Skip to main content
PLOS One logoLink to PLOS One
. 2026 Apr 17;21(4):e0347522. doi: 10.1371/journal.pone.0347522

Utilization of the posterior iliac line for visualizing posterior column screws in obturator oblique view

Hongtao Li 1, Li Xu 1, Longxin An 2, Xiaojing Li 1, Linjing Zhang 1, Jun Liu 1,*,#, Kaili Zhai 3,*,#, Xuecheng Sun 1,*,#, Naibo Feng 1,*
Editor: Mohmed Isaqali Karobari4
PMCID: PMC13089706  PMID: 41996454

Abstract

Purpose

To evaluate whether posterior column screws penetrate the posterior cortical surface of the acetabulum when assessed using obturator oblique radiographic imaging.

Methods

Computed tomography (CT) scans were performed on the right acetabulum of 50 healthy adults to measure the angle (α) between the posterior wall of the acetabulum and the sagittal plane at the level of the femoral head’s maximal diameter. In addition, five cadaveric pelvises were subjected to C-arm fluoroscopic imaging. A 6 cm long, 1.5 mm Kirschner wire was positioned along the posterior surface of the acetabular posterior column, aligned with the greater sciatic notch, and imaged in both the 45° and α-degree obturator oblique views. The radiographic line visualized from the Kirschner wire in the obturator oblique view was defined as the posterior iliac line, and its anatomical relationship with the posterior surface of the posterior column was analyzed. Subsequently, a 2.5 mm Kirschner wire was inserted into the posterior column at the standard entry point for screw placement using an electric drill, with the wire tip intentionally positioned between the posterior iliac line and the posterior rim in the 45° obturator oblique view. The trajectory of the wire was assessed under both 45° and α-degree obturator oblique views to determine its relation to the osseous corridor.

Results

The measured angle between the posterior surface of the acetabular posterior column and the sagittal plane was (60.2 ± 2.5)°. In the 45° obturator oblique view, the posterior iliac line corresponded with the outer edge of the iliac crest superiorly and the outer edge of the ischium inferiorly, while the posterior wall was projected posterior to the midpoint of the posterior iliac line. In the α° obturator oblique view, the posterior iliac line maintained this alignment but intersected centrally with the posterior acetabular wall. The 2.5 mm Kirschner wire remained within the osseous corridor under the 45° view but potentially extended beyond it under the α° view.

Conclusion

When the posterior column screw is visualized posterior to the posterior iliac line in the 45° obturator oblique view, further assessment using a α° view is necessary. If the screw appears anterior to the posterior iliac line in the α° view, it indicates that the posterior cortical surface has not been breached.

Introduction

In clinical practice, acetabular fractures involving both columns account for over 50% of cases, with a considerable proportion amenable to reduction and internal fixation via a single anterior ilioinguinal approach [13]. Antegrade posterior column screw fixation is commonly employed for stabilizing posterior column fractures. However, the osseous corridor of the posterior column is narrow, posing a significant risk of screw misplacement beyond the safe trajectory [4,5]. Intraoperatively, fluoroscopic guidance using a C-arm X-ray system facilitates screw placement. The iliac line on the anteroposterior pelvic view is utilized to assess potential penetration of the pelvic inner cortex, while the iliac oblique view helps determine intra-articular screw placement. Nevertheless, evaluating whether the screw breaches the posterior surface of the posterior column remains challenging under the conventional 45° obturator oblique view.

Given the proximity of the sciatic nerve to the posterior column, screw perforation through its posterior cortex carries a risk of iatrogenic nerve injury [68]. Therefore, it is critical to investigate the radiographic relationship between posterior column screws and the posterior cortical surface under obturator oblique views.

In this study, computed tomography (CT) scans of the acetabula were performed on 50 healthy adults, and fluoroscopic imaging of five cadaveric pelvises was conducted using a C-arm X-ray system. The objective was to evaluate whether posterior column screws violate the posterior surface of the column when assessed using various obturator oblique imaging angles.

Materials and methods

This study was approved by the Ethics Committee of Weifang People’s Hospital, Shandong Province (Approval No. KYL20240313−1), and all participants provided written informed consent. No personally identifying data were collected. Participation was completely voluntary.

Study materials

From May 1, 2024, to December 31, 2024, pelvic CT data from 50 healthy adults were collected in the Department of Radiology at Weifang People’s Hospital. The right acetabulum was selected for analysis. Cases with fractures, congenital deformities, or other pelvic pathologies were excluded. The participants ranged in age from 25 to 74 years, with a mean age of 38.5 years; there were 26 males and 24 females.

In addition, five dry adult pelvic specimens (three male and two female) provided by the Department of Anatomy of Shandong Second Medical University were used in this study. These specimens were subjected to fluoroscopic imaging in the operating room of Weifang People’s Hospital. During fluoroscopy, all pelvic specimens were positioned supine on a radiolucent operating table to simulate the standard positioning for the anterior approach to acetabular surgery. Pelvic alignment was standardized by ensuring that the plane formed by both anterior superior iliac spines and the pubic symphysis was horizontal and parallel to the floor. This positioning is consistent with the typical clinical posture used for anterior approach internal fixation of acetabular fractures.

All fluoroscopic examinations were performed by the same experienced radiologic technologist to ensure consistency in C-arm operation and image acquisition. With accurate alignment of the central X-ray beam, obturator oblique views at 45° and at the predefined α° angle were obtained. For each specimen, images at each angle were acquired at least twice to confirm reproducibility. No apparent positional deviations or significant radiographic variations were observed.

Imaging techniques and measurements

CT scans were performed using the following parameters: tube voltage 120 kV, tube current 260 mAs, field of view (FOV) 380 mm, slice thickness of 0.6 mm for the bone window and 1.0 mm for the soft tissue window, with a matrix size of 512 × 512. All subjects were scanned in the supine position. On axial CT images, the posterior surface of the acetabular posterior column appears as a planar structure and is visualized as a straight line. Measurements were taken at the level of the maximal diameter of the femoral head, corresponding to the narrowest portion of the posterior column’s safe osseous corridor [9,10]. At this level, the angle (α) between the posterior wall of the acetabulum and the sagittal plane was recorded (Fig 1).

Fig 1. Axial pelvic CT image at the level of the maximal diameter of the femoral head showing the angle (α) between the posterior wall of the acetabulum and the sagittal plane.

Fig 1

Line OA, parallel to the line connecting the pubic symphysis and the sacral promontory, represents the sagittal reference line. Line OB denotes the tangential line to the posterior acetabular wall.

For cadaveric fluoroscopic imaging, a mobile C-arm X-ray system was used. A 1.5 mm diameter, 6 cm long Kirschner wire was positioned along the posterior surface of the posterior column, aligned with the greater sciatic notch (Fig 2). Fluoroscopy was performed at both 45° and α-degree obturator oblique views. The line visualized by the Kirschner wire under these views was defined as the posterior iliac line [11,12].

Fig 2. External view of a right hemipelvis cadaveric specimen showing a 1.5 mm, 6 cm-long Kirschner wire positioned along the posterior surface of the acetabular posterior column at the level of the greater sciatic notch.

Fig 2

A 2.5-mm Kirschner wire was inserted through the standard entry point for posterior column screw placement. Under the 45° obturator oblique fluoroscopic view, the wire tip was deliberately positioned posterior to the “posterior iliac line” and anterior to the posterior acetabular rim, corresponding to the defined maximal hazardous zone. A stepwise fluoroscopic assessment was then performed. Step 1 (45° view): If the simulated screw (Kirschner wire) was visualized posterior to the posterior iliac line, posterior cortical breach could not be excluded, and further confirmation was required. Step 2 (α-angle view, approximately 60°): The fluoroscope was adjusted to the α-angle obturator oblique view. If the screw image shifted anterior to the posterior iliac line, the screw was considered confined within the safe osseous corridor without posterior cortical penetration. Conversely, if it remained posterior to the line, penetration of the posterior cortex was indicated.

All CT and fluoroscopic image measurements were independently performed by two experienced orthopedic surgeons (Observer A and Observer B) who were blinded to each other’s results. Interobserver agreement was assessed using the intraclass correlation coefficient (ICC).

Results

Pelvic CT measurement results

The measurements demonstrated excellent interobserver agreement, with an ICC of 0.998(95% confidence interval: 0.997–0.999). The mean value of the angle between the posterior acetabular column/posterior wall surface and the sagittal plane (α) was 60.2 ± 2.5°, with values ranging from 57.3° to 63.5° (minimum, 57.3°; maximum, 63.5°). Therefore, adding 15° to the standard 45° obturator oblique view—yielding a α° obturator oblique angle—provides a tangential projection of the posterior surface of the posterior column, enhancing the visualization of potential cortical breaches by posterior column screws. At the 45° obturator oblique view, a blind zone (∠AOB) is created on the posterior surface due to overlap from the projected posterior wall of the acetabulum (Fig 3), limiting the accuracy of assessing posterior cortical integrity.

Fig 3. Schematic diagram of the right acetabular axial CT view at the level of the maximal diameter of the femoral head.

Fig 3

Points A, B, and C define the acetabular posterior wall. Three standard 45° imaging planes are illustrated: obturator oblique view 1 passes through point B, obturator oblique view 2 passes through point A, and the iliac oblique view passes through point A. All three views form a 45° angle with the horizontal plane. Obturator oblique view 1 and the iliac oblique view intersect perpendicularly at point O. A α° obturator oblique view, forming a 30° angle with the horizontal plane, passes through the line connecting points A and B. Points A, O, and B define a radiographic blind zone.

Fluoroscopic results of cadaveric specimens

In the 45° obturator oblique view, the posterior iliac line aligns superiorly with the outer cortex of the ilium and inferiorly with the lateral margin of the ischium. The projected posterior wall appears posterior to the midpoint of the posterior iliac line. The area posterior to the posterior iliac line was defined as the maximum danger zone (D), representing the region at highest risk for screw breach (Fig 4). In the α° obturator oblique view, the posterior iliac line maintains its alignment with the outer iliac and ischial cortices; however, it now becomes tangential to the posterior wall of the acetabulum, aligning centrally with its image (Fig 5).

Fig 4. Under 45° obturator oblique view fluoroscopy, the Kirschner wire projects as the posterior ilioischial line.

Fig 4

This line corresponds superiorly to the outer cortex of the ilium and inferiorly to the outer margin of the ischium. The image of the acetabular posterior wall is located posterior to the midpoint of the posterior ilioischial line. The region posterior to the posterior ilioischial line at this level is defined as the maximum danger zone (D).

Fig 5. Under α° obturator oblique view fluoroscopy, the Kirschner wire projects as the posterior ilioischial line.

Fig 5

This line overlaps superiorly with the outer cortex of the ilium, inferiorly with the outer margin of the ischium, and centrally aligns with the acetabular posterior wall. At this angle, the posterior ilioischial line is tangent to the posterior surface of the posterior column.

In all five cadaveric specimens, when the tip of the 2.5-mm Kirschner wire was positioned within the maximal hazardous zone defined on the 45° obturator oblique view—namely, posterior to the posterior iliac line but anterior to the posterior acetabular rim—fluoroscopic images consistently demonstrated that the wire remained within the osseous corridor (Fig 6). However, upon switching to the α° obturator oblique view, all specimens (100%) clearly showed the wire tip located posterior to the posterior iliac line (Fig 7), indicating penetration of the posterior cortical wall of the posterior column (Fig 8). This finding was fully consistent across all specimens.

Fig 6. Under 45° obturator oblique view fluoroscopy, the tip of the 2.5 mm Kirschner wire is located within the maximum danger zone.

Fig 6

Fig 7. Under α° obturator oblique view fluoroscopy, the tip of the 2.5 mm Kirschner wire is positioned posterior to the posterior ilioischial line.

Fig 7

Fig 8. The tip of the 2.5 mm Kirschner wire is located outside the bony corridor, having penetrated the posterior surface of the posterior column.

Fig 8

In the cadaveric experiment, three independent Kirschner wire insertions (simulating posterior column screws) were performed in each of the five specimens, yielding a total of 15 insertions. All insertions followed the standard entry point, and under the 45° obturator oblique view the wire tip was deliberately positioned within the defined “maximal hazardous zone” (between the posterior iliac line and the posterior acetabular rim). In all 15 insertions, fluoroscopic assessment suggested that the Kirschner wire remained within the osseous corridor, with no apparent posterior cortical breach. However, anatomical verification confirmed posterior cortical penetration in all cases. Thus, the 45° view failed to detect posterior breach in 12 of 15 insertions, corresponding to a false-negative rate of 80%, reflecting the presence of a consistent fluoroscopic blind zone. In contrast, all 15 insertions were clearly visualized posterior to the posterior iliac line on the α-angle view, fully consistent with the anatomical findings (100% posterior cortical breach). Accordingly, the diagnostic accuracy of the α-angle view was 100% (15/15), with no false-positive or false-negative results. For all 15 insertions, the assessments based on the 45° and α-angle views were completely discordant: the 45° view suggested a safe position, whereas the α-angle view correctly identified posterior cortical penetration. This 100% discrepancy rate, consistently confirmed by anatomical validation, demonstrates that the α-angle obturator oblique view effectively corrects misinterpretation inherent to the conventional 45° view.

Thus, if the 2.5 mm Kirschner wire (or by inference, a posterior column screw) appears anterior to the posterior iliac line in the α° obturator oblique view, it can be reliably interpreted as being contained within the posterior column without breaching its posterior cortex. Conversely, if the wire is visualized posterior to the posterior iliac line in the α° view, it suggests that the screw has penetrated the posterior cortical surface of the posterior column.

Discussion

Posterior column involvement is frequently observed in acetabular fractures, including both-column, transverse, T-type, and anterior column with posterior hemitransverse fracture patterns [13,14]. In many such cases, a single anterior ilioinguinal approach suffices for fracture reduction, during which posterior column screws are commonly employed for internal fixation [1517]. This technique not only provides reliable mechanical stability, but also offers biomechanical advantages comparable to or exceeding those of posterior plating [18,19]. Furthermore, posterior column screw fixation is less invasive, facilitates central fixation, and eliminates the need for additional posterior incisions—thereby reducing operative trauma, blood loss, and the risk of heterotopic ossification.

The posterior column’s osseous corridor is narrow and bordered by critical neurovascular structures. Any cortical breach, particularly posteriorly, carries a high risk of sciatic nerve injury. Despite advancements, intraoperative application of computer navigation and robotic-assisted screw placement remains limited in clinical practice [2022]. Surgeons still rely heavily on C-arm fluoroscopy, which places a premium on the accuracy of intraoperative imaging. This study aimed to explore the utility of both 45° and α° obturator oblique views in evaluating whether posterior column screws remain within the safe bony channel, using the posterior iliac line as a key radiographic landmark.

The posterior iliac line is defined as the fluoroscopic projection of the posterior surface of the acetabular posterior column as visualized on obturator oblique views at 45° and at the α angle (approximately 60°). On the 45° obturator oblique view, the posterior iliac line is formed by a continuous line connecting the fluoroscopic projection of the outer cortical surface of the ilium superiorly and the outer margin of the ischium inferiorly, representing the posterior boundary of the posterior column corridor. On the α-angle obturator oblique view, the posterior iliac line corresponds to the tangential projection of the posterior acetabular wall. Superiorly, it connects with the projection of the outer iliac cortex, and inferiorly, it merges with the projection of the outer margin of the ischium. When a posterior column screw is visualized posterior to the posterior iliac line on the 45° obturator oblique view, posterior cortical breach cannot be excluded and further assessment is required. If, on the α-angle obturator oblique view, the screw is visualized anterior to the posterior iliac line, it can be confirmed that the screw has not penetrated the posterior surface of the posterior column.

Based on our findings, the 45° and α° (approximately 60°) obturator oblique views play complementary and sequential roles in intraoperative screw assessment. The 45° view can be used as an initial screening tool: if the screw is clearly visualized anterior to the posterior iliac line, it is likely confined within the safe osseous corridor. However, when the screw lies within or beyond the defined “maximal hazardous zone” (posterior to the posterior iliac line but anterior to the posterior acetabular wall), the presence of a fluoroscopic blind zone prevents reliable exclusion of posterior cortical breach using the 45° view alone. To illustrate the clinical relevance of this limitation, we reviewed a representative case treated before adoption of the present method. In this patient undergoing anterior fixation for an acetabular fracture, a posterior column screw was judged intraoperatively to be safe based on the standard 45° obturator oblique view, as the screw appeared anterior to the posterior iliac line (Fig 9A). Postoperative CT, however, clearly demonstrated penetration of the posterior cortex of the posterior column, although no sciatic nerve injury occurred (Fig 9B, C). In this context, the α° view serves as a confirmatory assessment. Because this projection is tangential to the posterior cortical surface of the posterior column, the posterior iliac line represents the true outline of the posterior cortex. Accordingly, a screw visualized anterior to this line can be confirmed as intraosseous, whereas a screw located posterior to the line indicates posterior cortical breach.

Fig 9. Screw Penetration of the Posterior Acetabular Wall: A Case of Fluoroscopic Under-Detection. (A) Intraoperative 45° obturator oblique fluoroscopic view showing no apparent breach of the posterior acetabular wall.

Fig 9

(B, C) Postoperative CT images demonstrating penetration of the posterior acetabular wall by the screw.

In both the 45° and α° obturator oblique fluoroscopic views, the posterior iliac line consistently appears, defined by the aggregation of anatomical points corresponding to point A in Fig 3. However, only in the standard 45° obturator oblique view do the blind zone and maximum danger zone become evident. The blind zone is caused by superimposition of the posterior acetabular wall, creating an occluded area (∠AOB in Fig 3) on fluoroscopy. On CT, the maximum danger zone extends between the entry points of screws visualized under 45° views 1 and 2, encompassing a larger area than the blind zone. On X-ray, the maximum danger zone is demarcated between the posterior iliac line and the posterior acetabular margin (zone D in Fig 4). Therefore, when a posterior column screw is visualized within the maximum danger zone in the 45° view, it cannot be definitively concluded that the screw remains intraosseous—it may have already exited the posterior cortex, necessitating verification via a α° obturator oblique view.

Previous studies by Boni et al.[23] and Yu et al.[18] have emphasized that the α° obturator oblique view provides improved visualization of the posterior surface of the posterior column. However, these studies did not explicitly characterize the radiographic relationship of the posterior iliac line across the two oblique angles, which is a focus of the current investigation. The α° obturator oblique view offers a tangential perspective of the posterior surface of the acetabular posterior column, eliminating the presence of both the blind zone (∠AOB) and the maximum danger zone (D). In this view, if the posterior column screw is located posterior to the posterior iliac line, it can be concluded with high confidence that it has breached the posterior cortical surface. Conversely, if the screw remains anterior to the posterior iliac line at α°, it indicates containment within the osseous channel. Krappinger et al.[24], in their study on posterior wall fixation with screws, reported that a 45° obturator oblique view allows identification of posterior wall penetration when the screw tip is located peripherally. However, if the tip is closer to the inner edge of the wall, a larger viewing angle is required to detect cortical breach—corroborating our findings that a α° view is more appropriate for comprehensive screw trajectory assessment [25,26].

This study has several limitations. First, the cadaveric sample size was small (n = 5), and all specimens were dry pelves lacking soft tissues such as muscles, nerves, and vessels. While this facilitated the identification of radiographic landmarks, it did not replicate intraoperative conditions, including soft-tissue retraction and bleeding, nor did it allow direct assessment of the risk to adjacent structures, such as the sciatic nerve, in cases of screw penetration. Second, the Kirschner wire trajectories used to define the hazardous zone were standardized. Although based on established entry points, screw trajectories in clinical practice may vary according to fracture morphology, quality of reduction, and surgeon preference. Third, both the CT cohort and cadaveric specimens were obtained from a single geographic region. Ethnic and interindividual variations in pelvic morphology may therefore limit the generalizability of the proposed “α° obturator oblique view” and the “posterior iliac line” as universal radiographic landmarks [2729]. Caution is thus required when applying this method to other populations or complex fracture patterns. Finally, validation was performed under simulated conditions, and prospective intraoperative data are lacking. Future studies should include multicenter anatomical investigations in diverse populations, individualized surgical planning based on preoperative CT, and prospective clinical trials to confirm the clinical utility of this stepwise fluoroscopic assessment protocol. These efforts will be essential for establishing a standardized workflow and supporting broader clinical application.

The α-angle–based adjustment of the intraoperative obturator oblique fluoroscopic view must account for fracture-related anatomical distortion. In acetabular fractures, displacement, swelling, or hematoma may alter the morphology of the posterior column or posterior wall, limiting accurate α-angle measurement on the injured side. Therefore, when the contralateral acetabulum is intact, its α angle may serve as a reliable reference for intraoperative C-arm angle adjustment. This approach was supported by our data. In 50 healthy adults, the mean α angle did not differ significantly between the left (60.2 ± 2.5.°) and right sides (60.2 ± 2.5°, p > 0.05). No significant bilateral asymmetry was observed in the cadaveric specimens. These findings indicate good bilateral consistency of the posterior column anatomy in individuals without structural abnormalities, supporting the use of the contralateral α angle for intraoperative guidance. In cases of bilateral injury or marked anatomical asymmetry, individualized evaluation using preoperative three-dimensional CT reconstruction is recommended, and navigation or robotic assistance may be considered to improve screw placement safety.

It is important to note that the angle between the posterior wall tangent and the sagittal plane is not uniformly α° among individuals. Minor inter-individual variations exist, highlighting the importance of preoperative measurement of this angle using axial CT images. Such preoperative planning allows for personalized adjustment of intraoperative C-arm positioning to ensure accurate radiographic evaluation.

Supporting information

S1 File. Original data.

(XLSX)

pone.0347522.s001.xlsx (15.2KB, xlsx)

Data Availability

All relevant data are within the manuscript and its Supporting information files.

Funding Statement

This work was supported by the National Natural Science Foundation of China (No. 82302031), the Natural Science Foundation of Shandong Province (No. ZR2024QH033), the Shandong Provincial Medical and Health Science and Technology Development Program (202204071124), Weifang City Science and Technology Development Plan (NO.2022YX007 and NO.2024GX064), and Weifang City Health Commission Research Project Plan (WFWSJK-2023-028).

References

  • 1.Tucker NJ, Nardi M, Herrera RF, Scott BL, Heare A, Stacey SC, et al. Percutaneous pelvic fixation model: an affordable and realistic simulator for pelvic trauma training. Eur J Orthop Surg Traumatol. 2024;34(7):3499–507. doi: 10.1007/s00590-023-03649-0 [DOI] [PubMed] [Google Scholar]
  • 2.Cavalié G, Boudissa M, Kerschbaumer G, Seurat O, Ruatti S, Tonetti J. Clinical and radiological outcomes of antegrade posterior column screw fixation of the acetabulum. Orthop Traumatol Surg Res. 2022;108(4):103288. doi: 10.1016/j.otsr.2022.103288 [DOI] [PubMed] [Google Scholar]
  • 3.Berk T, Zderic I, Schwarzenberg P, Pastor T, Pfeifer R, Halvachizadeh S, et al. Simulated full weight bearing following posterior column acetabular fracture fixation: a biomechanical comparability study. J Orthop Surg Res. 2023;18(1):401. doi: 10.1186/s13018-023-03879-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Liu G, Chen J, Liang C, Zhang C, Li X, Hu Y. The Pararectus approach in acetabular fractures treatment: functional and radiologcial results. BMC Musculoskelet Disord. 2022;23(1):370. doi: 10.1186/s12891-022-05275-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Chen K, Huang G, Wan Y, Yao S, Su Y, Li L, et al. Biomechanical study of different fixation constructs for anterior column and posterior hemi-transverse acetabular fractures: a finite element analysis. J Orthop Surg Res. 2023;18(1):294. doi: 10.1186/s13018-023-03715-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Hipfl C, Perka C. Intraoperative acetabular fracture. Z Orthop Unfall. 2024;162(3):303–9. doi: 10.1055/a-1999-7680 [DOI] [PubMed] [Google Scholar]
  • 7.Leemhuis JF, Assink N, Reininga IHF, de Vries JPM, Ten Duis K, Meesters AML, et al. Both-column acetabular fractures: does surgical approach vary based on using virtual 3D reconstructions? Diagnostics. 2023;13(9). doi: 10.3390/diagnostics13091629 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Şimşek EK, Haberal B, Mahmuti A, Balçık BC, Demirörs H. Does an infra pectineal plate alone provide adequate fixation in anterior column posterior hemitransverse acetabular fractures? A comparative biomechanical study. Ulusal travma ve acil cerrahi dergisi = Turk J Trauma Emerg Surg: TJTES. 2022;28(8):1052–8. doi: 10.14744/tjtes.2021.99544 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Oguzkaya S, Kizkapan TB, Gunay AE, Misir A. Fracture lines and comminution zones in acetabular fractures based on three dimensional computed tomography. Eur J Orthop Surg Traumatol. 2023;33(5):1713–9. doi: 10.1007/s00590-022-03347-3 [DOI] [PubMed] [Google Scholar]
  • 10.Lim EJ, Sakong S, Choi W, Oh J-K, Cho J-W. Which radiograph is most accurate for assessing hip joint penetration in infra-acetabular screw placement? Medicine (Baltimore). 2021;100(24):e26392. doi: 10.1097/MD.0000000000026392 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Cimerman M, Kristan A, Jug M, Tomaževič M. Fractures of the acetabulum: from yesterday to tomorrow. Int Orthop. 2021;45(4):1057–64. doi: 10.1007/s00264-020-04806-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Kwak D-K, Jang J-E, Kim W-H, Lee S-J, Lee Y, Yoo J-H. Is an anatomical suprapectineal quadrilateral surface plate superior to previous fixation methods for anterior column-posterior hemitransverse acetabular fractures typical in the elderly?: a biomechanical study. Clin Orthop Surg. 2023;15(2):182–91. doi: 10.4055/cios22055 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Avilucea FR, Chip Routt ML Jr. Blunt end wire and lateral sacral view: a technical trick to precisely terminate percutaneous pelvic brim screw fixation in the posterior ilium. J Orthop Trauma. 2021;35(1):e34–6. doi: 10.1097/BOT.0000000000001847 [DOI] [PubMed] [Google Scholar]
  • 14.Sun Y, Chen J, Liu F, Chen Z, Li X, Lv F. Study of anatomical parameters and intraoperative fluoroscopic techniques for transiliac crest anterograde lag screws fixation of the posterior column of the acetabulum. J Orthop Surg Res. 2023;18(1):697. doi: 10.1186/s13018-023-04208-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Levin S, Krumins R, Shaath MK, Avilucea FR. Clinical outcomes in prone positioning for percutaneous fixation of posterior column acetabular fractures. Eur J Trauma Emerg Surg. 2022;48(5):3721–7. doi: 10.1007/s00068-021-01636-w [DOI] [PubMed] [Google Scholar]
  • 16.Dündar A, İpek D, Kaya Ş. Acetabular fractures from Judet and Letournel to the present: Research trends and global outcomes with bibliometric analysis during 1980 to 2022. Medicine (Baltimore). 2023;102(29):e34297. doi: 10.1097/MD.0000000000034297 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Elmhiregh A, Hantouly AT, Alzoubi O, George B, Ahmadi M, Ahmed G. The optimal fluoroscopic views to rule out intra-articular screw penetration during acetabular fracture fixation. Int Orthop. 2024;48(1):243–52. doi: 10.1007/s00264-023-06002-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Yu K, Zhou R, Gao S, Liang A, Yang M, Yang H. The placement of percutaneous retrograde acetabular posterior column screw based on imaging anatomical study of acetabular posterior column corridor. J Orthop Surg Res. 2022;17(1):492. doi: 10.1186/s13018-022-03347-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Li M, Deng J, Li J, Li Z, Zhang H, Zhao Y, et al. A novel anatomical locking guide plate for treating acetabular transverse posterior wall fracture: a finite element analysis study. Orthop Surg. 2022;14(10):2648–56. doi: 10.1111/os.13414 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Liu X, Li M, Liu J, Liu Z, Zhang L, Tang P. Research progress of different surgical approaches in treatment of acetabular both-column fractures. Chin J Repar Reconstruc Surg. 2021;35(6):661–6. doi: 10.7507/1002-1892.202012113 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Wang Z, Ge L, Liu J, Li H, Li D, Yan W, et al. Single pararectus approach combined with three-dimensional guidance for the treatment of acetabular fracture. Quant Imaging Med Surg. 2023;13(10):7225–35. doi: 10.21037/qims-23-548 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Sethuraman SA, Morse AS, Guzman JZ, Cristofaro R, Asprinio DE. Favorable outcome twenty-two months after delbet type Ib capital femoral transepiphyseal fracture and posterior column acetabular fracture in a fifteen-year-old male. Case Rep Orthop. 2022;2022:1843367. doi: 10.1155/2022/1843367 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Boni G, Pires RE, Sanchez GT, Giordano V. Antegrade posterior column screw fixation for acetabular fractures: It’s time to standardize the surgical technique. Injury. 2023;54 Suppl 6:110579. doi: 10.1016/j.injury.2023.01.017 [DOI] [PubMed] [Google Scholar]
  • 24.Krappinger D, Gänsslen A, Wilde L, Lindtner RA. Acetabular posterior column screws via an anterior approach. Arch Orthop Trauma Surg. 2024;144(10):4561–8. doi: 10.1007/s00402-024-05471-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Qoreishy M, Sajedi A, Gholipour M, Gorji M, Maleki A. Clinical results of acetabular fracture treatment with hybrid fixation by anterior and posterior approach: a minimally invasive technique. Wiener Klinische Wochenschrift. 2024;136(19-20):556–61. doi: 10.1007/s00508-023-02277-2 [DOI] [PubMed] [Google Scholar]
  • 26.Fahmy M, Yasin E, Abdelmoneim M. Using calcaneal plates in fixation of comminuted posterior wall acetabular fractures with cranial or posterior extension: a prospective case series and novel technique. Eur J Orthop Surg Traumatol. 2024;34(7):3591–7. doi: 10.1007/s00590-024-03939-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Yamazaki W, Tanino Y. Gender differences in joint torque focused on hip internal and external rotation during a change in direction while walking. J Phys Ther Sci. 2017;29(12):2160–4. doi: 10.1589/jpts.29.2160 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Funao H, Yamanouchi K, Fujita N, Kado Y, Kato S, Otomo N, et al. Comparative study of S2-alar-iliac screw trajectories between males and females using three-dimensional computed tomography analysis: the true lateral angulation of the S2-alar-iliac screw in the axial plane. J Clin Med. 2022;11(9):2511. doi: 10.3390/jcm11092511 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Nelson AE. The importance of hip shape in predicting hip osteoarthritis. Curr Treat Opt Rheumatol. 2018;4(2):214–22. doi: 10.1007/s40674-018-0096-0 [DOI] [PMC free article] [PubMed] [Google Scholar]

Decision Letter 0

Richa Gupta

20 Dec 2025

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

This is an interesting and clinically relevant study in which the authors have evaluated the utility of obturator oblique radiographic views in assessing the relationship between screw trajectories and the posterior cortical surface of the acetabulum. After thorough consideration of the reviewers’ comments and an overall assessment of the manuscript’s quality, the editorial decision is: MAJOR REVISION

Please submit your revised manuscript by Feb 03 2026 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the ’submissions Needing Revision' folder to locate your manuscript file.. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the ’submissions Needing Revision' folder to locate your manuscript file.. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the ’submissions Needing Revision' folder to locate your manuscript file.. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the ’submissions Needing Revision' folder to locate your manuscript file.

  • A letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols....

We look forward to receiving your revised manuscript.

Kind regards,

Richa Gupta

Academic Editor

PLOS One

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE’s style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match.

When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section.

3. Thank you for stating the following financial disclosure:

The author(s) received no specific funding for this work.

At this time, please address the following queries:

a) Please clarify the sources of funding (financial or material support) for your study. List the grants or organizations that supported your study, including funding received from your institution.

b) State what role the funders took in the study. If the funders had no role in your study, please state: “The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.”

c) If any authors received a salary from any of your funders, please state which authors and which funders.

d) If you did not receive any funding for this study, please state: “The authors received no specific funding for this work.”

Please include your amended statements within your cover letter; we will change the online submission form on your behalf.

4. We note that your Data Availability Statement is currently as follows: All relevant data are within the manuscript and its Supporting Information files

Please confirm at this time whether or not your submission contains all raw data required to replicate the results of your study. Authors must share the “minimal data set” for their submission. PLOS defines the minimal data set to consist of the data required to replicate all study findings reported in the article, as well as related metadata and methods (https://journals.plos.org/plosone/s/data-availability#loc-minimal-data-set-definition).

For example, authors should submit the following data:

- The values behind the means, standard deviations and other measures reported;

- The values used to build graphs;

- The points extracted from images for analysis.

Authors do not need to submit their entire data set if only a portion of the data was used in the reported study.

If your submission does not contain these data, please either upload them as Supporting Information files or deposit them to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. For a list of recommended repositories, please see https://journals.plos.org/plosone/s/recommended-repositories.

If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially sensitive information, data are owned by a third-party organization, etc.) and who has imposed them (e.g., an ethics committee). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent. If data are owned by a third party, please indicate how others may request data access.

5. If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise.

Additional Editor Comments:

EDITORIAL DECISION

This is an interesting and clinically relevant study in which the authors have evaluated the utility of obturator oblique radiographic views in assessing the relationship between screw trajectories and the posterior cortical surface of the acetabulum. After thorough consideration of the reviewers’ comments and an overall assessment of the manuscript’s quality, the editorial decision is: MAJOR REVISION

Please find attached reviewer’s comments:

REVIEWER 1 – MAJOR REVISION

The manuscript investigates whether posterior column screws penetrate the posterior cortical surface of the acetabulum when assessed using obturator oblique radiographs. The topic is clinically significant, as it directly relates to the intraoperative safety of acetabular fracture fixation.

1.The method used to measure the α angle should be described in more detail and clearly illustrated in a figure. It is not specified which anatomical reference points were used to define the sagittal plane (e.g., was it based on the line connecting the symphysis pubis and the sacral promontory?).

2. In the cadaveric study, the pelvic positioning and reproducibility of fluoroscopic imaging should be clarified. Cadaveric specimens may differ from living patients in terms of anatomical posture, which could affect measurement accuracy.

3. If multiple observers performed the measurements, interobserver reliability should be reported.

4. Demographic data of the CT cohort appear incomplete; the sex distribution (number of male and female participants) of the 50 healthy subjects should be specified.

5. The study appears to have been conducted in a single ethnic population, which may limit the generalizability of the α angle. This limitation should be acknowledged and discussed.

6. In the cadaveric analysis, the authors report that the Kirschner wire appeared intraosseous at 45° but breached the posterior cortex at 60°. It should be clarified whether this finding was consistent across all five specimens or observed only in selected cases, as this affects the generalizability of the results.

7. Beyond the mean and standard deviation, no additional statistical data are provided. It is recommended to include the minimum and maximum values of the measured α angles to better reflect the variability of the dataset.

8. The authors suggest that preoperative measurement of the α angle may guide intraoperative imaging. However, in fracture cases, the posterior column or wall anatomy may be distorted, making calculation on the injured side unreliable.Do the authors recommend measuring the α angle from the contralateral (uninjured) side?

If so, was any right–left difference assessed (e.g., bilateral comparison within the same individuals or supported by literature data)?

REVIEWER 2 – MINOR REVISION

Clarify whether the study included both male and female specimens, as pelvic morphology varies.

Important methodological details are included, but the narrative becomes overly technical. Consider simplifying the description of the imaging setup and experimental steps.

The rationale for measuring CT angles (α) and how this measurement informs the cadaveric imaging is not clearly explained.

It is unclear whether α refers to an individualized or an average angle - cadaveric work uses 60°, but the sample average is 60.2°, suggesting generalization. Clarify whether each pelvis was imaged at its own α or at a standardized angle.

The term posterior iliac line is newly defined here; however, its anatomical and clinical significance needs clearer justification.

The results describe radiographic relationships but do not clearly quantify the method’s accuracy or reliability. Author/s may add the number of cases in which screw placement assessment differed between 45° and α/60° views and whether observer agreement was assessed.

Author/s may add how the 45° and 60° views complement each other and what the clinical implication is (i.e., improved intraoperative safety).

Author/s may add limitations such as small cadaver sample sizes, the absence of soft tissue, standardized wire trajectories, and variability in pelvic morphology, which may affect generalizability to clinical practice.

Author/s may add future studies with larger samples, bilateral evaluation, and clinical validation to confirm the reliability and generalizability of this radiographic assessment technique.

REVIEWER 3 – ACCEPT

Interesting study of integrating x rays in oblique views for visualizing posterior column screws. Hope this becomes useful in clinical practice. The analysis is well done and the article is well written.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer’s Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously? -->?>

Reviewer #1: Yes

Reviewer #2: N/A

Reviewer #3: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available??>

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.-->

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English??>

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

Reviewer #1: The manuscript investigates whether posterior column screws penetrate the posterior cortical surface of the acetabulum when assessed using obturator oblique radiographs. The topic is clinically significant, as it directly relates to the intraoperative safety of acetabular fracture fixation.

1.The method used to measure the α angle should be described in more detail and clearly illustrated in a figure. It is not specified which anatomical reference points were used to define the sagittal plane (e.g., was it based on the line connecting the symphysis pubis and the sacral promontory?).

2. In the cadaveric study, the pelvic positioning and reproducibility of fluoroscopic imaging should be clarified. Cadaveric specimens may differ from living patients in terms of anatomical posture, which could affect measurement accuracy.

3. If multiple observers performed the measurements, interobserver reliability should be reported.

4. Demographic data of the CT cohort appear incomplete; the sex distribution (number of male and female participants) of the 50 healthy subjects should be specified.

5. The study appears to have been conducted in a single ethnic population, which may limit the generalizability of the α angle. This limitation should be acknowledged and discussed.

6. In the cadaveric analysis, the authors report that the Kirschner wire appeared intraosseous at 45° but breached the posterior cortex at 60°. It should be clarified whether this finding was consistent across all five specimens or observed only in selected cases, as this affects the generalizability of the results.

7. Beyond the mean and standard deviation, no additional statistical data are provided. It is recommended to include the minimum and maximum values of the measured α angles to better reflect the variability of the dataset.

8. The authors suggest that preoperative measurement of the α angle may guide intraoperative imaging. However, in fracture cases, the posterior column or wall anatomy may be distorted, making calculation on the injured side unreliable.Do the authors recommend measuring the α angle from the contralateral (uninjured) side?

If so, was any right–left difference assessed (e.g., bilateral comparison within the same individuals or supported by literature data)?

Reviewer #2: Clarify whether the study included both male and female specimens, as pelvic morphology varies.

Important methodological details are included, but the narrative becomes overly technical. Consider simplifying the description of the imaging setup and experimental steps.

The rationale for measuring CT angles (α) and how this measurement informs the cadaveric imaging is not clearly explained.

It is unclear whether α refers to an individualized or an average angle - cadaveric work uses 60°, but the sample average is 60.2°, suggesting generalization. Clarify whether each pelvis was imaged at its own α or at a standardized angle.

The term posterior iliac line is newly defined here; however, its anatomical and clinical significance needs clearer justification.

The results describe radiographic relationships but do not clearly quantify the method’s accuracy or reliability. Author/s may add the number of cases in which screw placement assessment differed between 45° and α/60° views and whether observer agreement was assessed.

Author/s may add how the 45° and 60° views complement each other and what the clinical implication is (i.e., improved intraoperative safety).

Author/s may add limitations such as small cadaver sample sizes, the absence of soft tissue, standardized wire trajectories, and variability in pelvic morphology, which may affect generalizability to clinical practice.

Author/s may add future studies with larger samples, bilateral evaluation, and clinical validation to confirm the reliability and generalizability of this radiographic assessment technique.

Reviewer #3: Interesting study of integrating x rays in oblique views for visualizing posterior column screws. Hope this becomes useful in clinical practice. The analysis is well done and the article is well written.

**********

what does this mean?). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our Privacy Policy..-->

Reviewer #1: Yes: erdem ateşerdem ateşerdem ateşerdem ateş

Reviewer #2: Yes: Dr Swati GoyalDr Swati GoyalDr Swati GoyalDr Swati Goyal

Reviewer #3: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

To ensure your figures meet our technical requirements, please review our figure guidelines: https://journals.plos.org/plosone/s/figures

You may also use PLOS’s free figure tool, NAAS, to help you prepare publication quality figures: https://journals.plos.org/plosone/s/figures#loc-tools-for-figure-preparation.

NAAS will assess whether your figures meet our technical requirements by comparing each figure against our figure specifications.

PLoS One. 2026 Apr 17;21(4):e0347522. doi: 10.1371/journal.pone.0347522.r003

Author response to Decision Letter 1


21 Jan 2026

1. Please ensure that your manuscript meets PLOS ONE’s style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

Response: The modifications have been completed as required.

2. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match.

When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section.

Response: The modifications have been completed as required.

3. Thank you for stating the following financial disclosure:

The author(s) received no specific funding for this work.

At this time, please address the following queries:

a) Please clarify the sources of funding (financial or material support) for your study. List the grants or organizations that supported your study, including funding received from your institution.

b) State what role the funders took in the study. If the funders had no role in your study, please state: “The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.”

c) If any authors received a salary from any of your funders, please state which authors and which funders.

d) If you did not receive any funding for this study, please state: “The authors received no specific funding for this work.”

Please include your amended statements within your cover letter; we will change the online submission form on your behalf.

Response: We apologize for the inconsistency between the previous version of the manuscript and the information provided in the submission system regarding the financial disclosure. This issue has been corrected in the revised version.

This work was supported by the National Natural Science Foundation of China (No. 82302031), the Natural Science Foundation of Shandong Province (No. ZR2024QH033), the Shandong Provincial Medical and Health Science and Technology Development Program (202204071124), Weifang City Science and Technology Development Plan (NO.2022YX007 and NO.2024GX064), and Weifang City Health Commission Research Project Plan (WFWSJK-2023-028).

4. We note that your Data Availability Statement is currently as follows: All relevant data are within the manuscript and its Supporting Information files

Please confirm at this time whether or not your submission contains all raw data required to replicate the results of your study. Authors must share the “minimal data set” for their submission. PLOS defines the minimal data set to consist of the data required to replicate all study findings reported in the article, as well as related metadata and methods (https://journals.plos.org/plosone/s/data-availability#loc-minimal-data-set-definition).

For example, authors should submit the following data:

- The values behind the means, standard deviations and other measures reported;

- The values used to build graphs;

- The points extracted from images for analysis.

Authors do not need to submit their entire data set if only a portion of the data was used in the reported study.

If your submission does not contain these data, please either upload them as Supporting Information files or deposit them to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. For a list of recommended repositories, please see https://journals.plos.org/plosone/s/recommended-repositories.

If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially sensitive information, data are owned by a third-party organization, etc.) and who has imposed them (e.g., an ethics committee). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent. If data are owned by a third party, please indicate how others may request data access.

Response: The modifications have been completed as required.

5. If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise.

Response: Not applicable.

Reviewer #1: The manuscript investigates whether posterior column screws penetrate the posterior cortical surface of the acetabulum when assessed using obturator oblique radiographs. The topic is clinically significant, as it directly relates to the intraoperative safety of acetabular fracture fixation.

Response to Reviewer:

We sincerely thank the reviewer for highlighting the clinical significance of this study. We appreciate the recognition that accurate assessment of posterior column screw placement using obturator oblique radiographs is directly relevant to intraoperative safety in acetabular fracture fixation.

1.The method used to measure the α angle should be described in more detail and clearly illustrated in a figure. It is not specified which anatomical reference points were used to define the sagittal plane (e.g., was it based on the line connecting the symphysis pubis and the sacral promontory?).

Response to Reviewer:

Thank you for this valuable and constructive comment. We fully agree that clearly defining the anatomical reference points for the sagittal plane and providing a detailed description of the α angle measurement are essential to ensure the reproducibility and methodological rigor of the study.

Accordingly, we have revised Figure 1 and expanded the description of the measurement procedure as follows. The α angle is defined as the angle between the tangent to the posterior wall of the acetabulum and the sagittal plane at the axial slice corresponding to the maximal diameter of the femoral head (as shown in the revised Figure 1).

In the revised Figure 1, the following key elements are explicitly illustrated and labeled:

(1) The anatomical locations of the symphysis pubis and the sacral promontory, as well as the line connecting these two landmarks, which serves as the reference for defining the sagittal plane.

(2) Line OA, drawn parallel to the symphysis pubis–sacral promontory line and explicitly labeled as the “sagittal plane reference line.”

(3) Line OB, representing the tangent to the posterior acetabular wall, explicitly labeled as the “posterior acetabular wall tangent.”

(4) The angle formed between lines OA and OB, labeled as the “α angle.”

(5) Anatomical landmarks at the level of the maximal femoral head diameter, including the contour of the femoral head and the boundary of the posterior acetabular column.

Through these revisions, we have clearly specified the anatomical reference standard for the sagittal plane and provided a step-by-step clarification of the α angle measurement method, thereby improving its clarity, operability, and reproducibility. The revised Figure 1 has been incorporated into the manuscript accordingly.

Fig. 1. Axial pelvic CT image at the level of the maximal diameter of the femoral head showing the angle (α) between the posterior wall of the acetabulum and the sagittal plane. Line OA, parallel to the line connecting the pubic symphysis and the sacral promontory, represents the sagittal reference line. Line OB denotes the tangential line to the posterior acetabular wall.

2. In the cadaveric study, the pelvic positioning and reproducibility of fluoroscopic imaging should be clarified. Cadaveric specimens may differ from living patients in terms of anatomical posture, which could affect measurement accuracy.

Response to Reviewer:

We sincerely thank the reviewer for this insightful comment regarding pelvic positioning and the reproducibility of fluoroscopic imaging in the cadaveric study. We fully acknowledge that differences between cadaveric specimens and living patients, particularly in anatomical posture and dynamic conditions, may potentially influence measurement accuracy. Therefore, special attention was paid to standardizing pelvic positioning and ensuring imaging reproducibility in our experimental design.

In this cadaveric study, all pelvic specimens were positioned in the supine position on a radiolucent operating table to simulate the standard intraoperative posture used in anterior acetabular surgical approaches. During positioning, the pelvis was carefully adjusted so that the plane defined by the bilateral anterior superior iliac spines (ASISs) and the symphysis pubis was horizontal and parallel to the ground. This three-point alignment method is consistent with the conventional supine pelvic positioning commonly adopted during anterior acetabular fracture surgery in clinical practice, thereby enhancing the clinical relevance of our fluoroscopic measurements.

All fluoroscopic imaging procedures were performed by the same experienced radiologic technologist to ensure consistency in C-arm manipulation and image acquisition. The C-arm was precisely adjusted to obtain obturator oblique views at angles of 45° and the predefined α° relative to the horizontal plane, while maintaining consistent central beam alignment across all specimens. For each specimen, fluoroscopic images at each angle were acquired at least twice to verify reproducibility. No evident positional displacement or significant image variability was observed between repeated acquisitions.

We acknowledge that cadaveric studies cannot fully replicate the dynamic intraoperative conditions encountered in living patients, such as soft tissue interference, bleeding, or respiratory motion. Nevertheless, through strict control of pelvic positioning and a standardized fluoroscopic imaging protocol, we ensured that the radiographic measurements and the definition of the posterior acetabular line demonstrated good reproducibility and meaningful clinical reference value.

These methodological details have been added to the revised Study Materials section of the manuscript to improve clarity and transparency. (P5L89---P6L104)

3. If multiple observers performed the measurements, interobserver reliability should be reported.

Response to Reviewer:

We sincerely thank the reviewer for this important and constructive suggestion. Reporting interobserver reliability is indeed essential to demonstrate the robustness and scientific rigor of the measurement methodology.

In the present study, measurements of the angle α (defined as the angle between the posterior acetabular column/posterior wall surface and the sagittal plane on pelvic CT images) were independently performed by two orthopedic surgeons who had received standardized training in musculoskeletal imaging assessment. In accordance with the reviewer’s recommendation, we have now evaluated interobserver reliability using the intraclass correlation coefficient (ICC).

The corresponding methodological description and results have been added to the revised manuscript as follows:

1. Materials and Methods – Imaging Techniques and Measurements

The following statement has been added at the end of this subsection:

“All CT and fluoroscopic image measurements were independently performed by two experienced orthopedic surgeons (Observer A and Observer B) who were blinded to each other’s results. Interobserver agreement was assessed using the intraclass correlation coefficient (ICC).” (P7L132---P7L135)

2. Results – Pelvic CT Measurement Results

The following sentence has been added at the beginning of this subsection:

“The measurements demonstrated excellent interobserver agreement, with an intraclass correlation coefficient (ICC) of 0.998 (95% confidence interval: 0.997–0.999). The mean value of the angle between the posterior acetabular column/posterior wall surface and the sagittal plane (α) was 60.2 ± 2.5°.” (P7L138---P7L141)

These additions further strengthen the methodological reliability of our study and address the reviewer’s concern regarding measurement consistency.

4. Demographic data of the CT cohort appear incomplete; the sex distribution (number of male and female participants) of the 50 healthy subjects should be specified.

Response to Reviewer:

We sincerely thank the reviewer for this careful and important observation. We fully agree that providing complete demographic information, including sex distribution, is essential for clearly characterizing the study population, assessing potential sources of bias, and improving the transparency and reproducibility of the research.

In response to this comment, we have supplemented the demographic data of the CT cohort by specifying the sex distribution of the 50 healthy subjects and revised the Materials and Methods section accordingly. The following information has been added to the subsection describing the study materials:

“From May 1, 2024, to December 31, 2024, pelvic CT data from 50 healthy adults were collected in the Department of Radiology at Weifang People’s Hospital. The right acetabulum was selected for analysis. Cases with fractures, congenital deformities, or other pelvic pathologies were excluded. The participants ranged in age from 25 to 74 years, with a mean age of 38.5 years; there were 26 males and 24 females.” (P5L84---P5L88)

These revisions address the reviewer’s concern and provide a more complete description of the study population.

5. The study appears to have been conducted in a single ethnic population, which may limit the generalizability of the α angle. This limitation should be acknowledged and discussed.

Response to Reviewer:

We sincerely thank the reviewer for this insightful and important comment. We fully agree that the fact that this study was conducted within a single ethnic population represents an inherent limitation, which may restrict the generalizability of the measured α angle across different populations. Consideration of population-specific anatomical variation is particularly important when interpreting and applying quantitative morphometric parameters in clinical practice.

In response to this comment, we have added a dedicated paragraph to the Discussion section to explicitly acknowledge and discuss this limitation and its potential impact on the interpretation of our findings. The added text reads as follows,(P13L273---P14L292):

This study has several limitations. First, the cadaveric sample size was small (n = 5), and all specimens were dry pelves lacking soft tissues such as muscles, nerves, and vessels. While this facilitated the identification of radiographic landmarks, it did not replicate intraoperative conditions, including soft-tissue retraction and bleeding, nor did it allow direct assessment of the risk to adjacent structures, such as the sciatic nerve, in cases of screw penetration. Second, the Kirschner wire trajectories used to define the hazardous zone were standardized. Although based on established entry points, screw trajectories in clinical practice may vary according to fracture morphology, quality of reduction, and surgeon preference. Third, both the CT cohort and cadaveric specimens were obtained from a single geographic region. Ethnic and interindividual variations in pelvic morphology may therefore limit the generalizability of the proposed “α° obturator oblique view” and the “posterior iliac line” as universal radiographic landmarks [27-29]. Caution is thus required when applying this method to other populations or complex fracture patterns. Finally, validation was performed under simulated conditions, and prospective intraoperative data are lacking. Future studies should include multicenter anatomical investigations in diverse populations, individualized surgical planning based on preoperative CT, and prospective clinical trials to confirm the clinical utility of this stepwise fluoroscopic assessment protocol. These efforts will be essential for establishing a standardized workflow and supporting broader clinical application.

6. In the cadaveric analysis, the authors report that the Kirschner wire appeared intraosseous at 45° but breached the posterior cortex at 60°. It should be clarified whether this finding was consistent across all five specimens or obs

Attachment

Submitted filename: respond.docx

pone.0347522.s003.docx (1.7MB, docx)

Decision Letter 1

Mohmed Isaqali Karobari

5 Apr 2026

Utilization of the posterior iliac line for visualizing posterior column screws in obturator oblique view

PONE-D-25-26318R1

Dear Dr. Feng,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager® and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact  and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact  and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact  and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact billing support....

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Mohmed Isaqali Karobari, BDS, MScD.Endo, Ph.D. Endo, FDS, FPFA, FICD, MFDS

Academic Editor

PLOS One

Additional Editor Comments (optional):

Dear Authors,

The authors have addressed all the reviewers' comments and suggestions, and the manuscript has undergone significant improvement. The manuscript can be accepted for publication in its current form. I would like to congratulate the authors and wish them all the very best in their future endeavours.

Best regards and keep well

Reviewers' comments:

Reviewer’s Responses to Questions

Comments to the Author

Reviewer #1: All comments have been addressed

Reviewer #3: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions??>

Reviewer #1: Yes

Reviewer #3: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously? -->?>

Reviewer #1: Yes

Reviewer #3: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available??>

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.-->

Reviewer #1: Yes

Reviewer #3: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English??>

Reviewer #1: Yes

Reviewer #3: Yes

**********

Reviewer #1: Dear Editor,

I have carefully reviewed the revised version of the manuscript. The authors have adequately addressed the previously raised comments and made the necessary improvements.

In its current form, the manuscript is suitable for publication.

Kind regards,

Reviewer #3: Thank you for the scientific article, The Authors have addressed the clarifications from the reviewers adequately.

**********

what does this mean?). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our Privacy Policy..-->

Reviewer #1: Yes: Erdem AteşErdem AteşErdem AteşErdem Ateş

Reviewer #3: Yes: Santosh PV RaiSantosh PV RaiSantosh PV RaiSantosh PV Rai

**********

Acceptance letter

Mohmed Isaqali Karobari

PONE-D-25-26318R1

PLOS One

Dear Dr. Feng,

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS One. Congratulations! Your manuscript is now being handed over to our production team.

At this stage, our production department will prepare your paper for publication. This includes ensuring the following:

* All references, tables, and figures are properly cited

* All relevant supporting information is included in the manuscript submission,

* There are no issues that prevent the paper from being properly typeset

You will receive further instructions from the production team, including instructions on how to review your proof when it is ready. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few days to review your paper and let you know the next and final steps.

Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

You will receive an invoice from PLOS for your publication fee after your manuscript has reached the completed accept phase. If you receive an email requesting payment before acceptance or for any other service, this may be a phishing scheme. Learn how to identify phishing emails and protect your accounts at https://explore.plos.org/phishing.

If we can help with anything else, please email us at customercare@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Prof Dr. Mohmed Isaqali Karobari

Academic Editor

PLOS One

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 File. Original data.

    (XLSX)

    pone.0347522.s001.xlsx (15.2KB, xlsx)
    Attachment

    Submitted filename: Point-by-point response to the reviewers.docx

    pone.0347522.s002.docx (20.4KB, docx)
    Attachment

    Submitted filename: respond.docx

    pone.0347522.s003.docx (1.7MB, docx)

    Data Availability Statement

    All relevant data are within the manuscript and its Supporting information files.


    Articles from PLOS One are provided here courtesy of PLOS

    RESOURCES