Cardiac lead perforation: Mechanisms, detection, and therapeutic approaches

Ameer Awashra 1,4,∗, Mohammed AbuBaha 1,4, Hossam Salameh 1, Hammam Jallad 1, Aya Milhem 1, Anwar Zahran 1, Aseel Badwan 2, Fathi Milhem 1, Abdalhakim Shubietah 3
PMCID: PMC12902149  PMID: 41695519

Abstract

Cardiac lead perforation is an infrequent but serious complication of cardiac implantable electronic devices, with possibly lethal consequences if left untreated. Because of its very varied clinical appearance and lack of established criteria, perforation still presents diagnostic and therapeutic hurdles despite continuous improvements in lead design and insertion procedures. This review summarizes current research on the processes, risk factors, and clinical range of lead perforation. Thin myocardial walls, especially at the right ventricular apex; active-fixation leads; and operator-dependent elements, such as lead placement and torque, are important anatomic and procedural contributions. Steroid usage, low body mass index, and female sex all increase risk. A strong index of suspicion is often required for prompt diagnosis because presentations vary from asymptomatic instances to severe tamponade. Device interrogation must be integrated with imaging modalities, particularly computed tomography and echocardiography, which each have unique benefits based on the clinical setting, to achieve accurate detection. From cautious observation in stable patients to immediate percutaneous or surgical intervention in unstable or worsening situations, management approaches are highly customized. New technologies, including subcutaneous implantable cardioverter-defibrillators and leadless pacemakers, have the potential to lower the incidence of perforations, but they also present new difficulties that need to be further investigated. Importantly, the lack of consistent diagnostic standards hinders study/ comparability and postpones the creation of guidelines. Standardizing diagnostic criteria, establishing multicenter prospective registries, developing machine learning-based risk stratification tools, and conducting clinical trials to inform the treatment of asymptomatic or delayed perforations are among the top goals for the future. To improve patient outcomes and advance clinical practice, international cooperation and uniform reporting standards are crucial.

Keywords: Cardiac lead perforation, Pacemaker lead perforation, Lead dislodgement, Device-related complications, Transvenous pacing, Surgical revision, Lead extraction

Graphical abstract

graphic file with name ga1.jpg


Key Findings.

  • ▪

    Cardiac lead perforation is a rare but serious complication of cardiac implantable electronic device implantation, influenced by factors such as lead type, implant site, age, sex, and body mass index.

  • ▪

    Diagnosis requires high clinical suspicion; computed tomography imaging provides the best accuracy for identifying perforation and lead position.

  • ▪

    Management depends on patient stability, ranging from monitoring in asymptomatic cases to urgent lead extraction in unstable patients.

  • ▪

    Active fixation leads and apical placement increase perforation risk, whereas septal placement and passive leads are safer for high-risk patients.

  • ▪

    There is a need for standardized diagnostic criteria, multicenter registries, and machine learning models to improve detection and guide management.

Since their debut, cardiac implanted electronic devices, such as implantable cardioverter-defibrillators (ICDs), permanent pacemakers, and cardiac resynchronization therapy systems, have revolutionized the treatment of heart failure and cardiac rhythm abnormalities.1 These devices have seen substantial improvement in terms of usefulness, safety, and miniaturization since they were first created in the middle of the 20th century.2 With more than 1.5 million new implantations performed worldwide each year, these devices have become crucial therapeutic instruments.3

Even with advancements in technology, lead-related problems are still common; cardiac lead perforation is an uncommon but potentially fatal occurrence.4 Despite advancements in design and implantation methods, this complication, which was first identified in the early days of cardiac pacing, continues to pose diagnostic and treatment difficulties.5 When an implanted lead breaks through the myocardium and enters the pericardial space or nearby tissues, it is known as a cardiac lead perforation.6 Temporally, this issue is categorized as delayed/late (>1 month), subacute (24 hours to 1 month), or acute (≤24 hours after implantation).6 Because it influences the clinical presentation, diagnostic process, and therapeutic strategy, this temporal categorization is crucial.7

Pacemaker and ICD leads have recorded incidences ranging from 0.1% to 0.8%8 and 0.6% to 5.2%,8 respectively. The higher incidence rates for ICD leads are ascribed to their larger diameter and greater stiffness; cardiac resynchronization therapy systems could be involved, albeit there is currently little information available on these leads because of variations in lead location and setup.9

Pneumothorax, hemothorax, pericardial effusion, cardiac tamponade, and device malfunction are among the clinically significant outcomes that go beyond the acute procedural problems.10 Diagnostic delays may result from these problems’ presentations, which might mirror those of other illnesses and vary from asymptomatic patients to life-threatening situations.11 Depending on time, clinical presentation, hemodynamic stability, lead characteristics, and institutional competence, management strategies might vary from conservative monitoring to percutaneous lead extraction or surgical intervention.10

A mix of imaging modalities, such as computed tomography, transthoracic or transesophageal echocardiography, and chest radiography, is necessary for an accurate diagnosis.12 Depending on the clinical situation, each modality has a different diagnostic value.4

To raise clinician awareness, improve detection, optimize care, and lower related morbidity and mortality, this review summarizes the state of knowledge about cardiac lead perforation, with a focus on causes, risk factors, diagnostic techniques, and therapeutic measures.7

Anatomic considerations and risk assessment

Cardiac anatomy relevant to lead placement

The right ventricle (RV), which has 3 separate compartments, inlet, outflow, and apical, is the main chamber used for ventricular lead implantation. Every location has distinct features that influence the stability of lead and the likelihood of perforation.13

From the tricuspid valve to the papillary muscle insertions, the RV inlet is distinguished by conspicuous trabeculations that provide stability for passive-fixation leads but may make lead navigation more difficult.13 Because it is linked to more physiological activation patterns than standard apical pacing, the RV outflow tract, which is situated between the supraventricular crest and the pulmonary valve, is being used more and more.14

Owing to its much thinner myocardial walls (typically <1–2 mm) than other segments, the RV apex, traditionally the most popular location for ventricular lead placement, has a much higher risk of perforation.13 Research shows that RV apical leads had a 3.4-fold increased risk of perforation compared with nonapical placements.15

A popular target for modern lead implantation is the interventricular septum (IVS), especially the midsection close to the moderator band. Greater myocardial wall thickness (usually 6–12 mm), a lower risk of perforation, and the possibility of more physiological ventricular activation patterns are all benefits of the septum.13 In addition to providing mechanical support, structures such as the moderator band and septomarginal trabecula may aid in precisely directing septal lead attachment (shown in Figure 1).

Figure 1.

Figure 1

Right-heart cross-section with color-coded perforation-risk overlays. Red marks the RV apex (<2 mm wall, very high risk); yellow marks the RV outflow tract (3–4 mm, moderate risk); orange marks the RA free wall (∼2 mm, moderate to high risk); green marks the IVS (6–12 mm, low risk). Icons indicate typical CIED lead positions: traditional RV apical, preferred midseptal, and atrial appendage sites. CIED = cardiac implantable electronic device; IVS = interventricular septum; RA = right atrium; RV = right ventricle; RVOT = right ventricular outflow tract.

There are special concerns for lead insertion in the right atrium (RA). Because pectinate muscles provide stability, the RA appendage has historically been favored. The free wall is more vulnerable to perforation, and the RA wall is typically thinner than the RV (∼2 mm).16

Patient-related anatomic risk factors

Perforation risk is greatly influenced by a number of patient-specific variables. These include both inherent heart traits and more general physiological differences across groups. Progressive myocardial fibrosis, fatty infiltration, and thinning with age are examples of age-related cardiac abnormalities that might modify tissue mechanical characteristics and perhaps make a person more vulnerable to perforation.17

Differences by gender have also been reported. Owing to their smaller heart chambers and thinner myocardial walls, female patients have a much greater incidence of perforations—2–3 times higher.18 Another crucial factor to consider is body habitus. Reduced myocardial wall thickness and lower heart dimensions are probably the causes of the high correlation between low body mass index (BMI) (<20 kg/m2) and increased perforation risk.19

Congenital (such as Ebstein’s abnormality) and acquired structural anomalies (such as ventricular aneurysms) provide special anatomic obstacles and may need altered lead placement techniques. Lead trajectory planning may also be affected by disorders such as hypertrophic cardiomyopathy or postsurgical remodeling, which may skew typical anatomic markers.20

Preprocedural anatomic assessment

Planning structural and electrophysiological cardiac treatments to improve safety and efficacy requires preprocedural anatomic examination.21 This entails choosing the right imaging modalities according to clinical requirements: computed tomography (CT) for detailed anatomic visualization and vascular access planning21; cardiac MRI for tissue characterization, particularly in detecting fibrosis or thinning22; and transthoracic echocardiography (TTE) for baseline heart function.21 For valvular evaluation, transesophageal or 3D echocardiography may provide more information.23 To guide the procedural approach, it is necessary to identify key anatomic hazards, such as myocardial infarction at proposed device locations, myocardial thinning (<1 mm), aneurysms, calcification, complicated valve architecture, thrombi, tumors, or problematic vascular access.21 To prevent difficulties, precise lead or device trajectory planning is necessary, and 3D reconstruction for ideal placement is made possible by CT or cardiac MRI.22 When doing cardiac implanted electronic device (CIED) operations, it is also important to assess the patient’s cardiac rhythm and device history.

Mechanisms of cardiac lead perforation

Owing to the nature of the ever-changing anatomy of the heart chambers with each cardiac cycle and respiratory motion, the interaction between cardiac leads and the myocardium is complex.24

Cardiac lead perforations can be classified according to the time it takes to develop signs of perforation after surgery, with acute perforations developing within the first 24 hours, early perforations developing within 30 days, and delayed perforations developing after 30 days, respectively, according to Kumar et al.19

Cameron et al24 attributed cardiac wall perforation to the material of the used leads, given that there have been experiments where silicone rubber leads, instead of polyurethane-coated leads, yielded less stiff leads. However, the use of this material resulted in thinner leads with a higher penetration pressure owing to the force exerted on the wall by the thinner lead, which is further supported by Gasser et al25 in their experiment where lower pressure points resulted from a wider diameter used.

Tissue-related factors include myocardial tissue stiffness and elasticity, which, according to Gasser et al,25 are not the same throughout chambers and even within the chamber itself, described as heterogeneity of the myocardial tissue.

Although RA wall perforation still occur a lot of the times, recent data on almost 2400 patients show that RV anterior wall perforation was more common than the RA wall perforation, which is surprising because the RA is thinner,26 explained by the fact that the anterior wall is thin and can be easily perforated, in contrast with the septum, which can be 2 times in thickness.27

Perforation of the cardiac lead or migration into the left side of the heart is less common. It can rarely happen owing to various etiologies such as an atrial septal defect28 and a patent foramen ovale29 or even through the IVS.30 Even though these complications seem serious, according to the literature, they can be asymptomatic for a long time and might be discovered incidentally.28

With the increased frequency of surgeries using the coronary sinus as an access to left ventricular lead insertion, the rates of coronary sinus perforation/dissection have been reported to be as high as 2%29 and even extend to cause pericarditis and/or cardiac tamponade.30

As stated earlier, lead design and fixation types play into the role of cardiac perforation. Two types of lead fixation exist: active fixation, where the lead is inserted or “screwed into” the endocardium, and passive fixation, where the lead is passively fixed in the RV’s apex, where trabeculations function as an anchoring site for the lead.19 It seems that the active-fixation type leads demonstrate higher rates of cardiac perforation.26 Most cardiac lead perforations occur at the cardiac apex because it is a common site of lead insertion31 (shown in Figure 2).

Figure 2.

Figure 2

Mechanisms of cardiac lead perforation. A: Comparison between active screw-in and passive tine lead designs and their interaction with endocardial tissue. B: Effect of material stiffness—polyurethane exerts focal pressure, whereas silicone distributes force more safely. C: Torque, slack, and postprocedural migration contribute to progressive tissue damage and perforation. RV = right ventricle.

Risk factors for lead perforation

Risk factors for lead perforation can vary and have both intrinsic and extrinsic causes, such as myocardial wall thickness and physician expertise, respectively. Sterlinski et al,32 in their retrospective analysis on 2247 leads in 1419 patients, have categorized factors that may be involved in the process of cardiac perforation into lead design, surgeon experience, and patient-specific factors.

The use of helical screw leads has been associated with a higher risk of perforation, which was consistent with the findings in Sterlinski’s study,32 which showed that all perforations that occurred were related to the use of retractable screw leads.33

Even in highly skilled and experienced physicians, certain errors can still occur, such as overtorquing and leaving an excessive loop, which is associated with an increased risk of cardiac perforations owing to the forward movement of the lead and the associated excessive tension applied.31,32

Various patient-specific factors are associated with an increased risk of cardiac perforations and were categorized by Mahapatra et al33 according to postimplant effusion where a univariate analysis on 4280 implanted permanent pacemakers showed 2 strongly associated predictors such as the concomitant use of temporary transvenous pacemaker (hazard ratio [HR] 3.2; 95% confidence interval [CI] 1.6–6.2; P = .001) or recent steroid use (HR 4.1; 95% CI 1.1–10; P = .003). Other patient-specific predictors were a BMI of less than 20 kg/m2 (HR 2.5) and old age (HR 1.4).33

Other procedural risk factors that are considered patient specific included temporary pacemaker wire use (HR 2.7; 95% CI 1.4–3.9; P = .01) and longer fluoroscopy times (HR 1.3).33 However, the only variable that was independently associated with a reduced risk of cardiac perforation was an RV systolic pressure of >35 mm Hg (HR 0.70; 95% CI 0.50–0.92; P = .02) according to Mahapatra et al.33

In a study conducted by Hsu et al34 of more than 440,000 first-time implantable cardioverter-defibrillator (ICD) patients, cardiac lead perforation occurred in just 0.14% of cases. Older age, being female, having heart failure, or receiving a more complex device increased the risk. In contrast, atrial fibrillation, diabetes, past bypass surgery, and having a more experienced implanter reduced the risk. Patients who experienced perforation were much more likely to have serious complications, have longer hospital stays, and even die during their hospital stay (shown in Table 1).

Table 1.

Comparative summary of risk factors for cardiac lead perforation

Category Risk factor Impact on perforation risk
Anatomic location13 Right ventricular apex High (thin myocardial wall <1 to 2 mm)
Interventricular septum Lower (thicker wall 6–12 mm)
Patient characteristics18,19 Female sex Higher risk (2–3 times more frequent)
Low BMI (<20 kg/m2) Increased risk owing to a smaller heart size
Lead Type35,36 Active-fixation lead Higher perforation incidence
Passive-fixation lead Lower risk in stable anatomy

BMI = body mass index.

Clinical presentation and diagnosis

The clinical range of cardiac lead perforation is wide, spanning from life-threatening cardiac tamponade to asymptomatic electrical abnormalities. The degree of myocardial breach, lead type, anatomic location, and date of commencement are some of the variables that affect its clinical variability. High clinical suspicion and a methodical diagnostic approach are necessary for timely diagnosis, owing to the possibly subtle indications, to prevent intervention delays and minimize negative effects.37,38

Symptomatology spectrum

Lead perforation symptoms may manifest in a wide range of ways. With 25%–70% of patients, chest discomfort is still the most often reported symptom. It is usually characterized as positional, pleuritic, and acute.4,33 Approximately 30% of patients have respiratory problems, including dyspnea, which may vary from minor discomfort during exercise to severe respiratory distress linked to pericardial effusion or tamponade.15 Both pacing failure and abrupt hemodynamic compromise from tamponade physiology may cause syncope and presyncope, especially in patients who are pacemaker dependent.8 Furthermore, diaphragmatic or intercostal muscle activation brought on by extracardiac stimulation via lead migration may clinically show up as hiccups or obvious twitches of the chest or abdominal wall.39,40

The symptom profile is also influenced by the time of the perforation. Acute perforations often manifest suddenly with symptoms including pericardial tamponade, hypotension, or acute chest discomfort, and they frequently happen intraoperatively or within 24 hours after device installation.41 Subacute perforations usually manifest as gradual, nonspecific symptoms that often mirror pulmonary or musculoskeletal disorders and develop over days to weeks after implantation.41,42 In contrast, delayed perforations, which are often discovered by chance during device interrogation or imaging tests performed for unrelated purposes, could not show clinical symptoms for weeks or even months after implantation.43

Crucially, up to 15%–30% of cardiac lead perforations could not cause any symptoms at all and can only be found by accident by imaging or regular device follow-up.12,44 During regular pacemaker or ICD interrogations, these quiet holes highlight the need to be alert when interpreting electrical anomalies.45

Physical examination findings

Even though physical symptoms are sometimes vague, a careful examination might provide important hints that indicate a lead perforation. Hypotension, tachycardia, and raised jugular vein pressure are all signs of pericardial tamponade and may indicate hemodynamic instability.46 Pulsus paradoxus, which is characterized as an inspiratory decrease in systolic blood pressure of more than 10 mm Hg, is one of the most distinctive physical findings in this setting. It implies inadequate ventricular filling as a result of pericardial pressure.47

When a pericardial effusion is present, cardiac auscultation may show muffled heart sounds.9 Furthermore, a pericardial friction rub may sometimes be seen, signifying pericardial inflammation brought on by the displaced lead’s mechanical irritation.48 Rarely, pacing impulses may cause noticeable muscle twitches across the chest or upper abdomen, which would indicate the placement and stimulation of an extracardiac lead.49,50 Even though it is rare, subcutaneous emphysema may occur when a perforation is exacerbated by pneumothorax or pneumopericardium.51

Device interrogation and electrical parameter assessment

The earliest and most impartial proof of lead perforation is often provided via device questioning. According to reports, 60%–70% of perforations increase pacing threshold, which is defined as more than a 2-fold increase from the postimplant baseline. This is usually caused by the decreased myocardial contact when the lead tip leaves the endocardial surface.20,52 In contrast, because of the increased impedance mismatch and nonmyocardial stimulation, partial perforation may paradoxically lead to lower thresholds.53

Abnormalities in sensing are also frequent; 40%–60% of patients have R-wave or P-wave amplitudes that are less than 50% of baseline values, which indicates less than ideal contact with cardiac tissue.20,54 Inappropriate pacing inhibition or the administration of needless treatments in defibrillator devices may be symptoms of oversensing of extracardiac myopotentials or diaphragmatic signals.55 Although less dependable, lead impedance may change based on the medium in which the lead tip is located. When fluid builds up, as in pericardial effusion, or when air pressure rises, as in pneumothorax or pneumopericardium, it usually falls.21,48,56 Advanced perforation may result in exit block or capture failure, which means that the lead has completely passed through the myocardial barrier even when there is no myocardial depolarization.57,58

Imaging modalities

Given that no one imaging method is always conclusive, a multimodal imaging strategy is often required to confirm the diagnosis of lead perforation. Despite being readily accessible, chest radiography has a modest sensitivity (∼50%–60%) and may show indirect indicators such as aberrant lead curvature or extension outside of the cardiac profile.59, 60, 61

TTE is essential for quickly assessing heart chamber function and pericardial effusion. Acoustic shadowing often restricts accurate localization of the perforation site, despite having a sensitivity of approximately 70%–80% for seeing the actual lead tip.62, 63, 64 Widely regarded as the gold standard, chest CT has a diagnostic sensitivity of more than 95% and a specificity of more than 90%.65, 66, 67 Accurate lead location identification about the cardiac shape and surrounding thoracic structures is made possible by multiplanar reconstructions using CT.5 Intravenous contrast may improve the diagnosis of vascular damage or active extravasation in complicated situations, even though noncontrast CT is usually adequate.68

Although chest CT is highly accurate, some limitations should be accounted for, such as respiratory and cardiac contractile motion during imaging, which may overestimate the location, given that it has been recorded that cardiac motion varies between patients and could range from 1 mm to 15.2 mm, as evidenced by Wu et al.69

Transesophageal echocardiography, one of the other imaging methods, is especially helpful for identifying atrial lead perforations because of its proximity to cardiac structures and better resolution.70 When other modalities are unable to offer conclusive information and patient safety permits, cardiac magnetic resonance imaging (MRI) may be considered. Although traditionally contraindicated in patients with non-MRI-conditional leads, recent studies have reported the safe performance of MRI under controlled conditions in select cases. However, in the context of lead perforation, data are limited, and MRI is generally avoided owing to concerns over lead heating, dislodgment, or exacerbation of myocardial injury.71, 72, 73, 74, 75 For instance, Nazarian et al74 have demonstrated protocols for MRI in patients with pacemakers or defibrillators, but these typically excluded those with evidence of lead perforation (shown in Table 2).

Table 2.

Diagnostic modalities for cardiac lead perforation

Modality Sensitivity Advantages Limitations
Chest radiography12 50%–60% Quick and accessible Low sensitivity, indirect signs
Transthoracic echocardiography (TTE)12 70%–80% Identifies effusion, bedside use Poor visualization of lead tip
CT scan12 >95% High resolution, multiplanar Radiation exposure, not always immediate
Transesophageal echocardiography (TEE)16 High for atrial leads High resolution near RA Semi-invasive

CT = computed tomography; RA = right atrium.

Diagnostic algorithm

Clinical assessment, device interrogation, and imaging tests are all integrated in a step-by-step, tiered manner in an efficient diagnostic approach to suspected lead perforation. A comprehensive evaluation of the patient’s presenting symptoms, vital signs, physical examination, 12-lead electrocardiography, and prompt device interrogation is all part of the first assessment.76 An emergent bedside echocardiogram is necessary to evaluate lead placement and rule out tamponade when hemodynamic instability is present77 (shown in Figure 3).

Figure 3.

Figure 3

Diagnostic algorithm for suspected cardiac lead perforation, integrating symptom recognition, device interrogation, and multimodal imaging. Management depends on clinical stability and imaging confirmation. CT = computed tomography; MRI = magnetic resonance imaging; TEE = transesophageal echocardiography.

Clinical and imaging-based risk stratification aids in determining the urgency of further testing. High-risk patients need immediate imaging and may need surgery if they have unstable hemodynamics, significant pericardial effusions, or worsening symptoms. In contrast, low-risk individuals who have steady parameters and just slight deviations can be assessed less urgently.78,79

TTE is the first imaging modality used on stable patients. Noncontrast cardiac CT should be performed to confirm the diagnosis and direct treatment if the results are unclear, but suspicion is still high.37,80,81 Unless rapid vision is required to guide urgent treatment, conclusive imaging should be delayed until stabilization in unstable patients.82 Given that each modality has limits and a multimodal approach often produces the most accurate findings, the final diagnosis should ultimately include clinical context, device data, and imaging evidence.77,83

Management strategies

The timing of the rupture, the severity of the symptoms, the hemodynamic stability, and the institutional knowledge all influence the therapy of cardiac lead perforation, necessitating a patient-specific strategy.61 The spectrum of treatment includes anything from surgical intervention in life-threatening situations to conservative observation in clinically stable persons.41 Treatment routes must be guided by proper risk assessment and accurate diagnosis.

Conservative management

For individuals who are asymptomatic or just slightly symptomatic, conservative therapy is a good choice, especially when a delayed perforation is unintentionally found during regular follow-up. Patients who present more than a month after device implantation, have stable electrical parameters on device interrogation, and do not have a large pericardial effusion are the best prospects. The perforating lead may remain functionally quiescent in these individuals if there are no increasing sensing or pacing problems.10

Conservative management relies heavily on monitoring procedures. To identify arrhythmias or hemodynamic changes in real time, continuous telemetry is advised throughout the first observation period.21 Initially, serial device interrogations should be conducted every 6–12 hours to check for electrical degradation, such as changed sensitivity or increased capture thresholds. To check for new or expanding pericardial effusions, TTE should be performed every 24–48 hours, particularly during the first 48–72 hours after diagnosis.84

Modification of activity is also crucial. To reduce mechanical displacement of the lead, patients should briefly limit movement of the upper extremities on the implantation side.21 Because they might worsen lead migration or cause pericardial hemorrhage, Valsalva maneuvers and demanding exercises that increase intrathoracic pressure should be avoided.85

Reprogramming the device could be required to maximize performance and minimize any possible issues. Modifications include raising the pacing power to guarantee steady myocardial capture and altering the sensing thresholds to prevent unwarranted extracardiac signal detection.86 It can be necessary to temporarily disable rate-responsive characteristics in situations involving extracardiac stimulation, such as diaphragmatic pacing.

In 60%–85% of instances with stable, long-lasting perforations that do not exhibit severe electrical degradation or effusion, conservative measures may be effective.10 However, conservative treatment is often less successful in cases with acute or subacute perforations, particularly when symptoms change or electrical characteristics start to deteriorate. To prevent delayed escalation in care, timely re-evaluation is essential.40

Lead repositioning and extraction

Lead repositioning or extraction is the last treatment for individuals who are hemodynamically unstable, exhibiting escalating electrical abnormalities, or symptomatic,87,88 although according to Younis et al89 in their cohort of 38 patients, 72% of whom had iatrogenic perforated pacemaker leads in the RV, simple traction with or without a locking stylet was achieved in all cases with emphasis that, given the possible situational consequences, surgical backup should be present at all times; these findings support the safety of nonsurgical management of lead perforations. If there is little fibrosis and no indication of major problems, simple fluoroscopic retraction may be tried in situations involving leads that were inserted for less than 6 weeks.87 When the etiology of initial perforation is high apical pressure, leads may be moved to more secure cardiac regions, such as the IVS, instead of the apex.88,90 In these situations, passive-fixation leads could be better because they lower the chance of recurrent perforation.91

Different percutaneous extraction methods are used depending on the fibrotic load and lead age.87,91 Manual traction or stylet-based removal is often enough for freshly inserted leads.87 Securing the lead and making withdrawal easier via thick adhesions, locking stylets, and telescoping sheaths increase success for leads that have been in place for many months.87,91 Energy-based instruments such as laser sheaths or radiofrequency-powered devices could be required for leads with implantation times longer than a year to safely break up fibrosis and allow extraction.87,91

Safety throughout procedures is crucial.87,91 Throughout the intervention, hemodynamic monitoring should be continued, and echocardiography should be accessible right away to check for the formation of pericardial effusions.88,90 Equipment for pericardiocentesis should be at the bedside, and cardiac surgery support is crucial, especially in high-risk situations.87,88,91

Patient stability determines when to intervene.87,88 Patients with active tamponade or those who are hemodynamically unstable need to be treated right away.87,88 In contrast, lead extraction should be done within 24 hours for stable individuals who exhibit symptoms of a perforation.87,88 If the risk-benefit analysis favors intervention, chronic, asymptomatic perforations may be treated electively.87,88,91

With success rates of 90%–95%, lead repositioning and extraction provide very positive results in centers with expertise.87,91 Approximately 1%–3% of instances result in major consequences such as hemothorax, vascular damage, or cardiac tamponade. Percutaneous extraction-related mortality is uncommon, estimated to be between 0.1% and 0.3%, especially when suitable surgical support is present.87,91

Transvenous lead extraction in confirmed lead perforation cases seems to have high procedural success and reasonable safety. In a retrospective cohort, all 17 perforating leads undergoing transvenous lead extraction were removed successfully without procedure-related mortality or major adverse events.4 However, data on pericardial effusion and tamponade during extraction are limited. Among extraction cohorts, the incidence of new-onset pericardial effusion is approximately 6.6%, with severe effusion in 3.3% and some requiring pericardiocentesis or surgery.92 These findings suggest that, even in confirmed perforation, extraction is feasible but carries an elevated risk of effusion and tamponade, warranting readiness for prompt pericardial drainage and surgical backup.

Surgical management

When percutaneous approaches are not appropriate or have not worked or when consequences necessitate direct repair, surgical care is reserved for certain situations. Ineffective percutaneous extraction; extensive or loculated pericardial effusions that are not amenable to drainage; damage to nearby thoracic organs, including the lung or major arteries; complicated perforations involving several lines; and thick fibrosis are among the indications for surgery71,93, 94, 95, 96, 97 (shown in Figure 4).

Figure 4.

Figure 4

Side-by-side comparison of percutaneous transvenous and open surgical techniques for cardiac lead extraction. Method choice is shaped by lead dwell time, fibrotic burden, and imaging requirements. Percutaneous extraction relies on venous access and continuous fluoroscopy, making it ideal for relatively new leads with minimal scarring, whereas open surgical removal offers direct visualization and is reserved for chronic, fibrotic, perforated, or otherwise complicated cases. ICD = implantable cardioverter-defibrillator; RV = right ventricle.

The intricacy of the situation determines which surgical method is used.71,94,95,97 For isolated lead removal without further cardiac pathology, less invasive techniques such as thoracoscopic extraction or subxiphoid pericardial windows may be adequate.91,95 Full median sternotomy is still the gold standard for more severe injuries or situations that need concurrent cardiac treatments (eg, valve repair).94,95

Being prepared for temporary pacing is an essential part of perioperative treatment, particularly for patients who are pacemaker dependent.71,94,97 The necessity for thromboprophylaxis and hemostasis must be balanced while managing anticoagulation status, especially in patients with mechanical valves or significant thromboembolic risk.71,94 Depending on the new lead design and cardiac performance, postoperative device reprogramming should be customized.91

Excellent surgical results are obtained; in more than 99% of instances, a successful extraction is accomplished.97 Depending on the intricacy of the procedure and the patient’s level of recuperation, the average length of stay in the hospital is 3–7 days. Compared with sternotomy (8–12 weeks), minimally invasive procedures need less recovery time (4–6 weeks), and most patients resume their normal activity levels within this period.94,95

Special management considerations

Modified management methods are required in a number of therapeutic situations. To restore hemodynamic stability in tamponade instances, prompt echocardiography-guided pericardiocentesis is necessary. If the effusion returns, it could be necessary to place an indwelling pericardial drain, and depending on the risk of bleeding, anticoagulation should be reversed.98

Extraction has to be done very carefully when extracardiac tissues are implicated, such as pulmonary or vascular perforation. To assure safety, a multidisciplinary team including interventional radiology or thoracic surgery may be required, along with modified procedures.98

Planning for temporary pacing during lead removal is crucial for individuals who are pacemaker dependent.52,35 Depending on patient anatomy and urgency, this may include starting transcutaneous pacing or putting a temporary lead on the contralateral side. To avoid bradyarrhythmic problems, reimplantation of a new lead is sometimes done during the same surgery for these individuals.35

Recent studies have explored the safety profile of left bundle branch area (LBBa) pacing, particularly with regard to myocardial perforation risk. Although LBBa pacing involves deeper septal lead placement than traditional RV pacing, the incidence of perforation seems low, ranging from 0.2% to 0.8% in large series, as evidenced in studies such as those of Huang et al68 and Vijayaraman et al.99 The use of fixed-curve sheaths (eg, C315 His or Select Site C304) combined with electroanatomic and fluoroscopic guidance facilitates precise lead positioning and may reduce perforation risk by stabilizing engagement and limiting free-wall advancement. Some studies suggest that the fibrotic nature of the IVS, along with continuous pacing impedance and electrocardiographic monitoring during lead deployment, adds additional safety compared with apical or free-wall lead placements. Nonetheless, septal hematoma, delayed perforation, and coronary artery injury remain potential complications, especially in patients with thin septa or on anticoagulation.100

The mechanical and anatomic risk factors that caused the original hole must be considered in reimplantation techniques. Passive attachment techniques and placement in the septal wall rather than the apex are often recommended.35 Device-dependent patients often need same-day reimplantation, whereas other patients may be planned for a delayed approach. Reimplantation timing is contingent upon stability and risk.35

Prevention strategies and risk mitigation

From patient selection to long-term postimplantation monitoring, preventing cardiac lead perforation is an essential part of managing CIEDs.101 Vigilant follow-up, procedural optimization, and risk stratification are all components of a complete preventative approach.101,102

Preimplantation risk assessment

To reduce the risk of cardiac perforation during CIED operations, a thorough preimplantation risk assessment is essential. The clinical justification for device treatment and individual risk factors for consequences, including perforation, should be carefully evaluated before choosing a patient, according to the medical literature.7,21,103 Advanced age, female sex, low BMI, long-term corticosteroid treatment, and active-fixation lead consumption are key risk factors that have been repeatedly established.36,104 These elements are linked to heightened tissue brittleness and an elevated risk of cardiac damage with the lead implantation.

Using clinical and demographic factors such as those mentioned earlier, risk stratification techniques and scoring systems have been created to help identify individuals who are at a higher risk of perforation.36,104 Even though these technologies have potential, particularly when combined with machine learning techniques and big clinical datasets, they need further prospective validation before being widely used in clinical settings.103,104

Alternative therapy approaches should be taken into consideration for patients who have been classified as high risk. Pharmacologic treatment may be enough in some bradyarrhythmia instances, negating the requirement for device placement.21 Subcutaneous ICDs lower the risk of cardiac perforation in patients who need defibrillation but not pacing because they do not need transvenous connections.105 Given that leadless pacemakers completely eliminate the need for transvenous access, they are an appealing alternative for single-chamber ventricular pacing. However, it should be noted that although leadless systems lower the risk of infection, they may be slightly more susceptible to perforation than traditional transvenous pacemakers, especially in some populations.105

Although leadless pacemakers eliminate transvenous leads, cardiac perforation remains a recognized risk, often involving the inferior or apical RV wall. Real-world data show slightly higher rates of early perforation/pericardial effusion with leadless systems (∼0.8%) than transvenous pacemakers (∼0.4%) in the first 30 days.106 Despite this, leadless devices are associated with fewer overall complications, fewer reinterventions, and lower chronic complication rates over longer-term follow-up.107 Patients who experience perforation with these leadless devices may face more severe clinical outcomes, sometimes requiring urgent pericardiocentesis or surgical intervention.107,108 These findings suggest that although leadless pacing reduces lead-related issues, operator technique is critical, especially regarding implantation site selection to minimize RV wall injury.

Along with preoperative imaging and laboratory examination where necessary, preprocedural planning should also include thorough collaborative decision making that considers the patient’s comorbidities, frailty, and preferences.103 To reduce the risk of perforation, procedural considerations are also crucial. These include the kind of lead used (active-fixation leads have a greater risk of perforation) and the intraoperative evaluation of lead position and slack.36,104

In conclusion, current guidelines and recent clinical studies support the best practices for preventing perforations during CIED implantation: careful procedural technique, careful patient selection, individualized risk assessment, and consideration of alternative device strategies in high-risk patients.36,104,105

Procedural modifications

The risk of perforation during the surgery may be considerably decreased by making adjustments to the implantation method and the choice of device.35,63,109 Selecting the right lead is essential; if the anatomy allows for sufficient contact and stability, passive-fixation leads may be a better option for high-risk patients.35,37 Selecting leads with a shorter helix, steroid-eluting tips, and less stiffness may lessen the likelihood of perforation when active fixation is required.

Lead stiffness during implantation is affected by the stylet type and how far it is inserted. Bench testing shows that firmer stylets increase tip pressure, which may raise the risk of myocardial perforation, especially in thin septal areas. In contrast, soft or extra-soft stylets reduce tip force, potentially improving safety but may offer less control during implantation.110 Studies also show that stylet-driven leads (which rely on internal stylets) tend to be stiffer and faster to deploy, but may carry a slightly higher risk of complications such as dislodgement, although not necessarily perforation.111 Therefore, stylet selection and positioning should be adjusted carefully to balance control with safety in procedures such as LBBa pacing.

Procedural safety is also affected by lead handling and access site selection.37 Compared with conventional subclavian puncture, axillary venous access is linked to a more favorable lead trajectory and may result in less mechanical stress.112 To guarantee proper lead insertion and prevent unintentional progression beyond the cardiac outline, operators should use various fluoroscopic views. Myocardial mechanical damage may be reduced with careful lead manipulation that uses mild advancement and a suitable stylet shape.35,109,112

The risk of perforation is significantly influenced by the anatomic location of lead implantation. Despite its widespread usage, the RV apex is linked to a weaker myocardial wall and a higher risk of penetration. Other locations with thicker walls, including the low septum, the RV outflow tract septum, or the midseptum close to the moderator band, could provide more stable fixation with a decreased risk of perforation.13

To guarantee mechanical integrity and electrical function, intraoperative testing is necessary. Early identification of unstable lead placements is made possible by real-time evaluation of sensing, impedance, and capture thresholds. Intermittent instability may be detected by provocative testing in various body postures, and diaphragmatic or extracardiac stimulation, which may suggest that the lead tip is close to or beyond the myocardial boundary, can be revealed with high-output pacing (eg, at 10 V).57

Postimplantation management

The goal of postimplantation interventions is to lower the risk of negative outcomes and identify lead-related problems early.10,113,114

Without using too high voltage outputs that would cause progressive lead migration, device programming must be tuned to enable consistent myocardial capture with a sufficient safety margin. Inappropriate device behaviors resulting from partial or intermittent lead perforation may be avoided by modifying sensing vectors and thresholds.10,113

To track lead function over time, systematic follow-up is essential.87 It is advised to do an early clinical assessment 24–72 hours after implantation to detect any acute problems.115 Lead stability and function may be continuously monitored with follow-up assessments at 2–4 weeks, 3 months, and thereafter at regular intervals (usually every 6–12 months).87

The capacity to identify early indicators of lead-related problems has greatly increased, thanks to remote monitoring systems. These devices can identify small changes in lead performance that may occur before clinical symptoms appear, such as increasing capture thresholds, lowering impedance, or aberrant sensing. To improve early detection and reduce false alarms, customized alert levels have to be set.114

Finally, patient education is a critical component of prevention. The warning signals of a lead perforation, such as dizziness, syncope, persistent hiccups, chest discomfort, or dyspnea, should be explained in detail to patients. Both written and spoken educational resources should support this knowledge and include instructions on when to seek emergency medical care. Giving people the tools to identify and report any issues is a simple but very powerful way to speed up diagnosis and enhance results.10,113

Special populations

Certain populations must be managed cautiously owing to the risk of various complications such as device infection, which is estimated to occur at a rate of 0.5% after primary implantation and up to 7% after secondary manipulations,116 which can even be complicated by endocarditis, which is an indication for removal85 (shown in Figure 5).

Figure 5.

Figure 5

Illustrated overview of 3 high-risk populations in whom standard transvenous lead placement may pose elevated risks. Elderly patients with low body mass index have anatomic frailty, pediatric patients with congenital heart disease require growth- and anatomy-sensitive planning, and immunocompromised patients are prone to infection and tissue breakdown. Recommended strategies include leadless pacing, epicardial leads, and minimal-hardware configurations.

Therefore, leadless pacemakers emerge as a solution to the previously mentioned issue given that it has been shown to decrease the risk of such complications through encapsulation of the device; in addition, the position and the need to keep a foreign object in the venous circulation also add additional protective layers against infections.117, 118, 119 El-Chami et al119 state that, because leadless pacemakers have a small surface area, they convey a less theoretical risk of infection.

Emerging technologies and future directions

Conventional cardiac pacemakers pose a lot of risks to patients, considering the rates of complications associated with the traditional design, including mechanical, infectious, and device-related issues.120

As mentioned before, leadless pacemakers are promising in the world of cardiac pacing. As the field is rapidly advancing, we are seeing a potential for a future where leadless pacemakers are the first-line solution for patients in need, especially with the newly developed systems that can provide any type of pacing without the need for transvenous leads.121, 122, 123

With that said, complex pacing modalities, device interactions, longevity differences in multiple devices, and the missing developed techniques in transvenous pacing, which leadless pacemakers need to reach the level of, still require further research to understand and solve.

Guidelines and recommendations

A direct clear management guidelines in cardiac lead perforations is still needed to be developed, although many studies have been conducted; for instance, Rajkumar et al4 showed that CT imaging was attributed to be the imaging modality of choice when it comes to cardiac perforations, which was consistent with that of Zhou et al52 in which transvenous lead extraction was used and showed safe and efficient in managing cardiac lead perforations.

Chest CT is widely used for detecting lead perforation owing to its availability and high resolution. However, metallic leads often cause blooming and beam-hardening artifacts, which may mimic lead extension beyond the myocardium and result in false-positive diagnoses of perforation. To avoid overinterpretation, CT findings should be correlated with clinical context and, where possible, confirmed by echocardiography or fluoroscopic imaging.124,125

Recently, Kidoh et al125 tested a newly proposed imaging modality that uses the metal artifact reduction algorithm to lower diagnostic uncertainty and improve clinical management and showed that, in 47 consecutive cases, the metal artifact reduction algorithm was able to reduce metal artifacts and improve the produced images, as opposed to conventional CT where in a quarter of cases the results were equivocal and nonconclusive.

Given that no evidence-based management options exist for cardiac lead perforation, Rav Acha et al,10 in their retrospective study on 48 perforation cases, 22 of which were associated with recurrent symptoms, concluded that lead revision is a safer and more effective option in cardiac lead perforations, especially in patients on blood thinners (anticoagulants and antiplatelets). Therefore, we recommend that a consensus and clear guideline be sought and developed to lower the risk of devastating complications and improve patient care.

Conclusion and future research priorities

If not detected and treated right away, cardiac lead perforation, an uncommon but dangerous side effect of CIEDs, could be fatal. In addition to developments in diagnostic methods and therapy options ranging from conservative surveillance to surgical intervention, current research has greatly increased our awareness of the anatomic, procedural, and patient-specific risk factors involved.

Several significant gaps still exist despite these advancements. First, given that some instances are silent and diagnostic criteria are unclear, the real frequency of subclinical and delayed perforations is still underestimated. Furthermore, there is a dearth of prospective, high-quality data on the long-term results of conservatively handled patients as opposed to those who have extraction or relocation. Device-specific risk profiles are continually developing and should be continuously monitored and assessed, particularly for more recent technologies such as subcutaneous ICDs and leadless pacemakers.

There is an urgent need to standardize reporting and diagnostic criteria. Different definitions of lead perforation, such as those based on electrical characteristics, imaging thresholds, or clinical presentation, make it difficult to compare research and reduce the generalizability of results. The quality and comparability of future research will be improved by establishing consistent categorization schemes and core result sets.

Prospective multicenter registries to capture actual incidence and outcomes; machine learning-based risk stratification tools that integrate imaging, clinical, and procedural variables; randomized controlled trials assessing the best management practices for asymptomatic or minimally symptomatic perforations; and design optimization and postmarketing surveillance for leadless and alternative pacing systems are examples of high-priority areas for future scholarly research.85

Finally, it is critical to promote international cooperation via research networks and shared datasets. These initiatives may assist in tailoring risk-reduction plans for lead implantation, expediting the gathering of evidence, and influencing the creation of guidelines. The creation of surveillance registries that monitor perforation incidence, intervention results, and long-term device safety profiles, as well as an international agreement on best practices, would be very beneficial to the field.

Acknowledgments

Funding Sources

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

Disclosures

The authors have no conflicts of interest to disclose.

Authorship

All authors attest they meet the current ICMJE criteria for authorship.

Data Availability

This study did not involve the creation or analysis of any new data. Therefore, data sharing is not applicable.

References

  • 1.Mond H.G., Proclemer A. The 11th world survey of cardiac pacing and implantable cardioverter-defibrillators: calendar year 2009–a world society of arrhythmia’s project. Pacing Clin Electrophysiol. 2011;34:1013–1027. doi: 10.1111/j.1540-8159.2011.03150.x. [DOI] [PubMed] [Google Scholar]
  • 2.Epstein A.E., DiMarco J.P., Ellenbogen K.A., et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. Circulation. 2008;117:e350–e408. doi: 10.1161/CIRCUALTIONAHA.108.189742. [DOI] [PubMed] [Google Scholar]
  • 3.Han B., Li Sj, Feng Cg. Catheter ablation via the brachial artery of ventricular arrhythmia originating from the right coronary cusp. Europace. 2015;17 doi: 10.1093/europace/euv047. 670–670. [DOI] [PubMed] [Google Scholar]
  • 4.Rajkumar C.A., Claridge S., Jackson T., et al. Diagnosis and management of iatrogenic cardiac perforation caused by pacemaker and defibrillator leads. Europace. 2016 doi: 10.1093/europace/euw074. euw074. [DOI] [PubMed] [Google Scholar]
  • 5.Laborderie J., Barandon L., Ploux S., et al. Management of subacute and delayed right ventricular perforation with a pacing or an implantable cardioverter-defibrillator lead. Am J Cardiol. 2008;102:1352–1355. doi: 10.1016/j.amjcard.2008.07.025. [DOI] [PubMed] [Google Scholar]
  • 6.Allouche E., Chargui S., Fathi M., Bezdah L. Subacute right ventricle perforation: a pacemaker lead complication. BMJ Case Rep. 2021;14 doi: 10.1136/bcr-2021-242489. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Bolaji O.A., Markson F.E., Kwaku K.F., Schaller R.D. Lead perforation in cardiac implantable electronic devices: incidence, diagnosis, and management. Pacing Clin Electrophysiol. 2025;48:874–885. doi: 10.1111/pace.15210. [DOI] [PubMed] [Google Scholar]
  • 8.Akbarzadeh M.A., Mollazadeh R., Sefidbakht S., Shahrzad S., Bahrololoumi Bafruee N. Identification and management of right ventricular perforation using pacemaker and cardioverter-defibrillator leads: a case series and mini review. J Arrhythm. 2017;33:1–5. doi: 10.1016/j.joa.2016.05.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Welch A.R., Yadav P., Lingle K., et al. Subacute right ventricular pacemaker lead perforation: often talked about in consent forms but very rarely seen. J Innov Card Rhythm Manag. 2011;2:442–445. [Google Scholar]
  • 10.Rav Acha M., Rafael A., Keaney J.J., et al. The management of cardiac implantable electronic device lead perforations: a multicentre study. Europace. 2019;21:937–943. doi: 10.1093/europace/euz120. [DOI] [PubMed] [Google Scholar]
  • 11.Zhou X., Ze F., Li D., Li X.B. Percutaneous management of atrium and lung perforation: a case report. World J Clin Cases. 2019;7:4327–4333. doi: 10.12998/wjcc.v7.i24.4327. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Zhang X., Zheng C., Wang P., et al. Assessment of cardiac lead perforation: comparison among chest radiography, transthoracic echocardiography and electrocardiography-gated contrast-enhanced cardiac CT. Eur Radiol. 2019;29:963–974. doi: 10.1007/s00330-018-5633-6. [DOI] [PubMed] [Google Scholar]
  • 13.Israel C.W., Tribunyan S., Ho Y., JA Cabrera S., Cabrera J.A. Anatomy for right ventricular lead implantation. Herzschrittmacherther Elektrophysiol. 2022;33:319–326. doi: 10.1007/s00399-022-00872-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Erdogan O., Hunuk B. Right ventricular outflow tract pacing: an alternative, safe, and effective pacing site. J Long Term Eff Med Implants. 2010;20:13–21. doi: 10.1615/jlongtermeffmedimplants.v20.i1.20. [DOI] [PubMed] [Google Scholar]
  • 15.Cano Ó., Andrés A., Alonso P., Osca J., et al. Incidence and predictors of clinically relevant cardiac perforation associated with systematic implantation of active-fixation pacing and defibrillation leads: a single-centre experience with over 3800 implanted leads. Europace. 2017;19:96–102. doi: 10.1093/europace/euv410. [DOI] [PubMed] [Google Scholar]
  • 16.Zoppo F., Rizzo S., Corrado A., et al. Morphology of right atrial appendage for permanent atrial pacing and risk of iatrogenic perforation of the aorta by active fixation lead. Heart Rhythm. 2015;12:744–750. doi: 10.1016/j.hrthm.2014.12.023. [DOI] [PubMed] [Google Scholar]
  • 17.Ribeiro A.S.F., Zerolo B.E., Lopez-Espuela F., Sanchez R., Fernandes V.S. Cardiac system during the aging process. Aging Dis. 2023;14:1105–1122. doi: 10.14336/AD.2023.0115. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Moore K., Ganesan A., Labrosciano C., et al. Sex differences in acute complications of cardiac implantable electronic devices: implications for patient safety. J Am Heart Assoc. 2019;8 doi: 10.1161/JAHA.118.010869. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Kumar S., Yassin H., Esan O., SA T.J. In: Current Issues and Recent Advances in Pacemaker Therapy. Roka A., editor. IntechOpen; London, United Kingdom: 2012. Mechanism and management of pacing lead related cardiac perforation.http://www.intechopen.com/books/current-issues-and-recent-advances-in-pacemaker-therapy/mechanism-and-management-of-pacing-lead-related-cardiac-perforation [Google Scholar]
  • 20.Cecchin F., Atallah J., Walsh E.P., Triedman J.K., Alexander M.E., Berul C.I. Lead extraction in pediatric and congenital heart disease patients. Circ Arrhythm Electrophysiol. 2010;3:437–444. doi: 10.1161/CIRCEP.110.957324. [DOI] [PubMed] [Google Scholar]
  • 21.Kusumoto F.M., Schoenfeld M.H., Barrett C., et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice guidelines and the Heart Rhythm Society. J Am Coll Cardiol. 2019;74:e51–e156. doi: 10.1016/j.jacc.2018.10.044. [DOI] [PubMed] [Google Scholar]
  • 22.Christiansen M.N., Køber L., Weeke P., et al. Age-specific trends in incidence, mortality, and comorbidities of heart failure in Denmark, 1995 to 2012. Circulation. 2017;135:1214–1223. doi: 10.1161/CIRCULATIONAHA.116.025941. [DOI] [PubMed] [Google Scholar]
  • 23.Sasaki S., Watanabe H., Shibayama K., et al. Three-dimensional transesophageal echocardiographic evaluation of coronary involvement in patients with acute Type A aortic dissection. J Am Soc Echocardiogr. 2013;26:837–845. doi: 10.1016/j.echo.2013.05.001. [DOI] [PubMed] [Google Scholar]
  • 24.Cameron J., Mond H., Ciddor G., Harper K., Mckie J. Stiffness of the distal tip of bipolar pacing leads. Pacing Clin Electrophysiol. 1990;13:1915–1920. doi: 10.1111/j.1540-8159.1990.tb06916.x. [DOI] [PubMed] [Google Scholar]
  • 25.Gasser T.C., Gudmundson P., Dohr G. Failure mechanisms of ventricular tissue due to deep penetration. J Biomech. 2009;42:626–633. doi: 10.1016/j.jbiomech.2008.12.016. [DOI] [PubMed] [Google Scholar]
  • 26.Hirschl D.A., Jain V.R., Spindola-Franco H., Gross J.N., Haramati L.B. Prevalence and characterization of asymptomatic pacemaker and ICD lead perforation on CT. Pacing Clin Electrophysiol. 2007;30:28–32. doi: 10.1111/j.1540-8159.2007.00575.x. [DOI] [PubMed] [Google Scholar]
  • 27.Lang R.M., Bierig M., Devereux R.B., et al. Recommendations for chamber quantification: A report from the American Society of Echocardiography’s Guidelines and Standards Committee and the Chamber Quantification Writing Group, developed in conjunction with the European Association of Echocardiography, a branch of the European Society of Cardiology. J Am Soc Echocardiogr. 2005;18:1440–1463. doi: 10.1016/j.echo.2005.10.005. [DOI] [PubMed] [Google Scholar]
  • 28.Raghavan C., Cashion W.R., Spencer W.H. Malposition of transvenous pacing lead in the left ventricle. Clin Cardiol. 1996;19:335–338. doi: 10.1002/clc.4960190411. [DOI] [PubMed] [Google Scholar]
  • 29.Azizi M., Castel M.A., Behrens S., Rödiger W., Nägele H. Experience with coronary sinus lead implantations for cardiac resynchronization therapy in 244 patients. Herzschrittmacherther Elektrophysiol. 2006;17:13–18. doi: 10.1007/s00399-006-0502-4. [DOI] [PubMed] [Google Scholar]
  • 30.Kitamura K., Jorgensen C.R., Arthur H.L. Transvenous atrial and ventricular pacing from the coronary sinus complicated by perforation and cardiac tamponade. Chest. 1971;60:95–98. doi: 10.1378/chest.60.1.95. [DOI] [PubMed] [Google Scholar]
  • 31.Banaszewski M., Stępińska J. Editorial Right heart perforation by pacemaker leads. Arch Med Sci. 2012;1:11–13. doi: 10.5114/aoms.2012.27273. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Sterlinski M., Przybylski A., Maciag A., et al. Subacute cardiac perforations associated with active fixation leads. Europace. 2008;11:206–212. doi: 10.1093/europace/eun363. [DOI] [PubMed] [Google Scholar]
  • 33.Mahapatra S., Bybee K.A., Bunch T.J., et al. Incidence and predictors of cardiac perforation after permanent pacemaker placement. Heart Rhythm. 2005;2:907–911. doi: 10.1016/j.hrthm.2005.06.011. [DOI] [PubMed] [Google Scholar]
  • 34.Hsu J.C., Varosy P.D., Bao H., Dewland T.A., Curtis J.P., Marcus G.M. Cardiac perforation from implantable cardioverter-defibrillator lead placement: insights from the national cardiovascular data registry. Circ Cardiovasc Qual Outcomes. 2013;6:582–590. doi: 10.1161/CIRCOUTCOMES.113.000299. [DOI] [PubMed] [Google Scholar]
  • 35.Kusumoto F.M., Schoenfeld M.H., Wilkoff B.L., et al. HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 2017;14:e503–e551. doi: 10.1016/j.hrthm.2017.09.001. [DOI] [PubMed] [Google Scholar]
  • 36.Schwerg M., Stockburger M., Schulze C., et al. Clinical, anatomical, and technical risk factors for postoperative pacemaker or defibrillator lead perforation with particular focus on myocardial thickness. Pacing Clin Electrophysiol. 2014;37:1291–1296. doi: 10.1111/pace.12431. [DOI] [PubMed] [Google Scholar]
  • 37.Indik J.H., Gimbel J.R., Abe H., et al. HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm. 2017;14:e97–e153. doi: 10.1016/j.hrthm.2017.04.025. [DOI] [PubMed] [Google Scholar]
  • 38.Vamos M., Erath J.W., Benz A.P., Bari Z., Duray G.Z., Hohnloser S.H. Incidence of cardiac perforation with conventional and with leadless pacemaker systems: a systematic review and meta-analysis. J Cardiovasc Electrophysiol. 2017;28:336–346. doi: 10.1111/jce.13140. [DOI] [PubMed] [Google Scholar]
  • 39.Issa Z.F., Issa T.Z. Feasibility and safety of percutaneous lead revision for subacute and delayed cardiac device lead perforation. JACC Clin Electrophysiol. 2021;7:26–35. doi: 10.1016/j.jacep.2020.07.024. [DOI] [PubMed] [Google Scholar]
  • 40.Döring M., Müssigbrodt A., Ebert M., et al. Transvenous revision of leads with cardiac perforation following device implantation—safety, outcome, and complications. Pacing Clin Electrophysiol. 2020;43:1325–1332. doi: 10.1111/pace.14056. [DOI] [PubMed] [Google Scholar]
  • 41.Barron D.W., Rasmussen N.G., Auerbach M.H., Rappaport D., Martini W.A. Pacemaker lead perforation presenting as persistent abdominal pain: a case report. Cureus. 2025;17 doi: 10.7759/cureus.84116. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Munde K., Paliwal M., Jain A. Beneath the rhythm: deciphering the subtle perforation of the right ventricle by a pacemaker lead. Indones J Cardiol. 2024;12:183–186. [Google Scholar]
  • 43.Lin Y.S., Hung S.P., Chen P.R., et al. Risk factors influencing complications of cardiac implantable electronic device implantation: infection, pneumothorax and heart perforation: a nationwide population-based cohort study. Medicine (Baltimore) 2014;93 doi: 10.1097/MD.0000000000000213. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Henrikson C.A., Leng C.T., Yuh D.D., Brinker J.A. Computed tomography to assess possible cardiac lead perforation. Pacing Clin Electrophysiol. 2006;29:509–511. doi: 10.1111/j.1540-8159.2006.00385.x. [DOI] [PubMed] [Google Scholar]
  • 45.Computed tomography artefact finding of pacing lead perforation. Br J Cardiol. 2020;27:97–99. [Google Scholar]
  • 46.Chrispin J., Love C.J. Computed tomography imaging before lead extraction. JACC Clin Electrophysiol. 2019;5:1447–1449. doi: 10.1016/j.jacep.2019.10.005. [DOI] [PubMed] [Google Scholar]
  • 47.Sidhu B.S., Rajani R., Rinaldi C.A. Chronic ventricular lead perforation: expect the unexpected. Clin Case Rep. 2019;7:465–468. doi: 10.1002/ccr3.2005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Zhao Y., Su L., Gao Y., et al. Leadless pacemaker implantation after delayed atrial lead perforation and battery depletion: a case report. BMC Cardiovasc Disord. 2024;24:747. doi: 10.1186/s12872-024-04448-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Jain S., Clancy J., Schoenfeld M.H. An unusual presentation of delayed lead perforation: it’s never too late. HeartRhythm Case Rep. 2022;8:106–109. doi: 10.1016/j.hrcr.2021.11.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Maruya Y., Yamaura T., Mine H., Suzuki H. Successful pericardial repair and coverage for late pacemaker lead-related atrial perforation and pneumothorax: a case report. Gen Thorac CardioVasc Surg Cases. 2023;2:102. doi: 10.1186/s44215-023-00117-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Kinoshita M., Inaba S., Ota N., et al. Utility of surgical direct visualization in a high-risk patient of bleeding during extraction of perforating pacemaker lead. J Coron Artery Dis. 2022;28:19–23. [Google Scholar]
  • 52.Zhou X., Ze F., Li D., Wang L., Guo J., Li X. Outcomes of transvenous lead extraction in patients with lead perforation: a single-center experience. Clin Cardiol. 2020;43:386–393. doi: 10.1002/clc.23327. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Iacobazzi D., Alvino V.V., Caputo M., Madeddu P. Accelerated cardiac aging in patients with congenital heart disease. Front Cardiovasc Med. 2022;9 doi: 10.3389/fcvm.2022.892861. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Jenkins K.J., Botto L.D., Correa A., et al. Public health approach to improve outcomes for congenital heart disease across the life span. J Am Heart Assoc. 2019;8 doi: 10.1161/JAHA.118.009450. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Keir M., Borman M., Clegg R., et al. Caring for the aging patient with adult congenital heart disease: a review of cardiac and noncardiac comorbidities. CJC Pediatr Congenit Heart Dis. 2022;1:274–281. doi: 10.1016/j.cjcpc.2022.10.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Patsiou V., Haidich A.B., Baroutidou A., Giannopoulos A., Giannakoulas G. Epicardial versus endocardial pacing in paediatric patients with atrioventricular block or sinus node dysfunction: a systematic review and meta-analysis. Pediatr Cardiol. 2023;44:1641–1648. doi: 10.1007/s00246-023-03213-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Burri H., Starck C., Auricchio A., et al. EHRA expert consensus statement and practical guide on optimal implantation technique for conventional pacemakers and implantable cardioverter-defibrillators: endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), and the Latin-American Heart Rhythm Society (LAHRS) Europace. 2021;23:983–1008. doi: 10.1093/europace/euaa367. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Wilkoff B.L., Love C.J., Byrd C.L., et al. Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management. Heart Rhythm. 2009;6:1085–1104. doi: 10.1016/j.hrthm.2009.05.020. [DOI] [PubMed] [Google Scholar]
  • 59.Sridhar A.R.M., Lavu M., Yarlagadda V., et al. Cardiac implantable electronic device-related infection and extraction trends in the U.S. Pacing Clin Electrophysiol. 2017;40:286–293. doi: 10.1111/pace.13009. [DOI] [PubMed] [Google Scholar]
  • 60.Poole J.E., Gleva M.J., Mela T., et al. Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registry. Circulation. 2010;122:1553–1561. doi: 10.1161/CIRCULATIONAHA.110.976076. [DOI] [PubMed] [Google Scholar]
  • 61.Reynolds D., Duray G.Z., Omar R., et al. A leadless intracardiac transcatheter pacing system. N Engl J Med. 2016;374:533–541. doi: 10.1056/NEJMoa1511643. [DOI] [PubMed] [Google Scholar]
  • 62.Reddy V.Y., Exner D.V., Cantillon D.J., et al. Percutaneous implantation of an entirely intracardiac leadless pacemaker. N Engl J Med. 2015;373:1125–1135. doi: 10.1056/NEJMoa1507192. [DOI] [PubMed] [Google Scholar]
  • 63.Roberts P.R., Clementy N., Al Samadi F., et al. A leadless pacemaker in the real-world setting: the Micra transcatheter Pacing System Post-Approval Registry. Heart Rhythm. 2017;14:1375–1379. doi: 10.1016/j.hrthm.2017.05.017. [DOI] [PubMed] [Google Scholar]
  • 64.Sharma P.S., Dandamudi G., Naperkowski A., et al. Permanent His-bundle pacing is feasible, safe, and superior to right ventricular pacing in routine clinical practice. Heart Rhythm. 2015;12:305–312. doi: 10.1016/j.hrthm.2014.10.021. [DOI] [PubMed] [Google Scholar]
  • 65.Huang W., Su L., Wu S., et al. A novel pacing strategy with low and stable output: pacing the left bundle branch immediately beyond the conduction block. Can J Cardiol. 2017;33:1736.e1–1736.e3. doi: 10.1016/j.cjca.2017.09.013. [DOI] [PubMed] [Google Scholar]
  • 66.Vijayaraman P., Dandamudi G., Zanon F., et al. Permanent His bundle pacing: recommendations from a Multicenter His bundle Pacing Collaborative Working Group for standardization of definitions, implant measurements, and follow-up. Heart Rhythm. 2018;15:460–468. doi: 10.1016/j.hrthm.2017.10.039. [DOI] [PubMed] [Google Scholar]
  • 67.Zanon F., Ellenbogen K.A., Dandamudi G., et al. Permanent His-bundle pacing: a systematic literature review and meta-analysis. Europace. 2018;20:1819–1826. doi: 10.1093/europace/euy058. [DOI] [PubMed] [Google Scholar]
  • 68.Huang W., Chen X., Su L., Wu S., Xia X., Vijayaraman P. A beginner’s guide to permanent left bundle branch pacing. Heart Rhythm. 2019;16:1791–1796. doi: 10.1016/j.hrthm.2019.06.016. [DOI] [PubMed] [Google Scholar]
  • 69.Wu B., Atwood T., Mundt A.J., et al. Characterizing cardiac contractile motion for noninvasive radioablation of ventricular tachycardia. Heart Rhythm O2. 2024;5:131–136. doi: 10.1016/j.hroo.2023.12.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70.Vijayaraman P., Subzposh F.A., Naperkowski A., et al. Prospective evaluation of feasibility and electrophysiologic and echocardiographic characteristics of left bundle branch area pacing. Heart Rhythm. 2019;16:1774–1782. doi: 10.1016/j.hrthm.2019.05.011. [DOI] [PubMed] [Google Scholar]
  • 71.Kusumoto F.M., Calkins H., Boehmer J., et al. HRS/ACC/AHA expert consensus statement on the use of implantable cardioverter-defibrillator therapy in patients who are not included or not well represented in clinical trials. Circulation. 2014;130:94–125. doi: 10.1161/CIR.0000000000000056. [DOI] [PubMed] [Google Scholar]
  • 72.Acha M.R., Keaney J.J., Lubitz S.A., et al. Increased perforation risk with an MRI-conditional pacing lead: A single-center study. Pacing Clin Electrophysiol. 2015;38:334–342. doi: 10.1111/pace.12550. [DOI] [PubMed] [Google Scholar]
  • 73.Shah A.D., Morris M.A., Hirsh D.S., et al. Magnetic resonance imaging safety in nonconditional pacemaker and defibrillator recipients: a meta-analysis and systematic review. Heart Rhythm. 2018;15:1001–1008. doi: 10.1016/j.hrthm.2018.02.019. [DOI] [PubMed] [Google Scholar]
  • 74.Nazarian S., Beinart R., Halperin H.R. Magnetic resonance imaging and implantable devices. Circ Arrhythm Electrophysiol. 2013;6:419–428. doi: 10.1161/CIRCEP.113.000116. [DOI] [PubMed] [Google Scholar]
  • 75.Levine G.N., Gomes A.S., Arai A.E., et al. Safety of magnetic resonance imaging in patients with cardiovascular devices: an American Heart Association scientific statement from the committee on diagnostic and interventional cardiac catheterization, Council on Clinical Cardiology, and the Council on Cardiovascular Radiology and Intervention: endorsed by the American College of Cardiology Foundation, the North American Society for Cardiac Imaging, and the Society for Cardiovascular Magnetic Resonance. Circulation. 2007;116:2878–2891. doi: 10.1161/CIRCULATIONAHA.107.187256. [DOI] [PubMed] [Google Scholar]
  • 76.Brignole M., Auricchio A., Baron-Esquivias G., et al. 2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA) Eur Heart J. 2013;34:2281–2329. doi: 10.1093/eurheartj/eht150. [DOI] [PubMed] [Google Scholar]
  • 77.Slotwiner D., Varma N., Akar J.G., et al. HRS Expert Consensus Statement on remote interrogation and monitoring for cardiovascular implantable electronic devices. Heart Rhythm. 2015;12:e69–e100. doi: 10.1016/j.hrthm.2015.05.008. [DOI] [PubMed] [Google Scholar]
  • 78.Sticherling C., Marin F., Birnie D., et al. Antithrombotic management in patients undergoing electrophysiological procedures: a European Heart Rhythm Association (EHRA) position document endorsed by the ESC Working Group Thrombosis, Heart Rhythm Society (HRS), and Asia Pacific Heart Rhythm Society (APHRS) Europace. 2015;17:1197–1214. doi: 10.1093/europace/euv190. [DOI] [PubMed] [Google Scholar]
  • 79.Cronin E.M., Bogun F.M., Maury P., et al. HRS/EHRA/APHRS/LAHRS expert consensus statement on catheter ablation of ventricular arrhythmias. Europace. 2019;21:1143–1144. doi: 10.1093/europace/euz132. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 80.Task force chairs, Daubert J.C., Saxon L., et al. 2012 EHRA/HRS expert consensus statement on cardiac resynchronization therapy in heart failure: implant and follow-up recommendations and management: a registered branch of the European Society of Cardiology (ESC), and the Heart Rhythm Society; and in collaboration with the Heart Failure Society of America (HFSA), the American Society of Echocardiography (ASE), the American Heart Association (AHA), the European Association of Echocardiography (EAE) of the ESC and the Heart Failure Association of the ESC (HFA) Europace. 2012;14:1236–1286. doi: 10.1093/europace/eus222. [DOI] [PubMed] [Google Scholar]
  • 81.Zipes D.P., Calkins H., Daubert J.P., et al. 2015 ACC/AHA/HRS advanced training statement on clinical cardiac electrophysiology (A revision of the ACC/AHA 2006 update of the clinical competence statement on invasive electrophysiology studies, catheter ablation, and cardioversion) Heart Rhythm. 2016;13:e3–e37. doi: 10.1016/j.hrthm.2015.09.014. [DOI] [PubMed] [Google Scholar]
  • 82.Sequeira S., Jarvis C.I., Benchouche A., Seymour J., Tadmouri A. Cost-effectiveness of remote monitoring of implantable cardioverter-defibrillators in France: a meta-analysis and an integrated economic model derived from randomized controlled trials. Europace. 2020;22:1071–1082. doi: 10.1093/europace/euaa082. [DOI] [PubMed] [Google Scholar]
  • 83.Krahn A.D., Lee D.S., Birnie D., et al. Predictors of short-term complications after implantable cardioverter-defibrillator replacement: results from the Ontario ICD database. Circ Arrhythm Electrophysiol. 2011;4:136–142. doi: 10.1161/CIRCEP.110.959791. [DOI] [PubMed] [Google Scholar]
  • 84.Glikson M., Nielsen J.C., Kronborg M.B., et al. ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. Eur Heart J. 2021;42:3427–3520. doi: 10.1093/eurheartj/ehab364. [DOI] [PubMed] [Google Scholar]
  • 85.Corrigendum to: 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. Eur Heart J. 2022;43 doi: 10.1093/eurheartj/ehac075. 1651–1651. [DOI] [PubMed] [Google Scholar]
  • 86.Saliba W., Schliamser J.E., Lavi I., Barnett-Griness O., Gronich N., Rennert G. Catheter ablation of atrial fibrillation is associated with reduced risk of stroke and mortality: a propensity score–matched analysis. Heart Rhythm. 2017;14:635–642. doi: 10.1016/j.hrthm.2017.02.001. [DOI] [PubMed] [Google Scholar]
  • 87.Shah M.J., Silka M.J., Silva J.N.A., et al. PACES expert consensus statement on the indications and management of cardiovascular implantable electronic devices in pediatric patients. Heart Rhythm. 2021;18:1888–1924. doi: 10.1016/j.hrthm.2021.07.038. [DOI] [PubMed] [Google Scholar]
  • 88.Elbatran A.I., Akhtar Z., Bajpai A., et al. Percutaneous management of lead-related cardiac perforation with limited use of computed tomography and cardiac surgery. Pacing Clin Electrophysiol. 2021;44:614–624. doi: 10.1111/pace.14204. [DOI] [PubMed] [Google Scholar]
  • 89.Younis A., Awashra A., Matteo M., et al. Management of transvenous leads in patients with iatrogenic lead perforation. J Cardiovasc Electrophysiol. 2025 doi: 10.1111/jce.70006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 90.Archontakis S., Sideris K., Aggeli K., et al. Percutaneous lead extraction and repositioning: an effective and safe therapeutic strategy for early ventricular lead perforation with dislocation both inside and outside the pericardial sac following a cardiac device implantation. J Cardiovasc Electrophysiol. 2019;30:299–307. doi: 10.1111/jce.13804. [DOI] [PubMed] [Google Scholar]
  • 91.Silka M.J., Shah M.J., Silva J.N.A., et al. PACES expert consensus statement on the indications and management of cardiovascular implantable electronic devices in pediatric patients: executive summary. Heart Rhythm. 2021;18:1925–1950. doi: 10.1016/j.hrthm.2021.07.051. [DOI] [PubMed] [Google Scholar]
  • 92.Regoli F., D’Ambrosio G., Caputo M.L., et al. New-onset pericardial effusion during transvenous lead extraction: incidence, causative mechanisms, and associated factors. J Interv Card Electrophysiol. 2018;51:253–261. doi: 10.1007/s10840-018-0327-1. [DOI] [PubMed] [Google Scholar]
  • 93.Tułecki Ł., Czajkowski M., Targońska S., et al. The role of cardiac surgeon in transvenous lead extraction: experience from 3462 procedures. J Cardiovasc Electrophysiol. 2022;33:1357–1365. doi: 10.1111/jce.15510. [DOI] [PubMed] [Google Scholar]
  • 94.Miyagi Y., Oomori H., Maeda M., et al. Surgical management of cardiac implantable electronic device complications in patients unsuitable for transvenous lead extraction. Circ J. 2022;87:103–110. doi: 10.1253/circj.CJ-22-0456. [DOI] [PubMed] [Google Scholar]
  • 95.Gupta A.R., Power J.R., Yang Y., et al. Indications and outcomes of elective open chest lead extractions. J Cardiovasc Electrophysiol. 2024;35:2128–2133. doi: 10.1111/jce.16412. [DOI] [PubMed] [Google Scholar]
  • 96.Caniglia-Miller J.M., Bussey W.D., Kamtz N.M., et al. Surgical management of major intrathoracic hemorrhage resulting from high-risk transvenous pacemaker/defibrillator lead extraction: surgical management of major intrathoracic hemorrhage. J Card Surg. 2015;30:149–153. doi: 10.1111/jocs.12500. [DOI] [PubMed] [Google Scholar]
  • 97.Kratz J.M., Toole J.M. Pacemaker and internal cardioverter defibrillator lead extraction: a safe and effective surgical approach. Ann Thorac Surg. 2010;90:1411–1417. doi: 10.1016/j.athoracsur.2010.05.036. [DOI] [PubMed] [Google Scholar]
  • 98.Adler Y., Ristić A.D., Imazio M., et al. Cardiac tamponade. Nat Rev Dis Primer. 2023;9:36. doi: 10.1038/s41572-023-00446-1. [DOI] [PubMed] [Google Scholar]
  • 99.Vijayaraman P., Ponnusamy S., Cano Ó., et al. Left Bundle Branch Area Pacing for Cardiac Resynchronization Therapy: Results From the International LBBAP Collaborative Study Group. JACC Clin Electrophysiol. 2021;7:135–147. doi: 10.1016/j.jacep.2020.08.015. [DOI] [PubMed] [Google Scholar]
  • 100.Jastrzębski M., Moskal P., Hołda M.K., et al. Deep septal deployment of a thin, lumenless pacing lead: a translational cadaver simulation study. Europace. 2020;22:156–161. doi: 10.1093/europace/euz270. [DOI] [PubMed] [Google Scholar]
  • 101.Frausing M.H.J.P., Kronborg M.B., Johansen J.B., Nielsen J.C. Avoiding implant complications in cardiac implantable electronic devices: what works? Europace. 2021;23:163–173. doi: 10.1093/europace/euaa221. [DOI] [PubMed] [Google Scholar]
  • 102.Blomström-Lundqvist C., Traykov V., Erba P.A., et al. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections—endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), the Latin American Heart Rhythm Society (LAHRS), International Society for Cardiovascular Infectious Diseases (ISCVID), and the European Society of Clinical Microbiology and Infectious Diseases (ESCMID) in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS) Eur Heart J. 2020;41:2012–2032. doi: 10.1093/eurheartj/ehaa010. [DOI] [PubMed] [Google Scholar]
  • 103.Chung M.K., Patton K.K., Lau C.P., et al. HRS/APHRS/LAHRS guideline on cardiac physiologic pacing for the avoidance and mitigation of heart failure. Heart Rhythm. 2023;20:e17–e91. doi: 10.1016/j.hrthm.2023.03.1538. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 104.Celikyurt U., Acar B., Yavuz S., Agacdiken A., Vural A. Predictors of the right ventricular perforation caused by active-fixation pacing and defibrillator leads: a single-centre experience. J Cardiovasc Electrophysiol. 2024;35:399–405. doi: 10.1111/jce.16181. [DOI] [PubMed] [Google Scholar]
  • 105.Al-Khatib S.M. Cardiac implantable electronic devices. N Engl J Med. 2024;390:442–454. doi: 10.1056/NEJMra2308353. [DOI] [PubMed] [Google Scholar]
  • 106.Sarsenbayeva A., Baimbetov A., Puodziukynas A., Baimakhanov B., Sapunov A., Bizhanov K. Systematic review and meta-analysis of acute mortality and complication rates following leadless pacemaker placement using national-level data. Medicina (Mex) 2025;61:974. doi: 10.3390/medicina61060974. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 107.El-Chami M.F., Bockstedt L., Longacre C., et al. Leadless vs. transvenous single-chamber ventricular pacing in the Micra CED study: 2-year follow-up. Eur Heart J. 2022;43:1207–1215. doi: 10.1093/eurheartj/ehab767. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 108.Haddadin F., Majmundar M., Jabri A., et al. Clinical outcomes and predictors of complications in patients undergoing leadless pacemaker implantation. Heart Rhythm. 2022;19:1289–1296. doi: 10.1016/j.hrthm.2022.03.1226. [DOI] [PubMed] [Google Scholar]
  • 109.Ip J.E. Advanced helix-fixation leadless cardiac pacemaker implantation techniques to improve success and reduce complications. J Cardiovasc Electrophysiol. 2023;34:1268–1276. doi: 10.1111/jce.15918. [DOI] [PubMed] [Google Scholar]
  • 110.Yamazaki D. Measurement of tip load with a pacemaker lead stylet and guiding catheter using a silicone heart model. Cureus. 2024;16 doi: 10.7759/cureus.74655. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 111.Chen X., Dong J. Stylet-driven leads compared with lumenless leads for left bundle branch area pacing: a systematic review and meta-analysis. BMC Cardiovasc Disord. 2024;24:598. doi: 10.1186/s12872-024-04273-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 112.Deharo J.C., Dreyfus J., Bongiorni M.G., et al. Management of patients with transvalvular right ventricular leads undergoing transcatheter tricuspid valve interventions: a scientific statement of the European Heart Rhythm Association and the European Association of Percutaneous cardiovascular Interventions of the ESC endorsed by the Heart Rhythm Society, the Asian Pacific Heart Rhythm Society and the Canadian Heart Rhythm Society. Europace. 2025;27 doi: 10.1093/europace/euaf061. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 113.Tracy C.M., Epstein A.E., Darbar D., et al. 2012 ACCF/AHA/HRS Focused Update Incorporated into the ACCF/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities. J Am Coll Cardiol. 2013;61:e6–e75. doi: 10.1016/j.jacc.2012.11.007. [DOI] [PubMed] [Google Scholar]
  • 114.Ferrick A.M., Raj S.R., Deneke T., et al. HRS/EHRA/APHRS/LAHRS expert consensus statement on practical management of the remote device clinic. Heart Rhythm. 2023;20:e92–e144. doi: 10.1016/j.hrthm.2023.03.1525. [DOI] [PubMed] [Google Scholar]
  • 115.Knops R.E., Reddy V.Y., Ip J.E., et al. A dual-chamber leadless pacemaker. N Engl J Med. 2023;388:2360–2370. doi: 10.1056/NEJMoa2300080. [DOI] [PubMed] [Google Scholar]
  • 116.Wazni O., Wilkoff B.L. Considerations for cardiac device lead extraction. Nat Rev Cardiol. 2016;13:221–229. doi: 10.1038/nrcardio.2015.207. [DOI] [PubMed] [Google Scholar]
  • 117.Vamos M., Honold J., Duray G.Z., Hohnloser S.H. MICRA leadless pacemaker on autopsy. JACC Clin Electrophysiol. 2016;2:636–637. doi: 10.1016/j.jacep.2016.02.014. [DOI] [PubMed] [Google Scholar]
  • 118.Kypta A., Blessberger H., Lichtenauer M., Steinwender C. Complete encapsulation of a leadless cardiac pacemaker. Clin Res Cardiol. 2016;105 doi: 10.1007/s00392-015-0929-x. 94–94. [DOI] [PubMed] [Google Scholar]
  • 119.El-Chami M.F., Soejima K., Piccini J.P., et al. Incidence and outcomes of systemic infections in patients with leadless pacemakers: data from the Micra IDE study. Pacing Clin Electrophysiol. 2019;42:1105–1110. doi: 10.1111/pace.13752. [DOI] [PubMed] [Google Scholar]
  • 120.Bernard M.L. Pacing without wires: leadless cardiac pacing. Ochsner J. 2016;16:238–242. [PMC free article] [PubMed] [Google Scholar]
  • 121.Tjong F.V.Y., Koop B.E. The modular cardiac rhythm management system: the EMPOWER leadless pacemaker and the EMBLEM subcutaneous ICD. Herzschrittmacherther Elektrophysiol. 2018;29:355–361. doi: 10.1007/s00399-018-0602-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 122.Rashtian M., Banker R.S., Neuzil P., et al. Preclinical safety and electrical performance of novel atrial leadless pacemaker with dual-helix fixation. Heart Rhythm. 2022;19:776–781. doi: 10.1016/j.hrthm.2022.01.021. [DOI] [PubMed] [Google Scholar]
  • 123.Sidhu B.S., Gould J., Porter B., et al. Completely leadless cardiac resynchronization defibrillator system. JACC Clin Electrophysiol. 2020;6:588–589. doi: 10.1016/j.jacep.2020.02.012. [DOI] [PubMed] [Google Scholar]
  • 124.Tarkowski P., Siek E., Staśkiewicz G., Bielecki D.K., Czekajska-Chehab E. Angle dependence of electrode lead-related artifacts in single- and dual-energy cardiac ECG-gated CT scanning—a phantom study. J Clin Med. 2024;13:3746. doi: 10.3390/jcm13133746. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 125.Kidoh M., Oda S., Nakato K., et al. Assessment of cardiac implantable electric device lead perforation using a metal artifact reduction algorithm in cardiac computed tomography. Eur J Radiol. 2021;136 doi: 10.1016/j.ejrad.2021.109530. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

This study did not involve the creation or analysis of any new data. Therefore, data sharing is not applicable.


Articles from Heart Rhythm O2 are provided here courtesy of Elsevier

RESOURCES