Development of a multiplex real-time RT-PCR assay for simultaneous detection of 18 respiratory viruses

Aijuan Xu 1,2,3,4, Longxu Xie 3, Liping Zhong 3, Xiuli Zhao 5, Zhenzhou Wan 5,*, Jian-Hua Wang 2,*, Chiyu Zhang 4,*
PMCID: PMC12833288  PMID: 41602116

Abstract

Acute respiratory tract infections (ARTIs) are a major global health burden with high morbidity and mortality, mainly affecting children under 5 years old and people over 60 years old. The majority of ARTIs are caused by respiratory viruses, such as influenza virus (IFV), respiratory syncytial virus (RSV), human rhinovirus (HRV), human parainfluenza virus (HPIV), human adenovirus (HAdV), human coronavirus (HCoV), human bocavirus (HBoV) and human metapneumovirus (HMPV). Accurate and sensitive diagnosis of respiratory viruses is crucial for clinical treatment and management of patients with ARTIs. Here, we developed a multiplex RT-qPCR assay for detection of 18 common respiratory viruses. The in-house multiplex RT-qPCR assay possesses high sensitivity (limit of detection [LOD]: 5–75 copies/25 μL) and a good linear correlation between the logarithmic copy number and cycle threshold (Ct value, R²≥0.9944), along with excellent specificity characterized by no cross-reactivity within the 18 viruses and no non-specific amplification for other human pathogens. The in-house assay was assessed using 628 clinical samples and compared with Sansure and Easy Diagnosis kits. The results showed a sensitivity of 99.81%, a specificity of 100%, and an overall consistency rate of 99.84% for six target viruses (IAV, IBV, RSV, HRV, HAdV, and SARS-CoV-2). We also reported high prevalences of IAV, HAdV and HRV in Taizhou, Jiangsu, China, and a high potential of HRV co-infection with other respiratory viruses. The new multiplex RT-qPCR assay offers a sensitive, specific, and cost-effective tool for monitoring 18 common respiratory viruses to mitigate the prevalence of ARTIs.

Keywords: acute respiratory tract infections; co-infection; primers and probe; multiplex RT-qPCR; specificity, respiratory viruses; human rhinovirus; enterovirus

1. Introduction

Acute respiratory tract infections (ARTIs) constitute a major global health challenge, marked by high morbidity and mortality rates (Vos et al., 2020). Respiratory viruses are the leading cause of ARTIs with a particular impact on pediatric and geriatric populations (Marrie et al., 1989; Russell et al., 2015). According to the surveillance data from 2009–2019 by the Key Laboratory of Surveillance and Early-Warning on Infectious Diseases, Chinese Center for Disease Control and Prevention, the positive rate of respiratory viruses among children under five years old and school-aged children reached as high as 46.9% (Li et al., 2021). Influenza virus (IFV), respiratory syncytial virus (RSV), human adenovirus (HAdV), human rhinovirus (HRV), and human coronavirus (HCoV) are recognized as the primary etiological agents of ARTIs in the general population. Notably, IFV and RSV infections are responsible for over 300,000 annual deaths worldwide among children under five years of age (Azar et al., 2018). Furthermore, emerging coronaviruses, such as Middle East respiratory syndrome coronavirus (MERS-CoV) and severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), continue to pose substantial threats to global public health, highlighting the persistent risks associated with this viral genus (Zhang et al., 2020). In particular, co-infections of respiratory viruses with SARS-CoV-2 have been identified as significant contributors to morbidity and mortality in COVID-19 patients, presenting challenges for healthcare providers due to their overlapping clinical manifestations (Ma et al., 2020; Frediansyah et al., 2021; Swets et al., 2022).

Accurate and readily accessible molecular diagnostic techniques for screening and identifying respiratory viruses are crucial to the treatment and management of patients with ARTIs. Traditional diagnostic methods—including cell culture, hemagglutination inhibition tests, enzyme immunoassays, and direct fluorescent antibody assays—are often time-consuming, exhibit low sensitivity or specificity, and are highly dependent on operator expertise (Mahony, 2008; Ginocchio and McAdam, 2011; Vemula et al., 2016). Reverse transcription-quantitative PCR (RT-qPCR) not only offers high sensitivity and specificity but also enables multiplex detection of different pathogens (Mahony, 2010; Zhang et al., 2020; Horemheb-Rubio et al., 2022; Jiang et al., 2022). Therefore, RT-qPCR technology has been considered as a gold standard method for molecular diagnosis of pathogens globally.

In this study, we developed a multiplex RT-qPCR assay for detection of 18 common respiratory viruses, including influenza A virus (IAV), influenza B virus (IBV), influenza C virus (ICV), RSV, HRV, enterovirus (EV), HAdV, SARS-CoV-2, human coronavirus HKU1 (HCoV-HKU1), human coronavirus 229E (HCoV-229E), human coronavirus NL63 (HCoV-NL63), human coronavirus OC43 (HCoV-OC43), human metapneumovirus (HMPV), human bocavirus (HBoV), human parainfluenza virus type 1 (HPIV-1), HPIV-2, HPIV-3, and HPIV-4. The in-house multiplex RT-qPCR assay demonstrates high sensitivity with a limit of detection (LOD) ranging from 5–75 copies/25 μL, and exhibits excellent specificity with no cross-reactivity observed among 18 non-target viruses. Clinical evaluation using 628 clinical samples showed that the in-house multiplex RT-qPCR assay exhibited a sensitivity of 99.81%, a specificity of 100% and an overall consistency rate of 99.84% for the six target viruses (IAV, IBV, RSV, HRV, HAdV and SARS-CoV-2) when compared with commercially available kits.

2. Materials and methods

2.1. Design of primers and probes

To obtain optimal primers and probes for detection of respiratory viruses, 2–7 sets of primers and probes were designed for each virus using Oligo7 software (http://oligo.net) or retrieved from previous papers (Supplementary Table S1) (Sanghavi et al., 2011; Lee et al., 2021; Mancini et al., 2021; Yüce et al., 2021). The primers and probe were specifically tailored by selecting the most conserved genomic regions of the viruses. The specificities of primers and probes were preliminarily verified using Primer-BLAST (https://www.ncbi.nlm.nih.gov/tools/primer-blast/index.cgi?LINKLOC=BlastHome). To evaluate the adaptations to virus variants or quasispecies, the primers and probes were analyzed by aligning with corresponding viral sequences downloaded from GenBank (Supplementary Figure S1). The human B2M gene was selected as the internal control (IC) for the detection of 18 respiratory viruses. The probes were designed by incorporating fluorescent dyes (CY5/FAM/VIC/ROX) and quenchers (BHQ1/BHQ2) at the 5’ and 3’ ends, respectively. All primers and probes were synthesized by General Biotechnology Co., Ltd. (Anhui, China) and reconstituted in molecular-grade RNase-free water (Vazyme, China).

2.2. Preparation of RNA standards

Plasmids PUC-57 containing the target gene sequences of IAV, IBV, ICV, RSV, HRV, EV, HAdV, SARS-CoV-2, HCoV-HKU1, HCoV-229E, HCoV-NL63, HCoV-OC43, HMPV, HBoV, HPIV-1, HPIV-2, HPIV-3 and HPIV-4 were synthesized by General Biol (Anhui, China). For RNA viruses, their target gene fragments were amplified from the plasmids using specific PCR primers and Taq DNA Polymerase (Vazyme, China) following the manufacturer’s instructions. A T7 promoter was added to the 5’ end of the forward or reverse PCR primer according to the genomic strand of the virus (Supplementary Table S2). PCR amplicons were separated by agarose gel electrophoresis and purified using the FastPure Gel DNA Extraction Mini Kit (Vazyme, China). After quantification with a NanoDrop2000 spectrophotometer (Thermo Fisher Scientific, USA), 0.1-0.5 µg of the purified PCR products were used as templates for in vitro transcription with the T7 High Yield RNA Transcription Kit (Vazyme, China) at 37°C for approximately 16 hours. The newly synthesized RNA was treated with DNase I at 37°C for 15 minutes. RNA concentration was measured using a NanoDrop2000, and the RNA was diluted in TE solution (Sigma, USA). RNase inhibitor (Vazyme, China) was added to the RNA storage solution, which was then aliquoted and stored at -80°C. The copy number of RNA standards was calculated using the formula: RNA copies per mL = [RNA concentration (ng/mL)/(transcript length in nt × 340)] × 6.022 × 10²³ and the copy number of DNA standards was calculated using the formula: DNA copies per mL = [DNA concentration (ng/mL)/(transcript length in nt × 660)] × 6.022 × 10²³.

2.3. Establishment of singleplex and multiplex RT-qPCR assays

The singleplex RT-qPCR reaction system consists of 0.5× amplification buffer, 200 μM dNTPs, 1 U Taq DNA polymerase, 2 U reverse transcriptase, 20 U RNase inhibitor (Vazyme, China), 200 nM of each primer, and 100 nM probe. The multiplex RT-qPCR reaction system includes 1× amplification buffer, 200 μM dNTPs, 2 U Taq DNA polymerase, 40 U reverse transcriptase, 30 U RNase inhibitor, 0.15 U UNG enzyme (Vazyme, China), and the primers and probe for three viruses. 3 μL DNA/RNA standards or samples were added in each reaction. The thermal cycling conditions were as follows: reverse transcription at 55°C for 15 min; enzyme activation and initial denaturation at 95°C for 30 s; followed by 45 cycles of denaturation at 95°C for 10 s and combined annealing/extension at 60°C for 1 min. Fluorescence signals were collected during the annealing step using an ABI 7500 PCR Detection System (Thermo Fisher Scientific, USA).

2.4. Screening for optimal primer sets

To screen the optimal primer and probe set for each virus, parallel experiments were performed with the same reaction system and conditions except different sets of primers and probes. The primer and probe set yielding the lowest cycle threshold (Ct) value was selected for the RT-qPCR assay (Supplementary Figure S2). To improve the performance, some probes were modified by incorporating minor groove binder (MGB) to shorten the probe length.

2.5. Standard curves for quantification and limit of detection

The standard curves of the multiplex RT-qPCR assays were generated using 10-fold serial dilutions of RNA or DNA standards ranging from 106 to 100 copies/μL. The LODs of the multiplex assays were determined using 10-fold or 5-fold serial dilutions of RNA or DNA standards. Each dilution was tested with 20 replicates. The LOD was defined as a 95% probability of obtaining a positive result, using probit regression analysis with SPSS 26.0 software.

2.6. Specificity of the multiple RT-qPCR assays

As mentioned above, the specificities of primers and probes were firstly assessed by against all available genomic sequences of the 18 respiratory viruses in GenBank with Primer-BLAST (https://www.ncbi.nlm.nih.gov/tools/primer-blast/index.cgi?LINKLOC=BlastHome). Then, the specificity of the multiplex assay was evaluated using 24 virus isolates and 1 clinical sample previously confirmed positive for HBoV. The virus isolates were purchased from BeNa Culture Collection (BNCC) (Zhengzhou, China), including IAV-H1N1 (2009), IAV-H5N1, IAV-H3N2, IAV-H7N9, IBV-Yamagata (BY), IBV-Victoria (BV), RSV-A, RSV-B, HRV-A, HRV-B, EV-71, HAdV-1, HAdV-2, HAdV-3, HAdV-4, HAdV-5, HAdV-7, SARS-CoV-2, HCoV-229E, HCoV-OC43, HMPV, HPIV-1, HPIV-2 and HPIV-3. All these virus isolates have concentration of 104 copies/mL, except RSV-A and RSV-B, both of which had a concentration of 105 copies/mL. The HBoV-positive sample was determined to have a concentration of 2.14×105 copies/mL. These virus isolates and clinical sample had Ct values of 21–32 in the RT-qPCR assays (Figure 1). Because of lack of virus isolates and positive clinical samples for HPIV-4, HCoV-NL63, HCoV-HKU1 and ICV, we used four recombinant plasmids containing target gene fragments to evaluate analytical specificity of the assays and their concentrations were all 106 copies/mL. The target sequences were retrieved from GenBank (accession numbers: e.g., HPIV-4: ON778025.1; HCoV-NL63: LC687395.1; HCoV-HKU1: NC_006577.2; ICV: AB000611.1) and synthesized into pUC57 vectors by General Bio (Anhui) Co., Ltd.

Figure 1.

Six bar charts display fluorescence intensity (FI) levels for various viruses. Each chart is labeled with groups: IAV/IBV/RSV, HRV/HMPV/SARS-CoV-2, HAdV/HBoV/ICV, EV/HCoV-229E/HCoV-OC43, HPIV-1/HPIV-2/HPIV-3, and HPIV-4/HCoV-NL63/HKU1. The x-axis lists viruses, and the y-axis shows FI values. Colors differentiate virus types. Bars vary in height, indicating differing FI levels across viruses.

The specificity of the multiplex RT-qPCR assay. FI, fluorescence intensity.

2.7. Intra-assay and inter-assay reproducibility of the multiplex RT-qPCR assays

To evaluate the stability of the assays, intra-assay tests were respectively performed with 10 replicates at three different days. The intra-batch and inter-batch mean values, standard deviations, and coefficients of variation were calculated.

2.8. Evaluation with clinical samples

A total of 628 oropharyngeal swab samples were collected from outpatients, inpatients (with community-acquired pneumonia, upper respiratory tract infections, and other related diseases), and healthy individuals at Taizhou Fourth People’s Hospital during March 2024 to March 2025.

Viral DNA/RNA was extracted from 300 μL of each oral swab sample using a DNA/RNA extraction kit (Tianlong Biotech Inc., Shanxi Province, China) and eluted in 60 μL of nuclease-free water. The extracted DNA/RNA was stored at −80°C until use. 5 μL of DNA/RNA sample was added to the reaction.

A parallel test was performed using commercial RT-qPCR assays (Sansure Biotech Inc. and EasyDiagnosis) for each sample. The commercial kit from Sansure Biotech Inc. (Hunan Province) can detect influenza A virus, influenza B virus, RSV, HRV, HAdV, and Mycoplasma pneumoniae. The kit from EasyDiagnosis (Hubei Province) was used to detect SARS-CoV-2.

To validate the discrepant results between our in-house multiplex assay and the commercial kits, two complementary validation strategies were employed: first, nested RT-PCR assays were developed and conducted (details in Supplementary Table S3), with their amplification products targeting pathogens that are also detected by the commercial kits; second, specific primers for Sanger sequencing were designed, which targeted pathogens uniquely included in our in-house multiplex assay panel. For pathogens showing inconsistent detection results between the two platforms, Sanger sequencing was performed using an ABI 3730 Genetic Analyzer (Thermo Fisher Scientific, USA) to confirm their presence or absence. All sequences obtained from this validation process are provided in Supplementary Table S4.

2.9. Statistical analysis

The performance of the multiplex RT-qPCR assay was assessed by sensitivity (SE), specificity (SP) and consistency rate (CR). SE is calculated by SE = TP/(TP + FN) × 100%. SP is calculated by SP = TN/(FP +TN) × 100%. CR is calculated by CR = (TP + TN)/(TP + TN + FP + FN). A Kappa test was performed with SPSS v 26.0 software to estimate the degree of consistency between our multiplex RT-qPCR assays and the commercial RT-qPCR kits. The Kappa statistic is interpreted as follows: <0, as poor; 0–0.20, as slight; 0.21–0.40, as fair; 0.41–0.60, as moderate; 0.61–0.80, as substantial; and 0.81–1.00, as almost perfect agreement (Liu et al., 2014).

3. Results

3.1. Optimization of the multiple RT-qPCR assay

The reaction with the earliest amplification curve (i.e., the lowest Ct value) indicates the highest amplification efficiency. Two to seven sets of primers and probes were subjected to the screening of the optimal primer/probe set. Based on amplification efficiency and specificity, the primer/probe set with the lowest Ct value and without non-specific amplification (no amplification curve in non-template control) was selected for each virus. As a result, IAV-set5, IBV-set1, RSV-set5, HRV-set2, SARS-CoV-2-set1, HMPV-set1, ICV-set1, HBoV-set5, HAdV-set7, HCoV-OC43-set3, HCoV-229E-set2, EV-set3, HCoV-HKU1-set1, HCoV-NL63-set2, HPIV-4-set1, HPIV-1-set5, HPIV-2-set4, and HPIV-3-set3 were selected for the establishment of the singleplex and multiplex RT-qPCR assays since they exhibited the highest amplification efficiency and well specificity (Supplementary Figure S2). To cover 18 respiratory viruses, six independent triplex reactions (tubes) were established, including tube 1: IAV, IBV and RSV; tube2: HRV, SARS-CoV-2 and HMPV; tube3: ICV, HBoV and HAdV; tube4: EV, HCoV-229E and HCoV-OC43; tube5: HPIV-1, HPIV-2 and HPIV-3; tube6: HCoV-NL63, HCoV-HKU1 and HPIV-4 (Table 1). The optimal reaction mixture of each multiplex reaction was obtained by further optimizing the amounts of Taq DNA polymerase, RNA inhibitor, RT, UNG, as well as the concentrations of dNTPs, primers and probes (Supplementary Figure S3).

Table 1.

The information of the primers and probes used in the multiple RT-qPCR assay.

Reaction Viruses Primers/probes Sequences (5’-3’)’ Target gene Ref
Tube 1 IAV FRP GTCTTCTAACCGAGGTCGAAACGAGGTGACAGGATYGGTCTTGTCTTVIC-CRTCYGGCCCCCTCAAAGCCGA-BHQ1 M Thisstudy
IBV FRP TCCTCAACTCACTCTTCGAGCGCGGTGCTCTTGACCAAATTGGROX-CCAATTCGAGCAGCTGAAACTGCGGT-BHQ2 NS&NEP (Mancini et al., 2021)
RSV FRP GGGCAAATATGGAAACATACGGCACCCATATTGTTAGTGATGCFAM-CTTCACGARGGCTCCACATACACAG-3'BHQ2+TMP M (Lee et al., 2021)
Tube 2 HRV FRP TGTGAAGASYCGCGTGTGCGTCCCATCCCRCAATTRCTCFAM-TCCTCCGGCCCCTGAATGCGGCTAA-BHQ1 5’UTR This study
SARS-CoV-2 FRP TTCTCCTGCTAGAATGGCTGGCATTTTGCTCTCAAGCTGGTTCAATCVIC-TGATGCTGCTCTTGCTTTGCTGCTGCT-BHQ1 N This study
HMPV FRP TGGVAAAGCWTTAGGCTCATCTGTTAGATGAYCTGGCRATGACROX-CCCCACCTYAGCATTGTTTGACCAGCWC-BHQ2 N This study
Tube 3 ICV FRP CTGGAGACTTYTTGGGAGTGGAGTTCTCAACCAAGCTGTGATTGTTCCROX-ACCCAGACGACTACACACCAGACATCCG-BHQ2 M This study
HBoV FRP CCTATATAAGCCGATGCACTTCGCTTTTTCCAGAGATGTTCACVIC-TCAGACTGCATCCGGTCTC-BHQ1 NS1&NP1 This study
HAdV FRP CCAAAGGMCCCCATCCARGTCTACTTYCAYCACATCAACAGFAM-ATYGGCTCKCATCCTCGCAC- 3'MGB+TMP Hexon This study
Tube 4 HCoV-OC43 FRP TCAATACCCCGGCTGACATCCTGAGCCTTCAATATAGTAACCCFAM-TRCCAGGCGGAAACCTAGTCGGAAT-BHQ1 N&M This study
HCoV-229E FRP GTACCCCTAAGCCTTCTCGTAATTAGGCTGTCTTTTCCACCGTVIC-AGTCCTGCTTCTTCTCAATCTGCTGCC-BHQ1 N This study
EV FRP GCAASTCTGCRGCGGAACCATTGTCACCATAAGCAGCCAROX-CTACTTTGGGTGTCCGTGTTTCCT-BHQ2 5’UTR This study
Tube 5 HPIV-1 FRP CAACAGGAAATCATGTTCTGTAATAGCAAATACTAAGTCTTCTATACCTTCACFAM-AGTTATGCTCCTTGCCCACTGTGAATG-BHQ1 HN This study
HPIV-2 FRP GSRTTTCCAATCTTCAGGACTATGCAAGTCTCAGTTCAGCTAGRTCVIC-AATCAATCGSAAAAGCTGTTCAGTCAC-BHQ1 HN This study
HPIV-3 FRP TCTCGGGTATGGAGGTCTTGAACAATGCYTGATTRCAGTCTCTCTGTGTROX-CCAGGTCACCCAGTTGTGTTGCAGA-BHQ2 HN This study
Tube 6 HCoV-HKU1 FRP GCGGAGGAGTTGGCTAGTAATGCAGCGACAATCTATCTTCAROX-ATGCCCCTCTACTGGTCAAGCGATGG-BHQ1 ORF 1ab This study
HCoV-NL63 FRP TAGTGGTAAGTCACATTGTTCCATCAATGCTATAAACAGTCATAGCTTFAM-CTTGTGCCCATGCTGCTGTTGATTCCT-BHQ1 ORF 1ab This study
HPIV-4 FRP AGATCACAATGAACTACWGATGCCAYTCTGATGGGTGGAKTCTGGTGVIC-TTCTGGGATGAGTGGTGYTCTGRCTGT-BHQ1 N This study

The adenine (A) residue marked in red is modified with the TMP probe, aiming to enhance detection sensitivity and accelerate the reaction.

3.2. Specificity for the RT-qPCR assays

A total of 29 pathogens, including 24 culture strains (IAV-H1N1 (2009), IAV-H5N1, IAV-H3N2, IAV-H7N9, IBV-Yama (BY), IBV-Victoria (BV), RSV-A, RSV-B, HRV-A, HRV-B, EV-71, HAdV-1, HAdV-2, HAdV-3, HAdV-4, HAdV-5, HAdV-7, SARS-CoV-2, HCoV-OC43, HCoV-229E, HMPV, HPIV-1, HPIV-2, HPIV-3), 1 clinical sample positive for HBoV and 4 Plasmids (HPIV-4, HCoV-NL63, HCoV-HKU1 and ICV) were used to determine the specificity of the multiplex RT-qPCR in this study. Neither non-specific amplification nor cross-reaction was observed among these pathogens, indicating high clinical and analytical specificity of the RT-qPCR assays (Figure 1, Supplementary Figure S4).

3.3. LOD for RT-qPCR assay

The LODs of the single RT-qPCR assay for IAV, IBV, ICV, RSV, HRV, EV, HAdV, SARS-CoV-2, HCoV-HKU1, HCoV-229E, HCoV-NL63, HCoV-OC43, HMPV, HPIV-1, HPIV-2, HPIV-3, HPIV-4 and HBoV were determined as 8, 44, 75, 17, 21, 50, 68, 9, 16, 25, 39, 15, 36, 17, 40, 30, 14 and 152 copies per 25 μL reaction, respectively (Table 2). The LODs of the multiple-RT-qPCR assays for IAV, IBV, ICV, RSV, HRV, EV, HAdV, SARS-CoV-2, HCoV-HKU1, HCoV-229E, HCoV-NL63, HCoV-OC43, HMPV, HPIV-1, HPIV-2, HPIV-3, HPIV-4 and HBoV were 5, 50, 18, 11, 38, 50, 75, 11, 17, 15, 54, 15, 31, 7, 31, 71, 9 and 41 copies per 25 μL reaction, respectively (Table 2).

Table 2.

LOD of the multiple RT-qPCR assay.

Panels Viruses LOD (copies/25 µL reaction)
Multiplex reaction Singleplex reaction
Tube 1 IAV 5 8
IBV 50 44
RSV 11 17
Tube 2 HRV 38 21
SARS-CoV-2 11 9
HMPV 31 36
Tube 3 ICV 18 75
HBoV 41 152
HAdV 75 68
Tube 4 HCoV-OC43 13 15
HCoV-229E 15 25
EV 50 50
Tube 5 HPIV-1 7 17
HPIV-2 31 40
HPIV-3 71 30
Tube 6 HCoV-HKU1 17 16
HCoV-NL63 54 39
HPIV-4 9 14

3.4. Standard curves for the RT-qPCR assays

A good linear relationship between the log copy number and the Ct value in the standard curve was observed for each virus within the scope from 1 × 101 to 1 × 106 copies/μL, regardless of the singleplex (Supplementary Figure S5) and multiplex RT-qPCR assays (Figure 2). The correlation coefficients (R2) of the singleplex RT-qPCR assays for IAV, IBV, ICV, RSV, HRV, EV, HAdV, SARS-CoV-2, HCoV-HKU1, HCoV-229E, HCoV-NL63, HCoV-OC43, HMPV, HPIV-1, HPIV-2, HPIV-3, HPIV-4 and HBoV were 0.9988, 0.9987, 0.9993, 0.997, 0.9994, 0.9988, 0.9971, 0.9944, 0.9995, 0.9965, 0.9989, 0.9986, 0.9998, 0.9997, 0.9976, 0.9996, 0.9993 and 0.9993, respectively. TheR2 of the multiplex RT-qPCR assays for IAV, IBV, ICV, RSV, HRV, EV, HAdV, SARS-CoV-2, HCoV-HKU1, HCoV-229E, HCoV-NL63, HCoV-OC43, HMPV, HPIV-1, HPIV-2, HPIV-3, HPIV-4 and HBoV were 0.9989, 0.998, 0.9994, 0.9998, 0.9986, 0.9992, 0.9992, 0.9973, 0.9978, 0.9989, 0.9994, 0.9999, 0.9998, 0.9978, 0.9972, 0.9989, 0.9994 and 0.9995, respectively. These results indicate that the assays can be well used for absolute quantification of viral load.

Figure 2.

Multiple line graphs show the relationship between viral copies number (log10) on the x-axis and Ct value on the y-axis for various viruses. Each graph includes a linear regression equation and R-squared value, indicating the fit. Graphs are arranged in tubes labeled one to six, representing different viruses like IAV, IBV, RSV, HBoV, SARS-CoV-2, and others. The dotted trend lines slope downward in each graph, demonstrating a decrease in Ct value as viral copies increase.

Standard curves of the multiplex RT-qPCR assay for 18 respiratory viruses. There was a good linear correlation between the logarithmic copy number and cycle threshold (Ct value) for each virus (R² ≥ 0.9972).

3.5. Intra-assay and inter-assay reproducibility of the multiplex RT-qPCR assays

The intra-assay reproducibility of the multiplex RT-qPCR assays were assessed using 10 replicates of 3000 copies template input per 25 μL reaction. Very small intra coefficient of variations of 0.18-1.68% were observed for the multiplex RT-qPCR assays for the 18 respiratory viruses (Table 3). Similarly, inter coefficient of variation was 0.36-1.71%. These results indicate good consistency in the detection performance of the multiplex assays (Table 3).

Table 3.

Intra-assay and inter-assay reproducibility tests of the multiplex RT-qPCR assay.

Panels Concentration of standard plasmids (3000copies/25 µL reaction) Intra-coefficient of variation Inter-coefficient of variation
n Mean ± SD CV% n Mean ± SD CV%
Tube 1 IAV 10 27.72 ± 0.47 1.68% 3 27.33 ± 0.47 1.71%
IBV 10 27.33 ± 0.12 0.43% 3 27.33 ± 0.12 0.45%
RSV 10 20.27 ± 0.11 0.56% 3 20.17 ± 0.13 0.64%
Tube 2 HRV 10 26.35 ± 0.15 0.56% 3 26.37 ± 0.13 0.49%
HMPV 10 27.63 ± 0.11 0.40% 3 27.66 ± 0.10 0.36%
SARS-CoV-2 10 28.73 ± 0.15 0.51% 3 28.94 ± 0.20 0.68%
Tube 3 HAdV 10 27.99 ± 0.15 0.54% 3 27.9 ± 0.14 0.52%
HBoV 10 30.36 ± 0.16 0.54% 3 30.41 ± 0.15 0.50%
ICV 10 27.18 ± 0.11 0.42% 3 26.72 ± 0.36 1.34%
Tube 4 EV 10 24.64 ± 0.14 0.58% 3 24.7 ± 0.17 0.67%
HCoV-OC43 10 28.35 ± 0.34 1.19% 3 28.71 ± 0.38 1.33%
HCoV-29E 10 28.49 ± 0.15 0.53% 3 28.52 ± 0.18 0.62%
Tube 5 HPIV-1 10 26.76 ± 0.30 1.13% 3 26.89 ± 0.26 0.98%
HPIV-2 10 29.68 ± 0.25 0.85% 3 30.13 ± 0.42 1.38%
HPIV-3 10 28.57 ± 0.16 0.55% 3 28.51 ± 0.14 0.50%
Tube 6 HCoV-HKU1 10 27.8 ± 0.11 0.40% 3 27.47 ± 0.29 1.06%
HCoV-NL63 10 27.52 ± 0.33 1.18% 3 27.45 ± 0.30 1.10%
HPIV-4 10 26.4 ± 0.05 0.18% 3 26.32 ± 0.10 0.39%

3.6. Clinical performance of the new in-house multiplex RT-qPCR assay

To evaluate the clinical performance of the multiplex RT-qPCR assays, we performed a comparative analysis of the new multiplex RT-qPCR assays with the commercial kits in detection of six respiratory viruses (IAV, IBV, RSV, HRV, HAdV, and SARS-Cov-2) using 628 oropharyngeal swab samples. Relevant information regarding the specimens included in this study is summarized in Table 4. Furthermore, the Ct values of IBV, HAdV, and SARS-CoV-2 detected by the in-house RT-qPCR assay showed a significant linear correlation with those obtained using commercial RT-qPCR kits, with R² values of 0.82, 0.89, and 0.92, respectively (all P-values < 0.05; specifically, P < 0.01, P < 0.0001, and P < 0.05) (Figure 3).

Table 4.

Socio-demographics information of 628 patients for screening of respiratory viruses.

Patients (n=628)
Age, years (Mean ± SD) 35.6 ± 32.8
Range of age (years old) 2月-98岁
<5 109 (17.4%)
5-18 217(34.7%)
19-35 18 (2.9%)
36-60 68 (10.8%)
>60 216 (34.4%)
Gender
Male 332(52.9%)
Female 296 (47.1%)
Diseases
Upper Respiratory Tract Infection 195 (31.0%)
Respiratory Failure and Dyspnea 8 (1.3%)
Acute Bronchitis, Tonsillitis 115 (18.0%)
Non-severe Community-Acquired Pneumonia, Pulmonary Infection 189 (30.1%)
Severe Community-Acquired Pneumonia Complicated with Acute Exacerbation of Chronic Obstructive Pulmonary Disease 26(4.1%)
Healthy Individuals 6 (1.0%)
Other (Not Related to Lung Infection) 89 (14.2%)

Figure 3.

Two sets of analyses are shown: Panel A displays scatter plots for correlation analyses comparing commercial kits to new assays for viruses: IAV, IBV, SARS-CoV-2, RSV, HRV, and HAdV. Equations, R-squared, and p-values indicate correlation strength. Panel B shows Bland-Altman plots for each virus, illustrating the difference between commercial kits and new assays against their average Ct values. Each plot includes mean differences and limits of agreement.

Comparison of the new multiplex RT-qPCR assay with commercial kits for six respiratory viruses. (A) Ct value scatter plots. (B) Bland-Altman analyses.

Notably, the primary clinical manifestations of the patients from whom these specimens were collected included symptoms associated with upper respiratory tract infection (URTI) and pneumonia. Of these samples, 532 (84.7%) samples were detected as positive for any one of the six viruses by the new in-house multiplex RT-qPCR assay and the commercial kits (Table 5). Both methods had consistent detection rates for RSV (3.1%), IBV (1.1%) and SARS-CoV-2 (0.6%). The commercial kit detected 309 (49.2%) and 92 (14.6%) positives for IAV and HRV, respectively, with one more positive than the new in-house method for each virus. One IAV-positive sample only by the commercial kit had a Ct value of 38, and failed to be confirmed by a nested RT-PCR assay, suggesting a very low viral load. One HRV-positive sample only by the commercial kit was detected positive to EV and further confirmed to EVA71 by Sanger sequencing (Supplementary Table S4). We performed in-silico cross-reactivity analysis by aligning HRV-specific primers and probe with 18 representative strains of five enteroviruses, including 4 EVA71, 4 EVD68, 3 Coxsackievirus A16 (CVA16), 4 CVA6, and 3 echovirus 30 (accession numbers listed in Supplementary Table S5). No significant homologous sequences to the HRV-primers and probe were found in the genomes of the enteroviruses, supporting high specificity of the in-house assay and low possibility of cross-reactivity to enteroviruses. The in-house method detected 138 positive samples for HAdV, obviously more than 132 by the commercial kit. All 6 HAdV-positive samples only by the in-house method were confirmed by Sanger sequencing. Then we determined true positive and true negative samples for each virus based on the consistent detection results by both the in-house assay and commercial kits, and confirmation results for inconsistent samples experiments by Sanger sequencing, and then assessed the performance of the in-house method. The results showed that the new in-house method exhibited 100% specificity and sensitivity in detection of IAV, IBV, RSV, HAdV and SARS-CoV-2 (Table 6). The new in-house assay demonstrated a better performance for HADV and HRV than the commercial kit, but a slightly lower sensitivity for IAV than the commercial kit.Co-infection of respiratory viruses are frequent in patients with ARTIs. The in-house multiplex assay detected 33 (5.3%) dual infections and 2 triplex infections among above six respiratory viruses, respectively, more than 30 (4.8%) and 1 (0.16%) by the commercial kits (Table 5). These results indicate that new in-house method demonstrated have better performance than the commercial RT-qPCR kits in detection of respiratory viruses especially for the detection of coinfections.

Table 5.

Detection of common respiratory viruses in 628 clinical samples by the new in-house and commercial RT-qPCR assays.

Infection pattern Viruses New multiplex RT-qPCR assays Commercial RT-qPCR kits
Pos.(n) PR(%) Pos.(n) PR(%)
Overall IAV 308 49.0% 309# 49.2%
IBV 7 1.1% 7 1.1%
RSV 19 3.0% 19 3.0%
HRV 91 14.5% 92* 14.6%
HAdV 138 22.0% 132 21.0%
SARS-Cov-2 4 0.6% 4 0.6%
Sum 1: Any one of six viruses 532 84.7% 532 84.7%
HPIV-1 4 0.6% NA NA
HPIV-2 1 0.2% NA NA
HPIV-3 2 0.3% NA NA
HPIV-4 1 0.2% NA NA
EV 4 0.6% NA NA
HCoV-229E 1 0.2% NA NA
HBoV 1 0.2% NA NA
HMPV 1 0.2% NA NA
Sum 2: Any one of above 14 respiratory viruses 535 85.2% NA NA
Dual infection IAV、HRV 16 2.6% 17¶ 2.7%
HAdV、HRV 7 1.1% 6 1.0%
IAV、HAdV 4 0.6% 3 0.5%
HAdV、RSV 2 0.3% 1 0.2%
HRV、RSV 1 0.2% 1 0.2%
IAV、RSV 1 0.2% 1 0.2%
Sum 3: any one of above dual infection patterns 33 5.3% 30 4.8%
HAdV、HPIV-1 1 0.2% NA NA
HAdV、HPIV-4 1 0.2% NA NA
HRV、EV 3 0.5% NA NA
IAV、HPIV-1 2 0.3% NA NA
IAV、HPIV-3 1 0.2% NA NA
RSV、HBoV 1 0.2% NA NA
IAV、HCoV-229E 1 0.2% NA NA
Sum 4: all dual infection patterns 41 6.5% NA NA
Triplex infection IAV、HAdV、HRV 2 0.3% 1 0.2%
IAV、HAdV、HMPV 1 0.2% NA NA
Total 3 0.5%

#The Ct value by the commercial kit was 38, a value very close to the cutoff of positive. *One HRV positive sample identified by the commercial kit was detected positive for EV by the new in-house RT-qPCR assay and further confirmed as EVA71 by Sanger sequencing and blast search (Supplementary Table S4). One dual infection with IAV and HRV by the commercial kit was detected as triplex infection with IAV, HRV and HAdV by the new in-house RT-qPCR assay with confirmation of HAdV coinfection by Sanger sequencing and blast search. NA, not applicable. Pos., positive. PR, prevalence rate.

Table 6.

Comparison between the commercial RT-qPCR kits and the new multiplex RT-qPCR assay.

Viruses Methods TP TN FP FN Se (%) Sp (%) ACC (%)
IAV Commercial kit 309 319 0 0 100% 100% 100%
In-house 308 320 0 1 99.7% 100% 99.8%
IBV Commercial kit 7 621 0 0 100% 100% 100%
In-house 7 621 0 0 100% 100% 100%
RSV Commercial kit 19 609 0 0 100% 100% 100%
In-house 19 609 0 0 100% 100% 100%
HRV Commercial kit 91 536 1 0 100% 99.8% 99.8%
In-house 91 537 0 0 100% 100% 100%
HAdV Commercial kit 132 490 0 6 95.6% 100% 99.0%
In-house 138 490 0 0 100% 100% 100%
SARS-CoV-2 Commercial kit 4 624 0 0 100% 100% 100%
In-house 4 624 0 0 100% 100% 100%

TN, true negative; FP, false positive; FN, false negative; Se, sensitivity; Sp, specificity; Acc, accuracy.

Apart from above 6 respiratory viruses, the new in-house RT-qPCR assay also detected 15 (2.4%) positive for other respiratory viruses, including 4 HPIV-1, 1 HPIV-2, 2 HPIV-3, 1 HPIV-4, 4 EV, 1 HCoV-229E, 1 HBoV, and 1 HMPV. Of 15 positive samples, 11 were coinfected with one or two of IAV, HAdV, HRV, and RSV.

In summary, 14 respiratory viruses were detected among 536 (85.4%) of 628 oropharyngeal samples. IAV was the predominant respiratory virus (49.2%) circulating in this cohort, followed by HAdV (22.0%) and HRV (14.5%). It was not surprising that the top three dominant respiratory viruses were involved in almost all 44 coinfection events except one dual infection by RSV and HBoV, including IAV in 27 (61.4%), HADV in18 (40.9%) and HRV in 29 (65.9%). Furthermore, we found that HRV (29/91, 31.9%) had higher possibility of coinfecting with other respiratory viruses than HAdV (18/138, 13.0%) and IAV (27/309, 8.7%).

4. Discussion

In this study, we developed an in-house multiplex RT-qPCR assay for detection of 18 common respiratory viruses, encompassing influenza A/B/C viruses (IAV/IBV/ICV), RSV, HRV, EV, HAdV, SARS-CoV-2, four human coronaviruses (HCoV-HKU1, HCoV-229E, HCoV-NL63, and HCoV-OC43), HMPV, HBoV, and four human parainfluenza viruses (HPIV-1/2/3/4). The multiplex assay demonstrated high analytical sensitivity with LOD of 5–75 copies/25 μL reaction, great linear relationship (R² ≥ 0.9944) between log copy numbers and Ct values, and high analytical reproducibility with intra-assay CV of 0.18–1.68%, and inter-assay CV of 0.36–1.71%. No cross-reactivity among these 18 viruses and no non-specific amplification for other human pathogens indicate good specificity. Clinical evaluation with 628 clinical samples showed a better performance of the in-house multiplex RT-qPCR assay than the commercial kits at least for IAV, IBV, RSV, HRV, HAdV and SARS-CoV-2.

qPCR is widely recognized as the gold standard for molecular diagnosis due to its high sensitivity, specificity, and capacity to enable multiplex detection. Developments of multiplex qPCR assay for various pathogens play a crucial role in the prevention and control of infectious diseases. An inherent challenge for the development of a multiplex qPCR assay is mutual interference and non-specific hybridization among the primers and probes targeting different pathogens, which may reduce detection sensitivity and cause cross-reactivity, therefore leading to false results. In this study, we implemented three strategies to address these issues. First, 2–7 sets of primers and probes were designed for each virus and then subjected to preliminary screening experiments to obtain the primer and probe set with the highest amplification efficiency and high specificity with no non-specificity amplification in non-template control. Second, we performed sequence alignment to assess the identity of each primer and probe to the viral target using all available viral sequences from GenBank. To improve the coverage to highly heterogeneous viral sequences, degenerate bases were incorporated to corresponding primers and probes (e.g. the primers for HRV and HAdV) to minimize mismatches with viral variants. Furthermore, AutoDimer was used to evaluate the thermodynamic stability of the primers and probes and Hetero-Dimer Analysis was performed to exclude the non-specific hybridization among the primers and probes. Third, minor groove binder (MGB) was used to improve the hybridization affinity and specificity of probe (e.g. the probe for HAdV). These strategies enabled the same detection sensitivity (almost the same LOD) and specificity of the multiplex RT-qPCR assays to the singleplex assays.

The majority of ARTIs are caused by a diverse array of respiratory viruses, many of which have multiple genotypes or subtypes. For instance, HRV comprise over 150 subtypes, and HMPV has two major genotypes (A, B). The in-house multiplex RT-qPCR assay covers 18 common respiratory viruses, and show good specificity to detect and distinguish these viruses. Both HRV and EV belong to the family Picornaviridae, and share high genetic similarity especially at the 5’ untranslated region (5’UTR) of their genomes. As the relatively conserved region in viral genome, 5’-UTRs of HRV and EV are often used as a target genomic region for designing primers and probe, thus resulting in cross-amplification between both viruses (Ng et al., 2025). Of 628 clinical samples, one EV-positive sample was misclassified as positive for HRV by the commercial assay, but was correctly identified by the in-house assay and further confirmed by Sanger sequencing, indicating the in-house RT-qPCR assay has high specificity to detect and distinguish these respiratory viruses.

Coinfections with two or more respiratory viruses are often observed in children with ARTIs, and are associated with more severe outcomes, including increased hospitalization duration, respiratory failure, and mortality (Le Glass et al., 2021; Adams et al., 2022; Abo et al., 2023; Cao et al., 2024; de Hoog et al., 2025). In this study, we detected 44 (7.0%) coinfection cases, including 41 dual infections and 3 triple infections. The most commonly observed coinfection pattern was dual infections with IAV and HRV (16 cases, 2.6%), followed by dual infections with HAdV and HRV (7 cases, 1.1%) and IAV and HAdV (4 cases, 0.6%). A multicenter study in China (Li et al., 2024) reported that IAV and HRV co-infection was associated with a 2.3-fold higher risk of severe pneumonia in adults, likely due to synergistic immune dysregulation. HRV infections can induce epithelial damage that facilitates IAV entry, and the latter can suppress type I interferon responses, promoting HRV replication (Di Maio et al., 2025). Furthermore, HRV (29/91, 31.9%) appeared to have a very high possibility of coinfecting with other respiratory viruses. Most respiratory viral infections cause mild and self-limiting diseases. HRV co-infections with other respiratory viruses (e.g. HCoV-229E) can aggravate the disease (Madewell et al., 2023), such as increased cough severity, longer symptom duration (≥7 days), and higher rates of hospital admission in children and older adults. Therefore, surveillance of HRV coinfection is very important for clinical treatment and management of patients with ARTIs especially in the spring and winter seasons when influenza and other respiratory viruses are prevalent. Compared with commercial RT-qPCR kits, the in-house multiplex RT-qPCR assay demonstrates comparable detection sensitivity, but offers broader coverage of respiratory viruses, lower assay costs, and superior capability for identifying respiratory virus coinfections (Supplementary Table S6). Therefore, the new in-house multiplex RT-qPCR assay can be used as a robust tool for surveillance of respiratory viral infections.

This study has three limitations. First, the in-house RT-qPCR assay focuses exclusively on respiratory viruses and does not cover bacterial and atypical bacterial etiologies of ARTIs (e.g. Streptococcus pneumoniae, Mycoplasma pneumoniae, Chlamydia pneumoniae). S. pneumoniae and M. pneumoniae are the top bacterial causes of pneumonia in China, and their co-infection with respiratory viruses (e.g. IAV) can exacerbate disease severity. The RT-qPCR assay can be updated to contain these bacterial pathogens in future. Second, IAV, HAdV and HRV are among the top respiratory viruses circulating in China. However, the three viruses are separately tested in three independent reactions. A new combination pattern containing the three viruses in a single reaction may facilitate the monitoring of these respiratory viruses. Third, the clinical performance of the in-house RT-qPCR assay was assessed only for six respiratory viruses (i.e., SARS-CoV-2, IAV, IBV, RSV, HRV, and HAdV), but not for other respiratory viruses. The main reason is lack of commercial kits approved by the National Medical Products Administration (NMPA) for other respiratory viruses. Furthermore, analytical specificity of the new assay to HPIV-4, HCoV-NL63, HCoV-NL63, and ICV was determined by plasmids containing target gene fragments, rather than full-length genome due to lack of virus isolates and clinical samples positive for these viruses. Experiments should be performed to further determine the specificity and clinical performance of the new assay in future.

In conclusion, the in-house multiplex RT-qPCR assay offers a sensitive, specific, and comprehensive tool for detecting 18 common respiratory viruses. Its key strengths include high detection sensitivity (LOD: 5–75 copies per 25 μL reaction), specificity, and reproducibility, broad coverage of respiratory viruses, and excellent performance in clinical samples. Furthermore, we reported high prevalences of IAV, HAdV and HRV in Taizhou, Jiangsu, China, and a high potential of HRV coinfection with other respiratory viruses.

Acknowledgments

We thank Liejun Li at Guangzhou Hybribio Medical Ltd. for his technical support.

Funding Statement

The author(s) declared financial support was received for this work and/or its publication. The author(s) declared that this work received funding from Chaozhou Hybribio Biochemistry Ltd. The funder was not involved in the study design, collection, analysis, interpretation of data, the writing of this article or the decision to submit it for publication.

Footnotes

Edited by: Jozef Dingemans, Jessa Hospital, Belgium

Reviewed by: Eduardo Becerril, National Institute of Respiratory Diseases-Mexico (INER), Mexico

Christian von Wintersdorff, Maastricht University Medical Centre, Netherlands

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material. Further inquiries can be directed to the corresponding authors.

Ethics statement

The studies involving humans were approved by Taizhou Fourth People’s Hospital. The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation in this study was provided by the participants’ legal guardians/next of kin.

Author contributions

AX: Data curation, Formal Analysis, Investigation, Writing – original draft. LX: Funding acquisition, Writing – review & editing. LZ: Data curation, Investigation, Writing – review & editing. XZ: Investigation, Resources, Writing – review & editing. ZW: Investigation, Resources, Supervision, Writing – review & editing. J-HW: Supervision, Writing – review & editing. CZ: Conceptualization, Methodology, Project administration, Supervision, Writing – review & editing.

Conflict of interest

The authors AX, LX, and LZ are the employees at Guangzhou Hybribio Medical Ltd.

The remaining author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The author(s) declare that Generative AI was not used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcimb.2025.1727780/full#supplementary-material

Table1.docx (1.6MB, docx)

References

  1. Abo Y.-N., Oliver J., McMinn A., Osowicki J., Baker C., Clark J. E., et al. (2023). Increase in invasive group A streptococcal disease among Australian children coinciding with northern hemisphere surges. Lancet Regional Health - Western Pacific 41, 100873. doi:  10.1016/j.lanwpc.2023.100873, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Adams K., Tastad K. J., Huang S., Ujamaa D., Kniss K., Cummings C., et al. (2022). Prevalence of SARS-coV-2 and influenza coinfection and clinical characteristics among children and adolescents aged <18 years who were hospitalized or died with influenza — United states 2021–22 influenza season. Morbidity Mortality Weekly Rep. 71, 1589–1596. doi:  10.15585/mmwr.mm7150a4, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Azar M. M., Landry M. L., Kraft C. S. (2018). Detection of influenza A and B viruses and respiratory syncytial virus by use of clinical laboratory improvement amendments of 1988 (CLIA)-waived point-of-care assays: a paradigm shift to molecular tests. J. Clin. Microbiol. 56, e00367–18. doi:  10.1128/jcm.00367-18, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Cao N., Cai Y., Huang X., Jiang H., Huang Z., Xing L., et al. (2024). Inhibition of influenza A virus and SARS-CoV-2 infection or co-infection by griffithsin and griffithsin-based bivalent entry inhibitor. mBio 15, 1–16. doi:  10.1128/mbio.00741-24, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. de Hoog M. L. A., Hauser-van Westrhenen E., Winkel A., de Jong M. D., van Houten M. A., van Lelyveld S. F. L., et al. (2025). Impact of co-infection with SARS-CoV-2 and other respiratory viruses on illness: Pooled analyses of 11 COVID-19 cohorts. J. Infect. 90, 9. doi:  10.1016/j.jinf.2025.106501, PMID: [DOI] [PubMed] [Google Scholar]
  6. Di Maio V. C., Scutari R., Mastropaolo M., Colagrossi L., Linardos G., Coltella L., et al. (2025). Viral burden of respiratory syncytial virus and viral coinfections as factors regulating paediatric disease severity. Viruses 17, 1236. doi:  10.3390/v17091236, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Frediansyah A., Tiwari R., Sharun K., Dhama K., Harapan H. (2021). Antivirals for COVID-19: A critical review. Clin. Epidemiol. Global Health 9, 90–98. doi:  10.1016/j.cegh.2020.07.006, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Ginocchio C. C., McAdam A. J. (2011). Current best practices for respiratory virus testing. J. Clin. Microbiol. 49, S44–S48. doi:  10.1128/jcm.00698-11 [DOI] [Google Scholar]
  9. Horemheb-Rubio G., Eggeling R., Schmeiβer N., Pfeifer N., Lengauer T., Gärtner B. C., et al. (2022). Respiratory viruses dynamics and interactions: ten years of surveillance in central Europe. BMC Public Health 22, 1167. doi:  10.1186/s12889-022-13555-5, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Jiang X.-W., Huang T.-S., Xie L., Chen S.-Z., Wang S.-D., Huang Z.-W., et al. (2022). Development of a diagnostic assay by three-tube multiplex real-time PCR for simultaneous detection of nine microorganisms causing acute respiratory infections. Sci. Rep. 12, 13306. doi:  10.1038/s41598-022-15543-6, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Lee J. S., Ahn J. J., Kim S. J., Yu S. Y., Koh E. J., Kim S. H., et al. (2021). POCT detection of 14 respiratory viruses using multiplex RT-PCR. BioChip J. 15, 371–380. doi:  10.1007/s13206-021-00037-w, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  12. Le Glass E., Hoang V. T., Boschi C., Ninove L., Zandotti C., Boutin A., et al. (2021). Incidence and outcome of coinfections with SARS-coV-2 and rhinovirus. Viruses 13, 2528. doi:  10.3390/v13122528, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. Li Y., Yu J., Wang Y., Yi J., Guo L., Wang Q., et al. (2024). Cocirculation and coinfection of multiple respiratory viruses during autumn and winter seasons of 2023 in Beijing, China: A retrospective study. J. Med. Virol. 96, e29602. doi:  10.1002/jmv.29602, PMID: [DOI] [PubMed] [Google Scholar]
  14. Li Z.-J., Zhang H.-Y., Ren L.-L., Lu Q.-B., Ren X., Zhang C.-H., et al. (2021). Etiological and epidemiological features of acute respiratory infections in China. Nat. Commun. 12, 5026. doi:  10.1038/s41467-021-25120-6, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. Liu J., Kabir F., Manneh J., Lertsethtakarn P., Begum S., Gratz J., et al. (2014). Development and assessment of molecular diagnostic tests for 15 enteropathogens causing childhood diarrhoea: a multicentre study. Lancet Infect. Dis. 14, 716–724. doi:  10.1016/s1473-3099(14)70808-4, PMID: [DOI] [PubMed] [Google Scholar]
  16. Ma S., Lai X., Chen Z., Tu S., Qin K. (2020). Clinical characteristics of critically ill patients co-infected with SARS-CoV-2 and the influenza virus in Wuhan, China. Int. J. Infect. Dis. 96, 683–687. doi:  10.1016/j.ijid.2020.05.068, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  17. Madewell Z. J., Wang L. P., Dean N. E., Zhang H. Y., Wang Y. F., Zhang X. A., et al. (2023). Interactions among acute respiratory viruses in Beijing, Chongqing, Guangzhou, and Shanghai, China 2009–2019. Influenza Other Respir. Viruses 17, e13212. doi:  10.1111/irv.13212, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Mahony J. B. (2008). Detection of respiratory viruses by molecular methods. Clin. Microbiol. Rev. 21, 716–747. doi:  10.1128/cmr.00037-07, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  19. Mahony J. B. (2010). Nucleic acid amplification-based diagnosis of respiratory virus infections. Expert Rev. Anti-Infective Ther. 8, 1273–1292. doi:  10.1586/eri.10.121, PMID: [DOI] [PubMed] [Google Scholar]
  20. Mancini F., Barbanti F., Scaturro M., Fontana S., Di Martino A., Marsili G., et al. (2021). Multiplex real-time reverse-transcription polymerase chain reaction assays for diagnostic testing of severe acute respiratory syndrome coronavirus 2 and seasonal influenza viruses: A challenge of the phase 3 pandemic setting. J. Infect. Dis. 223, 765–774. doi:  10.1093/infdis/jiaa658, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  21. Marrie T. J., Durant H., Yates L. (1989). Community-acquired pneumonia requiring hospitalization 5-year prospective study. Rev. Infect. Dis. 11, 586–599. doi:  10.1093/clinids/11.4.586, PMID: [DOI] [PubMed] [Google Scholar]
  22. Ng T. F. F., Nix W. A., Rogers S. L., Emery B., Chern S.-W., Butler K., et al. (2025). Type-specific EV-D68 real-time RT-PCR assay for the detection of all extant enterovirus D68 strains. J. Clin. Microbiol. 63, 1–11. doi:  10.1128/jcm.01492-22, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  23. Russell C. J., Liu J., Ai H., Xiong Y., Li F., Wen Z., et al. (2015). Prevalence and correlation of infectious agents in hospitalized children with acute respiratory tract infections in central China. PloS One 10, e119170. doi:  10.1371/journal.pone.0119170, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  24. Sanghavi S. K., Bullotta A., Husain S., Rinaldo C. R. (2011). Clinical evaluation of multiplex real-time PCR panels for rapid detection of respiratory viral infections. J. Med. Virol. 84, 162–169. doi:  10.1002/jmv.22186, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  25. Swets M. C., Russell C. D., Harrison E. M., Docherty A. B., Lone N., Girvan M., et al. (2022). SARS-CoV-2 co-infection with influenza viruses, respiratory syncytial virus, or adenoviruses. Lancet 399, 1463–1464. doi:  10.1016/s0140-6736(22)00383-x, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  26. Vemula S., Zhao J., Liu J., Wang X., Biswas S., Hewlett I. (2016). Current approaches for diagnosis of influenza virus infections in humans. Viruses 8, 96. doi:  10.3390/v8040096, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  27. Vos T., Lim S. S., Abbafati C., Abbas K. M., Abbasi M., Abbasifard M., et al. (2020). Global burden of 369 diseases and injuries in 204 countries and territories 1990–2019: a systematic analysis for the Global Burden of Disease Study 2019. Lancet 396, 1204–1222. doi:  10.1016/s0140-6736(20)30925-9, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  28. Yüce M., Filiztekin E., Özkaya K. G. (2021). COVID-19 diagnosis —A review of current methods. Biosensors Bioelectronics 172, 112752. doi:  10.1016/j.bios.2020.112752, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  29. Zhang N., Wang L., Deng X., Liang R., Su M., He C., et al. (2020). Recent advances in the detection of respiratory virus infection in humans. J. Med. Virol. 92, 408–417. doi:  10.1002/jmv.25674, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Table1.docx (1.6MB, docx)

Data Availability Statement

The original contributions presented in the study are included in the article/Supplementary Material. Further inquiries can be directed to the corresponding authors.


Articles from Frontiers in Cellular and Infection Microbiology are provided here courtesy of Frontiers Media SA

RESOURCES