. 2026 Feb 25;98(13):9753–9763. doi: 10.1021/acs.analchem.5c07120

Detection of Gaseous Mercuric Halides Using Acetate and Iodide Chemical Ionization Mass Spectrometry

Mohammad Borna Bahramsari †, Alexei F Khalizov †,‡,*
PMCID: PMC13063219  PMID: 41738485

Abstract

Predicting the fate of gaseous oxidized mercury (GOM) in the atmosphere remains a challenge because no direct analytical methods exist to determine its molecular speciation. Conventional preconcentration techniques produce speciation losses and induce ligand exchange, obscuring the original composition of GOM. Direct analysis by chemical ionization mass spectrometry (CIMS) offers a path toward addressing this analytical challenge, but its success depends critically on the use of an appropriate ion–molecule chemistry. Previously, we showed the applicability of the negative-ion CIMS for detection of gaseous HgCl2. Here, we demonstrate the first systematic evaluation of the detection of gaseous mercuric halides (HgCl2, HgBr2, and HgI2) by acetate (CH3COO–·CH3COOH) and iodide (I–) reagent ions, which have been successfully used in the past for the detection of atmospheric trace chemicals. As soft Lewis bases, these reagent ions react promptly with mercuric halides through either complexation or ligand switching. Using a calibrated permeation source, we establish the sensitivity and limit of detection (LOD) for HgBr2 with these and previously reported reagent ions. Measured sensitivities range from 3.8 to 21.0 cps/ppbv (for 1 Mcps reagent ion signal), and in the case of ligand switching, they agree with theoretical sensitivities. The current LOD of 53–268 pptv, obtained at a 2.5 Torr ion–molecule reaction pressure, is sufficient for laboratory studies of Hg chemistry and speciating GOM from combustion flue gases. Operating at higher ion–molecule reaction pressures will significantly enhance ionization efficiency, extending the applicability of CIMS toward direct detection of GOM in the atmosphere.


graphic file with name ac5c07120_0006.jpg


graphic file with name ac5c07120_0004.jpg

1. Introduction

Mercury is a global, persistent neurotoxic pollutant, and its concentration in the atmosphere has increased 7 times due to anthropogenic activities since preindustrial times. In the atmosphere, it can be present as elemental mercury (gaseous elemental mercury, GEM), oxidized mercury (gaseous oxidized mercury, GOM), and particle-bound mercury (PBM). In contrast to GEM, which can persist and be transported globally for 0.5–2 years, GOM and PBM are removed more rapidly due to their higher reactivity and solubility. − Evaluating the deposition of these forms is critical, as oxidized mercury in water and soil can be converted into methylmercury, , which bioaccumulates and biomagnifies to high levels in fish, an important protein source for humans. , The evaluation of the fate of GOM remains challenging because its exact chemical composition is not well-defined. In most cases, the identities of oxidized mercury species are inferred from theoretical considerations rather than confirmed by direct experimental evidence. −

Current methods for chemical speciation of GOM involve its preconcentration using appropriate media such as cation exchange membranes, nylon membranes, metal oxide glasses, and custom-made collectors. , The preconcentrated samples are thermally desorbed to release either the original mercuric compounds or elemental mercury, which are analyzed using various methods. For instance, mass spectrometry-based techniques, such as a commercial atmospheric pressure chemical ionization mass spectrometer (APCI-MS) and a gas chromatograph coupled with an electron impact ionization mass spectrometer (GC–MS), have been used to detect preconcentrated HgCl2 and HgBr2. , In contrast to explicit speciation of the evolving Hg­(II) by MS-based techniques, in the pyrolysis-based methods the chemical speciation is lost, and so they rely on an indirect identification based on the comparison against pure Hg­(II) standards. Specifically, the collected mercuric compounds are desorbed under a controlled temperature ramp, completely pyrolyzed to elemental mercury, and subsequently analyzed using a Tekran analyzer. For chemical speciation, the thermal desorption profiles of field samples are deconvoluted by comparing them with reference profiles obtained from the thermal desorption of permeation-derived pure GOM compounds, including HgBr2, HgCl2, Hg­(NO3)2·H2O, and HgSO4. − A major disadvantage of all preconcentration techniques is that subsequent thermal desorption can introduce artifacts, e.g., many proposed oxidized mercury species (e.g., BrHgONO, BrHgONO2, BrHgOH, and BrHgOOH) may be labile, have low volatility, or undergo exchange reactions with other chemicals upon collection, thereby altering the original chemical speciation. Unsurprisingly, only mercuric halides were successfully detected using the combination of preconcentration coupled with thermal desorption and mass spectrometry analysis. , Furthermore, preconcentration is inapplicable for the detection of reactive species, such as free radicals, making preconcentration-based methods unsuitable for kinetic and mechanistic studies of transient intermediates.

Chemical ionization mass spectrometry (CIMS) can be used to ionize and analyze neutral molecules directly from the gas phase, offering a way for chemical speciation and quantification of both stable and transient GOM molecules in laboratory and field studies. , In CIMS, ionization occurs through an ion–molecule reaction (IMR) of a gaseous analyte with a reagent ion produced in corona discharge or using a radioactive or X-ray source. The sensitivity and limit of detection of CIMS depend on the IMR mechanism, reaction pressure, cleanliness of the system, and other factors, giving rise to various method implementations that have been used for the direct detection of various atmospheric constituents. − Our research group has previously explored the use of three reagent ions, SF6 –, NO3 –·HNO3, and CO n – (n = 3 and 4), reporting their relative detection sensitivities toward HgCl2. Later, we employed two of those ions, SF6 – and NO3 –·HNO3, for the investigations of the heterogeneous uptake , and surface-catalyzed exchange reactions of mercuric halides. While SF6 – showed a significantly higher detection sensitivity than NO3 –·HNO3, its performance was affected by water vapor, prompting the use of the nitrate complex ion in the experiments where water vapor concentration was elevated. Other studies also reported the interfering effect of water on SF6 –, caused by the irreversible reactions to form other ion products and limiting the use of SF6 – in atmospheric measurements. , Unlike SF6 –, ions such as iodide (I–) and acetate (CH3COO–) undergo only a reversible reaction with water, making them suitable for atmospheric detection of trace organic and inorganic compounds, including hyperoxides, organic and inorganic acids, and vaporized particle constituents in secondary organic aerosols. −

This work evaluates the suitability of the iodide and acetate reagent ions for the detection of gaseous mercuric halides HgX2 (where X is Cl, Br, or I) in the laboratory and potentially for atmospheric measurements. We investigate the IMR mechanism, and using a calibrated HgBr2 source, we establish the absolute sensitivity and the limit of detection for these reagent ions, as well as for SF6 – used previously.

2. Methods

2.1. Chemicals

Mercuric chloride (HgCl2, Honeywell, >99.5%), mercuric bromide (HgBr2, Alfa Aesar, >99%), and mercuric iodide (HgI2, Sigma-Aldrich, >99%) were used as surrogates for GOM. Iodomethane (CH3I, Alfa Aesar, >99.5%), trifluoro iodomethane (CF3I, Sigma-Aldrich, 99%), acetic anhydride (C4H6O3, Sigma-Aldrich, >99%), and SF6 (15% in argon, Airgas) were used as precursors for the generation of reagent ions. Helium and nitrogen compressed gases were of ultrahigh purity (UHP) and provided by Airgas.

2.2. Ion Drift-Chemical Ionization Mass Spectrometer

The ion drift-chemical ionization mass spectrometer (ID-CIMS) used in this work is similar to that described in our previous studies. ,,, It consists of an ion drift tube (DT) and a collision deactivation chamber (CDC) connected to a differentially pumped vacuum chamber that houses a quadrupole mass spectrometer (Figure S1). Chemical ionization takes place in the ion drift tube upon mixing of a gas flow containing neutral analyte with a flow containing reagent ion produced from an appropriate precursor in the afterglow of a negative corona discharge. The drift tube is a stack of five stainless steel rings (1/2 in. aperture and 1/8 in. thickness) separated by five static-dissipating fluoropolymer insulators (1 in. aperture and 3/4 in. thickness) sealed by Viton O-rings. The CDC is a stack of four stainless steel ring electrodes (1/2 in. aperture and 1/16 in. thickness) and four static dissipating fluoropolymer insulators (1/2 in. aperture and 3/8 in. thickness). In this work, the pinhole between DT and CDC is removed to maintain the same 1–3 Torr pressure in both parts, essentially resulting in a longer DT. The ring electrodes in DT and CDC are electrically connected via resistors, and a small electric current is sent through the resistors to develop a predefined voltage drop along the electrode stack. The DT resistors are fixed at 1 MΩ. The CDC resistors are single-turn potentiometers adjusted to about 30% of their maximum value of 2 MΩ to set a uniform electric field through both regions, DT and CDC. The electric field focuses the ions in a tight beam and also allows control of the ion–molecule reaction time and ion clustering. At the end of CDC, ions are focused through a 0.5 mm pinhole into the differentially pumped chamber, where most of the neutral gas molecules are removed in the first compartment while the ions are refocused and guided by lenses into the second compartment, where they are detected using a quadrupole mass filter with a counting channeltron electron multiplier. Compared to our previous work, the corona discharge–DT region, pumping, and quadrupole system were upgraded to improve performance, as described below.

First, we changed the location where the reagent ions are mixed with the analyte. Previously, the mixing point was inside DT, where the analyte flow entered DT coaxially through the front ring electrode and the reagent ion flow entered at a right angle to DT through the fluoropolymer insulator between the first and second ring electrodes. Such a configuration required the use of a strong electric field to refocus the ions in a tight beam to maintain a high ion transmission. In the new configuration (Figure S1), while the corona discharge source is kept transverse to the drift tube axis, the reagent ion and analyte flows are mixed in a glass tee and only then introduced axially into DT. By moving the mixing region outside of DT, this new design allowed achieving sufficient ion beam focusing even with electric fields as low as 5 V cm–1, resulting in longer ion–molecule reaction times. It should be noted that positioning the corona discharge source coaxially with DT results in a significant mass-independent background signal (100–150 kcps) when detecting negative ions despite the use of an off-axis electron multiplier. The shortwave ultraviolet emitted by corona discharge enters the vacuum chamber through the pinhole and strikes the metal parts, causing the photoelectric effect. The emitted electrons are then attracted to the electron multiplier entrance, which is held at a high positive potential (+3 kV) during negative ion detection, producing secondary electrons that result in a high background count. Additionally, we found that placing the corona discharge source at a longer distance from the DT axis significantly reduces the mass-independent background count without affecting the reagent ion signal intensity. The mass-independent background is caused by the shortwave ultraviolet emitted by the metastables emerging from the corona discharge and reaching DT.

Second, we found that optimum refocusing of ions in DT can be achieved by keeping the first and second ring electrodes at the same potential, resulting in zero electric field in that region. As shown by Hanson et al. using SIMION simulations, the fringe electric field penetrating into this zero-field region from the region between the second and third ring electrodes helps to refocus the ions into a tight beam.

Third, two turbopumps were added to the first chamber to triple the pumping speed from 250 L/s to 750 L/s. Those pumps are an Agilent TwisTorr 304 FS and a split-flow Agilent TwisTorr 305 SF. The two regular turbo pumps are backed by a single Agilent DS302 rotary vane pump. The split flow turbo pump is backed by an Agilent DS402 rotary vane pump. In future applications, the split flow turbopump will be used to provide 10 L/s pumping of the CDC region, but currently the side port is sealed off. A larger pumping speed allows increasing the maximum DT pressure from 3 to 5 Torr concurrently with enlarging the sampling pinhole diameter from 0.3 mm to 0.5 mm to allow more ions to enter the detection chamber. The rear chamber, holding the quadrupole and multiplier, is pumped by a single Agilent TwisTorr 304 FS backed by a single Agilent DS302 rotary vane pump, allowing it to maintain a 10–6 Torr vacuum with a full gas load.

Fourth, the original old-generation 19 mm quadrupole was replaced with a 19 mm segmented quadrupole equipped with prefilter and postfilter rods. These RF-only rods reduce the fringing electric field at the entrance and exit of the mass filter, improving both ion transmission and resolving power.

2.3. Sources of Gaseous Mercuric Halides

Three different sources were used to generate gaseous mercuric halides: a commercial permeation tube, a glass tube packed with small glass beads coated by solid mercuric halides, and a glass tube with a thin coat of solid mercuric halides on its inner wall. A detailed description of each source is provided below.

The permeation source was based on a certified, gravimetrically calibrated permeation tube supplied by VICI Metronics, producing 556.3 ng min–1 HgBr2 at 100 °C and atmospheric pressure. The 9.5 mm diameter and 8 cm length permeation tube was placed at the bottom of one arm of a temperature-controlled U-shaped glass tube of 1.7 cm diameter and 20 cm height. The other arm was filled with 0.6 cm diameter spherical glass beads to ensure a fast temperature equilibration of the gas flow. The two arms of the U-tube are divided at the bottom by a glass separator with multiple holes. The HgBr2-laden gas flow leaving the U-tube was diluted with a flow of nitrogen, using a tee adaptor (Figure S2). A critical orifice installed between the second tee and ID-CIMS ensured a constant sampling flow of 300 sccm. To maintain a uniform temperature, the U-tube was encased in a thin, pliable copper sheet, and the space between the glass and copper was filled with copper wool to ensure efficient heat transport. The entire assembly was wrapped with a heating tape and two layers of fiberglass insulation. A constant temperature was maintained using a PID controller with a thermocouple attached to the outside glass wall next to the region where the permeation tube was located. Separate experiments showed that the temperature in the arm holding the permeation tube was within 0.5 °C of the target temperature (Figure S3). The mixing ratio of HgBr2 was varied between 0.31 and 4.2 ppbv by setting the carrier flow through the U-tube to 500 sccm and adjusting the dilution flow between 4 and 60 slpm. The lower limit on the carrier flow is imposed at about 360 sccm to prevent condensation of HgBr2 upon cooling of the gas to room temperature.

Although the permeation tube produces a known mercury output, it is not a convenient source for everyday experimental work, as it must be continuously purged by a flow of gas at ambient pressure and at a precisely controlled elevated temperature (100 °C). Hence, most qualitative measurements were performed using the other two sources, whereas the permeation source was used only to determine the sensitivity of the ID-CIMS toward HgBr2 with different reagent ions. The first alternative source consisted of glass beads (0.5 mm diameter, Cat. No. 11079105, BioSpec Products) coated with a solid mercuric halide and packed inside a 1/4″ outer diameter (OD) glass tube between a cotton wool plug upstream and a porous polyethylene plug (made from a 1/8″ OD rod, 15–40 μm pore size, Porex-5520, Interstate Specialty Products) downstream (Figure S4). The second alternative source was made of a 1/4″ OD glass tube coated on the inside with a solid mercuric halide layer. A filter holder (Savillex, 25 mm Single Stage Filter Assembly) loaded with a PTFE membrane (Savillex, 25 mm Filter Membrane, 5–6 μm pore size) was connected downstream to prevent accidentally detached mercuric halide particles from reaching the drift tube (Figure S4). The first source performed without any reduction in its output over a year but created a significant pressure drop, up to 20 Torr at a 4 sscm flow. The second source produced only a small pressure drop (1.2 Torr) with the same flow but had to be recharged after a few days due to the smaller amount of mercuric halide that could be securely deposited on the tubing wall. Stability of both sources was verified by repeated measurements of the HgX2 signal under identical operating conditions, with no systematic drift observed within experimental uncertainty.

2.4. Calculations

Density Functional Theory (DFT) calculations were performed using Gaussian 16 (Revision C.01). Geometry optimizations and vibrational frequency calculations were carried out at the M06-2X level of theory using the aug-cc-pVTZ , basis set for H, C, O, and Cl and the aug-cc-pVTZ-PP basis set for the outer electrons of Hg, Br, and I. M06-2X is empirically parameterized to account for medium-range electron correlation, which is particularly relevant for ion–molecule interactions, and has been previously used by our research group and others for mercury-containing molecules, showing reliable performance. ,,,− To account for scalar relativistic effects in Br, I, and Hg, the Stuttgart/Köln effective core potentials (ECPs) were used, replacing the 10, 28, and 60 innermost electrons, respectively. − Although CCSD­(T) single-point energy calculations improve accuracy by 1–5 kcal/mol, , the larger errors occur mostly for nonisogyric reactions where the multiplicity is not preserved, such as when calculating electron affinity. Most reactions of interest involve clustering between a closed-shell neutral molecule and a radical anion. In electron affinity calculations, several species exhibit open-shell electronic configurations. Closed-shell species were treated using restricted DFT, whereas open-shell radical anions and the resulting ion–molecule complexes were treated using unrestricted DFT. For all open-shell systems, wave function stability checks were performed after geometry optimizations and frequency calculations. Geometry optimizations were performed without imposing symmetry constraints. Binding energies of the ion–molecule complexes were estimated from total electronic energies obtained at the DFT level.

The Ion Mobility Spectrometry Suite v1.13 (IMoS) software was used to calculate the reduced ion mobility of reagent ions under standard ambient temperature and pressure (SATP) conditions. , The input file for IMoS included the Cartesian coordinates, van der Waals radii, and Mulliken charges of atoms in molecules obtained from DFT calculations (Table S2). The trajectory method with Lennard-Jones and quadrupole potentials was applied using N2 gas. Default Lennard-Jones parameters and N2 quadrupole parameters implemented in IMoS were used without modification (Table S3). To validate the accuracy of this method, the reduced ion mobilities of SF6 –, CO3 –, NO3 –·HNO3, and H3O+ were calculated and compared with experimental results. For all ions, the relative error was less than 10% (Table ).

1. Reduced Ion Mobilities (cm2 V–1 s–1) Used for Calculation of Ion Velocities and Ion–Molecule Reaction Times.

ion calculated literature relative error %
SF6 – 2.17 1.99 9.0
CO3 – 2.39 2.55 6.2
NO3 –·HNO3 2.00 1.86 7.5
H3O+ 2.95 2.76 6.8
CH3COO–·CH3COOH 1.72 n/a  
I– 2.13 n/a  

Ion–molecule reaction rates were calculated using the Average Dipole Orientation (ADO) theory, which is based on the collision rate between reagent ions and analyte molecules. , The ADO rate constant is given by

kADO=(2πqμ)[α+CμD(2πkBT)] 1

where μ is the reduced mass of the analyte and reagent ion, q is the charge of the reagent ion, k B is the Boltzmann constant, α is the isentropic component of the polarizability tensor, and μD is the dipole moment; μD and α were obtained by DFT calculation of the analyte molecules. C is a parameter, ranging between 0 and 1, which describes the effectiveness of the charge locking in the dipole. Since the dipole moment of mercuric halides is zero, eq reduces to the Langevin equation

kL=(2πqμ)α 2

3. Results and Discussion

3.1. Ion Chemistry

To study ion–molecule reactions, we selected three mercuric halides, HgCl2, HgBr2, and HgI2, as surrogates for GOM because of their commercial availability. The relatively high saturation vapor pressures of these chemicals at room temperature allowed the construction of simple sources with stable outputs in the range of 396–5911 ng/h, as estimated from the measured sensitivities (Section ). Although these halides are nonpolar, unlike proposed atmospheric GOM molecules such as BrHgNO2, varying the halogen substituent X in HgX2 enables exploration of the effects of polarizability and Hg–X bond strength on the efficiency of ion–molecule reactions. From HgCl2 to HgI2, polarizability increases while Hg–X bond strength decreases. We used three methods of delivery of gaseous mercuric halides: two semiquantitative methods rely on the saturation vapor pressure of HgX2, and a fully quantitative method is based on a calibrated HgBr2 permeation source.

Ion–molecule reactions of HgX2 with the iodide and acetate reagent ion families can proceed through three pathways: electron transfer, ion clustering, and ion transfer. According to DFT calculations (Table ), the energy required to remove an electron (−ΔH) from all reagent ions is greater than the electron affinities of all mercuric halides, where the calculated electron affinities agree with values available in the literature. Therefore, electron transfer is not a thermodynamically feasible pathway. However, ion clustering and ion transfer reactions between reagent ions and mercuric halides are all exothermic (Tables and ). Notably, the enthalpy changes between different mercuric halides are close to each other for the same reagent ion, within 1 kcal/mol. Differences in Gibbs free energy are somewhat larger (within 3 kcal/mol) and arise due to variations in the reaction entropy between different mercuric halides (Tables and ). DFT calculations also indicate that acetate forms a significantly stronger complex with HgX2 compared to iodide. The bond dissociation enthalpies for acetate clusters with mercuric halides are in the range of 42–43 kcal/mol, whereas for iodide, they are in the range of 35–36 kcal/mol. Optimized structures of iodide and acetate clusters with mercuric halides are shown in Figure . Upon complex formation, significant geometric transformations take place in HgX2: the initial X–Hg–X 180° angle decreases to 117–140°, depending on the ion, and the Hg–X bond distances elongate by 0.13–0.18 Å.

2. Calculated and Literature Values of Electron Affinities (EA) of Reagent Ions and Mercuric Halides.

    ΔH (kcal/mol)  
# reaction calculation literature
1.1
I+e−→I−
–70.62 –70.54
1.2
I2+e−→I2−
–61.91 –55.34
1.3
I3+e−→I3−
–98.26 –97.45
1.4
(CH3CO)2O+e−→(CH3CO)2O−
11.51 n/a
1.5
CH3COO+e−→CH3COO−
–78.40 –74.94
1.6
HgCl2+e−→HgCl2−
–42.12 –36.67
1.7
HgBr2+e−→HgBr2−
–44.45 –37.36
1.8
HgI2+e−→HgI2−
–46.41 n/a

3. Enthalpies and Free Energies of Reactions between Mercuric Halides and the Acetate Family of Reagent Ions Produced Using (CH3CO)2O as a Precursor.

# reaction ΔH (kcal/mol) ΔG (kcal/mol)
2.1
HgCl2+CH3COO−→HgCl2·CH3COO−
–42.73 –37.52
2.2
HgBr2+CH3COO−→HgBr2·CH3COO−
–43.35 –35.46
2.3
HgI2+CH3COO−→HgI2·CH3COO−
–43.77 –38.73
2.4
HgCl2+CH3COO−·CH3COOH→HgCl2·CH3COO−+CH3COOH
–10.43 –14.72
2.5
HgBr2+CH3COO−·CH3COOH→HgBr2·CH3COO−+CH3COOH
–11.04 –12.67
2.6
HgI2+CH3COO−·CH3COOH→HgI2·CH3COO−+CH3COOH
–11.46 –15.94

4. Enthalpies and Free Energies of Reactions between Mercuric Halides and the Iodide Family of Reagent Ions Produced Using CF3I as a Precursor.

# reaction ΔH (kcal/mol) ΔG (kcal/mol)
3.1 HgCl2 + I– → HgCl2I– –35.79 –34.28
3.2 HgBr2 + I– → HgBr2I– –36.36 –34.21
3.3 HgI2 + I– → HgI2I– –36.48 –34.07
3.4a HgCl2 + I2 – → HgCl2I2 – –23.15 –19.38
3.4b HgCl2 + I2 – → HgCl2I– + I –10.70 –14.12
3.5a HgBr2 + I2 – → HgBr2I2 – –23.43 –17.34
3.5b HgBr2 + I2 – → HgBr2I– + I –11.27 –11.81
3.6a HgI2 + I2 – → HgI2I2 – –23.41 –20.07
3.6b HgI2 + I2 – → HgI2I– + I –11.38 –13.91
3.7a HgCl2 + I3 – → HgCl2I3 – –16.25 –16.47
3.7b HgCl2 + I3 – → HgCl2I– + I2 –4.65 –13.28
3.8a HgBr2 + I3 – → HgBr2I3 – –16.39 –15.62
3.8b HgBr2 + I3 – → HgBr2I– + I2 –5.22 –10.97
3.9a HgI2 + I3 – → HgI2I3 – –16.39 –16.81
3.9b HgI2 + I3 – → HgI2I– + I2 –5.34 –13.07
3.13 HgCl2 + CF3I2 – → HgCl2I– + CF3I –17.85 –22.85
3.14 HgBr2 + CF3I2 – → HgBr2I– + CF3I –18.42 –20.54
3.15 HgI2 + CF3I2 – → HgI2I– + CF3I –18.54 –22.64

1.

1

Calculated structures of the experimentally observed products of reactions of the three neutrals (HgCl2, HgBr2, and HgI2) with two reagent ions, CH3COO–·CH3COOH and I–. Panels: (a) HgCl2·CH3COO– (C1/Cs), (b) HgBr2·CH3COO– (C1/Cs), (c) HgI2·CH3COO– (C1/Cs), (d) HgCl2I– (C1/C2v), (e) HgBr2I– (C1/C2v), and (f) HgI2I– (C3v/C3v). Atom colors: Hg (gray), C (dark gray), O (red), H (white), Cl (green), Br (brownish red), and I (violet). All reported internuclear distances are given in angstroms (Å). Geometry optimizations were performed without symmetry constraints. The symmetries are given in parentheses after the complexes, where the first symbol denotes the point group reported by Gaussian, whereas the second symbol denotes an approximate symmetry inferred from the optimized geometries.

Depending on the drift tube’s electric field and the concentration of reagent ion precursor, the following reagent ions were identified in our experimental system: CH3I produced I–, I2 –, and I3 –; CF3I produced I–, I2 –, CF3I2 –, and I3 –; and (CH3CO)2O produced CH3COO–·CH3COOH, CH3COO–, and (CH3CO)2O–. The electric field strength was optimized to the lowest possible value, allowing the focusing of ions into a narrow beam, thereby maximizing product-ion signal output while maintaining high reagent-ion signal intensity. Figures S5 and S6 show the detuned spectrum of I– (m/z = 127) in the CF3I/N2 and CH3I/N2 systems, respectively. Detuning was required to prevent the saturation of the multiplier and was achieved by adjusting the voltage of the quadrupole exit lens; this method allowed the broadest control of the ion intensity and the lowest mass discrimination. However, detuning affected the ion energy in the quadrupole, resulting in the peak shape deviating from a Gaussian profile and appearing distorted. The signal intensity of iodide at a full multiplier voltage (2100 V) was estimated indirectly, using the experiments described in Section . In addition to the bare iodide, other ions were formed from CF3I and CH3I in relatively high concentrations. Figure S7 shows the mass spectra of the ions produced from CF3I, including I2 – (m/z = 254 amu), I3 – (m/z = 381 amu), and CF3I2 – (m/z = 323 amu) with signal intensities of 150, 350, and 750 kcps, respectively. The relatively strong signal of CF3I2 – is likely due to the high concentration of CF3I in the system, resulting in its clustering with I–. Figure S8 shows the mass spectra of the ions produced from CH3I, including I2 – (m/z = 254 amu), I3 – (m/z = 381 amu), and an unknown ion at m/z = 280 amu with signal intensities of 40, 60, and 100 kcps, respectively. Figure S9 presents the mass spectra of the ions observed using (CH3CO)2O as a precursor. The signal intensities of CH3COO– (m/z = 59 amu), (CH3CO)2O– (m/z = 102 amu), and CH3COO–·CH3COOH (m/z = 199 amu) were 200, 250, and 1100 kcps, respectively. When possible, in most of our experiments, the parameters of the system were adjusted to have I– and CH3COO–·CH3COOH as the major reagent ions.

When mercuric halides were introduced into the drift tube, the only observed ion products in the iodide and acetate systems were HgX2·I– and HgX2·CH3COO–, respectively (Figure ). The signal intensities of the iodide complexes HgCl2·I– (m/z = 395–405 amu), HgBr2·I– (m/z = 483–493 amu), and HgI2·I– (m/z = 577–585 amu) were 14.0, 1.1, and 0.6 kcps, respectively. The signal intensities of the acetate complexes HgCl2·CH3COO– (m/z = 325–337 amu), HgBr2·CH3COO– (m/z = 413–426 amu), and HgI2·CH3COO– (m/z = 509–519 amu) were 25.0, 1.4, and 1.4 kcps, respectively. The difference in signal intensities between various mercuric halides for the same reagent ion can be attributed to differences in their saturation vapor pressures, producing different concentrations of these halides in the drift tube. The difference in signal intensities between the iodide and acetate systems for the same mercuric halides is mostly due to the different product ion stabilities, as described in Section .

2.

2

Mass spectra of product ions produced in the reaction of HgCl2, HgBr2, and HgI2 neutrals with CH3COO–·CH3COOH and I– reagent ions. The black curves are for the experimental spectra, and the red sticks indicate the theoretical spectra.

When comparing the product ions observed in the experiments against the predictions based on the DFT calculations, one can see the absence of complexes such as HgX2·I2 – and HgX2·I3 –. We surmise that these complexes are unstable under the conditions of our experiment. Either they are not thermolyzed and dissociate immediately after formation, or they dissociate during collisions promoted by the electric field in the drift tube or near the pinhole separating the drift tube from the lower pressure region in CIMS. As described in Section , even HgX2·I– complexes, which are bonded by about 36 kcal/mol, were not fully thermalized, partially dissociating at the 2–3 Torr pressure of the drift tube. By taking the difference between a and b channels in Table , one can see that the elimination of the I and I2 neutrals from HgX2·I2 – and HgX2·I3 – requires only 11–12 kcal/mol. Thus, the decomposition of these complexes due to inefficient thermalization is plausible. Similarly, no complexes of HgX2 with CH3COO–·CH3COOH and (CH3CO)2O– were observed.

3.2. Sensitivity and Limit of Detection

In our previous study, the sensitivity for detection of HgCl2 by several reagent ions has been estimated, assuming that the helium flow passing through a bed made of powdered HgCl2 was fully saturated with HgCl2 vapor. The resulting experimental sensitivity was found to be significantly lower than theoretical sensitivity, indicating that the saturation assumption may have been invalid. Here, we relied on a gravimetrically calibrated permeation source to produce a gas flow with a known concentration of HgBr2. This source was used to measure the sensitivity in the iodide and acetate systems. Additionally, we remeasured the sensitivity with the SF6 –, using HgBr2 as the analyte. All calibrations were performed in a 0.3–3.2 ppbv range of HgBr2 concentrations, using selected-ion monitoring (SIM). The experimentally measured sensitivities ranged from 17 to 83 cps/ppbv, depending on the reagent ion (Table ), which was a result of not only different reagent ion signal intensities but also of different product ion stabilities. To make the comparison possible, the sensitivities were normalized to the reagent ion count rate of 1 Mcps, as also shown in Table .

5. Measured Experimental Sensitivities, Limits of Detection, and Normalized Theoretical Sensitivities for the Detection of HgBr2 Using Different Reagent Ions.

  reagent ion signal intensity S exp LOD S exp  
reagent ion Mcps cps ppb–1 ppt cps ppb–1 @ 1 Mcps reagent ion S t /S exp
CH3COO–·CH3COOH 1.1 22 216 21.0 ± 0.9 1.3
I– 4.5 17 53 3.8 ± 0.1 6.6
I– 7.6 18 268 2.3 ± 0.2 10.7
SF6 – 4.5 83 102 18.5 ± 0.5 1.3
a

Calculated for a 60 s integration time and assuming a signal-to-noise ratio S/N = 3.

b

Theoretical sensitivity normalized to 1 Mcps reagent ion signal intensity.

c

Produced from CF3I.

d

Produced from CH3I.

To measure the signal intensity of CH3COO–·CH3COOH during calibration experiments while preventing the saturation of the electron multiplier, the less abundant isotopic peak with an m/z of 120 was used. Since iodine has only a single isotope, this approach could not be used for I– and instead its signal intensity was calculated from the ratio of detuned to tuned H2C2O4·I– signals, using a small addition of H2C2O4. Since the detuned-to-tuned signal ratio depends on the ion’s mass-to-charge ratio (m/z), a fit was constructed based on the measurements across a broad spectral range (Figure S10). From the fit, the mass correction coefficient (C mass) for H2C2O4·I– (m/z = 217) relative to I– (m/z = 127) was determined to be 2.02. The tuned iodide signal intensity was then obtained using the following equation,

IT(I−)=IDT(I−)×Cmass×IT(H2C2O4·I−)IDT(H2C2O4·I−) 3

where I T(I–) is the tuned signal intensity of iodide, I DT(I–) is the detuned signal intensity of iodide, C mass is the mass correction coefficient H2C2O4·I– (m/z = 217) relative to I– (m/z = 127), I T(H2C2O4·I–) is the tuned signal intensity of the oxalic acid cluster with iodide, and I DT(H2C2O4·I–) is the detuned signal intensity of the oxalic acid cluster with iodide.

Oxalic acid was intentionally added to the system to provide a reference for the iodide signal. It was selected because its concentration could be readily controlled and the mass of its product ion (m/z = 217) was relatively close to that of iodide (m/z = 127), producing a lower error from mass discrimination. The detuned-to-tuned ratios of H2C2O4·I– were 20.2 and 17.5 for the CF3I/N2 and CH3I/N2 systems, respectively, and this ratio was assumed to remain constant under tuned ion transmission conditions. Detuning was performed by changing the quadrupole exit lens from +40 V to −30 V in Merlin Automation software. The detuning method was validated using SF6 – as the reagent ion, since its signal intensity could be independently calculated from the less abundant (4.4%) isotopic peak at m/z = 148 amu. When oxalic acid was used as the reference to calculate the signal intensity of the most abundant isotopic peak of SF6 – (m/z = 146 amu) based on its isotopic peak at m/z = 148 amu, the relative error of the tuning method was 3.1%.

Figure shows calibration plots where product ion intensities are normalized per million total reagent ion cps. The plots indicate a good linearity; the observed scatter in data points can be attributed to variations in the mercuric bromide concentration and reagent ion signal. For mercuric bromide, the variability is caused jointly by a ±1 °C temperature oscillation of the PID-controlled heating system that creates a 7–10% variation in the emission rate from the permeation source, flow and pressure variability introduced by mass flow controllers (about 2%), and incomplete signal stabilization caused by wall adsorption/desorption of mercuric bromide when its concentration is increased or decreased. As for the reagent ion, due to the high intensity, its signal was measured indirectly at the end of each concentration point, introducing some uncertainty, as described above. The sensitivities derived from the slopes of the plots are listed in Table , decreasing in the order CH3COO–·CH3COOH, SF6 –, I– (generated from CF3I), and I– (generated from CH3I). The difference between the sensitivities obtained using the iodide produced by the two different precursors arises from the presence of a high concentration of CF3I2 – in the CF3I/N2 system (Figure S7). Although the signal intensity of CF3I2 – is only about one-sixth that of iodide, it can outcompete the iodide by forming a fully thermalized HgBr2·I– in an ion exchange reaction 3.14 (Table ), where collisional thermalization of the product is not required, as the energy produced upon complex formation is removed in the form of kinetic energy of the two products. On the contrary, in clustering reaction 3.2 between iodide and mercuric bromide, the resulting HgBr2·I– complex is roto-vibrationally hot and must be thermalized through collisions with the surrounding gas. Since the pressure in our system was relatively low, the dissociation rate was higher than the thermalization rate (see Section S1 in Supporting Information describing the calculation of the thermalization and decomposition rates), meaning the net collision rate with the bath gas determined the outcome. Consequently, fewer iodide–mercury clusters were detected with CH3I, leading to a lower sensitivity where only bare iodide was present (Table ). Similar results were reported by Thornton et al., where the sensitivity of detection of formic acid by I– increased with relative humidity because when I­(H2O)− reacted with formic acid, the leaving water molecule carried away a large fraction of the reaction energy, requiring no collisional thermalization of the formic acid–iodide complex.

3.

3

Experimental calibration curves obtained using selected ion monitoring, showing the dependence of standardized product ion intensity (corresponding to 1 Mcps reagent ion signal) on HgBr2 mixing ratio. Reagent ions SF6 –, CH3COO–·CH3COOH, and I– were used to form HgBr2F–, HgBr2CH3COO–, and HgBr2I–, respectively. The iodide reagent ion was produced from two different precursors, CF3I and CH3I, as indicated in the legend.

One can expect that the rate of the ion exchange reactions that produce thermalized ion products can be well described by the ion–molecule collision rates, such as those predicted by the ADO (or Langevin) theory. Then, for analyte mixing ratios below 100 ppbv, where reagent-ion depletion is negligible and pseudo-first-order ion–molecule reaction conditions apply, theoretical detection sensitivity (S t) can be calculated based on the ion–molecule reaction rate equation, ,

St=R×k×t×fiso×C1ppbv(cpsppbv−1) 4

where R is the reagent ion count rate, k is the ion–molecule reaction rate constant, t is the ion drift time, f iso = 0.22 is the ratio of the relative intensity of the monitored isotope peak to the sum of relative intensities of all isotope peaks in the product ion, and C 1 ppbv = 8.42 × 107 molecules cm–3 is the concentration of the analyte gas at a mixing ratio of 1 ppbv at a total pressure of 2.56 Torr and temperature of 298 K. Table shows values of the ion–molecule reaction rate constants between reagent ions and HgBr2, as well as the parameters needed for the Langevin theory. Table summarizes the theoretical sensitivities for HgBr2 detection using different reagent ions, together with the parameters used in these calculations. Table shows agreement between the normalized theoretical and experimental sensitivities for CH3COO–·CH3COOH and SF6 –, with the ratio of theoretical to experimental sensitivity being 1.3 in both cases. Some of this remaining discrepancy may be due to combined uncertainties in the calculated ion–molecule reaction rates and reduced ion mobilities, but the majority is caused by ion mass discrimination. Our instrument employs a 19 mm segmented quadrupole equipped with both prefilter and postfilter rods, where transmission efficiency is 41.2% at 131 amu and 29.2% at 414 amu. A factor of 1.4 lower transmission efficiency for the higher-m/z product ions nearly matches the 1.3-fold lower experimental sensitivity relative to theoretical sensitivity.

6. Ion–molecule Rate Constants for the Reaction of HgBr2 with Several Reagent Ions Calculated by the Langevin Theory, along with Parameters Used for the Calculation.

    α k Lan
reagent ion μm Å3 10–10 cm3 s–1 molecule–1
CH3COO–·CH3COOH 89.46 13.69 9.16
I– 93.85 13.69 8.94
SF6 – 103.91 13.69 8.50
a

Reduced mass of analyte and reagent ion.

b

Isentropic component of the polarizability tensor.

7. Theoretical Sensitivities for Detection of HgBr2 by Different Reagent Ions, along with the Reduced Ion Mobility, Electric Field Strength, and Reaction Time Used as Inputs to Eq .

  μ E t S t
reagent ion cm2 V–1 s–1 V cm–1 ms ncps ppb–1
I– 686.6 –21 1.52 25
CH3COO–·CH3COOH 555.8 –21 1.63 28
SF6 – 706.4 –21 1.5 24
a

Ion mobility at 298 K and 2.56 Torr, calculated using IMoS.

b

Electric field in the drift tube.

c

Normalized to 1 Mcps reagent ion signal intensity.

The sensitivity toward HgBr2 with SF6 – determined in this study using a permeation source is about an order of magnitude higher than the sensitivity toward HgCl2 with the same reagent ion in our previous study that employed a glass tube filled with solid HgCl2, indicating that the assumption of the fully saturated vapor was incorrect. The ratios of normalized theoretical to experimental sensitivity for the iodide ion produced by CF3I/N2 and CH3I/N2 are 6.6 and 10.7, respectively. The significantly lower experimental sensitivities are mainly due to insufficient thermalization of the product ion, HgBr2·I–, as discussed above. A better performance in the case of iodide generated from CF3I is due to a significant contribution from the ion exchange reaction with CF3I2 –, where the product does not require thermalization.

The limit of detection (LOD) was calculated based on the standard deviation of the background signal measured with the calibration source bypassed to eliminate the residual HgBr2. With a 60 s integration time and assuming a signal-to-noise ratio of 3, the LOD was 216 pptv for CH3COO–·CH3COOH, 53 pptv for I– produced from CF3I, 268 pptv for I– produced from CH3I, and 102 pptv for SF6 – (Table ). Although the limit of detection is directly related to sensitivity, it is also significantly influenced by the background signal. Therefore, the cleanliness of the system plays an important role in achieving a lower LOD. Another factor is the mass of the detected ion, because as the ion mass increases, the background noise contributed by unrelated ions typically decreases. This occurs because, at higher masses, the probability of naturally existing interfering molecules becomes progressively lower. Indeed, as Table demonstrates, lower LODs are achieved using the iodide reagent ion produced from CF3I as a precursor, even when the sensitivity was lower than that with SF6. This improvement results from the cleaner CF3I-based background, which minimizes chemical noise and spectral interferences. In contrast, SF6 – exhibits a higher sensitivity but yields a worse LOD due to a broader chemical background, producing greater interference.

4. Conclusions

In this study, the use of the CH3COO–·CH3COOH and I– reagent ions for the detection of gaseous mercuric halides was investigated theoretically and experimentally. The results show that the corresponding ion–molecule reactions occur as ion exchange and ion–neutral clustering. The acetate generally forms product ions in a higher yield than the iodide because its complexes with mercuric halides are 7 kcal/mol more stable and formed primarily through ion exchange, requiring no thermalization. On the contrary, most of the iodide–mercuric halide complexes are produced via clustering, which requires collisional thermalization to dissipate the energy released during complex formation. Such thermalization is inefficient under the low-pressure and low-humidity experimental conditions of our study. Accordingly, the normalized experimental sensitivities decrease in the order CH3COO–·CH3COOH, I– (generated from CF3I), and I– (generated from CH3I). The higher sensitivity obtained with the iodide generated from CF3I versus CH3I is attributed to the relatively large amount of CF3I2 – clusters present in that system that can react with HgX2 through ion exchange, whereas bare iodide generated from CH3I reacts solely through clustering. The discrepancy between theoretical and experimental sensitivities increases in the order CH3COO–·CH3COOH, I– (generated from CF3I), and I– (generated from CH3I) because as the reaction mechanism shifts from primarily ion exchange (CH3COO–·CH3COOH and CF3I2 –) to primarily clustering (I–), the product ions require more thermalization. The Langevin theory, used here to calculate theoretical sensitivities, assumes that the product ions are fully thermalized, and it performs well for the cases involving ion exchange. However, in the case of the bare iodide, this assumption is not valid, explaining the observed significant discrepancy between experimental and theoretical results. Although iodide is less suitable than CH3COO–·CH3COOH or SF6 – for the detection of mercuric halides in the experiments with lower IMR pressures, at higher pressures the thermalization will occur more efficiently, becoming competitive with thermal decay at ambient pressure. Furthermore, at typical ambient humidities, when iodide forms clusters with water, the reaction with mercuric halides will switch from clustering to ion exchange, requiring no thermalization. If the product ion is stable and thermalized, our results show that theoretical and experimental sensitivities agree with each other. In such cases, theoretical sensitivity can be used with confidence to estimate the concentration of Hg-containing free radicals and transient molecules measured by CIMS, where calibration using standards is impossible or impractical.

The soft ionization capability of the iodide CIMS provides a powerful analytical tool that has been used for detecting atmospheric trace gases, such as hydroperoxides and bicyclic peroxy radicals. − The current LOD of a few hundred pptv is sufficient for laboratory experiments focused on mercury chemistry, making iodide and acetate CIMS applicable for identifying the products of HgBr reactions with NO2 and O3, whose kinetics have been studied by the Pulsed Laser Photolysis-Laser-Induced Fluorescence. This LOD may be sufficient for sampling from high-concentration sources, such as flue gas from the stacks of cement plants, coal-fired power plants, and garbage incinerators, where GOM concentration commonly reaches tens of μg m–3 (a few ppbv). , A higher-pressure implementation of CIMS can significantly improve LOD through the combination of more efficient thermalization, lower analyte and reagent ion dilution in the IMR region, and longer IMR time. Additional improvements can be achieved by employing a high-efficiency atmospheric-pressure ion interface, as described by Junninen et al., which guides ions more effectively, reducing ion losses associated with multistage pumping, as well as by using a flow-opposed drift tube to concentrate and extract product ions prior to sampling. , Molecular oxygen was not explicitly added to the bath gas in this study because O2 is not expected to react with or significantly deplete the reagent ions based on previous studies. ,, Furthermore, O2 is not anticipated to interfere with the formation of mercuric compound–reagent ion clusters and introduce detectable interferences. Using CIMS at ambient pressure may allow for the direct detection of the atmospheric GOM, skipping the preconcentration step, similar to the detection of the atmospheric hydroxyl radical, sulfuric acid, and highly oxidized organic molecules. ,− Further research will be needed to elucidate and address possible matrix effects not present in laboratory studies, such as the reduction in sensitivity or changes in ion chemistry caused by abundant ambient species and worsening of the limit of detection from the increase in the background signal in the mass region of mercuric product ions.

Supplementary Material

ac5c07120_si_001.pdf (1.1MB, pdf)

The Supporting Information is available free of charge at https://pubs.acs.org/doi/10.1021/acs.analchem.5c07120.

  • Schematics of the ion drift–chemical ionization mass spectrometer, calibration setup, and mercuric halide semiquantitative generation sources; representative reagent-ion mass spectra; tuned-to-detuned signal ratio as a function of ion mass; QRRK dissociation rate calculations; ion mobility input parameters (Cartesian coordinates, van der Waals radii, Mulliken charges, Lennard–Jones, and electrostatic parameters); and optimized geometries, electronic energies, zero-point energies, and vibrational frequencies for all calculated species (PDF)

The authors declare no competing financial interest.

References

  1. Geyman B. M., Thackray C. P., Jacob D. J., Sunderland E. M.. Impacts of Volcanic Emissions on the Global Biogeochemical Mercury Cycle: Insights From Satellite Observations and Chemical Transport Modeling. Geophys. Res. Lett. 2023;50:e2023GL104667. doi: 10.1029/2023gl104667. [DOI] [Google Scholar]
  2. Lyman S. N., Cheng I., Gratz L. E., Weiss-Penzias P., Zhang L.. An updated review of atmospheric mercury. Sci. Total Environ. 2020;707:135575. doi: 10.1016/j.scitotenv.2019.135575. [DOI] [PubMed] [Google Scholar]
  3. Schroeder W. H., Munthe J.. Atmos. Environ. 1998;32:809. doi: 10.1016/S1352-2310(97)00293-8. [DOI] [Google Scholar]
  4. Sather M. E., Mukerjee S., Smith L., Mathew J., Jackson C., Callison R., Scrapper L., Hathcoat A., Adam J., Keese D., Ketcher P., Brunette R., Karlstrom J., Van der Jagt G.. Gaseous oxidized mercury dry deposition measurements in the Four Corners area and Eastern Oklahoma, U. S. A. Atmos. Pollut. Res. 2013;4:168–180. doi: 10.5094/APR.2013.017. [DOI] [Google Scholar]
  5. Wright L. P., Zhang L., Marsik F. J.. Overview of mercury dry deposition, litterfall, and throughfall studies. Atmos. Chem. Phys. 2016;16:13399–13416. doi: 10.5194/acp-16-13399-2016. [DOI] [Google Scholar]
  6. Han D., Zhang J., Hu Z., Ma Y., Duan Y., Han Y., Chen X., Zhou Y., Cheng J., Wang W.. Particulate mercury in ambient air in Shanghai, China: Size-specific distribution, gas–particle partitioning, and association with carbonaceous composition. Environ. Pollut. 2018;238:543–553. doi: 10.1016/j.envpol.2018.03.088. [DOI] [PubMed] [Google Scholar]
  7. Lin C.-J., Pongprueksa P., Lindberg S. E., Pehkonen S. O., Byun D., Jang C.. Scientific uncertainties in atmospheric mercury models I: Model science evaluation. Atmos. Environ. 2006;40:2911–2928. doi: 10.1016/j.atmosenv.2006.01.009. [DOI] [Google Scholar]
  8. Heyes A., Mason R. P., Kim E.-H., Sunderland E.. Mercury methylation in estuaries: Insights from using measuring rates using stable mercury isotopes. Mar. Chem. 2006;102:134–147. doi: 10.1016/j.marchem.2005.09.018. [DOI] [Google Scholar]
  9. Monperrus M., Tessier E., Amouroux D., Leynaert A., Huonnic P., Donard O. F. X.. Mercury methylation, demethylation and reduction rates in coastal and marine surface waters of the Mediterranean Sea. Mar. Chem. 2007;107:49–63. doi: 10.1016/j.marchem.2007.01.018. [DOI] [Google Scholar]
  10. Hammerschmidt C. R., Fitzgerald W. F.. Methylmercury in Freshwater Fish Linked to Atmospheric Mercury Deposition. Environ. Sci. Technol. 2006;40:7764–7770. doi: 10.1021/es061480i. [DOI] [PubMed] [Google Scholar]
  11. Kumar V., Umesh M., Shanmugam M. K., Chakraborty P., Duhan L., Gummadi S. N., Pasrija R., Jayaraj I., Dasarahally Huligowda L. K.. A Retrospection on Mercury Contamination, Bioaccumulation, and Toxicity in Diverse Environments: Current Insights and Future Prospects. Sustainability. 2023;15:13292. doi: 10.3390/su151813292. [DOI] [Google Scholar]
  12. Dibble T. S., Zelie M. J., Mao H.. Thermodynamics of reactions of ClHg and BrHg radicals with atmospherically abundant free radicals. Atmos. Chem. Phys. 2012;12:10271–10279. doi: 10.5194/acp-12-10271-2012. [DOI] [Google Scholar]
  13. Jiao Y., Dibble T. S.. Quality Structures, Vibrational Frequencies, and Thermochemistry of the Products of Reaction of BrHg• with NO2, HO2, ClO, BrO, and IO. J. Phys. Chem. A. 2015;119:10502–10510. doi: 10.1021/acs.jpca.5b04889. [DOI] [PubMed] [Google Scholar]
  14. Jiao Y., Dibble T. S.. Structures, Vibrational Frequencies, and Bond Energies of the BrHgOX and BrHgXO Species Formed in Atmospheric Mercury Depletion Events. J. Phys. Chem. A. 2017;121:7976–7985. doi: 10.1021/acs.jpca.7b06829. [DOI] [PubMed] [Google Scholar]
  15. Lindqvist O., Rodhe H.. Atmosphere mercury - a review. Tellus. 1985;37:136–159. doi: 10.3402/tellusb.v37i3.15010. [DOI] [Google Scholar]
  16. Gustin M. S., Dunham-Cheatham S. M., Lyman S., Horvat M., Gay D. A., Gačnik J., Gratz L., Kempkes G., Khalizov A., Lin C.-J., Lindberg S. E., Lown L., Martin L., Mason R. P., MacSween K., Vijayakumaran Nair S., Nguyen L. S. P., O’Neil T., Sommar J., Weiss-Penzias P., Zhang L., Živković I.. Measurement of Atmospheric Mercury: Current Limitations and Suggestions for Paths Forward. Environ. Sci. Technol. 2024;58:12853–12864. doi: 10.1021/acs.est.4c06011. [DOI] [PubMed] [Google Scholar]
  17. Lown L., Dunham-Cheatham S. M., Murray P., Lyman S. N., Carlson K. L., Gustin M. S.. Feasibility of Metal Oxide Glasses and Polymer Membranes as Sorbents for Gaseous Oxidized Mercury. ACS Omega. 2025;10:47676–47687. doi: 10.1021/acsomega.5c05401. [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Jones C. P., Lyman S. N., Jaffe D. A., Allen T., O’Neil T. L.. Detection and quantification of gas-phase oxidized mercury compounds by GC/MS. Atmos. Meas. Tech. 2016;9:2195–2205. doi: 10.5194/amt-9-2195-2016. [DOI] [Google Scholar]
  19. Deeds D. A., Ghoshdastidar A., Raofie F., Guérette E. ´.-A., Tessier A., Ariya P. A.. Development of a Particle-Trap Preconcentration-Soft Ionization Mass Spectrometric Technique for the Quantification of Mercury Halides in Air. Anal. Chem. 2015;87:5109–5116. doi: 10.1021/ac504545w. [DOI] [PubMed] [Google Scholar]
  20. Gustin M. S., Dunham-Cheatham S. M., Allen N., Choma N., Johnson W., Lopez S., Russell A., Mei E., Magand O., Dommergue A., Elgiar T.. Observations of the chemistry and concentrations of reactive Hg at locations with different ambient air chemistry. Sci. Total Environ. 2023;904:166184. doi: 10.1016/j.scitotenv.2023.166184. [DOI] [PubMed] [Google Scholar]
  21. Luippold A., Gustin M. S., Dunham-Cheatham S. M., Zhang L.. Improvement of quantification and identification of atmospheric reactive mercury. Atmos. Environ. 2020;224:117307. doi: 10.1016/j.atmosenv.2020.117307. [DOI] [Google Scholar]
  22. Huang J., Miller M. B., Edgerton E., Sexauer Gustin M.. Deciphering potential chemical compounds of gaseous oxidized mercury in Florida, USA. Atmos. Chem. Phys. 2017;17:1689–1698. doi: 10.5194/acp-17-1689-2017. [DOI] [Google Scholar]
  23. Huang J., Miller M. B., Weiss-Penzias P., Gustin M. S.. Comparison of Gaseous Oxidized Hg Measured by KCl-Coated Denuders, and Nylon and Cation Exchange Membranes. Environ. Sci. Technol. 2013;47:7307–7316. doi: 10.1021/es4012349. [DOI] [PubMed] [Google Scholar]
  24. Mao N., Khalizov A.. Exchange Reactions Alter Molecular Speciation of Gaseous Oxidized Mercury. ACS Earth Space Chem. 2021;5:1842–1853. doi: 10.1021/acsearthspacechem.1c00178. [DOI] [Google Scholar]
  25. Dutt M., Arigò A., Famiglini G., Palma P., Cappiello A.. Chemical Ionization Mass Spectrometry: Fundamental Principles, Diverse Applications, and the Latest Technological Frontiers. Mass Spectrom. Rev. 2025:70007. doi: 10.1002/mas.70007. [DOI] [PubMed] [Google Scholar]
  26. Li W., Chen J., Ji Y., Zheng J., An T.. Recent progress in chemical ionization mass spectrometry and its application in atmospheric environment. Atmos. Environ. 2024;325:120426. doi: 10.1016/j.atmosenv.2024.120426. [DOI] [Google Scholar]
  27. Berresheim H., Elste T., Plass-Dülmer C., Eiseleb F. L., Tannerb D. J.. Chemical ionization mass spectrometer for long-term measurements of atmospheric OH and H2SO4. Int. J. Mass Spectrom. 2000;202:91–109. doi: 10.1016/S1387-3806(00)00233-5. [DOI] [Google Scholar]
  28. Fortner E. C., Zhao J., Zhang R. Y.. Development of ion drift-chemical ionization mass spectrometry. Anal. Chem. 2004;76:5436–5440. doi: 10.1021/ac0493222. [DOI] [PubMed] [Google Scholar]
  29. Zheng J., Khalizov A., Wang L., Zhang R.. Atmospheric Pressure-Ion Drift Chemical Ionization Mass Spectrometry for Detection of Trace Gas Species. Anal. Chem. 2010;82:7302–7308. doi: 10.1021/ac101253n. [DOI] [PubMed] [Google Scholar]
  30. Khalizov A. F., Guzman F. J., Cooper M., Mao N., Antley J., Bozzelli J.. Direct detection of gas-phase mercuric chloride by ion drift - chemical ionization mass spectrometry. Atmos. Environ. 2020;238:117687. doi: 10.1016/j.atmosenv.2020.117687. [DOI] [Google Scholar]
  31. Arnold S. T., Viggiano A. A.. Turbulent Ion Flow Tube Study of the Cluster-Mediated Reactions of SF6- with H2O, CH3OH, and C2H5OH from 50 to 500 Torr. J. Phys. Chem. A. 2001;105:3527–3531. doi: 10.1021/jp003967y. [DOI] [Google Scholar]
  32. Viggiano A. A., Fernandez A. I., Troe J.. Ion–molecule kinetics at 15–700 Torr. Phys. Chem. Chem. Phys. 2005;7:1533–1539. doi: 10.1039/B417454B. [DOI] [PubMed] [Google Scholar]
  33. Ye C., Yuan B., Lin Y., Wang Z., Hu W., Li T., Chen W., Wu C., Wang C., Huang S., Qi J., Wang B., Wang C., Song W., Wang X., Zheng E., Krechmer J. E., Ye P., Zhang Z., Wang X., Worsnop D. R., Shao M.. Chemical characterization of oxygenated organic compounds in the gas phase and particle phase using iodide CIMS with FIGAERO in urban air. Atmos. Chem. Phys. 2021;21:8455–8478. doi: 10.5194/acp-21-8455-2021. [DOI] [Google Scholar]
  34. Lee B. H., Lopez-Hilfiker F. D., Mohr C., Kurtén T., Worsnop D. R., Thornton J. A.. An Iodide-Adduct High-Resolution Time-of-Flight Chemical-Ionization Mass Spectrometer: Application to Atmospheric Inorganic and Organic Compounds. Environ. Sci. Technol. 2014;48:6309–6317. doi: 10.1021/es500362a. [DOI] [PubMed] [Google Scholar]
  35. Inomata S., Hirokawa J.. Non-radioactive Chemical Ionization Mass Spectrometry Using Acetic Acid–Acetate Cluster as a Reagent Ion for the Real-time Measurement of Acids and Hydroperoxides. Chem. Lett. 2017;46:38–41. doi: 10.1246/cl.160828. [DOI] [Google Scholar]
  36. Aljawhary D., Lee A. K. Y., Abbatt J. P. D.. High-resolution chemical ionization mass spectrometry (ToF-CIMS): application to study SOA composition and processing. Atmos. Meas. Tech. 2013;6:3211–3224. doi: 10.5194/amt-6-3211-2013. [DOI] [Google Scholar]
  37. Veres P., Roberts J. M., Burling I. R., Warneke C., de Gouw J., Yokelson R. J.. Measurements of gas-phase inorganic and organic acids from biomass fires by negative-ion proton-transfer chemical-ionization mass spectrometry. J. Geophys. Res.:Atmos. 2010;115:D23302. doi: 10.1029/2010jd014033. [DOI] [Google Scholar]
  38. Mao N., Antley J., Cooper M., Shah N., Kadam A., Khalizov A.. Heterogeneous Chemistry of Mercuric Chloride on Inorganic Salt Surfaces. J. Phys. Chem. A. 2021;125:3943–3952. doi: 10.1021/acs.jpca.1c02220. [DOI] [PubMed] [Google Scholar]
  39. Khalizov A. F., Mao N.. Heterogeneous Reaction of Gaseous Mercuric Chloride with Atmospherically Relevant Organic Films. ACS Earth Space Chem. 2023;7:2593–2602. doi: 10.1021/acsearthspacechem.3c00268. [DOI] [Google Scholar]
  40. Hanson D. R., Greenberg J., Henry B. E., Kosciuch E.. Proton transfer reaction mass spectrometry at high drift tube pressure. Int. J. Mass Spectrom. 2003;223–224:507–518. doi: 10.1016/S1387-3806(02)00924-7. [DOI] [Google Scholar]
  41. Frisch, M. J. ; Trucks, G. W. ; Schlegel, H. B. ; Scuseria, G. E. ; Robb, M. A. ; Cheeseman, J. R. ; Scalmani, G. ; Barone, V. ; Petersson, G. A. ; Nakatsuji, H. ; Li, X. ; Caricato, M. ; Marenich, A. V. ; Bloino, J. ; Janesko, B. G. ; Gomperts, R. ; Mennucci, B. ; Hratchian, H. P. ; Ortiz, J. V. ; Izmaylov, A. F. ; Sonnenberg, J. L. ; Williams-Young, D. ; Ding, F. ; Lipparini, F. ; Egidi, F. ; Goings, J. ; Peng, B. ; Petrone, A. ; Henderson, T. ; Ranasinghe, D. ; Zakrzewski, V. G. ; Gao, J. ; Rega, N. ; Zheng, G. ; Liang, W. ; Hada, M. ; Ehara, M. ; Toyota, K. ; Fukuda, R. ; Hasegawa, J. ; Ishida, M. ; Nakajima, T. ; Honda, Y. ; Kitao, O. ; Nakai, H. ; Vreven, T. ; Throssell, K. ; Montgomery, J. A., Jr. ; Peralta, J. E. ; Ogliaro, F. ; Bearpark, M. J. ; Heyd, J. J. ; Brothers, E. N. ; Kudin, K. N. ; Staroverov, V. N. ; Keith, T. A. ; Kobayashi, R. ; Normand, J. ; Raghavachari, K. ; Rendell, A. P. ; Burant, J. C. ; Iyengar, S. S. ; Tomasi, J. ; Cossi, M. ; Millam, J. M. ; Klene, M. ; Adamo, C. ; Cammi, R. ; Ochterski, J. W. ; Martin, R. L. ; Morokuma, K. ; Farkas, O. ; Foresman, J. B. ; Fox, D. J. . Gaussian 16; Gaussian, Inc.: Wallingford, CT, 2019. [Google Scholar]
  42. Zhao Y., Truhlar D. G.. The M06 suite of density functionals for main group thermochemistry, thermochemical kinetics, noncovalent interactions, excited states, and transition elements: two new functionals and systematic testing of four M06-class functionals and 12 other functionals. Theor. Chem. Acc. 2008;120:215–241. doi: 10.1007/s00214-007-0310-x. [DOI] [Google Scholar]
  43. Dunning T. H. Jr. Gaussian basis sets for use in correlated molecular calculations. I. The atoms boron through neon and hydrogen. J. Chem. Phys. 1989;90:1007–1023. doi: 10.1063/1.456153. [DOI] [Google Scholar]
  44. Woon D. E., Dunning T. H. Jr. Gaussian basis sets for use in correlated molecular calculations. III. The atoms aluminum through argon. J. Chem. Phys. 1993;98:1358–1371. doi: 10.1063/1.464303. [DOI] [Google Scholar]
  45. Peterson K. A., Puzzarini C.. Systematically convergent basis sets for transition metals. II. Pseudopotential-based correlation consistent basis sets for the group 11 (Cu, Ag, Au) and 12 (Zn, Cd, Hg) elements. Theor. Chem. Acc. 2005;114:283–296. doi: 10.1007/s00214-005-0681-9. [DOI] [Google Scholar]
  46. Jiao Y., Dibble T. S.. First kinetic study of the atmospherically important reactions BrHġ + NO2 and BrHġ + HOO. Phys. Chem. Chem. Phys. 2017;19:1826–1838. doi: 10.1039/C6CP06276H. [DOI] [PubMed] [Google Scholar]
  47. Lam K. T., Wilhelmsen C. J., Schwid A. C., Jiao Y., Dibble T. S.. Computational Study on the Photolysis of BrHgONO and the Reactions of BrHgO• with CH4, C2H6, NO, and NO2: Implications for Formation of Hg­(II) Compounds in the Atmosphere. J. Phys. Chem. A. 2019;123:1637–1647. doi: 10.1021/acs.jpca.8b11216. [DOI] [PubMed] [Google Scholar]
  48. Hewa Edirappulige D. T., Kirby I. J., Beckett C. K., Dibble T. S.. Atmospheric Chemistry of HOHg­(II)­O• Mimics That of a Hydroxyl Radical. J. Phys. Chem. A. 2023;127:8392–8403. doi: 10.1021/acs.jpca.3c04159. [DOI] [PubMed] [Google Scholar]
  49. Peterson K. A., Figgen D., Goll E., Stoll H., Dolg M.. Systematically convergent basis sets with relativistic pseudopotentials. II. Small-core pseudopotentials and correlation consistent basis sets for the post-d group 16–18 elements. J. Chem. Phys. 2003;119:11113–11123. doi: 10.1063/1.1622924. [DOI] [Google Scholar]
  50. Peterson K. A., Shepler B. C., Figgen D., Stoll H.. On the Spectroscopic and Thermochemical Properties of ClO, BrO, IO, and Their Anions. J. Phys. Chem. A. 2006;110:13877–13883. doi: 10.1021/jp065887l. [DOI] [PubMed] [Google Scholar]
  51. Figgen D., Rauhut G., Dolg M., Stoll H.. Energy-consistent pseudopotentials for group 11 and 12 atoms: adjustment to multi-configuration Dirac–Hartree–Fock data. Chem. Phys. 2005;311:227–244. doi: 10.1016/j.chemphys.2004.10.005. [DOI] [Google Scholar]
  52. Montero-Campillo M. M., Lamsabhi A. M., Mó O., Yáñez M.. Alkyl mercury compounds: an assessment of DFT methods. Theor. Chem. Acc. 2013;132:1328. doi: 10.1007/s00214-012-1328-2. [DOI] [Google Scholar]
  53. Shrivastav V., Nahin M., Hogan C. J., Larriba-Andaluz C.. Benchmark Comparison for a Multi-Processing Ion Mobility Calculator in the Free Molecular Regime. J. Am. Soc. Mass Spectrom. 2017;28:1540–1551. doi: 10.1007/s13361-017-1661-8. [DOI] [PubMed] [Google Scholar]
  54. Wu T., Derrick J., Nahin M., Chen X., Larriba-Andaluz C.. Optimization of long range potential interaction parameters in ion mobility spectrometry. J. Chem. Phys. 2018;148:074102. doi: 10.1063/1.5016170. [DOI] [PubMed] [Google Scholar]
  55. Su T., Bowers M. T.. Ion-Polar molecule collisions: the effect of ion size on ion-polar molecule rate constants; the parameterization of the average-dipole-orientation theory. Int. J. Mass Spectrom. Ion Phys. 1973;12:347–356. doi: 10.1016/0020-7381(73)80104-4. [DOI] [Google Scholar]
  56. Su T., Bowers M. T.. Theory of ion-polar molecule collisions. Comparison with experimental charge transfer reactions of rare gas ions to geometric isomers of difluorobenzene and dichloroethylene. J. Chem. Phys. 1973;58:3027–3037. doi: 10.1063/1.1679615. [DOI] [Google Scholar]
  57. Gandhi V. D., Short K., Hua L., Rodríguez I., Larriba-Andaluz C.. A numerical tool to calculate ion mobility at arbitrary fields from all-atom models. J. Aerosol Sci. 2023;169:106122. doi: 10.1016/j.jaerosci.2022.106122. [DOI] [Google Scholar]
  58. Okuyama Y., Sabo M., Itoh H., Matejčík S. ˇ.. Formations of negative ions in Sf6/N2 mixtures and their transport at atmospheric pressure. Eur. Phys. J.:Appl. Phys. 2013;61:24320. doi: 10.1051/epjap/2012120448. [DOI] [Google Scholar]
  59. Shumate C., St. Louis R. H., Hill H. H.. Table of reduced mobility values from ambient pressure ion mobility spectrometry. J. Chromatogr. 1986;373:141–173. doi: 10.1016/s0021-9673(00)80211-3. [DOI] [PubMed] [Google Scholar]
  60. Lichvanová Z., Ilbeigi V., Sabo M., Tabrizchi M., Matejčík S. ˇ.. Using corona discharge-ion mobility spectrometry for detection of 2,4,6-Trichloroanisole. Talanta. 2014;127:239–243. doi: 10.1016/j.talanta.2014.04.021. [DOI] [PubMed] [Google Scholar]
  61. Dotan I., Albritton D. L., Lindinger W., Pahl M.. Mobilities of CO2+, N2H+, H3O+, H3O+·H2O, and H3O+· (H2O)­2 ions in N2. J. Chem. Phys. 1976;65:5028–5030. doi: 10.1063/1.432943. [DOI] [Google Scholar]
  62. Myers R. T.. The periodicity of electron affinity. J. Chem. Educ. 1990;67:307. doi: 10.1021/ed067p307. [DOI] [Google Scholar]
  63. Person W. B.. Electron Affinities of Some Halogen Molecules and the Charge-Transfer Frequency. J. Chem. Phys. 1963;38:109–116. doi: 10.1063/1.1733447. [DOI] [Google Scholar]
  64. Taylor T. R., Asmis K. R., Zanni M. T., Neumark D. M.. Characterization of the I3 radical by anion photoelectron spectroscopy. J. Chem. Phys. 1999;110:7607–7609. doi: 10.1063/1.478672. [DOI] [Google Scholar]
  65. Wang X.-B., Woo H.-K., Wang L.-S., Minofar B., Jungwirth P.. Determination of the Electron Affinity of the Acetyloxyl Radical (CH3COO) by Low-Temperature Anion Photoelectron Spectroscopy and ab Initio Calculations. J. Phys. Chem. A. 2006;110:5047–5050. doi: 10.1021/jp060138p. [DOI] [PubMed] [Google Scholar]
  66. Dibble T. S., Zelie M. J., Mao H.. Corrigendum to “Thermodynamics of reactions of ClHg and BrHg radicals with atmospherically abundant free radicals” published in. Atmos. Chem. Phys. 2012;12:10271–10279. doi: 10.5194/acp-12-10271-2012. [DOI] [Google Scholar]
  67. Warneke C., van der Veen C., Luxembourg S., de Gouw J. A., Kok A.. Measurements of benzene and toluene in ambient air using proton-transfer-reaction mass spectrometry: calibration, humidity dependence, and field intercomparison. Int. J. Mass Spectrom. 2001;207:167–182. doi: 10.1016/S1387-3806(01)00366-9. [DOI] [Google Scholar]
  68. Pedder, R. E. ; Wei, J. . Quadrupole Performance: Do Quadrupoles Discriminate Against High Masses?. 46th ASMS Conference on Mass Spectrometry and Allied Topics; ASMS: Orlando, Florida, 1998. [Google Scholar]
  69. Iyer S., He X., Hyttinen N., Kurtén T., Rissanen M. P.. Computational and Experimental Investigation of the Detection of HO2 Radical and the Products of Its Reaction with Cyclohexene Ozonolysis Derived RO2 Radicals by an Iodide-Based Chemical Ionization Mass Spectrometer. J. Phys. Chem. A. 2017;121:6778–6789. doi: 10.1021/acs.jpca.7b01588. [DOI] [PubMed] [Google Scholar]
  70. Song M., He S., Li X., Liu Y., Lou S., Lu S., Zeng L., Zhang Y.. Optimizing the iodide-adduct chemical ionization mass spectrometry (CIMS) quantitative method for toluene oxidation intermediates: experimental insights into functional-group differences. Atmos. Meas. Tech. 2024;17:5113–5127. doi: 10.5194/amt-17-5113-2024. [DOI] [Google Scholar]
  71. Liu Y., Li X., Liu Y., He S., Qiu Y., Song M., Ye J., Lou S., Lu S., Zeng L., Zhang Y.. Speciated Measurement of Bicyclic Peroxy Radicals via Iodide-CIMS and its Implication on OH-Initiated Aromatic Oxidation. Atmos. Chem. Phys. 2025;25:15819–15834. doi: 10.5194/acp-25-15819-2025. [DOI] [Google Scholar]
  72. Wu R., Wang C., Dibble T. S.. First experimental kinetic study of the atmospherically important reaction of BrHg + NO2. Chem. Phys. Lett. 2020;759:137928. doi: 10.1016/j.cplett.2020.137928. [DOI] [Google Scholar]
  73. Gómez Martín J. C., Lewis T. R., Douglas K. M., Blitz M. A., Saiz-Lopez A., Plane J. M. C.. The reaction between HgBr and O3: kinetic study and atmospheric implications. Phys. Chem. Chem. Phys. 2022;24:12419–12432. doi: 10.1039/D2CP00754A. [DOI] [PMC free article] [PubMed] [Google Scholar]
  74. Wang S. X., Zhang L., Li G. H., Wu Y., Hao J. M., Pirrone N., Sprovieri F., Ancora M. P.. Mercury emission and speciation of coal-fired power plants in China. Atmos. Chem. Phys. 2010;10:1183–1192. doi: 10.5194/acp-10-1183-2010. [DOI] [Google Scholar]
  75. Li G., Wang S., Wu Q., Li J., Chen Z., Li J., Wang F., Han D., Li Z., Tang Y., Ouyang D., Liu K.. Mercury emission characteristics and mechanism in the raw mill system of cement clinker production. J. Hazard. Mater. 2022;430:128403. doi: 10.1016/j.jhazmat.2022.128403. [DOI] [PubMed] [Google Scholar]
  76. Junninen H., Ehn M., Petäjä T., Luosujärvi L., Kotiaho T., Kostiainen R., Rohner U., Gonin M., Fuhrer K., Kulmala M., Worsnop D. R.. A high-resolution mass spectrometer to measure atmospheric ion composition. Atmos. Meas. Tech. 2010;3:1039–1053. doi: 10.5194/amt-3-1039-2010. [DOI] [Google Scholar]
  77. Eisele F. L.. Direct tropospheric ion sampling and mass identification. Int. J. Mass Spectrom. Ion Processes. 1983;54:119–126. doi: 10.1016/0168-1176(83)85011-3. [DOI] [Google Scholar]
  78. Perkins M. D., Eisele F. L.. First mass spectrometric measurements of atmospheric ions at ground level. J. Geophys. Res.:Atmos. 1984;89:9649–9657. doi: 10.1029/JD089iD06p09649. [DOI] [Google Scholar]
  79. Davern M. J., Hu Y., West G. V., Kim Y., Francini M. H., Morrison G. C., Zhang Y., Turpin B. J., Surratt J. D.. Online Iodide Chemical Ionization Mass Spectrometry (I-CIMS) Enables Occupational Inhalation Exposure Assessment of 6:2 Fluorotelomer Alcohol (6:2 FTOH) Emitted to Air during Floor Waxing. Environ. Sci. Technol. Lett. 2025;12:607–612. doi: 10.1021/acs.estlett.5c00258. [DOI] [Google Scholar]
  80. Lopez-Hilfiker F. D., Iyer S., Mohr C., Lee B. H., D’Ambro E. L., Kurtén T., Thornton J. A.. Constraining the sensitivity of iodide adduct chemical ionization mass spectrometry to multifunctional organic molecules using the collision limit and thermodynamic stability of iodide ion adducts. Atmos. Meas. Tech. 2016;9:1505–1512. doi: 10.5194/amt-9-1505-2016. [DOI] [Google Scholar]
  81. Heard D. E., Pilling M. J.. Measurement of OH and HO2 in the Troposphere. Chem. Rev. 2003;103:5163–5198. doi: 10.1021/cr020522s. [DOI] [PubMed] [Google Scholar]
  82. Eisele F. L., Tanner D. J.. Measurement of the gas-phase concentration of H2SO4 and methane sulfonic-acid and estimates of H2SO4 production and loss in the atmosphere. J. Geophys. Res. 1993;98:9001–9010. doi: 10.1029/93JD00031. [DOI] [Google Scholar]
  83. Bianchi F., Kurtén T., Riva M., Mohr C., Rissanen M. P., Roldin P., Berndt T., Crounse J. D., Wennberg P. O., Mentel T. F., Wildt J., Junninen H., Jokinen T., Kulmala M., Worsnop D. R., Thornton J. A., Donahue N., Kjaergaard H. G., Ehn M.. Highly Oxygenated Organic Molecules (HOM) from Gas-Phase Autoxidation Involving Peroxy Radicals: A Key Contributor to Atmospheric Aerosol. Chem. Rev. 2019;119:3472–3509. doi: 10.1021/acs.chemrev.8b00395. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

ac5c07120_si_001.pdf (1.1MB, pdf)

Articles from Analytical Chemistry are provided here courtesy of American Chemical Society

RESOURCES