Impact of slice thickness on CACS calculation with virtual non-contrast in photon-counting CT

Qiuju Hu 1,#, Huixin Zhang 1,#, Bangjun Guo 2,#, Dongsheng Jin 1, Meirong Sun 1, Jiliang Chen 4, Song Luo 1,✉, Yane Zhao 1,✉, Guang-ming Lu 3
PMCID: PMC12903409  PMID: 41559569

Abstract

Background

This study aims to investigate the feasibility of coronary artery calcium scoring (CACS) calculating from PureCalcium virtual non-iodine algorithm on photon-counting detector CT (PCD-CT) and the potential impact of different section thickness, level of virtual monoenergetic images (VMIs), and quantum iterative reconstruction (QIR) on the accuracy of CACS quantification.

Materials and methods

A total of 123 patients who underwent coronary CT angiography on PCD-CT with a separate true non-contrast CACS (CACSTNC) scan were prospectively included. Agatston scores were calculated from the PureCalcium algorithm (CACSPC) using a section thickness of 3 mm–1.5 mm, different VMI (55–75 kilo-electron volt (keV)) and QIR (strength 1,4) levels, respectively. CACSTNC at 70 keV and QIR 2 were used as reference standards. Differences in CACS of different reconstructions section thicknesses, various keV levels, and QIR strength were compared using the Wilcoxon rank sum test with Bonferroni correction. The intraclass correlation coefficients (ICCs) and Bland-Altman analysis were conducted to assessed the agreement. The agreement of plaque burden groups (based on CACS) at different reconstruction parameters was evaluated using weighted Cohen kappa.

Results

At all investigated section thickness, VMI, and QIR levels, the CACSPC were strongly correlated with CACSTNC (ICC: 0.94–0.98, P < 0.001 for all). There were no statistical differences in CACS between CACSPC at 3 mm section thickness, 60/65 keV (QIR1/4), and at 1.5 mm section thickness with 55 keV (QIR1/4), compared with CACSTNC. The smallest CACS bias was observed at a 1.5 mm section thickness, 55 keV, QIR 1, with mean bias of 2.4; LoA (IQR: −182.7, 187.4). CACSPC correctly identified 105 of 123 participants (85.4%) into the corresponding plaque burden group using CACSTNC as the referent standard (excellent agreement, κ = 0.904).

Conclusion

CACS derived from the PureCalcium algorithm with optimized reconstruction parameters shows excellent correlation with true non-contrast scans derived values. Thus, it is may possible to use the PureCalcium virtual non-iodine algorithm to replace the true non-contrast scans for CACS quantification, without additional radiation dose exposure.

Keywords: Coronary artery calcium scoring, Spectral imaging, Virtual non-iodine algorithm, Coronary artery disease, Photon counting CT

Introduction

Coronary artery calcium score (CACS), which is obtained from non-contrast CT scan, has been demonstrated as an independent prognostic predictor for all-cause mortality in asymptomatic individuals [1]. The progression of coronary calcification remained as an independent predictor for all-cause mortality after adjusting for baseline CACS, interval time of scan, and clinical risk factors [2]. Nowadays, CACS is recommended at a class IIa level of the 2019 AHA guidelines for patients with borderline/intermediate risk and is helpful in guiding clinical management and treatment [3, 4].

In routine clinical examinations, a separate non-contrast CT scan before coronary CT angiography (CCTA), was conducted to quantify the CACS, which may increase the examination time and lead to extra radiation dose. With the advent of dual-energy and spectral imaging techniques, several studies [5–7] have explored the assessment of the CACS on virtual non-contrast (VNC) images. These studies using dual-energy techniques, and have demonstrated high correlations but underestimated CACS compared to the true non-contrast (TNC) images. With the advent of photon-counting detector CT (PCD-CT), it has been shown to have higher spatial resolution and reduced radiation dose. Additionally, PCD-CT data contain differentiated spectral information, which can be used for multi-material decomposition [8].

In this context, a novel virtual noniodine (VNI) algorithm has been developed to create PureCalcium reconstructions on PCD-CT data. Previous studies have demonstrated that the PureCalcium algorithm achieves significantly higher accuracy in CACS quantification than VNC reconstructions [9]. Of note, current virtual non-iodine reconstructions consistently demonstrate a systematic underestimation of coronary artery calcium scores compared to true non-contrast reference standards. While previous investigations have examined the individual and combined effects of virtual monoenergetic image energy levels and iterative reconstruction strength, these studies have critically overlooked the fundamental parameter of slice thickness [10, 11]. This omission represents a substantial limitation in the existing literature, given that slice thickness directly governs partial volume effects and significantly impacts the detectability of small calcifications. Consequently, the optimal tripartite combination of slice thickness, keV level, and iterative reconstruction strength for minimizing discrepancies between virtual and true non-contrast calcium scoring remains undetermined. The systematic investigation of these three interdependent parameters is therefore essential for establishing standardized reconstruction protocols that can enable reliable cardiovascular risk assessment and facilitate the clinical adoption of virtual non-contrast calcium scoring.In this study, we aimed to assess the impact of different VMIs, QIR levels, and section thickness on the accuracy of CACS quantification on the PureCalcium algorithm on PCD-CT.

Materials and methods

Study patients

A total of 123 patients who underwent separate TNC and contrast-enhanced photon-counting CCTA due to suspected coronary artery disease between January 2024 and April 2024 were prospectively enrolled. Inclusion criteria included subjects older than 18 who underwent CCTA examinations by standard spatial resolution scanning. Exclusion criteria were as follows: (1) patients with CACS = 0; (2) patients who had a history of coronary revascularization (percutaneous coronary intervention or coronary artery bypass graft). This study was approved by our ethics committee (YJ-2024-044-1), and written informed consent was obtained.

Data acquisition

All patients were scanned using a PCD-CT (Naeotom Alpha, software version VA50, Siemens Healthineers). TNC scans were performed using the prospective ECG-gated sequential technique, the tube voltage was set at 120 kVp, and the tube current-time product was set to an IQ level of 19. CCTA scans were selected as prospective and retrospective based on the participants’ heart rate and rhythm. The tube voltage was set at 120 kVp, and the tube current-time product was set to an IQ level of 64. The contrast materials were applied using bolus tracking with an enhancement threshold of 100 HU in the descending aorta and a time delay of 6 s. A total of 45 to 60 mL of an iodinated contrast agent (350 mg iohexol/mL, Omnipaque; GE HealthCare) was injected at a flow rate of 4.5 mL/sec. Subsequently, 40 mL of saline was injected with the same rate. If not clinically contraindicated, 0.4 mg of nitroglycerin was sublingually administered approximately 5 min before the scan [12].

Image reconstruction

TNC images were reconstructed using a section thickness of 3.0 mm with an increment of 1.5 mm at 70 keV, QIR 2, and a recommended kernel Qr36, as defined by the vendor’s clinical calcium scoring protocol. The CCTA images were post-processed using a novel VNI algorithm (PureCalcium, Siemens Healthineers) with Qr36 kernel. For all of the patient data, VMIs were reconstructed at 5-keV intervals from 55 to 75 keV with QIR level 1 or 4 for each keV level and section thickness of 3.0 mm with an increment of 1.5 mm or section thickness of 1.5 mm with an increment of 1.0 mm.

Calcium scoring analysis

The TNC and PureCalcium datasets were evaluated on a semi-automated workflow (CT CaScoring, syngo.via VB60; Siemens Healthineers). Calcified lesion attenuation exceeding 130 HU and a minimum area of 0.5 mm2 was automatically identified as calcium. Manual adjustments were made as necessary. The calcium scoring calculation was performed by two radiologists (H.X.Z. and Y.E.Z., with > 10 years of experience in cardiovascular imaging) on anonymized data sets provided on a syngo.via server. They were blinded to clinical information and independently assessed all data. Disagreements in calcium scoring between the two readers were resolved by consensus reading involving a third senior radiologist (with > 15 years of experience). A direct comparison was performed between the PureCalcium algorithm (CACSPC) using prior reconstruction parameters (55 keV/QIR 1 and 60 keV/QIR 4, 3.0 mm section thickness) [9] and true non-contrast CT (CACSTNC, 70 keV/QIR 2) to determine the patients with (CACSTNC> 0) that was not detected by the previous reconstruction protocol. The smallest mean difference of CACS determined the combination of optimal reconstruction parameters for calcium detection compared to the true non-contrast scan.

The burden of calcium plaque in each patient was classified according to the Coronary Artery Disease Reporting System (P0 = score of 0 [none]; P1 = 1–100 [mild]; P2 = 101–300 [moderate]; P3 = 301–999 [severe]; P4 > = 1000 [extensive]) [10].

Statistical analysis

Continuous data were described as mean ± SD, while categorical data were expressed as frequencies and proportions. The Kolmogorov-Smirnov test was used to test the continuous data for normality. Spearman’s correlation coefficient (r) was used to assess the correlation between CACSTNC and CACSPC. The differences in calculated CACS with different reconstruction methods were compared with the referent standard (CACSTNC) using the Friedman test with Bonferroni correction. Intraclass correlation coefficient (ICC) and Bland-Altman analysis were used for agreement assessment. The agreement of plaque burden classification between CACSTNC and CACSPC was analyzed using Cohen’s kappa (κ) statistics. Statistical analysis was performed using SPSS 27 (IBM Corporation, Armonk, NY) and MedCalc 22.0 (MedCalc, Ostend, Belgium).

Results

Patient characteristics

A total of 123 patients, including 78 males, were enrolled in this study. The mean age was 69.9 ± 9.5 years, with a range of 45–89 years. Table 1 summarizes the demographic characteristics of the included patients in detailed.

Table 1.

Characteristics of the enrolled patients

Characteristic Value
Male/Female 78(63)/45(37)
Age, years 69.9 ± 9.5 (45–89)
Heart rate during CCTA, bpm 68.0 ± 17.3 (58–123)
Body mass index, kg/m2 25.1 ± 2.4
Total CACS 109.7 (36.95, 485.95)
Risk factors, n (%)
Hypertension 96 (78)
Hyperlipidemia 77 (63)
Diabetes mellitus 41 (33)
Smoking 76 (62)
Family history of coronary artery disease 32 (42)

Values are mean ± SD, median (interquartile range), or n (%). CCTA, coronary CT angiography; CACS, coronary artery calcium scoring

Comparison of CACSTNC and CACSPC using different reconstruction parameters

The median value of CACSTNC was 109.7 (IQR: 36.9–496.9). With the reconstruction parameters of 3 mm section thickness and an increment of 1.5 mm, the value of median CACSPC ranged from 83.2 (IQR: 5.5–399.7; 75 keV, QIR 4) to 132.5 (IQR: 20.2–570.1; 55 keV, QIR 1) depending on different level of VMI and QIR. For the reconstruction parameters of 1.5 mm section thickness and an increment of 1 mm, the median value of CACSPC ranged from 76.9 (IQR: 10.8–364.5; 75 keV, QIR 4) to 107.4 (IQR: 17.0–474.6; 55 kV, QIR 1). Regardless of whether the reconstruction section thickness is 3 mm–1.5 mm, CACSPC strongly correlates with CACSTNC (r = 0.93 and r = 0.96, respectively, P < 0.001) and shows excellent agreement (ICC between 0.97 and 0.98 for all) for different keV and QIR levels. No statistical differences were observed in CACSPC at 3 mm section thickness, 55 keV (QIR4), 60/65 keV (QIR1/4), and at 1.5 mm section thickness with 55 keV (QIR1/4), 60 keV (QIR1) compared with CACSTNC (Fig. 1). Detailed results of correlation and agreement analysis are illustrated in Table 2. The smallest mean bias was obtained at 1.5 mm section thickness, 55 keV with QIR 1 (CACS: 107.5 [IQR: 18.9–480.4] compared to CACSTNC: 109.7 [IQR: 36.9–496.9] mean bias, 2.3; LoA, (− 182.7/187.4)) (Fig. 2). Regardless of section thickness and QIR level, CACSPC values were gradually decreased with increased keV values (P < 0.001 for all) (Fig. 1).

Fig. 1.

Fig. 1

Impact of different section thickness, VMI and QIR levels compared with CACSTNC. Both CACSTNC and CACSpc are illustrated as median. ns indicates not significant. ***P < 0.001; *P < 0.05; ns > = 0.05. QIR, quantum iterative reconstruction; VMI, virtual monoenergetic image; CACS, coronary artery calcium scoring

Table 2.

Comparison of CACSTNC and CACSPC with different section thicknesses, VMI, and QIR

Reconstruction methods Section thickness /increment(mm/mm) VMI (keV) QIR Agatston Score (IQR) Friedman P r ICC Bias LoA
TNC 3/1.5 70 2 109.7 (36.87,496.92) - - - - -
Pure Calcium 3/1.5 55 1 132.5 (20.25,570.15) 0.002 0.96 0.97 48.63 -206.86/304.15
4 126.4 (18.87,552.82) >0.99 0.95 0.98 43.77 -197.32/284.87
60 1 121.6 (17.12,513.02) >0.99 0.96 0.98 19.89 -185.98/225.77
4 112.5 (11.37,502.67) >0.99 0.95 0.98 15.21 -180.13/210.57
65 1 108.4 (13.47,473.87) >0.99 0.95 0.98 -5.56 -187.44/176.31
4 94.1 (7.47,470.07) 0.025 0.93 0.98 -13.78 -203.21/175.65
70 1 100.2 (11.72,438.65) <0.001 0.95 0.98 -28.07 -214.63/158.48
4 89.1 (5.50,438.20) <0.001 0.94 0.98 -34.86 -228.95/159.23
75 1 89.7 (9.65, 408.85) <0.001 0.95 0.98 -47.60 -256.27/161.07
4 83.2 (5.52,399.70) <0.001 0.94 0.98 -54.33 -271.89/163.23
1.5/1.0 55 1 107.5 (18.90,480.37) >0.99 0.96 0.98 2.35 -182.68/187.39
4 107.4 (17.02, 474.55) >0.99 0.95 0.98 -7.31 -186.34/171.71
60 1 99.8 (17.12,452.80) >0.99 0.96 0.98 -18.45 -190.94/154.04
4 99.4 (15.90,438.70) 0.001 0.95 0.98 -27.86 -213.66/157.93
65 1 93.1 (15.05,429.60) 0.001 0.96 0.98 -36.89 -226.26/152.48
4 90.1 (14.72,412.90) 0.001 0.95 0.98 -46.17 -248.44/156.10
70 1 89.3 (63.16,441.24) <0.001 0.96 0.98 -52.38 -264.78/160.00
4 80.7 (13.55,381.72) <0.001 0.95 0.98 -62.29 -287.98/163.39
75 1 86.3 (11.92,389.10) <0.001 0.95 0.98 -64.67 -297.70/168.34
4 76.9(10.77, 364.47) <0.001 0.96 0.98 -75.24 -326.30/175.82

Values are median (interquartile range)

TNC indicates true noncontrast; VMI, virtual monoenergetic images; QIR, quantum iterative reconstruction; ICC, intraclass correlation; LoA, limits of agreement

P value < 0.05 was considered statistically significant

Fig. 2.

Fig. 2

Bland-Altman analysis of CACS compared between TNC and PureCalicum algorithm at 1.5 mm,55 keV, QIR1. QIR, quantum iterative reconstruction; TNC, true noncontrast; CACS, coronary artery calcium scoring

Agreement between the 2 readers was excellent for both CACSPC (ICC = 0.99 for all) and CACSTNC (ICC = 0.98 for all).

Comparison of standard purecalcium and optimized purecalcium algorithms for tiny calcium detection

Of the 123 participants included, 21 (17.7%) participants had CACSTNC >0 but were not detected on standard PureCalcium parameters reconstructions. The median CACSTNC was 12.4 (IQR: 6.4–19.35). Of those, 13 (61.9%) participants were detected using optimized PureCalcium reconstructions with a section thickness of 1.5 mm and an increment of 1.0 mm, 55 keV, and QIR 1. Figure 3 shows representative cases of CACSTNC and section thickness 3.0 mm standard CACSPC compared with section thickness 1.5 mm CACSPC using the optimized reconstruction.

Fig. 3.

Fig. 3

Example cases of axial CT reconstructions of CACS. Comparing CACSTNC (left column: 70 keV, quantum iterative reconstruction QIR 2), CACSPC at standard reconstruction (middle column: 3 mm/1.5 mm,55 keV, QIR 1 and 60 keV, QIR 4, as previously recommended) and CACSPC at the optimized reconstruction (right column: 1.5 mm/1 mm,55 keV, QIR 1). The upper row (patient 1) shows CT scans in a 75-year-old male and the lower row (patient 2) in a 61-year-old male. The optimized reconstruction improved the detection of tiny calcified plaques (circles) initially detected in TNC scans but not in standard PureCalcium reconstructions. LCX, left Circumflex Artery; RCA, right coronary artery; CACS, coronary artery calcium scoring; QIR, quantum iterative reconstruction; TNC, true noncontrast

Impact of differences in CACS on plaque burden classification

In the P classification of the Coronary Artery Disease Reporting System, 56, 26, 32, and 9 patients were classified as P1, P2, P3, and P4, respectively, using CACSTNC as the reference standard. No statistical difference was observed in P classification overall between CACSTNC and CACSPC images (P = 0.06). Of note, 18 patients had different P classification results between CACSPC and CACSTNC. On PureCalcium images, 8 patients were misdiagnosed as P0 instead of P1 on TNC images. In these patients, the median CACSPC was 3.95 (range: 0.3 to 10.2). Additionally, 3 patients were misclassified as P1 instead of P2, and 1 was misclassified as P3 instead of P2. P2 and P4 were misdiagnosed instead of P3 in 1 patient and 4 patients, respectively. Furthermore, 1 patient was misclassified as P3 instead of P4 (Tables 3 and 4) (Fig. 4). The categorical agreement of plaque classification between CACSTNC and CACSPC was excellent (κ = 0.87).

Table 3.

CACS distribution in TNC and purecalcium reconstruction

Scan Type and CACS
CACS burden PureCalcium images TNC images
0 8 (6) 0
1-100 51 (41) 56 (46)
101–300 23 (19) 26 (21)
301–999 29 (24) 32 (26)
> 1000 12 (10) 9 (7)

Data are numbers of patients, with percentages in parentheses. TNC, true noncontrast; CACS, coronary artery calcium scoring

Table 4.

CAD-RADS P classification based on CACSTNC images or CACSPC images

CAD-RADS P Classification TNC P1 TNC P2 TNC P3 TNC P4 Total
PureCalcium P0 8 0 0 0 8
PureCalcium P1 48 3 0 0 51
PureCalcium P2 0 22 1 0 23
PureCalcium P3 0 1 27 1 29
PureCalcium P4 0 0 4 8 12
Total 56 26 32 9 123

CAD-RADS, Coronary Artery Disease Reporting and Data System; CACS, coronary artery calcium scoring

Fig. 4.

Fig. 4

Sankey diagrams of change in Plaque Burden Classification at TNC and PureCalcium. A total of 8(6%) patients were misdiagnosed as P0 instead of P1 on Purecalcium images. 3 (2%)patients were misclassified as P1 instead of P2, and 1(0.8%) patient were misclassified as P3 instead of P2. P2 and P4 were misdiagnosed instead of P3 in 1(0.8%) and 4 (3%)participants, respectively. 1(0.8%) participant was misclassified as P3 instead of P4.TNC, true noncontrast

Discussion

Our study aimed to evaluate the feasibility of CACS quantification based on the PureCalcium algorithm from contrast-enhanced CCTA and the effects of different section thickness, VMI, and QIR levels during imaging reconstruction on the accuracy of CACS calculation. The findings of our study are as follows: First, the value of CACSPC showed excellent correlation and agreement compared to CASCTNC. Second, the value of CACSPC at a reconstruction parameter of a section thickness of 1.5 mm, 55 kV (QIR 1), is closest to CACSTNC.

In this study, only QIR1 and QIR4 were used for imaging reconstruction [7], consistent with previous studies that used different imaging techniques to evaluate the accuracy and consistency of calcium imaging. Compared with the energy integral detector CT, PCD-CT has higher spatial resolution and can improve dose efficiency. PCD-CT can utilize higher spectral separation technology for multi-substance separation and generate advanced spectral post-processed images, such as pure calcium images, which can significantly improve the accuracy of CACS calculation both in vitro and in vivo studies [10, 12– 13]. However, standard PureCalcium reconstruction is unable to detect very low-density calcifications accurately. This may be attributed to the following reasons: the limited detectability of low-density calcifications and partial volume effects due to the section thickness of 3 mm [14–17]. Vliegenthart et al. showed that the calcium volumes were higher in 1.5 mm slices than in 3 mm in vitro and in vivo [18]. Georg et al. observed that a thinner section was associated with increased CACS values [19]. Our study demonstrated that the 1.5 mm slice thickness can detect more tiny calcification, which can guide the use of this reconstruction parameter to detect small and low-density plaques.

A key finding regarding the thickness-VMIs that minimises the bias in CACS has been identified. For 3-mm slice thickness, 65 keV resulted in the lowest bias; while 1.5-mm thickness were optimised at 55 keV. This elucidates the unique physical trade-offs in the virtual non-contrast technique for CACS. For 3-mm slices, partial volume effects is a dominant limiting factor. Small or low-attenuation calcifications are susceptible to being averaged with surrounding soft tissue, reducing their apparent CT attenuation [20]. A higher monoenergetic level (65 keV) provides crucial benefits: better suppression of residual intravascular iodine and an improved signal-to-noise ratio. These effects enhance the specificity of calcium identification by mitigating false-positive classifications arising from misregistration or noise. Conversely, 1.5-mm slices have better spatial resolution, allowing for more accurate representation of small calcifications. Lowering the monoenergetic energy to 55 keV significantly elevates the contrast-to-noise ratio, optimising the sensitivity for detecting subtle and low-density calcifications.

According to CACSPC, the agreement on risk categorization was acceptable. In this study, 85% of patients’ CACSPC burdens were correctly categorized using CACSTNC as the reference. It is essential for risk stratification and management recommendations. Although most CACSPC misdiagnoses occur at P1, which may result in false-negative CACS, our study partially addressed this issue by employing 1.5 mm sections. So, CACS calculation on the PureCalcium algorithm with a reconstruction section of 1.5 mm may be a more accurate way for CACS quantification.

The observed discrepancies in individual patient CACS grading—encompassing both over- and under-estimation—are inherent to the spectral material decomposition process.The predominance of overestimation can be attributed to incomplete spectral separation in regions exhibiting high iodine concentrations. In such cases, residual contrast signal may be erroneously classified as calcification. Concurrently, elevated image noise at lower virtual monoenergetic levels and edge-blurring effects from high-strength iterative reconstruction have been shown to artificially amplify calcium measurements. Conversely, underestimation occurs as a result of inadvertent signal subtraction of low-attenuation calcifications whose spectral profiles overlap with iodine-enhanced soft tissue, particularly at lower energy levels. Furthermore, the derived CACSPC imaging are susceptible to signal depression from decomposition inaccuracies or image noise. This can cause the calculated attenuation of truly calcified voxels to fall below the 130-HU threshold, thereby leading to underestimation.

These technical trade-offs directly inform differentiated clinical protocols.When the primary clinical objective is the detection of early or minimal coronary calcification—such as in high-risk primary prevention screening—the 1.5/1 mm and 55-keV reconstruction protocol is recommended.

Although this study has demonstrated the excellent correlation between CACSPC and CACSTNC and showed improved identification of tiny coronary calcifications by using 1.5 mm section thickness images, several limitations should be noted. Firstly, there were no detailed analytical results based on different QIR levels; only QIR1 and QIR4 were used in this study. The CACS calculated from other levels of QIRs should be further investigated. Secondly, the quantification of CACS can be affected by various factors; only VMIs, iterative reconstruction levels, and slice parameters were investigated; other parameters, such as different densities of calcification in terms of position, breath-hold depth, and heart rate, require further investigation. Thirdly, a relatively small number of participants were included for analysis; the consistency between CACSPC and CACSTNC in larger-scale populations with different severities of calcification should be further investigated. Fourthly, this study is a single center and only involved one CT vendor; the results should be interpreted carefully in other CT vendors with different virtual noniodine algorithms.

In conclusion, the CACS calculated from the PureCalcium algorithm showed excellent correlations with true non-contrast images. The slice thickness, VMI, and QIR levels can affect the results of CACS. Compared to CACSTNC, the lower CACSPC can be mitigated by using thinner sections and lower keV reconstructions. This method facilitates accurate PureCalcium-based assessment of CACS in clinical practice and potentially eliminates the need for separate non-enhanced scans, thereby reducing radiation dose.

Acknowledgements

Not applicable.

Abbreviations

CACS

Coronary artery calcium scoring

PCD-CT

Photon-counting detector CT

VMIs

Virtual monoenergetic images

QIR

Quantum iterative reconstruction

VNC

Virtual non-contrast

TNC

True non-contrast

CCTA

Coronary CT angiography

ICC

Intraclass correlation coefficient

Author contributions

Qiuju hu: Writing original draft, Conceptualization, Project administration, Writing - review & editing. Huixin zhang: Data curation, Formal analysis, Bangju guo: Conceptualization, review & editing. Dongsheng Jin: Resource, Meirong Sun: Data curation JiliangJilang Chen: MethodologySong Luo: Investigation, Resources, Software, Writing - review & editing. Yane Zhao: Formal analysis, Methodology, Writing - review & editingGuang-ming Lu: Investigation, Resource.

Funding

This research received no external funding.

Data availability

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Declarations

Ethical approval

This study was approved by the Ethics Committee of Geriatric Hospital of Nanjing Medical University(YJ-2024-044-1) in accordance with the Declaration of Helsinki.

Consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Qiuju Hu, Huixin Zhang and Bangjun Guo contributed equally to this work.

Contributor Information

Song Luo, Email: hnldls@163.com.

Yane Zhao, Email: 15335137@qq.com.

References

  • 1.Parsa S, Saleh A, Raygor V, et al. Measurement and application of incidentally detected coronary calcium: JACC review topic of the week. J Am Coll Cardiol. 2024;83(16):1557–67. 10.1016/j.jacc.2024.01.039. [DOI] [PubMed] [Google Scholar]
  • 2.Golub IS, Termeie OG, Kristo S, et al. Major global coronary artery calcium guidelines. JACC Cardiovasc Imaging. 2023;16(1):98–117. 10.1016/j.jcmg.2022.06.018. [DOI] [PubMed] [Google Scholar]
  • 3.Arnett DK, Blumenthal RS, Albert MA et al. (2019) 2019 ACC/ AHA guideline on the primary prevention of cardiovascular disease: a report of the American College of Cardiology/American Heart Association Task Force on clinical practice guidelines. Circulation, 2019;140:e596-e646. 10.1161/CIR.0000000000000678. [DOI] [PMC free article] [PubMed]
  • 4.Jiang Q, Zhang N, Zhang H, et al. Impact of off-hour admission on the maces of patients with acute myocardial infarction. Clin Exp Hypertens. 2023;45(1):2186317. 10.1080/10641963.2023.2186317. [DOI] [PubMed] [Google Scholar]
  • 5.Gassert FG, Schacky CE, Müller-Leisse C, et al. Calcium scoring using virtual non-contrast images from a dual-layer spectral detector CT: comparison to true non-contrast data and evaluation of proportionality factor in a large patient collective. Eur Radiol. 2021;31(8):6193–9. 10.1007/s00330-020-07677-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Yang P, Zhao R, Deng W, et al. Feasibility and accuracy of coronary artery calcium score on virtual non-contrast images derived from a dual-layer spectral detector CT: A retrospective multicenter study. Front Cardiovasc Med. 2023;10:1114058. 10.3389/fcvm.2023.1114058. Published 2023 Mar 2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Langenbach IL, Wienemann H, Klein K, et al. Coronary calcium scoring using virtual non-contrast reconstructions on a dual-layer spectral CT system: feasibility in the clinical practice. Eur J Radiol. 2023;159:110681. 10.1016/j.ejrad.2022.110681. [DOI] [PubMed] [Google Scholar]
  • 8.Sandfort V, Persson M, Pourmorteza A, et al. Spectral photon-counting CT in cardiovascular imaging. J Cardiovasc Comput Tomogr. 2021;15(3):218–25. 10.1016/j.jcct.2020.12.005. [DOI] [PubMed] [Google Scholar]
  • 9.Emrich T, Aquino G, Schoepf UJ, et al. Coronary computed tomography Angiography-Based calcium scoring: in vitro and in vivo validation of a novel virtual noniodine reconstruction algorithm on a Clinical, First-Generation Dual-Source photon Counting-Detector system. Invest Radiol. 2022;57(8):536–43. 10.1097/RLI.0000000000000868. [DOI] [PubMed] [Google Scholar]
  • 10.Fink N, Zsarnoczay E, Schoepf UJ, et al. Photon counting detector CT-Based virtual noniodine reconstruction algorithm for in vitro and in vivo coronary artery calcium scoring: impact of virtual monoenergetic and quantum iterative reconstructions. Invest Radiol. 2023;58(9):673–80. 10.1097/RLI.0000000000000959. [DOI] [PubMed] [Google Scholar]
  • 11.Leipsic J, Abbara S, Achenbach S, et al. SCCT guidelines for the interpretation and reporting of coronary CT angiography: a report of the society of cardiovascular computed tomography guidelines committee. J Cardiovasc Comput Tomogr. 2014;8(5):342–58. 10.1016/j.jcct.2014.07.003. [DOI] [PubMed] [Google Scholar]
  • 12.Cury RC, Leipsic J, Abbara S, et al. CAD-RADS™ 2.0–2022 Coronary Artery Disease - Reporting and Data System. An expert consensus document of the Society of Cardiovascular Computed Tomography (SCCT), the American College of Cardiology (ACC), the American College of Radiology (ACR), and the North American Society of Cardiovascular Imaging (NASCI). J Am Coll Radiol. 2022;19(11):1185–212. 10.1016/j.jacr.2022.09.012. [DOI] [PubMed]
  • 13.Sandstedt M, Marsh J Jr, Rajendran K, et al. Improved coronary calcification quantification using photon-counting-detector CT: an ex vivo study in cadaveric specimens. Eur Radiol. 2021;31(9):6621–30. 10.1007/s00330-021-07780-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Van der Werf NR, Booij R, Greuter MJW, et al. Reproducibility of coronary artery calcium quantification on dual-source CT and dual-source photon-counting CT: a dynamic Phantom study. Int J Cardiovasc Imaging. 2022;38(7):1613–9. 10.1007/s10554-022-02540-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Mergen V, Higashigaito K, Allmendinger T, et al. Tube voltage-independent coronary calcium scoring on a first-generation dual-source photon-counting CT proof-of-principle Phantom study. Int J Cardiovasc Imaging. 2022;38(4):905–12. 10.1007/s10554-021-02466-y. [DOI] [PubMed] [Google Scholar]
  • 16.Skoog S, Henriksson L, Gustafsson H, Sandstedt M, Elvelind S, Persson A. Comparison of the Agatston score acquired with photon-counting detector CT and energy-integrating detector CT: ex vivo study of cadaveric hearts. Int J Cardiovasc Imaging. 2022;38(5):1145–55. 10.1007/s10554-021-02494-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.van Praagh GD, Wang J, van der Werf NR, et al. Coronary artery calcium scoring: toward a new standard. Invest Radiol. 2022;57(1):13–22. 10.1097/RLI.0000000000000808. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Vliegenthart R, Song B, Hofman A, et al. Coronary calcification at electron-beam CT: effect of section thickness on calcium scoring in vitro and in vivo. Radiology. 2003;229(2):520–5. 10.1148/radiol.2292021305. [DOI] [PubMed] [Google Scholar]
  • 19.Mühlenbruch G, Thomas C, Wildberger JE, et al. Effect of varying slice thickness on coronary calcium scoring with multislice computed tomography in vitro and in vivo. Invest Radiol. 2005;40(11):695–9. 10.1097/01.rli.0000179523.07907.a6. [DOI] [PubMed] [Google Scholar]
  • 20.Kim SY, Suh YJ, Lee HJ, et al. Influence of computed tomography slice thickness on deep learning-based, automatic coronary artery calcium scoring software performance. Quant Imaging Med Surg. 2023;13(7):4257–67. 10.21037/aims-22-835. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Citations

  1. Cury RC, Leipsic J, Abbara S, et al. CAD-RADS™ 2.0–2022 Coronary Artery Disease - Reporting and Data System. An expert consensus document of the Society of Cardiovascular Computed Tomography (SCCT), the American College of Cardiology (ACC), the American College of Radiology (ACR), and the North American Society of Cardiovascular Imaging (NASCI). J Am Coll Radiol. 2022;19(11):1185–212. 10.1016/j.jacr.2022.09.012. [DOI] [PubMed]

Data Availability Statement

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.


Articles from BMC Medical Imaging are provided here courtesy of BMC

RESOURCES