Optimizing the dissection of small-diameter pulmonary vessels using vessel-sealing systems

Yuichiro Ueda 1,✉, Jun-ichi Wakahara 1, So Miyahara 1, Hiroyasu Nakashima 1, Yoshiko Masuda 1, Fumihiro Shoji 1, Takeshi Shiraishi 1, Toshihiko Sato 1
PMCID: PMC12972124  PMID: 41673261

Abstract

Although vessel-sealing systems have been shown useful for dissection of small-diameter vessels, no studies to date have evaluated the optimal angle of dissection. Lateral thermal damage to a canine pulmonary artery was assessed microscopically. Burst pressure was compared in swine carotid arteries undergoing orthogonal and oblique dissections to determine the safer dissection technique. Histological analysis of a canine model revealed a thermal spread of approximately 1.3 mm. Burst pressure comparison showed no significant differences between orthogonal and oblique dissection angles (829.2 mmHg [range 608–1214 mmHg] vs. 949.1 mmHg [range, 593–1306 mmHg], P = 0.206). In a surrogate arterial model, burst pressure did not differ between orthogonal and oblique cuts. Given histologically limited lateral thermal spread on PA, angle may be less critical than avoiding vessel tension. These findings may contribute to the safer application of vessel-sealing systems in patients undergoing minimally invasive thoracic surgery.

Supplementary Information

The online version contains supplementary material available at 10.1038/s41598-026-39741-8.

Keywords: Vessel-sealing system, Vessel dissection, Burst pressure

Subject terms: Medical research, Engineering

Introduction

Vessel-sealing systems are utilized surgically to seal blood vessels and achieve hemostasis. This technology has been found to safely assist tissue and vascular dissection in patients undergoing endoscopic surgery1. Advanced bipolar devices, such as the LigaSure vessel sealing system (Valleylab FT10, Boulder, Colorado), employ a combination of pressure and precisely controlled bipolar energy to denature collagen and elastin within vessel walls, thereby creating a permanent seal. LigaSure has been reported to prevent postoperative chylothorax and bleeding in patients undergoing thoracic surgery2. The LigaSure system has been recommended for the safe dissection of small-diameter (≤ 7 mm) vessels without proximal ligation3,4. Despite the advantages of the LigaSure system, it also presents risks, particularly the potential for thermal damage to surrounding tissues5. Studies indicate that the extent of lateral thermal damage caused by LigaSure is approximately 1.5 mm6–8, suggesting the importance of preventing thermal damage, especially when operating in the perivascular region.

Energy devices must be handled with care, particularly when introduced through a small incision. The insertion of energy devices is complicated in patients undergoing uniportal video-assisted thoracoscopic surgery (U-VATS), as the surgical space is more limited using U-VATS than more conventional approaches9. To date, however, no studies have evaluated the optimal position and angle for vessel dissection.

The present study therefore evaluated the effects of lateral thermal damage from the LigaSure Xp on the pulmonary artery, measured the burst pressure of a dissected edge cut orthogonally or obliquely to the vessel, and determined the most appropriate position and angle for vascular dissection. Optimizing the position and angle of dissection may enhance the safety of endoscopic surgery.

Materials and methods

Pulmonary artery dissection and histologic examination

All animals received humane care in compliance with the “Guide for the Care and Use of Laboratory Animals” published by the US National Institutes of Health (NIH Publication No. 85 − 3, revised 1985). The study is reported in accordance with the ARRIVE guidelines (https://arriveguidelines.org). The experimental protocol was approved by the Animal Experimental Committee of Fukuoka University (approval No. 2314111). A dog was sedated with an intramuscular injection of 0.05 mg/kg atropine sulfate, 10 mg/kg ketamine hydrochloride, and 7 mg/kg xylazine and subsequently intubated with an 8-mm endotracheal tube. Mechanical ventilation was maintained using inhalational sevoflurane. Through the thoracotomy incision, the main pulmonary artery was exposed and cut using a LigaSure Xp vessel-sealing system (Fig. 1). LigaSure Xp was connected with the Valleylab FT10 energy platform. The current and voltage change automatically depending on the condition of the tissue, and the maximum rated voltage is 244 V and the sealing time is approximately 4 s. The dog was euthanized using deep anesthesia with an overdose of thiopental sodium. The main pulmonary artery (vascular diameter 7 mm) was resected en bloc, immersed in 10% formaldehyde, embedded in paraffin blocks, and sectioned into 5-mm-thick slices. The sections were stained with hematoxylin and eosin or immunohistochemically with Elastica van Gieson(EVG)stain and examined by light microscopy. Thermal injury was defined as collagen fibers in the vessel wall that exhibited hyalinization and as smooth muscle cells in the tunica media that demonstrated shrinkage and loss of nuclear detail, consistent with coagulative necrosis.

Fig. 1.

Fig. 1

Intraoperative view of the resection of the main pulmonary artery of a dog using the LigaSure vessel-sealing system.

Burst pressure evaluation

Twenty-four excised swine internal carotid arteries (vascular diameter 5–7 mm), purchased from a meat processing company (TOKYOSHIBAURAZOUKI Co., Ltd., Tokyo, Japan), were cut orthogonally (n = 12, Fig. 2 A) or obliquely (n = 12, Fig. 2B) using a LigaSure Xp. Dissection at an angle of 90° and 45° to the long axis of the vessel was defined as orthogonal and oblique, respectively. The arteries were individually connected to a handy manometer (Nidec components INC., PG-100B) and fixed with Vicryl sutures (Ethicon, Inc). Each vessel stump was immersed in a container of saline solution and pressure was applied to each using a syringe. The pressure at which bubbles were first observed in the solution was defined as the burst pressure (Fig. 3, Video 1). The swine carotid artery was selected as a surrogate medium-sized artery to enable controlled angle testing.

Fig. 2.

Fig. 2

Picture showing the resection of swine internal carotid arteries using LigaSure orthogonally (A) and obliquely (B).

Fig. 3.

Fig. 3

Schema of this experiment. (Arrow: swine vessel)

Analysis and evaluation

Means of continuous variables, such as burst pressure in vessel stumps cut orthogonally and obliquely, were compared by t- tests. All statistical analyses were performed using EZR (Saitama Medical Center, Jichi Medical University, Saitama, Japan) software, a modified, graphical user interface of R commander (The R Foundation for Statistical Computing, Vienna, Austria) designed to include statistical functions frequently used in biostatistics.

Results

Microscopic findings

Hematoxylin and Eosin staining and elastica Van Gieson staining

The pulmonary artery was completely sealed and the elastic fibers within the vascular lumen were preserved. However, coagulation of denatured proteins, indicative of lateral thermal damage, was observed along the vessel wall. The length of thermal damage was measured at five points (1.2, 1.2, 1.3, 1.3, and 1.5 mm), and the average value was within 1.3 mm of the sealed section (Fig. 4).

Fig. 4.

Fig. 4

A Hematoxylin and eosin staining of a vessel stump. The stump (triangles) was completely closed by coagulation of denatured proteins. B Elastica Van Gieson staining of a vessel stump. Coagulation of denatured proteins, indicative of lateral thermal damage, was observed along the vessel walls from the stump. Scale bars: 200 μm (A and B).

Burst pressure of swine carotid arteries

Average vascular diameter and burst pressure did not differ significantly in swine carotid arteries that underwent orthogonal and oblique resection (6.25 mm vs. 6.17 mm, P = 0.813; 829.2 mmHg [range 608–1214 mmHg] vs. 949.1 mmHg [range, 593–1306 mmHg], P = 0.206) (Fig. 5).

Fig. 5.

Fig. 5

Burst pressure of swine internal carotid arteries resected orthogonally and obliquely. *P = 0.206.

Discussion

The present study found that the extent of lateral thermal damage to a canine pulmonary artery when utilizing LigaSure XP was approximately 1.3 mm. Additionally, burst pressure did not differ significantly in swine carotid arteries that underwent orthogonal and oblique dissections. Although the insertion angle of the device is limited during thoracoscopic surgery, particularly during U-VATS, oblique vascular resection may be a viable option.

The safety and efficacy of vessel-sealing systems have been extensively documented10. Energy sealing without reinforcement was found to enable secure treatment during the resection of pulmonary arteries up to 5 mm in diameter and pulmonary veins up to 7 mm11. Although small-diameter vessels patients undergoing thoracic surgery are generally dissected using vessel-sealing systems without ligation, vertical dissection can be challenging in minimally invasive procedures such as U-VATS, where instrument maneuverability is limited. Vertical dissection under these conditions may lead to excessive tension or thermal injury, thereby increasing the risk of inadvertent bleeding. The present study was designed to determine the appropriate vascular dissection technique utilizing vessel-sealing systems, thereby enhancing the safety of vessel dissection in patients undergoing minimally invasive surgery.

Thermal spread is an unavoidable consequence when tissues are dissected using energy devices, such as ultrasonic devices and vessel-sealing systems. Evaluation of the temperature and thermal spread associated with these devices found that thermal spread was greater using LigaSure than ultrasonic devices, despite the latter having significantly higher temperatures12,13. Although the use of LigaSure for pulmonary vascular dissection necessitates careful consideration of thermal spread, no reports to date have evaluated the extent of thermal damage in the pulmonary arteries of living animals. Examination of the effects of thermal damage on living tissues is important, as thermal spread is dependent on the moisture content within these tissues13. Due to ethical considerations, histological evaluation following the dissection of one pulmonary artery in a living dog was performed at five points within the vessel in the present study. The average thermal spread extended approximately 1.3 mm from the dissected site. The length of the five points did not exceed 2 mm; therefore, it is important that the pulmonary artery is dissected at least 2 mm from the main trunk regardless of the angle of dissection when utilizing LigaSure. To reduce cutting distance, small-diameter vessels are frequently cut vertically when using vessel-sealing systems. However, the angle of insertion of surgical instruments is constrained by small incisions, with forced insertion for vertical dissection frequently leading to excessive tension on the vessel, thereby increasing the risk of bleeding.

The present study investigated the burst pressure of orthogonal and oblique cutting angles on swine carotid arteries, as conducting pressure testing directly on pulmonary arterial specimens was technically and ethically challenging. Although the carotid artery possesses a thicker and more elastic wall than the pulmonary artery due to its role in high-pressure systemic circulation, previous reports have demonstrated comparable burst pressures after ligation of these two vessels14. Considering the results and anatomical similarity of the three-layer structure, the influence of the cutting angle on burst pressure is anticipated to exhibit analogous trends between carotid and pulmonary artery ligations.

Another previous study by Voegele et al. reported that perpendicular vessel seals were mechanically stronger than angled seals15. The apparent discrepancy between their results and the present findings may be attributable to differences in vessel size and experimental conditions. In particular, their study evaluated relatively larger vessels, whereas the present study focused on small-diameter vessels comparable to pulmonary artery branches. Vessel diameter, wall thickness, and elastic properties may influence the effect of sealing angle on burst pressure. The angle of artery dissection did not significantly affect burst pressure in the present study, suggesting that, when arteries are sealed uniformly, the pressure per unit area at the vascular stump remains constant. Maintaining vessel integrity is therefore more critical than the angle of dissection is minimizing excessive tension during the cutting and sealing process.

The current study has several limitations. First, a sample from a single dog was subjected to histological analysis, because of ethical considerations and concerns about animal welfare, as obtaining this sample required dissection of the main pulmonary artery trunk. Second, the pulmonary artery would have been the most appropriate vessel for the burst-pressure test. However, performing such testing on the pulmonary artery was technically and ethically difficult. Consequently, swine carotid arteries were used as a surrogate model. Finally, although it is desirable to resect vessels without applying tension to the vessel walls during energy sealing, tension was not measured in this study.

In conclusion, cutting angle has a negligible effect on burst pressure when small-diameter vessels are dissected using a vessel-sealing system without ligation. More critically, it is essential to minimize excessive tension on the vessel and to take precautions to prevent thermal injury.

Supplementary Information

Below is the link to the electronic supplementary material.

Supplementary Material 1 (24.5MB, mp4)

Acknowledgements

The authors wish to thank Takahisa Orita, Yukihiro Suzuki and Taichi Orikasa for contributing to the analysis and assisting with the preparation of the experiment.

Author contributions

Yuichiro Ueda conceived and designed the study, performed the experiments, analyzed the data, and drafted the manuscript. Jun-ichi Wakahara, So Miyahara, Hiroyasu Nakajima, and Yoshiko Masuda contributed to data acquisition and interpretation. Fumihiro Shoji and Takeshi Shiraishi provided critical revisions for important intellectual content. Toshihiko Sato supervised the study and approved the final version of the manuscript. All authors read and approved the final manuscript.

Funding

The authors have no funding sources to disclose.

Data availability

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Declarations

Competing interests

The authors declare no competing interests.

Footnotes

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Yoshida, K. et al. Feasibility of using a vessel sealing system in a human pulmonary lobectomy: a retrospective comparison of this procedure with or without a vessel sealing system. Ann. Thorac. Cardiovasc. Surg.20, 353–358 (2014). [DOI] [PubMed] [Google Scholar]
  • 2.Miyajima, M. et al. Vessel sealing system for video-assisted lung resection for cancer reduces chylothorax and bleeding. J. Thorac. Dis.13, 3458–3466 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Timm, R. W. et al. Sealing vessels up to 7 mm in diameter solely with ultrasonic technology. Med. Devices (Auckl). 7, 263–271 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Toishi, M. et al. Usefulness of vessel-sealing devices for ≤ 7 mm diameter vessels: a randomized controlled trial for human thoracoscopic lobectomy in primary lung cancer. Interact. Cardiovasc. Thorac. Surg.19, 448–455 (2014). [DOI] [PubMed] [Google Scholar]
  • 5.Oyama, S. et al. A new method using a vessel-sealing system provides coagulation effects to various types of bleeding with less thermal damage. Surg. Endosc. 35, 1453–1464 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Hefermehl, L. J. et al. Lateral temperature spread of monopolar, bipolar and ultrasonic instruments for robot-assisted laparoscopic surgery. BJU Int.114, 245–252 (2014). [DOI] [PubMed] [Google Scholar]
  • 7.Smith, C. T., Zarebczan, B., Alhefdhi, A. & Chen, H. Infrared thermographic profiles of vessel sealing devices on thyroid parenchyma. J. Surg. Res.170, 64–68 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Družijanić, N., Pogorelić, Z., Perko, Z., Mrklić, I. & Tomić, S. Comparison of lateral thermal damage of the human peritoneum using monopolar diathermy, harmonic scalpel and LigaSure. Can. J. Surg.55, 317–321 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Özyurtkan, M. O., Kaba, E. & Toker, A. Technological innovation in video-assisted thoracic surgery. J. Vis. Surg.3, 20 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Miyajima, M. et al. Interlobar division using vessel-sealing system in robot-assisted pulmonary lobectomy. JTCVS Tech.13, 211–216 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Okada, M. et al. Prospective feasibility study of sealing pulmonary vessels with energy in lung surgery. J. Thorac. Cardiovasc. Surg.157, 388–395 (2019). [DOI] [PubMed] [Google Scholar]
  • 12.Hayami, M. et al. Lateral thermal spread induced by energy devices: a Porcine model to evaluate the influence on the recurrent laryngeal nerve. Surg. Endosc. 33, 4153–4163 (2019). [DOI] [PubMed] [Google Scholar]
  • 13.Hayami, M. et al. Steam induced by the activation of energy devices under a wet condition May cause thermal injury. Surg. Endosc. 34, 2295–2302 (2020). [DOI] [PubMed] [Google Scholar]
  • 14.Lesser, T. G., Wolfram, F. & Boltze, C. Sealing of pulmonary arteries with ligasure: in vivo and ex vivo examinations. J. Thorac. Cardiovasc. Surg.145, 1525–1528 (2013). [DOI] [PubMed] [Google Scholar]
  • 15.Voegele, A. C., Korvick, D. L., Gutierrez, M., Clymer, J. W. & Amaral, J. F. Perpendicular blood vessel seals are stronger than those made at an angle. J. Laparoendosc. Adv. Surg. Tech.23, 669–672 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Material 1 (24.5MB, mp4)

Data Availability Statement

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.


Articles from Scientific Reports are provided here courtesy of Nature Publishing Group