 110.Gray, S. A., J. K. Kusel, K. M. Shaffer, Y. S. Shubin, D. A. Stenger, and J. J. Pancrazio. 2001. Design and demonstration of an automated cell-based biosensor. Biosensor. Bioelect. 16:535-542. [DOI] [PubMed] [Google Scholar]
  • 111.Grimm, D. 2004. Standards set for anthrax detection kits. Science 306:1113. [Google Scholar]
  • 112.Grow, A. E., L. L. Wood, J. L. Claycomb, and P. A. Thompson. 2003. New biochip technology for label-free detection of pathogens and their toxins. J. Microbiol. Methods 53:221-233. [DOI] [PubMed] [Google Scholar]
  • 113.Hall, R. H. 2002. Biosensor technologies for detecting microbiological foodborne hazards. Microbes Infect. 4:425-432. [DOI] [PubMed] [Google Scholar]
  • 114.Hampl, J., M. Hall, N. A. Mufti, Y.-m. M. Yao, D. B. MacQueen, W. H. Wright, and D. E. Cooper. 2001. Upconverting phosphor reporters in immunochromatographic assays. Anal. Biochem. 288:176-187. [DOI] [PubMed] [Google Scholar]
  • 115.Hawley, R. J., and E. M. Eitzen, Jr. 2001. Biological weapons: A primer for microbiologists. Annu. Rev. Microbiol. 55:235-253. [DOI] [PubMed] [Google Scholar]
  • 116.Hayhurst, A., S. Happe, R. Mabry, Z. Koch, B. L. Iverson, and G. Georgiou. 2003. Isolation and expression of recombinant antibody fragments to the biological warfare pathogen Brucella melitensis. J. Immunol. Methods 276:185-196. [DOI] [PubMed] [Google Scholar]
  • 117.Heden, C. G. 1967. Defences against biological warfare. Annu. Rev. Microbiol. 21:639-676. [DOI] [PubMed] [Google Scholar]
  • 118.Heid, C., J. Stevens, K. Livak, and P. Williams. 1996. Real time quantitative PCR. Genome Res. 6:986-994. [DOI] [PubMed] [Google Scholar]
  • 119.Hesselberth, J. R., M. P. Robertson, S. M. Knudsen, and A. D. Ellington. 2003. Simultaneous detection of diverse analytes with an aptazyme ligase array. Anal. Biochem. 312:106-112. [DOI] [PubMed] [Google Scholar]
  • 120.Hibbitts, S., A. Rahman, R. John, D. Westmoreland, and J. D. Fox. 2003. Development and evaluation of NucliSensR Basic Kit NASBA for diagnosis of parainfluenza virus infection with ‘end-point’ and ‘real-time’ detection. J. Virol. Methods 108:145-155. [DOI] [PubMed] [Google Scholar]
  • 121.Higgins, J. A., M. Cooper, L. Schroeder-Tucker, S. Black, D. Miller, J. S. Karns, E. Manthey, R. Breeze, and M. L. Perdue. 2003. A field investigation of Bacillus anthracis contamination of U.S. Department of Agriculture and other Washington, D.C., buildings during the anthrax attack of October 2001. Appl. Environ. Microbiol. 69:593-599. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 122.Higgins, J. A., S. Nasarabadi, J. S. Karns, D. R. Shelton, M. Cooper, A. Gbakima, and R. P. Koopman. 2003. A handheld real time thermal cycler for bacterial pathogen detection. Biosensor. Bioelect. 18:1115-1123. [DOI] [PubMed] [Google Scholar]
  • 123.Hindson, B. J., S. B. Brown, G. D. Marshall, M. T. McBride, A. J. Makarewicz, D. M. Gutierrez, D. K. Wolcott, T. R. Metz, R. A. Madabhushi, J. M. Dzenitis, and B. Colston. 2004. Development of an automated sample preparation module for environmental monitoring of biowarfare agents. Anal. Chem. 76:3492-3497. [DOI] [PubMed] [Google Scholar]
  • 124.Hindson, B. J., M. T. McBride, A. J. Makarewicz, B. D. Henderer, U. S. Setlur, S. M. Smith, G. D. M., M. T. R., S. L. Nasarabadi, K. S. Venkateswaran, S. W. Farrow, B. W. Colston, Jr., and J. M. Dzenitis. 2005. Autonomous detection of aerosolized biological agents by multiplexed immunoassay with polymerase chain reaction confirmation. Anal. Chem. 77:284-289. [DOI] [PubMed] [Google Scholar]
  • 125.Hock, B., M. Seifert, and K. Kramer. 2002. Engineering receptors and antibodies for biosensors. Biosensor. Bioelect. 17:239-249. [DOI] [PubMed] [Google Scholar]
  • 126.Hoffmaster, A. R., R. F. Meyer, M. P. Bowen, C. K. Marston, R. S. Weyant, G. A. Barnett, J. J. Sejvar, J. A. Jernigan, B. A. Perkins, and T. Popovic. 2002. Evaluation and validation of a real-time polymerase chain reaction assay for rapid identification of Bacillus anthracis. Emerg. Infect. Dis. 8:11781182. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 127.Holland, P., R. Abramson, R. Watson, and D. Gelfand. 1991. Detection of specific polymerase chain reaction product by utilizing the 5′ to 3′ exonuclease activity of Thermus aquaticus DNA polymerase. Proc. Natl. Acad. Sci. USA 88:7276-7280. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 128.Holmberg, M., F. Gustafsson, E. G. Hoernsten, F. Winquist, L. E. Nilsson, L. Ljung, and I. Lundstroem. 1998. Bacteria classification based on feature extraction from sensor data. Biotechnol. Technol. 12:319-324. [Google Scholar]
  • 129.Homola, J., J. Dostalek, S. Chen, A. Rasooly, S. Jiang, and S. S. Yee. 2002. Spectral surface plasmon resonance biosensor for detection of staphylococcal enterotoxin B in milk. Int. J. Food Microbiol. 75:61-69. [DOI] [PubMed] [Google Scholar]
  • 130.Horisaka, T., K. Fujita, T. Iwata, A. Nakadai, A. T. Okatani, T. Horikita, T. Taniguchi, E. Honda, Y. Yokomizo, and H. Hayashidani. 2004. Sensitive and specific detection of Yersinia pseudotuberculosis by loop-mediated isothermal amplification. J. Clin. Microbiol. 42:5349-5352. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 131.Hsih, H. Y., and H. Y. Tsen. 2001. Combination of immunomagnetic separation and polymerase chain reaction for the simultaneous detection of Listeria monocytogenes and Salmonella spp. in food samples. J. Food Prot. 64:1744-1750. [DOI] [PubMed] [Google Scholar]
  • 132.Huang, Y., K. L. Ewalt, M. Tirado, R. Haigis, A. Forster, D. Ackley, M. J. Heller, J. P. O'Connell, and M. Krihak. 2001. Electric manipulation of bioparticles and macromolecules on microfabricated electrodes. Anal. Chem. 73:1549-1559. [DOI] [PubMed] [Google Scholar]
  • 133.Ibekwe, A. M., and C. M. Grieve. 2003. Detection and quantification of Escherichia coli O157:H7 in environmental samples by real-time PCR. J. Appl. Microbiol. 94:421-431. [DOI] [PubMed] [Google Scholar]
  • 134.Ibrahim, S. M., D. A. Kulesh, S. S. Saleh, I. K. Damon, J. J. Esposito, A. L. Schmaljohn, and P. B. Jahrling. 2003. Real-time PCR assay to detect smallpox virus. J. Clin. Microbiol. 41:3835-3839. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 135.Ilic, B., D. Czaplewski, and H. G. Craighead. 2000. Mechanical resonant immunospecific biological detector. Appl. Phys. Lett. 77:450-452. [Google Scholar]
  • 136.Inglis, T. J. J., M. Aravena-Roman, S. Ching, K. Croft, V. Wuthiekanun, and B. J. Mee. 2003. Cellular fatty acid profile distinguishes Burkholderia pseudomallei from avirulent Burkholderia thailandensis. J. Clin. Microbiol. 41:4812-4814. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 137.Iqbal, S. S., M. W. Mayo, J. G. Bruno, B. V. Bronk, C. A. Batt, and J. P. Chambers. 2000. A review of molecular recognition technologies for detection of biological threat agents. Biosensor. Bioelect. 15:549-578. [DOI] [PubMed] [Google Scholar]
  • 138.Ivnitski, D., I. Abdel-Hamid, P. Atanasov, and E. Wilkins. 1999. Biosensors for detection of pathogenic bacteria. Biosensor. Bioelect. 14:599-624. [DOI] [PubMed] [Google Scholar]
  • 139.Ivnitski, D., D. J. O'Neil, A. Gattuso, R. Schlicht, M. Calidonna, and R. Fisher. 2003. Nucleic acid approaches for detection and identification of biological warfare and infectious disease agents. BioTechniques 35:862-869. [DOI] [PubMed] [Google Scholar]
  • 140.Iwamoto, T., T. Sonobe, and K. Hayashi. 2003. Loop-mediated isothermal amplification for direct detection of Mycobacterium tuberculosis complex, M. avium, and M. intracellulare in sputum samples. J. Clin. Microbiol. 41:2616-2622. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 141.Jaffer, A. S. 2004. Biohazard detection system deployment resumes. U. S. Postal Service Postal News No. 38.
  • 142.Jarman, K. H., S. T. Cebula, A. J. Saenz, C. E. Petersen, N. B. Valentine, M. T. Kingsley, and K. L. Wahl. 2000. An algorithm for automated bacterial identification using matrix-assisted laser desorption/ionization mass spectrometry. Anal. Chem. 72:1217-1223. [DOI] [PubMed] [Google Scholar]
  • 143.Jarvis, R. M., and R. Goodacre. 2004. Discrimination of bacteria using surface-enhanced Raman spectroscopy. Anal. Chem. 76:40-47. [DOI] [PubMed] [Google Scholar]
  • 144.Jean, J., B. Blais, A. Darveau, and I. Fliss. 2002. Simultaneous detection and identification of hepatitis A virus and rotavirus by multiplex nucleic acid sequence-based amplification (NASBA) and microtiter plate hybridization system. J. Virol. Methods 105:123-132. [DOI] [PubMed] [Google Scholar]
  • 145.Jean, J., D. H. D'Souza, and L.-A. Jaykus. 2004. Multiplex nucleic acid sequence-based amplification for simultaneous detection of several enteric viruses in model ready-to-eat foods. Appl. Environ. Microbiol. 70:6603-6610. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 146.Johansson, M. K., and R. M. Cook. 2003. Intramolecular dimers:A new design strategy for fluorescence-quenched probes. Chem. Eur. J. 9:3466-3471. [DOI] [PubMed] [Google Scholar]
  • 147.Jones, J. J., M. J. Stump, R. C. Fleming, J. O. Lay, Jr., and C. L. Wilkins. 2003. Investigation of MALDI-TOF and FT-MS techniques for analysis of Escherichia coli whole cells. Anal. Chem. 75:1340-1347. [DOI] [PubMed] [Google Scholar]
  • 148.Karch, H., C. Janetzki-Mittmann, S. Aleksic, and M. Datz. 1996. Isolation of enterohemorrhagic Escherichia coli O157 strains from patients with hemolytic-uremic syndrome by using immunomagnetic separation, DNA-based methods, and direct culture. J. Clin. Microbiol. 34:516-519. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 149.Keshri, G., and N. Magan. 2000. Detection and differentiation between mycotoxigenic and non-mycotoxigenic strains of two Fusarium spp. using volatile production profiles and hydrolytic enzymes. J. Appl. Microbiol. 89:825-833. [DOI] [PubMed] [Google Scholar]
  • 150.Keshri, G., P. Voysey, and N. Magan. 2002. Early detection of spoilage moulds in bread using volatile production patterns and quantitative enzyme assays. J. Appl. Microbiol. 92:165-172. [DOI] [PubMed] [Google Scholar]
  • 151.Kessler, H. H., A. M. Clarici, E. Stelzl, G. Muhlbauer, E. Daghofer, B. I. Santner, E. Marth, and R. E. Stauber. 2002. Fully automated detection of Hepatitis C virus RNA in serum and whole-blood samples. Clin. Diagn. Lab. Immunol. 9:1385-1388. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 152.Khan, A. S., and D. A. Ashford. 2001. Ready or not -preparedness for bioterrorism. N. Engl. J. Med. 345:287-289. [DOI] [PubMed] [Google Scholar]
  • 153.Khan, A. S., R. Thompson, C. Cao, and J. J. Valdes. 2003. Selection and characterization of peptide memitopes binding to ricin. Biotechnol. Lett. 25:1671-1675. [DOI] [PubMed] [Google Scholar]
  • 154.Kijek, T. M., C. A. Rossi, D. Moss, R. W. Parker, and, H. E. A. 2000. Rapid and sensitive immunomagnetic-electrochemiluminescent detection of staphyloccocal enterotoxin B. J. Immunol. Methods 236:9-17. [DOI] [PubMed] [Google Scholar]
  • 155.Kim, G.-H., A. G. Rand, and S. V. Letcher. 2003. Impedance characterization of a piezoelectric immunosensor, part II: Salmonella typhimurium detection using magnetic enhancement. Biosensor. Bioelect. 18:91-99. [DOI] [PubMed] [Google Scholar]
  • 156.Kim, H.-Y., C. R. Estes, A. G. Duncan, B. D. Wade, F. C. Cleary, C. R. Lloyd, W. R. Ellis, Jr., and L. S. Powers. 2004. Real-time detection of microbial contamination. Eng. Med. Biol. Mag. IEEE 23:122-129. [DOI] [PubMed] [Google Scholar]
  • 157.King, D., V. Luna, A. Cannons, J. Cattani, and P. Amuso. 2003. Performance assessment of three commercial assays for direct detection of Bacillus anthracis spores. J. Clin. Microbiol. 41:3454-3455. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 158.King, K. D., G. P. Anderson, K. E. Bullock, M. J. Regina, E. W. Saaski, and F. S. Ligler. 1999. Detecting staphylococcal enterotoxin B using an automated fiber optic biosensor. Biosensor. Bioelect. 14:163-170. [DOI] [PubMed] [Google Scholar]
  • 159.Kirby, R., E. J. Cho, B. Gehrkem, T. Bayer, Y. S. Park, D. P. Neikirk, J. T. McDevitt, and A. D. Ellington. 2004. Aptamer-based sensor arrays for the detection and quantitation of proteins. Anal. Chem. 76:4066-4075. [DOI] [PubMed] [Google Scholar]
  • 160.Klietmann, W. F., and K. L. Ruoff. 2001. Bioterrorism: Implications for the clinical microbiologist. Clin. Microbiol. Rev. 14:364-381. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 161.Koshkin, A. A., S. K. Singh, P. Nielsen, V. K. Rajwanshi, R. Kumar, M. Meldgaard, C. E. Olsen, and J. Wengel. 1998. LNA (locked nucleic acids): synthesis of the adenine, cytosine, guanine, 5-methylcytosine, thymine and uracil bicyclonucleoside monomers, oligomerisation, and unprecedented nucleic acid recognition. Tetrahedron 54:3607-3630. [Google Scholar]
  • 162.Kovacs, G. T. A. 2003. Electronic sensors with living cellular components. Proc. IEEE 91:915-929. [Google Scholar]
  • 163.Kramer, K., M. Fiedler, A. Skerra, and B. Hock. 2002. A generic strategy for subcloning antibody variable regions from the scFv phage display vector pCANTAB 5 E into pASK85 permits the economical production of Fab fragments and leads to improved recombinant immunoglobulin stability. Biosensor. Bioelect. 17:305-313. [DOI] [PubMed] [Google Scholar]
  • 164.Kramer, M. F., and D. V. Lim. 2004. A rapid and automated fiber optic-based biosensor assay for the detection of Salmonella in spent irrigation water used in the sprouting of sprout seeds. J. Food Prot. 67:46-52. [DOI] [PubMed] [Google Scholar]
  • 165.Kramer, M. F., T. B. Tims, and D. V. Lim. 2002. Recovery of Escherichia coli O157:H7 from optical waveguides used for rapid biosensor detection. J. Rapid Methods Automat. Microbiol. 10:93-106. [Google Scholar]
  • 166.Krebs, B., R. Rauchenberger, S. Reiffert, C. Rothe, M. Tesar, E. Thomassen, M. Cao, T. Dreier, D. Fischer, and A. Hoöβ. 2001. High-throughput generation and engineering of recombinant human antibodies. J. Immunol. Methods 254:67-84. [DOI] [PubMed] [Google Scholar]
  • 167.Kulesh, D. A., R. O. Baker, B. M. Loveless, D. Norwood, S. H. Zwiers, E. Mucker, C. Hartmann, R. Herrera, D. Miller, D. Christensen, L. P. Wasieloski, Jr., J. Huggins, and P. B. Jahrling. 2004. Smallpox and pan-orthopox virus detection by real-time 3′-minor groove binder TaqMan assays on the Roche LightCycler and the Cepheid Smart Cycler platforms. J. Clin. Microbiol. 42:601-609. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 168.Lambert, A. J., R. S. Nasci, B. C. Cropp, D. A. Martin, B. C. Rose, B. J. Russell, and R. S. Lanciotti. 2005. Nucleic acid amplification assays for detection of La Crosse virus RNA. J. Clin. Microbiol. 43:1885-1889. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 169.Lampel, K., D. Dyer, L. Kornegay, and P. Orlandi. 2004. Detection of Bacillus spores using PCR and FTA filters. J. Food Prot. 67:1036-1038. [DOI] [PubMed] [Google Scholar]
  • 170.Lapizco-Encinas, B., B. Simmons, E. Cummings, and Y. Fintschenko. 2004. Insulator-based dielectrophoresis for the selective concentration and separation of live bacteria in water. Electrophoresis 25:1695-1704. [DOI] [PubMed] [Google Scholar]
  • 171.Lay, J. O., Jr. 2001. MALDI-TOF mass spectrometry of bacteria. Mass Spectrom. Rev. 20:172-194. [DOI] [PubMed] [Google Scholar]
  • 172.Leclercq, A., A. Guiyoule, M. El Lioui, E. Carniel, and J. Decallonne. 2000. High homogeneity of the Yersinia pestis fatty acid composition. J. Clin. Microbiol. 38:1545-1551. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 173.Lee, H., and S. K. R. Williams. 2003. Analysis of whole bacterial cells by flow field-flow fractionation and matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. Anal. Chem. 75:2746-2752. [DOI] [PubMed] [Google Scholar]
  • 174.Lee, J., and R. A. Deininger. 2004. A rapid screening method for the detection of viable spores in powder using bioluminescence. Luminescence 19:209-211. [DOI] [PubMed] [Google Scholar]
  • 175.Lee, L. Y., S. L. Ong, J. Y. Hu, W. J. Ng, Y. Feng, X. Tan, and S. W. Wong. 2004. Use of semiconductor quantum dots for photostable immunofluorescence labeling of Cryptosporidium parvum. Appl. Environ. Microbiol. 70:5732-5736. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 176.le Roux, W. J., D. Masoabi, C. M. de Wet, and S. N. Venter. 2004. Evaluation of a rapid polymerase chain reaction based identification technique for Vibrio cholerae isolates. Water Sci. Technol. 50:229-232. [PubMed] [Google Scholar]
  • 177.Lichlyter, D. J., S. A. Grant, and O. Soykan. 2003. Development of a novel FRET immunosensor technique. Biosensor. Bioelect. 19:219-226. [DOI] [PubMed] [Google Scholar]
  • 178.Ligler, F. S., M. Breimer, J. P. Golden, D. A. Nivens, J. P. Dodson, T. M. Green, D. P. Haders, and O. A. Sadik. 2002. Integrating waveguide biosensor. Anal. Chem. 74:713-719. [DOI] [PubMed] [Google Scholar]
  • 179.Ligler, F. S., C. R. Taitt, L. C. Shriver-Lake, K. E. Sapsford, Y. Shubin, and J. P. Golden. 2003. Array biosensor for detection of toxins. Anal. Bioanal. Chem. 377:469-477. [DOI] [PubMed] [Google Scholar]
  • 180.Lim, D. V. 2003. Detection of microorganisms and toxins with evanescent wave fiber-optic biosensors. Proc. IEEE 91:902-907. [Google Scholar]
  • 181.Lim, D. V. 2000. Rapid pathogen detection in the new millennium. Nat. Food Process. Assoc. J. 2:13-17. [Google Scholar]
  • 182.Lin, B., M. T. Vahey, D. Thach, D. A. Stenger, and J. J. Pancrazio. 2003. Biological threat detection via host gene expression profiling. Clin. Chem. 49:1045-1049. [DOI] [PubMed] [Google Scholar]
  • 183.Linde, H.-J., H. Neubauer, H. Meyer, S. Aleksic, and N. Lehn. 1999. Identification of Yersinia species by the Vitek GNI card. J. Clin. Microbiol. 37:211-214. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 184.Lindqvist, R. 1997. Preparation of PCR samples from food by a rapid and simple centrifugation technique evaluated by detection of Escherichia coli O157:H7. Int. J. Food Microbiol. 37:73-82. [DOI] [PubMed] [Google Scholar]
  • 185.Lindqvist, R., B. Norling, and S. Lambertz. 1997. A rapid sample preparation method for PCR detection of food pathogens based on buoyant density centrifugation. Lett. Appl. Microbiol. 24:306-310. [DOI] [PubMed] [Google Scholar]
  • 186.Liu, R. H., J. Yang, R. Lenigk, J. Bonanno, and P. Grodzinski. 2004. Self-contained, fully integrated biochip for sample preparation, polymerase chain reaction amplification, and DNA microarray detection. Anal. Chem. 76:1824-1831. [DOI] [PubMed] [Google Scholar]
  • 187.Livak, K., S. Flood, J. Marmaro, W. Giusti, and K. Deetz. 1995. Oligonucleotides with fluorescent dyes at opposite ends provide a quenched probe system useful for detecting PCR product and nucleic acid hybridization. PCR Methods Appl. 4:357-362. [DOI] [PubMed] [Google Scholar]
  • 188.Logan, N. A., J. A. Carman, J. Melling, and R. C. Berkeley. 1985. Identification of Bacillus anthracis by API tests. J. Med. Microbiol. 20:75-85. [DOI] [PubMed] [Google Scholar]
  • 189.Longoria, C. C., and G. A. Gonzalez. 1987. FiltraCheck-UTI, a rapid, disposable system for detection of bacteriuria. J. Clin. Microbiol. 25:926-928. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 190.Luna, V. A., D. King, C. Davis, T. Rycerz, M. Ewert, A. Cannons, P. Amuso, and J. Cattani. 2003. Novel sample preparation method for safe and rapid detection of Bacillus anthracis spores in environmental powders and nasal swabs. J. Clin. Microbiol. 41:1252-1255. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 191.Magan, N., A. Pavlou, and I. Chrysanthakis. 2001. Milk-sense: a volatile sensing system recognises spoilage bacteria and yeasts in milk. Sensor Acutuator B Chem. 72:28-34. [Google Scholar]
  • 192.Mahbubani, M. H., F. W. Schaefer, III, D. D. Jones, and A. K. Bej. 1998. Detection of Giardia in environmental waters by immuno-PCR amplification methods. Curr. Microbiol. 36:107-113. [DOI] [PubMed] [Google Scholar]
  • 193.Makino, S.-i., and H.-i. Cheun. 2003. Application of the real-time PCR for the detection of airborne microbial pathogens in reference to the anthrax spores. J. Microbiol. Methods 53:141-147. [DOI] [PubMed] [Google Scholar]
  • 194.Maruyama, F., T. Kenzaka, N. Yamaguchi, K. Tani, and M. Nasu. 2003. Detection of bacteria carrying the stx2 gene by in situ loop-mediated isothermal amplification. Appl. Environ. Microbiol. 69:5023-5028. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 195.Mason, H.-Y., C. Lloyd, M. Dice, R. Sinclair, W. Ellis, Jr., and L. Powers. 2003. Taxonomic identification of microorganisms by capture and intrinsic fluorescence detection. Biosensor. Bioelect. 18:521-527. [DOI] [PubMed] [Google Scholar]
  • 196.McBride, M. T., S. Gammon, M. Pitesky, T. W. O'Brien, T. Smith, J. Aldrich, R. G. Langlois, B. Colston, and K. S. Venkateswaran. 2003. Multiplexed liquid arrays for simultaneous detection of simulants of biological warfare agents. Anal. Chem. 75:1924-1930. [DOI] [PubMed] [Google Scholar]
  • 197.McBride, M. T., D. Masquelier, B. J. Hindson, A. J. Makarewicz, S. Brown, K. Burris, T. Metz, R. G. Langlois, K. W. Tsang, R. Bryan, D. Anderson, K. S. Venkateswaran, F. P. Milanovich, and B. W. Colston, Jr. 2003. Autonomous detection of aerosolized Bacillus anthracis and Yersinia pestis. Anal. Chem. 75:5293-5299. [DOI] [PubMed] [Google Scholar]
  • 198.McFadden, P. 2002. Biosensors: broadband biodetection: Holmes on a chip. Science 297:2075-2076. [DOI] [PubMed] [Google Scholar]
  • 199.McMillan, W. 2002. Real-time point-of-care molecular detection of infectious disease agents. Am. Clin. Lab. 21:29-31. [PubMed] [Google Scholar]
  • 200.Mehrvar, M., C. Bis, J. M. Scharer, M. M. Young, and J. H. Luong. 2000. Fiber-optic biosensors-trends and advances. Anal. Sci. 16:677-692. [Google Scholar]
  • 201.Mengelle, C., K. Sandres-Saune, C. Pasquier, L. Rostaing, J. M. Mansuy, M. Marty, I. Da Silva, M. Attal, P. Massip, and J. Izopet. 2003. Automated extraction and quantification of human cytomegalovirus DNA in whole blood by real-time PCR assay. J. Clin. Microbiol. 41:3840-3845. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 202.Mittelmann, A. S., E. Z. Ron, and J. Rishpon. 2002. Amperometric quantification of total coliforms and specific detection of Escherichia coli. Anal. Chem. 74:903-907. [DOI] [PubMed] [Google Scholar]
  • 203.Monk, D. J., and D. R. Walt. 2004. Optical fiber-based biosensors. Anal. Bioanal. Chem. 379:931-945. [DOI] [PubMed] [Google Scholar]
  • 204.Moore, C., S. Hibbitts, O. Neil, S. A. Corden, G. Harrison, J. Fox, C. Gelder, and D. Westmoreland. 2004. Development and evaluation of a real-time nucleic acid sequence based amplification assay for rapid detection of influenza A. J. Med. Virol. 74:619-628. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 205.Morales-Morales, H. A., G. Vidal, J. Olszewski, C. M. Rock, D. Dasgupta, K. H. Oshima, and G. B. Smith. 2003. Optimization of a reusable hollow-fiber ultrafilter for simultaneous concentration of enteric bacteria, protozoa, and viruses from water. Appl. Environ. Microbiol. 69:4098-4102. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 206.Morata, P., M. I. Queipo-Ortuno, and J. de Dios Colmenero. 1998. Strategy for optimizing DNA amplification in a peripheral blood PCR assay used for diagnosis of human brucellosis. J. Clin. Microbiol. 36:2443-2446. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 207.Moreno-Bondi, M., J. Alarie, and T. Vo-Dinh. 2003. Multi-analyte analysis system using an antibody-based biochip. Anal. Bioanal. Chem. 375:120-124. [DOI] [PubMed] [Google Scholar]
  • 208.Mumy, K. L., and R. H. Findlay. 2004. Convenient determination of DNA extraction efficiency using an external DNA recovery standard and quantitative-competitive PCR. J. Microbiol. Methods 57:259-268. [DOI] [PubMed] [Google Scholar]
  • 209.Murray, P., E. J. Baron, J. H. Jorgensen, M. A. Pfaller, and R. H. Yolken (ed.). 2003. Manual of clinical microbiology, 8th ed. ASM Press, Washington, D.C.
  • 210.Mygind, T., L. Ostergaard, S. Birkelund, J. Lindholt, and G. Christiansen. 2003. Evaluation of five DNA extraction methods for purification of DNA from atherosclerotic tissue and estimation of prevalence of Chlamydia pneumoniae in tissue from a Danish population undergoing vascular repair. BMC Microbiol. 3:19. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 211.Naimushin, A. N., S. D. Soelberg, D. K. Nguyen, L. Dunlap, D. Bartholomew, J. Elkind, J. Melendez, and C. E. Furlong. 2002. Detection of Staphylococcus aureus enterotoxin B at femtomolar levels with a miniature integrated two-channel surface plasmon resonance (SPR) sensor. Biosensor. Bioelect. 17:573-584. [DOI] [PubMed] [Google Scholar]
  • 212.Narang, U., G. P. Anderson, F. S. Ligler, and J. Burans. 1997. Fiber optic-based biosensor for ricin. Biosensor. Bioelect. 12:937-945. [DOI] [PubMed] [Google Scholar]
  • 213.Nazarenko, I., B. Lowe, M. Darfler, P. Ikonomi, D. Schuster, and A. Rashtchian. 2002. Multiplex quantitative PCR using self-quenched primers labeled with a single fluorophore. Nucleic Acids Res. 30:e37. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 214.Neufeld, T., A. Schwartz-Mittelmann, D. Biran, E. Z. Ron, and J. Rishpon. 2003. Combined phage typing and amperometric detection of released enzymatic activity for the specific identification and quantification of bacteria. Anal. Chem. 75:580-585. [DOI] [PubMed] [Google Scholar]
  • 215.Ni, J., R. J. Lipert, G. B. Dawson, and M. D. Porter. 1999. Immunoassay readout method using extrinsic Raman labels adsorbed on immunogold colloids. Anal. Chem. 71:4903-4908. [DOI] [PubMed] [Google Scholar]
  • 216.Niedbala, R. S., H. Feindt, K. Kardos, T. Vail, J. Burton, B. Bielska, S. Li, D. Milunic, P. Bourdelle, and R. Vallejo. 2001. Detection of analytes by immunoassay using up-converting phosphor technology. Anal. Biochem. 293:22-30. [DOI] [PubMed] [Google Scholar]
  • 217.Notomi, T., H. Okayama, H. Masubuchi, T. Yonekawa, K. Watanabe, N. Amino, and T. Hase. 2000. Loop-mediated isothermal amplification of DNA. Nucleic Acids Res. 28:e63. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 218.Obika, S., D. Nanbu, Y. Hari, K. Morio, Y. In, T. Ishida, and T. Imanishi. 1997. Synthesis of 2′-O,4′-C-methyleneuridine and -cytidine. Novel bicyclic nucleosides having a fixed C3,-endo sugar puckering. Tetrahedron Lett. 38:8735-8738. [Google Scholar]
  • 219.Odumeru, J. A., M. Steele, L. Fruhner, C. Larkin, J. Jiang, E. Mann, and W. B. McNab. 1999. Evaluation of accuracy and repeatability of identification of food-borne pathogens by automated bacterial identification systems. J. Clin. Microbiol. 37:944-949. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 220.Pancrazio, J. J., P. P. Bey, Jr., D. S. Cuttino, J. K. Kusel, D. A. Borkholder, K. M. Shaffer, G. T. A. Kovacs, and D. A. Stenger. 1998. Portable cell-based biosensor system for toxin detection. Sensor Acutuator B Chem. 53:179-185. [Google Scholar]
  • 221.Pancrazio, J. J., S. A. Gray, Y. S. Shubin, N. Kulagina, D. S. Cuttino, K. M. Shaffer, K. Eisemann, A. Curran, B. Zim, G. W. Gross, and T. J. O'Shaughnessy. 2003. A portable microelectrode array recording system incorporating cultured neuronal networks for neurotoxin detection. Biosensor. Bioelect. 18:1339-1347. [DOI] [PubMed] [Google Scholar]
  • 222.Parida, M., G. Posadas, S. Inoue, F. Hasebe, and K. Morita. 2004. Real-time reverse transcription loop-mediated isothermal amplification for rapid detection of West Nile virus. J. Clin. Microbiol. 42:257-263. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 223.Park, T., S. Lee, G. H. Seong, J. Choo, E. K. Lee, Y. S. Kim, W. H. Ji, S. Y. Hwang, D.-G. Gweon, and S. Lee. 2005. Highly sensitive signal detection of duplex dye-labelled DNA oligonucleotides in a PDMS microfluidic chip: confocal surface-enhanced Raman spectroscopic study. Lab. Chip 5:437-442. [DOI] [PubMed] [Google Scholar]
  • 224.Pavlou, A., A. P. F. Turner, and N. Magan. 2002. Recognition of anaerobic bacterial isolates in vitro using electronic nose technology. Lett. Appl. Microbiol. 35:366-369. [DOI] [PubMed] [Google Scholar]
  • 225.Pavlou, A. K., N. Magan, J. M. Jones, J. Brown, P. Klatser, and A. P. F. Turner. 2004. Detection of Mycobacterium tuberculosis (TB) in vitro and in situ using an electronic nose in combination with a neural network system. Biosensor. Bioelect. 20:538-544. [DOI] [PubMed] [Google Scholar]
  • 226.Pavlou, A. K., N. Magan, C. McNulty, J. M. Jones, D. Sharp, J. Brown, and A. P. F. Turner. 2002. Use of an electronic nose system for diagnoses of urinary tract infections. Biosensor. Bioelect. 17:893-899. [DOI] [PubMed] [Google Scholar]
  • 227.Payne, M. J., and R. G. Kroll. 1991. Methods for the separation and concentration of bacteria from foods. Trends Food Sci. Technol. 2:315-319. [Google Scholar]
  • 228.Perch-Nielsen, I. R., D. D. Bang, C. R. Poulsen, J. El-Ali, and A. Wolff. 2003. Removal of PCR inhibitors using dielectrophoresis as a selective filter in a microsystem. Lab. Chip 3:212-216. [DOI] [PubMed] [Google Scholar]
  • 229.Peruski, A. H., L. H. Johnson III, and L. F. Peruski,, Jr. 2002. Rapid and sensitive detection of biological warfare agents using time-resolved fluorescence assays. J. Immunol. Methods 263:35-41. [DOI] [PubMed] [Google Scholar]
  • 230.Peruski, A. H., and L. F. Peruski, Jr.2003. Immunological methods for detection and identification of infectious disease and biological warfare agents. Clin. Diagn. Lab. Immunol. 10:506-513. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 231.Peruski, L. F., Jr., and A. H. Peruski. 2003. Rapid diagnostic assays in the genomic biology era: Detection and identfication of infectious disease and biological weapon agents. BioTechniques 35:840-846. [DOI] [PubMed] [Google Scholar]
  • 232.Petrenko, V. A., and I. B. Sorokulova. 2004. Detection of biological threats. A challenge for directed molecular evolution. J. Microbiol. Methods 58:147-168. [DOI] [PubMed] [Google Scholar]
  • 233.Petrenko, V. A., and V. J. Vodyanoy. 2003. Phage display for detection of biological threat agents. J. Microbiol. Methods 53:253-262. [DOI] [PubMed] [Google Scholar]
  • 234.Pham, H. M., C. Nakajima, K. Ohashi, and M. Onuma. 2005. Loop-mediated isothermal amplification for rapid detection of Newcastle disease virus. J. Clin. Microbiol. 43:1646-1650. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 235.Poon, L. L. M., C. S. W. Leung, K. H. Chan, J. H. C. Lee, K. Y. Yuen, Y. Guan, and J. S. M. Peiris. 2005. Detection of human influenza A viruses by loop-mediated isothermal amplification. J. Clin. Microbiol. 43:427-430. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 236.Radstrom, P., R. Knutsson, P. Wolffs, M. Lovenklev, and C. Lofstrom. 2004. Pre-PCR processing: strategies to generate PCR-compatible samples. Mol. Biotechnol. 26:133-146. [DOI] [PubMed] [Google Scholar]
  • 237.Raggam, R., E. Leitner, G. Muhlbauer, J. Berg, M. Stocher, A. Grisold, E. Marth, and H. Kessler. 2002. Qualitative detection of Legionella species in bronchoalveolar lavages and induced sputa by automated DNA extraction and real-time polymerase chain reaction. Med. Microbiol. Immunol. 191:119-125. [DOI] [PubMed] [Google Scholar]
  • 238.Rau, D., K. Kramer, and B. Hock. 2002. Single-chain Fv antibody-alkaline phosphatase fusion proteins produced by one-step cloning as rapid detection tools for ELISA. Immunoassay Immunochem. 23:129-143. [DOI] [PubMed] [Google Scholar]
  • 239.Relman, D. A. 2003. Shedding light on microbial detection. N. Engl. J. Med. 349:2162-2163. [DOI] [PubMed] [Google Scholar]
  • 240.Reschiglian, P., A. Zattoni, L. Cinque, B. Roda, F. Dal Piaz, A. Roda, M. H. Moon, and B. R. Min. 2004. Hollow-fiber flow field-flow fractionation for whole bacteria analysis by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. Anal. Chem. 76:2103-2111. [DOI] [PubMed] [Google Scholar]
  • 241.Rider, T. H., M. S. Petrovick, F. E. Nargi, J. D. Harper, E. D. Schwoebel, R. H. Mathews, D. J. Blanchard, L. T. Bortolin, A. M. Young, J. Chen, and M. A. Hollis. 2003. A B cell-based sensor for rapid identification of pathogens. Science 301:213-215. [DOI] [PubMed] [Google Scholar]
  • 242.Riegler, J., and T. Nann. 2004. Application of luminescent nanocrystals as labels for biological molecules. Anal. Bioanal. Chem. 379:913-919. [DOI] [PubMed] [Google Scholar]
  • 243.Rijpens, N., L. Herman, F. Vereecken, G. Jannes, J. De Smedt, and L. De Zutter. 1999. Rapid detection of stressed Salmonella spp. in dairy and egg products using immunomagnetic separation and PCR. Int. J. Food Microbiol. 46:37-44. [DOI] [PubMed] [Google Scholar]
  • 244.Rippin, K. P., W. C. Stinsin, J. Eisenstadt, and J. A. Washington. 1995. Clinical evaluation of the slide centrifuge (cytospin) gram's stained smear for the detection of bacteria and comparison with the FiltraCheck-UTI and UTIscreen. Am. J. Clin. Pathol. 103:316-319. [DOI] [PubMed] [Google Scholar]
  • 245.Rodriguez-Lazaro, D., J. Lloyd, A. Herrewegh, J. Ikonomopoulos, M. D'Agostino, M. Pla, and N. Cook. 2004. A molecular beacon-based real-time NASBA assay for detection of Mycobacterium avium subsp. paratuberculosis in water and milk. FEMS Microbiol. Lett. 237:119-126. [DOI] [PubMed] [Google Scholar]
  • 246.Romaschin, A. D., D. M. Harris, M. B. Ribeiro, J. Paice, D. M. Foster, P. M. Walker, and J. C. Marshall. 1998. A rapid assay of endotoxin in whole blood using autologous neutrophil dependent chemiluminescence. J. Immunol. Methods 212:169-185. [DOI] [PubMed] [Google Scholar]
  • 247.Rose, L., B. Jensen, A. Peterson, S. N. Banerjee, and M. J. Aruduino. 2004. Swab materials and Bacillus anthracis spore recovery from nonporous surfaces. Emerg. Infect. Dis. 10:1023-1029. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 248.Ruan, C., K. Zeng, O. K. Varghese, and C. A. Grimes. 2003. Magnetoelastic immunosensors: amplified mass immunosorbent assay for detection of Escherichia coli O157:H7. Anal. Chem. 75:6494-6498. [DOI] [PubMed] [Google Scholar]
  • 249.Ruan, C., K. Zeng, O. K. Varghese, and C. A. Grimes. 2004. A staphylococcal enterotoxin B magnetoelastic immunosensor. Biosensor. Bioelect. 20:585-591. [DOI] [PubMed] [Google Scholar]
  • 250.Sanderson, W., M. Jein, L. Taylor, B. Curwin, G. Kinnes, T. Seitz, T. Popovic, H. Holmes, M. Kellum, S. McAllister, D. Whaley, E. Tupin, T. Walker, J. Freed, D. Small, B. Klusaritz, and J. Bridges. 2002. Surface sampling methods for Bacillus anthracis spore contamination. Emerg. Infect. Dis. 8:1145-1151. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 251.Sapsford, K. E., A. Rasooly, C. R. Taitt, and F. S. Ligler. 2004. Detection of Campylobacter and Shigella species in food samples using an array biosensor. Anal. Chem. 76:433-440. [DOI] [PubMed] [Google Scholar]
  • 252.Schiffman, S. S., D. W. Wyrick, R. Gutierrez-Osuna, and H. T. Nagle. 2000. Effectiveness of an electronic nose for monitoring bacterial and fungal growth, p. 173-180. In J. W. Gardner and K. C. Persaud (ed.), Electronic noses and olfaction 2000: Proceedings of the Seventh International Symposium on Olfaction and Electronic Noses. Institute of Physics, Brighton, United Kingdom.
  • 253.Schmaljohn, C., Y. Cui, S. Kerby, D. Pennock, and K. Spik. 1999. Production and characterization of human monoclonal antibody Fab fragments to vaccinia virus from a phage-display combinatorial library. Virology 258:189-200. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 254.Schmid, M., R. Oehme, G. Schalasta, S. Brockmann, P. Kimmig, and G. Enders. 2004. Fast detection of noroviruses using a real-time PCR assay and automated sample preparation. BMC Infect. Dis. 4:15. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 255.Schneider, P., L. Wolters, G. Schoone, H. Schallig, P. Sillekens, R. Hermsen, and R. Sauerwein. 2005. Real-time nucleic acid sequence-based amplification is more convenient than real-time PCR for quantification of Plasmodium falciparum. J. Clin. Microbiol. 43:402-405. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 256.Schulze, H., S. Vorlova, F. Villatte, T. T. Bachmann, and R. D. Schmid. 2003. Design of acetylcholinesterases for biosensor applications. Biosensor. Bioelect. 18:201-209. [DOI] [PubMed] [Google Scholar]
  • 257.Seki, M., Y. Yamashita, H. Torigoe, H. Tsuda, S. Sato, and M. Maeno. 2005. Loop-mediated isothermal amplification method targeting the lytA gene for detection of Streptococcus pneumoniae. J. Clin. Microbiol. 43:1581-1586. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 258.Sharma, N. K., P. W. Doyle, S. A. Gerbasi, and J. H. Jessop. 1990. Identification of Yersinia species by the API 20E. J. Clin. Microbiol. 28:1443-1444. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 259.Sharma, V., A. Narayanan, T. Rengachari, G. C. Temes, F. Chaplen, and U.-K. Moon. 2005. A low-cost, portable generic biotoxicity assay for environmental monitoring applications. Biosensor. Bioelect. 20:2218-2227. [DOI] [PubMed] [Google Scholar]
  • 260.Shelton, D. R., and J. S. Karns. 2001. Quantitative detection of Escherichia coli O157 in surface waters by using immunomagnetic electrochemiluminescence. Appl. Environ. Microbiol. 67:2908-2915. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 261.Slavik, R., J. Homola, and E. Brynda. 2002. A miniature fiber optic surface plasmon resonance sensor for fast detection of staphylococcal enterotoxin B. Biosensor. Bioelect. 17:591-595. [DOI] [PubMed] [Google Scholar]
  • 262.Smith, D., B. D. Collins, J. Heil, and T. H. Koch. 2003. Sensitivity and specificity of photoaptamer probes. Mol. Cell Proteomics 2:11-18. [DOI] [PubMed] [Google Scholar]
  • 263.Snyder, J. W., and W. Check. 2001. Bioterrorism threats to our future: the role of the clinical microbiology laboratory in detection, identification, and confirmation of biological agents. ASM Press, Washington, D.C. [PubMed]
  • 264.Song, J. M., J. Mobley, and T. Vo-Dinh. 2003. Detection of bacterial pathogen DNA using an integrated complementary metal oxide semiconductor microchip system with capillary array electrophoresis. J. Chromatogr. B 783:501-508. [DOI] [PubMed] [Google Scholar]
  • 265.Song, T., C. Toma, N. Nakasone, and M. Iwanaga. 2005. Sensitive and rapid detection of Shigella and enteroinvasive Escherichia coli by a loop-mediated isothermal amplification method. FEMS Microbiol. Lett. 243:259-263. [DOI] [PubMed] [Google Scholar]
  • 266.Song, Y., R. Yang, Z. Guo, M. Zhang, X. Wang, and F. Zhou. 2000. Distinctness of spore and vegetative cellular fatty acid profiles of some aerobic endospore-forming bacilli. J. Microbiol. Methods 39:225-241. [DOI] [PubMed] [Google Scholar]
  • 267.Spanova, A., B. Rittich, R. Karpiskova, L. Cechova, and D. Skapova. 2000. PCR identification of Salmonella cells in food and stool samples after immunomagnetic separation. Bioseparation 9:379-384. [DOI] [PubMed] [Google Scholar]
  • 268.Srinivasan, A., C. N. Kraus, D. DeShazer, P. M. Becker, J. D. Dick, L. Spacek, J. G. Bartlett, W. R. Byrne, and D. L. Thomas. 2001. Glanders in a military research microbiologist. N. Engl. J. Med. 345:256-258. [DOI] [PubMed] [Google Scholar]
  • 269.Stenger, D. A., J. D. Andreadis, G. J. Voraa, and J. J. Pancrazio. 2002. Potential applications of DNA microarrays in biodefense-related diagnostics. Curr. Opin. Biotechnol. 13:208-212. [DOI] [PubMed] [Google Scholar]
  • 270.Stenger, D. A., G. W. Gross, E. W. Keefer, K. M. Shaffer, J. D. Andreadis, W. Ma, and J. J. Pancrazio. 2001. Detection of physiologically active compounds using cell-based biosensors. Trends Biotechnol. 19:304-309. [DOI] [PubMed] [Google Scholar]
  • 271.Stetzenbach, L. D., M. P. Buttner, and P. Cruz. 2004. Detection and enumeration of airborne biocontaminants. Curr. Opin. Biotechnol. 15:170-174. [DOI] [PubMed] [Google Scholar]
  • 272.Stevens, K. A., and L. A. Jaykus. 2004. Bacterial separation and concentration from complex sample matrices: a review. Crit. Rev. Microbiol. 30:7-24. [DOI] [PubMed] [Google Scholar]
  • 273.Stopa, P. J., D. Tieman, P. A. Coon, M. M. Milton, and D. Paterno. 1999. Detection of biological aerosols by luminescence techniques. Field Anal. Chem. Technol. 3:283-290. [Google Scholar]
  • 274.Stratis-Cullum, D. N., G. D. Griffin, J. Mobley, A. A. Vass, and T. Vo-Dinh. 2003. A miniature biochip system for detection of aerosolized Bacillus globigii spores. Anal. Chem. 75:275-280. [DOI] [PubMed] [Google Scholar]
  • 275.Su, X. L., and Y. Li. 2004. Quantum dot biolabeling coupled with immunomagnetic separation for detection of Escherichia coli O157:H7. Anal. Chem. 76:4806-4810. [DOI] [PubMed] [Google Scholar]
  • 276.Taitt, C. R., G. P. Anderson, B. M. Lingerfelt, M. J. Feldstein, and F. S. Ligler. 2002. Nine-analyte detection using an array-based biosensor. Anal. Chem. 74:6114-6120. [DOI] [PubMed] [Google Scholar]
  • 277.Taitt, C. R., J. P. Golden, Y. S. Shubin, L. C. Shriver-Lake, K. E. Sapsford, A. Rasooly, and F. S. Ligler. 2004. A portable array biosensor for detecting multiple analytes in complex samples. Microb. Ecol. 47:175-185. [DOI] [PubMed] [Google Scholar]
  • 278.Tempelman, L. A., K. D. King, G. P. Anderson, and F. S. Ligler. 1996. Quantitating staphylococcal enterotoxin B in diverse media using a portable fiber-optic biosensor. Anal. Biochem. 233:50-57. [DOI] [PubMed] [Google Scholar]
  • 279.Thai, H. T. C., M. Q. Le, C. D. Vuong, M. Parida, H. Minekawa, T. Notomi, F. Hasebe, and K. Morita. 2004. Development and evaluation of a novel loop-mediated isothermal amplification method for rapid detection of severe acute respiratory syndrome coronavirus. J. Clin. Microbiol. 42:1956-1961. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 280.Tims, T. B., and D. V. Lim. 2003. Confirmation of viable E. coli O157:H7 by enrichment and PCR after rapid biosensor detection. J. Microbiol. Methods 55:141-147. [DOI] [PubMed] [Google Scholar]
  • 281.Tims, T. B., and D. V. Lim. 2004. Rapid detection of Bacillus anthracis spores directly from powders with an evanescent wave fiber-optic biosensor. J. Microbiol. Methods 59:127-130. [DOI] [PubMed] [Google Scholar]
  • 282.Toye, B., W. Woods, M. Bobrowska, and K. Ramotar. 1998. Inhibition of PCR in genital and urine specimens submitted for Chlamydia trachomatis testing. J. Clin. Microbiol. 36:2356-2358. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 283.Tucker, J. B. 1999. Historical trends related to bioterrorism: an empirical analysis. Emerg. Infect. Dis. 5:498-504. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 284.Turnbough, C. L., Jr. 2003. Discovery of phage display peptide ligands for species-specific detection of Bacillus spores. J. Microbiol. Methods 53:263-271. [DOI] [PubMed] [Google Scholar]
  • 285.Turnbull, W., and P. Abhayaratne. 2003. 2002 WMD terrorism chronology: incidents involving sub-national actors and chemical, biological, radiological, and nuclear materials. Monterey Institute of International Studies: Center for Nonproliferation Studies, Washington, D.C.
  • 286.Turner, A. P., and N. Magan. 2004. Electronic noses and disease diagnostics. Nat. Rev. Microbiol. 2:161-166. [DOI] [PubMed] [Google Scholar]
  • 287.Tyagi, S., and F. R. Kramer. 1996. Molecular beacons: Probes that fluoresce upon hybridization. Nat. Biotechnol. 14:303-308. [DOI] [PubMed] [Google Scholar]
  • 288.Uhl, J. R., C. A. Bell, L. M. Sloan, M. J. Espy, T. F. Smith, J. E. Rosenblatt, and F. R. Cockerill III. 2002. Application of rapid-cycle real-time polymerase chain reaction for the detection of microbial pathogens: the Mayo-Roche rapid anthrax test. Mayo Clin. Proc. 77:673-680. [DOI] [PubMed] [Google Scholar]
  • 289.U.S. Environmental Protection Agency. 1999. Method 1622: Cryptosporidium in water by filtration/IMS/FA EPA-821-R-01-026. U.S. Environmental Protection Agency; Office of Water, Washington, D.C.
  • 290.U.S. Environmental Protection Agency. 1999. Method 1623: Cryptosporidium and Giardia in water by filtration/IMS/FA EPA-821-R-99-006. U.S. Environmental Protection Agency, Office of Water, Washington, D.C.
  • 291.Uyttendaele, M., I. Van Hoorde, and J. Debevere. 2000. The use of immuno-magnetic separation (IMS) as a tool in a sample preparation method for direct detection of L. monocytogenes in cheese. Int. J. Food Microbiol. 54:205-212. [DOI] [PubMed] [Google Scholar]
  • 292.Valentine, N. B., J. H. Wahl, M. T. Kingsley, and K. L. Wahl. 2002. Direct surface analysis of fungal species by matrix-assisted laser desorption/ionization mass spectrometry. Rapid Comm. Mass Spectrom. 16:1352-1357. [DOI] [PubMed] [Google Scholar]
  • 293.van Courtland Moon, J. E. (ed.). 1991. Biological and toxin weapons: research, development, and use from the middle ages to 1945. Stockholm International Peace Research Institute, Stockholm, Sweden.
  • 294.van de Rijke, F., H. Zijlmans, S. Li, T. Vail, A. K. Raap, R. S. Niedbala, and H. J. Tanke. 2001. Up-converting phosphor reporters for nucleic acid microarrays. Nat. Biotechnol. 19:273-276. [DOI] [PubMed] [Google Scholar]
  • 295.Van Ert, M. N., S. A. Hofstadler, Y. Jiang, J. D. Busch, D. M. Wagner, J. J. Drader, D. J. Ecker, J. C. Hannis, L. Y. Huynh, J. M. Schupp, T. S. Simonson, and P. Keim. 2004. Mass spectrometry provides accurate characterization of two genetic marker types in Bacillus anthracis. BioTechniques 37:642-651. [DOI] [PubMed] [Google Scholar]
  • 296.Van Kessel, J. S., J. S. Karns, and M. L. Perdue. 2003. Using a portable real-time PCR assay to detect Salmonella in raw milk. J. Food. Prot. 66:1762-1767. [PubMed] [Google Scholar]
  • 297.Verkooyen, R., A. Luijendijk, W. Huisman, W. Goessens, J. Kluytmans, J. van Rijsoort-Vos, and H. Verbrugh. 1996. Detection of PCR inhibitors in cervical specimens by using the AMPLICOR Chlamydia trachomatis assay. J. Clin. Microbiol. 34:3072-3074. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 298.Vignali, D. A. 2000. Multiplexed particle-based flow cytometric assays. J. Immunol. Methods 243:243-255. [DOI] [PubMed] [Google Scholar]
  • 299.Villatte, F., H. Schulze, R. Schmid, and T. Bachmann. 2002. A disposable acetylcholinesterase-based electrode biosensor to detect anatoxin-a(s) in water. Anal. Bioanal. Chem. 372:322-326. [DOI] [PubMed] [Google Scholar]
  • 300.Wahl, K. L., S. C. Wunschel, K. H. Jarman, N. B. Valentine, C. E. Petersen, M. T. Kingsley, K. A. Zartolas, and A. J. Saenz. 2002. Analysis of microbial mixtures by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. Anal. Chem. 74:6191-6199. [DOI] [PubMed] [Google Scholar]
  • 301.Warscheid, B., and C. Fenselau. 2004. A targeted proteomics approach to the rapid identification of bacterial cell mixtures by matrix-assisted laser desorption/ionization mass spectrometry. Proteomics 4:2877-2892. [DOI] [PubMed] [Google Scholar]
  • 302.Warscheid, B., K. Jackson, C. Sutton, and C. Fenselau. 2003. MALDI analysis of Bacilli in spore mixtures by applying a quadrupole ion trap time-of-flight tandem mass spectrometer. Anal. Chem. 75:5608-5617. [DOI] [PubMed] [Google Scholar]
  • 303.Watanabe, T., S. Tomita, M. Kudo, M. Kurokawa, A. Orino, A. Todo, and T. Chiba. 1998. Detection of Helicobacter pylori gene by means of immunomagnetic separation-based polymerase chain reaction in feces. Scand. J. Gastroenterol. 33:1140-1143. [DOI] [PubMed] [Google Scholar]
  • 304.Westin, L., C. Miller, D. Vollmer, D. Canter, R. Radtkey, M. Nerenberg, and J. P. O'Connell. 2001. Antimicrobial resistance and bacterial identification utilizing a microelectronic chip array. J. Clin. Microbiol. 39:1097-1104. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 305.Wheelis, M. 1991. Biological warfare before 1914, p. 8-34. In J. E. van Courtland Moon (ed.), Biological and toxin weapons: research, development, and use from the middle ages to 1945, vol. 1. Stockholm International Peace Research Institute, Stockholm, Sweden. [Google Scholar]
  • 306.Whitcombe, D., J. Theaker, S. P. Guy, T. Brown, and S. Little. 1999. Detection of PCR products using self-probing amplicons and fluorescence. Nat. Biotechnol. 17:804-807. [DOI] [PubMed] [Google Scholar]
  • 307.Williams, D. D., O. Benedek, and C. L. Turnbough, Jr. 2003. Species-specific peptide ligands for the detection of Bacillus anthracis spores. Appl. Environ. Microbiol. 69:6288-6293. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 308.Williams, T. L., D. Andrzejewski, J. O. Lay, Jr., and S. M. Musser. 2003. Experimental factors affecting the quality and reproducibility of MALDI TOF mass spectra obtained from whole bacteria cells. J. Am. Soc. Mass Spectrom. 14:342-351. [DOI] [PubMed] [Google Scholar]
  • 309.Wilmoth, B., M. Chu, and T. Quan. 1996. Identification of Yersinia pestis by BBL Crystal Enteric/Nonfermenter identification system. J. Clin. Microbiol. 34:2829-2830. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 310.Wittwer, C., K. Ririe, R. Andrew, D. David, R. Gundry, and U. Balis. 1997. The LightCycler™: A microvolume multisample fluorimeter with rapid temperature control. BioTechniques 22:176-181. [DOI] [PubMed] [Google Scholar]
  • 311.Wolffs, P., R. Knutsson, B. Norling, and P. Radstrom. 2004. Rapid quantification of Yersinia enterocolitica in pork samples by a novel sample preparation method, flotation, prior to real-time PCR. J. Clin. Microbiol. 42:1042-1047. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 312.Wolk, D. M., S. K. Schneider, N. L. Wengenack, L. M. Sloan, and J. E. Rosenblatt. 2002. Real-time PCR method for detection of Encephalitozoon intestinalis from stool specimens. J. Clin. Microbiol. 40:3922-3928. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 313.Woodbury, R. G., C. Wendin, J. Clendenning, J. Melendez, J. Elkind, D. Bartholomew, S. Brown, and C. E. Furlong. 1998. Construction of biosensors using a gold-binding polypeptide and a miniature integrated surface plasmon resonance sensor. Biosensor. Bioelect. 13:1117-1126. [DOI] [PubMed] [Google Scholar]
  • 314.Xu, S., X. Ji, W. Xu, X. Li, L. Wang, Y. Bai, B. Zhao, and Y. Ozaki. 2004. Immunoassay using probe-labelling immunogold nanoparticles with silver staining enhancement via surface-enhanced Raman scattering. Analyst 129:63-68. [DOI] [PubMed] [Google Scholar]
  • 315.Yang, J. M., J. Bell, Y. Huang, M. Tirado, D. Thomas, A. H. Forster, R. W. Haigis, P. D. Swanson, R. B. Wallace, B. Martinsons, and M. Krihak. 2002. An integrated, stacked microlaboratory for biological agent detection with DNA and immunoassays. Biosensor. Bioelect. 17:605-618. [DOI] [PubMed] [Google Scholar]
  • 316.Yao, Z.-P., P. A. Demirev, and C. Fenselau. 2002. Mass spectrometry-based proteolytic mapping for rapid virus identification. Anal. Chem. 74:2529-2534. [DOI] [PubMed] [Google Scholar]
  • 317.Yoo, J.-H., J.-H. Choi, S.-M. Choi, D. Lee, W.-S. Shin, W. Min, and C.-C. Kim. 2005. Application of nucleic acid sequence-based amplification for diagnosis of and monitoring the clinical course of invasive aspergillosis in patients with hematologic diseases. Clin. Infect. Dis. 40:392-398. [DOI] [PubMed] [Google Scholar]
  • 318.Yu, H. 1996. Enhancing immunoelectrochemiluminescence (IECL) for sensitive bacterial detection. J. Immunol. Methods 192:63-71. [DOI] [PubMed] [Google Scholar]
  • 319.Yu, H., and J. G. Bruno. 1996. Immunomagnetic-electrochemiluminescent detection of Escherichia coli O157 and Salmonella typhimurium in foods and environmental water samples. Appl. Environ. Microbiol. 62:587-592. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 320.Yu, L. S. L., S. A. Reed, and M. H. Golden. 2002. Time-resolved fluorescence immunoassay (TRFIA) for the detection of Escherichia coli O157:H7 in apple cider. J. Microbiol. Methods 49:63-68. [DOI] [PubMed] [Google Scholar]
  • 321.Zaytseva, N. V., R. A. Montagna, E. M. Lee, and A. J. Baeumner. 2004. Multi-analyte single-membrane biosensor for the serotype-specific detection of dengue virus. Anal. Bioanal. Chem. 380:46-53. [DOI] [PubMed] [Google Scholar]
  • 322.Zhang, X., M. A. Young, O. Lyandres, and R. P. Van Duyne. 2005. Rapid detection of an anthrax biomarker by surface-enhanced Raman spectroscopy. J. Am. Chem. Soc. 127:4484-4489. [DOI] [PubMed] [Google Scholar]
  • 323.Zhou, B., P. Wirsching, and K. D. Janda. 2002. Human antibodies against spores of the genus Bacillus: a model study for detection of and protection against anthrax and the bioterrorist threat. Proc. Natl. Acad. Sci. USA 99:5241-5246. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 324.Zhu, L., S. Ang, and W.-T. Liu. 2004. Quantum dots as a novel immunofluorescent detection system for Cryptosporidium parvum and Giardia lamblia. Appl. Environ. Microbiol. 70:597-598. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 325.Zuiderwijk, M., H. J. Tanke, R. S. Niedbala, and P. L. A. M. Corstjens. 2003. An amplification-free hybridization-based DNA assay to detect Streptococcus pneumoniae utilizing the up-converting phosphor technology. Clin. Biochem. 36:401-403. [DOI] [PubMed] [Google Scholar]

Articles from Clinical Microbiology Reviews are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES