Skip to main content
Health Science Reports logoLink to Health Science Reports
. 2026 Mar 10;9(3):e72063. doi: 10.1002/hsr2.72063

Assessment of Clinical Hematology Laboratory Performance Using Sigma Metrics and Associated Factors in Total Testing Process at the Dessie Comprehensive Specialized Hospital, Northeast, Ethiopia: A Cross‐Sectional Study

Zewudu Mulatie 1,, Endris Ebrahim 1, Mihreteab Alebachew 1, Alemu Gedefie 1, Bruktawit Eshetu 1, Mihret Tilahun 1, Habtu Debash 1, Yeshimebet Kassa 1, Ermiyas Alemayehu 1, Tesfaye Gessese 1, Dereje Mengesha Berta 2
PMCID: PMC12975641  PMID: 41822766

ABSTRACT

Background and Aims

Clinical laboratory test results play a fundamental role in clinical decision making, influencing approximately 70% of medical decisions. Numerous studies have shown that errors can occur throughout the total testing process. This study aimed to assess the performance of the Clinical Hematology Laboratory and identify associated factors using sigma metrics across the total testing process at Dessie Comprehensive Specialized Hospital.

Methods

A cross‐sectional study was conducted from July1, 2023, to September 30, 2023. The study included all eligible laboratory samples and corresponding test requests during the study period using a consecutive sampling technique. Data for each variable were collected using a pre prepared checklist and record formats by trained laboratory professionals. The data were entered into EPI data version 3.1 and exported to Stata version 17 for analysis. Logistic regression analysis was performed, and the laboratory's total testing process performance was evaluated using the sigma scale.

Result

The overall prevalence of clinical hematology laboratory errors in the total testing process was 12.44%. Most errors occurred during the pre‐analytical phase (77.8%), followed by the post‐analytical phase (19.77%) and the analytical phase (2.36%). The overall sigma value of the hematology laboratory was 2.7, indicating substandard performance. The sigma values for the pre‐analytical, analytical, and post‐analytical phases were 2.6, 3.2, and 2.8, respectively.

Conclusion and Recommendation

This study found that errors in the clinical hematology laboratory were common, and overall performance was below the acceptable sigma threshold. Therefore, hospital administration should provide need‐based training for all staff involved in hematology laboratory sample collection and processing. Strengthening adherence to standard operating procedures and implementing continuous quality improvement strategies are essential to enhance laboratory performance and reduce errors.

Keywords: analytical errors, hematology laboratory performance, post‐analytical errors, pre‐analytical errors, sigma metrics, sigma metricssigma metricssigma metrics


Abbreviations

CBC

complete blood count

CLIA

clinical laboratory improvement amendment

HCT

hematocrit

IPD

Inpatient Department

ISO

International Organization for Standardization

MRN

medical record number

OPD

outpatient department

QIs

quality indicators

TAT

turnaround times

TTP

Total testing process

WHO

World Health Organization

1. Introduction

1.1. Background

According to research, the utilization of clinical laboratory test results in clinical decision making has become an essential component of medical practice [1, 2]. Approximately 70% of clinical decisions related to hospitalization, admission, prescription, and release depend on laboratory data; however, slight variations between 60% and 80% have been reported [1, 3, 4]. In clinical medicine, screening, diagnosis, and therapeutic monitoring for many diseases and disorders rely heavily on hematology laboratory results. Therefore, ensuring quality in laboratory services is crucial throughout the total testing process (TTP) to produce accurate, precise, and timely results for optimal patient care [5, 6, 7]. The total laboratory testing process is complex and encompasses all steps from the medical decision to perform a test to the delivery of the results. It consists of three main phases: pre‐analytical, analytical, and post‐analytical [8, 9].

Patient safety is a major global public health concern [10]. A Study in Malaysia reported a diagnostic error rate of 3.6% [11]. Similarly, data from the United States indicate that diagnostic errors occur in approximately 5% of outpatient cases [12, 13]. Moreover, evidence from low and middle‐income countries suggests that about 134 million adverse events and 2.6 million deaths occur annually due to medical errors. Among low income countries, South Asia and Western Sub‐Saharan Africa had the highest mortality rates, with 1.9 million and 1.2 million deaths, respectively [14, 15].

Laboratory errors contribute to several clinical problems, including delayed diagnosis, additional laboratory testing, misdiagnosis, overmedication, under‐medication, incorrect treatment, increased healthcare costs, decreased patient satisfaction, injury, and death [16, 17]. Consequently, the quality of laboratory services has become a primary concern for many governments worldwide [18, 19].

In Africa, the magnitude of laboratory errors is higher than in other continents due to inadequate health care infrastructure, a lack of trained personnel, poor management systems, substandard quality control measures, equipment shortages, weak government policies, low staff motivation, and poverty [20, 21]. As a result, most clinical laboratory services in Africa remain below standard, and laboratory participation in accreditation programs is below the expected level [22]. According to ISO15189 reports, the performance of African laboratories is poor. Similarly, the majority of laboratories in Ethiopia are not accredited and fail to meet quality standards. However, laboratory errors still receive little attention and remain difficult to address [22, 23, 24, 25].

Recently, several laboratories worldwide have begun applying sigma metrics principles across all phases of TTP to improve quality, reduce defects, streamline the registration process, decrease patients' wait times, lower costs, and enhance customer satisfaction [26]. During pre‐analytical phases, sigma metrics help improve the quality of information on test requisitions, specimen collection, patient identification, and transportation. In the analytical phase, it aids e in minimizing laboratory testing errors, misinterpretations, misreading and misjudgments of results [27]. Although quality control (QC) programs and quality assurance measures have been established to identify defects, these alone are insufficient for providing a direct and integrated assessment of laboratory process performance [3, 28]. Therefore, sigma metrics serve as an essential tool for evaluating laboratory performance and achieving intended quality goals [28].

Sigma metrics is a quantitative tool used to assess laboratory process performance by measuring defects per‐million opportunities (DPMO) [29]. It is more effective for evaluating laboratory errors than simply counting defects, as sigma metrics analysis assesses the quality of the TTP by providing mechanisms for error correction and performance improvement through models such as the Define, Measure, Analyze, Improve, and Control model. Additionally, sigma metrics help determine the appropriate frequency and number of quality control measures required in a laboratory [30, 31].

Despite efforts to improve performance in African countries, significant challenges persist. Many clinical laboratories continue to provide sub‐optimal services, and the overall quality of their outputs remains poor [25, 32]. Evidence from Ethiopia also supports this concern [25]. However, very few studies have been conducted to assess the overall distribution of clinical hematology laboratory errors, and limited research has evaluated process performance by using sigma metric in Ethiopia, specifically in the study area. We hypothesized that the clinical hematology laboratory's TTP demonstrates suboptimal performance, with sigma metrics below the acceptable threshold, indicating the need for targeted quality improvement interventions. Therefore, this study aimed to assess the performance of the clinical hematology laboratory's TTP using sigma metrics.

2. Materials and Methods

2.1. Study Area, Design and Period

A cross‐sectional study was conducted at the Clinical Hematology Laboratory of Dessie Comprehensive Specialized Hospital from July 1, 2023, to September 30, 2023. The Hospital laboratory consists of multiple sections, including Phlebotomy, Hematology, Parasitology, Clinical Chemistry, Urinalysis, Serology, Emergency and Microbiology. Among these, the Hematology laboratory serves as the primary unit, performing tests such as complete blood cell count (CBC), Hematocrit (HCT), peripheral morphology, body fluid analysis, erythrocyte sedimentation rate (ESR) and coagulation tests. The Clinical Hematology Laboratory processes approximately 70,000 samples annually.

2.2. Study Population

The study included all phlebotomists, laboratory professionals, and laboratory test requests along with their respective blood samples submitted for analysis.

2.3. Inclusion and Exclusion Criteria

2.3.1. Inclusion Criteria

Test requests accompanied by blood samples for Clinical Hematology Laboratory analysis were included.

2.3.2. Exclusion Criteria

Tests requested for non‐routine Hematology Laboratory tests were excluded. In addition, tests requiring reagents that were inconsistently available during the study period were also excluded.

2.4. Variables

2.4.1. Dependent Variables

Clinical Hematology laboratory performance and frequency of errors.

2.4.2. Independent Variables

Sample collection site, work shift and adherence to SOP.

2.5. Operational Definitions

Clinical laboratory error: Any defect or mistake occurring at any phase of the total testing process.

Pre‐analytical errors: Mistakes that occur before sample analysis.

Analytical errors: Errors that arise during testing or analysis.

Post‐analytical errors: Mistake that occur after analysis or testing [33].

Total error: The sum of all errors that may occur throughout the entire testing process [34].

Critical values: Results that fall significantly above or below the reference range and require immediate medical intervention.

Hemolysis: The destruction of red blood cells either in vitro or vivo, leading to visibly red plasma in an EDTA‐anticoagulated, settled or centrifuged blood sample.

Lipemic sample: Plasma with a visible milky appearance due to high lipid content.

Clotted sample: Plasma in a solidified state that may obstruct analyzer probes [35].

In a sufficient sample: A collected sample with a volume lower than the acceptable limit.

Overfilled sample: A collected sample exceeding the acceptable volume.

Delayed sample: A sample left at room temperature for more than 2 h without being analyzed.

Turnaround time: The duration between sample collection and the release of test results to the physicians.

Phlebotomist: A laboratory or healthcare professional responsible for collecting blood specimens.

Work shift: The period during which the Clinical Laboratory is fully operational. It follows a two shift system, each lasting 4 h (1st shift from 8:00 to 12:30 AM and 2nd shift from 1:00 to 5:00 PM).

Sigma metrics: A measure of laboratory performance that indicates how many standard deviations (SDs) fit within the tolerance limit from the assay mean. A test is considered unacceptable if the total observed error exceeds the total allowable error. In this study, laboratory performance was evaluated based on the sigma value, as detailed in the table below [36, 37] (Table 1).

Table 1.

Sigma value interpretations table.

Sigma value Defect per million opportunities Yield Level
1 690,000 30.9 Unacceptable
2 308,537 69.1 Poor
3 66,807 93.3 Marginal
4 6210 99.4 Good
5 233 99.98 Excellent
6 3.4 99.99966 World class

Unacceptable Test Performance: The average sigma value is less than or equal to three [37].

Acceptable Test Performance: The average sigma value will be greater than three [37].

2.6. Sample Size Determination and Sampling Technique

The study included all eligible Clinical Hematology Laboratory samples and their corresponding test requests collected during the study period. A consecutive sampling technique was used to include all test requests submitted for Clinical Hematology Laboratory testing.

2.7. Data Collection and Laboratory Methods

Data were collected using a standardized, pre‐tested checklist designed to evaluate errors in the TTP within the Clinical Hematology Laboratory. The checklist was developed based on variables from previous studies and accreditation standards from the College of American Pathologists [17, 35, 38]. After finalizing the checklist, trained data collectors and supervisors were assigned.

All data collectors were laboratory professionals with training in laboratory quality management systems. They were instructed on how to collect relevant data to assess the completeness of standard laboratory request forms, including patient age, gender, physician's signature, location, date and authorized request format. Additionally, they were trained on how to evaluate specimen quality, as well as analytical and post‐analytical quality indicators (QIs) based on the checklist. Data collectors were assigned to sample collection and Hematology laboratory sections, where they systematically recorded observations across all TTP phases. Additionally, face‐to‐face interviews were conducted to collect socio demographic data from phlebotomists and laboratory personnel. Pre‐analytical variables related to laboratory specimen handling were assessed through direct observation upon sample arrival at the Clinical Hematology Laboratory. Post‐analytical variables were recorded after analysis, ensuring that all checklist data were systematically entered.

2.8. Quality Assurance

To ensure data quality, a standardized observational checklist was used. A pre‐test was conducted on 40 samples with their corresponding test requests at Borumeda Hospital to assess feasibility and ensure the validity of study tools. Based on the pre‐test findings, necessary modifications were made to the checklist. Furthermore, data quality was ensured by having data collectors verify the completeness of the checklist in real time while gathering relevant information at each phase of the Clinical Hematology laboratory testing process. Supervisors provided daily feedback and corrections to investigators throughout the data collection period. Additionally, the collected data was regularly reviewed for completeness, accuracy, and clarity. Data analysis, reporting, and interpretation adhered to the Guidelines for Reporting Statistics for Clinical Research in Urology [39], SAMPL guidelines [40], and the STROBE checklist for observational studies [41].

2.9. Data Analysis and Interpretation

Data was coded and entered to EPI data version 3.1 for completeness and validation before being exported to Stata version 17 for analysis. Descriptive statistics were used to summarize the overall distribution of errors in the Clinical Hematology Laboratory. The association between laboratory performance and the explanatory variable was assessed using bivariable and multivariable logistic regression models. Explanatory variables with a p‐value < 0.25 in bivariable analysis were included in the multivariable logistic regression model to control for potential confounders. Model fitness was assessed using the Hosmer–Lemeshow goodness‐of‐fit test. Crude odds ratio (COR) and adjusted odds ratio (AOR) within the 95% confidence interval (CI) were used to determine the strength of associations. The p‐value < 0.05 was considered statistically significant.

The laboratory test process was assessed using the Sigma scale, which quantifies errors and defects in the total testing process. This approach monitors each phase, counts errors per million opportunities, and converts them into sigma metrics using a standard sigma conversion table. The calculation steps included:

  • 1.

    DPMO = (Number of defects × 1,000,000)/number of defect opportunities

  • 2.

    Convert DPMO to Sigma level: first, find the probability of a defect occurring: P = DPMO/1,000,000

  • 3.

    Compute Z‐score: Z= Φ−1(1 − P). Find the Z value from a Z‐table.

  • 4.

    Convert Z‐score to Sigma level: Σ=1 + Z

  • 5.

    0.8406

Additionally, sigma values were calculated using the formula: Sigma= (TEa – Bias %)/CV. So, the bias of all parameters was calculated using the following formula: Bias (%) = [(our laboratory mean of IQC data ‐ target mean of IQC data)/target mean of IQC data] ×100. The CV for all parameters was calculated from the observed mean and SD of the daily IQC in the study period by using the following formula: CV (%) = (SD/mean) x 100. In addition, the overall observed error committed within the laboratory was calculated as: TE= Bias (%) + CV (%) × 2. For, improved accuracy, sigma scores were calculated using an online sigma calculator (Defects Per Million Calculator, available at

https://www.westgard.com/six-sigma-calculators.htm).

2.10. Ethical Consideration

Ethical approval was obtained from the College of Medicine and Health Sciences Research and Ethics Review Committee of Wollo University. The objective and purpose of the study were explained to the medical director, and a permission letter was obtained from Dessie comprehensive specialized Hospital before the actual data collection began. After providing an explanation of the possible benefits and risks, written informed consent was obtained from study participants. To ensure confidentiality of data, study subjects were identified using codes, and only authorized people had access to the collected data. Identified errors were linked to the responsible people for better patient management and quality improvement.

3. Result

3.1. General Information

During the study period, a total of 12,402 hematology tests were requested. Complete blood count tests accounted for the majority, comprising 9728 (78.44%) of the total, followed by coagulation tests with 1547 (12.47%). Of all the test requests, 6207 (50.05%) were requested from outpatient department (OPD), while 2550 (20.56%) were from the MCH unit. Most test requests were ordered during the morning shift (57.32%) and, 8255 (66.56%) of samples were collected by laboratory professionals. All sample records were documented manually (Table 2).

Table 2.

General information for the hematology laboratory at Dessie Comprehensive Specialized Hospital, 2023.

Variables Frequency Percentage
Total test request assessed 12,402 100
Test requested for CBC 9728 78.44
Test requested for ESR 680 5.48
Test requested for Manual HCT 217 1.75
Test requested for PM 230 1.64
Test request for coagulation 1547 12.47
Sample requested site OPD 6207 50.05
IPD 1193 9.62
Emergency 730 5.89
MCH 2,550 20.56
Unknown addresses 1722 13.88
Sample collection shift First (morning) 7109 57.32
Second (afternoon) 5293 42.68
Type of sample collectors Laboratory professionals 8255 66.56
Non‐Laboratory healthy professionals 4147 33.44
Recording system Manually 12,402 100
laboratory information system 0 0

3.2. Overall Prevalence of Errors and Performance Levels By Sigma Metrics in Hematology Laboratory

The overall prevalence of hematology laboratory errors was 12.44%. Most errors occurred in the pre‐analytical phase with 43,148 errors (77.87%), followed by the post‐analytical phase with 10,954 (19.77%) and the analytical phase had 1305 (2.36%). The overall sigma value for the hematology laboratory was 2.7, with mean sigma values of 2.6, 3.2, and 2.8 for the pre‐analytical, analytical, and post‐analytical phases, respectively.

3.3. The Sigma Metrics Levels of the Pre‐Analytical Phase Related to Missed Information on Hematology Laboratory Requests

The sigma values for inappropriate and unauthorized test requests, missing MRN, and missed test orders were greater than three. The average sigma value for request form completeness was below three (Table 3).

Table 3.

Sigma metrics performance levels of Hematology Laboratory request forms at the Dessie Comprehensive Specialized Hospital, Northeast Ethiopia, 2023.

Variables No. of Error Percentage Total Sigma value
Appropriate and authorized requests 721 5.81 12,402 3.1
MRN 268 2.16 12,402 3.5
Patient age 2671 21.54 12,402 2.3
Sex of patient 2277 18.36 12,402 2.5
Signature of the physician 4917 39.65 12,402 1.8
Clinical history of the patient 12402 100 12,402 < 3
Patients address 6239 50.31 12,402 1.5
Name of sender address/ward 1747 14.08 12,402 2.6
Date of order 4867 39.24 12,402 1.8
Test ordered 0 0 12,402 > 6.0
Time of sample collection 1854 14.95 12,402 2.6
Handwriting legible 984 7.93 12,402 3.0
Total 38,947 26.17 148,824 2.2

3.4. Medical Record Number (MRN)

3.4.1. The Sigma Metrics Levels of Pre‐Analytical Phase Related to Specimen Quality, Collection, Preparation, Storage and Transportation in the Hematology Lab

In the assessment of specimen quality, 704 (5.68%) samples were deemed unsuitable for analysis. The frequencies of hemolyzed, incorrectly labeled, clotted, and insufficient samples were 173 (1.39%), 226 (1.82%), 258 (2.08%), and 363 (2.93%), respectively, each with sigma values greater than three. Additionally, the sigma metrics for lost test requests and lost samples were ≥ 4.2. Overall, the pre‐analytical performance level was below the three sigma value (Table 4).

Table 4.

Sigma metrics levels of Hematology laboratory in pre‐analytical phases related to specimen quality, collection, preparation, storage and transportation, Dessie Comprehensive Specialized Hospital, Northeast, Ethiopia, 2023.

Variables No. of error Percentage Total (N/%) Sigma value
Hemolyzed samples 173 1.39 12,402 3.7
Clotted samples 258 2.08 12,402 3.6
Insufficient volume 363 2.93 12,402 3.4
Incorrect containers 0 0 12,402 > 6
Incorrectly labeled specimen 226 1.82 12,402 3.6
Delayed samples 116 0.94 12,402 3.9
Wrong sample transportation 52 0.42 12,402 4.2
Sample lost 45 0.36 12,402 4.2
Request lost 32 0.26 12,402 4.3
Unacceptable quality smear 24 10.43 230 2.8
Wrong sample storage 35 100 35 < 3
Blood mixed with an anticoagulant improperly 568 4.75 11,955 3.2
Improperly sealed capillary tube 21 9.68 217 2.9
Incorrect anticoagulant‐to‐blood ratio 1416 11.84 11,955 2.7
Patient identified improperly 234 2.32 10,076 3.5
Incorrect tourniquet application time 572 5.88 9721 3.1
Blood is unmixed before analysis 66 0.63 10,408 4.0
Total 4201 2.53 166,215 3.5
Sample unsuitable for analysis 704 5.68 12,402 3.1
Grand total pre‐analytical errors 43,148 13.70 315,039 2.6

3.5. Total Frequency (N)

3.5.1. Sigma Metrics Levels of Analytical Phase in Hematology Laboratory

Among the 27,432 QIs assessed during the analytical phase, 1305 (4.79%) errors were identified. Nonlinear and questionable results with sigma values below three were reported without further testing or morphological examination. Internal quality control results passed consistently, with sigma values above 6. Overall, the analytical phase achieved a sigma value greater than 3 (Table 5).

Table 5.

Sigma metrics levels of Hematology Laboratory in analytical phase at Dessie Comprehensive Specialized Hospital, Northeast Ethiopia, 2023.

Variables No. of error Percentage Total Sigma value
Daily IQC was not performed 42 72.73 66 1.2
IQC result failed 0 0 14 > 6
Background not checked 5 7.58 66 3.0
Preventive maintenance was not performed 17 25.76 81 2.1
Equipment malfunction observed 0 0 66 > 6
Reference range unavailable for parameters 0 0 18 > 6
Electric power inconsistency during analysis 521 4.81 10,823 3.2
Non‐linear results released without retesting 18 100 18 < 3
Reagents expired 2 3.03 66 3.4
Inappropriate reagent storage condition 0 0 66 > 6
Reagent stockout during analysis 218 2.06 10,606 3.6
Methods not updated upon the new reagent 0 0 12 > 6
Improperly filled ESR tube 44 8.37 526 2.9
The position of ESR tube is wrong 31 5.89 526 3.1
Delay in ESR results reading 38 7.22 526 3.0
ESR sample analyzed at the wrong temperature 0 0 526 > 6
Questionable results were not retested 79 26.16 302 2.2
Critical results were not checked by PM 98 68.53 143 1.1
HCT tube leaked 32 17.78 180 2.5
HCT tube broken 14 7.78 180 3.0
The speed of the centrifuge was adjusted improperly 6 3.33 180 3.4
The time of the centrifuge was adjusted improperly 7 3.89 180 3.3
HCT results were measured incorrectly 10 5.56 180 3.1
The smear was not properly air dried 5 2.34 213 3.5
Incorrect preparation of the working solution for PM 4 6.06 66 3.1
Smear stained at the incorrect time 79 37.09 213 1.9
Incorrectly washed smear 17 7.98 213 3.0
Cuvettes are not clean for coagulation 18 1.51 1189 3.7
Total 1305 4.79 27,245 3.2

3.5.2. Sigma Metrics Level of Post‐Analytical Phase in Hematology Laboratory

Among the 103,261 QIs assessed in the post‐analytical phase, 10,954 (10.61%) errors were identified. Sigma values for failures to communicate critical results to physicians and releasing without verification were below 3. Overall, the mean sigma value for the post‐analytical phase was less than three (Table 6).

Table 6.

Sigma metrics performance level of Hematology Laboratory in post analytical phase at Dessie Comprehensive Specialized Hospital, Northeast Ethiopia, 2023.

Variables No. of error Percentage Total Sigma value
Critical values were not communicated to the physician immediately 106 71.62 148 1.0
Results released without result verification 9867 85.90 11,487 0.5
Test results unrecorded 152 1.32 11,487 3.8
Results released without TAT 596 5.19 11,487 3.2
Result reported without standard unit 13 0.11 11,352 4.6
Samples were not retained/stored as per the policy 89 0.77 11,487 4.0
Laboratory results lost 49 0.43 11,487 4.2
Results reported with incorrect standard unit 0 0 11,352 > 6
Results reported without reference range 65 0.57 11,487 4.1
Result reported by unauthorized personnel 17 0.15 11,487 4.5
Total 10954 10.60807 103,261 2.8

3.6. Turnaround Time (TAT), Defect Per Million Opportunities (DPMO)

3.6.1. Location of Sample Collection with Pre and Post‐Analytical Errors

Out of the 12,402 hematology test requests evaluated, 755 (6.09%) lacked patient age information. Among 704 rejected specimens, 71 (0.6%) hemolyzed samples requested from the emergency section. Of the 596 test results released beyond the expected TAT, 273 (2.38) were reported to the OPD (Table 7).

Table 7.

Association of patient location and shift with pre‐analytical and post‐analytical errors at Dessie Comprehensive Specialized Hospital, 2023.

Variables Location of patients (Clinic/Ward)
OPD IPD Emergency MCH Unknown
Patient age missed 753 (6.07) 232 (1.87) 471 (3.80) 460 (3.71) 755 (6.09)
Gender of patient missed 873 (7.04) 234 (1.89) 277 (2.23) 4 (0.03) 889 (7.71)
Signature of the physicians missed 1645 (13.26) 706 (5.69) 478 (3.85) 994 (8.01) 1094 (8.82)
Date of ordered missed 1690 (13.63) 550 (4.43) 481 (3.88) 902 (7.27) 1334 (10.76)
Hemolyzed specimen 38 (0.31) 71 (0.57) 32 (0.26) 12 (0.097) 19 (0.15)
Mislabeled specimen 34 (0.44) 89 (0.12) 61 (0.53) 24 (0.20) 18 (0.29)
Specimen rejection 293 (2.36) 107 (0.86) 45 (0.36) 133 (1.07) 126 (1.02)
Unrecorded results 42 (0.36) 28 (0.24) 64 (0.56) 6 (0.05) 12 (0.10)
Results released out of TAT 273 (2.38) 24 (0.21) 20 (0.17) 177 (1.54) 102 (0.89)

3.7. Factors Associated with Prolonged TAT

Factors associated with prolonged TAT were analyzed using a logistic regression model, with model goodness‐of‐fit was assessed by the Hosmer–Lemshow test (p = 0.056). In the bivariate logistic regression analysis, the first work shift (8.00 to 12:30 AM), patient location (IPD, OPD and unknown), and lack of adherence to SOP were significantly associated with the prolonged TAT. After adjusting for potential confounders, multivariate analysis identified that first shift (AOR: 1.92; 95% CI: 1.60–2.30; p < 0.001), IPD (AOR: 1.89; 95% CI: 1.62–2.32; p < 0.001), OPD (AOR: 1.98; 95% CI: 1.23–2.64; p < 0.001), unknown address (AOR: 2.23; 95% CI: 1.37–3.64; p < 0.001), and lack of adherence to SOPs (AOR: 2.88; 95% CI: 2.34–3.54; p < 0.001) as significant predictors of delayed TAT (Table 8).

Table 8.

Logistic regression analysis of prolonged TAT and explanatory variables in clinical Hematology Laboratory at DCSH, Northeast Ethiopia, 2023.

Variable Category Prolonged TAT COR (95% CI) AOR (95% CI) P‐value
Yes No
Work shift First 421 6061 2.14 (1.67–3.54) 1.92 (1.60–2.30) < 0.001*
Second 175 4830 1a 1a 1a
Ward IPD 26 1044 2.56 (2.21–2.95) 1.89 (1.62–2.32) < 0.001*
OPD 272 5562 2.14 (1.39–3.257) 1.98 (1.23–2.64) < 0.001*
MCH 176 2141 1.45 (0.79–2.67) 1.35 (0.74–2.46) 0.23
Unknown 102 1499 3.59 (2.33–5.54) 2.23 (1.37–3.64) < 0.001*
Emergency 20 645 1a 1a 1a
Lack of adherence to SOP Yes 483 6509 3.35 (2.70–4.15) 2.88 (2.34–3.54) < 0.001*
No 113 4382 1a 1a 1a
*

Stastical significance.

Reference Categories [1a], Crude Odds Ratio (COR), Confidence Interval (CI), Adjusted Odds ratio (AOR), Laboratory Information System (LIS), Inpatient Department (IPD), Outpatient Department (OPD) and Standard Operating Procedure (SOP).

3.8. Factors Associated with Sample Rejection

Factors contributing to specimen rejection were analyzed using a logistic regression model, with the model's goodness‐of‐fit assessed through the Hosmer–Lemeshow test (p = 0.631). In the bivariate logistic regression analysis, patient location (IPD and Unknown) and non‐adherence to SOPs were significantly associated with specimen rejection. The multivariate logistic regression analysis confirmed an independent association between IPD (AOR: 1.99; 95% CI: 1.58–2.50; p < 0.001), unknown address (AOR: 1.59; 95% CI: 1.28–1.98; p < 0.001), and non‐adherence to SOPs (AOR: 1.60; 95% CI: 1.38–1.86; p < 0.001) with specimen rejection (Table 9).

Table 9.

Logistic regression analysis of sample rejection and explanatory variables in the clinical Hematology Laboratory at DCSH, Northeast Ethiopia, 2023.

Variable Category Rejection COR (95% CI) AOR (95% CI) P‐value
Yes No
Work shift First 405 6704 1.01 (0.87–1.18) 0.95 (0.79–1.13) 0.58
Second 299 4994 1a 1a 1a
Ward IPD 107 1086 2.10 (1.66–2.66) 1.99 (1.58–2.50) < 0.001*
MCH 133 2417 1.10 (0.89–1.37) 1.11 (0.90–1.37) 0.33
Emergency 45 685 1.93 (1.37–2.73) 1.33 (0.96–1.83) 0.09*
Unknown 126 1596 2.25 (1.76–2.88) 1.59 (1.28–1.98) < 0.001*
OPD 293 5914 1a 1a 1a
Lack of adherence to SOP Yes 302 4541 2.00 (1.69–2.37) 1.60 (1.38–1.86) < 0.001*
No 352 7207 1a 1a 1a
*

Stastical significance.

Reference categories [1a], Crude odds ratio (COR), Confidence interval (CI) Adjusted Odds ratio (AOR), Inpatient Department (IPD), Outpatient Department (OPD) and Standard Operating Procedure (SOP).

4. Discussion

This study aimed to evaluate the performance of hematology laboratory in identifying and managing errors throughout the TTP. The findings revealed that the overall performance of the laboratory was suboptimal, with a sigma value of 2.7. This result is comparable to a study conducted in Gondar, Ethiopia, which reported a sigma value of 2.2 [42]. However, the performance of our laboratory was lower than that reported in studies from Tehran, Iran [1], Guangdong, China [43], India [44], Mysore, India [45] and Turkey [3], all of which reported performance levels exceeding three sigma. This discrepancy may be attributed to variations in QIs, test ordering systems, laboratory infrastructure, and the expertise and experience of healthcare professionals. Notably, the laboratories in these studies included fewer comprehensive QIs, had more advanced laboratories, and utilized an electronic system for ordering all tests, unlike the current study.

The poor performance of the Dessie hematology laboratory (sigma value < 3) may be attributed to several operational and systemic factors. Non‐adherence to standard operating procedures, limited training opportunities for laboratory personnel, the frequent collection of blood specimens by non‐laboratory professionals, and a high daily workload likely contribute to increased human errors. Moreover, the continued reliance on manual record‐keeping and did not use laboratory information systems increase the risk of mislabeling, sample loss, and delayed result reporting. The absence of regular internal quality control for hematology tests and the laboratory's limited participation in external quality assessment programs further hinder continuous quality improvement. To enhance laboratory performance, several practical and sustainable interventions can be implemented. Regular in service training should be provided to strengthen staff competency and ensure strict adherence to standard operating procedures. Assigning specimen collection exclusively to trained laboratory professionals would reduce pre‐analytical errors. Introducing a computerized laboratory information management system can minimize transcription errors and improve traceability of samples and results. Furthermore, establishing a robust internal quality control system for all hematological tests, coupled with consistent participation in national or international external quality assessment schemes, would enhance analytical reliability. Furthermore, adopting Lean and Six Sigma quality management approaches could help reduce variation and optimize workflow efficiency.

In this study, the overall prevalence of clinical hematology laboratory errors in the TTP was 12.44%, with pre‐analytical, analytical, and post‐analytical error rates of 9.68%, 0.29%, and 2.46%, respectively. The finding is comparable to studies conducted in Dessie, Ethiopia [11] and Aurangabad, India [46], which reported overall error rates of 15.3% and14.9%, respectively. However, it is higher than studies conducted in Anokye, Ghana [47], Assah, Saudi Arabia [48], Teerthanker Mahaveer, India [49], Imam, Iran [50] and Lahore, Pakistan [51], which reported total error rates of 4.7%, 4.35%, 0.17%, 6.3%, and 1.2%, respectively. Conversely, the overall error frequency in our study was lower than hat reported in Gondar, Ethiopia [42], Gondar, Ethiopia [52], Addis Ababa [24] and Wolega, Ethiopia [53], which reported defect rates of 26%, 36.8, 28.5% and 58.2%, respectively. The discrepancy among these findings could be attributed to differences in the QIs included and the operational definitions used to assess errors. Additionally, variations in the sample size, technical capacity and the inclusion of samples from different laboratory units rather than focusing solely on hematological tests, may have contributed to the observed differences.

In the pre‐analytical phase, the mean sigma value was below the acceptable limit ( < 3 sigma). Possible root causes for this lower performance include inadequate patient preparation, improper sample collection techniques, insufficient mixing of anticoagulated blood, and improper transportation or storage of specimens. Additionally, factors such as limited experience, negligence, lack of attention, high workload, and non‐compliance with standard operating procedures (SOPs) may have contributed to inconsistencies. In contrast, the performance of our laboratory in the analytical and post‐analytical phases was marginal (3.2) and poor (2.8), respectively. The lower performance in the analytical phase may be attributed to a shortage of trained personnel, inadequate reagent supply, insufficient preventive maintenance, a small sample size, and a lack of verification and awareness regarding critical value checks. Poor performance in the post‐analytical phase could be due to failure to communicate critical test results to physicians, unverified test results, and delayed release of results beyond the established turnaround time (TAT). Additional contributing factors may include increased workload, weak implementation of laboratory policies, an inefficient reporting system, inadequate infrastructure, and electrical fluctuations. In this study, the sigma value for incomplete filling of test request forms was below the acceptable limit ( < 3). This finding is similar to a study done Gondar, Ethiopia, which also reported a sigma value below 3 [42]. However, our performance was lower than that reported in studies conducted in India [54] and Romanian [43], which reported optimal performance ( > 3) which observed optimal performance ( > 3). This variation could be attributed to differences in the experience and consistency of physicians attending to patients at the study site, as most of the physicians were medical students.

According to the current study, the laboratory achieved undesirable performance in specimen quality, with a sigma value of 2.6, indicating that performance in this area was lower than that reported in a study conducted in eastern India ( > 4 sigma) [55]. The lower performance in our laboratory may be attributed to non‐adherence to standard operating procedures (SOPs), reliance on manual recording systems, work overload, and sample collection by untrained personnel.

On the other hand, the laboratory achieved marginal performance (3.4–3.7 sigma) for hemolysis, wrong labeling, clotting, and incorrect sample volume. Although these values are within the acceptable limit, they do not represent world‐class performance. Studies conducted in India [54], China [43], Romania [43] and the United States [56] reported better performance (very good to world‐class) for these parameters. The lower performance observed in our study could be due to improper tourniquet application during sample collection, insufficient mixing of blood with anticoagulant, negligence, and suboptimal sample transportation conditions.

In this study, the performance of verifying nonlinear results through retesting and examining questionable results by morphology was undesirable ( < 3 sigma). This indicates that our laboratory's performance is lower than studies conducted in Guangdong, China [43], Mysore, India [45] and Turkey [3], where performance levels were reported as desirable. The poor performance in our laboratory may be due to a lack of verification and awareness regarding the importance of rechecking critical values.

The current study also indicated poor performance ( ≤ 3 sigma) in communicating critical test results to physicians, verifying test results, and releasing results within the established turnaround time (TAT). Our laboratory's performance in these areas is lower than that reported in China [43], India [45] and Turkey [3]. Contributing factors may include increased workload, weak laboratory policies, inefficient reporting systems, inadequate infrastructure, and electrical fluctuations.

Multivariate logistic regression analysis revealed that prolonged TAT was significantly associated with sample collection from unknown wards (AOR = 2.23, 95% CI: 1.37–3.64), IPD (AOR = 1.89, 95% CI: 1.62–2.32), OPD (AOR = 1.98, 95% CI: 1.23–2.64), the first work shift (AOR = 1.92, 95% CI: 1.60–2.30), and lack of adherence to SOPs (AOR = 2.88, 95% CI: 2.34–3.54). These findings are consistent with studies conducted in Gondar, Ethiopia (2023) [42] and the Armed Force hospital, Ethiopia [57]. Prolonged TAT during the first (morning) shift may be due to work overload and fatigue, while delays associated with samples from unknown wards, IPD, and OPD may relate to the physical distance between sample collection sites and the analysis room. In the study area, samples were collected from wards located far from the laboratory.

Furthermore, the study showed that patient address (IPD and unknown) and non‐adherence to SOPs were significantly associated with unsuitable samples. Lack of adherence to SOPs nearly doubled the likelihood of sample rejection (AOR = 1.67, 95% CI: 1.38–1.86) compared to adherence, consistent with findings from Gondar, Ethiopia (2023) [42] and Kenya [58]. Poor SOP adherence in our setting may be due to weak standardization among laboratory personnel, sample collection by non‐laboratory staff, work overload, and inadequate supervision. Similarly, IPD samples had nearly twice the rejection rate (AOR = 1.99, 95% CI: 1.58–2.50) compared to OPD samples, a finding supported by studies in Gondar, Ethiopia (2023) [42], Hawassa, Ethiopia [59], Nepal [60] and India [46]. Samples from unknown addresses also had nearly twice the rejection rate (AOR = 1.59, 95% CI: 1.28–1.98) compared to OPD. This may be related to sample collection by untrained non‐laboratory personnel.

4.1. Strengths and Limitations of the Study

The strength of this study lies in its comprehensive assessment of the TTP performance of the Clinical Hematology Laboratory using a relatively detailed checklist. Additionally, sigma metrics were employed as the performance assessment tool, providing an objective measure of laboratory quality across all TTP phases. However, the study has some limitations. Hemolysis was assessed visually, which may have introduced interpersonal bias. The study was conducted over a relatively short period, and samples collected outside normal working hours were not included. Furthermore, not all potential root causes of low performance were evaluated. Finally, the performance of the automated hematology analyzer could not be assessed using sigma metrics due to the absence of regular quality control.

5. Conclusion and Recommendation

5.1. Conclusion

This study concluded that clinical hematology laboratory errors occurred at a high frequency, with the majority detected during the pre‐analytical phase, followed by the post‐analytical phase. Analytical errors were the least frequent. Among pre‐analytical errors, incomplete request forms and unsuitable samples were the most common. Lack of adherence to SOPs and patient addresses from IPD and unknown wards were significantly associated with poor sample quality. Similarly, the first work shift, IPD, and unknown addresses were significantly associated with prolonged turnaround time (TAT). Overall, the Hematology Laboratory demonstrated poor sigma metric values (< 3 sigma), indicating that the quality of services provided was not assured.

5.2. Recommendations

We recommend conducting further studies with an extended study period and involving a larger number of laboratory and healthcare professionals. For Dessie Comprehensive Specialized Hospital, we suggest providing targeted training for healthcare professionals on proper sample handling and implementing immediate corrective actions in the TTP, prioritizing improvements in the pre‐analytical phase. The hospital should consider sigma metrics as a tool for ongoing quality monitoring and improvement. Laboratory personnel are advised to strictly follow SOPs across all phases of the testing process. Overall, the laboratory should implement regular and comprehensive internal and external audits to identify deficiencies and take evidence‐based corrective measures.

Author Contributions

Zewudu Mulatie: conceptulization, project administration, writing – orginal draft and writing – review and editing the manuscript. Endris Ebrahim: formal analysis and validation. Mihreteab Alebachew: methodology, softwarre and investigation. Alemu Gedefie: formal analysis, supervision. Bruktawit Eshetu: formal analysis, validation and software. Mihret Tilahun: software, supervision, data curation. Habtu Debash: investigation, resources, software. Yeshimebet Kassa: visualization, investigation and resource. Ermiyas Alemayehu: formal analysis, validation and data curation. Tesfaye Gessese: Investigation, data curation and methodology. Dereje Mengesha Berta: formal analysis, project administration, validation and writing – review and editing the manuscript. All authors reviewed and approved the manuscript.

Funding

The authors received no specific funding for this work.

Ethics Statement

Ethical approval was obtained from the Ethical Review Committee of the College of Medicine and Health Sciences, according to the declaration of Helsinki. The corresponding author, Zewudu Mulatie, had full access to all of the data in this study and took complete responsibility for the integrity of the data and the accuracy of the data analysis.

Conflicts of Interest

The authors declare that there are no conflicts of interest regarding the publication of this manuscript.

Consent for Publication

The authors have nothing to report.

Transparency Statement

The lead author, Zewudu Mulatie, affirms that this manuscript is an honest, accurate, and transparent account of the study being reported; that no important aspects of the study have been omitted; and that any discrepancies from the study as planned (and, if relevant, registered) have been explained.

Supporting information

Supporting 2 STROBE‐checklist.

Supporting file 1: Raw data summary.

HSR2-9-e72063-s001.docx (24.6KB, docx)

Acknowledgments

First, we would like to express our appreciation for the College of Medicine and Health Sciences, Wollo University, for giving us the opportunity to conduct research on this topic. Secondly, we would also like to thank and appreciate the staff of Dessie Comprehensive Specialized Hospital for their commitment during data collection.

Data Availability Statement

The data supporting these findings are contained within the manuscript. If additional data are needed, it can be obtained from the corresponding authors upon request.

References

  • 1. Abdollahi A., Saffar H., and Saffar H., “Types and Frequency of Errors During Different Phases of Testing at a Clinical Medical Laboratory of a Teaching Hospital in Tehran, Iran,” North American Journal of Medical Sciences 6, no. 5 (2014): 224. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Arul P., Pushparaj M., Pandian K., et al., “Prevalence and Types of Preanalytical Error in Hematology Laboratory of a Tertiary Care Hospital in South India,” Journal of Laboratory Physicians 10, no. 02 (2018): 237–240. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Charuruks N., “Sigma Metrics Across the Total Testing Process,” Clinics in Laboratory Medicine 37, no. 1 (2017): 97–117. [DOI] [PubMed] [Google Scholar]
  • 4. Laposata M. and Dighe A., “Pre‐Pre” and “Post‐Post” Analytical Error: High‐Incidence Patient Safety Hazards Involving the Clinical Laboratory,” Clinical Chemical Laboratory Medicine 45, no. 6 (2007): 712–719. [DOI] [PubMed] [Google Scholar]
  • 5. Plebani M., “Quality Indicators to Detect Pre‐Analytical Errors in Laboratory Testing,” Clinical Biochemist Reviews 33, no. 3 (2012): 85–88. [PMC free article] [PubMed] [Google Scholar]
  • 6. Plebani M., Sciacovelli L., Aita A., and Chiozza M. L., “Harmonization of Pre‐Analytical Quality Indicators,” Biochemia Medica 24, no. 1 (2014): 105–113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Fuadi R., “Using Six Sigma to Evaluate Analytical Performance of Hematology Analyzer,” Indonesian journal of clinical pathology and medical laboratory 25, no. 2 (2019): 165–169. [Google Scholar]
  • 8. Najat D., “Prevalence of Pre‐Analytical Errors in Clinical Chemistry Diagnostic Labs in Sulaimani City of Iraqi Kurdistan,” PLoS One 12, no. 1 (2017): e0170211. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Aakre K. M., Langlois M. R., Watine J., et al., “Critical Review of Laboratory Investigations in Clinical Practice Guidelines: Proposals for the Description of Investigation,” Clinical Chemistry and Laboratory Medicine (CCLM) 51, no. 6 (2013): 1217–1226. [DOI] [PubMed] [Google Scholar]
  • 10. World Health Organization . Patient Safety: Making Health Care Safer. (World Health Organization, 2017). [Google Scholar]
  • 11. Teshome M., Worede A., and Asmelash D., “Total Clinical Chemistry Laboratory Errors and Evaluation of the Analytical Quality Control Using Sigma Metric for Routine Clinical Chemistry Tests,” Journal of Multidisciplinary Healthcare 14 (2021): 125–136. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Lippi G., Simundic A.‐M., and Mattiuzzi C., “Overview on Patient Safety in Healthcare and Laboratory Diagnostics,” Biochemia Medica 20, no. 2 (2010): 131–143. [Google Scholar]
  • 13. Wears R. and Sutcliffe K., Still Not Safe: Patient Safety and the Middle‐Managing of American Medicine (Oxford University Press, 2019). [Google Scholar]
  • 14. Stokowski L. A., “Who Believes That Medical Error is the Third Leading Cause of Hospital Deaths,” Medscape Medical News (2016): 1–16. [Google Scholar]
  • 15. Flott K., Fontana G., and Darzi A.. The Global State of Patient Safety. (Imperial College London, 2019). [Google Scholar]
  • 16. Howanitz P. J., “Errors in Laboratory Medicine: Practical Lessons to Improve Patient Safety,” Archives of Pathology & Laboratory Medicine 129, no. 10 (2005): 1252–1261. [DOI] [PubMed] [Google Scholar]
  • 17. World Health Organization . Laboratory Quality Management System: Handbook. (World Health Organization, 2011). [Google Scholar]
  • 18. Weingart S. N., “Epidemiology of Medical Error,” BMJ 320, no. 7237 (2000): 774–777. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Hammerling J. A., “A Review of Medical Errors in Laboratory Diagnostics and Where We Are Today,” Laboratory Medicine 43, no. 2 (2012): 41–44. [Google Scholar]
  • 20. Petti C. A., Polage C. R., Quinn T. C., Ronald A. R., and Sande M. A., “Laboratory Medicine in Africa: A Barrier to Effective Health Care,” Clinical Infectious Diseases 42, no. 3 (2006): 377–382. [DOI] [PubMed] [Google Scholar]
  • 21. Mesfin E. A., Taye B., Belay G., Ashenafi A., and Girma V., “Factors Affecting Quality of Laboratory Services in Public and Private Health Facilities in Addis Ababa, Ethiopia,” EJIFCC 28, no. 3 (2017): 205–223. [PMC free article] [PubMed] [Google Scholar]
  • 22. Asmelash D., Worede A., and Teshome M., “Extra‐Analytical Clinical Laboratory Errors in Africa: A Systematic Review and Meta‐Analysis,” EJIFCC 31, no. 3 (2020): 208–224. [PMC free article] [PubMed] [Google Scholar]
  • 23. Adane K., Girma M., and Deress T., “How Does ISO 15189 Laboratory Accreditation Support the Delivery of Healthcare in Ethiopia? A Systematic Review,” Ethiopian Journal of Health Sciences 29, no. 2 (2019): 259–264. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Tadesse H., Desta K., Kinde S., Hassen F., and Gize A., “Errors in the Hematology Laboratory at St. Paul's Hospital Millennium Medical College, Addis Ababa, Ethiopia,” BMC Research Notes 11, no. 1 (2018): 420. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Girma M., Desale A., Hassen F., Sisay A., and Tsegaye A., “Survey‐Defined and Interview‐Elicited Challenges That Faced Ethiopian Government Hospital Laboratories as They Applied ISO 15189 Accreditation Standards in Resource‐Constrained Settings in 2017,” American Journal of Clinical Pathology 150, no. 4 (2018): 303–309. [DOI] [PubMed] [Google Scholar]
  • 26. Gras J. M. and Philippe M., “Application of the Six Sigma Concept in clinical Laboratories: A Review,” Clinical Chemistry and Laboratory Medicine 45 (2007): 789–796. [DOI] [PubMed] [Google Scholar]
  • 27. Cian F., Villiers E., Archer J., Pitorri F., and Freeman K., “Use of Six Sigma Worksheets for Assessment of Internal and External Failure Costs Associated With Candidate Quality Control Rules for an ADVIA 120 Hematology Analyzer,” Veterinary Clinical Pathology 43, no. 2 (2014): 164–171. [DOI] [PubMed] [Google Scholar]
  • 28. Levey S. and Jennings E. R., “The Use of Control Charts in the Clinical Laboratory,” American Journal of Clinical Pathology 20, no. 11 (1950): 1059–1066. [DOI] [PubMed] [Google Scholar]
  • 29. Shaikh M. S. and Moiz B., “Analytical Performance Evaluation of a High‐Volume Hematology Laboratory Utilizing Sigma Metrics as Standard of Excellence,” International journal of laboratory hematology 38, no. 2 (2016): 193–197. [DOI] [PubMed] [Google Scholar]
  • 30. Westgard S. and Westgard Q., “Six Sigma Metric Analysis for Analytical Testing Processes,” Abott Lab MS 9 (2009): 4. [Google Scholar]
  • 31. Nevalainen D., Berte L., Kraft C., Leigh E., Picaso L., and Morgan T., “Evaluating Laboratory Performance on Quality Indicators With the Six Sigma Scale,” Archives of Pathology & Laboratory Medicine 124, no. 4 (2000): 516–519. [DOI] [PubMed] [Google Scholar]
  • 32. Afrifa J., Gyekye S., Owiredu W., et al., “Application of Sigma Metrics for the Assessment of Quality Control in Clinical Chemistry Laboratory in Ghana: A Pilot Study,” Nigerian Medical Journal 56, no. 1 (2015): 54. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33. Aggarwal K., Patra S., Acharya V., Agrawal M., and Mahapatra S. K., “Application of Six Sigma Metrics and Method Decision Charts in Improvising Clinical Chemistry Laboratory Performance Enhancement,” International Journal of Advances in Medicine 6, no. 5 (2019): 1524. [Google Scholar]
  • 34. Westgard J. O., “Useful Measures and Models for Analytical Quality Management in Medical Laboratories,” Clinical Chemistry and Laboratory Medicine (CCLM) 54, no. 2 (2016): 223–233. [DOI] [PubMed] [Google Scholar]
  • 35. Collage of American Pathologists . “Laboratory Accreditation Program, Hematology and Coagulation Checklist,” 22 (2022): 2–10, https://documents.cap.org/documents/cap-hem-sample-checklist.pdf. [Google Scholar]
  • 36. Westgard S., Bayat H., and Westgard J. O., “Analytical Sigma Metrics: A Review of Six Sigma Implementation Tools for Medical Laboratories,” Biochemia Medica 28, no. 2 (2018): 174–185. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37. Coskun A., Six Sigma: Projects and Personal Experiences. (Intech, 2011). [Google Scholar]
  • 38. Pathologists CoA, CAP accreditation program , “Laboratory General Checklist,” College of American Pathologists (2016);40(19).
  • 39. Assel M., Sjoberg D., Elders A., et al., “Guidelines for Reporting of Statistics for Clinical Research in Urology,” BJU International 123, no. 3 (2019): 401–410. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Lang T. A. and Altman D. G., “Basic Statistical Reporting for Articles Published in Biomedical Journals: The “Statistical Analyses and Methods in the Published Literature” or the SAMPL Guidelines,” International Journal of Nursing Studies 52, no. 1 (2015): 5–9. [DOI] [PubMed] [Google Scholar]
  • 41. Cuschieri S., “The STROBE Guidelines,” Saudi Journal of Anaesthesia 13, no. Suppl 1 (2019): S31–S34. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42. Berta D. M., Grima M., Melku M., et al., “Assessment of Hematology Laboratory Performance in the Total Testing Process Using Quality Indicators and Sigma Metrics in the Northwest of Ethiopia: A Cross‐Sectional Study,” Health Science Reports 7, no. 1 (2024): e1833. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43. Grecu D. S., Vlad D. C., and Dumitrascu V., “Quality Indicators in the Preanalytical Phase of Testing in a Stat Laboratory,” Laboratory Medicine 45, no. 1 (2014): 74–81. [DOI] [PubMed] [Google Scholar]
  • 44. Mukhopadhyay T., Shekhar S., Dagar V. K., and Mukhopadhyay A. K., “Characterization of Pre‐Analytical Errors Using Six Sigma Metrics and Process Capability Index in a Clinical Biochemistry Laboratory,” International Journal of Health Science Research 11, no. 2 (2021): 171–176. [Google Scholar]
  • 45. Swetha N. K., Kusuma K. S., Sahana K. R., et al., “Sigma Metric Analysis of Quality Indicators Across the Testing Process as an Effective Tool for the Evaluation of Laboratory Performance,” Medical Journal, Armed Forces India 79 (2023): S150–S155. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46. Kale S., Gumber R., Mahajan M., and Mulay S., “Identifying Errors Involving Clinical Laboratory: A 1 Year Study,” International Journal of Health Science Research 4, no. 8 (2014): 48–53. [Google Scholar]
  • 47. Sakyi A., Laing E., Ephraim R., Asibey O., and Sadique O., “Evaluation of Analytical Errors in a Clinical Chemistry Laboratory: A 3 Year Experience,” Annals of Medical and Health Sciences Research 5, no. 1 (2015): 8–12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48. Pothula Y., Al‐Marzooq Y. M., Salem R., Al‐Jasem W., and Al‐Hajji A., “A Retrospective Study of Quality Improvement in Clinical Biochemistry Laboratory,” Parameters 2016 (2017): 2018. [Google Scholar]
  • 49. Aadil S. and Vidyapeeth J., “Study of the Errors in Hematology Laboratory in a Tertiary Care Hospital,” Eur J Mol Clin Med 7 (2020): 1366–1381. [Google Scholar]
  • 50. Abdollahi A., Saffar H., and Saffar H., “Types and Frequency of Errors During Different Phases of Testing at a Clinical Medical Laboratory of a Teaching Hospital in Tehran, Iran,” North American Journal of Medical Sciences 6, no. 5 (2014): 224–228. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51. Sadiq F., Yasmeen F., Mumtaz A., et al., “Frequency of Errors in Clinical Laboratory Practice,” Iranian Journal of Pathology 9, no. 1 (2014): 45–49. [Google Scholar]
  • 52. Ambachew S., Adane K., Worede A., et al., “Errors in the Total Testing Process in the Clinical Chemistry Laboratory at the University of Gondar Hospital, Northwest Ethiopia,” Ethiopian Journal of Health Sciences 28, no. 2 (2018): 235–244. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53. Tola E. K., Dabi Y. T., and Dano G. T., “Assessment of Types and Frequency of Errors in Diagnostic Laboratories Among Selected Hospitals in East Wollega Zone, Oromia, Ethiopia,” Pathology and Laboratory Medicine International 14 (2022): 1–6. [Google Scholar]
  • 54. Chaudhuri S., Das A., Das S. K., and Tanmay Saha S., “Evaluation of Performance in the Pre‐Analytical Phase of a Clinical Biochemistry Laboratory in a Tertiary Medical College Hospital,” Asian Journal of Medical Sciences 13, no. 6 (2022): 62–67. [Google Scholar]
  • 55. Bir A., Ghosh A., Sinha S., and Banerjee A., “Quality Indicators Are Effective to Monitor the Performance Level of Preanalytical Phase‐A Study in a Clinical Laboratory of Eastern India,” Journal of Evidence Based Medicine and Healthcare (2018), https://www.researchgate-net/publication/324064048. [Google Scholar]
  • 56. Chen A., Anderson J., and Frater J. L., “Preanalytical Errors in a Satellite Stat Laboratory: A Six Sigma Analysis of Seven Years' Data,” Clinica Chimica Acta 523 (2021): 26–30. [DOI] [PubMed] [Google Scholar]
  • 57. Mebrat Gebreyes G., Abay Sisay S., Dilargachew Tegen T., Abushet Asnake A., and Mistire Wolde W., “Evaluation of Laboratory Performance, Associated Factors and Staff Awareness Towards Achieving Turnaround Time in Tertiary Hospitals, Ethiopia,” Ethiopian Journal of Health Sciences 30, no. 5 (2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58. Kipkulei J. C. and Lotodo T. C., “Evaluation of the Completeness in the Filling of Laboratory Request Forms Submitted to the Haematology Laboratory at a Tertiary Hospital in Kenya,” Health 11, no. 7 (2019): 862–868. [Google Scholar]
  • 59. Tesfaw H. M. and Hassen S. A., “F. Frequency of Specimen Rejection and A Ssociated F Actors at St. Paul's Hospital Millennium Medical College, Addis Ababa Ethiopia,” Journal of Multidisciplinary Healthcare 2, no. 1 (2015): 1–16. [Google Scholar]
  • 60. Pande K., Dahal P., and Pokharel L., “Identification of Types and Frequency of Pre‐Analytical Errors in Hematology Laboratory at a Tertiary Hospital of Nepal,” Journal of Pathology of Nepal 11, no. 1 (2021): 1842–1846. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supporting 2 STROBE‐checklist.

Supporting file 1: Raw data summary.

HSR2-9-e72063-s001.docx (24.6KB, docx)

Data Availability Statement

The data supporting these findings are contained within the manuscript. If additional data are needed, it can be obtained from the corresponding authors upon request.


Articles from Health Science Reports are provided here courtesy of Wiley

RESOURCES