Abstract
Objectives: This study aims to evaluate the intrasession repeatability of the HBM-1 (Huvitz) optical biometer and to assess its agreement with the IOLMaster 700 (Zeiss) for the main biometric parameters used in intraocular lens (IOL) power calculation and in myopia management. Methods: A cross-sectional observational study was conducted in 82 eyes of 82 patients with the following age distribution: pediatric 26.8%, young adults 35.4%, and older adults 37.8% (total range 6–79 years). Optical biometry was performed three consecutive times with both biometers in the same session by a single trained examiner. HBM-1 repeatability was assessed using the within-subject standard deviation (Sw), the repeatability coefficient, and the intraclass correlation coefficient (ICC). Agreement between biometers was analyzed using Bland–Altman plots (limits of agreement, LoA). Results: The HBM-1 showed excellent intrasession repeatability, with very low Sw values—on the order of hundredths of a millimeter for axial length (AL), anterior chamber depth (ACD), and lens thickness (LT), and hundredths of a diopter for keratometry—with ICC ≥ 0.97 for most parameters. The mean bias (HBM-1 vs. IOLMaster 700) was small: AL 0.012 ± 0.052 mm (p = 0.045; LoA: −0.09 to 0.11 mm), ACD 0.059 ± 0.068 mm (p < 0.001; −0.07 to 0.19 mm), LT 0.052 ± 0.090 mm (p < 0.001; −0.12 to 0.23 mm), and central corneal thickness 0.82 ± 7.12 μm (p = 0.301; −13.1 to 14.8 μm). For corneal diameter and corneal curvature, mean differences were small (≤0.07 D) and not statistically significant in most cases. Age was not associated with discrepancies in AL but showed weak correlations with some anterior segment differences, without clear clinical relevance. Conclusions: The HBM-1 demonstrated excellent intrasession repeatability and a good level of clinical agreement with the IOLMaster 700 in a broad population that included children, young adults, and older adults.
Keywords: optical biometry, swept-source OCT, axial length, intraocular lens power calculation, myopia management
1. Introduction
High-precision ocular biometry constitutes a fundamental component of contemporary ophthalmology and optometry, as it allows for improved accuracy in intraocular lens (IOL) power calculation, facilitates the planning of personalized refractive strategies, and supports axial length monitoring in myopia management programs [1,2,3,4]. Optical biometers based on partial coherence interferometry and, more recently, swept-source optical coherence tomography (SS-OCT) have largely superseded ultrasound biometry, owing to their superior accuracy and reproducibility in the measurement of axial length and anterior segment parameters [2,5,6].
The IOLMaster 700 (Carl Zeiss Meditec, Jena, Germany), based on SS-OCT, has become an established clinical reference standard in ocular biometry, supported by robust evidence of its repeatability, reliability, and agreement with other optical biometers, including in cataract populations [6,7,8,9,10]. In parallel with advances in optical biometry, novel devices integrating biometry with additional functionalities for anterior segment analysis and/or myopia monitoring have been introduced, underscoring the need to benchmark their performance against consolidated standards using repeatability and agreement metrics [4,11,12].
The HBM-1 (Huvitz, Anyang, Republic of Korea) is an optical low coherence reflectometry (OLCR)-based optical biometer that provides axial length and key anterior segment parameters. It integrates pupillometry and corneal analysis features for routine biometry workflows. However, published evidence on this device is very limited, and to the best of our knowledge, no paired study has directly compared it with the IOLMaster 700.
In addition, interest in axial length monitoring in children has increased markedly with the expansion of myopia management strategies and the need to detect small axial changes with high reliability [4,12,13,14]. Recent studies have shown that different optical biometers can be used in pediatric populations with good repeatability, but also that inter-device biases may exist, discouraging alternating devices during longitudinal follow-up [14,15]. Validating a general-purpose biometer such as the HBM-1 in a cohort that includes children allows its indication to be extended to myopia management programs, enabling the use of a single device for both phacorefractive surgery planning and refractive follow-up.
The aim of this study was to compare the intrasession repeatability of the HBM-1 and its agreement with the IOLMaster 700 for the main biometric parameters (AL, ACD, LT, CCT, WTW, and keratometry) in a mixed-age cohort assessed in a real-world clinical practice setting, including patients enrolled in myopia management programs.
2. Methods
2.1. Study Design and Participants
A cross-sectional observational study was conducted in a private optometry practice located within the facilities of Ophthalmology Clinic San Diego in Medellín (Colombia). The practice is independently operated and is not institutionally affiliated with the clinic. Eighty-two eyes from 82 consecutive patients requiring optical biometry for cataract surgery planning, pre- or postoperative refractive evaluation, or myopia follow-up were included. Inclusion criteria were sufficient ocular media clarity to allow AL acquisition using SS-OCT or OLCR, adequate fixation, and age of 5 years or older. Eyes were excluded if they had prior intraocular surgery (other than uncomplicated cataract surgery in the fellow eye), advanced corneal pathology, retinal pathology compromising fixation, or repeatedly poor measurement quality with either device.
For patients who had both eyes measured, only one eye (the right eye) was included to avoid statistical dependence. The study adhered to the principles of the Declaration of Helsinki, and all participants, or their legal guardians in the case of minors, provided written informed consent.
Each patient was informed about the characteristics and justification of the present study. All patients signed an informed consent form following the tenets of the Declaration of Helsinki. This research was approved by the Ethics Committee of the University of Alicante (No. UA-2021-01-14, and date of approval 14 January 2021).
2.2. Optical Biometers
The HBM-1 (Huvitz, Anyang, Republic of Korea) is an optical biometer based on optical low-coherence reflectometry (OLCR) that provides measurements of AL, ACD, LT, CCT, WTW, and dynamic pupillometry, fully integrated with a Placido-disc corneal topography system that provides corneal aberrometry and axial and tangential multi-annular keratometry. It includes automated alignment, internal fixation, and rapid acquisition of multiple examinations along the visual axis.
The IOLMaster 700 (Carl Zeiss Meditec, Jena, Germany) is a widely validated SS-OCT biometer that measures AL, ACD, LT, CCT, pupillometry, keratometry, total keratometry, and other anterior segment parameters and integrates with modern formulas for IOL power calculation [6,7,8,9,10]. In this study, it was used as the reference to evaluate the clinical agreement of the HBM-1.
2.3. Measurement Protocol
Following a comprehensive optometric and ophthalmological examination, biometric examinations were performed in separate rooms under controlled ambient illumination by a single experienced examiner, adhering to a standardized sequence and each device’s user manual. For each patient, three consecutive measurements were obtained with the HBM-1 and three with the IOLMaster 700. In approximately half of the cases, the device order was reversed to minimize order effects. Adequate blinking was ensured, and any measurement that did not meet a device’s built-in quality criteria was repeated. For inter-biometer agreement analysis, the average of the three measurements from each device was used. Axial length acquisition was successful in all included eyes with both devices (82/82; 100%).
2.4. Study Variables
The following parameters were analyzed: AL (mm), ACD (mm), LT (mm), CCT (µm), WTW (mm), flat and steep keratometry (Kf and Ks, D), mean keratometry (Avg K, D), and keratometric cylinder (Cyl, D). The astigmatism was further expressed using the vector components J0 and J45 (D). For consistency, all CCT values from the HBM-1 (reported in millimeters) were converted to micrometers (mm × 1000) to enable direct comparison with the IOLMaster 700 outputs. The J0 and J45 components were derived from cylinder power and axis using the Thibos vector transformation, which facilitates conventional parametric statistical analysis [16].
2.5. Statistical Analysis
Statistical analyses were performed using SPSS version 23 (IBM Corp., Armonk, NY, USA). The sample’s age was described using the mean, standard deviation, median, range, and 95% confidence interval. The intrasession repeatability of the HBM-1 was evaluated for each parameter using repeated-measures analysis of variance (ANOVA). From this model, the within-subject standard deviation (Sw) was derived as the square root of the within-subject error mean square (MSwithin) [6,11,17]. The repeatability coefficient (CR = 2.77 × Sw) was then calculated, which estimates the interval within which 95% of the differences between two repeated measurements of the same eye are expected to lie. Furthermore, the intraclass correlation coefficient (ICC) was computed using a two-way mixed-effects model for absolute agreement to quantify measurement reliability [6,11,18].
For comparisons between the HBM-1 and IOLMaster 700, the difference HBM-1 − IOLMaster 700 was calculated for each parameter and analyzed using a paired Student’s t test. Bland–Altman plots were constructed to visualize agreement and to estimate the mean bias with 95% limits of agreement (mean ± 1.96 × SD) [5,11,19,20,21]. Additionally, Pearson’s correlation coefficient was calculated to assess the relationship between patient age and each inter-device difference. A two-sided p value < 0.05 was defined as the threshold for statistical significance.
2.6. Sample Size and Power Calculation
An a priori sample size calculation was performed using G*Power software (version 3.1) for a two-tailed paired t-test (difference between two dependent means). The inter-device difference in AL was prespecified as the primary outcome for this estimation because AL is critical for IOL power calculation and myopia monitoring. Assuming a moderate standardized effect size (Cohen’s dz) of 0.5 (dz = Δ/SDdiff), which corresponds to detecting an inter-device AL bias of approximately 0.02 mm with an assumed standard deviation (SD) of the paired differences of 0.04 mm, a sample of 54 paired eyes was required for α = 0.05 (two-sided) and 95% power. The final sample included 82 paired eyes, exceeding this a priori requirement. A post hoc power calculation confirmed an achieved power of 0.994 for n = 82 under the same assumptions. Because standard sample size formulas do not directly apply to ICC and Bland–Altman agreement analyses, these metrics were not used for the a priori calculation; thus, any related subgroup analyses are considered exploratory.
3. Results
3.1. Sample Characteristics
Eighty-two cases were included, with no missing data for age. Age group distribution was: pediatric 22 (26.8%), young adult 29 (35.4%), and older adult 31 (37.8%). Mean ages of these subgroups ages (mean ± SD; min–max) were: pediatric 11.64 ± 3.84 (6–17), young adult 24.52 ± 6.67 (18–39), and older adult 61.65 ± 9.52 (41–79) (Table 1). Overall, the cohort covered a broad age range (6–79 years), consistent with a real-world clinical practice population.
Table 1.
Sex and age-group distribution. Abbreviations: SD, standard deviation; IQR, interquartile range; CI, confidence interval.
| Age Group | n (%) | Age, Mean ± SD (Min–Max) | Median (IQR) | 95% CI of the Mean |
|---|---|---|---|---|
| Pediatric | 22 (26.8%) | 11.64 ± 3.84 (6–17) | 11.00 (7) | 9.94–13.34 |
| Young adult | 29 (35.4%) | 24.52 ± 6.67 (18–39) | 25.00 (9) | 21.98–27.06 |
| Older adult | 31 (37.8%) | 61.65 ± 9.52 (41–79) | 64.00 (14) | 58.15–65.14 |
| Eye analyzed | One eye per patient | |||
| Female sex, n (%) | 46 (56.1%) | |||
| Male sex, n (%) | 36 (43.9%) | |||
| Total | 82 (100%) | |||
3.2. HBM-1 Intrasession Repeatability
Repeated-measures analyses demonstrated excellent intrasession repeatability of the HBM-1, with very low within-subject variability for all parameters (Table 2). Sw values were in the hundredths of a millimeter for AL, ACD, and LT; approximately 10–15 μm for CCT; and in the tenths of a diopter for keratometry. ICCs for individual measurements exceeded 0.95 for all parameters, ranging from 0.97 to 0.99 for AL, ACD, LT and CCT.
Table 2.
HBM-1 intrasession repeatability indices. Abbreviations: Sw, within-subject standard deviation; CoV, coefficient of variation; ICC, intraclass correlation coefficient; D, diopters; CI, confidence interval; AL, axial length; ACD, anterior chamber depth; LT, lens thickness; CCT, central corneal thickness; WTW, white-to-white corneal diameter; Kf and Ks, flat and steep keratometry; Avg K, mean keratometry; Cyl, keratometric cylinder; J0 and J45, astigmatism vector components.
| Parameter | Unit | HBM-1 Mean | Sw | CoV (%) | Repeatability (±2.77 × Sw) | ICC (Single Measure), 95% CI |
|---|---|---|---|---|---|---|
| AL | mm | 24.8918 | 0.063 | 0.25 | ±0.17 | 0.999 (0.997–0.999) |
| ACD | mm | 3.5573 | 0.009 | 0.25 | ±0.03 | 0.999 (0.999–1.000) |
| LT | mm | 3.8855 | 0.012 | 0.31 | ±0.03 | 0.999 (0.998–1.000) |
| CCT | µm | 545.37 | 1.430 | 0.26 | ±3.96 | 0.998 (0.995–0.999) |
| WTW | mm | 12.1435 | 0.032 | 0.26 | ±0.09 | 0.993 (0.984–0.996) |
| Cyl | D | −1.6432 | 0.005 | 0.31 | ±0.01 | 1.000 (≈1.000–1.000) |
| Avg K | D | 43.6867 | 0.118 | 0.27 | ±0.33 | 0.997 (0.993–0.998) |
| Kf (K1) | D | 42.8856 | 0.118 | 0.28 | ±0.33 | 0.997 (0.994–0.998) |
| Ks (K2) | D | 44.4880 | 0.120 | 0.27 | ±0.33 | 0.997 (0.994–0.998) |
| J0 | D | 0.5789 | 0.032 | — * | ±0.09 | 0.997 (0.994–0.998) |
| J45 | D | −0.0016 | 0.071 | — * | ±0.20 | 0.974 (0.955–0.985) |
* CoV is not reported for J0/J45 because these are vector components with a mean potentially close to zero and positive/negative sign, making the CoV poorly interpretable.
Table 2 summarizes the HBM-1 intrasession repeatability indices (Sw, CR, and ICC for each variable).
3.3. Differences Between HBM-1 and IOLMaster 700
The mean values, standard deviations, mean inter-device difference (Δ = HBM-1 − IOLMaster 700), range of differences, and mean absolute difference for all parameters are presented in Table 3. Bland–Altman plots illustrating the agreement for each parameter are shown in Figure 1, with the corresponding mean bias and 95% limits of agreement. Paired-t tests revealed statistically significant inter-device mean differences for AL (p = 0.045), ACD (p < 0.001), and LT (p < 0.001). In contrast, no statistically significant differences were observed for CCT, WTW, corneal cylinder, mean keratometry, K1, K2, or the astigmatism vectors (J0 and J45) (Table 3).
Table 3.
Comparison between HBM-1 and IOLMaster 700 (n = 82 pairs). Values are reported as mean ± SD. Δ = HBM-1 − IOLMaster 700. Range of Δ corresponds to the observed minimum and maximum inter-device differences. |Δ| is the mean absolute difference. Paired-samples t test was used. CI, confidence interval.
| Parameter | Unit | HBM-1 (Mean ± SD) | IOLMaster 700 (Mean ± SD) | Δ (Mean ± SD) | Range of Δ (Min to Max) | |Δ| (Mean ± SD) | 95% CI for Δ | t | df | p | r |
|---|---|---|---|---|---|---|---|---|---|---|---|
| AL | mm | 24.8918 ± 2.29622 | 24.8802 ± 2.28493 | 0.01159 ± 0.05158 | −0.10 to 0.13 | 0.0421 ± 0.03169 | 0.00025 to 0.02292 | 2.034 | 81 | 0.045 | 1.000 |
| ACD | mm | 3.5573 ± 0.43531 | 3.4980 ± 0.45136 | 0.05927 ± 0.06769 | −0.10 to 0.30 | 0.0734 ± 0.05179 | 0.04439 to 0.07414 | 7.929 | 81 | <0.001 | 0.989 |
| LT | mm | 3.8855 ± 0.50285 | 3.8340 ± 0.54898 | 0.05146 ± 0.08982 | −0.11 to 0.48 | 0.0720 ± 0.07421 | 0.03173 to 0.07120 | 5.189 | 81 | <0.001 | 0.989 |
| CCT | µm | 545.37 ± 36.876 | 544.55 ± 37.757 | 0.817 ± 7.115 | −11 to 12 | 6.33 ± 3.28 | −0.746 to 2.380 | 1.040 | 81 | 0.301 | 0.982 |
| WTW | mm | 12.1435 ± 0.45609 | 12.1293 ± 0.40077 | 0.01427 ± 0.24334 | −0.86 to 0.78 | 0.1935 ± 0.16856 | −0.03920 to 0.06774 | 0.531 | 81 | 0.597 | 0.846 |
| Cyl | D | −1.6432 ± 1.09891 | −1.7127 ± 1.22728 | 0.06951 ± 0.39776 | −1.08 to 1.99 | 0.2734 ± 0.29568 | −0.01789 to 0.15691 | 1.583 | 81 | 0.117 | 0.947 |
| Avg K | D | 43.6867 ± 2.54609 | 43.6401 ± 2.58907 | 0.04659 ± 0.40451 | −1.05 to 1.20 | 0.2661 ± 0.30683 | −0.04230 to 0.13547 | 1.043 | 81 | 0.300 | 0.988 |
| K1 | D | 42.8856 ± 2.59744 | 42.8333 ± 2.60803 | 0.05232 ± 0.39573 | −1.22 to 1.31 | 0.2706 ± 0.29196 | −0.03463 to 0.13927 | 1.197 | 81 | 0.235 | 0.988 |
| K2 | D | 44.5288 ± 2.61514 | 44.5283 ± 2.72344 | 0.00049 ± 0.47964 | −1.65 to 1.43 | 0.3071 ± 0.36687 | −0.10490 to 0.10588 | 0.009 | 81 | 0.993 | 0.985 |
| J0 | D | 0.5789 ± 0.64770 | 0.6081 ± 0.65497 | −0.02918 ± 0.15366 | −0.50 to 0.39 | 0.1192 ± 0.10046 | −0.06294 to 0.00458 | −1.720 | 81 | 0.089 | 0.972 |
| J45 | D | −0.0016 ± 0.47588 | 0.0221 ± 0.56136 | −0.02370 ± 0.19655 | −0.44 to 1.00 | 0.1350 ± 0.14407 | −0.06689 to 0.01948 | −1.092 | 81 | 0.278 | 0.941 |
Figure 1.
Bland–Altman plots of the agreement between the HBM-1 and the IOLMaster 700 for the different biometric parameters: AL, axial length (A); ACD, anterior chamber depth (B); LT, lens thickness (C); CCT, central corneal thickness (D); WTW, white-to-white corneal diameter (E); K1, flat keratometry (Kf) (F); K2, steep keratometry (Ks) (G); Avg K, mean keratometry (H); Cyl, corneal cylinder (I); J0 (J) and J45 (K). The central horizontal line represents the mean difference (bias) and the upper and lower lines represent the 95% limits of agreement; values are annotated on each panel.
3.4. Correlation Between Age and Inter-Device Differences
Pearson correlations between patient age and inter-device differences for each biometric parameter are shown in Table 4 and visually in Figure 2. Age was not significantly associated with differences in axial length (Diff_AL: r = −0.162; p = 0.145; Figure 2A), keratometric cylinder (Diff_Cyl: r = −0.175; p = 0.116; Figure 2I), Avg K (Diff_AvgK: r = 0.101; p = 0.368; Figure 2H), or the astigmatism vector component J0 and J45 (Diff_J0: r = 0.126; p = 0.259; Figure 2J; Diff_J45: r = 0.023; p = 0.835; Figure 2K). A weak positive correlation was observed between age and the ACD difference (Diff_ACD: r = 0.284; p = 0.010; Figure 2B). Conversely, age demonstrated moderate negative correlations with differences in LT, CCT, and WTW diameter (Diff_LT: r = −0.547; p < 0.001; Figure 2C; Diff_CCT: r = −0.432; p < 0.001; Figure 2D; Diff_WTW: r = −0.371; p = 0.001; Figure 2E). For keratometry, a weak positive correlation was found for K1 (flat) (Diff_K1: r = 0.236; p = 0.033; Figure 2F), while no significant association was observed for K2 (steep) (Diff_K2: r = 0.054; p = 0.628; Figure 2G).
Table 4.
Pearson correlations between age and inter-device differences. Abbreviations: AL, axial length; ACD, anterior chamber depth; LT, lens thickness; CCT, central corneal thickness; WTW, white-to-white corneal diameter; Kf and Ks, flat and steep keratometry; Avg K, mean keratometry; Cyl, keratometric cylinder; J0 and J45, astigmatism vector components; D, diopters; r, coefficient of correlation.
| Parameter (HBM − IOL Difference) | Unit | r (Age, Difference) | p (Two-Tailed) |
|---|---|---|---|
| Diff_AL | mm | −0.162 | 0.145 |
| Diff_ACD | mm | 0.284 | 0.010 |
| Diff_LT | mm | −0.547 | <0.001 |
| Diff_CCT | µm | −0.432 | <0.001 |
| Diff_WTW | mm | −0.371 | 0.001 |
| Diff_Cyl | D | −0.175 | 0.116 |
| Diff_K1 | D | 0.054 | 0.628 |
| Diff_K2 | D | 0.236 | 0.033 |
| Diff_AvgK | D | 0.101 | 0.368 |
| Diff_J0 | D | 0.126 | 0.259 |
Figure 2.
Relationship between age and inter-device differences (HBM-IOL) for the different biometric parameters: AL, axial length (A); ACD, anterior chamber depth (B); LT, lens thickness (C); CCT, central corneal thickness (D); WTW, white-to-white corneal diameter (E); K1, flat keratometry (Kf) (F); K2, steep keratometry (Ks) (G); Avg K, mean keratometry (H); Cyl, corneal cylinder (I); J0 (J) and J45 (K). Each panel represents a specific parameter, with the fitted linear regression line and coefficient of determination (R2).
4. Discussion
4.1. Main Interpretation
This study demonstrates that the HBM-1 exhibits excellent intrasession repeatability and high agreement with the IOLMaster 700 in a cohort representative of clinical practice, including a wide age range (6–79 years), supporting the clinical applicability of the findings in both pediatric and adult populations; however, age subgroup results should be interpreted in light of the sample size within each category. Mean bias for axial length was only 0.0116 mm and limits of agreement were approximately ±0.10 mm, values considered clinically acceptable for IOL power calculation and axial length monitoring [4,5,6,8,11,14,21].
Importantly, a small mean bias at the cohort level can mask larger individual discrepancies. The range of observed differences and the mean absolute difference (|Δ|) presented in Table 3 provide a more informative view of dispersion and help identify parameters with sporadic outliers (e.g., ACD differences up to 0.30 mm, Cyl up to 1.99 D, and CCT up to 12 µm). In older adults, greater variability may be attributable to factors such as reduced optical quality, tear film instability, fixation variability, or lens opacity. Consequently, direct interchangeability of the two devices should be approached with caution, particularly when measurements fall near clinical decision thresholds, even in the context of negligible overall mean bias.
The statistically significant inter-device differences observed for ACD and LT, while methodologically relevant, are unlikely to be clinically significant for IOL power calculation when using fifth-generation formulas [3], especially if the same device is used consistently for preoperative planning. However, in eyes with extreme biometric values or in candidates for premium IOLs, it would be prudent to verify the consistency of biometric data and consider employing multiple formulas if substantial inter-device discrepancies arise [3,22,23].
As shown in Table 4 and Figure 2, inter-device differences were not age-invariant for all parameters. Moderate negative correlations were found between age and the differences for LT (r = −0.547, p < 0.001), CCT (r = −0.432, p < 0.001), and WTW (r = −0.371, p = 0.001), indicating a tendency for the HBM-1 to yield progressively lower values relative to the IOLMaster 700 in older individuals. Clinically, this age-dependent variation suggests that device interchangeability is not uniform across the lifespan, particularly for parameters reliant on boundary detection or image contrast (e.g., LT, WTW). Potential contributors include age-related changes in crystalline lens morphology affecting LT segmentation, reduced limbal contrast influencing WTW estimation, and inherent algorithmic differences in pachymetry measurements. Therefore, even with modest cohort-level mean differences, clinicians should avoid alternating devices when serially monitoring age-sensitive parameters, especially in older adults or when measurements approach critical clinical thresholds.
4.2. Comparison with Literature
Our results are consistent with the broader literature comparing SS-OCT, PCI, and OLCR biometers, which typically report small, clinically acceptable AL biases and narrow limits of agreement [5,6,7,8,18,19,20,24]. More recent large-scale and multi-device comparative studies have confirmed similar agreement for corneal and anterior segment parameters, although somewhat larger discrepancies have been noted for variables such as WTW or corneal cylinder, particularly when measurements involve different keratometry technologies or segmentation algorithms [11,21,25,26,27,28].
The high repeatability (low within-session variability and high ICCs) of SS-OCT biometers is well-established in both adult and cataract populations [9,10,17]. In pediatric cohorts, studies indicate good feasibility and repeatability but emphasize the importance of using the same device for longitudinal myopia monitoring [4,12,13,14]. Our findings contribute to this literature by demonstrating that a widely available clinical biometer, theHBM-1, achieves comparable levels of repeatability and agreement with the IOLMaster 700, even within a pediatric subgroup.
4.3. Implications for IOL Power Calculation
In IOL power calculation, a 0.1 mm difference in AL can induce a refractive error of approximately 0.25–0.30 D in eyes of average length [1,3]. Consequently, it is clinically desirable for the 95% limits of agreement between biometers to remain within ±0.10 mm. Our results confirm that the HBM-1 meets this criterion against the IOLMaster 700, supporting its use for planning both monofocal and premium IOLs. While the observed differences in ACD and LT are small, they could have a greater impact in formulas that rely heavily on effective lens position (ELP) estimation, particularly in eyes with extreme biometrics. In such cases, cross-checking calculations with multiple formulas is advisable [3].
4.4. Implications for Myopia Control
The cross-sectional agreement in AL between the HBM-1 and IOLMaster 700, coupled with its excellent intrasession repeatability in pediatric eyes, supports the potential of the HBM-1 for AL monitoring in myopia management, provided the same device is used for follow-up [4,13,14]. However, as this was a single-visit study, our findings establish device comparability at a single timepoint rather than validating the tracking of axial elongation over time. Prospective longitudinal studies with repeated measurements over intervals relevant to myopia progression (e.g., 6–12 months) are required to confirm inter-device agreement for change in AL over time and to define the smallest detectable change for myopia control applications [12]. Until such longitudinal validation is available, alternating devices during AL follow-up for the same patient is not recommended, in order to avoid interpreting device-related changes as true progression [4,12,14,15].
4.5. Strengths and Limitations
The strengths of this study include its rigorous methodology and broad clinical applicability. Key features are the acquisition of repeated measurements per device, the inclusion of a wide age spectrum (6–79 years) encompassing a pediatric subgroup, and the comprehensive evaluation of biometric parameters relevant to both cataract surgery and myopia management. Furthermore, the analysis incorporated robust assessments of intrasession repeatability (Sw, CR, ICC) and inter-device agreement (paired t-tests, Bland–Altman analysis, and correlation with patient age) [1,3,4,6,11]. This study has several limitations. First, its cross-sectional design precludes longitudinal assessment of AL change. Second, while the sample size is moderate and comparable to other biometer validation studies, the single-center setting may limit the generalizability of the findings. Third, although the cohort was stratified by age, other clinically relevant covariates—such as refractive status, extreme axial lengths (e.g., <22 mm or >26 mm), and lens opacity density—were not prespecified for subgroup analysis. Future studies should incorporate these variables, ideally including standardized lens opacity grading (e.g., LOCS III) and device signal-quality metrics, and should be adequately powered to evaluate extreme-eye subgroups.
It is also important to note that the study was not statistically powered for stratified comparisons. Consequently, subgroup analyses involving small samples should be interpreted with caution. The precision of agreement metrics (e.g., Bland–Altman limits of agreement, ICC confidence intervals) is dependent on sample size and measurement variability. For the overall cohort (n = 82), the observed SD of AL differences (0.0516 mm) yields a 95% confidence interval half-width of approximately 0.02 mm around each limit of agreement endpoint (calculated as 1.96 × √3 × SDdiff/√n), indicating reasonably precise estimates for AL agreement. However, precision is reduced for parameters with greater variability and within smaller subgroups.
5. Conclusions
The HBM-1 optical biometer demonstrated excellent intrasession repeatability and strong clinical agreement with the IOLMaster 700 for axial length and key anterior segment parameters in a cohort spanning adults and children. Mean inter-device differences were small, and the 95% limits of agreement for axial length fell within the clinically acceptable range for modern IOL power calculation and for cross-sectional monitoring in myopia management. While these findings support the use of the HBM-1 as a reliable alternative to the IOLMaster 700 in routine clinical practice, consistent use of the same device for individual patient follow-up is essential. Future studies are needed to confirm device interchangeability in eyes with extreme biometrics and to validate its performance for tracking axial elongation over time.
Acknowledgments
The authors thank the patients and their families for participating in the study.
Author Contributions
Conceptualization, F.J.B. and D.P.P.; methodology, F.J.B. and D.P.P.; validation, F.J.B. and D.P.P.; formal analysis, F.J.B. and D.P.P.; investigation, F.J.B. and D.P.P.; resources, F.J.B. and D.P.P.; data curation, F.J.B.; writing—original draft preparation, F.J.B.; writing—review and editing, F.J.B. and D.P.P.; visualization, F.J.B. and D.P.P.; supervision, D.P.P.; project administration, F.J.B. and D.P.P.; funding acquisition, F.J.B. and D.P.P. All authors have read and agreed to the published version of the manuscript.
Institutional Review Board Statement
The study was conducted in accordance with the Declaration of Helsinki, and approved by the Ethics Committee of the University of Alicante (protocol code No. UA-2021-01-14 and date of approval 14 January 2021).
Informed Consent Statement
Informed consent was obtained from all subjects involved in the study.
Data Availability Statement
The data that support the findings of this study are available from the corresponding author, DPP, upon reasonable request.
Conflicts of Interest
Author Farid J. Bedrán García serves as a Key Opinion Leader (KOL) for US Ophthalmic (a distributor of ophthalmic medical equipment) for the Huvitz product line, including the HBM-1 biometer evaluated in this study. None of the other authors declare conflicts of interest related to the content of this work. No generative artificial intelligence has been used for the preparation of this document.
Funding Statement
This research received no external funding.
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
References
- 1.Nazm N., Chakrabarti A. Update on optical biometry and intraocular lens power calculation. TNOA J. Ophthalmic Sci. Res. 2017;55:196–210. doi: 10.4103/tjosr.tjosr_44_17. [DOI] [Google Scholar]
- 2.Montés-Micó R., Pastor-Pascual F., Ruiz-Mesa R., Tañá-Rivero P. Ocular biometry with swept-source optical coherence tomography. J. Cataract Refract. Surg. 2021;47:802–814. doi: 10.1097/j.jcrs.0000000000000551. [DOI] [PubMed] [Google Scholar]
- 3.Kane J.X., Chang D.F. Intraocular Lens Power Formulas, Biometry, and Intraoperative Aberrometry: A Review. Ophthalmology. 2021;128:e94–e114. doi: 10.1016/j.ophtha.2020.08.010. [DOI] [PubMed] [Google Scholar]
- 4.Jones D., Chow A., Fadel D., Meijome J.M.G., Grzybowski A., Kollbaum P., Loughman J., Wolffsohn J. IMI—Instrumentation for Myopia Management. Investig. Ophthalmol. Vis. Sci. 2025;66:7. doi: 10.1167/iovs.66.9.7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Arriola-Villalobos P., Almendral-Gómez J., Garzón N., Ruiz-Medrano J., Fernández-Pérez C., Martínez-De-La-Casa J.M., Díaz-Valle D. Agreement and clinical comparison between a new swept-source optical coherence tomography-based optical biometer and an optical low-coherence reflectometry biometer. Eye. 2016;31:437–442. doi: 10.1038/eye.2016.241. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Kunert K.S., Peter M., Blum M., Haigis W., Sekundo W., Schütze J., Büehren T. Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry. J. Cataract Refract. Surg. 2016;42:76–83. doi: 10.1016/j.jcrs.2015.07.039. [DOI] [PubMed] [Google Scholar]
- 7.Akman A., Asena L., Güngör S.G. Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500. Br. J. Ophthalmol. 2016;100:1201–1205. doi: 10.1136/bjophthalmol-2015-307779. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Kanclerz P., Hoffer K.J., Przewłócka K., Savini G. Comparison of an upgraded optical biometer with 2 validated optical biometers. J. Cataract Refract. Surg. 2021;47:859–864. doi: 10.1097/j.jcrs.0000000000000541. [DOI] [PubMed] [Google Scholar]
- 9.Michael R., Wirkner K., Engel C., Loeffler M., Kirsten T., Rauscher F.G. Feasibility and repeatability of ocular biometry measured with IOLMaster 700 in a large population-based study. Ophthalmic Physiol. Opt. 2023;43:860–873. doi: 10.1111/opo.13148. [DOI] [PubMed] [Google Scholar]
- 10.Langenbucher A., Szentmáry N., Cayless A., Hoffmann P., Wendelstein J., Cooke D. Repeatability of biometric measures from the IOLMaster 700 in a cataractous population. PLoS ONE. 2024;19:e0297869. doi: 10.1371/journal.pone.0297869. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Alberquilla I.M., Svensson S., Ruiz-Alcocer J., Madrid-Costa D., Dominguez-Vicent A., Venkataraman A.P. Evaluation of repeatability and agreement of two optical biometers for intraocular lens power calculation. Sci. Rep. 2024;14:22151. doi: 10.1038/s41598-024-73206-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Chow A.Y.H., Mungalsingh M.A., Thai D., Selimos Z., Yan S.K., Xu H., Jones D.A. Suitability of multifunction devices Myah and Myopia Master for monitoring myopia progression in children and adults. Ophthalmic Physiol. Opt. 2024;44:1017–1030. doi: 10.1111/opo.13332. [DOI] [PubMed] [Google Scholar]
- 13.Martínez-Plaza E., Molina-Martín A., Arias-Puente A., Piñero D.P. Clinical validation of a new optical biometer for myopia control in a healthy pediatric population. Children. 2022;9:1713. doi: 10.3390/children9111713. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Mattern A.I., Neller K., Devenijn M., Schwahn H., Langenbucher A., Seitz B., Kaymak H. A comparison of optical biometers used in children for myopia control. Klin. Monatsblätter Augenheilkd. 2023;240:1306–1313. doi: 10.1055/a-2117-9335. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Garcia Ardoy E., Mateos N., Roda L., Torrado Sierra O., Baptista A.M., Serra P.M. Repeatability and agreement of swept-source optical coherence tomography and partial coherence interferometry biometers in myopes. Clin. Exp. Optom. 2023;106:783–792. doi: 10.1080/08164622.2022.2152312. [DOI] [PubMed] [Google Scholar]
- 16.Thibos L.N., Horner D. Power vector analysis of the optical outcome of refractive surgery. J. Cataract Refract. Surg. 2001;27:80–85. doi: 10.1016/S0886-3350(00)00797-5. [DOI] [PubMed] [Google Scholar]
- 17.Ferrer-Blasco T., Domínguez-Vicent A., Esteve-Taboada J.J., Aloy M.A., Adsuara J.E., Montés-Micó R. Evaluation of the repeatability of a swept-source ocular biometer for measuring ocular biometric parameters. Graefes Arch. Clin. Exp. Ophthalmol. 2017;255:343–349. doi: 10.1007/s00417-016-3555-z. [DOI] [PubMed] [Google Scholar]
- 18.Wylęgała A., Bolek B., Mazur R., Wylęgała E. Repeatability, reproducibility, and comparison of ocular biometry using a new optical coherence tomography-based system and another device. Sci. Rep. 2020;10:14440. doi: 10.1038/s41598-020-71192-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Yang C.M., Lim D.H., Kim H.J., Chung T.Y. Comparison of two swept-source optical coherence tomography biometers and a partial coherence interferometer. PLoS ONE. 2019;14:e0223114. doi: 10.1371/journal.pone.0223114. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Zarei-Ghanavati S., Nikpayam M., Namdari M., Bakhtiari E., Hassanzadeh S., Ziaei M. Agreement between a spectral-domain ocular coherence tomography biometer with a swept-source ocular coherence tomography biometer and an optical low-coherence reflectometry biometer in eyes with cataract. J. Curr. Ophthalmol. 2023;35:153–158. doi: 10.4103/joco.joco_30_23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Lender R., Mirsky D., Greenberger R., Boim Z., Ben-Yaakov L., Kashtan C., Naffar I., Shine S., Chowers I., Ben-Eli H. Evaluation of three biometric devices: Ocular parameters and calculated intraocular lens power. Sci. Rep. 2022;12:19478. doi: 10.1038/s41598-022-24017-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Lee C.Y., Yang S.F., Chen H.C., Lian I.B., Huang J.Y., Chang C.K. The refractive accuracy between topographic keratometry and biometric keratometry for extended depth-of-focus intraocular lens implantation. Int. J. Med. Sci. 2024;21:2912–2918. doi: 10.7150/ijms.99907. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Multack S., Plummer N., Smits G., Hall B. Randomized trial comparing prediction accuracy of two swept source optical coherence tomography biometers. Clin. Ophthalmol. 2023;17:2423–2428. doi: 10.2147/OPTH.S407538. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Cho Y.J., Lim T.H., Choi K.Y., Cho B.J. Comparison of ocular biometry using new swept-source optical coherence tomography-based optical biometer with other devices. Korean J. Ophthalmol. 2018;32:257–264. doi: 10.3341/kjo.2017.0091. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Ruan X., Yang G.Y., Xia Z., Zhang J., Gu X., Tan Y., Liu Z., Luo L. Agreement of anterior segment parameter measurements with CASIA 2 and IOLMaster 700. Front. Med. 2022;9:777443. doi: 10.3389/fmed.2022.777443. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Omoto M.K., Torii H., Masui S., Ayaki M., Tsubota K., Negishi K. Ocular biometry and refractive outcomes using two swept-source optical coherence tomography-based biometers with segmental or equivalent refractive indices. Sci. Rep. 2019;9:6557. doi: 10.1038/s41598-019-42968-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Porwolik M., Porwolik A., Mrukwa-Kominek E. Evaluation of selected biometric parameters in cataract patients—A comparison between Argos® and IOLMaster 700®: Two swept-source optical coherence tomography-based biometers. Medicina. 2024;60:1057. doi: 10.3390/medicina60071057. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Románek J., Sluková K. Comparison of optical biometers ARGOS and IOL Master 700. Ceska Slov. Oftalmol. 2021;77:296–300. doi: 10.31348/2021/35. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data that support the findings of this study are available from the corresponding author, DPP, upon reasonable request.