Skip to main content
Springer logoLink to Springer
. 2025 Jul 22;36(1):194–202. doi: 10.1007/s00330-025-11814-8

Multiphase aortic valve calcium scoring on true-non-contrast and calcium-preserving spectral reconstructions using dual-source photon-counting detector CT

Judith van der Bie 1,#, Mark M P van den Dorpel 2,#, Marcel van Straten 1, Daniel Bos 1,3, Alexander Hirsch 1,2, Nicolas M van Mieghem 2, Ricardo P J Budde 1,
PMCID: PMC12711920  PMID: 40696218

Abstract

Objectives:

This study investigated differences between aortic valve calcium (AVC) scores derived from true non-contrast (TNC) and virtual-non-contrast reconstructions acquired with photon-counting detector CT (PCD-CT) and the impact of ECG-phase variability on AVC scores.

Materials and methods

A hundred patients undergoing PCD-CT for transcatheter aortic valve implantation (TAVI) planning were retrospectively analyzed. Scores were computed using the Agatston methodology for TNC and virtual-non-iodine (VNI) reconstruction at scanner-selected optimal phase (best) and a fixed ECG-phase (300 ms). For VNI reconstructions, additional phases from 150 ms to 450 ms with 50 ms increments were reconstructed. AVC scores of TNCbest vs TNC300, VNIbest vs TNCbest, VNI300 vs TNC300, and all VNI phases vs VNIbest were compared using Wilcoxon signed-rank tests. The agreement was assessed using scatter plots, Bland–Altman plots, and intra-class coefficients. AVC scores were also categorized based on the likelihood of severe aortic stenosis. Differences between reconstructions were evaluated as percentages (reclassification) and analyzed using Cohen’s kappa coefficients.

Results

TNCbest and TNC300 differed significantly (mean bias: 226; LoA: [−820, 1300]; p < 0.001, reclassification 17%). VNIbest vs TNCbest resulted in a mean bias of −512 (LoA: [−1900, 860]; p < 0.001) and reclassification of 17%. TNC300 vs VNI300 demonstrated a bias of −200 and reclassification of 14% (κ = 0.72). VNI reconstructions showed less variability across phases than the difference between TNCbest and TNC300 (range, mean bias: 22–146).

Conclusion

VNI is a feasible alternative for AVC scoring but tends to overestimate compared to TNC. While phase-dependent variability in TNC underscores the need for standardization, further optimization of VNI is necessary for routine clinical use.

Key Points

Question What is the performance of AVC score calculation from virtual non-contrast images with PCD-CT and the impact of the reconstructed ECG-phase?

Findings VNI reconstructions tend to overestimate the scores compared to non-enhanced acquisitions. ECG phase significantly impacts AVC scores for non-enhanced acquisitions and VNI reconstructions.

Clinical relevance Utilizing VNI reconstructions to calculate AVC scores might reduce radiation dose, and understanding the influence of ECG-phase on these scores might improve reliability.

Graphical Abstract

graphic file with name 330_2025_11814_Figa_HTML.jpg

Keywords: Photon-counting detector CT, Computed tomography angiography, Aortic stenosis, Aortic valve calcium scores, Transcatheter aortic valve replacement

Introduction

Aortic valve stenosis (AS) is a prevalent and clinically significant cardiovascular condition characterized by progressive narrowing of the aortic valve orifice, most often due to calcium build-up [1]. Transcatheter aortic valve implantation (TAVI) has proven superior to optimal medical therapy and non-inferior to surgical aortic valve replacement across a range of patient categories with AS [24]. AS is diagnosed by echocardiography, but the computed tomography-derived aortic valve calcium (AVC) score (which quantifies the extent of calcium deposition on the aortic valve) provides additional information to stratify patients with inconclusive echocardiography results while also offering prognostic value [5]. The AVC score is determined on a non-contrast CT of the heart (‘true non-contrast’, TNC) using the Agatston method [6].

CT plays an essential role in the preprocedural planning of TAVI [7]. Photon-counting detector CT (PCD-CT) encompasses the next generation of CT scanners and allows for spectral reconstructions and increased spatial resolution compared to conventional CT [8]. Spectral CT systems enable reconstruction of virtual non-contrast-enhanced images from CT angiography (CTA) acquisitions without additional scanning. PCD-CT provides two types of virtual-non-contrast (VNC) images: conventional VNC images, primarily used for iodine subtraction in soft tissues, and calcium-preserving VNC images, which allow for the visualization and quantification of calcified structures such as the aortic valve or coronary arteries [911]. Before the introduction of PCD-CT systems, conventional techniques such as dual-source, kV-switching, and dual-layer CT were used to generate VNC reconstructions. However, PCD-CT offers enhanced spectral separation, which may improve the accuracy of iodine subtraction and the preservation of calcified structures. Additionally, when implemented on dual-source systems, PCD-CT maintains high temporal resolution, potentially leading to more precise and diagnostically reliable VNC reconstructions. Currently, four studies investigated the performance of VNI reconstructions computed with PCD-CT for AVC score calculations [5, 1214].

Although the Agatston method requires adherence to the end-diastolic phase (60–80% of the RR interval) for AVC score calculation, these four studies reported different reconstructed phases (Table S-1) [6, 15]. Since changes within the cardiac cycle impact both the aortic annulus geometry and the leaflet position, ECG phase variability may affect the AVC score [16, 17]. This has not previously been examined in the context of the aortic valve Agatston score, changes in the cardiac cycle have been demonstrated to affect the coronary Agatston score [18].

In this study, we examined differences between AVC scores derived from TNC and calcium preserving reconstructions and correlated the magnitude of these differences with the different ECG phases used.

Method

Patient selection

Consecutive adult patients (age > 18 years) who underwent a clinically indicated CT scan for TAVI planning on a dual-source PCD-CT system (NAEOTOM ALPHA, Siemens Healthineers) were retrospectively included from March to December 2023. Patients were excluded if they had undergone a previous surgical aortic valve or TAVI. Because of the observational and retrospective nature of the study, the institutional review board waived the need for ethics committee approval. General written consent was obtained from all included patients for data usage. The study adhered to the principles outlined in the Declaration of Helsinki, and it was not subject to the regulations of the Medical Research Involving Human Subjects Act, as determined by the institutional review board.

Image acquisition and reconstruction parameters

First, a prospectively ECG-triggered (padding: 310 milliseconds (ms) from the r-wave) TNC calcium scan was acquired at 120 kV, image quality level setting of 16, and 144 × 0.4 mm collimation. Images were reconstructed at a slice thickness of 3.0 mm with 1.5 mm increments using the Qr36 kernel at 70 keV without applying iterative reconstruction. [19]. Two phases of the cardiac cycle were reconstructed: a phase deemed best by the scanner itself (in TNCbest) and at a fixed time-point of 300 ms after the R-peak (TNC300).

Subsequently, a prospective ECG-triggered CT angiography (15–45% of the R–R interval) of the heart was performed, including the ascending aorta to the apex. A tube voltage of 120 kV, image quality setting of 34, and a 144 × 0.4 mm collimation were employed. The acquired scans were reconstructed with 0.4 mm slices, 0.2 mm increments at 55 keV using a Bv48 kernel and an iterative reconstruction strength of 4. The contrast protocol used was as follows: 45 mL contrast agent (Iodixanol 320), flow rate: 3.0 mL/s), was administered for the CTA of the heart. Followed by 25 mL (flow rate: 2.5 mL/s) contrast agent and a saline flush (25 mL, flow rate 2.5 mL/s) for the prospectively ECG-triggered high-pitch acquisition from the skull base to the groin.

From the CTA of the heart, calcium-preserving virtual non-contrast (VNC) images, referred to as virtual non-iodine (VNI) images, were reconstructed using the same parameters as the TNC reconstructions (70 keV, Qr36, iterative reconstruction off) [20]. VNI reconstructions utilize a manufacturer-specific algorithm (PureCalcium, Siemens) that virtually subtracts iodine-based contrast media while preserving the calcium signal. Unlike conventional VNC images, which are reconstructed from polychromatic datasets, VNI images are based on monoenergetic reconstructions and can be generated at various energy levels, improving material differentiation. In addition to identifying the best VNI reconstruction in the systolic phase (VNIbest), images were also reconstructed at absolute delay phases ranging from 150 ms to 450 ms in 50-ms increments.

AVC score computation

Each image series was loaded into a dedicated imaging segmentation software program (3Mensio Structural Heart, Pie Medical). First, the aortic annulus plane was obtained by marking the nadir of the 3 cusps, after which the software displayed the reference plane. Using a dedicated ‘Agatston’ module within the software, a 3D view of the aortic valve and aortic root was automatically reconstructed, displaying all areas containing calcium (≥ 130 Hounsfield Units) located between 20 mm above and 5.0 mm below the annulus. The selection of calcium areas was subsequently adjusted manually, if required (Fig. 1). All analyses were performed by a single observer with experience in calculating AVC scores (M.v.d.D., > 5 years of experience) and was blinded to the type of reconstruction/acquisition during the assessment.

Fig. 1.

Fig. 1

Aortic valve Agatston score computation in 3Mensio. A Example of aortic valve calcifications (‘hockey puck’ view). B Example of selected AVC in blue. All calcium with an HU ≥ 130 has been selected (‘hockey puck’ view). C Example of aortic valve calcifications (3-chamber view). D Example of selected AVC in blue. All calcium with an HU ≥ 130 has been selected (3-chamber view)

Data analysis

Categorical variables are expressed as frequencies and percentages, and continuous variables are expressed as median with 25th to 75th percentile. AVC scores were compared with a Wilcoxon signed-rank test following the following strategy. First, the AVC scores of TNCbest and TNC300. Next, AVC scores calculated from VNIbest and VNI300 were compared to TNCbest and TNC300, respectively. Lastly, AVC scores from all VNI reconstructed phases were compared to the scanner-selected best phase of VNI. Scatter plots and Bland–Altman plots supported by intra-class coefficients (ICC) were computed to assess the agreement of AVC scores as well. AVC scores of all reconstructions were divided into the following categories used to diagnose severe aortic stenosis: (a) highly likely ≥ 1600 for women ≥ 3000 for men, (b) likely ≥ 1200 for women and ≥ 2000 for men, and (c) unlikely < 800 for women and < 1600 for men [4]. Differences in category distributions were evaluated between TNCbest vs VNIbest, TNC300 vs VNI300, and TNCbest vs TNC300. These differences were assessed as percentages and analyzed using Cohen’s kappa coefficients to measure agreement. All statistical analyses were performed using SPSS statistical software (IBM Corp. Released 2016. IBM SPSS Statistics for Windows, Version 28.0.1.0) and Python (Python Software Foundation. Python Language Reference, version 3.9).

Results

Study population

A total of 100 patients were included (57 male, 43 female), which resulted in 1000 reconstructions. Baseline characteristics and CT dose details are displayed in Table 1.

Table 1.

Patient and scan characteristics (n = 100)

Age [years] 78 [76–83]
Male [%] 57%
Body mass index [kg/m2] 33.7 [27.3-40.0]
Heart rate [beats per min]
 True-non-contrast 72 [61–84]
 CT angiography 71 [62–81]
Volume CT dose index [mGy]
 True-non-contrast 2.51 [2.1–3.1]
 CT angiography 11.2 [9.0–14.7]
Dose length product [mGy cm]
 True-non-contrast 33.7 [27.3–40.0]
 CT angiography 164.5 [134.3–215.8]
Size-specific dose estimate [mGy]
 True-non-contrast 3.4 [3.1–3.9]
 CT angiography 14.8 [11.9–18.1]
True-non-contrast
 Phase “best” [ms] 322 [320–324]
Virtual-non-iodine
 Phase “best” [ms] 330 [304–351]

Data is presented as median [25th–75th percentile) or percentages

AVC scores

A median AVC score of 2364 [1633–3977] was observed for TNCbest compared to 2743 [1765–4286] for TNC300 (Table 2). Despite a time difference of only 22 ms, these scores were significantly different (Fig. 2: mean bias: 226, LoA: [−820:1300], p < 0.001). The direct comparison between TNCbest (2364 [1633–3977]) vs VNIbest (3018 [1815–4804]) also showed significantly different AVC scores (p < 0.001, Table 3). VNIbest overestimated AVC scores compared to TNCbest, with a mean bias of −512 (LoA: −1900; 860) (Fig. 2). AVC scores differed significantly between TNC300 (2743 [1765–4286]) and VNI300 (2946 [1884–4685]), with a mean bias of −200 and LoA of [−1200; 820].

Table 2.

Median AVC scores per reconstruction compared to TNCbest

Median calcium score Mean difference (%)
TNCbest 2364 [1633–3977] REF
TNC300 2743 [1765–4286] +10% (±21%)
VNIbest 3018 [1815–4804] +19% (±28%)
VNI150 3006 [1858–4661] +17% (±29%)
VNI200 2900 [1846–4561] +20% (±29%)
VNI250 2984 [1828–4478] +21% (±29%)
VNI300 2946 [1884–4685] +18% (±27%)
VNI350 2948 [1789–4444] +17% (±28%)
VNI400 2895 [1705–4422] +15% (±28%)
VNI450 2901 [1746–4400] +16% (±28%)

AVC aortic valve calcium, TNC true-non-contrast, VNI virtual-non-iodine

Fig. 2.

Fig. 2

Scatter and corresponding Bland–Altman plots. VNI reconstructions overestimated AVC scores computed from TNC reconstructions. By selecting a fixed phase (300 ms) for TNC and VNI, the mean bias was reduced by approximately 50%. Examining the variation between phases, the bias for TNCbest and TNC300 is roughly 65% higher than for VNIbest compared to VNI300

Table 3.

p-values of Wilcoxon signed rank tests (white) and ICCs (gray) per reconstruction

graphic file with name 330_2025_11814_Tab1_HTML.gif

ICC intra-class correlation coefficient, TNC true-non-contrast, VNI virtual-non-iodine

Variability of AVC scores

The variability of AVC scores was assessed by comparing scores computed from VNIbest (median t = 326 ms) with the obtained different phases of the VNI reconstructions as illustrated in Fig. 3. Mean biases ranged from 22 (VNI250ms) to 146 (VNI400ms). VNI150, VNI250ms, VNI300ms, and VNI350ms demonstrated similar calcium scores to VNIbest (p > 0.05).

Fig. 3.

Fig. 3

Blant-Altman plots illustrate comparisons between VNIbest and all other VNI reconstructions

Reclassification of AS severity

Between TNCbest to TNC300, 17 (17%, κ = 0.69) patients with a median difference of −201 [−378; 142] were assigned to a new class (Fig. 4). Again, 17 (17%, κ = 0.66) patients were reclassified when comparing AVC scores from TNCbest to VNIbest. Among the misclassified patients, the median difference in AVC scores for misclassified patients was −303 [−753; −75]. The lowest reclassification rate was obtained with TNC300 to VNI300, with 14 (14%, κ = 0.72) of the patients reclassified with a median difference of 281 [−790; −206]. For the various phases of VNI compared to VNIbest, the reclassification rate ranged from 8% at 350 ms to 18% at 150 ms. The lowest overall reclassification rate, at 3%, occurred between VNI350 and VNI450.

Fig. 4.

Fig. 4

Blue is unlikely severe AS (< 800 for women and < 1600 for men), yellow is unspecified (800–1200 for woman and 1600–2000 for men), light red is likely severe AS (≥ 1200 for women and ≥ 2000 for men) and dark red is highly likely severe AS (≥ 1600 for women ≥ 3000 for men). Both TNCbest to VNIbest/TNC300 showed a misclassification rate of 17% of the patients. For VNI reconstruction, patients were more often classified in a higher risk group. For all comparisons, the largest shift was seen in the “likely severe AS” group

Discussion

Our study compared AVC scores of TNC images to VNI images reconstructed from PCD-CT scans and AVC scores computed from multiple reconstructed phases of the heart cycle. The most accurate AVC scores were observed using a fixed phase of 300 ms for both TNC and VNI reconstructions. Compared to TNC, VNI consistently overestimated AVC scores, with a mean bias of −200 and a reclassification rate of 14% (κ = 0.72). When using scanner-selected optimal phases, VNI reconstructions overestimated (mean bias: −512) AVC scores obtained from TNC reconstructions. TNC reconstructions with a minimal time difference, also demonstrated a significant difference in AVC scores with an equal number of patients reclassified as TNC/VNI optimal phases. Indicating AVC scores substantially vary within scans from the same acquisition if the reconstructed phase is different. VNI reconstructions showed less variability of AVC scores over the reconstructed phases.

Four studies investigated the feasibility of AVC scoring with PCD-CT compared to TNC acquisitions (Table S-1). All studies employed standard Agatston reconstruction, similar to the approach in the present study. Feldle et al investigate VNI reconstructions acquired with prospective high-pitch and retrospective acquisitions [14]. AVC scores of VNI and TNC demonstrated high correlations, however, no exact scores of VNI were reported. Diagnostic accuracy was investigated only by categorizing patients in severe or non-severe AS, leading to 85% diagnostic accuracy in high pitch mode and 81% with retrospective gating.

Risch et al investigated AVC scores with standard Agatston reconstruction and reconstructions with thinner slice thicknesses and found an underestimation of AVC scores computed with VNI to TNC from −25% up to −32% [12]. The smallest difference was achieved with thin slices (0.4 mm) and an iterative reconstruction strength of four. We found a smaller difference of +19% but an overestimation. This difference might be explained by the use of high-pitch acquisitions and the difference in the reconstructed phase.

In addition to the standard Agatston reconstruction at 70 keV, Mergen et al generated multiple virtual monoenergetic images at 60 keV, 70 keV, 80 keV, and 90 keV, and utilized iterative reconstruction with strengths of two, three, and four to reconstruct VNI images computed from cardiac late enhancement scans. Mean difference biases ranged from 258 to 267 at 70 keV, which were lower than our study; however, they also indicated a significant overestimation of AVC scores. Mergen et al reconstructed the scans at 280 ms relative to the R-wave, which could account for the discrepancies observed in comparison to our study.

Additionally, Mergen et al explained that residual iodine might be in the center of macro calcifications even after using the VNI algorithm to subtract iodine from the image. This phenomenon may also have contributed to the overestimation of AVC by VNI compared to TNC in our study. Some voxels containing iodine were mistakenly identified as true calcium voxels due to the simple counting method used in the Agatston score. By increasing the keV levels to 80 keV, results similar to those obtained by TNC acquisition were achieved, with mean bias differences ranging from 3 to 11. Switching to VNI images at 80 keV might alter the contribution of pixels to the score, but rather slightly reduce the total score. This reduction is due to how the Agatston score is calculated, which places more weight on higher attenuation values. Utilizing higher KeV reconstructions does not improve the performance of the VNI algorithm, it only adjusts the score by a different weighting. Because Agatston scoring is validated at 120 kV (corresponding to ~70 keV), and algorithms are optimized for this energy range, we recommend reconstructing VNI images at 70 keV [21, 22]. However, even in the absence of visible iodine, slight overestimation was observed compared to non-enhanced scans. This indicates that inter-scan and intra-scan variability, arising from factors such as patient positioning and cardiac motion, may also play a role [23].

Sartoretti et al investigated the influence of temporal resolution on AVC scores and found that lower temporal resolution leads to the overestimation of AVC scores caused by motion artifacts and blurring of calcifications [13]. In our study, we utilized a high temporal resolution of 66 ms, which helped minimize motion artefacts and blurring that may lead to overestimation of the AVC score. Nonetheless, we observed noticeable variability in TNC-derived AVC scores across different cardiac phases, even when reconstructed at the same energy level and with identical parameters. This suggests that AVC quantification is inherently sensitive to the timing of image acquisition within the cardiac cycle, and that variability can persist despite high temporal resolution [23].

There are several limitations to this study. First, it is retrospective and conducted at a single center, involving a restricted number of patients. Thus, the findings may not fully represent the broader population. Secondly, to limit our data, we selected one additional phase for 300 ms for the non-enhanced scans. As a result, we could only perform a single comparison of variability. This may introduce bias to our study, as we reconstructed multiple phases for VNI. However, we observed a significant difference between the two phases of TNC, emphasizing the importance of standardization, which is the central message of our manuscript. Third, this study only investigated the Agatston reconstruction method with a 3.0 mm slice thickness. Risch et al showed improved results with thin slices (0.4 mm) and a high iterative reconstruction strength of four due to the sharp delineation of calcifications. Employing reconstructions with higher spatial resolution might aid the performance of VNI reconstructions in computing AVC scores. Fourth, like Mergen et al, relatively high heart rates (median: 71 beats per min TNC, median: 72 beats per min VNI) were included in this study. Lower heart rates may lead to more accurate calcium quantification on both TNC and VNI reconstructions. However, Beta-blockers cannot be administered preventively in this patient population. Finally, no outcome measures were available in the current study to examine the prognostic value of both VNI- and TNC-derived AVC scores. Follow-up studies may aim to examine potential differences in the prognostic value of the different types of reconstructions.

In conclusion, VNI reconstructions are feasible for computing AVC scores, but they demonstrated an overestimation compared to AVC scores computed from TNC reconstructions. VNI reconstructions showed less variation between reconstructed phases than TNC reconstructions. However, AVC-scoring is not very robust since only a small change in ECG-phase results in a significant change in score. Fixed phases demonstrated the smallest difference (mean bias = −200) between the two reconstructions and the lowest reclassification of 14%, which may be related to less anatomic variation due to less motion difference between the scans. Therefore, standardization is needed to determine at which phase TNC or VNI reconstruction should be performed to enable comparison of AVC scores. Further improvements are required to implement the VNI reconstruction algorithm in daily clinical practice.

ELECTRONIC SUPPLEMENTARY MATERIAL

Abbreviations

AS

Aortic valve stenosis

AVC

Aorta valve calcium

CTA

Computed tomography angiography

PCD-CT

Photon-counting detector CT

TAVI

Transcatheter aortic valve implantation

TNC

True non-contrast

VNC

Virtual-non-contrast

VNI

Virtual-non-iodine

Funding

The authors state that this work has not received any funding.

Compliance with ethical standards

Guarantor

The scientific guarantor of this publication is R.P.J. Budde.

Conflict of interest

The authors of this manuscript declare relationships with the following companies: Institutional support to Erasmus MC by Siemens Healthineers. R.P.J. Budde: Speaker's bureau, Siemens and Bayer, and advisory board, Bayer, payment to Erasmus MC.

Statistics and biometry

D. Bos kindly provided statistical advice for this manuscript.

Informed consent

Written informed consent was obtained from all subjects (patients) in this study.

Ethical approval

Institutional Review Board approval was obtained.

Study subjects or cohorts overlap

Not applicable.

Methodology

  • Retrospective

  • Diagnostic accuracy, performance

  • Performed at one institution

Footnotes

Publisher’s Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Judith van der Bie and Mark M. P. van den Dorpel contributed equally to this work.

Supplementary information

The online version contains supplementary material available at 10.1007/s00330-025-11814-8.

References

  • 1.Kanwar A, Thaden JJ, Nkomo VT (2018) Management of patients with aortic valve stenosis. Mayo Clin Proc 93:488–508 [DOI] [PubMed] [Google Scholar]
  • 2.Leon MB, Smith CR, Mack M et al (2010) Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. N Engl J Med 363:1597–1607 [DOI] [PubMed] [Google Scholar]
  • 3.Smith CR, Leon MB, Mack MJ et al (2011) Transcatheter versus surgical aortic-valve replacement in high-risk patients. N Engl J Med 364:2187–2198 [DOI] [PubMed] [Google Scholar]
  • 4.Baumgartner H, Falk V, Bax JJ et al (2018) 2017 ESC/EACTS guidelines for the management of valvular heart disease. Rev Esp Cardiol 71:110 [DOI] [PubMed] [Google Scholar]
  • 5.Mergen V, Ghouse S, Sartoretti T et al (2023) Cardiac virtual noncontrast images for calcium quantification with photon-counting detector CT. Radiol Cardiothorac Imaging 5:e220307 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Pawade T, Sheth T, Guzzetti E, Dweck MR, Clavel MA (2019) Why and how to measure aortic valve calcification in patients with aortic stenosis. JACC Cardiovasc Imaging 12:1835–1848 [DOI] [PubMed] [Google Scholar]
  • 7.van der Bie J, Sharma SP, van Straten M et al (2023) Photon-counting detector CT in patients pre- and post-transcatheter aortic valve replacement. Radiol Cardiothorac Imaging 5:e220318 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.van der Bie J, van Straten M, Booij R et al (2023) Photon-counting CT: review of initial clinical results. Eur J Radiol 163:110829 [DOI] [PubMed] [Google Scholar]
  • 9.Fink N, Zsarnoczay E, Schoepf UJ et al (2023) Photon counting detector CT-based virtual noniodine reconstruction algorithm for in vitro and in vivo coronary artery calcium scoring: impact of virtual monoenergetic and quantum iterative reconstructions. Invest Radiol 58:673–680 [DOI] [PubMed] [Google Scholar]
  • 10.Sharma SP, van der Bie J, van Straten M et al (2023) Coronary calcium scoring on virtual non-contrast and virtual non-iodine reconstructions compared to true non-contrast images using photon-counting computed tomography. Eur Radiol 34:3699–3707. 10.1007/s00330-023-10402-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Emrich T, Aquino G, Schoepf UJ et al (2022) Coronary computed tomography angiography-based calcium scoring: in vitro and in vivo validation of a novel virtual noniodine reconstruction algorithm on a clinical, first-generation dual-source photon counting-detector system. Invest Radiol 57:536–543 [DOI] [PubMed] [Google Scholar]
  • 12.Risch F, Harmel E, Rippel K et al (2024) Virtual non-contrast series of photon-counting detector computed tomography angiography for aortic valve calcium scoring. Int J Cardiovasc Imaging 40:723–732. 10.1007/s10554-023-03040-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Sartoretti T, Mergen V, Dzaferi A et al (2024) Effect of temporal resolution on calcium scoring: insights from photon-counting detector CT. Int J Cardiovasc Imaging 41:615–625. 10.1007/s10554-024-03070-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Feldle P, Scheuber M, Grunz JP et al (2024) Virtual non-iodine photon-counting CT-angiography for aortic valve calcification scoring. Sci Rep 14:4724 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Blaha MJ, Mortensen MB, Kianoush S, Tota-Maharaj R, Cainzos-Achirica M (2017) Coronary artery calcium scoring: Is it time for a change in methodology?. JACC Cardiovasc Imaging 10:923–937 [DOI] [PubMed] [Google Scholar]
  • 16.Petersen J, Voigtlander L, Schofer N et al (2018) Geometric changes in the aortic valve annulus during the cardiac cycle: impact on aortic valve repair. Eur J Cardiothorac Surg 54:441–445 [DOI] [PubMed] [Google Scholar]
  • 17.Alkadhi H, Leschka S, Trindade PT et al (2010) Cardiac CT for the differentiation of bicuspid and tricuspid aortic valves: comparison with echocardiography and surgery. AJR Am J Roentgenol 195:900–908 [DOI] [PubMed] [Google Scholar]
  • 18.Lee HKJG, Park JK, Han YM, Lee YS, Kwon KS, Kim SS (2012) Variation of the degree of coronary artery stenosis during the cardiac cycle: influence of heart rate or calcium score on coronary CT angiography. J Korean Soc Radiol 66:221–228 [Google Scholar]
  • 19.Agatston AS, Janowitz WR, Hildner FJ, Zusmer NR, Viamonte JrM, Detrano R (1990) Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol 15:827–832 [DOI] [PubMed] [Google Scholar]
  • 20.Flohr T, Schmidt B (2023) Technical basics and clinical benefits of photon-counting CT. Invest Radiol 58:441–450 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Eberhard M, Mergen V, Higashigaito K et al (2021) Coronary calcium scoring with first generation dual-source photon-counting CT-first evidence from phantom and in-vivo scans. Diagnostics (Basel) 11:1708 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.van der Werf NR, Booij R, Greuter MJW et al (2022) Reproducibility of coronary artery calcium quantification on dual-source CT and dual-source photon-counting CT: a dynamic phantom study. Int J Cardiovasc Imaging 38:1613–1619 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.McCollough CH, Rajendran K, Leng S (2023) Standardization and quantitative imaging with photon-counting detector CT. Invest Radiol 58:451–458 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials


Articles from European Radiology are provided here courtesy of Springer

RESOURCES