Abstract
TAVR is an alternative to surgical aortic valve [AoV] replacement. In adults, rates of atrioventricular block [AVB] requiring permanent pacemaker [PPM] placement in the modern era are 4–24%. Post-TAVR conduction abnormality incidence and risk factors are unknown in children and young adults. Describe post-TAVR conduction abnormalities in children and young adults. Retrospective single-center review of patients undergoing TAVR (9/2014 to 6/2021). Patients with pre-existing complete AVB or PPM were excluded (N = 1). The relationship between described adult risk factors for AVB and primary outcome of new conduction abnormality was assessed. Of 28 cases (Ages 3.5–22 y), 50% were male and 43% had isolated AoV disease, and the remainder with multilevel obstructive left-sided heart disease (29%) or complex congenital heart disease (29%). Baseline conduction abnormality was present in 57% (16/28), with right bundle branch block the most prevalent (9, 56%). Post-TAVR, acute- and late-onset conduction abnormalities occurred in 9 and 2 patients, respectively, and resolved in 8/11 patients during follow-up. One patient required PPM for complete heart block. There was no association between new conduction abnormality and previously reported adult risk factors—baseline RBBB, membranous septum length, valve implantation depth, or degree of valve oversizing. There was no relationship between outcome and baseline conduction abnormality nor history of multiple AoV interventions. In our pediatric series, AVB requiring PPM was rare following TAVR, with an incidence 3.6%—lower than average rates reported in adult literature. We identified no association of conduction abnormality with described adult risk factors.
Supplementary Information
The online version contains supplementary material available at 10.1007/s00246-025-03889-3.
Keywords: TAVR, Pacemaker, Congenital heart disease, Pediatrics
Introduction
Since publication of the sentinel paper on transcatheter aortic valve replacement (TAVR) in 2010 [1], TAVR has gained increasing popularity and has become the standard of care for treatment of calcific aortic stenosis in elderly adults regardless of their surgical risk [2]. Our center has pioneered TAVR in pediatric patients with congenital heart disease (CHD). However, despite the abundance of adult literature on outcomes and complications, there are only small case series describing center-specific experiences with TAVR in the pediatric and young adult populations [3].
Conduction abnormalities, including complete heart block (CHB) requiring permanent pacemaker placement (PPM), are well-described complications of TAVR in the adult population with a variable incidence of pacemaker placement generally ranging from 4–24% in the modern era, with historical data as high as 51% [4–7]. This risk is due to the vulnerable course of the penetrating bundle of His. Based on autopsy specimens, the His bundle travels along the border of the membranous septum located inferiorly between the right and noncoronary aortic cusps, rendering it susceptible to mechanical trauma with TAVR [8]. Risk factors for CHB with the TAVR procedure—both modifiable and nonmodifiable—have been identified in the adult population. The most consistently reported risk factor for CHB is pre-existing right bundle branch block (RBBB) [4]. Other described risk factors include deep valve implantation within the left ventricular outflow tract (LVOT), prosthesis oversizing (relative to valve annulus or LVOT), shorter membranous septum, advanced age, non-coronary cusp (NCC) thickness, and the degree of NCC calcification (with the latter two potentially leading to direct compression and damage of the local conduction system) [4, 7, 9–15].
Due to the relative paucity of children and CHD patients who have undergone TAVR, conduction abnormalities after TAVR have not been described in this population. Consequently, there is a lack of data regarding risk factors in this group of patients. In fact, these patients represent a unique clinical group, exhibiting a preponderance of baseline conduction abnormalities relative to the healthy pediatric population. For this reason, some of the risk factors not directly related to advanced age, as identified in adult studies, may more accurately predict CHB within this pediatric population. Additional unknowns in pediatric TAVR patients include how peri- and post-procedural changes in ventricular loading conditions affect ectopy burden. We aim to describe our institutional outcomes concerning conduction abnormalities and ectopy burden after TAVR in pediatric and young adult patients with CHD.
Methods
Patient Population
All patients treated with a SAPIEN 3 valve (Edwards Lifesciences, Irvine, CA, United States) at our institution from the first case in 2014 to June 2021 were included in this study (n = 29). The decision to undergo TAVR was determined via a multidisciplinary discussion within our heart center involving cardiothoracic surgeons along with referring and interventional cardiologists. All patients met our institution’s clinical criteria for aortic valve replacement. All patients were provided written informed consent for the procedure. TAVR was performed via standard technique under general anesthesia using the transfemoral approach in all but n = 1 patients, in which a hybrid, trans-aortic approach was chosen due to the small size of the patient. To determine the incidence of conduction abnormalities after TAVR, patients with CHB and previous pacemaker were excluded. This study was approved by the local IRB.
Prosthesis Sizing
Valve size and type were selected by the primary interventionalist and was based on careful pre-procedural measurements obtained via multi-slice computed tomography and the manufacturer’s recommendations, along with intra-procedural angiography.
Definition of Risk Factor Variables
Risk factor variables were drawn from seminal studies on TAVR in the adult population and included: (1) pre-existing RBBB, (2) membranous septum length, (3) implantation depth, (4) membranous septum length relative to implantation depth (∆MSID) where a value less than one indicates implantation below the membranous septum, and (5) oversizing ratio—two separate variables for the ratio of the prosthesis diameter relative to (1) the aortic valve (AoV) and, (2) the LVOT diameters, respectively.
To capture risk factors which may be unique to our patient population, we included the following novel factors as part of our risk model:
1) Prior aortic valve intervention >1 (including both transcatheter and surgical), and
2) Pre-existing—or baseline—conduction abnormalities (types listed in Table 1 under baseline CA).
Table 1.
Baseline patient characteristics
| All (n = 28) | Post-TAVR CA | No CA | p-value | |
|---|---|---|---|---|
| Total no. patients, n | 28 | 9 | 19 | |
| Age range, years (median) | 14.9 (3.5–22) | 17.4 (15.1, 18.7) | 14.4 (12.2, 15.6) | 0.015 |
| Female (%) | 14 (50%) | 4 (44.4%) | 10 (52.6%) | 1.000 |
| Race, non-Hispanic white | 24 (85.7%) | 9 (100%) | 15 (78.9%) | 1.000 |
| BSA, median (IQR) | 1.6 (1.4–1.8) | 1.7 (1.6, 1.8) | 1.6 (1.4, 1.7) | 0.099 |
| Congenital heart disease€ | 28 (100%) | 9 (100%) | 19 (100%) | |
| Isolated aortic valve disease | 12 (42.8%) | 5 (55.6%) | 7 (36.8%) | 0.432 |
| Multilevel obstructive left-sided disease | 8 (28.6%) | 2 (22.2%) | 6 (31.6%) | 1.000 |
| Complex CHD | 9 (32.1%) | 2 (22.2%) | 7 (36.8%) | 0.670 |
| Prior transcatheter or surgical AoV intervention | ||||
| Only 1 prior transcatheter | 5 (17.9%) | 3 (33.3%) | 2 (10.5%) | 0.290 |
| Only 1 prior surgical | 1 (3.6%) | 0 (0%) | 1 (5.3%) | 1.000 |
| Multiple AoV interventionsƚ | 17 (60.7%) | 4 (44.4%) | 13 (68.4%) | 0.223 |
| Multiple transcatheter¥ | 3 (10.7%) | 1 (11.1%) | 2 (10.5%) | 1.000 |
| Multiple surgical | 9 (32.1%) | 1 (11.1%) | 8 (42.1%) | 0.195 |
| Prior sternotomies > 1 | 13 (46.4%) | 2 (22.2%) | 11 (57.9%) | 0.148 |
| Baseline coronary anomalies | 4 (14.3%) | 1 (11.1%) | 3 (15.8%) | 1.000 |
| History of coronary reimplantation | 1 (5%) | 0 (0%) | 1 (5.3%) | 1.000 |
| Anomalous origin | 2 (7.1%) | 0 (0%) | 2 (10.5%) | 1.000 |
| Dilated coronary | 1 (5.3%) | 1 (11.1%) | 0 (0%) | 0.321 |
| Comorbidities | ||||
| Genetic syndrome | 2 (7.1%) | 1 (11.1%) | 1 (5.3%) | 1.000 |
| Liver dysfunction | 1 (5.3%) | 0 (0%) | 1 (5.3%) | 1.000 |
| Renal dysfunction | 1 (5.3%) | 0 (0%) | 1 (5.3%) | 1.000 |
| Neurological disorder | 4 (14.3%) | 0 (0%) | 4 (21.1%) | 0.273 |
| Patients with baseline CA* | 16 (57.1%) | |||
| Types of baseline CA (n = 16) | 20 | |||
| PR prolongation | 4 (20%) | |||
| NIVCD | 5 (25%) | |||
| Incomplete RBBB | 4 (20%) | |||
| RBBB | 5 (25%) | |||
| Incomplete LBBB | 1 (5%) | |||
| LBBB | 1 (5%) |
Values expressed as median (IQR) or n (%)
€One patient qualified as both multilevel obstructive left-sided disease and complex CHD
ƚEither transcatheter or surgical AoV interventions
¥Excluding surgical AoV interventions
¶Excluding transcatheter AoV interventions
*Some patients exhibited multiple levels of conduction abnormalities
CA Conduction Abnormalities, BSA Body Surface Area, AoV Aortic Valve, CHD Congenital Heart Disease
Procedure-specific data were also collected but were not included as primary risk factors in the forest plot. These variables included the incidence of valve-in-valve procedures, whether the surgical bioprosthetic valve frame was intentionally fractured prior to valve implantation, the final valve diameter at time of implantation, and the presence of device malposition during the case.
Multislice Computed Tomography Data Analysis
Multi-slice computed tomography (MSCT) datasets from patients were transmitted to a dedicated workstation for multiplanar reconstructions. The membranous septum length was measured from the aortic annular plane to the most superior portion of the muscular interventricular septum in the coronal view as described in prior studies [9, 16]. From this view, the AoV annular diameter and the LVOT diameter (within 2–5 mm of the aortic valve annular plane) were measured. Oversizing ratio was calculated as a ratio of prosthesis diameter to both the LVOT and the aortic valve annular diameters, respectively, as measured on MSCT images. No correction was performed for eccentricity.
Determination of Implantation Depth
The implantation depth was determined via angiographic data measured offline using Siemens Syngo Dynamics Workplace (Siemen Healthcare GmbH, Siemens Medical Solutions USA Inc.). The measurements were obtained, as described in prior studies [17], as the distance from base of the noncoronary cusp to the prosthesis stent inflow from aortic root injection post-TAVR final deployment. Angiographic data were missing for two patients. Implantation depth was reported both as an absolute measurement (mm) and as a ratio of this angiographic measurement (i.e., valve projection into the LVOT) to the membranous septum length as measured on pre-procedure MSCT (see prior section).
Electrocardiographic and Arrhythmia Analysis
Twelve or 15-lead electrocardiograms (ECGs) were recorded on admission, immediately post-procedure, before discharge, and at follow-up. These were reviewed by 2 physicians (CN, SYA) according to recommendations issued by the 2009 American Heart Association/American College of Cardiology Foundation, and Heart Rhythm Society [18]. Patients’ medical records were reviewed for documentation of arrhythmias while inpatient, as telemetry data is not stored after discharge.
Echocardiographic Analysis
Echocardiograms performed at three time-points: (1) pre-procedure, (2) during post-procedure admission, and (3) at follow-up visit (1–5 months) were reviewed. Patient-specific, pre-procedural, and intra-procedural parameters were collected, including left ventricular (LV) dysfunction, LV end diastolic dimension and Z-score, presence of LV hypertrophy, aortic valve peak and mean gradients, as well as the degree of valvar and paravalvar insufficiency.
Statistical Analysis
Demographics and clinical outcomes of interest were summarized using mean and standard deviation (SD) or median and interquartile range (IQR) for continuous variables. Counts and percentages were reported for categorical variables. Given the small sample size, hypothesis testing was not conducted. Instead, 83.7% confidence intervals (CIs) were generated for continuous variables with Wilson score-based method [19]. Binary variables were represented by their proportion of events and 83.7% CI of the proportion. Two groups were statistically non-significant (p-value > 0.05) if their corresponding 83.7% CI intervals overlap. Statistical significance was assessed at the 0.05 level. Statistical analyses were implemented using R v. 4.0.3 [20].
Results
Patient Demographics
Over the study period, 29 patients underwent TAVR—using the Edwards Sapien 3 (Edwards Lifesciences, Irvine, CA, USA) balloon-expandable valve. One patient with pre-existing AVB and PPM was excluded. In the study population (age range 3.5–22 years [median 14.9 years], 50% male), 42.9% had isolated AoV disease, and the remaining had left-sided cardiac lesions in-series, or complex CHD. The majority of patients had undergone at least one prior transcatheter (n = 18, 64%) or surgical (n = 15, 53.6%]) AoV intervention. A minority of patients had baseline coronary anomalies (n = 4, 14.3%), but without evidence of coronary stenosis. Two patients (7.1%) had genetic syndromes: Mosaic Turner syndrome and 22q11.2 deletion syndrome. Patients with new-onset post-TAVR conduction abnormalities were older (median 17.4 vs. 14.4 years, p = 0.015) but otherwise there were no statistical differences in pre-procedural characteristics between patients with and without post-TAVR conduction abnormality. Comprehensive patient demographics are listed in Table 1.
Conduction Characteristics Pre- and Post-TAVR
At baseline, 16 (57.1%) patients demonstrated conduction abnormalities (Table 1). Conduction abnormalities included 1st degree AVB (n = 4, 20%), nonspecific intraventricular conduction delay (NIVCD, n = 5, 25%), incomplete RBBB (iRBBB, n = 4, 20%), RBBB (n = 5, 25%), incomplete LBBB (iLBBB, n = 1, 5%), and LBBB (n = 1, 5%). Note that four patients had multilevel abnormalities consisting of 1st degree AVB in addition to an intraventricular delay vs. bundle branch block (see Table 1).
Immediately following TAVR, 9/28 (32.1%) patients had new-onset conduction abnormalities, two of whom (22.2%) had multiple levels of conduction abnormalities (Table 2). First degree AVB (n = 4) and LBBB (n = 6) were the most frequent acquired conduction abnormalities, with one instance each of 2nd and 3rd degree AVB. Two patients had late-onset LBBB which was first noted one-week post-TAVR.
Table 2.
Post-TAVR conduction abnormalities (CA)
| Patients with CA post-TAVR | 9 (32.1%) |
|---|---|
| Types of post-TAVR CA* | |
| PR prolongation | 4 (33.3%) |
| 2nd degree, type 2 AVB | 1 (8.3%) |
| 3rd degree AVB | 1 (8.3%) |
| LBBB | 6 (50.0%) |
| Patients with multilevel CA | 2 (22.2%) |
| PR prolongation and LBBB (n = 1) | |
| PR prolongation, 2nd degree, type 2 AVB, and LBBB (n = 1) | |
| Patients with late-onset CA (> 1 week post-TAVR) LBBB | 2 (22.2%) |
| Patients without follow-up at our institution | 2 (22.2%) |
| Resolution of CAƚ | |
| PR prolongation | 3/3 (100%) |
| 2nd degree, type 2 AVB⁋ | 1/1 (100%) |
| 3rd degree AVB¥ | 0/1 (0%) |
| LBBB | 3/5 (60.0%) |
Values expressed as n (%)
NIVCD Non-Specific Intraventricular Conduction delay; RBBB Right Bundle Branch Block, LBBB Left Bundle Branch Block, AVB Atrioventricular Block
*Some patients exhibited multiple levels of conduction abnormalities
ƚTwo patients had late-onset LBBB at initial follow-up post-TAVR with resolution within 1 month, n = 1, and n = 1 was lost to follow-up
⁋Resolution on Holter completed at 1 month follow-up
¥Patient undergoing valve-in-valve procedure (prior valve: Sapien XT), presence of baseline LBBB (QRS 160 ms), prior sternotomies: 3, prior transcatheter AoV interventions: 6, history of two prior mechanical MVR
Among patients with follow-up data available (n = 26), 100% of 1st and 2nd degree AVB, and 60% of LBBB, had resolved by one-year post-procedure. Persistent conduction abnormalities were seen in 3 patients and included LBBB (n = 2) and 3rd degree AVB requiring PPM the following day (n = 1, 3.6% total cases). The latter patient had a valve-in-valve procedure which included overexpansion of their prior Sapien XT, in addition to baseline LBBB (QRS 160 ms) with a history of 3 prior sternotomies, 6 transcatheter AoV interventions, 2 prior mechanical mitral valve replacements.
Of the 4 patients with multiple conduction abnormalities at baseline, only one patient experienced new post-TAVR conduction abnormality—1st degree AVB—which then resolved at follow-up.
Risk Factors for Conduction Abnormalities Post-TAVR
Due to the small study population, we used a forest plot to assess for the effect of risk factors on the primary outcome of any new conduction abnormalities including PPM. There was no significant association between post-TAVR conduction abnormality incidence and the risk factors reported in the adult literature, including presence of baseline RBBB, membranous septum length, valve implantation depth, or degree of valve oversizing– relative to both valve annulus and LVOT. Additionally, there was no relationship between outcome and any degree of baseline conduction abnormality or history of > 1 prior surgical or transcatheter AoV intervention. See Fig. 1 and Table 3.
Fig. 1.
Forest plot of risk factors for post-TAVR conduction abnormalities (CA)
Table 3.
Risk factors for conduction abnormalities (CA)
| Post-TAVR CA (n = 9) | No CA (n = 19) | p-value | |
|---|---|---|---|
| Baseline Characteristics | |||
| Baseline RBBB, n | 1 (11.1%) | 4 (21.1%) | 1.000 |
| Pre-TAVR AoV diameter†, mm | 26.1 (3.2) | 24.7 (5.0) | 0.316 |
| Pre-TAVR LVOT diameter†, mm | 27.6 (3.1) | 27.3 (6.5) | 0.457 |
| Intra-Procedural Characteristics, n | |||
| ViV | 1 (11.1%) | 8 (42.1%) | 0.195 |
| Intentional BPV frame fracture | 0 (0%) | 4 (21.1%) | 0.338 |
| Sapien 3 Valve size | – | ||
| 20 mm | 0 (0%) | 4 (21.1%) | 0.195 |
| 23 mm | 2 (22.2%) | 8 (42.1%) | – |
| 26 mm | 2 (22.2%) | 3 (15.8%) | – |
| 29 mm | 5 (55.6%) | 4 (21.1%) | – |
| Device Malposition | 0 (0%) | 0 (0%) | N/A |
| Device Positioning Characteristics | |||
| Membranous septum length†, mm | 9.9 (4.5) | 8.7 (2.7) | 0.664 |
| Implantation depth* | 8.1 (3.3) | 7.7 (3.1) | 0.782 |
| ΔMSID | 1.1 (0.6) | 1.1 (0.6) | 0.843 |
| Degree of valve oversizing | |||
| Prosthesis d./LVOT d.† | 1.0 (0.1) | 0.9 (0.2) | 0.211 |
| Prosthesis d./AoV annulus d.† | 1.0 (0.1) | 1.0 (0.2) | 0.455 |
Values expressed as mean (SD) or n (%)
†CTA
*Angiography
AoV Aortic Valve, LVOT Left Ventricular Outflow Tract, ViV Valve-In-Valve, BPV Bioprosthetic Valve, ΔMSID Ratio of Membranous Septum Length to the Implantation Depth, a value < 1 indicates implantation below the membranous septum; d diameter
There was no measurable difference between groups with and without post-TAVR conduction abnormalities in terms of native vs valve-in-valve procedures, intentional bioprosthetic valve frame fracture, or the absolute size to which the valve was dilated. There were no incidences of valve malposition or embolization Table 3.
Echocardiographic Data
Prior to TAVR, 3 (10.7%) patients demonstrated LV dysfunction (EF < 55%) on echocardiogram and more than half of subjects (n = 16, 57.1%) had LV hypertrophy. At follow-up several months after TAVR (median 98 days, IQR 28.5, 139.5 days), LV dysfunction had resolved in all patients, LV hypertrophy had resolved in 50% (n = 8), and the aortic valve mean gradient improved from a median of 42 (IQR 34.5, 56.5) to 16.5 mmHg (IQR 13.8, 24.0). No patients had significant AI and only one patient had greater than mild paravalvar leak (Table 4).
Table 4.
Echocardiographic data before and after TAVR
| Pre-TAVR (n = 28) | Immediately Post-TAVR (n = 28) | Follow-up (n = 27) | |
|---|---|---|---|
| Days before/after procedure | 66.0 (105.8, 31.5) | 1 (1.0, 1.0) | 98 (28.5, 139.5) |
| LV dysfunction (EF < 55%) | 3 (10.7%) | 3 (10.7%) | 0 (0%) |
| 40–55% | 1 (3.6%) | 2 (7.1%) | 0 (0%) |
| 30–39% | 1 (3.6%) | 1 (3.6%) | 0 (0%) |
| < 30% | 1 (3.6%) | 0 (0%) | 0 (0%) |
| LV size (LVIDd, cm) | 5.2 (4.7, 5.7) | 4.8 (4.3, 5.3) | 4.9 (4.3, 5.4) |
| LV size, indexed* | 1.4 (−0.1, 2.4) | 0.1 (−0.8, 1.7) | 0.2 (−1.1, 1.2) |
| LV hypertrophy (Y) | 16 (57.1%) | 12 (42.9%) | 8 (29.6%) |
| If yes, LV mass Z-score | 4.4 (3.6, 5.1) | 3.2 (2.8, 4.4) | 4.2 (2.6, 5.3) |
| AoV stenosis (Y) | 24 (85.7%) | 23 (82.1%) | 25 (92.6%) |
| AoV mean gradient | 42.0 (34.5, 56.5) | 15.5 (12.0, 23.0) | 16.5 (13.8, 24.0) |
| AoV peak gradient | 71.5 (58.5, 100.0) | 26.0 (21.2, 35.0) | 25.5 (20.0, 34.5) |
| AoV insufficiency | |||
| None | 2 (7.1%) | 21 (75%) | 18 (66.7%) |
| Trace | 1 (3.6%) | 7 (25%) | 8 (29.6%) |
| Mild | 5 (17.9%) | 0 (0%) | 1 (3.7%) |
| Mild to moderate | 7 (25%) | 0 (0%) | 0 (0%) |
| Moderate | 8 (28.6%) | 0 (0%) | 0 (0%) |
| Moderate to severe | 4 (14.3%) | 0 (0%) | 0 (0%) |
| Severe | 1 (3.6%) | 0 (0%) | 0 (0%) |
| Paravalvar leak | |||
| None | 25 (89.3%) | 23 (82.1%) | 22 (81.5%) |
| Trace | 0 (0%) | 1 (3.6%) | 2 (7.4%) |
| Mild | 1 (3.6%) | 3 (10.7%) | 2 (7.4%) |
| Mild to moderate | 1 (3.6%) | 1 (3.6%) | 0 (0%) |
| Moderate | 0 (0%) | 0 (0%) | 1 (3.7%) |
Values expressed as median (IQR) or n (%)
*LVIDd Z-score
LV Left Ventricle, EF Ejection Fraction, Y Yes or Present, AoV Aortic Valve
Arrhythmias and Ectopy Pre- and Post-TAVR
Three patients had documented arrhythmias prior to TAVR including one with atrial ectopic tachycardia (AET), a second individual with intra-atrial reentrant tachycardia and atrial fibrillation (AF), and a third subject with history of intermittent accelerated ventricular rhythm (AVR). The patient with AVR was treated with low dose nadolol prior to TAVR, but other patients were not started on antiarrhythmics prior to TAVR.
Following TAVR, no patients experienced clinically significant arrhythmias including no sustained, hemodynamically significant, or frequent arrhythmias. Transient post-procedure arrhythmias were seen in 17.8% (n = 5) of patients including AVR or nonsustained ventricular tachycardia (n = 3, 10.7%), and AET (n = 2, 7.1%). Holter monitors were performed in 39.3% (n = 11) of patients. No new conduction abnormalities were identified on post-discharge ambulatory ECG monitoring.
Short to Mid-Term Outcomes
At follow-up (median 98 days), up to one-year, there were no new or progressive conduction abnormalities, reinterventions, or deaths.
Discussion
In our pediatric and young adult population with CHD undergoing TAVR, CHB and the need for PPM was rare, with an incidence (3.6%), lower than that reported in adult patients (4–24%) [5, 13, 21, 22]. Less advanced conduction abnormalities sustained following TAVR largely resolved within a year. Selected risk factors well-described in adult patients [9, 14, 23, 24], including baseline presence of RBBB, membranous septum length, implantation depth, and oversizing ratio, were not significantly associated with PPM (n = 1) or new conduction abnormalities (n = 9) in our population. Additionally, pre-procedural risks including any degree of baseline conduction abnormalities and > 1 AoV interventions, were not associated with new post-TAVR conduction abnormality. To our knowledge, this is the first study in pediatric and young adult patients, and in those with complex CHD, assessing electrical consequences following TAVR.
To date, studies investigating predictors for post-TAVR PPM have exclusively involved adult populations. TAVR patients have a mean age of 70–80 years, with high rates of baseline subaortic and leaflet calcification, and thus represent a distinctly different patient population than children and young adults with CHD [4, 13, 22]. Studies focusing on coronary artery calcium deposition in the “young” (i.e., ≤ 45 years old) adult population have excluded those with CHD [25]. Furthermore, the molecular mechanisms of valve disease may be different—i.e., excessive extracellular matrix deposition without calcification—between adult and pediatric aortic valve stenosis [26]. This has led to TAVR risk-stratification systems specific to the unaltered aging process, including the “calcification score” which quantifies valvar calcium deposition [27].
The existing data creates a map of risk factors for post-TAVR PPM in which procedure-specific, cardiac-specific, and aging-specific factors are interwoven and inextricable—highlighting the need for focused research in the pediatric and young adult population. We lack appropriate pediatric risk models for post-TAVR CHB. This study establishes a baseline for this specific patient population and offers an initial reference point to create a foundational framework for developing risk models.
Longer-term prediction tools with improved accuracy and applicability will become increasingly relevant for the pediatric cardiologist when counseling younger patients and their families regarding the various treatment options for aortic valve disease. It has been almost a decade since the FDA approved the TAVR procedure, initially restricted to patients with high surgical risk. Although use has expanded to patients < 65 years old with low surgical risk, its use remains relatively rare in the pediatric population. We are beyond the need for studies of procedural efficacy and must pivot to the long-term effects of the TAVR procedure, including valvar longevity and the morbidity and mortality associated with eventual explantation [28]. Patients undergoing TAVR at a young age anticipate decades of survival, highlighting the importance of longitudinal data. Therefore, long-term safety and efficacy data—especially in these evolving young adult populations—could be practice-changing.
The pediatric cardiac conduction system differs from that of older adults in several ways, including faster conduction velocity, shorter refractory periods, and less fibrosis and sclerosis. Acquired injury to the conduction system from ischemia or invasion of aortic valve calcification into the His bundle and infrahisian conduction system are more common in adults. Even in those who did experience post-TAVR conduction abnormalities in our series, the changes were almost always transient. Additionally, the younger patients were less likely to have new conduction abnormalities, which may be secondary to the health of the conduction system or the compliance and adaptability of the LVOT in younger children. No patients experienced progressive conduction abnormality at final follow-up visit.
In addition to identifying risk factors, the literature in adult patients aims to dictate who should receive an ambulatory ECG monitor at the time of discharge. However, in our population, none of the remote monitors (n = 11) demonstrated new evidence of abnormal conduction, despite a skew towards placement of a monitor on patients who experienced immediate post-procedural arrhythmias. Additionally, the benign, transient nature of the arrhythmias that were identified suggest that there may be no routine indication for monitors in this younger population. Future research will help to define these practice recommendations.
Limitations
This study was limited by the small population size (n < 30), which prohibited performing a traditional regression analysis of risk factor modification of PPM and conduction abnormality incidence.
Data accuracy for some variables was limited by the retrospective study design. Intra-procedural documentation typically does not contain granular data regarding periprocedural issues with conduction abnormalities or arrhythmias unless they were persistent and significant. Unreported brief abnormalities likely have no clinical significance. Ultimately, detailed evaluation of the periprocedural, dynamic effect of TAVR placement on conduction, as described in more recent papers [29, 30], was not plausible and warrants further study in this population. Likewise, measurement of implantation depth on angiography was challenging in this pediatric cohort, where the priority is often given to limiting contrast exposure making it challenging to discern the border of the NCC for measurements.
Finally, this was a retrospective study with limited follow-up. By necessity, patients requiring TAVR in childhood or young adulthood will be committed to a greater number of life-time interventions than an elderly patient. The presence of CHD and prior surgical palliation likely predisposes patients to earlier-onset, and increased incidence, of calcific valve disease, potentially further shortening the lifespan of a bioprosthetic valves [31]. It is unknown how earlier TAVR will impact TAVR versus surgical valve candidacy in late adulthood, when adult risk factors coalesce with risks secondary to chronic heart disease. Further longitudinal studies are needed to fully elucidate the long-term outcomes in this population.
Conclusion
Permanent heart block and PPM placement were rare in a population of children and young adults undergoing TAVR for treatment of CHD. Presence of baseline RBBB, membranous septum length, implantation depth, oversizing ratio, baseline conduction abnormalities, and > 1 prior aortic valve intervention were not associated with new post-TAVR conduction abnormality.
Supplementary Information
Below is the link to the electronic supplementary material.
Acknowledgements
Study data were collected and managed using REDCap electronic data capture tools hosted at the University of Utah.
Author Contribution
C.N. and Y.A. primarily wrote the manuscript, but all authors contributed throughout and provided invaluable feedback. C.N., Y.A, and M.M. created Fig. 1. E.B. completed the CTA measurements. C.N. and Y.A. reviewed the ECGs. Z.O. completed the statistics. All authors reviewed the manuscript.
Data Availability
No datasets were generated or analysed during the current study.
Declarations
Conflict of interests
Dr. Pilcher has served as a consultant for Abbott. Dr. Asaki has served as a consultant for Philips. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose.
Ethical Approval
This study received proper ethical oversight from the University of Utah institutional review board.
Consent for Publication
None of the paper’s contents have been previously published.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Leon MB, Smith CR, Mack M et al (2010) Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. N Engl J Med 363(17):1597–1607. 10.1056/NEJMoa1008232 [DOI] [PubMed] [Google Scholar]
- 2.Mack MJ, Leon MB, Thourani VH et al (2023) Transcatheter Aortic-Valve replacement in low-risk patients at 5 years. N Engl J Med 389(21):1949–1960. 10.1056/NEJMoa2307447 [DOI] [PubMed] [Google Scholar]
- 3.Sinha S, Khan A, Qureshi AM et al (2020) Application of transcatheter valves for aortic valve replacement in pediatric patients: a case series. Catheter Cardiovasc Interv 95(2):253–261. 10.1002/ccd.28505 [DOI] [PubMed] [Google Scholar]
- 4.Sammour Y, Krishnaswamy A, Kumar A et al (2021) Incidence, predictors, and implications of permanent pacemaker requirement after transcatheter aortic valve replacement. JACC Cardiovasc Interv 14(2):115–134. 10.1016/j.jcin.2020.09.063 [DOI] [PubMed] [Google Scholar]
- 5.Lilly SM, Deshmukh AJ, Epstein AE et al (2020) 2020 ACC expert consensus decision pathway on management of conduction disturbances in patients undergoing transcatheter aortic valve replacement: a report of the American college of cardiology solution set oversight committee. J Am Coll Cardiol 76(20):2391–2411. 10.1016/j.jacc.2020.08.050 [DOI] [PubMed] [Google Scholar]
- 6.Droppa M, Rudolph TK, Baan J et al (2020) Risk factors for permanent pacemaker implantation in patients receiving a balloon-expandable transcatheter aortic valve prosthesis. Heart Vessels 35(12):1735–1745. 10.1007/s00380-020-01653-6 [DOI] [PubMed] [Google Scholar]
- 7.Husser O, Pellegrini C, Kessler T et al (2016) Predictors of permanent pacemaker implantations and new-onset conduction abnormalities with the SAPIEN 3 Balloon-Expandable transcatheter heart valve. JACC Cardiovasc Interv 9(3):244–254. 10.1016/j.jcin.2015.09.036 [DOI] [PubMed] [Google Scholar]
- 8.Kawashima T, Sato F (2014) Visualizing anatomical evidences on atrioventricular conduction system for TAVI. Int J Cardiol 174(1):1–6. 10.1016/j.ijcard.2014.04.003 [DOI] [PubMed] [Google Scholar]
- 9.Hamdan A, Guetta V, Klempfner R et al (2015) Inverse relationship between membranous septal length and the risk of atrioventricular block in patients undergoing transcatheter aortic valve implantation. JACC Cardiovasc Interv 8(9):1218–1228. 10.1016/j.jcin.2015.05.010 [DOI] [PubMed] [Google Scholar]
- 10.Tsoi M, Tandon K, Zimetbaum PJ, Frishman WH (2021) Conduction disturbances and permanent pacemaker implantation after transcatheter aortic valve replacement: predictors and prevention. Cardiol Rev. 10.1097/CRD.0000000000000398 [DOI] [PubMed] [Google Scholar]
- 11.Maier O, Piayda K, Afzal S et al (2021) Computed tomography derived predictors of permanent pacemaker implantation after transcatheter aortic valve replacement: a meta-analysis. Catheter Cardiovasc Interv 98(6):E897–E907. 10.1002/ccd.29805 [DOI] [PubMed] [Google Scholar]
- 12.Tretter JT, Mori S, Anderson RH et al (2019) Anatomical predictors of conduction damage after transcatheter implantation of the aortic valve. Open Heart 6(1):e000972. 10.1136/openhrt-2018-000972 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Siontis GC, Juni P, Pilgrim T et al (2014) Predictors of permanent pacemaker implantation in patients with severe aortic stenosis undergoing TAVR: a meta-analysis. J Am Coll Cardiol 64(2):129–140. 10.1016/j.jacc.2014.04.033 [DOI] [PubMed] [Google Scholar]
- 14.Nazif TM, Dizon JM, Hahn RT et al (2015) Predictors and clinical outcomes of permanent pacemaker implantation after transcatheter aortic valve replacement: the PARTNER (Placement of AoRtic TraNscathetER Valves) trial and registry. JACC Cardiovasc Interv. 10.1016/j.jcin.2014.07.022 [DOI] [PubMed] [Google Scholar]
- 15.Kiani S, Kamioka N, Black GB et al (2019) Development of a risk score to predict new pacemaker implantation after transcatheter aortic valve replacement. JACC Cardiovasc Interv 12(21):2133–2142. 10.1016/j.jcin.2019.07.015 [DOI] [PubMed] [Google Scholar]
- 16.Chen YH, Chang HH, Liao TW et al (2020) Membranous septum length predicts conduction disturbances following transcatheter aortic valve replacement. J Thorac Cardiovasc Surg. 10.1016/j.jtcvs.2020.07.072 [DOI] [PubMed] [Google Scholar]
- 17.Aslan S, Demir AR, Celik O et al (2020) Usefulness of membranous septum length in the prediction of major conduction disturbances in patients undergoing transcatheter aortic valve replacement with different devices. Kardiol Pol 78(10):1020–1028. 10.33963/KP.15538 [DOI] [PubMed] [Google Scholar]
- 18.Surawicz B, Childers R, Deal BJ et al (2009) AHA/ACCF/HRS recommendations for the standardization and interpretation of the electrocardiogram: part III: intraventricular conduction disturbances: a scientific statement from the American Heart Association Electrocardiography and Arrhythmias Committee, Council on Clinical Cardiology; the American college of cardiology foundation; and the heart rhythm Society. Endorsed by the International society for computerized electrocardiology. J Am Coll Cardiol 53(11):976–981. 10.1016/j.jacc.2008.12.013 [DOI] [PubMed] [Google Scholar]
- 19.Goldstein H, Healy MJ (1995) The graphical presentation of a collection of means. J Royal Stat Soc: Series A (Statistics in Society) 158(1):175–177. 10.2307/2983411 [Google Scholar]
- 20.A language and environment for statistical computing. R Foundation for Statistical Computing. https://www.R-project.org/
- 21.van Rosendael PJ, Delgado V, Bax JJ (2018) Pacemaker implantation rate after transcatheter aortic valve implantation with early and new-generation devices: a systematic review. Eur Heart J 39(21):2003–2013. 10.1093/eurheartj/ehx785 [DOI] [PubMed] [Google Scholar]
- 22.Ullah W, Zahid S, Zaidi SR et al (2021) Predictors of permanent pacemaker implantation in patients undergoing transcatheter aortic valve replacement—a systematic review and meta-analysis. J Am Heart Assoc 10(14):e020906. 10.1161/JAHA.121.020906 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Mauri V, Reimann A, Stern D et al (2016) Predictors of permanent pacemaker implantation after transcatheter aortic valve replacement with the SAPIEN 3. JACC Cardiovasc Interv 9(21):2200–2209. 10.1016/j.jcin.2016.08.034 [DOI] [PubMed] [Google Scholar]
- 24.Maan A, Refaat MM, Heist EK et al (2015) Incidence and predictors of pacemaker implantation in patients undergoing transcatheter aortic valve replacement. Pacing Clin Electrophysiol 38(7):878–886. 10.1111/pace.12653 [DOI] [PubMed] [Google Scholar]
- 25.Mortensen MB, Dzaye O, Bodtker H et al (2021) Interplay of risk factors and coronary artery calcium for CHD risk in young patients. JACC Cardiovasc Imaging 14(12):2387–2396. 10.1016/j.jcmg.2021.05.003 [DOI] [PubMed] [Google Scholar]
- 26.Clift CL, Saunders J, Drake RR, Angel PM (2022) Perspectives on pediatric congenital aortic valve stenosis: extracellular matrix proteins, post translational modifications, and proteomic strategies. Front Cardiovasc Med 9:1024049. 10.3389/fcvm.2022.1024049 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Maeno Y, Abramowitz Y, Kawamori H et al (2017) A highly predictive risk model for pacemaker implantation after TAVR. JACC Cardiovasc Imaging 10:1139–1147. 10.1016/j.jcmg.2016.11.020 [DOI] [PubMed] [Google Scholar]
- 28.Bapat VN, Zaid S, Fukuhara S et al (2021) Surgical explantation after TAVR failure: mid-term outcomes from the EXPLANT-TAVR International registry. JACC Cardiovasc Interv 14(18):1978–1991. 10.1016/j.jcin.2021.07.015 [DOI] [PubMed] [Google Scholar]
- 29.Bourenane H, Galand V, Boulmier D et al (2021) Electrophysiological study-guided permanent pacemaker implantation in patients with conduction disturbances following transcatheter aortic valve implantation. Am J Cardiol 149:78–85. 10.1016/j.amjcard.2021.03.014 [DOI] [PubMed] [Google Scholar]
- 30.Reiter C, Lambert T, Kellermair J et al (2021) Intraprocedural dynamics of cardiac conduction during transcatheter aortic valve implantation: assessment by simultaneous electrophysiological testing. Heart Rhythm 18(3):419–425. 10.1016/j.hrthm.2020.10.018 [DOI] [PubMed] [Google Scholar]
- 31.Halpern DG, Steigner ML, Prabhu SP, Valente AM, Sanders SP (2015) Cardiac calcifications in adults with congenital heart defects. Congenit Heart Dis Sep-Oct 10(5):396–402. 10.1111/chd.12243 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
No datasets were generated or analysed during the current study.