Troubleshooting: intraperitoneal migration of a dedicated plastic stent during endoscopic ultrasound-guided hepaticogastrostomy

Hirotsugu Maruyama 1,✉, Yuki Ishikawa-Kakiya 1, Yuji Kawata 1, Tatsuya Kurokawa 1, Yoshinori Shimamoto 1, Kojiro Tanoue 1, Yasuhiro Fujiwara 1
PMCID: PMC12912853  PMID: 41702584

Complications of endoscopic ultrasound-guided hepaticogastrostomy (EUS-HGS) have decreased with improved techniques and devices, with cholangitis and bleeding now being the most common 1 . Although fatal stent migration was previously a major concern, anti-migration stents have made it rare 2 3 . Nevertheless, severe complications can still occur due to endoscopist inattention or misjudgment.

In this case, a dedicated plastic stent migrated intraperitoneally because the placement relied solely on fluoroscopy alone. An additional EUS-HGS via B3 was performed, allowing the successful retrieval of the migrated stent through the new tract.

A 66-year-old woman with prior gastrojejunostomy for duodenal obstruction due to cholecystitis was referred for cholangitis caused by intrahepatic stones. EUS-HGS was attempted by placing a dedicated plastic stent (PS). However, the stones prevented deep stent advancement, and it was inserted using fluoroscopic guidance alone. The stent was not seen endoscopically, leading us to conclude that it had been inadvertently deployed into the peritoneal cavity ( Fig. 1 ). We then punctured B3 and placed a metal stent (MS) to create a wider drainage tract and reduce intraperitoneal leakage. Through the MS tract, we attempted to locate the PS fluoroscopically but could not identify it. Using a 22-gauge needle and a 0.018-inch guidewire, we accessed the PS in the hepatic parenchyma, but could not direct the guidewire toward the MS tract. Finally, we succeeded in advancing a GW into the lumen of the migrated PS via the MS route. Using a drill dilator, we were able to retrieve the stent into the stomach ( Video 1 ). The postoperative course was uneventful ( Fig. 2 ).

Fig. 1.

Fig. 1

A fluoroscopic image of endoscopic ultrasound-guided hepaticogastrostomy. a A fluoroscopic image of cholangiography showing intrahepatic stones. b A fluoroscopic image after the migration of the dedicated plastic stent into the peritoneal cavity.

Fig. 2.

Fig. 2

The clinical course after the migration of the dedicated plastic stent. a Endoscopic ultrasound-guided hepaticogastrostomy was performed at B3, and a fully covered metal stent was placed. b A fluoroscopic image showing the insertion of the drill dilator into the stent lumen. c After pulling the stent proximally toward the endoscope using the drill dilator, it was grasped with a snare. d A post-endoscopic procedural CT image. No obvious bile and contrast leakage were observed. CT, computed tomography.

Download video file (65MB, mp4)

Troubleshooting: intraperitoneal migration of a dedicated plastic stent during endoscopic ultrasound-guided hepaticogastrostomy.

Video 1

This case serves as an important reminder that endoscopic and fluoroscopic imaging must be fully utilized. It also presents a method for retrieving a stent even if it migrates into the abdominal cavity, making it a very rare and valuable report.

Endoscopy_UCTN_Code_TTT_1AS_2AH

Footnotes

Conflict of Interest The authors declare that they have no conflict of interest.

Contributorsʼ Statement Hirotsugu Maruyama: Conceptualization, Writing – original draft. Yuki Ishikawa-Kakiya: Writing – review & editing. Yuji Kawata: Writing – review & editing. Tatsuya Kurokawa: Writing – review & editing. Yoshinori Shimamoto: Writing – review & editing. Kojiro Tanoue: Writing – review & editing. Yasuhiro Fujiwara: Writing – review & editing.

Reference

  • 1.Binda C, Dajti E, Giuffrida P et al. Efficacy and safety of endoscopic ultrasound-guided hepaticogastrostomy: a meta-regression analysis. Endoscopy. 2024;56:694–705. doi: 10.1055/a-2282-3350. [DOI] [PubMed] [Google Scholar]
  • 2.Ishii S, Isayama H, Sasahira N et al. A pilot study of Spring Stopper Stents: Novel partially covered self-expandable metallic stents with anti-migration properties for EUS-guided hepaticogastrostomy. Endosc Ultrasound. 2023;12:266–272. doi: 10.4103/EUS-D-22-00104. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Matsunami Y, Itoi T, Sofuni A et al. EUS-guided hepaticoenterostomy with using a dedicated plastic stent for the benign pancreaticobiliary diseases: A single-center study of a large case series. Endosc Ultrasound. 2021;10:294–304. doi: 10.4103/EUS-D-20-00232. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Endoscopy are provided here courtesy of Thieme Medical Publishers

RESOURCES