. 2025 Sep 24;95(2):975–986. doi: 10.1002/mrm.70097

Scanner‐based real‐time automated volumetry reporting of the fetus, amniotic fluid, placenta, and umbilical cord for fetal MRI at 0.55T

Sara Neves Silva 1,2,, Alena Uus 1,2, Hadi Waheed 1, Simi Bansal 1, Kamilah St Clair 1, Wendy Norman 1, Jordina Aviles Verdera 1,2,3, Daniel Cromb 1, Tomas Woodgate 1, Milou van Poppel 1, Johannes K Steinweg 1, Jacqueline Matthew 1,2, Kuberan Pushparajah 1,2,4, David Lloyd 1,2,4, Vanessa Kyriakopoulou 1, Dimitris Siassakos 5, Anna David 5, Joseph V Hajnal 1,6, Lisa Story 1,7,8, Mary A Rutherford 1, Jana Hutter 1,2,3
PMCID: PMC12681313  PMID: 40991800

Abstract

Purpose

This work aims to enable real‐time automated intra‐uterine volumetric reporting and fetal weight estimation for fetal MRI, deployed directly on the scanner.

Methods

A multi‐region segmentation nnUNet was trained on 146 images of 73 fetal subjects (coronal and axial orientations) for the parcellation of the fetal head, fetal body, placenta, amniotic fluid, and umbilical cord from whole uterus bSSFP stacks. A reporting tool was then developed to integrate the segmentation outputs into an automated report, providing volumetric measurements, fetal weight estimations, and z‐score visualizations. The complete pipeline was subsequently deployed on a 0.55T MRI scanner, enabling real‐time inference and fully automated reporting in the duration of the acquisition.

Results

The segmentation pipeline was quantitatively and retrospectively evaluated on 36 stacks of 18 fetal subjects and demonstrated sufficient performance for all labels, with high scores (>0.98) for the fetus, placenta, and amniotic fluid, and 0.91 for the umbilical cord. The prospective evaluation of the scanner deployment step was successfully performed on 50 cases, with the regional volumetric reports available directly on the scanner.

Conclusions

This work demonstrated the feasibility of multi‐regional intra‐uterine segmentation, fetal weight estimation, and automated reporting in real‐time. This study provides a robust baseline solution for the integration of fully automated scanner‐based measurements into fetal MRI reports.

Keywords: Fetal MRI, fetal weight estimation, scanner deployment, uterus segmentation

1. INTRODUCTION

While antenatal imaging is routinely performed using ultrasound, fetal MRI is increasingly employed as a complementary modality for complex fetal anomalies. 1 In addition to offering superior soft tissue contrast, it provides full‐view 3D spatial information, reduces operator dependency, and allows fetal assessment until late gestation. Biometric and volumetric measurements are thereby a crucial aspect of radiological reporting, particularly allowing quantitative assessment of fetal development and comparison to normal ranges, enabling the early detection of deviations.

In the third trimester, accurate assessment of fetal growth and the intrauterine environment is critical for optimal perinatal outcomes. Estimation of fetal weight is essential for identifying growth abnormalities, such as fetal growth restriction (FGR) and macrosomia, both linked to increased risks of stillbirth, operative delivery, and neonatal morbidity. 2 , 3 Fetal volume is additionally used to normalize organ volumes (e.g., lungs 4 ) in cases with pathologies. Placental and amniotic fluid volumes are key indicators of fetal well‐being. Small placental volume may suggest placental insufficiency and is associated with FGR and pre‐eclampsia, while large placental size may indicate gestational diabetes or fetal anomalies. 5 , 6 Similarly, abnormal amniotic fluid levels—oligohydramnios or polyhydramnios—may indicate placental dysfunction, fetal renal/gastrointestinal anomalies, or maternal‐fetal transfer issues. 7 Umbilical cord measurements, including insertion site, volume, coiling index, and length, provide additional information on fetal circulation and risk factors. 8 However, they cannot be assessed using conventional clinical ultrasound. Short cords are associated with growth restriction and adverse outcomes, while excessively long cords increase the risk of entanglement or cord accidents. 9 Together, these metrics are important for assessing fetal health and planning delivery. They contribute to risk stratification and inform clinical decisions regarding timing and mode of birth, especially in high‐risk pregnancies.

However, in routine clinical practice, ultrasound‐based estimation of fetal weight 10 and amniotic fluid volume 11 are typically derived indirectly from 2D biometry measurements, which can be inconsistent and inaccurate. Placental volume is not currently routinely clinically assessed. Furthermore, ultrasound accuracy in late gestation is reduced due to fetal head engagement in the pelvis and the lack of consideration of the contribution of fetal fat. It is further limited in cases of high maternal BMI, oligohydramnios, or unfavorable fetal positioning. In contrast, fetal MRI enables direct volumetric measurements based on true 3D information, and the measurements derived from segmentations are operator‐independent, thus offering high reproducibility and accuracy in estimation of fetal weight. 12 3D segmentation of the fetus in MRI stacks has been used for fetal weight estimation 13 and global organ volume normalization. 4 , 14 Although highly relevant, 15 , 16 placenta and amniotic fluid volumes are not often measured with MRI. These large structures may be segmented in either balanced steady‐state free‐precession (bSSFP) or T2‐weighted single‐shot Turbo Spin Echo (ssTSE) sequences, using a large region‐of‐interest (ROI) coverage. Conventional clinical practice and research studies rely mainly on 2D slice‐wise manual segmentations 4 , 13 , 17 for calculation of fetal body volumes, which is particularly time‐consuming for larger fetuses at later gestation age (GA).

Recently, several research works relying on UNet‐style deep learning (DL) networks 18 proposed automated solutions for segmentation of large intrauterine structures, including: Fetus, 19 , 20 , 21 placenta, 22 , 23 and amniotic fluid 24 , 25 in bSSFP, ssTSE, and EPI fetal MRI stacks. Yet, combined segmentation of these structures is still lacking, which could be beneficial for combined relative volumetry analysis, as well as more consistent segmentation between ROI interfaces.

Furthermore, such frameworks were performed offline post‐scan, foregoing the ability to react (e.g., by performing further dedicated sequences) to any measured deviations while the patient is still in the scanner. Recent studies have integrated real‐time processing tools for fetal MRI scans, including quality control and re‐acquisition of low‐quality slices and stacks, 26 , 27 brain tracking, 28 3D reconstruction, 29 and stack acquisition planning. 30 However, real‐time volumetric analysis and reporting have not yet been performed.

1.1. Contributions

In this work, we present the first scanner‐based framework for interactive assessment in fetal MRI by developing a real‐time automated DL internal uterine parcellation and volumetry tool for the fetus, placenta, amniotic fluid, and umbilical cord. The segmentation results are combined into an automated report with 3D visualization, calculated centiles, and estimated fetal weight (EFW)—all available during the scan. The method is implemented and tested for 0.55T late GA fetal MRI datasets and evaluated on a cohort of 50 prospective scans.

2. METHODS

An overview of the proposed framework is presented in Figure 1. Immediately upon acquisition, a whole uterus bSSFP stack (maternal coronal orientation) is exported via FIRE 31 (Step 1) to an external Gadgetron PC. Next, DL‐based segmentation is performed for 3D parcellation of the fetus, placenta, amniotic fluid, and umbilical cord (Step 2). Based on the segmentation output, the volumetric measurements are performed, including fetal weight estimation (Step 3), calculated centiles, 3D models, and growth charts, and all results are compiled into a .pdf report (Step 4), available for viewing on the scanner console.

FIGURE 1.

FIGURE 1

Overview of the automated volumetry reporting pipeline. Following the bSSFP acquisition, the data is immediately exported to an external Gadgetron PC using the FIRE framework (Step 1); then, DL‐based segmentation is performed (Step 2), enabling fetal weight estimation (Step 3) and automated report generation (Step 4).

2.1. Cohort, datasets and acquisition parameters

The fetal MRI datasets used in this study were acquired at St. Thomas' Hospital, London, as part of the ethically approved MEERKAT [REC: 21/LO/0742], MiBirth [REC: 23/LO/0685], and NANO [REC: 22/YH/0210] studies. All experiments were performed in accordance with relevant guidelines and regulations. Informed written consent was obtained from all participants.

The fetal scans were performed on a 0.55T clinical MRI scanner (MAGNETOM Free.Max, Siemens Healthcare, Germany) in the supine position, using a 6‐element blanket coil and a 9‐element spine coil. A bSSFP sequence was acquired in coronal orientation to the whole‐uterus ROI with TR per shot = 669.7 ms, 86 segments, TE = 4.2 ms, FA = 120°, GRAPPA acceleration factor 2, Partial Fourier 6/8, in‐plane resolution = 0.77×0.77 mm, slice thickness = 2.4 mm, matrix size = 544×544, and 64–80 slices per dataset. At 0.55T, higher flip angles are permissible due to reduced SAR constraints, and this adjustment helped compensate for the inherently lower SNR at low field.

The gestational ages of the 50 prospective datasets ranged between 25.4 and 39.3 weeks, and primarily included late gestation cases between 36–40 weeks due to the specific aims of the main study (“MiBirth: MRI imaging at term for prediction of the mode of birth”, https://www.mibirthstudy.com/). The main inclusion criteria for all cases were: Singleton pregnancy, no severe structural anomalies of the fetus, and acceptable image quality with clear visibility of the uterus and fetus, without extreme signal artefacts.

2.2. Multi‐regional internal uterine segmentation

The proposed parcellation protocol for the bSSFP whole uterus acquisitions, summarized in Figure 2, was designed to include regions that are relevant to both volumetric analysis and 3D relative position assessment. It includes five labels in total: The fetus (subdivided into body and head labels), placenta, umbilical cord, and amniotic fluid.

FIGURE 2.

FIGURE 2

Proposed multi‐regional intrauterine segmentation protocol for bSSFP stacks: Fetal body (red), fetal head (cyan), placenta (green), umbilical cord (blue), and amniotic fluid (yellow).

The nnUNet 32 framework was selected for the DL segmentation task. It was trained on 146 labeled bSSFP images, acquired in axial and coronal whole uterus orientations from 73 fetal subjects (GA 37–40 weeks). The ground‐truth labels were created using outputs of in‐house pre‐trained DL networks for individual structures (developed for other projects within the department), followed by manual refinement (AU, SNS, SB, KSC, WN).

A 3D U‐Net architecture was implemented via the nnU‐Net, 32 with pre‐processing, architecture, and training parameters automatically configured based on the characteristics of the dataset. The model architecture integrated a 3D PlainConvUNet design with six encoder and five decoder stages, each composed of two convolutional layers. Most kernels were 3×3×3, except the first layer (1×3×3), with feature maps ranging from 32 up to 320. Pooling was asymmetrical, using six downsampling stages with kernel sizes: [1,1,1], [1,2,2], [2,2,2], [2,2,2], [2,2,2], and [1,2,2]. Images were resampled to 2.4×0.77×0.77 mm, and z‐score normalization was applied per case. The input patch size was 56×224×160 voxels. Training used stochastic gradient descent with Nesterov momentum (initial learning rate 0.01) and a combined Dice and cross‐entropy loss. A batch size of 2 was chosen. From the initial 146 datasets, 117 cases (5‐fold cross‐validation) were used for training, and 29 cases for validation (1 fold). Standard nnU‐Net data augmentations were applied, and model performance was assessed using the pseudo‐Dice similarity coefficient across all segmented regions. The model was trained for 1 000 epochs.

As a part of the retrospective quantitative evaluation of the pipeline, the model was subsequently tested on 36 images, 18 in each orientation. All 18 network predictions extracted from the coronal whole uterus images were refined by a fetal radiographer (R1, 1.5 years of experience), while a subset was additionally refined by R2 (3 years of experience) to measure inter‐observer variability. The average volume differences were calculated for all labels and normalized to the manually refined volume.

2.3. Fetal weight estimation

Similarly to previous studies performing whole fetus segmentation (e.g., 20 ), EFW was computed from the total fetal label volume based on the classical model formulas derived by Baker et al. (EFWBaker(kg)=1.031·Vfetus+0.12) 33 and Kacem et al. (EFWKacem(kg)=0.989·Vfetus(L)+0.147). 34

2.4. Normative ranges for late gestation datasets

To create normative growth centiles for automated reporting, the trained network was applied to segment an additional dataset comprising 90 coronal bSSFP stacks from MiBirth control subjects (35–39 weeks GA). The inclusion criteria were: No reported fetal and maternal anomalies. All DL segmentations were reviewed and manually refined when required. The extracted volumes for the fetus and placenta, along with the estimated fetal weight, 33 were then used to generate normative growth charts. The Baker 33 approach was chosen as both formulae yield comparable estimations, and thus, integrating only one was considered sufficient for this study. The mean 50th, 5th and 95th centile models were created using classical linear fitting. 35

2.5. Automated reporting

The volumetry report was structured to include all information relevant for comprehensive clinical interpretation. It comprises: Visualization of the original bSSFP stack slices with the respective segmentations overlaid, visualization of the 3D model of the fetus, a table presenting extracted volumes and EFW with computed centiles and z‐scores, and corresponding plots derived from normative growth charts. The report is automatically generated in .pdf based on a Python script.

2.6. Scanner‐based deployment

The 0.55T scanner was connected to an external GPU‐accelerated Gadgetron PC (NVIDIA GEFORCE RTX 2080 Ti, NVIDIA Corporate), and the prototype Siemens framework for image reconstruction environments (FIRE) 31 was used as the interface for real‐time inline processing of the scanner‐reconstructed images, as described above in Figure 1. Each prospective dataset was streamed and stored in the workstation in real‐time, and the online pipeline was configured using an XML file, stored in the scanner console. This file, linked with the respective sequence, specified the path to the Python processing script located on the workstation. Once the complete image matrix was collected, the processing pipeline was triggered automatically—the multi‐label nnUNet segmentation model was immediately loaded and applied to the acquired image matrix, volumes were extracted for the segmented ROIs, the 3D model of the fetus and EFW were computed, and the report was generated.

3. RESULTS

3.1. Multi‐regional internal uterine segmentation

Predicted segmentations for three randomly chosen exemplary fetal subjects from the retrospective testing cohort are shown in Figure 3. All 18 coronal stacks from 18 subjects were visually reviewed and deemed to be of acceptable quality. Quantitative results comparing network predictions to the average of manually refined segmentations—performed by two radiographers (R1 and R2) and one researcher (R3)—are summarized in Table 1. The model showed excellent performance on large ROIs, with high Dice scores and minimal volume discrepancies: 1% for the fetus, 3% for the placenta, and 2% for the amniotic fluid. Performance was lower for the umbilical cord (12% average volume difference), due to its small size, variable shape and position, and limited visibility. Despite these challenges, the average Dice score for the cord remained high at 0.9.

FIGURE 3.

FIGURE 3

Retrospective testing: Examples of predicted segmentations for three fetal subjects with different positions within the uterus. The calculated volume measurements (cc) of the key structures: Placenta (P, green), umbilical cord (UC, dark blue), amniotic fluid (AF, yellow), and whole fetus (F, red and light blue).

TABLE 1.

Retrospective testing: Quantitative evaluation for 18 datasets.

Fetus Placenta Umbilical cord Amniotic fluid
Dice (predictions vs. average R1–R3)
0.99±0.01
0.98±0.02
0.91±0.08
0.99±0.01
Absolute volume difference (cc) (predictions vs. average R1–R3)
8.57±8.88
19.34±25.74
10.10±11.23
7.03±8.61
Relative volume difference (%) (predictions vs. average R1–R3)
0.30±0.30
2.20±2.83
11.02±12.67
1.48±1.74

Note: Comparison between the predicted segmentations from the nnUNet vs. the manually refined segmentations by three manual observers (average of all refinements).

3.2. Normative ranges for late gestation datasets

Figure 4 presents the generated normative centiles for intrauterine regional volumetry and estimated fetal weight across the late GA range, based on the 90 control subjects (above‐mentioned) from the MiBirth study cohort. Segmentations required only minor refinements in fewer than 20% of cases—primarily in challenging ROIs such as the umbilical cord, placenta, and fetal limbs, with minimal impact on the overall centile trends. As expected, fetal volume and estimated weight increased with gestational age, consistent with reported ranges. 20 Placental and amniotic fluid volumes exhibited greater variability in the later GA range, in line with findings from previous studies. 36 These results underscore the need for further research into the influence of maternal and fetal factors on normative modeling.

FIGURE 4.

FIGURE 4

Normative ranges (50th, 5th and 95th centiles) for fetal, placenta and amniotic fluid volume and estimated fetal weight (Baker et al.), generated from 90 control subjects.

3.3. Scanner‐based automated reporting

The complete segmentation and reporting pipeline, deployed on the 0.55T scanner using the FIRE framework, was prospectively evaluated in 50 fetal cases between 25–39 weeks GA. The segmentation network processed each stack in real‐time, with an inference time of 45 s for multi‐ROI label prediction. Volumetric segmentations were successfully generated in all cases, and structured reports were automatically produced and made available directly on the scanner workstation during the imaging session, enabling immediate review by clinical teams. Figure 5 shows an example of a report displayed on the scanner console.

FIGURE 5.

FIGURE 5

Automated volumetry report displayed on the scanner console immediately after the scan, shown with the coronal bSSFP acquisition used in the real‐time volumetry pipeline on the MR viewer.

Figure 6 presents results from the full pipeline in four fetal cases, demonstrating the real‐time extraction of key intrauterine structures and automated volumetry reporting. In several anecdotal cases, deviations from normative volumetric ranges correlated with ultrasound findings. For example, in Fetus 1, the placenta volume (461.7 cc) was below average for gestational age, and the amniotic fluid volume (189.4 cc) was at the lower end of normal. Ultrasound findings corroborated the low AFI, but no placenta anomalies were reported. In contrast, all volumetric measurements for Fetus 2 fell within expected ranges. Fetus 3 showed a total volume and estimated fetal weight (3748.4 cc) exceeding the 95th centile, while Fetus 4 measured below the 5th centile (2345.7 cc), aligning with the birthweight centiles for both of these cases: 97th and 7th, respectively.

FIGURE 6.

FIGURE 6

Prospective testing: Examples of complete automated volumetry .pdf reports for four fetal subjects, depicting the acquired bSSFP images overlaid with the real‐time segmentations (A), the 3D fetal model (B), and the real‐time generated volumetry results (C).

Figure 7 summarizes qualitative scoring in 50 prospective cases. All segmentations were successfully generated using the integrated FIRE pipeline, with output labels and .pdf reports automatically saved to the scanner workstation. Over 95% of segmented anatomical structures were rated as “excellent” or “good” in quality. The umbilical cord segmentations were characterized by lower scores due to minor errors caused by their inherent shape complexity and limited visibility. Importantly, no segmentation failures were recorded across any of the evaluated structures.

FIGURE 7.

FIGURE 7

Prospective testing: Qualitative evaluation results for 50 datasets. Grading of fetus, placenta, amniotic fluid, and umbilical cord segmentation labels (fail, poor, acceptable, good).

The clinical utility of the structured report was supported by verbal feedback from fetal MRI and obstetrics specialists (HW, SB, LS, MR). The added value of the comprehensive visualization of all structures within the uterus, the inclusion of a 3D fetal model, and the integration of z‐score‐based quantification was particularly highlighted for the early identification of potential anomalies. Radiographers reported a positive user experience, emphasizing the seamless integration of the pipeline into the existing MRI acquisition protocols. The tool was described as intuitive and easy to launch, requiring minimal changes to standard workflows.

4. DISCUSSION

This work introduces the first workflow for real‐time, automated intrauterine volumetry reporting for fetal MRI, performed directly on the scanner during acquisition. The approach was integrated into the scanner environment (0.55T) via the FIRE framework and involves (1) an nnUNet network trained to automatically parcellate bSSFP whole uterus stacks into 5 regions: Fetal body and head, placenta, amniotic fluid and umbilical cord; (2) a processing script that uses the segmentation results to subsequently generate a .pdf report that includes volumetric measurements, fetal weight estimation, and computed percentiles and z‐scores against normative ranges, alongside visualization of the segmented regions. The Gadgetron PC, connected to the scanner, launches the complete pipeline directly after stack acquisition. The resulting .pdf report is available for review directly on the scanner console during the scanning session.

Whilst deep learning segmentation of the whole fetus has already been addressed in several works (e.g., 20 ), this is the first pipeline that combines segmentation of the fetus, placenta, umbilical cord, and amniotic fluid in one network. It also includes the corresponding normative ranges specific to the late GA range and 0.55T fetal MRI. Joint volumetric analysis of these ROIs, in addition to fetal weight estimation, could potentially provide valuable information to radiologists and obstetricians to guide and support decision planning. Retrospective quantitative evaluation of the network demonstrated good performance for large structures, with less than 3% relative volume differences after manual refinement. However, the umbilical cord was the most challenging for the model, due to suboptimal visibility, complex shape, and high variability, leading to imperfect labeling in the training and testing datasets. These findings support the need for further training on a larger dataset to improve performance in anatomically complex structures.

Accurate volumetry of the fetus, placenta, amniotic fluid, and umbilical cord supports the identification of growth deviations (e.g., FGR or macrosomia), placental insufficiency, altered fluid dynamics, and potential umbilical cord‐related risks. Multi‐ROI segmentation also allows 3D rendering for visual assessment of the relative position information (e.g., umbilical cord vs. fetal neck, placenta location, etc.). When combined into an automated report with centiles and z‐scores, this integrated analysis aids in stratifying risk, guiding delivery planning, and supporting clinical decision‐making, particularly in complex or borderline cases.

The prospective evaluation of the scanner‐deployed pipeline was successfully performed on 50 datasets, with the output reports and segmentations classified as acceptable/good for >95% of all ROIs in all cases. The feedback from both clinicians and radiographers was very positive, with clear added value of having estimated fetal weight and volumetry z‐scores for identification of potential high‐risk cases, as well as seamless integration into the scanning workflow.

This confirms both the feasibility and utility of the proposed solution. One of the major advantages of scanner‐integrated segmentation is the possibility to transfer the parcellation label files directly to the local archiving system as a part of the case files, without any additional processing on external workstations.

This prototype pipeline is the first step towards real‐time scanner‐based fetal MRI volumetric analysis. Automated reporting for fetal weight estimation and intrauterine regional volumetry could potentially reduce the radiologist's workload, thus paving the way for performing high‐standard fetal reporting more widely in non‐specialist centers.

4.1. Limitations and future work

Although the segmentation network showed good performance on the test dataset, the main testing and training cohort was predominantly from the late GA range, singleton pregnancy, normal fetal anatomy, and from the same 0.55T acquisition protocol. Further retraining of the model on early GA datasets, fetal structural anomalies, and higher‐field strength protocols is required for translation to clinical practice. This will additionally require the generation of normative growth models for the whole duration of the second and third trimesters.

In addition, integration of automated image quality control 37 and image restoration via correction of motion and intensity artefacts (e.g., shading and banding) 38 as a preprocessing step can further improve the accuracy of segmentations. For extreme motion corruption, detection and reacquisition of bSSFP stacks might also be required. 27

Further sub‐parcellation of the fetal ROI into brain tissue regions 39 and body organs 40 could potentially enable full volumetric reporting from a single stack. This will also allow organ‐specific density weight estimation, potentially improving accuracy compared to the approach of Baker et al., which relies on an approximated average density value. Segmentation of additional maternal anatomy structures (myometrium and cervix), as well as optimization of the pipeline for multiple gestation pregnancies, could also be beneficial. The report structure will be extended with a 3D visualization of all structures and a relative position assessment.

Finally, while the report is currently sent to the scanner console upon generation, it is not yet transferred to the reporting database and therefore remains detached from the clinical workflow. Future work will focus on converting the report into a DICOM object to enable export alongside the acquisitions.

5. CONCLUSION

We presented the first scanner‐deployed automated fetus, placenta, amniotic fluid, and umbilical cord volumetry reporting pipeline for fetal MRI. It enables fast DL segmentation of the whole uterus directly from 0.55T bSSFP stacks during acquisition, with a .pdf report with z‐scores and visualization of the results available on the scanner console. The complete pipeline for scanner‐based automated volumetry reporting is available at https://github.com/saranevessilva/automated‐fetal‐mri.git, including scripts for scanner integration and the trained multi‐label segmentation model. The pipeline was successfully evaluated both retrospectively (18 cases) and prospectively (50 cases), and demonstrated good segmentation quality and positive user feedback from clinicians and radiographers. Real‐time fetal weight estimation and multi‐ROI volumetry reporting have the potential to provide valuable information to radiologists, thus opening the possibility for high‐standard fetal MRI assessment in non‐specialist centers. Future work will focus on extending the GA range, optimizing the pipeline for 1.5T and 3T, and further sub‐parcellating the fetal brain and body ROIs. We also plan to integrate image restoration and quality control measures.

ACKNOWLEDGMENTS

We thank everyone involved in the acquisition and analysis of the datasets at the Department of Early Life Imaging at King's College London and St Thomas' Hospital. The authors thank all participants and their families.

This work was supported by the MRC grant [MR/X010007/1], the ERC Starting Grant [ERC101165242], the Sir Henry Wellcome Fellowship to Jana Hutter [201374/Z/16/Z], DFG Heisenberg funding [502024488], the High‐Tech Agenda of the Free State of Bavaria, the UKRI FLF to Jana Hutter [MR/T018119/1], the NIHR Advanced Fellowship awarded to Lisa Story [NIHR30166], the Wellcome Trust and EPSRC IEH award [102431] for the iFIND project [WT 220160/Z/20/Z], the Wellcome/ EPSRC Center for Medical Engineering at King's College London [WT 203148/Z/16/Z], the NIHR Clinical Research Facility (CRF) at Guy's and St Thomas' and by the National Institute for Health Research Biomedical Research Center based at Guy's and St Thomas' NHS Foundation Trust and King's College London.

The views expressed are those of the authors and not necessarily those of the NHS, the NIHR, or the Department of Health.

Neves Silva S., Uus A., Waheed H., et al., “Scanner‐based real‐time automated volumetry reporting of the fetus, amniotic fluid, placenta, and umbilical cord for fetal MRI at 0.55T,” Magnetic Resonance in Medicine 95, no. 2 (2026): 975–986, 10.1002/mrm.70097.

Sara Neves Silva and Alena Uus contributed equally to this work.

DATA AVAILABILITY STATEMENT

The data that support the findings of this study are available on request from the corresponding author. The data are not publicly available due to privacy or ethical restrictions.

REFERENCES

  • 1. Prayer D, Malinger G, De Catte L, et al. ISUOG Practice Guidelines (updated): Performance of fetal magnetic resonance imaging. Ultrasound Obstet Gynecol. 2023;61:278‐287. doi: 10.1002/uog.26129 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Aucott SW, Donohue PK, Northington FJ. Increased morbidity in severe early intrauterine growth restriction. J Perinatol. 2004;24:435‐440. doi: 10.1038/SJ.JP.7211116 [DOI] [PubMed] [Google Scholar]
  • 3. Weissmann‐Brenner A, Simchen MJ, Zilberberg E, et al. Maternal and neonatal outcomes of large for gestational age pregnancies. Acta Obstet Gynecol Scand. 2012;91:844‐849. doi: 10.1111/J.1600-0412.2012.01412.X [DOI] [PubMed] [Google Scholar]
  • 4. Story L, Zhang T, Steinweg JK, et al. Foetal lung volumes in pregnant women who deliver very preterm: a pilot study. Pediatr Res. 2020;87:1066‐1071. doi: 10.1038/s41390-019-0717-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. González‐González NL, González‐Dávila E, González Marrero L, Padrón E, Conde JR, Plasencia W. Value of placental volume and vascular flow indices as predictors of intrauterine growth retardation. Eur J Obstet Gynecol Reprod Biol. 2017;212:13‐19. doi: 10.1016/J.EJOGRB.2017.03.005 [DOI] [PubMed] [Google Scholar]
  • 6. Soongsatitanon A, Phupong V. First trimester 3D ultrasound placental volume for predicting preeclampsia and/or intrauterine growth restriction. J Obstet Gynaecol. 2019;39:474‐479. doi: 10.1080/01443615.2018.1529152 [DOI] [PubMed] [Google Scholar]
  • 7. Morris RK, Meller CH, Tamblyn J, et al. Association and prediction of amniotic fluid measurements for adverse pregnancy outcome: Systematic review and meta‐analysis. BJOG Int J Obstet Gynaecol. 2014;121:686‐699. doi: 10.1111/1471-0528.12589 [DOI] [PubMed] [Google Scholar]
  • 8. De Laat MWM, Franx A, Bots ML, Visser GHA, Nikkels PGJ. Umbilical coiling index in normal and complicated pregnancies. Obstet Gynecol. 2006;107:1049‐1055. doi: 10.1097/01.AOG.0000209197.84185.15 [DOI] [PubMed] [Google Scholar]
  • 9. Krakowiak P, Smith EN, De Bruyn G, Lydon‐Rochelle MT. Risk factors and outcomes associated with a short umbilical cord. Obstet Gynecol. 2004;103:119‐127. doi: 10.1097/01.AOG.0000102706.84063.C7 [DOI] [PubMed] [Google Scholar]
  • 10. Milner J, Arezina J. The accuracy of ultrasound estimation of fetal weight in comparison to birth weight: A systematic review. Ultrasound. 2018;26:32‐41. doi: 10.1177/1742271X17732807 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Hughes DS, Magann EF, Whittington JR, Wendel MP, Sandlin AT, Ounpraseuth ST. Accuracy of the ultrasound estimate of the amniotic fluid volume (amniotic fluid index and single deepest pocket) to identify actual low, normal, and high amniotic fluid volumes as determined by quantile regression. J Ultrasound Med. 2020;39:373‐378. doi: 10.1002/jum.15116 [DOI] [PubMed] [Google Scholar]
  • 12. Kadji C, Cannie MM, Resta S, et al. Magnetic resonance imaging for prenatal estimation of birthweight in pregnancy: review of available data, techniques, and future perspectives. Am J Obstet Gynecol. 2019;220:428‐439. doi: 10.1016/j.ajog.2018.12.031 [DOI] [PubMed] [Google Scholar]
  • 13. Liao K, Tang L, Peng C, et al. Two new models for the estimation of foetal weight more than a week before delivery: An MRI study. Eur J Radiol. 2019;121:108729. doi: 10.1016/j.ejrad.2019.06.026 [DOI] [PubMed] [Google Scholar]
  • 14. Cannie MM, Jani JC, Van Kerkhove F, et al. Fetal body volume at MR imaging to quantify total fetal lung volume: Normal ranges. Radiology. 2008;247:197‐203. doi: 10.1148/radiol.2471070682 [DOI] [PubMed] [Google Scholar]
  • 15. Story L, Hutter J, Zhang T, Shennan AH, Rutherford M. The use of antenatal fetal magnetic resonance imaging in the assessment of patients at high risk of preterm birth. Eur J Obstet Gynecol Reprod Biol. 2018;222:134‐141. doi: 10.1016/j.ejogrb.2018.01.014 [DOI] [PubMed] [Google Scholar]
  • 16. Li R, Song F, Zhou Q, et al. A hybrid model for fetal growth restriction assessment by automatic placental radiomics on T2‐weighted MRI and multifeature fusion. J Magn Reson Imaging. 2024;61:494‐504. doi: 10.1002/jmri.29399 [DOI] [PubMed] [Google Scholar]
  • 17. Bouachba A, Bartin R, De Jesus Neves J, et al. Normative range of MRI‐based fetal body volume and association with ultrasonographically estimated fetal weight at 16–36 weeks: Prospective study. Ultrasound Obstet Gynecol. 2025;66:81‐88. doi: 10.1002/uog.29234 [DOI] [PubMed] [Google Scholar]
  • 18. Ronneberger O, Fischer P, Brox T. U‐Net: Convolutional networks for biomedical image segmentation. International Conference on Medical Image Computing and Computer‐Assisted Intervention. Springer International Publishing; 2015:234‐241. [Google Scholar]
  • 19. Lo J, Nithiyanantham S, Cardinell J, et al. Cross attention squeeze excitation network (Case‐net) for whole body fetal MRI segmentation. Sensors (Basel). 2021;21:4490. doi: 10.3390/s21134490 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Specktor‐Fadida B, Link‐Sourani D, Rabinowich A, et al. Deep learning–based segmentation of whole‐body fetal MRI and fetal weight estimation: assessing performance, repeatability, and reproducibility. Eur Radiol. 2024;34:2072‐2083. doi: 10.1007/s00330-023-10038-y [DOI] [PubMed] [Google Scholar]
  • 21. Bischoff LM, Nowak S, Mader M, et al. Fetal MRI deep learning segmentation of body and lung in congenital diaphragmatic hernia. Radiol Adv. 2025;2:umae034. doi: 10.1093/radadv/umae034 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Pietsch M, Ho A, Bardanzellu A, et al. APPLAUSE: Automatic prediction of placental health via U‐net segmentation and statistical evaluation. Med Image Anal. 2021;72:102145. doi: 10.1016/j.media.2021.102145 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Li J, Shi Z, Zhu J, et al. Placenta segmentation in magnetic resonance imaging: Addressing position and shape of uncertainty and blurred placenta boundary. Biomed Signal Process Control. 2024;88:105680. doi: 10.1016/j.bspc.2023.105680 [DOI] [Google Scholar]
  • 24. Costanzo A, Ertl‐Wagner B, Sussman D. AFNet algorithm for automatic amniotic fluid segmentation from fetal MRI. Bioengineering. 2023;10:783. doi: 10.3390/bioengineering10070783 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Csillag D, Monteiro Paes L, Ramos T, et al. AmnioML: Amniotic fluid segmentation and volume prediction with uncertainty quantification. Proc AAAI Conf Artif Intell. 2023;37:15494‐15502. doi: 10.1609/aaai.v37i13.26837 [DOI] [Google Scholar]
  • 26. Gagoski B, Xu J, Wighton P, et al. Automated detection and reacquisition of motion‐degraded images in fetal HASTE imaging at 3 T. Magn Reson Med. 2022;87:1914‐1922. doi: 10.1002/mrm.29803 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Aviles Verdera J, Bortolazzi A, Neves Silva S, et al. HERON: High‐efficiency real‐time motion quantification and re‐acquisition for fetal diffusion MRI. IEEE Trans Med Imaging. 2025;1. doi: 10.1109/TMI.2025.3569853 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Neves Silva S, Aviles Verdera J, Tomi‐Tricot R, et al. Real‐time fetal brain tracking for functional fetal MRI. Magn Reson Med. 2023;90:2306‐2320. doi: 10.1002/mrm.29803 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Uus A, Neves Silva S, Aviles Verdera J, et al. Scanner‐based real‐time three‐dimensional brain+body slice‐to‐volume reconstruction for T2‐weighted 0.55‐T low‐field fetal magnetic resonance imaging. Pediatr Radiol. 2025;55:556‐569. doi: 10.1007/s00247-025-06165-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Neves Silva S, McElroy S, Aviles Verdera J, et al. Fully automated planning for anatomical fetal brain MRI on 0.55T. Magn Reson Med. 2024;92:1263‐1276. doi: 10.1002/mrm.30122 [DOI] [PubMed] [Google Scholar]
  • 31. Chow K, Kellman P, Xue H. Prototyping image reconstruction and analysis with FIRE. SCMR 24th Annual Scientific Sessions. Virtual Meeting; 2021. [Google Scholar]
  • 32. Isensee F, Jaeger PF, Kohl SAA, Petersen J, Maier‐Hein KH. nnU‐Net: a self‐configuring method for deep learning‐based biomedical image segmentation. Nat Methods. 2021;18:203‐211. doi: 10.1038/s41592-020-01008-z [DOI] [PubMed] [Google Scholar]
  • 33. Baker PN, Johnson IR, Gowland PA, et al. Fetal weight estimation by echo‐planar magnetic resonance imaging. Lancet. 1994;343:644‐645. doi: 10.1016/S0140-6736(94)92638-7 [DOI] [PubMed] [Google Scholar]
  • 34. Kacem Y, Cannie MM, Kadji C, et al. Fetal weight estimation: comparison of two‐dimensional US and MR imaging assessments. Radiology. 2013;267:902‐910. doi: 10.1148/radiol.12121374 [DOI] [PubMed] [Google Scholar]
  • 35. Royston P, Wright EM. How to construct ‘normal ranges’ for fetal variables. Ultrasound Obstet Gynecol. 1998;11:30‐38. doi: 10.1046/j.1469-0705.1998.11010030.x [DOI] [PubMed] [Google Scholar]
  • 36. Peterson HF, Eskild A, Sommerfelt S, et al. Percentiles of intrauterine placental volume and placental volume relative to fetal volume: A prospective magnetic resonance imaging study. Placenta. 2022;121:40‐45. doi: 10.1016/j.placenta.2022.02.023 [DOI] [PubMed] [Google Scholar]
  • 37. Sanchez T, Esteban O, Gomez Y, et al. FetMRQC: A robust quality control system for multi‐centric fetal brain MRI. Med Image Anal. 2024;97:103282. doi: 10.1016/j.media.2024.103282 [DOI] [PubMed] [Google Scholar]
  • 38. Lim A, Lo J, Wagner MW, Ertl‐Wagner B, Sussman D. Motion artifact correction in fetal MRI based on a generative adversarial network method. Biomed Signal Process Contr. 2023;81:104484. doi: 10.1016/j.bspc.2022.104484 [DOI] [Google Scholar]
  • 39. Uus AU, Kyriakopoulou V, Makropoulos A, et al. BOUNTI: Brain volumetry and automated parcellatioN for 3D fetal MRI. elife. 2023;12:RP88818. doi: 10.7554/eLife.88818.1 [DOI] [Google Scholar]
  • 40. Uus AU, Hall M, Grigorescu I, et al. Automated body organ segmentation, volumetry and population‐averaged atlas for 3D motion‐corrected T2‐weighted fetal body MRI. Sci Rep. 2024;14:6637. doi: 10.1038/s41598-024-57087-x [DOI] [PMC free article] [PubMed] [Google Scholar]