1. Introduction
The electrical properties of human tissue (i.e., electrical conductivity σ and permittivity ε) can be measured quantitatively in vivo with standard MR sequences and standard MR systems using a technique called Electrical Properties Tomography (EPT) [1, 2]. The full implementation of EPT involves first measuring the complex RF transmit field component B1 + (i.e., its magnitude |B1 +| and phase ϕ+) as modulated by the tissue electrical properties, and then reconstructing σ and ε from the modulated B1 +. The reconstructed electrical properties correspond to the Larmor frequency of the applied main field strength B0.
This article is meant as a short guideline for measuring complex B1 + for EPT. A comprehensive review on how to perform the subsequent reconstruction of σ and ε from the measured complex B1 + was reported earlier [3]. Software packages performing such reconstructions are available publicly [4, 5, 6], and so are EPT example datasets [7, 8].
This guideline will be complemented by two further EPT‐related guidelines, that is on standardization and phantom building, provided in separate articles. Together, these three guidelines are thought to parallel corresponding guidelines on mapping magnetic properties of tissue [9].
2. Measuring Complex B1
Although sequences with simultaneous measurements of |B1 +| and ϕ+ are possible [10], current practice is to use separate measurements for |B1 +| and ϕ+, as described in the following.
3. Measuring B1 ‐Magnitude
Numerous B1‐mapping techniques for measuring |B1 +| have been published, independent of EPT (e.g., Actual Flip angle Imaging (AFI) [11], Bloch‐Siegert mapping [12], and Dual Refocusing Echo Acquisition Mode (DREAM) [13]). A comparison of B1‐mapping techniques on the determination of ε primarily highlighted the need for high SNR [14]. On the other hand, reviews of B1‐mapping techniques without reference to EPT were published [15].
Although relative units of |B1 +| are sufficient for the majority of EPT approaches, such as differential equation‐based methods, absolute scaling of |B1 +| (e.g., in μT) is required for other EPT approaches, like some integral equation‐based methods, as these rely on accurate field magnitudes to compute the necessary integrals.
4. Measuring B1 ‐Phase
A standard MR system does not allow ϕ+ to be measured. Consequently, most EPT studies make use of the so‐called transceive phase ϕ±, which is the superposition of the B1‐phases from RF transmission and RF reception, assuming ϕ+~ϕ±/2 (so‐called “Transceive Phase Assumption”, TPA). Thus, this subsection describes the measurement of ϕ±.
In principle, the phase image of any clinical sequence can be taken as ϕ±; however, it must be free of components unrelated to RF penetration. Unwanted phase contributions originate mainly from three sources: (A) B0‐inhomogeneities, (B) gradient‐induced eddy currents, (C) patient/tissue motion.
(A) Phase due to B0‐inhomogeneities can be avoided by using spin‐echo‐based sequences, gradient‐echo‐based sequences with balanced gradients, or sequences with ultrashort/zero TE [16]. Gradient‐echo‐based sequences without gradient balancing can be used to estimate ϕ± if multiple echoes with different TEs are acquired and the observed phase for each voxel is extrapolated back to TE = 0 [17]. In the presence of chemical shift effects (e.g., due to fat), one might not be able to resolve the additional phase component.
(B) Unwanted phase due to eddy currents typically appears as artificially increased conductivity along one side of the field‐of‐view (FOV) and artificially decreased conductivity along the other side of the FOV. This too can be eliminated by using gradient‐echo‐based sequences with balanced gradients or by repeating spin‐echo sequences with inverted gradient polarization and averaging the results [18].
(C) As motion affects all types of sequences, standard motion correction schemes should be applied to minimize motion‐induced phase contributions. Due to the applied numerical derivatives, reconstructed maps are typically more corrupted by motion than the original scan's magnitude/phase image itself.
5. General Remarks
The optimal field strength B0 for conductivity mapping involves a trade‐off between sufficient SNR (i.e., high B0) and the validity of the TPA, which is increasingly violated with increasing B0 and object size. For brain applications (and with today's reported EPT techniques), B0 = 3T appears to provide a good trade‐off [18]. B0 = 1.5T might be a better choice for other in vivo applications, and B0 > 3T for imaging small animals. For permittivity, TPA plays a minor role, and the quest for optimal B0 is dominated by SNR, with accuracy increasing at least up to B0 = 7T [18]. Future studies for the upcoming systems with B0 > 7T might reveal an upper bound for optimal permittivity imaging.
The optimal voxel size for EPT is chosen by trading‐off sufficient SNR (i.e., large voxels) against the need for a sufficient number of voxels per spatial direction (i.e., small voxels) for stable numerical calculations of the Laplacian operator [19], which requires a minimum of N = 3 voxels per spatial direction within a tissue compartment. Although N = 3 is sufficient for an ideal noise‐free case, in realistic measurements including noise, N > 10 is frequently recommended. This aspect should be taken into account while defining the applied sequence's voxel size, as anatomic compartments smaller than N voxels cannot be reconstructed reliably. For conductivity studies, this consideration applies mainly to measuring ϕ±, as |B1 +| acts here mainly as a spatially smooth correction term.
A volumetric image (i.e., a 3D or multi‐slice 2D scan) is required for the full version of EPT. 3D scans may be preferred as they avoid through‐plane inconsistencies when measuring ϕ±. Alternatively, the EPT reconstructions can be restricted to 2D in‐plane, thus eliminating through‐plane inconsistencies and introducing the additional model assumption of constant B1 + in the through‐plane direction.
If an RF receive coil array is used, care should be taken when combining the different array elements due to the possibility of violating the TPA [20]. RF transmission is based on quadrature (body or head) coils in the majority of EPT studies published yet.
Acceleration techniques are compatible with EPT, at least using moderate acceleration factors [21].
6. Summary—Recommendations
To support brain conductivity imaging, |B1 +| can be acquired using low‐resolution B1‐mapping (e.g., voxel‐size = 4 (RO) × 8 (PE) × 8 (PE) mm3), followed by interpolation to the higher resolution of the ϕ± map (see below). For permittivity imaging, a higher resolution of B1‐mapping is needed (e.g., voxel‐size = 2 × 2 × 2 mm3).
If available, low‐resolution B1‐mapping might be acquired using dual TR sequences (e.g., AFI with TR1/TR2 = 50/200 ms, TE = 2.3 ms, scan‐time = 2.3 min) [11], and high‐resolution B1‐mapping using dual TE sequences (e.g., DREAM with TR = 7 ms, TE1/TE2 = 2.5/3.4 ms, 2 averages, scan‐time = 2.2 min) [13]. Other vendor‐specific sequences may also be available. However, the specific choice of the B1‐mapping sequence does not seem to be of utmost importance, as long as the main demand for high SNR is fulfilled [14].
To image ϕ±, we recommend a 3D acquisition (e.g., 3D balanced Fast Field Echo (bFFE)/balanced Steady State Free Precession (bSSFP), voxel‐size = 1 × 1 × 1 mm3, TR = 3 ms, TE = 1.5 ms, dynamics = 2, scan‐time = 3.5 min). We recommend the acquisition of multiple, separate dynamics rather than signal averages to better resolve artifacts arising for example from flow, and allow co‐registration to reduce motion effects. Potential banding artifacts might be reduced via phase cycling or acquisitions with different TR. A 2D acquisition may be used if fast temporal changes in conductivity are to be examined.
Acknowledgments
The authors thank the ISMRM‐EMTP study group for reviewing both the original and the revised version of the guideline. In addition to the authors, the following members of the ISMRM‐EMTP Study Group explicitly recommend this guideline: A. Arduino, T. Cuhna, R. Dadarwal, X. Li, C. Rae, K. Shmueli, K.K. Tha, L. Zilberti. SM received funding from NWO‐VENI 18078 for the coordination of this work and acknowledges funding from the 24DIT01 APULEIO project. The project 24DIT01 APULEIO has received funding from the European Partnership on Metrology, co‐financed from the European Union's Horizon Europe Research and Innovation Programme and by the Participating States. IIG acknowledges funding from NIH K99 EB035163.
Mandija S., van den Berg C. A. T., Giannakopoulos I., et al., “ MR Electrical Properties Tomography Acquisitions: A Guideline From the ISMRM Electro‐Magnetic Tissue Properties Study Group,” Journal of Magnetic Resonance Imaging 63, no. 1 (2026): 279–281, 10.1002/jmri.70057.
Cornelis A.T. van den Berg, Ilias Giannakopoulos, Zhongzheng He, Yusuf Ziya Ider, Kyu‐Jin Jung, Nitish Katoch, Dong‐Hyun Kim, Riccardo Lattanzi, and Paul Soullié contributed equally to this study.
References
- 1. Katscher U. and van den Berg C. A. T., “Electric Properties Tomography: Biochemical, Physical and Technical Background, Evaluation and Clinical Applications,” NMR in Biomedicine 30, no. 8 (2017): 3729. [DOI] [PubMed] [Google Scholar]
- 2. Zhang X., Liu J., and He B., “Magnetic‐Resonance‐Based Electrical Properties Tomography: A Review,” IEEE Reviews in Biomedical Engineering 7 (2014): 87–96. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Leijsen R., Brink W., van den Berg C., Webb A., and Remis R., “Electrical Properties Tomography: A Methodological Review,” Diagnostics 11, no. 2 (2021): 176–213. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Arduino A., “EPTlib: An Open‐Source Extensible Collection of Electric Properties Tomography Techniques,” Applied Sciences 11, no. 7 (2021): 3237. [Google Scholar]
- 5. “Vision Transformer‐Based Network and Synthetic Phantom Data for MR Electrical Properties Tomography,” https://cai2r.net/resources/vision‐transformer‐network‐for‐mr‐electrical‐properties‐tomography/.
- 6. “MRI Conductivity,” https://xip.uclb.com/product/MRI_conductivity.
- 7. Meerbothe T. G., Meliado E. F., Stijnman P. R. S., van den Berg C. A. T., and Mandija S., “A Database for MR‐Based Electrical Properties Tomography With In Silico Brain Data‐ADEPT,” Magnetic Resonance in Medicine 91, no. 3 (2024): 1190–1199. [DOI] [PubMed] [Google Scholar]
- 8. Cao J. and Rae C., “MRI Collection of Functional Conductivity Imaging, UNSW, Sydney, Dataset,” 10.26190/M3F9-1N36. [DOI]
- 9. QSM Consensus Organization Committee , Bilgic B., Costagli M., et al., “Recommended Implementation of Quantitative Susceptibility Mapping for Clinical Research in the Brain: A Consensus of the ISMRM Electro‐Magnetic Tissue Properties Study Group,” Magnetic Resonance in Medicine 91, no. 5 (2024): 1834–1862. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Iyyakkunnel S., Weigel M., Ganter C., and Bieri O., “Complex B1 + Mapping With Carr‐Purcell Spin Echoes and Its Application to Electrical Properties Tomography,” Magnetic Resonance in Medicine 87, no. 3 (2022): 1250–1260. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Yarnykh V. L., “Actual Flip‐Angle Imaging in the Pulsed Steady State: A Method for Rapid Three‐Dimensional Mapping of the Transmitted Radiofrequency Field,” Magnetic Resonance in Medicine 57, no. 1 (2007): 192–200. [DOI] [PubMed] [Google Scholar]
- 12. Sacolick L. I., Wiesinger F., Hancu I., and Vogel M. W., “B1 Mapping by Bloch–Siegert Shift,” Magnetic Resonance in Medicine 63, no. 5 (2010): 1315–1322. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Nehrke K. and Börnert P., “DREAM‐A Novel Approach for Robust, Ultrafast, Multislice B1 Mapping,” Magnetic Resonance in Medicine 68, no. 5 (2012): 1517–1526. [DOI] [PubMed] [Google Scholar]
- 14. Gavazzi S., van den Berg C. A. T., Sbrizzi A., et al., “Accuracy and Precision of Electrical Permittivity Mapping at 3T: The Impact of Three B1 Mapping Techniques,” Magnetic Resonance in Medicine 81, no. 6 (2019): 3628–3642. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Pohmann R. and Scheffler K., “A Theoretical and Experimental Comparison of Different Techniques for B1 Mapping at Very High Fields,” NMR in Biomedicine 26, no. 3 (2013): 265–275. [DOI] [PubMed] [Google Scholar]
- 16. Lee S. K., Bulumulla S., Wiesinger F., Sacolick L., Sun W., and Hancu I., “Tissue Electrical Properties Mapping From Zero Echo‐Time Magnetic Resonance Imaging,” IEEE Transactions on Medical Imaging 34, no. 2 (2015): 541–550. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Ryu K., Shin J., Lee H., Kim J. H., and Kim D. H., “Multi‐Echo GRE‐Based Conductivity Imaging Using Kalman Phase Estimation,” Magnetic Resonance in Medicine 81, no. 1 (2019): 702–710. [DOI] [PubMed] [Google Scholar]
- 18. van Lier A. L. H., Raaijmakers A., Voigt T., Lagendijk J. J., Katscher U., and van den Berg C. A. T., “Electric Properties Tomography in the Human Brain at 1.5, 3, and 7T: A Comparison Study,” Magnetic Resonance in Medicine 71, no. 1 (2014): 354–363. [DOI] [PubMed] [Google Scholar]
- 19. Mandija S., Sbrizzi A., Katscher U., Luijten P., and van den Berg C. A. T., “Error Analysis of Helmholtz‐Based MR‐Electrical Properties Tomography,” Magnetic Resonance in Medicine 80, no. 1 (2018): 90–100. [DOI] [PubMed] [Google Scholar]
- 20. Lee J. S., Shin J. W., and Kim D. H., “MR‐Based Conductivity Imaging Using Multiple Receiver Coils,” Magnetic Resonance in Medicine 76, no. 2 (2016): 530–539. [DOI] [PubMed] [Google Scholar]
- 21. Cao J., Ball I., Humburg P., Dokos S., and Rae C., “Repeatability of Brain Phase‐Based Magnetic Resonance Electric Properties Tomography Methods and Effect of Compressed SENSE and RF Shimming,” Physical and Engineering Sciences in Medicine 46, no. 2 (2023): 753–766. [DOI] [PMC free article] [PubMed] [Google Scholar]