. 2026 Feb 11;47(3):243–249. doi: 10.1002/elps.70073

Determination of AAV Capsid Protein Ratios by Sodium Dodecyl Sulfate Capillary Electrophoresis (CE‐SDS)—UV Absorption Method

Yiran Li 1,2, Yufei Sun 3, Xiang Li 1,2, Hua Bi 1,2, Dening Pei 1,2, Zhihao Fu 1,2, Youxue Ding 1,2, Yue Pan 4, Xi Qin 1,2,✉, Chenggang Liang 1,2,✉, Lan Wang 1,2,✉
PMCID: PMC13005713  PMID: 41674186

ABSTRACT

Recombinant adeno‐associated viruses (AAVs) have become a leading platform for in vivo gene therapy. The AAV capsid is assembled from 60 individual protein subunits derived from three overlapping viral proteins (VPs): VP1, VP2, and VP3. The VP's ratio distribution affects the capsid trafficking and transgene expression. Quantitative analysis of the ratio is important in monitoring the potency of the AAVs. Sodium dodecyl sulfate capillary electrophoresis (CE‐SDS) offers automated and quantitative determination of the VP ratio distribution, and the method has been widely used in the application. The CE‐SDS method described was designed to ensure the consistent determination of AAV VP ratios across various sample concentrations. UV‐absorption detection was chosen for this purpose as it negated the need for any labeling in the process. To compensate for the lower sensitivity associated with UV‐absorption detection, methanol precipitation was employed during sample preparation, and a stacking injection technique was optimized for the samples in the study under CE conditions. These two techniques were found to be highly compatible. The method demonstrated robust linearity for AAV serotype 9 concentrations and exhibited good multiday precision. The VP ratios were successfully determined in AAV9 samples across a concentration range of 6.15 × 1012 to 1.85 × 1013 GC/mL. The relative standard deviation (RSD%) for VP ratio determination within this concentration range was maintained below 7% for VP3, VP2, and VP1. Additionally, this method was further applied to other AAV serotypes, including AAV2 and AAV8, illustrating its versatility and broad applicability.

Keywords: adeno‐associated virus (AAV), methanol precipitation, sodium dodecyl sulfate capillary electrophoresis (CE‐SDS), viral protein (VP)

1. Introduction

Recombinant adeno‐associated viruses (AAVs) have become a leading platform for in vivo gene therapy [1]. The clinical translation of AAV‐based therapies has accelerated in recent years, marked by regulatory approvals of several therapies, with numerous candidates advancing through clinical trials for various genetic and acquired diseases [2]. As for all therapeutics, to ensure the therapeutic efficacy and safety of AAV vectors, rigorous analytical characterization is required. Requirements must adhere to the guidance of various regulatory agencies throughout product process development and for final product release [3].

The AAV particle itself is a non‐enveloped virion, approximately 25 nm in diameter, with a T = 1 icosahedral capsid structure [4, 5]. The capsid packages a single‐stranded DNA genome (up to ∼4.7 kb) [6, 7]. The capsid is assembled from 60 individual protein subunits derived from three overlapping viral proteins (VPs): VP1 (81.2 kDa), VP2 (66.2 kDa), and VP3 (59.7 kDa) [8, 9]. The typical stoichiometric ratio of the three proteins is approximately VP1:VP2:VP3 = 1:1:10. The VP's ratio distribution affects the capsid trafficking and transgene expression [10, 11]. Quantitative analysis of the ratio is important in monitoring the potency of AAVs.

Sodium dodecyl sulfate capillary electrophoresis (CE‐SDS) offers automated and quantitative determination of the VP ratio distribution [12, 13, 14, 15, 16, 17]. Two detection modes have been used for CE‐SDS determination of AAV VP ratio: UV absorption detection (UV absorption) and laser‐induced fluorescence (LIF) detection [15, 17]. UV absorption provides reliable quantitative data, as it is a direct method to quantify proteins without bias. However, its sensitivity may not be sufficient for some applications. In contrast, LIF detection has higher sensitivity but requires covalent dye labeling of proteins sample before separation [15]. The labeling may generate method‐related impurities and may be bias against certain components when different dyes are used [16]. Reviewing the literature, we found limited study of VP ratio determination within a certain sample concentration range by CE‐SDS. For a method used to determine the VP ratio of AAVs, the measured values in a given AAV concentration range need to be consistent. Capsid protein concentration linearity has been shown for total VP or individual VPs; however, to our knowledge, none demonstrate consistency of the measured VP ratios in this linear range.

In this study, the determination of the VP ratio via CE‐SDS was investigated throughout the sample concentration linear range. Both sensitivity and reproducibility were prioritized during the development of our analytical method. To ensure reproducible VP ratios across a defined range of sample concentrations, an optimal total AAV VP sample loading concentration was first established for VP ratio assessments. The signal linear range for total VPs of the method should be within ±50% of this predetermined loading concentration. Furthermore, the VP ratios calculated within this range were consistent.

To meet these criteria, UV absorption detection was implemented to negate the necessity for sample labeling, thereby eliminating variations in labeling across different sample concentrations. Additionally, the method's sensitivity was augmented by incorporating an organic solvent precipitation technique, which is widely utilized in proteomic research [18]. This precipitation strategy facilitated the application of sample stacking method in electrokinetic (EK) injection within our CE‐SDS protocol, further enhancing the method's sensitivity [19].

2. Materials and Methods

2.1. Materials and Instruments

AAV9 (5.8 × 1014 GC/mL), AAV2, and AAV8 products were obtained from their manufacturers. Methanol (HPLC plus grade) and 2‐mercaptoethanol (β‐ME, M3148) were purchased from Sigma (St. Louis, MO, USA). Reagents for CE‐SDS runs were from the Reagent Kit of Maurice CE‐SDS plus (PS‐MRK01‐S, ProteinSimple, San Jose, CA, USA). The reagents in the Reagent Kit included column condition solutions, running buffer for filling capillary column, and wash solution.

Maurice CE instrument and its CE‐SDS Plus cartridge (catalog # 090‐000 and PS‐MC02‐SP, ProteinSimple, San Jose, CA, USA) were used in the method.

2.2. Sample Preparation

A volume of 20 µL AAV sample solution was transferred to a 500‐µL centrifuge tube, whereas 180 µL ice‐cold methanol was added to the tube. The mixture was vortexed for 15 s. The mix was stored overnight for about 12 h at −80°C. The sample in the tube was centrifuged for 30 min at 13 200 rpm at −10°C. After the centrifuging, the supernatant was carefully removed from the tube. The precipitated pellet in the bottom of the tube was let air dry for 10 min. The pellet was dissolved with 10 µL 1× CE‐SDS Plus Buffer. The tube was vortexed for 10 s, then it was centrifuged for 5 s. The above two steps were repeated for three times. β‐ME (at 5% V/V in final 10 µL solution) and CE‐SDS internal standard (IS, 4% final) were added to the solution. The sample mixture was heated at 70°C for 10 min and was set cool to room temperature. The sample tube was centrifuged firefly for 10 s. Finally, 40 µL water was added into the tube, and it was mixed well.

2.3. CE‐SDS Method

In CE‐SDS methods run on Maurice CE instrument, all steps were predetermined by the software with minimum requirement for operator interference. Before sample injection, the capillary column was conditioned by Condition Solutions 1 and 2 with deionized water between the Condition Solution washes. The column was filled with running buffer by pressure. The samples were loaded onto the CE‐SDS Plus cartridge using EK injection for 20 s at 4600 kV. The separation of samples was conducted at 5750 V over 35 min, with detection accomplished through 220 nm absorbance, using Compass for ICE software for data acquisition.

3. Results and Discussion

3.1. Method Optimization Description

Three organic solvents, acetone, ethanol and methanol, were tested for precipitating VPs. For the AAV9 sample, methanol showed slightly higher recovery rate (data are not shown).

To keep a high portion of methanol in sample precipitation step, the ratio of 1 to 9 (sample volume to methanol volume) was used in the precipitation step. However, different sample volumes were tested: 5, 10, 20, 30, and 40 µL. These sample volumes gave the same recovery rate for different AAV9 sample concentrations. To balance the sample volume and precision, 20 µL sample:180 µL methanol ratio was used in the AAV precipitation.

3.2. Method Sensitivity

To circumvent the complications associated with sample labeling, such as impurities and introduced bias, UV absorption detection at 220 nm was opted for determining the VP ratio in this study. To address the lower sensitivity associated with UV detection, strategies like methanol precipitation of the AAV particles before denaturing the sample, followed by a streamlined sample stacking technique during injection, were applied.

The ratio of methanol volume to sample volume was fixed at 9:1. The precipitated solid pellet was then dissolved in 10 µL of sample buffer. This maintains that the final sample volume remains consistently at 10 µL in undiluted sample buffer, allowing the SDS and other compounds to stay at optimal concentrations for effective protein denaturation. The 10 µL of solution was subsequently heated for denaturing the AAV particles. This volume also matched well with the stacking injection method utilized under our CE conditions.

For the sample stacking in EK injection, after denaturation, to achieve the good stacking effect, the sample buffer in the sample needed a five‐fold dilution with a low conductivity solvent, such as water. After the precipitation step, the precipitated VP pellet was dissolved in 10 µL sample buffer (total 10 µL volume). This precipitation‐re‐dissolving step ensured that there was adequate room for the dilution by 40 µL water, resulting in the required final sample volume of 50 µL by the method for the injection into the instrument. The stacking effect during the EK injection was achieved by this diluted sample buffer in the sample. The method aligns closely with the process reported in reference [19]. In our application, although the separation capillary column was filled with undiluted sample buffer, the sample solution contained the same sample buffer but was diluted five times with water. The conductivity of the sample solution was much lower than that of the conductivity of the undiluted sample buffer that filled in the separation capillary. During the EK injection, a greater voltage drop happened in the low conductivity section (diluted sample solution in sample around the capillary column inlet) compared to that throughout the separation column, owing to the diluted solution's lower conductivity. This led to the accumulation of sample molecules at the capillary column inlet.

Figure 1 displays a comparison between two sample injections: one with stacking, A and the other without, B. Each sample was positioned at 50 µL volume and 2.0 × 1012 GC/mL concentration under identical electrophoresis conditions. The primary difference was in sample buffer concentration; Sample A featured a five‐fold diluted sample buffer concluding at a 50 µL volume (sample stacking), whereas Sample B maintained an undiluted sample buffer also reaching a volume of 50 µL (for normal sample injection method). With the implementation of our stacking method, the peak area increased six‐fold (Figure 1) with stacking sample injection.

FIGURE 1.

FIGURE 1

(A) 2.0 × 1012 GC/mL AAV9 concentration in 50 µL volume with 0.2× sample buffer concentration; (B) the same sample concentration as that in (A), 50 µL sample volume with 1× sample buffer concentration. All other electrophoresis conditions are the same for (A) and (B), as described in CE‐SDS condition.

Combined with methanol precipitation and stacking in sample injection, the limit of quantitation (LOQ) of the method, defined as 10× peak height/base line noise ratio, was 1.2 × 1011 GC/mL for the VP3 peak.

3.3. Total VP Concentration Linear Range

The combined method of methanol precipitation and stacking injection had good linearity for total VPs, which expanded from 2.0 × 1012 to 3.0 × 1013 GC/mL with R 2 = 0.988, as shown in Figure 2.

FIGURE 2.

FIGURE 2

Linear range of total VP concentration. Each data point was the results of three injections (error bar in each point).

Despite the linearity across the relatively wide concentration range, the determined VP ratio was found to be inconsistent due to two primary factors. Initially, at the lower end of the total VP range, the peaks for VP1 and VP2 fell below the LOQ, resulting in a decreased representation in their peak areas. Conversely, at the upper end of the range, the detection response for the principal component, VP3, exhibited slight saturation, which consequently diminished the percentage of the VP3 peak area. Therefore, the sample concentration range yielding a consistent VP ratio should be more restrictive than the overall linear range.

3.4. Typical Sample Loading Concentration and Concentration Range for Consistent VP Ratio Determination

To find the concentration range that had the consistent determined VP ratio, we first determined the typical sample loading concentration by reviewing the data. The concentration was determined to be 1.23 × 1013 GC/mL for the AAV9 sample. The method precision at this concentration point, as shown in Table 1, was evaluated by running the concentration for 3 days, with three injections in each day. Figure 3 shows the representative electropherograms of the sample at the concentration obtained in 3 days. The method was reproducible.

TABLE 1.

Precision of the method at the typical loading concentration (1.23 × 1013 GC/mL).

VP3 VP2 VP1
Day 1 ave 91.4 4.8 3.7
Day 2 ave 91.7 4.7 3.6
Day 3 ave 91.3 5.1 3.7
Ave 91.5 4.8 3.7
RSD% 0.2 4.2 1.4

Abbreviation: RSD%, relative standard deviation.

FIGURE 3.

FIGURE 3

Three‐day representative electropherograms of VPs at AAV concentration of 1.23 × 1013 GC/mL (B‐D) and the reagent blank (A).

After the typical loading concentration was determined, we expanded the concentration range for test surrounding this concentration point in 1.23 × 1013 ± 50% GC/mL (6.15 × 1012 to 1.85 × 1013 GC/mL) of the AAV9 sample to see if the concentration range showed consistent VP ratio. This AAV9 concentration range ensured both the VP1 and VP2 lowest concentrations were above LOQ, and the VP3 highest concentration was in the linear range as shown in Figure 2. Figure 4 shows the electropherograms of the AAV9 samples in the concentration range of 1.23 × 1013 ± 50% GC/mL. The three VPs had good linearity in the concentration range with R 2 > 0.998. In this AAV9 concentration range, the VP ratio was consistent as shown in Table 2. The relative standard deviation (RSD%) for three VP's percentage determination was within 7%, even for minor components VP2 and VP1.

FIGURE 4.

FIGURE 4

AAV9 in the concentration range 6.15 × 1012 to 1.85 × 1013 GC/mL.

TABLE 2.

Viral protein (VP) ratio at different adeno‐associated viruses (AAV) concentration in the concentration range of Figure 4.

Concentration ×1012 GC/mL VP3 (%) VP2 (%) VP1 (%)
6.2 92.2 4.6 3.2
6.2 92.3 4.2 3.5
6.2 92.1 4.5 3.5
9.3 91.4 4.9 3.7
9.3 91.6 4.8 3.7
12.3 91.1 5.1 3.7
12.3 91.4 5.0 3.6
15.4 91.0 5.2 3.9
15.4 91.2 5.1 3.7
15.4 90.8 5.3 3.9
18.5 90.9 5.3 3.9
18.5 91.1 5.1 3.8
18.5 91.2 5.1 3.8
Ave 91.4 4.9 3.7
RSD% 0.5 6.7 5.4

Abbreviation: RSD%, relative standard deviation.

3.5. Application to Other Serotypes

The method was applied to AAV2 and AAV8 samples. Figure 5 shows typical electropherograms of the two AAV samples. For the two samples, AAV2 and AAV8, a front peak was separated before VP3 peak as reported [15]. It was identified as VP3 clip by reversed‐phase liquid chromatography–mass spectrometry method [20]. The measured VP ratios of the two samples by our method are VP3:VP2:VP1 = 12.4:1.0:1.3 for AAV2 and 7.1:1.4:1.0 for AAV8.

FIGURE 5.

FIGURE 5

Typical electropherograms: (A) AAV2 (5.0 × 1012 GC/mL), (B) AAV8 (1.0 × 1013 GC/mL), and (C) blank run only with internal standard (IS).

4. Conclusions

The CE‐SDS method developed in this study was designed to ensure the consistent determination of AAV VP ratios across various sample concentrations. UV‐absorption detection was chosen for this purpose as it negated the need for any labeling in the process. To compensate for the lower sensitivity associated with UV‐absorption detection, methanol precipitation was employed during sample preparation, and a stacking injection technique was adapted under CE conditions. These two techniques were found to be highly compatible.

The precipitation method effectively concentrated the samples into a consistent volume of 10 µL in the sample buffer, irrespective of the initial sample volume and concentration. This concentrated sample was then diluted to the standard analysis volume of 50 µL with water, creating a solution with low conductivity, which is crucial for effective stacking injections.

The method demonstrated robust linearity for AAV serotype 9 concentrations and exhibited good multiday precision. The VP ratios were successfully determined in AAV9 samples across a concentration range of 6.15 × 1012 to 1.85 × 1013 GC/mL. The RSD% for VP ratio determination within this concentration range was maintained below 7% for VP3, VP2, and VP1. Additionally, this method was further applied to other AAV serotypes, including AAV2 and AAV8, illustrating its versatility and broad applicability.

Conflicts of Interest

The authors declare no conflicts of interest.

Acknowledgments

This work was supported by the Fundamental Research Funds for the Central Public‐interest Scientific Research Institutes, Chinese Academy of Medical Sciences (Grant Number 2023‐PT350‐01), “Quality Evaluation and Related Mechanisms of Novel Biologics including Gene Therapy Drugs and mRNA Vaccines.”

Contributor Information

Xi Qin, Email: qinxi@nifdc.org.cn.

Chenggang Liang, Email: liangchenggang@nifdc.org.cn.

Lan Wang, Email: wanglan@nifdc.org.cn.

Data Availability Statement

The data that support the findings of this study are available on request from the corresponding author. The data are not publicly available due to privacy or ethical restrictions.

References

  • 1. Naso M. F., Tomkowicz B., Perry W. L., and Strohl W. R., “Adeno‐Associated Virus (AAV) as a Vector for Gene Therapy,” Biodrugs 31 (2017): 317–334, 10.1007/s40259-017-0234-5.dfd. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Wang J. H., Gessler D. J., Zhan W., Gallagher T. L., and Gao G., “Adeno‐Associated Virus as a Delivery Vector for Gene Therapy of Human Diseases,” Signal Transduction and Targeted Therapy 9 (2024): 78, 10.1038/s41392-024-01780-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Wright J. F., “Product‐Related Impurities in Clinical‐Grade Recombinant AAV Vectors: Characterization and Risk Assessment,” Biomedicines 2 (2014): 80–97, 10.3390/biomedicines2010080. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Girod A., Wobus C. E., Zádori Z., et al., “The VP1 Capsid Protein of Adeno‐Associated Virus Type 2 Is Carrying a Phospholipase A2 Domain Required for Virus Infectivity,” Journal of General Virology 83 (2002): 973–978, 10.1099/0022-1317-83-5-973. [DOI] [PubMed] [Google Scholar]
  • 5. Grieger J. C., Snowdy S., and Samulski R. J., “Separate Basic Region Motifs Within the Adeno‐Associated Virus Capsid Proteins Are Essential for Infectivity and Assembly,” Journal of Virology 80 (2006): 5199–5210, 10.1128/JVI.02723-05. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Mietzsch M., Jose A., Chipman P., et al., “Completion of the AAV Structural Atlas: Serotype Capsid Structures Reveals Clade‐Specific Features,” Viruses 13 (2021): 101, 10.3390/v13010101. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Xie Q., Bu W., Bhatia S., et al., “The Atomic Structure of Adeno‐Associated Virus (AAV‐2), a Vector for Human Gene Therapy,” PNAS 99 (2002): 10405–10410, 10.1073/pnas.162250899. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Becerra S. P., Koczot F., Fabisch P., and Rose J. A., “Synthesis of Adeno‐Associated Virus Structural Proteins Requires Both Alternative mRNA Splicing and Alternative Initiations From Single Transcript,” Journal of Virology 62 (1988): 2745–2754, 10.1128/jvi.62.8.2745-2754.1988. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Becerra S. P., Rose J. A., Hardy M., Baroudy B. M., and Anderson C. W., “Direct Mapping of Adeno‐Associated Virus Capsid Proteins B and C: A Possible ACG Initiation Codon,” PNAS 82 (1985): 7919–7923, 10.1073/pnas.82.23.7919. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Venkatakrishnan B., Yarbrough J., Domsic J., et al., “Structure and Dynamics of Adeno‐Associated Virus Serotype 1 VP1‐Unique N‐Terminal Domain and Its Role in Capsid Trafficking,” Journal of Virology 87 (2013): 4974–4984, 10.1128/JVI.02524-12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Powell S. K. and McCown T. J., “Adeno‐Associated Virus 9 (AAV9) Viral Proteins VP1, VP2, and Membrane‐Associated Accessory Protein (MAAP) Differentially Influence In Vivo Transgene Expression,” Journal of Virology 98 (2024): e0168124, 10.1128/jvi.01681-24. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Aebischer M. K., Bouvarel T., Barrozo E., et al., “Boosting the Separation of Adeno‐Associated Virus Capsid Proteins by Liquid Chromatography and Capillary Electrophoresis Approaches,” International Journal of Molecular Sciences 24 (2023): 2583, 10.3390/ijms24108503. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Rodionova N., Peng Z., Miller E., and Nilsson M., PerkinElmer Poster Presentation (2020), https://www.mabdesign.fr/wp‐content/uploads/2020/09/Characterization‐of‐AAV‐structure.pdf.
  • 14. Smith J., Carillo S., Kulkarni A., Redman E., Yu E., and Bones J., “Rapid Characterization of Adeno‐Associated Virus (AAV) Capsid Proteins Using Microchip ZipChip CE‐MS,” Analytical and Bioanalytical Chemistry 416 (2024): 1069–1084, 10.1007/s00216-023-05097-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Li T., Santos M., and Guttman A., Ultrahigh Sensitivity Analysis of Adeno‐Associated Virus Chromatography Today 2021; Buyers’ Guide , https://www.labmate‐online.com/article/bioanalytical/40/sciex/ultrahigh‐sensitivity‐analysis‐of‐adeno‐associated‐virus‐aavbr‐capsid‐proteins‐by‐sodium‐dodecyl‐sulphate‐capillary‐gel‐electrophoresis/2852.
  • 16. Li X., Gao T., Shi X.‐C., et al., “Analysis of the Proportion Content of Recombinant Adeno‐Associated Virus (rAAV) Capsid Protein by CGE‐LIF*,” Chinese Journal of Pharmaceutical Analysis 43 (2023): 1833–1839, 10.16155/j.0254-1793.2023.11.04. [DOI] [Google Scholar]
  • 17. Li T., Malik M., Yowanto H., and Mollah S., Purity Analysis of Adeno‐Associated Virus (AAV) Capsid Proteins Using CE‐SDS Method , Sciex Application Note RUO‐MKT‐02‐9761‐A (SCIEX, 2019). [Google Scholar]
  • 18. Pérez‐Rodriguez S., Ramírez O., Trujillo‐Roldán M. A., and Valdez‐Cruz N. A., “Comparison of Protein Precipitation Methods for Sample Preparation Prior to Proteomic Analysis of Chinese Hamster Ovary Cell Homogenates,” Electronic Journal of Biotechnology 48 (2020): 86–94, 10.1016/j.ejbt.2020.09.006. [DOI] [Google Scholar]
  • 19. Zhang C.‐X. and Meagher M. M., “Sample Stacking Provides Three Orders of Magnitude Sensitivity Enhancement in SDS Capillary Gel Electrophoresis of Adeno‐Associated Virus Capsid Proteins,” Analytical Chemistry 89 (2017): 3285–3292, 10.1021/acs.analchem.6b02933. [DOI] [PubMed] [Google Scholar]
  • 20. Zhao X., Jin X., Liu L., et al., “Optimized Reversed‐Phase Chromatography/Mass Spectrometry Methods for Intact Protein Analysis and Peptide Mapping of Adeno‐Associated Virus Proteins,” Human Gene Therapy 32 (2021): 23–24, 10.1089/hum.2021.046. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data that support the findings of this study are available on request from the corresponding author. The data are not publicly available due to privacy or ethical restrictions.