Construction of Phantoms for MR Electrical Properties Tomography (From Structure to Composition): A Guideline From the ISMRM Electro‐Magnetic Tissue Properties Study Group

Ilias I Giannakopoulos 1,, Alessandro Arduino 2, Cornelis A T van den Berg 3, Zhongzheng He 4, Kyu‐Jin Jung 5, Dong‐Hyun Kim 5, Riccardo Lattanzi 1, Jessica A Martinez 6,7, Thierry Meerbothe 3, Freddy Odille 4,8, Adriano Troia 2, Luca Zilberti 2, Stefano Mandija 3
PMCID: PMC12706702  PMID: 40831322

MR‐based Electrical Properties (EPs) Tomography (MR‐EPT) denotes non‐invasive electrical conductivity (σe ) and permittivity (ε r ) mapping methods using MR measurements. The lack of standardized tissue‐mimicking phantoms hinders reproducibility studies and method comparisons. For example, the NIST/ISMRM MRI system phantom has contributed significantly to standardization efforts in relaxometry, underscoring the importance of having similar benchmarks for MR‐EPT. The guidelines presented herein outline the importance of MR‐EPT phantom design and construction, focusing on structure, composition, and reliability.

1. Phantom Structure

The structure should mimic the shape and size of the anatomy of interest. For example, the head can be modeled using a simple spherical geometry or with anatomically realistic designs such as the one presented in [1]. The abdomen can be represented using large elliptic cylindrical phantoms or specialized holders that conform to human anatomy and are commonly used for training purposes [2, 3]. This allows for testing compatibility with standard receive coils and allows the use of established methods to mitigate B 0 inhomogeneity artifacts and spurious phase contributions. Phantoms that are too small to sufficiently load the coil could detune the coil elements or couple opposite receivers. Moreover, phantoms with translational symmetry along the longitudinal axis and significantly larger geometry in the Frankfort horizontal orientation (B 0 direction) than the anterior‐to‐posterior and right‐to‐left (e.g., cylinders with length ≥ 4× radius) orientations are ideal for 2D MR‐EPT methods since the derivatives along z of B 1 in the midplane of a birdcage coil are approximately null. However, they do not provide a good validity test for human anatomies.

To discriminate between different tissue EPs, heterogeneous phantoms should be used. The phantom's compartments can be separated with plastic boundaries (e.g., for liquid‐based compounds) or kept in direct contact (gel‐based compounds). Liquid‐based compounds may be suboptimal due to the risk of flow artifacts in images, which are avoided in gel‐based compounds. Gel‐based compounds may be put in direct contact [4], but electrolyte diffusion between compartments may alter the conductivity and internal geometry near the interfaces, as highlighted in [5, 6, 7]. To avoid these issues, solid structures can be used to separate different compartments. However, due to the significant mismatch in EP between the plastic dividers and adjacent tissue‐mimicking materials, these interfaces introduce localized artifacts [1]. In particular, at least one voxel in the magnitude transmit magnetic field is consistently corrupted by noise near the boundary. Figure 1 illustrates this boundary effect in a two‐compartment cylindrical phantom, demonstrating how these image artifacts around the plastic dividers limit EP mapping assessment in small structures. Precise quantification of EP reconstruction errors related to plastic boundary thickness and material composition has not yet been established in MR‐EPT literature.

FIGURE 1.

FIGURE 1

Magnitude of the transmit magnetic field (left) and transceive phase (right) for an axial slice of a two‐compartment cylindrical phantom (details on compartment EP values are provided in the next sections). The phantom was scanned at 2.5 mm isotropic resolution using the RF map sequence [8]. The plastic boundary separating the two compartments was approximately 2.5 mm thick. This boundary region is noisy due to the lack of MR signal in both the magnitude and phase maps.

2. Phantom Composition

Phantoms should use deionized water as the solvent. Sodium chloride (NaCl) can be added to raise conductivity [5, 9, 10, 11]. Permittivity can be reduced using low‐permittivity materials such as polyvinylpyrrolidone (PVP) and sucrose. Although more costly, PVP is preferred, as sucrose can lead to stability issues over time, degradation in highly concentrated solutions, caramelization when exposed to high temperatures, and T 2* shortening that may lower the SNR [12]. Ethanol and other organic solvents can also be used for permittivity reduction, but they have either poor water solubility, are volatile, or require light‐proof containers [5]. Glycerol can also be used, but it can induce chemical shift artifacts [11] and its high viscosity limits its useful concentration to around 30% and 50% for gel‐based and liquid‐based phantoms, respectively. Barium titanate (BaTiO3) or calcium titanate (CaTiO3) can increase permittivity, and due to their low solubility, osmosis is negligible. BaTiO3 is often preferred since it requires a lower concentration to achieve the desired permittivity value [13]. Agar or gelatin can be utilized for phantoms that require low heat diffusivity (to assess RF coil safety). Preservatives such as CuSO4, Proclin, or benzoic acid can enhance stability over time. Toxic and reactive preservatives like sodium azide should be avoided.

Relaxation times, particularly T 1, can be adjusted with paramagnetic salts such as gadolinium chloride (GdCl3), copper(II) sulfate (CuSO4), or manganese(II) chloride (MnCl2) [14]. GdCl3 requires a lower concentration as a T 1‐shortening agent [14], which helps minimize diffusion effects in phantoms without plastic boundaries, but it is toxic and has a pronounced impact on EP due to its high ionic charge. Agar and gelatin can also be used as T 1 and T 2 shortening agents [1, 11, 14, 15].

Phantom components should be weighed before construction to minimize errors in concentrations and potential mass loss during the high‐temperature steps of the mixing process. Once the Larmor frequency and target EPs are established according to the organ or the tissue of interest (e.g., brain, prostate, liver, etc.), the required amounts of PVP and NaCl may be calculated using, for example, the software outlined in [12].

Realistic tissue EPs should be targeted in phantom preparations, for example, relative permittivity < 120, conductivity < 2.5 S/m for 3 T [1]. When the conductivity is negligible, high permittivity can lead to interfering B 1 patterns due to its inverse relationship with the wavelength, especially at fields ≥ 3 T [9]. High conductivity may give rise to B 1 inhomogeneities due to eddy currents [9], which at fields ≥ 3 T can impair the performance of EPT methods that are based on specific assumptions that may no longer hold (namely, the transceive phase assumption, or negligible B z gradient).

3. Example Procedure for Phantom Preparation

First, weigh the appropriate amount of water in a sealable borosilicate glass bottle. Second, dissolve the NaCl in the water. Third, gradually add the PVP (if used) while stirring vigorously. If the PVP concentration exceeds 30%, heat the solution to 60°C in a microwave oven. Typically, 500 g of solution is heated for 4 min at 900 W in a commercial microwave oven. Depending on PVP concentration, this procedure could be repeated, keeping the solution under agitation until it becomes fully transparent with a slight amber tint [16]. After this stage, agar or gelatin can be added for gel‐based phantoms. The mixture should be heated up to allow them to fully dissolve (agar: > 95°C, gelatine: > 70°C). Finally, during the cooling step under agitation (at room temperature), benzoic acid (or other preservatives) and the paramagnetic salt (if used) must be added for long‐term stability to prevent organic degradation and to tune the phantom relaxation times, respectively. Pouring into the phantom scaffold should be done carefully when the solution temperature is > 65°C, since the higher viscosity traps air bubbles at lower temperatures. Gentle shaking of the hot solution helps raise the bubbles to the surface. Alternatively, one can vacuum the material [17] or keep it in a water bath for 45 min before allowing it to cool to room temperature [18]. For gel‐based heterogeneous phantoms, solidification of the first gel is required before pouring the next one [4]. The application of this procedure with the quantities reported in Table 1 would result in two materials (C‐1 and C‐2) mimicking the electromagnetic behavior of cerebrospinal fluid and gray matter when exposed to the RF field of a 3 T MRI scanner.

TABLE 1.

Example recipe for a two‐compartment (C‐1, C‐2) phantom used for EPT reconstruction at 3 T MRI.

Recipe Targets Measurements
PVP NaCl MnCl2 ε r σ e ε r σ e T 1 T 2
Units g g mg S/m S/m ms ms
C‐1 131.19 18.32 3.5 73.6 2.17 74.9 ± 3.8 1.97 ± 0.10 1080 ± 203 242 ± 12
C‐2 433.96 9.41 5.3 60.9 0.60 63.8 ± 3.2 0.63 ± 0.03 875 ± 148 423 ± 8

Note: The recipe is normalized to 1 kg of water. The relative permittivity (ε r ) and the conductivity (σ e ) were measured with a dielectric probe (Agilent, Santa Clara, CA), at 22.1°C and 128 MHz. The phantom's holder was constructed using cast acrylic. T 1 and T 2 values were measured with MP2RAGE and CPMG sequences, respectively.

4. Phantom Lifetime

A quality assurance protocol should be put in place to monitor the temporal stability of the phantom. However, since the criteria for determining when a phantom should be decommissioned are yet to be defined, reporting the lifetime of the phantom in publications can help understand the possible discrepancies between expected (from constructions) and measured EPs.

5. Reporting

When reporting phantom recipes, all concentrations should be expressed in terms of mass/mass (g/kg, mg/g), as it is independent of temperature fluctuations. Considering the variability of EPs with Larmor frequency and temperature, we recommend reporting the target EP values for a given frequency and temperature (e.g., as calculated using [12]), and when feasible, the measured EPs with an open‐ended coaxial probe [5, 19] (see an example in Table 1), including probe model. The sample characterized with the coaxial probe should be large enough to mitigate measurement errors. To improve the precision, we suggest repeating probe measurements five times and reporting the average and standard deviation values. If measurements last longer than 1 h, it is advisable to report the sample temperature before and after the experiment. If the T 1 and T 2 were modulated, it is also beneficial to report them. The type of plastic used for the phantom scaffold should also be reported. Finally, we recommend including a figure that shows the geometry and dimensions (compartment size and boundary thickness) to facilitate reproducibility, as in Figure 2.

FIGURE 2.

FIGURE 2

Schematic representation of a two‐compartment cylindrical liquid phantom used for EP mapping in [20]. The phantom consists of an outer hollow cylindrical and an inner cylindrical compartment. The total length of the phantom is 21.5 cm. The inner radius of the hollow cylinder is 6.875 cm, while the inner cylinder has an inner radius of 3.1 cm and is offset by 0.5 cm from the central axis of the hollow cylinder. The scaffold that contains the liquid solutions has a uniform thickness of 0.5 cm, except at the interface between the two compartments, where it is reduced to 0.3 cm.

Disclosure

Certain commercial equipment, instruments, software, or materials are identified in this paper in order to specify the experimental procedure adequately. Such identification is not intended to imply recommendation or endorsement by NIST, nor is it intended to imply that the materials or equipment identified are necessarily the best available for the purpose.

Acknowledgments

I.I.G. was supported in part by National Institutes of Health K99 EB035163, and this work was performed under the rubric of the Center for Advanced Imaging Innovation and Research (CAI2R, www.cai2r.net), an NIBIB National Center for Biomedical Imaging and Bioengineering (NIH P41 EB017183). A.A., T.M., A.T., L.Z. and S.M. acknowledge funding from the 24DIT01 APULEIO project. The project 24DIT01 APULEIO has received funding from the European Partnership on Metrology, co‐financed from the European Union’s Horizon Europe Research and Innovation Programme and by the Participating States. J.A.M. would like to acknowledge support from NIST‐PREP (Professional Research Experience Program), performed under the following financial assistance award 70NANB18H006 from the U.S. Department of Commerce, National Institute of Standards and Technology. S.M. received funding from the NWO VENI Grant 18078. The following members of the ISMRM‐EMTP Study Group explicitly recommend this guideline: C. Rae, Y.Z. Ider, K.K. Tha, K. Shmueli, U. Katscher, R. Dadarwal, X. Li. We also thank the whole ISMRM‐EMTP study group for reviewing both the original and the revised version of the guideline.

Giannakopoulos I. I., Arduino A., van den Berg C. A. T., et al., “Construction of Phantoms for MR Electrical Properties Tomography (From Structure to Composition): A Guideline From the ISMRM Electro‐Magnetic Tissue Properties Study Group,” Journal of Magnetic Resonance Imaging 63, no. 1 (2026): 282–285, 10.1002/jmri.70059.

Funding: This work was supported by National Institutes of Health (K99 EB035163, P41 EB017183); NIST‐PREP (70NANB18H006); NWO‐VENI (18078) and by the European Partnership on Metrology (24DIT01 APULEIO).

Ilias I. Giannakopoulos and Alessandro Arduino contributed equally to this work.

References

  • 1. Meerbothe T. G., Floczak S., van den Berg C. A. T., Levato R., and Mandija S., “A Reusable 3D Printed Brain‐Like Phantom for Benchmarking Electrical Properties Tomography Reconstructions,” Magnetic Resonance in Medicine 92 (2024): 2271–2279. [DOI] [PubMed] [Google Scholar]
  • 2. Westhoff N., Siegel F. P., Hausmann D., et al., “Precision of MRI/Ultrasound‐Fusion Biopsy in Prostate Cancer Diagnosis: An Ex Vivo Comparison of Alternative Biopsy Techniques on Prostate Phantoms,” World Journal of Urology 35 (2017): 1015–1022. [DOI] [PubMed] [Google Scholar]
  • 3. Hoffmans D., Niebuhr N., Bohoudi O., Pfaffenberger A., and Palacios M., “An End‐To‐End Test for MR‐Guided Online Adaptive Radiotherapy,” Physics in Medicine and Biology 65, no. 12 (2020): 125012. [DOI] [PubMed] [Google Scholar]
  • 4. Zilberti L., Arduino A., Zanovello U., et al., “Magnetic Resonance‐Based Electric Properties Tomography via Green's Integral Identity,” IEEE Access 13 (2025): 42029–42044. [Google Scholar]
  • 5. He Z., Lefebvre P. M., Soullié P., et al., “Phantom Evaluation of Electrical Conductivity Mapping by MRI: Comparison to Vector Network Analyzer Measurements and Spatial Resolution Assessment,” Magnetic Resonance in Medicine 91, no. 6 (2024): 2374–2390. [DOI] [PubMed] [Google Scholar]
  • 6. Fieremans E. and Lee H. H., “Physical and Numerical Phantoms for the Validation of Brain Microstructural MRI: A Cookbook,” NeuroImage 182 (2018): 39–61. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Hamamura M. J., Muftuler L. T., Birgul O., and Nalcioglu O., “Measurement of Ion Diffusion Using Magnetic Resonance Electrical Impedance Tomography,” Physics in Medicine and Biology 51, no. 11 (2006): 2753–2762. [DOI] [PubMed] [Google Scholar]
  • 8. Akoka S., Florence F., Seguin F., and Le Pape A., “Radiofrequency Map of an NMR Coil by Imaging,” Magnetic Resonance Imaging 11 (1993): 437–441. [DOI] [PubMed] [Google Scholar]
  • 9. Vaidya M. V., Collins C. M., Sodickson D. K., Brown R., Wiggins G. C., and Lattanzi R., “Dependence of and Field Patterns of Surface Coils on the Electrical Properties of the Sample and the MR Operating Frequency,” Concepts in Magnetic Resonance Part B: Magnetic Resonance Engineering 46, no. 1 (2016): 25–40. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Stogryn A., “Equations for Calculating the Dielectric Constant of Saline Water,” IEEE Transactions on Microwave Theory and Techniques 19, no. 8 (1971): 733–736. [Google Scholar]
  • 11. Yusuff H., Chatelin S., and Dillenseger J.‐P., “Narrative Review of Tissue‐Mimicking Materials for MRI Phantoms: Composition, Fabrication, and Relaxation Properties,” Radiography 30, no. 6 (2024): 1655–1668. [DOI] [PubMed] [Google Scholar]
  • 12. Ianniello C., de Zwart J. A., Duan Q. I., et al., “Synthesized Tissue‐Equivalent Dielectric Phantoms Using Salt and Polyvinylpyrrolidone Solutions,” Magnetic Resonance in Medicine 80, no. 1 (2018): 413–419. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Vaidya M. V., Deniz C. M., Collins C. M., Sodickson D. K., and Lattanzi R., “Manipulating Transmit and Receive Sensitivities of Radiofrequency Surface Coils Using Shielded and Unshielded High‐Permittivity Materials,” Magma 31 (2018): 355–366. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Antoniou A. and Christakis D., “MR Relaxation Properties of Tissue‐Mimicking Phantoms,” Ultrasonics 119 (2022): 106600. [DOI] [PubMed] [Google Scholar]
  • 15. Hattori K., Ikemoto Y., Takao W., et al., “Development of MRI Phantom Equivalent to Human Tissues for 3.0‐T MRI,” Medical Physics 40, no. 3 (2013): 032303. [DOI] [PubMed] [Google Scholar]
  • 16. Nobre P., Gaborit G., Troia A., Zanovello U., Duvillaret L., and Beuf O., “Electric Field Measurements in Preclinical MRI at 11.7 T and 7 T for Experimental SAR Comparison,” Journal of Magnetism and Magnetic Materials 593 (2024): 171818. [Google Scholar]
  • 17. Hacker L., Ivory A. M., Joseph J., et al., “A Stable Phantom Material for Optical and Acoustic Imaging,” Journal of Visualized Experiments 196 (2023): e65475. [DOI] [PubMed] [Google Scholar]
  • 18. Jerome N. P., Papoutsaki M. V., Orton M. R., et al., “Development of a Temperature‐Controlled Phantom for Magnetic Resonance Quality Assurance of Diffusion, Dynamic, and Relaxometry Measurements,” MedPhys 43, no. 6 (2016): 2998–3007. [DOI] [PubMed] [Google Scholar]
  • 19. La Gioia A., Porter E. R., Merunka I., et al., “Open‐Ended Coaxial Probe Technique for Dielectric Measurement of Biological Tissues: Challenges and Common Practices,” Diagnostics 8, no. 2 (2018): 40. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Giannakopoulos I. I., Carluccio G., Keerthivasan M. B., et al., “MR Electrical Properties Mapping Using Vision Transformers and Canny Edge Detectors,” Magnetic Resonance in Medicine 93, no. 3 (2025): 1117–1131. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Magnetic Resonance Imaging are provided here courtesy of Wiley

RESOURCES