ABSTRACT
Boron neutron capture therapy (BNCT) utilizes high linear energy transfer (LET) α-particles and 7Li ions generated through the 10B(n, α)7Li reaction. Precise dosimetry is essential for maximizing therapeutic efficacy while minimizing normal tissue adverse events, considering the microscopic distribution of 10B and cellular structures. Recently, the photon isoeffective dose (DisoE) has been proposed as a more appropriate metric for BNCT treatment planning and can be evaluated using the stochastic microdosimetric kinetic (SMK) model. However, clinical implementation of the SMK model remains challenging due to the difficulty of evaluating its input parameters, which requires computationally intensive radiation transport simulations at the cellular scale. To address this issue, we developed LISMEC (Linear Interpolation System for Stochastic Microdosimetric Kinetic model parameters Evaluated from Cellular-scale simulation), a rapid estimation framework based on precomputed cellular-scale PHITS (Particle and Heavy Ion Transport code System) simulations covering various cell geometries and boron distributions. By applying a linear interpolation algorithm, LISMEC enables the retrieval of SMK model parameters without the need for computationally intensive cellular-scale simulations. The utility of LISMEC, in conjunction with PHITS, was demonstrated through simulations of various irradiation scenarios in reactor-based BNCT. The results showed that DisoE values ranged from 7.4 to 32.7 Gy, even under a fixed macroscopic 10B concentration of 60 ppm. These findings emphasize the importance of incorporating a microscopic distribution of 10B and cellular structures into BNCT treatment planning.
Keywords: BNCT, microdosimetry, boron distribution, cell morphology
INTRODUCTION
Boron neutron capture therapy (BNCT) is a radiation modality that leverages high linear energy transfer (LET) alpha particles and 7Li ions emitted via the 10B(n, α)7Li reaction [1]. Because these charged particles have an extremely short range—<10 μm—they selectively deposit substantial energy within tumor cells, producing a potent cytotoxic effect. In 2020, BNCT for unresectable head and neck cancers obtained national health insurance approval in Japan [2, 3]. This approval has encouraged research and clinical studies aimed at expanding the types of diseases treatable with BNCT. Drugs containing boron that are currently available have limited applicability to certain tumors. Therefore, developing novel boron-containing agents is essential to realize broader clinical applications [4]. A commonly used metric for comparing different radiation types is the relative biological effectiveness (RBE), defined as the ratio of doses required to achieve the same biological endpoint, typically benchmarked against X-rays [5, 6]. In BNCT, however, an additional factor called the compound biological effectiveness (CBE) has been introduced to reflect the in-cell behavior of boron agents [7–10]. Although CBE values incorporate parameters such as boron compound type, biological endpoints and cell characteristics, they often rely on empirical assumptions and may insufficiently address dose dependence, synergistic effects among multiple radiation components or the highly localized nature of dose distributions [11–14].
To overcome these limitations, the photon isoeffective dose (DisoE) has been proposed as a more theoretically grounded index [15, 16]. However, these studies cannot explicitly consider cell morphology and intracellular 10B distribution. Then, we previously developed a method to apply the stochastic microdosimetric kinetic (SMK) model [17, 18] to BNCT dosimetry, enabling the direct calculation of DisoE [13, 19] while taking these effects into consideration. The SMK model predicts cell survival by evaluating the probability distributions of energy deposition in both the cell nucleus and subnuclear domains—an approach well suited for high-LET therapies like BNCT, carbon-ion therapy [20] and targeted alpha therapy [21], which exhibit significant dose heterogeneity at microscopic scales.
Nevertheless, harnessing the SMK model for BNCT requires detailed cellular-scale Monte Carlo simulations that account for cell morphology and intracellular 10B distribution, followed by the computation of DisoE. Constructing specialized cell models for each boron agent or clinical scenario, and then running resource-intensive simulations, has hindered the broader adoption of this approach.
In this study, we performed comprehensive cellular-scale simulations under various conditions using the general-purpose Particle and Heavy Ion Transport code System (PHITS) [22]. From these precomputed results, we developed an Excel-based system—termed ‘LISMEC’ (Linear Interpolation system for Stochastic Microdosimetric kinetic model parameters Evaluated from Cellular-scale simulation)—that permits rapid and straightforward estimation of SMK model parameters. With LISMEC, users only need to specify seven key parameters related to cell morphology, 10B distribution and microdosimetric properties to obtain DisoE over a wide range of conditions without additional complex simulations.
In this article, we detail the development and implementation of LISMEC and demonstrate its utility for BNCT dosimetry by applying it to radiation field data from the Kyoto University Reactor (KUR) at the Institute for Integrated Radiation and Nuclear Science. Specifically, we compute DisoE for diverse cellular structures and boron distributions, and then analyze the resultant dose profiles. Finally, we discuss challenges for clinical application and propose directions for future work.
MATERIALS AND METHODS
Procedure for calculating DisoE using the SMK model
Figure 1 provides an overview of the workflow for calculating DisoE in this study. Parameters shown in gray boxes represent user inputs. The process consists of the following steps:
Fig. 1.
Flowchart for calculating DisoE developed in this study.
-
Human-body-scale PHITS simulation:
In this step, a macroscale PHITS simulation is conducted to calculate the absorbed doses for four BNCT dose components: boron, nitrogen, hydrogen and photon, respectively [23, 24].
-
Retrieval of SMK model parameters via LISMEC:
Building on the results of a precomputed ‘cellular-scale PHITS simulation’, LISMEC provides the SMK model parameters for the specified cell morphology and 10B distribution.
-
Calculation of the cell survival fraction S (D):
Using the dose data from steps 1 and 2, the cell survival fraction S (D) is determined.
-
Calculation of DisoE:
Finally, given user-specified values for α (Gy−1), β (Gy−2) (the reference radiation’s linear and quadratic parameters) and X (Gy per fraction), the photon isoeffective dose DisoE is calculated.
Input parameters for the SMK model
The following parameters must be provided to the SMK model:
: conversion factors that translate the macroscopic boron dose into the cell nuclear dose [25].
applies to 10B located inside the cell, whereas
applies to 10B outside the cell.
: the dose-mean cell-nucleus dose for each dose component i.
: the dose-mean domain dose within the cell nucleus.
: the saturation-corrected domain dose, accounting for the overkill effect.
In this study, we precomputed these parameters across various cell morphologies and 10B distributions, then incorporated them into an Excel-based system (LISMEC) where they can be efficiently retrieved via linear interpolation. Consequently, when clinical conditions or boron agents change, complex cellular-scale simulations need not be repeated. DisoE can thus be computed promptly and with minimal computational burden using the SMK model.
Cellular-scale PHITS simulation
In this study, we constructed 512 cell models by systematically varying three parameters that characterize the cellular morphology: (1) the nuclear surface area (Snucl), (2) the ratio of nuclear to total cell surface area (Snucl/Scell) and (3) the fraction of total volume occupied by cells (Vcell/Vtotal). Each model comprises an 11 × 11 × 11 array of cubic cells surrounded by an extracellular region. Specifically, we set (Snucl) to 9, 36, 81 or 144 μm2; Snucl/Scell to 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7 or 0.8; and Vcell/Vtotal to ~0.3–0.95 by adjusting the intercellular distance (0.1, 0.5, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13 and 14 μm). All the regions were assigned a density of 1 g cm−3 and an elemental composition of 10.7% hydrogen, 12.1% carbon, 2% nitrogen, 71.4% oxygen [26]. The remaining 3.8% other elements was neglected in the calculations, resulting in a total of 96.2%.
For each of the 512 models, we considered three scenarios for the 10B(n, α)7Li reaction: (1) in the extracellular space, (2) in the cytoplasm or (3) in the nucleus. Radiation transport calculations were then performed using PHITS versions 3.33 and 3.341.
Figure 2 illustrates the source-generation procedure. In the nucleus scenario, a random reaction point was sampled within the nucleus, emitting alpha particles and 7Li ions via a correlated source function. The correlated source function is a built-in feature of PHITS enabling simultaneous sampling of multiple radiations from the same spatial coordinate. In this study, we utilized isocorr = 3. For cytoplasmic and extracellular scenarios, we accounted for spatial configuration by randomizing the relative positions of the nucleus and the source region in each history (Fig. 2). Since the default PHITS capabilities cannot accommodate this, we employed a user-defined source function. The 10B(n, α)7Li reaction proceeds via two branching channels, one accompanied by gamma emission and one without, each yielding alpha particles and 7Li ions of different energies. We conducted separate simulations for each channel and combined them according to their branching ratios.
Fig. 2.
Illustration of the source regions and the overall dose distribution for the extracellular region, cytoplasm, and nucleus.
In Fig. 2, the nine small squares represent the cell nucleus, where dose calculations are performed. Each of the four images within each scenario shows an example of one possible arrangement. The dose distribution on the right is obtained by aggregating many such randomized arrangements, including the four cases illustrated here.
To model the dose by neutron-induced reactions by nitrogen and hydrogen, as well as the gamma-rays produced by these reactions, we uniformly generated source particles from cells and interstitial regions in the center of the cell model. The source information was derived from a macroscale simulation of a 1 cm3 International Commission on Radiation Units and Measurements (ICRU) phantom under neutron irradiation [13]. All the particles were transported down to 1 keV, and electrons were managed using the Electron Gamma Shower (EGS) mode [27]. We selected a sufficient number of histories to keep the statistical uncertainty < 1% in each ‘bin’. Here, a ‘bin’ is a small interval of energy or dose used to categorize simulation results for statistical analysis. From these simulations, we extracted the mean energy deposited in the cell nuclei and in the entire cell (nuclei, cytoplasm and extracellular space). We also tallied the frequency distributions of specific energies (nuclear and domain). To calculate domain-specific energies, we used PHITS’s microdosimetric function [t-sed] tally [14] with eight domain diameters: 0.05, 0.1, 0.2, 0.3, 0.4, 0.5, 0.7 and 1 μm.
Development of LISMEC
Outline of the system
To streamline the retrieval of
—the parameters required for SMK model calculations under any cell morphology or 10B distribution—we developed an Excel-based system called ‘LISMEC’. This system applies linear interpolation to the precomputed simulation data described in the previous section. The user interface is shown in Fig. 3.
Fig. 3.

User interface of the LISMEC system.
By specifying inputs in three categories—(1) SMK model parameters, (2) boron distribution parameters and (3) cellular morphologic parameters—LISMEC yields the ‘boron dose correction coefficient’, ‘dose-mean cell-nucleus specific energy’, ‘dose-mean domain-specific energy’ and ‘saturation-corrected dose-mean domain-specific energy’ for each BNCT dose component. Note that specific energy is defined as the quotient of the energy imparted to a target volume by its mass [28]. Although it shares the same units as the absorbed dose, it is a stochastic quantity that should be expressed by its probability density. In Fig. 3, the pale blue fields represent user inputs, while the pale red fields show the outputs. If certain parameters are unknown or difficult to measure, default values can be substituted.
Furthermore, LISMEC automatically arranges these outputs on a separate Excel sheet in a format compatible with PHITS. By simply copying and pasting the generated data into a PHITS input file, users can seamlessly incorporate the LISMEC-derived parameters for DisoE calculations.
Input parameters of LISMEC
The input parameters in LISMEC (Fig. 3) are designed to capture morphological features and boron distribution data that closely reflect actual cellular conditions. By integrating experimentally measured or literature-based morphological characteristics and boron distribution profiles, BNCT dose assessments can more accurately reflect real biological scenarios.
Domain radius: rd (mm)
This parameter represents the domain radius in the MK model. Initially, it was defined as the distance within the cell nucleus over which DNA damage can diffuse. In practice, it is typically determined by least-squares fitting of cell survival data obtained from multiple radiation types.
Saturation parameter: y0 (keV mm−1)
This parameter sets the threshold for the overkill effect, in which cells receive more energy than necessary to induce cell death. As with rd, it is typically determined by least-squares fitting of cell survival data from multiple radiation types.
Boron uptake ratio into cell: Ccell/Cextra or Acell/Aextra
This parameter reflects the extent of boron uptake by cells relative to the extracellular space. It can be expressed either as a concentration ratio (Ccell/Cextra) or an abundance ratio (Acell/Aextra). These values can be determined via immunofluorescence imaging [29] or by combining CR-39 track-etch and H&E-stained imaging [30, 31].
Boron uptake ratio into cell nucleus: Cnucl/Ccyto or Anucl/Acyto
This parameter reflects the extent of boron accumulation in the nucleus relative to the cytoplasm. It can be expressed as either a concentration ratio (Ccell/Cextra) or an abundance ratio (Acell/Aextra). These values are also typically determined via immunofluorescence imaging [29] or by combining CR-39 track-etch with Hematoxylin Eosin (H&E)-stained imaging [30, 31].
Volume ratio of cells to total tissue: Vcell/Vtotal
This parameter indicates the fraction of the total tissue volume occupied by cells. It can be determined through immunofluorescence or H&E-stained image analysis [29], as well as by MRI-based methods [32–34].
Mean cross-sectional area of cell nucleus and its ratio to cell: Snucl (mm2) and Snucl/Scell
These parameters represent the average cross-sectional area of the nucleus (Snucl [μm2]) and the fraction of the cell’s total cross-sectional area occupied by the nucleus (Snucl/Scell). They can be measured via immunofluorescence or H&E-stained image analysis [11, 29, 35].
Output parameters of LISMEC
Conversion factor from the macroscopic boron dose to cell nucleus dose: κintra and κextra
In the SMK model, the cell survival fraction is estimated based on the dose delivered to the cell nucleus. Hence, a conversion factor κ is required to translate the macroscale kerma into the cell-nucleus dose. In this study, κ is defined as follows:
![]() |
(1) |
where Nn, Nc and Ne are the relative proportions of 10B in the nucleus, cytoplasm and extracellular space, respectively. Meanwhile, Rn, Rc and Re represent the ratio of the cell-nucleus dose when 10B is localized in the nucleus, cytoplasm or extracellular space, respectively, to the cell-nucleus dose with uniformly distributed 10B. The total proportion of 10B across all compartments is normalized to 1.0:
![]() |
(2) |
Furthermore,
![]() |
(3) |
![]() |
(4) |
By these definitions, Equation (1) can be rewritten as
![]() |
(5) |
Dose-mean cell-nucleus specific energy is
Using the [t-deposit] function in the cellular-scale PHITS simulation, we obtain the probability density
for the specific energy
imparted by each radiation component
to the cell nucleus. From this distribution, we then calculate the dose-mean cell-nucleus specific energy
according to equation (6).
![]() |
(6) |
Dose-mean and saturation-corrected domain specific energy:
and
Using the microdosimetric function in the cellular-scale PHITS simulation, we obtained the probability density
for the specific energy imparted by each radiation component
within the domain. From this distribution, we then calculated the dose-mean and saturation-corrected domain-specific energies
and
according to Equations (7) and (8).
![]() |
(7) |
![]() |
(8) |
In this study, we treated
(with a saturation parameter y0 = 1 × 104 [keV mm−1]) as equivalent to
.
We then compiled the cellular-scale PHITS simulation results in Excel and used the VLOOKUP, INDEX and MATCH functions to perform linear interpolation, enabling efficient retrieval of SMK model parameters for arbitrary conditions.
Validation of the interpolation method accuracy
We assessed the accuracy of the linear interpolation method in two steps. First, under a uniform boron distribution (boron concentration ratio in tissue: cell/extracellular = 1, nucleus/cytoplasm = 1), κall is theoretically equal to 1. We used this as our accuracy benchmark. Second, we quantified the deviation of κall from 1 at linearly interpolated locations within the parameter space. In nonuniform conditions, we conducted several validation simulations (four points) for scenarios that had not been evaluated during LISMEC’s development and compared the outcomes with LISMEC’s linear-interpolation results.
Applications of LISMEC
Sensitivity analysis of SMK model parameters using LISMEC
We used LISMEC to investigate how the boron uptake ratio (Ccell/Cextra) and cellular morphometric parameters (Vcell/Vtotal, Snucl/Scell and Snucl) affect the SMK model parameters. We varied four parameters (Ccell/Cextra, Vcell/Vtotal, Snucl/Scell and Snucl) and evaluated the resulting values of κ
. The sensitivity analysis was performed under the following conditions: ・Snucl/Scell = 0.2, 0.8 ・Vcell/Vtotal = 0.3, 0.9 ・Ccell/Cextra = 0, 0.1, 0.5, 1, 3.5 ・Snucl = 60, 120 mm2. We assumed a uniform boron concentration inside the cell (Cnucl/Ccyto = 1) [36]. Reference values for the nuclear cross-sectional area were drawn from basal cell carcinoma (65 μm2 [37]) and breast carcinoma (150 μm2 [38]). The upper limit of 3.5 for the boron uptake ratio was adopted from a previous study [39] as a representative example.
DisoE evaluation in BNCT radiation fields
In this study, we employed the epithermal neutron irradiation mode (CO-0000-F mode) of the Heavy Water Neutron Irradiation Facility (HWNIF) installed at Kyoto University Reactor (KUR) as our BNCT radiation field [40, 41]. Figure 4 illustrates the computational setup used in the simulation and shows the fluence distribution obtained via PHITS. The PHITS simulations were conducted under the following conditions:
Fig. 4.
KUR–HWNIF radiation field. (Left) Cross-sectional view of the computational geometry representing the KUR irradiation port and the phantom. (Right) Neutron fluence distribution obtained via a PHITS simulation of the same geometry under the KUR thermal–epithermal neutron mode (CO0000F). The color scale indicates the magnitude of the neutron fluence.
· Irradiation geometry: Neutrons were directed from a 12-cm-diameter beam port onto a cubic phantom measuring 20 × 20 × 20 cm3.
· Phantom: Composed of ICRU soft tissue enriched with 60 ppm of 10B [42, 43].
· Calculation: Using the [t-track] tally, we calculated absorbed doses for each BNCT component from the neutron fluence using the kerma approximation at 1 cm intervals from the phantom’s surface to a depth of 20 cm. We ran 20 000 000 histories to keep the statistical uncertainty <1%.
Using these results, we converted the absorbed doses into DisoE under various conditions specified in Section 2.5.1 (Sensitivity Analysis of the SMK Model Parameters Using LISMEC) and evaluated the resulting DisoE (Fig. 6).
Fig. 6.
D isoE evaluation using LISMEC.
We also compared DisoE with the CBE- and RBE-weighted doses, as well as the absorbed dose (Fig. 7). For the DisoE calculations in Fig. 7, we adopted: ・Ccell/Cextra = 1 ・Snucl/Scell = 0.2 ・Vcell/Vtotal = 0.9 ・Snucl = 120 μm2. These settings were chosen simply to illustrate how DisoE can be evaluated in a BNCT radiation field. Specifically, Ccell/Cextra = 1 minimizes the influence of the other parameters (Vcell/Vtotal, Snucl/Scell, Snucl), so these particular values are not intended to have any special significance.
Fig. 7.

Comparison of DisoE, CBE–RBE-weighted dose and absorbed dose.
For the CBE- and RBE-weighted dose calculations in Fig. 7, we assigned CBE and RBE values of 3.8, 2.5, 2.5 and 2.5 to 10B, N, H and γ, respectively [13]. Because the absolute absorbed dose is critical for evaluating DisoE, we assumed a blood boron concentration of 20 ppm [42] and a skin-to-blood ratio of 1.1 [44], resulting in 22 ppm in the skin. We then adjusted the normalization so that the CBE- and RBE-weighted dose at the skin reached 12 Gy-Eq [9]. At this point, we set CBE and RBE for 10B, N, H and γ at 2.5, 2.5, 2.5 and 1, respectively [13]. These choices ensured proper scaling of the absolute dose under realistic BNCT irradiation conditions. Furthermore, we employed the following SMK model parameters:
·r0 = 0.24 [μm] [13, 19] ·y0 = 125 [keV μm−1] [28] ·α0 = 0.0422 [Gy−1] [13, 19].
·β0 = 0.00822 [Gy−2] [13, 19] ·α = 0.0633 [Gy−1] [13, 19] ·β = 0.00822 [Gy−2] [13, 19].
Except for y0 [keV μm−1], these parameters were derived from survival data for SCC VII squamous cell carcinomas [13, 19, 45].
RESULTS
Sensitivity analysis of SMK model parameters using LISMEC
Figure 5 illustrates the analysis results for κ as well as
. On the horizontal axis, we plotted the boron uptake ratio Ccell/Cextra, while the vertical axis displays κ,
and
. These results compare how changes in cellular morphologic parameters (Vcell/Vtotal, Snucl/Scell and Snucl) and the boron uptake ratio (Ccell/Cextra) affect these quantities. From Fig. 5, it is evident that κ and
strongly depend on the boron uptake ratio and cellular morphologic parameters. In contrast,
and
show only minor dependence on those same factors.
Fig. 5.
SMK model parameter analysis using LISMEC.
D isoE evaluation in BNCT radiation fields
Figure 6 shows the depth-dose distributions of DisoE when varying the cellular morphologic parameters (Vcell/Vtotal, Snucl/Scell, Snucl) and the boron uptake ratio (Ccell/Cextra). Even under the same macro boron concentration (60 ppm), these parameter settings can substantially alter the DisoE. The lower-left panel of Fig. 6 (10B uptake ratio = 1) can also be regarded as the result of a DisoE calculation that neglects both the boron distribution and the cellular structure. If the boron distribution and cell structure are not taken into account, all the other calculation results would likewise be homogenized and evaluated as equivalent to the lower-left panel of Fig. 6 (10B uptake ratio = 1). However, by considering the boron distribution and the cellular structure, the dose under each condition can be separated and evaluated.
Meanwhile, Fig. 7 compares DisoE with both the CBE- and RBE-weighted dose and absorbed dose. In regions of high dose, the CBE- and RBE-weighted doses tend to be significantly greater than the DisoE.
Validation result of the interpolation method accuracy
We confirmed that, overall, the error was within ~+5%, and that in the region Snucl/Scell = 0.1–0.2 it was within +15%. Comparing the computed results for conditions not evaluated during LISMEC’s development with LISMEC’s estimates, we confirmed that the error was generally within +5%. However, for nuclear localization in the Snucl/Scell range of 0.1–0.2, an error of +21% was observed. In contrast, for cytoplasmic and extracellular localization within the same range, the errors were within 5%.
DISCUSSION
Sensitivity of SMK model parameters to boron uptake ratio and cellular morphologic parameters
In Fig. 5, we showed how microscopic parameters—cell occupancy (Vcell/Vtotal), the ratio of nuclear to cellular cross-sectional area (Snucl/Scell), nuclear cross-sectional area (Snucl) and boron uptake ratio (Ccell/Cextra)—affect the conversion factor κ (which translates macroscale kerma to cell-nucleus dose) as well as
. The findings indicate that these microscopic parameters significantly influence the behavior of κ.
First, κ approached 1 when Ccell/Cextra = 1, regardless of the cell occupancy (Vcell/Vtotal), the ratio of nuclear to cellular cross-section area (Snucl/Scell) and nuclear cross-section area (Snucl). This likely occurs because the intracellular and extracellular boron concentrations become equal, minimizing differences in boron dose between the macro and microscales.
Second, when cell occupancy is low (Vcell/Vtotal is small), κ tends to increase in proportion to Ccell/Cextra. This can be interpreted as the larger extracellular space rendering a high uptake ratio more impactful, effectively concentrating boron within cells. On the other hand, when cell occupancy is high (Vcell/Vtotal is large), κ saturates at a comparatively early stage even with an elevated uptake ratio, likely because most of the tissue is composed of cells, so the boron distribution becomes nearly homogeneous at the microscopic level, reducing discrepancies in dose evaluation.
Third, the effect of Snucl/Scell is particularly evident for Ccell/Cextra < 1, where higher values of Snucl/Scell correlate with higher κ. This is presumably because charged particles originating from extracellular boron more readily reach the nucleus due to geometric considerations. In contrast, when Ccell/Cextra > 1 and the cell occupancy is high, Snucl/Scell has little influence because the tissue is occupied by cells. On the other hand, when Ccell/Cextra > 1 and the cell occupancy is low, a smaller Snucl/Scell ratio tended to result in a higher cell-nucleus dose. One possible explanation is that reducing Snucl/Scell increases the relative cytoplasmic volume, thereby placing more boron within the range of BNCT-generated charged particles that reach the nucleus.
D isoE evaluation
Based on the mechanisms discussed in the previous section (4.1, Sensitivity of SMK Model Parameters to Boron Uptake Ratio and Cellular Morphologic Parameters), we evaluated the actual DisoE (Fig. 6) and observed a trend similar to the changes in κ shown in Fig. 5. This suggests that κ is the dominant factor influencing DisoE.
For instance, at a depth of 2 cm, DisoE values span ~7.4–19.0 Gy (Ccell/Cextra = 0), 10.5–20.3 Gy (Ccell/Cextra = 0.1), 17.9–21.7 Gy (Ccell/Cextra = 0.5), 22.0–23.0 Gy (Ccell/Cextra = 1) and 21.6–32.7 Gy (Ccell/Cextra = 3.5).
Comprehensively, even with the same macroscale boron concentration (60 ppm), the estimated DisoE can vary widely between ~7.4 and 32.7 Gy depending on the microscopic parameters. This wide range highlights the significance of microscopic factors that cannot be captured by evaluations based solely on macroscale concentrations.
Limitations and future work
It is important to emphasize that the conclusions of this study were drawn under the assumption that a macroscopic boron concentration has already been observed and remains fixed. So, our findings do not suggest that increasing the uptake ratio itself lacks efficacy.
Our cell model is simple compared to actual structures. In reality, cell systems are much more complex, incorporating necrotic regions and vascular networks. A mesh-type cell model could be desirable in future system developments to better represent these complexities [46]. On the other hand, making the model overly complex significantly increases the workload, so it is important to consider a balance between the detail of the model and the effort required to build and use it.
In this study, the dosimetric evaluation is conducted using a homogeneous cell model constructed on the basis of cellular structural parameters. However, actual tumor tissue is heterogeneous: cellular structures vary by location within the tumor, and cells with diverse morphologies are mixed. Therefore, this model which uniformly assumes an identical cell shape for all cells may not faithfully reflect real conditions. This is a limitation of models that rely on simplifications.
Based on the validation results for the accuracy of the interpolation method, it should be noted that when using LISMEC, especially for Snucl/Scell = 0.1–0.2, the error is +15%∽21%. The increased error in the Snucl/Scell range of 0.1–0.2 was considered to be largely attributable to the linear interpolation of the nuclear-localization results.
The findings revealed in this study emphasize that, in BNCT, dose assessments can vary significantly not only with changes in macroscopic boron concentration, but also with boron distribution at the cellular level and morphological features. However, to apply these insights in clinical practice, it will be necessary to develop methods for accurately measuring and estimating microscale boron dynamics and cellular morphological parameters.
Toward clinical implementation, we will first validate the concordance between experimentally measured survival rates and LISMEC predictions. Once confirmed, we will enhance usability. At present, the parameters determined by LISMEC are applied uniformly across the entire dose distribution; in future iterations, we aim to enable region-specific parameterization. In the future, we aim to support not only tumor assessment but also the evaluation of radiation-induced adverse events in normal tissues.
In experimental validation, we plan to conduct in vitro studies. Although in vivo experiments more closely approximate actual conditions, it is difficult to impose the controlled conditions required to test the validity of our system parameters. Therefore, we chose an in vitro design. Validation will be based on the concordance between the surviving fraction predicted by LISMEC and that measured in irradiation experiments. To assess the boron-distribution parameters, we will use multiple boron agents with distinct intracellular distributions—BPA (boronophenylalanine), boric acid and BSH (sodium borocaptate). To assess the cell-structure parameters, we will employ cells with differing structural characteristics. The resulting survival data will be compared with LISMEC-predicted surviving fractions to evaluate the model’s validity.
In this study, we computed DisoE with LISMEC in the radiation field at KUR; however, the precomputed data required by LISMEC are applicable to other neutron sources as well. By performing a macroscale radiation-transport simulation for the neutron source of interest and subsequently applying LISMEC to the simulation results, DisoE can be computed using the same procedure. Therefore, the precomputed data developed in this work are generalizable.
This study clearly demonstrates the significant impact of microscopic parameters on the photon isoeffective dose (DisoE) in BNCT dosimetry, highlighting variations exceeding a factor of 4 under identical macroscopic boron concentrations. By utilizing LISMEC, microscale parameter analyses, which would conventionally require cumulative computational efforts spanning several months to potentially years if performed individually, can now be executed within seconds. In more detail, by using LISMEC, microdosimetry calculations that previously required 10 minutes per condition can now be completed in 1 second (600-fold speedup per condition). The precomputed database and linear interpolation approach employed by LISMEC provide rapid and efficient retrieval of the necessary microdosimetric parameters, greatly enhancing practical applicability in clinical settings. This evaluation framework, integrating the SMK model with LISMEC, facilitates rapid, flexible and accurate dose calculations, opening the way for personalized BNCT treatment planning.
ACKNOWLEDGEMENTS
The authors express their gratitude to Dr T. Kai and Dr T. Furuta for their valuable guidance on PHITS. The authors gratefully acknowledge Mr J. Prateepkaew for providing data on the KUR collimator. The authors used artificial intelligence tools for language editing, translation and writing code for graph plotting in the preparation of this manuscript.
Contributor Information
Takafumi Shigehira, Particle Radiation Oncology Research Center, Institute for Integrated Radiation and Nuclear Science, Kyoto University, 2-1010, Asashiro-nishi, Kumatori-cho, Sennan-gun, Osaka 590-0494, Japan.
Tubasa Watanabe, Particle Radiation Oncology Research Center, Institute for Integrated Radiation and Nuclear Science, Kyoto University, 2-1010, Asashiro-nishi, Kumatori-cho, Sennan-gun, Osaka 590-0494, Japan.
Minoru Suzuki, Particle Radiation Oncology Research Center, Institute for Integrated Radiation and Nuclear Science, Kyoto University, 2-1010, Asashiro-nishi, Kumatori-cho, Sennan-gun, Osaka 590-0494, Japan.
Yuho Hirata, Research Group for Radiation Transport Analysis, Nuclear Science and Engineering Center, Japan Atomic Energy Agency (JAEA), 2-4 Shirakata, Tokai, Ibaraki 319-1195, Japan.
Tatsuhiko Ogawa, Research Group for Radiation Transport Analysis, Nuclear Science and Engineering Center, Japan Atomic Energy Agency (JAEA), 2-4 Shirakata, Tokai, Ibaraki 319-1195, Japan.
Atsushi Fujimura, Department of Cellular Physiology, Neutron Therapy Research Center, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences, 2-5-1 Shikata-cho, Kita-ku, Okayama 700-8558, Japan.
Yoshinori Sakurai, Particle Radiation Oncology Research Center, Institute for Integrated Radiation and Nuclear Science, Kyoto University, 2-1010, Asashiro-nishi, Kumatori-cho, Sennan-gun, Osaka 590-0494, Japan.
Tatsuhiko Sato, Research Group for Radiation Transport Analysis, Nuclear Science and Engineering Center, Japan Atomic Energy Agency (JAEA), 2-4 Shirakata, Tokai, Ibaraki 319-1195, Japan.
FUNDING
This work was supported by JSPS KAKENHI 23K21426 and 23K27554.
REFERENCES
- 1. Suzuki M. Boron neutron capture therapy (BNCT): a unique role in radiotherapy with a view to entering the accelerator-based BNCT era. Int J Clin Oncol 2020;25:43–50. 10.1007/s10147-019-01480-4. [DOI] [PubMed] [Google Scholar]
- 2. Matsumura A, Asano T, Hirose K et al. Initiatives toward clinical boron neutron capture therapy in Japan. Cancer Biother Radiopharm 2023;38:201–7. 10.1089/cbr.2022.0056. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Takeno S, Yoshino Y, Aihara T et al. Preliminary outcomes of boron neutron capture therapy for head and neck cancers as a treatment covered by public health insurance system in Japan: real-world experiences over a 2-year period. Cancer Med 2024;13:e7250. 10.1002/cam4.7250. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Pagar KR, Khandbahale SV. A review on novel drug delivery system: a recent trend. Asian J Pharm Tech 2019;9:135–40. 10.5958/2231-5713.2019.00023.0. [DOI] [Google Scholar]
- 5. Neshasteh-Riz A, Mahmoud Pashazadeh A, Mahdavi SR. Relative biological effectiveness (RBE) of (131)I radiation relative to (60)Co gamma rays. Cell J 2013;15:224–9. [PMC free article] [PubMed] [Google Scholar]
- 6. Hu N, Suzuki M, Masunaga SI et al. Experimentally determined relative biological effectiveness of cyclotron-based epithermal neutrons designed for clinical BNCT: in vitro study. J Radiat Res 2023;64:811–5. 10.1093/jrr/rrad056. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Ono K. An analysis of the structure of the compound biological effectiveness factor. J Radiat Res 2016;57:i83–9. 10.1093/jrr/rrw022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Iwasaki R, Yoshikawa R, Umeno R et al. The effects of BPA-BNCT on normal bone: determination of the CBE value in mice. J Radiat Res 2023;64:795–803. 10.1093/jrr/rrad054. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Igaki H, Murakami N, Nakamura S et al. Scalp angiosarcoma treated with linear accelerator-based boron neutron capture therapy: a report of two patients. Clin Transl Radiat Oncol 2022;33:128–33. 10.1016/j.ctro.2022.02.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Hattori Y, Andoh T, Kawabata S et al. Proposal of recommended experimental protocols for in vitro and in vivo evaluation methods of boron agents for neutron capture therapy. J Radiat Res 2023;64:859–69. 10.1093/jrr/rrad064. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Takeno S, Tanaka H, Ono K et al. Analysis of boron neutron capture reaction sensitivity using Monte Carlo simulation and proposal of a new dosimetry index in boron neutron capture therapy. J Radiat Res 2022;63:780–91. 10.1093/jrr/rrac038. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Sato T, Masunaga SI, Kumada H et al. Depth distributions of RBE-weighted dose and photon-isoeffective dose for boron neutron capture therapy. Radiat Prot Dosim 2019;183:247–50. 10.1093/rpd/ncy235. [DOI] [PubMed] [Google Scholar]
- 13. Sato T, Masunaga SI, Kumada H et al. Microdosimetric modeling of biological effectiveness for boron neutron capture therapy considering intra- and intercellular heterogeneity in 10B distribution. Sci Rep 2018;8:988. 10.1038/s41598-017-18871-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Sato T, Watanabe R, Niita K. Development of a calculation method for estimating specific energy distribution in complex radiation fields. Radiat Prot Dosim 2006;122:41–5. 10.1093/rpd/ncl407. [DOI] [PubMed] [Google Scholar]
- 15. González SJ, Santa Cruz GA. The photon-isoeffective dose in boron neutron capture therapy. Radiat Res 2012;178:609–21. 10.1667/RR2944.1. [DOI] [PubMed] [Google Scholar]
- 16. International Atomic Energy Agency . Advances in Boron Neutron Capture Therapy. Vienna: International Atomic Energy Agency, 2023. [Google Scholar]
- 17. Sato T, Kase Y, Watanabe R et al. Biological dose estimation for charged-particle therapy using an improved PHITS code coupled with a microdosimetric kinetic model. Radiat Res 2009;171:107–17. 10.1667/RR1510.1. [DOI] [PubMed] [Google Scholar]
- 18. Sato T, Furusawa Y. Cell survival fraction estimation based on the probability densities of domain and cell nucleus specific energies using improved microdosimetric kinetic models. Radiat Res 2012;178:341–56. 10.1667/rr2842.1. Erratum in: Radiat Res 2012;178:622. [DOI] [PubMed] [Google Scholar]
- 19. Sato T, Hashimoto S, Inaniwa T et al. Implementation of simplified stochastic microdosimetric kinetic models into PHITS for application to radiation treatment planning. Int J Radiat Biol 2021;97:1450–60. 10.1080/09553002.2021.1956003. [DOI] [PubMed] [Google Scholar]
- 20. Inaniwa T, Kanematsu N. Adaptation of stochastic microdosimetric kinetic model for charged-particle therapy treatment planning. Phys Med Biol 2018;63:095011. 10.1088/1361-6560/aabede. [DOI] [PubMed] [Google Scholar]
- 21. Sato T, Furuta T, Sasaki H et al. Establishment of a practical methodology for evaluating equieffective dose of individual patients based on RT-PHITS. EJNMMI Phys 2025;12:28. 10.1186/s40658-025-00743-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Sato T, Iwamoto Y, Hashimoto S et al. Recent improvements of the particle and heavy ion transport code system–PHITS version 3.33. J Nucl Sci Technol 2024;61:127–35. 10.1080/00223131.2023.2275736. [DOI] [Google Scholar]
- 23. Kumada H, Yamamoto K, Matsumura A et al. Verification of the computational dosimetry system in JAERI (JCDS) for boron neutron capture therapy. Phys Med Biol 2004;49:3353–65. 10.1088/0031-9155/49/15/003. [DOI] [PubMed] [Google Scholar]
- 24. Menzel HG, Clement C, DeLuca P. ICRP publication 110. Realistic reference phantoms: an ICRP/ICRU joint effort. A report of adult reference computational phantoms. Ann ICRP 2009;39:1–164. 10.1016/j.icrp.2009.09.001. Erratum in: Ann ICRP 2009;39:165. [DOI] [PubMed] [Google Scholar]
- 25. Santa Cruz GA, Zamenhof RG. The microdosimetry of the B-10 reaction in boron neutron capture therapy: a new generalized theory. Radiat Res 2004;162:702–10. 10.1667/RR3257. [DOI] [PubMed] [Google Scholar]
- 26. Masunaga S, Sakurai Y, Tanaka H et al. The dependency of compound biological effectiveness factors on the type and the concentration of administered neutron capture agents in boron neutron capture therapy. Springerplus 2014;3:128. 10.1186/2193-1801-3-128. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Hirayama H, Namito Y, Bielajew AF et al. The EGS5 Code System. SLAC-R-730 (2005) and KEK Report 2005–8. Department Associate in the Radiation Physics Group (retired) Radiation Protection Department Stanford Linear Accelerator Center Stanford University 2575 Sand Hill Road, Menlo Park, CA 94025, USA; Radiation Science Center Advanced Research Laboratory High Energy Accelerator Research Organization (KEK) 1-1 Oho Tsukuba-shi Ibaraki-ken 305-0801 Japan.
- 28. International Commission on Radiation Units and Measurements (ICRU) . Microdosimetry, ICRU REPORT 36. International Commission on Radiation Units and Measurements. Bethesda, MD (United States), 1993.
- 29.LATB Staff, Thermo Fisher Scientific Inc, Thermo Fisher Scientific 168 Third Avenue Waltham, MA 02451, USA, 2 January 2026. https://www.thermofisher.com/blog/learning-at-the-bench/cell-image-analyzing2/.
- 30. Takeno S, Tanaka H, Watanabe T et al. Quantitative autoradiography in boron neutron capture therapy considering the particle ranges in the samples. Phys Med 2021;82:306–20. 10.1016/j.ejmp.2021.02.012. [DOI] [PubMed] [Google Scholar]
- 31. Kodaira S, Morokoshi Y, Li HK et al. Evidence of local concentration of α-particles from 211At-labeled antibodies in liver metastasis tissue. J Nucl Med 2019;60:497–501. 10.2967/jnumed.118.216853. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Jiang X, Li H, Devan SP et al. MR cell size imaging with temporal diffusion spectroscopy. Magn Reson Imaging 2021;77:109–23. 10.1016/j.mri.2020.12.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Jiang X, Harkins KD, Xie J et al. Joint estimation of compartment-specific T2 relaxation and tumor microstructure using multi-TE IMPULSED MRI. Magn Reson Med 2025;93:96–107. 10.1002/mrm.30254. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Choi BK, Katoch N, Park JA et al. Measurement of extracellular volume fraction using magnetic resonance-based conductivity tensor imaging. Front Physiol 2023;14:1132911. 10.3389/fphys.2023.1132911. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Ono K, Tanaka H, Tamari Y et al. Proposal for determining absolute biological effectiveness of boron neutron capture therapy-the effect of 10B(n,α)7Li dose can be predicted from the nucleocytoplasmic ratio or the cell size. J Radiat Res 2019;60:29–36. 10.1093/jrr/rry080. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Kirihata M, Hattori Y, DOJINDO LABORATORIES, Kumamoto Techno Research Park, 2025-5 Tabaru, Mashiki-machi, Kamimashikigun, Kumamoto 861-2202, Japan, 2 January 2026. https://www.dojindo.co.jp/letterj/182/review/01.html.
- 37. Mendaçolli PJ, Brianezi G, Schmitt JV et al. Nuclear morphometry and chromatin textural characteristics of basal cell carcinoma. An Bras Dermatol 2015;90:874–8. 10.1590/abd1806-4841.20154076. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Watanabe T, Kimijima I, Furukawa Y et al. Nuclear morphometry and clinicopathological variables in breast cancer. J Jpn Soc Clin Cytol 1993;32:860–6. 10.5795/jjscc.32.860. [DOI] [Google Scholar]
- 39. Capala J, Makar MS, Coderre JA. Accumulation of boron in malignant and normal cells incubated in vitro with boronophenylalanine, mercaptoborane or boric acid. Radiat Res 1996;146:554–60. 10.2307/3579556. [DOI] [PubMed] [Google Scholar]
- 40. Sakurai Y, Kobayashi T. Spectrum evaluation at the filter-modified neutron irradiation field for neutron capture therapy in Kyoto University Research Reactor. Nucl Instrum Methods Phys Res 2004;531:585–95. 10.1016/j.nima.2004.05.084. [DOI] [Google Scholar]
- 41. Prateepkaew J, Takata T, Tanaka H et al. Experimental verification of liquid multilayer spectrometer in neutron irradiation field for boron neutron capture therapy. Nucl Instrum Methods Phys Res 2024b;1065:169534. 10.1016/j.nima.2024.169534. [DOI] [Google Scholar]
- 42.STELLA PHARMA, Borofalan, Interview Form, STELLA PHARMA, ORIX Koraibashi Building, 3-2-7 Koraibashi, Chuo-ku, Osaka 541-0043, Japan, 2 January 2026. https://image.packageinsert.jp/pdf.php?mode=1=4291453A1027.
- 43. Fukuda H. Boron neutron capture therapy (BNCT) for cutaneous malignant melanoma using 10B-p-boronophenylalanine (BPA) with special reference to the radiobiological basis and clinical results. Cells 2021;10:2881. 10.3390/cells10112881. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Watanabe T, Hattori Y, Ohta Y et al. Comparison of the pharmacokinetics between L-BPA and L-FBPA using the same administration dose and protocol: a validation study for the theranostic approach using [18F]-L-FBPA positron emission tomography in boron neutron capture therapy. BMC Cancer 2016; 16:859. 10.1186/s12885-016-2913-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45. Masunaga S, Sakurai Y, Tanaka H et al. The dependency of compound biological effectiveness factors on the type and the concentration of administered neutron capture agents in boron neutron capture therapy. Springerplus. 2014;3:128. 10.1186/2193-1801-3-128. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46. Wang Y, Kong D, Gao H et al. Multiple mesh-type real human cell models for dosimetric application coupled with Monte Carlo simulations. Radiat Res 2023;200:176–87. 10.1667/RADE-23-00020.1. [DOI] [PubMed] [Google Scholar]