Abstract
Soft bioelectronic devices can be interfaced with anatomically curved organs, such as the heart, brain and skin, to provide continuous analysis of physiological information. However, body movements and physiological activities may induce motion artefacts, which can adversely affect signal accuracy and stability. Importantly, motion artefact management is key to promoting the clinical translation of soft bioelectronics to ensure that soft bioelectronic devices can selectively detect target biological signals with high accuracy. In this Review, we discuss how body activities can affect the soft bioelectronic–tissue interface and result in motion artefact signals, including interface impedance instability motion artefacts, biopotential motion artefacts and mechanical motion artefacts. We then investigate different motion artefact management strategies, including materials engineering, device and circuit design, and algorithmic intervention, to reduce the contribution of motion artefacts to signal acquisition, processing and interpretation.
Introduction
Soft bioelectronics can continuously produce clinical-grade data for the personalized and remote monitoring of health and disease1–3. In contrast to rigid silicon microelectronics, which fail to establish long-lived and robust contacts with soft biological surfaces, soft bioelectronics are designed for seamless and conformal integration with the human body4, particularly for the management of diseases currently limited by delayed or inefficient medical services such as infectious diseases and care for the ageing population5–8. Soft bioelectronics-based body area networks with monitoring, therapeutic and energy functions9,10 can consistently collect and analyse physiological information, providing continuous and real-time health-care data to identify dynamic physiological changes and potential health issues at an early stage8,11,12. Therefore, soft bioelectronics may enable personalized, predictive and participatory health care for disease prevention and health promotion.
A soft bioelectronic system usually encompasses sensing units and electrode components for the acquisition of diverse physiological information, spanning electrophysiological, biochemical and biomechanical signals as well as circuits for signal processing and transmission (Fig. 1a). By establishing a seamless and stable bioelectronic–tissue interface, soft bioelectronics can provide conformability and monitoring accuracy13,14. However, the stability of the bioelectronic–tissue interface is susceptible to human physiological activities, which can cause motion artefacts that may disruptively modify target signals, thereby challenging the accuracy and reliability of soft bioelectronic measurements. Depending on the causes of formation, motion artefacts can be classified into three categories: interface impedance instability motion artefacts, biopotential motion artefacts and mechanical motion artefacts.
Fig. 1 |. Soft bioelectronic systems and bioelectronic–tissue interfaces.

a, A typical soft bioelectronic system includes electrodes and sensors for signal acquisition as well as circuits for signal processing and transmission. b, Soft bioelectronics interact with the human body to extract physiological information, including continuous electrophysiological, biochemical and biomechanical signal monitoring. ECG, electrocardiogram.
Interface impedance instability motion artefacts originate from impedance instability at the bioelectronic–tissue interface, attributed to soft bioelectronic detachment and displacement-induced changes in the electrode contact state15. Interface impedance instability motion artefacts can induce signal baseline drift and severe signal distortion16.
Biopotential motion artefacts stem from biopotential signals, produced by the electrochemical activities of excitable cells in nervous and muscular tissues. These biopotential signals may be captured and can interfere with the acquisition and interpretation of the target signals, which may limit physiological monitoring, in particular, of weak or closely spaced signals17–19.
Mechanical motion artefacts can be caused by biomechanical activities such as body movements and physiological functions. These activities lead to mechanical disturbances that propagate to soft bioelectronics, thereby introducing irrelevant biomechanical signal contributions from the device itself to the target signals20,21.
Such motion artefacts place challenges to the clinical translation of soft bioelectronics and thus need to be addressed. In this Review, we introduce the configuration of a typical soft bioelectronic system and bioelectronic–tissue interfaces for electrophysiological, biochemical and biomechanical signal measurements. We then analyse the three types of motion artefact and examine methods for their elimination and management. Specifically, to mitigate motion artefacts from interface impedance instability, we delve into the construction of soft, stretchable and adhesive soft bioelectronics, proffering multifaceted solutions, including material–tissue mechanical matching, device structure engineering and bioelectronic–tissue adhesion. For biopotential and mechanical motion artefact management, we underscore the pivotal role of algorithmic intervention from the software perspective, alongside hardware-based strategies, collectively encompassing both sensor and circuit designs. By exploring these solutions to motion artefact management strategies, the clinical translation of soft bioelectronics for personalized health care may be greatly accelerated.
Soft bioelectronic–tissue interfaces
Soft bioelectronics, owing to their mechanical conformability, can seamlessly integrate with curved tissues. The bioelectronic–tissue interface between soft bioelectronics and living tissues enables stable mechanical coupling and the efficient capture of physiological signals13,22–29. Depending on the nature and source of physiological signals, this interface facilitates the monitoring of electrophysiological, biochemical and biomechanical signals (Fig. 1b). For electrophysiological monitoring, soft electrodes (with or without an adhesive layer) establish an electrical interface with living tissues to accurately and imperceptibly capture weak electrophysiological activities through various biopotential recordings, including electrical activities in the brain (electroencephalogram (EEG)), heart (electrocardiogram (ECG)), eye (electrooculogram (EOG)) and muscle (electromyogram (EMG)). Soft microelectrodes can further be implanted to establish brain–machine interfaces30,31.
Biochemical monitoring is typically employed for the analysis of biomarkers in easily accessible body fluids such as sweat, saliva and blood. For example, biochemical sensors can initiate chemical reactions with body fluids on the skin, such as electrolytes (for example, Na+, K+ and Ca2+) and organic molecules (for example, glucose, lactate, hormones), generating electrical signals for real-time biochemical parameter measurement3.
For biomechanical monitoring, soft bioelectronics employ mechanical sensors to capture biophysical parameters, such as stress, pressure and deformation, enabling the detection of biomechanical processes within the body, including wrist pulse, respiration movements and body-bending signals, through biomechanical-to-electrical conversion32.
However, the coupling state of bioelectronic–tissue interfaces can be influenced by various factors. Over time, changes may occur in the structure and properties of soft bioelectronics, leading to alterations in their sensing performance. For example, structural instability in bioelectronics may deteriorate sensing performance, including dislodgement of the conductive layer on the electrode33,34 and corrosion within the ionic biofluid environment35,36. Similarly, ion-selective, gel-based sensors may exhibit property fluctuations in response to varying environmental ion concentrations. The impact of such device changes at the bioelectronic–tissue interface may be reduced by material design and sensor replacement. Nonetheless, motion artefacts, the most prevalent factor affecting the bioelectronic–tissue interface, can result in distortions or unreliability in monitoring data and remain difficult to address.
Mechanisms of motion artefacts
Interface impedance instability motion artefacts
The electrical characteristics of the bioelectronic–tissue interface play a crucial role in electrophysiological monitoring35,37. To capture high-quality electrophysiological signals from living tissues using flexible electrodes, a stable interface has to be maintained. A simplified equivalent circuit model consisting of resistance and capacitance can be applied to represent the interface impedance characteristics at different frequencies38,39 (Fig. 2a). Depending on the contact state of the interface, the electrode can establish direct contact (as a dry electrode) or engage through a conductive gel layer (as a wet gel electrode) with the tissue. The interface impedance of these contact modes can be quantitatively expressed as:
| (1) |
| (2) |
| (3) |
where Rt, Re and Rd are the resistance of tissue, electrode and gel, respectively; Ce and Cd are the capacitance of the electrode–epidermis-or-gel interface and gel–epidermis interface, respectively; j is the imaginary component; ω is the angular frequency; and Z and Z′ are the impedance of the bioelectronic–tissue interface in dry electrode and wet gel electrode contact modes, respectively. For capacitance, ε0 and εr are the relative permittivity of free space and material, respectively; A is the contact area of the plate; and d is the distance between two plates. Resistance primarily influences the transmission of low-frequency signals by attenuating them as they traverse the circuit, whereas capacitance predominantly facilitates the transmission of high-frequency signals by storing and releasing electrical energy. Compared to the dry electrode, the gel layer of the wet gel electrode acts as an electrolyte between the electrode and epidermis, providing adhesion and reducing the interface impedance from 200 Ω of the dry electrode to 10 Ω of the wet gel electrode. This configuration offers stable tissue integration and accurate bioelectrical signal capture40–42.
Fig. 2 |. Motion artefacts in soft bioelectronic systems.

a, Interface impedance instability at electrode–tissue interfaces. b, Biopotential signals are generated by excitable cells through ion transmembrane transport. Overlapping biopotential signals can disturb the acquisition of target electrophysiological signals. c, Signal artefact disruptions to target signals can result from biomechanical activities, including body movements and physiological functions. Re, resistance of electrode; Rt, resistance of tissue; Rd, resistance of gel; Ce, capacitance of the electrode–epidermis interface; Cd, capacitance of the electrode–gel interface; U, transmembrane potential.
Soft bioelectronics continuously slide and are displaced owing to body activities such as skin deformation or sweat accumulation at the interface. The resulting contact-state changes of the interface lead to contact capacitance variations within the circuit model, causing fluctuations of interface impedance (at the order of 100–1,000 Ω for dry electrodes). By contrast, the adhesive gel layer in wet gel electrodes reduces electrode movement and fluctuations in interface impedance (in the order of 10–100 Ω)43. However, during body activities, mechanical mismatch between the conductive gel and electrode can lead to electrode delamination and fragmentation as well as to polymer fatigue from the conductive gel, impacting the stability of electrode–gel interface impedance16,23. The consequential fluctuations in interface impedance instability considerably impact current transmission at the interface, causing distortions to the recorded signals and thereby hindering the acquisition of accurate and reliable electrophysiological and impedance-related signals such as impedance-based skin hydration, electrical impedance tomography and bioimpedance-based blood pressure measurement.
Biopotential motion artefacts
Biopotentials are electrical signals generated by excitable cells during physiological processes44 inherent in daily activities. At the cellular level, these signals arise from variations in ionic concentrations across cell membranes (Fig. 2b). Biopotential signals include EOG from the movement of eyeballs and eyelids, EMG from muscular movements, and ECG from heart muscle during heartbeats45. These biopotential signals can be captured simultaneously through bioelectronic–tissue interfaces; however, overlapping signals may lead to signal distortion during electrophysiological monitoring20.
EEG is a non-linear and non-stationary signal pattern, characterized by complex components and reflecting important information about brain electrical activity. EEG signals exhibit microvolt-level amplitudes and weak intensities, rendering them susceptible to interference from other physiological signals originating within the human body46,47. During EEG measurements, eye blinks or movements can produce spike-like signal waveforms, with peak amplitudes reaching up to 800 μV (ref. 48). These waveforms, referred to as EOG signals, produce biopotential motion artefacts for EEG signals owing to their considerably higher magnitude18. Similarly, higher-amplitude ECG signals can substantially influence EMG signals, in particular, when recording from upper trunk muscles49–51. This influence results in an augmentation of power content and distortion of the frequency characteristics of EMG signals. Therefore, such biopotential motion artefacts stemming from body activities should be addressed. By contrast, the detection of biochemical and biomechanical signals follows mechanisms that do not require the direct interface of electrodes with biological tissues, rendering them unaffected by biopotential motion artefacts and offering a pathway to uncontaminated and robust data acquisition.
Mechanical motion artefacts
Mechanical motion artefacts are induced by biomechanical activities, causing mechanical disturbances that propagate through organs and tissues, ultimately reaching the soft bioelectronics in contact with the body. This cascade of motion artefacts introduces irrelevant biomechanical signals to the target signals, which often occurs in wearable sensing applications such as biomechanical, biochemical, electrophysiological and temperature sensors. These mechanical disturbances lead to interference and signal distortion, impacting the accuracy and reliability of physiological monitoring.
Mechanical motion artefacts can originate from various sources, including body movements, such as walking, running or jumping, as well as physiological functions such as heartbeat and respiration52–54 (Fig. 2c). For example, interference from gait motions can introduce signal artefacts with high amplitudes and irregular frequencies, exceeding those of most physiological signals, thereby disturbing physiological monitoring. Consequently, the low signal quality impacts data extraction and analysis, preventing the detection of valuable physiological information, particularly from self-powered soft bioelectronics based on biomechanical-to-electrical conversion55. By contrast, interference sources from physiological functions typically exhibit small amplitudes and stable frequencies. The generated signal artefacts often behave at similar levels of frequencies or amplitudes to the target physiological signals, exerting a consistent influence on signal acquisition. Physiological signals and mechanical motion artefact signals usually coexist across different frequency bands, allowing the possibility of separating the contributions of these overlapping functions based on their characteristic frequencies56,57. Thus, signals with motion artefacts can still provide valuable physiological information through various post-signal processing algorithms.
Interface impedance instability motion artefact management
Interface impedance instability motion artefacts originate from changes in contact status caused by the detachment and displacement of soft bioelectronics, resulting in subsequent fluctuations in interface impedance. The management of interface impedance instability motion artefacts revolves around strategies to ensure a robust connection of soft bioelectronics with irregularly curvilinear and dynamically changing living tissue surfaces. Managing interface impedance instability motion is particularly important for implantable bioelectronics used in electrocorticography, local field potentials and electroneurogram to ensure accurate monitoring of brain and peripheral nerve activity. Maintaining stable contact at the bioelectronic–tissue interface during body movements is key to suppressing interface impedance variations. To enhance conformability, bioelectronics should be stretchable, soft and adhesive to dissipate strain energy and rapidly adapt to dynamic tissue surfaces. This can be achieved by ensuring material–tissue mechanical matching and device structure engineering to enhance structural deformation. Additionally, bioelectronic–tissue adhesion should be ensured to facilitate the seamless integration of bioelectronics with living tissues.
Material–tissue mechanical matching
The mechanical properties of electronic materials, particularly their intrinsic stretchability and softness (modulus matching), are crucial for achieving contact stability with living tissues. Matching the mechanical properties of the bioelectronics with those of the tissue permits conformability under dynamic changes to manage interface impedance instability of the bioelectronic–tissue interface.
Intrinsic stretchability.
The stretchability variation of different living tissues during various body activities reflects their physiological characteristics and functional requirements (Fig. 3a) and needs to be accounted for in the design of soft bioelectronics. Adapting to the surface topography of the tissue is paramount in eliminating the strain mismatch from impedance instability motion artefacts induced by bioelectronic–tissue interfaces58. For example, the nervous system, including the brain and spinal cord59,60, is characterized by low stretchability. The brain and spinal cord can experience tensile strains of 5% and 10–20%, respectively, during normal postural movements. Conversely, muscles, including skeletal and cardiac muscles, exhibit a moderate strain range, typically between 20% and 40%, which corresponds to their muscle activities and cardiac physiological functions61. Furthermore, joints, such as the knee, fingers and elbows, which perform a range of movements, including bending, stretching and twisting, exhibit the greatest strain range among tissues owing to their need for flexibility and accommodation of large-scale motions62.
Fig. 3 |. Interface impedance instability motion artefact management.

a, Stretchability and softness of human tissues. b, Stretchability and softness of soft bioelectronic materials. c, Interface impedance instability motion artefacts can be managed by reducing device thickness. To provide conformal contact with the irregular shape of living tissues during human body activities, the thickness of soft bioelectronics can be reduced to an ultrathin structure to increase the interfacial contact energy (Uinterface). d, Device strain can be adjusted by modifying the structure of the device, for example, by including an island bridge design, in which a connecting bridge is used for stretching while not affecting functional rigid islands. To increase the mechanical stability of soft bioelectronics, an elastiff layer can be introduced to divert strain to the deformable, low-modulus interspace. e, Deformation-tolerant designs provide reversible stretchability of soft bioelectronics; a wrinkle structure can be introduced into the device through pre-strain during fabrication; serpentine traces achieve tuneable deformation and tissue-level conformability, depending on the geometry-yielding feature; fractal structures can be applied across the entire device to increase stretchability; kirigami patterns, created from 2D starting materials, achieve device deformation involving structural stretching and bending; knitted textiles, with interlocking yarn loops, can achieve high stretchability and provide breathability to epidermal bioelectronics. f, Bioelectronic–tissue adhesion can be improved to manage interface impedance instability motion artefacts. Physical interactions can be formed at the bioelectronic–tissue interface to ensure that soft bioelectronics conformally attach to dry skin surfaces. Physical interactions, including van der Waals (vdW) forces, hydrogen bonding and electrostatic interactions, result from an electron density transient shift, atoms with differences in electronegativity (H and X, X including O, F, N), and oppositely charged polar group attraction, respectively. g, To provide long-term integration and non-invasive debonding of soft bioelectronics in wet conditions, a robust bioelectronic–tissue interface can be constructed through covalent chemical bonding. h, Capillary forces can achieve conformal wrapping of soft bioelectronics with tissue surfaces in a liquid environment. i, Surface microstructures can be implemented to improve adhesion on wet and rough skin surfaces. Strong physical forces can be introduced through surface microstructures, for example, a hierarchical structure with a large contact area and a micro-sucker with negative pressure. δ+, positive charge; δ–, negative charge; θ, contact angle; EGaIn, eutectic gallium-indium; GO, graphene oxide; NHS, N-hydroxysuccinimide; NWs, nanowires; P3HT-NFs, poly(3-hexylthiophene-2,5-diyl) nanofibrils; PAA, poly(acrylic acid); PAAm, polyacrylamide; PDMS, polydimethylsiloxane; PEDOT:PSS, poly(3,4-ethy lenedioxythiophene):poly(styrenesulfonate); PET, polyethylene terephthalate; PI, polyimide; PU, polyurethane; PVA, polyvinyl alcohol; PVDF, polyvinylidene difluoride; SBS, poly(styrene-butadiene-styrene); SEBS, polystyrene-ethylenebutylene-styrene; r, radius; a, contact circle radius.
To achieve tissue-level stretchability in soft bioelectronics, polymer materials are commonly used owing to their long-chain molecular structure, intermolecular slippages between chains63 and the reversible dissipation of strain energy through dynamic bonds64–68. Thus, polymer materials can be designed with a high strain range to facilitate the seamless adaptation of bioelectronics to the dynamic surfaces of tissues. For example, elastomers, such as polydimethylsiloxane (PDMS)69, silicone70,71, polystyrene-ethylene-butylene-styrene16, polyurethane72 and PDMS–isophorone bisurea0.6–methylenebis(phenyl urea)0.4 (refs. 67,73) can serve as substrates in soft bioelectronics (Table 1 and Fig. 3b). Polymer-based hydrogels, such as polyacrylamide16 and poly(3,4-ethylenedioxythiophene):poly(styrenesulfonate)74 hydrogels, can be modified for conductivity and adhesion to function as substrates, adhesive layers, electrodes or interconnections in soft bioelectronics. Additionally, inorganic materials with certain electrical conductivity, such as graphene75, silver nanowires35 and carbon nanotubes11, can be structured into networks on elastomeric surfaces to serve as electrodes, interconnections and functional units. In addition, the physical interactions between the nanosheets of molybdenum disulfide, not relying on bonding, can provide high mechanical stretchability63. Furthermore, conductive fillers (for example, liquid metal, silver nanowires and silver flakes) can be embedded into polymer matrices to form elastic composites that exhibit both electrical conductivity and mechanical stretchability76–82.
Table 1 |.
The mechanical properties of materials in soft bioelectronics
| Material class | Examples | Mechanical properties | Functional components | Fabrication scalability | Biocompatibility | Refs. | |
|---|---|---|---|---|---|---|---|
| Stretchability (%) | Softness (kPa) | ||||||
| Plastics | PI, PET, PVA, PVDF | <2 | 1.9–3.4×106 | Substrate | High | Moderate | 91,92,101,219,220 |
| Elastomers | PDMS, SEBS, PU, silicone, PDMS–IU0.6–MPU0.4 | 170–1,000 | 60–3.8×103 | Substrate | High | Moderate | 16,67,69–73 |
| Hydrogels | PAAm, PEDOT:PSS, GO/PVA/PAA/NHS ester | 20–150 | 2.6–460 | Substrate, adhesive layer, electrode, interconnections | Moderate | High | 15,16,62,74 |
| Inorganic materials | MoS2, graphene, AgNWs, CNTs | 100–200 | NA | Electrode, interconnections, functional unit | Moderate | Poor | 11,34,63,75,221 |
| Composites | AgNWs/AuNWs/SBS, P3HT-NFs/PDMS, Ag flasks/PAAm/alginate, EGaIn/PU | 200–1,170 | 10–3.28×103 | Electrode, interconnections, functional unit | Poor | Moderate | 77,79–82 |
CNTs, carbon nanotubes; EGaIn, eutectic gallium-indium; GO, graphene oxide; IU, isophorone bisurea; MPU, methylenebis(phenyl urea); NA, not applicable; NHS, N-hydroxysuccinimide; NWs, nanowires; P3HT-NFs, poly(3-hexylthiophene-2,5-diyl) nanofibrils; PAA, poly(acrylic acid); PAAm, polyacrylamide; PDMS, polydimethylsiloxane; PEDOT:PSS, poly(3,4-ethylenedioxythiophene): poly(styrenesulfonate); PET, polyethylene terephthalate; PI, polyimide; PU, polyurethane; PVA, polyvinyl alcohol; PVDF, polyvinylidene difluoride; SBS, poly(styrene-butadiene-styrene); SEBS, polystyrene-ethylene-butylene-styrene.
Softness.
Soft tissues are typically characterized by a low Young’s modulus, reflecting their intrinsic elasticity to maintain tissue structure and support various physiological processes. For example, neural and muscle tissues have low moduli of around 1–100 kPa (refs. 83,84), whereas skin has a modulus in the range of 0.85–4.5 MPa (refs. 85,86). Therefore, to establish a compliant and durable connection, bioelectronics should have low modulus levels to ensure uniform strain distribution between soft bioelectronics and the tissue surface and prevent concentrated strains and displacement or detachment of bioelectronics under human body activities, thereby improving their stability and eliminating interface impedance instability motion artefacts. Moreover, tissue-level softness allows conformal contact with living tissues while minimizing mechanical damage of tissue, inflammation and skin irritation during soft bioelectronic integration87.
A range of low-modulus materials has been employed in the development of tissue-level soft bioelectronics. For example, hydrogels with tissue-level moduli can be applied in implantable bioelectronics intended to monitor physiological activities, such as neural, cardiac and muscular signals, to allow conformal integration with neural and muscle tissues. Wearable bioelectronics can be based on elastic materials, such as silicone (~60 kPa), PDMS (0.2–3 MPa), polystyrene-ethylene-butylene-styrene (2.83 MPa) and polyurethane (0.5–3.8 MPa), to match the moduli of the skin88,89 and enable conformal and long-term stable bioelectronic–tissue interfaces for continuous physiological monitoring.
Device structure engineering
The conformability and stretchability of soft bioelectronic systems can be greatly enhanced through device structure engineering, ensuring a dynamic conformal contact between soft bioelectronics and living tissues during human body activities. Device conformability and structural stretchability can be improved through thickness reduction, strain adjustment and deformation-tolerant design.
Thickness reduction.
Reducing the thickness of soft bioelectronics improves adhesion and thus contact at the bioelectronic–tissue interface (Fig. 3c) as the mechanical properties and energy dynamics at the bioelectronic–tissue interface determine the interfacial contact states. To investigate the impact of device thickness reduction at the bioelectronic–tissue interface, an analytical mechanical model can be applied that delineates the morphology of the interface90. The total energy related to the interfacial contact (Uinterface) can be mathematically expressed as:
| (3) |
where Udevice_bending, Uskin_elasticity and Uadhesion represent the bending energy of the device, the skin elastic energy and the adhesion energy of the bioelectronic–tissue interface, respectively. Among these, Udevice_bending is influenced by device thickness. If the thickness of the device reduces to a point at which the adhesion energy exceeds the sum of bending energy and skin elastic energy, the gap at the bioelectronic–tissue interface is eliminated, thereby enabling conformal contact and allowing the device to adapt to irregular shapes of living tissues. Thus, the flexibility of bioelectronics based on intrinsically stiff materials (for example, polyimide and polyethylene terephthalate) can be increased by designing an ultrathin membrane structure (~10 μm)91,92 to engineer and maintain a conformal and adaptable interface with dynamically evolving microscopic surface topographies, thereby reducing the impact of motion artefacts on impedance stability.
Strain adjustment.
Strain adjustment refers to the manipulation of strain distribution in soft bioelectronics through device structure engineering, aiming to achieve structural stretchability and bioelectronic–tissue interface stability. Typically, applied strain induces geometric changes and performance variations in the active regions of the device. Device structure engineering can be applied to distribute the strain experienced by active regions to certain deformable regions, resulting in a patterned strain distribution throughout the entire soft bioelectronic system. Implementing strain-adjusting structures, such as an island bridge or elastiff layer, can thus increase structural stretchability, accommodating external strains exceeding 100%.
The island bridge structure seamlessly couples highly stretchable ‘bridges’ with rigid multifunctional ‘island’ devices93–95 (Fig. 3d). In this configuration, the strain exerted on soft bioelectronics predominantly affects the thin and narrow bridge structures, while not impacting the functional rigid islands. Consequently, if the island bridge structure is subjected to tensile or compressive stress, these deformable connecting bridges are primed for mechanical responses that accommodate complex strain distributions and the active device islands ensure stable performance characterized by high bending stiffness.
Mechanical stability and stretchability can also be improved by including a patterned region of mechanical heterogeneity, referred to as the elastiff layer, to adjust strain distribution16,96. The patterned elastiff layer provides areas of local stiffness and thus strong adhesion to the substrate, enabling the system to withstand substantial shear stresses during strain-induced deformation. The strain can be adjusted to the lower-modulus interspaces between the active devices, thereby enabling soft bioelectronics to achieve conformable and enduring interfaces while maintaining long-term stability and high stretchability.
Deformation-tolerant design.
Soft bioelectronics that contain rigid and brittle materials, such as metals, silicon and plastics, can be made stretchable through patterning of deformation-tolerant device structures, including wrinkles, serpentines, fractals, kirigami structures and textiles. Wrinkles can be engineered to introduce pre-strain into the device during fabrication97–100 (Fig. 3e). Upon relaxation of the pre-strain, compressive forces generate complex and wavy patterns of relief through a non-linear buckling process, thereby providing reversible stretchability. Geometric yielding of ultrathin filamentary serpentine traces under stress allows tunability and conformability to living tissues70,77,101,102. Similarly, fractal structures, such as self-similar serpentine geometries, can be included throughout the device to accommodate high levels of stretchability and conformability103–106. The art of kirigami involves patterns created through cutting and folding, enabling 2D starting materials to undergo 3D structural deformations. This technique achieves high stretchability through structure stretching and bending107–109. Finally, by leveraging interlocking yarn loops crafted through knitting, a soft and stretchable fabric can be designed that exhibits high shape conformability and stretchability110,111. In addition, textiles can endow the device with breathability and improve the comfort of wearable bioelectronics7,8,112–114. Such deformation-tolerant designs help mitigate interface impedance instability motion artefacts.
Bioelectronic–tissue adhesion
The constant movements of living organisms result in dynamically evolving microscopic surface topographies that affect the adhesion of soft bioelectronics, which can be addressed by introducing diverse forces and bonds across the device. Soft bioelectronics can adapt and adhere to living tissues through physical interactions, capillary forces, surface microstructures and covalent chemical bonds. Improving adhesion ensures a robust connection with the tissue, thereby eliminating interface impedance instability motion artefacts.
Physical interactions.
Physical interactions, including van der Waals forces, hydrogen bonds and electrostatic interactions, can form at the device–tissue interface through non-covalent bonds115,116. The nature of this adhesion is determined by the bioelectronic–tissue interfacial energy90. If the adhesion energy exceeds the sum of bending energy and elastic energy, physical interactions drive the contact adhesion of ultrathin soft bioelectronics, ensuring their conformal attachment while mitigating impedance instability motion artefacts at the interface.
Van der Waals forces arise from the attraction and repulsion of transient shifts in electron density between molecules (Fig. 3f). Once in contact with living tissues, soft bioelectronics establish stable connections through van der Waals adhesion forces. For example, a molybdenum disulfide-assembled membrane directly adheres to soft living tissues through highly conformal interfaces via van der Waals forces, imparting electronic functions to living organisms63. In ECG monitoring, conventional Ag/AgCl electrodes typically suffer a reduction in signal-to-noise ratio from 44.3 dB to 28.5 dB owing to motion artefacts63. By establishing conformal gate transistors, the membrane maintains stable ECG signals without deviations, achieving signal-to-noise ratios of 49.8 dB and 49.2 dB before and during human motion, respectively. In addition, the adhesive layer can seamlessly integrate with living tissue surfaces through hydrogen bonds, which are primarily formed through interactions with functional groups, such as amino, carboxylic and thiol groups, present on the adhesive surface117,118. Such interactions facilitate the long-term adherence and functionality of soft bioelectronics.
Living tissue surfaces usually carry either negative or positive charges and polar groups, offering various anchoring sites for adhesive materials equipped with oppositely charged and/or polar groups119–121. The positive and negative charges induce the generation of localized electric fields, thereby driving the electronic materials with opposite charges to form a robust adhesive interface with living tissues through electrostatic interactions. The introduction of these physical interactions to the bioelectronic–tissue interface holds the potential to bridge bioelectronic–tissue gaps and bestow interface impedance stability122.
The adhesive interface, featuring multiple physical interactions, is achieved through material design in the adhesive layer. For example, hydrogels incorporating polydopamine exhibit an adhesive strength to tissues that relies on physical interactions formed between the adhesive and tissue surface23,123,124 (Fig. 3g). Physical adhesion, including hydrogen bonding between catechol groups and amino groups, as well as π–π stacking interactions with aromatic groups, can drive the stable adhesion of thin bioelectronics to tissues125–127. Similarly, polymers doped with tannic acid containing gallol groups can provide stable contact between bioelectronics and tissue through a series of physical interactions, including hydrogen bonds, electrostatic interactions and cation–π interactions128.
Capillary forces.
Physical interactions play a crucial role in facilitating the adhesion of bioelectronics under dry conditions. However, in a wet environment (for example, implantable devices on moist dynamic organs or wearable devices on perspiring skin), physical interactions are attenuated owing to the charge screening effect from highly polarized water molecules, resulting in a notable deterioration of adhesive properties129. By contrast, capillary forces, resulting from surface tension at the solid–liquid–gas interface and the Laplace pressure within the liquid130, drive the adhesive behaviour of bioelectronics in wet environments. A model to analyse capillary forces involves the formation of a liquid bridge between two parallel flat plates131–133 (Fig. 3h). In this model, the parallel flat plates and the liquid bridge represent the living tissue or device and the liquid, respectively. The capillary forces (Fcap) can be quantified as:
| (4) |
| (5) |
where a and θ represent the contact circle radius and the contact angle at the bridge–plate intersection, respectively; Δp is the local pressure change at the interface, as determined by the Young–Laplace equation; r is the principal radius of curvature; and γ is the surface tension at the liquid–gas interface. Based on this analysis, capillary forces can drive the conformal wrapping process of a thin film on moist tissue surfaces, achieving a high degree of conformality through physical and stable adhesion of the bioelectronic device to complex curvilinear surfaces134.
Surface microstructures.
Attaching wearable bioelectronics to the skin remains challenging owing to the rough morphology of skin, the presence of hair and on-surface body fluids such as sweat, oil and blood135. Inspired by adhesive phenomena observed in nature, such as gecko footpads and octopus suckers, surface microstructures have been designed for soft bioelectronics that harness strong physical forces, offering reversible and robust adhesion properties on wet and rough skin surfaces, and thereby eliminating the risk of device detachment.
For example, a micropillar structure, inspired by micro-hierarchical and nano-hierarchical structures in the toe pads of geckos, provides surface roughness for reversible adhesion136,137. In particular, micropillar structures with a high aspect ratio and wide spatula tips maximize the contact area between bioelectronics and the skin (Fig. 3i), achieving a cyclically high normal adhesive force (~1.3 N cm–2) through collective van der Waals forces. Consequently, such modified bioelectronics can adhere to human skin, even in dynamic conditions and wet environments138. Similarly, inspired by the octopus suction cup, which enables reversible adhesion in dry and wet conditions, micro-sucker structures have been engineered to facilitate stable adhesion through the generation of negative pressure by volume changes in the suction chamber combined with van der Waals forces (in dry contact) or capillary action (in wet contact) at the interface139,140. Soft bioelectronics based on a micro-sucker structure exhibit a high normal adhesive capability (~1.89 N cm–2) to the skin, enabling accurate biomonitoring without requiring external pressure141,142.
Covalent chemical bonds.
Establishing a tissue interface through physical adhesion has several limitations. Physical interactions are unstable in dynamic tissues owing to their weak strength, and typically only provide adhesion for thin devices of less than 5 μm in thickness134,143,144. Surface microstructures depend on external preloads and pull-off adhesion forces to achieve repeatable adhesion, typically around 35 kPa in the normal direction139. Alternatively, sufficient adhesion force and minimal detachment in wet conditions can be achieved by covalent chemical bonding at the bioelectronic–tissue interface145–147, also enabling on-demand and non-invasive debonding. In particular, low-modulus hydrogels can be applied to form covalent chemical bonds, involving both physical adhesion and chemical bonding. Upon contact with the wet tissue surface, the carboxylic acid groups in the hydrogel can rapidly remove interfacial water while forming physical adhesion through hydrogen bonds and electrostatic interactions with the tissue surface. Subsequently, the functional groups in contact with the tissue surface can further form covalent chemical bonds. For example, N-hydroxysuccinimide ester groups can react with primary amine groups to form amide covalent bonds14,15,148 or two thiol groups can form disulfide bonds147. These covalent bonds provide long-term stability for the integration of soft bioelectronics in wet and dynamic physiological environments.
Biopotential motion artefact management
In contrast to the unpredictable and irregular frequency variations associated with interface impedance instability motion artefacts, biopotential motion artefacts typically occur within specific frequency spectra, often distributed in higher frequency ranges. Biopotential signals from physiological activities of nervous and muscular tissues typically have an amplitude ranging from a few millivolts to several hundred microvolts149,150, including 0.1–100 Hz for ECG signals151, 1–30 Hz for EEG signals152,153 and 20–500 Hz for EMG signals154,155 (Fig. 4a).
Fig. 4 |. Biopotential and mechanical motion artefact management.

a, Frequency spectrum of various biopotential and mechanical signals. b, Electrophysiological signal sensing mechanism. c, To address motion artefacts and reduce the computational burden of post-signal processing, a signal compensation circuit can be designed for artefact signal reduction and a soft functional transistor allows filtering of mechanical motion artefacts based on vibration frequencies. d, To provide artefact signal post-processing, filtering algorithms can be used to selectively eliminate unwanted frequency components associated with artefacts. To isolate target signals from artefacts in complex patterns and varying conditions, machine learning models can be trained to extract physiological features from biosignals. e, Biochemical signal sensing mechanism. f, Biomechanical signal sensing mechanism. g, To eliminate motion artefacts from the initial sensing stage, sensors can be designed using dampers with band-pass filtering functions and strain isolators to prevent irrelevant mechanical motions from propagating into the sensing area. ΔB, magnetic flux density changes; ΔC, capacitance change; ΔP, pressure change; ΔQ, transferred charge; ΔR, resistance change; ECG, electrocardiogram; EEG, electroencephalogram; EMG, electromyogram; GOx, glucose oxidase; IDS, drain-source current; ox, oxidized analyte; red, reduced analyte; VGS, gate-source voltage.
Biopotential motion artefacts are often observed in the sensing of electrophysiological signals, such as ECG, EMG, EOG or EEG signals, with flexible electrodes17,18,51 because the bioelectronic–tissue interface, responsible for capturing the target electrophysiological signals, is susceptible to interference by biopotential motion artefacts arising from body activities (Fig. 4b), which may be managed through circuit design and algorithmic intervention.
Circuit design
Biopotential motion artefacts are inevitably collected by large-area, skin-interfaced interconnections and electrodes within the circuit but can be mitigated through tailored circuit design, for example, through signal compensation, which is the process of removing noise from target signals by noise reduction strategies. A typical signal compensation circuit contains two functional modules, the acquisition channel and the compensation channel (Fig. 4c). The acquisition channel can detect both the target signals and irrelevant biopotential signals55,156,157. The compensation channel is strategically deployed at specific locations on the human body to exclusively capture motion artefact signals. The differential between these two channels results in the reduction of artefact signals, thereby diminishing the interference of motion artefacts with target signals. This strategy can be applied to eliminate EMG biopotential artefact signals during ECG monitoring; by introducing a parallel compensation channel positioned closely to the ECG acquisition channel, similar EMG artefact signals can be detected in both channels55,156. As grip-induced EMG artefacts increase, the Pearson correlation coefficient of the compensated ECG remains at 0.93, while the Pearson correlation coefficient of the uncompensated ECG drops to 0.77 (ref. 156). Therefore, the compensation circuit design strategy cannot only diminish biopotential motion artefacts but can also reduce the computational burden of the system, thereby simplifying the processing of biosignals before their final interpretation.
Algorithmic intervention
The specific frequency spectra of biopotential motion artefacts render algorithmic intervention an effective post-processing strategy for artefact signal management158. Algorithmic intervention can be categorized into two types, defined by their fundamental approach and intended purpose: filtering algorithms and machine learning.
Filtering algorithms.
Filtering algorithms play a crucial role in electrophysiological signal processing, systematically enhancing the quality of recorded biosignals by selectively modifying or eliminating unwanted frequency components associated with artefacts159. Various transformative algorithms, including fast Fourier transform, Laplace transform and wavelet transform, are fundamental for comprehending signal properties and characteristics (such as frequency, time and complexity), thereby allowing efficient and effective filtering160–162. Different types of filtering algorithm, such as Butterworth, finite impulse response, infinite impulse response and elliptic filters, possess specific characteristics and applications for signal filtering, and filtering tasks typically include low-pass, high-pass, band-pass and band-stop filtering163 (Fig. 4d). For example, through well-designed cascade-adaptive finite impulse response filters, ECG, EOG and EMG artefacts in EEG records can be considerably attenuated without compromising the core information embedded in these EEG signals164,165.
Machine learning.
Machine learning represents a more dynamic and data-driven approach in biosignal post-processing166–168. Instead of relying on predefined rules, machine learning models are trained on labelled or unlabelled data sets to learn patterns and relationships within data169. These models use the acquired knowledge to make predictions, classify signals or identify anomalies associated with artefacts. Independent component analysis is a commonly used machine learning method capable of isolating electrophysiological signals, such as EEG, EMG and ECG, even when mixed with biopotential motion artefacts, into their intrinsic and independent source components18. In addition, neural networks have been explored for motion artefact management; in particular, deep learning-based neural networks are effective in dealing with artefacts that exhibit complex and non-linear patterns (such as some biopotential motion artefacts from EMG and ECG), enabling the algorithm to adapt and generalize across varying conditions17,170,171. Such approaches improve biosignal quality by learning and extracting relevant features from data, ultimately leading to more precise post-processed biosignals.
Mechanical motion artefact management
Mechanical motion artefacts typically exhibit frequency spectra below 10 Hz and can impact diverse sensing mechanisms; however, they can be addressed by systematic management strategies, including sensor design, functional circuit design and algorithmic intervention.
Artefact impact analysis
Mechanical motion artefacts, stemming from biomechanical activities such as body movements (for example, walking, running, jumping and speaking) and physiological functions (for example, breathing, heartbeat and eye blinks), typically occur below 10 Hz (Fig. 4a) and can influence the measurement performance of soft bioelectronics. The degree and nature of their impact depends on the sensing mechanism, including biochemical, biomechanical and other soft sensors.
Biochemical sensors.
Biochemical sensors, including enzymatic or ion-selective, voltammetric, and bioaffinity sensors, can be integrated into soft bioelectronics to enable continuous monitoring of human health at the biomolecular level (Fig. 4e). However, their operational stability is susceptible to mechanical motions172–176; for example, mechanical motion-induced device deformation may alter the geometry of the flexible microchannel and affect the stability of microfluidic flow. Furthermore, current signal acquisition techniques (potentiometry, amperometry, voltammetry and impedance spectroscopy) are affected by the dynamic state of biofluids that contain the biomarkers, which may cause signal drift and performance deterioration during body activities3,176.
Biomechanical sensors.
Biomechanical sensors can monitor respiratory activity, heart rate, pulse wave and muscle contraction by establishing a conformal bioelectronic–tissue interface with living tissues7 (Fig. 4f). In particular, self-powered sensors leveraging triboelectric, piezoelectric and magnetoelastic effects177,178 can convert mechanical energy into electric current or voltage signals, for example, for self-powered electrotherapeutics and energy harvesting from human body motions7,112,114. Nonetheless, owing to their self-powered sensing mechanisms, these sensors are particularly vulnerable to extraneous mechanical movements, which can introduce electric current or voltage signals that overlap with the target biomechanical signals, resulting in the occurrence of mechanical motion artefacts. For example, mechanical sensors designed to monitor artery pulse waves on the wrist54,179,180 are affected by hand movements; in arteries, the variation in vessel length during a pressure pulse is negligible whereas the circumferential strain varies significantly, ranging from 2% to 18%181. However, the skin proximal to the wrist joint undergoes 30–40% strain when the joint is moved from full extension to full flexion182. Therefore, disturbances from hand movements can lead to skin deformation at the location of the sensor, thereby introducing mechanical motion artefacts that overlap with the actual pulse waves. This interference can degrade signal quality or render it unusable.
Mechanical motion artefacts are also frequently observed in electro physiological, temperature and optical sensors. Deformations from biomechanical activities are inevitably transmitted to soft bioelectronics, and these strain-induced disturbances along the interconnection and sensing areas lead to baseline drift and abnormal spikes in the signal. However, the artefact signals related to specific body activities exhibit consistent frequency features, including frequency spectra and propagation directions that are distinct from target signals.
Sensor design
Eliminating motion artefacts at the initial sensing stage through specific sensor design strategies may reduce post-signal processing tasks.
Material-enabled damper.
A material-enabled damper is capable of absorbing specific mechanical vibrations within predefined frequency ranges183 using a band-pass-filtering gelatin–chitosan hydrogel inspired by the viscoelastic cuticular pad of spiders. The hydrogel can eliminate the transmission of mechanical artefact signals arising from biomedical activities such as gait motion, heartbeat and respiration, which occur within a frequency range below 30 Hz. Its frequency-dependent phase transition, shifting from a rubbery state for the attenuation of low-frequency noise to a glassy state for the transmission of high-frequency signals (Fig. 4g), allows it to function as a versatile filter, facilitating the acquisition of mechanical motion artefact-free signals, including electrophysiological signals (such as ECG and EEG) and high-frequency biomechanical signals (such as acoustic signals), thereby minimizing post-signal processing.
Strain isolator.
The effects of mechanical motion artefacts can also be mitigated by combining a strain isolator with the sensors to isolate interfering motions. Skin deformations, caused by body movement, can propagate as mechanical signals to the sensing area, causing signal artefacts. The integration of a strain isolator into soft bioelectronics enables the maintenance of conformal contact with the skin while safeguarding against excessive skin strain and vibration during body motion21,184,185. In this way, irrelevant mechanical motions can be isolated from the sensor, ensuring the accurate measurement of target signals (Fig. 4g). This approach can be applied for physiological monitoring, such as ECG, heart rate and respiratory rate monitoring during a range of human activities21,184 as well as for wearable biochemical and biomechanical sensors3,54,186.
In biochemical sensing devices, increasing material stretchability provides a partial solution to mitigate motion artefacts; however, long-term stability in wet environments remains a challenge. A strain isolation strategy can be applied to design motion artefact-free biochemical sensors. For example, a biomarker signal can be vertically transmitted from a deformed region to a different layered strain-isolated region through an interconnection to prevent the sensing of motion artefacts resulting from biomechanical activities186. Similarly, sensor units can be constructed within unaffected rigid islands, while the strain is adjusted by stretchable bridges187.
Circuit design
The circuits of soft bioelectronics enable the propagation and preprocessing of signals. By incorporating signal compensation circuits and soft functional transistors in the circuits, mechanical motion artefacts can be diminished, reducing the computational burden of the monitoring system.
Signal compensation circuit.
Similar to the compensation strategy for biopotential motion artefacts, a signal compensation circuit for mechanical motion artefacts reduces artefact signals through the differential between the compensation channel and the acquisition channel. Mechanical motion artefacts generated in local areas of the human body during physical activities have almost identical characteristics. To minimize artefact signals, the compensation channel should be located at a position away from the target signal sensing area but where it can capture the same mechanical motion artefacts. In this configuration, mechanical motion artefacts can be separately reduced through the differential process between the acquisition channel and the compensation channel55. For example, during cardiopulmonary monitoring by wearable electronics with accelerometers, the acquisition channel and compensation channel are located at the suprasternal notch and sternal manubrium, respectively. These dual-sensing devices allow measurements of diverse physiological parameters, including heart rate, sounds, respiratory activities and body temperature, while maintaining insensitivity to body motions during routine daily activities55. During movements (such as walking, running and jumping), the ratio of the mean difference to the standard deviation of heart rate (beats per minute (bpm)), extracted using a compensation circuit, improves from 2.23/13.92 bpm (normal) to 0.01/2.71 bpm (differential), providing accurate monitoring during daily body activities55.
Soft functional transistor.
Soft transistors are central components in soft bioelectronics, serving as switches or amplifiers82,91,188,189. Unlike transistors made from rigid materials (such as metals and silicon semiconductors), soft transistors are typically constructed from intrinsically stretchable materials190, including polymers, carbon-based compounds or other organic molecules11,191–193. Soft transistors, such as organic field-effect transistors, can be incorporated into soft bioelectronics to amplify weak biopotential signals and conduct analogue filtering of irrelevant biomechanical signals. Combined with other functional components, such as capacitors and resistors96,194 (Fig. 4c), these transistors undertake amplifier and filter functions that are vital in boosting signal strength while preserving its inherent characteristics, thereby suppressing signal artefacts at specific frequencies188. For example, a mechanically flexible amplifier made of organic thin-film transistors and thin-film capacitors can be attached to human skin, amplifying differential input signals while suppressing common-mode noise and mechanical motion artefacts during EEG or ECG monitoring195. Soft functional transistors also play a key role in the analogue-to-digital conversion of bioelectronics, which refers to the encoding of sensory information at the front end of the sensor. By converting artefact-sensitive analogue signals into discrete, uninfluenced digital signals, the resilience of soft bioelectronics against interference from mechanical motion artefacts can be further enhanced17.
Algorithmic intervention
The distinct characteristics between mechanical motion artefacts and target signals render algorithmic intervention an effective post-processing strategy for artefact signal management179,196. Algorithmic intervention can be categorized into two types: filtering algorithms and machine learning (Fig. 4d).
Filtering algorithms.
In contrast to analogue filtering, which requires physical hardware to conduct signal filtering tasks, digital filters, based on various filtering algorithms, have higher adaptability and tunability when dealing with complex and variable mechanical motion artefacts. Assisted by transformative algorithms, the cutoff frequency range can be chosen to achieve precise artefact signal mitigation. For example, in electrophysiological signal monitoring, mechanical motion artefacts, resulting from body movements and physiological functions with frequencies below 30 Hz, can be addressed using Butterworth-based band-pass or high-pass filters127. These filters, equipped with cutoff frequencies that cover the target signals, block undesired mechanical motion artefacts196. Biomechanical signal monitoring can respond to full-body kinematics but is also susceptible to interference from mechanical motion artefacts, which disrupt the target information. A high-pass filter (>10 Hz) can selectively capture biomechanical signals across high-frequency ranges, including acoustic vibrations and cardiac activity, while minimizing interference from respiration and locomotion at low frequencies197,198. For example, based on the filtered signals, the data classification accuracy reaches 0.9 ± 0.08 for coughing and 0.88 ± 0.1 for speaking with a convolutional neural network model196.
Machine learning.
Principal component analysis (PCA), which is an unsupervised machine learning algorithm, also enables mechanical motion artefact management. Through a dimensionality reduction calculation, the target signals and artefact signals can be differentiated into distinct principal components. For example, during photoplethysmography for the continuous monitoring of metrics related to vascular resistance, PCA can separate target signals for pulsatile capillary blood volume changes and mechanical motion artefacts into the first and second principal components, respectively199,200. Employing PCA for feature extraction from blood pressure following pulse reconstruction improves the mean blood pressure from a mean difference and standard deviation of −2.26 and 13.79 mm Hg (in the presence of motion) to −0.19 and 5.65 mm Hg (after artefact elimination), thereby enhancing signal estimation quality200.
More complex data characterized by non-linear and high-dimensional features may be better addressed by neural networks. For example, wearable pressure sensors can achieve continuous and precise blood pressure measurements with the assistance of a neural network trained against mechanical motion artefacts54,179. In addition, convolutional neural networks have shown high diagnostic performance for shockable and non-shockable arrhythmia classifications with robust tolerance against mechanical motion artefacts from muscular activities201. However, if target signals and mechanical motion artefact signals share similar frequency bandwidth, algorithmic intervention strategies may be less effective. Under these circumstances, sensor and circuit design approaches, such as compensation circuits, strain isolators and material-enabled dampers, should be considered.
In addition to biomechanical signals, various other physical parameters, including temperature, vessel volume and tissue impedance, can be detected for physiological activity monitoring. Sensing mechanisms for these physical parameters typically exhibit lower sensitivity to motion artefacts and could thus be leveraged to achieve motion artefact-free detection. For example, physiological monitoring based on thermal signal transmission could substantially reduce the sensitivity to motion artefacts, providing quantitative monitoring of blood flow velocity and direction within the microvascular system202,203. Additionally, the interaction between light and living tissue could be harnessed to develop photoplethysmography sensors for the monitoring of cardiac output and pulse waveform features199. Moreover, bioimpedance measurements could exploit the penetration of high-frequency current into tissues for haemodynamic parameter monitoring55. These deep tissue monitoring platforms, which harness the interaction between various physical signals and living tissues, offer promising avenues for the advancement of physiological monitoring while mitigating the impact of motion artefacts.
Outlook
Motion artefacts pose a major challenge to the advancement of soft bioelectronics for point-of-care applications but can be managed by the aforementioned strategies (Table 2). Other approaches related to the signal detection stage, including intelligent materials, functional circuits and customized service, as well as to the signal processing stage, including distributed information storage and computing, may further mitigate motion artefacts to improve the accuracy and reliability of bioelectric signal measurements (Box 1).
Table 2 |.
Quantitative analysis of management strategies for motion artefacts
| Artefact category | Strategy | Signal category | Frequency (Hz) | Performance | Ref. |
|---|---|---|---|---|---|
| Interface impedance instability motion artefacts | Ultrathin, physical interactions and mechanical matching | Electrophysiology: ECG/EEG | 0–3 | SNR: with strategy, 49.2 dB; without strategy, 28.5 dB | 63 |
| Physical interactions | Electrophysiology: ECG | 0–3 | SNR: with strategy 19.6 ± 1.8 dB; without strategy, 102.7 ± 9.4 dB | 127 | |
| Ultrathin, physical interactions and manufacturing | Electrophysiology: ECG | 0–3 | SNR: with/without strategy approximately the same, ~45 dB | 213 | |
| Ultrathin, capillary forces and mechanical matching | Electrophysiology: neural mapping | 0–50 | Root mean square amplitude ratio: with strategy, 3.6 ± 1.8; without strategy, 5.2 ± 3.9 | 134 | |
| Surface microstructures | Electrophysiology: ECG | 0–3 | Signal amplitude: with strategy in wet condition, 74.4%; without strategy, 0% (detach) | 138 | |
| Mechanical motion artefacts | Machine learning regression algorithm | Arterial blood pressure: Bio-Z waveform | 0–3 | With strategy: error: 0.06 ± 2.5 mm Hg (diastolic blood pressure), 0.2 ± 3.6 mm Hg (systolic blood pressure) | 56 |
| Supervised feedforward neural network architecture | Pulse waveform monitoring | 0–3 | SNR: with strategy, 23.3 dB | 179 | |
| PCA | Cardiac monitoring: PPG | 0.5–5 | Precision–recall curve, with strategy: baseline 0.82, exercise 0.98, cold pressor/breath hold 0.73 | 199 | |
| PCA | Cardiac monitoring: PPG | 0.5–3 | MD/SD: without strategy, −2.26/13.79 mmHg; with strategy, −0.19/5.65 mmHg | 200 | |
| Digital filtering and CNN | Mechanoacoustic: cardiac and respiration signs | 20–55, 0.1–1 | Accuracy with strategy: 0.9 ± 0.08 for coughing; 0.88 ± 0.1 for speaking | 198 | |
| Compensation circuit design | Cardiopulmonary monitoring: heart rate, acceleration | 1–10 | MD/SD: without strategy, 2.23/13.92 bpm; with strategy, 0.01/2.71 bpm | 55 | |
| Damper materials | Electrophysiological and voice: acoustic signal | 80–200 | SNR: without strategy, <20 dB; with strategy, 60–120 dB | 183 | |
| Biopotential motion artefacts | Independent component analysis | Electrophysiology: EEG/EOG | 0.5–10 | SNR: without strategy, −15 to 0 dB; with strategy, 6.65–7.55 dB | 18 |
| Non-linear filtering | Electrophysiology: EMG/ECG | 50–150, 0–3 | EMG relative error, with strategy: −0.20% | 17 | |
| 1D-CNN | Electrophysiology: EEG/EOG, ECG, and EMG | 0–256 | SNR: without strategy, −5 dB; with strategy: 19.8, 19.3, 20.6 dB | 170 | |
| Compensation circuit design | Electrophysiology: ECG | 5–50 | PCC: without strategy, 0.77; with strategy, 0.93 | 156 | |
| Low-pass filter algorithm | Electrophysiology: EEG | 0.1–35 | SNR: without strategy, 10 dB; with strategy, 22.05 dB | 164 |
bpm, beats per minute; CNN, convolutional neural network; ECG, electrocardiogram; EEG, electroencephalogram; EMG, electromyogram; EOG, electrooculogram; MD/SD, ratio of mean difference to the standard deviation; PCA, principal component analysis; PCC, Pearson correlation coefficient; PPG, photoplethysmography; SNR, signal-to-noise ratio
Box 1. Translational considerations.
Scale-up manufacturing of soft bioelectronics
Manufacturing scalability and device reliability are paramount to transferring soft bioelectronics from laboratory prototypes to industrial production10,208. Efficient workflows, cost reduction through material and process optimization, and meeting market demands are achievable through scaled-up manufacturing222,223. In addition, device reliability is fundamental for integrating soft bioelectronics into clinical monitoring83,224,225. Improving reliability across material selection, manufacturing processes and performance validation ensures extended lifespan and consistent performance. Furthermore, soft bioelectronics may be integrated into existing wearable systems; for example, electronic functionalities may be embedded in common textiles to create smart textiles with biomonitoring capabilities8,114.
Customized functions for personalized needs
Customizing the functionality of soft bioelectronics based on specific needs may contribute to personalized and precise health-care monitoring226–228. The development of personalized soft bioelectronics can be fostered through collaboration with patient communities and health-care experts, for example, to design interfacing strategies based on auditory, tactile or visual principles227–231.
Establishing clinical gold standards for evaluation
Establishing clinical gold standards for the evaluation of soft bioelectronics requires the continuous monitoring of conventional clinical indicators to create a dynamic physiological data base232–234. These data can then be analysed in terms of feature extraction, pattern recognition and machine learning to extract physiological insights235. Calibration on established clinical assessments ensures data alignment and accuracy. Defining and validating measurement standards for various physiological indicators is fundamental to ensuring stable and repeatable performance.
Identification and prediction of clinical needs
Soft bioelectronics should be developed to address key clinical needs, for example, for newborns or critically ill patients236,237, who require close monitoring of vital signs. Similarly, acute disease prevention has driven the development of continuous and non-invasive monitoring devices. However, prioritizing encryption and security is crucial for medical monitoring, particularly in the context of cloud services and big data, to ensure patient privacy.
Intelligent materials
Current sensor units in soft bioelectronics typically lack the capacity to selectively detect target biosignals while filtering out irrelevant motion artefact signals. Intelligent materials with selective response capabilities or motion-insensitive features may allow the accurate extraction of intended signals within precise frequency ranges and propagation directions. For example, selective dampers could be integrated that capture vibrations within designated frequency ranges to inhibit the propagation of signal artefacts within the sensor183. Materials with anisotropic responses may further enable the regulation of sensitivity in relation to biological information across different propagation directions. Thus, the sensor could maintain high sensitivity in the direction of the target physiological signals whilst not detecting disturbances from artefact signal directions.
Motion-insensitive materials could be integrated into sensors to selectively detect target signals at the front end. For example, by introducing conjugated polymer nanostructures into elastomer-based semiconducting materials, the mobility of soft transistors is maintained stable during stretching193. Similarly, composite nanofibers of carbon nanotubes and graphene are insensitive to bending in pressure measurements204. These materials can thus isolate interfering mechanical motion artefacts from target biosignals at the sensing stage. Furthermore, the electrical and mechanical properties of polymer materials could be improved for bioelectronic–tissue interfaces. For example, during dynamic motion, fatigue from polymer deformation may lead to detection errors205. Tough and self-healing hydrogel–elastomer materials, featuring multi-strength hydrogen bonding interactions, in which strongly crosslinked hydrogen bonds provide rigidity and elasticity and weakly crosslinked hydrogen bonds facilitate strain energy dissipation through reversible breaking and reformation23,73,206, may enable a stable interface with tissue. In addition, modulus and electrical property matching between polymer materials and electrodes may be achieved using soft liquid metal composites and adhesive hydrogels to provide interface stability through chemical bonding between polymer layers and electrodes, thereby improving signal recording performance23. Moreover, adhesive and conductive hydrogels may simultaneously act as conformable adhesive and biosignal recording electrodes124,128 in soft bioelectronics.
Functional circuits
Motion artefacts may also be addressed and signal detection optimized by using functional circuit components with tissue-level softness and compensation circuits, reducing the need for post-signal processing. Soft circuit components ensure a uniform distribution of stress within the bioelectronic system, establishing a stable and consistent connection with the dynamic surfaces of living tissues. Moreover, functional circuit components, such as soft transistors, capacitors and resistors, equipped with built-in amplification and signal filtering capabilities, can be integrated into the circuit design188–192,207,208 to endow the system with artefact signal filtering capabilities and reduce post-processing of data. Soft functional circuits can also encode detected biosignals from analogue to digital, offering anti-interference capabilities against motion artefacts.
Integrating a compensation channel into the monitoring system allows the distinct detection of biopotential or mechanical signal artefacts. The detected artefacts can then be employed to filter out motion artefact signals in the acquisition channel by a straightforward differential process between the two channels. This approach is particularly effective in eliminating motion artefacts from electrophysiological (EMG, ECG, EOG) and biomechanical (breathing, eye blinking, heartbeat) signals, given their specific frequencies and regular waveforms. Similarly, closed-loop circuit design can provide real-time feedback and adjustments209,210; owing to their energy efficiency and customization potential, closed-loop designs are particularly suitable for individualized health monitoring and fitness tracking, ensuring both the accuracy and reliability of soft bioelectronics.
The clinical translation of soft bioelectronics will further benefit from miniaturized, implantable designs, which require amplifiers with low power consumption and minimal noise, such as complementary metal oxide semiconductor-based amplifiers211,212. These amplifiers show beneficial power-to-noise tradeoff in biosignal amplifying, which is a key factor in miniaturizing implantable bioelectronic devices and may also improve the signal processing capabilities of soft bioelectronics.
Customization
The local dynamic surface morphology of biological tissues varies among individuals, which is not accounted for in standardized soft bioelectronic devices and may cause motion artefacts. Therefore, individual surface morphology characteristics may be collected for the customized optimization of soft bioelectronics. Furthermore, different parts of the body exhibit distinct surface features and varying physiological signals. Such differences should be considered in the design of soft bioelectronics213–218.
Distributed information storage and computing
In addition to hardware optimization, information storage, processing and interpretation can be improved in soft bioelectronics. Distributed information storage and computing may be applied to manage physiological data for personalized health care, integrating a network of nodes. Physiological data of individuals could be transmitted to a dedicated cloud server that functions as a distributed information storage system, overcoming the constraint of limited information storage capacity in soft bioelectronics. These data could then be analysed using distributed information computing to match with data bases, thereby improving the accuracy of data classification and allowing the extraction of physiological parameters from biosignals. The data could also be shared with health-care professionals for medical guidance and advice.
Citation diversity statement
We acknowledge that papers authored by scholars from historically excluded groups are systematically under-cited. Here, we have made every attempt to reference relevant papers in a manner that is equitable in terms of racial, ethnic, gender and geographical representation.
Key points.
Soft bioelectronic systems for the monitoring of human health rely on a stable and conformal bioelectronic–tissue interface.
Motion artefacts can occur due to body movements and physiological activities, and can negatively affect signal detection and interpretation in bioelectronic measurements.
Motion artefact management is crucial to the clinical translation of soft bioelectronics to ensure measurement with high accuracy.
Materials usage, device design, bioelectronic–tissue adhesion, sensor and circuit designs, and algorithmic intervention are effective motion artefact management strategies.
Acknowledgements
The authors acknowledge the Henry Samueli School of Engineering & Applied Science and the Department of Bioengineering at the University of California, Los Angeles, for their startup support. J.C. acknowledges the Vernroy Makoto Watanabe Excellence in Research Award at the UCLA Samueli School of Engineering, the Office of Naval Research Young Investigator Award (award ID: N00014-24-1-2065), NIH grant (award ID: R01 CA287326), the American Heart Association Innovative Project Award (award ID: 23IPA1054908), the American Heart Association Transformational Project Award (award ID: 23TPA1141360), the American Heart Association’s Second Century Early Faculty Independence Award (award ID: 23SCEFIA1157587), the Brain & Behavior Research Foundation Young Investigator Grant (grant number: 30944), and the NIH National Center for Advancing Translational Science UCLA CTSI (grant number: KL2TR001882).
Footnotes
Competing interests
The authors declare no competing interests.
References
- 1.Williams GJ et al. Wearable technology and the cardiovascular system: the future of patient assessment. Lancet Digit. Health 5, e467–e476 (2023). [DOI] [PubMed] [Google Scholar]
- 2.Sunwoo S-H et al. Soft bioelectronics for the management of cardiovascular diseases. Nat. Rev. Bioeng 2, 8–24 (2024). [Google Scholar]
- 3.Sempionatto JR, Lasalde-Ramírez JA, Mahato K, Wang J & Gao W Wearable chemical sensors for biomarker discovery in the omics era. Nat. Rev. Chem 6, 899–915 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]; This article summarizes biochemical sensor platforms that can be integrated with human skin for wearable health-care monitoring.
- 4.Cho KW et al. Soft bioelectronics based on nanomaterials. Chem. Rev 122, 5068–5143 (2022). [DOI] [PubMed] [Google Scholar]
- 5.Someya T, Bao Z & Malliaras GG The rise of plastic bioelectronics. Nature 540, 379–385 (2016). [DOI] [PubMed] [Google Scholar]
- 6.Haque A, Milstein A & Fei-Fei L Illuminating the dark spaces of healthcare with ambient intelligence. Nature 585, 193–202 (2020). [DOI] [PubMed] [Google Scholar]
- 7.Chen G et al. Electronic textiles for wearable point-of-care systems. Chem. Rev 122, 3259–3291 (2022). [DOI] [PubMed] [Google Scholar]
- 8.Libanori A, Chen G, Zhao X, Zhou Y & Chen J Smart textiles for personalized healthcare. Nat. Electron 5, 142–156 (2022). [Google Scholar]
- 9.Meng K et al. A wireless textile-based sensor system for self-powered personalized health care. Matter 2, 896–907 (2020). [Google Scholar]
- 10.Luo Y et al. Technology roadmap for flexible sensors. ACS Nano 17, 5211–5295 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Wang S et al. Skin electronics from scalable fabrication of an intrinsically stretchable transistor array. Nature 555, 83–88 (2018). [DOI] [PubMed] [Google Scholar]
- 12.Gao Q et al. Highly stretchable sensors for wearable biomedical applications. J. Mater. Sci 54, 5187–5223 (2019). [Google Scholar]
- 13.Wang C, Yokota T & Someya T Natural biopolymer-based biocompatible conductors for stretchable bioelectronics. Chem. Rev 121, 2109–2146 (2021). [DOI] [PubMed] [Google Scholar]
- 14.Yang Q et al. Photocurable bioresorbable adhesives as functional interfaces between flexible bioelectronic devices and soft biological tissues. Nat. Mater 20, 1559–1570 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Deng J et al. Electrical bioadhesive interface for bioelectronics. Nat. Mater 20, 229–236 (2021). [DOI] [PubMed] [Google Scholar]
- 16.Li Y et al. Achieving tissue-level softness on stretchable electronics through a generalizable soft interlayer design. Nat. Commun 14, 4488 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Liang H, Lin Z & Yin F Removal of ECG contamination from diaphragmatic EMG by nonlinear filtering. Nonlinear Anal. Theory Methods Appl 63, 745–753 (2005). [Google Scholar]
- 18.Wang G, Teng C, Li K, Zhang Z & Yan X The removal of EOG artifacts from EEG signals using independent component analysis and multivariate empirical mode decomposition. IEEE J. Biomed. Health Inform 20, 1301–1308 (2016). [DOI] [PubMed] [Google Scholar]
- 19.Lai CQ et al. Artifacts and noise removal for electroencephalogram (EEG): a literature review. In 2018 IEEE Symposium on Computer Applications & Industrial Electronics (ISCAIE) 326–332 (IEEE, 2018). [Google Scholar]
- 20.Webster JG Reducing motion artifacts and interference in biopotential recording. IEEE Trans. Biomed. Eng 31, 823–826 (1984). [DOI] [PubMed] [Google Scholar]
- 21.Rodeheaver N et al. Strain-isolating materials and interfacial physics for soft wearable bioelectronics and wireless, motion artifact-controlled health monitoring. Adv. Funct. Mater 31, 2104070 (2021). [Google Scholar]
- 22.Nan K et al. Mucosa-interfacing electronics. Nat. Rev. Mater 7, 908–925 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Choi H et al. Adhesive bioelectronics for sutureless epicardial interfacing. Nat. Electron 6, 779–789 (2023). [Google Scholar]
- 24.Boehler C, Carli S, Fadiga L, Stieglitz T & Asplund M Tutorial: guidelines for standardized performance tests for electrodes intended for neural interfaces and bioelectronics. Nat. Protoc 15, 3557–3578 (2020). [DOI] [PubMed] [Google Scholar]
- 25.Jiang Y & Tian B Inorganic semiconductor biointerfaces. Nat. Rev. Mater 3, 473–490 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]; This article discusses fundamental semiconductor physics and operation principles with a focus on biointerface establishment under physiological conditions.
- 26.Driscoll N et al. MXene-infused bioelectronic interfaces for multiscale electrophysiology and stimulation. Sci. Transl. Med 13, eabf8629 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Strakosas X et al. Metabolite-induced in vivo fabrication of substrate-free organic bioelectronics. Science 379, 795–802 (2023). [DOI] [PubMed] [Google Scholar]
- 28.Kim J, Ghaffari R & Kim D-H The quest for miniaturized soft bioelectronic devices. Nat. Biomed. Eng 1, 0049 (2017). [Google Scholar]
- 29.Xiao X, Zhou Y, Yin J, Zhao X & Chen J In-ear electrophysichochemical sensing. Nat. Biomed. Eng 7, 1207–1209 (2023). [DOI] [PubMed] [Google Scholar]
- 30.Tang X, Shen H, Zhao S, Li N & Liu J Flexible brain-computer interfaces. Nat. Electron 6, 109–118 (2023). [Google Scholar]
- 31.Lee S et al. Ultrasoft electronics to monitor dynamically pulsing cardiomyocytes. Nat. Nanotechnol 14, 156–160 (2019). [DOI] [PubMed] [Google Scholar]
- 32.Ates HC et al. End-to-end design of wearable sensors. Nat. Rev. Mater 7, 887–907 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Liu Z et al. High-adhesion stretchable electrodes based on nanopile interlocking. Adv. Mater 29, 2 (2017). [Google Scholar]
- 34.Qi D et al. Highly stretchable, compliant, polymeric microelectrode arrays for in vivo electrophysiological interfacing. Adv. Mater 29, 40 (2017). [Google Scholar]
- 35.Zhao Y et al. Soft strain-insensitive bioelectronics featuring brittle materials. Science 378, 1222–1227 (2022). [DOI] [PubMed] [Google Scholar]
- 36.Gablech I & Głowacki ED State-of-the-art electronic materials for thin films in bioelectronics. Adv. Electron. Mater 9, 2300258 (2023). [Google Scholar]
- 37.He H et al. Hybrid assembly of polymeric nanofiber network for robust and electronically conductive hydrogels. Nat. Commun 14, 759 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Wang Y et al. Skin bioelectronics towards long-term, continuous health monitoring. Chem. Soc. Rev 51, 3759–3793 (2022). [DOI] [PubMed] [Google Scholar]
- 39.Yuk H, Wu J & Zhao X Hydrogel interfaces for merging humans and machines. Nat. Rev. Mater 7, 935–952 (2022). [Google Scholar]
- 40.Yuk H, Lu B & Zhao X Hydrogel bioelectronics. Chem. Soc. Rev 48, 1642–1667 (2018). [Google Scholar]
- 41.Alizadeh-Meghrazi M et al. Evaluation of dry textile electrodes for long-term electrocardiographic monitoring. Biomed. Eng. Online 20, 68 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Buxi D et al. Correlation between electrode-tissue impedance and motion artifact in biopotential recordings. IEEE Sens. J 12, 3373–3383 (2012). [Google Scholar]
- 43.Murphy BB et al. Time evolution of the skin–electrode interface impedance under different skin treatments. Sensors 21, 5210 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Neuman M In: Medical Instrumentation: Application and Design 4th edn (ed Webster JG) 189–240 (Wiley, 1998). [Google Scholar]
- 45.Reilly RB & Lee TC Electrograms (ECG, EEG, EMG, EOG). Technol. Health Care 18, 443–458 (2010). [DOI] [PubMed] [Google Scholar]
- 46.Teplan M Fundamental of EEG measurement. Meas. Sci. Rev 2, 1–11 (2002). [Google Scholar]
- 47.Elul R The genesis of the EEG. Int. Rev. Neurobiol 15, 227–272 (1972). [Google Scholar]
- 48.Woestenburg JC, Verbaten MN & Slangen JL The removal of the eye-movement artifact from the EEG by regression analysis in the frequency domain. Biol. Psychol 16, 127–147 (1983). [DOI] [PubMed] [Google Scholar]
- 49.Kher R Signal processing techniques for removing noise from ECG signals. J. Biomed. Eng. Res 3, 212573348 (2019). [Google Scholar]
- 50.Mithun P, Pandey PC, Sebastian T, Mishra P & Pandey VK A wavelet based technique for suppression of EMG noise and motion artifact in ambulatory ECG. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc 2011, 7087–7090 (2011). [DOI] [PubMed] [Google Scholar]
- 51.Marouf M, Saranovac L & Vukomanovic G Algorithm for EMG noise level approximation in ECG signals. Biomed. Signal Process. Control 34, 158–165 (2017). [Google Scholar]
- 52.Meng K et al. Wearable pressure sensors for pulse wave monitoring. Adv. Mater 34, 2109357 (2022). [Google Scholar]
- 53.Su Y et al. Self-powered respiration monitoring enabled by a triboelectric nanogenerator. Adv. Mater 33, 2101262 (2021). [Google Scholar]
- 54.Zhang S et al. Leveraging triboelectric nanogenerators for bioengineering. Matter 4, 845–887 (2021). [Google Scholar]
- 55.Jeong H et al. Differential cardiopulmonary monitoring system for artifact-canceled physiological tracking of athletes, workers, and COVID-19 patients. Sci. Adv 7, eabg3092 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]; This article reports a circuit compensation design to accurately measure various physiological and movement indicators while mitigating motion artefacts during human activities.
- 56.Kireev D et al. Continuous cuffless monitoring of arterial blood pressure via graphene bioimpedance tattoos. Nat. Nanotechnol 17, 864–870 (2022). [DOI] [PubMed] [Google Scholar]
- 57.Sel K, Osman D & Jafari R Non-invasive cardiac and respiratory activity assessment from various human body locations using bioimpedance. IEEE Open J. Eng. Med. Biol 2, 210–217 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Kang J et al. Tough-interface-enabled stretchable electronics using non-stretchable polymer semiconductors and conductors. Nat. Nanotechnol 17, 1265–1271 (2022). [DOI] [PubMed] [Google Scholar]
- 59.Minev IR et al. Electronic dura mater for long-term multimodal neural interfaces. Science 347, 159–163 (2015). [DOI] [PubMed] [Google Scholar]
- 60.Afanasenkau D et al. Rapid prototyping of soft bioelectronic implants for use as neuromuscular interfaces. Nat. Biomed. Eng 4, 1010–1022 (2020). [DOI] [PubMed] [Google Scholar]
- 61.Bootsma K et al. 3D printing of an interpenetrating network hydrogel material with tunable viscoelastic properties. J. Mech. Behav. Biomed. Mater 70, 84–94 (2017). [DOI] [PubMed] [Google Scholar]
- 62.Li G et al. Highly conducting and stretchable double-network hydrogel for soft bioelectronics. Adv. Mater 34, e2200261 (2022). [DOI] [PubMed] [Google Scholar]
- 63.Yan Z et al. Highly stretchable van der Waals thin films for adaptable and breathable electronic membranes. Science 375, 852–859 (2022). [DOI] [PubMed] [Google Scholar]; This article describes an adaptable and breathable electronic membrane with consistent bioelectronic–tissue adhesion for interface impedance instability motion artefact management.
- 64.Song K-I et al. Adaptive self-healing electronic epineurium for chronic bidirectional neural interfaces. Nat. Commun 11, 4195 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Seo H et al. Durable and fatigue-resistant soft peripheral neuroprosthetics for in vivo bidirectional signaling. Adv. Mater 33, e2007346 (2021). [DOI] [PubMed] [Google Scholar]
- 66.Liu Y et al. Morphing electronics enable neuromodulation in growing tissue. Nat. Biotechnol 38, 1031–1036 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Tringides CM et al. Viscoelastic surface electrode arrays to interface with viscoelastic tissues. Nat. Nanotechnol 16, 1019–1029 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Koo JH, Song J-K, Kim D-H & Son D Soft implantable bioelectronics. ACS Mater. Lett 3, 1528–1540 (2021). [Google Scholar]
- 69.Yan Z et al. Thermal release transfer printing for stretchable conformal bioelectronics. Adv. Sci 4, 1700251 (2017). [Google Scholar]
- 70.Xu L et al. 3D multifunctional integumentary membranes for spatiotemporal cardiac measurements and stimulation across the entire epicardium. Nat. Commun 5, 3329 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.Yu Y, Sanchez D & Lu N Work of adhesion/separation between soft elastomers of different mixing ratios. J. Mater. Res 30, 2702–2712 (2015). [Google Scholar]
- 72.Choi YS et al. Stretchable, dynamic covalent polymers for soft, long-lived bioresorbable electronic stimulators designed to facilitate neuromuscular regeneration. Nat. Commun 11, 5990 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Kang J et al. Tough and water-insensitive self-healing elastomer for robust electronic skin. Adv. Mater 30, e1706846 (2018). [DOI] [PubMed] [Google Scholar]
- 74.Lu B et al. Pure PEDOT:PSS hydrogels. Nat. Commun 10, 1043 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75.Liu N et al. Ultratransparent and stretchable graphene electrodes. Sci. Adv 3, e1700159 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76.Dickey MD Stretchable and soft electronics using liquid metals. Adv. Mater 29, 1606425 (2017). [Google Scholar]
- 77.Choi S et al. Highly conductive, stretchable and biocompatible Ag-Au core-sheath nanowire composite for wearable and implantable bioelectronics. Nat. Nanotechnol 13, 1048–1056 (2018). [DOI] [PubMed] [Google Scholar]
- 78.Li Y et al. Ultrasensitive and ultrastretchable electrically self-healing conductors. Proc. Natl Acad. Sci. USA 120, e2300953120 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 79.Ohm Y et al. An electrically conductive silver-polyacrylamide-alginate hydrogel composite for soft electronics. Nat. Electron 4, 185–192 (2021). [Google Scholar]
- 80.Xu Y et al. Porous liquid metal-elastomer composites with high leakage resistance and antimicrobial property for skin-interfaced bioelectronics. Sci. Adv 9, eadf0575 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81.Zheng L et al. Conductance-stable liquid metal sheath-core microfibers for stretchy smart fabrics and self-powered sensing. Sci. Adv 7, eabg4041 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Sim K et al. An epicardial bioelectronic patch made from soft rubbery materials and capable of spatiotemporal mapping of electrophysiological activity. Nat. Electron 3, 775–784 (2020). [Google Scholar]
- 83.Song E, Li J, Won SM, Bai W & Rogers JA Materials for flexible bioelectronic systems as chronic neural interfaces. Nat. Mater 19, 590–603 (2020). [DOI] [PubMed] [Google Scholar]
- 84.Zhang Y et al. Climbing-inspired twining electrodes using shape memory for peripheral nerve stimulation and recording. Sci. Adv 5, eaaw1066 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85.Guimarães CF, Gasperini L, Marques AP & Reis RL The stiffness of living tissues and its implications for tissue engineering. Nat. Rev. Mater 5, 351–370 (2020). [Google Scholar]
- 86.Peñuela L et al. Atomic force microscopy for biomechanical and structural analysis of human dermis: a complementary tool for medical diagnosis and therapy monitoring. Exp. Dermatol 27, 150–155 (2018). [DOI] [PubMed] [Google Scholar]
- 87.Choi C, Lee Y, Cho KW, Koo JH & Kim D-H Wearable and implantable soft bioelectronics using two-dimensional materials. Acc. Chem. Res 52, 73–81 (2019). [DOI] [PubMed] [Google Scholar]
- 88.Lee YK et al. Chemical sensing systems that utilize soft electronics on thin elastomeric substrates with open cellular designs. Adv. Funct. Mater 27, 1605476 (2017). [Google Scholar]
- 89.Song E et al. Miniaturized electromechanical devices for the characterization of the biomechanics of deep tissue. Nat. Biomed. Eng 5, 759–771 (2021). [DOI] [PubMed] [Google Scholar]
- 90.Jeong J et al. Materials and optimized designs for human-machine interfaces via epidermal electronics. Adv. Mater 25, 6839–6846 (2013). [DOI] [PubMed] [Google Scholar]; This article reports soft bioelectronics with a stretchable serpentine structure and thickness reduction engineering strategies for interface impedance instability motion artefact management.
- 91.Viventi J et al. A conformal, bio-interfaced class of silicon electronics for mapping cardiac electrophysiology. Sci. Transl. Med 2, 24ra22 (2010). [Google Scholar]
- 92.Lee S et al. A strain-absorbing design for tissue-machine interfaces using a tunable adhesive gel. Nat. Commun 5, 5898 (2014). [DOI] [PubMed] [Google Scholar]
- 93.Kim D-H et al. Materials and noncoplanar mesh designs for integrated circuits with linear elastic responses to extreme mechanical deformations. Proc. Natl Acad. Sci. USA 105, 18675–18680 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94.Rogers JA, Someya T & Huang Y Materials and mechanics for stretchable electronics. Science 327, 1603–1607 (2010). [DOI] [PubMed] [Google Scholar]
- 95.Silva CA et al. Liquid metal based island-bridge architectures for all printed stretchable electrochemical devices. Adv. Funct. Mater 30, 2002041 (2020). [Google Scholar]
- 96.Wang W et al. Strain-insensitive intrinsically stretchable transistors and circuits. Nat. Electron 4, 143–150 (2021). [Google Scholar]
- 97.Zhang Y, Huang Y & Rogers JA Mechanics of stretchable batteries and supercapacitors. Curr. Opin. Solid State Mater. Sci 19, 190–199 (2015). [Google Scholar]
- 98.Kim D-H et al. Stretchable and foldable silicon integrated circuits. Science 320, 507–511 (2008). [DOI] [PubMed] [Google Scholar]
- 99.Ma Y et al. Design of strain-limiting substrate materials for stretchable and flexible electronics. Adv. Funct. Mater 26, 5345–5351 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 100.Wang M et al. Ultrastretchable MXene microsupercapacitors. Small 19, e2300386 (2023). [DOI] [PubMed] [Google Scholar]
- 101.Kim D-H et al. Epidermal electronics. Science 333, 838–843 (2011). [DOI] [PubMed] [Google Scholar]
- 102.Yeo W et al. Multifunctional epidermal electronics printed directly onto the skin. Adv. Mater 25, 2773–2778 (2013). [DOI] [PubMed] [Google Scholar]
- 103.Zhang Y et al. Mechanics of ultra-stretchable self-similar serpentine interconnects. Acta Mater. 61, 7816–7827 (2013). [Google Scholar]
- 104.Xu S et al. Stretchable batteries with self-similar serpentine interconnects and integrated wireless recharging systems. Nat. Commun 4, 1543 (2013). [DOI] [PubMed] [Google Scholar]
- 105.Norton JJS et al. Soft, curved electrode systems capable of integration on the auricle as a persistent brain-computer interface. Proc. Natl Acad. Sci. USA 112, 3920–3925 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 106.Xu L et al. Materials and fractal designs for 3D multifunctional integumentary membranes with capabilities in cardiac electrotherapy. Adv. Mater 27, 1731–1737 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 107.Blees MK et al. Graphene kirigami. Nature 524, 204–207 (2015). [DOI] [PubMed] [Google Scholar]
- 108.Meng K et al. Kirigami-inspired pressure sensors for wearable dynamic cardiovascular monitoring. Adv. Mater 34, 2202478 (2022). [Google Scholar]; This article discusses a kirigami-inspired anisotropic response sensor design strategy for mechanical motion artefact management.
- 109.Brooks AK, Chakravarty S, Ali M & Yadavalli VK Kirigami-inspired biodesign for applications in healthcare. Adv. Mater 34, e2109550 (2022). [DOI] [PubMed] [Google Scholar]
- 110.Luo Y et al. Learning human-environment interactions using conformal tactile textiles. Nat. Electron 4, 193–201 (2021). [Google Scholar]
- 111.Chen G, Fang Y, Zhao X, Tat T & Chen J Textiles for learning tactile interactions. Nat. Electron 4, 175–176 (2021). [Google Scholar]
- 112.Chen G, Li Y, Bick M & Chen J Smart textiles for electricity generation. Chem. Rev 120, 3668–3720 (2020). [DOI] [PubMed] [Google Scholar]
- 113.Fang Y, Chen G, Bick M & Chen J Smart textiles for personalized thermoregulation. Chem. Soc. Rev 50, 9357–9374 (2021). [DOI] [PubMed] [Google Scholar]
- 114.Yin J et al. Smart textiles for self-powered biomonitoring. Med-X 1, 3 (2023). [Google Scholar]
- 115.Creton C Pressure-sensitive adhesives: an introductory course. MRS Bull. 28, 434–439 (2003). [Google Scholar]
- 116.Chung JY & Chaudhury MK Soft and hard adhesion. J. Adhes 81, 1119–1145 (2005). [Google Scholar]
- 117.Cong Y & Fu J Hydrogel-tissue interface interactions for implantable flexible bioelectronics. Langmuir 38, 11503–11513 (2022). [DOI] [PubMed] [Google Scholar]
- 118.Bai S et al. Healable, transparent, room-temperature electronic sensors based on carbon nanotube network-coated polyelectrolyte multilayers. Small 11, 5807–5813 (2015). [DOI] [PubMed] [Google Scholar]
- 119.Sun TL et al. Physical hydrogels composed of polyampholytes demonstrate high toughness and viscoelasticity. Nat. Mater 12, 932–937 (2013). [DOI] [PubMed] [Google Scholar]
- 120.Long T, Li Y, Fang X & Sun J Salt-mediated polyampholyte hydrogels with high mechanical strength, excellent self-healing property, and satisfactory electrical conductivity. Adv. Funct. Mater 28, 1804416 (2018). [Google Scholar]
- 121.Roy CK et al. Self-adjustable adhesion of polyampholyte hydrogels. Adv. Mater 27, 7344–7348 (2015). [DOI] [PubMed] [Google Scholar]
- 122.Pan L et al. A compliant ionic adhesive electrode with ultralow bioelectronic impedance. Adv. Mater 32, 2003723 (2020). [Google Scholar]
- 123.Hou Y et al. Tuning water-resistant networks in mussel-inspired hydrogels for robust wet tissue and bioelectronic adhesion. ACS Nano 17, 2745–2760 (2023). [DOI] [PubMed] [Google Scholar]
- 124.Kim S et al. Injection-on-skin granular adhesive for interactive human–machine interface. Adv. Mater 35, e2307070 (2023). [DOI] [PubMed] [Google Scholar]
- 125.Saiz-Poseu J, Mancebo-Aracil J, Nador F, Busqué F & Ruiz-Molina D The chemistry behind catechol-based adhesion. Angew. Chem. Int. Ed. Engl 58, 696–714 (2019). [DOI] [PubMed] [Google Scholar]
- 126.Han L et al. A mussel-inspired conductive, self-adhesive, and self-healable tough hydrogel as cell stimulators and implantable bioelectronics. Small 13, 1601916 (2017). [Google Scholar]
- 127.Liu J et al. Intrinsically stretchable electrode array enabled in vivo electrophysiological mapping of atrial fibrillation at cellular resolution. Proc. Natl Acad. Sci. USA 117, 14769–14778 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 128.Yang S et al. Stretchable surface electromyography electrode array patch for tendon location and muscle injury prevention. Nat. Commun 14, 6494 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 129.Cadirov N, Booth JA, Turner KL & Israelachvili JN Influence of humidity on grip and release adhesion mechanisms for gecko-inspired microfibrillar surfaces. ACS Appl. Mater. Interfaces 9, 14497–14505 (2017). [DOI] [PubMed] [Google Scholar]
- 130.Ma X et al. Capillary-force-assisted clean-stamp transfer of two-dimensional materials. Nano Lett. 17, 6961–6967 (2017). [DOI] [PubMed] [Google Scholar]
- 131.Cheng S & Robbins MO Nanocapillary adhesion between parallel plates. Langmuir 32, 7788–7795 (2016). [DOI] [PubMed] [Google Scholar]
- 132.Fan J, Coninck JD, Wu H & Wang F Microscopic origin of capillary force balance at contact line. Phys. Rev. Lett 124, 125502 (2020). [DOI] [PubMed] [Google Scholar]
- 133.Liu Y et al. Capillary-force-induced cold welding in silver-nanowire-based flexible transparent electrodes. Nano Lett. 17, 1090–1096 (2017). [DOI] [PubMed] [Google Scholar]
- 134.Kim D-H et al. Dissolvable films of silk fibroin for ultrathin conformal bio-integrated electronics. Nat. Mater 9, 511–517 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 135.Pokrajac L et al. Nanotechnology for a sustainable future: addressing global challenges with the international network4sustainable nanotechnology. ACS Nano 15, 18608–18623 (2021). [DOI] [PubMed] [Google Scholar]
- 136.Bae WG et al. Enhanced skin adhesive patch with modulus-tunable composite micropillars. Adv. Healthc. Mater 2, 109–113 (2013). [DOI] [PubMed] [Google Scholar]
- 137.Kwak MK, Jeong H & Suh KY Rational design and enhanced biocompatibility of a dry adhesive medical skin patch. Adv. Mater 23, 3949–3953 (2011). [DOI] [PubMed] [Google Scholar]
- 138.Kim T, Park J, Sohn J, Cho D & Jeon S Bioinspired, highly stretchable, and conductive dry adhesives based on 1D-2D hybrid carbon nanocomposites for all-in-one ECG electrodes. ACS Nano 10, 4770–4778 (2016). [DOI] [PubMed] [Google Scholar]
- 139.Baik S et al. A wet-tolerant adhesive patch inspired by protuberances in suction cups of octopi. Nature 546, 396–400 (2017). [DOI] [PubMed] [Google Scholar]
- 140.Baik S, Kim J, Lee HJ, Lee TH & Pang C Highly adaptable and biocompatible octopus-like adhesive patches with meniscus-controlled unfoldable 3D microtips for underwater surface and hairy skin. Adv. Sci 5, 1800100 (2018). [Google Scholar]
- 141.Choi MK et al. Cephalopod-inspired miniaturized suction cups for smart medical skin. Adv. Healthc. Mater 5, 80–87 (2016). [DOI] [PubMed] [Google Scholar]
- 142.Chun S et al. Conductive and stretchable adhesive electronics with miniaturized octopus-like suckers against dry/wet skin for biosignal monitoring. Adv. Funct. Mater 28, 1805224 (2018). [Google Scholar]
- 143.Park J et al. Electromechanical cardioplasty using a wrapped elasto-conductive epicardial mesh. Sci. Transl. Med 8, 344ra86 (2016). [Google Scholar]
- 144.Sim K et al. Metal oxide semiconductor nanomembrane-based soft unnoticeable multifunctional electronics for wearable human-machine interfaces. Sci. Adv 5, eaav9653 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 145.Yang Q, Hu Z & Rogers JA Functional hydrogel interface materials for advanced bioelectronic devices. Acc. Mater. Res 2, 1010–1023 (2021). [Google Scholar]
- 146.Kim J, Oh J, Park Y, Kim JJ & Jeong U Soft conductive interfacing for bioelectrical uses: adhesion mechanisms and structural approaches. Macromolecules 56, 4431–4446 (2023). [Google Scholar]
- 147.Tian G et al. A nonswelling hydrogel with regenerable high wet tissue adhesion for bioelectronics. Adv. Mater 35, 2212302 (2023). [Google Scholar]
- 148.Chen X, Yuk H, Wu J, Nabzdyk CS & Zhao X Instant tough bioadhesive with triggerable benign detachment. Proc. Natl Acad. Sci. USA 117, 15497–15503 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 149.Zhan Y, Halliday D, Jiang P, Liu X & Feng J Detecting time-dependent coherence between non-stationary electrophysiological signals-a combined statistical and time-frequency approach. J. Neurosci. Methods 156, 322–332 (2006). [DOI] [PubMed] [Google Scholar]
- 150.Williams WJ, Zaveri HP & Sackellares JC Time-frequency analysis of electrophysiology signals in epilepsy. IEEE Eng. Med. Biol. Mag 14, 133–143 (1995). [Google Scholar]
- 151.Murthy VK, Grove TM, Harvey GA & Haywood LJ Clinical usefulness of ECG frequency spectrum analysis. Proc. Annu. Symp. Comput. Appl. Med. Care 610–612 (1978). [Google Scholar]
- 152.Kostyunina MB & Kulikov MA Frequency characteristics of EEG spectra in the emotions. Neurosci. Behav. Physiol 26, 340–343 (1996). [DOI] [PubMed] [Google Scholar]
- 153.Al-Fahoum AS & Al-Fraihat AA Methods of EEG signal features extraction using linear analysis in frequency and time-frequency domains. ISRN Neurosci. 2014, 730218 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 154.Komi PV & Tesch P EMG frequency spectrum, muscle structure, and fatigue during dynamic contractions in man. Eur. J. Appl. Physiol. Occup. Physiol 42, 41–50 (1979). [DOI] [PubMed] [Google Scholar]
- 155.Grosse P, Cassidy MJ & Brown P EEG-EMG, MEG-EMG and EMG-EMG frequency analysis: physiological principles and clinical applications. Clin. Neurophysiol 113, 1523–1531 (2002). [DOI] [PubMed] [Google Scholar]
- 156.Wang Y et al. Electrically compensated, tattoo-like electrodes for epidermal electrophysiology at scale. Sci. Adv 6, eabd0996 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]; This article presents a signal compensation circuit using the differential method for biopotential and mechanical motion artefact management.
- 157.Sugiyama M et al. An ultraflexible organic differential amplifier for recording electrocardiograms. Nat. Electron 2, 351–360 (2019). [Google Scholar]
- 158.Faisal AA, Selen LPJ & Wolpert DM Noise in the nervous system. Nat. Rev. Neurosci 9, 292–303 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 159.van Drongelen W Signal Processing for Neuroscientists 169–175 (Elsevier Science, 2018). [Google Scholar]
- 160.Lin C-F & Zhu J-D Hilbert-Huang transformation-based time-frequency analysis methods in biomedical signal applications. Proc. Inst. Mech. Eng. H 226, 208–216 (2012). [DOI] [PubMed] [Google Scholar]
- 161.Challis RE & Kitney RI Biomedical signal processing, part 2: the frequency transforms and their inter-relationships. Med. Biol. Eng. Comput 28, 509–524 (1990). [DOI] [PubMed] [Google Scholar]
- 162.Gómez-Echavarría A, Ugarte JP & Tobón C The fractional Fourier transform as a biomedical signal and image processing tool: a review. Biocybern. Biomed. Eng 40, 1081–1093 (2020). [Google Scholar]
- 163.Challis RE & Kitney RI The design of digital filters for biomedical signal processing part 3: the design of Butterworth and Chebychev filters. J. Biomed. Eng 5, 91–102 (1983). [DOI] [PubMed] [Google Scholar]
- 164.Sadasivan PK & Dutt DN Use of finite wordlength FIR digital filter structures with improved magnitude and phase characteristics for reduction of muscle noise in EEG signals. Med. Biol. Eng. Comput 33, 306–312 (1995). [DOI] [PubMed] [Google Scholar]
- 165.Correa AG, Laciar E, Patiño HD & Valentinuzzi ME Artifact removal from EEG signals using adaptive filters in cascade. J. Phys. Conf. Ser 90, 012081 (2007). [Google Scholar]
- 166.Krittanawong C et al. Integration of novel monitoring devices with machine learning technology for scalable cardiovascular management. Nat. Rev. Cardiol 18, 75–91 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 167.Li Y et al. Learning hand kinematics for Parkinson’s disease assessment using a multimodal sensor glove. Adv. Sci 10, 2206982 (2023). [Google Scholar]
- 168.Xu J et al. Deep learning assisted ternary electrification layered triboelectric membrane sensor for self-powered home security. Nano Energy 113, 108524 (2023). [Google Scholar]
- 169.Dillen A et al. Deep learning for biosignal control: insights from basic to real-time methods with recommendations. J. Neural Eng 19, 011003 (2022). [Google Scholar]
- 170.Sun W, Su Y, Wu X & Wu X A novel end-to-end 1D-ResCNN model to remove artifact from EEG signals. Neurocomputing 404, 108–121 (2020). [Google Scholar]
- 171.Xiong D, Zhang D, Zhao X & Zhao Y Deep learning for EMG-based human-machine interaction: a review. IEEE/CAA J. Autom. Sin 8, 512–533 (2021). [Google Scholar]
- 172.Luca CJD, Gilmore LD, Kuznetsov M & Roy SH Filtering the surface EMG signal: movement artifact and baseline noise contamination. J. Biomech 43, 1573–1579 (2010). [DOI] [PubMed] [Google Scholar]; This article reports on a signal processing strategy based on a Butterworth filtering algorithm in electrophysiological monitoring for mechanical motion artefact management.
- 173.Wang M et al. A wearable electrochemical biosensor for the monitoring of metabolites and nutrients. Nat. Biomed. Eng 6, 1225–1235 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 174.Yang Y & Gao W Wearable and flexible electronics for continuous molecular monitoring. Chem. Soc. Rev 48, 1465–1491 (2018). [Google Scholar]
- 175.Flynn CD et al. Biomolecular sensors for advanced physiological monitoring. Nat. Rev. Bioeng 1, 560–575 (2023). [Google Scholar]
- 176.Chen S et al. Wearable flexible microfluidic sensing technologies. Nat. Rev. Bioeng 1, 950–971 (2023). [Google Scholar]
- 177.Che Z, Wan X & Xu J et al. Speaking without vocal folds using a machine-learningassisted wearable sensing-actuation system. Nat. Commun 15, 1873 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 178.Zhou Y et al. A multimodal magnetoelastic artificial skin for underwater haptic sensing. Sci. Adv 10, eadj8567 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 179.Fang Y et al. Ambulatory cardiovascular monitoring via a machine-learning-assisted textile triboelectric sensor. Adv. Mater 33, 2104178 (2021). [Google Scholar]; This article presents a machine learning-assisted signal features extraction strategy for biopotential and mechanical motion artefact management.
- 180.Yi Z et al. Piezoelectric dynamics of arterial pulse for wearable continuous blood pressure monitoring. Adv. Mater 34, 2110291 (2022). [Google Scholar]
- 181.Rowland EM, Bailey EL & Weinberg PD Estimating arterial cyclic strain from the spacing of endothelial nuclei. Exp. Mech 61, 171–190 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 182.Edin BB & Johansson N Skin strain patterns provide kinaesthetic information to the human central nervous system. J. Physiol 487, 243–251 (1995). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 183.Park B et al. Cuticular pad-inspired selective frequency damper for nearly dynamic noise-free bioelectronics. Science 376, 624–629 (2022). [DOI] [PubMed] [Google Scholar]
- 184.Jeong C et al. Motion artifact-resilient zone for implantable sensors. Adv. Funct. Mater 32, 2206461 (2022). [Google Scholar]
- 185.Rodeheaver N, Kim H, Herbert R, Seo H & Yeo W-H Breathable, wireless, thin-film wearable biopatch using noise-reduction mechanisms. ACS Appl. Electron. Mater 4, 503–512 (2022). [Google Scholar]
- 186.Zhao Y et al. A wearable freestanding electrochemical sensing system. Sci. Adv 6, eaaz0007 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 187.Xu S et al. Soft microfluidic assemblies of sensors, circuits, and radios for the skin. Science 344, 70–74 (2014). [DOI] [PubMed] [Google Scholar]
- 188.Wang W et al. Neuromorphic sensorimotor loop embodied by monolithically integrated, low-voltage, soft e-skin. Science 380, 735–742 (2023). [DOI] [PubMed] [Google Scholar]
- 189.Rivnay J et al. Organic electrochemical transistors. Nat. Rev. Mater 3, 17086 (2018). [Google Scholar]
- 190.Dai Y, Hu H, Wang M, Xu J & Wang S Stretchable transistors and functional circuits for human-integrated electronics. Nat. Electron 4, 17–29 (2021). [Google Scholar]
- 191.Molina-Lopez F et al. Inkjet-printed stretchable and low voltage synaptic transistor array. Nat. Commun 10, 2676 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 192.Shim H et al. An elastic and reconfigurable synaptic transistor based on a stretchable bilayer semiconductor. Nat. Electron 5, 660–671 (2022). [Google Scholar]
- 193.Xu J et al. Highly stretchable polymer semiconductor films through the nanoconfinement effect. Science 355, 59–64 (2017). [DOI] [PubMed] [Google Scholar]
- 194.Liu K, Ouyang B, Guo X, Guo Y & Liu Y Advances in flexible organic field-effect transistors and their applications for flexible electronics. NPJ Flex. Electron 6, 1 (2022). [Google Scholar]
- 195.Uemura T & Sekitani T Ultraflexible Biosignal Amplifier Based on Organic Thin-film Transistors. 2019 Compound Semiconductor Week (CSW). (IEEE, 2019). [Google Scholar]
- 196.Krishnan S & Athavale Y Trends in biomedical signal feature extraction. Biomed. Signal Process. Control 43, 41–63 (2018). [Google Scholar]
- 197.Lee K et al. Mechano-acoustic sensing of physiological processes and body motions via a soft wireless device placed at the suprasternal notch. Nat. Biomed. Eng 4, 148–158 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 198.Ni X et al. Automated, multiparametric monitoring of respiratory biomarkers and vital signs in clinical and home settings for COVID-19 patients. Proc. Natl Acad. Sci. USA 118, e2026610118 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 199.Franklin D et al. Synchronized wearables for the detection of haemodynamic states via electrocardiography and multispectral photoplethysmography. Nat. Biomed. Eng 7, 1229–1241 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 200.Liu J et al. PCA-based multi-wavelength photoplethysmography algorithm for cuffless blood pressure measurement on elderly subjects. IEEE J. Biomed. Health Inform 25, 663–673 (2020). [Google Scholar]
- 201.Shen CP et al. Convolution neural network algorithm for shockable arrhythmia classification within a digitally connected automated external defibrillator. J. Am. Heart Assoc 12, e026974 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 202.Li S, Lu D, Pet M, Rogers JA & Huang Y An analytical model for sensing microvascular blood flow in flaps and organ grafts. J. Mech. Phys. Solids 170, 105119 (2023). [Google Scholar]
- 203.Webb RC et al. Epidermal devices for noninvasive, precise, and continuous mapping of macrovascular and microvascular blood flow. Sci. Adv 1, e1500701 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 204.Lee S et al. A transparent bending-insensitive pressure sensor. Nat. Nanotechnol 11, 472–478 (2016). [DOI] [PubMed] [Google Scholar]
- 205.Song J-K et al. Stretchable colour-sensitive quantum dot nanocomposites for shape-tunable multiplexed phototransistor arrays. Nat. Nanotechnol 17, 849–856 (2022). [DOI] [PubMed] [Google Scholar]
- 206.Kim SH et al. A bioinspired stretchable sensory-neuromorphic system. Adv. Mater 33, e2104690 (2021). [DOI] [PubMed] [Google Scholar]
- 207.Shi X et al. Large-area display textiles integrated with functional systems. Nature 591, 240–245 (2021). [DOI] [PubMed] [Google Scholar]
- 208.He J et al. Scalable production of high-performing woven lithium-ion fibre batteries. Nature 597, 57–63 (2021). [DOI] [PubMed] [Google Scholar]
- 209.Ouyang W et al. A wireless and battery-less implant for multimodal closed-loop neuromodulation in small animals. Nat. Biomed. Eng 7, 1252–1269 (2023). [DOI] [PubMed] [Google Scholar]
- 210.Jin S et al. Injectable tissue prosthesis for instantaneous closed-loop rehabilitation. Nature 623, 58–65 (2023). [DOI] [PubMed] [Google Scholar]
- 211.Luan L et al. Recent advances in electrical neural interface engineering: minimal invasiveness, longevity, and scalability. Neuron 108, 302–321 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 212.Harrison RR & Charles C A low-power low-noise CMOS amplifier for neural recording applications. IEEE J. Solid State Circuits 38, 958 (2003). [Google Scholar]
- 213.Ershad F et al. Ultra-conformal drawn-on-skin electronics for multifunctional motion artifact-free sensing and point-of-care treatment. Nat. Commun 11, 3823 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 214.Zhang Y et al. Printing, folding and assembly methods for forming 3D mesostructures in advanced materials. Nat. Rev. Mater 2, 17019 (2017). [Google Scholar]
- 215.Liu S et al. Conformability of flexible sheets on spherical surfaces. Sci. Adv 9, eadf2709 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 216.Choi C et al. Human eye-inspired soft optoelectronic device using high-density MoS2-graphene curved image sensor array. Nat. Commun 8, 1664 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 217.Khan Y et al. A new frontier of printed electronics: flexible hybrid electronics. Adv. Mater 32, e1905279 (2020). [DOI] [PubMed] [Google Scholar]
- 218.Xu Y et al. In-ear integrated sensor array for the continuous monitoring of brain activity and of lactate in sweat. Nat. Biomed. Eng 7, 1307–1320 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 219.Balakrishnan G, Song J, Mou C & Bettinger CJ Recent progress in materials chemistry to advance flexible bioelectronics in medicine. Adv. Mater 34, 2106787 (2022). [Google Scholar]
- 220.Han M et al. Three-dimensional piezoelectric polymer microsystems for vibrational energy harvesting, robotic interfaces and biomedical implants. Nat. Electron 2, 26–35 (2019). [Google Scholar]
- 221.Zhou Y et al. Stretchable high-permittivity nanocomposites for epidermal alternating-current electroluminescent displays. ACS Mater. Lett 1, 511–518 (2019). [Google Scholar]
- 222.Luo Y et al. Devising materials manufacturing toward lab-to-fab translation of flexible electronics. Adv. Mater 32, 2001903 (2020). [Google Scholar]
- 223.Shi HH et al. Sustainable electronic textiles towards scalable commercialization. Nat. Mater 22, 1294–1303 (2023). [DOI] [PubMed] [Google Scholar]
- 224.Kiran DR In: Total Quality Management: Key Concepts and Case Studies 391–404 (Elsevier Science, 2017). [Google Scholar]
- 225.Zheng Y et al. Environmentally stable and stretchable polymer electronics enabled by surface-tethered nanostructured molecular-level protection. Nat. Nanotechnol 18, 1175–1184 (2023). [DOI] [PubMed] [Google Scholar]
- 226.Bierman AS & Tinetti ME Precision medicine to precision care: managing multimorbidity. Lancet 388, 2721–2723 (2016). [DOI] [PubMed] [Google Scholar]
- 227.Zhou Z et al. Sign-to-speech translation using machine-learning-assisted stretchable sensor arrays. Nat. Electron 3, 571–578 (2020). [Google Scholar]
- 228.Bouton CE et al. Restoring cortical control of functional movement in a human with quadriplegia. Nature 533, 247–250 (2016). [DOI] [PubMed] [Google Scholar]
- 229.Hou B et al. An interactive mouthguard based on mechanoluminescence-powered optical fibre sensors for bite-controlled device operation. Nat. Electron 5, 682–693 (2022). [Google Scholar]
- 230.Yu X et al. Skin-integrated wireless haptic interfaces for virtual and augmented reality. Nature 575, 473–479 (2019). [DOI] [PubMed] [Google Scholar]
- 231.Bai H et al. Stretchable distributed fiber-optic sensors. Science 370, 848–852 (2020). [DOI] [PubMed] [Google Scholar]
- 232.Kerr EA & Hayward RA Patient-centered performance management: enhancing value for patients and health care systems. JAMA 310, 137–138 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 233.Rose N Personalized medicine: promises, problems and perils of a new paradigm for healthcare. Procedia Soc. Behav. Sci 77, 341–352 (2013). [Google Scholar]
- 234.Li Z, Tian X, Qiu C-W & Ho JS Metasurfaces for bioelectronics and healthcare. Nat. Electron 4, 382–391 (2021). [Google Scholar]
- 235.Schwalbe N & Wahl B Artificial intelligence and the future of global health. Lancet 395, 1579–1586 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 236.Kim J et al. Skin-interfaced wireless biosensors for perinatal and paediatric health. Nat. Rev. Bioeng 1, 631–647 (2023). [Google Scholar]
- 237.Xu S et al. Wireless skin sensors for physiological monitoring of infants in low-income and middle-income countries. Lancet Digit. Health 3, e266–e273 (2021). [DOI] [PubMed] [Google Scholar]
