Abstract
Background:
Gait impairment is a primary driver of disability in multiple sclerosis (MS). However, traditional disability measures are nonlinear and lack sensitivity to subtle changes in motor function. Objective, scalable tools for gait assessment are critically needed.
Objective:
To evaluate the agreement between a depth-sensing video system and wearable inertial measurement unit (IMU) sensors in quantifying gait metrics in people with MS (pwMS), to assess the relationship to patient-reported disability.
Methods:
Twenty-six pwMS completed a 6-meter walk wearing IMU sensors (Clario OPAL) and were simultaneously recorded with a depth-sensing camera (Microsoft Kinect). Disability was self-reported using the Patient Determined Disease Step (PDDS) scale.
Results:
Participants had a median age of 54 years (range 23 – 71), 69% were female, and the median disease duration was 65 months (0 – 382). Strong correlations with low differences were observed between devices for cadence (r>0.9, p<0.001; mean difference −5 steps/min), stride duration (r>0.9, p<0.001, mean difference 0.03-0.06 s), and gait speed (r=0.7 – 0.9, p<0.001, mean difference 0.04 – 0.15 m/s). Gait cycle phases were moderately correlated (r=0.4 – 0.7). PDDS scores correlated most strongly with speed (r= −0.61), with weaker correlations across gait parameters.
Interpretation:
Wearable and depth video systems yielded highly correlated spatiotemporal gait data obtained in a clinical setting and are more sensitive to motor abnormalities than the PDDS. These tools could potentially be widely employed for clinical gait assessment, democratizing access to sensitive qualitative measurement of disease progression in MS.
Keywords: Disease Steps (DS) scale, patient-reported outcome measurement, gait dysfunction, wearable inertial measurement unit (IMU) sensors, video monitoring
Introduction
Multiple sclerosis (MS) is a common neurologic disease, affecting approximately 300 per 100,000 individuals in the US.1,2 It is associated with progressive disability, with most patients requiring a gait aid within 15 years of diagnosis, reducing independence, opportunity for employment, and quality of life.3,4 Reliable and sensitive measures of motor impairment are critical to identify and track progression early in the disease course, to monitor treatment response, and to individualize rehabilitation interventions. The current gold standard for monitoring disability in people with MS (pwMS) is the Expanded Disability Status Score (EDSS)5, which is limited in its utility due to the time and training needed to administer the test, poor inter-rater reliability, insensitivity to small but significant changes over time, and a discontinuous qualitative scale that is heavily skewed toward assessment of severe motor disability. 6-8 The Disease Steps (DS) scale was developed as a metric of disability in pwMS that emphasizes motor and gait impairments but can be administered by neurologists who are not MS specialists. This metric is well correlated with the EDSS.9 10,11 The Patient Determined Disease Steps (PDDS) assessment is a patient-reported outcome measurement that is an adaptation of the physician-reported DS scale.11,12 The PDDS is strongly correlated with the EDSS,13-15 and has been used broadly as a disability metric.15-19
Research lab-based testing using specialized equipment has demonstrated important appreciable differences in gait metrics in pwMS20 that correlate with level of disability and progression of MS, including walking speed, stride length, cadence and double support time.21-23 These research studies offer important insights into gait dysfunction in pwMS. There is a growing body of evidence validating wearable inertial measurement unit (IMU) based devices for gait in MS against gold standard marker-based camera systems,24-31 which offer the potential for efficient in-clinic assessments. Video based options using consumer-grade marker-less “3D” cameras are also of interest for remote clinical assessments and because of the potential for in home monitoring. However, limited findings are available regarding the utility of video-based gait analysis in pwMS. In this study, we evaluate the correlation between gait metrics derived from a wearable IMU-based system and an infrared depth-sensing camera system,32 and we assess the relationship of these measurements to the PDDS score to demonstrate that these systems could be integrated into clinical and remote monitoring workflow to detect important gait features in pwMS.33-35
Methods
Participants
This study was approved by the Institutional Review Board of Mayo Clinic, Rochester, Minnesota (Institutional Review Board 08-007846), and all patients gave written consent for the use of their medical records and gait analysis for research purposes. Patients seen in the MS clinic at Mayo Clinic in Rochester, Minnesota with a diagnosis of MS were consecutively offered enrollment into the Center for MS and Autoimmune (CMSAN) biorepository for gait analysis. The CMSAN biorepository is a biobank that prospectively collects patient gait analysis, serum, spinal fluid, and clinical data. All 26 pwMS that had completed gait assessments were included.
Survey Evaluation
Patients filled out surveys using tablet devices at the time of gait assessment. Surveys included the Patient Determined Disease Steps (PDDS)9,10,12,13 instrument and quantification of falls in the last year.
Walking assessment
Participants completed gait tasks in clinic while wearing OPAL Devices (Clario, Pennsylvania, USA) over the sternum (1 sensor), lumbar back (1 sensor), bilateral wrists (2 sensors), and dorsal surface of the feet (2 sensors) per the manufacturer guidelines for gait analysis in Mobility Lab Version 2 (Clario, Pennsylvania, USA). The OPAL device (Clario, Pennsylvania, USA) is a commercially available IMU device for measuring metrics of gait and balance validated in healthy36 and neurologic populations.36,37 Study coordinators instructed the participants to walk 6 meters, turn 180 degrees, and walk 6 meters back at their normal self- selected speed.24 Two trials were performed. Gait metrics were recorded and computed in Mobility Lab software version 2 (Clario, Pennsylvania, USA) which uses data from all sensors to create a three-dimensional motion matrix to calculate gait parameters. One Azure HD Kinect (Microsoft, Washington, USA) camera was placed behind the starting point and was synchronized using iPi Recorder software (iPi Soft LLC, Lamarjean Group, Inc, Poway, CA) to record depth information related to the subject’s walking. The camera recorded normal video and depth video at 30 frames per second (FPS). Layout of the recording set up is shown in Figure 1A and was consistent between all subjects.
Figure 1:

A. Schematic of gait assessment setup. Participants walked a 6-meter path while wearing six inertial measurement unit (IMU) devices: two on the wrists, two on the ankles, one on the sternum, and one on the lumbar spine. A depth camera simultaneously recorded their movement. The light gray rectangle depicts the walking area. B. Representative still frames from the depth camera. Frames show the participant walking toward the depth camera (left to right, top to bottom). The outline color transitions from yellow to magenta as the participant approaches the camera. Data were analyzed using iPi Mocap Studio 4 Software (iPi Soft LLC, Lamarjean Group, Inc, Poway, CA), with an overlaid orange skeleton used to track body movements.
Depth video data preprocessing and extracting gait metrics.
Gait metrics were extracted from the depth video as previously described.32 Depth recordings were processed offline using the iPi Mocap Studio 4 Software (iPi Soft LLC, Lamarjean Group, Inc, Poway, CA). A skeleton was fit to the depth point cloud data to track the person walking during the assessment (Figure 1B). The turn was excluded from analyses because camera coverage was insufficient for accurate tracking at the walk apogee (optimal camera depth 0.5 – 4 m). The skeleton trajectory was smoothed using the iPiSoft jitter removal tool to remove potential outlier datapoints. Once the depth data were processed, the joint coordinates were exported using the biomech add-in to obtain absolute positions relative to the ground in the tracking software. The default skeleton available in the iPi Motion Studio provided the following marker coordinates: center of mass (com), center of mass projection to ground, hip, lower spine, middle spine, chest, neck, head, eye, effector head, clavicle (Left/Right), shoulder (L/R), forearm (L/R), hand (L/R), thigh (L/R), shin (L/R), foot (L/R), toe (L/R), effector toe (L/R). Effector terminology is defined within the iPi Motion Studio software and was not used as a parameter in the final analysis. The cartesian values of the joint coordinates were outputted in tabular format text files which were processed in python for subsequent analysis. These text files contained header information to select specific joint marker information, such as toe or com coordinates, for subsequent stride detection.
Center of mass (com) and left and right toe markers were used for detecting strides during the session. Missing frames were interpolated using the piecewise cubic hermite interpolating polynomial (pchip) algorithm.38 The x-axis was aligned to the medial-lateral axis, the y-axis was aligned to the inferior-superior axis, and the z-axis was aligned to the anterior-posterior axis during gait. The comx and comz positions were used for subsequent gait detection and the first frame was subtracted to ensure the person was centered and moving in the positive comz direction. The comz and toez positions were used to detect strides. Strides were defined from toe-down to toe-down (tdown[n+1] – tdown[n]) of the ipsilateral leg. First, the toez position was subtracted from the comz position to find the relative change in the foot from the body. Toe down was defined as the positive peak of toez from the comz (find_peaks, prominence 0.5 standard deviations of the overall toez). Similarly, toe up was defined as the negative peak of toez from the comz (in the backward direction). These events were then labeled as Rdown, Rup, Ldown, and Lup depending on which toez marker was being analyzed. This resulted in a sequence of events that was then used to detect “valid strides.” We defined a valid right stride as sequence of Rdown – Lup – Ldown – Rup – Rdown and left stride as Ldown – Rup – Rdown – Lup – Ldown. These events also defined the gait cycle for which subsequent metrics were calculated.
Using the detected strides, the following gait metrics were calculated:39 1) stride duration: time between toe down events (tdown[n+1] – tdown[n]); 2) stride distance: Euclidean distance between toe down events, ∣toex,z(tdown[n+1])– toex,z(tdown[n])∣ 3) step duration: time from the ipsilateral toe down to the contralateral toe down (tcontr,down[n] – tips,down[n]); 4) step distance: Euclidean distance between the ipsilateral and contralateral toe down events ∣contr,toex,z(tdown[n+1]) – ips,toex,z(tdown[n])∣. The following time based metrics were represented as a percentage of the gait cycle: 1) double support: duration both feet were on the ground 100 × ((tcontr,up[n] – tips,down[n]]) + (tips,up[n] – tcontr,down[n]))/(tips,down[n+1] – tips,down[n]); 2) swing duration: time from ipsilateral toe lifts up to touches down (tips,up[n] – tips,down[n+1])/(tips,down[n+1] – tips,down[n]); 3) stance duration: time from the contralateral toe lifts up to touches down (tcontr,up[n] – tcontr,down[n+1])/(tips,down[n+1] – tips,down[n]); 4) gait speed: stride length divided by stride time; 5) cadence: the sum of the strides for both legs divided by the sum of the stride durations calculated for the entire session, 6) maximum swing elevation: the maximal toey position between tips,up[n] and tips,down[n+1].
Statistical Analyses
The mean and standard deviation of each metric were reported for each experimental session. Summary statistics across sessions were calculated by taking the mean of the means or mean of the standard deviations of each metric. No standard deviation was calculated for the cadence metric derived from the video data because it was calculated across the entire session.
Each gait metric was compared between the depth video and the corresponding wearable data. Spearman correlation coefficient was used to analyze the gait testing metrics (wearable, depth video) and disability survey scores (PDDS). Pearson correlations and Bland-Altman plots were calculated to determine how well the values aligned for each metric. Bland-Altman results were presented as the bias of the difference between the two modalities (wearable, depth video) along with the lower and upper limits of agreement (LOAs, 1.96 x standard deviation). 40,41 We considered correlation coefficients >0.9 as “strongly correlated,” coefficients >=0.4 and <0.9 as “moderately correlated,” and >=0.2 and <0.4 as “weakly correlated.” A p value of 0.05 was considered significant.
Results
Gait and depth video assessments of 26 pwMS were analyzed, including 13 people with relapsing remitting MS (RRMS), 9 with secondary progressive MS (SPMS), 2 with primary progressive MS (PPMS), 1 with clinically isolated syndrome and 1 with radiographic isolated syndrome (Table 1). Gait analysis was performed at median age of 54 years (range 23-71) and 65 months from initial diagnosis (range 0-382); 69% were female (18/26). The PDDS score for the cohort was median of 1 (range 0-5), with the most severe disability in those with PPMS (table 1).
Table 1.
Demographic and clinical features of participants.
| Multiple Sclerosis Diagnosis |
Number of participants |
Number Female (%) |
Age (years) at gait analysis (Median, Range) |
Time (months) since diagnosis (Median, Range) |
PDDS Score (Median, Range) |
Number using gait aid |
|---|---|---|---|---|---|---|
| Radiographic Isolated Syndrome | 1 | 1 (100) | [40-45] * | 240 | 0 | 0 |
| Clinically Isolated Syndrome | 1 | 1 (100) | [70-75] * | 6 | 0 | 0 |
| Relapsing Remitting MS | 13 | 11 (85) | 53 (23-64) | 89 (11-382) | 0 (0-2) | 0 |
| Primary progressive MS | 2 | 0 (0) | 50 (41-60) | 38 (28-48) | 5 (5-5) | 2 |
| Secondary Progressive MS | 9 | 5 (56) | 58 (28-66) | 50 (0-382) | 3 (0-5) | 2 |
| Total Cohort | 26 | 18 (69) | 54 (23-71) | 65 (0-382) | 1 (0-5) | 4 |
Abbreviations: MS, Multiple Sclerosis; PDDS, Patient Determined Disease Step.
Single individuals identified by 5-year age epochs to avoid identification.
Correlation and accuracy of spatiotemporal gait metrics between devices
Cadence was strongly correlated between devices for both right (r=0.95, Figure 2, Table 2) and left (r=0.94) legs, with a mean difference of −5.6 steps/min and −4.6 steps/min respectively based on Bland-Altman plots (where a negative value indicates that the depth video metrics underestimate the OPAL metrics). Spatiotemporal based metrics are summarized in Figure 3 and Table 2. There was very strong correlation between devices for time-based metrics, while distance-based metrics had moderate correlation. Stride duration (r=0.97 both legs, mean difference left=0.03 s, right=0.06 s) (Figure 3) was strongly correlated between the wearable data and depth video data. Gait speed was more strongly correlated between platforms for the left leg (r=0.94), compared to the right leg (r=0.78), with small differences between the devices (mean difference 0.04 m/s left, 0.15 m/s right). Stride length measurements were correlated between platforms for the left leg (r=0.8) but not the right leg (r=0.41), with small mean differences between the two devices (0.08 m left, 0.22 m right).
Figure 2:

Cadence metrics correlation plots and Bland-Altman plots comparing wearable device and depth video. a) Correlation plot comparing cadence for the OPAL wearable device (y-axis) vs the depth video (x-axis) measured in the left leg. Black dashed line indicates y=x. Correlation, p-value, and RMSE annotated in the plot. b) Bland-Altman plot indicating the difference between OPAL wearable device and the depth video vs the average value of the two modalities for the left leg. Black dashed lines indicate the lower and upper LOAs, solid lines indicate mean difference. Values are depicted to the right of the plot. c-d) Correlation plot and Bland-Altman plot for right leg cadence metrics.
Table 2.
Summary statistics of gait metrics measured by wearable and video devices.
| Wearable | Video | |||
|---|---|---|---|---|
| Mean of Means |
Mean of Std |
Mean of Means |
Mean of Std |
|
| Cadence (steps/min), L | 97.93 | 3.65 | 94.23 | NA |
| Cadence (steps/min), R | 98.87 | 5.37 | 93.24 | NA |
| Stride length (m), L | 0.97 | 0.08 | 1.05 | 0.19 |
| Stride length (m), R | 0.82 | 0.23 | 1.04 | 0.18 |
| Stride duration (s), L | 1.27 | 0.05 | 1.30 | 0.26 |
| Stride duration (s), R | 1.26 | 0.07 | 1.32 | 0.26 |
| Step duration (s), L | 0.63 | 0.03 | 0.66 | 0.21 |
| Step duration (s), R | 0.63 | 0.05 | 0.67 | 0.19 |
| Gait speed (m/s), L | 0.80 | 0.06 | 0.85 | 0.12 |
| Gait speed (m/s), R | 0.69 | 0.16 | 0.84 | 0.14 |
| Swing phase (%), L | 36.41 | 1.61 | 39.54 | 4.98 |
| Swing phase (%), R | 35.56 | 3.20 | 39.64 | 4.44 |
| Stance phase (%), L | 63.59 | 1.61 | 60.46 | 4.98 |
| Stance phase (%), R | 64.44 | 3.20 | 60.36 | 4.44 |
| Double support (%), L | 26.65 | 1.96 | 21.68 | 13.22 |
| Double support (%), R | 27.71 | 2.85 | 20.18 | 5.23 |
| Elevation at midswing (cm), L | 1.57 | 0.48 | 3.04 | 4.17 |
| Elevation at midswing (cm), R | 1.66 | 0.59 | 2.78 | 3.23 |
Abbreviations: L, left; R, right; Std, standard deviation.
Figure 3:

Correlation and Bland-Altman plots of distance and time-based gait metrics comparing wearable device and depth video. a) Stride length, b) stride duration, c) step duration, and d) gait speed for left leg and right leg. For each metric, the left-most plot depicts the Pearson correlation (r), p-value, and RMSE value, and the right-most plot depicts the Bland-Altman plot with mean difference and LOAs. Conventions similar to figure 2.
Correlation and accuracy of gait cycle metrics between devices
Phases of the gait cycle were measured as percentages of total gait cycle time. Correlation between gait cycle measurements including swing, stance, and double support time, was moderate between the two devices (range 0.4-0.7 across phases) (Figure 4). There was a small overestimation of the swing phase relative to standing phases in the depth video device compared to the wearable devices, with mean differences ranging from −7.5% to 4.1% for the different phase measurements. Elevation at mid-swing was not well-correlated (r=0.44, Figure 5), though the mean difference between devices was less than 1.5 cm on either side.
Figure 4:

Correlation and Bland-Altman plots of gait phases comparing wearable device and depth video. a) Swing phase percent, b) stance phase percent, and c) double support time for left leg and right leg. For each metric, the left-most plot depicts the Pearson correlation (r), p-value, and RMSE value, and the right-most plot depicts the Bland-Altman plot with mean difference and LOAs. Conventions similar to figure 2.
Figure 5:

Correlation and Bland-Altman plots of elevation at mid-swing comparing OPAL wearable device and depth video. a) Correlation plot for elevation at mid-swing comparing wearable device and depth video of left leg, b) Bland-Altman plot indicating the difference between OPAL wearable device and the depth video vs the average value of the two modalities for the left leg. c-d) Correlation and Bland-Altman plots of elevation at mid-swing for right leg. Conventions similar to figure 2.
Gait metric correlation with patient reported disability status
Given the correlation between the wearable device and the depth video measurements, we assessed the correlation of these metrics with the PDDS using principal component analysis (PCA) (Figure 6A) and pairwise Spearman's rank correlation analysis (Figure 6B). These analyses confirm the strong correlation for most metrics between the wearable devices and the depth video (short separation distance in the PCA) and indicate that stride duration and step duration clustered with PDDS and were strongly correlated to this metric. Of note, cadence, gait speed, stride length, and swing phase were clustered separately from the PDDS and were anti-correlated with this metric. The strongest correlations were observed for gait speed, with a Spearman coefficient >0.9 between video and wearable measurements and Spearman coefficients <−0.6 between each of these modalities and the PDDS. The correlations between devices were stronger than with the PDDS.
Figure 6.

Correlation between wearable device and depth video measurements with PDDS. (A) Principal component analysis (PCA) of gait metrics demonstrates close clustering of wearable and depth video measurements, confirming their strong agreement. Stride duration and step duration cluster with PDDS, indicating a strong correlation with disability level. In contrast, cadence, gait speed, stride length, and swing phase cluster separately and are anti-correlated with PDDS. (B) Pairwise Spearman’s rank sum correlation heat map demonstrates a three-way comparison of strength of correlation in color (range: dark red, r=−1; dark blue, r=+1) and significance in black and white (range: black, p=0, white p=1). Gait speed demonstrates the strongest correlations (>0.9 between video and wearable measurements) and strong inverse correlations with PDDS (Spearman coefficients <−0.6).
Discussion
This study demonstrates that both wearable IMUs and depth video systems can detect meaningful differences in gait within a clinical testing environment, potentially offering greater sensitivity to change than traditional stepwise disability scales. In a clinical setting, both technologies reliably replicated key spatiotemporal gait parameters. Metrics such as cadence, stride duration, and gait speed showed high correlation and minimal error across the two modalities. In contrast, measures related to gait cycle phases and mid-swing elevation exhibited lower correlation. Gait speed showed a moderate correlation with the stepwise Patient-Determined Disease Steps (PDDS) score, whereas other gait parameters were only weakly associated. In contrast, measures related to gait cycle phases and mid-swing elevation exhibited lower correlation between devices. Gait speed showed a moderate negative correlation with the stepwise Patient-Determined Disease Steps (PDDS) score. These findings highlight the potential for wearable and depth-sensing technologies to quantitatively assess disability and monitor disease progression in individuals with multiple sclerosis in the clinical setting.
The lower correlation between PDDS and gait metrics in our study was anticipated given the variability in functional status across different PDDS levels. The PDDS is a practical and easy to administer clinical tool, but its non-linear and discrete structure limits its sensitivity, particularly in capturing the full spectrum of dysfunction in patients who report normal or advanced disability. Prior studies have found weaker correlation between PDDS and EDSS in patients who have mild impairment (0-2),13 and more variability in measures of gait speed at PDDS of 5-6.42 Quantifiable gait assessments will be particularly helpful in patients with mild disability to detect and monitor subtle changes over time.
Prior work evaluating gait changes in pwMS has shown differences in velocity, stride length, step length, double support time, and swing phase duration of gait compared to healthy controls.43 Consistently, these studies show smaller steps, slower speed, and less time in swing phase for MS patients relative to healthy controls. PwMS also exhibit greater variability in these spatial-temporal parameters overall compared to controls. Studies have varied extensively in methodology for assessing and measuring these metrics,44-47 and therefore the acceptable range of clinically significant changes has not yet been established. It is important to directly compare gait analysis methodologies and clinical scales, as we have done here, to better understand what constitutes a clinically significant change in these gait metrics. Longitudinal assessment will be needed to further clarify this. Prior work has primarily measured patients in optimized gait lab settings, often with multiple trials for each test. While optimal for scientific purposes, these approaches are expensive, time-consuming, and require sophisticated devices and highly trained personnel. In contrast, the use of wearable devices – and perhaps most importantly, the use of simple depth cameras – offers a less expensive, high bandwidth approach for reliably measuring gait in patients within a clinical setting. However, these devices require appropriate validation. In the current study the data were collected by study coordinator staff using a mobile cart, with on-the-spot study enrollment and assessment in the hallway outside of the patient exam room. Notably, our findings also suggest that depth video imaging may be readily migrated from the clinic to the patient’s home, opening up a tremendous opportunity for daily longitudinal assessment of gait in a large cohort of MS patients.
We hypothesize that the differences in gait phase measurements between the two devices are primarily due to variations in detection methods: OPAL devices rely on foot-ground contact, while depth video systems track toe marker positions relative to the body’s center of mass. This pattern is consistent with a recent IMU-vs-camera validation performed by Zahn et al., which reported that foot-worn IMUs underestimate swing time and overestimate stance time relative to a marker-based video system, leading to a relative error in the swing time calculation in healthy controls that exceeded 7%.30 The authors attribute this bias to differences in detection of the heel strike and toe-off events used to separate stance phases and swing phases, with the camera system using a heel-strike definition based on a force-plate threshold and the IMU defining the end of the swing phase as the transition from dorsiflexion to plantar flexion. This methodological asymmetry mirrors our findings, as the OPAL pipeline anchors phase separations to foot–ground contact, while the depth video derives toe-down/up from the toe kinematics relative to the body COM. Kinematic toe events can precede or lag true load acceptance or toe-off when light contact persists between the foot and ground, increasing swing time and reducing stance time in the video analysis. Zahn et al. also found the weakest agreement between measurement modalities for toe clearance, with the IMU underestimating the value by ~1.9 cm relative to the video measurement. Similarly, our mid-swing elevation showed the poorest cross-device agreement, with mean elevations differing by up to 1.5 cm between the devices. With further optimization and validation against a pressure-sensing gait mat, these methods have the potential for improvement. Across metrics, we observed stronger correlations for the left leg compared to the right. We suspect this is because most patients initiate walking with their right leg, requiring the depth camera to recalibrate the skeletal model after patients completed turns. This recalibration may result in inaccuracies in the first step of the gait cycle. Potential solutions include instructing patients to pause after turning or refining the algorithm to exclude the first step after a turn. Additionally, some variability may arise from patients wearing their usual clothing and footwear during testing, as well as conducting assessments in a clinical hallway with potential interruptions. However, we view these conditions as strengths, as they reflect a real-world clinical environment and demonstrate the robustness of these methods in practical settings.
This study was limited by utilizing the patient-reported PDDS, rather than rater reported EDSS. Future studies should incorporate both, particularly for individuals with low disability where there is more variability in the correlation of these metrics. In the future, important next steps include validation of these devices with lab-based motion capture or pressure-sensitive walkways with patients walking at difference paces to understand any bias based on gait speed introduced due to video sampling rate.48,49 In home remote gait monitoring is an attractive solution for enhancing and democratizing access to clinical trials and care for pwMS. Evaluating feasibility and validity of in-home monitoring with wearable sensors or depth-camera systems will be required to achieve that goal. While a 20% change in timed 25-foot walk time has been proposed as clinically meaningful,50 robust longitudinal studies are needed to determine meaningful differences in other gait metrics.
Our study demonstrates a strong correlation between spatiotemporal measurements obtained from a wearable IMU device and a depth-camera system in pwMS. These metrics are more sensitive to disease-relevant functional changes than the PDDS score, particularly for those with mild disability. These findings highlight the potential for either system to be widely adopted in clinical settings, offering a quantitative and accessible method for evaluating disability and tracking disease progression in MS patients.
Highlights.
Gait impairment is a measure of disease progression in multiple sclerosis
Validated gait-assessment tools in multiple sclerosis are critically needed
Spatiotemporal metrics by depth video correlate with previously validated wearable sensors
Patient reported disability correlated most strongly with speed
Study implemented in clinic setting, demonstrating feasibility of practical clinical use
Funding and conflicts of interest
This study was supported by the Mayo Clinic Center for Multiple Sclerosis and Autoimmune Neurology.
SAB: was supported by Grant Number UL1 TR002377 from the National Center for Advancing Translational Sciences (NCATS), its contents are solely the responsibility of the authors and do not necessarily represent the official views of the NIH.
KM: Nothing to disclose.
LB: Nothing to disclose.
WOT: Receives research funding from the National institutes of Health and the Mayo Clinic Center for Multiple Sclerosis and Autoimmune. He receives royalties from the publication of “Mayo Clinic Cases in Neuroimmunology” (OUP).
FA: Receives research funding from the National institutes of Health.
CLH: was supported by R01 NS115126 and R01 NS92882 from the NIH, H011674201 from the Minnesota Partnership for Biotechnology and Medical Genomics, NORD ID 18008 from the National Organization for Rare Disorders, and by funding from Adimune.
Role of the Funding Source
The funding source played no role in the preparation of this manuscript nor in the decision to offer to this journal for publication.
Footnotes
Declaration of interests
The authors declare the following financial interests/personal relationships which may be considered as potential competing interests:
Samantha Banks reports a relationship with National Center for Advancing Translational Sciences that includes: funding grants. W. Oliver Tobin reports a relationship with National Institutes of Health that includes: funding grants. Farwa Ali reports a relationship with National Institutes of Health that includes: funding grants. Charles L. Howe reports a relationship with National Institute of Neurological Disorders and Stroke that includes: funding grants. Charles L. Howe reports a relationship with National Organization for Rare Disorders that includes: funding grants. Charles L. Howe reports a relationship with Minnesota Partnership for Biotechnology and Medical Genomics that includes: funding grants. Charles L. Howe reports a relationship with Adimmune Corporation that includes: funding grants. Book royalties from Oxford University Press - W.O.T. If there are other authors, they declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- 1.Briggs FB, Hill E. Estimating the prevalence of multiple sclerosis using 56.6 million electronic health records from the United States. Multiple Sclerosis Journal. 2020;26(14):1948–1952. [DOI] [PubMed] [Google Scholar]
- 2.Mayr WT, Pittock SJ, McClelland RL, Jorgensen NW, Noseworthy JH, Rodriguez M. Incidence and prevalence of multiple sclerosis in Olmsted County, Minnesota, 1985–2000. Neurology. 2003;61(10):1373–1377. [DOI] [PubMed] [Google Scholar]
- 3.Tedeholm H, Skoog B, Lisovskaja V, Runmarker B, Nerman O, Andersen O. The outcome spectrum of multiple sclerosis: disability, mortality, and a cluster of predictors from onset. Journal of neurology. 2015;262:1148–1163. [DOI] [PubMed] [Google Scholar]
- 4.Tremlett H, Zhao Y, Devonshire V. Natural history of secondary-progressive multiple sclerosis. Multiple Sclerosis Journal. 2008;14(3):314–324. [DOI] [PubMed] [Google Scholar]
- 5.Kurtzke JF. Rating neurologic impairment in multiple sclerosis: an expanded disability status scale (EDSS). Neurology. 1983;33(11):1444–1444. [DOI] [PubMed] [Google Scholar]
- 6.Cohen M, Bresch S, Rocchi OT, et al. Should we still only rely on EDSS to evaluate disability in multiple sclerosis patients? A study of inter and intra rater reliability. Multiple Sclerosis and Related Disorders. 2021;54:103144. [DOI] [PubMed] [Google Scholar]
- 7.Noseworthy J, Vandervoort M, Wong C, Ebers G. Interrater variability with the Expanded Disability Status Scale (EDSS) and Functional Systems (FS) in a multiple sclerosis clinical trial. Neurology. 1990;40(6):971–971. [DOI] [PubMed] [Google Scholar]
- 8.Meyer-Moock S, Feng Y-S, Maeurer M, Dippel F-W, Kohlmann T. Systematic literature review and validity evaluation of the Expanded Disability Status Scale (EDSS) and the Multiple Sclerosis Functional Composite (MSFC) in patients with multiple sclerosis. BMC neurology. 2014;14:1–10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Hohol M, Orav E, Weiner H. Disease steps in multiple sclerosis: a longitudinal study comparing disease steps and EDSS to evaluate disease progression. Multiple Sclerosis Journal. 1999;5(5):349–354. [DOI] [PubMed] [Google Scholar]
- 10.Hohol M, Orav E, Weiner H. Disease steps in multiple sclerosis: a simple approach to evaluate disease progression. Neurology. 1995;45(2):251–255. [DOI] [PubMed] [Google Scholar]
- 11.Schwartz CE, Vollmer T, Lee H, Group NARCoMSOS. Reliability and validity of two self-report measures of impairment and disability for MS. Neurology. 1999;52(1):63–63. [DOI] [PubMed] [Google Scholar]
- 12.Marrie R, Goldman M. Validity of performance scales for disability assessment in multiple sclerosis. Multiple Sclerosis Journal. 2007;13(9):1176–1182. [DOI] [PubMed] [Google Scholar]
- 13.Learmonth YC, Motl RW, Sandroff BM, Pula JH, Cadavid D. Validation of patient determined disease steps (PDDS) scale scores in persons with multiple sclerosis. BMC neurology. 2013;13:1–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Kobelt G, Berg J, Atherly D, Hadjimichael O. Costs and quality of life in multiple sclerosis: a cross-sectional study in the United States. Neurology. 2006;66(11):1696–1702. [DOI] [PubMed] [Google Scholar]
- 15.Hadjimichael O, Kerns RD, Rizzo MA, Cutter G, Vollmer T. Persistent pain and uncomfortable sensations in persons with multiple sclerosis. Pain. 2007;127(1-2):35–41. [DOI] [PubMed] [Google Scholar]
- 16.Lo AC, Hadjimichael O, Vollmer TL. Treatment patterns of multiple sclerosis patients: a comparison of veterans and non-veterans using the NARCOMS registry. Multiple Sclerosis Journal. 2005;11(1):33–40. [DOI] [PubMed] [Google Scholar]
- 17.Hadjimichael O, Vollmer T, Oleen-Burkey M. Fatigue characteristics in multiple sclerosis: the North American Research Committee on Multiple Sclerosis (NARCOMS) survey. Health and quality of life outcomes. 2008;6:1–11. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Motl RW, McAuley E, Snook EM, Gliottoni RC. Physical activity and quality of life in multiple sclerosis: intermediary roles of disability, fatigue, mood, pain, self-efficacy and social support. Psychology, health & medicine. 2009;14(1):111–124. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Gulick EE, Goodman S. Physical activity among people with multiple sclerosis. International Journal of MS Care. 2006;8(4):121–129. [Google Scholar]
- 20.Cofré Lizama LE, Khan F, Lee PV, Galea MP. The use of laboratory gait analysis for understanding gait deterioration in people with multiple sclerosis. Multiple Sclerosis Journal. 2016;22(14):1768–1776. [DOI] [PubMed] [Google Scholar]
- 21.Lizrova Preiningerova J, Novotna K, Rusz J, Sucha L, Ruzicka E, Havrdova E. Spatial and temporal characteristics of gait as outcome measures in multiple sclerosis (EDSS 0 to 6.5). Journal of neuroengineering and rehabilitation. 2015;12:1–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Zörner B, Hostettler P, Meyer C, et al. Prognosis of walking function in multiple sclerosis supported by gait pattern analysis. Multiple Sclerosis and Related Disorders. 2022;63:103802. [DOI] [PubMed] [Google Scholar]
- 23.Filli L, Sutter T, Easthope CS, et al. Profiling walking dysfunction in multiple sclerosis: characterisation, classification and progression over time. Scientific reports. 2018;8(1):4984. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Banks SA, Howe CL, Mandrekar J, et al. Assessing fall risk in multiple sclerosis using patient-reported outcomes and wearable gait metrics. Multiple Sclerosis Journal–Experimental, Translational and Clinical. 2025;11(2):20552173251329825. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Gervasoni E, Anastasi D, Di Giovanni R, et al. Uncovering Subtle Gait Deterioration in People with Early-Stage Multiple Sclerosis Using Inertial Sensors: A 2-Year Multicenter Longitudinal Study. Sensors. 2023;23(22):9249. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Rudisch J, Jöllenbeck T, Vogt L, et al. Agreement and consistency of five different clinical gait analysis systems in the assessment of spatiotemporal gait parameters. Gait & posture. 2021;85:55–64. [DOI] [PubMed] [Google Scholar]
- 27.Spain RI, Mancini M, Horak FB, Bourdette D. Body-worn sensors capture variability, but not decline, of gait and balance measures in multiple sclerosis over 18 months. Gait & posture. 2014;39(3):958–964. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Fang X, Liu C, Jiang Z. Reference values of gait using APDM movement monitoring inertial sensor system. Royal Society open science. 2018;5(1):170818. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Müller R, Hamacher D, Hansen S, Oschmann P, Keune PM. Wearable inertial sensors are highly sensitive in the detection of gait disturbances and fatigue at early stages of multiple sclerosis. BMC neurology. 2021;21:1–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Zahn A, Koch V, Schreff L, et al. Validity of an inertial sensor-based system for the assessment of spatio-temporal parameters in people with multiple sclerosis. Front Neurol. 2023;14:1164001. doi: 10.3389/fneur.2023.1164001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Müller R, Schreff L, Koch L-E, Oschmann P, Hamacher D. Measuring gait stability in people with multiple sclerosis using different sensor locations and time scales. Sensors. 2021;21(12):4001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Mazurek KA, Barnard L, Botha H, Christianson T, Graff-Radford J, Petersen R, Vemuri P, Windham BG, Jones DT and Ali F A validation study demonstrating portable motion capture cameras accurately characterize gait metrics when compared to a pressure-sensitive walkway. . Scientific reports. 2024;14(1):17464. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Gholami F, Trojan DA, Kövecses J, Haddad WM, Gholami B. A microsoft kinect-based point-of-care gait assessment framework for multiple sclerosis patients. IEEE journal of biomedical and health informatics. 2016;21(5):1376–1385. [DOI] [PubMed] [Google Scholar]
- 34.Shah VV, McNames J, Mancini M, et al. Laboratory versus daily life gait characteristics in patients with multiple sclerosis, Parkinson’s disease, and matched controls. Journal of neuroengineering and rehabilitation. 2020;17:1–12. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Arpan I, Shah VV, McNames J, et al. Fall prediction based on instrumented measures of gait and turning in daily life in people with multiple sclerosis. Sensors. 2022;22(16):5940. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Morris R, Stuart S, McBarron G, Fino PC, Mancini M, Curtze C. Validity of Mobility Lab (version 2) for gait assessment in young adults, older adults and Parkinson’s disease. Physiological measurement. 2019;40(9):095003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Lanotte F, Shin SY, O’Brien MK, Jayaraman A. Validity and reliability of a commercial wearable sensor system for measuring spatiotemporal gait parameters in a post-stroke population: the effects of walking speed and asymmetry. Physiological Measurement. 2023;44(8):085005. [DOI] [PubMed] [Google Scholar]
- 38.Fritsch FN, Butland J. A method for constructing local monotone piecewise cubic interpolants. SIAM journal on scientific and statistical computing. 1984;5(2):300–304. [Google Scholar]
- 39.Huxham F, Gong J, Baker R, Morris M, Iansek R. Defining spatial parameters for non-linear walking. Gait & posture. 2006;23(2):159–163. [DOI] [PubMed] [Google Scholar]
- 40.Bland JM, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. The lancet. 1986;327(8476):307–310. [PubMed] [Google Scholar]
- 41.Bunce C. Correlation, agreement, and Bland–Altman analysis: statistical analysis of method comparison studies. American journal of ophthalmology. 2009;148(1):4–6. [DOI] [PubMed] [Google Scholar]
- 42.Motl R, Neal W, Backus D, et al. Middle-range scores from the patient determined disease steps scale reflect varying levels of walking dysfunction in multiple sclerosis. BMC neurology. 2024;24(1):383. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Comber LGR, Coote S. Gait deficits in people with multiple sclerosis: A systematic review and meta-analysis. Gait & posture. 2017;51:25–35. [DOI] [PubMed] [Google Scholar]
- 44.Kalron A, Dvir Z, Achiron A. Walking while talking—difficulties incurred during the initial stages of multiple sclerosis disease process. Gait & Posture. 2010;32(3):332–335. [DOI] [PubMed] [Google Scholar]
- 45.Allali G, Laidet M, Assal F, Armand S, Lalive P. Walking while talking in patients with multiple sclerosis: the impact of specific cognitive loads. Neurophysiologie Clinique/Clinical Neurophysiology. 2014;44(1):87–93. [DOI] [PubMed] [Google Scholar]
- 46.Benedict RH, Holtzer R, Motl RW, et al. Upper and lower extremity motor function and cognitive impairment in multiple sclerosis. Journal of the International Neuropsychological Society. 2011;17(4):643–653. [DOI] [PubMed] [Google Scholar]
- 47.Burschka JM, Keune PM, Menge U, Oy UH-v, Oschmann P, Hoos O. An exploration of impaired walking dynamics and fatigue in multiple sclerosis. BMC neurology. 2012;12(1):1–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Sanders O, Wang B, Kontson K. Concurrent Validity Evidence for Pressure-Sensing Walkways Measuring Spatiotemporal Features of Gait: A Systematic Review and Meta-Analysis. Sensors. 2024;24(14):4537. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.D’haene M, Chorin F, Colson SS, Guérin O, Zory R, Piche E. Validation of a 3d markerless motion capture tool using multiple pose and depth estimations for quantitative gait analysis. Sensors. 2024;24(22):7105. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Hobart J, Blight AR, Goodman A, Lynn F, Putzki N. Timed 25-foot walk: direct evidence that improving 20% or greater is clinically meaningful in MS. Neurology. 2013;80(16):1509–1517. [DOI] [PubMed] [Google Scholar]