Phantom for standardization in functional near-infrared spectroscopy, part 2: optical properties and Monte Carlo simulations

Caterina Amendola a, Antonio Pifferi a,b, Alessandro Torricelli a,b, Dirk Grosenick c,*, Heidrun Wabnitz c, David R Busch d, Michele Lacerenza e, Marianne Floor-Westerdijk f, Sanathana Konugolu Venkata Sekar g,h,i, Claudia Nunzia Guadagno g, Yukari Tanikawa j, Hiroshi Kawaguchi j
PMCID: PMC12978593  PMID: 41821524

Abstract.

Significance

The international standard IEC 80601-2-71 defines requirements for the basic safety and essential performance of medical continuous wave functional near-infrared spectroscopy (fNIRS) equipment, including a set of tests based on a dedicated tissue-equivalent phantom consisting of switchable apertures sandwiched between two diffusing media.

Aim

In a companion paper, we described a first implementation of this phantom developed by AIST, its basic characterization of attenuation properties, and results of its application in the testing of commercial fNIRS devices. In the present paper, we provide the absorption and reduced scattering spectra of the AIST phantom and of alternative silicone phantom matrices, complemented with extensive simulations on the effect of the phantom optical properties and geometry on the features of the phantom relevant for the fNIRS standard test.

Approach

All spectral measurements were performed with time-domain diffuse optical spectrometers for accurate quantification of the absorption and reduced scattering coefficient, whereas simulations were obtained using a Monte Carlo code.

Results

The POM-C matrix of the AIST phantom exhibits negligible absorption (<0.01  cm1) and a monotonically decreasing reduced scattering coefficient (from 10.5 down to 8.5  cm1) in the 650 to 850 nm range. Monte Carlo simulations explain the independence from wavelength, geometry, and partially also optical properties in the change in attenuation when switching between two apertures of 6 and 4 mm diameter within the phantom, with <0.1  dB spectral variation. Extended spectral characterization of POM-C and silicone matrices in the 600 to 1100 nm range showed additional marked absorption peaks around 900 and 1020 nm.

Conclusions

We provide valid support for researchers and manufacturers to understand the operation of the phantom described in the fNIRS standard and to replicate it in their labs. In addition, we provide a basis for further evolution of the standard itself to encompass alternative probe geometries or functional tests.

Keywords: functional near-infrared spectroscopy, phantom, international standard, absorption, scattering

1. Introduction

Functional near-infrared spectroscopy (fNIRS) is an optical method to noninvasively assess changes in oxy- and deoxyhemoglobin in the brain cortex associated with cerebral activation (due to, e.g., cognitive, visual, or motor stimuli).1,2 Alteration in the absorption spectrum of hemoglobin when it binds oxygen and relatively low absorption of biological tissues in the 650 to 900 nm range allows one to track cerebral activation by measuring attenuation of light between injection and collection points on the scalp at a few wavelengths. Several different approaches have been realized that use continuous wave (CW), time-domain (TD), or frequency domain (FD) light illumination and detection.3 The applications of fNIRS encompass basic neuroscience research, psychopedagogical studies, neuroeconomics, brain–computer interface, sport, and leisure.4 Clinically approved commercial devices have been deployed, initially in Japan, and are used in neurosurgical preoperative testing and to assist in the differential diagnosis of depressive symptoms.5,6

The international standard IEC 80601-2-71 defines requirements for the basic safety and essential performance of medical CW fNIRS equipment. It was developed by the IEC/ISO Joint Working Group (JWG) “Oximeters” (IEC/TC 62/SC 62D/JWG 5 and ISO/TC 121/SC 3/JWG 10). In 2025, following a systematic review, the second edition of this standard was published.7 Several performance tests for fNIRS equipment in this standard are based on a dedicated phantom that is equipped with a mechanism to change the light intensity detected by the fNIRS device under test. Although the major specifications of the phantom are defined in the standard, its specific technical implementation is not detailed there. Therefore, it was advisable to provide fNIRS manufacturers with information about a valid example of the practical implementation of this phantom. In 2022/2023, during the revision of this standard, an international measurement campaign was organized among members of the JWG “Oximeters.” It was led by the National Institute of Advanced Industrial Science and Technology (AIST) and was carried out as a round-robin campaign, circulating one of the original phantom prototypes developed at AIST. The participants included various European fNIRS manufacturers and research institutions that were interested in gaining experience with the phantom prototype and its application in fNIRS device testing. The development of the fNIRS standard has previously been reported.8

In a companion paper,9 the fNIRS standard phantom developed by AIST (hereafter briefly called AIST phantom) was presented together with the basic characterization of its attenuation properties and results of its application in the testing of commercial fNIRS devices. The goal of the present paper is to help researchers and manufacturers understand the requirements of the fNIRS standard and replicate the tests in their labs. More specifically, we provide: (i) the absorption (μa) and reduced scattering (μs) spectra of the AIST phantom in the range 600 to 850 nm; (ii) a Monte Carlo (MC) study of the effect of the optical properties and geometrical parameters of the phantom on its total attenuation and on the magnitude of the attenuation change, which are key requirements of the fNIRS standard; (iii) the extended spectral characterization in the 600 to 1100 nm range of suitable materials to construct a fNIRS standard phantom.

Implications of the findings for the implementation of the fNIRS tests will be presented. Furthermore, the paper will address some subtleties in the accurate characterization of phantom materials, particularly when dealing with low absorption and finite dimensions.

2. fNIRS Phantom

The international standard for fNIRS equipment, IEC 80601-2-71, describes tests to evaluate the performance of CW fNIRS devices. Measuring changes in detected light intensities is a key capability of CW fNIRS devices; therefore, the fNIRS standard prescribes a phantom test to assess this capability. The standard defines the basic specifications of the phantom, i.e., ranges for the absorption coefficient (μa), reduced scattering coefficient (μs), and size, for example. White polyoxymethylene copolymer (POM-C) is recommended as the main turbid material for the phantom due to the similarity of its μs to biological tissues. The μa of white POM-C is much lower than that of biological tissues and is almost constant in the wavelength range of the light sources used by fNIRS devices. In addition, POM-C has excellent mechanical and processing properties. The fNIRS standard does not prevent the use of epoxy resin or silicone resin, which are often applied as the base material of traditional fNIRS phantoms. Further characterizations of the various materials will be presented in Sec. 3.

The AIST phantom meets the specifications of the standard and consists of three main parts: an emission side attenuator, a detection side attenuator, and a circular aperture of changeable size sandwiched between the two attenuators. The two attenuators are cylindrical pieces of POM-C with a diameter of 60 mm and thicknesses of 20 and 10 mm on the injection and detection sides, respectively. The light is injected and detected at the centers of the circular surfaces. The light travels between the two attenuators through the aperture only. The active aperture is 4 or 6 mm in diameter, small compared with the diameter of the attenuators, and can be easily mechanically exchanged. The center of the active aperture is designed to coincide with the central axis of the cylindrical POM-C attenuators. By switching one aperture to the other, the detected light intensity is expected to change approximately proportionally to the areas of the apertures, i.e., by a factor of 2.25. See the part 1 paper for more details.9 Replicas of the prototype became commercially available from a phantom manufacturer (BioPixS Ltd., Cork, Ireland), which is one of the major outcomes of this work.

3. Materials and Methods

The AIST phantom was characterized at AIST using a CW spectrometer to provide high spectral resolution and a TD diffuse spectrometer at Physikalisch-Technische Bundesanstalt (PTB, Germany) to separate absorption from scattering contribution. An extended spectral characterization of POM-C and silicone for phantom construction was performed with TD diffuse spectrometers at Politecnico di Milano (POLIMI, Italy) and Tyndall National Institute (TYNDALL, Ireland), respectively. POLIMI conducted MC simulations on different phantom implementations. Each institution adopted the best practices for measurement procedures and data analysis established in their lab, as detailed below.

3.1. CW Spectroscopy

A POM-C slice from the same bulk as the fNIRS standard phantom (Duracon NE25X, Polyplastics, Tokyo, Japan) was cut into a rectangular shape of 40×40×5  mm. The surface roughness (Ra) of the sample was 0.8. The sample was measured with a spectrophotometer equipped with an integrating sphere (UV-2600i and ISR-2600Plus for UV-2600, Shimadzu, Kyoto, Japan). The widest surface of the sample was held in close contact with the aperture of the integrating sphere to measure diffuse transmittance and reflectance. The measurement wavelength range was 650 to 900 nm, and automatic sweeps were repeated seven times at 1 nm intervals to calculate the mean and standard deviation for each wavelength. To check the effect of bias due to different sample attachment positions, the sample was attached and removed five times and measured in the same manner, and the mean was taken.

3.2. TD Diffuse Spectroscopy

The PTB time-domain system utilizes a SuperK Fianium FIU-15 PP supercontinuum laser combined with a SuperK Varia tunable filter (NKT Photonics, Germany) for wavelength selection. The pulse repetition rate was set to 39 MHz. To estimate the optical properties of the turbid blocks of the fNIRS standard phantom, the 2 cm thick POM-C cylinder (diameter 60 mm) was taken out of the housing of the phantom and placed between two plates. These plates were covered with black Neoprene and contained holes to hold the ferrules of optical fibers. The measurements were performed in transmission geometry. A graded index fiber of 62.5  μm core diameter was used to guide the laser pulses to the center of the front face of the cylinder. On the rear face, the transmitted light was collected on axis by a graded-index fiber with a core diameter of 600  μm. The distributions of times of flight of photons (DTOF) were recorded by time-correlated single photon counting (TCSPC) with a cooled fast hybrid detector (HPM-100-07C, Becker & Hickl, Germany) and an SPC-150NX board (Becker & Hickl, Germany). A variable neutral density filter in front of the detector was employed to adjust the photon count rate.

The measurements on the cylinder were done at wavelengths from 600 to 845 nm in steps of 10 nm, with a spectral bandwidth of 10  nm. The average light power on the surface of the phantom was a few milliwatt at each wavelength. Using an automatic wavelength sweep with 25 s steps, a collection time interval of 20 s was available for analysis at each wavelength. The maximum photon count rate over all wavelengths was adjusted not to exceed 500 kHz. The full width at half maximum (FWHM) of the instrument response function (IRF) was 50  ps at 600 nm and decreased with increasing wavelength to 35  ps in the 800 to 845 nm range. For each wavelength, the measured DTOF was analyzed by a white MC model (e.g., simulations were performed with μa set to zero, and absorption effects were added a posteriori according to Beer’s law) of light transport in a homogeneous cylinder. For this purpose, a database of DTOFs with a step size of Δμs=0.1  cm1 was created, with μa set to zero. DTOFs for intermediate values of μs were determined by linear interpolation. When fitting the optical properties, absorption was taken into account by multiplying the DTOFs of the database by the Lambert Beer factor exp(μacMt) where cM and t denote the speed of light in the phantom and the time of flight, respectively. In addition to μa and μs, a small temporal offset between the theoretical DTOF (convolved with the IRF) and the measured DTOF was used as a third fitting parameter. Including the offset in the fitting improves the matching of the DTOF’s leading edge by the model and also corrects for a possible temporal drift between the phantom and IRF measurements. After analyzing the DTOFs at each wavelength independently, the μs were smoothed by applying the scattering power law

μs(λ)=μs(λ0)(λ/λ0)b, (1)

with λ0 set to 800 nm. Then, μa and the temporal offsets were fitted a second time to obtain the final result for absorption.

The POLIMI system10 operates a supercontinuum white light laser source (SuperK Extreme, NKT Photonics, Denmark) emitting pulses with a width on the order of picoseconds at 40 MHz repetition rate as an illuminating source. Spectral scanning of the source wavelength was achieved by first dispersing the white laser light using a Pellin Broca prism (Bernhard Halle Nachfl, Germany) placed on a rotating stage for wavelength tuning, and then selecting a narrow band by a small core (50  μm) graded index fiber, resulting in a spectral bandwidth of 3 to 9 nm over the 600 to 1100 nm wavelength range. After attenuation by a variable neutral density filter (Edmund Optics, Barrington, New Jersey, United States), light was injected into and collected from the phantom by a pair of 1 mm core, 1.5 m long glass step-index fibers in reflection geometry. Light gathered from the phantom at a source-detector distance of 2 cm was delivered to a home-made detection module based on a silicon photomultiplier (SiPM) (S10362-11-050C, Hamamatsu, Japan) with custom driving electronics.11 The DTOF of re-emitted photons was recorded using a TCSPC board (SPC-130, Becker & Hickl, Berlin, Germany). All processes involved in the spectral measurement (wavelength scanning, laser attenuation, data visualization, and saving) were automated by a custom ANSI C software implemented in the LabWindows CVI framework. A full spectrum, from 600 to 1100 nm at steps of 5 nm, required around 5 min. The dark background of the instrument was 2·104 counts per second, and the FWHM of the IRF was around 90 ps. The maximum power impinging onto the sample was 4 mW, whereas the count rate was kept <106 counts per second. The absorption and reduced scattering spectra were derived by fitting the DTOFs to the convolution of the IRF with a solution of the photon diffusion equation for a cylindrical geometry. Only μa and μs were used as free-fitting parameters, whereas the time origin was defined from the IRF.

The TYNDALL system is based on a supercontinuum broadband laser source (Fianium SC450, Southampton, United Kingdom) with a 20 MHz repetition rate and pulse width in the tens of picoseconds range. White light was dispersed using a rotatable Pellin Broca prism (UV Fused Silica, Thorlabs, Newton, New Jersey, United States, ADBU-20). The wavelength selection was achieved by coupling light into a graded index fiber (0.1 NA, 25  μm, M67L02, Thorlabs) using a convex lens. The fiber tip, which was in contact with the sample, acted as a source. Due to the nonlinear dispersion of the prism, the FWHM of the selected light wavelength band varied with wavelength, ranging from 5 nm at 650 nm to 12 nm at 1100 nm. A multimode fiber (0.39 NA, 300  μm, M67L02, Thorlabs) was employed for light collection in transmission geometry. The detection system included a single-photon avalanche diode (SPAD, Micro Photon Devices, Bolzano, Italy, PDM freespace) and custom-made front-end optics to focus the light onto the 100  μm active area. The detector’s signal was processed by a laser-synchronized TCSPC board (Picoharp 300, PicoQuant GmbH, Berlin-Adlershof, Germany), which acquired and stored the DTOF data for further analysis. To control irradiance at the sample and prevent detector saturation, a pair of continuously variable neutral density filters (NDC-100C-4, Ø100 mm, OD: 0.04 to 4.0, Thorlabs) was used, ensuring photon count rates remained within acceptable limits to avoid pulse pile-up. The aim was to maintain a count rate of 3·105 counts per second, which is <2% of the laser repetition rate.

3.3. Monte Carlo Simulations of Phantom Properties

Numerical simulations were performed by the MC method using the MMCLAB OpenCL mesh-based MC tool.12 The domain (see Fig. 1) was designed to reproduce the real geometry of the fNIRS phantom (see also Fig. 1 in the part 1 paper9). The geometry consists of two cylindrical attenuators (Atn1 and Atn2, with thickness t1=2  cm and t2=1  cm, respectively, and diameter dc=6  cm), a slide bar (with ts=0.05  cm thickness) containing an aperture with diameter da=0.6  cm, and a slide bar holder containing an aperture with diameter dh=0.8  cm. The distance between Atn1 and Atn2 is tsh=0.12  cm. The empty space between the two cylindrical attenuators is filled with air. The mesh’s smallest edge was set to 0.05 cm to enhance geometrical accuracy while maintaining reasonable computational and memory requirements.

Fig. 1.

Fig. 1

Geometry of the domain for MC simulations. Atn1 and Atn2: cylindrical attenuators; t1 and t2: thickness of Atn1 and Atn2; dc: diameter of Atn1 and Atn2; ts: thickness of slide bar; da: diameter of aperture in the slide bar; dh: diameter of aperture in the slide bar holder; tsh: distance between Atn1 and Atn2; dd: diameter of the detector.

Simulations were performed in transmittance geometry, mimicking the source as a pencil beam, and the detector as a disk with diameter dd=0.8  cm. Each simulation was performed by injecting 109 photons. The total number of photons detected (Nd) through the 8 mm diameter aperture was used to calculate the optical loss (OL=log10(Nd/Ni), in dB). Five wavelengths (650, 700, 750, 800, and 850 nm) were simulated. The input optical properties of the cylindrical attenuators Atn1 and Atn2 are reported in Table 1 (derived from the broadband time domain diffuse reflectance measurements, see Sec. 3.2). The optical properties of the slide bar and slide bar holder were assumed as μa=10,000  cm1, μs=100  cm1, with refractive index 1.0. The optical properties of air were assumed as μa=0  cm1, μs=0  cm1, with a refractive index of 1.0. These parameters were kept fixed for all wavelengths. In the simulations, the μa was initially set to 0  cm1, and added a posteriori, decreasing the photon weight according to the path-length traveled by the photon in each compartment.13

Table 1.

Optical properties of the cylindrical attenuators Atn1 and Atn2 in the MC simulations. Data for μa and μs were taken from Fig. 2(b).

λ (nm) μa (cm1) μs (cm1) n
650 0.002 10.6 1.5
700 0.002 9.9 1.5
750 0.007 9.4 1.5
800 0.006 8.8 1.5
850 0.005 8.4 1.5

Starting from the initial configuration, different simulations were repeated by varying: (a) the diameters of the cylinders: dc=4  cm, and 8 cm; (b) the thickness of Atn1 (or the position of the internal aperture at the center of the slide bar): t1=1.0  cm, and 1.5 cm (keeping fixed the total thickness of Atn1 and Atn2 t1+t2=3  cm and the distance between Atn1 and Atn2 tsh=0.12  cm); (c) the diameter of the internal aperture at the center of the slide bar: da=0.4  cm, and 0.8 cm diameter; (d) the size of the incident beam using a uniform disk with 0.5 cm diameter; (e) the bulk absorption and reduced scattering coefficients. Simulations were performed on a Windows 10 Pro workstation equipped with 2 processors of 12 cores at 2.3 GHz, and a GPU NVIDIA GeForce RTX 3080 Ti. The time needed for a simulation was 80  s.

3.4. Phantom Matrices for Extended Spectral Characterization

BioPixS manufactured a phantom suitable for use in the fNIRS standard based on the same concept as the AIST phantom. BioPixS started from a large block of POM-C material,14 50 mm in thickness. Sheet-extruded POM-C was used because it provides uniform optical properties across the lateral dimensions of the material. This block was then machined down to create uniform samples of thickness 20 and 10 mm, as required by the fNIRS standard. The 20 mm thick block was used for characterizing the optical properties. The samples were then assembled into a 3D-printed casing to create the BioPixS fNIRS standard phantom. The sliding aperture is created using 3D printing of PLA material (Prusa, PLA Prusament Jet black, Prague, Czech Republic) with high carbon black concentration. The input and output sides of the fNIRS standard phantom are designed to permit the attachment of a wide array of optodes.

Silicone was considered an alternative material for the phantom construction (POLIMI). TiO2 was added to Silicone Elastomer Sylgard 184 (Base 10 and Curing Agent 1, Dow Corning), for a final concentration of 0.1% (in weight). The mixture was stirred for 3 min and homogenized for 2 min before adding the curing agent (10%). Again, following 2-min stirring and 1-min homogenizing, air bubbles were removed from the solution by a vacuum pump system. Curing was obtained by keeping the samples in the oven (UT 20P, Heraeus, Hanau, Germany) at 65°C overnight. The phantom was cast in a similar shape as the AIST phantom attenuators, namely, a 6 cm diameter, 2.2 cm height cylinder. More details and procedures are reported in detail elsewhere.10

4. Spectral Characterization of the AIST Phantom

Figure 2(a) shows the transmission spectrum of a thin (2 mm) slice of POM-C material taken from the same batch as used for the AIST Phantom and measured using a standard spectrometer equipped with an integrating sphere as described in Sec. 3.1. Diffuse transmittance of light is largely uniform over the whole 600 to 850 nm spectral range, with only very feeble valleys around 740 and 815 nm. Above 850 nm, the attenuation increases, hinting at a more prominent absorption peak (localized at 900 nm, data not shown). The spectrum shows a slight increase in transmittance with wavelength (10% over 200 nm).

Fig. 2.

Fig. 2

(a) Diffuse transmittance of a 5 mm thin slice of the POM-C matrix from the same batch of the AIST phantom was measured using a spectrometer equipped with an integrating sphere in transmission geometry. (b) Spectra of μa (red, left) and μs (blue, right) of the 2 cm thick AIST phantom matrix measured in transmittance geometry. The lines indicate the final (smoothed) results when using Eq. (1) for μs(λ). The solid squares show the underlying raw data. Error bars are plotted for selected wavelengths.

Figure 2(b) shows the absorption and the reduced scattering coefficients of the 2 cm thick POM-C cylinder of the AIST phantom between 600 and 850 nm measured at PTB. Below 700 nm, the absorption is 0.002  cm1. At 740 and 815 nm, the absorption spectrum shows two peaks, whereby the maxima do not exceed 0.01  cm1. These peaks are responsible for the small dips in the transmission spectrum in Fig. 2(a). Overall, the absorption is significantly lower than that of most types of biological tissue. The μs is within the general range of values for tissue. The decrease with increasing wavelength is comparatively flat. The decay coefficient b in Eq. (1) amounts to 0.86. The application of Eq. (1) leads to only minor corrections of μs and μa compared with the raw data. The uncertainty of the optical properties is dominated by the influence of the refractive index, as addressed in Sec. 7.

5. Monte Carlo Simulations

The effect of the phantom geometry and optical properties on OL as well as on ΔOL—that is, the change in OL for switching the internal aperture in the slide bar—was studied using MC simulations. Figure 3(a) shows OL as a function of wavelength and for a pencil beam injection with different sizes of the internal aperture da. The estimated OL for the configuration with da=0.6  cm diameter of the internal aperture (green triangles) is between 40.6 and 41.3 dB. A larger OL from 44.4 to 45.1 dB is obtained if the diameter of the internal aperture is reduced to da=0.4  cm, as expected. Conversely, a larger internal aperture with da=0.8  cm results in lower OL between 38.3 and 38.9 dB (data not shown). Switching the aperture diameter from da=0.6  cm to da=0.4  cm increases the OL by 3.8  dB as shown in Fig. 3(b). Increasing the diameter of the aperture from da=0.6  cm to da=0.8  cm decreases the OL by 2.4  dB (data not shown).

Fig. 3.

Fig. 3

(a) Optical loss of the simulated phantom as a function of wavelength for different internal aperture diameters da=0.4, 0.6 cm. (b) Change in optical loss when changing the internal aperture diameter from da=0.6  cm to da=0.4  cm.

Changing the diameter of the cylinder has a negligible effect on the OL. Reducing the diameter of the cylinders from dc=6  cm to dc=4  cm yields an increase in OL of 0.1  dB, whereas increasing the diameter from dc=6  cm to dc=8  cm leads to a maximal decrease of OL by 0.03 dB [see Fig. 4(a)]. The latter (former) results are consistent with an increased (decreased) number of photons lost at the lateral boundary of the cylinder.

Fig. 4.

Fig. 4

Optical loss of the simulated phantom as a function of wavelength for (a) different cylinder diameters dc=4, 6, 8 cm and (b) different positions of the internal aperture t1=1.0, 1.5, 2.0 cm while keeping t1+t2=3  cm (b).

To study the effect of the distance of the internal aperture from the source, we have changed t1 and t2, the thicknesses of cylindrical attenuators Atn1 and Atn2, while keeping fixed the overall thickness t1+t2=3  cm and da=0.6  cm. The change in OL ranges from 0.35 to 0.38  dB when moving from the configuration with t1=1  cm and t2=2  cm to that with t1=2  cm and t2=1  cm, whereas it ranges from 1.12 to 1.35 dB for the configuration with t1=t2=1.5  cm [see Fig. 4(b)].

Figure 5(a) systematically shows that the OL—as expected—is influenced by the bulk μa of the cylindrical attenuators Atn1 and Atn2. In the range for the bulk μa 0.0005 to 0.1  cm1 (a 200× change), the OL increases from 40  dB up to 50 dB. However, the change in OL when passing from the internal aperture diameter da=0.6  cm to da=0.4  cm varies less than 0.1  dB when the bulk μa spans the range 0.0005 to 0.1  cm1, as shown in Fig. 5(b). The effect of μs on the OL is minimal when the absorption is low (<1  dB differences between minimum and maximum μs values at μa=0.0005  cm1) and becomes more evident at larger absorption values (<2.1  dB differences between minimum and maximum μs values at μa=0.1  cm1).

Fig. 5.

Fig. 5

(a) Optical loss of the simulated phantom for different values of the bulk μa and bulk μs of the cylindrical attenuators Atn1 and Atn2. (b) Change in optical loss of the simulated phantom when the internal aperture is reduced from da=0.6  cm to da=0.4  cm as a function of wavelength for different values of the bulk μa and bulk μs of the cylindrical attenuators Atn1 and Atn2.

Figure 6(b) shows the OL for different spectra of μs each characterized by a different b parameter, as described in Fig. 6(a). As expected, a larger b corresponds to steeper spectral dependence, resulting in a reduced OL for longer wavelengths.

Fig. 6.

Fig. 6

(a) Different spectra of the bulk μs of the cylindrical attenuators Atn1 and Atn2. Equation (1) was used keeping fixed μs(λ0)=10.6  cm1 and λ0=650  nm while changing b. (b) The corresponding optical loss of the simulated phantom for the different spectra of the bulk μs of the cylindrical attenuators Atn1 and Atn2 shown in panel (a).

Finally, the results are not affected if the pencil beam is replaced by a 0.5 cm-diameter uniform beam (data not shown).

6. Extended Spectral Characterization of Various Materials for Phantom Construction

The material (POM-C) used for the construction of the AIST phantom has the great advantage of providing intrinsically a μs similar to what is observed in biological tissues in the 650 to 900 nm range. For that reason, it was used at the early stages of diffuse optics as a viable phantom material.15 Yet, the very fact that POM is naturally scattering makes the exact definition of optical properties less reliable as they depend on the blend and fabrication, different from lab-made phantoms, where absorbers and scatterers can be mixed in due quantities to yield planned properties. In addition, different extrusion and processing of POM-C could provide different optical properties and potential heterogeneity within the material. For example, cylindrical extrusion of POM-C seems to lead to a highly scattering region along the center line. Therefore, we investigated different extrusions of POM-C, and also other materials, frequently used in diffuse optics, with fabrication that is well documented and reproducible in the lab.

In addition, we extended the spectral range to 600 to 1100 nm, because some fNIRS instruments are adopting higher wavelengths and because the potential use of 1064 nm to assess oxygenated hemoglobin changes has been proposed.16

A first alternative option to POM-C is the use of a silicone (Sylgard) matrix, as described above, with TiO2 particles added to control scattering and with no additional absorbers. Figure 7 shows μa and μs of a silicone matrix with the same dimensions as those of the 2 cm thick block of the AIST phantom. The optical properties of the silicone matrix were measured in the spectral range from 600 nm up to 1100 nm. The absorption is negligible up to 870 nm, apart from a minor peak at 745 nm. Above 870 nm, there are two additional peaks at 910 and 1025 nm. The scattering spectrum is flat and decreasing upon increasing wavelength. As the Sylgard matrix has negligible scattering contributions, the required μs values can be finely tuned using TiO2 concentration. Compared with the AIST phantom, this matrix shows a steeper scattering spectrum [i.e., a larger b coefficient in Eq. (1): b=1.48].

Fig. 7.

Fig. 7

Spectra of μa (left axis) and μs (right axis) of a Sylgard cylinder [similar dimensions to the AIST phantom in Fig. 2(b): 6 cm of diameter, and 2.2 cm thickness] measured in transmittance geometry.

Among POM-C materials, the best option is sheet-extruded POM-C, which can provide uniform optical properties across the lateral dimensions of the material. For this purpose, BioPixS adopted sheet-extruded POM-C. For the time domain characterization, a larger block of POM-C material of 50 mm in thickness was machined down to create a sample of 20 mm thickness. Measurements were performed in transmittance geometry. Figure 8 shows the spectral characteristics of the machined POM-C material. The POM-C material exhibits similar scattering properties and negligible absorption, as seen in the AIST POM-C material below 860 nm, whereas above that limit, μa increases rapidly with two major peaks at 900 and 1020 nm, respectively.

Fig. 8.

Fig. 8

Spectra of μa (left axis) and μs (right axis) of a BioPixS POM block [thickness 2 cm, i.e., same as the AIST POM block characterized in Fig. 2(b)], measured in transmittance geometry.

7. Discussion

In general, there is an increasing need to adopt standard procedures and reference phantoms in fNIRS to permit verification of instruments, inter-laboratory comparison, and assessment of clinical studies. In addition, the push for Open Science and deployment of Open Data demands that test procedures and reference data be made available together with experimental data. There are major activities in the development of reference phantoms,17 including those for basic testing of devices18 and those designed to mimic complex biological structures,1921 performance assessment procedures,2224 standardization of analysis,2527 and inter-laboratory comparison.28 This paper will not go into detail on these different approaches and solutions: we will rather focus on the specific requirements of the phantom for the fNIRS standard.

The spectral characterization of the AIST phantom revealed a very low absorption—apart from two minor peaks at 740 and 815 nm, which do not exceed 0.01  cm1—and an almost flat scattering spectrum around 10  cm1 with a Mie-related decay coefficient b in Eq. (1) of 0.86. These two properties make the total optical loss OL and even more its change ΔOL to be largely wavelength independent. This specific feature is not strictly required by the standard because manufacturers are instructed to characterize ΔOL using a power meter for each wavelength of the particular fNIRS device, to use this information when calculating the expected changes in deoxygenated and oxygenated hemoglobin. Nevertheless, wavelength independence is a useful condition for a fair application to different devices relying on distinct choices of wavelengths so as to test them under similar attenuation conditions.

Measuring the very small absorption coefficients of the 2 cm thick POM phantom is difficult because they have only little impact on the shape of the DTOF. The results are strongly dependent on boundary effects under these conditions. The uncertainty of the optical properties is dominated by the influence of the refractive index. The data in Fig. 2(b) were obtained by assuming the refractive index of POM as n=1.5 (not specified by the manufacturer), which is a typical value for plastics in the near-infrared spectral range. A change of n of ±0.01 in the MC model shifts the fitting results for μa by about ±0.0005  cm1, which corresponds to a relative change in μa of 20% in the 600 to 700 nm range. The associated change in the results for μs is ±0.1  cm1, which corresponds to a relative change in μs of only 1%. The error bars in Fig. 2(b) reflect an assumed uncertainty in the refractive index of ±0.04. Published values of the refractive index for POM are smaller (n=1.435 to 1.414 from 650 to 800 nm).29 However, these values were obtained from films of POM-C pressed from granulate. These films could contain inclusions of air being responsible for a reduction of the observed refractive index. If these smaller values were adopted in the fitting procedure based on the MC model, μa would become negative and therefore unrealistic. This observation supports the assumption that the refractive index must be higher for the extruded POM used in the AIST phantom than the values reported for the pressed granulate. To reduce the uncertainty of μa, the refractive index of POM-C should be measured. A suitable method appears to be the determination of the Brewster angle based on reflection measurements on a flat, polished surface, as was done in Ref. 29 for the pressed granulate. The fNIRS standard does not contain any specifications regarding the refractive index of the phantom material. It defines a range for μs and μa to ensure optical properties similar to those of brain tissue. However, a model-based determination of the scattering and absorption properties requires knowledge of the refractive index.

As discussed in the companion paper part 1,9 there is a wide variety of white POM bulk products available. For the fabrication of a standard phantom, the reproducibility in the optical properties, the optical homogeneity, and the surface roughness are relevant features to be considered. For instance, “center line porosity” is a known issue in POM-H (polyoxymethylene homopolymer) such as Delrin® and should be avoided. Conversely, POM-C materials as Duracon were observed to provide a good level of optical homogeneity and were selected for the AIST phantom, as well as for the BioPixS phantom.

The MC simulations permitted to identify some remarkable properties of the designed phantom. The spectral variations in OL are limited due to the rather flat scattering and very low absorption, with an increase of 0.7 dB from 650 to 800 nm due to a small absorption peak in POM. The corresponding variation in ΔOL is negligible (<0.08  dB). Furthermore, though the phantom is rather small (just 6 cm in diameter), no significant finite-size effect is observed. Indeed, OL remains unchanged for a larger diameter (8 cm) and is unaltered for an even smaller size (4 cm). It is still true—as commented above—that within the bulk materials, key losses are caused by boundaries for μa0  cm1, but with the insertion of the black plate with the hole in between the two attenuators, the main light losses are governed by the aperture geometry rather than by the lateral boundaries of the phantom.

In agreement with this design, ΔOL shows only a weak dependence on μa within the explored range (0 to 0.5  cm1), remaining confined within 3.5 to 3.8 dB, well inside the 3 to 4 dB limits prescribed by the standard. This behavior is consistent with the fact that ΔOL is primarily determined by the ratio of the aperture areas in the black plate.

In addition, the position of the black plate along the axis of the phantom is not so critical, with just <1  dB difference when the plate is placed in the middle of the phantom or with a displacement of 0.5 cm from the center (i.e., the actual realization, with 2 and 1 cm thickness, respectively).

MC simulations further confirm that ΔOL is very close to the ratio of the areas of the apertures, at least up to the largest tested diameter of 0.8 cm. The simulated OL at 700 nm was 42 dB, slightly different from the measured value of 40 dB. At this level, inaccuracies in the estimate of the optical properties, in the modeling of the complex geometry of the plate with the slider, or the nonideal surface roughness could create the discrepancy. The 40 dB upper limit for attenuation as prescribed in the standard could be too low to replicate the effective attenuation observed on the human head when dealing with large source-detector distances. The addition of a 20 dB neutral density filter can raise the OL to 60 dB, as discussed in part 1.9 It is also possible to increase OL by adding absorption in a fabricated phantom matrix. As shown from MC simulations, increasing μa up to 0.5  cm1 raises OL from 40 to 70 dB. ΔOL is rather independent from μa, being confined in the 3.5 to 3.8 dB range over the explored range from 0 to 0.5  cm1 of μa, well within the 3 to 4 dB limits prescribed by the standard.

The phantom can be constructed with alternative materials, such as silicone. Example spectra of a Sylgard matrix with no added absorption are reported in Sec. 6. As compared to the AIST phantom, these materials exhibit a more steeply descending scattering spectrum (b=1.48 against b=0.83 of POM). As shown in Sec. 5, the greater slope in scattering increases the wavelength dependence of OL, whereas ΔOL is substantially unaffected. As discussed above, wavelength independence is not a requirement of the standard, but rather a helpful attribute that facilitates fair comparisons between instruments operating at different wavelengths. The adoption of such materials is advantageous because it permits fine adjustment of optical properties, with the addition of black toner for absorption and TiO2 particles for scattering. This can be useful—for instance—to yield a 60 dB OL. In addition, good homogeneity and reproducibility can be assured. However, reliable phantom fabrication is a critical task that requires mastery; therefore, Duracon POM-C remains a good practical solution.

The AIST phantom, implementing the fNIRS standard, can be operated only in transmittance geometry. As some devices might not permit detachment of source from detector (e.g., in many biomedical optical applications, sources and detectors are integrated within a single probe to ensure a fixed source–detector separation and stable optical coupling), a phantom operated in reflectance geometry is advisable. However, keeping ΔOL constant independently of source-detector distances is not trivial. There is certainly room for further research on this issue and potential expansion of the standard. In addition, the present standard neither addresses depth sensitivity nor time- or frequency-domain instruments. In a future revision of the standard, such expansion could be desirable, implying updates to test phantom designs.

8. Conclusion

The international standard for fNIRS devices, IEC 80601-2-71, foresees specific tests based on a 3 dB to 4 dB change in a dynamic phantom with a baseline 40 dB optical loss in the 650 to 850 nm range. A companion paper9 describes the AIST phantom, which implements the requirements of the standard, and presents the characterization of its attenuation properties as well as its practical application in tests of commercial devices.

In the present paper, we provide the spectral characterization in the 600 to 850 nm range of the turbid materials of the first prototypes of such a phantom produced in Japan by AIST. The absorption coefficient of this POM-C material is extremely low, with minor peaks at 740 and 815 nm not exceeding 0.01  cm1. The reduced scattering coefficient is rather flat and monotonically decreasing from 11.5 down to 8.5  cm1. An extended spectral characterization up to 1100 nm of a different slab of POM-C material showed marked absorption peaks at 900 and 1020 nm, which should be considered when operating above 860 nm.

MC simulations confirmed several notable features of the AIST phantom, attributable to its specific optical properties and geometry. In particular, both OL and ΔOL were found to be largely independent of: (i) wavelength within the 600 to 850 nm range; (ii) the lateral dimensions of the phantom cylinders, even for diameters as small as 4 cm; and (iii) the axial position of the black plate, i.e., the relative thickness of the two attenuator cylinders. The MC simulations could be easily extended to reproduce a multi-aperture design and to evaluate the influence of increased spacing between two attenuators due to thick apertures, which are features of the phantom concept.18

The AIST phantom was constructed with a POM-C (Duracon) matrix. Alternative materials for phantom production with well-documented provision and fabrication information are also presented. This option permits to finely tune the optical properties with the addition of TiO2 for scattering and black toner for absorption. A key benefit could be the option to raise OL to 60 dB, closer to the typical attenuation in real applications of the device. MC simulations demonstrated that μa can be increased up to 0.1cm1 with minor effects on ΔOL still confined in the 3.5 to 3.8 dB range, well within the 3 to 4 dB limits of the standard.

Overall, these data can be useful to identify a suitable phantom for the implementation of the fNIRS standard. In addition, they provide additional information to complement the standard. Furthermore, this work can facilitate the manufacturing of fNIRS standard phantoms by commercial providers of turbid phantoms.

Acknowledgments

The authors are grateful to the members of fNIRS manufacturers Hamamatsu Photonics K. K. (Hamamatsu, Japan), Hitachi, Ltd. (Tokyo, Japan), and Shimadzu Corporation (Kyoto, Japan) for cooperation in developing the AIST phantom. This work was partly supported by internal competitive funding for standardization from the Department of Information Technology and Human Factors [Grant No. AIST R2-S-04 (HK)]. The authors acknowledge financial support by the European Union’s NextGenerationEU Programme with the I-PHOQS Infrastructure (Grant Nos. IR0000016, ID D2B8D520, and CUP B53C22001750006) “Integrated infrastructure initiative in Photonic and Quantum Sciences.”

Biographies

Caterina Amendola is a researcher specializing in the field of Diffuse Optics. She works in the Physics Department of Politecnico di Milano (Italy), where she earned her PhD in March 2022. Her research primarily focuses on time-domain near-infrared spectroscopy (TD-NIRS) and diffuse correlation spectroscopy (DCS) techniques. With expertise in developing hybrid TD-NIRS and DCS devices, she applies these advancements in clinical settings while conducting data analysis and physical modelling.

Antonio Pifferi is a professor of Physics at Politecnico di Milano, Department of Physics, and Director of the “Center for Ultrafast Science and Biomedical Optics” (CUSBO). His research activity is focused on the study of photon propagation through diffusive media (diffuse optics) using time-resolved techniques. Diffuse optics has been explored vertically—from basic research up to exploitation of new application directions—and horizontally—covering diverse applications, including optical mammography, functional brain imaging, food analysis, and wood optics.

Alessandro Torricelli is a full-time professor in the Department of Physics, Politecnico di Milano, Italy. He received his MS degree in electronic engineering from Politecnico di Milano in 1994 and his PhD in physics from Politecnico di Torino, Italy, in 1999. He is the author of more than 200 papers in international peer-reviewed journals. His current research interests include photon migration in diffusive media, functional near-infrared spectroscopy, and noninvasive diffuse spectroscopy with time-domain systems.

Dirk Grosenick heads the working group Optical Medical Imaging and Cytometry at the German National Metrology Institute PTB. He received his PhD in physics from Humboldt University of Berlin, Germany, in 1994. His research interests in near-infrared spectroscopy include the development of time-resolved, spatially resolved, and continuous wave methods for accurate measurement of tissue optical properties and for fluorescence imaging, related models of diffuse light propagation, and the design and characterization of tissue-simulating phantoms.

Heidrun Wabnitz received her Dr. rer. nat. degree from Friedrich Schiller University in Jena in 1982 and continued her research in molecular spectroscopy and laser scanning microscopy. She joined Physikalisch-Technische Bundesanstalt (PTB) in Berlin in 1991 and focused on optical mammography and optical brain imaging. Her activities in these fields of diffuse optical imaging and spectroscopy included the development of time-domain instrumentation, modeling, and data analysis, clinical applications, performance characterization of instruments, and standardization.

David R. Busch is an assistant professor in the Departments of Anesthesiology and Pain Management, Biomedical Engineering, and Neurology at the University of Texas Southwestern Medical Center. His work focuses on developing non- and minimally invasive optical tools to monitor deep tissues in health and disease, as well as throughout therapy.

Michele Lacerenza received his MSc in physics engineering, nano-optics and photonics from Politecnico di Milano, where he pursued his PhD in physics (2022), focusing on the development and exploitation of compact photonic platforms for cerebral and muscle oximetry monitoring. He served as a reviewer, director, chair/organizer of international scientific events and participated in several high-impact international research projects. He is now CTO and co-founder of PIONIRS s.r.l. aiming to foster novel discoveries and improve diagnostics in the medical sectors.

Marianne Floor-Westerdijk is the CEO at Artinis Medical Systems, a Netherlands-based leader in innovative solutions for near-infrared spectroscopy (NIRS) and electrophysiology (ExG). She has a background in R&D and biomedical engineering, with degrees from the University of Twente. At Artinis, she drives innovation in wearable brain and muscle monitoring technologies used worldwide in research and clinical applications. Her work focuses on combining science, usability, and scalable impact in neurophysiological measurement tools.

Sanathana Konugolu Venkata Sekar is a scientist and entrepreneur with over a decade of experience in biophotonics. He leads the Integrated Fast Biophotonics Group at Tyndall National Institute and is an associate professor at the School of Physics, as well as a recipient of an ERC grant. At IPIC, he coordinates biomedical research across nine PIs and over 25 researchers. He also co-founded BioPixS Ltd, which drives standardization in biophotonics. His research focuses on photonic integration and microprobes for healthcare, time-domain diffuse optics, GASMAS, and FLIM, with an emphasis on standardization, miniaturization, and real-world clinical impact.

Claudia Nunzia Guadagno is CTO at BioPixS Ltd. She received her master’s degree in physics Engineering in 2016. Her thesis topic was in time-domain diffuse optics for in vivo and phantom applications. Her area interests are time domain diffuse optics, phantoms, and standardization in biophotonics. She formerly worked for different companies and also received her BSc degree in communication in 2007.

Yukari Tanikawa is the deputy director of the Innovative Human Resources Department and concurrently serves at the Intelligent Systems Research Institute at the National Institute of Advanced Industrial Science and Technology (AIST), Japan. She received her PhD in engineering from the University of Electro–Communications, Japan. Her work focuses on the development of human resources in science and engineering, as well as promoting research on individual well-being in the workplace.

Hiroshi Kawaguchi is a senior research scientist at the Human Informatics and Interaction Research Institute, National Institute of Advanced Industrial Science and Technology (AIST), Japan. He received his PhD in engineering from Keio University, Japan. His current research interests include the development of human neuroimaging techniques, such as near-infrared spectroscopy and diffuse optical tomography. He was the project leader for the second revision of the international standard IEC 80601-2-71.

Funding Statement

This work was partly supported by internal competitive funding for standardization from the Department of Information Technology and Human Factors [Grant No. AIST R2-S-04 (HK)]. The authors acknowledge financial support by the European Union’s NextGenerationEU Programme with the I-PHOQS Infrastructure (Grant Nos. IR0000016, ID D2B8D520, and CUP B53C22001750006) “Integrated infrastructure initiative in Photonic and Quantum Sciences.”

Contributor Information

Caterina Amendola, Email: caterina.amendola@polimi.it.

Antonio Pifferi, Email: antonio.pifferi@polimi.it.

Alessandro Torricelli, Email: alessandro.torricelli@polimi.it.

Dirk Grosenick, Email: dirk.grosenick@ptb.de.

Heidrun Wabnitz, Email: heidrun.wabnitz@ptb.de.

David R. Busch, Email: david.busch@utsouthwestern.edu.

Michele Lacerenza, Email: lacerenza@pionirs.com.

Marianne Floor-Westerdijk, Email: marianne@artinis.com.

Sanathana Konugolu Venkata Sekar, Email: Sanathana. Konugolu@tyndall.ie.

Claudia Nunzia Guadagno, Email: claudia@biopixstandards.com.

Yukari Tanikawa, Email: yukari.tanikawa@aist.go.jp.

Hiroshi Kawaguchi, Email: hiroshi.kawaguchi@aist.go.jp.

Disclosures

M.F.W. is CTO of Artinis Medical Systems. M.L. is CTO and co-founder of PIONIRS. A.P. and A.T. are cofounders of PIONIRS S.r.l. C.N.G. is CTO of BioPixS Ltd. S.K.V.K. is co-founder and shareholder of BioPixS Ltd. with an interest in tissue-mimicking optical phantoms.

Code and Data Availability

The data that support the findings of this study are available from the corresponding author upon reasonable request.

References

  • 1.Ferrari M., Quaresima V., “A brief review on the history of human functional near-infrared spectroscopy (fNIRS) development and fields of application,” NeuroImage 63(2), 921–935 (2012). 10.1016/j.neuroimage.2012.03.049 [DOI] [PubMed] [Google Scholar]
  • 2.Pinti P., et al. , “The present and future use of functional near-infrared spectroscopy (fNIRS) for cognitive neuroscience,” Ann. N. Y. Acad. Sci. 1464(1), 5–29 (2020). 10.1111/nyas.13948 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Durduran T., et al. , “Diffuse optics for tissue monitoring and tomography,” Rep. Progr. Phys. 73(7), 076701 (2010). 10.1088/0034-4885/73/7/076701 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Quaresima V., Ferrari M., Scholkmann F., “Best practices for simultaneous measurement of NIRS-based cerebral and muscle oximetry during exercise,” J. Sport Health Sci. 14, 100997 (2024). 10.1016/j.jshs.2024.100997 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Miyai I., et al. , “Premotor cortex is involved in restoration of gait in stroke,” Ann. Neurol. 52(2), 188–194 (2002). 10.1002/ana.10274 [DOI] [PubMed] [Google Scholar]
  • 6.Ehlis A.-C., et al. , “Application of functional near-infrared spectroscopy in psychiatry,” NeuroImage 85, 478–488 (2014). 10.1016/j.neuroimage.2013.03.067 [DOI] [PubMed] [Google Scholar]
  • 7.IEC 80601-2-71:2025, Medical Electrical Equipment Part 2-71: Particular Requirements for the Basic Safety and Essential Performance of Functional Near-infrared Spectroscopy (Functional NIRS) Equipment, International Standard, International Organization for Standardization; (2025). [Google Scholar]
  • 8.Kawaguchi H., Eda H., “Development of the international standard for functional near-infrared spectroscopy as medical equipment,” Biomed. Opt. Express 15(9), 5272–5279 (2024). 10.1364/BOE.531664 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Kawaguchi H., et al. , “A phantom for standardization in functional near-infrared spectroscopy, part 1: implementation and usage,” Neurophotonics 12(4), 045010 (2025). 10.1117/1.NPh.12.4.045010 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Zhao F., et al. , “Reproducibility of identical solid phantoms,” J. Biomed. Opt. 27(7), 074713 (2022). 10.1117/1.JBO.27.7.074713 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Dalla Mora A., et al. , “The SiPM revolution in time-domain diffuse optics,” Nucl. Instrum. Methods Phys. Res. Sect. A: Accel. Spectrom. Detect. Assoc. Equip. 978, 164411 (2020). 10.1016/j.nima.2020.164411 [DOI] [Google Scholar]
  • 12.Fang Q., Yan S., “Graphics processing unit-accelerated mesh-based Monte Carlo photon transport simulations,” J. Biomed. Opt. 24(11), 115002 (2019). 10.1117/1.JBO.24.11.115002 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Boas D. A., et al. , “Three dimensional Monte Carlo code for photon migration through complex heterogeneous media including the adult human head,” Opt. Express 10(3), 159–170 (2002). 10.1364/OE.10.000159 [DOI] [PubMed] [Google Scholar]
  • 14.RS Components, “Solid plastic sheets—white acetal plastic sheet, 500 mm × 300 mm × 20 mm, rs stock no. 139-580,” https://uk.rs-online.com/.
  • 15.Berg R., et al. , “Time-gated viewing studies on tissue like phantoms,” Appl. Opt. 35(19), 3432–3440 (1996). 10.1364/AO.35.003432 [DOI] [PubMed] [Google Scholar]
  • 16.Di Sieno L., et al. , “Oxyhemoglobin measurements using 1064 nm light,” IEEE J. Sel. Top. Quantum Electron. 31(4), 1–6 (2025). 10.1109/JSTQE.2024.3512776 [DOI] [Google Scholar]
  • 17.Hacker L., et al. , “Criteria for the design of tissue-mimicking phantoms for the standardization of biophotonic instrumentation,” Nat. Biomed. Eng. 6(5), 541–558 (2022). 10.1038/s41551-022-00890-6 [DOI] [PubMed] [Google Scholar]
  • 18.Firbank M., Delpy D. T., “A phantom for the testing and calibration of near infra-red spectrometers,” Phys. Med. Biol. 39(9), 1509–1513 (1994). 10.1088/0031-9155/39/9/015 [DOI] [PubMed] [Google Scholar]
  • 19.Funane T., et al. , “Dynamic phantom with two stage-driven absorbers for mimicking hemoglobin changes in superficial and deep tissues,” J. Biomed. Opt. 17(4), 047001 (2012). 10.1117/1.JBO.17.4.047001 [DOI] [PubMed] [Google Scholar]
  • 20.Martelli F., et al. , “Phantoms for diffuse optical imaging based on totally absorbing objects, part 2: experimental implementation,” J. Biomed. Opt. 19(7), 076011 (2014). 10.1117/1.JBO.19.7.076011 [DOI] [PubMed] [Google Scholar]
  • 21.Pifferi A., et al. , “Mechanically switchable solid inhomogeneous phantom for performance tests in diffuse imaging and spectroscopy,” J. Biomed. Opt. 20(12), 121304 (2015). 10.1117/1.JBO.20.12.121304 [DOI] [PubMed] [Google Scholar]
  • 22.Pifferi A., et al. , “Performance assessment of photon migration instruments: the MEDPHOT protocol,” Appl. Opt. 44(11), 2104–2114 (2005). 10.1364/AO.44.002104 [DOI] [PubMed] [Google Scholar]
  • 23.Wabnitz H., et al. , “Performance assessment of time-domain optical brain imagers, part 2: nEUROPt protocol,” J. Biomed. Opt. 19(8), 086012 (2014). 10.1117/1.JBO.19.8.086012 [DOI] [PubMed] [Google Scholar]
  • 24.Wabnitz H., et al. , “Performance assessment of time-domain optical brain imagers, part 1: basic instrumental performance protocol,” J. Biomed. Opt. 19(8), 086010 (2014). 10.1117/1.JBO.19.8.086010 [DOI] [PubMed] [Google Scholar]
  • 25.Tak S., Ye J. C., “Statistical analysis of fNIRS data: a comprehensive review,” NeuroImage 85, 72–91 (2014). 10.1016/j.neuroimage.2013.06.016 [DOI] [PubMed] [Google Scholar]
  • 26.Yücel M. A., et al. , “Best practices for fNIRS publications,” Neurophotonics 8(1), 012101 (2021). 10.1117/1.NPh.8.1.012101 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Schroeter M. L., et al. , “Towards a standard analysis for functional near-infrared imaging,” NeuroImage 21(1), 283–290 (2004). 10.1016/j.neuroimage.2003.09.054 [DOI] [PubMed] [Google Scholar]
  • 28.Lanka P., et al. , “Multi-laboratory performance assessment of diffuse optics instruments: the bitmap exercise,” J. Biomed. Opt. 27(7), 074716 (2022). 10.1117/1.JBO.27.7.074716 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Whittet D. C. B., et al. , “The optical constants of polyoxymethylene,” Mon. Not. R. Astron. Soc. 175, 197–207 (1976). 10.1093/mnras/175.1.197 [DOI] [Google Scholar]