Analysis of chemiluminescence and liquid chromatography-mass spectrometry in 25-hydroxyvitamin D detection using fuzzy logic

Hongkun Liu 1, Sirui Li 2, Kok Wai Wong 2, Yujie Li 1, Xinyi He 1, Shuhao Liang 1, Hui Deng 1, Linlin Zhang 3, Lei Zhang 1,, Jianli Cui 1,2,
PMCID: PMC13066485  PMID: 41776217

Abstract

Vitamin D is an essential nutrient closely associated with the prevention of multiple diseases, including osteoporosis, diabetes, and cardiovascular disorders. The serum level of 25-hydroxyvitamin D [25(OH)D] is the primary biomarker for assessing vitamin D status, and its precise quantification is critical for clinical diagnosis and treatment. Chemiluminescence immunoassay (CLIA) and liquid chromatography–tandem mass spectrometry (LC-MS/MS) are widely used analytical methods; however, methodological discrepancies often lead to inconsistent results that may be further influenced by demographic factors such as age and gender. In this study, we evaluated the consistency and correlation between CLIA and LC-MS/MS measurements using a Generative Fuzzy Inference System (GENFIS). Analysis of 138 serum samples showed that LC-MS/MS produced significantly higher 25(OH)D concentrations than CLIA (p < 0.01; mean difference = 1.33 ± 3.71; 95% CI: − 5.95 to 8.61), though the two methods exhibited strong linear correlation and agreement (Cohen’s Kappa = 0.8257; R² = 0.9075; intraclass correlation coefficient = 0.93). GENFIS analysis indicated a possible 30–40-year female pattern of larger between the detection differences of two methods. To verify this finding, an additional 59 samples were analyzed, revealing a relative risk (RR) of 3.18 (95% CI: 1.71–5.93; p < 0.05) for this subgroup compared with other populations, supporting the GENFIS inference. These results highlight age- and gender-related differences in 25(OH)D measurement between CLIA and LC-MS/MS, providing valuable insights for improving the standardization and clinical interpretation of vitamin D testing.

Supplementary Information

The online version contains supplementary material available at 10.1038/s41598-026-41793-9.

Keywords: Fuzzy Inference System, 25-Hydroxyvitamin D (25(OH)D), Chemiluminescence immunoassay (CLIA), Liquid chromatography-tandem mass spectrometry (LC-MS/MS)

Subject terms: Biochemistry, Biomarkers, Diseases, Endocrinology, Health care, Medical research

Introduction

Vitamins are essential nutrients vital to physiological function and overall health. Among them, vitamin D has attracted considerable attention for its role in maintaining bone integrity, modulating immune responses, and preventing a range of diseases1. Recent studies2,3 suggest that adequate vitamin D levels may reduce the risk of conditions such as Alzheimer’s disease, respiratory infections in infants, and asthma in offspring when supplemented during pregnancy. Additionally, insufficient vitamin D may result in overactivation of the adaptive immune system, thereby increasing the risk of diseases such as type 1 diabetes and atherosclerosis1,4. Despite its importance, vitamin D deficiency remains widespread globally, particularly in developing regions and among low-income populations5,6, imposing a considerable burden on public health systems. Therefore, accurate and timely assessment of vitamin D status is essential to guide supplementation and support preventive healthcare. Testing of serum 25(OH)D has been the focus of global harmonization efforts through the Vitamin D Standardization Program (VDSP), which promotes traceability to the reference measurement procedures listed by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) and the Joint Committee for Traceability in Laboratory Medicine (JCTLM), and calibration against National Institute of Standards and Technology (NIST) standard reference materials7.

Serum 25-hydroxyvitamin D [25(OH)D] concentration is recognized as the most reliable biomarker of vitamin D status1. Among the available analytical methods, chemiluminescence immunoassay (CLIA) is widely used in clinical laboratories and population screenings due to its simplicity, high throughput, automation, and efficiency8,9. Electrochemiluminescence immunoassay (ECLIA), an advanced form of CLIA, further enhances sensitivity, reproducibility, and detection range10. However, CLIA results may vary across laboratories and regions because of inconsistent calibration systems and standardization8. Moreover, CLIA is prone to antibody cross-reactivity and cannot differentiate between vitamin D₂and D₃8,9.

Recently, Liquid chromatography–tandem mass spectrometry (LC-MS/MS) has been recognised as the gold standard for 25(OH)D testing11, addresses some of CLIA’s limitations—particularly its ability to distinguish between vitamin D₂ and D₃. Yet, LC-MS/MS also presents challenges: it demands expert operators11, and cross-reactivity from vitamin D metabolites can interfere with total 25(OH)D measurement8. For instance, Fraser and Milan12 reported that vitamin D₂ can affect the accuracy of results. Given the complementary strengths and limitations of these methods, a systematic comparison is necessary to improve measurement reliability and standardisation.

Although previous studies have compared CLIA and LC-MS/MS, most have focused on specific subpopulations—such as children—and their findings have been inconsistent or even contradictory8,13,14. These discrepancies suggest that the factors underlying inter-method differences are not fully understood. Expanding the study population and integrating advanced analytical approaches, such as artificial intelligence (AI)-based models, may help elucidate these complex relationships and improve standardization efforts.

With rapid advances in AI, its applications in biomedical research are growing rapidly. Although AI has been scarcely applied to vitamin D analysis, it holds great promise for improving interpretability and managing uncertainty. Differences between CLIA and LC-MS/MS measurements may be affected by nonlinear factors such as individual variability and matrix effects—sources of uncertainty that traditional statistical methods struggle to address. AI techniques, particularly those based on fuzzy logic, are well-suited to handle such vagueness and nonlinear relationships15. While neural network models dominate current AI research, they typically require large datasets and often lack interpretability under uncertain conditions15. In contrast, fuzzy logic is effective with smaller datasets and can integrate expert knowledge, making it particularly suitable for studies involving ambiguous variables and limited data15,16. Unlike classical logic, which operates on binary true–false decisions, fuzzy logic accommodates degrees of truth (Fig. 1), allowing nuanced modeling of real-world biological phenomena16.

Fig. 1.

Fig. 1

The differences between classical logic and fuzzy logic; In classical logic, decision boundaries are typically clear and well-defined, as shown in the left part of the figure (classical), where the boundary distinctly separates true (T, 1) and false (F, 0). In fuzzy logic, decision boundaries are usually ambiguous, as illustrated in the right.

In medical and life sciences, many analytical concepts are inherently uncertain. For example, in differential analysis, statements such as “matrix effects may affect measurement accuracy” contain vague qualifiers that traditional models cannot quantify. Liu et al.17 demonstrated that fuzzy logic–based AI models can effectively mitigate such uncertainties. Fuzzy logic has been successfully applied to various domains of biomedical data analysis and differential diagnosis18,19. For instance, Taylan et al.19 used fuzzy neural inference to assess clinical differences among 17 cardiovascular diseases, illustrating its strong potential in medical data interpretation20.

In summary, to improve differential analysis between CLIA and LC-MS/MS, this study measures serum 25(OH)D levels in a diverse population under defined inclusion and exclusion criteria using advanced detection instruments. By integrating fuzzy inference modeling with conventional data analysis, this research aims to uncover interpretable patterns underlying method discrepancies, providing new insights for standardizing vitamin D detection and enhancing clinical decision-making.

Materials and methods

Datasets

This study utilized the datasets obtained from Sichuan Taikang Hospital, with formal permission granted by the hospital’s authorized personnel. The data were retrospectively and randomly selected from a diverse population who underwent health checkups and received 25(OH)D testing by chemiluminescence immunoassay (CLIA) between October 2024 and February 2025. On the same day, these samples were reanalyzed using liquid chromatography–tandem mass spectrometry (LC-MS/MS) at the Precision Laboratory of the hospital’s Medical Laboratory Department.

The dataset excluded inpatients and individuals with cardiovascular diseases, severe immune deficiencies, or major infectious diseases. All personal identifiers were removed before analysis, ensuring complete anonymity and eliminating privacy concerns.

CLIA testing was performed using the Roche cobas e 801 analyzer with the Elecsys Vitamin D total II kit (Roche Diagnostics GmbH). The assay is based on a competitive binding principle, with a quantification range of 3–100 ng/mL. LC-MS/MS analyses were conducted using the AB Sciex QTRAP 6500 instrument, with reagents provided by Guangdong Zhongke Qingzi Medical Technology Co., Ltd. All procedures followed manufacturer protocols, and quality control was performed for each analytical batch. The detection instruments used in this study are certified by the China Food and Drug Administration (CFDA) and represent the most advanced testing systems currently approved in China.

Data analysis

Data were processed and analyzed using MATLAB R2020b, Microsoft Excel 2016, and PyCharm 2022.3. Linear regression analysis was used to evaluate the correlation between CLIA and LC-MS/MS measurements, while a paired t-test assessed the mean difference between the two methods for the same cohort. The regression equation is expressed as:

graphic file with name d33e431.gif 1

where f(x) indicates the dependent variable (LC-MS/MS testing results), x represents the independent variable (CLIA testing results), p1 is the slope, and p2 is the intercept.

The agreement between CLIA and LC-MS/MS is evaluated employing two complementary methods. First, the intraclass correlation coefficient (ICC) is calculated to assess measurement consistency, where ICC < 0.40 indicates poor consistency and ICC > 0.75 means excellent consistency.The ICC is calculated from between-group (MSB) and within-group (MSW) variances according to the formula:

graphic file with name d33e439.gif 2

where k represents the number of evaluated columns. Since this study assessed the result differences between two testing methods, k is equal to 2.

Second, the Bland-Altman analysis was performed to assess the mean bias and the limits of agreement (LOA) between the two testing methods. The LOA were calculated as:

graphic file with name d33e447.gif 3

where SD represents the standard deviation of the differences. This method allows illustration of systematic bias between CLIA and LC-MS/MS results.

The classification consistency between the two methods was evaluated utilizing Cohen’s Kappa analysis, with a Kappa coefficient > 0.8 indicating strong consistency. The collected quantitative data were expressed as mean ± standard deviation (x̄±s).

Fuzzy logic analysis

In biomedical analyses, relationships among variables are often nonlinear and influenced by uncertain factors such as age and gender. Fuzzy logic offers a powerful framework to address such uncertainties by allowing continuous rather than binary reasoning. Its flexibility enables adjustment of variable types and weights, thereby enhancing the interpretability of complex patterns underlying measurement differences.

This study employed a Generative Fuzzy Inference System (GENFIS) integrated with Fuzzy C-Means (FCM) clustering to automatically generate fuzzy rules from the data. The FCM algorithm partitions the dataset into n clusters, assigning each data point a membership degree for each cluster. The objective function is given by:

graphic file with name d33e461.gif 4

where D denotes the total number of data points, N the number of clusters, m > 1 the fuzzy overlap parameter controlling the degree of cluster fuzziness, xi the ith data point, cⱼ the center of the jth cluster, and µiⱼ the membership degree of xi in cluster j21. Following standard practice for biomedical modeling, the number of clusters (n) was set to 3 and the overlap parameter (m) to 2.

In this model, CLIA and LC-MS/MS results, age, and gender were input variables, while the difference between CLIA and LC-MS/MS measurements was defined as the output variable. The structure of the fuzzy inference system is illustrated in Fig. 2. All fuzzy modeling was conducted using MATLAB R2020b.

Fig. 2.

Fig. 2

The structure of fuzzy logic model.

GENFIS summarizes the relationships between inputs (CLIA, LC-MS/MS, age, sex) and the output (method difference) through interpretable IF–THEN rules. FCM identifies graded data patterns, which GENFIS then translates into weighted fuzzy rules that quantify how specific input combinations contribute to observed inter-method variations. This approach enhances explainability while maintaining predictive precision.

Results

Samples collection & testing results

According to the inclusion and exclusion criteria outlined above (Sect. 2.1), this study collected a total of 138 samples, including 60 males (43.5%) and 78 females (56.5%), with a mean age of 40.75 ± 15.59 (Table 1). The detection range of CLIA is from 3.0 ng/mL to 55.2 ng/mL, with a mean of 20.71 ± 9.22 ng/mL; the testing range of LC-MS/MS is between 5.09 ng/mL and 56.24 ng/mL, with a mean of 22.04 ± 11.23 ng/mL(Fig. 3). The difference range between the two methods is −4.56 ng/mL to 13.19 ng/mL, with a mean difference of 1.33 ± 3.71 ng/mL (Table 1).

Table 1.

Description of Sample Information and Testing Results.

Items Value (total number or n (%)/Mean + SD)
Total samples 138
Age 40.75 ± 15.59
Male 43.5% (60)
Female 56.5% (78)
CLIA (ng/mL) 20.71 ± 9.22
LC-MS/MS (ng/mL) 22.04 ± 11.23
Difference (ng/mL) 1.33 ± 3.71

Fig. 3.

Fig. 3

LC-MS/MS Chromatograms of Internal Standards in Different Samples; Subfigures A, B, and C display the chromatograms of the internal standard 25(OH)D3, while subfigures D, E, and F display the chromatograms of the internal standard 25(OH)D2.

Data analysis of the testing differences

According to Table 2, the paired t-test results for the two detection methods provide a P-value < 0.05, indicating statistically significant differences. The average of the LC-MS/MS testing results are higher than those of CLIA (Table 1).

Table 2.

The results of the bland-altman analysis and t-test.

Items Results value
MSB 102.08
MSW 3.87
ICC 0.93
The mean difference 1.33
LOA range −5.95-8.61
P-value (t-test) < 0.01

The result of ICC was calculated as 0.93 (> 0.75), meaning high consistency.between the two detection methods. Bland-Altman analysis shows the mean difference of 1.33 ng/mL between LC-MS/MS and CLIA detection, with a SD of 3.71 ng/mL. The range of LOA between − 5.95 ng/mL and 8.61 ng/mL (Fig. 4).

Fig. 4.

Fig. 4

Bland-Altman analysis for the differences between the testing method of CLIA and LC-MS/MS;The confidence interval equals the mean (1.33) plus or minus 1.96 times the standard deviation (SD). The SD of the differences is 3.71.

Using the clinical threshold of 20 ng/mL for 25(OH)D deficiency as the cutoff for Cohen’s Kappa analysis22, the agreement rate of the two methods is calculated as the proportion of samples (126) where both methods produce results ≥ 20 ng/mL or < 20 ng/mL out of the total sample size (138), with the formula:

graphic file with name d33e667.gif 5

The calculated agreement rate is 126/138 = 91.3%, and the Cohen’s Kappa coefficient is 0.83(> 0.80), suggesting strong consistency between the two methods at the 20 ng/mL threshold (Table 3).

Table 3.

The results of classification agreement rate between the two methods and Cohen’s Kappa.

CLIA ≥ 20 ng/mL LC-MS/MS ≥ 20 ng/mL LC-MS/MS < 20 ng/mL
60 (43.48%) 5 (3.62%)
CLIA < 20 ng/mL 7 (5.07%) 66 (47.83%)
Cohen’s Kappa 0.83

Table 4; Fig. 5 show information related to the linear regression. The adjusted R² indicates a strong linear relationship, with approximately 91% of the variance in the dependent variable explained by the independent variable. The sum of squared residuals (SSE) and root mean square error (RMSE) are small, suggesting low prediction errors.

Table 4.

The general results of linear regression analysis.

Regression Statistics
The regression equation LC-MS/MS = 1.161×CLIA-2.009
0.9082
Adjusted R² 0.9075
Coefficients (with 95% CI) p1 = 1.161(1.099, 1.224)
p2=−2.009 (−3.427, −0.5901)
SSE 1586
RMSE 3.415

Fig. 5.

Fig. 5

The linear relationships between the testing results of LC-MS/MS and CLIA.

Clinical threshold agreement and reclassification

Utilizing decision thresholds of 20 and 30 ng/mL22, cross-classification of CLIA and LC-MS/MS showed that 12/138 (8.70%; Wilson 95% CI 5.0–14.6.0.6%) individuals would change diagnostic category depending on the method. At 20 ng/mL, the number 5 and 7 (CLIA ≥ 20 and LC–MS/MS < 20 = 5; CLIA < 20 & LC-MS/MS ≥ 20 = 7). At 30 ng/mL, the number is 2 and 10 (CLIA ≥ 30 and LC–MS/MS < 30 = 2; CLIA < 30 & LC–MS/MS ≥ 30 = 12). The overall agreement was 91.3%, with Cohen’s kappa is 0.83.

Fuzzy logic establishment and analysis

Both the fuzzy membership functions and fuzzy rules were automatically derived from the observed input-output data rather than manually defined, ensuring an objective, data-driven rule generation process. After GENFIS identified the input and output data, it generated fuzzy membership functions and three data-driven fuzzy rules (as shown in Fig. 6), each corresponding to a cluster in the input-output space (Table 5). The input variables Input1, Input2, Input3, and Input4 represent the CLIA testing result, LC-MS/MS testing result, gender, and age, respectively, while the Output represents the differences of CLIA and LC-MS/MS testing. The differences are normalized to a range of [0, 1] based on the equation:

Fig. 6.

Fig. 6

The membership function of the fuzzy logic analysis for the differences between the two testing methods. Low is represented by the area under the blue line, Medium by the orange, and High by the yellow.

Table 5.

The GENFIS rules information.

Rules CLIA Level LC-MS/MS Level Gender Age Output (Difference) Normalized Weight
R1 Low Low Female 30–40 Medium 0.386
R3 Medium Medium Female 40–60 Medium 0.32
R2 High High Female 30–40 High 0.294
graphic file with name d33e899.gif 6

The following are the fuzzy rules extracted from the input-output data:

R1: If (CLIA (in1) testing results is low) and (LC-MS/MS testing results (in2) is low) and (gender (in3) is low) and (age (in4) is low) then (difference (output) is low).

R2: If (CLIA testing results is medium) and (LC-MS/MS testing results is medium) and (gender is medium) and (age is medium) then (difference is medium).

R3: If (CLIA testing result is high) and (LC-MS/MS testing result is high) and (gender is high) and (age is high) then (difference is high).

As a result, when 25(OH)D levels exceed the threshold (20 ng/mL), they are more likely to result in greater differences, contributing significantly to the variability. The differences across age groups in males are not significant, whereas in females, the differences are most pronounced in the 30–40 age group (Table 6). The rule surface visualization in Fig. 7 further shows that higher differences are concentrated within this demographic group.

Table 6.

The results of fuzzy logic output in the women’s differences between different ages.

Age The output of GENFIS
20 0.299
30 0.348
35 0.349
40 0.342
50 0.285
> 60 ≦ 0.27

Note: The detection threshold for both methods is set at 20 ng/mL; when the results of the two methods are consistent, the trends of the remaining results are the similar.

Fig. 7.

Fig. 7

The three-dimensional plot of GEMFIS rule surface.

Fuzzy Reasoning Verification

To validate the output results of GENFIS, this study collected an additional 59 samples for verification. The experimental group consisted of women aged 30–40 years (28 samples), while the control group included men or women outside the 30–40 age range (31 samples). The average difference between the detection methods was calculated as 1.70 ng/ml, which was used as the threshold to classify the differences into high (≥ 1.70 ng/ml) and low (< 1.70 ng/ml) categories. Table 7 shows the proportion of high differences in the experimental group was 3.18 times that of the control group (RR = 3.18, 95% CI: 1.71–5.93). This result provides preliminary support for the GENFIS hypothesis that women aged 30–40 may amplify the differences between the detection methods. The CI excludes 1 (p < 0.05), indicating a statistically significant difference between the groups.

Table 7.

Relative Risk (RR) of High Differences Between Detection Methods.

Experinental group High difference Low difference
0.536 (15/28) 0.226 (7/31)
Control group 0.464 (13/28) 0.774 (24/31)
Relative Risk (RR) 3.31
95% CI 1.71–5.93

Disucussion

As one of the essential fat-soluble vitamins in the human body, studies have shown that vitamin D is not only involved in bone health but also plays a key role in various diseases23,24. Research shows25 that prediabetic individuals with vitamin D levels above the sufficiency threshold (25(OH)D > 30ng/ml) have a 38% lower risk of developing type 2 diabetes (T2D) compared to those with severe deficiency (25(OH)D < 12ng/ml). During the COVID-19 pandemic, vitamin D has played a positive role in preventing infections26. Thus, monitoring vitamin D levels is important for promoting overall health development. 25(OH)D is the primary metabolic form of vitamin D in the body, and CLIA and LC-MS/MS are the two most commonly used methods for detecting 25(OH)D14, each with its own advantages and limitations. Through comparing the differences between these two widely used testing methods, the consistency of 25(OH)D detection can be improved, thereby facilitating further advancements in these two detection technologies.

The results of this study indicated that the average detection values of LC-MS/MS are higher than those of CLIA, with a statistically significant difference (P < 0.05) (see Tables 1 and 2). This is attributed to the differing testing principles of the two methods: CLIA cannot effectively distinguish between D2 and D3, leading to overall lower detection levels compared to LC-MS/MS. The findings of this study differ from those of other related studies8,14, which may be due to differences in the study population, reagents, and detection equipment; this will be discussed in detail in later sections. Linear regression analysis revealed an R² > 0.90, indicating a high degree of fit between the two detection methods. Bland-Altman analysis demonstrated an ICC > 0.9 for the two methods, suggesting excellent consistency. Additionally, a Cohen’s Kappa coefficient > 0.8 indicated near-perfect agreement in the classification results of the two methods, with an agreement rate exceeding 90%.

The average 25(OH)D levels detected in this study are all below 30 ng/mL, slightly above 20 ng/mL. In some studies25,27, 25(OH)D levels below 30 ng/mL are considered indicative of vitamin D insufficiency. Given that medical checkups and treatments typically follow the principle of proximity, this suggests that vitamin D levels in this region still warrant attention. Other related studies8,13,14 have shown similar patterns in vitamin D levels, whereas the study by Biswasarma et al. revealed a more concerning situation: the average vitamin D level among women in Bangladesh was below 20 ng/mL28. A commonality between these studies and the present study is that they were all conducted in Asian countries, though not all of these regions are low-income areas. For example, the region in this study has a per capita Gross Domestic Product (GDP) exceeding 15,000 USD29, reaching the level of high-income countries. It suggests that the relatively low vitamin D levels observed in this study may be more closely associated with the region’s dietary habits and climatic conditions. The research by Lips et al.30 also confirmed that vitamin D levels in Asian countries are lower than those in the Americas and Europe. They proposed that this is related to the more frequent consumption of dairy products among Western populations30. Consequently, Asian countries should continue to monitor the vitamin D levels of their populations. Although the mean bias is small, a limited percentage of cases (8.70%, 95% CI 5.0%–14.6%) near clinical thresholds may be reclassified depending on the method used, showing some affections for threshold-based decision-making. In addition, small bias It may arise when assays are not fully aligned to IFCC reference procedures or uniformly calibrated to NIST Standard Reference Materials (SRMs)31. Routine participation in Vitamin D External Quality Assessment Scheme (DEQAS) and alignment with VDSP recommendations can reduce such discrepancies and mitigate threshold‑dependent reclassification. In practice, laboratories should verify traceability, monitor external quality performance, and interpret results near decision thresholds with particular caution, especially when changing methods7.

GENFIS algorithm produces fuzzy rules by extracting the relationships between input and output variabless—in other words, fuzzy rules are defined based on those input–output patterns rather than being manually specified—and then uses these rules to describe the relationships among fuzzy sets (clusters)32. Due to its ability to handle ambiguous concepts15, fuzzy logic is frequently applied in biomedical research. Another key feature of fuzzy logic is its capacity to perform analysis with limited data types and quantities. In light of these advantages and the project’s requirements, fuzzy logic is employed to further explore the distinctions between the two detection techniques. In this study, inference analysis was conducted based on data from 138 samples. The membership functions for age and gender reveal that all clusters tend toward 1 (female) in gender, while in age, high level overlaps significantly with low and medium in the 30–40 age range. According to GENFIS analysis, a tentative pattern of greater between the differences of two testing methods in women aged 30–40 emerged but this hypothesis-generating observation requires confirmation in larger cohorts with multivariable adjustment. In addition,, female samples outnumber male ones, and although this imbalance is modest, it may subtly influence model inference. Compared with traditional methods8, which typically rely on linear relationships, fuzzy logic is more capable of capturing ambiguous or uncertain situations. For example, in this study, potential influences of gender and age on the differences between LC-MS/MS and CLIA detection methods can be better addressed employing fuzzy logic. This advantage has already been showed in previous research33. However, another advantage of fuzzy logic is its ability to adapt inference based on changes in variables without altering the overall model structure34. This will be discussed in detail in the next section on the study’s limitations.

The study is limited by the challenges of the sample collection, resulting in certain limitations. The gender imbalance in the samples could introduce slight bias into the results and have a limited impact on the validation of the fuzzy logic analysis. Furthermore, the large number of female samples and their overlap increase the model’s sensitivity to this group, potentially causing overfitting33. This phenomenon may be attributed to women’s greater attention to health indicators. According to the China Health and Nutrition Survey35, the proportion of women participating in physical examination is significantly higher than that of men, which may be a key factor contributing to the gender imbalance in the samples. The validation results provide preliminary support for the fuzzy logic inference, the aged 30–40 women’s testing differences 30–40 are more pronounced, which may be associated with greater fluctuations in hormone levels during this age range, thereby affecting vitamin D metabolism. Zhao et al.36 investigated the impact of sex hormones on vitamin D levels, suggesting that fluctuations in sex hormone-binding globulin (SHBG) levels in women may interfere with the accuracy of vitamin D level measurements. It means that the GENFIS-identified 30–40 female pattern should be interpreted cautiously, as the dataset lacked other affection factors such as BMI, hormonal status, and dietary intake, and the validation set was limited. In future studies, a larger sample size and additional variables should be included to further validate these findings. This study employs fuzzy logic to analyze the variables that are difficult to be handled by traditional methods, such as the potential influence of gender and age on testing differences. These variables lack clear boundaries, and fuzzy logic proves more effective in modeling underlying trends. According to the results of the fuzzy reasoning verification, despite an RR of 3.18 with a 95% CI of 1.71–5.93—meaning the interval excludes 1 (p < 0.05) and thus demonstrates statistical significance—further enlarging the sample size would help to strengthen the robustness of the fuzzy logic model’s validation. Future studies should expand collaboration efforts and collect samples through multicenter and perspectives to better train the model, further enhancing the precision and reliability of the results. It will promote advancements in detection technology and contribute positively to the development of precision medicine and public health3740.

This study introduced a novel use of fuzzy logic, to analyze and compare the differences between LC-MS/MS and CLIA in 25(OH)D testing, providing new findings and insights for differential analysis. The data analysis in this study is based on the detection results from the Roche e 801 and AB Sciex QTRAP 6500 instruments. Compared to previous similar studies, these two devices exhibit greater technological advancement. For instance, as mentioned earlier, the AB Sciex QTRAP 6500 is currently the most advanced instrument certified by the CFDA. To the best of our knowledge, these two instruments are at the forefront of medical laboratory platforms in China. This indicates that the results obtained in this study are more precise and accurate, significantly enhancing the reliability of the analytical conclusions and providing a solid foundation for elucidating the differences between the detection methods.

Conclusion

By broadening the study population and integrating fuzzy logic–based analytical techniques, this research provides a novel perspective on the methodological differences between CLIA and LC-MS/MS for measuring serum 25(OH)D. The results indicate that LC-MS/MS consistently yielded significantly higher values than CLIA (p < 0.01), while both methods demonstrated strong linear correlation and high consistency. GENFIS further revealed a distinct pattern in which women aged 30–40 years exhibited greater inter-method discrepancies, suggesting potential demographic influences on measurement variation. By incorporating fuzzy logic into the comparative analysis of two advanced detection techniques, this study enhances both the reliability and interpretability of differential assessment. Nonetheless, the gender imbalance in the sample population introduces a potential source of bias that may influence the conclusions. Future research should therefore focus on achieving more balanced sampling and validating the observed patterns across larger and more diverse cohorts to strengthen the generalizability and robustness of vitamin D detection standardization.

Supplementary Information

Below is the link to the electronic supplementary material.

Supplementary Material 1 (24.7KB, docx)
Supplementary Material 2 (15.8KB, xlsx)

Author contributions

All authors have approved the manuscript and declare that it is not under consideration elsewhere. There are no conflicts of interest to disclose. The specific contributions of each author are as follows: Hongkun Liu served as the principal investigator and was responsible for study conception and manuscript drafting.Jianli Cui and Lei Zhang, as department professor, provided overall supervision, critical guidance, and final approval of the research.Kok Wai Wong and Sirui Li, professors and lecturers at the School of Information Technology, Murdoch University, advised on model architecture, refined the fuzzy-logic framework, and assisted in selecting appropriate AI methodologies.Hui Deng, Xinyi He, and Shuhao Liang managed sample acquisition, performed statistical analyses, and compiled the data.Yujie Li oversaw the quality control of the LC-MS/MS assays.Linlin Zhang contributed to manuscript revision and offered recommendations on experimental design.

Data availability

All data generated or analysed during this study are included in this published article and its supplementary information files.

Declarations

Competing interests

The authors declare no competing interests.

Ethic statement

The study was reviewed and approved by the Ethics Committee of Sichuan Taikang Hospital (approval number: SCTK-IRB-2025-003). This research was conducted as a retrospective study using laboratory testing data from Sichuan Taikang Hospital. The Ethics Committee granted a waiver of informed consent in accordance with the Declaration of Helsinki and relevant national guidelines, as the study involved the analysis of anonymized data and posed no risk to participants. All methods were performed in accordance with the relevant guidelines and regulations.

Footnotes

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Contributor Information

Lei Zhang, Email: 23172495@qq.com.

Jianli Cui, Email: 1439471329@qq.com.

References

  • 1.Liu, X., Baylin, A. & Levy, P. D. Vitamin D deficiency and insufficiency among US adults: Prevalence, predictors and clinical implications. Br. J. Nutr.119, 928–936 (2018). [DOI] [PubMed] [Google Scholar]
  • 2.Ertilav, E., Barcin, N. E. & Ozdem, S. Comparison of serum free and bioavailable 25-hydroxyvitamin D levels in Alzheimer’s disease and healthy control patients. Lab. Med.52, 219–225 (2021). [DOI] [PubMed] [Google Scholar]
  • 3.Camargo, C. A. Jr et al. Cord-blood 25-hydroxyvitamin D levels and risk of respiratory infection, wheezing, and asthma. Pediatrics127, e180–e187 (2011). [DOI] [PubMed] [Google Scholar]
  • 4.Carbone, F. et al. Vitamin D in atherosclerosis and cardiovascular events. Eur. Heart J.44, 2078–2094 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Niedermaier, T. et al. Vitamin D food fortification in European countries: The underused potential to prevent cancer deaths. Eur. J. Epidemiol.37, 309–320 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Radonsky, V. et al. Alert for the high prevalence of vitamin D deficiency in adolescents in a large Brazilian sample. J Pediatr (Rio J)100, 360–366 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Centers for Disease Control and Prevention (CDC). Participant Protocol for Vitamin D Standardization-Certification Program. CDC. (2023). Available at: https://www.cdc.gov/clinical-standardization-programs/media/pdfs/2024/04/Vitamin-D-Protocol-508.pdf (accessed 24 Oct 2025).
  • 8.Wang, J. et al. Comparison of the serum total 25-hydroxyvitamin D concentrations using chemiluminescent immunoassay and liquid chromatography-tandem mass spectrometry in children. J Nutr Sci Vitaminol (Tokyo)68, 181–188 (2022). [DOI] [PubMed] [Google Scholar]
  • 9.Wise, S. A. et al. Interlaboratory comparison of 25-hydroxyvitamin D assays—impact of 3-epi-25-hydroxyvitamin D3 on assay performance. Anal. Bioanal Chem.414, 333–349 (2022). [DOI] [PubMed] [Google Scholar]
  • 10.Keyfi, F. et al. Evaluation of 25-OH vitamin D by high performance liquid chromatography: Validation and comparison with electrochemiluminescence. J. Anal. Sci. Technol.9, 1–6 (2018). [Google Scholar]
  • 11.Shin, S. Y. et al. Measurement of serum total vitamin D (25-OH) using automated immunoassay in comparison with liquid chromatography tandem-mass spectrometry. J. Clin. Lab. Anal.27, 284–289 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Fraser, W. D. & Milan, A. M. Vitamin D assays: Past and present debates, difficulties, and developments. Calcif. Tissue Int.92, 118–127 (2013). [DOI] [PubMed] [Google Scholar]
  • 13.Jamali, M. A. et al. Prevalence of vitamin D deficiency among healthy young adults at Kuwait University. Cureus16, e66844 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Ma, S. J. et al. Consistency study of chemiluminescence immunoassay and liquid chromatography-tandem mass spectrometry in detecting serum 25-hydroxyvitamin D in children. J. Clin. Exp. Med.23, 91–94 (2024). (in Chinese). [Google Scholar]
  • 15.Sarker, P. & Roy, S. A comprehensive review and analysis of the plethora of branches of medical science and bioinformatics based on fuzzy logic. In Fuzzy Logic Applications in Computer Science and Mathematics 267–278 (2023).
  • 16.Suzuki, A. & Negishi, E. Fuzzy logic systems for healthcare applications. J Biomed Sustain Healthc Appl4, 1 (2024). [Google Scholar]
  • 17.Liu, H. et al. A fuzzy risk assessment model used for assessing the introduction of African swine fever into Australia from overseas. Artif. Intell. Agric.7, 27–34 (2023). [Google Scholar]
  • 18.Singh, S. K., Abolghasemi, V. & Anisi, M. H. Fuzzy logic with deep learning for detection of skin cancer. Appl. Sci.13, 8927 (2023). [Google Scholar]
  • 19.Taylan, O. et al. Early prediction in classification of cardiovascular diseases with machine learning, neuro-fuzzy and statistical methods. Biology12, 117 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Khushal, R. & Fatima, U. A novel fuzzy three-valued logic computational framework in machine learning for medicine dataset. Comput. Biol. Med.186, 109636 (2025). [DOI] [PubMed] [Google Scholar]
  • 21.Gafa, C. Fuzzy systems from data. Towards Data Science (2022). Available at: https://towardsdatascience.com/fuzzy-systems-from-data-51a53bcbc9b1 (accessed 16 Apr 2025).
  • 22.Holick, M. F. et al. Evaluation, treatment, and prevention of vitamin D deficiency: An Endocrine Society clinical practice guideline. J. Clin. Endocrinol. Metab.96, 1911–1930 (2011). [DOI] [PubMed] [Google Scholar]
  • 23.White, J. H. Emerging roles of vitamin D-induced antimicrobial peptides in antiviral innate immunity. Nutrients14, 284 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Wimalawansa, S. J. Physiology of vitamin D—focusing on disease prevention. Nutrients16, 1666 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.di Filippo, L. & Giustina, A. Vitamin D deficiency and type 2 diabetes: The dangerous link between two modern pandemics. J. Clin. Endocrinol. Metab.110, e905–e906 (2025). [DOI] [PubMed] [Google Scholar]
  • 26.Charoenngam, N., Shirvani, A. & Holick, M. F. Vitamin D and its potential benefit for the COVID-19 pandemic. Endocr. Pract.27, 484–493 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Herrick, K. A. et al. Vitamin D status in the United States, 2011–2014. Am. J. Clin. Nutr.110, 150–157 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Biswasarma, S. C. et al. Prevalence and determinants of vitamin D deficiency at a CLIA laboratory in Dhaka. Bangladesh J. Nucl. Med.27, 89–94 (2024). [Google Scholar]
  • 29.2022, G. D. P. Top 100 Cities List: All 13 Cities in Jiangsu Exceed 400 Billion, Top 10 Sees Changes. Yicai.com (2024). Available at: https://web.archive.org/web/20240131000000/https://www.yicai.com/news/101703850.html (accessed 27 Mar 2024).
  • 30.Lips, P., de Jongh, R. T. & van Schoor, N. M. Trends in vitamin D status around the world. JBMR Plus5, e10585 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Wise, S. A. et al. Commutability assessment of new standard reference materials (SRMs) for determining serum total 25-hydroxyvitamin D using ligand binding and liquid chromatography–tandem mass spectrometry (LC-MS/MS) assays[J]. Anal. Bioanal. Chem.417 (12), 2539–2561 (2025). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Bardossy, A. & Duckstein, L. Fuzzy rule-based modeling with applications to geophysical, biological, and engineering systems (CRC Press, 2022). [Google Scholar]
  • 33.Liu, K. Application of artificial intelligence (AI) techniques to risk assessment of African swine fever (ASF) for Australia. Doctoral dissertation, Murdoch University (2023).
  • 34.Zadeh, L. A. Fuzzy logic, granular, fuzzy, and soft computing. In Granular, Fuzzy, and Soft Computing 19–49 (Springer, 2023).
  • 35.Zhang, B., Wang, H. & Du, S. China health and nutrition survey, 1989–2019. In Encyclopedia of Gerontology and Population Aging 943–948 (Springer International Publishing, 2022).
  • 36.Zhao, D. et al. Serum vitamin D and sex hormones levels in men and women: The Multi-Ethnic Study of Atherosclerosis (MESA). Maturitas96, 95–102 (2017). [DOI] [PMC free article] [PubMed]
  • 37.Hou, H., Zhang, R. & Li, J. Artificial intelligence in the clinical laboratory. Clin. Chim. Acta559, 119724 (2024). [DOI] [PubMed] [Google Scholar]
  • 38.Reyes-García, C. A. & Torres-García, A. A. Fuzzy logic and fuzzy systems. In Biosignal Processing and Classification Using Computational Learning and Intelligence 153–176 (Academic Press, 2022).
  • 39.Moodley, K. Artificial intelligence (AI) or augmented intelligence? How big data and AI are transforming healthcare: Challenges and opportunities. S. Afr. Gen. Pract.5, 170–175 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Xie, H., Jia, Y. & Liu, S. Integration of artificial intelligence in clinical laboratory medicine: Advancements and challenges. Interdiscip. Med.2, e2023005 (2024).

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Material 1 (24.7KB, docx)
Supplementary Material 2 (15.8KB, xlsx)

Data Availability Statement

All data generated or analysed during this study are included in this published article and its supplementary information files.


Articles from Scientific Reports are provided here courtesy of Nature Publishing Group

RESOURCES