The effects of measurement errors on minimum ablative margins after thermal ablation of liver tumors: a simulation study

Iwan Paolucci 1,, Jessica Albuquerque 1, Noreen S Siddiqi 1, A Kyle Jones 2, Kristy K Brock 2, Bruno C Odisio 1
PMCID: PMC12804325  PMID: 41551239

Abstract

Purpose

To develop a mathematical framework to estimate the in silico A0 threshold based on the technical specifications of a specific ablation confirmation software package for thermal ablation of liver tumors that can then be used to identify the impact of different sources of error.

Methods

To estimate in silico A0 thresholds, we developed a simulation framework incorporating technical parameters and biological effects. Technical parameters were segmentation error, registration error, and slice thickness, and biological effects were tissue shrinkage and microscopic satellite lesions; these parameters and effects were all modeled using statistical distributions. For each permutation of parameters, a logistic regression was fitted to determine the observed MAM required to achieve ≥ 99% probability of true complete tumor coverage (i.e., the A0 threshold). The mathematical framework was integrated into a web application to estimate the A0 threshold and the reliability of the commonly used 5-mm A0 threshold based on several software performance characteristics.

Results

A total of 15,000,000 simulations (10,000 simulations × 1500 parameter permutations) were run and summarized. Tumor and ablation zone segmentation most greatly influenced the A0 threshold, with thresholds of 3.4 and 8.4 mm for 1- and 5-mm errors, whereas slice thickness had a relatively small effect, with A0 thresholds of 2.9 and 3.4 mm for thicknesses of 1 and 5 mm, respectively.

Conclusion

This framework provides a method to determine software-specific in silico A0 thresholds and evaluate the reliability of existing 5-mm criteria based on software performance metrics. The results further show that ablation confirmation software should have registration and segmentation errors of ≤ 3 mm to reliably use a 5-mm A0 threshold.

Keywords: Liver malignancies, Thermal ablation, Minimum ablative margin, Simulation

Introduction

Thermal ablation is a minimally invasive local curative-intent treatment option for patients with small (≤ 3–5 cm) primary and secondary liver tumors [1]. It is part of national guidelines for hepatocellular carcinomas, intrahepatic cholangiocarcinoma, and colorectal liver metastases [26] and also frequently used in other metastatic settings. Over the past decade, image guidance for thermal ablation has drastically improved with the introduction of stereotactic guidance, robotics, and software-based ablation confirmation. Software-based ablation confirmation is used to ensure that the minimum ablative margin (MAM), which has been repeatedly shown to be the most important predictor of local tumor progression [710], is achieved during the procedure.

Similar to R0 resection, A0 ablation has been defined as ablation with complete tumor coverage that includes a pre-specified margin (e.g., MAM ≥ 5 mm) and a low risk of local tumor progression [11]. However, the true MAM cannot be measured microscopically or histologically because the tumor is not surgically removed; instead, it is destroyed by the ablation in situ and only necrosis remains within the liver. Thus, MAM is an imaging-based surrogate of histopathologic margin measured by co-registering the pre-ablation and post-ablation images. A recent systematic review reinforced 5 mm as a minimum requirement but stated 10 mm as optimal outcomes [12]. The study also identified high heterogeneity among studies which could be due to significant differences in measurement methodologies and their accuracies. For quantitative and objective measurements, segmentation methods are used to contour the tumor and ablation zone and compute the MAM in all 3 dimensions. Because this image processing is time consuming, automated algorithms based on convolutional neural networks have been developed and integrated into dedicated MAM confirmation software (also called ablation confirmation [AC] software) packages [13, 14].

Many AC software platforms lack robust clinical evidence establishing a clear and reproducible A0 margin threshold within the measurement accuracy of their respective systems. Specific thresholds for each AC software are often determined in retrospective series. Previously published and ongoing trials often use a MAM of 5 or 10 mm as sufficient (A0) margins [1518]. However, newer studies using more advanced and accurate imaging processing techniques have identified lower thresholds [7]—indicating the influence of measurement accuracy on the A0 threshold. Conducting a clinical study to first find and then validate A0 in each software package is impractical, as these packages evolve and provide newer and more accurate image processing methods regularly. Thus, in silico methods for initial technical validation would be a valuable tool.

In this study, we present a mathematical framework to estimate the in silico A0 threshold based on the technical specifications of the specific AC software package being used. The primary aim was to estimate the effect of registration and segmentation inaccuracies, as well as the slice thickness, on the threshold of A0 margins to further guide efforts to improve these technologies. The secondary aim was to use those estimates, to identify the minimum accuracy requirements needed to reliably use a 5-mm MAM threshold.

Materials and methods

A recent review article identified 5 key sources of inaccuracy affecting A0 thresholds: image resolution, segmentation error, registration error, deformation, and image artifacts [19]. The specific parameters used in the simulation are presented in Table 1. For this study, the effects of image artifacts were considered part of segmentation error, and deformation was considered part of tissue shrinkage and registration error and not explicitly modeled. Further, for image resolution, only the slice thickness was included in this study since the in-plane resolution is typically < 1 mm.

Table 1.

Parameter sets used in this study

Parameter Notation Values
Tissue shrinkage, % φShrinkage 0, 10, 20, 30
Microscopic satellite lesions, % pSatellite 0, 25, 50, 75, 100
Registration error, mm εRegistration 1, 2, 3, 4, 5
Segmentation error, mm εSegmentation 1, 2, 3, 4, 5
Slice thickness, mm φSlice thickness 1, 2, 3, 4, 5
Tumor size, mm dTumor size 10 – 30
Minimum ablative margin, mm MAMtrue  −5 – 10

Biological effects

We examined 2 major biological effects on MAM measurement: (i) tissue shrinkage due to the application of microwave ablation energy and (ii) the presence of microscopic satellite lesions not visible on imaging. Both effects influence the accuracy of the depiction of the actual tumor and ablation zone (Fig. 1).

Fig. 1.

Fig. 1

The effects of visibility of microscopic tumor on cross-sectional imaging (A), microscopic satellite lesions (B), and tissue shrinkage (C) on the required margins

Tissue shrinkage

Tissue shrinkage during microwave ablation has been studied mainly in ex vivo experiments, as in vivo studies involving patients undergoing microwave ablation are unfeasible. Radial tissue contraction has been reported to be up to 40% in ex vivo studies in bovine tissue [20]. This finding likely represents an upper limit because ex vivo bovine tissue is non-perfused and therefore dehydrates faster than perfused tissue. Tissue contraction within the ablation zone can cause the MAM to be overestimated. In addition, tumor tissue contraction cannot be observed during the procedure, and thus, the exact tumor-specific shrinkage is unknown. In our model, we assumed that the tissue shrinkage is normally distributed.

eShrinkageNormalx,0.05

Microscopic satellite lesions

The accuracy of the measurement relies on the accurate depiction of microscopic disease on macroscopic imaging, which has an average spatial resolution of approximately 1 mm. Therefore, microscopic satellite lesions adjacent to the ablated tumor may be present but not visible on cross-sectional imaging. In addition, the transition zone between microscopic tumor and healthy tissue might not be accurately depicted on imaging, and therefore, the tumor size might be underestimated, resulting in an overestimated MAM. In our model, we assumed that microscopic satellite lesions are directly adjacent, the sizes are uniformly distributed between 0.5 and 2.5 mm, and the probability of the presence of microscopic satellite lesions follows a binomial distribution.

pSatelliteBinomial1,p
dSatellite=0,pSatellite=0Uniform(0.5,2.5),pSatellite=1

Technical parameters

We examined the following sources of technical error: image registration, image segmentation variation, and slice thickness (Fig 2). These errors directly affect the accuracy of the MAM.

Fig. 2.

Fig. 2

Intermediate results from an example case showing the true margin (A), application of the registration error (B), and application of random noise to the segmentation masks (C). The tumor is shown in white, the tumor contour in green, the true ablation zone in red, and the ablation zone influenced by measurement error in yellow

Registration error

Registration error significantly affects the accuracy of MAM measurements and is inherent to each specific registration method. In addition, it is often not possible to improve the registration accuracy during an ablation procedure with a given set of images. In our model, the registration error consists of a magnitude and a direction. The magnitude is normally distributed with a mean of 0 with varying standard deviation. The direction is chosen randomly in either X, Y, or Z direction, with each axis having the same probability of being chosen.

emagNormal0,eRegistration
edirX,Y,Z
eShift=edir×emag

The shift (eshift) is then applied to the ablation center, whereas the tumor center remains constant.

Segmentation error

Segmentation of tumors and ablation zones varies between observers and can also be influenced by the available segmentation tools and artificial intelligence–based automated algorithms. In contrast to registration error, segmentation error can be minimized by having an experienced radiologist review artificial intelligence–based segmentation results and make necessary adjustments using manual tools such as brushes. However, for intra-procedural MAM assessment, the time to make such adjustments is limited, increasing the risk for segmentation errors. Segmentation errors are typically calculated using the Dice similarity coefficient, measuring the overlap of the contours, or the Hausdorff distance, measuring the distance between the contours [21, 22]. For our model, the segmentation error was estimated to be randomly distributed and added as random noise to the segmentation mask. In addition, the segmentation errors for the tumor and the ablation zone were assumed to be independent.

eTumorNormal0,1
eAblationNormal(0,1)

Slice thickness

Recent studies on AC software have used slice thicknesses between 1 and 5 mm [2325]. The slice thickness interacts with the segmentation and registration process because thin slices typically result in a lower signal-to-noise ratio, reducing tumor depiction. Slice thickness also limits the precision of the measurement in the cranio-caudal direction. In addition, thinner slices result in more slices and thus a higher computational cost and more manual work. In our model, the interactions of slice thickness with other errors and computing time were ignored. To analyze the effect of slice thickness, the tumor and ablation segmentation mask were reconstructed with an in-plane resolution of 1 mm but varying resolution in cranio-caudal direction denoted as φSlice thickness.

Simulations

Approach and parameters

For simulations, we considered a set of parameters for each influencing factor (Table 1). The data were processed as follows in each simulation:

  1. Randomly sample tumor size and ablative margin

  2. Create 2 images with a tumor and ablation zone based on the size drawn in step 1

  3. Apply tissue shrinkage and microscopic satellite lesions to tumor

  4. Measure true MAM (MAMTrue)

  5. Apply random registration error to ablation zone

  6. Apply random noise to the tumor and ablation segmentation masks

  7. Resample both images with given slice thickness

  8. Measure observed MAM (MAMObserved)

For each permutation of parameters (N = 1500), 10,000 simulations were performed, and the results were recorded. The simulations were implemented in Python (version 3.12). Synthetic image generation and processing were implemented using NiBabel (version 5.2.0) and NumPy (version 1.24.4). The “qam” package was used to quantify the minimum ablative margins [26].

Estimating A0 and probability of complete coverage

We defined the A0 threshold as the observed MAM where the true MAM is > 0 (complete microscopic tumor coverage) at least 99% of the time. Mathematically, this is denoted as:

PMAMtrue>0MAMobserved>xA0)99%

To calculate the A0 threshold, we fitted a logistic regression model to estimate the sigmoid function on a continuous scale of the observed MAM from − 5 mm to 10 mm. We then located the observed MAM where the probability of true MAM is greater than 99% and defined it as A0. In addition, we estimated the probability of complete coverage (> 0-mm MAMtrue) if the commonly used 5-mm MAM threshold is observed as follows:

PMAMtrue>0MAMobserved5)

Web application for A0 and 5-mm MAM simulation

The simulations and A0 definition were integrated into a web application to estimate the A0 threshold for a given software and to estimate the probability of achieving a > 0-mm true MAM if a > 5-mm MAM threshold was observed. The web application was implemented in Python and Streamlit (version 1.29.0) and is available at https://acs-accuracy.streamlit.app/. After entering the technical parameters of the software and anticipated biological effects, the simulations were performed, and the thresholds were estimated (Fig 3).

Fig. 3.

Fig. 3

Web application for simulating the influence of measurement errors on A0 for a given software. MAM, minimum ablative margin

Statistical analysis

The A0 cutoff for each permutation of the simulation parameters was computed on a high-performance computing cluster. The sigmoid functions were plotted using ggpubr and ggplot2 [27, 28]. A contour plot was further used to plot the A0 threshold for segmentation and registration errors between 1 and 5 mm.

The analysis was performed in R (version 4.4.0) and RStudio (version 2024.04.0). The complete code for the statistical analysis is available in the Github code repository (https://github.com/ipa/acs-accuracy).

Results

A total of 10,000 simulations for each of the 1500 parameter permutations were performed (N = 15,000,000 total simulations).

Registration error

Higher registration errors led to higher required A0 thresholds, as the A0 threshold needs to incorporate the margin of error of the registration process itself. For registration errors of 1, 3, and 5 mm, the A0 thresholds were 3.4, 4.9, and 7.0 mm, respectively (Fig. 4a). Most studies trying to identify optimal MAM thresholds for A0 exclude cases with inaccurate registrations, commonly using a 3-mm threshold measured by visual inspection at anatomical landmarks [8, 9]. Figure 4b shows the effects of excluding inaccurate registrations for registration errors of 1, 3, and 5 mm and demonstrates that the A0 threshold found with this practice was lower than when all registration measurements were used. Overall, the difference in A0 thresholds increased with increasing registration errors.

Fig. 4.

Fig. 4

Comparison of how different measurement inaccuracies influence the definition of A0: registration error (a), registration error (when excluding registrations with > 3 mm of error) (b), segmentation error (c), slice thickness (d), tissue shrinkage (e), and microscopic satellite lesions (f). MAM, minimum ablative margin

Segmentation error

A greater segmentation error led to a higher A0 threshold, and the segmentation error more greatly influenced the A0 threshold than the registration error. For segmentation errors of 1, 3, and 5 mm, the A0 thresholds were 3.4, 5.2, and 8.4 mm, respectively (Fig. 4c).

Slice thickness

With larger slice thickness, the A0 threshold marginally decreased (≤ 0.5 mm, which is below the resolution of the simulation and could be attributed to noise in the measurement or simulation). For slice thicknesses of 1, 3, and 5 mm, the A0 thresholds were 3.4, 3.2, and 2.9 mm, respectively (Fig. 4d).

Tissue shrinkage

Tissue shrinkage during microwave ablation reduced the A0 threshold. The A0 thresholds for 0%, 10%, 20%, and 30% tissue shrinkage were 3.4, 2.8, 2.2, and 1.8 mm, respectively (Fig. 4e).

Microscopic satellite lesions

The presence of microscopic satellite lesions, which were undetectable on radiologic imaging, increased the A0 threshold, even if not all tumors presented with microscopic satellite lesions. The A0 thresholds for 0%, 25%, 50%, 75%, and 100% probability of the presence of satellite lesions were 3.4, 5.8, 7.4, 7.9, and 7.7 mm, respectively (Fig. 4f).

Validity of 5-mm A0 threshold

Figure 5 shows the A0 threshold depending on segmentation and registration errors—the 2 most important sources of error. Segmentation and registration errors between 1 and 3 mm resulted in A0 thresholds of ≤ 5 mm. Errors larger than 3 mm resulted in larger A0 thresholds, and thus the commonly used 5-mm threshold would be unreliable in those cases.

Fig. 5.

Fig. 5

Contour plot of the A0 threshold depending on registration and segmentation errors

Discussion

In this study, we developed a simulation framework to determine how measurement inaccuracies impact A0 thresholds based on cross-sectional imaging. We integrated this simulation into an interactive web application designed to help clinicians estimate the appropriate A0 threshold—and assess the validity of a 5-mm MAM threshold—for their specific AC software. We found that tumor and ablation zone segmentation and registration accuracy greatly affect the definition of A0, whereas slice thickness has a negligible effect. Thus, significant effort should be invested in those related algorithms, especially in registration methods, as errors in these parameters can often not be corrected during the procedure. While the results suggest thicker slices lead to better results, this is likely noise from the simulations as the difference is below the image resolution. Based on our simulations, to achieve an A0 threshold of ≤ 5 mm, an AC system must have a segmentation error of ≤ 3 mm and a registration error of ≤ 3 mm. However, if both errors were exactly 3 mm, a 5-mm A0 threshold would be unreliable. Further, because the exclusion of inaccurate registrations affects the A0 definition, the number of excluded cases needs to be disclosed in scientific studies.

The current literature on the factors influencing A0 thresholds is limited, as each study uses one specific software implementation to process all images. Thus, comparing different software implementations with sufficient sample sizes (N > 1000) to capture the large variation in tumor sizes, margins, and measurement errors is impractical. Furthermore, biological effects like microscopic satellite lesions are undepictable on clinically available cross-sectional imaging and therefore only known probabilistically from histological studies. Therefore, for rare tumor types, which are more prone to have microscopic satellite lesions (e.g., intrahepatic cholangiocarcinoma), it is impossible to collect the required sample size to study the effects of microscopic satellite lesions on A0 thresholds. The same challenge occurs with tissue shrinkage, which is known only from ex vivo studies and cannot be individually measured. The results of our study showed that these biological effects have a large influence on the definition of A0. Thus, in clinical practice, assumptions about these effects should be carefully considered when deciding on MAM requirements.

This study has limitations. First, as a simulation study, certain influences might not have been modeled. For example, the real tissue shrinkage during thermal ablation is still unknown, but in this study, we assumed it to be uniform and of specific magnitudes. Tumors and ablation zones were also modeled as spheres and ellipsoids which represent a simplification of these irregularly shaped structures. Second, we ignored the interactions between the sources of error, which vary depending on the algorithm. For example, a biomechanical deformable registration algorithm would be less accurate if the underlying segmentation algorithms is inaccurate. Similarly, an intensity-based registration might be affected by lower signal-to-noise ratios owing to smaller slice thickness. Differences in slice thickness and other image reconstruction parameters (e.g., mAs and kV), which were not modeled, could also influence segmentation and registration accuracies. Third, segmentation and registration errors were modeled as following a normal distribution with a mean of 0 but varying standard deviation, assuming no systematic errors. Although this might be the case for certain software packages, we assumed that such systematic errors would be corrected during the development of those algorithms, which may not be true for all software packages. We also did not incorporate differences in registration types such rigid vs. deformable as these are heavily implementation dependent.

Conclusions

In conclusion, this study showed how measurement errors and biological effects influence A0 cutoffs. Our simulation represents an important tool for estimating the required MAM depending on technical errors of the measurement method and biological effects. This in silico model could be used a priori to verify technical performance requirements. In addition, we provided a benchmark for the minimum accuracy necessary for AC software.

Acknowledgements

The authors acknowledge the High Performance Computing for Research facility at The University of Texas MD Anderson Cancer Center for providing computational resources that have contributed to this research. We thank Madison Semro, Associate Scientific Editor, and Dawn Chalaire, Associate Director of Editing Services, in the Research Medical Library at The University of Texas MD Anderson Cancer Center, for editing this article.

Authors’ contributions

Conceptualization: IP, BCO; methodology: IP; formal analysis: IP; resources: IP, KKB; visualization: IP; writing—original draft: IP; funding acquisition: KKB, BCO; writing—review and editing: IP, JA, NSS, KJ, KKB, BCO.

Funding

NSS was supported by an Image Guided Cancer Therapy (IGCT) T32 Training Program Fellowship (T32CA261856).

Data availability

Code and data are available on GitHub: https://github.com/ipa/acs-accuracy.

Declarations

Ethics approval and consent to participate

This is a simulation study. Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Crocetti L, de Baere T, Pereira PL, Tarantino FP. CIRSE standards of practice on thermal ablation of liver tumours. Cardiovasc Intervent Radiol. 2020;43(7):951–62. [DOI] [PubMed] [Google Scholar]
  • 2.National Comprehensive Cancer Network. NCCN Clinical Practice Guidelines in Oncology (NCCN Guidelines®): Hepatocellular Carcinoma (Version 1.2025). 2025.
  • 3.Cervantes A, Adam R, Rosello S, et al. Metastatic colorectal cancer: ESMO clinical practice guideline for diagnosis, treatment and follow-up. Ann Oncol. 2023;34(1):10–32. [DOI] [PubMed] [Google Scholar]
  • 4.Vogel A, Martinelli E, ESMO Guidelines Committee. Updated treatment recommendations for hepatocellular carcinoma (HCC) from the ESMO clinical practice guidelines. Ann Oncol. 2021;32(6):801–5. [DOI] [PubMed] [Google Scholar]
  • 5.National Comprehensive Cancer Network. NCCN Clinical Practice Guidelines in Oncology (NCCN Guidelines®): Biliary Tract Cancer (Version 2.2025). 2025.
  • 6.National Comprehensive Cancer Network. NCCN Clinical Practice Guidelines in Oncology (NCCN Guidelines®): Colon Cancer (Version 4.2025). 2025.
  • 7.Ruiter SJS, Tinguely P, Paolucci I, et al. 3D quantitative ablation margins for prediction of ablation site recurrence after stereotactic image-guided microwave ablation of colorectal liver metastases: a multicenter study. Front Oncol. 2021;11:757167. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Lin YM, Paolucci I, O’Connor CS, et al. Ablative margins of colorectal liver metastases using deformable CT image registration and autosegmentation. Radiology. 2023;307(2):e221373. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Laimer G, Jaschke N, Schullian P, et al. Volumetric assessment of the periablational safety margin after thermal ablation of colorectal liver metastases. Eur Radiol. 2021;31(9):6489–99. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Vasiniotis Kamarinos N, Gonen M, Sotirchos V, et al. 3D margin assessment predicts local tumor progression after ablation of colorectal cancer liver metastases. Int J Hyperthermia. 2022;39(1):880–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Shady W, Petre EN, Do KG, et al. (2018) Percutaneous microwave versus radiofrequency ablation of colorectal liver metastases: ablation with clear margins (A0) provides the best local tumor control. J Vasc Interv Radiol. 29(2):268–275 e261 [DOI] [PMC free article] [PubMed]
  • 12.Chlorogiannis DD, Sotirchos VS, Georgiades C, et al. The importance of optimal thermal ablation margins in colorectal liver metastases: a systematic review and meta-analysis of 21 studies. Cancers (Basel). 2023. 10.3390/cancers15245806. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Solbiati M, Muglia R, Goldberg SN, et al. A novel software platform for volumetric assessment of ablation completeness. Int J Hyperthermia. 2019;36(1):337–43. [DOI] [PubMed] [Google Scholar]
  • 14.Anderson BM, Rigaud B, Lin YM, et al. Automated segmentation of colorectal liver metastasis and liver ablation on contrast-enhanced CT images. Front Oncol. 2022;12:886517. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Sofocleous CT, Callstrom MR, Petre EN, et al. Real-time 3D confirmation of complete ablation with margins as a local cure for colorectal liver metastases: the ACCLAIM trial. Trials. 2025;26(1):360. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Paolucci I, Albuquerque Marques Silva J, Lin YM, et al. Study protocol STEREOLAB: stereotactic liver ablation assisted with intra-arterial CT hepatic arteriography and ablation confirmation software assessment. Cardiovasc Intervent Radiol. 2023;46(12):1748–54. [DOI] [PubMed] [Google Scholar]
  • 17.van der Velden AL, Verhagen CAM, Gholamiankhah F, et al. Prospective registry study on thermal liver ablation of primary and secondary liver tumours named the A-IMAGIO study. Cardiovasc Intervent Radiol. 2025;48(8):1193–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Odisio BC, Albuquerque J, Lin YM, et al. Software-based versus visual assessment of the minimal ablative margin in patients with liver tumours undergoing percutaneous thermal ablation (COVER-ALL): a randomised phase 2 trial. Lancet Gastroenterol Hepatol. 2025;10(5):442–51. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Paolucci I, Silva JAM, Lin Y-M, et al. Quantitative ablation confirmation methods in percutaneous thermal ablation of malignant liver tumors: technical insights, clinical evidence, and future outlook. Radiol Imaging Cancer: In press; 2025. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Lee J, Rhim H, Lee MW, Kang TW, Song KD, Lee JK. Direction of tissue contraction after microwave ablation: a comparative experimental study in ex vivo bovine liver. Korean J Radiol. 2022;23(1):42–51. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Zou KH, Warfield SK, Bharatha A, et al. Statistical validation of image segmentation quality based on a spatial overlap index. Acad Radiol. 2004;11(2):178–89. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Huttenlocher DP, Klanderman GA, Rucklidge WJ. Comparing images using the Hausdorff distance. IEEE Trans Pattern Anal Mach Intell. 1993;15(9):850–63. [Google Scholar]
  • 23.Laimer G, Verdonschot KHM, Kopf L, et al. Multicenter and inter-software evaluation of ablative margins after thermal ablation of colorectal liver metastases. Eur Radiol. 2025;35(2):1046–56. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Paolucci I, Albuquerque Marques Silva J, Lin YM, et al. (2024) Identification of A0 minimum ablative margins for colorectal liver metastases: multicentre, retrospective study using deformable CT registration and artificial intelligence-based autosegmentation. Br J Surg. 111(9) [DOI] [PMC free article] [PubMed]
  • 25.Zirakchian Zadeh M, Sotirchos VS, Kirov A, et al. (2024) Three-dimensional margin as a predictor of local tumor progression after microwave ablation: intraprocedural versus 4–8-week postablation assessment. J Vasc Interv Radiol. 35(4):523–532 e521 [DOI] [PMC free article] [PubMed]
  • 26.Sandu RM, Paolucci I, Ruiter SJS, et al. Volumetric quantitative ablation margins for assessment of ablation completeness in thermal ablation of liver tumors. Front Oncol. 2021;11:623098. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Kassambara A. ggpubr: 'ggplot2' Based Publication Ready Plots. R package version 0.6.1. 2025.
  • 28.Wickham H. ggplot2: elegant graphics for data analysis. New York: Springer-Verlag; 2016. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Code and data are available on GitHub: https://github.com/ipa/acs-accuracy.