Abstract
Measures of membrane capacitance offer insight into a variety of cellular processes. Unfortunately, popular methodologies rely on model simplifications that sensitize them to interference from inevitable changes in resistive components of the traditional cell-clamp model. Here I report on a novel method to measure membrane capacitance that disposes of the usual simplifications and assumptions, yet is immune to such interference and works on the millisecond timescale. It is based on the exact empirical determination of the elusive partial derivative, ∂Y/∂Cm, which heretofore had been approximated. Furthermore, I illustrate how this method extends to the vesicle fusion problem by permitting the determination of ∂Yv/∂Cv, thereby providing estimates of fusion pore conductance and vesicle capacitance. Finally, I provide simulation examples and physiological examples of how the method can be used to study processes that are routinely interrogated by measures of membrane capacitance.
INTRODUCTION
Biological cells are typically modeled as simple parallel combinations of an electrical resistance and capacitance. Measures of membrane capacitance (Cm) based on this model under voltage clamp provide an informative view of such varied processes as fertilization (Jaffe and Schlichter, 1985), ionic channel gating (Fernandez et al., 1982), synaptic vesicle fusion (Neher and Marty, 1982), gap junctional coupling (Santos-Sacchi, 1991a), and receptor cell activity (Zhao and Santos-Sacchi, 1999).
A powerful feature of capacitance measures is the ability to monitor interactions between membrane-bound compartments, e.g., fusion of vesicles with the plasma membrane or cell coupling via gap junctions (Neher and Marty, 1982; Joshi and Fernandez, 1988; Santos-Sacchi, 1991a; Zhao and Santos-Sacchi, 1998), because these measures provide an estimate of membrane surface area (Fricke, 1925; Cole, 1971). Indeed, electrical measures of vesicular fusion (Alvarez De Toledo et al., 1993) have played a significant role in the development of one of the most controversial issues facing neurobiology today, namely, the occurrence of kiss-and-run synaptic vesicle transmitter release (Gandhi and Stevens, 2003; Aravanis et al., 2003). Single-frequency admittance analysis of membrane fusion, pioneered by Neher and Marty (1982), has been used by a number of investigators to gauge the time course of vesicular membrane fusion by monitoring the changes in the conductance of the fusion pore that initiates membrane fusion (Breckenridge and Almers, 1987; Alvarez De Toledo et al., 1993; Debus and Lindau, 2000; Klyachko and Jackson, 2002). Estimates of the pore's initial conductance hover near single-channel conductance values (Breckenridge and Almers, 1987; Lollike et al., 1995).
Arguably the most widely used method to measure membrane capacitance (Cm) is the phase-angle method, variously termed, but all employing approaches to extract the current at a particular phase angle arising from small changes in Cm (Neher and Marty, 1982; Gillis, 1995). Such AC methods ostensibly provide excellent time resolution and sensitivity. Multifrequency approaches, which we and others have used (Fernandez et al., 1982; Santos-Sacchi, 1989; Zhao and Santos-Sacchi, 1999), are less in vogue, but offer potential protection from changes in admittance that can plague single-sine methods. Implementations of step, single-, and dual-sine techniques are available in a few software packages (jClamp, www.SciSoftCo.com; pClamp, www.Axon.com; Pulse Control, www.Instrutech.com, chroma.med.miami.edu/cap).
One of the key problems that challenges the validity of Cm measures, and thus our understanding of those cellular processes that are interrogated by Cm measurements, is the interference of those measures by changes in either patch electrode resistance (Rs) or cell membrane resistance (Rm), the latter often a natural corollary of those cellular processes (Gillis, 1995; Debus et al., 1995). An extensively discussed topic of the single-sine camp concerns the determination of the proper phase angle to utilize in measuring small changes in membrane capacitance. There are various mechanisms to determine an appropriate phase angle, but all estimates of capacitance employ simplifying assumptions and requirements, usually devised to limit the effects of uncontrollable Rs and Rm changes on Cm measures. Indeed, the traditional methodology of single-sine admittance analysis suffers from a number of assumptions and simplifications, especially those methods used to extract vesicle capacitance and conductance, and the fusion pore conductance. These assumptions cannot be met in all cases, and this quandary is clearly indicated by the posthoc manipulations of data that are often required (Lindau, 1991; Ratinov et al., 1998). In detailing these assumptions, I refer to the traditional electrical circuit models that are used to evaluate admittance data (Fig. 1). A nearly universal assumption is that Rm ≫ Rs; additionally, for fusion vesicle analysis, Rv ≫ Rp, and in some cases Rv = ∞. Below I illustrate through circuit analysis and modeling how these assumptions can lead to errors, and provide a method, based on the exact empirical identification of the partial derivatives ∂Y/∂Cm and ∂Yv/∂Cv, that offers solutions without compromising assumptions. The approach is referred to as the eCm method.
Theory
The phase of the capacitive current components evoked by a sinusoidal command voltage (V) across the simple models presented in Fig. 1 can be used to arrive at measures of the capacitive components themselves. Once the capacitive component values are obtained, the resistive component values are easily deduced. This can be appreciated from the following evaluation, which first deals with the model without a fusion vesicle and without capacitance compensation components (Fig. 1 A).
Two standard conditions must be met for successful use of the methods detailed here: 1), the patch-clamp system transfer function (which derives from all equipment that contributes to shaping the output current) must be corrected for. This is a trivial matter, where a pure resistor, say 1 MΩ, is used at the input of the headstage to determine the correction phase and magnitude at each frequency (see Gillis, 1995 for details); and 2), the pipette stray capacitance must be neutralized before establishing whole-cell configuration.
Absent vesicle fusion
The current measurable under voltage clamp (Fig. 1) is that through Rs, namely, IRs. According to Ohm's law, IRs is defined as the product of the clamp voltage (V) and input admittance (Y) of the circuit,
(1) |
where AY and BY are the real and imaginary components of the admittance. Closer inspection of the circuit reveals that IRs is the sum of a resistive and a capacitive component induced by only that portion (Vm) of the clamp voltage (V), which drops across the cell membrane (Rm ‖ Cm),
(2) |
These equations show that IRs can be represented as a frequency (ω = 2πf) dependent complex number,
(3) |
where Ym is the membrane admittance, AIrs is the real component, and BIrs is the imaginary component of IRs. Plainly, this signifies that the phase relationship between capacitive (ICm) and resistive (IRm) membrane current components is 90°. It is important to realize that the magnitude of the series resistance (Rs), or its existence (i.e., Rs = 0), is irrelevant, and will not affect the phase relationship between ICm and IRm. Of course, for Rs > 0, the phase of IRs, relative to that of V, will vary as a result of any circuit parameter change, because the phase of IRs is dependent on that of Vm,
(4) |
where τclamp, the actual clamp time constant, is
(5) |
Viewed relative to the phase of the command voltage (V), the capacitive current phase angle, Φ, will always reside at
(6) |
Thus, in the absence of Rs (Rs = 0), the angles of Vm and V are identical, and Φ will be orthogonal to the angle of V. However, as pointed out by Neher and Marty (1982), the introduction of Rs causes Φ to shift away from orthogonality with respect to the applied command voltage (V). The phase shift, θ, in the angle of IRs that accompanies changes in the circuit parameters is defined as
(7) |
and thus
(8) |
Although it is apparent from Eq. 2 that the magnitude of the current at Φ, namely, ICm, can be used to extract Cm
(9) |
(10) |
neither absolute angle, θ nor Φ, can be determined from direct manipulation of the empirically obtained input admittance at any one frequency. Thus, in the face of changing circuit parameters, estimating Cm by directly measuring raw current at Φ appears unfeasible. Nevertheless, inspection of the current at Φ can be achieved by operating on the imaginary component of IRs, whose phase follows those phase shifts that result from parameter changes.
(11) |
Thus, the imaginary component of Y (or IRs) can be used to find Cm. To obtain direct estimates of Cm from one needs to correct with a noncomplex gain factor (Hc) that accounts for the relationship between inevitable changes in Rs, Rm, and Cm that occur during the course of an experiment.
(12) |
where, after substitution and rearrangement,
(13) |
and
(14) |
Note that τcell is the membrane time constant, and most importantly,
(15) |
Thus, to extract Cm from we must precisely evaluate the partial derivative, ∂Y/∂Cm. This is one of the enigmas facing cell capacitance aficionados.
There are two methods that are routinely used to provide estimates of ∂Y/∂Cm and its phase angle; both involve dithering of circuit parameters, and typically require hardware modifications and/or manipulations (Fidler and Fernandez, 1989; Joshi and Fernandez, 1988; Neher and Marty, 1982). Series resistance dithering and capacitance compensation dithering each have shortcomings that have been discussed in detail (Joshi and Fernandez, 1988). Ultimately, these shortcomings stem from the untoward effects of series resistance, because as the circuit of Fig. 1 A approaches a simple parallel Rm Cm configuration, and the estimates ultimately provide precise solutions. Regrettably, we are rarely so lucky during our experimental manipulations of cells to guarantee such relationships (Debus et al., 1995). Can this problem be resolved? That is, can the true magnitude and phase angle of ∂Y/∂Cm be obtained, so that estimates of Cm are impervious to cell and clamp parameter changes? Furthermore, can this be accomplished without hardware additions and manipulations? Actually, it is a simple task, requiring, in addition to measures of the determination of the partial derivative, ∂Y/∂ω or correspondingly, ∂IRs/∂ω. As we shall see, such a determination ensures accurate measures of membrane capacitance.
(16) |
where the magnitude (αω) and phase (βω) are
(17) |
(18) |
Significantly, βω, the phase angle of ∂IRs/∂ω and ∂Y/∂ω, is identical to that of ∂Y/∂Cm, and also to the angle of ∂Y/∂Rm + 90°. This can be appreciated immediately by inspection of the partial derivatives.
(19) |
(20) |
(21) |
where
(22) |
It is this equivalence between the angles of ∂Y/∂ω and ∂Y/∂Cm that forms the basis of the new eCm method of capacitance measurement that I report here. Given the angle, βω, and the real and imaginary components of Y, the two time constants, τclamp and τcell, can be found. Each has two solutions, one of which will be realizable (positive) depending on the circuit parameters. For subsequent calculations the positive values must be used.
(23) |
(24) |
where
(25a) |
(25b) |
Finally, with τclamp and τcell in hand, Hc (Eqs. 13 and 15) can be obtained, thus revealing the exact solution for all clamp parameters, Rs, Rm, and Cm.
(26) |
(27) |
(28) |
For completeness, the pertinent partial derivatives are provided in polar form (all parameters obtainable through a combination of measurement (that of Y and βω; see following) and calculation (as above)),
(29) |
(30) |
The exact determination of βω
Several practical approaches to obtain βω via frequency manipulations are available. Ordinarily, the empirical determination of an angle of a partial derivative with respect to angular frequency, such as βω, is found by introducing a second voltage sinusoid (ω1), in addition to the primary, ω0, at an infinitesimal interval of dω, which must be approximated with Δω. The angle of the resulting difference current will closely approximate βω. However, the determination of βω using digital techniques necessitates that the two frequencies be no closer than the minimum frequency interval, ωmin. This interval is determined by the number of points (npts) of the stimulus (ωmin = 1/(dt npts), where dt = sample clock) (see Oppenheim and Schafer, 1975). Unfortunately, the rather large size of ωmin offered by a stimulus composed of, say, 128 points at 4 μs clock, will provide for a poor approximation of dω by Δω (fmin would be 1953.1 Hz). Of course, npts can be increased such that ωmin would be very small and permit a reasonable approximation of βω, but time resolution suffers. If high time resolution is not an issue, or if discrete determinations (as in “pause and dither” approaches) are acceptable, then determining the angle of the difference current at the primary frequency (ω0) and secondary frequency (ω1) at a very small interval will give βω.
Alternatively, a nondiscrete hardware implementation of this technique could easily allow dω to be approached; furthermore, the magnitude of the second frequency could be quite small, because only the phase would be required. In fact, the stimulus could very well resemble a single sinusoid, thus limiting the cell's stimulus exposure.
Notwithstanding these approaches to the determination of βω, our overriding intent is to obtain a method for Cm measurement without additional hardware or hardware manipulations, and to correct for parameter changes on a point-by-point (real time) basis. To this end, it is fortuitous that the need for an infinitesimal interval, dω, can be circumvented by acknowledging that for the linear component model of Fig. 1 a, Rs or Rm may be considered frequency independent. That is, we may use either of the derivations of Rs or Rm provided by Pusch and Neher, 1988 (their Eqs. 5 and 6 reproduced below); also see Lindau and Neher (1988) to obtain an exact solution of βω at ω0.
(31) |
(32) |
where b = 1 / (Rs + Rm) and n = 0,1.
Equating Rs (or Rm, either of which produce the exact same result) determined at two angular frequencies, ω0 and ω1, provides two solutions for b (b0, b1; solutions not shown), and therefore two solutions for τclamp at ω0 (obtained by utilizing Pusch and Neher's (1988) solution for Cm; their Eq. 7),
(33) |
Notably, however, either solution of or when used to derive the angle of ∂Y/∂ω, i.e., βω0, from Eq. 18 produces only one, unique result, namely
(34) |
where the subscripts denote the real (A) and imaginary (B) admittance components at any two angular frequencies, ω0 and ω1. It should be pointed out that whereas two frequencies are required to obtain βω, the value itself is derived from ∂Y/∂ω0, whose location is exactly at ω0.
(35) |
(36) |
that is, all three pieces of information, BY, AY, and βω that are used to measure Cm theoretically derive from single (ω0), not dual, frequency inspection.
Present vesicle fusion
The fusion of independent membrane-bound compartments (vesicle to the plasma membrane or gap junctional coupling between two cells) can be evaluated with input capacitance measures, and modeled to obtain component values of the applicable circuit (Fig. 1 B) (Zhao and Santos-Sacchi, 1998; Bigiani and Roper, 1995; Santos-Sacchi, 1991a). Below, I focus on vesicular fusion. If experiments are performed using the cell-attached mode it must be determined that the cell's input admittance is far greater than the patch admittance. Additionally, data must be corrected posthoc after determination of stray pipette capacitance (perhaps using the Sylgard bead-sealing technique (Sakmann and Neher, 1995)), which may be greater than patch capacitance. Thus, the application of this method under cell-attached mode is not straightforward. Nevertheless, in that case, or if vesicle fusion is monitored under whole-cell mode, the following applies; the subscript m denotes either patch or whole-cell characteristics. The input admittance before (Yb) and after (Ya) vesicle fusion differs.
(37) |
(38) |
where
(39) |
Typically, access to the circuit parameters defining the vesicle admittance (Yv: Rp, Rv, and Cv; see Fig. 1 B) is obtained experimentally by making an inescapable assumption – that at the onset of fusion Rs, Rm, and Cm remain unchanged. Given the model, without this assumption estimates cannot be obtained with confidence.
The original and still common method to calculate vesicle parameters, including fusion pore conductance, employs capacitance compensation to estimate vesicle admittance (Breckenridge and Almers, 1987; Lindau, 1991; Ales et al., 1999). In this case, membrane admittance before fusion is removed by subtracting the Ym baseline determined by capacitance compensation dithering (Neher and Marty, 1982); the subsequent change in admittance after vesicle fusion is used to calculate vesicle components based on a simplified vesicle circuit model of Cv in series with Rp (Breckenridge and Almers, 1987). We can employ the eCm method to arrive at Yv, in a similar manner. First, Ym is extracted from Yb, the exact value of Rs having been obtained by analysis (see above) before fusion.
(40) |
Consequently, the vesicle admittance, Yv, which includes the fusion pore, can be calculated (by substitution of Eq. 40 into Eq. 38) as a function of admittances measured before and after fusion,
(41) |
From inspection of the vesicle circuit (Fig. 1 B), it is clear that
(42) |
that is, we have the analogous circuit conditions that we analyzed above in the absence of fusion. The parameters Rp, Rv, and Cv simply and, respectively, replace Rs, Rm, and Cm. Using the same analysis logic as above, the magnitude and angle of ∂Yv/∂Cv is obtained, thus enabling the solution of all vesicle circuit parameters. The following equations detail fusion event analysis.
(43) |
(44) |
(45) |
where
(46) |
(47) |
and
(48) |
Simulations and biological cell examples
I provide several approaches to evaluate the eCm method of parameter estimation, including mathematical modeling, physical modeling, and tests on biological cells. It is important to realize that tests on biological cells may provide important information on real-world capabilities, but cannot be used to truly verify the validity of an approach. For example, it is impossible to know if a certain time course of Cm change is truly occurring in cells, unless, of course, confirmation is obtained through other measures. However, physical and mathematical model evaluations can confirm validity of an approach, and do provide confidence that the results from biological cells are sound. In the first set of evaluations, MatLab (Natick, MA) was used to model the circuit admittance and implement eCm strictly in the frequency domain. In the following mathematical and physical simulations, measures were made at an f0 of 390.6 Hz and f1 of twice that, delivered simultaneously, unless otherwise noted. Amplitudes at each frequency were set equal, at 10–20 mV peak. Fig. 2 presents a simple whole-cell mathematical model simulation where the circuit parameters are stepped to evaluate component interactions on Cm measures; comparisons are made of calculated component values using either the eCm method or other approaches that contain approximations. Clearly, eCm is superior, and is impervious to fluctuations in circuit component values. Fig. 3 presents another mathematical simulation in Matlab where vesicle fusion under whole-cell mode is evaluated, and compared with the standard methodology (Breckenridge and Almers, 1987; Lollike et al., 1995). Both methods are excellent even in the face of substantial Rs changes, but even with changes in vesicle resistance the eCm method produces an exact reconstruction of the model parameter changes. Although these model data highlight the exact nature of the eCm method, more demanding modeling efforts show the method's strengths. For the tests that follow, capacitance measures were obtained with eCm as implemented in the software package jClamp using a time resolution of 2.56 ms (i.e., f0 of 390.6 Hz), unless otherwise noted.
Fig. 4 shows the results of a physical implementation of the simple model in Fig. 1 A under voltage clamp during a large step in capacitance, that was evoked by physically dislodging an additional capacitor that was placed parallel with Cm. Simultaneous measures of Rs and Rm remain essentially unperturbed. To evaluate large changes in Rm, this parallel approach did not work because of associated parasitic capacitances of the large resistances involved. Instead, Fig. 5 shows two separate measures where all circuit and measurement details remained exactly the same except that either a 50-MΩ or 800-MΩ Rm resistor was soldered into place. Again, there are minimal changes in Cm measures. Fig. 6 shows the effects of maladjustments of the amplifier's fast capacitive compensation on parameter estimates. The model cell was measured over 10 seconds at a resolution of 2.56 ms while changes in compensation were made. Stray Rs capacitance of ±0.15 pF had little effect on Cm (±0.4%) or Rm (<±0.01%) parameter estimation. Rs estimates varied ±1.8%.
Finally, in Figs. 7 and 8 biological cells are investigated. Fig. 7 shows measures of outer hair cell (OHC) capacitance obtained with jClamp using a time resolution of 512 μs. Detrending of data before analysis was performed (Santos-Sacchi, 2002). The rapid relaxation in capacitance during a step voltage change is due to the shift of the cell's nonlinear Q-V function along the voltage axis (Santos-Sacchi et al., 1998). In that article, we confirmed through alternative measures that the time course of Cm relaxation corresponded to the shift. The shift signifies the very rapid change in the gain of OHC electromotility, a process that likely drives the mammalian cochlear amplifier (Santos-Sacchi, 2003). The solid shaded line is a triple exponential fit to the data. The three time constants are 1.04, 5.96, and 36 ms. This time course is over an order of magnitude faster than has been previously reported (Santos-Sacchi et al., 1998). The substantial noise levels in this trace derive from the huge nonlinear capacitance (>20 pF) and low input resistance (<300 MΩ) of the OHC. For comparison, Fig. 8 shows capacitance measures in a HEK cell transfected with prestin (Meltzer and Santos-Sacchi, 2001), where Cm noise was 5.5 fF (rms; 0.256 ms time resolution (f0: 390.6 Hz); see discussion on noise effects below). This cell possesses a nonlinear Qmax (by fitting the Q-V curve with a two-state Boltzmann function (Santos-Sacchi, 1991b)) of 63.7 fC, giving rise to a peak voltage-dependent capacitance at −101 mV of 420 fF. The figure shows a 30-fF step in capacitance when voltage was stepped from +50 mV to +30 mV. These data indicate that the sensitivity of the eCm method is well within the range required for fusion event analysis.
DISCUSSION
Three pieces of information are required to determine the circuit parameters, Rs, Rm, and Cm of the traditional patch-clamp-cell model circuit (Fig. 1 A). The typical information obtained in a one-frequency (ω0) measurement, namely, the real and imaginary components of the admittance, is insufficient. A third piece of information is required, and is usually obtained by exploring beyond ω0, that is, by utilizing information obtained at frequencies other than ω0. One popular solution is the “sine+dc” method (Pusch and Neher, 1988; Gillis, 2000), where a third piece of information, the input resistance (Rin = Rs + Rm), is extracted based on a priori knowledge of membrane reversal potential. This is basically a measurement at ω1 = 0. Alternatively, another non-dc piece of information, the admittance at can be used to extract Rin (Santos-Sacchi et al., 1998; Santos-Sacchi, 2002). The latter approach benefits from statistical modeling to obtain an optimal result (Barnett and Misler, 1997). In this report, I have uncovered an additional piece of information at ω0, namely, βω, which can be used to determine all circuit parameters. Despite the location of βω precisely at ω0, the method does require two-frequency stimulation, and therefore analogous to the standard dual-sine method, it is reasonable to expect that optimal results could arise from further refinement (Barnett and Misler, 1997). I should also note that because βω depends on τclamp, an alternative approach would be fitting of current transients induced by step voltage commands. Of course, any method that does not measure βω synchronously with the primary frequency component will be unable to correct for changes in cell parameters on a point-by-point basis. This will limit the utility of the technique, and this limitation is inherent in all “pause and dither” techniques. The eCm method benefits from point-by-point correction because gain and angle are redetermined and applied point-by-point at the f0 rate.
Relation of the eCm method to traditional phase-tracking methods
The eCm method that I have developed here simply estimates absolute Cm by dividing the imaginary component of the input admittance (BY) at a single frequency by the absolute value of the partial derivative ∂Y/∂Cm at that same frequency. This method has an unforeseen relation to traditional phase-tracking techniques. The original phase-tracking method was devised to account for and limit the effects of resistive components on Cm measures (Neher and Marty, 1982; Joshi and Fernandez, 1988). It is a method that theoretically strives to determine Cm changes through measurement of changes in input admittance (ΔY) as a function of changes in the circuit parameter, Cm.
(49) |
(50) |
(51) |
(52) |
The latter equation is related to the nonsimplified form of Eq. 43 in Gillis (1995) and is addressed in a subsequent article of his (Gillis, 2000). However, heretofore, approximations of both βω and Hc have only been possible. The benefits and drawbacks of using particular parameter approximations have been discussed in detail (Joshi and Fernandez, 1988). Thus, phase-tracking paradigms that use estimates of the magnitude and angle of ∂Y/∂Cm are recognized as better suited to systems where Rs is stable and Rm may fluctuate (Joshi and Fernandez, 1988). However, where the resistive components have the converse behavior, the angle of ∂Y/∂Gs − 90°, namely, 2 tan−1 (BY/AY) − 90° as nicely provided by Zierler (1992), may be a better choice (Joshi and Fernandez, 1988). Unfortunately, where both resistive components simultaneously fluctuate, phase tracking is problematic. Despite these caveats, and even with the correct quantities I now provide, implementation of the phase tracking either by feeding the raw current output of the patch clamp into a hardware lock-in amplifier or a software phase detector (PSD) appears to be an inappropriate implementation of the phase analysis concept (Joshi and Fernandez, 1988; Neher and Marty, 1982; Fidler and Fernandez, 1989). That is, if the analyzed signal is Y, then in contradistinction to Eq. 52,
(53) |
Consequently, the result of evaluating Y (or IRs) directly should not provide an accurate picture of changes in Cm (i.e., ΔCm).
The surprising fact is, however, that implementations using direct evaluations of Y can, under certain conditions, provide very good estimates of absolute Cm.
It should be noted that applying capacitance and series resistance compensation at the onset of an experiment cannot be construed as providing a ΔIRs (i.e., ΔY V) output from the amplifier. This is simply a one-time subtraction that would have to be repeated, along with a new angle determination, every time a data point is collected. Additionally, only a two-element model is used by amplifiers for such compensation.
One reason for success with this direct approach would be if Rs were very small, i.e., as Rs approaches zero, and absolute values of ICm will be evaluated. Unfortunately, it is unlikely that Rs will be so low (and maintained so low) so as to approach this condition. Notwithstanding this apparent quandary, it is possible to implement phase-tracking paradigms with direct evaluations of Y and still obtain excellent estimates of Cm in the face of very large changes in Rs and Rm. Surprising, the angle returned by series resistance dithering works better than that returned by capacitance compensation dithering.
Series resistance dithering
Dithering of the series resistance, championed by Joshi and Fernandez (1988), is used to return a close approximation of βω, namely ∠∂Y/∂Gs − 90°. This approximation, however, when used with a direct analysis of Y, rather than ΔY, and point-by-point (i.e., real time) gain and angle correction is surprisingly exact (see the simulations in Fig. 2 B). The outcome of such an implementation does not suffer from the caveats normally expected with the phase-tracking approach (Joshi and Fernandez, 1988)—namely, large changes in Rm (or Rs) have no effect on the measure of Cm. The reason for this is because the angle of ∂Y/∂Gs is twice the angle of Y (Zierler, 1992), and the PSD measurements are simply returning BY,
(54) |
(55) |
Thus, whereas Eq. 53 fails to provide ΔCm estimates, simply replacing the angle with results in an exact solution for absolute Cm, as shown previously in Eq. 26,
(56) |
Any complex number can be evaluated at twice its angle to return its real component and at twice its angle −90° to return its imaginary component. Thus, in this case, it turns out that the eCm method is simply being employed.
Capacitance compensation dithering
The initial work of Neher and Marty (1982) utilized capacitance compensation dithering to obtain an angle at which to evaluate, via PSD, changes in Cm. The angle provides an excellent approximation of βω when capacitance compensation is in effect. This is shown by inspection of ∂Y/∂Cd and its angle, based on the model of Fig. 1 A, now including the capacitance compensation components Cd and Rd (in series with each other, from headstage to ground).
(57) |
(58) |
where
βω, the angle of ∂Y/∂Cm, remains unaffected by the addition of the capacitance compensation circuitry,
(59) |
thus indicating that when the clamp time constant (τclamp) and compensation time constant (τd) are equal, i.e., when compensation is in effect, the angle returned by dithering equals βω. Unlike the use of the angle returned by series resistance dithering, the use of this angle to measure Cm with direct evaluations of Y is not as accurate (Fig. 2 E). It should be noted that these analyses only apply to direct measures of Y that are corrected on a point-by-point basis for phase and gain changes, and not necessarily to the traditional use of the angle (Neher and Marty, 1982).
Validity, accuracy, and sensitivity of Cm measurement methods
There are several factors, including model validity, approximations, and noise that contribute to errors in the measurement of capacitance with sinusoidal analysis. Obviously, approximations lead to errors only if they fail; however, because approximations often remove a particular component, e.g., Rs or Rv, from the system of equations, it is impossible to gauge failure during an experiment. Thus, it is prudent to make as few simplifications and approximations as possible. The mathematical and physical modeling that I have presented clearly indicate that the eCm method benefits from this rule.
Noise, on the other hand, is an issue that affects all electrical measures of cell activity regardless of the method's validity. Even for a method that provides a valid measure of some electrical quantity in the absence of noise, the introduction of noise can prove devastating, or at least render it useless for a given task. Not surprisingly, some methods may be able to deal with system noise better than others.
The design of sinusoidal stimulus protocols has enormous impact on Cm measurement sensitivity in the face of real-world noise. Important issues such as stimulus frequency, stimulus magnitude, stray capacitance, and unexpected conductance changes have been discussed in detail (Gillis, 1995; Barnett and Misler, 1997; Thompson et al., 2001; Debus et al., 1995). Careful attention to these issues can lead to incredible sensitivity; for example, with very high-frequency stimulation and very large voltages, very good signal/noise ratios can be obtained, providing noise floors well below 1 fF (Klyachko and Jackson, 2002). One of the consequences of the equivalency between the eCm method and the phase-tracking method as typically employed, is that effects of noise on performance might be expected to be similar for each method. To evaluate issues like this, comparisons were made under varying noise conditions in model simulations.
In Fig. 9, a modeling approach is used to compare the effects of Gaussian noise levels on Cm estimate noise among a few common methods of Cm estimation. In this approach, the simple circuit of Fig. 1 A was modeled in the time domain using the RK4 method hardcoded in C within jClamp; the results were then independently analyzed by transient step analysis, single-sine analysis, dual-sine analysis, or the eCm method. Step analysis follows the method outlined in a prior publication (see appendix in Huang and Santos-Sacchi, 1993). Such an analysis is multifrequency, providing extended information similar to that obtained from “chirp” stimuli (Santos-Sacchi, 1991b; Santos-Sacchi, 1989), and wideband noise stimuli (Fernandez et al., 1982). Features of our step method included automatic clocking to reach steady state and utilization of exponential fitting after the initial onset transient, methodology recently reemphasized as important for the robust use of transient analysis for Cm measurement (Thompson et al., 2001). In our original implementation and in jClamp, a correction for errors introduced by the Rs/Rm voltage divider on Q, namely that is employed (Huang and Santos-Sacchi, 1993). To my knowledge, no other implementation accounts for this potential error. Single-sine analysis follows the approximation (for βω and Hc) method outlined in the Fig. 2 legend. The dual-sine method (Santos-Sacchi, 1992; Santos-Sacchi et al., 1998), follows the same fast Fourier transform-based determination of the real and imaginary components employed by the eCm method. In both cases estimates derived from both ω0 and ω1 are averaged (f0: 390.6 Hz; f1: 781.2 Hz). This averaging benefit must be kept in mind when comparing single-sine estimates. All methodologies are standard features in jClamp. In Fig. 9 the relation between measured Cm noise and imposed current noise is examined for each of the methodologies with either a 100-MΩ or 500-MΩ Rm. All methods work well under low-noise conditions, and this simply indicates that the true limitation for Cm estimation is the noise of the system at those frequencies used by the method. Indeed, under real-world conditions where system noise levels are well minimized, step, single-sine and dual-sine techniques can provide roughly equivalent Cm estimate noise levels (within ±10–30%) close to theoretical limits (Gillis, 1995; Barnett and Misler, 1997; Thompson et al., 2001). However, under forced noise levels as in Fig. 9 the eCm and single-sine techniques appear better able to deal with higher noise levels than this step implementation. Because of the exact same implementation of frequency analysis and the absence of approximations for eCm and dual-sine approaches, sensitivities turn out to be essentially equivalent. Had βω been determined in another manner, e.g., by the dω approach, the results may have differed. Given the relation between the single-sine phase-tracking approach and the eCm method detailed above, it is not surprising that these two methods are roughly equivalent in handling noise. For each method, slightly better noise levels are achieved when βω is fixed during analysis to simulate a “pause and dither” approach; thus, recalculating βω for each sample introduces additional noise in the analysis that must be balanced with the increased validity afforded by point-by-point correction. The step analysis is particularly prone to high noise levels; this is especially obvious when low frequency (e.g., 60 Hz) interference is included, and is in marked contrast to sinusoidal approaches that can avoid extraneous spectral interference. The conclusions based on this mathematical modeling effort were essentially confirmed with a physical implementation (see Fig. 9). From these results, it is clear that given a sufficiently low system noise or actually a sufficiently high signal/noise level, the eCm method will have the sensitivity to analyze fusion events, as clearly suggested by the physiological results shown in Fig. 8.
SUMMARY
The new eCm method that I outlined here provides another tool for analysis of cellular processes that can be interrogated by Cm measures. It is a method that works in the absence of usual simplifying assumptions and approximations of the traditional patch-clamp-cell model circuit. In detailing the approach, several new observations have been made. They include: 1), the imaginary component of the input admittance Y, namely, BY, can be worked on by ∂Y/∂Cm to obtain absolute values of Cm; 2), the angle of ∂Y/∂Cm, namely, βω, is the same as that of ∂Y/∂ω and therefore can be obtained from frequency manipulation alone; 3), given βω, the absolute value of ∂Y/∂Cm can be determined; 4), frequency manipulation need not require an infinitesimal dω to extract βω; and 5), the traditional phase-tracking technique that evaluates raw amplifier currents, rather than difference currents, at the angle provided by series resistance dithering, works similar to the eCm method by simply returning estimates of BY, not ΔICm, and ultimately provides absolute, not delta, Cm estimates. Finally, it might be expected that enhancements of the eCm technique may arise from future implementations in hardware and by statistical optimizations.
Acknowledgments
I thank Peter Dallos, Fred Sigworth, and Alan Segal for looking at early drafts. I especially thank Peter Dallos for confirming some of the concepts presented in the manuscript, and identifying others that needed rethinking. Finally, I thank the reviewers for many helpful suggestions.
This work was supported in part by the National Institutes of Health (National Institute on Deafness and Other Communication Disorders grant no. DC00273).
References
- Ales, E., L. Tabares, J. M. Poyato, V. Valero, M. Lindau, and D. T. Alvarez. 1999. High calcium concentrations shift the mode of exocytosis to the kiss-and-run mechanism. Nat. Cell Biol. 1:40–44. [DOI] [PubMed] [Google Scholar]
- Alvarez De Toledo, G. A., R. Fernandez-Chacon, and J. M. Fernandez. 1993. Release of secretory products during transient vesicle fusion. Nature. 363:554–558. [DOI] [PubMed] [Google Scholar]
- Aravanis, A. M., J. L. Pyle, and R. W. Tsien. 2003. Single synaptic vesicles fusing transiently and successively without loss of identity. Nature. 423:643–647. [DOI] [PubMed] [Google Scholar]
- Barnett, D. W., and S. Misler. 1997. An optimized approach to membrane capacitance estimation using dual-frequency excitation. Biophys. J. 72:1641–1658. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bigiani, A., and S. D. Roper. 1995. Estimation of the junctional resistance between electrically coupled receptor cells in Necturus taste buds. J. Gen. Physiol. 106:705–725. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Breckenridge, L. J., and W. Almers. 1987. Currents through the fusion pore that forms during exocytosis of a secretory vesicle. Nature. 328:814–817. [DOI] [PubMed] [Google Scholar]
- Cole, K. S. 1971. Some aspects of electrical studies of the squid giant axon membrane. In Biophysics and Physiology of Excitable Membranes. W. J. Adelman, editor. Van Nostrand Reinhold Co., New York. 125–42.
- Debus, K., J. Hartmann, G. Kilic, and M. Lindau. 1995. Influence of conductance changes on patch clamp capacitance measurements using a lock-in amplifier and limitations of the phase tracking technique. Biophys. J. 69:2808–2822. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Debus, K., and M. Lindau. 2000. Resolution of patch capacitance recordings and of fusion pore conductances in small vesicles. Biophys. J. 78:2983–2997. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fernandez, J. M., F. Bezanilla, and R. E. Taylor. 1982. Distribution and kinetics of membrane dielectric polarization. II. Frequency domain studies of gating currents. J. Gen. Physiol. 79:41–67. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fidler, N., and J. M. Fernandez. 1989. Phase tracking: an improved phase detection technique for cell membrane capacitance measurements. Biophys. J. 56:1153–1162. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fricke, H. 1925. The electric capacity of suspensions with special reference to blood. J. Gen. Physiol. 9:137–152. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gandhi, S. P., and C. F. Stevens. 2003. Three modes of synaptic vesicular recycling revealed by single-vesicle imaging. Nature. 423:607–613. [DOI] [PubMed] [Google Scholar]
- Gillis, K. D. 1995. Techniques for membrane capacitance measurements. In Single-Channel Recording. B. Sakmann and E. Neher, editors. Plenum Press, New York. 155–98.
- Gillis, K. D. 2000. Admittance-based measurement of membrane capacitance using the EPC-9 patch-clamp amplifier. Pflugers Arch. 439:655–664. [DOI] [PubMed] [Google Scholar]
- Huang, G., and J. Santos-Sacchi. 1993. Mapping the distribution of the outer hair cell motility voltage sensor by electrical amputation. Biophys. J. 65:2228–2236. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jaffe, L. A., and L. C. Schlichter. 1985. Fertilization-induced ionic conductances in eggs of the frog, Rana pipiens. J. Physiol. 358:299–319. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Joshi, C., and J. M. Fernandez. 1988. Capacitance measurements. An analysis of the phase detector technique used to study exocytosis and endocytosis. Biophys. J. 53:885–892. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Klyachko, V. A., and M. B. Jackson. 2002. Capacitance steps and fusion pores of small and large-dense-core vesicles in nerve terminals. Nature. 418:89–92. [DOI] [PubMed] [Google Scholar]
- Lindau, M. 1991. Time-resolved capacitance measurements: monitoring exocytosis in single cells. Q. Rev. Biophys. 24:75–101. [DOI] [PubMed] [Google Scholar]
- Lindau, M., and E. Neher. 1988. Patch-clamp techniques for time-resolved capacitance measurements in single cells. Pflugers Arch. 411:137–146. [DOI] [PubMed] [Google Scholar]
- Lollike, K., N. Borregaard, and M. Lindau. 1995. The exocytotic fusion pore of small granules has a conductance similar to an ion channel. J. Cell Biol. 129:99–104. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Meltzer, J., and J. Santos-Sacchi. 2001. Temperature dependence of non-linear capacitance in human embryonic kidney cells transfected with prestin, the outer hair cell motor protein. Neurosci. Lett. 313:141–144. [DOI] [PubMed] [Google Scholar]
- Neher, E., and A. Marty. 1982. Discrete changes of cell membrane capacitance observed under conditions of enhanced secretion in bovine adrenal chromaffin cells. Proc. Natl. Acad. Sci. USA. 79:6712–6716. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Oppenheim, A., and R. Schafer. 1975. Digital Signal Processing. Prentice-Hall, Englewood Cliffs, NJ
- Pusch, M., and E. Neher. 1988. Rates of diffusional exchange between small cells and a measuring patch pipette. Pflugers Arch. 411:204–211. [DOI] [PubMed] [Google Scholar]
- Ratinov, V., I. Plonsky, and J. Zimmerberg. 1998. Fusion pore conductance: experimental approaches and theoretical algorithms. Biophys. J. 74:2374–2387. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sakmann, B., and E. Neher. 1995. Geometric parameters of pipettes and membrane patches. In Single-Channel Recording. B. Sakmann and E. Neher, editors. Plenum Press, New York. 637–50.
- Santos-Sacchi, J. 1989. Asymmetry in voltage-dependent movements of isolated outer hair cells from the organ of Corti. J. Neurosci. 9:2954–2962. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Santos-Sacchi, J. 1991a. Isolated supporting cells from the organ of Corti: some whole cell electrical characteristics and estimates of gap junctional conductance. Hear. Res. 52:89–98. [DOI] [PubMed] [Google Scholar]
- Santos-Sacchi, J. 1991b. Reversible inhibition of voltage-dependent outer hair cell motility and capacitance. J. Neurosci. 11:3096–3110. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Santos-Sacchi, J. 1992. On calculating series resistance: step and two-sine approaches. Poster presentation at the Yale Neuroscience Retreat, Feburary 21–23. Marine Biological Laboratory, Woods Hole, MA.
- Santos-Sacchi, J. 2002. Scisoft Technical Note. http://www.scisoftco.com/detrending_for_2_sine_cm_measures.pdf.
- Santos-Sacchi, J. 2003. New tunes from Corti's organ: the outer hair cell boogie rules! Curr. Opin. Neurobiol. 13:459–468. [DOI] [PubMed] [Google Scholar]
- Santos-Sacchi, J., S. Kakehata, and S. Takahashi. 1998. Effects of membrane potential on the voltage dependence of motility-related charge in outer hair cells of the guinea-pig. J. Physiol. 510:225–235. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Thompson, R. E., M. Lindau, and W. W. Webb. 2001. Robust, high-resolution, whole cell patch-clamp capacitance measurements using square wave stimulation. Biophys. J. 81:937–948. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhao, H. B., and J. Santos-Sacchi. 1998. Effect of membrane tension on gap junctional conductance of supporting cells in Corti's organ. J. Gen. Physiol. 112:447–455. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhao, H. B., and J. Santos-Sacchi. 1999. Auditory collusion and a coupled couple of outer hair cells. Nature. 399:359–362. [DOI] [PubMed] [Google Scholar]
- Zierler, K. 1992. Simplified method for setting the phase angle used in capacitance measurements in studies of exocytosis. Biophys. J. 63:854–856. [DOI] [PMC free article] [PubMed] [Google Scholar]