Skip to main content
British Heart Journal logoLink to British Heart Journal
. 1988 Jan;59(1):12–19. doi: 10.1136/hrt.59.1.12

Digital subtraction contrast echocardiography: a new method for the evaluation of regional myocardial perfusion.

M J Monaghan 1, P J Quigley 1, J M Metcalfe 1, S D Thomas 1, D E Jewitt 1
PMCID: PMC1277066  PMID: 3277654

Abstract

The potential of contrast enhanced digital subtraction echocardiography to demonstrate and quantitate myocardial perfusion was evaluated in 36 patients undergoing routine coronary arteriography or angioplasty. In 24 technically successful studies, multiple cross sectional echocardiographic images, obtained before and after intracoronary (sonicated contrast) injection, were stored by high speed, real time data transfer to an on line minicomputer. Subsequent digital subtraction processing of the stored image data provided composite images in which the distribution of myocardial perfusion was easily seen. Quantitative analysis of peak enhanced myocardial grey level and washout half time successfully differentiated between myocardial segments in which angiography had suggested normal, reduced, and grossly impaired or absent perfusion. The results suggest that this new method of digital image capture and quantitative processing has substantial advantages over previous off line qualitative techniques. It is likely to be of considerable value for routine coronary arteriography, angioplasty, and coronary thrombolysis.

Full text

PDF
12

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Feinstein S. B., Shah P. M., Bing R. J., Meerbaum S., Corday E., Chang B. L., Santillan G., Fujibayashi Y. Microbubble dynamics visualized in the intact capillary circulation. J Am Coll Cardiol. 1984 Sep;4(3):595–600. doi: 10.1016/s0735-1097(84)80107-2. [DOI] [PubMed] [Google Scholar]
  2. Feinstein S. B., Ten Cate F. J., Zwehl W., Ong K., Maurer G., Tei C., Shah P. M., Meerbaum S., Corday E. Two-dimensional contrast echocardiography. I. In vitro development and quantitative analysis of echo contrast agents. J Am Coll Cardiol. 1984 Jan;3(1):14–20. doi: 10.1016/s0735-1097(84)80424-6. [DOI] [PubMed] [Google Scholar]
  3. Gillam L. D., Kaul S., Fallon J. T., Levine R. A., Hedley-Whyte E. T., Guerrero J. L., Weyman A. E. Functional and pathologic effects of multiple echocardiographic contrast injections on the myocardium, brain and kidney. J Am Coll Cardiol. 1985 Sep;6(3):687–694. doi: 10.1016/s0735-1097(85)80132-7. [DOI] [PubMed] [Google Scholar]
  4. Goldman M. E., Mindich B. P. Intraoperative cardioplegic contrast echocardiography for assessing myocardial perfusion during open heart surgery. J Am Coll Cardiol. 1984 Nov;4(5):1029–1034. doi: 10.1016/s0735-1097(84)80067-4. [DOI] [PubMed] [Google Scholar]
  5. Ong K., Maurer G., Feinstein S., Zwehl W., Meerbaum S., Corday E. Computer methods for myocardial contrast two-dimensional echocardiography. J Am Coll Cardiol. 1984 May;3(5):1212–1218. doi: 10.1016/s0735-1097(84)80179-5. [DOI] [PubMed] [Google Scholar]
  6. Sakamaki T., Tei C., Meerbaum S., Shimoura K., Kondo S., Fishbein M. C., Y-Rit J., Shah P. M., Corday E. Verification of myocardial contrast two-dimensional echocardiographic assessment of perfusion defects in ischemic myocardium. J Am Coll Cardiol. 1984 Jan;3(1):34–38. doi: 10.1016/s0735-1097(84)80427-1. [DOI] [PubMed] [Google Scholar]
  7. Ten Cate F. J., Drury J. K., Meerbaum S., Noordsy J., Feinstein S., Shah P. M., Corday E. Myocardial contrast two-dimensional echocardiography: experimental examination at different coronary flow levels. J Am Coll Cardiol. 1984 May;3(5):1219–1226. doi: 10.1016/s0735-1097(84)80180-1. [DOI] [PubMed] [Google Scholar]

Articles from British Heart Journal are provided here courtesy of BMJ Publishing Group

RESOURCES