Abstract
Major challenges remain to precisely detect low-abundance proteins rapidly and cost-effectively from diverse biofluids. Here we present a gold nanoparticle (AuNP)-supported, rapid electronic detection (NasRED) platform with sub-femtomolar sensitivity and high specificity. Surface-functionalized AuNPs act as multivalent detectors to recognize target antigens and antibodies through high-affinity binding, subsequently forming aggregates precipitated in a microcentrifuge tube and producing a solution color change. The residual floating AuNPs’ optical extinction is digitized using customized circuitry incorporating inexpensive optoelectronics and feedback mechanisms for stabilized readout. NasRED introduces active fluidic forces through engineered centrifugation and vortex agitation, effectively promoting low-concentration protein detection and accelerating signal transduction. Using SARS-CoV-2 as a demonstration, NasRED enables detection of both antibodies and antigens from a small sample volume , distinguishes the viral antigens from those of human coronaviruses, and delivers test results in <15 min. The limits of detection (LoDs) for antibody detection are approximately 49 aM (7 fg/mL) in phosphate-buffered saline (PBS), or >3,000 times more sensitive than Enzyme-Linked Immunosorbent Assay (ELISA), ~76 aM (11 fg/mL) in human pooled serum and in the femtomolar range in diluted whole blood. For nucleocapsid protein detection, NasRED LoDs are ~190 aM (10 fg/mL) in human saliva and ~2 fM (100 fg/mL) in nasal fluid. Unlike centralized platforms, NasRED is a one-pot, in-solution assay without the needs for washing, labeling, expensive instrumentation or highly trained operators. With low reagent costs and a compact system footprint, this modular digital platform is well-suited for accurate, near-patient diagnosis and screening of a wide range of infectious and chronic diseases.
Keywords: SARS-CoV-2, Gold nanoparticles, Point of care, Antibody detection, Rapid antigen detection, Sub-femtomolar sensitivity
INTRODUCTION
Infectious diseases pose a serious threat to human health and the world economy. For example, as of March 1st, 2025, COVID-19 alone has resulted in more than 777 million infections and 7 million deaths,1 while causing an estimated $14 trillion in economic losses in the U.S. by the end of 2023.2 In addition, the pandemic also seriously strained the global healthcare system3,4 and caused delayed or missed diagnosis or treatment for many other patients. To better prepare for future pandemics, which could be caused by a currently unknown disease ‘X’,5,6 it is of paramount importance to continue investing in innovative diagnostic technologies. These technologies should enable early and accurate disease detection and quantification in a portable, cost-effective, and user-friendly manner, ensuring accessibility for all patients in need, including those in rural areas and low- to middle-income countries where medical resources are scarce. To complement this, various new approaches, including label-free high-sensitivity diagnostic technologies, have been proposed recently.7,8
Conventionally, molecular and immunochemical tests are widely used to diagnose infectious diseases, including COVID-19.9,10 These tests look for molecules of different genetic information to diagnose diseases and predict the risk of unknown diseases.11 Nucleic Acid Amplification Tests (NAATs), including Polymerase Chain Reaction (PCR) and Loop-mediated Isothermal Amplification (LAMP), leverage the feasibility of DNA or RNA amplification to maximize the sensitivity. However, as the gold standard for laboratory diagnostics, PCR requires several hours to days to amplify and confirm the results due to the multiple steps of sample handling and the thermal cycling process.12 Additionally, it requires expensive, bulky equipment, making it suitable for centralized testing but inaccessible in resource-limited settings. Isothermal amplification methods, such as LAMP, while offering a simpler and more cost-effective alternative to PCR, face challenges in practical applications, including the complexity in primer design, the limited stability of enzymatic reagents, and the need of careful sample preparation when working with complex biological fluids such as blood and saliva.
Immunochemical tests diagnose diseases by confirming the presence or absence of a specific protein based on antigen-antibody reactions. The most commonly used methods include Enzyme-Linked Immunosorbent Assay (ELISA) and Lateral Flow Assay (LFA).13 ELISA typically achieves picomolar-level sensitivity in biological fluids.14–16 However, like PCR, ELISA is most suited for high-sensitivity analysis of large amounts of samples in laboratory settings since it requires a complex, multistep workflow, making it highly labor-intensive and expensive.17,18 In addition, Rapid Antigen Tests (RATs), such as LFA, are portable and suitable for field applications but generally have lower sensitivity (e.g., 65.3% for COVID-19). This limitation makes them less reliable for accurate detection, particularly at early-stage infection when the viral load is low.19–21 Here, using SARS-CoV-2 as an example, we report a prototyped, inexpensive, miniaturized sensing system, i.e., nanoparticle-supported rapid electronic detection (NasRED), to detect the antibodies and antigens directly from serum, whole blood, saliva, and nasal fluid within a short time (15 to 30 min, comparable to LFA). NasRED achieves femtomolar or better sensitivity (or fg/mL in mass concentration), about 2 to 4 orders of magnitude better than ELISA22 and about 5 to 6 orders better than LFA,23 enabling rapid and accurate diagnostics of disease biomarkers in different clinical settings. Specifically, our digital sensing platform was designed to detect the target proteins by simply mixing bodily fluids with our prepared sensing solution comprising functionalized gold nanoparticles (AuNPs) in a testing tube.24,25 The entire process, including mixing, reaction, and signal readout, occurs within the same tube without washing, labeling, or fluorescent imaging, significantly reducing operation complexity. The sensing signals were collected, digitized, and transmitted through our customized semiconductor devices and circuits, which minimized signal fluctuations and enable the accurate detection of antigens and antibodies at ultralow concentrations. Further, we comprehensively engineered the sensing protocol, i.e. the centrifugation and vortex agitation speed and time, to evaluate their impact on the limits of detection (LoDs) of SARS-CoV-2 antibody (AS35) and antigen (N-protein). The LoDs were found <100 fg/mL from a wide range of bodily fluids, including in diluted whole blood without requiring plasma separation. This high sensitivity is attributed to the application of active fluidic forces during centrifugation and vortex agitation, which enhance protein collisions, biochemical binding, and signal transduction, even at low concentrations. Additionally, vortex agitation creates an AuNP concentration gradient along the liquid column, thus enhancing optical signals and enabling detection of low-abundance proteins. Lastly, using inactivated SARS-CoV-2 virion particles, NasRED successfully detected N-protein from lysate products in saliva with high sensitivity (190 aM or 10 fg/mL), outperforming ELISA. Based on parallel qRT-PCR measurements, this protein detection level corresponds to an estimated RNA sensitivity of 3×105 copies/mL, demonstrating performance comparable to portable nucleic acid amplification tests. Together, our preliminary data demonstrates the strong promise of NasRED as a new and broadly applicable platform for highly sensitive, rapid, and accessible detection of infectious diseases as well as chronic diseases.
RESULTS
NasRED sensing platform
The NasRED protein sensing platform can be applied to the detection of both the antigens and the antibodies (Figure 1). The sensors are prepared by attaching biotinylated proteins to streptavidin-coated AuNPs through a biotin-streptavidin reaction, where the protein binders can be antigens (e.g. SARS-CoV-2 spike (S) or receptor binding domain (RBD) proteins) for antibody (neutralizing antibody AS35) sensing or antibodies (mouse IgG1 AS47 and chimeric IgG1 AM223) for antigen (nucleocapsid (N) protein) sensing (Figure 1a). Using antibody sensing as an example (Figure 1b), of the RBD-functionalized AuNPs were mixed with of medium (e.g., phosphate-buffered saline (PBS) buffer, human pooled serum (HPS), or human whole blood (WB)) containing target SARS-CoV-2 antibodies diluted to desired concentrations. A sample of blank medium without antibodies was used as a negative control (NC). For rapid detection, the mixture was centrifuged at 1,200 g for 5 minutes and incubated for 5 minutes to precipitate the functionalized AuNPs to the bottom of the microcentrifuge tube, where their concentration was greatly enhanced. Such a localized and boosted reactant concentration significantly promoted the localized reaction with target proteins, thus accelerating the antibody-antigen reaction at low concentrations to achieve an ideal LoD and reduce assay time. To preserve the detection specificity, a vortex agitation was introduced to disperse unbound AuNPs without disturbing the precipitated AuNP clusters, verified by fully dispersed AuNPs in the NC reference (Figure 1c). The optical absorption signal from the supernatant, attributed to free-floating AuNPs after vortexing, was digitized using a portable electronic detector (PED) for quantification (Figure 1d–e). The electronic signals of the antigen or antibody samples collected from the PED represented the optical transmission of light-emitting diode (LED) light through the supernatant in the microcentrifuge tube and were normalized (Methods). The NasRED system has proven to be feasible to detect antigens and antibodies in complex biological fluids with high sensitivity and specificity (Figure 1f–g).
Figure 1.

Nanoparticle-supported, rapid electronic detection (NasRED) of antigens and antibodies for infectious diseases diagnostics. (a) Nanosensor preparation by biotinylating antigens or antibodies and attaching such proteins to streptavidin-coated AuNPs. (b-d) Schematics and optical images illustrating the detection process, using antibody testing as an example. (b) The AuNP sensing solution was mixed with the medium solution . Negative controls (NCs) did not contain any target molecules but otherwise contained the same biological fluids and AuNP sensors. (c) The AuNPs were precipitated by centrifugation, incubated, and resuspended after stirring vortex. (d) The readout was collected directly from the tube, which was inserted in a tube holder as part of the NasRED readout system. The amount of AuNP precipitation at the tube bottom correlated with the number of test molecules, while no AuNP precipitates formed in the NC sample. (e) Schematic of a typical sensing curve showing the normalized PED readout signal . (f) Exemplary optical images and bar plots showing sensitivity and specificity in SARS-CoV-2 antigen detection in PBS. Tubes containing N-proteins of two other coronaviruses (HCoV-NL63 and HCoV-229E at 400 nM in PBS buffer) displayed very similar colors and electronic signals comparable to that of the NC sample. In contrast, tubes containing SARS-CoV-2 N-proteins showed much higher transparency and differentiable electronic signals at 4 fM and 400 nM. (g) Exemplary optical images and sensing curve plots showing antibody detection at high sensitivity in diluted whole blood.
Signal-stabilizing circuit for low-noise detection
The NasRED platform consists of a portable electronic detector (PED) for signal readout, which is composed of two customized components (Figure S1a). The first component is a semiconductor-based, digital, signal-collecting device, which includes an LED, a photodetector (PD), and a microcentrifuge tube chamber designed to block ambient light. This device functions to probe the free-floating AuNP concentration in the supernatant of the testing tube and to digitize the optical signals of the AuNPs to electronic current signals. The second component is a circuit board designed to stabilize the LED and PD signals for reliable antibody or antigen detection. Electronically, the stabilizing circuit board (Figure 2a–b) had three key components: a voltage regulator that adjusts the power supply to the desired value for the LED circuit and microprocessor, a constant current LED driver circuit that stabilizes the light intensity, and a microprocessor for signal processing. The regulator circuit includes two electrolytic capacitors, two ceramic capacitors, and one protective Zener diode (Figure 2a (i)). The constant current LED driver circuit utilizes a feedback loop that coordinates double bipolar junction transistors (BJTs, T1 and T2 in Figure 2a (ii)) to stabilize the current supplied to the LED. In addition, a customized microprocessor (Figure 2a (iii)) processes the signals and enables communication with a laptop through USB, Wi-Fi, or Bluetooth modules. Without the stabilization circuit, the LED displayed a large deviation in current (average 0.14 mA over 10 minutes but significantly increased after the 40-min operation, possibly due to heating, Figure 2c), and the PED produced random spike noises when used to test AS35 antibodies in PBS (Figure 2d). This instability can be attributed to the fact the LED light emission is proportional to its electrical current and exponentially depends on the bias voltage, and therefore, a slight voltage fluctuation from a power glitch or device heating can introduce a systematic error, ultimately seriously limiting the sensing performance at low analyte concentrations. In comparison, our improved readout system demonstrated significantly reduced current fluctuation (0.12 mA within the first two minutes during circuit initiation and decreased to 0.02 mA during the rest 60-min testing period, Figure 2c) and greatly minimized system error in antibody detection (Figure 2d), indicating a substantial improvement in stability.
Figure 2.

Electronic circuit design for stabilized NasRED readout. (a) Schematic showing an integrated printed circuit board (PCB), which includes: (i) a built-in voltage regulator, (ii) a constant current LED driver circuit, and (iii) a microcontroller, which processes the signals and communicates with computers through USB, Wi-Fi, or Bluetooth modules. (b) Optical image showing the PCB board for signal processing. (c) The stability of LED currents over time for a circuit without the stabilization circuit (black line) and our integrated PCB board with the stabilization circuit (red line). (d) PED readout signal plotted against antibody AS35 concentrations spiked in PBS, with (red) and without (black) the stabilization circuit. (e) LED irradiance recorded for 60 minutes using four different devices #1 (black line), #2 (red line), #3 (green line), and #4 (blue line). (f) Three, each three-hour-long, replicate measurements of LED irradiance from one PED reader.
To further evaluate the reproducibility and reliability of the PED circuit design, four different PED devices were constructed (Figure S2). The LED output power from each PED circuit was measured by a PD and found stable at over a 60-minute test period (Figure 2e) with a very small (~2.5%) coefficient of variation (CV). Then, one PED circuit was randomly selected, and its LED power was measured three times, 3 hours each (Figure 2f). The measured irradiance was again very stable at with a CV of 2.4%. These inter- and intra-device tests proved the high reproducibility and suitability of our NasRED reading system for continuous use or repeated operation.
Impacts of centrifugation and vortex agitation on sensing performance
The centrifugation and vortex agitation actively modulate the reagent and AuNP distribution in the buffer solutions, providing additional flexibility to optimize the sensitivity and specificity. To examine their impact on the NasRED assay performance, we chose mAb AS35 (, from to 40 fM in PBS) as the target antibody, using AuNPs functionalized with WT-RBD (Wuhan-Hu-1, in1×PBS dilution buffer). Electronic signals were collected using our stabilized readout system. Here, a series of tests (Figure 3a–d) were performed by modifying four variables: the centrifugate force gravity (g) (Figure 3a), centrifugation time (Figure 3b), vortex agitation speed (Figure 3c), and the vortex time (Figure 3d). The normalized assay signals, , were plotted as a function of AS35 concentration (Figure 3a–d). The default parameters were centrifugation at 1,200 g for 5 minutes, incubation for 5 minutes, and vortex agitation at 34.5 rps for 5 seconds unless otherwise varied. It was clear that the AS35 antibodies could be detected with a large contrast in signal , i.e. a high value at low concentrations and a small signal at high concentrations , or . A high signal contrast indicated that the AuNP sensors effectively participated in the biochemical reaction at high protein concentrations and subsequently precipitated, which was ideal for maximizing the sensitivity and dynamic range (about 9 logs in this test). Clearly, the sensing performance of NasRED could be optimized within a range of centrifugation conditions, i.e., from 900 g to 1500 g and from 3 to 5 minutes, and at moderate vortex conditions.
Figure 3.

Optimization of NasRED sensing protocol under different centrifugation and vortex conditions. (a-d) The PED readout signal plotted as a function of the antibody concentration, with: (a) the centrifugate force from 0 to 1,500 gravity (g), (b) the centrifugation time from 0 to 5 minutes, (c) the vortex agitation speed from 32 to 40 round per second (rps), and (d) the vortex time from 3 to 15 seconds. (e-f) Schematic representations of the AuNP cluster formation and precipitation mechanism during the sensing process: (1) uniform mixing of AuNPs with target proteins, (2) AuNP sensors react to and bind to target proteins, thus forming clusters, during centrifugation, (3) after centrifugation and during incubation, the AuNP clusters form larger precipitates at the tube bottom, and (4) vortex agitation releasing loose AuNPs back into the solution. (g) Representative Cryo-TEM images of 80 nm AuNP clusters at AS35 concentrations of 40 pM, 4 pM, 400 fM, 40 fM, 40 aM, and NC. (h) Statistical analysis of AuNP cluster sizes formed at different AS35 concentrations. NC: negative control using blank sample. (i–k) Modeling and analysis of vertical AuNP concentration distribution. (i) Schematic of the optical detection setup with dual-position PED readout, with green arrows indicating upper (5.5 mm) and lower (3.5 mm) sensing locations and black arrow indicating the z direction in the mathematical model (z=0 at the tube bottom). (j) Reconstructed AuNP concentration profiles using an exponential decay model in the z direction, fitted to the measured concentrations at both PED readout positions (highlighted by green). The profiles were generated using Python under a mass conservation constraint. (k) Extracted model parameters and as a function of analyte concentration. refers to the AuNP monomer concentrations at z=0 and is the characteristic decay length of AuNP concentrations in the model.
NasRED sensing mechanism
The dependence of NasRED’s performance on the centrifugation and vortex parameters can be understood from the unique sensing mechanism. First, fundamentally NasRED is an in-solution assay, where diffusion of AuNP sensors and reagents drives signal transduction, leading to AuNP precipitation and ultimately determining the assay performance. In NasRED, the sedimentation time of AuNPs linearly scales with the precipitation length.26 Unlike conventional assays such as ELISA, which rely on the passive diffusion of target molecules to a stationary reaction surface and require long incubation time to maximize the signal-to-noise ratio, NasRED supports external controls, such as centrifugation, incubation, and vortex agitation, to actively accelerate diffusion, reaction, and readout (Figures 3 and S3). Centrifugation pellets AuNPs dispersed in the tube to the bottom, with a thickness estimated as <0.5 mm (supplementary section 4), thus eliminating slow sedimentation and accordingly reducing the readout time to less than 30 minutes (Figure 3a–b). In comparison, passive precipitation of AuNP clusters was previously found to require 3 to 24 hours of assay time from mixing to readout24 (Figure S4a–c). Additionally, the PED signal was found stable for hours after vortex agitation (Figure S4 d–e), indicating the establishment of a quasi-equilibrium state desirable for reliable readout.
Further, the introduction of active fluidic forces is an integral part of NasRED’s unique signal transduction process, which begins with biochemical protein reaction and is dynamically modulated by fluidic interactions (Figure 3e–f). Centrifugation significantly increases the kinetic velocity of AuNPs, enhancing their collision with and effective capture of the target proteins, which in turn promotes the growth of AuNP clusters. Protein binding and AuNP cluster growth are expected to occur throughout the centrifugation process, ultimately leading to precipitation of the majority of the AuNPs (Figure 3e). The dimensionless Peclet number (Pe), which compares nanoparticle centrifugal sedimentation to Brownian motion, scales linearly with the relative centrifugal force (RCF) and quantically with the AuNP diameter (supplementary section 4). For example, the Pe number was found ~1.92 for 80 nm AuNPs at an RCF of 1200×g, while significantly lower values were obtained for smaller nanoparticles (~0.01, 0.12, and 0.61 for 20 nm, 40 nm, and 60 nm AuNPs, respectively). The estimates suggest that the centrifugal forces dominate over Brownian motion during the sedimentation process of 80 nm and larger AuNPs, aligning with our prior observation that 80 nm AuNPs precipitated faster than smaller ones.24 On the other hand, the particle Reynolds number , which quantifies the ratio of inertial to viscous forces, increases proportionally with the AuNP diameter. was estimated ~0.04 for 80 nm AuNPs, suggesting that fluidic drag forces remain a significant factor in governing the particle motion and dispersion within biological fluids. Consequently, the NasRED sensing protocol should be individually optimized for fluid media with different viscosities.
The AuNPs are expected to capture target proteins while following distinct sedimentation paths, eventually forming precipitates composed of heterogeneous clusters, whose size distributions statistically correlate with protein concentrations (Figure 3f left). Importantly, centrifugation greatly localizes AuNPs and their clusters into a much smaller volume (estimated <1 nL, or 3 to 4 orders more condensed, supplementary section 4) and drastically boost the AuNP concentration at the bottom of the reaction tube. This effectively modulates the dynamic equilibrium of antigen-antibody reaction to favor AuNP cluster formation both during centrifugation and incubation, even at ultralow reagent concentration in the tube.24 During incubation, AuNPs deposited on the tube sidewalls gradually settle to the tube bottom, thus minimizing the nonspecific particle interference with the LED-PD signal-collecting optical path in the upper solution. Additionally, AuNPs may possibly rearrange during incubation, forming more stable clusters that enhance sensitivity.
After incubation, sufficient vortex agitation is required to fully resuspend monomer AuNPs that have not incorporated into clusters, ensuring high specificity, i.e., the negative control should return to their initial states. However, excessively high agitation risks breaking the AuNP clusters biochemically “glued” together, even at moderate or high protein concentrations. This would diminish the signal contrast (Figure 3c–d arrows on the right) and negatively affect the NasRED sensitivity. Therefore, optimizing the agitation force and duration (e.g. 5 sec at 34.5 rps for the chosen centrifugation conditions) is necessary to maximize the signal contrast and to best differentiate samples at low concentrations, for example from attomolar to picomolar for AS35 (Figure 3e optical images). Interestingly, the 40 aM and 4 pM samples could be clearly differentiated from their precipitate formation in the tubes, indicating feasible attomolar detection. For more precise analysis, the AuNP precipitates were collected from the bottom of the tubes and imaged by Cryogenic Electron Microscopy (Cryo-EM) (Figure 3g and Figure S5). Then, the AuNP clusters at each AS35 concentration were measured and analyzed (Figure 3h). Evidently, large clusters (e.g. , up to ) formed at moderate AS35 concentrations (e.g., 40 pM, 4 pM, and 400 fM) consisted of densely packed and stacked AuNPs. In comparison, smaller clusters were found at low antibody concentrations ( at 4 fM and 40 aM), consisting of loosely connected AuNPs. Notably, AuNP clusters of were only formed in the presence of AS35, whereas very small AuNP clusters or oligomers were detected from the NC (blank) sample, likely resulting from minor nonspecific inter-AuNP interactions during sample preparation (such as pipetting, TEM grid loading, and freezing). These Cryo-EM imaging provided direct physical evidence to confirm that the NasRED sensing protocol effectively dispersed the AuNPs for blank sample while differentiating antibodies at attomolar levels from higher-concentration and blank samples.
The fluidic dynamic behavior of AuNPs is complex in a turbulent flow during vortex agitation, potentially involving both shear- and spin-related fluidic drag forces,27 and is strongly dependent on the value. Both shear- and spin-induced lift forces may diminish at higher values, and are highly dependent on the AuNP size, shape, flow velocity, rotation, and position within the fluid. As such, it is reasonable to expect that AuNP clusters, which exhibit a substantially higher effective , experience reduced lift forces compared to AuNP monomers with smaller . Therefore, AuNP clusters are far less likely to move upwards during vortex agitation. Precise physical modeling of these behaviors is challenging due to the complex fluidic dynamics, the protein-concentration-dependent, inhomogeneous mixture of AuNP monomers and clusters in the precipitates, and the three-dimensional conical geometry of the microcentrifuge tubes. Therefore, a detailed quantitative model will require extensive future studies and thus beyond the scope of this work.
To provide an intuitive understanding of the signal transduction mechanisms, we first established a correlation between the PED signal and AuNP concentration, by cross-validating results using the manufacturer-provided data, NanoSight calibration (Figure S6), and AuNP extinction calculations based on Beer–Lambert law (supplementary section 4). Then we built a PED device capable of measuring the AuNP extinction signals at two different fluidic levels, and converted the measured two-channel signals into the AuNP concentrations at the two positions, for each tested AS35 antibody concentration (Figure 3i and Figure S7). We hypothesize that the AuNP concentration profile after vortex agitation follows a simplified exponential decay, consistent with the Mason-Weaver theory, and assume mass conservation of AuNPs. To test this hypothesis, we modeled the AuNP concentration distribution (Figure 3j) in Python26 as: . Here and are empirical, analyte-concentration-dependent fitting variables (Figure 3k) that characterize the AuNP concentration at tube bottom (z=0) and characteristic decay length, respectively. The PED-measured and theoretically calculated values showed strong overall agreement (<20% error, Figure 3j). This physics-based model supports our hypothesis that the AuNPs tend to aggregate more at the tube bottom and diffuse back less to the suspension at higher , also in agreement with our Cryo-EM analyses.
The optical tube images, PED readout signals, TEM analyses, and AuNP concentration modelling collectively demonstrated the success of attomolar detection from a small sample volume . While further investigations are needed to fully elucidate the mechanisms behind the high-sensitivity of NasRED detection, a few factors possibly contribute. First, a strong antigen-antibody binding was achieved through multiple mechanism: (1) Multivalent binding on the AuNP sensor surfaces promotes high avidity, which has been demonstrated in pentameric IgM and nanobody oligomers to enhance the effective affinity by three to four orders of magnitude.24,28 (2) Protein capture is enhanced during centrifugation, and their disassociation from AuNPs is suppressed. (3) The reagent concentration at the tube bottom was enhanced about 4 orders of magnitude (supplementary section 4.3). Second, unlike fluorescent imaging that suffers from a high background noise, the PED readout has minimized background noise owing to successful strategic assay system design: (i) The stabilized PED readout device successfully suppresses signal fluctuations. (ii) NasRED collects the subtractive signals from a large population of AuNPs, thus minimizing stochastic background noises. (iii) Nonspecific interparticle interactions and tube wall adsorption are minimized by using inert AuNPs with covalently coated streptavidin as the signal carrier and by introducing glycerol and BSA in the AuNP sensor solution. (iv) A post-centrifugation incubation step allows the AuNP precipitates to settle at the tube bottom, thereby reducing unintentional interference to signal collection from free-floating AuNP monomers. Third, AuNP clusters bound by proteins may recruit non-protein-binding AuNP monomers through physical trapping, thus enlarging the AuNP precipitates and enhancing the PED signals. This could occur through disruption of the fluidic flow during centrifugation and/or vortex agitation, or by long-range attractive forces such as electrostatic interactions. TEM images (Figure S5) revealed inter-particle distance (e.g., >40 nm) far exceeding typical protein dimensions between AuNPs at the perimeters of clusters, suggesting possible long-range electrostatic interactions between protein analyte and protein-coated AuNPs.29 Lastly, larger AuNP clusters (Figure 3h) not only better resist fluidic lift force than AuNP monomers during vortex agitation, but also influence the resuspension of smaller AuNPs. Consequently, the analyte concentration , which determines the size and amount of AuNP precipitates, ultimately plays an important role in the redistribution of AuNPs. The significant differences in the fitting parameters and (Figure 3j) between sub-femtomolar analyte concentrations and blank samples resulted in a detectable change in AuNP concentration (about 4 × 109). This change exceeded the minimum PED-resolvable concentration difference in (3 × 108), thus enabling sub-femtomolar protein detection on the NasRED assay.26
NasRED comparison to ELISA
To benchmark the sensitivity of NasRED, the AS35 antibody was spiked in PBS at concentration range from 40 aM to 400 nM and analyzed using identical reagents on both standard ELISA assay (Figure 4a–b, Figure S8, and Table S2) and NasRED (Figures 4b and Figure S9) platforms. The LoDs, following two slightly different definitions, and LoQs were determined as described in the Method section and summarized in Table 1. For consistency, the LoDs reported in this work refer to the protein concentration at which the PED signal on the fitted sensing curve is distinguishable from the blank (NC) sample by three standard deviations of the NC measurements, i.e. . Here intra-assay imprecision (or within-test error, ) represents the systematic readout errors from the PED device at each . In contrast, inter-assay error (or between-test variation, ) captures the variability across replicate experiments and is used to assess assay reproducibility.30 For NasRED assay LoD reporting, was determined from replicate blank samples (for example 20 samples in PBS).
Figure 4.

ELISA and NasRED detection of mAb AS35 in PBS and biological fluids. (a) Schematics of ELISA assay, with antibody concentrations ranging from 400 nM to 40 aM. NC: PBS only. (b) The PED readout signal (black line) plotting the average of the 6 replicate experiments in PBS, with a LoD of 49 aM (0.007 pg/mL). The ELISA absorbance (red line), averaged from the 3 replicate experiments in PBS, were plotted against antibody concentration, showing a LoD of ~155 fM (23.3 pg/mL). ELISA values were normalized between 0 and 1 for direct comparison with . Error bars indicated the standard deviation (SD) across replicates for both experiments. (c-d) Comparison of the ELISA and NasRED sensing performance of AS35 in PBS. The CV analysis showed a larger variation (as large as 52%) in ELISA absorbance and a smaller variation (<10%) in PED signals across the entire tested concentration range. The recovery of ELISA was out of the desired range (100±30%) at moderate and low concentrations (4 pM - 40 aM), while NasRED produced stable recovery signals. (e-f) Optical images showing samples ready for NasRED readout: (e) in human pooled serum and (f) in 1% diluted human whole blood. Upon mixing with the AuNP sensing solution, the HPS sample was 25% diluted, and the WB was 0.25% diluted. (g) Extracted PED signals plotted against antibody concentration in serum (blue line, LoD=76 aM, or 0.011 pg/mL), and whole blood (green line, LoD=360 aM, or 0.054 pg/mL). The samples were analyzed using a standard testing condition, i.e., centrifugation at 1,200 g for 5 minutes, incubation for 5 minutes, and vortex agitation at 34.5 rps for 5 seconds.
Table 1.
NasRED design and performance in detecting SARS-CoV-2 antigens and antibodies
| Target Molecules | Medium | Protein binders on AuNPs | Centrifuge Incubation, Replicates # | LoD (1.645) | LoD ## | LoQ ## | LoQ/LoD ratio | |||
|---|---|---|---|---|---|---|---|---|---|---|
| aM | pg/mL | aM | pg/mL | fM | pg/mL | |||||
| Antibody AS35 | PBS | WT-RBD | C12, I5** | 51 | 0.008 | 49 | 0.007 | 0.7 | 0.11 | 14.29 |
| HPS | WT-RBD | C12, I5** | 90 | 0.014 | 76 | 0.011 | 0.86 | 0.129 | 11.32 | |
| 1% WB | WT-RBD | C15, I20** | 570 | 0.086 | 360 | 0.054 | 33 | 5 | 91.67 | |
| 20% WB | WT-RBD | C12, I20* | 7×104 | 10 | 17×104 | 25.5 | 14×104 | 21×103 | 823.53 | |
| WT-RBD | C15, I20** | 223 | 0.034 | 154 | 0.023 | 4×103 | 600 | 25974 | ||
| SARS-CoV-2 N-protein | PBS | AS47 | C12, I20* | 4 ×109 | 2×105 | 4×109 | 2×105 | - | - | - |
| AM223 | C12, I20* | 8 ×108 | 4×104 | 8.2×108 | 4.1×104 | 1.5×106 | 7.5×104 | 1.83 | ||
| AS47/AM223 | C12, I20* | 200 | 0.010 | 180 | 0.009 | 743 | 37 | 4127.78 | ||
| Saliva | AS47/AM223 | C12, I20* | 216 | 0.011 | 190 | 0.010 | 50 | 2.5 | 263.16 | |
| Nasal fluid | AS47/AM223 | C12, I20* | 850 | 0.043 | 2×103 | 0.1 | 17×103 | 850 | 8500 | |
| SARS-CoV-2 virion particles | Saliva | AS47/AM223 | C12, I20** | 12×106 | 600 | 4×106 | 200 | 49×103 | 2.5×103 | 12.25 |
| 9.4×105 RNA copies/mL | 2.9×105 RNA copies/mL | 3.9×106 RNA copies/mL | ||||||||
| 5.5×103 TCID50/mL | 1.7×103 TCID50/mL | 2.2×104 TCID50/mL | ||||||||
C12: centrifuge 1200g for 5 min; C15: centrifugate 1500 g for 5 min. I5: incubation for 5 min; I20: incubation for 20 min. The vortex agitation was 5 seconds for all tests, but an Analog Mixer (at 2,100 rpm) was used for antibody and virion particle detection, and a Digital Mixer (at 1550 rpm) was used for antigen detection (Methods section).
The LoD and LoQ values listed here were from the as-tested biological media, and need to be multiplexed with the dilution factor to estimate the values from undiluted WB. For example, the LoQ of antibody sensing was 33 fM in diluted 1% blood and 4 pM in 20% blood, corresponding to 3.3 pM and 20 pM, respectively, in undiluted 100% blood.
Intra-assay imprecision ( or ) was determined from multiple measurements of the same sample along different tube orientations.
Inter-assay error ( or ) was calculated based on the measured variability across replicate samples.
Remarkably, NasRED (LoD ~49 aM, or 7 fg/mL, from six replicates) outperformed ELISA (LoD 155 fM, or 23.3 pg/mL, from three replicates) in sensitivity by about 3,000 times while requiring only 16 times less sample volume ( compared to ) and 30 time shorter assay time (10 min compared to 5 hours). Additionally, the coefficient of variation (CV), an indicator of the variability of test results,31,32 and recovery analysis, measurement of assay precision in quantification,32 were also analyzed using for both NasRED and ELISA replicate tests. Our analyses showed repeated NasRED tests produced comparable LoDs (90 aM, 45 aM, 93 aM, 58 aM, 71 aM, and 60 aM, respectively, Figure S9a–b), determined by using the measurement error of the same NC sample along different tube orientations, or . In addition, NasRED exhibited <10% CV values and moderate recovery rates mostly within a range of 100±30% (Figure S9c–d), suggesting consistent and reproducible detection with acceptable assay precision across the 10 logs of concentrations. In comparison, ELISA analyses exhibited significantly higher variation in absorbance intensity, CV, and recovery values, particularly at low to moderate antibody concentrations (attomolar to picomolar range) (Figure 4c–d). The comparison demonstrated that NasRED outperformed ELISA in both sensitivity and accuracy, particularly at low protein concentrations.
Additionally, unlike ELISA or other assays that usually are characterized by a sigmoidal dose response (Figure 4b), NasRED data fitting was best modeled using a biphasic dose response33 (solid line, Figure S9e). The biphasic response yielded improved fitting (adjusted R2 = 0.999 versus 0.991) and recovery values compared to sigmoidal fitting (dash line, Figure S9f). While further theoretical and experimental investigations in the future are needed to completely decipher the underlying mechanisms, it is hypothesized that the affinity-driven, biochemical protein binding and the centrifugation/vortex-driven fluidic dynamic interactions work as two mechanisms that affect the AuNP size distribution in the precipitates, concentration gradient in the supernatant, and the subsequent signal transduction in the NasRED assay. These two mechanisms are thought to contribute to the two-phase transitions observed in the response curves. Furthermore, 20 replicates of NC samples in PBS were analyzed after mixing and the completion of the sensing protocol, displaying comparable signals within a 6% error (Figure S9g–h). These results confirm that the NasRED assay exhibits minimal non-specific background signals.
Antibody detection in biological fluids
SARS-CoV-2 antibody AS35 was then spiked into undiluted human serum in three replicate tests (Figure S10) and 1% diluted whole blood (WB) in two replicate tests (Figure S11), and analyzed following our above-established sensing protocol (Figure 4e–f). Unlike PBS buffer, human serum contains proteins, amino acids, and antibodies and is more viscous.34,35 WB presents additional challenges due to the presence of various blood cells and platelets, which may hinder protein binding. In addition, hemoglobin in the WB strongly absorbs blue and green light, producing a red color36 that could interfere with the solution color and electronic signals. To mitigate these physical and chemical impacts, WB was diluted to 1% or 20% (Figure S12) to ensure sufficient light transmission. Further, the sensing protocols (Table 1 and S3) were adjusted to increase the centrifugation speed to 1,500 g, the incubation time to 20 minutes, and the vortex force to 37.2 rps. This modified protocol facilitated the formation of AuNP precipitates at the tube bottoms for the antibody samples in HPS (Figure S10a), 1% WB (Figure S11a), and 20% WB (Figure S12 a–e), similar to those formed in PBS (Figure 4e–f).
The NasRED LoD for antibody detection in undiluted serum as found 76 aM (11 fg/mL) from three replicates, comparable to the estimated LoDs from each individual test (70 aM, 87 aM, and 200 aM, respectively). In comparison, the NasRED LoD was found to be 360 aM or 54 fg/mL in 1% WB (corresponding to 36 fM in undiluted blood) (Figure 4g), where was obtained from 10 replicates of NC samples in 1% WB (Figure S11d–e). Additionally, NasRED LoD was found 154 aM or 23 fg/mL in 20% WB (corresponding to 770 aM in undiluted blood) using a triplicate test (Figure S12d–g). Further, of the WB supernatants were collected and loaded in a PDMS plate to quantify the optical extinction on a laboratory spectrometer (Horiba) (Figure S12h–j). Clearly, high-sensitivity (femtomolar) antibody detection was evidently feasible without blood separation, as shown from both the PED signals and the spectral intensity modulation. Therefore, NasRED supports simplified sample preparation for near-patient use. Noticeably, the signal contrast was close to 1 in 1% WB (similar to tests in PBS), which is typically associated with a steeper sensing curve slope, broad dynamic range, and potentially more accurate quantification. However, dropped to only ~0.23 in 20% WB, possibly attributed to biological matrix effect in thicker blood due to the presence of diverse cells, proteins and other biomolecules. In case of antibody detection in 20% WB, blood cells also precipitated to the tube bottom after centrifugation (Figure S13). They might also contribute to fluidic and electrostatic interactions with AuNP clusters and monomers that could modulate the AuNP concentration in the supernatant to affect the sensitivity. Future closer examination of the dilution effect and protocol engineering may be necessary in applications where optimal LoDs and precise quantification are both critical. These analyses proved the feasibility of NasRED in semi-quantitative, high-sensitivity antibody detection in complex biological fluids for disease diagnostics and immunology studies.
N-protein antigen detection in biological fluids
NasRED was also demonstrated for SARS-CoV-2 antigen detection by functionalizing the AuNPs, using commercially available monoclonal antibodies (AS47 and AM223, both from Acrobiosystems) targeting the N-protein in PBS (Figure 5a–c, and Figure S14a–c). Interestingly, sensing using AS47-functionalized AuNP sensors displayed a poor color contrast ( and worst sensitivity (LoD 4 nM, or 200 ng/mL), AM223-AuNP sensors produced more differentiable optical contrast and better LoD (820 pM, or 41 ng/mL), and finally the use of two sets of AuNPs, functionalized with AS47 and AM223, respectively, produced the best signal contrast and best sensitivity (LoD 180 aM, or 9 fg/mL). The co-binding AuNP sensors also produced LoDs of 190 aM (10 fg/mL) and 2 fM (100 fg/mL) for N-proteins spiked in saliva and nasal fluid (Figure 5e, and Figure S14d–e). This greatly enhanced sensitivity is attributed to the improved binding avidity because the two mAbs cooperatively target distant, non-overlapping epitopes of the N-protein (AS47 on the C-terminus side and AM223 on the N-terminus side), which favors much more stable protein complex formation and accordingly larger AuNP clusters upon antibody-antigen binding24. On the other hand, the cobinder-based NasRED sensors easily distinguished SARS-CoV-2 from HCoV-NL63 and HCoV-229E, which cause the common cold and result in similar symptoms to SARS-CoV-2 infection (Figure 5f–g). In fact, even 4 fM SARS-CoV-2 N-protein produces a signal (, Figure 1g) differentiable from 400 nM HCoV-NL63 and HCoV-229E (, Figure 1g), indicating a very high specificity of the NasRED system.
Figure 5.

SARS-CoV-2 N-protein antigen detection in biological fluids. (a-c) Optical images showing antigen samples using AuNPs functionalized with different mAbs in PBS buffer: (a) only AS47, (b) only AM223, (c) co-binder (AS47 and AM223) on two sets of AuNPs. The antigen concentrations were 400 nM, 400 pM, 400 fM, and 40 aM. NC: buffer only. (d) Extracted PED signals in PBS buffer using different AuNP sensors, i.e. monobinder AS47/AuNP (black dashed line, LoD=4 nM), monobinder AM223/AuNP (black dotted line, LoD=820 pM), and cobinders (AS47/AuNP + AM223/AuNP, black solid line, LoD=180 aM). (e) Optical images showing samples for specificity test in PBS against human coronavirus (HCoV) (HCoV-NL63 and 229E) and SARS-CoV-2 N-proteins. All N-proteins were at 400 nM. (f) Extracted PED signals for NasRED tests against HCoV and SARS-CoV-2 antigens, as shown in figure e. (g) Extracted PED signal for SARS-CoV-2 antigen testing with the cobinders in saliva (red solid line, LoD=190 aM, or 0.01 pg/mL) and nasal fluid (blue solid line, LoD=2 fM, or 0.1 pg/mL). (h-k) The PED readout signal plotted as a function of the N-protien concentration, with: (h) the centrifugate force from 0 to 1,500 gravity (g), (i) the centrifugation time from 0 to 5 minutes, (j) the vortex agitation speed from 30 to 36 round per second (rps), and (k) the vortex time from 0 to 5 seconds. (l) Schematics of preparing inactive virus sample in saliva: lysis using triton x-100 to produce product comprising viral RNA and proteins, and mixing the lysis product with NasRED antigen sensor solution for viral N-protein detection. (m) Optical image showing a titration replicate of the inactivated virus samples. (n) PED signals of viral particles plotted against the virus N-protein concentration (bottom x-axis, determined by ELISA) and the viral particle RNA count (top x-axis, determined by qPCR) in saliva (LoD=4.1 pM determined from inter-assay variations, corresponding to 3×105 RNA copies/mL or 1.7×103 TCID50/mL). Solid black line: the average from the three replicates. Dash lines: individual tests with their LoDs calculated from the intra-assay imprecision.
To validate the performance of NasRED in real physiological samples, we systematically analyzed key physical parameters affecting detection sensitivity after spiking SARS-CoV-2 N-protein into saliva. In particular, we quantified the change in PED signal according to the N-protein concentration by controlling the centrifugation conditions (gravitational force, time) and vortex agitation conditions (speed, time), respectively (Figure 5h–k), and visualized the signal intensity difference for each condition . In saliva, N-protein showed a high PED signal at low concentration and a low signal at high concentration , showing a clear signal contrast of approximately . This indicated that the biochemical reactions between target proteins and the AuNPs, along with the subsequent physical AuNP precipitation and redistribution, occurred efficiently even in the saliva matrix. Similar to protocol optimization in PBS, a broad range of centrifugation and vortex conditions were able to yield high sensitivity and signal reproducibility. These findings further demonstrate that NasRED has the potential to maintain stable performance after optimization even in complex biological matrices. We further demonstrated direct detection of inactivated SARS-CoV-2 virion particles in saliva using the antigen sensors (Figure 5l–n). This was achieved by mixing the AuNP sensor buffer with PBS-diluted viruses (by 10 log titrations) and lysis reagents (Triton X-100, 0.3% v/v concentration) to extract N-proteins (Methods). By referencing the initial concentrations of N-proteins (2995 ng/mL or ~30 nM by ELISA) and RNA copy numbers (4.7 × 109 copies/mL by quantitative PCR, Table S5), the LoDs were determined as 4 pM (200 pg/mL) for the N-protein in three replicate experiments. This corresponds to an estimated RNA sensitivity of 2.9 × 105 copies/ml without amplification, and 1.7 × 103 TCID50/mL for the virus particles. In saliva samples, the presence of enzymes such as protease degrade N-protein molecules, thus lowering the sensitivity.37,38 As a reference, qPCR achieves a sensitivity of about 3.24 Log10 IU/mL in saliva or about 2×103 copies/mL.39 Importantly, while this simplified one-pot reaction is desired for point-of-care applications, the lysate products and buffer present in the NasRED sensing solution potentially negatively affected sensitivity. It is possible that separating the protein extraction and NasRED sensing into two different steps could further improve sensitivity.
DISCUSSION
In the NasRED platform, AuNP sensors play multiple roles in promoting the assay performance. They act as multivalent sensors to enhance protein binding28,40 and also effective optoelectronic beacons to convert the optical extinction to electronic signals, thus enabling reliable detection using a simple PED readout device for portable diagnostics. Through controlled fluidic dynamic interactions, AuNPs actively capture proteins during centrifugation-induced sedimentation, generating a concentration-boosting effect to greatly strengthen the protein-antibody binding, and then redistribute upon vortex agitation, establishing a concentration gradient profile that allows sensitive signal differentiation at sub-femtomolar level analyte concentrations.
To compare NasRED with existing assay formats, LoDs were calculated to represent the assay’s sensitivity in differentiating the minimum target SARS-CoV-2 antibody or antigen concentration from the background. Here we assessed the NasRED analytical sensitivity following two commonly used LoD definitions (Table 1), and , where and denote the standard deviation in measuring the blank and least-concentrated protein samples, respectively. The first definition incorporates the signal variations from both blank sample and the presence of analyte, thereby providing a more empirical and data-driven determination of the minimal concentration of analyte that can be distinguished from its absence.41 In comparison, the second approach estimates the LoD based solely on the measurement errors from blank sample without requiring additional data from low-concentration analyte, and therefore is much more widely adopted for its simplicity. While the approach was ultimately used to report the NasRED LoD values for consistency with standard practices, our comprehensive analyses demonstrated that the two definitions indeed yielded comparable results for both SARS-CoV-2 antibodies and antigens across all the tested biological fluids. Notably, the approach provided a slightly more optimistic estimate while the approach was more conservative.
Additionally, we introduced the limit of quantitative (LoQ), defined as , to assess NasRED assay’s capability for reliable quantification. We also calculated the ratio as a simplified reference metric that correlates with the slope of the assay sensing curve. Sub-femtomolar (<0.01 fM) LoD and 0.5–1 fM LoQ values were calculated for antibody sensing in PBS and HPS, with moderate LoQ/LoD ratios (<20), indicating strong sensitivity and quantification performance in HPS and PBS. In comparison, for antibody tests in 1% WB and antigen tests in saliva, sub-femtomolar (0.2 to 0.4 fM) LoDs and 30–50 fM LoQs were obtained, with notably larger LoQ/LoD ratio (100 to 250), indicating increased biological matrix effect. Further, sensing in highly complex and viscose fluids, for example antibody detection in 20% WB and antigen detection in nasal fluids, yielded sub-femtomolar to femtomolar LoDs, picomolar LoQs and very large LoQ/LoD ratios (~104). These were consistent with the small sensing slopes and signal contrasts of the PED readout data, attributed to the fact that these tests were not optimized due to resource constraint and more challenging biological matrix effect. Collectively, these above analyses underscore the importance of sample dilution and tailored sensing protocol optimization in individual biological fluids towards optimal NasRED performance.
We also attempted to compare NasRED with the reported performance from other traditional assay formats. LFAs usually have much poorer sensitivity (e.g., LoD of about 3 ng/mL, Table S6) compared to ELISA (e.g., literature reported LoD of 8.4 pg/mL, Table S7). In comparison, NasRED achieved exceptional antigen sensitivity (10 fg/mL and 100 fg/mL in saliva and nasal fluid), which is about 102 to 105 times more sensitive than LFA42 and 102 to 103 times better than ELISA.43 Importantly, NasRED also directly detected virion particles from saliva in a one-step, single-tube reaction format, achieving an estimated LoD of ~105 RNA copies/mL, comparable to that of some commercially available, isothermal NAAT tests such as Abbott ID NOW (~105 RNA copies/mL).
Besides the analytical performance, many other factors, such as the assay time, cost, complexity of the readout instrumentation, sample volume, etc., also strongly influence the success of antigen tests44–48 in practical use. For example, despite its low sensitivity, LFA became widely used during the COVID-19 pandemic due to its relatively low cost (~$10), rapid turnaround time (~15 min), and simpler sample preparation and detection schemes.49 In comparison, the use of ELISA and qPCR (Table S8) in POC applications is limited due to their higher test cost,50 bulky and costly instrument (e.g. ~$10,000 or more for ELISA absorbance reader or a PCR machine), long turnaround time (typically hours), and the need for specially trained personnel. In contrast, NasRED performs tests in microcentrifuge tubes without any labeling, washing, or enzymatic reaction, therefore simplifying the sensing process, reducing turnaround time (15 to 30 minutes from sample mixing to readout), and minimizing both sample volume (only , much smaller than a single drop of blood) and reagent usage24,25. Additionally, the overall sensing cost could be under $2 per NasRED test (~$0.1 per AuNP sensing solution, or ~$1.8 per test) using commercially available AuNPs (Cytodiagnostics) and can be further decreased to <$1 if the AuNPs are produced at scale by solution synthesis24,25. With performance better than ELISA but significantly reduced assay time, sample volume, system footprint, and reagent cost, NasRED is an ideal candidate for rapid, precise point-of-care testing, which is currently unavailable. Future development, such as integration and automation of centrifugation, vortex, and data analysis processes, will support the deployment of the NasRED platform not only in clinics, hospitals, and community centers but also in the field next to the users.
METHODS
Sources of reagents and instrumentation
Biotinylated WT-RBD (Wuhan-Hu-1 Catalog No. SPD-C82E9), AS35 anti-SARS-CoV-2 antibody (Catalog No. SAD-S35), biotinylated anti-SARS-CoV-2 Nucleocapsid antibody, Chimeric mAb, Human IgG (AM223 Catalog No. NUN-BM272), mouse IgG (AS47 Catalog No. NUN-S47L8) and SARS-COV-2 Nucleocapsid protein (Catalog No. NUN-C5227) were commercially purchased from ACROBiosystems. Triton™ X-100 was purchased from Millipore Sigma. The microcentrifuge (accuSpin Micro 17) and the vortexer (Analog Vortex Mixer Catalog No. 02-215-365 for antibody detection and Digital Vortex Mixer Catalog No. 02-215-418 for antigen detection) were both purchased from Thermo Fisher.
Electronic readout system
The NasRED readout system (Figure S1) included a reading device and a circuit board for signal processing. The reading device consisted of three main components: an LED (WP7113PGD, Kingbright), a photodiode-integrated circuit (SEN-12787, SparkFun Electronics, integrated with a digital light-sensor APDS-9960 from Broadcom), and a microcentrifuge tube holder that was 3D printed (Qidi Tech X-Plus 3D Printer) using black carbon fiber polycarbonate filament to fit snugly into a standard 0.5 mL Labcon microcentrifuge tube. An alkaline battery (9 V) was converted to 5V by a voltage regulator circuit as the power supply. The photodiode APDS-9960 was biased at 3.3 V and interfaced with a microcontroller (Atmega328) to convert the output into a digital signal. The regulator circuit was comprised of two electrolytic capacitors , two ceramic capacitors , and one protective Zener diode (Figure 2a (i)). The constant current LED driver circuit utilized a feedback loop to stabilize the current supplied to the LED. For example, a sudden increase in the base voltage of transistor T2 ) would increase its base-to-emitter bias and, therefore, conduct a higher amount of current through the LED and the T2 collector, which would tend to cause the LED current to rise. However, the higher would also boost the voltage drop across resistor R3, or the base-to-emitter bias voltage of transistor T1 . This, in turn, would lead to enhanced emitter current and collector current . This increase in would cause larger voltage to drop across resistor R2, and, therefore lower to counter the voltage fluctuation and stabilize the LED signals. --All the above electronic components, including microcontroller, capacitors, resistors, transistors, etc., were purchased from DigiKey unless specified otherwise.
Optical power calculations
The output power of a green LED (peak emission wavelength 557 nm, full-width-at-half maximum (FWHM) 30 nm) was measured and digitized using an APDS-9960 sensor. The APDS-9960 sensor operates at a gain of 16 and an integration time (ATIME) of 175 msec, measures the light intensity in Clear, Blue, Green, and Red channels (Figure S1c), and outputs the emission power in the Clear channel with a responsibility of . The LED output power (irradiance) was calculated from the counts obtained via the inter-integrated circuit (I2C) interface. The same LED and APDS-9960 sensor were used throughout the measurements for consistency.
Polydimethylsiloxane (PDMS) well plate fabrication
Sylgard 184 silicone elastomer base was thoroughly mixed with the curing aagent (mass ratio 10:1) for 30 minutes in a petri dish and placed in a vacuum container for 2 hours to degas. The mixture was then fully cured at room temperature into a PDMS membrane, which was then cut to the desired size and punched to create 2 mm wells. The as-prepared PDMS membrane and a diced fused silica chip ( thick) were rinsed with isopropyl alcohol, dried by nitrogen, and treated with oxygen plasma. Immediately after, the two were bonded to form a PDMS well plate. The PDMS plate was treated with 1 wt% PVA in water solution for 10 minutes to prevent the non-specific binding of proteins, adapted from previously reported methods.51 Finally, it was dried with nitrogen, heated on a 110°C hotplate for 15 minutes, and cooled to room temperature.
1×PBS dilution buffer
The 10× Phosphate-buffered saline (PBS, Fisher Scientific) stock buffer was mixed with glycerol (Sigma-Aldrich) and BSA (Sigma-Aldrich) and deionized water (Fisher Scientific) to create a 1×PBS dilution buffer with a final concentration of 1×PBS, 20 % v/v glycerol, and 1 wt % BSA. This dilution buffer (pH~7.4) was used to prepare the AuNP sensors, antigen and antibody solutions, and diluted biological media.
Sensing solution preparation and quantification
Streptavidin-functionalized AuNPs (~0.13 nM, 80 nm, OD10) were purchased from Cytodiagnostics, and quantified by Nanosight (NS3000, Malvern Panalytical). To create AuNP sensing solution (e.g. , ODPED = 0.5), the streptavidin-coated AuNPs (, OD10) were mixed with an excessive amount of biotinylated antigens or antibodies (e.g. about ), incubated for 2 hours, diluted by adding of 1×PBS dilution buffer and purified by centrifugation at 9,600 g for 10 minutes, followed by removing supernatant and adding 1×PBS dilution buffer. The above purification was repeated another time to remove unbound biotinylated antigens or antibodies. The purified sensing solutions were measured by Nanodrop 2000 (Thermo Fisher) to determine the AuNP concentration and then adjusted to the desired optical extinction level (e.g., in our case ~0.019 nM for 80 nm AuNPs) by mixing with 1×PBS dilution buffer. Prior to each sensing experiment, the stock sensing solutions were aliquoted into Eppendorf tubes of each.
Antibody and antigen detection in biological buffer
Sensing in PBS:
The target antigen (N-protein) or antibody was adjusted to a target concentration (e.g., 40 aM to ) by mixing with 1×PBS dilution buffer. Then of such antigen or antibody solutions were mixed with an AuNP sensing solution, followed by vortex agitation at 34.5 rps for 5 seconds. The final antigen or antibody concentration in the mixed reaction buffer was therefore diluted four times from the original solutions, e.g. 40 aM to 10 aM and to .
Biological media preparation:
Human Pooled Serum (HPS) was used as purchased. Human whole blood (WB) was diluted with 1×PBS dilution buffer to 20% and 1% to minimize optical and biochemical interference. Saliva and nasal fluid were purified by centrifugation at 9,600 g for 5 minutes to precipitate the mucus, and the supernatant was used. Their corresponding final concentration in the sensing mixture was 25% HPS, 25% nasal fluid, 25% saliva, and 5% WB. Single donor human WB, HPS, human nasal fluid, and saliva were purchased from Innovative Research, Inc.
Inactivated virus N-protein extraction:
PROtrol SARS-CoV-2 (Isolate: USA-WA1/2020) was purchased from ZeptoMetrix LLC in standard Vero E6 culturing medium (2% minimum essential medium (MEM)). The N-Protein was extracted by adding 0.6% v/v Triton X-100 (non-ionic surfactant, from Millipore Sigma) at a 1:1 v/v ratio to the SARS-CoV-2 sample, followed by 5 minutes incubation to release the N-protein via viral envelope micellization.52,53W
qPCR assay was performed by ZeptoMetrix, and the protocol and the data were shared with their permission. The SARS-CoV-2 viral RNA was extracted from of PROtrol sample (part # PROSARS(COV2)-587) using QIAmp MinElute Virus extraction. Briefly, the sample was lysed using the lysis buffer (NaCl, Tris-HCL, SDS), followed by RNA binding to silica in a spin column, sequentially washed by guanidine hydrochloride and ethanol to remove impurities and salts, and the viral RNA was eluted into of aqueous viral elution (AVE) buffer (RNase-free water, 0.04% Sodium Azide (NaN3)) and immediately used to measure genomic copies in real-time quantitative PCR assay. Furthermore, the RNA from working quantitative standard and extraction control were extracted alongside the PROtrol samples. The extracted SARS-CoV-2 RNA was tested in the SAR-CoV-2 qRT-PCR assay that targets the N protein-coding gene. The PCR reaction buffer included TaqMan Fast Virus 1-Step Multiplex Master Mix for qPCR (carboxy rhodamine dye free), of each forward and reverse primers at 500 nM final concentration, fluorescent probe at 200 nM final concentration (Table S4), of extracted RNA as PCR template, and RNase-free water ( total mixture volume). PCR was run in a QuantStudio 5 System. A standard curve was generated using 10-fold serial dilutions of the extracted RNA from a working standard with known copies/mL, determined by droplet digital PCR (ddPCR). The efficiency of the PCR reaction was found to be 93.4%, derived from a slope of −3.492 from the standard CT curve. The same 10-fold serial dilutions were made with the extracted RNA from the PROtrol sample (n=2 sets) and tested against the standard curve to determine the quantity of viral RNA in the PROtrol sample. The working standard and the PROtrol qPCR data were analyzed using the QuantStudio Design & Analysis Software (V1.4.3).
NasRED data processing
The NasRED data analysis involved signal collection, calibration, and normalization. First, the target proteins in biological media were measured on the designed PED device, together with a negative control (NC, functionalized AuNPs mixed with biological medium without target proteins). Then, each sample tube ( AuNP sensing solution mixed with target protein solution, in total ) and the NC tube underwent measurements along five orientations relative to a tube holder position to account for tube variations and inhomogeneous matrix optical impacts such as turbidity and color. The measured digital photodiode sensor signals for each sample tube were averaged, collected to a datasheet (.csv), and saved to a predefined local folder on a computer, all automated by Python scripts. The averaged sample signals ( for the five measurements, at protein concentration ) represented the total optical transmission through the sample supernatant and the microcentrifuge tube right after mixing. After testing (through centrifugation, incubation, and vortex agitation), another five measurements were performed for each tube along the same orientations, and the electronic signals were again recorded as . Subsequently, the calibrated electronic signals, representing the difference in optical extinction caused only by the decrease of free-floating AuNPs due to precipitation, were calculated as , and . The difference of NC signals before and after the test was also calculated, and <6% , i.e. most AuNPs redisperse and return to their original state right after mixing, was used as a criteria to validate the sensing protocol (Figure S9). Finally, considering the batch-to-batch difference in AuNP sensor concentrations as well as physical variations of biological fluids on the NasRED signals, the collected PED signals were further normalized to a range from 0 to 1 as . Here, a positive control signal (PC) was chosen for normalization purpose to represent the highest possible optical transmission, i.e. AuNPs completely precipitate. For mostly clear biological media, including PBS, saliva, nasal fluid, and serum, a mixture sample of 1×PBS dilution buffer (representing sensing solution after AuNPs precipitate) and biological media was used as a reference for simplicity, and the transmission signal was collected as the . To account for the complex physical and chemical interactions between blood and AuNPs, the PC reference in WB was designed by mixing WT RBD-coated AuNP sensing solution with 20% WB spiked with high-concentration antibodies. In this case, the PC signals were defined as .
In addition, the intra-assay imprecision, or within-test error , of the measurements were calculated as the standard deviation of the 5 normalized signals for the same sample (, for the 5 measurement orientations). The normalized sensing signals were plotted against the antigen or antibody concentrations in each tested biological media. For the reproducibility experiment performed in this work, the average of each concentration measurement was calculated as (j=1 to N for the N replicate measurements) and the inter-assay error (or between-test variation) was calculated as . Here characterizes the systematic readout errors from the PED device and human operations, and accounts for the variability of the repeated experiments30. and were found to be quite close in our experiments in different biological media, often leading to comparable sensitivity values.
Data fitting, LoD, LoQ, CV and Recovery calculations
The normalized signals were fitted using Origin 2024b software (OriginLab, USA) by orthogonal distance regression algorithm to consider the data and error weight in fitting calculations. Biphasic dose-response model was used for fitting, following
where two sets of EC50, , and slope values were used. and were the highest and lowest sensing signals obtained from NC (or lowest concentrations) and the highest protein concentrations, respectively, EC50 was the half maximal effective concentration, and describes the steepness of the sensing curve. In general, the EC50 and values would determine the LoD. A large signal contrast , resulting from a small (i.e. high level of AuNP precipitation), in combination with a small slope , would be desired for obtaining a large dynamic range. In addition, a dose-response model (or sigmoidal function) was explored to produce acceptable fitting, following
where one sets of EC50, , and slope values were used.
Fitting parameters were adjusted so that the goodness of fit parameters such as reduced Chi-square, R-squared coefficient of determination (COD), and reduced sum of squares (RSS) were calculated close to 1,1, and 0, respectively. The LoD and LoQ values were then calculated such that the PED signal at this concentration on the fitted curve was distinguished from the NC sample in consideration of the intra-assay imprecision ) or inter-assay errors of the NC sample measurements, respectively.54–56
We also included LoD calculations following another definition, where 1.645 times the standard deviation of the NC and the lowest concentration sample were considered:
where or and was the lowest protein (antibody, antigen or virion particle) concentration.54–56
The coefficient of variation (CV) was calculated as the ratio of the standard deviation to the mean value of each data set, following (j=1 to N for the N replicate measurements). The recovery (%) was calculated as the ratio of measured PED signals to the values from standard biphasic dose-response fitting at different protein concentrations , following .57
Direct ELISA assay
The MICROLON® 96-well polystyrene microplate (Cat# 5665-5061, USA Scientific) was coated with of streptavidin solution (Sigma-Aldrich) and incubated overnight at 4°C. The plate was then washed three times using of PBS containing 0.5% (v/v) Tween-20. To block non-specific surface adsorption, of a 5% (w/v) BSA solution (Sigma-Aldrich) was added and incubated for 2 hours at room temperature and washed four times with the same washing solution. Biotinylated RBD was added to each well and incubated for 2 hours at room temperature followed by four times washing. Serially diluted samples in three replicates, ranging from 400 nM to 40 aM, including a negative control, were added to the wells ( per well) and incubated for 2 hours at room temperature and washed four times using the washing solution. Subsequently, of secondary antibody (Peroxidase AffiniPure™ Goat Anti-Human IgG (H+L), Jackson ImmunoResearch, Cat# 109-035-088) was added to the wells and incubated for 2 hours at room temperature and washed four times with the washing solution. To develop the reaction, of 1-Step Ultra TMB-ELISA substrate solution (Cat# 34028, Thermo Fisher Scientific) was added to the wells and incubated for 15 minutes. The reaction was stopped by adding of 2 M H2SO4 (Thermo Fisher Scientific) to each well. The absorbance was measured on a BioTek Synergy Neo2 Hybrid Multimode Reader at 450 nm (Agilent Technologies).
The ELISA absorbance signals at 450 nm were calculated as (j=1 to 3 for the 3 replicates). To facilitate comparison with , the ELISA signals were also normalized, with 0 representing the minimum absorption at negative controls and 1 representing maximum absorption (approaching device limit at high analyte concentrations such as 40 nM). The error bars, i.e. (j=1 to 3), represented the standard deviations of the ELISA signals between the 3 replicate experiments. The LoD was calculated based on the same principle described in the NasRED data processing, where the assay can differentiate the sample from blank sample following the equation, .
Spectrometry measurement in blood sample
AS35 was detected in 20% diluted human WB, following the modified protocol, and analyzed by spectroscopic measurements. Here, the samples’ supernatant liquids were loaded into a PDMS well plate and measured by the spectrometer (iHR-320, Horiba Instruments Inc.). To achieve the best contrast, the focal plane was adjusted at the well plate surface, and then three spots were selected from within each well (one at the center and two near the edges) to account for possible inhomogeneity of AuNPs’ distribution. The transmitted light was collected by a 50×objective lens (NA=0.8) from each spot, and the transmission spectra were measured 64 times from 350 to 750 nm range with 0.01 seconds integration time. The signals collected from the three spots were then averaged and normalized against the reflectance from a silver-coated mirror as , where was the wavelength, and was the protein concentration. The opticalextinction spectra were obtained as , from which the values at were selected as the signal . In addition, was similarly collected for the positive control ( WT RBD-coated AuNP sensing solution mixed with high-concentration () antibodies spiked in 20% WB), and was determined for the NC sample (sensing solution mixed with the biological medium at 3:1 ratio). The spectrometer signals were calculated as and the error of the measurement was calculated as the standard deviation (SD) of the 3 normalized signals (, for the 3 measurement spots).
Supplementary Material
Device design details, AuNP sedimentation and quantification, reproducibility tests, ELISA comparison, antibody and antigen detection in biological fluids, qPCR validation, and diagnostic performance comparison.
ACKNOWLEDGMENT
We acknowledge support from Dr. D. Williams at Arizona State University for Cryo-EM inspections, and Dr. Q. Xu at National Institutes of Health for helpful discussions. This project was supported in part by the National Science Foundation (NSF) under grant no. 1847324, U.S. Department of Agriculture AFRI 2022-67021-37013, and by the National Institute of Health under grant no. R21AI169098 and DP2GM149552. Access to the Cryo-EM was supported, in part, by NSF grant no. ECCS-1542160.
Footnotes
Declaration of Generative AI and AI-Assisted Technologies in the Writing Process
After completing the manuscript revision, the authors used OpenAI’s ChatGPT-4o as a reference in selected sections of the manuscripts to enhance clarity. The authors thoroughly reviewed and edited the content as necessary and take full responsibility for the final published version.
Competing Interests Statement
C.W. and S.M. are co-founders of REDX Diagnostics, LLC. Y.C., S.M., and C.W. are inventors listed on a patent application relevant to some of the data included in this article. While the underlying technology may be subject to future commercialization, the affiliation did not influence the reported research findings. The remaining authors declare no competing interests.
REFERENCES
- (1).World Health Organization. WHO COVID-19 dashboard. https://data.who.int/ (accessed 2025-01-02). [Google Scholar]
- (2).Walmsley T; Rose A; John R; Wei D; Hlávka JP; Machado J; Byrd K Macroeconomic Consequences of the COVID-19 Pandemic. Econ Model 2023, 120, 106147. 10.1016/j.econmod.2022.106147. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (3).Myers LC; Liu VX The COVID-19 Pandemic Strikes Again and Again and Again. JAMA Netw Open 2022, 5 (3), e221760. 10.1001/jamanetworkopen.2022.1760. [DOI] [PubMed] [Google Scholar]
- (4).Filip R; Puscaselu RG; Anchidin-Norocel L; Dimian M; Savage WK Global Challenges to Public Health Care Systems during the COVID-19 Pandemic: A Review of Pandemic Measures and Problems. J Pers Med 2022, 12 (8), 1295. 10.3390/jpm12081295. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (5).World Health Organization. Prioritizing diseases for research and development in emergency contexts. https://www.who.int/activities/prioritizing-diseases-for-research-and-development-in-emergency-contexts (accessed 2024-01-06).
- (6).UK Health Security Agency. Emerging infections: how and why they arise. https://www.gov.uk/government/publications/emerging-infections-characteristics-epidemiology-and-global-distribution/emerging-infections-how-and-why-they-arise (accessed 2024-01-06).
- (7).Ahmadivand A; Gerislioglu B; Ahuja R; Kumar Mishra Y Terahertz Plasmonics: The Rise of Toroidal Metadevices towards Immunobiosensings. Materials Today. 2020. 10.1016/j.mattod.2019.08.002. [DOI] [Google Scholar]
- (8).Mahshid SS; Flynn SE; Mahshid S The Potential Application of Electrochemical Biosensors in the COVID-19 Pandemic: A Perspective on the Rapid Diagnostics of SARS-CoV-2. Biosens Bioelectron 2021, 176. 10.1016/j.bios.2020.112905. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (9).Kubina R; Dziedzic A Molecular and Serological Tests for COVID-19. A Comparative Review of SARS-CoV-2 Coronavirus Laboratory and Point-of-Care Diagnostics. Diagnostics (Basel) 2020, 10 (6), 434. 10.3390/diagnostics10060434. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (10).Reta DH; Tessema TS; Ashenef AS; Desta AF; Labisso WL; Gizaw ST; Abay SM; Melka DS; Reta FA Molecular and Immunological Diagnostic Techniques of Medical Viruses. Int J Microbiol 2020, 2020 (1), 8832728. 10.1155/2020/8832728. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (11).Katsanis SH; Katsanis N Molecular Genetic Testing and the Future of Clinical Genomics. Nat Rev Genet 2013, 14 (6), 415–426. 10.1038/nrg3493. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (12).Binny RN; Priest P; French NP; Parry M; Lustig A; Hendy SC; Maclaren OJ; Ridings KM; Steyn N; Vattiato G; Plank MJ Sensitivity of Reverse Transcription Polymerase Chain Reaction Tests for Severe Acute Respiratory Syndrome Coronavirus 2 through Time. Journal of Infectious Diseases 2023, 227 (1), 9–17. 10.1093/infdis/jiac317. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (13).Eyzaguirre EJ; Walker DH; Zaki SR Immunohistology of Infectious Diseases. In Diagnostic Immunohistochemistry; Dabbs DJ, Ed.; 2006; pp 43–64. [Google Scholar]
- (14).Girt GC; Lakshminarayanan A; Huo J; Dormon J; Norman C; Afrough B; Harding A; James W; Owens RJ; Naismith JH The Use of Nanobodies in a Sensitive ELISA Test for SARS-CoV-2 Spike 1 Protein. R Soc Open Sci 2021, 8 (9), 211016. 10.1098/rsos.211016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (15).Vernet R; Charrier E; Grogg J; Mach N A Quantitative ELISA Protocol for Detection of Specific Human IgG against the SARS-CoV-2 Spike Protein. Vaccines (Basel) 2021, 9 (7), 770. 10.3390/vaccines9070770. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (16).MacMullan MA; Ibrayeva A; Trettner K; Deming L; Das S; Tran F; Moreno JR; Casian JG; Chellamuthu P; Kraft J; Kozak K; Turner FE; Slepnev VI; Le Page LM ELISA Detection of SARS-CoV-2 Antibodies in Saliva. Sci Rep 2020, 10 (1), 20818. 10.1038/s41598-020-77555-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (17).Shah K; Maghsoudlou P Enzyme-Linked Immunosorbent Assay (ELISA): The Basics. Br J Hosp Med 2016, 77 (7), 98–101. 10.12968/hmed.2016.77.7.C98. [DOI] [PubMed] [Google Scholar]
- (18).Aydin S A Short History, Principles, and Types of ELISA, and Our Laboratory Experience with Peptide/Protein Analyses Using ELISA. Peptides (N.Y.) 2015, 72, 4–15. 10.1016/j.peptides.2015.04.012. [DOI] [PubMed] [Google Scholar]
- (19).Jegerlehner S; Suter-Riniker F; Jent P; Bittel P; Nagler M Diagnostic Accuracy of a SARS-CoV-2 Rapid Antigen Test in Real-Life Clinical Settings. International Journal of Infectious Diseases 2021, 109, 118–122. 10.1016/j.ijid.2021.07.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (20).Ndwandwe D; Mathebula L; Kamadjeu R; Wiysonge CS Cochrane Corner: Rapid Point-of-Care Antigen and Molecular-Based Tests for the Diagnosis of COVID-19 Infection. Pan Afr Med J 2020, 37, 10. 10.11604/pamj.supp.2020.37.10.25982. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (21).Lin Q; Wen D; Wu J; Liu L; Wu W; Fang X; Kong J Microfluidic Immunoassays for Sensitive and Simultaneous Detection of IgG/IgM/Antigen of SARS-CoV-2 within 15 Min. Anal Chem 2020, 92 (14), 9454–9458. 10.1021/acs.analchem.0c01635. [DOI] [PubMed] [Google Scholar]
- (22).Brandmeier JC; Jurga N; Grzyb T; Hlaváček A; Obořilová R; Skládal P; Farka Z; Gorris HH Digital and Analog Detection of SARS-CoV-2 Nucleocapsid Protein via an Upconversion-Linked Immunosorbent Assay. Anal Chem 2023, 95 (10), 4753–4759. 10.1021/acs.analchem.2c05670. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (23).Zhang GQ; Gao Z; Zhang J; Ou H; Gao H; Kwok RTK; Ding D; Tang BZ A Wearable AIEgen-Based Lateral Flow Test Strip for Rapid Detection of SARS-CoV-2 RBD Protein and N Protein. Cell Rep Phys Sci 2022, 3 (2). 10.1016/j.xcrp.2022.100740. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (24).Chen X; Kang S; Ikbal MA; Zhao Z; Pan Y; Zuo J; Gu L; Wang C Synthetic Nanobody-Functionalized Nanoparticles for Accelerated Development of Rapid, Accessible Detection of Viral Antigens. Biosens Bioelectron 2022, 202, 113971. 10.1016/j.bios.2022.113971. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (25).Ikbal MDA; Kang S; Chen X; Gu L; Wang C Picomolar-Level Sensing of Cannabidiol by Metal Nanoparticles Functionalized with Chemically Induced Dimerization Binders. ACS Sens 2023, 8 (12), 4696–4706. 10.1021/acssensors.3c01758. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (26).Midelet J; El-Sagheer AH; Brown T; Kanaras AG; Werts MHV The Sedimentation of Colloidal Nanoparticles in Solution and Its Study Using Quantitative Digital Photography. Particle and Particle Systems Characterization 2017, 34 (10), 1700095. 10.1002/ppsc.201700095. [DOI] [Google Scholar]
- (27).Loth E Lift of a Solid Spherical Particle Subject to Vorticity and/or Spin. AIAA Journal 2008, 46 (4). 10.2514/1.29159. [DOI] [Google Scholar]
- (28).Hornick CL; Karush F Antibody Affinity-III the Role of Multivalence. Immunochemistry 1972, 9 (3), 325–340. 10.1016/0019-2791(72)90096-1. [DOI] [PubMed] [Google Scholar]
- (29).Leckband DE; Schmitt FJ; Israelachvili JN; Knoll W Direct Force Measurements of Specific and Nonspecific Protein Interactions. Biochemistry 1994, 33 (15). 10.1021/bi00181a023. [DOI] [PubMed] [Google Scholar]
- (30).Monach PA Repeating Tests: Different Roles in Research Studies and Clinical Medicine. Biomark Med 2012, 6 (5), 691–703. 10.2217/bmm.12.57. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (31).Reed GF; Lynn F; Meade BD Use of Coefficient of Variation in Assessing Variability of Quantitative Assays. Clin Diagn Lab Immunol 2002, 9 (6). 10.1128/CDLI.9.6.1235-1239.2002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (32).Cox KL; Devanarayan V; Kriauciunas A; Manetta J; Montrose C; Sittampalam S Immunoassay Methods. In The Assay Guidance Manual; Eli Lilly & Company and the National Center for Advancing Translational Sciences, 2012. [Google Scholar]
- (33).Calabrese EJ Biphasic Dose Responses in Biology, Toxicology and Medicine: Accounting for Their Generalizability and Quantitative Features. Environmental Pollution 2013, 182, 452–460. 10.1016/j.envpol.2013.07.046. [DOI] [PubMed] [Google Scholar]
- (34).Leeman M; Choi J; Hansson S; Storm MU; Nilsson L Proteins and Antibodies in Serum, Plasma, and Whole Blood—Size Characterization Using Asymmetrical Flow Field-Flow Fractionation (AF4). Anal Bioanal Chem 2018, 410 (20), 4867–4873. 10.1007/s00216-018-1127-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (35).Rosencranz R; Bogen SA Clinical Laboratory Measurement of Serum, Plasma, and Blood Viscosity. Am J Clin Pathol 2006, 125, 78–96. 10.1309/fff7u8rrpk26vapy. [DOI] [PubMed] [Google Scholar]
- (36).Shenk JH; Hall JL; King HH Spectrophotometric Characteristics of Hemoglobins. Journal of Biological Chemistry 1934, 105 (4), 753–760. 10.1016/s0021-9258(18)75508-7. [DOI] [Google Scholar]
- (37).Nishibata Y; Koshimoto S; Ogaki K; Ishikawa E; Wada K; Yoshinari M; Tamura Y; Uozumi R; Masuda S; Tomaru U; Ishizu A RNase in the Saliva Can Affect the Detection of Severe Acute Respiratory Syndrome Coronavirus 2 by Real-Time One-Step Polymerase Chain Reaction Using Saliva Samples. Pathol Res Pract 2021, 220. 10.1016/j.prp.2021.153381. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (38).Thomadaki K; Helmerhorst EJ; Tian N; Sun X; Siqueira WL; Walt DR; Oppenheim FG Whole-Saliva Proteolysis and Its Impact on Salivary Diagnostics. J Dent Res 2011, 90 (11). 10.1177/0022034511420721. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (39).Ko H-Y; Li Y-T; Li Y-Y; Chiang M-T; Lee Y-L; Liu W-C; Liao C-C; Chang C-S; Lin Y-L Optimization and Improvement of QPCR Detection Sensitivity of SARS-CoV-2 in Saliva. Microbiol Spectr 2023, 11 (3). 10.1128/spectrum.04640-22. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (40).Greenspan NS; Cooper LJN Cooperative Binding by Mouse IgG3 Antibodies: Implications for Functional Affinity, Effector Function, and Isotype Restriction. Springer Semin Immunopathol 1993, 15 (2–3), 275–291. 10.1007/BF00201107. [DOI] [PubMed] [Google Scholar]
- (41).Armbruster DA; Pry T Limit of Blank, Limit of Detection and Limit of Quantitation. Clin Biochem Rev 2008, 29 Suppl 1. [PMC free article] [PubMed] [Google Scholar]
- (42).Ang GY; Chan KG; Yean CY; Yu CY Lateral Flow Immunoassays for SARS-CoV-2. Diagnostics. 2022. 10.3390/diagnostics12112854. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (43).Humbert MV; Opurum PC; Brendish NJ; Poole S; He P; Katis I; Quaye J; Bediako Y; Duriez PJ; Eason RW; Sones C; Quaye O; Awandare GA; Christodoulides M; Clark TW; Quashie PK; McCormick CJ A SARS-CoV-2 Nucleocapsid ELISA Represents a Low-Cost Alternative to Lateral Flow Testing for Community Screening in LMI Countries. Journal of Infection 2022, 84 (1), 48–55. 10.1016/j.jinf.2021.08.049. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (44).iHealth. iHealth COVID-19 antigen rapid test. https://ihealthlabs.com/products/ihealth-covid-19-antigen-rapid-test?srsltid=AfmBOopp4XjzvuEyn7R7XCppH7XUWWWm0S3PSYqq57E2ldIP8bumQRS (accessed 2024-08-16).
- (45).Genabio. Genabio COVID-19 rapid self-test kit. https://www.genabio.com/covid/ (accessed 2024-08-16).
- (46).Abbott. Abbott m2000 real-time. https://www.molecular.abbott/us/en/products/infectious-disease/RealTime-SARS-CoV-2-Assay (accessed 2024-08-16).
- (47).Viviani T DiaSorin Simplex COVID-19 direct. https://us.diasorin.com/en/molecular-diagnostics/kits-reagents/simplexa-covid-19-flu-ab-direct (accessed 2024-08-16).
- (48).Matsumura Y; Shimizu T; Noguchi T; Nakano S; Yamamoto M; Nagao M Comparison of 12 Molecular Detection Assays for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2). Journal of Molecular Diagnostics 2021, 23 (2), 164–170. 10.1016/j.jmoldx.2020.11.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (49).Peeling RW; Heymann DL; Teo YY; Garcia PJ Diagnostics for COVID-19: Moving from Pandemic Response to Control. The Lancet 2022, 399 (10326), 757–768. 10.1016/S0140-6736(21)02346-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (50).Humbert MV; Opurum PC; Brendish NJ; Poole S; He P; Katis I; Quaye J; Bediako Y; Duriez PJ; Eason RW; Sones C; Quaye O; Awandare GA; Christodoulides M; Clark TW; Quashie PK; McCormick CJ A SARS-CoV-2 Nucleocapsid ELISA Represents a Low-Cost Alternative to Lateral Flow Testing for Community Screening in LMI Countries. Journal of Infection 2022, 84 (1), 48–55. 10.1016/j.jinf.2021.08.049. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (51).Trantidou T; Elani Y; Parsons E; Ces O Hydrophilic Surface Modification of PDMS for Droplet Microfluidics Using a Simple, Quick, and Robust Method via PVA Deposition. Microsyst Nanoeng 2017, 3, 16091. 10.1038/micronano.2016.91. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (52).Dalgarno PA; Juan-Colás J; Hedley GJ; Piñeiro L; Novo M; Perez-Gonzalez C; Samuel IDW; Leake MC; Johnson S; Al-Soufi W; Penedo JC; Quinn SD Unveiling the Multi-Step Solubilization Mechanism of Sub-Micron Size Vesicles by Detergents. Sci Rep 2019, 9 (1), 12897. 10.1038/s41598-019-49210-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (53).Roberts PL Virus Inactivation by Solvent/Detergent Treatment Using Triton X-100 in a High Purity Factor VIII. Biologicals 2008, 36 (5), 330–335. 10.1016/j.biologicals.2008.06.002. [DOI] [PubMed] [Google Scholar]
- (54).Clinical and Laboratory Standards Institute. Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline. NCCLS Document EP17-A. In CLSI Document EP17-A; 2004; Vol. 24. [Google Scholar]
- (55).Shrivastava A; Gupta V Methods for the Determination of Limit of Detection and Limit of Quantitation of the Analytical Methods. Chronicles of Young Scientists 2011, 2 (1). 10.4103/2229-5186.79345. [DOI] [Google Scholar]
- (56).Armbruster DA; Pry T Limit of Blank, Limit of Detection and Limit of Quantitation. Clin Biochem Rev 2008, 29 (1), 49–52. [PMC free article] [PubMed] [Google Scholar]
- (57).Validation BM Guidance for Industry Bioanalytical Method Validation. Vet Med 2001, No. May. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.