Traceable stiffness calibration of colloidal AFM probes for biomechanical measurements

Zhi Li 1,, Valeriya Cherkasova 2, Sai Gao 1, Thomas Fröhlich 2, Uwe Brand 1
PMCID: PMC12880979  PMID: 41644719

Abstract

The accurate calibration of bending stiffness of colloidal atomic force microscopy (AFM) probes is essential for reliable nanomechanical measurements, especially when large micro-spheres are used in biological applications. This study investigates the influence of frictional contact between an AFM spherical tip and the load button on stiffness measurements obtained via bending tests and proposes a new analytical model to account for this effect. Finite element simulations of frictional sliding contact between colloidal spheres and load button were conducted to validate the proposed model. A proof-of-principle experimental setup was developed to traceably acquire force-deflection curves of several typical colloidal AFM probes, and results showed good agreement (within 1.5 % deviation) with a validated stiffness calibration system. Experimental data for large-sphere colloidal probes confirmed the presence of a transition phase in the unloading curve due to frictional contact and demonstrated that accurate stiffness results can be obtained when friction is properly considered. Additionally, friction coefficients for four tip-surface material combinations were experimentally determined, providing broadly relevant data that can be effectively applied in AFM nanomechanics, especially in investigations of tip-sample interactions.

Keywords: Cantilever stiffness calibration; Atomic force microscopy (AFM); Bio-AFM, biomechanical measurement; AFM colloidal probes tip-surface interaction, Quasi-static stiffness measurement; Nanofriction; Nano-sliding contact; Nano-force metrology

Subject terms: Engineering, Materials science, Nanoscience and technology, Physics

Introduction

Atomic force microscopy (AFM)-based nanomechanical measurements play a crucial role in various scientific and industrial fields, particularly in the development of advanced soft materials1 including biomaterials25. In recent years, bio-AFM nanomechanical phenotyping has emerged as a powerful tool for early disease detection and diagnosis6,7, thanks to its exceptional spatial resolution down to the sub-nanometer scale and its ultra-high force sensitivity in the piconewton (Inline graphic N) range. As a result, it has been widely adopted in cell and tissue biology, as well as biomedicine, to address critical questions related to human health and disease810 and future pandemics11. However, achieving reliable and quantitative measurements with bio-AFM requires careful attention to critical instrumentation issues1214. Among these, one of the key challenges is the traceable and accurate calibration of AFM cantilever bending stiffness1521.

Established calibration methods for conventional AFM probes, such as rectangular cantilevers with relatively sharp tips, have been well standardized22. One widely adopted traceable quasi-static approach involves using a traceably calibrated micro-force measurement facility23, e.g. a precision compensation balance with a force resolution down to nanoNewton (nN, Inline graphic N)21 to measure the reaction force of an AFM probe under incremental deflection. This method enables high sensitivity and low measurement uncertainty, even for cantilevers with nominal stiffness as low as 0.01 N/m24.

In the meantime, bio-AFM applications often utilize probes with spherical tips, commonly referred to as colloidal probes, with radii ranging from a few microns to tens of microns25,26. These are particularly suited for quantitative mechanical characterization of soft biomaterials such as biological tissues. However, applying the standard traceable method to calibrate the bending stiffness of colloidal probes presents significant challenges, primarily due to adhesion and frictional interactions between the probe tip and the load button surface. In addition, it is worthwhile to mention that there exist non-contact indirect stiffness calibration approaches for large colloidal probes on basis of the thermal noise method27 or optical tweezers28. However, such measurement results need generally to be carefully verified to ensure traceability.

To address this, we propose in this manuscript an analytical model for extracting the effective bending stiffness of colloidal AFM probes from their force–deflection curves. Finite element analysis (FEA) is employed to experimentally investigate the influence of tip-surface friction during bending and to validate the analytical model. A proof-of-principle experimental setup has been developed to investigate the bending stiffness of representative colloidal probes, including custom-fabricated designs. The experimental results support the validity of this proposed model.

Materials and methods

Determination of the bending stiffness of AFM cantilevers using a micro-force measurement facility

The quasi-static method for the determination of the bending stiffness Inline graphic of an AFM cantilever is illustrated in Fig. 1(a): An AFM cantilever under calibration is positioned by a nano-positioning system (not shown) and engaged onto the load button mounted on a weighing scale, e.g. a compensation balance. For simplicity, in the schematic drawing, the AFM cantilever beam is horizontally mounted. After engagement, the cantilever is further moved downwards, and the reaction force Inline graphic of the cantilever beam under deflection will be directly acquired by the weighing scale. In case that, (1) the AFM tip should be smooth, have relatively small tip radius, be well centered with respect to the load button, and (2) the cantilever bending deflection Inline graphic is small, the force-increasing (or loading) and force-decreasing (unloading) curves should overlap well, and demonstrate excellent linearity. The bending stiffness of the cantilever Inline graphic can be finally determined by evaluation of the slope of the loading and unloading (Inline graphic) curves.

Fig. 1.

Fig. 1

Determination of the quasi-static bending stiffness of colloidal AFM probes using a traceably calibrated compensation balance based microforce measurement facility.

On the contrary, typical colloidal AFM probes usually have glued spheres with relatively large tip radius and poor surface quality for material testing, leading to evident frictional (sliding) contact with the load button for bending stiffness measurement, as illustrated in Fig. 1(b). And, in practice, the tip radius Inline graphic of the load button is usually far larger than the AFM tip radius Inline graphic, i.e. Inline graphic. And quite often, instead of a rounded load button, a flat punch will be utilized for AFM cantilever bending measurements.

Theoretical analysis of the frictional tip-surface sliding contact for stiffness evaluation

Under the applied normal force Inline graphic, the friction force Inline graphic between the AFM spherical tip and the load button can be estimated by

graphic file with name d33e345.gif 1

where Inline graphic denotes the Coulomb’s coefficient of friction between the colloidal tip and the load button. Notably, the sign before Inline graphic depends on the direction of tip sliding. Specially, during unloading Inline graphic.

Under the steady-state frictional sliding contact between the AFM spherical tip and the load button, the vertical deflection Inline graphic of colloidal probe due to the normal contact force Inline graphic and the friction force Inline graphic can be deduced as follows:

graphic file with name d33e377.gif 2

where Inline graphic is the equivalent moment on the cantilever beam caused by friction force Inline graphic, Inline graphic the equivalent tip height of the AFM spherical tip, and

graphic file with name d33e395.gif 3
graphic file with name d33e399.gif 4

where L is the equivalent cantilever length at the contact point, E the elastic modulus of the cantilever beam, and I is the second moment of area of the beam’s cross section29.

Finally, we have the total deflection of the cantilever at the contact point

graphic file with name d33e420.gif 5

Given the cantilever stiffness Inline graphic, the measured stiffness Inline graphic from the Inline graphic curves can be estimated as follows:

graphic file with name d33e437.gif 6

It can be seen from Eq. 6 that the measured cantilever stiffness Inline graphic suffers from the influence of tip-surface contact friction. The influence will become stronger, when the ratio of tip height to cantilever length Inline graphic increases. In addition, from Eq. 6 the measured stiffness from the steady-state part of the loading curve Inline graphic and the measured unloading stiffness Inline graphic can be deduced as follows:

graphic file with name d33e466.gif 6a

Obviously, the measured Inline graphic is usually larger than Inline graphic, as shown in Fig. 1(c). The cantilever stiffness Inline graphic can be therefore conveniently deduced from Eq. 6a as

graphic file with name d33e490.gif 7

From Eq. 6a we obtain

graphic file with name d33e498.gif 7a

Combing Eq. 7a with Eq. 7, the friction coefficient Inline graphic can be further derived as follows:

graphic file with name d33e514.gif 8

It is also very interesting to note that at the beginning of the unloading curve, there should exist a transition phase, since the frictional force will change its direction. From Eq. (4), the transition range Inline graphic can be estimated as follows:

graphic file with name d33e526.gif 9

where Inline graphic is the maximum deflection of the cantilever beam during measurement.

To evaluate the unloading stiffness Inline graphic precisely, it is therefore recommended that only measurement data out of the transition phase should be utilized for data evaluation.

Finite element analysis of frictional tip-surface sliding contact during bending measurement

To validate the theoretical analysis in Eqs. (1)-(9), the frictional contact between a colloidal sphere and the load button during the stiffness measurement process has been numerically investigated with the help of finite element (FE) analysis. The geometrical model used in FE simulation is illustrated in Fig. 2(a). The fundamental parameters of the cantilever beam, glued sphere, and the load button are listed in Table 1. And the ratio Inline graphic amounts to about 1/3.

Fig. 2.

Fig. 2

FEA-based numerical investigation of the tip-surface interaction during the stiffness calibration process.

Table 1.

Parameter table of the colloidal AFM probe and the load button used in the finite element analysis.

Dimensions Materials Young’s modulus, GPa Poisson’s ratio
Colloidal AFM probe* Beam Length L, Inline graphicm 150 Si 169 0.28
Thickness T, Inline graphicm 2.5
Width w, Inline graphicm 30
Tip Height Inline graphic, Inline graphicm 50 Glass 63 0.17
Radius Inline graphic, Inline graphicm 20
Load button Radius Inline graphic, Inline graphicm 250 Ruby 400 0.3

*Cantilever beam dimensions come from the datasheet of AFM probe HQ: XSC11 series, Inline graphicMasch.

For simplicity, a two-dimensional FE analysis was carried out using Ansys Mechanical APDL version 19.2 (Ansys Inc., USA). During the loading (and unloading) process, the load button is fixed, only the left end of the cantilever is moved downwards (and upwards). Typical displacement vector sum of the colloid probe and the load button under maximum deflection is illustrated in Fig. 2(b).

The numerical simulation results are presented in Fig. 3. As shown in Fig. 3(a), when the friction coefficient Inline graphic, the loading and unloading curves overlapped quite well, displaying excellent linearity across the entire deflection range of 4 Inline graphicm. As Inline graphic increasing, the loading stiffness Inline graphic becomes noticeably greater than the unloading stiffness Inline graphic. In particular for Inline graphic, the transition range calculated using Eq. (9) is Inline graphic Inline graphicm, which agrees well with the numerical analysis illustrated in Fig. 3(b).

Fig. 3.

Fig. 3

Numerical investigation of the force-deflection curves for AFM cantilever stiffness measurement under different frictional tip-surface sliding.

Results

To experimentally validate the data analysis model proposed above, a proof-of-principle experimental setup has been developed, with which home-made and commercial colloidal AFM probes have been measured.

A proof-of-principle micro-force measurement facility for traceable measurement of the normal stiffness of colloidal AFM probes with relatively large spheres

The home-developed micro-force measurement facility (MFMF) consists of a traceably calibrated compensation balance (Sartorius AG, WZ2P-CW, linearity Inline graphic 20 nN, standard deviation Inline graphic 20 nN) for reading the reaction force of a cantilever under calibration, a nanopositioning system (P-620, Physik Instrumente GmbH) for moving the cantilever, and a three-axis coarse positioning system, as illustrated in Fig. 4(a).

Fig. 4.

Fig. 4

Home-developed micro-force measurement facility (MFMF) for traceable calibration of the bending stiffness of AFM probes.

A custom-built data acquisition system with a LabView-based user interface has been developed to drive the motorized z-axis stage for coarse positioning of the cantilever under calibration. Fine engagement of the cantilever with the load button on the balance, along with the subsequent bending measurement, is achieved by actuating the z-axis piezo stage. To enable in-situ calibration of the combined z-positioning system, including both the motorized z-axis stage and the nanopositioning piezo-stage, a home-built laser interferometer (not depicted in Fig. 4(b)) is employed.

Typical measurements of classical AFM probes

The MFMF has been firstly employed to measure the normal stiffness of classical AFM probes with a diamond flat punch load button (20 Inline graphicm in diameter, fabricated by Synton AG, Switzerland). Using the motorized z-stage, the cantilever tip was positioned approximately 15 Inline graphicm above the top surface of the load button. The z-piezo stage was then used to move the AFM probe incrementally toward the load button until the maximum displacement Inline graphic was reached. Each step included a typical waiting time of 5 seconds, corresponding to five times the balance response time, to ensure reliable readings. The actual position of the AFM probe and the corresponding balance output were recorded simultaneously. All measurements were typically performed under controlled environmental conditions, with a temperature of Inline graphic C and a relative humidity of Inline graphic. The experimental setup was installed on a stable granite table without air damping. Alignment between the AFM probe and the load button was achieved with the help of a home-built auxiliary microscope, resulting in a misalignment below 5 Inline graphicm.

Figure 5(a) shows one of the force-deflection curves obtained by the MFMF for a contact-mode AFM cantilever with rounded tip (SD-R30-CONT, NanoSensors, Inline graphic = 30 nm, L = 450 Inline graphicm and Inline graphic = 15 Inline graphicm). The maximum cantilever deflection Inline graphic reaches approximately 16 Inline graphicm during the measurement, and the total duration of one measurement cycle amounts to 3.5 minutes.

Fig. 5.

Fig. 5

Typical measurement results for determining the normal stiffness of a contact-mode AFM cantilever with a rounded tip (SD-R30-CONT, NanoSensors™).

According to Eq. (4), the ratio Inline graphic is relatively small, indicating that the effect of frictional contact between the AFM tip and the load button on stiffness measurements is minimal and can be considered negligible. This is well supported by Fig. 5(a), where the loading and unloading curves overlap closely over a cantilever bending range of 16 Inline graphicm. As shown in Fig. 5(b), the residual linear deviation of the loading and unloading curves within the evaluation region for stiffness calculation is less than 10 nN. Figure 5(c) further demonstrates excellent repeatability of the stiffness measurements, with a variation of less than Inline graphic across 50 measurement cycles (approximately 3 hours in total). The measured stiffness of this cantilever is Inline graphic (a detailed uncertainty analysis is available in21,24).

For comparison, the bending stiffness of the same cantilever has also been measured using a traceable force-displacement measurement device developed at TU Ilmenau, employing both electromagnetic (EMFC) and electrostatic force (ESFC) compensation principles24,30. These methods yielded results of Inline graphic N/m and Inline graphic N/m, respectively. The stiffness measured using the proof-of-principle MFMF clearly agrees well with those obtained by the validated measurement system at TU Ilmenau.

Typical stiffness measurements of colloidal AFM probes

A home-made colloidal probe was fabricated by attaching a Soda Lime glass microsphere (Inline graphic Inline graphicm in diameter, Goodfellow GmbH) to a HQ AFM cantilever (HQ: XSC11, Cantilever C, Inline graphicMasch Europe), as illustrated in Fig. 6(a). The three-dimensional topography of the contact surface of the microsphere was characterized using a laser scanning confocal microscope (LEXT OLS4100, Olympus), yielding a surface roughness of Inline graphic nm in the central contact region. This colloidal probe has an equivalent cantilever length of Inline graphic Inline graphicm, and a total tip height of approximately 50 Inline graphicm.

Fig. 6.

Fig. 6

Traceable calibration of the bending stiffness of a home-made colloidal AFM probe.

A home-made load button with a ruby half-ball (Edmund Optics GmbH, 500 Inline graphicm in diameter) as the testing stylus has been used to measure the bending stiffness of the colloidal probe, as shown in Fig. 6(b). Topography measurement of the ruby half-ball by the aforementioned confocal microscope reveals that the top surface of the load button has a roughness of Inline graphic nm.

Figure 7(a) presents a representative force-deflection (Inline graphic) curve acquired using the MFMF for this home-fabricated colloidal probe. Notably, two key features are observed: (1) the loading Inline graphic curve exhibits a different slope compared to the unloading curve, and (2) a distinct transition phase is evident at the onset of unloading.

Fig. 7.

Fig. 7

Typical measurement results for determining the normal stiffness of a home-fabricated colloidal probe.

By performing linear fitting in the respective linear regions of the loading and unloading curves, as indicated in Fig. 7(a), the measured loading stiffness Inline graphic and unloading stiffness Inline graphic are quantitatively determined. Figure 7(b) details the residual deviation from linearity within these regions. Interestingly, the magnitude of the residual deviation is approximately Inline graphic Inline graphicN, which is nearly 20 times larger than the deviations observed in Fig. 5(b). This larger deviation is attributed to the relatively high surface roughness of the microsphere (Inline graphic and Inline graphic ) attached to the cantilever. Figure 7(c) summarizes the measured values of Inline graphic, Inline graphic, along with the calculated cantilever stiffness Inline graphic derived using Eq. (7). Additionally, Figure 7(d) reports the measured friction coefficient Inline graphic between the glass microsphere and the ruby ball load button. During the sequence of contact bending measurements, slight variation in Inline graphic were observed, resulting in the minor fluctuations in Inline graphic and Inline graphic. Despite this, the cantilever stiffness remains stable, yielding Inline graphic (repeatability, 1Inline graphic) N/m, demonstrating excellent consistency throughout the measurement series.

A commercial colloidal AFM probe (CP-CONT-BSG, sQube, Borosillicate Glass sphere, 20 Inline graphicm in diameter, cantilever length Inline graphic Inline graphicm, nominal stiffness 0.36 N/m) has also been calibrated using the MFMF with a diamond flat punch, as shown in Fig. 8(a). With the same measurement procedure and data analysis method detailed above, the measured loading, unloading stiffness and deduced stiffness of this commercial colloidal probe is depicted in Fig. 8(b). The measured friction coefficient between a glass microsphere and a diamond flat punch is illustrated in Fig. 8(c). We can conclude that the stiffness of this commercial colloidal cantilever Inline graphic N/m and Inline graphic.

Fig. 8.

Fig. 8

Typical measurement results for determining the normal stiffness of a contact-mode colloidal AFM probe (CP-CONT-BSG, sQube).

Finally, a commercial colliodal AFM probe with glued PMMA microsphere (CP-CONT-PM-E, sQube, 15 Inline graphic in diameter) has also been measured with the above mentioned MFMF and procedure. For convenience, the measurement results of the cantilevers investigated in this manuscript are summarized in Table 2.

Table 2.

Summary of the experimental results reported in this manuscript.

AFM probes Reference cantilever Colloidal probes
SD-R30-CONT, NanoSensorsTM HQ: XSC11, Inline graphicMasch (glued microsphere, R = 20 Inline graphicm) CP-CONT-BSG, (R = 10 Inline graphicm), sQube CP-CONT-PM, (R = 7.5 Inline graphicm), sQube
Beam Dimensions, Inline graphic m L 450 150 450 450
Inline graphic 15 50 20 15
Stiffness ± 4%, N/m Inline graphic 0.394 5.718 0.366 0.487
Inline graphic 0.394 5.129 0.360 0.470
Inline graphic 0.394 5.408 0.362 0.478
Friction coefficient, Inline graphic Inline graphic = 0.00 Inline graphic = 0.11 Inline graphic = 0.02 Inline graphic = 0.04
Stiffness by TU Ilmenau, N/m Inline graphic 0.3930 ± 0.0046 na 0.3570 ± 0.0055 na
Inline graphic 0.3968 ± 0.0022 na 0.3568 ± 0.0016 na

Conclusion

In this manuscript, we have systematically analyzed and modeled the influence of frictional contact between AFM tips and load buttons on stiffness measurement results obtained using the quasi-static bending-based calibration method. A new data evaluation model has been developed for the quantitative determination of the bending stiffness of colloidal AFM probes, particularly those equipped with large micro-spheres. The proposed analytical model has been validated through finite element simulations of representative frictional sliding contacts between a colloidal probe and a load button.

A proof-of-principle experimental setup was constructed to traceably acquire force-deflection (Inline graphic) curves from several representative AFM probes. Initial results indicate that, for stiffness calibration of standard AFM probes, the home-built micro-force measurement facility yields results that are in excellent agreement with those obtained from the validated force-displacement measurement device, with deviations of less than Inline graphic.

Force-deflection measurements on colloidal AFM probes with relatively large spheres confirm the key predictions of the analytical model: (1) a distinct transition phase appears at the beginning of the unloading curve due to frictional contact effects, and (2) accurate stiffness results can still be achieved when these frictional contributions are properly accounted for. Additionally, friction coefficients for several typical tip-surface combinations, including silicon-diamond (nanoscale), glass-diamond (microscale), PMMA-diamond (microscale), and glass-ruby (submillimeter scale), were experimentally determined. These coefficients are of broad relevance and can be utilized in future research and development efforts in AFM nanomechanics, particularly in studies of tip-sample interactions.

It should be noted that the present theoretical analysis does not account for the influence of geometrical form errors of the colloidal probe spheres, which may contribute to measurement uncertainty. Future work will address this limitation through quantitative modeling. Moreover, the current analysis does not include the effects of adhesion forces Inline graphic31, which may become significant when calibrating the stiffness of ultra-soft cantilevers. In the regime where Inline graphic, the friction force in Eq. (1) becomes Inline graphic, and the transition range Inline graphic can be estimated as

graphic file with name d33e1390.gif 10

Meaningful measurements can only be obtained when the total deflection Inline graphic. Further development of the analytical model to encompass these effects will be pursued to enable accurate characterization of ultra-soft AFM colloidal probes with large spheres, particularly for biomechanical measurement applications.

In addition, stiffness calibration in this manuscript was performed using quasi-static measurements; therefore, only quasi-static tip–button interactions were considered. In contrast, for fast stiffness calibration approaches-such as those employing high-speed piezostages and high-speed force sensors, including reference springs-the friction force between the AFM tip and the load button becomes dependent on sliding velocity and exhibits generally nonlinear behavior32,33. Under these conditions, the analytical model proposed here is no longer valid and requires further extension.

Finally, it is worth noting that most commercial AFMs operate with a tilted cantilever. However, the analytical model proposed in this manuscript is formulated for stiffness measurements of untilted AFM cantilevers. For in-situ stiffness calibration of tilted colloidal AFM probes, for example, using an in-situ nanoforce sensor, an extension of the proposed analytical model will be required to accommodate this commonly used configuration.

Supplementary Information

Acknowledgements

The authors thank Ole Willke from Department 5.1, Surface Metrology, PTB Braunschweig, for preparing the colloidal probes.

Author contributions

Conceptualization, Zhi Li; methodology, Zhi Li; measurements and data analysis, Zhi Li, Valeriya Cherkasova, and Sai Gao; formal analysis, Zhi Li; data curation, Zhi Li, Valeriya Cherkasova; Finite Element simulation and validation, Sai Gao; writing—original draft preparation, Zhi Li and Sai Gao; writing—review and editing, all authors. All authors have read and agreed to the published version of the manuscript.

Funding

Open Access funding enabled and organized by Projekt DEAL.

Data availability

The data obtained and used in this contribution can be provided by the corresponding author upon request.

Declarations

Competing interests

The authors declare no competing interests.

Footnotes

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

The online version contains supplementary material available at 10.1038/s41598-026-38158-7.

References

  • 1.Sharma, V., Shi, X., Yao, G., Pharr, G. M. & Wu, J. Y. Atomic force microscopy of biomaterials surfaces and interfaces. Surf. Sci.491, 303–332. 10.1016/S0039-6028(01)01296-1 (2001). [Google Scholar]
  • 2.Jandt, K. D. Surface characterization of an ultra-soft contact lens material using an atomic force microscopy nanoindentation method. Sci. Reports12, 20013. 10.1016/S0039-6028(01)01296-1 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Alsteens, D. et al. Atomic force microscopy of biomaterials surfaces and interfaces. Nat. Rev. Mater.2, 17008. 10.1038/natrevmats.2017.8 (2017). [Google Scholar]
  • 4.Stylianou, A. Atomic force microscopy for collagen-based nanobiomaterials. J. Nanomater.2017, 9234627. 10.1155/2017/9234627 (2009). [Google Scholar]
  • 5.Mittelviefhaus, M., Müller, D. B., Zambelli, T. & Vorholt, J. A. A modular atomic force microscopy approach reveals a large range of hydrophobic adhesion forces among bacterial members of the leaf microbiota. The ISME J.13, 1878–1882. 10.1038/s41396-019-0404-1 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Lekka, M. Discrimination between normal and cancerous cells using afm. BioNanoScience6, 65–80. 10.1155/2017/9234627 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Deng, X. et al. Application of atomic force microscopy in cancer research. J. Nanobiotechnology16, 102. 10.1186/s12951-018-0428-0 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Krieg, M. et al. Atomic force microscopy-based mechanobiology. Nat. Rev. Phys.1, 41–57. 10.1038/s42254-018-0001-7 (2019). [Google Scholar]
  • 9.Najera, J., Rosenberger, M. R. & Datta, M. Atomic force microscopy methods to measure tumor mechanical properties. Cancers (Basel)15, 3285. 10.3390/cancers15133285 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Cho, D. H., Aguayo, S. & Cartagena-Rivera, A. X. Atomic force microscopy-mediated mechanobiological profiling of complex human tissues. Biomaterials303, 122389. 10.1016/j.biomaterials.2023.122389 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.dos Santos Natividade, R., Danzer, B., Somoza, V. & Koehler, M. Atomic force microscopy at the forefront: unveiling foodborne viruses with biophysical tools. npj Viruses3, 25. 10.1038/s44298-025-00107-y (2025). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Emerson, R. & Camesano, T. On the importance of precise calibration techniques for an atomic force microscope. Ultramicroscopy106, 413–22. 10.1016/j.ultramic.2005.11.008 (2006). [DOI] [PubMed] [Google Scholar]
  • 13.Schillers, H. et al. Standardized nanomechanical atomic force microscopy procedure (snap) for measuring soft and biological samples. Sci. Reports7, 5117. 10.1038/s41598-017-05383-0 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Weber, A., Iturri, J., Benitez, R. & Toca-Herrera, J. Measuring biomaterials mechanics with atomic force microscopy . 1. in-fluence of the loading rate and applied force (pyramidal tips). Microsc. research technique82, 1392–1400. 10.1002/jemt.23291 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Sader, J. E. & White, L. Theoretical analysis of the static deflection of plates for atomic force microscope applications. J. Appl. Phys.74, 1–9. 10.1002/jemt.23291 (1993). [Google Scholar]
  • 16.Clifford, C. A. & Seah, M. P. Improved methods and uncertainty analysis in the calibration of the spring constant of an atomic force microscope cantilever using static experimental methods. Meas. Sci. Technol.20, 125501. 10.1002/jemt.23291 (2009). [Google Scholar]
  • 17.te Riet, J. et al. Interlaboratory round robin on cantilever calibration for afm force spectroscopy. Ultramicroscopy111, 1659–1669. 10.1016/j.ultramic.2011.09.012 (2011). [DOI] [PubMed] [Google Scholar]
  • 18.Slattery, A., Blanch, A., Quinton, J. & Gibson, C. Accurate measurement of atomic force microscope cantilever deflection excluding tip-surface contact with application to force calibration. Ultramicroscopy131, 46–55. 10.1016/j.ultramic.2013.03.009 (2013). [DOI] [PubMed] [Google Scholar]
  • 19.Gao, S., Z-K., Z., Y., W. & Herrmann, K. Towards quantitative determination of the spring constant of a scanning force microscope cantilever with a microelectromechanical nano-force actuator. Measurement Science and Technology21, 015103, 10.1088/0957-0233/21/1/015103 (2009).
  • 20.Gao, S. & Brand, U. In-situ nondestructive characterization of the normal spring constant of afm cantilevers. Meas. Sci. Technol.25, 044014. 10.1088/0957-0233/25/4/044014 (2014). [Google Scholar]
  • 21.Brand, U., Gao, S., Engl, W., Sulzbach, T. & S.W., S. Comparing AFM cantilever stiffness measured using the thermal vibration and the improved thermal vibration methods with that of an SI traceable method based on MEMS. Meas. Sci. Technol. 28, 034010, 10.1088/1361-6501/28/3/034010 (2017).
  • 22.Clifford, A. Summary of ISO/TC 201 standard: ISO 11775:2015 – surface chemical analysis –scanning probe microscopy – determination of cantilever normal spring constants. Surf. Interface Anal.49, 171–172 (2017). [Google Scholar]
  • 23.Kim, M.-S., Pratt, J., Brand, U. & Jones, C. Report on the first international comparison of small force facilities: a pilot study at the micronewton level. Metrologia49, 70–81. 10.1088/0026-1394/49/1/011 (2012). [Google Scholar]
  • 24.Cherkasova, V., Li, Z., Popadic, R., Froehlich, T. & Brand, U. Calibration of the stiffness of soft cantilevers based on electromagnetic and electrostatic force compensation principles. Meas. Sensors 38, 101335, 10.1016/j.measen.2024.101335 (2025). Proceedings of the XXIV IMEKO World Congress.
  • 25.Dörig, P. et al. Exchangeable colloidal afm probes for the quantification of irreversible and long-term interactions. Biophys. J.105, 463–472. 10.1016/j.bpj.2013.06.002 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Chighizola, M., Puricelli, L., Bellon, L. & Podestá, A. Large colloidal probes for atomic force microscopy: fabrication and calibration issues. J. Mol. Recognit.34, e2879. 10.1002/jmr.2879 (2021). [DOI] [PubMed] [Google Scholar]
  • 27.Gates, R. & Osborn, W. S. G. Accurate flexural spring constant calibration of colloid probe cantilevers using scanning laser doppler vibrometry. Nanotechnology26, 235704. 10.1088/0957-4484/26/23/235704 (2015). [DOI] [PubMed] [Google Scholar]
  • 28.Witko, T., Baster, Z., Rajfur, Z., Sofińska, K. & Barbasz, J. Increasing AFM colloidal probe accuracy by optical tweezers. Sci. Reports11, 509. 10.1038/s41598-020-79938-z (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Young, W. C. & Budynas, R. G. Roark’s Formulas for Stress and Strain (McGraw-Hill, New York, 2002), 7th edn.
  • 30.Cherkasova, V. Review and perspectives of the force-displacement measurement system with electromagnetic and electrostatic force compensation principles. tm - Technisches Messen92, 141–151. 10.1515/teme-2025-0005 (2025). [Google Scholar]
  • 31.Maugis, D. Adhesion of spheres: The JKR-DMT transition using a dugdale model. J. Colloid Interface Sci.150, 243–269. 10.1016/0021-9797(92)90285-T (1992). [Google Scholar]
  • 32.Stark, R. W., Schitter, G. & Stemmer, A. Velocity dependent friction laws in contact mode atomic force microscopy. Ultramicroscopy100, 309–317. 10.1016/j.ultramic.2003.11.011 (2003). [DOI] [PubMed] [Google Scholar]
  • 33.Lee, H., Lee, N., Seo, Y., Eom, J. & Lee, S. Comparison of frictional forces on graphene and graphite. Nanotechnology20, 325701. 10.1088/0957-4484/20/32/325701 (2009). [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Data Availability Statement

The data obtained and used in this contribution can be provided by the corresponding author upon request.


Articles from Scientific Reports are provided here courtesy of Nature Publishing Group