Abstract
In magnetic resonance imaging (MRI), arrays of small loop antennas/coils are extensively used for signal reception. In such arrays, the loop current must be suppressed to prevent transmit antennas or other elements in a receive array from being detuned, thereby preserving image quality. While suppressing the current in conventional loop antennas/coils is straightforward, achieving effective suppression in shielded loops has remained an unresolved challenge. In this article, we derive theoretical principles for optimally suppressing the loop current of shielded loops at fundamental resonance and verify them experimentally. Maximal current suppression is achieved by carefully selecting the electrical load at the antenna outputs. We identify a critical relationship between loop inductance and load inductance specific to shielded loops. Our results demonstrate that the optimal suppression method reduces loop current by an additional 31–36 dB compared with the straightforward yet suboptimal approach of shorting the antenna outputs. These findings can facilitate the development of more robust antenna/coil arrays for MRI applications.
Keywords: Decoupling, Loop antenna, Coil, Magnetic resonance imaging (MRI), Shielded loop
Subject terms: Biomedical engineering, Electrical and electronic engineering
Introduction
A loop antenna consists of one or more turns of conductive wire wound around a frame of any shape, which is typically circular1. Loop antennas may be formed from single-conductor wires or two-conductor coaxial cables. While electrically small loop antennas are inefficient far-field radiators, they are valuable for measuring magnetic fields and current distributions. In MRI, electrically small loop antennas are widely used for signal reception and are commonly referred to as “coils”. In this article, loop antennas formed from single-conductor wires are termed “loop antennas” or “coils”, often written together as “loop antennas/coils”. Loop antennas formed from two-conductor coaxial cables are called “shielded loops”.
Loop antennas have existed since 19002 and have been used in MRI since its inception in the 1970s3. Modern coaxial-cable shielded loops were invented in 19414 and introduced into MRI in 20185. Prior to their adoption in MRI, shielded loops had already been widely utilized for magnetic field sensing6–8 and could take various forms, including multi-turn multi-gap9,10 and Möbius loops11. Since their introduction into MRI, shielded loops have demonstrated greater resilience to mechanical deformation and exhibit lower noise correlation than loop antennas/coils when assembled into arrays5,12–15. When used in MRI, shielded loops are typically designed to operate at fundamental resonance, where their impedance characteristics closely resemble those of parallel R-L-C resonant circuits and exhibit high impedance at their terminals. The MRI community therefore refers to them as “high-impedance coils” or “parallel-resonant coils”.
The imaging process of MRI involves a transmission stage and a reception stage. To facilitate transmission and reception, at least one radio frequency (RF) transmitter and one RF receiver are needed, the latter of which is typically an electrically small loop antenna/coil or a loop antenna/coil array as described above. During transmission, the RF transmitter generates an alternating magnetic field that excites the nuclear spins within the sample. This causes the macroscopic magnetic moment to rotate away from its thermal equilibrium orientation. During reception, the transmitter is deactivated. The sample magnetization relaxes toward equilibrium while inducing a voltage across the RF receiving loops, and accordingly, currents on these loops16,17. For standalone loops, there are no constraints on the loop current, but for loops in an array, the loop current must be suppressed. There are three reasons. First, in a receiving loop array, the currents of the loop antennas/coils interfere with one another through their magnetic fields. This interference couples the electrical characteristics of each element and merges their individual sensitivity profiles, which in practice significantly complicates the tuning of receiving circuitry and degrades image quality18. Second, the loop current may be induced by the high-power RF transmitter during the transmission phase. This can interfere with the nuclear spins in the sample, resulting in artifacts and a reduced signal-to-noise ratio (SNR) in the images19. Third, the current may pose a safety hazard.
By carefully overlapping loops, the mutual inductance between them can be minimized. However, this approach does not mitigate interference between non-adjacent loops, nor does it suppress loop current during the transmit phase. For these reasons, loop current suppression must be addressed in MRI. Such techniques have been described extensively in the literature. One technique is “preamplifier decoupling”, which employs specially designed preamplifiers (low-noise amplifiers) and antenna/coil-to-amplifier matching circuitry to suppress loop current18,20, thereby reducing inter-element coupling in antenna/coil arrays. Another technique is “active detuning”, which incorporates current-blocking circuits that activate during the transmit phase. In both cases, the key principle is to create a high-impedance block along the loop current path.
For loop antennas/coils, the method to suppress current is straightforward: connect a high-impedance load to the antenna/coil terminals18,21. For shielded loops, however, the method remains unclear. Short-circuiting the antenna/coil output terminals initially seemed intuitive5. However, whether through “reverse preamplifier decoupling”5 or through more rigorous analyses22,23, inductors have emerged as central to shielded loop current suppression. To date, discussions have only encompassed shielded loops with a single gap and without tuning components. For multi-gap loops9 or loops with tuning components14,24, a solution has yet to be proposed.
To tackle the problem of optimal loop current suppression, we propose suitable conditions for suppressing the loop current of a shielded loop at fundamental resonance. The main propositions are:
Although a shielded loop behaves like a parallel R-L-C circuit with a current source, it is not a parallel R-L-C circuit. The proper method to suppress its loop current is by neutralizing the reactance of the shielded loop at the output and presenting a low resistance to the shielded loop, rather than neutralizing the susceptance of the shielded loop and presenting a high conductance.
For a multi-gap shielded loop as shown in Fig. 1, regardless of whether tuning components (e.g.,
in Fig. 2(a)) are present or whether the gaps are evenly distributed along the shielded loop, the reactance of the shielded loop
at the resonant frequency can be calculated by disconnecting all loop inductors formed by the outer shield. This disconnection does not change the reactance of the shielded loop at the resonant frequency. To completely suppress the loop current, the component
placed at the output of the shielded loop must have low equivalent series resistance (ESR) and a reactance of
.If the shielded loop has no tuning component, or if tuning components are not directly connected to the cable sections containing the output of the shielded loop, then the current-suppression component is or resembles a low-loss inductor, regardless of whether the gaps are evenly distributed.
- For an
-gap shielded loop (
) with evenly distributed gaps and no tuning components, a low-loss inductor is required to completely suppress the current. Its inductance is
, where
is the loop inductance formed by a loop antenna of the same dimensions as the shielded loop. For a wire radius
and a loop radius
1, 
1
Fig. 1.
(a) Dissection of a shielded loop: iC—inner conductor, oSi—inner surface of the outer shield, oSo—outer surface of the outer shield. (b) Shielded loop terminated by a matching network
and an amplifier
. The nodes of the antenna output are labeled
,
. Gaps are labeled
, etc. The current branches flowing on oSo are labeled
, etc. (c) One-turn 1-gap, 2-gap, and 3-gap shielded loops, whose the loop diameter is
and the wire diameter is
. A wire loop of loop diameter
and wire diameter
is drawn for comparison. The inductance and resistive loss of such a wire loop are denoted as
and
respectively. Subfigure (c) is re-used from Wang et al.35.
Fig. 2.
(a) Equivalent circuit of a 1-gap shielded loop.
is an optional tuning component that lowers the resonant frequency
.
is the Thévenin impedance of
.
is the antenna output. The transmission lines and
constitute a network
, of which the half is denoted as
. (b) Half of the equivalent circuit of a 1-gap shielded loop. (c) Half of the equivalent circuit of a 2-gap shielded loop. (d) Half of the equivalent circuit of a 3-gap shielded loop. In (b), (c), and (d), transmission line
has length
. All
’s need not be equal. Transmission lines
and
constitute a network
. Because all the currents flowing into and out of the network are equal to
, the network can be deemed as having two ports the nodes of port 1 are labeled
, and the nodes of port 2 are labeled
; cf. Supplementary Information §2.4. Node
is also node
. For even
, the voltage at
and
is 0. When
is absent,
is not connected to ground. The other half of the equivalent circuit is obtained by reversing the directions of
and
, and replacing
. (e) General full-circuit cascaded-network representation of
-gap shielded loops, including the 1-gap in (b), the 2-gap in (c), and the 3-gap in (d). Half is denoted without a prime; the other half is denoted with a prime. The input impedance of the half of
at port 1, i.e.
or
, equals half the impedance of
at the port where the loop impedance
is connected, i.e.
; the same applies to
and
.
and
are the same node as exemplified in (b), (c), and (d). Disconnecting all blue components does not change
at
. The tuning component
, if used, must remain connected because
belongs to
, as exemplified in (b), (c), and (d). Subfigures (a), (b) are re-used from Wang et al.35.
All the propositions above are applicable whether a sample is present or absent. However, in the presence of a sample, the loop inductance
can deviate significantly from the value given by (1).
Below exemplifies the interpretation of Propositions 1–4.
-
I.For a single-gap shielded loop where
, the propositions are interpreted as follows, whether a sample is present or absent: - The correct approach to suppressing loop current is to neutralize the reactance
and present a low resistance
, rather than to neutralize the susceptance
and present a high conductance
. - Construct an equivalent circuit as shown in Fig. 2(b) to obtain the output reactance
. Remove the loop inductor
to obtain the output reactance
. At the resonant frequency,
. To fully suppress the loop current, the component
at the loop output must have low ESR and satisfy
, regardless of whether tuning components are present. From Propositions 1 and 2, it can be shown that at the resonant frequency
and in the absence of tuning components,
and
. - In the absence of tuning components, the current-suppression component
is, or closely resembles, a low-loss inductor. - In the absence of tuning components, a low-loss inductor with inductance
is required to fully suppress the loop current, where
is given by (1) in the absence of a sample. This is consistent with the analysis of Mollaei et al.22. In the presence of a sample,
can deviate significantly from the value given by (1).
-
II.For a three-gap shielded loop where
, the propositions are interpreted as follows, whether a sample is present or absent: - The correct approach to suppressing loop current is to neutralize the reactance
and present a low resistance
, rather than to neutralize the susceptance
and present a high conductance
. - In Fig. 2(d), remove the loop inductor
to obtain the output reactance
. At the resonant frequency,
. To fully suppress the loop current, the component
at the loop output must have low ESR and satisfy
, regardless of whether tuning components are present. From Propositions 1 and 2, it can be shown that at the resonant frequency
, when tuning components do not exist on
,
and
. In the complete absence of tuning components,
. - When tuning components do not exist on
, the current-suppression component
is, or closely resembles, a low-loss inductor.
Proposition 1 facilitates the optimal design of a matching network for minimal loop current. Proposition 2 can be applied to calculate the component values for current suppression while avoiding rapid impedance changes near the resonant frequency of a parallel R-L-C circuit. Propositions 3 and 4 enable the rapid calculation of inductor values for current suppression. A comparison between brute-force calculation and the propositions presented in this article can be found in §1 in the Supplementary Information.
The formal statements are presented in the “Propositions” section within the “Theory” section based on the models in the “Structures and Working Mechanism of Shielded Loops” and “Definitions of Resonance” sections within “Theory”. Their derivation is provided in the “Theory” section and in the Supplementary Information. These propositions are experimentally validated, which requires a calibration method for double-loop probe measurements25, of which a brief overview can be found in Section “Calibrating a Double-Loop H-field Probe”.
Results
To validate Propositions 1–4, five shielded loops are tested: a 1-gap loop without tuning components, a 2-gap loop without tuning components, a 3-gap loop without tuning components, a 1-gap loop with a tuning capacitor, a 2-gap loop with a tuning capacitor at position
(refer to Fig. 1 and Fig. 3(a), (b)). Samples are not used. Inductor–capacitor
–
parallel combinations are employed as tunable inductors with the capacitor tuning up the equivalent inductance
as shown in Fig. 3(d)26, of which the equivalent inductance
is measured with a vector network analyzer (VNA). Detailed specifications and measurement steps are provided the “Methods” section. The results are in Table 1, Fig. 4, Fig. 5.
Fig. 3.
(a) Experimental setup. A double-loop probe measures the loop current. The double-loop probe is connected to a VNA. (b) Tuning capacitors. (c) PCB. An RF shield is installed to suppress the mutual inductance between a loop and the inductor on the board. (d) Circuit schematic of (c).
Table 1.
Resonant frequencies (“Rs. Freq.”), nuclei with the closest Larmor frequencies to the resonant frequencies (“Nucleus”), inferred values of inductors
and trimmer capacitors
in experiments, equivalent inductance
, inferred loop inductance
, and theoretical inductance
. For “no tuning”,
. All loop diameters are 16 cm. Calculated
.
|
No Tuning | Tuned | |||
|---|---|---|---|---|---|
| 1 | 2 | 3 | 1 | 2 | |
[MHz] |
63.97 | 126.62 | 190.03 | 32.15 | 106.27 |
| Nucleus |
H, 1.5T |
H, 3.0T |
— |
C, 3.0T |
Na, 9.4T |
[pF] |
— | — | — | 43.89 | 9.89 |
[nH] |
327.01 | 131.34 | 99.71 | 305.47 | 211.92 |
[pF] |
2.47 | 2.93 | — | 10.89 | 1.03 |
[nH] |
452.63 | 235.27 | 151.71 | 396.40 | 331.95 |
[nH] |
483.30 | 471.61 | 460.46 | 440.83† | 445.55† |
[nH] |
483.30 | 235.81 | 153.49 | 375.32 | 341.89 |
[ ] |
194.26 | 187.60 | 183.26 | 75.82 | 228.28 |
Errors ,
|
6.8% | 0.2% | 1.2% | 5.3% | 3.0% |
†: Has no direct relation with
.
Fig. 4.
Loop current when
are open, shorted, terminated by an inductor or L–C parallel combination (“Induc.”), for (a) 1-gap, (b) 2-gap, and (c) 3-gap shielded loops without any tuning components, and simulated
of (d) 1-gap, (e) 2-gap, and (f) 3-gap shielded loops. Experimental
values are converted to shield current by (2). “Floor” is the detection limit of
of the VNA. “Induc.” and “Floor” are smoothed. The original data of “Induc.” and “Floor” are shown translucent. “L-on” means all
or
are connected to
in the simulation setup, as illustrated in Fig. 2. “L-off” means all
or
are disconnected from
in the simulation setup, i.e. all blue components in Fig. 2 are erased. By unplugging all
,
becomes more numerically stable. The IMax-resonant frequencies
are in the middle of the plots,
, as marked by
and annotated as
. The X0-resonant frequencies
are where “L-on” curves cross
, as marked by
. In all cases, the IMax-resonant frequencies
are almost the same as X0-resonant frequencies
. At
, the “L-on” curves intersect with the “L-off” curves at the square marker, as stated by Proposition 2: disconnecting all
’s does not change
.
Fig. 5.
Loop current when antenna outputs
are open, shorted, terminated by an inductor or L–C parallel combination, for (a) 1-gap and (b) 2-gap shielded loops with a tuning capacitor at
, as illustrated in Fig. 2, and
for (c) 1-gap and (d) 2-gap shielded loops.
values are converted to shield current by (2). “Induc.”, “Floor”, “L-on”, “L-off” mean the same as in Fig. 4.
and
are marked in the same way as in Fig. 4. By unplugging all
,
becomes numerically more stable. In both cases, the IMax-resonant frequencies
are almost the same as X0-resonant frequencies
. At
, the “L-on” curves intersect with the “L-off” curves, as stated by Proposition 2.
Table 1 presents the resonant frequencies
, the values of tuning capacitors
used to tune the 1-gap and the 2-gap shielded loops, the values of the inductors
and the values of the trimmer capacitors
inferred from experiments, the equivalent inductance at the resonant frequency
of the
–
parallel combination, the theoretical inductance
to completely suppress loop current, and the relative difference between
and
.
For 1-, 2- and 3-gap shielded loops without tuning components (
is “—”), with a loop diameter 16 cm, and a cable outer diameter 2.10 mm, the loop inductance is calculated as
according to (1). After accounting for loop length and gap variations in Advanced Design System, it is determined that
for the 1-gap loop in the experiment,
for the 2-gap loop in the experiment, and
for the 3-gap loop in the experiment. For shielded loops with a tuning capacitor
, Proposition 4—which pertains to shielded loops without tuning components—does not apply, and the loop inductance
is not directly related to the theoretical inductance
required to suppress the loop current
.
For 1-, 2-, and 3-gap shielded loops without tuning components (
is “—”), according to Proposition 4,
; for example for the 3-gap loop
, so
. In theory, for all three cases, the component that completely suppresses the loop current is an inductor. In experiments, as shown in Fig. 4(a)–(c), the
–
parallel combinations listed in Table 1 are also equivalent to inductors at the resonance frequencies and suppress loop current to a minimum. This is further demonstrated in Fig. 4(a), (b), (c): installing inductor-equivalent
–
parallel combinations at the loop outputs (refer to Fig. 1(b), (c)) results in a significantly greater level of current suppression compared with shorting the outputs, both theoretically and experimentally. Thus Proposition 3 is verified. The theoretical inductance values
as derived from Proposition 4 are close to the equivalent inductance values
of the L–C parallel combinations used in experiments, with a maximum error of 6.8%. This validates Proposition 4.
Furthermore, as shown in Fig. 4(d), (e), and (f) for 1-, 2-, 3-gap shielded loops without tuning components, the output reactance near resonance typically changes very rapidly (“L-on”), with rates exceeding 22.9
/kHz; after unplugging loop inductors
’s (“L-off”) in the equivalent circuits Fig. 2(b), (c), and (d), the output reactance
changes only slightly, with rates below 3.6m
/kHz. Nevertheless, the “L-on” and “L-off” curves intersect at the resonant frequency
, where
, as indicated in Table 1. This verifies Proposition 2. This also demonstrates that unplugging inductors in the equivalent circuits makes
significantly less susceptible to steep numerical gradients near the resonant frequency; it can be used as a reliable method to determine the inductance required to suppress loop current. Since Proposition 2 is validated, Proposition 1 is also confirmed.
Furthermore, in Fig. 4(d), (e), and (f) the difference between the X0-resonant frequency
and the IMax-resonant frequency
is small. For 1-, 2-, 3-gap loops the differences are
,
,
, respectively (
). Hence in practice X0-resonance and IMax-resonance are practically indistinguishable. However, as demonstrated in the “Theory” section, IMax-resonance provides a theoretical framework through which Propositions 1–4 can be derived.
For shielded loops with a tuning capacitor
at position
, as written in Table 1, the 1-gap (
) shielded loop is tuned to 32.15MHz by
, and the 2-gap (
) shielded loop is tuned to 106.27MHz by
. The loop current measurements are in Fig. 5. In the case of the 1-gap shielded loop, the tuning capacitor
is directly connected to the cable sections containing the antenna output, which is represented as
in Fig. 2(b), so Proposition 3 does not apply. In the case of the 2-gap shielded loop, the tuning capacitor
is not directly connected to the cable sections containing the antenna output represented as
in Fig. 2(c), so Proposition 3 applies. In this case the component required to completely suppress the loop current
is an inductor, as predicted by Proposition 3; thus Proposition 3 is validated for the 2-gap shielded loop with a tuning capacitor.
However, Propositions 1 and 2 are applicable to all shielded loops near fundamental resonance. As shown in Fig. 5(b), (d) for 1-, 2-gap shielded loops with
, the output reactance near resonance typically changes faster than 122
/kHz (“L-on”); after unplugging loop inductors
’s (“L-off”) in the equivalent circuits shown in Fig. 2(b), (c), the output reactance
changes more slowly than 3.4 m
/kHz. The “L-on” and “L-off” curves intersect at the resonant frequency
where
as written in Table 1. This verifies Proposition 2. Again, for shielded loops with tuning components, unplugging inductors in the equivalent circuits makes
significantly less susceptible to steep numerical gradients near the resonant frequency; it can be used as a reliable method to determine the inductance needed to suppress loop current. Since Proposition 2 is verified for shielded loops with tuning components, Proposition 1 is also true.
The difference between X0-resonance and IMax-resonance is
for the 1-gap tuned shielded loop, and
for the 2-gap tuned shielded loop.
In all the cases experimentally validated, whether with or without tuning components, as shown in Fig. 4 and Fig. 5(a), (b), by installing inductors at antenna outputs
, the loop current amplitude
is reduced by an additional 31dB to 36dB. Clearly, for suppressing the loop current
, shorting the antenna outputs is not optimal.
Discussion
Table 1 shows fitted
values vary around
. This is likely because of errors in coil construction. The coaxial cable is severed at the loop output and bridged by copper filling zones of printed circuit boards (PCBs), resulting in a loop that is not perfectly circular. The gap sizes, gap positions, and the contacts between coaxial cables and PCBs vary. The simulations cannot fully account for these factors, so
values determined from experimental data vary.
During simulation, propagation loss is modeled into coaxial cables as
and conductor resistivity. However, in our work, propagation loss can also be neglected. The loop perimeter is 0.50m, the cables are RG-316, and, referring to Supplementary Table S1, the wave propagation loss is at most 0.09dB at 50MHz, 0.13dB at 100MHz, and 0.20dB at 200MHz. Had the shielded loop been constructed from RG-178 or thinner cables, the losses would have increased and might no longer be negligible.
In the experiments, samples are not used so that the loop inductance is not affected. In practical MRI, samples are present near shielded loops. This will change loop inductance
. Notwithstanding, Propositions 1–4 still apply. Although it is difficult to know
in advance, the change of
is mild near resonance, as exemplified in Fig. 4(d), (e), and (f) and Fig. 5(c), (d). This means that in practice substituting sample-free
in
provides a good starting point for fine-tuning
to achieve minimum
at
.
As Jabbari et al.27 note, mutual inductance can exist between different loop segments
. In this article, however, mutual inductance is not incorporated into the theoretical analysis or simulation modeling. This has not led to significant errors. As Table 1 indicates, the errors between the theoretical inductance values and the actual inductance values are below 7% for the coils at fundamental resonance. As frequency increases and the coil enters the second resonance, it may become necessary to incorporate the inter-segment mutual inductance.
“High-impedance coils” can be made from other types of cables such as twisted pairs28 and striplines27. A stripline-based shielded loop has a similar structure to a coaxial shielded loop. In this case Propositions 1–4 are applicable. Twisted pairs, however, have different working mechanisms as both wires are exposed to the external magnetic field and contribute to the loop current. In this case Propositions 1–4 may not apply. Moreover, shielded loops can be made resonant when the antenna outputs are short-circuited29. In this case Propositions 1–4 may not apply.
Conclusion
Systematic methods to determine the termination for suppressing the loop current of shielded loops at fundamental resonance are presented: introduce a component that cancels the antenna reactance; unplug loop inductors to determine the antenna reactance; and, when no tuning components are present and all gaps are evenly distributed, use an inductive component with an inductance of
. The difference between shorting antenna outputs and maximal current suppression is also confirmed experimentally. These results can assist designers in developing loop arrays that produce higher-quality images.
Methods
Materials
Five RG-316 coaxial cables are made into five shielded loops with a diameter of 16 cm: a 1-, a 2-, and a 3-gap loop without tuning components, and a 1- and a 2-gap loop with tuning capacitors
. The experimental setup is shown in Fig. 3. Tuning capacitors are installed at
, as shown in Fig. 2(a)–(d) and Fig. 3(a), (b). The resonant frequencies of these shielded loops are measured using a custom-built double-loop H-field probe connected to a Rohde & Schwarz ZNL3 vector network analyzer (VNA) (Rohde & Schwarz; Munich, Germany). Printed circuited boards (PCBs) are fabricated with FR-4. Coilcraft air-core inductors (Coilcraft Inc.; Cary, IL, USA) are used to suppress loop current. For the 1- and 2-gap shielded loops with tuning capacitors
, low-ESR fixed-value (Passive Plus Inc.; Huntington, NY 11743, USA) and trimmer capacitors (Knowles Electronics; Itasca, IL 60143, USA) are used at gap
to tune the shielded loops. For all shielded loops, inductors
are installed at the loop outputs, as shown in Fig. 1(c) and Fig. 2(a)–(d), to suppress the loop current. In some cases, inductor-trimmer
–
parallel combinations as shown in Fig. 3(d) are used as tunable inductors with the capacitors
increasing the equivalent inductance
of the parallel combination26. The frequency of maximum current suppression decreases with increasing equivalent inductance. After each loop current measurement, a VNA is used to measure the equivalent inductance
of the parallel inductor-capacitor combination
–
. To eliminate the influence of mutual inductance between oSo of the shielded loops and the inductors
on the PCBs, RF shields (Würth Elektronik; 74638 Waldenburg, Germany) are installed on PCBs, as shown in Fig. 3(c).
No samples are used in the experiment for two reasons. First, samples can significantly alter the loop inductance in the frequency range of interest, making it challenging to model in circuit simulations. Second, samples modify the current sensing path of a double-loop H-field probe25, causing the calibrated
of a double-loop H-field probe to deviate substantially from
, thereby introducing additional errors in loop current measurements.
Experiments
The circuit simulator is Advanced Design System (ADS) (Keysight Technologies; Santa Rosa, CA, USA). In simulations the coaxial cable’s parameters are taken from the standard specifications of RG-316 coaxial cables: iC’s diameter 0.454 mm, oSi’s diameter 1.52 mm,
,
. The conductor resistivity is
copper resistivity to account for outer shield braiding. The experiment for a shielded loop is performed as follows: (1) If there is no tuning capacitor, measure the resonance directly; otherwise install fixed-value and trimmer capacitors at
, keep the coil output open-circuited, and tune the resonance to the desired frequency; (2) Calibrate the double-loop H-field probe following the method described in Section “Calibrating a Double-loop H-field Probe”; (3) Install
,
and an RF shield onto the PCB and tune for maximum current suppression at the resonant frequency, with the
and
values determined according to Propositions 1–4; (4) Remove the PCB from the shielded loop, install the 0
jumper R1, and measure the equivalent impedance using a VNA at J3 as shown in Fig. 3(d); (5) Determine the inductor
and the capacitor
values from the measurements.
Calibrating a double-loop H-field probe
A calibration method25 is employed to correct for residual probe coupling:
Connect a probe to a VNA. Tighten all connections. Secure the probe position. Ensure that all shielded loops remain in the same place throughout all measurements.
Remove the shielded loop. Record the
measurement results as
.Put back the shielded loop. Record the
measurement results as
.Take the complex-number difference
.
There is
, where
is a constant related to the specifics of a setup. The exact value value of
need not be known25. In our experiments
is used as the magnitude of measured loop current relative to that at
. This quantity is compared against simulation. In decibels the expression is
![]() |
2 |
Theory
Structure and working mechanism of shielded loops
In this article, only axially symmetric shielded loops made of coaxial cables are addressed. Typically, such an
gap shielded loop is formed by alternating outer gaps and inner gaps. At an outer gap, outer conductors are cut while inner conductors join. At an inner gap, outer conductors join while inner conductors are cut. The gaps can be unevenly or evenly distributed. Evenly distributed gaps are spaced every
. These gaps are denoted as
and
,
. In practice, a tuning component can also be placed at gap
30. Shielded loops of
gaps are shown in Fig. 1(c). The output port is located midway, where outer conductors join and inner conductors are exposed. The two nodes of the exposed inner conductors are labelled as
and
. The antenna output port
can be connected to a network
. The impedance looking toward
at port
is denoted as
. An amplifier
with an input impedance
where
and
can be further connected to
.
presents impedance
to amplifier
. Usually,
is chosen to be the optimal input impedance of amplifier
. Other values may be used if required. The cross-section of a shielded loop, shown in Fig. 1(a), consists of three parts: the inner conductor iC, the inner surface of the outer shield oSi, and the outer surface of the outer shield oSo.
The following are assumed regarding the working mechanism of a shielded loop:
-
(i)
The frequency is sufficiently high that the skin depth31 is far shallower than the shield thickness, so oSi and oSo can be deemed independent.
-
(ii)
Each loop segment is no longer than 0.1 wavelengths in air so the current on oSo can be deemed constant.
-
(iii)
Following assumption (ii), the loop antenna formed by oSo can be well approximated by an inductor.
-
(iv)
Each transmission line is shorter than 1/4 wavelength in the dielectric.
-
(v)
All transmission lines are in TEM mode.
-
(vi)
The coaxial cable of which the shielded loop is made has negligible losses.
-
(vii)
Network
has negligible losses.
For a typical RG-316 coaxial cable, the thickness of the outer shield is about 0.25 mm. Above 2.44 MHz, the skin depth is less than 1/6 of shield thickness. Most MRI applications are above this frequency, and Assumption (i) holds. Assumptions (ii) and (iii) are typical requirements for loop antennas/coils in MRI32, as uneven current distribution will induce electric dipole modes1 and degrade the signal-to-noise ratio (SNR) of signal reception. Whether Assumption (vi) holds depends on the cable type, frequency, and length; for example, the attenuation a typical RG-316 coaxial cable is
at
and
at
, and that of a typical RG-178 cable is
at
and
at
, as shown in Supplementary Table S1. Cables thinner than RG-178, such as Huber+Suhner K_01152 (Herisau, Appenzell Ausserrhoden, Switzerland), can exhibit losses as high as
. Since cable loss degrades SNR, coil designers should minimize cable loss to ensure Assumption (vi) remains valid. Similarly, since matching network loss also degrades SNR, designers should minimize matching network loss to ensure Assumption (iv) remains valid. However, the criterion for “low loss” varies by application; therefore, a universal threshold cannot be specified.
When a shielded loop receives an electromagnetic wave, voltages are induced at the gaps
and
,
, which subsequently generate alternating currents on the oSo segments. These current branches, referred to as “loop current” and denoted as
and
,
, flow into coaxial cable sections, and then to the electrical load
. On oSo, gap voltages can be modeled as voltage sources in series with loop antenna/coil segments that combine to form a loop with the same dimensions as the shielded loop8, as shown in Fig. 1(c). For MRI applications, the loop can be well approximated by a lossy inductor. Denote the inductance as
and the resistance as
. Following this model, the equivalent circuits of a 1-gap shielded loop can be drawn as shown in Fig. 2(a). This is the same model as that in Jabbari et al.27 At the output port of the shielded loop,
, the shielded loop presents impedance
to its electrical load. Since the shielded loop is reflectionally symmetric, the equivalent half circuit can be drawn as Fig. 2(b). The equivalent half circuits of 2- and 3-gap shielded loops can be drawn as Fig. 2(c) and (d) following the same procedure.
Definitions of resonance
Assumption (ii) implies the current is constant on oSo:
![]() |
3 |
due to charge conservation at each gap. Henceforth they are all denoted as
.
Resonance of shielded loops is defined as in Definition 1.
Definition 1
(IMax-resonant) A shielded loop is IMax-resonant if, at angular frequency
, with the port
is open-circuited, i.e.
,
reaches its maximum possible value.
is called the “IMax-resonant frequency”, or “resonant frequency” for short.
As revealed by Lemma 2 in the “Derivation” section, our definition of resonance is equivalent to that used by Nohava et al.9,13
Some literature defines resonance as the condition in which the output reactance or susceptance of the shielded loop at port
is zero5,22. This is formally stated in Definition 2.
Definition 2
(X0-resonant) A shielded loop is X0-resonant if, at angular frequency
,
or equivalently
.
is called the “X0-resonant frequency”.
Near
, an
-gap shielded loop behaves similarly to a parallel R-L-C circuit5, where high impedance is created at port
. An
-gap shielded loop is sometimes referred to as a “high-impedance coil”5,12,22.
Our definition of coil resonance typically yields an
that is nearly identical to
, as shown in Fig. 4 and Fig. 5:
while
. Their relative difference is below 40 parts per million (ppm). Thus the current distributions at the two frequencies are virtually the same, an example of which is given in Supplementary Fig. S1 of a one-gap
shielded loop made of RG-316 coaxial cable with no tuning components. From a purely practical standpoint, there is no need to distinguish between these two definitions of resonance. However, Definition 1 (
) allows direct measurement of the resonance using a double-loop H-field probe25, and it also provides a theoretical framework that directly supports all Propositions 1–4, whereas Definition 2 (
), though numerically close, does not.
Propositions
The propositions in Section “Introduction” can be restated rigorously as follows. For an
-gap IMax-resonant shielded loop, to fully suppress the loop current
at
,
Proposition 1
is low and positive, and
.
Proposition 2
Whether or not tuning components exist,
does not change when
, i.e.
does not change when the loop inductors
are disconnected from the circuit.
Proposition 3
When tuning components are not directly connected to
or
,
. This means
resembles an inductor.
Proposition 4
When the shielded loop has no tuning components, and all
’s and
’s have the same length,
resembles a low-loss inductor of which
.
The meaning of “low” in Proposition 1 depends on the context:
-
I.When a shielded loop is terminated solely by components intended for complete suppression of
, e.g., for current blocking during the transmit phase, low
means all components have high Q or low ESR. The lower the ESR, the lower the resulting
. In this case, at
,
where
4
denotes the functional dependence of
on
. -
II.When a shielded loop is terminated by a matching network
that is connected to an amplifier
, low
means all components in
have low ESR and are tuned such that
and
. In this case,
cannot be arbitrarily low but is bounded by
,
, and
21:
where
5
, and
is the power wave reflection coefficient33,34. Accordingly, 
6
Proof of the propositions
Refer to the Supplementary Information for the derivation of Propositions 1–4.
Supplementary Information
Acknowledgements
The authors would like to thank Professor Lars G. Hanson at the Technical University of Denmark for improving the manuscript. The authors would like to thank Lindsey Barber at ALK (2970 Hørsholm, Denmark) and Thomas Benjamin Wareham-Mathiassen at the Technical University of Denmark for proofreading the “Introduction” section.
Author contributions
W.W. derived the theory, conceived the experiments, conducted the experiments, and analyzed the results. R.A.J. derived part of the theory. J.D.S. conceived the experiments. V.Z. and J.H.A. are the supervisors. All authors reviewed the manuscript.
Funding
This work was supported by ERC Synergy grant 856432—HyperQ and Innovation Fund Denmark grant E2409.
Data availability
The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.
Declarations
Competing interests
The authors declare no competing interests.
Footnotes
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
The online version contains supplementary material available at 10.1038/s41598-026-36956-7.
References
- 1.King, R. W. P. The Loop Antenna for Transmission and Reception. In Collin, R. E. & Zucker, F. J. (eds.) Antenna Theory Part 1, chap. 11, 458–482 (McGraw-Hill, 1969).
- 2.Balanis, C. A. The Evolution of Antenna Technology: History, dipoles, and loops. IEEE Antennas and Propagation Magazine66, 42–51. 10.1109/MAP.2024.3393889 (2024). [Google Scholar]
- 3.Grannell, P. K. & Mansfield, P. Microscopy in vivo by nuclear magnetic resonance. Physics in Medicine and Biology20, 477–482. 10.1088/0031-9155/20/3/011 (1975). [DOI] [PubMed] [Google Scholar]
- 4.Finch, W. G. H. & Wakefield, J. M. Coaxial Cable Loop Antenna (1944). US Patent, 2349154.
- 5.Zhang, B., Sodickson, D. K. & Cloos, M. A. A high-impedance detector-array glove for magnetic resonance imaging of the hand. Nature Biomedical Engineering2, 570–577. 10.1038/s41551-018-0233-y (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Heebl, J. D., Thomas, E. M., Penno, R. P. & Grbic, A. Comprehensive analysis and measurement of frequency-tuned and impedance-tuned wireless non-radiative power-transfer systems. IEEE Antennas and Propagation Magazine56, 44–60. 10.1109/MAP.2014.6931657 (2014). [Google Scholar]
- 7.Dyson, J. D. Measurement of Near Fields of Antennas and Scatterers. IEEE Transactions on Antennas and Propagation21, 446–460. 10.1109/TAP.1973.1140518 (1973). [Google Scholar]
- 8.Libby, L. L. Special Aspects of Balanced Shielded Loops. Proceedings of the IRE34, 641–646. 10.1109/JRPROC.1946.230887 (1946). [Google Scholar]
- 9.Nohava, L. et al. Flexible Multi-Turn Multi-Gap Coaxial RF Coils: Design Concept and Implementation for Magnetic Resonance Imaging at 3 and 7 Tesla. IEEE Transactions on Medical Imaging40, 1267–1278. 10.1109/TMI.2021.3051390 (2021). [DOI] [PubMed] [Google Scholar]
- 10.Lindsay, J. E. & Münter, K. Distributed Parameter Analysis of Shielded Loops Used for Wide-Band H-Field Measurements. IEEE Transactions on Instrumentation and Measurement32, 241–244. 10.1109/TIM.1983.4315051 (1983). [Google Scholar]
- 11.Aluf, O. Moebius loop antenna system stability analysis under parameters variation. In 2017 IEEE International Conference on Microwaves, Antennas, Communications and Electronic Systems, COMCAS 2017, 1–5, 10.1109/COMCAS.2017.8244713 (IEEE, 2017).
- 12.Zhang, B. et al. Twenty-four-channel high-impedance glove array for hand and wrist MRI at 3T. Magnetic Resonance in Medicine87, 2566–2575. 10.1002/mrm.29147 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Ruytenberg, T., Webb, A. & Zivkovic, I. Shielded-coaxial-cable coils as receive and transceive array elements for 7T human MRI. Magnetic Resonance in Medicine83, 1135–1146. 10.1002/mrm.27964 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Zhurbenko, V., Sanchez-Heredia, J. D., Wang, W. & Ardenkjær-Larsen, J. H. Flexible Self-Resonant Detector Coil for Magnetic Resonance Imaging of Carbon-13. In 2020 50th European Microwave Conference, EuMC 2020, 112–115, 10.23919/EuMC48046.2021.9338182 (Utrecht, Netherlands, 2021).
- 15.Obermann, M. et al. Panoramic Magnetic Resonance Imaging of the Breast With a Wearable Coil Vest. Investigative Radiology58, 799–810. 10.1097/RLI.0000000000000991 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Nishimura, D. G. Principles of Magnetic Resonance Imaging (Lulu Press, Inc., 2010).
- 17.Hanson, L. G. Is quantum mechanics necessary for understanding magnetic resonance?. Concepts in Magnetic Resonance Part A32A, 329–340. 10.1002/cmr.a.20123 (2008). [Google Scholar]
- 18.Roemer, P. B., Edelstein, W. A., Hayes, C. E., Souza, S. P. & Mueller, O. M. The NMR phased array. Magnetic Resonance in Medicine16, 192–225. 10.1002/mrm.1910160203 (1990). [DOI] [PubMed] [Google Scholar]
- 19.Marhabaie, S., Labbé, A., Quesson, B. & Poirier-Quinot, M. The Minimum Admissible Detuning Efficiency of MRI Receive-Only Surface Coils. Journal of Magnetic Resonance Imaging60, 777–788. 10.1002/jmri.29209 (2024). [DOI] [PubMed] [Google Scholar]
- 20.Reykowski, A., Wright, S. M. & Porter, J. R. Design of Matching Networks for Low Noise Preamplifiers. Magnetic Resonance in Medicine33, 848–852. 10.1002/mrm.1910330617 (1995). [DOI] [PubMed] [Google Scholar]
- 21.Wang, W., Zhurbenko, V., Sánchez-Heredia, J. D. & Ardenkjær-Larsen, J. H. Trade-off between preamplifier noise figure and decoupling in MRI detectors. Magnetic Resonance in Medicine89, 859–871. 10.1002/mrm.29489 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Mollaei, M. S. M., Van Leeuwen, C. C., Raaijmakers, A. J. & Simovski, C. R. Analysis of High Impedance Coils Both in Transmission and Reception Regimes. IEEE Access8, 129754–129762. 10.1109/ACCESS.2020.3009367 (2020). [Google Scholar]
- 23.Sharifian Mazraeh Mollaei, M., Mollaei, A. S. M. & Simovski, C. R. Dual-Band Transceiver High Impedance Coil Array for Ultrahigh Field Magnetic Resonance Imaging. IEEE Transactions on Antennas and Propagation70, 1033–1044, 10.1109/TAP.2021.3111347 (2022).
- 24.Payne, K., Zhao, Y., Bhosale, A. A. & Zhang, X. Dual-Tuned Coaxial-Transmission-Line RF Coils for Hyperpolarized13 C and Deuterium2 H Metabolic MRS Imaging at Ultrahigh Fields. IEEE Transactions on Biomedical Engineering71, 1521–1530. 10.1109/TBME.2023.3341760 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Wang, W., Sánchez-Heredia, J. D., Maurouard, T., Zhurbenko, V. & Ardenkjær-Larsen, J. H. Calibrating Double-Loop H-Field Probe Measurements of RF Coil Current for MRI. IEEE Journal of Electromagnetics, RF and Microwaves in Medicine and Biology7, 266–272. 10.1109/JERM.2023.3274742 (2023). [Google Scholar]
- 26.Hoult, D. I. & Kolansky G. A Magnetic-Field-Tolerant Low-Noise SiGe Pre-amplifier and T/R Switch. In Proceedings of the 18th Annual Meeting of International Society of Magnetic Resonance in Medicine, vol. 5470, 649 (Stockholm, Sweden, 2010).
- 27.Jabbari, A., Simovski, C., & Mollaei Sharifian Mazraeh M, Tunable Dual-Band High-Impedance Coil for Wireless Power Transfer Applications. IEEE Transactions on Antennas and Propagation71, 9467–9476. 10.1109/TAP.2023.3318837 (2023).
- 28.Czerny, R., Frass-Kriegl, R., Cap, V., Laistler, E. & Nohava, L. Twisted pair coils as flexible receive elements for 7T-SNR and active detuning efficiency. In Proceedings of the 31st Annual Meeting of the International Society for Magnetic Resonance in Medicine, 3730 (Toronto, ON, Canada, 2023).
- 29.Fuqua, H., Taracila, V., Giancola, M. & Stormont, R. A Conforming Posterior Radio Frequency (RF) Coil Array for a Magnetic Resonance Imaging (MRI) System (2018).
- 30.Zhurbenko, V., Sánchez-Heredia, J. D., Wang, W. & Ardenkjær-Larsen, J. H. Flexible Receive Array of Parallel-Resonance High-Impedance Coaxial Coils for13C Imaging. In Proceedings of the International Society for Magnetic Resonance in Medicine 30, 2261 (London, UK, 2022).
- 31.Pozar, D. M. Chapter 1: Electromagnetic Theory. In Microwave Engineering, chap. 1, 1–47 (John Wiley & Sons, Inc., 2011), 4 edn.
- 32.Collins, C. M. & Webb, A. G. Quadrature Surface Coils. In Vaughan, J. T. & Griffiths, J. R. (eds.) RF Coils for MRI, 17–26 (John Wiley & Sons Ltd., Chichester, United Kingdom, 2012).
- 33.Kurokawa, K. Power Waves and the Scattering Matrix. IEEE Transactions on Microwave Theory and Techniques13, 194–202. 10.1109/TMTT.1965.1125964 (1965). [Google Scholar]
- 34.Pozar, D. M. Chapter 4: Microwave Network Analysis. In Microwave Engineering, chap. 4, 165–227 (John Wiley & Sons, Inc., 2011), 4 edn.
- 35.Wang, W., Sanchez-Heredia, J. D., Zhurbenko, V. & Ardenkjaer-Larsen, J. H. Condition for Optimal Preamplifier Decoupling in One-turn Single- and Multi-gap Shielded Loop MRI Detectors. In 2022 IEEE 2nd Ukrainian Microwave Week, UkrMW 2022 - Proceedings, 150–154, 10.1109/UkrMW58013.2022.10037007 (IEEE, 2022).
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.