Abstract
Correlative microscopy techniques are used for many different applications in the biological sciences because the comparison of different imaging methods allows researchers to gain more insight and data from samples. Correlative light and electron microscopy (CLEM) methods have been developed to preserve biological samples to withstand the harsh environments necessary for electron microscopy. After first being imaged using widefield (WF) and super-resolution structured illumination fluorescence microscopy (SIM), a NanoSuit chemical treatment was applied to a mammalian testis sample before imaging with scanning electron microscopy (SEM). This was done to compare the image quality and resolution of each technique. SEM yields higher resolution and offers validation of results from SIM.
Keywords: Structured illumination microscopy, SIM, Scanning electron microscopy, SEM, surface shield enhancer, SSE, NanoSuit, correlative light and electron microscopy, CLEM, super-resolution
Introduction
Scanning electron microscopy (SEM) is a method used to obtain high-resolution images of microscopic samples by coating the sample in a thin layer of conductive material and scanning the surface with an electron beam, which generates an image. A high vacuum environment is necessary for SEM but simultaneously creates undesirable conditions for biological samples by drying, dehydrating, or otherwise distorting them [1,2]. To combat these unfavorable outcomes, samples must be specially prepared prior to imaging by SEM. One such chemical treatment is the NanoSuit [3,4], which offers a layer of protection against the high vacuum environment. A modification of the original NanoSuit method, called surface shield enhancer (SSE), consists of electrolytes and glycerol and protects the sample from moisture loss and other damage from the high vacuum conditions and electron beam without compromising the quality and resolution of SEM imaging [5].
Although SEM provides information regarding the structure and surface of a sample, important functional details and insights into dynamics can only be obtained from other techniques, such as fluorescence microscopy [6]. However, optical microscopy is unable to attain the same high-resolution images that SEM does. The cooperative use of electron and optical microscopy—correlative light and electron microscopy (CLEM) allows researchers to observe and relate structural and functional characteristics of biological samples by utilizing both SEM and optical microscopy on the same region of a biological sample [7]. The SSE NanoSuit is an ideal chemical preparation for samples that undergo CLEM, as it provides the conductive and protective layer for SEM, preserves the quality of the sample, and can be removed to restore slides to their original state for further optical microscopy [8,9].
The choice of a specific optical microscopy technique is central to the success of CLEM in research applications. Fluorescence microscopy allows for 3D imaging of the inside of cells and tissues that have been tagged with fluorescent labels to gain functional information in sample but is hindered by the diffraction limit, thus generating low-resolution images. To combat this, structured illumination micrscopy (SIM) [10–14], a fluorescence micrscopy technique that uses shifting illumination patterns to generate a collection of images, was designed to accomplish the acquisition of optically sectioned and/or super-resolution images. Maximum a posteriori (MAP)-SIM [15] is an evolution of standard SIM methods and is useful for imagng past the diffraction limit at incresed sample depths [10]. Used to process SIM data, SIMToolbox [16] is a MATLAB software that generates widefield (WF), optically sectioned SIM (OS-SIM) [14], and super-resolution MAP-SIM images from a single set of data.
These techniques were used to image a rat epididymis tissue sample with SEM, widefield fluorescence microscopy (WF), and MAP-SIM. The high-resolution structual data from SEM and the information derived from super-resolution fluorescence imaging via SIM can be combined to enhance the strengths of each technique while minimizing the weaknesses to get as much information as possible from one sample. This can be applied to live biological samples [17–20] with use of the NanoSuit method and has applications in biomedical research, histology, pathology, and potentially numerous other research fields that utilize high resolution imaging.
Methods
Surface shield enhancer (SSE) NanoSuit preparation:
The SSE solution was prepared by combining 5.00 g of each sucrose, fructose, and sodium chloride which were dissolved in 500 mL of deionized water. Additionally, 1.25 g of citric acid and 0.05 g of sodium glutamate were included in the solution and stirred until all solids dissolved. To create a 1:2 ratio for the final solution, 150 mL SSE was combined with 300 mL glycerol [1], the final pH was 7.4.
Slide preparation:
A few prepared slides were selected (Carolina Biological, item 31-6464, rat epididymis) and placed in toluene until the coverslip detached. The samples were rehydrated in 100, 90, and 70% ethanol, respectively, for five minutes each, and finally in deionized water for a further five minutes. The sample slides were then dipped in the SSE/glycerol solution for 10 minutes at room temperature [8]. The tissue samples were stained by the manufacturer with hematoxylin and eosin (H&E) and are brightly fluorescent.
SEM imaging:
One of the sample slides was coated with gold for 5 minutes using a sputter coating machine (Desk V, Denton Vacuum, New Jersey). Once the gold conductive layer was applied, the slide was cut (to reduce its size for the SEM chamber) with a diamond tip glass cutter, then placed into the SEM. The sample was then imaged at 30 kV using a Vega 3 SEM (Tescan, Brno, Czech Republic).
SIM imaging:
We used a home-built SIM setup based on the same design as described previously [10,11,15] (Fig. 1). The SIM system is based on a IX83 microscope (Olympus, Tokyo, Japan) equipped with a Zyla 4.2+ sCMOS camera (Andor, Belfast, Northern Ireland) under the control of IQ software (Andor). The following Olympus objectives were used: UPLSAPO 20×/0.85 NA oil immersion, and UPLSAPO 100×/1.4 NA oil immersion. The microscope is equipped with a filter set appropriate for TRITC (ET series, Chroma, Bellows Falls, VT) and a motorized X, Y, piezo Z stage (Applied Scientific Instrumentation, Eugene, OR). Figure 1 shows a simplified diagram of the SIM system.
Figure 1:
Simplified diagram of SIM system. LCOS, liquid crystal on silicon.
The SIM system uses a ferroelectric liquid crystal on silicon (LCOS) microdisplay (type SXGA-3DM, Forth Dimension Displays, Dalgety Bay, Fife, UK). This device has been used previously in SIM and fluorescence microscopy, [15,16,21–27] (among others) allowing one to produce patterns of illumination on the sample which can be manipulated by changing the image displayed on the microdisplay. The fiber-coupled light source (Lumencor Spectra-X, Beaverton, OR, USA) is switched off in between SIM patterns and during camera readout. Synchronization between the microdisplay, light source, and camera ensures rapid image acquisition, reduces artifacts, and decreases light exposure to the sample. The optical fiber was a 1.5 mm diameter multi-mode fiber (Thor Labs, Newton, NJ). The acquired SIM data was processed using SIMToolbox software [16], which generated widefield, OS-SIM, and MAP-SIM images. In some experiments (Figure 2) we used a BX53 microscope with a UPLXAPO 10×/0.40 NA objective and DP28 camera (Olympus). To evaluate the resolution of the acquired SIM images we calculated the power spectral density [28].
Figure 2:
Representative transmitted light and fluorescence images of a rat testis sample using a UPLXAPO 10×/0.4 NA objective and DP28 color camera.
Results
Both transmitted light and widefield fluorescence images of a rat testis sample stained with H&E were acquired using a DP28 color camera. This is shown in Figure 2 and offers an overview image of the type of sample used in this study.
A region of interest of a second rat testis sample was imaged with SEM (with SSE NanoSuit and gold sputtering treatment), conventional widefield fluorescence microscopy, and MAP-SIM to observe the different methods side-by-side. This is shown in Figure 3. WF and SIM imaging occurred first, followed by NanoSuit-SSE preparation and SEM.
Fig 3:
Correlative fluorescence and scanning electron microscopy of rat testis tissue stained with H&E. (A) conventional WF; (B) MAP-SIM; 20X/0.85 NA objective. (C) SEM after treatment with SSE and gold coating.
Figure 4 shows a comparison of WF and MAP-SIM acquired with a 100×/1.4 NA oil immersion objective and with SEM.
Fig 4:
Correlative widefield, super-resolution structured illumination, and scanning electron microscopy (A) WF, (B) MAP-SIM (both 100×/1.4 NA). (C) SEM, region of interest indicated by the red box in panel B. (D) MAP-SIM of the same region of interest.
Table 1 shows resolution measurements for the images in figure 4. The resolution of each image was measured by evaluation of the power spectral density [28] as applied by us previously [10,11,29].
Table 1:
resolution measurements of the images in Fig. 4.
| WF | SIM | SEM | |
|---|---|---|---|
| Image pixel size | 65 nm | 65 nm | 24 nm |
| Resolution measurement | 338 nm | 142 nm | 88 nm |
| Electron beam spot size (nm) | N/A | N/A | 88 nm |
Discussion
It has been difficult to obtain “ground truth” images when considering super-resolution microscopy. This is an important topic because both structured illumination microscopy and single molecule super-resolution microscopy (PALM/STORM type imaging [30–33]) rely on reconstruction algorithms to obtain the final image. These algorithms are under constant development, and their reliability and possible potential to introduce artifacts are of concern. Typically, conventional widefield images are shown in comparison to super-resolution images that were obtained by whichever method is being developed or applied. However, comparison of super-resolution microscopy images with SEM provides a higher resolution image for comparison and helps prove that the super-resolution optical microscopy image is accurate and free from artifacts. The results shown here indicate that super-resolution structured illumination microscopy, as implemented by us (MAP-SIM), is showing the correct sample structure and is free from apparent artifacts. Other SIM methods are more prone to pattern-related artifacts [34].
Other correlative microscopy techniques differ in procedure but serve a similar function in obtaining information about a sample from various methods. Another type of CLEM involves the use of laser scanning confocal microscopy (LSCM) followed by transmission electron microscopy (TEM) [35]. The images captured from TEM and LSCM can be overlaid into an integrated image [36], providing, for example, ultrastructural imaging of cells rapidly frozen at a moment in time at which a particular cellular function is taking place.
Conclusion
Procedures for correlative microscopy techniques have been in development for many years and are constantly being altered to maximize the benefits from their use. The utilization of super-resolution SIM with SEM is an especially promising form of CLEM due to the relative ease of the method and surprisingly good results as shown here.
Funding
Research reported in this publication was supported by the National Institute of General Medical Sciences of the National Institutes of Health under award number 2R15GM128166-02. This work was also supported by the UCCS BioFrontiers center. The funding sources had no involvement in study design; in the collection, analysis and interpretation of data; in the writing of the report; or in the decision to submit the article for publication.
Abbreviations
- H&E
hematoxylin and eosin
- MAP-SIM
maximum a posteriori probability SIM
- SSE
surface shield enhancer
- NA
numerical aperture
- LCOS
liquid crystal on silicon
- CLEM
correlative light and electron microscopy
- SEM
scanning electron microscopy
- SIM
structured illumination microscopy
- WF
wide field
References
- 1.Takaku Y., Suzuki H., Ohta I., Tsutsui T., Matsumoto H., Shimomura M., and Hariyama T., “A ‘NanoSuit’ surface shield successfully protects organisms in high vacuum: Observations on living organisms in an FE-SEM,” Proc. R. Soc. B Biol. Sci. 282, (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Ohta I., Takaku Y., Suzuki H., Ishii D., Muranaka Y., Shimomura M., and Hariyama T., “Dressing living organisms in a thin polymer membrane, the nanosuit, for high-vacuum FE-SEM observation,” Microscopy 63, 295–300 (2014). [DOI] [PubMed] [Google Scholar]
- 3.Takaku Y., Suzuki H., Ohta I., Ishii D., Muranaka Y., Shimomura M., and Hariyama T., “A thin polymer membrane, nano-suit, enhancing survival across the continuum between air and high vacuum,” Proc. Natl. Acad. Sci. U. S. A. 110, 7631–7635 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Takaku Y., “NanoSuit Method for the Observation of Living/Hydrous Organisms in an SEM,” Microsc. Microanal. 28, 1388–1388 (2022). [Google Scholar]
- 5.Takaku Y., Suzuki H., Kawasaki H., Ohta I., Ishii D., Hirakawa S., Tsutsui T., Matsumoto H., Takehara S., Nakane C., Sakaida K., Suzuki C., Muranaka Y., Kikuchi H., Konno H., Shimomura M., and Hariyama T., “A modified ‘NanoSuit’ preserves wet samples in high vacuum: Direct observations on cells and tissues in field-emission scanning electron microscopy,” R. Soc. Open Sci. 4, (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Liv N., Zonnevylle A. C., Narvaez A. C., Effting A. P. J., Voorneveld P. W., Lucas M. S., Hardwick J. C., Wepf R. A., Kruit P., and Hoogenboom J. P., “Simultaneous Correlative Scanning Electron and High-NA Fluorescence Microscopy,” PLoS One 8, e55707 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Begemann I., Viplav A., Rasch C., and Galic M., “Stochastic Micro-Pattern for Automated Correlative Fluorescence - Scanning Electron Microscopy,” Sci. Reports 2015 51 5, 1–12 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Kawasaki H., Itoh T., Takaku Y., Suzuki H., Kosugi I., Meguro S., Iwashita T., and Hariyama T., “The NanoSuit method: a novel histological approach for examining paraffin sections in a nondestructive manner by correlative light and electron microscopy,” Lab. Investig. 2019 1001 100, 161–173 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Shinmura K., Kawasaki H., Baba S., Ohta I., Kato H., Yasuda H., Yamada S., Misawa K., Sugimoto K., Osawa S., Sato M., Hariyama T., and Sugimura H., “Utility of Scanning Electron Microscopy Elemental Analysis Using the ‘NanoSuit’ Correlative Light and Electron Microscopy Method in the Diagnosis of Lanthanum Phosphate Deposition in the Esophagogastroduodenal Mucosa,” Diagnostics 2020, Vol. 10, Page 1 10, 1 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Paul T. C., Johnson K. A., and Hagen G. M., “Super-Resolution Imaging of Neuronal Structures with Structured Illumination Microscopy,” Bioengineering 10, 1081 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Johnson K. A. and Hagen G. M., “Artifact-free whole-slide imaging with structured illumination microscopy and Bayesian image reconstruction,” Gigascience 9, 1–14 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Gustafsson M. G. L., “Surpassing the lateral resolution limit by a factor of two using structured illumination microscopy,” J. Microsc. 198, 82–87 (2000). [DOI] [PubMed] [Google Scholar]
- 13.Heintzmann R. and Cremer C., “Laterally modulated excitation microscopy: improvement of resolution by using a diffraction grating,” Proc. SPIE 3568, 185–196 (1998). [Google Scholar]
- 14.Neil M. A. A., Juškaitis R., and Wilson T., “Method of obtaining optical sectioning by using structured light in a conventional microscope,” Opt. Lett. 22, 1905 (1997). [DOI] [PubMed] [Google Scholar]
- 15.Lukeš T., Křížek P., Švindrych Z., Benda J., Ovesný M., Fliegel K., Klíma M., and Hagen G. M., “Three-dimensional super-resolution structured illumination microscopy with maximum a posteriori probability image estimation,” Opt. Express 22, 29805–17 (2014). [DOI] [PubMed] [Google Scholar]
- 16.Křížek P., Lukeš T., Ovesný M., Fliegel K., and Hagen G. M., “SIMToolbox: A MATLAB toolbox for structured illumination fluorescence microscopy,” Bioinformatics 32, 318–320 (2015). [DOI] [PubMed] [Google Scholar]
- 17.Takehara S., Takaku Y., Suzuki H., Ohta I., Shimomura M., and Hariyama T., “A “NanoSuit” successfully protects petals of cherry blossoms in high vacuum: Examination of living plants in an FE-SEM,” Sci. Rep. 8, (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Fujita M., Blanke A., Nomura S., and Machida R., “Simple, Artifact-free SEM Observations of Insect Embryos: Application of the Nano-suit Method to Insect Embryology,” Proc Arthropod. Embryol. Soc. Jpn 50, 7–10 (2016). [Google Scholar]
- 19.Takaku Y., Takehara S., Suzuki C., Suzuki H., Shimomura M., and Hariyama T., “In situ elemental analyses of living biological specimens using ‘NanoSuit’ and EDS methods in FE-SEM,” Sci. Rep. 10, (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Kikuchi H., Matsumoto T., Hiraide T., Ozaki Y., A. H.-C. Research, and undefined 2017, “Direct observation of colorectal cancers using field-emission scanning electron microscopy with a thin polymer membrane, the NanoSuit,” AACR (n.d.). [Google Scholar]
- 21.Shao L., Kner P., Rego E. H., and Gustafsson M. G. L., “Super-resolution 3D microscopy of live whole cells using structured illumination,” Nat. Methods 8, 1044–1048 (2011). [DOI] [PubMed] [Google Scholar]
- 22.Hagen G. M., Caarls W., Lidke K. A., De Vries A. H. B., Fritsch C., Barisas B. G., Arndt-Jovin D. J., and Jovin T. M., “Fluorescence recovery after photobleaching and photoconversion in multiple arbitrary regions of interest using a programmable array microscope,” Microsc. Res. Tech. 72, 431–440 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Kantelhardt S. R., Caarls W., de Vries A. H. B., Hagen G. M., Jovin T. M., Schulz-Schaeffer W., Rohde V., Giese A., and Arndt-Jovin D. J., “Specific visualization of glioma cells in living low-grade tumor tissue,” PLoS One 5, 1–11 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Fiolka R., Shao L., Rego E. H., Davidson M. W., and Gustafsson M. G. L., “Time-lapse two-color 3D imaging of live cells with doubled resolution using structured illumination,” Proc. Natl. Acad. Sci. U. S. A. 109, 5311–5315 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Chen B. C., Legant W. R., Wang K., Shao L., Milkie D. E., Davidson M. W., Janetopoulos C., Wu X. S., Hammer J. A., Liu Z., English B. P., Mimori-Kiyosue Y., Romero D. P., Ritter A. T., Lippincott-Schwartz J., Fritz-Laylin L., Mullins R. D., Mitchell D. M., Bembenek J. N., Reymann A. C., Böhme R., Grill S. W., Wang J. T., Seydoux G., Tulu U. S., Kiehart D. P., and Betzig E., “Lattice light-sheet microscopy: Imaging molecules to embryos at high spatiotemporal resolution,” Science 346, (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Hagen G. M., Caarls W., Thomas M., Hill A., Lidke K. A., Rieger B., Fritsch C., van Geest B., Jovin T. M., and Arndt-Jovin D. J., “Biological applications of an LCoS-based programmable array microscope (PAM),” Imaging, Manip. Anal. Biomol. Cells, Tissues V 6441, 64410S (2007). [Google Scholar]
- 27.Křížek P., Raška I., and Hagen G. M., “Flexible structured illumination microscope with a programmable illumination array,” Opt. Express 20, 24585 (2012). [DOI] [PubMed] [Google Scholar]
- 28.Pospíšil J., Fliegel K., and Klíma M., “Assessing resolution in live cell structured illumination microscopy,” in Proceedings of SPIE - The International Society for Optical Engineering, Páta P. and Fliegel K., eds. (SPIE, 2017), Vol. 10603, p. 39. [Google Scholar]
- 29.Pospíšil J., Lukeš T., Bendesky J., Fliegel K., Spendier K., and Hagen G. M., “Imaging tissues and cells beyond the diffraction limit with structured illumination microscopy and Bayesian image reconstruction,” Gigascience 8, 1–12 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Ovesný M., Křížek P., Borkovec J., Švindrych Z., and Hagen G. M., “ThunderSTORM: A comprehensive ImageJ plug-in for PALM and STORM data analysis and super-resolution imaging,” Bioinformatics 30, 2389–2390 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Shtengel G., Wang Y., Zhang Z., Goh W. I., Hess H. F., and Kanchanawong P., “Imaging cellular ultrastructure by PALM, iPALM, and correlative iPALM-EM,” Methods Cell Biol. 123, 273–294 (2014). [DOI] [PubMed] [Google Scholar]
- 32.Rust M. J., Bates M., and Zhuang X., “Sub-diffraction-limit imaging by stochastic optical reconstruction microscopy (STORM),” Nat. Methods 3, 793–795 (2006). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Betzig E., Patterson G. H., Sougrat R., Lindwasser O. W., Olenych S., Bonifacino J. S., Davidson M. W., Lippincott-Schwartz J., and Hess H. F., “Imaging intracellular fluorescent proteins at nanometer resolution,” Science 313, 1642–1645 (2006). [DOI] [PubMed] [Google Scholar]
- 34.Pospíšil J., Fliegel K., and Klíma M., “Analysis of image reconstruction artifacts in structured illumination microscopy,” in Applications of Digital Image Processing XL, Tescher A. G., ed. (SPIE, 2017), Vol. 10396, pp. 1–12. [Google Scholar]
- 35.Verkade P. and Kingdom U., “Moving EM: the Rapid Transfer System as a new tool for correlative light and electron microscopy and high throughput for high-pressure freezing,” J. Microsc. 230, 317–328 (2008). [DOI] [PubMed] [Google Scholar]
- 36.Keene D. R., Tufa S. F., Wong M. H., Smith N. R., Sakai L. Y., and Horton W. A., “Correlation of the Same Fields Imaged in the TEM, Confocal, LM, and MicroCT by Image Registration: From Specimen Preparation to Displaying a Final Composite Image,” Methods Cell Biol. 124, 391–417 (2014). [DOI] [PubMed] [Google Scholar]