Abstract
Introduction
Three-dimensional (3D) computed tomography planning improves preoperative quantification of glenoid morphology when planning total shoulder arthroplasty, but measurement reliability across commonly used planning platforms remains uncertain.
Methods
A retrospective within-subject comparison of 30 shoulders was planned on four systems: Stryker Blueprint (BP), Zimmer Signature (ZS), Exactech Equinoxe (EE), and SurgiCase Materialise (MS). Measurements of retroversion and inclination were analysed to evaluate measurement variability and correlations between systems.
Results
Retroversion medians (IQR) were BP −6.5 (10.75), ZS −6.0 (7.0), EE −5.5 (7.0), and MS −3.5 (19.5); overall p=0.37. MS showed greater dispersion and weak correlations with other systems (ρ≈0.03-0.07). Inclination means (SD) were BP 5.93 (8.13), ZS 1.28 (7.27), EE 3.53 (7.57), MS 6.00 (5.07); overall p<0.001. All paired contrasts were significant except for BP vs. MS (−0.07, 95% CI −2.14 to 2.01; p=0.948). Inter-system inclination correlations were strong among BP, EE, and ZS (r≈0.80-0.90) and moderate to strong for MS vs. others (r≈0.57-0.75).
Conclusion
Platforms were not fully interchangeable. Retroversion showed no group-level bias but reduced precision, particularly for MS, while inclination was highly precise yet displayed platform-specific level differences (notably lower with ZS). Based on this, threshold-adjacent decisions (augment selection, baseplate tilt) may vary by platform. Verification of retroversion outliers and confirmation of inclination near boundaries are recommended.
Keywords: glenoid morphology, glenoid version, measurement reliability, shoulder arthroplasty/ replacement, three-dimensional preoperative planning, total shoulder arthroplasty
Introduction
Total shoulder arthroplasty is an established long-term treatment for glenohumeral arthritis and rotator cuff disease, with 9,809 shoulder replacements performed in the United Kingdom in 2024 alone [1-4]. As the population continues to age, utilisation is projected to rise at a faster rate than for hip or knee arthroplasty [5]. In this context, minimising complications, revision surgery, and length of stay is essential to limit the burden upon healthcare systems.
Accurate and reproducible planning is therefore central to this aim [6-8]. In particular, robust characterisation of glenoid orientation - the geometric relationship between the glenoid cavity and the scapular plane - underpins the correction of bone loss, implant/augment selection, and restoration of joint biomechanics. This assessment is frequently complicated by asymmetric wear, osteophytes, and cavitary or segmental bone loss that distort native landmarks and the scapular reference frame [9]. Although debate persists regarding the ideal component positioning (version and inclination), appropriate positioning is essential for balancing forces across the prosthetic joint and ensuring the long-term success of shoulder arthroplasty [10]. Conversely, suboptimal component positioning and excessive glenoid reaming (loss of bone stock) can affect longevity via scapular notching, eccentric loading, instability, glenoid loosening, and eventual failure [7,8,11-13].
Planning has evolved significantly from plain radiography to the introduction of two-dimensional (2D) computed tomography (CT) and, more recently, to three-dimensional (3D) CT-based platforms. While 2D CT improved visualisation of bony anatomy, measurements are sensitive to slice orientation and uncorrected scapular rotation, introducing variability [14,15]. Multiple contemporary 3D planning software programs have been introduced, allowing reorientation of the scapula plane and modelling of bone loss, demonstrating greater accuracy and reproducibility for native glenoid morphology than 2D methods [9,16]. This enhanced anatomical characterisation supports more consistent planning, more appropriate implant selection, and reduced intra-operative variability [17-19]. Alongside improvements in the understanding of arthroplasty biomechanics, the increase in planning accuracy has seen improvements in surgical outcomes, particularly early implant failure [11,12,20,21]. However, despite widespread adoption, limited information exists regarding the measurement variability between commonly used systems, raising practical questions about reliability and interchangeability between platforms [22,23]. Accordingly, this study aims to evaluate and compare four widely available 3D CT-guided templating platforms for their agreement on native glenoid retroversion and inclination measurements. By doing so, we aim to provide surgeons with practical insight into the degree of agreement and interchangeability between these systems for shoulder arthroplasty planning.
Materials and methods
A retrospective review was conducted of 30 consecutive patients who underwent pre-operative CT in preparation for shoulder arthroplasty. All studies were performed at the same institution using a standardised protocol, with thin-slice imaging of the entire scapula to ensure comprehensive visualisation and consistent scapular-plane referencing.
Each shoulder was planned on four commonly used pre-operative planning systems to obtain measurements of native glenoid version and inclination. Three of the systems employ automated, manufacturer-developed preoperative planning solutions: Stryker's Blueprint 3D planning (BP), Zimmer Signature ONE Surgical Planning System (ZS), and Materialise SurgiCase by Lima-Enovis (MS) (Figure 1).
Figure 1. Planning software systems used: (a) Blueprint 3D by Stryker – BP, (b) Zimmer Signature – ZS, (c) Exactech Equinoxe GPS – EE, (d) Materialise SurgiCase – MS.
BP and ZS derive scapular planes and native orientation using proprietary automated algorithms that collect thousands of data points on the scapula body, calculating the scapula planes according to those boundary median points. These automated vector calculations of the model are verified by technical engineers. MS generates automated measurements and analysis, with a human post-processing review. The authors do not have access to the underlying algorithm used to determine the scapula plane and glenoid morphology for these systems. The final platform utilised was Exactech Equinoxe GPS (EE), a surgeon-directed system. The surgeon uploads the CT, selects reference points, including the glenoid centre, trigonum, and glenoid limits, and adjusts the Friedman axis (glenoid centre to the medial scapular border). The resulting 3D reconstruction is then confirmed by the surgeon, from which implant selection can then be performed.
Statistical analysis
Distributional assumptions were assessed with the Shapiro-Wilk test. Retroversion deviated from normality, and therefore, the Friedman's test for the omnibus repeated-measures comparison was used, and where applicable, Wilcoxon signed-rank tests for within-subject pairwise contrasts. For inclination, normality was verified with the Shapiro-Wilk test, and sphericity was acceptable on Mauchly’s W test. Therefore, a repeated-measures ANOVA (rANOVA) was used for the omnibus comparison across systems. When the omnibus test was significant, within-subject paired t-tests quantified pairwise differences. Where adjustment for multiplicity was desired, the Holm correction was applied. Agreement between systems for inclination was quantified with Pearson’s correlation (r). Correlation was judged as very strong from 1 to 0.9, strong from 0.9 to 0.7, moderate from 0.7 to 0.5, low from 0.5 to 0.3, and poor from 0.3 to 0. The alpha risk was set to 0.05. All analyses used EasyMedStat (version 3.44; www.easymedstat.com).
Results
A total of 30 shoulders were analysed (nine male, 21 female), at a mean of 73 (range, 45-88). The right shoulders comprised 10 patients, with the remaining 20 left-sided. Median retroversion (IQR) was −6.5 (10.75) for BP, −6.0 (7.0) for ZS, −5.5 (7.0) for EE, and −3.5 (19.5) for MS (Table 1). The repeated-measures omnibus (Friedman) test showed no overall difference among systems (χ² = 3.14, p = 0.37). Observed median paired differences were small, and in keeping with the non-significant omnibus, inferential post-hoc testing was not interpreted. Distributions are shown in Figure 2, with paired analyses presented in Table 2 for completeness.
Table 1. Measurements of retroversion and inclination for the four software systems.
| Mean | Median | Interquartile Range (Q1,3) | Standard Deviation (SD) | 95% Confidence Interval (CI) | |
| Retroversion | |||||
| Blueprint | -6.33 | -6.50 | -12.75, -2.00 | 9.64 | -9.93, -2.73 |
| Exactec | -5.83 | -5.50 | -10.00, -3.00 | 9.15 | -9.25, -2.42 |
| Zimmer | -5.10 | -6.00 | -9.25, -2.25 | 9.58 | -8.68, -1.52 |
| Materialise | -0.73 | -3.50 | -9.75, 9.75 | 11.62 | -5.07, 3.61 |
| Inclination | |||||
| Blueprint | 5.93 | 7 | -1.00, 12.00 | 8.13 | 2.90, 8.97 |
| Exactec | 3.53 | 3 | -1.75, 9.75 | 7.57 | 0.71, 6.36 |
| Zimmer | 1.28 | 2 | -4.12, 6.38 | 7.27 | -1.43, 4.00 |
| Materialise | 6 | 6 | 4.90, 9.75 | 5.07 | 4.11, 7.89 |
Table 2. Groupwise comparison of retroversion and inclination measurements.
| Native Retroversion (Mean Paired Difference, Degrees) | p-value | Native Inclination (Median Paired Difference, Degrees) | p-value | |
| Blueprint vs. Materialise | 0.00 | 0.361 | 0.07 | 0.948 |
| Exactec vs. Materialise | -1.50 | 0.492 | 2.47 | 0.039 |
| Zimmer vs. Materialise | -2.50 | 0.948 | 4.72 | <0.001 |
| Blueprint vs. Zimmer | -0.50 | 0.068 | -4.65 | <0.001 |
| Blueprint vs. Exactec | 0.00 | 0.427 | 2.40 | 0.009 |
| Zimmer vs. Exactec | 0.50 | 0.186 | -2.25 | 0.013 |
Figure 2. Distribution of (A) retroversion and (B) inclination across systems (box & whisker).
Mean inclination (SD) measured by BP was 5.93 (8.13), EE 3.53 (7.57), ZS 1.28 (7.27), and MS 6.00 (5.07). Repeated-measures ANOVA demonstrated a significant main effect of system (p<0.001). Pre-specified within-subject post-hoc paired t-testing showed that BP yielded a greater inclination than ZS by 4.65 (95% CI 3.33 to 5.97; p<0.001) and greater than EE by 2.40 (95% CI 0.64 to 4.16; p=0.009) with no difference from MS (−0.07; 95% CI −2.14 to 2.01; p=0.948). ZS produced a lower inclination than EE by 2.25 (95% CI −3.99 to −0.51; p=0.013) and lower than MS by 4.72 (95% CI −6.51 to −2.92; p<0.001). MS exceeded EE by 2.47 (95% CI 0.13 to 4.80; p=0.039). Findings were unchanged when applying a Holm adjustment across the six paired comparisons. These contrasts indicate systematically lower inclination estimates with ZS compared with BP and MS (and lower than EE), with BP and MS similar and EE intermediate. The inclination distribution is shown in Figure 2, and full pairwise results are provided in Table 2.
Correlations
For retroversion (Spearman ρ), inter-system agreement was strong among three platforms: BP vs. ZS (ρ=0.91; r²=0.89), BP vs. EE (ρ=0.84; r²=0.82), and ZS vs. EE (ρ=0.89; r²=0.90); all p<0.001 (Table 3). In contrast, correlations between MS and each of the other platforms were weak (ρ≈0.03-0.07; r²≈0.00-0.01; p>0.70). For inclination (Pearson r), correlations were strong between BP and ZS (r≈0.90; r²≈0.81), BP and EE (r≈0.82; r²≈0.68), and ZS and EE (r≈0.80; r²≈0.64), all p<0.001. Associations involving MS were moderate to strong (MS-BP r≈0.74; r²≈0.54; MS-ZS r≈0.75; r²≈0.57; MS-EE r≈0.57; r²≈0.33; all p≤0.001).
Table 3. Groupwise correlation coefficient values for retroversion and inclination.
| p-value | r2 value | p-value | |
| Retroversion | |||
| Blueprint vs. Materialise | 0.026 | <0.001 | 0.893 |
| Exactec vs. Materialise | 0.087 | 0.008 | 0.646 |
| Zimmer vs. Materialise | 0.053 | 0.001 | 0.779 |
| Blueprint vs. Zimmer | 0.910 | 0.885 | <0.001 |
| Blueprint vs. Exactec | 0.840 | 0.824 | <0.001 |
| Zimmer vs. Exactec | 0.890 | 0.904 | <0.001 |
| Inclination | |||
| Blueprint vs. Materialise | 0.740 | 0.544 | <0.001 |
| Exactec vs. Materialise | 0.570 | 0.329 | <0.001 |
| Zimmer vs. Materialise | 0.750 | 0.565 | <0.001 |
| Blueprint vs. Zimmer | 0.910 | 0.810 | <0.001 |
| Blueprint vs. Exactec | 0.820 | 0.676 | <0.001 |
| Zimmer vs. Exactec | 0.800 | 0.644 | <0.001 |
Discussion
Accurate pre-operative characterisation of glenoid version and inclination underpins the correction of bone loss, implant/augment selection, restoration of biomechanics, and reproducible reporting across centres [12,13]. Malposition, particularly retroversion >10° or abnormal inclination, has been associated with eccentric loading, micromotion, inferior impingement and notching, erosion, and early loosening [12,21,24,25]. Excessive glenoid inclination is likewise linked to inferior impingement and notching, glenoid neck erosions, and prosthesis loosening [7,26,27].
In this within-subject comparison across four planning platforms, retroversion did not differ at the group level, and the paired median differences were small, indicating no systematic inter-platform bias. However, MS demonstrated markedly greater dispersion (IQR 19.5 vs. 7-11 for other platforms) and weak inter-system correlations with the others (ρ≈0.03-0.07). This combination of no average shift but lower precision suggests that MS measurements are not consistently higher or lower relative to other systems, yet for an individual patient, the value may be unusually high or low and may not align with the value another platform would report.
By contrast, inclination showed a clear system effect. Repeated-measures ANOVA identified significant differences (p<0.001), and within-subject tests showed that all pairwise contrasts were significant except BP vs. MS (p=0.948). BP was higher than ZS by 4.65 and EE by 2.40, MS was higher than ZS by 4.72 and than EE by 2.47, and EE exceeded ZS by 2.25. Despite these level differences, inter-system correlations were strong, indicating good precision with non-zero bias such that the systems track patients similarly but with differing absolute values. These findings are consistent with prior work, showing that inter-platform measurements can be highly correlated yet differ in absolute level, particularly for inclination. Sabesan et al. reported a reliable cross-platform assessment of version (p>0.05) with strong correlation (r≈0.7) but a significant difference in inclination between automated and surgeon-directed workflows (p=0.039) despite only moderate correlation (r≈0.6) [23]. Our results mirror this pattern in which inclination comparisons between automated systems were significant (p<0.001), with smaller yet still significant differences when a surgeon-directed workflow was involved (BP vs. EE p=0.009; ZS vs. EE p=0.013), whereas the version showed no group-level difference.
Taken together, these data suggest that while 3D planning outperforms 2D slice measurement for version, inclination, and bone loss, platforms may not be truly interchangeable. They tend to track patients similarly (high precision) yet sit at different absolute levels (bias). Because inclination thresholds often guide augment selection and baseplate tilt, platform-specific bias can move a case across a threshold when decisions hinge on a few degrees. Observed differences likely reflect variations in calculation parameters or landmark selection. Reference points for inclination and retroversion may be distorted by glenoid cavity changes such as osteophytes, erosions, or superior-to-inferior twists of the glenoid [28,29]. Re-orienting CT data into a reproducible scapular plane reduces variability, but divergence between proprietary algorithms and user-directed steps remains [27]. Surgeon-directed systems offer more flexibility but may introduce surgeon error, whereas automated systems rely on their underlying algorithms, making the process of glenoid mapping less transparent. Finally, as implantation typically deviates little from the pre-operative plan, platform-related differences are not only academic. They can translate into different augment choices or baseplate tilt for the same shoulder [16]. Although variation exists between implant design manufacturers, there should be no statistical difference in measurement parameters to allow consistency in the implementation of the biomechanical principles of shoulder replacement. Therefore, it is important to recognise that platform-specific differences exist and should be factored into decision-making. Two practical points follow. First, for retroversion, particularly when an MS value is near a decision threshold or appears incongruent with the CT anatomy, internal verification is appropriate with repetition or refinement of segmentation, confirmation of rim landmarks, and re-checking of scapular plane derivation, then considering the value within a narrow sensitivity range rather than as a rigid cut-off. Secondly, for inclination, known offsets between platforms mean that threshold-adjacent plans should explicitly confirm the reference frame and landmarking within the chosen platform, and cases that are close to a decision threshold should be reviewed carefully.
The strengths of this study include a within-subjects design across four widely used systems (both fully automated and surgeon-directed) and the use of a standardised whole-scapula CT protocol at a single site, minimising inter-patient variation in imaging technique. Limitations include a modest sample size, lack of access to proprietary algorithms, and absence of an external ground truth (e.g., cadaveric validation). Similarly, CT scan variation, artefact, or differences in how reference points used to calculate version and inclination may introduce measurement variability. Moreover, surgeon-directed software may introduce subjectivity and human error, leading to inter-user variability. Additionally, intra-operative execution or clinical outcomes were not assessed, noting prior evidence that implantation typically follows the pre-operative plan closely [16]. Accordingly, our results reflect planning agreement rather than implant placement or patient-reported results, and further research is needed to better understand the downstream effects of such variability.
Conclusions
Across four contemporary planning systems, retroversion showed no group-level bias but reduced inter-system precision, particularly for MS, whereas inclination was highly precise yet systematically different between pairs. Surgeons should account for these platform-specific behaviours during pre-operative planning by verifying retroversion values that drive correction and confirming inclination decisions close to thresholds so that implant selection and baseplate orientation remain guided by anatomy and informed by known characteristics of the planning platform.
Disclosures
Human subjects: Informed consent for treatment and open access publication was obtained or waived by all participants in this study. NHS Health Research Authority/East and North Hertfordshire Clinical Governance Department issued approval 18751. All data were handled in accordance with local NHS information governance requirements and analysed in anonymised form.
Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue.
Conflicts of interest: In compliance with the ICMJE uniform disclosure form, all authors declare the following:
Payment/services info: All authors have declared that no financial support was received from any organization for the submitted work.
Financial relationships: All authors have declared that they have no financial relationships at present or within the previous three years with any organizations that might have an interest in the submitted work.
Other relationships: All authors have declared that there are no other relationships or activities that could appear to have influenced the submitted work.
Author Contributions
Acquisition, analysis, or interpretation of data: George Matheron, Joshua Ong
Drafting of the manuscript: George Matheron, Joshua Ong
Critical review of the manuscript for important intellectual content: George Matheron, Joshua Ong, Adam Rumian, Angelos Assiotis, Harpal S. Uppal
Concept and design: Adam Rumian, Angelos Assiotis, Harpal S. Uppal
Supervision: Harpal S. Uppal
References
- 1.Clinical outcomes after reverse shoulder arthroplasty in patients 60 years old and younger; medium-term results. Chamberlain AM, Aleem AW, Sefko JA, Steger-May K, Keener JD. JSES Int. 2023;7:277–284. doi: 10.1016/j.jseint.2022.12.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Results of a convex-back cemented keeled glenoid component in primary osteoarthritis: multicenter study with a follow-up greater than 5 years. Walch G, Young AA, Melis B, Gazielly D, Loew M, Boileau P. J Shoulder Elbow Surg. 2011;20:385–394. doi: 10.1016/j.jse.2010.07.011. [DOI] [PubMed] [Google Scholar]
- 3.Impact of total shoulder arthroplasty on generic and shoulder-specific health-related quality-of-life measures: a systematic literature review and meta-analysis. Carter MJ, Mikuls TR, Nayak S, Fehringer EV, Michaud K. J Bone Joint Surg Am. 2012;94:0. doi: 10.2106/JBJS.K.00204. [DOI] [PubMed] [Google Scholar]
- 4.National Joint Registry for England, Wales Wales, Northern Ireland and the Isle of Man. National Joint Registry - annual report 2024. National Joint Registry 21st Annual Report. [ Sep; 2025 ]. 2025. https://www.njrcentre.org.uk/national-joint-registry-annual-report-2024/ https://www.njrcentre.org.uk/national-joint-registry-annual-report-2024/
- 5.The incidence of shoulder arthroplasty: rise and future projections compared with hip and knee arthroplasty. Wagner ER, Farley KX, Higgins I, Wilson JM, Daly CA, Gottschalk MB. J Shoulder Elbow Surg. 2020;29:2601–2609. doi: 10.1016/j.jse.2020.03.049. [DOI] [PubMed] [Google Scholar]
- 6.Avoiding superior tilt in reverse shoulder arthroplasty: a review of the literature and technical recommendations. Laver L, Garrigues GE. J Shoulder Elbow Surg. 2014;23:1582–1590. doi: 10.1016/j.jse.2014.06.029. [DOI] [PubMed] [Google Scholar]
- 7.Scapular notching in reverse shoulder arthroplasty. Lévigne C, Boileau P, Favard L, Garaud P, Molé D, Sirveaux F, Walch G. J Shoulder Elbow Surg. 2008;17:925–935. doi: 10.1016/j.jse.2008.02.010. [DOI] [PubMed] [Google Scholar]
- 8.Predictors of scapular notching in patients managed with the Delta III reverse total shoulder replacement. Simovitch RW, Zumstein MA, Lohri E, Helmy N, Gerber C. J Bone Joint Surg Am. 2007;89:588–600. doi: 10.2106/JBJS.F.00226. [DOI] [PubMed] [Google Scholar]
- 9.Accuracy of CT-based measurements of glenoid version for total shoulder arthroplasty. Hoenecke HR Jr, Hermida JC, Flores-Hernandez C, D'Lima DD. J Shoulder Elbow Surg. 2010;19:166–171. doi: 10.1016/j.jse.2009.08.009. [DOI] [PubMed] [Google Scholar]
- 10.Effects of glenoid component version on humeral head displacement and joint reaction forces: an experimental study. Nyffeler RW, Sheikh R, Atkinson TS, Jacob HA, Favre P, Gerber C. J Shoulder Elbow Surg. 2006;15:625–629. doi: 10.1016/j.jse.2005.09.016. [DOI] [PubMed] [Google Scholar]
- 11.Glenoid replacement in total shoulder arthroplasty. Keller J, Bak S, Bigliani LU, Levine WN. Orthopedics. 2006;29:221–226. doi: 10.3928/01477447-20060301-05. [DOI] [PubMed] [Google Scholar]
- 12.Risks of loosening of a prosthetic glenoid implanted in retroversion. Farron A, Terrier A, Büchler P. J Shoulder Elbow Surg. 2006;15:521–526. doi: 10.1016/j.jse.2005.10.003. [DOI] [PubMed] [Google Scholar]
- 13.Effect of glenoid deformity on glenoid component placement in primary shoulder arthroplasty. Iannotti JP, Greeson C, Downing D, Sabesan V, Bryan JA. J Shoulder Elbow Surg. 2012;21:48–55. doi: 10.1016/j.jse.2011.02.011. [DOI] [PubMed] [Google Scholar]
- 14.Two-dimensional glenoid version measurements vary with coronal and sagittal scapular rotation. Bryce CD, Davison AC, Lewis GS, Wang L, Flemming DJ, Armstrong AD. J Bone Joint Surg Am. 2010;92:692–699. doi: 10.2106/JBJS.I.00177. [DOI] [PubMed] [Google Scholar]
- 15.Variability of measurement of glenoid version on computed tomography scan. Bokor DJ, O’Sullivan MD, Hazan GJ. J Shoulder Elb Surg. 1999;8:595–598. doi: 10.1016/s1058-2746(99)90096-4. [DOI] [PubMed] [Google Scholar]
- 16.What is the deviation in 3D preoperative planning software? A systematic review of concordance between plan and actual implant in reverse total shoulder arthroplasty. Lilley BM, Lachance A, Peebles AM, et al. J Shoulder Elbow Surg. 2022;31:1073–1082. doi: 10.1016/j.jse.2021.12.006. [DOI] [PubMed] [Google Scholar]
- 17.Use of three-dimensional computed tomography for the analysis of the glenoid anatomy. Kwon YW, Powell KA, Yum JK, Brems JJ, Iannotti JP. J Shoulder Elbow Surg. 2005;14:85–90. doi: 10.1016/j.jse.2004.04.011. [DOI] [PubMed] [Google Scholar]
- 18.The influence of three-dimensional planning on decision-making in total shoulder arthroplasty. Werner BS, Hudek R, Burkhart KJ, Gohlke F. J Shoulder Elbow Surg. 2017;26:1477–1483. doi: 10.1016/j.jse.2017.01.006. [DOI] [PubMed] [Google Scholar]
- 19.Computerized 3D morphological analysis of glenoid orientation. Ghafurian S, Galdi B, Bastian S, Tan V, Li K. J Orthop Res. 2016;34:692–698. doi: 10.1002/jor.23053. [DOI] [PubMed] [Google Scholar]
- 20.Innovations in shoulder arthroplasty. Leafblad N, Asghar E, Tashjian RZ. J Clin Med. 2022;11 doi: 10.3390/jcm11102799. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Prosthetic positioning in total shoulder arthroplasty. Iannotti JP, Spencer EE, Winter U, Deffenbaugh D, Williams G. J Shoulder Elbow Surg. 2005;14:111–121. doi: 10.1016/j.jse.2004.09.026. [DOI] [PubMed] [Google Scholar]
- 22.Proper benefit of a three dimensional pre-operative planning software for glenoid component positioning in total shoulder arthroplasty. Jacquot A, Gauci MO, Chaoui J, et al. Int Orthop. 2018;42:2897–2906. doi: 10.1007/s00264-018-4037-1. [DOI] [PubMed] [Google Scholar]
- 23.Reliability and accuracy of 3D preoperative planning software for glenoid implants in total shoulder arthroplasty. Sabesan VJ, Lima DJL, Rudraraju RT, Wilneff M, Sheth B, Yawman J. Semin Arthroplast JSES. 2020;30:375–382. [Google Scholar]
- 24.The glenoid in shoulder arthroplasty. Strauss EJ, Roche C, Flurin PH, Wright T, Zuckerman JD. J Shoulder Elbow Surg. 2009;18:819–833. doi: 10.1016/j.jse.2009.05.008. [DOI] [PubMed] [Google Scholar]
- 25.Glenoid component failure in total shoulder arthroplasty. Matsen FA 3rd, Clinton J, Lynch J, Bertelsen A, Richardson ML. J Bone Joint Surg Am. 2008;90:885–896. doi: 10.2106/JBJS.G.01263. [DOI] [PubMed] [Google Scholar]
- 26.Outcomes of anatomic shoulder arthroplasty in primary osteoarthritis in type B glenoids. Chin PC, Hachadorian ME, Pulido PA, Munro ML, Meric G, Hoenecke HR Jr. J Shoulder Elbow Surg. 2015;24:1888–1893. doi: 10.1016/j.jse.2015.05.052. [DOI] [PubMed] [Google Scholar]
- 27.Radiographic characterization of the B2 glenoid: the effect of computed tomographic axis orientation. Chalmers PN, Salazar D, Chamberlain A, Keener JD. J Shoulder Elbow Surg. 2017;26:258–264. doi: 10.1016/j.jse.2016.07.021. [DOI] [PubMed] [Google Scholar]
- 28.Reliability of the glenoid plane. De Wilde LF, Verstraeten T, Speeckaert W, Karelse A. J Shoulder Elbow Surg. 2010;19:414–422. doi: 10.1016/j.jse.2009.10.005. [DOI] [PubMed] [Google Scholar]
- 29.Evaluation of three-dimensional glenoid structure using MRI. Inui H, Sugamoto K, Miyamoto T, Machida A, Hashimoto J, Nobuhara K. J Anat. 2001;199:323–328. doi: 10.1046/j.1469-7580.2001.19930323.x. [DOI] [PMC free article] [PubMed] [Google Scholar]