Abstract
Continuous monitoring of the technical and physiological function of cochlear implants (CI) is a central part of the care process. Despite worldwide efforts to standardise procedures, there is still considerable variation between CI centres, particularly in terms of the methods used, their practical implementation and the definition of meaningful target parameters. A standardised structured test procedure is needed for reliable quality assurance and better comparability. Against this background, the ADANO Working Group for Evoked Response Audiometry (AG-ERA), in close cooperation with the Cochlear Implants and Implantable Hearing Systems Committee of the German Society of Audiology (DGA), developed a minimum standard for audiological and technical functional testing of CIs in an open consensus process. This standard defines basic requirements for performance and documentation and serves as a practical recommendation for CI centres. It is intended to improve interdisciplinary cooperation, increase the quality of care and enable structured long-term optimised care for CI patients.
Keywords: Quality assurance, Minimum standard, Measurement methods, Evoked potentials, Impedances
Abstract
Die kontinuierliche Kontrolle der technischen und physiologischen Funktion von Cochlea-Implantaten (CI) stellt einen zentralen Baustein im gesamten Versorgungsprozess dar. Trotz weltweiter Bestrebungen zur Vereinheitlichung der Verfahren zeigen sich nach wie vor erhebliche Unterschiede zwischen den CI-versorgenden Einrichtungen – insbesondere hinsichtlich der eingesetzten Methoden, ihrer praktischen Umsetzung und der Festlegung aussagekräftiger Zielgrößen. Für eine verlässliche Qualitätssicherung und verbesserte Vergleichbarkeit ist ein einheitlicher, strukturierter Prüfprozess erforderlich. Vor diesem Hintergrund wurde in einem offenen Konsensverfahren der Arbeitsgruppe Elektrische Reaktionsaudiometrie (AG-ERA) der ADANO, gemeinsam mit dem Fachausschuss „Cochlea-Implantate und implantierbare Hörsysteme“ der DGA, ein Minimalstandard für die audiologisch-technische Funktionsprüfung von CI entwickelt. Dieser definiert grundlegende Anforderungen an Durchführung und Dokumentation und dient als praxisnahe Empfehlung für CI-versorgende Einrichtungen. Ziel ist eine standardisierte, nachvollziehbare Vorgehensweise, die die interdisziplinäre Zusammenarbeit verbessert, die Versorgungsqualität erhöht und eine strukturierte, langfristig optimierte Betreuung von CI-Tragenden ermöglicht.
Schlüsselwörter: Qualitätssicherung, Minimalstandard, Messverfahren, Evozierte Potentiale, Impedanzen
Preamble
Throughout the process of cochlear implant (CI) provision, as much information as possible about the technical and physiological function of the implant must be available. In addition, the correct position of the electrode array should be checked regularly. For this purpose, extensive CI-mediated technical and electrophysiological measurements are usually performed intra- and postoperatively [11, 42, 57, 58]. They are an integral part of audiological diagnostics.
Despite international efforts to standardize procedures, there is still great heterogeneity between CI centers in the selection and use of appropriate methods and the associated choice of target parameters [2, 3]. Therefore, the authors believe that there is a need for a standardized process of functional audiological-technical functional testing of the cochlear implant, the determination of the measurement procedures to be performed, and a definition of relevant target parameters. As a result of an open process within the German Working Group for Evoked Response Audiometry (AG-ERA) of the Working Group of German-Speaking Audiologists, Neurotologists, and Otologists (ADANO) of the German Society of Oto-Rhino-Laryngology, Head and Neck Surgery (DGHNO-KHC) and in close cooperation with the Cochlear Implants and Implantable Hearing Systems expert committee of the German Society of Audiology (DGA), a minimum standard protocol for functional performance of audiological-technical functional tests of cochlear implants throughout the care process has been defined and a consensus has been reached.
This minimum standard does not apply only to CI centers but also explicitly to CI companies. It is essential that appropriate measurement and evaluation procedures are provided by the manufacturers and that these procedures are integrated into the respective clinical software solutions. These measures should ensure that the measurements of the defined target parameters can be reliably recorded, reproducibly evaluated, and made comparable between different CI centers.
The use of electrode impedance measurements [1, 29, 30, 37, 40, 41, 55, 57], measurements of the stimulation current-induced non-stimulating electrode voltage [5, 6, 17, 22, 25, 45, 54, 59], electrically evoked stapedius reflexes [8, 16, 31, 53, 56, 57], electrically evoked compound action potentials and auditory brainstem responses [7, 9, 13, 14, 18–21, 26, 27, 31, 32, 34–36, 38, 39, 43, 47, 51, 52, 60], or late cortical auditory evoked potentials [15, 27, 33, 46, 48, 58] and extra- and intracochlear electrocochleography during and after electrode insertion [10, 23, 24, 28] offers a wide range of possible target variables to be measured. In addition, there are various manufacturer-specific options and implementations of the aforementioned measurement methods in the respective clinical software. However, some of these methods are not currently integrated as standard by all manufacturers. Therefore, these recommendations also include information on the respective measurement procedures. On the one hand, the defined target parameters and process recommendations are intended to enable a profound assessment of the measurement results in order to confirm the regular technical and physiological function of the implant. On the other hand, they should contribute to ensure the quality of care [49] in accordance with the currently valid S2k guideline “Cochlea-Implantat Versorgung” of the Association of Scientific Medical Societies (AWMF Registry No.: 017-071; [4]) and the white paper “Cochlea-Implantat (CI)-Versorgung” of the DGHNO-KHC [12] and should also be used for the German CI Register, if applicable [50].
The application of other methods and the selection of additional target parameters for the investigation of individual questions, e.g., in the case of a relative contraindication to CI treatment (see [4, p. 33]), should explicitly not be restricted by this recommendation.
Recommendations
If audiological–technical functional tests of the CI or electrophysiological functional tests of the hearing system are carried out during and after electrode insertion in accordance with these recommendations, the methods listed in Table 1 (column 1) should be used for various questions and the specified target parameters (column 3) should be measured, determined, and reported during the surgery in accordance with the process description in Fig. 1. The listed process recommendations (column 4) should also be observed.
Table 1.
Target parameter and process recommendations for the different measurement methods
| Method | Objective | Target parameter recommendation (end points) | Process recommendation | Regular findings (minimum consensus) |
|---|---|---|---|---|
| Standard (provision required) | ||||
| Coupling and implant short-circuit test |
Testing for bidirectional telemetric data transmission Checking that the implant functions correctly in accordance with the specifications, including the electrode contacts Indications of defects in the implant |
Coil-implant coupling [yes/no] Impedances [Z in kΩ] |
Measurement in all available stimulation modes on all electrode contacts of the implant pre-OP: before the start of surgery in the sterile packaging (alternatively: check in NaCl solution directly before insertion) intra/post-OP: monitoring of the coupling during the entire audiological-technical diagnostics |
(1) Stable coupling and bidirectional data transmission (2) No short circuits between the electrode contacts |
| Electrode impedance measurements |
Checking that the implant is functioning correctly according to specifications Check all electrode contacts on the electrode array and the external reference electrode(s) Indications of implant failure |
Impedances [Z in kΩ] Ground path impedance [Z in kΩ] |
Measurement in all available stimulation modes on all electrode contacts of the implant intra-OP: directly after insertion (if necessary, after electrode fixation) and after wound closure post-OP: regularly (e.g., before each fitting) |
(1) No indication of open circuits (no noticeable impedances) (2) No short circuits between the electrode contacts (3) Impedances of all intracochlear electrodes within the expected range of the respective implant model or type of electrode array |
| Measurements of the stimulation current-induced non-stimulating electrode voltage (SCINSEV) |
Checking that the implant is functioning correctly according to specifications Estimation of the electrode position(s) and identification of a defect such as tip fold-over, base kinking Research is currently focusing on the insertion depth and position of the electrode array, e.g., for migration of the electrode array [5, 6] |
Voltage per current unit [Rn,m in kΩ] |
Measurement of SCINSEV and visualization in a 12 × 12, 16 × 16 or 22 × 22 matrix (depending on the implant model) intra-OP: after electrode array insertion (if necessary, after electrode array fixation) post-OP: regularly in the course of the fitting procedure |
(1) No indication of open circuits (no noticeable impedances) (2) No short circuits between the electrode contacts (3) Continuous decrease in voltage per unit current at the recording non-stimulating electrode with increasing distance from the stimulating electrode |
| Electrically evoked stapedius reflexes (eSR) |
Physiological functional testing of the signal transmission (lower auditory pathway—olive complex—brainstem level) and reflex triggering by the facial nerve Confirmation of complete insertion or exclusion of a possible incorrect insertion of the electrode array Determination of the upper limit of the electrical dynamic range (over- or understimulation) Plausibility check of stimulation parameters and eCAP |
Reflex triggering [can be evoked/cannot be evoked] eSR thresholds [Q in nC] |
Stimulation on several electrode contacts distributed over the electrode array (basal—medial-apical) intra-OP: visual detection (ipsilateral) after initial impedance and measurement of SCINSEV under visual control (intraoperative situs) during electrical stimulation post-OP: if required and if feasible, measurement using an impedance audiometer with direct electrical stimulation (ipsilateral) at selected electrodes or determination of the ESR thresholds in the free sound field |
(1) Reflex can be evoked and identified as movement of the staple tendon and/or head |
| Electrically evoked compound action potentials of the auditory nerve (eCAP) |
Stimulus response threshold diagnostics/threshold profile (initial findings and data for CI fitting) Monitoring of neuronal parameters/retrocochlear diagnostics in the case of unclear integrity of the auditory nerve Position check of the electrode array (e.g., exclusion of tip fold-over, electrode migration) |
Detectability of the eCAP [yes/no] eCAP thresholds [Q in nC] eCAP growth functions (AGF), N1-P1 amplitude [U in µV] Absolute latencies of N1 and P1 [t in µs] SoE profile, N1-P1 amplitudes [U in µV] |
Determination of eCAP-AGF, eCAP latencies and eCAP thresholds on all intracochlear electrodes (12-22, depending on implant model) Measurement of one or more SoE profiles on several electrodes (apical), depending on the implant model and electrode array intra-OP: after impedance measurement, if necessary, also after electrode conditioning, followed by SoE measurement if required post-OP: regularly (before/at fitting) after impedance measurement, if required |
(1) eCAP detectable on all electrodes (2) eCAP thresholds are within the expected range of the respective implant model (depending on the values of the stimulation parameters) [7, 19, 26, 31, 32, 36, 38, 47] (3) In the case of SoE measurement: while maintaining a defined test electrode position (test and recording electrode), the influence of the displacement of the masking position (masking electrode) along the electrode carrier decreases continuously (depending on the values of the stimulation parameters and the respective implant model) |
| Electrically evoked auditory brainstem responses (eABR) |
Retrocochlear diagnostics for auditory synaptopathy/neuropathy (AS/AN) and unclear integrity of the auditory nerve and lower brainstem, detection of maturation and degeneration processes Assessment of stimulus processing and transmission in special cases, e.g., when eCAP measurement is not possible (in the case of acquired changes to the inner ear, e.g., following chronic inflammation, traumatic changes) Stimulus response threshold diagnostics |
Absolute latencies of the waves eIII and eV and inter-peak latency (IPL) as a function of the stimulus intensity [t in ms] Smallest stimulus with detectable and reproducible stimulus response threshold [Q in nC] Potential amplitudes [U in nV] |
Potential measurement on selected (intracochlear) electrode contacts or areas, detection of the potential latencies and determination of the stimulus threshold intra-OP: if required, measurement via an AEP system after impedance/eSR/eCAP measurement, position of the recording electrodes: Cz or Fpz vs. mastoid or earlobe (contralateral) or in the lower area of the sternocleidomastoid muscle (ipsilateral) post-OP: if required, measurement via an AEP system, position of the recording electrodes: Cz or Fpz vs. mastoid or earlobe (ipsi- and contralateral) |
(1) Brainstem potentials can be triggered above threshold, waves eIII and eV can be identified, absolute latencies and interpeak latencies are within the expected range (depending on the values of the stimulation parameters) (2) eABR thresholds are within the expected range (depending on the values of the stimulation parameters) |
| For EAS/hybrid (provision useful in individual cases) | ||||
| Electrocochleography (ECochG) | Control and monitoring of the cochlear function of hair cells and afferent auditory nerve fibers (residual auditory function) before, during, and after insertion of the electrode array |
Detectability of the 3 signal components of the ECochG (CM, CAP, SP), at least “cochlear microphonic” (CM) Determination of the (largest) CM amplitude [U in µV] |
During insertion: Acoustic stimulation, e.g., at 500 Hz (tone burst) with approx. 40 dB SL and potential measurement on selected electrode contacts intra-OP: if required recording of the ECochG extracochlear: electrode on the round window (before insertion) or on the promontory (during insertion), intracochlear: apical electrode; in each case in reference to a lead electrode (Cz or Fpz) post-OP: if required, recording of the ECochG intracochlear: apical electrode (22 or 1 depending on implant model); in each case in reference to a lead electrode (Cz or Fpz) |
No (rapid) drop in amplitude in the ECochG during insertion of the electrode carrier Intra- and post-OP stable CM amplitudes |
| Optional (mainly of scientific interest) | ||||
| Electrically evoked auditory cortical responses (eACR) |
Integrity testing of the auditory pathway up to the level of the cerebral cortex Assessment of auditory pathway function and maturation at cortical level |
P1-N1-P2-N2 complex [t in ms] |
Potential measurement on selected electrode contacts or areas, detection of the potential latencies and determination of the stimulus response threshold post-OP: if required, measurement via an AEP system, position of the recording electrodes: Cz or Fpz vs. mastoid or earlobe (ipsi- and contralateral) |
(1) Cortical potentials can be triggered above threshold, P1-N1-P2-N2 complex well detectable, absolute latencies and interpeak latencies are within the expected range (depending on the values of the stimulation parameters) |
AEP auditory evoked potentials, AGF amplitude growth function, CAP compound action potential, CI cochlear implant, CM cochlear microphonic, Cz electrode positioned at vertex, eABR electrically evoked auditory brainstem responses , eACR electrically evoked auditory cortical responses, EAS electrically acoustic stimulation, eCAP electrically evoked compound action potential, ECochG electrocochleography, eSR electrically evoked stapedius reflex, eSRT electrically evoked stapedius reflex threshold, Fpz frontocentral placement of electrode, NaCl sodium chloride, SCINSEV stimulation current-induced non-stimulating electrode voltage, SL sensation level, SoE spread of excitation, SP summation potential
Fig. 1.
Description of the audiological-technical functional testing of the cochlear implant during surgery: After the coupling and implant short-circuit testing in the package and/or in sodium chloride solution (1), the electrode array is inserted, if necessary, with real-time electrocochleography (ECochG) monitoring in the case of residual hearing preservation surgery (EAS). The electrical (technical) function of the cochlear implant is checked by impedance measurement and measurement of the impedance or voltage matrix (SCINSEV) (2). If an electrode problem is detected, the possible cause should be investigated immediately and a replacement implant should be used if necessary. If possible, electrical stimulation of the stapedius reflexes (eSR), if necessary, with the determination of the stimulation threshold (eSRT) (3) and measurement of the electrically evoked compound action potentials (eCAP) (4) is then performed under observation of the surgical site. In the case of difficult questions or unclear results, electrically evoked brainstem potentials (eABR) and/or the spread of excitation (SoE) is also measured, and intraoperative imaging is performed if required. If necessary, surgical intervention and/or placement of the backup implant will follow. The impedance measurement is repeated after wound closure (5)
The highest possible quality of the examination results is ensured by guaranteeing the functional integrity of all medical device components, creating optimal measurement conditions (appropriate examination room) by reducing artifacts at the lowest possible noise levels and/or eliminating/shielding electromagnetic interference fields, minimizing residual noise and residual interference (e.g., by relaxing/sleeping/sedating/anesthetizing the person to be examined), and the professional qualifications and experience of the operators.
Definitions
Stimulation current-induced non-stimulating electrode voltage.
Indirect measure of the electric field propagation along the electrode array and out of the cochlea (in the text also abbreviated as SCINSEV [44]), measured as voltage per unit of current and represented in an m × n matrix, i.e., the electric voltage U induced by the input current I at a stimulation electrode (m) as the potential difference between a (non-stimulating) recording electrode (n) and a ground electrode, expressed in the unit of measurement kΩ with
![]() |
Stimulus response threshold.
The lowest applied electrical charge Q with (still) identifiable and reproducible stimulus response (also referred to as “threshold” in Table 1) as the product of stimulation current and stimulation time (pulse phase time), specified in the unit of measurement nC with
![]() |
If this unit of measurement is not available, the manufacturer-dependent intensity unit should be specified and labeled accordingly.
Declarations
Conflict of interest
A. Müller, M. Blümer, O.C. Dziemba, A. Elsholz, L. Fröhlich, U. Hoppe, D. Polterauer, T. Rader, T. Rahne, T. Steffens, M. Walger, T. Weißgerber, T. Wesarg and S. Zirn declare that there is no conflict of interest.
For this article no studies with human participants or animals were performed by any of the authors. All studies followed the ethical standards indicated in each case.
The supplement containing this article is not sponsored by the industry.
Footnotes
ADANO: Working Group of German-Speaking Audiologists, Neurotologists, and Otologists of the German Society of Oto-Rhino-Laryngology, Head and Neck Surgery (DGHNO-KHC)
AG-ERA: German Working Group for Evoked Response Audiometry under the umbrella of ADANO, in which audiologists of different professional origin come together for an exchange of experience and scientific projects
DGA: German Society of Audiology
Scan QR code & read article online
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Alahmadi A, Abdelsamad Y, Yousef M, Almuhawas F, Hafez A, Alzhrani F, Hagr A (2024) Cochlear Implantation: Long-Term Effect of Early Activation on Electrode Impedance. J Clin Med. 10.3390/jcm13113299 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Aljazeeri I, Abdelsamad Y, Alsanosi A et al (2024) Minimum intraoperative testing battery for cochlear implantation: the international practice trend. Eur Arch Otorhinolaryngol. 10.1007/s00405-024-08944-y [DOI] [PubMed] [Google Scholar]
- 3.Alzhrani F, Aljazeeri I, Abdelsamad Y et al (2024) International Consensus Statements on Intraoperative Testing for Cochlear Implantation Surgery. Ear Hear 45(6):1418–1426. 10.1097/AUD.0000000000001526 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Aschendorff A, Baumann U, Delank K, Ernst A, Hoppe U et al (2020) Cochlea-Implantat Versorgung: S2k-Leitlinie 017/071 der Deutschen Gesellschaft für Hals-Nasen-Ohren-Heilkunde, Kopf und Hals-Chirurgie e. V. (DGHNO KHC), Bonn. https://register.awmf.org/assets/guidelines/017-071l_S2k_Cochlea-Implantat-Versorgung-zentral-auditorische-Implantate_2020-12.pdf
- 5.Ayas M, Muzaffar J, Borsetto D, Eitutis S, Phillips V, Tam YC, Salorio-Corbetto M, Bance ML (2024) A scoping review on the clinical effectiveness of Trans-Impedance Matrix (TIM) measurements in detecting extracochlear electrodes and tip fold overs in Cochlear Ltd devices. PLoS ONE 19(3):e299597. 10.1371/journal.pone.0299597 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Ayas M, Tam YC, Bysouth-Young D, Eitutis ST, Salorio-Corbetto M, Axon PR, Donnelly NP, Tysome JR, Borsetto D, Smith ME, Bance ML (2024) Intraoperative Detection of Extracochlear Electrodes Using Stimulation Current Induced Non-Stimulating Electrode Voltage (SCINSEV) Measures (Transimpedance Measures)-A Case Series. Clin Otolaryngol 49(6):827–832. 10.1111/coa.14212 [DOI] [PubMed] [Google Scholar]
- 7.Berger K, Hocke T, Hessel H (2017) Lautheitsoptimierte Messung von Summenaktionspotentialen bei Cochlea Implantat Trägern. Laryngorhinootologie 96(11):780–786. 10.1055/s-0043-119292 [DOI] [PubMed] [Google Scholar]
- 8.Chai B, Holland ML, Camposeo EL, King K, Schvartz-Leyzac KC (2024) Patient and Device Factors Contributing to Electrically Evoked Stapedial Reflex Thresholds in Cochlear Implanted Adults. Audiol Neurootol 29(4):263–270. 10.1159/000535058 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Cinar BC, Özses M (2024) How differ eCAP types in cochlear implants users with and without inner ear malformations: amplitude growth function, spread of excitation, refractory recovery function. Eur Arch Otorhinolaryngol. 10.1007/s00405-024-08971-9 [DOI] [PubMed] [Google Scholar]
- 10.Dalbert A, Weder S (2024) Anwendung der extra- und intracochleären Elektrocochleographie während und nach der Cochleaimplantation. HNO. 10.1007/s00106-024-01481-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Deutsche Gesellschaft für Audiologie e. V. (2015) Audiologische Leistungen nach der CI-Indikation: Empfehlungen der Deutschen Gesellschaft für Audiologie (DGA). Z Audiol 54(1):36–37 [Google Scholar]
- 12.Deutsche Gesellschaft für Hals-Nasen-Ohren-Heilkunde, Kopf- und Hals-Chirurgie e. V. (2021) Weißbuch Cochlea-Implantat(CI)-Versorgung: Empfehlungen zur Struktur, Organisation, Ausstattung, Qualifikation und Qualitätssicherung inder Versorgung von Patienten mit einem Cochlea-Implantat in der Bundesrepublik Deutschland. https://www.ci-register.de/wp-content/uploads/DGHNO_Weissbuch_CI-Versorgung.pdf (2. Aufl. Deutsche Gesellschaft für Hals-Nasen-Ohren-Heilkunde, Kopf- und Hals-Chirurgie e. V., Bonn)
- 13.Dziemba OC, Hocke T, Müller A (2022) EABR on cochlear implant—measurements from clinical routine compared to reference values. Gms Z Audiol Acoust. 10.3205/zaud000023 [Google Scholar]
- 14.Dziemba OC, Salim MP, Müller A (2024) References for electrophysiological measurements for intraoperative verification of placement for several types of CI electrodes. Gms Z Audiol – Audiol Acoust. 10.3205/ZAUD000049 [Google Scholar]
- 15.Firszt JB, Chambers RD, Kraus And N, Reeder RM (2002) Neurophysiology of cochlear implant users I: effects of stimulus current level and electrode site on the electrical ABR, MLR, and N1-P2 response. Ear Hear 23(6):502–515. 10.1097/00003446-200212000-00002 [DOI] [PubMed] [Google Scholar]
- 16.Franke-Trieger A, Mattheus W, Seebacher J, Zahnert T, Neudert M (2021) Stapedius reflex evoked in free sound field in cochlear implant users compared to normal-hearing listeners. Int J Audiol 60(9):695–703. 10.1080/14992027.2020.1866780 [DOI] [PubMed] [Google Scholar]
- 17.Franke-Trieger A, Lailach S, Strahl SB, Murrmann K, Schebsdat E, Neudert M, Zahnert T (2024) Voltage matrix algorithm for intraoperative detection of cochlear implant electrode misplacement. Audiol Neurootol. 10.1159/000543264 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Gabr T, Debis H, Hafez A (2024) Electric Auditory Brainstem Response Audiometry in Cochlear Implants: New Recording Paradigm. Audiol Res 14(4):581–592. 10.3390/audiolres14040049 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Gärtner L, Bräcker T, Kals M, Penninger RT, Billinger-Finke M, Lenarz T, Büchner A (2022) ARTFit‑A Quick and Reliable Tool for Performing Initial Fittings in Users of MED-EL Cochlear Implants. Life. 10.3390/life12020269 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Gordon KA, Papsin BC, Harrison RV (2006) An evoked potential study of the developmental time course of the auditory nerve and brainstem in children using cochlear implants. Audiol Neurootol 11(1):7–23. 10.1159/000088851 [DOI] [PubMed] [Google Scholar]
- 21.Grolman W, Maat A, Verdam F, Simis Y, Carelsen B, Freling N, Tange RA (2009) Spread of excitation measurements for the detection of electrode array foldovers: a prospective study comparing 3‑dimensional rotational x‑ray and intraoperative spread of excitation measurements. Otol Neurotol 30(1):27–33. 10.1097/mao.0b013e31818f57ab [DOI] [PubMed] [Google Scholar]
- 22.Hans S, Arweiler-Harbeck D, Kaster F, Ludwig J, Hagedorn E, Lang S, Meyer M, Holtmann LC (2021) Transimpedance Matrix Measurements Reliably Detect Electrode Tip Fold-over in Cochlear Implantation. Otol Neurotol 42(10):e1494–e1502. 10.1097/MAO.0000000000003334 [DOI] [PubMed] [Google Scholar]
- 23.Harris MS, Koka K, Thompson-Harvey A, Harvey E, Riggs WJ, Saleh S, Holder JT, Dwyer RT, Prentiss SM, Lefler SM, Kozlowski K, Hiss MM, Ortmann AJ, Nelson-Bakkum ER, Büchner A, Salcher R, Harvey SA, Hoffer ME, Bohorquez JE, Alzhrani F, Alshihri R, Almuhawas F, Danner CJ, Friedland DR, Seidman MD, Lenarz T, Telischi FF, Labadie RF, Buchman CA, Adunka OF (2024) Amplitude Parameters Are Predictive of Hearing Preservation in a Randomized Controlled Trial of Intracochlear Electrocochleography During Cochlear Implant Surgery. Otol Neurotol 45(8):887–894. 10.1097/MAO.0000000000004286 [DOI] [PubMed] [Google Scholar]
- 24.Haumann S, Mynarek Née Bradler M, Maier H, Helmstaedter V, Büchner A, Lenarz T, Teschner MJ (2024) Does Intraoperative Extracochlear Electrocochleography Correlate With Postoperative Audiometric Hearing Thresholds in Cochlear Implant Surgery? A Retrospective Analysis of Cochlear Monitoring. Trends Hear 28:23312165241252240. 10.1177/23312165241252240 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Hoppe U, Brademann G, Stöver T, Ramos de Miguel A, Cowan R, Manrique M, Falcón-González JC, Hey M, Baumann U, Huarte A, Liebscher T, Bennett C, English R, Neben N, Ramos Macías A (2022) Evaluation of a Transimpedance Matrix Algorithm to Detect Anomalous Cochlear Implant Electrode Position. Audiol Neurootol 27(5):347–355. 10.1159/000523784 [DOI] [PubMed] [Google Scholar]
- 26.Hoth S, Herisanu I, Praetorius M (2016) Objektive Maße bei der Anpassung der Prozessoren von Cochleaimplantatsystemen : Nutzung von Diskriminationsfunktionen und Betrachtung von Elektrodenprofilen. HNO 64(12):870–879. 10.1007/s00106-016-0281-0 [DOI] [PubMed] [Google Scholar]
- 27.Hoth S, Dziemba OC (2017) The Role of Auditory Evoked Potentials in the Context of Cochlear Implant Provision. Otol Neurotol 38(10):e522–e530. 10.1097/MAO.0000000000001480 [DOI] [PubMed] [Google Scholar]
- 28.Kim J‑S (2020) Electrocochleography in Cochlear Implant Users with Residual Acoustic Hearing: A Systematic Review. Int J Environ Res Public Health. 10.3390/ijerph17197043 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Konrad S, Büchner A, Lenarz T, Paasche G (2023) Impedance development after implantation of hybrid-L24 cochlear implant electrodes. Int J Audiol 62(12):1137–1144. 10.1080/14992027.2022.2125914 [DOI] [PubMed] [Google Scholar]
- 30.Körtje M, Stöver T, Leinung M, Baumann U (2022) The Importance of a Preinsertion Integrity Device Test of Cochlear Implants. Otol Neurotol 43(6):e641–e644. 10.1097/MAO.0000000000003548 [DOI] [PubMed] [Google Scholar]
- 31.Kosaner J, Spitzer P, Bayguzina S, Gultekin M, Behar LA (2018) Comparing eSRT and eCAP measurements in pediatric MED-EL cochlear implant users. Cochlear Implants Int 19(3):153–161. 10.1080/14670100.2017.1416759 [DOI] [PubMed] [Google Scholar]
- 32.Lambriks L, van Hoof M, Debruyne J, Janssen M, Hof J, Hellingman K, Devocht E, George E (2023) Toward neural health measurements for cochlear implantation: The relationship among electrode positioning, the electrically evoked action potential, impedances and behavioral stimulation levels. Front Neurol 14:1093265. 10.3389/fneur.2023.1093265 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Lamminmäki S, Cormier K, Davidson H, Grigsby J, Sharma A (2023) Auditory Cortex Maturation and Language Development in Children with Hearing Loss and Additional Disabilities. Children. 10.3390/children10111813 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Lassaletta L, Calvino M, Díaz M, Morales-Puebla JM, Sánchez-Cuadrado I, Varela-Nieto I, Gavilán J (2024) Intraoperative assessment of cochlear nerve functionality in various vestibular schwannoma scenarios: Lessons learned. Hear Res 446:108997. 10.1016/j.heares.2024.108997 [DOI] [PubMed] [Google Scholar]
- 35.Liebscher T, Mewes A, Hoppe U, Hornung J, Brademann G, Hey M (2021) Electrode Translocations in Perimodiolar Cochlear Implant Electrodes: Audiological and Electrophysiological Outcome. Z Med Phys 31(3):265–275. 10.1016/J.ZEMEDI.2020.05.004 [DOI] [PubMed] [Google Scholar]
- 36.Liebscher T, Hornung J, Hoppe U (2023) Electrically evoked compound action potentials in cochlear implant users with preoperative residual hearing. Front Hum Neurosci 17:1125747. 10.3389/fnhum.2023.1125747 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Lin DP‑Y, Chen JK‑C, Tung T‑H, Li LP‑H (2019) Differences in the impedance of cochlear implant devices within 24 hours of their implantation. PLoS ONE 14(9):e222711. 10.1371/journal.pone.0222711 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Müller A, Hocke T, Mir-Salim P (2015) Intraoperative findings on ECAP-measurement: normal or special case? Int J Audiol 54(4):257–264. 10.3109/14992027.2014.969410 [DOI] [PubMed] [Google Scholar]
- 39.Müller A, Kropp MH, Mir-Salim P, Aristeidou A, Dziemba OC (2021) Intraoperatives Tip-Foldover-Screening mittels Spread of Excitation Messungen. Z Med Phys 31(3):276–288. 10.1016/J.ZEMEDI.2020.07.002 [DOI] [PubMed] [Google Scholar]
- 40.Mussoi BS, Meibos A, Woodson E, Sydlowski S (2024) The association between electrode impedance and short-term outcomes in cochlear implant recipients of slim modiolar and slim straight electrode arrays. Cochlear Implants Int 25(1):59–68. 10.1080/14670100.2023.2290768 [DOI] [PubMed] [Google Scholar]
- 41.Oberhoffner T, Mlynski R, Schraven S, Brademann G, Dierker A, Spitzer P, Hey M (2024) Effects of Intraoperative Cochlear Implant Electrode Conditioning on Impedances and Electrically Evoked Compound Action Potentials. IEEE Trans Biomed Eng 71(2):650–659. 10.1109/TBME.2023.3313198 [DOI] [PubMed] [Google Scholar]
- 42.Rader T, Baumann U, Stöver T, Weissgerber T, Adel Y, Leinung M, Helbig S (2016) Management of Cochlear Implant Electrode Migration. Otol Neurotol 37(9):e341–e348. 10.1097/MAO.0000000000001065 [DOI] [PubMed] [Google Scholar]
- 43.Rader T, Nachtigäller P, Linke T, Weißgerber T, Baumann U (2023) Exponential fitting of spread of excitation response measurements in cochlear implants. J Neurosci Methods 391:109854. 10.1016/j.jneumeth.2023.109854 [DOI] [PubMed] [Google Scholar]
- 44.de Rijk SR, Tam YC, Carlyon RP, Bance ML (2020) Detection of Extracochlear Electrodes in Cochlear Implants with Electric Field Imaging/Transimpedance Measurements: A Human Cadaver Study. Ear Hear 41(5):1196–1207. 10.1097/AUD.0000000000000837 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.de Rijk SR, Hammond-Kenny A, Tam YC, Eitutis ST, Garcia C, Carlyon RP, Bance M (2022) Detection of Extracochlear Electrodes Using Stimulation-Current-Induced Non-Stimulating Electrode Voltage Recordings With Different Electrode Designs. Otol Neurotol 43(5):e548–e557. 10.1097/MAO.0000000000003512 [DOI] [PubMed] [Google Scholar]
- 46.Saravanan P, Devi N, Geetha C (2024) Electrically evoked late latency response using single electrode stimulation and its relation to speech perception among paediatric cochlear implant users. Front Hum Neurosci 18:1441854. 10.3389/fnhum.2024.1441854 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Schrank L, Nachtigäller P, Müller J, Hempel J‑M, Canis M, Spiegel JL, Rader T (2024) ART and AutoART ECAP measurements and cochlear nerve anatomy as predictors in adult cochlear implant recipients. Eur Arch Otorhinolaryngol 281(7):3461–3473. 10.1007/s00405-023-08444-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Sharma A, Dorman MF, Spahr AJ (2002) A sensitive period for the development of the central auditory system in children with cochlear implants: implications for age of implantation. Ear Hear 23(6):532–539. 10.1097/00003446-200212000-00004 [DOI] [PubMed] [Google Scholar]
- 49.Stöver T, Leinung M, Loth A (2020) Welche Qualität macht den Unterschied in der Cochlea-Implantat-Versorgung? Laryngorhinootologie 99(S 01):S107–S164. 10.1055/a-1019-9381 [DOI] [PubMed] [Google Scholar]
- 50.Stöver T, Plontke SK, Guntinas-Lichius O, Welkoborsky H‑J, Zahnert T, Delank KW, Deitmer T, Esser D, Dietz A, Wienke A, Loth A, Dazert S (2023) Struktur und Einrichtung des Deutschen Cochlea-Implantat-Registers (DCIR). HNO 71(12):767–778. 10.1007/s00106-023-01309-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Thai-Van H, Cozma S, Boutitie F, Disant F, Truy E, Collet L (2007) The pattern of auditory brainstem response wave V maturation in cochlear-implanted children. Clin Neurophysiol 118(3):676–689. 10.1016/j.clinph.2006.11.010 [DOI] [PubMed] [Google Scholar]
- 52.Thangaraj M, Arunachalam R, Gore M, AjithKumar U (2023) Prediction of behavioral MCL using electrophysiological responses in children using MED-EL implant. Int J Pediatr Otorhinolaryngol 172:111696. 10.1016/j.ijporl.2023.111696 [DOI] [PubMed] [Google Scholar]
- 53.Tyagi Y, Chatterjee I (2023) The effect of cochlear implant age and duration of intervention on ESRT in children with cochlear implant. Cochlear Implants Int 24(4):224–234. 10.1080/14670100.2023.2221048 [DOI] [PubMed] [Google Scholar]
- 54.Vanpoucke FJ, Boermans P‑P, Frijns JH (2012) Assessing the placement of a cochlear electrode array by multidimensional scaling. IEEE Trans Biomed Eng 59(2):307–310. 10.1109/TBME.2011.2173198 [DOI] [PubMed] [Google Scholar]
- 55.Wei JJ‑F, Tung T‑H, Li LP‑H (2021) Evolution of impedance values in cochlear implant patients after early switch-on. PLoS ONE 16(2):e246545. 10.1371/journal.pone.0246545 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Weiss NM, Óvári A, Oberhoffner T, Demaret L, Bicer A, Schraven S, Ehrt K, Dahl R, Schneider A, Mlynski R (2021) Automated detection of electrically evoked stapedius reflexes (eSR) during cochlear implantation. Eur Arch Otorhinolaryngol 278(6):1773–1779. 10.1007/s00405-020-06226-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Wesarg T, Arndt S, Aschendorff A, Laszig R, Zirn S (2014) Intraoperative audiologisch-technische Diagnostik bei der Cochleaimplantatversorgung. HNO 62(10):725–734. 10.1007/s00106-014-2936-z [DOI] [PubMed] [Google Scholar]
- 58.Wesarg T, Arndt S, Aschendorff A, Laszig R, Beck R, Jung L, Zirn S (2017) Intra- und postoperative elektrophysiologische Diagnostik. HNO 65(4):308–320. 10.1007/s00106-016-0195-x [DOI] [PubMed] [Google Scholar]
- 59.Zhang L, Schmidt FH, Oberhoffner T, Ehrt K, Cantré D, Großmann W, Schraven SP, Mlynski R (2024) Transimpedance Matrix Can Be Used to Estimate Electrode Positions Intraoperatively and to Monitor Their Positional Changes Postoperatively in Cochlear Implant Patients. Otol Neurotol 45(4):e289–e296. 10.1097/MAO.0000000000004145 [DOI] [PubMed] [Google Scholar]
- 60.Zirn S, Arndt S, Aschendorff A, Wesarg T (2015) Interaural stimulation timing in single sided deaf cochlear implant users. Hear Res 328:148–156. 10.1016/j.heares.2015.08.010 [DOI] [PubMed] [Google Scholar]