ABSTRACT
This case report describes the successful use of an augmented reality (AR) guidance system (XR90, MediView XR Inc., Cleveland Ohio) to assist in the percutaneous biopsy of a splenic mass performed in a minor procedure room. Traditional two‐dimensional (2D) imaging modalities such as ultrasound and computed tomography (CT) present limitations in depth perception, spatial orientation, and procedural ergonomics, particularly in anatomically complex and mobile regions like the spleen. The AR system integrates real‐time ultrasound, electromagnetic (EM) tracking, and pre‐procedural CT data to generate a 3D holographic overlay within a heads‐up display (HUD), enhancing operator spatial awareness and needle control. A 69‐year‐old male with a 6 cm splenic mass underwent biopsy using the AR platform alongside ultrasound. The system enabled accurate needle placement, reduced cognitive load, and provided continuous visual confirmation during the procedure. All biopsy samples were successfully obtained on the first pass without complications, and pathology confirmed high‐grade B‐cell lymphoma. This case highlights the feasibility and clinical utility of deploying AR‐guided interventions in outpatient procedure rooms, supporting broader adoption of AR technologies in settings beyond the operating room or angiography suite.
Keywords: 3‐dimensional, augmented reality, computed tomography, electromagnetic, head up display, head‐mounted display
Key Clinical Message
Augmented reality (AR) provided enhanced 3D visualization and precision guidance during biopsy of a splenic mass, helping to overcome limitations of traditional 2D imaging. By improving spatial awareness in complex anatomy, AR reduced cognitive load and supported safe, efficient needle placement in a minimally invasive procedural setting.
First‐person intraprocedural view of XR90 providing a live holographic 3D rendering of a spleen, splenic mass, and nearby lung parenchyma is used to drive procedural planning, allow for biopsy instrument tracking, define optimal trajectory, and subsequently visualize parenchymal tract closure with SinglePass biopsy tract electrocautery.

1. Introduction
1.1. Background
Clinicians performing percutaneous soft tissue biopsies rely on traditional medical imaging tools, predominantly 2D displays from ultrasound (US) and computed tomography (CT) to navigate through complex three‐dimensional (3D) anatomical spaces. This reliance on two‐dimensional views during three‐dimensional procedures often leads to increased cognitive and physical loads on proceduralists, potentially impacting precision and workflow efficiency [1]. Augmented reality (AR) technologies have emerged as promising tools to address these challenges by enhancing spatial orientation and visualization during minimally invasive needle‐based procedures [2, 3]. Recent cadaver and porcine preclinical studies using an AR system in percutaneous needle‐based procedures have shown that AR systems are as accurate and safe as procedures with traditional 2D image guidance, while significantly reducing needle repositions [2, 3, 4]. AR systems have also shown comparable accuracy, increased visualization, decreased radiation dose, and decreased procedure times in the neuro and orthopedic space [5, 6, 7, 8, 9]. Additional demonstrated benefits of AR system utilization are improved ergonomics, reduced operative time, and decreased blood loss [5]. Nearly all existing AR systems are designed to work within traditional operating rooms, interventional angiography suites, and/or hybrid operating room environments. The inherent complexity of intervention in these settings is an excellent match for the technological advances provided by AR systems. Furthermore, operating rooms and angiography suites typically feature tables built without metal components, a feature that is required for radiofrequency energy emission and interpretation that drives AR registration.
While AR systems have been proven to work in such settings for the purposes of supporting complex procedures, there is also a desire to perform AR procedures in minor procedure room settings, defined as designated spaces for performing outpatient procedures. Procedures in these settings are typically performed in short duration and may only require local anesthesia or moderate sedation. These rooms allow for faster procedural and patient turnover as compared with operating room or angiography suites. In interventional radiology, common procedures include paracentesis, thoracentesis, and biopsy procedures. In these settings, many ultrasound‐guided procedures are performed including those with complex or less familiar anatomy. The ability to bring AR into minor procedure room settings will allow for further democratization of this technology.
We present a case in which we performed a splenic biopsy utilizing an AR System as an adjunct to ultrasound (US)‐guided needle biopsy. Splenic biopsies are uncommon and present risks to the patient and challenges to the physician [3]. The spleen's vascularity introduces an increased risk for life‐threatening bleeding and hemorrhage following the procedure. Additionally, pneumothorax, pleural effusion, and colonic injury are known complications [10, 11]. The anatomical position of the spleen, situated in the upper abdominal cavity and partially obscured by the rib cage, complicates access and visualization, making the procedure technically demanding for the physician. The AR system supported precision‐guided needle biopsy along with 3D visualization to overcome the restrictive nature of the organ's location.
1.2. Technology Overview
Recent technological advances such as lightweight head‐mounted displays, along with tracked instruments (optical or EM) have led to an increased application of AR devices for clinical interventions across multiple specialties within the surgical and interventional space. AR devices have been successfully used for orthopedic, neurologic, surgical, urologic, and interventional radiology. Table 1 demonstrates a brief overview of recent literature published demonstrating the diverse implications for AR for intraprocedural guidance.
TABLE 1.
Representative applications of augmented reality in spine surgery, neurosurgery, orthopedics, and interventional radiology.
| Clinical specialty | Clinical procedure | Augmented reality (AR) application | Key findings | Citation |
|---|---|---|---|---|
| Spine surgery | Percutaneous vertebroplasty | AR/AI guided trocar insertion with motion compensation | Procedure was found to be accurate, safe, and lower patient radiation exposure as compared to SOC | Auloge et al. [6] |
| Spine surgery | Thoracolumbar pedicle screw instrumentation | Intraoperative 3D imaging and registration with AR‐Head Mounted Display (HMD) computer navigation for screw placement | Clinically accurate insertion of pedicle screws | Liu et al. [12] |
| Neurosurgery | Neurosurgical navigation for brain tumors located on the brain surface | AR based neurosurgical navigation with 3D MRI images viewed through smart glasses | Clear visualization of the surgical field with a clinically acceptable target registration error | Maruyama et al. [9] |
| Neurosurgery | Craniotomy for intracranial tumors and vascular lesions | AR neuro navigation with 3D MRI/CT reconstructions overlaying the surgical field via microscope‐ or HMD‐based visualization | AR facilitated visualization of cortex and critical structures. Preserved neurological function with no AR related complications in gross and near total resection | Gómez et al. [13] |
| Orthopedics | Minimally invasive Scapula fracture surgery | Pre‐operative virtual simulation combined with intra‐operative navigation assisted fixation | Decreased procedure times and less blood loss as compared to SOC | Guo et al. [5] |
| Orthopedics | Acetabular cup placement during total hip arthorplasty | AR system produced superimposed image in the surgical field via smarthphone | The AR system provided more accurate information as compared to the conventional method | Ogawa et al. [14] |
| Interventional Radiology | Percutaneous tumor ablation of soft tissue tumors | AR‐HMD intraprocedural holographic guidance and navigation | Intraprocedural 3D guidance agreed with the standard imaging guidance | Gadodia et al. [2] |
| Interventional Radiology | Percutaneous Ct‐guided bone biopsy | AR‐HMD intraprocedural lesion targeting and needle trajectory planning | CT passes and radiation does were significantly lower for the AR‐HMD group as compared to SOC | Albano et al. [15] |
The benefits for AR have been well charactered for specialties with fixed locations such as neurologic or orthopedic applications. However utilizing AR for percutaneous soft tissue with which are often affected by soft tissue deformation and respiratory motion require complex solutions. One such solution is the XR90 (MediView, Cleveland, OH) system which integrates live ultrasound imaging with multimodal image fusion into a head mounted display along with tracked tools for soft tisse applications.
This platform integrates several cutting‐edge technologies to deliver its unique capabilities. The system utilizes HoloLens 2 (Microsoft, Redmond, WA) as its foundational AR Head‐Mounted Display (HMD), providing an immersive visual interface for clinicians. For real‐time tracking and precise navigation, the AR system incorporates the Aurora Electromagnetic Tracking system (Northern Digital Inc., Waterloo, Canada), which works in conjunction with electromagnetically (EM) sensor‐equipped instruments, including eTRAX (CIVCO Medical Solutions, Coralville, IA) needles and ultrasound probe bracket. Integration with commercially available ultrasound systems, such as the Vivid IQ (GE Healthcare, Chicago, IL), allows for the streaming of live ultrasound imaging to a HUD that is displayed in space. Through a registration process, 3D CT‐based models of segmented patients' anatomy are projected and registered in the procedural space.
2. Case Presentation
2.1. Patient Profile
A 69‐year‐old male presented with left‐sided abdominal pain and was found to have a 6 cm heterogeneous mass within the spleen. The patient had no significant past medical history and no history of malignancy. Other than the splenic mass, the only other finding on physical examination and imaging was nonspecific, unenlarged retroperitoneal lymph nodes. Medical oncology consulted and a splenic mass biopsy was requested to be performed in interventional radiology.
Interventional radiology review of the case revealed a normal sized spleen tucked posteriorly within the upper abdominal cavity. There was a 6 cm hypoattenuating, hypoenhancing mass within the splenic parenchyma (Figure 1a). On non‐contrast imaging, however, the splenic mass was poorly defined (Figure 1b).
FIGURE 1.

(a) Hypovascular, hypoattenuating splenic mass (white arrows) is seen within the splenic parenchyma. The spleen is not enlarged and is located posteriorly within the upper abdominal cavity. (b) Noncontrast CT imaging reveals that the splenic mass is isoattenuating with splenic parenchyma and cannot easily be visualized. Large white arrow indicates one of three MediView XR90 skin markers required for registration.
While a CT‐guided biopsy could be performed using landmarks, given the splenic location, vascularity, and depth, precise needle trajectory planning and real‐time visualization were deemed important to minimize procedural risks and maximize technical success. The patient provided informed consent for the procedure and the use of the AR system.
2.2. Procedure Overview
This procedure was completed by an Interventional Radiologist with less than 1 year of AR experience. AR system specific benchtop training was completed prior to clinical procedures. The proceduralist participated in only 2 clinical cases utilizing the AR system prior to this highlighted case, highlighting ease of use. Training was performed rapidly using a phantom model provided by the vendor and consisted of an approximately 20‐min walk‐through of a pre‐existing example case.
Pre‐procedural imaging for the procedure was conducted using a conventional CT scanner. Three radio‐opaque markers were placed on the patient in the left upper quadrant (Figure 1b). Axial CT images with 1.25 mm thick reconstructions were obtained and transferred to the Picture Archiving and Communications System. Following the scan, the radio‐opaque markers were replaced with permanent surgical skin markings, after which the patient was transported to the minor procedure room.
On a separate workstation, the axial CT images were downloaded and imported into a 3D medical imaging post‐processing software (Tera Recon, Concert AI, Durham, NC). Within this workflow, the three radio‐opaque markers, the lung parenchyma of the left lower lobe, the spleen, and the splenic mass were individually contoured. Each of these segmented 3D datasets was then exported in an OBJ file format and subsequently uploaded to the AR system. With the pre‐procedural data prepared, the intervention began.
In the minor procedure room, the patient was placed onto a SurgiGraphic 6000 image‐guided surgical table (Steris, Mentor, OH), followed by sterile prep and draping of the left upper quadrant. Sterile registration markers were then placed on the skin markings. The operator donned the headset and AR registration was initiated and completed using verbal registration commands to the system.
Prior to biopsy initiation, a 3D rendering of the spleen, splenic mass, and lungs was visualized holographically for operator familiarization and treatment planning in open space (Figure 2). Subsequently, this 3D rendering and tracked tools were then registered to the electromagnetic (EM) coordinate space overlaying the 3D structures' patient's anatomy within the EM field to guide biopsy. Upon identifying the optimal trajectory, extraneous structures were removed with verbal command to localize the mass (Figure 3a). A 17‐gauge biopsy guide needle was then advanced into the splenic mass using combined AR and ultrasound guidance (Figure 3b). Core needle biopsies were subsequently performed using sonographic guidance via the head up display feature, allowing for continuous dual confirmation of needle positioning with both sonography and AR throughout the remainder of the procedure.
FIGURE 2.

Treatment planning is pursued external to the patient. AR visualization of structures of interest in this case, spleen (blue), splenic mass (purple—partially visualized), and the lower lung (green) allows for dynamic manipulation of structures and review of potential procedural trajectories and plans.
FIGURE 3.

(a) Splenic mass (purple) is visualized in AR onto and within the patient's anatomy. (b) 17‐gauge biopsy guide needle is advanced into the mass under sonographic and AR guidance. (c) Real‐time sonographic guidance of a percutaneous biopsy pass closure with SinglePass Kronos electrocautery is now visualized via a “heads up display” while retaining AR guidance of the splenic mass within the patient torso.
Three core needle biopsy passes were performed and three core needle specimens were successfully obtained using an 18‐gauge biopsy device (Bard Max‐Core, Becton Dickinson, Franklin Lakes, NJ). The needle track was closed with an electrocautery device (Single Pass Kronos, Single Pass, Lake Forest, CA), with clear delineation of the cauterization tract on final imaging (Figure 3C). Total procedural duration was 15 min.
The patient was observed for 4 h under bedrest restrictions and then discharged home without complications. Biopsy specimens were diagnostic. Final pathology confirmed adequate and diagnostic specimens, revealing high‐grade B‐cell lymphoma, leading to the initiation of treatment. No intra‐procedural or immediate post‐procedural complications were observed.
3. Discussion
This case report demonstrates the successful application of the AR system for a percutaneous splenic mass biopsy performed in a minor procedure room. The procedure, which can be challenging due to the spleen's anatomical location, mobility, and vascularity, was executed with precision and efficiency.
The integration of AR technology provided distinct advantages over conventional 2D ultrasound guidance. The real‐time holographic superposition of pre‐procedural CT data with live ultrasound offered an enhanced spatial understanding of the target and critical adjacent structures as compared to traditional techniques. This “X‐ray vision” capability allowed the proceduralist to navigate the needle path with increased confidence and precision, potentially reducing the risk of complications associated with inadvertent organ or vessel puncture. Furthermore, this added confidence resulted in anecdotally faster procedural time and, therefore, reduced sedation time and room utilization. This aligns with findings from other studies exploring AR in percutaneous procedures, which suggest comparable accuracy and reduced cognitive load compared to traditional guidance methods [3, 16].
A key implication of this case is the demonstrated feasibility of operationalizing advanced AR navigational technology in a minor procedure room. Historically, such image‐guidance systems have been confined to highly equipped operating rooms or angiography suites due to infrastructure requirements coupled with perceived complexity. This successful application in a less specialized setting suggests broader applicability for AR systems, potentially democratizing access to precision guidance for a wider range of percutaneous interventions and treatment settings.
4. Conclusion
The AR system effectively guided a percutaneous splenic mass biopsy in a minor procedure room, demonstrating enhanced anatomical visualization, improved procedural ergonomics, and successful target acquisition without complications. This case highlights the system's potential to improve precision and efficiency in challenging interventions and suggests a broader utility of AR technology in diverse clinical environments beyond conventional specialized suites.
Author Contributions
Alexander S. Misono: conceptualization, data curation, formal analysis, funding acquisition, investigation, methodology, project administration, resources, software, supervision, validation, visualization, writing – original draft, writing – review and editing.
Funding
MediView provided funding for publication fees.
Ethics Statement
As a single case report with the patient's signed consent, no other ethical review was required.
Consent
Written informed consent was obtained from the patient for the publication of this case report.
Conflicts of Interest
Author is a member of the Scientific Advisory Board for MediView XR Inc.
Misono A. S., “Augmented Reality Visualization and Navigation Operationalized in Biopsy of Indeterminate Splenic Mass,” Clinical Case Reports 14, no. 1 (2026): e71850, 10.1002/ccr3.71850.
Data Availability Statement
Data sharing is not applicable to this article as no datasets were generated or analyzed for this case report. All relevant data have been shared in this manuscript.
References
- 1. Lusch A., Bucur P. L., Menhadji A. D., et al., “Evaluation of the Impact of Three‐Dimensional Vision on Laparoscopic Performance,” Journal of Endourology 28, no. 2 (2013): 261–266. [DOI] [PubMed] [Google Scholar]
- 2. Gadodia G., Yanof J., Hanlon A., et al., “Early Clinical Feasibility Evaluation of an Augmented Reality Platform for Guidance and Navigation During Percutaneous Tumor Ablation,” Journal of Vascular and Interventional Radiology 33, no. 3 (2022): 333–338. [DOI] [PubMed] [Google Scholar]
- 3. Gadodia G., Evans M., Weunski C., Ho A., Cargill A., and Martin C., “Evaluation of an Augmented Reality Navigational Guidance Platform for Percutaneous Procedures in a Cadaver Model,” Journal of Medical Imaging 11, no. 6 (2024): 062602. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Datta S., Short R. F., Milsom J. W., et al., “In‐Vivo Evaluation of an Augmented Reality Enhanced Ultrasound Needle Guidance System for Minimally Invasive Procedures in Porcine Models: A Preclinical Comparative Study,” Journal Minimum Invasive Surgery 28, no. 3 (2025): 122–129. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Guo Q., Li X., Tang Y., Huang Y., and Luo L., “Augmented Reality and Three‐Dimensional Plate Library‐Assisted Posterior Minimally Invasive Surgery for Scapula Fracture,” International Orthopaedics 46, no. 4 (2022): 875–882. [DOI] [PubMed] [Google Scholar]
- 6. Auloge P., Cazzato R. L., Ramamurthy N., et al., “Augmented Reality and Artificial Intelligence‐Based Navigation During Percutaneous Vertebroplasty: A Pilot Randomised Clinical Trial,” European Spine Journal 29, no. 7 (2020): 1580–1589. [DOI] [PubMed] [Google Scholar]
- 7. Faiella E., Castiello G., Bernetti C., et al., “Impact of an Augmented Reality Navigation System (SIRIO) on Bone Percutaneous Procedures: A Comparative Analysis With Standard CT‐Guided Technique,” Current Oncology 28, no. 3 (2021): 1751–1760. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Montemurro N., Condino S., Carbone M., et al., “Brain Tumor and Augmented Reality: New Technologies for the Future,” International Journal of Environmental Research and Public Health 19 (2022): 6347. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Maruyama K., Watanabe E., Kin T., et al., “Smart Glasses for Neurosurgical Navigation by Augmented Reality,” Oper Neurosurgery 15, no. 5 (2018): 551–556. [DOI] [PubMed] [Google Scholar]
- 10. Razak S. and Khan A., “Ultrasound Guided Percutaneous Splenic Biopsy: Diagnostic Accuracy and Complication,” EMJ Radiology (2025), 10.33590/emjradiol/JMVR8857. [DOI] [Google Scholar]
- 11. Quinn S. F., vanSonnenberg E., Casola G., Wittich G. R., and Neff C. C., “Interventional Radiology in the Spleen,” Radiology 161, no. 2 (1986): 289–291. [DOI] [PubMed] [Google Scholar]
- 12. Liu A., Jin Y., Cottrill E., et al., “Clinical Accuracy and Initial Experience With Augmented Reality–Assisted Pedicle Screw Placement: The First 205 Screws,” Journal of Neurosurgery. Spine 36, no. 3 (2021): 351–357. [DOI] [PubMed] [Google Scholar]
- 13. Gómez D. F., Vargas‐Osorio M., RamírezSanabria A. D., et al., “Feasibility of Utilizing Augmented Reality in Neurosurgery: Insights From a Single‐Center Experience,” Journal of Neurosciences in Rural Practice 16, no. 2 (2025): 184–191. [Google Scholar]
- 14. Ogawa H., Hasegawa S., Tsukada S., and Matsubara M., “A Pilot Study of Augmented Reality Technology Applied to the Acetabular Cup Placement During Total Hip Arthroplasty,” Journal of Arthroplasty 33, no. 6 (2018): 1833–1837. [DOI] [PubMed] [Google Scholar]
- 15. Albano D., Messina C., Gitto S., Chianca V., and Sconfienza L. M., “Bone Biopsies Guided by Augmented Reality: A Pilot Study,” European Radiology Experimental 7, no. 1 (2023): 40. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Borde T., Saccenti L., Li M., et al., “Smart Goggles Augmented Reality CT‐US Fusion Compared to Conventional Fusion Navigation for Percutaneous Needle Insertion,” International Journal of Computer Assisted Radiology and Surgery 20, no. 1 (2025): 107–115. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data sharing is not applicable to this article as no datasets were generated or analyzed for this case report. All relevant data have been shared in this manuscript.