What If Each Voxel Were Measured With a Different Diffusion Protocol?

Santiago Coelho 1,, Gregory Lemberskiy 1, Ante Zhu 2, Hong‐Hsi Lee 3, Nastaren Abad 2, Thomas K F Foo 2, Els Fieremans 1, Dmitry S Novikov 1
PMCID: PMC12850600  PMID: 41387990

ABSTRACT

Purpose

Expansion of diffusion MRI (dMRI) both into the realm of strong gradients and into accessible imaging with portable low‐field devices brings about the challenge of gradient nonlinearities. Spatial variations of the diffusion gradients make diffusion weightings and directions non‐uniform across the field of view, and deform perfect shells in the q‐space designed for isotropic directional coverage. Such imperfections hinder parameter estimation: Anisotropic shells hamper the deconvolution of the fiber orientation distribution function (fODF), while brute‐force retraining of a nonlinear regressor for each unique set of directions and diffusion weightings is computationally inefficient.

Methods

Here, we propose a protocol‐independent parameter estimation (PIPE) method that enables fast parameter estimation for the most general case where each voxel is measured with a different protocol in q‐space. PIPE applies to any spherical convolution‐based dMRI model, irrespective of its complexity, which makes it suitable both for white and gray matter in the brain or spinal cord, and for other tissues where fiber bundles have the same properties (fiber response) within a voxel, but are distributed with an arbitrary fODF. We also derive a parsimonious representation that isolates isotropic and anisotropic effects of gradient nonlinearities on multidimensional diffusion encodings.

Results

Applied to in vivo human MRI with linear tensor encoding on a high‐performance gradient system, PIPE evaluates fiber response and fODF parameters for the whole brain in the presence of significant gradient nonlinearities in under 3 min.

Conclusions

PIPE enables fast parameter estimation in the presence of arbitrary gradient nonlinearities, eliminating the need to arrange dMRI in shells or to retrain the estimator for different protocols in each voxel. PIPE applies to any model based on a convolution of a voxel‐wise fiber response and fODF, and data from varying b‐values, diffusion/echo times, and other scan parameters.

Keywords: diffusion MRI, gradient nonlinearities, machine learning, microstructure, spherical convolution

1. Introduction

The ability to non‐invasively probe the random motion of water molecules within tissues makes diffusion MRI (dMRI) sensitive to tissue micro‐architecture [1, 2, 3, 4, 5, 6]. Typical experimental settings detect such motion at a scale of 1–50 micrometers, making dMRI sensitive, and possibly specific, to disease processes originating at this scale, and thereby provide biomarkers of pathological processes [7, 8, 9, 10, 11]. Biophysical models of diffusion have the potential to deliver the specificity that would aid in early diagnosis [9]. This prompts the development of dMRI acquisitions and models that strive not only for sensitivity but also for specificity [11].

The development of ultra‐high gradients for tissue microstructure imaging [12, 13, 14, 15, 16], and the availability of cost‐effective, portable scanners [17, 18, 19, 20], has created unprecedented opportunities for dMRI [21]. At the high‐performance end, in vivo dMRI exhibits an improved resolution [22], as well as the sensitivity to axon diameters [23], diffusion time‐dependence [24], water exchange [25], and the magnetization localized near cell walls [26]. At the opposite end of the spectrum, ultra‐low‐field systems [27], gaining traction due to their affordability and accessibility, offer the possibility of diffusion tensor imaging and fiber tracking [28, 29].

Head‐only MRI systems offer the opportunity to optimize gradient performance at both high‐performance and low field ends. These gradients are designed to have good linearity in about 25 cm‐diameter sphere, compared to 40–50 cm for whole‐body gradients [22]. Therefore, in the peripheral parts of the brain, gradient nonlinearities are non‐negligible [21] (Figure 1). Away from the bore isocenter, applied diffusion weightings deviate from the nominal settings; this deviation further depends on the gradient direction [30]. Thus, actual diffusion weightings and measurement protocols can vary significantly across the field of view (FOV), and may even not consist of conventional shells in the diffusion q‐space designed for isotropic directional sampling. The anisotropy of shells, and the spatially varying diffusion weightings and directions, make diffusion processing [31] and parameter estimation of nonlinear models challenging.

FIGURE 1.

FIGURE 1

Deformation of b‐value and direction by the gradient nonlinearities across the FOV based on the measured Lij(r) for a high‐performance head‐only system [12]. Linear tensor encoding (LTE) weighting b(r,g^), Equation (9), varies differently across the FOV depending on its nominal direction g^. (a) An example of the deformation L(r)g^ of a generic direction g^, and the corresponding dimensionless ratio b(r,g^)/b (color). (b) The mean over all possible LTE directions g^𝕊2 of the ratio b(r,g^)/b, Equation (9), represented by the isotropic component N0 of N(r), Equation (10). The brain contour of the volunteer is drawn for reference. (c) Relative directional variations of the LTE b‐value (9), characterized by the standard deviation of N(r,g^) over all possible LTE directions g^ (square root of the variance (11)). This quantity is determined by the anisotropic part N(2)(r) of N(r) and is proportional to the invariant N2(r), Equation (12).

As the logarithm of the DTI signal linearly depends on the diffusion encoding, estimated diffusion tensors can be corrected for gradient nonlinearities a posteriori in each voxel [30]. For essentially nonlinear multi‐compartment models, however, post‐hoc parameter corrections are challenging, and the entire estimation process must be reconsidered. In this work, we focus on the overarching spherical convolution framework for modeling diffusion in brain tissue [2, 32, 33, 34, 35], which assumes that each voxel contains a collection of identical fiber bundles with arbitrary orientations described by the fiber Orientation Distribution Function (fODF), Figure 2. The dMRI signal can be expressed as a convolution over the unit sphere 𝕊2 of the fiber bundle response function (the kernel) and the fODF. The generality of this framework allows an arbitrary number of Gaussian or non‐Gaussian diffusion compartments, possibility of water exchange, diffusion time‐, inversion time‐, or echo time‐dependence, and so forth. This has enabled its application to both white matter (WM) [36, 37, 38, 39, 40, 41, 42, 43] and gray matter (GM) [44, 45, 46].

FIGURE 2.

FIGURE 2

PIPE central assumption: dMRI signal is a spherical convolution of an axially‐symmetric fiber response (kernel 𝒦), characterized by scalar model parameters ξ, and an arbitrary fODF 𝒫(n^) characterized by its spherical harmonics coefficients pm.

Conventional fiber responses depend nonlinearly on microstructural parameters, making it difficult to estimate these parameters robustly using standard maximum likelihood approaches. Additionally, many of these regression‐based techniques assume that the data is acquired in spherical shells in the q‐space, to ensure isotropic coverage and a robust fODF deconvolution. Gradient nonlinearities can distort these shells, causing the actual diffusion weighting to vary with the direction and the spatial location across the imaging volume. At typical signal‐to‐noise ratios (SNR), data‐driven regression methods have been shown to achieve lower mean‐squared error (MSE) than likelihood‐based methods [47, 48]. However, retraining such machine learning (ML)‐based estimators independently for each voxel to accommodate protocol variations would result in a prohibitively high computational cost.

Here we propose a protocol‐independent parameter estimation (PIPE) ML framework that enables fast parameter estimation of convolution‐based biophysical models, for which every voxel can be acquired with a different dMRI protocol, thereby enabling applications on MRI systems with arbitrary gradient nonlinearities [49]. PIPE applies to any biophysical model where fiber bundles in a given voxel have the same properties, but are distributed with an arbitrary fODF. Both the fiber response and fODF can vary between voxels, thereby ensuring the applicability of PIPE to practically all white matter and gray matter dMRI models [36, 37, 38, 39, 40, 41, 42, 43, 44, 45, 46].

PIPE relies on the singular value decomposition (SVD) based separation between the protocol and model parameters in any model expression [50], such that the training is done only once on the model‐parameter part, while the protocol‐dependent part is allowed to vary from voxel to voxel. Furthermore, the data is not constrained to be acquired in any fashion (such as shells), and there are no limits on the gradient nonlinearities as long as the protocol is well defined for each voxel. PIPE incorporates varying diffusion times, echo times, and other scan parameters within the protocol, and is naturally suited for the most widespread linear tensor encoding (attained, e.g., via PGSE [51]), being generalizable to any axially symmetric multidimensional diffusion encodings.

In what follows, we will describe the method, evaluate its accuracy and precision via synthetic noise propagation, and apply it to in vivo human brain PGSE‐dMRI acquired on a high‐performance system with notable gradient nonlinearities.

2. Methods

2.1. Machine Re‐Learning?

Why should one worry about having different sets of diffusion directions and b‐values in every voxel? Theoretically, one can use an unbiased estimator for each voxel separately; after all, a well‐validated model is the best way to factor out the differences in protocol [11]. Conventional maximum likelihood estimators [52] are asymptotically unbiased and can be employed. Given the dMRI noise model, maximizing the likelihood across all measurements defines a mapping from the space of noisy measurements to model parameters.

While statistically rigorous, maximum‐likelihood estimation suffers from a few drawbacks. First, finding a global maximum for hundreds of measurements and at least a few model parameters is computationally intense for strongly nonlinear dMRI models, and must be performed in each voxel separately. Second, to speed up the search for the global likelihood maximum, one has to initialize the search within the right “basin of attraction” around the ground truth, which is a priori unknown; this can make the outcome dependent on the initialization. Third, even with the appropriate initialization, the strongly nonlinear nature of diffusion models and a relatively low SNR make the estimated parameters veer off quite far from the ground truth along the “shallow directions” in the likelihood landscape [53], often into unphysical domains; this is typically cured by introducing priors or regularization terms into the objective function [54], making the outcome dependent on their assumptions.

Alternatively, data‐driven supervised ML regressions “learn” the mapping from noisy measurements to model parameters. The mapping is provided by a sufficiently flexible regressor, such as a neural network or a high‐order multivariate polynomial, based on the training data generated with the forward signal and noise models of interest. The regression coefficients are then chosen such that mean‐squared error (MSE) or other error metrics are minimized over the training data [41, 55]. While the training can be slow and computationally intense, the estimation is typically fast, with the learned regression evaluated once per voxel.

For realistic dMRI voxel‐wise SNR (∼25–100 for unweighted images), ML regressions outperform conventional estimation approaches, even though the estimated parameters may be biased by the training set at low SNR [47]. Multiple research groups have applied variations of these regressions to various biophysical models aimed at data from clinical scanners [41, 44, 56, 57, 58]. However, applying ML regressions in the case of significant gradient nonlinearities is nontrivial: Each voxel within the field of view may end up with a distinct set of diffusion weightings and directions. Thus, the dimensionality of the mapping one needs to learn increases dramatically: for example, in a FOV of 100×100×100 voxels, one ends up with 1 million unique protocols. Even when the gradient nonlinearities are slowly varying, one would have to “re‐learn” the mapping for thousands of distinct protocols for local patches with sufficiently different sets of gradient directions and weightings. Naively training an ML estimator for each voxel would take too long to compute. The PIPE framework addresses this challenge.

2.2. Gradient Nonlinearities

Nonuniform magnetic field gradients introduce artifacts in dMRI. This causes not only image distortions [59] but also spatially varying errors in the direction and strength of the applied diffusion encoding [30]. Thus, the actual gradient G(r)=Bz(r) affecting the local Larmor frequency (and responsible for the diffusion encoding) differs both in its strength and orientation from the nominal gradient G. Since the Maxwell's equations are linear, the deviations from the nominal G can be parametrized by the gradient coil tensor field [30] Lij(r), which relates the actual to the nominal gradient in the coordinate system of the gradient coils:

Gi(r)=Lij(r)Gj, (1)

where the summation over repeated indices is implied, and

Lij(r)Gi(r)Gj=Bz(r)xi·1Gj. (2)

Introducing the wave vector q(t) as the antiderivative,

qi(t)0tdτγGi(τ)=Lijqj(t),qj=0tdτγGj(τ),

where γ is gyromagnetic ratio, one obtains the diffusion‐encoding B‐tensor [60, 61, 62, 63] as a spatially varying field [30]

B(r)dtq(t)q(t)=L(r)BLt(r) (3)

in terms of the nominal B‐tensor [64, 65]

B=dtq(t)q(t). (4)

The rank of B reflects how many dimensions of the diffusion process are being simultaneously probed; rankB>1 means probing the diffusion along more than one dimension. Let us consider linear tensor encoding (LTE, rankB=1), for which pulsed gradient spin echo (PGSE) [51] is the most common gradient waveform. The nominal LTE tensor B=bg^g^, defined by its nominal unit direction |g^|=1 and b‐value b, becomes

B(r)=bL(r)g^g^Lt(r). (5)

Remarkably, the LTE rank remains unchanged by the gradient nonlinearities: rankB(r)=1 for any L(r), when rankB=1. Both the LTE direction Lg^ and the b‐value

br,g^=trB(r)=bg^tN(r)g^,N(r)=Lt(r)L(r), (6)

get modified by L(r). Here, we introduced the nonlinearity tensor field Nij(r)=Lki(r)Lkj(r), which at every point r is a symmetric 3×3 matrix, determined by 6 independent parameters, instead of 9 parameters in Lij. It admits the irreducible decomposition (we follow the notations of Reference [66]):

Nij=Nij(0)+Nij(2)N0δij+m=22N2m𝒴ij2m. (7)

Decomposition (7) separates the isotropic part N(0), defined by a single parameter N0=13trN, from its anisotropic counterpart N(2), defined by five N2m. Here, δij is Kronecker's delta, and 𝒴ij2m are the symmetric trace‐free basis tensors [66, 67], such that 𝒴ij2mĝiĝj=Y2m(g^) for a unit vector g^𝕊2 yield complex spherical harmonics (SH) Y2m(g^). We employ Racah normalization for SH and for the basis tensors 𝒴i1im, as in Reference [66]:

𝕊2dn^Ym(n^)Ym(n^)=δδmm2+1,dn^=dΩ4π, (8)

where dΩ=sinθdθdϕ is the standard measure on the unit sphere 𝕊2, such that 𝕊2dn^1, and Y00(n^)1. As a result, the LTE b‐value (6) can be recast as

b(r,g^)=bN(r,g^),N=N(0)(r)+N(2)(r,g^),N(2)(r,g^)=m=22N2m(r)Y2m(g^). (9)

Namely, at each point r in the FOV, the LTE b‐value depends on the nominal direction g^ via an ellipsoid N(r,g^)=Nij(r)ĝiĝj, such that its average over all g^𝕊2 is

meang^N(g^)=𝕊2dg^N(g^)=N0, (10)

and the directional variance

2.2. (11)

where we dropped the r‐dependence for brevity, and

2.2. (12)

is the =2 invariant of tensor N, compare Reference [66]. Figure 1a shows the dimensionless ratio b(r,g^)/b=N(r,g^) for a generic g^, as well as the vector field L(r)g^ distorted relative to g^, for the system [12] used in this study. Figure 1b,c shows maps of mean and standard deviation of b(r,g^)/b, Equations ((9), (10), (11), (12)).

Throughout the main text, we limit ourselves to LTE, as the shape of the B‐tensor (5) is determined by a single unit direction

g^(r)=L(r)g^|L(r)g^|, (13)

and remains LTE (rank‐1) for arbitrary L(r). In Appendix A, we outline how the b‐value and shape of an arbitrary B tensor are affected by L(r). As one may expect, a generic L(r) removes any special symmetries of the nominal B tensor. Hence, the moment we use beyond‐LTE encodings (4) with rankB>1, we should assume that the actual B‐tensor (3) is neither axially‐ nor spherically‐symmetric; working with generic‐shape B tensors would take us away from a familiar spherical convolution, as described below.

The gradient coil tensor field Lij(r) is obtained in a calibration experiment, which needs to be performed only once. A typical approach is to measure the static field produced by each gradient coil, represent these fields via harmonic polynomials [68] rYm(r^) (as the solutions of the Laplace equation), and compute the derivatives (2) to get the r‐dependent Lij(r).

2.3. Spherical Convolution: Kernel – fODF Factorization

PIPE applies to any microstructure model that can be represented as a spherical convolution. This functional form, empirically introduced at the level of signal representations [32, 33, 34, 35], encompasses the vast majority of biophysical models for white matter [36, 37, 38, 39, 40, 41, 42, 43] and gray matter [44, 45, 46]. Namely, a voxel is assumed to contain a collection of identical fiber bundles with arbitrary fODF (Figure 2). Under the assumption of every fiber bundle in a voxel having the same microstructure (justified in Reference [69]), one introduces the fiber response kernel 𝒦 that depends on the mutual orientation of the fiber direction n^ and the direction g^ of the diffusion encoding. Because this response does not depend on n^ and g^ separately, but only on their relative “distance” g^·n^ on a unit sphere, we can represent the dMRI signal as a convolution over the unit sphere 𝕊2 of the kernel with the fODF, Figure 2:

S(b,g^|ξ,pm)=𝕊2dn^𝒦(b,g^·n^|ξ)𝒫(n^|pm). (14)

Here b and g^, Equations (9) and (13), define the measurement, ξ are the scalar tissue parameters describing the kernel 𝒦 (e.g., compartment fractions and diffusivities), and

𝒫(n^|pm)==0,2,4,m=pmYm(n^) (15)

is the fODF parametrized using coefficients pm in the SH basis Ym(n^), Equation (8), with p001 (fODF normalization).

In Equation (14), we can further include the diffusion time t, echo time TE, and inversion time TI dependence of the signal. From now on, for brevity we will write 𝒦(b,g^·n^|ξ), implying that b stands for all scalar protocol parameters, b{b,TE,TI,t,} defining the measurement, such that, for example, compartments within the fiber bundle can have different T2 and T1 values [43, 70, 71] (with the respective compartment relaxation times included as tissue parameters ξ). The dependence on PGSE diffusion time t and pulse duration δ allows one to incorporate different classes of structural disorder in different compartments [72, 73, 74, 75, 76], as well as exchange between them [45, 46, 77, 78], with the corresponding model parameters added to ξ.

Much like convolutions become products under the ordinary Fourier transform, the convolution (14) becomes a product [32, 33, 41, 42] in the SH basis:

Sm(b|ξ,pm)=𝒦(b|ξ)pm, (16)

where Sm are the signal's SH coefficients

Sb,g^|ξ,pm=mSm(b|ξ,pm)Ym(g^) (17)

with respect to the directions (13), and 𝒦𝒦0 are the projections of the kernel (aligned with the z^ axis) onto Legendre polynomials P(ζ)=Y0(ζ), where ζ=g^·z^:

𝒦(b|ξ)=01dζ𝒦(b,ζ|ξ)P(ζ), (18)

such that the kernel is expanded as

𝒦(b,g^·n^ξ)==0,2,4,(2+1)𝒦(b|ξ)P(g^·n^). (19)

The factorization (16), essential for what follows, is valid as long as the kernel possesses axial symmetry (𝒦m=0 for nonzero m in the fiber basis n^=z^). Note that beyond this assumption, the microstructure of the fiber bundle is not constrained—it can have an arbitrary number of compartments (with Gaussian or non‐Gaussian diffusion), they can be coupled by exchange, have distinct relaxation properties, and so forth.

We note in passing that the convolution for generic B‐tensors and non‐axially symmetric kernels has to be defined on the SO(3) rotation group manifold 𝕊3/2 (the 3‐dimensional sphere 𝕊3 with the antipodal points identified, 2={1,1}), rather than on the 2‐dimensional sphere 𝕊2. The corresponding Fourier transform involves a product of non‐commuting matrices (the coefficients in the Wigner functions' basis), as the SO(3) group is non‐abelian. For axially symmetric response kernels, invariant under the SO(2) rotation around the fiber bundle axis, the convolution becomes effectively defined [79] on a quotient space 𝕊2SO(3)/SO(2) that is equivalent to the 2‐dimensional sphere 𝕊2, which explains the integration over fODF directions in Equation (14).

2.4. PIPE: Tissue – Protocol Factorization

The idea behind PIPE is to take the factorization (16) a step further and fully decouple the kernel's dependence on tissue parameters ξ and pm, and protocol parameters b{b,TE,TI,t,}. This can be done numerically up to any desired accuracy by splitting the kernel's rotational invariants (18) for a chosen tissue model (specified by a particular functional form of the kernel) into orthogonal functions using singular value decomposition (SVD):

𝒦(b|ξ)n=1Nsn()un()(b)vn()(ξ) (20)

up to arbitrary precision set by the number N. Substituting Equation (20) into Equations (16) and (17), we obtain:

S(b,g^|ξ,pm)n,,mun()(b)Ym(g^)αnmsn()vn()(ξ)pmγnmn=1N=0maxm=αnm(b,g^)γnm(ξ,pm) (21)

which allows us to write the dMRI signal as an expansion in the basis of data‐driven functions, factorizing the protocol (b,g^) and tissue (ξ,pm) dependencies. Indeed, the set of basis functions αnm(b,g^) depends purely on the protocol parameters, while the set γnm(ξ,pm) depends only on tissue (model) parameters. Including sn() in the definition of γnm(ξ,pm) ensures that the noise in the latter is approximately homoscedastic.

The factorization in Equation (21) decouples voxel‐to‐voxel variations in the protocol and tissue. Knowing the actual gradients (1) everywhere in the FOV, we interpolate the local basis functions αnm(b(r),g^(r)) (as described below) onto the unique set of diffusion encodings and directions for a voxel at position r, and linearly estimate tissue‐dependent signal components

γ^nm(r)=αnm(b,g^)S(b,g^) (22)

using standard Moore‐Penrose pseudoinverse αmn applied to the set of measurements S(b,g^)=[S(b1,g^1)S(bk,g^k)]t in a given voxel r.

Finally, we map the estimated γ^nm(r) onto the microstructural parameters of interest:

γ^nm(r){ξ^(r),p^m(r)}. (23)

Although such mapping is highly nonlinear, it is fully decoupled from the spatially varying protocol. Hence, it can use the same trained regressor as r varies throughout the FOV. This allows us to learn the regression (23) only once, and apply it to γ^nm from all brain voxels simultaneously in virtually no time. Practically, the SVD over the library of 𝒦(b|ξ) values used to compute un()(b) in Equation (20) also provides us with the tissue basis functions vn()(ξ).

From this point, we can proceed in two different ways. First, we combine vn()(ξ) and sn() with randomly sampled fODF SH coefficients pm to generate sets (γnm;{ξ,pm}) for the mapping (23), and use these as training data. Alternatively, we can form rotational invariants γn by factoring out pm (akin to References [41, 42]): The highest‐SNR ones

γ1=γ1mm=s1()v1()(ξ)p,p=pmm (24a)

for n=1, where m is the 2‐norm over m=,, divided by 2+1, and p are the fODF invariants; and the ones for n>1 by forming the ratios averaged over m=,,:

γn|n>1=γ1γnmγ1mm=sn()vn()(ξ)p. (24b)

The invariants (24) are analogous to those used in References [41, 42], with the dependence on discrete b shells substituted by the SVD index n. Like for (23), we can learn the mapping

γ^n(r){ξ^(r),p^(r)} (25)

from invariants (24) to tissue parameters and fODF invariants. To solve either (23) or (25), one needs to apply a sufficiently flexible regression. In this work, we employ the polynomial regression up to degree W=3:

ξ^=j1+j2++jNWaj1,j2,,jNy1j1y2j2yNjN, (26)

where ξ^ is the ML estimator, {yi}i=1N are the estimated γnm, W is the degree of the polynomial, and aj1,j2,,jN are the regression coefficients computed during training.

We observe that, for realistic SNR, mappings (23) and (25) perform similarly, although (25) is simpler due to the reduced dimensionality of γn compared to γnm. Note that the mapping (25) relies on having sufficient SNR to guarantee an unbiased γ^1 after taking the 2‐norm over m in Equation (24a). An alternative to Equations (24b) could be to define γn=γ^nmm=sn()vn()(ξ)p for all n, resembling the rotationally‐invariant mapping [42]. Practically, averaging the ratios γ^nm/γ^1m over m in Equation (24b) helps mitigate the non‐central‐χ bias in the 2‐norms γ^nmm, as their SNR can become quite low with increasing n.

The interpolation of the protocol‐dependent functions αnm(b,g^), Equation (21), for the specific directions g^ is performed using spherical harmonics Ym(g^). Chebyshev interpolation [80] is used for the continuum of b‐values (as well as other protocol parameters if necessary) in the smooth functions un()(b), which we pre‐compute at the Chebyshev polynomials' roots when generating the SVD factorization (20) for a large matrix containing a range of 𝒦(b|ξ) values (separately for each ), for physically meaningful combinations of tissue parameters ξ.

Figure 3 illustrates the SVD factorization (20) for the Standard Model [2] of diffusion in white matter, where

𝒦(b,g^·n^|ξ)=febDa(g^·n^)2+fwebDw+(1ffw)eb(DeDe)(g^·n^)2bDe (27)

includes the intra‐axonal compartment (a zero‐radius “stick” with a single tensor eigenvalue Da along the fascicle), the extra‐axonal compartment (axially‐symmetric tensor with axial and radial diffusivities De and De oriented along the fascicle), and isotropic free water of diffusivity Dw fixed at 3μm2/ms, with the fractions f+fe+fw=1, such that ξ is a set of the above diffusivities and fractions. Specifically, we generated a library of 𝒦(bξ) up to max=2, consisting of 50 000 random sets of SM parameters (uniformly sampling f[0.05,0.95],Da[1,3],De[1,3],De[0.1,1.2],fw[0,1]) and 1000 b‐values sampled at the Chebyshev roots [0,bmax], with bmax=10ms/μm2. This enabled accurate Chebyshev interpolation of un()(b) and subsequent approximation of 𝒦(b|ξ), see Figure 3b (note that 𝒦2<0). This interpolation is performed for each voxel to obtain an exact protocol‐dependent pseudoinverse αnm(b,g^) entering Equation (22). Since both the interpolation and pseudoinversion are linear operations, this is not a computationally intensive step; it takes about the same time as the conventional DTI estimation.

FIGURE 3.

FIGURE 3

(a) Chebyshev interpolation of basis functions un()(b) from SVD of a pre‐computed library for the Standard Model, with =0,2, and N0=N2=5 components for each in Equation (20). (b) Computation of rotational invariants 𝒦(b) for b‐values outside the library based on interpolated un()(b). It can be seen that: (i) just a few SVD components already provide sufficient overall accuracy, (ii) the bias propagation from the interpolation of the basis functions to the rotational invariants is negligible for typical SNR conditions (SNR=50 at b=0, and 60 directions in a shell).

2.5. MRI Experiments

2.5.1. Imaging

This study was performed under a local IRB‐approved protocol. After providing informed consent, a 44‐year‐old male volunteer underwent MRI in a 3T‐system (GE HealthCare, WI, USA) using an investigational MAGNUS head‐only gradient insert [12] and a 32‐channel head coil (NOVA Medical, Wilmington, MA, USA). The system was powered by a 1 MVA gradient driver per gradient axis, operating at a maximum gradient amplitude of 200mT/m and slew rate of 500T/m/s [12]. A monopolar PGSE diffusion weighting sequence was used for acquiring shells at (b[ms/μm2], Ndirs)={(1,25);(2,60);(8,50)}, with Δ=23ms, δ=12ms. Imaging parameters: Voxel size Inline graphic, TE=46ms, TR=5s, echo‐spacing =424μs, in‐plane acceleration =2, partial Fourier=0.65.

2.5.2. Image Reconstruction and Processing

Diffusion‐weighted complex MR images were reconstructed using GE Orchestra SDK tools, including k‐space filling approaches such as homodyne partial Fourier and ASSET. Denoising was applied in the complex domain [81, 82], therefore, reducing Rician bias significantly. Geometric distortions due to EPI were minimized due to the 500T/m/s slew rate, enabling a rapid readout, eliminating the need for distortion correction.

In our experiments, subject motion was minimal and therefore no motion correction was applied. However, when motion is more pronounced, gradient nonlinearities interact with rigid‐body head motion, modifying the alteration to the local B‐tensors differently for each diffusion‐weighted image (DWI) [31, 83]. This effect can be corrected by applying the estimated rigid‐body transformation of each DWI to the gradient coil tensor field, that is, by evaluating Lij at the original position r0=R1(rt), where R and t are the rotation and translation from the acquired frame to the final frame. The motion‐aware effective B‐tensor at voxel r is then given by L(r0)BLt(r0), which then needs to be further rotated to match the updated image coordinate frame. Incorporating such corrections is a natural extension of our framework and may be particularly important in studies involving pediatric or clinical populations where subject motion is unavoidable.

2.5.3. Model Estimation and Training Data

The linear estimation of γ^nm from the set of diffusion measurements was done using a voxel‐specific Moore‐Penrose pseudoinverse, Equation (22). For the mapping (23), a cubic polynomial regression (W=3) was used for each model parameter in ξ, as it was deemed optimal for 25SNR500. Training data γnm=sn()vn()pm was used tor the regression to capture the factorization. Independent uniform priors were used for the kernel parameters f𝒰(0.05,0.95),Da𝒰(0.5,3),De𝒰(0.5,3),De𝒰(0.1,1.5),fw𝒰(0,1), with f+fw1. The fODFs used in the training were random collections of 2 fiber lobes with exponentially decaying p, as in Reference [84]:

p=Cλ,>0,p2𝒰(0.02,0.9),λ𝒰(0.5,0.9),pm=wpm(1)+(1w)pm(2),w𝒰(0,1).

The spherical harmonics coefficients of each lobe, pm(1,2) with 6, were independently rotated in 3D using Wigner rotation matrices parametrized by ZYZ Euler angles α,β,γ [85] with rotations uniformly sampled from the SO(3) group manifold 𝕊3/2 according to the invariant measure sinβdαdβdγ. Training took under 3 min for all kernel parameters. Computations were performed on a 3.7 GHz 6‐Core i5 CPU with 32GB of RAM. All codes for PIPE were implemented in MATLAB (R2022a, MathWorks, Natick, Massachusetts). These are publicly available as part of the PIPE toolbox at https://github.com/NYU‐DiffusionMRI/PIPE.

3. Results

Field measurements from the head‐only system used here show non‐negligible gradient nonlinearities away from the isocenter, affecting peripheral parts of the brain. These can be observed by computing the re‐scaling of the diffusion weighting b(r,g^)/b as we traverse the FOV (this profile is specific to each nominal diffusion direction g^), see Figure 1a. With the irreducible decomposition in Equation (7), we can separate isotropic and anisotropic modifications to the nominal b‐value due to nonlinearities, see Figure 1b,c.

The accuracy of the Chebyshev interpolation of un()(b) library values is shown in Figure 3a. Additionally, Figure 3b indicates that N0=4 and N2=3 basis functions suffice for accurate factorization of the Standard Model kernel in white matter. Errors due to SVD truncation and interpolation of un()(b) are below typical SNR levels and, therefore, can be considered negligible. This is possible due to the exponentially fast decrease of sn() with n.

Noise propagation experiments emphasize the flexibility of our proposed approach to work on shelled and non‐shelled data, see Figure 4. This showcases that PIPE can account for arbitrarily large gradient nonlinearities, applying even to cases where the nominal shelled structure of data is lost due to nonlinearities. In this work we have focused on the Standard Model of diffusion in white matter (27) as an illustration, but the PIPE framework can be applied to any spherical convolution‐based approach [32, 33, 34, 35, 36, 37, 38, 39, 40, 41, 42, 43, 44, 45, 46] irrespective of the functional dependence of the kernel on diffusion weighting parameters.

FIGURE 4.

FIGURE 4

Noise propagation simulation highlighting that PIPE can handle arbitrary protocols. Note that due to suboptimal experimental design (LTE data only), parallel diffusivities are highly dominated by the prior distribution used for training the machine learning estimator [47].

In vivo full brain maps of the linearly estimated γ^nm parameters are shown in Figure 5 for a healthy volunteer. These parameters approximate the nonlinear b‐value dependence of 𝒦0(b) and 𝒦2(b) rotational invariants of the Standard Model kernel (27). The resemblance between γ^n,2,m and maps of spherical harmonics is due to the fODF p2m coefficients being part of the former. Neglecting the spatial variation of the diffusion encoding causes a bias dependent on the distance of the voxel to the bore's isocenter.

FIGURE 5.

FIGURE 5

Estimated γ^nm maps for a healthy volunteer from non‐shelled voxelwise protocols. Anatomical patterns similar to spherical harmonics Sm(b) are observed, with the difference that the γ^nm do not depend on the protocol parameters because the combination of multiple n captures the b‐dependence. Larger values have fewer significant components n, and larger n elements are noisier due to having smaller relative contributions (dictated by roughly exponentially decreasing sn()). γn denote the rotational invariants described in Equation (24).

Figure 6a shows signal rotational invariants computed by resampling αnm(b,g^) at b=1,2,4ms/μm2 uniformly over the unit sphere and applying Equation (21) to get S(b,g^) and subsequently S(b). The effect of accounting for gradient nonlinearities is minor in the center of the brain because it coincided with the bore's isocenter, while its significance gradually increases up to values of the order 10%. Corresponding parameter biases will likely be greater due to a highly nonlinear parameter estimation for typical brain microstructure models. Figure 7 shows SM parametric maps obtained after applying a polynomial regression to γ^nm.

FIGURE 6.

FIGURE 6

(a) Rotational invariants S0(b) and S2(b), normalized with b=0, computed from γ^nm and resampling αnm(b,g^) at b=1,2,4ms/μm2. (b) Relative errors in % between accounting for gradient nonlinearities and considering a flat gradient field.

FIGURE 7.

FIGURE 7

Standard model microstructure maps (ξ=[f,fw,Da,De||,De]) for transversal and coronal slices. These maps were obtained by applying a fully connected neural network on the γ^nm shown in Fig. 5. Note that this model only applies to white matter.

4. Discussion

Gradient nonlinearities are ubiquitous in MRI systems, a natural consequence of Maxwell's equations. A design variable is the linearity region, which determines the size of the organs that can be scanned (head‐only vs. whole‐body scanners). Smaller linearity regions for MRI gradient coils offer advantages for achieving higher gradient performance, particularly in terms of gradient strength and slew rate. This has motivated recent hardware advances in head‐only [13, 16] or insertable gradient coils [12, 15], optimized for brain imaging. These provide access to the frontier of tissue microstructure imaging through the lens of diffusion; however, this comes at the cost of FOV spatial variations in diffusion weightings.

This work addresses the practical challenge of efficiently handling variations in diffusion weightings due to gradient nonlinearities. Previous work exploited the linearity of the DTI log‐signal to correct diffusion tensors post hoc [30]. Such corrections, however, are not applicable to more general nonlinear models. In these cases, B‐tensors must be corrected in a voxelwise fashion before the estimation process, making data‐driven estimators extremely inefficient. The proposed PIPE framework enables fast parameter estimation of a wide class of biophysical models (spherical convolution models) from dMRI data, where each voxel is acquired with a different diffusion weighting and directions.

The key ingredient behind PIPE is the factorization of the signal dependence on experimental and tissue parameters. This is achieved through SVD and a projection into a spherical harmonics basis. Therefore, allowing us to split the estimation process into two parts:

  • i.

    a linear step that handles the spatial variations of experimental parameters while remaining independent of tissue parameters;

  • ii.

    a nonlinear step that converts SVD coefficients into model parameters.

Training compute time for step (ii) was fast (1‐minute) because PIPE needs to be trained only once, irrespective of the magnitude of gradient nonlinearities. Naively retraining conventional fixed‐protocol ML estimators for each voxel would take a prohibitive amount of compute (100 days, considering a full brain at 2mm resolution has in the order of 105 voxels in the brain). Furthermore, accuracy and memory consumption are significantly improved with respect to computing a brute‐force regression where the full experimental and model parameter space are sampled jointly.

The PIPE framework can be tailored to any model based on a spherical convolution of a fiber response with an fODF [32, 33, 34, 35, 36, 37, 38, 39, 40, 41, 42, 43, 44, 45, 46], thereby allowing its application in white and gray matter. Additionally, PIPE is readily extendable for simultaneously varying diffusion times, TE, and so forth. PIPE does not require the data to be sampled in shells. Furthermore, it provides optimal data resampling for those algorithms that do need shells, for example, References [34, 35]. This feature is shared with previously proposed approaches focused on data resampling for artifact correction [86]. However, the main advantage of PIPE is that it is not constrained to shells and that it allows for spatially varying protocols throughout the FOV.

We have tested the feasibility of our method on simulations and in vivo brain dMRI data acquired with a head‐only gradient insert with non‐negligible gradient nonlinearities. Results showed no trace of gradient nonlinearities in output maps, see Figure 7. Our results will stimulate the development of high‐performance gradient hardware where nonlinearities are allowed to be large by design.

The proposed estimator, like any other data‐driven algorithm, is influenced by the nature of its training data. The dependence on the training set is tied to the quality of acquisitions (sampling of the q‐space and the SNR). In scenarios where comprehensive protocols are employed, the algorithm's performance is less susceptible to variations in training data [47]. Conversely, in cases with limited protocols containing less information, the performance is more biased towards the mean of the training set [47]. Here, we used independent uniform distributions for training to minimize spurious parameter correlations.

The impact of gradient nonlinearities on multidimensional diffusion encodings to date has focused on their effect on Maxwell compensation for asymmetric gradient waveforms [87]. In this work, we provide the first in‐depth analysis of the impact of gradient nonlinearities on arbitrary multidimensional diffusion encodings, extending previous work on LTE [30]. Our result provides a parsimonious way to separate isotropic and anisotropic contributions of gradient nonlinearities to the b‐value and B‐tensor shape. Both isotropic and anisotropic components affect the b‐value, whereas only anisotropic components alter the B‐tensor shape, see Appendix A. The advantage of acquiring LTE diffusion encodings is that while gradient nonlinearities change the actual intensity and direction of the diffusion weighting, they do not affect the B‐tensor shape. One should be cautious when acquiring non‐LTE data on systems with strong gradient nonlinearities, as these will introduce deformations in the B‐tensor, removing any pre‐set symmetries. This would require a re‐definition of the convolution on the SO(3) group manifold rather than on a 2‐dimensional sphere, which is beyond the scope of this work. Approaches like Reference [88] could be used to design encodings that remain maximally symmetric for specific nonlinearities.

5. Conclusions

We proposed a two‐step machine learning PIPE parameter estimation framework that enables fast parameter estimation of convolution‐based diffusion MRI models from data where each voxel is acquired with a different protocol. This method allows for the straightforward application of a large class of biophysical tissue models to data acquired with arbitrarily large gradient nonlinearities. Furthermore, the data is not constrained to be acquired in any fashion, for example, in shells, and there are no limits to the gradient nonlinearities as long as the protocol is well defined for each voxel. This method is readily extendable for simultaneously varying diffusion times, echo, or inversion times.

Conflicts of Interest

G.L., E.F., D.S.N., and N.Y.U. School of Medicine are stockholders of MicSi Inc.—post‐processing tools for advanced MRI methods. S.C., E.F., and D.S.N. are co‐inventors in technology related to this research; a PCT patent application has been filed. A.Z., N.A., and T.K.F.F. are employees of GE HealthCare.

Acknowledgments

This work has been supported by NIH under NINDS award R01 NS088040 and NIBIB awards R01 EB027075, P41 EB017183, K99 EB036080, and NIH OD and NIDCR award DP5OD031854. The authors are grateful to Sune Jespersen for fruitful discussions, and to Eric Fiveland, Maggie Fung, and Chitresh Bhushan for their help with the MRI experiments.

Appendix A. How an Arbitrary B‐Tensor is Modified by Gradient Nonlinearities

The most general (nominal) B‐tensor, a symmetric 3×3 matrix with 6 independent parameters, admits the irreducible decomposition (in the notations of Reference [66]):

Bij=13bδij+m=22B2m𝒴ij2m. (A1)

Here the first term defines the isotropic part, and the remaining five parameters B2m, coupled to symmetric trace‐free tensors 𝒴ij2m, characterize the anisotropy of encoding while not contributing to the overall diffusion weighting b=trB. Substituting Equation (A1) into Equation (3), the actual B‐tensor for this general case becomes

B=b3LLt+m=22B2mL𝒴2mLt. (A2)

We can see that the tensor N=LtL introduced in the main text is not enough to describe the deformation of a generic B.

Let us find the =0 and =2 spherical tensor components of the symmetric tensor (A2). The =0 isotropic part b3=13trB, using cyclic permutation under the trace, can be cast via the components of the N tensor:

b=trB=bN0+m=22B2mtr𝒴2mN=bN0+32m=22B2mN2m, (A3)

where we used the fact that N is real‐valued, and the orthogonality [66] 𝒴ij2m𝒴ij2m=32δmm. The geometric meaning of Equation (A3) is as follows. The =0 component of (A2) comes from two sources: The =0 component of N, and the =0 component of the “addition of angular momenta”, 220, of the =2 components of B and N. Indeed, using N2m=(1)mN2,m, the sum in Equation (A3) can be understood as [66]

15m=22B2mN2m=15m(1)mB2mN2,m=2020|00m,m2m2m|00B2mN2m=B(2)N(2)=0, (A4)

where the Clebsch‐Gordan coefficients 2020|00=1/5 and 2m2m|00=(1)mδm,m/5.

The =2 components are found via B2m=23𝒴ij2mBij:

B2m=29btr𝒴2mLLt+23mB2mtr𝒴2mL𝒴2mLt (A5)

They stem both from the =2 anisotropy of symmetric matrix LLt, and from the =2 part of the object LBLt, where tensors L and Lt have the same =0,2 components, and the =1 components ai and ai, respectively, where the pseudo‐vector ai=ϵijkLjk.

From the result (A5) we observe that the presence of generic matrices L and LtL, which do not possess any symmetry and have all their 1+3+5=9 components corresponding to the irreducible representations with =0,1,2, removes any special symmetry of the nominal B tensor (A1). For example, even if the original B=b3δij were isotropic, B2m0, the anisotropy of the symmetric tensor LLt (nonzero m=22 components for =2) means that the actual B‐tensor will be of a generic form, without any particular symmetries. The same applies to the widely used axially symmetric family [64, 65]

B=bβg^g^+1β3I (A6)

parametrized by the unit vector g^ along the symmetry axis and by the shape parameter β. In the basis where g^=z^, only B20 will be nonzero, out of five B2m. However, for a generic L, Equation (A5) will have all the actual B2m components tr𝒴2mL𝒴20Lt present. This has important implications, since sampling with generic B‐tensors takes us away from the convolution on a 2‐sphere 𝕊2 onto the 3‐dimensional SO(3) group manifold.

Coelho S., Lemberskiy G., Zhu A., et al., “What If Each Voxel Were Measured With a Different Diffusion Protocol?,” Magnetic Resonance in Medicine 95, no. 4 (2026): 2277–2290, 10.1002/mrm.70144.

Funding: This work was partially supported by the National Institutes of Health, National Institute of Biomedical Imaging and Bioengineering (Grant/Award Nos: P41‐EB017183, R01‐EB027075, K99‐EB036080), the National Institute of Neurological Disorders and Stroke (Grant/Award No: R01‐NS088040), and the Office of the Director and National Institute of Dental & Craniofacial Research (Grant/Award No: DP5‐OD031854).

Data Availability Statement

The data that support the findings of this study are available on request from the corresponding author. The data are not publicly available due to being the property of GE Healthcare. The code used for the analyses and experiments in this study is publicly available on GitHub at: https://github.com/NYU‐DiffusionMRI/PIPE.

References

  • 1. Kiselev V. G., “Fundamentals of Diffusion MRI Physics,” NMR in Biomedicine 30 (2017): 1–18. [DOI] [PubMed] [Google Scholar]
  • 2. Novikov D. S., Fieremans E., Jespersen S. N., and Kiselev V. G., “Quantifying Brain Microstructure With Diffusion MRI: Theory and Parameter Estimation,” NMR in Biomedicine 32 (2019): e3998. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Alexander D. C., Dyrby T. B., Nilsson M., and Zhang H., “Imaging Brain Microstructure With Diffusion MRI: Practicality and Applications,” NMR in Biomedicine 32, no. 4 (2019): e3841. [DOI] [PubMed] [Google Scholar]
  • 4. Novikov D. S., “The Present and the Future of Microstructure MRI: From a Paradigm Shift to Normal Science,” Journal of Neuroscience Methods 351 (2021): 108947. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Kiselev V. G., “Microstructure With Diffusion MRI: What Scale we Are Sensitive to?,” Journal of Neuroscience Methods 347 (2021): 108910. [DOI] [PubMed] [Google Scholar]
  • 6. Weiskopf N., Edwards L. J., Helms G., Mohammadi S., and Kirilina E., “Quantitative Magnetic Resonance Imaging of Brain Anatomy and in Vivo Histology,” Nature Reviews Physics 3, no. 8 (2021): 570–588. [Google Scholar]
  • 7. Assaf Y., “Can we Use Diffusion MRI as a Bio‐Marker of Neurodegenerative Processes?,” BioEssays 30, no. 11–12 (2008): 1235–1245. [DOI] [PubMed] [Google Scholar]
  • 8. Jones D. K., Diffusion MRI (Oxford University Press, 2010). [Google Scholar]
  • 9. Jelescu I. O. and Budde M. D., “Design and Validation of Diffusion MRI Models of White Matter,” Frontiers in Physics 5 (2017): 61. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Nilsson M., Englund E., Szczepankiewicz F., et al., “Imaging Brain Tumour Microstructure,” NeuroImage (2018): 232–250. [DOI] [PubMed] [Google Scholar]
  • 11. Novikov D. S., Kiselev V. G., and Jespersen S. N., “On Modeling,” Magnetic Resonance in Medicine 79 (2018): 3172–3193. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Foo T. K., Tan E. T., Vermilyea M. E., et al., “Highly Efficient Head‐Only Magnetic Field Insert Gradient Coil for Achieving Simultaneous High Gradient Amplitude and Slew Rate at 3.0T (MAGNUS) for Brain Microstructure Imaging,” Magnetic Resonance in Medicine 83, no. 6 (2020): 2356–2369. [DOI] [PubMed] [Google Scholar]
  • 13. Huang S. Y., Witzel T., Keil B., et al., “Connectome 2.0: Developing the Next‐Generation Ultra‐High Gradient Strength Human MRI Scanner for Bridging Studies of the Micro‐, Meso‐ and Macro‐Connectome,” NeuroImage 243 (2021): 118530. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Feng J., Littin S., Amrein P., et al., “Design of a High‐Performance non‐Linear Gradient Coil for Diffusion Weighted MRI of the Breast,” Journal of Magnetic Resonance 331 (2021): 107052. [DOI] [PubMed] [Google Scholar]
  • 15. Feinberg D. A., Beckett A. J., Vu A. T., et al., “Next‐Generation MRI Scanner Designed for Ultra‐High‐Resolution Human Brain Imaging at 7 Tesla,” Nature Methods 20, no. 12 (2023): 2048–2057. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. Ramos‐Llorden G., Lee H.‐H., Davids M., et al., “Ultra‐High Gradient Connectomics and Microstructure MRI Scanner for Imaging of Human Brain Circuits Across Scales,” Nature Biomedical Engineering (2025), 10.1038/s41551-025-01457-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Sarracanie M., LaPierre C. D., Salameh N., et al., “Low‐Cost High‐Performance MRI,” Scientific Reports 5, no. 1 (2015): 15177. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Zimmerman Cooley C., McDaniel P. C., Stockmann J. P., et al., “A Portable Scanner for Magnetic Resonance Imaging of the Brain,” Nature Biomedical Engineering 5, no. 3 (2021): 229–239. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Liu Y., Leong A. T. L., Zhao Y., et al., “A Low‐Cost and Shielding‐Free Ultra‐Low‐Field Brain MRI Scanner,” Nature Communications 12, no. 1 (2021): 7238. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Zhao Y., Ding Y., Lau V., et al., “Whole‐Body Magnetic Resonance Imaging at 0.05 Tesla,” Science 384, no. 6696 (2024): 7168. [DOI] [PubMed] [Google Scholar]
  • 21. Jones D. K., Alexander D. C., Bowtell R., et al., “Microstructural Imaging of the Human Brain With a ‘Super‐Scanner’: 10 Key Advantages of Ultra‐Strong Gradients for Diffusion MRI,” NeuroImage 182 (2018): 8–38. Microstructural Imaging. [DOI] [PubMed] [Google Scholar]
  • 22. McNab J. A., Edlow B. L., Witzel T., et al., “The Human Connectome Project and Beyond: Initial Applications of 300mT/m Gradients,” NeuroImage 80 (2013): 234–245. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Veraart J., Nunes D., Rudrapatna U., et al., “Noninvasive Quantification of Axon Radii Using Diffusion MRI,” eLife 9 (2020): e49855. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Dai E., Zhu A., Yang G. K., et al., “Frequency‐Dependent Diffusion Kurtosis Imaging in the Human Brain Using an Oscillating Gradient Spin Echo Sequence and a High‐Performance Head‐Only Gradient,” NeuroImage 279 (2023): 120328. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Chan K.‐S., Ma Y., Lee H., et al., “In Vivo Human Neurite Exchange Time Imaging at 500 mT/m Diffusion Gradients,” Imaging Neuroscience 3 (2025): 00544, 10.1162/imag_a_00544. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Lee H.‐H., Novikov D. S., and Fieremans E., “Localization Regime of Diffusion in Human Gray Matter on a High‐Gradient MR System: Sensitivity to Soma Size,” in Proceedings of the International Society of Magnetic Resonance in Medicine (International Society of Magnetic Resonance in Medicine, 2022). [Google Scholar]
  • 27. Campbell‐Washburn A. E., Keenan K. E., Hu P., et al., “Low‐Field MRI: A Report on the 2022 ISMRM Workshop,” Magnetic Resonance in Medicine 90, no. 4 (2023): 1682–1694. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Abate F., Adu‐Amankwah A., Ae‐Ngibise K. A., et al., “UNITY: A Low‐Field Magnetic Resonance Neuroimaging Initiative to Characterize Neurodevelopment in Low and Middle‐Income Settings,” Developmental Cognitive Neuroscience 69 (2024): 101397. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Gholam J., Schmid P., Ametepe J., et al., Diffusion Tensor MRI and Spherical‐Deconvolution‐Based Tractography on an Ultra‐Low Field Portable MRI System (2025), 10.48550/arXiv.2506.04473. [DOI]
  • 30. Bammer R., Markl M., Barnett A., et al., “Analysis and Generalized Correction of the Effect of Spatial Gradient Field Distortions in Diffusion‐Weighted Imaging,” Magnetic Resonance in Medicine 50, no. 3 (2003): 560–569. [DOI] [PubMed] [Google Scholar]
  • 31. Rudrapatna U., Parker G. D., Roberts J., and Jones D. K., “A Comparative Study of Gradient Nonlinearity Correction Strategies for Processing Diffusion Data Obtained With Ultra‐Strong Gradient MRI Scanners,” Magnetic Resonance in Medicine 85, no. 2 (2021): 1104–1113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Tournier J. D., Calamante F., Gadian D. G., and Connelly A., “Direct Estimation of the Fiber Orientation Density Function From Diffusion‐Weighted MRI Data Using Spherical Deconvolution,” NeuroImage 23 (2004): 1176–1185. [DOI] [PubMed] [Google Scholar]
  • 33. Anderson A. W., “Measurement of Fiber Orientation Distributions Using High Angular Resolution Diffusion Imaging,” Magnetic Resonance in Medicine 54, no. 5 (2005): 1194–1206. [DOI] [PubMed] [Google Scholar]
  • 34. Tournier J. D., Calamante F., and Connelly A., “Robust Determination of the Fibre Orientation Distribution in Diffusion MRI: Non‐Negativity Constrained Super‐Resolved Spherical Deconvolution,” NeuroImage 35, no. 4 (2007): 1459–1472. [DOI] [PubMed] [Google Scholar]
  • 35. Jeurissen B., Tournier J. D., Dhollander T., Connelly A., and Sijbers J., “Multi‐Tissue Constrained Spherical Deconvolution for Improved Analysis of Multi‐Shell Diffusion MRI Data,” NeuroImage 103 (2014): 411–426. [DOI] [PubMed] [Google Scholar]
  • 36. Jespersen S. N., Kroenke C. D., Østegaard L., Ackerman J. J. H., and Yablonskiy D. A., “Modeling Dendrite Density From Magnetic Resonance Diffusion Measurements,” NeuroImage 34 (2007): 1473–1486. [DOI] [PubMed] [Google Scholar]
  • 37. Fieremans E., Jensen J. H., and Helpern J. A., “White Matter Characterization With Diffusional Kurtosis Imaging,” NeuroImage 58 (2011): 177–188. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38. Zhang H., Schneider T., Wheeler‐Kingshott C. A., and Alexander D. C., “NODDI: Practical in Vivo Neurite Orientation Dispersion and Density Imaging of the Human Brain,” NeuroImage 61 (2012): 1000–1016. [DOI] [PubMed] [Google Scholar]
  • 39. Kaden E., Kelm N. D., Carson R. P., Does M. D., and Alexander D. C., “Multi‐Compartment Microscopic Diffusion Imaging,” NeuroImage 139 (2016): 346–359. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Jensen J. H., Russell G. G., and Helpern J. A., “Fiber Ball Imaging,” NeuroImage 124 (2016): 824–833. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41. Reisert M., Kellner E., Dhital B., Hennig J., and Kiselev V. G., “Disentangling Micro From Mesostructure by Diffusion MRI: A Bayesian Approach,” NeuroImage 147 (2017): 964–975. [DOI] [PubMed] [Google Scholar]
  • 42. Novikov D. S., Veraart J., Jelescu I. O., and Fieremans E., “Rotationally‐Invariant Mapping of Scalar and Orientational Metrics of Neuronal Microstructure With Diffusion MRI,” NeuroImage 174 (2018): 518–538. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43. Veraart J., Novikov D. S., and Fieremans E., “TE Dependent Diffusion Imaging (TEdDI) Distinguishes Between Compartmental T2 Relaxation Times,” NeuroImage 182 (2018): 360–369. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44. Palombo M., Ianus A., Guerreri M., et al., “SANDI: A Compartment‐Based Model for Non‐Invasive Apparent Soma and Neurite Imaging by Diffusion MRI,” NeuroImage 215 (2020): 116835. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45. Jelescu I. O., de Skowronski A., Geffroy F., Palombo M., and Novikov D. S., “Neurite Exchange Imaging (NEXI): A Minimal Model of Diffusion in Gray Matter With Inter‐Compartment Water Exchange,” NeuroImage 256 (2022): 119277. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46. Olesen J. L., Østegaard L., Shemesh N., and Jespersen S. N., “Diffusion Time Dependence, Power‐Law Scaling, and Exchange in Gray Matter,” NeuroImage 251 (2022): 118976. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47. Coelho S., Fieremans E., and Novikov D. S., “How Do we Know we Measure Tissue Parameters, Not the Prior?,” in Proceedings of the International Society of Magnetic Resonance in Medicine (International Society of Magnetic Resonance in Medicine, 2021). [Google Scholar]
  • 48. Liao Y., Coelho S., Chen J., et al., “Mapping Tissue Microstructure of Brain White Matter in Vivo in Health and Disease Using Diffusion MRI,” Imaging Neuroscience 2 (2024): 1–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49. Coelho S., Lemberskiy G., Zhu A., et al., “What if Every Voxel Was Measured With A Different Diffusion Protocol?,” in Proceedings of the International Society of Magnetic Resonance in Medicine (International Society of Magnetic Resonance in Medicine, 2023). [Google Scholar]
  • 50. Coelho S., Fieremans E., and Novikov D. S., “Zero‐Shell Diffusion MRI: Focus on Microstructure by Decoupling Fiber Orientations,” in Proceedings of the International Society of Magnetic Resonance in Medicine (International Society of Magnetic Resonance in Medicine, 2023). [Google Scholar]
  • 51. Stejskal E. O. and Tanner T. E., “Spin Diffusion Measurements: Spin Echoes in the Presence of a Time‐Dependent Field Gradient,” Journal of Chemical Physics 42 (1965): 288–292. [Google Scholar]
  • 52. Kay S. M., Fundamentals of Statistical Signal Processing: Estimation Theory (PTR Prentice‐Hall, 1993). [Google Scholar]
  • 53. Jelescu I. O., Veraart J., Fieremans E., and Novikov D. S., “Degeneracy in Model Parameter Estimation for Multi‐Compartmental Diffusion in Neuronal Tissue,” NMR in Biomedicine 29 (2016): 33–47. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54. Mozumder M., Pozo J. M., Coelho S., and Frangi A. F., “Population‐Based Bayesian Regularization for Microstructural Diffusion MRI With NODDIDA,” Magnetic Resonance in Medicine 84 (2019): 1553–1565. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55. Golkov V., Dosovitskiy A., Sperl J. I., et al., “Q‐Space Deep Learning: Twelve‐Fold Shorter and Model‐Free Diffusion MRI Scans,” IEEE Transactions on Medical Imaging 35, no. 5 (2016): 1344–1351. [DOI] [PubMed] [Google Scholar]
  • 56. de Almeida Martins J. P., Nilsson M., Lampinen B., et al., “Neural Networks for Parameter Estimation in Microstructural MRI: Application to a Diffusion‐Relaxation Model of White Matter,” NeuroImage 244 (2021): 118601. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57. Gyori N. G., Palombo M., Clark C. A., Zhang H., and Alexander D. C., “Training Data Distribution Significantly Impacts the Estimation of Tissue Microstructure With Machine Learning,” Magnetic Resonance in Medicine 87, no. 2 (2022): 932–947. [DOI] [PubMed] [Google Scholar]
  • 58. Coelho S., Baete S. H., Lemberskiy G., et al., “Reproducibility of the Standard Model of Diffusion in White Matter on Clinical MRI Systems,” NeuroImage 257 (2022): 119290. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59. Sumanaweera T., Glover G., Song S., Adler J., and Napel S., “Quantifying MRI Geometric Distortion in Tissue,” Magnetic Resonance in Medicine 31, no. 1 (1994): 40–47. [DOI] [PubMed] [Google Scholar]
  • 60. Mitra P. P., “Multiple Wave‐Vector Extensions of the NMR Pulsed‐Field‐Gradient Spin‐Echo Diffusion Measurement,” Physical Review B 51 (1995): 15074–15078. [DOI] [PubMed] [Google Scholar]
  • 61. Mori S. and Van Zijl P. C. M., “Diffusion Weighting by the Trace of the Diffusion Tensor Within a Single Scan,” Magnetic Resonance in Medicine 33, no. 1 (1995): 41–52. [DOI] [PubMed] [Google Scholar]
  • 62. Wong E. C., Cox R. W., and Song A. W., “Optimized Isotropic Diffusion Weighting,” Magnetic Resonance in Medicine 34, no. 2 (1995): 139–143. [DOI] [PubMed] [Google Scholar]
  • 63. Cheng Y. and Cory D. G., “Multiple Scattering by NMR,” Journal of the American Chemical Society 121 (1999): 7935–7936. [Google Scholar]
  • 64. Westin C.‐F., Knutsson H., Pasternak O., et al., “Q‐Space Trajectory Imaging for Multidimensional Diffusion MRI of the Human Brain,” NeuroImage 135 (2016): 345–362. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65. Topgaard D., “Multidimensional Diffusion MRI,” Journal of Magnetic Resonance 275 (2017): 98–113. [DOI] [PubMed] [Google Scholar]
  • 66. Coelho S., Chen J., Szczepankiewicz F., et al., “Geometry of the Cumulant Series in Diffusion MRI,” 2025. Preprint arXiv:2409.03010.
  • 67. Thorne K. S., “Multipole Expansions of Gravitational Radiation,” Reviews of Modern Physics 52 (1980): 299–339. [Google Scholar]
  • 68. Roméo F. and Hoult D. I., “Magnet Field Profiling: Analysis and Correcting Coil Design,” Magnetic Resonance in Medicine 1, no. 1 (1984): 44–65. [DOI] [PubMed] [Google Scholar]
  • 69. Christiaens D., Veraart J., Cordero‐Grande L., et al., “On the Need for Bundle‐Specific Microstructure Kernels in Diffusion MRI,” NeuroImage 208 (2020): 116460. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70. Lampinen B., Szczepankiewicz F., Martensson J., et al., “Towards Unconstrained Compartment Modeling in White Matter Using Diffusion‐Relaxation MRI With Tensor‐Valued Diffusion Encoding,” Magnetic Resonance in Medicine 84, no. 3 (2020): 1605–1623. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71. Coelho S., Liao Y., Szczepankiewicz F., et al., “Assessment of Precision and Accuracy of Brain White Matter Microstructure Using Combined Diffusion MRI and Relaxometry,” Human Brain Mapping 45 (2024): e26725. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 72. Novikov D. S., Jensen J. H., Helpern J. A., and Fieremans E., “Revealing Mesoscopic Structural Universality With Diffusion,” Proceedings of the National Academy of Sciences of the United States of America 111, no. 14 (2014): 5088–5093. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73. Lee H.‐H., Fieremans E., and Novikov D. S., “LEMONADE (t): Exact Relation of Time‐Dependent Diffusion Signal Moments to Neuronal Microstructure,” in Proceedings of the International Society of Magnetic Resonance in Medicine (International Society of Magnetic Resonance in Medicine, 2018). [Google Scholar]
  • 74. Lee H.‐H., Papaioannou A., and Kim S.‐L., et al., “A Time‐Dependent Diffusion MRI Signature of Axon Caliber Variations and Beading,” Communications Biology 3, no. 354 (2020): 1–13. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 75. Lee H.‐H., Papaioannou A., Novikov D. S., and Fieremans E., “In Vivo Observation and Biophysical Interpretation of Time‐Dependent Diffusion in Human Cortical Gray Matter,” NeuroImage 222 (2020): 117054. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76. Chan K.‐S., Ma Y., Lee H.‐H., et al., “Time‐Dependent Standard Model of Diffusion in Human Brain White Matter Evaluated in Vivo on the High Gradient Performance Connectome 2.0 Scanner,” in Proceedings of the International Society of Magnetic Resonance in Medicine (International Society of Magnetic Resonance in Medicine, 2024). [Google Scholar]
  • 77. Novikov D. S., Coronado‐Leija R., and Fieremans E., “Exchange Between Structurally‐Disordered Compartments,” in Proceedings of the International Society of Magnetic Resonance in Medicine (International Society of Magnetic Resonance in Medicine, 2023). [Google Scholar]
  • 78. Lee H.‐H., Novikov D. S., Fieremans E., and Huang S. Y., “Revealing Membrane Integrity and Cell Size From Diffusion Kurtosis Time Dependence,” Magnetic Resonance in Medicine 93, no. 3 (2025): 1329–1347. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 79. Healy D. M., Hendriks H., and Kim P. T., “Spherical Deconvolution,” Journal of Multivariate Analysis 67, no. 1 (1998): 1–22. [Google Scholar]
  • 80. Trefethen L. N., Approximation Theory and Approximation Practice, Extended ed. (Society for Industrial and Applied Mathematics, 2019). [Google Scholar]
  • 81. Veraart J., Fieremans E., and Novikov D. S., “Diffusion MRI Noise Mapping Using Random Matrix Theory,” Magnetic Resonance in Medicine 76, no. 5 (2016): 1582–1593. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 82. Lemberskiy G., Baete S. H., Veraart J., et al., “Achieving Sub‐Mm Clinical Diffusion MRI Resolution by Removing Noise During Reconstruction Using Random Matrix Theory,” in Proceedings 27th Scientific Meeting, 0770, International Society for Magnetic Resonance in Medicine (International Society of Magnetic Resonance in Medicine, 2019) 2019. [Google Scholar]
  • 83. Guo F., de Luca A., Parker G., et al., “The Effect of Gradient Nonlinearities on Fiber Orientation Estimates From Spherical Deconvolution of Diffusion Magnetic Resonance Imaging Data,” Human Brain Mapping 42, no. 2 (2021): 367–383. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 84. Coronado‐Leija R., Abdollahzadeh A., Lee H.‐H., et al., “Volume Electron Microscopy in Injured Rat Brain Validates White Matter Microstructure Metrics From Diffusion MRI,” Imaging Neuroscience 2 (2024): 1–20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 85. Tinkham M., Group Theory and Quantum Mechanics (Dover Books on Chemistry) (Dover Publications, 2003). [Google Scholar]
  • 86. Christiaens D., Cordero‐Grande L., Hutter J., et al., “Learning Compact q‐Space Representations for Multi‐Shell Diffusion‐Weighted MRI,” IEEE Transactions on Medical Imaging 38, no. 3 (2019): 834–843. [DOI] [PubMed] [Google Scholar]
  • 87. Szczepankiewicz F., Eichner C., Anwander A., Westin C.‐F., and Paquette M., “The Impact of Gradient Non‐Linearity on Maxwell Compensation When Using Asymmetric Gradient Waveforms for Tensor‐Valued Diffusion Encoding,” in Proceedings of the International Society of Magnetic Resonance in Medicine (International Society of Magnetic Resonance in Medicine, 2020). [Google Scholar]
  • 88. Szczepankiewicz F., Westin C.‐F., and Nilsson M., “Maxwell‐Compensated Design of Asymmetric Gradient Waveforms for Tensor‐Valued Diffusion Encoding,” Magnetic Resonance in Medicine 82, no. 4 (2019): 1424–1437. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data that support the findings of this study are available on request from the corresponding author. The data are not publicly available due to being the property of GE Healthcare. The code used for the analyses and experiments in this study is publicly available on GitHub at: https://github.com/NYU‐DiffusionMRI/PIPE.


Articles from Magnetic Resonance in Medicine are provided here courtesy of Wiley

RESOURCES