Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2026 Apr 27.
Published in final edited form as: Magn Reson Med. 2025 Dec 30;95(5):2611–2626. doi: 10.1002/mrm.70234

In vivo meso-scale whole-brain quantitative imaging with tailored MRF on the NexGen 7T scanner

Xiaozhi Cao 1,2,*, Alexander Beckett 3,4,*, Congyu Liao 5,#, Erica Walker 3,4, Zheren Zhu 5, Yurui Qian 5, Mengze Gao 1,2, Nan Wang 1,2, Yimeng Lin 1,2, Lisong Gong 1,6, Matthew A McCready 2, Zhixing Wang 7, Zhitao Li 8, An Vu 5,9, Samantha Ma 10, Gabriel Ramos-Llordén 11, Qiyuan Tian 12, Adam Kerr 2, Yang Yang 5, David A Feinberg 3,4,+, Kawin Setsompop 1,2,+
PMCID: PMC13110358  NIHMSID: NIHMS2165228  PMID: 41472402

Abstract

Purpose:

To push the speed and resolution limit of in-vivo quantitative imaging and enable estimation of quantitative tissue parameters of subtle brain structures that were previously difficult to assess.

Methods:

This study implemented an efficient quantitative imaging approach, 3D-SPI MRF, on the NexGen 7T scanner equipped with a high-performance head-only gradient and 96-channel receiver array. To address challenges associated with performing rapid mesoscale MRF on this system, acquisition and reconstruction mitigation methods were developed and incorporated into the MRF framework, including: i) flip-angle-aware dictionary fitting to account for both B1+ inhomogeneity and voxel-specific RF frequency response, ii) gradient imperfection corrections via Skope measurements that incorporates a new per-TR trajectory rewinder compensation, iii) incorporation of rapid B1+ and B0 mappings into the MRF sequence, and iv) high-temporal motion navigation.

Results:

Whole-brain T1 and T2 maps were obtained at 560-μm isotropic resolution within 4 minutes, where ablation studies demonstrated the necessity of the various mitigation methods implemented in removing bias and artifacts. For comparison, MRF data were acquired using current state-of-the-art method but limited to typical whole-body gradient specifications to demonstrate that the proposed developments resulted in ~3× shorter scan time while producing more accurate parameter maps. Data were also acquired at ~3.8× smaller voxel size, 360-μm isotropic, using the developed technique, to achieve mesoscale multi-parameter quantitative mapping in vivo.

Conclusion:

Tailored 3D MRF acquisition and reconstruction were developed to enable fast and accurate T1 and T2 mapping across the whole-brain at mesoscale resolution on the NexGen 7T scanner.

Keywords: MRF, quantitative mapping, mesoscale, high field

1. Introduction

High-resolution quantitative imaging presents both a challenging and promising frontier in MRI research1. Submillimeter resolution can enable the detection of subtle quantitative tissue parameter changes in small brain structures during early disease pathogenesis2, as well as developmental and aging processes3. These quantitative changes could be used as biomarkers for many applications including myelination abnormality across cortical layers, micro plaques, early-stage fibrosis and iron deposition. However, such quantitative imaging is often limited by long scan time and limited SNR. Compared to traditional contrast-weighted images that can achieve submillimeter-level resolutions (e.g., 0.4-mm or higher4), this level of resolution in quantitative imaging remains difficult to perform, particularly across multiple tissue parameters.

As a rapid, multiparametric quantitative imaging technique, MR Fingerprinting (MRF5) holds the potential to push the limits of resolution in quantitative imaging when effectively implemented on high-performance, high-field scanners. In particular, 7T systems can provide substantially higher SNR, which is crucial for overcoming the signal limitations inherent to ultra-high-resolution imaging. Prior studies have leveraged the increased SNR at 7T for MRF applications targeting nuclei with inherently low signal, such as 23Na6 and 31P7. Other investigations have proposed novel strategies to measure and correct the severe B1+ inhomogeneity commonly observed at 7T8,9 by optimizing MRF acquisitions that enjoy the benefit of high-field SNR avoiding the quantification biases associated with B1+ inhomogeneity. Additionally, the use of shorter readout trajectories, such as radial10,11 or Cartesian12 sampling, has been explored to mitigate distortion and blurring induced by the severe B0 inhomogeneity at 7T. Yet, whole-brain, ultra-high-resolution quantitative imaging at 7T with the use of MRF remains largely unexplored.

Other simultaneous multi-parametric approaches have also been developed at ultra-high field, including QRAGE13, MR-STAT14 and QUICS15. These studies demonstrate the technical feasibility of concurrent quantitative mapping at ultra high fields and highlight the potential of advanced reconstruction and modeling strategies to overcome the intrinsic field inhomogeneities of 7T. Several studies have leveraged parallel transmission coils available on 7T systems to achieve more accurate RF excitation, thereby improving the accuracy of quantitative imaging at 7T16,17. In addition, methods such as MP2RAGE18,19 and 3D-QALAS20 adapted for 7T have pushed the spatial resolution of multi-parametric imaging to the 500–600 μm range, which is the highest reported to date. However, they typically require comparatively long acquisition time of 16–20 minutes for whole-brain coverage.

This study aims to implement a tailored MRF acquisition on a recently-developed Siemens NexGen 7T MAGNETOM Terra Impulse edition scanner21 to shorten the scan time required for submillimeter quantitative imaging and to push the spatial resolution limits. The NexGen 7T scanner features a head-only gradient system with PNS optimization2123 achieving a maximum gradient amplitude (Gmax) of 200 mT/m and a maximum slew rate (SRmax) of 900 T/m/s along with a high-density 96-channel receiver coil that should enable faster k-space traversal, better parallel imaging performance, and higher SNR. The scanner has enabled high resolution fMRI with whole brain coverage24, however, the full exploitation of MRF’s capability on this NexGen 7T scanner faces several challenges, as outlined below:

  1. B0 and B1+ Field Inhomogeneities: in addition to the image blurring and distortions caused by significant B0 inhomogeneities, it also impacts tissue quantification due to changes in the effective flip angle with the limited bandwidth of typical RF pulses used in MRF, especially when water-only excitation is used to avoid fat artifacts. Moreover, local B0 inhomogeneities can cause the frequency of fat to deviate from its theoretical value, making spectral suppression of fat signal more challenging.

  2. Gradient Imperfections: the ability to reach higher slew rates also comes with the risk of introducing additional image artifacts. Moreover, as part of this work, we observed that imperfect k-space trajectory rewinding in the MRF sequence can cause a reduction in signal refocusing, leading to a bias in T2 quantification.

  3. Motion Artifacts: subtle motions can lead to noticeable blurring in ultra-high-resolution imaging, even for a motion-robust sequence like MRF.

To address these challenges, this study integrates fast B1+ and B0 field mapping and high resolution motion-navigation modules into the 3D spiral-projection MRF sequence (3D-SPI MRF2527), along with a redesigned water-only excitation RF pulse for improved fat suppression. External field probes (Skope field camera28) measurements were also employed to characterize gradient imperfections. Notably, this work is the first quantitative imaging study to highlight and correct the FA attenuation bias that is jointly caused by RF frequency response and local off-resonance shift, while also proposing the use of compensation gradients to improve spiral trajectory rewinding for better signal refocusing and thus more accurate quantitative estimation.

The technologies developed in this work were shown to enable accurate whole-brain MRF at 560-μm isotropic resolution in less than 4 minutes. Data were also acquired at 360-μm isotropic resolution with motion correction, to enable, for the first time, multi-parameter quantitative mapping of the human brain in vivo at this high a resolution.

2. Method

2.1. Sequence design

The FISP-based MRF sequence29 used in this work is based on our recently developed 3D-spiral projection-imaging (3D-SPI) MRF sequence25 that contains a 3D-SPI motion-navigator module30. Using this motion-corrected 3D-SPI MRF sequence as a base, we tailored the acquisition to allow it to work well at high resolution with the high-performance and high-field system. A brief description of sequence is provided below, with detailed descriptions in our previous work25. Following an inversion preparation using a 20-ms adiabatic tanh pulse31 with an inversion time (TI) set at 15 ms, there are 500 TRs acquired with a range of varying flip angles (FAs) spanning between 10 and 90 degrees, using different spiral projection interleaves and a constant TR of 12.5 ms. Once the data acquisition for a particular acquisition group across the 500 TRs is completed, there is a 1.2-s sequence deadtime to allow for Mz recovery, resulting in a total acquisition time of 8.7 s per group. The next acquisition group is then performed to gather additional spiral interleaves at each TR for a more comprehensive spatiotemporal k-space encoding. A constant low flip-angle (10 degree) FISP acquisition is also performed during the sequence deadtime across 100 TRs using 3D-SPI encoding to provide 2-mm navigator volume with each acquisition group, where the low-flip angle excitations used still allow for good Mz recovery.

Specific improvements to this sequence performed in this work include:

  1. A rapid joint B1+ and B0 calibration scan at 1.6-mm resolution (Figure 1A) is incorporated into the MRF sequence in place of the conventional 8.7 s dummy scan used to reach steady-state Mz signal (typically the first acquisition group is discarded). This calibration scan is 15 s in total (9 s for B1+ module and 6 s for B0 module), thereby adding only 6.3 s of scan time to the MRF sequence.

    The B1+ estimation module (Figure 2A) employs a 6-ms 240-degree Bloch-Siegert (B-S) adiabatic pulse32 following the excitation pulse, where the RF phase was alternated by 180 degrees between odd and even TRs (Figure 2B). Images from the odd and even TRs are then reconstructed separately, with the B1+ maps derived from their phase difference. This B1+ estimation module employs a 1.6-mm isotropic 3D-SPI spiral readout over 480 TRs with 18.5 ms per TR. The adiabatic B-S pulse design is used here to enable accurate B1+ mapping over a large range of B1+ inhomogeneities. To evaluate its performance, a Fermi-shaped B-S pulse is also implemented and acquired for comparison, as well as the standard Siemens B1+ mapping sequence using TurboFLASH33, which takes about 80 s for the same resolution.

    The sequence diagram for the B0 estimation module is shown in Figure 2D. This module follows a similar acquisition scheme to the B1+ estimation module, using the same k-space sampling trajectory, number of TRs, and reconstruction framework, but now without the B-S pulse. Different TEs (1.7ms and 2.7ms, respectively, resulting in a ΔTE of 1ms) are used across the odd and even TRs to generate phase differences needed to calculate the B0 field map. To evaluate its performance, data were also acquired with the standard Siemens B0 mapping sequence using dual-echo GRE for reference, which takes about 40 s for the same resolution. To ensure the same steady-state longitudinal magnetization Mz is reached at the end of acquisition like the regular MRF acquisition group, the same TR (12.5 ms) and flip angle (10 degree constantly) settings as those in the final 100 TRs of the MRF acquisition are used. This eliminates the need for a separate dummy acquisition group typically required in conventional 3D MRF, thereby saving the acquisition time associated with one entire group (8.7 s).

  2. Compensation of imperfect gradient rewinder (Figure 1B): Gradient imperfections can cause incomplete gradient rewinding in the 3D-SPI MRF acquisition, where the amount of residual gradient area can vary across different interleaves and rotations of the 3D-SPI spiral readouts and hence across TRs. Based on Skope measurements (Supporting Figure S1A), the observed residual gradient moment can reach 50 μs·mT/m. We observed that the variations in this residual gradient moment across TRs could lead to imperfect signal refocusing in the FISP signal pathway, affecting the signal evolution and thus introducing large biases in T2 quantitative estimation. To compensate for this, a TR-specific rewinder compensation gradient blip is added in each TR, where Skope field measurement confirms that this correction is able to reduce the rewinding imperfection by more than 10 fold to less than 5 μs·mT/m (Supporting Figure S1B).

  3. Improved fat suppression: A water-only excitation RF pulse (2.86-ms duration, passband / stopband ripple = 0.46 / 0. 0001) was designed using the SLR algorithm34 (Figure 3A) to provide improved water-only excitation with less fat contamination35 at 7T. Compared to the hard pulse design previously employed in MRF at 3T25,35 and to Gaussian pulse designs (the RF pulse duration was designed to ensure the presence of a zero-crossing around the fat resonance frequency), the frequency response of this SLR pulse (shown in Figure 3B) exhibited markedly lower excitation around the fat frequency (from −900Hz to −1100Hz for example), resulting in reduced fat signal contamination in the presence of local B0 inhomogeneities. Furthermore, the broader Full-Width-at-Half-Maximum (FWHM) of its frequency response also reduces the flip-angle (FA) attenuation effect on the water signal caused by B0 inhomogeneity. This FA attenuation effect and its correction will be described in Section 2.2 below (FA-corrected dictionary fitting).

  4. A Skope trigger was added prior to the spiral readout. This allows the Skope field camera to measure system imperfections, such as the actual spiral trajectory, B0-eddy-current induced phase shifts, and residual gradient moments, which are used for determining the amplitude of trajectory rewinder compensation gradients in ii), and for incorporating into the spiral reconstruction to improve reconstruction fidelity.

  5. To provide higher resolution volume navigators needed for our target mesoscale imaging, the MRF motion navigator module30,36 was modified to increase the volume navigator resolution from 4mm in our previous work to 2 mm here. This was achieved via faster k-space traversal with the head insert gradient as well as increasing the number of navigator acquisitions from 40 TRs to 100 TRs during the sequence deadtime after the MRF acquisition group to allow the encoding of the volume with 2-mm isotropic resolution to be completed during a single deadtime period and maintain motion navigation temporal resolution of 8.7s.

Figure 1.

Figure 1.

(A) Global sequence diagram including an upfront calibration (B1+ and B0 estimation) module, and MRF acquisition and motion-navigator modules that get repeated across MRF acquisition groups.

(B) Sequence diagram of the modified MRF acquisition within a specific TR, where a redesigned SLR pulse is used to reduce fat signal contamination and a TR-specific rewinder compensation gradient is introduced to help keep gradient area constant for each TR, thus getting better signal refocus.

(C) Reconstruction pipeline, including i) Gradient imperfection correction, ii) Motion correction, iii) Subspace reconstruction with B0 correction, and iv) Dictionary fitting with FA correction.

Figure 2.

Figure 2.

(A) Sequence diagram of the B1+ estimation module.

(B) RF phases of the B-S adiabatic pulses alternated in sign between odd and even TRs.

(C) B1+ and T2 maps acquired from a uniform phantom, demonstrating that the use of B-S adiabatic pulses yields a more uniform T2 map.

(D) B0 estimation module using interleaved TE gaps (ΔTE = 1 ms) between odd and even TRs.

(E) Comparison of B0 maps obtained using the proposed B0 estimation module and the standard B0 mapping sequence

Figure 3.

Figure 3.

(A) The waveform of 4 different RF excitation pulses and (B) its corresponding frequency responses.

(C) Zoomed-in the first coefficient maps of subspace reconstruction (top row) and the T1 maps (bottom row) using different RF excitation pulse.

Spiral trajectory design

In this work, spiral trajectories for both 560-μm and 360-μm isotropic resolutions were generated. For the 560-μm setting, two trajectories were designed using different SRmax of 500 T/m/s and 100 T/m/s, respectively. The latter was intended to mimic the performance of a whole-body gradient system (Siemens 7T MAGNETOM Terra) operating at ~85% of the peripheral nerve stimulation (PNS) threshold. Detailed design parameters are provided in Supporting Figure S2. This work employs a variable density spiral (VDS) design37, where the undersampling factor at the outer k-space is twice that at the k-space center. The MRF data acquired using the 100 T/m/s trajectory, which features slower k-space traversal and smaller gradient-induced imperfections, will be used as a comparison baseline to assess the effectiveness of the proposed gradient imperfection correction methods applied to the faster 500 T/m/s design.

To ensure consistency in scan parameters, particularly the minimum TR used, we maintained the same total duration of ~8.8 ms for the spiral readout plus rewinder and spoiler gradients. Consequently, for the same 560-μm resolution, the trajectory with lower SRmax (100 T/m/s) requires longer durations for rewinder and spoiler gradients, resulting in a need to employ a slightly shorter spiral readout (6.24 ms vs. 7.20 ms for 100 T/m/s and 500 T/m/s, respectively).

Therefore, a higher slew rate enables faster traversal of k-space and a slightly longer spiral readout duration. As a result, each single spiral interleaf is substantially less undersampled. For example, at the center of k-space, the in-plane undersampling factor is 16 for the SRmax of 500 T/m/s, compared with 48 for the SRmax of 100 T/m/s. This indicates that, to cover the same k-space extent, the latter would require three times as many TRs.

To balance SNR and scan efficiency, 24 acquisition groups were scanned for the 560-μm acquisition with slew rate of 500 T/m/s, which was subsequently shown in the Results section to be sufficient for achieving high-quality reconstructions. With the lower slew rate of 100 T/m/s, achieving comparable k-space coverage requires increasing the number of acquisition groups to 72. This resulted in total scan times of 3.5 minutes and 10.5 minutes, respectively (8.7 s per acquisition group).

For the 360-μm design, 168 acquisition groups and 24-minute scan time are implemented. Given that the voxel size at 360-μm isotropic is 3.8 times smaller than at 560-μm isotropic resolution, achieving similar SNR at 360-μm resolution would theoretically require approximately 50 minutes, as SNR scales with the square of the scan time (i.e., 3.82 × 3.5 ≈ 50 minutes). Therefore, two repetitions (2 × 24 minutes) were acquired for 360-μm to achieve a comparable SNR.

2.2. Reconstruction pipeline

In this work, the subspace reconstruction with a locally low-rank (LLR) constraint3944 is employed, which can be described as follows:

mincPFSΦc-y22+λRr(c) [1]

where P is the sampling pattern of spiral-projection trajectory, F is the non-uniform Fourier transform (NUFFT), S are the coil sensitivity maps, y are the acquired k-space data, and λ is the regularization parameter for LLR constraint Rr(c), and Φ is the subspace basis. c are the reconstructed subspace coefficient maps. In this work, λ of 1e-5 was selected for the LLR constraint. Specifically, for each voxel r, a local block of size 8 × 8 × 8 centered around r was extracted using the operator Rr, and each block is reshaped to form a matrix whose columns correspond to the subspace coefficient images. SVD is then applied to each block, followed by soft-thresholding of the singular values according to λ. Further details and the rationale behind the selection of λ and the number of subspace coefficients can be found in Supporting Figure S3 and Figure S4.

To improve reconstruction quality, the following additional components are added to the reconstruction (as shown in Figure 1C):

  1. Gradient imperfection correction: Skope field measurements were obtained in a phantom to provide 0th, 1st and 2nd order spatial field information. Significant deviation can be observed in the 0th order likely from B0 eddy current45,46 at high slew rates, as well as some minor 1st order deviations, with minimal 2nd order field variations observed. The reconstruction was modified to incorporate 0th and 1st order field information to improve performance. 1st order was incorporated by modifying the k-space trajectory used in the reconstruction, while 0th order was corrected via signal demodulation. For the 0th order, since the vendor’s provided k-space data has already undergone a default B0 eddy current compensation (B0 ECC) that is less accurate than what is measured via Skope, this effect has to be first removed prior to applying the updated B0 eddy current correction. Supporting Figure S5A shows the simulated B0-eddy-currents phase term, (obtained using Siemens sequence simulation software) and the measured one of one specific TR.

  2. Motion correction: The 2-mm motion navigator images are registered using the AFNI toolbox4749 to estimate motion parameters. The motion correction strategy in this work is to apply rotation matrix on the k-space trajectory and additional phase modulation on the k-space data to reflect the translation motion30.

  3. B0 correction: Severe B0 Field inhomogeneities at 7T can lead to image blurring and distortion, even in a highly segmented spiral acquisition of a 6-ms duration. To correct B0-induced distortion and blurring, the MFI approach50 was employed. Due to the severe B0 inhomogeneity at 7T, 11 demodulation frequencies ω1~11 = [−500:100:500] Hz were used to cover the typical ΔB0 range in the human brain at 7T.

  4. FA-corrected dictionary fitting: Through an acquisition of a B1+ map and an expanded dictionary fitting, the effect of B1+ inhomogeneity in creating a spatially varying FA can be accounted for in MRF tissue quantification, as we have performed in prior work25,51. However, large B0 inhomogeneity and limited RF excitation bandwidth can also cause a spatially varying FA, which also needs to be accounted for to achieve accurate tissue quantification. As shown in Figure 4C, the pronounced B0 inhomogeneity at 7T, could reach up to ±300Hz, leading to significant FA attenuation due to the finite bandwidth of the applied RF pulse. Figure 4B shows the frequency response of the SLR pulse, which indicates a FA attenuation of 0.63 at a voxel where the local ΔB0 is +300Hz, indicated by the red and blue arrows.

Figure 4.

Figure 4.

(A) The waveform of the 2.86-ms SLR pulse and (B) its corresponding frequency response.

(C) The ΔB0 map (Column 1) was used to determine the ΔB0-induced FA attenuation map (Column 2) via the RF frequency response. Incorporated with the B1+ map (Column 3), the apparent FA map (4th column) could be obtained.

(D) The T1 (top row) and T2 (bottom row) maps without any correction (Column 1), with only B0 MFI correction (Column 2), with both B0 & B1+ correction (Column 3), and with B0 & B1+ & frequency response correction (Column 4).

To correct this FA attenuation effect, this work proposed to use the frequency response of the excitation RF pulse as a look-up table for deriving the FA attenuation map based on the ΔB0 map. Subsequently, this FA attenuation map (the second column in Figure 4C) was then multiplied with the B1+ map (the third column) to obtain the actual FA factor map (the fourth column). Based on the FA factor maps, the T1 and T2 maps were then obtained by matching the reconstructed signal evolutions with the MRF dictionary entries corrected by the corresponding FA factor at each voxel.

After obtaining quantitative maps, a series of contrast-weighted images, including T1-MPRAGE, T2-weighted (T2W), T2 Fluid Attenuated Inversion Recovery (FLAIR) and Double Inversion Recovery (DIR), were synthesized using a Bloch simulation52. It’s worth highlighting that this study employed a local head-only transmission coil. Thus, the spatially-non-selective RF excitation trains in the MRF sequence operated in a similar fashion to slab-selective excitations used in MR angiography (MRA) sequence. As a result, synthesized MRA-like images could be obtained by applying a maximum intensity projection (MIP) to the PD maps.

All computations were conducted on an Ubuntu 20.04 server, equipped with 32 Core i7 Intel Xeon 2.8 GHz CPUs, an Nvidia A6000 GPU, and a memory of 1-TB RAM. The primary softwares utilized for these computations were MATLAB R2022a (The MathWorks, Inc., Natick, MA) and Python 3.8 (Python Software Foundation, Wilmington, DE).

2.3. In-vivo validation

Four healthy volunteers were scanned with approval from the Institutional Review Board. These scans were conducted on a Siemens NexGen 7T MAGNETOM Terra Impulse edition scanner (Erlangen, Germany) using a 96-channel-receive plus 16-channel-transmit coil and a 64-channel-receive plus 8-channel-transmit coil53 (MrCoilTech, Glasgow, UK). Additionally, to validate the stability of the gradient rewinder compensation based on Skope field camera measurements, five scan sessions were conducted over the same six-month period.

3. Results

3.1. Methodological validation

Since this study involves multiple correction methods, this section shows the results of ablation studies for each correction method.

3.1.1. Field inhomogeneity correction

Figure 4D shows the T1 and T2 mapping results, subsequent to field inhomogeneity corrections, including B0, B1+, and frequency response corrections. For the T1 map, the zoom-in region in the frontal lobe demonstrates the ability of B0 MFI correction in mitigating image blurring. As expected, B1+ and frequency response corrections have minimal effect on T1 quantification in this MRF sequence, since it utilizes an adiabatic inversion pulse to create a spatially uniform inversion flip angle. The effects of these corrections are much more significant for T2 quantification where it can be seen that the frequency response correction is critical in removing large biases.

Figure 2C shows T2 mapping results from the proposed MRF sequence on a uniform phantom and in an in vivo case, where comparison of B1+ corrections using B1+ maps obtained via i) proposed fast B1+ mapping with adiabatic B-S pulse, ii) proposed mapping but with Fermi-shaped B-S pulse and iii) with Siemens’s TurboFLASH mapping sequence is provided. The use of adiabatic B-S pulse in our proposed fast mapping produces T2 maps that are most spatially uniform, without observable large spatial biases seen with the other approaches. Figure 2E shows the consistency between the B0 maps obtained using the proposed B0 estimation module and those from the standard Siemens B0 sequence of a much longer scan time.

3.1.2. Fat ring suppression

Figure 3C shows the 1st coefficient maps from the subspace reconstruction and the T1 maps obtained from MRF acquisitions with the following RF excitation pulse types: a 2.94-ms hard pulse, 3-ms Gaussian pulse, 5.12-ms Gaussian pulse, and 2.86-ms SLR pulse. As indicated by the red arrows, the 2.94-ms hard pulse exhibits a pronounced fat ring artifact even though it was designed to minimize excitation around −1030 Hz, namely the predominant fat frequency at 7T. This is because the presence of significant B0 inhomogeneity affects the local resonance frequency of fat causing fat in some areas to still be excited by this RF pulse. The 3-ms Gaussian pulse mitigates some of these fat artifacts (yellow arrows), while the 5.12-ms Gaussian and 2.86-ms SLR pulses further diminish these artifacts, but with SLR pulse being significantly shorter in duration to allow for shorter TR and faster imaging. Importantly, the SLR pulse offers a broader FWHM within the ΔB0 range, which helps reduce the FA attenuation effect at off-resonance for the water signal.

3.1.3. Gradient imperfection correction

Figure 5 shows the T2 maps at 560-μm resolution acquired via an MRF acquisition using SRmax of 500 T/m/s where significant gradient imperfections exist. As indicated by the green arrows, the use of gradient rewinder compensation improves the uniformity of the T2 maps. The pink arrows indicate improved image sharpness that was achieved by incorporating the measured gradient trajectory into the reconstruction. T1 and T2 maps obtained using a low SRmax of 100 T/m/s with much reduced gradient imperfections are shown as reference. Figure S5B shows a zoomed-in view of the T1 and T2 maps reconstructed under three different conditions: 1) without any B0 ECC, 2) using the simulated B0 ECC (vendor’s default setting, obtained using Siemens sequence simulation software), and 3) using the measured B0 ECC. The measured B0 ECC shows a better visualization of PVS (perivascular spaces, red arrows), which is less discernible in images reconstructed using the simulated B0 ECC.

Figure 5.

Figure 5.

(A) T2 maps at 560-μm isotropic resolution using MRF with SRmax = 500 T/m/s: Column 1) without any correction, Column 2) with rewinder compensation, Column 3) with rewinder compensation and measured trajectory correction, compare to Column 4) reference acquisition with SRmax = 100 T/m/s.

(B) and (C) show the zoom-in T2 maps corresponding to the red and yellow boxes in (A), respectively.

Supporting Figure S1C shows the residual gradient areas across five scan sessions when using the rewinder compensation gradient settings measured from the first scan session. The sum-of-squared residual gradient area of all TRs are listed, which indicates that the gradient compensation could consistently remove 85% ~ 95% of the residual gradient moment compared to the one without implementing rewinder compensation gradient. Figure S1D shows both the measured and designed trajectories. When compared to the first session, the subsequent four sessions exhibit notable consistency. This demonstrated good stability of the gradient hardware, indicating that a single Skope calibration measurement can yield compensation parameters that remain valid for at least several months.

3.2. Comparison to baseline

Figure 6 shows results exhibiting how the proposed refinements and modifications to the MRF acquisition and reconstruction have enabled 560-μm whole-brain T1 and T2 mapping to be performed 3 times faster with significantly less biases compared to current state-of-the-art method. As a baseline, the first column in Figure 6B and 6C shows T1 and T2 maps in sagittal and coronal views from an MRF acquisition performed using conventional body gradient specifications (slew rate limited to ~100 T/m/s due to PNS constraint) along with Siemens’s standard B1+ and B0 mapping sequences, resulting in a total scan time of 12 minutes and 26 seconds. Such an acquisition should not contain significant gradient imperfections, but the resulting T2 map still contains large spatial biases from i) not employing the proposed RF frequency response correction (red arrow), and ii) imperfect B1+ correction from biases in the B1+ map from the standard B1+ mapping sequence (yellow arrows). The second column of Figure 6B and 6C shows the results from the same setup, but with three times the reduction in MRF k-space encoding, thereby facilitating a three-fold reduction in scan time, with additional aliasing artifacts observed (pink arrow). The third column shows that such aliasing artifacts can be overcome via faster k-space traversal with a higher slew rate of 500 T/m/s, but at a cost of added T2 biases and blurring from gradient imperfections and incomplete gradient rewinding (green arrows). The last column shows results obtained using an MRF acquisition at 500 T/m/s with gradient rewinder compensation and fast B1+ and B0 mapping, along with all the proposed reconstruction modifications. Here, high quality T1 and T2 maps can be obtained in just 3 minutes 44 seconds. Additional slices and orientations are presented in Supporting Figure S6.

Figure 6.

Figure 6.

(A) The proposed method enables quantitative imaging at 560-μm resolution with a total acquisition time of less than 4 minutes on the NexGen 7T with Impulse gradient, which is more than three times faster than conventional gradient systems combined with product field mapping sequences.

(B) Without the proposed correction methods, even a 12-minute acquisition leads to noticeable T2 bias, caused by either uncorrected RF frequency response (red arrows), or inaccurate B1+ estimation (yellow arrows). Additionally, simply using a higher acceleration factor to match the same acquisition time introduces severe undersampling artifacts (pink arrow), while increasing slew rate without employing the proposed compensation methods leads to significant T2 shading caused by imperfect signal refocusing.

(C) Coronal view further reveals T2 bias caused by inaccurate B1+ estimation, as indicated by the yellow arrow.

3.3. 360-μm MRF

3.3.1. motion correction

Figure 7 shows comparisons of quantitative maps, with and without applying motion correction, for both one-repetition and two-repetitions scans at 360-μm isotropic resolution across 24 and 48 minutes respectively. It can be observed that even for a dedicated healthy volunteer, the prolonged scan time inevitably leads to observable motion. The maximum translation is only around 0.5 mm (Figure 7A), which may not cause noticeable issues at lower resolution, however, this translation corresponds to a displacement of approximately 1.4 pixels at the 360-μm resolution. The reconstructed quantitative maps with and without motion correction (Figure 7B) validated that even such small motion can lead to observable blurring, making subtle structures (PVS as indicated by red arrows) invisible. Notably, the motion-induced blurring is more pronounced for the two-repetitions scan, highlighting the necessity of motion correction in this case.

Figure 7.

Figure 7.

(A) Motion estimation (rotation and translation) of a 2-repetition scan (48 minutes) using the 2-mm navigators every 8.7 s.

(B) One-repetition and two-repetitions quantitative maps at 360-μm isotropic resolution, reconstructed with and without motion correction (MOCO). With motion correction, the subtle structures, such as PVS (red arrows), become more clearly visible.

3.3.2. Quantitative maps and synthesis contrasts

Figure 8 presents three orthogonal views of the T1 and T2 maps at 360-μm of a healthy volunteer, reconstructed using all the correction methods described above. The hippocampus, highlighted within the red boxes, is used to demonstrate the image quality of these high-resolution quantitative images, where the complex internal structures of the hippocampus are clearly visible. Black arrows indicate other subtle structures, such as the optic radiations (Figure 8A), cerebellar vermis (Figure 8B), and pontes grisei (caudato-lenticular gray matter bridges, Figure 8C). Additionally, as indicated by the yellow arrows, some nucleus exhibit distinct differences from their surrounding regions in the T2 maps, whereas these differences are not as easily distinguishable in the T1 maps. In addition, Supporting Video S1 provides dynamic visualizations of T1, T2, and PD maps across three orthogonal views for all four subjects. The corresponding quantitative values of T1 and T2 for WM and GM are reported in Supporting Table S1, along with reference values from previous studies14,15,18,19,53,54.

Figure 8.

Figure 8.

Whole-brain T1 and T2 maps at 360-μm isotropic resolution. The red box highlights the hippocampus, where the red arrow indicates the exceptional image quality and structural detail achieved in the ultra-high-resolution quantitative maps. Black arrows, from top to bottom, point to the optic radiations, cerebellar vermis, and pontes grisei, while the yellow arrows point to deep-brain nucleis.

Figure 9 shows the synthesized T1 MPRAGE, T2W, T2 FLAIR, and DIR images at 360-μm isotropic resolution, where high quality results with detailed structural information can be observed. Furthermore, we also observed that, with the head-only RF transmission coil, simply applying MIP to the proton density map can achieve a MRA-like image, which provides additional information beyond conventional T1 and T2 contrasts.

Figure 9.

Figure 9.

Contrast-weighted images are synthesized from the quantitative maps using the proposed method, including T1 MPRAGE, T2W, DIR, FLAIR and MRA-like.

Figure 10A and 10B show the T1 and T2 maps of the slice containing the substantia nigra (SN) and subthalamic nucleus (STN). It can be observed that the T1 values of SN and STN (critical for the early diagnosis and intervention of Parkinson disease) are very similar to those of the surrounding white matter, making them difficult to distinguish on T1 maps. However, their T2 values show a significant difference against the surrounding white matter, where in the zoom-in sub-figure with compressed color scaling, it can be seen that STN exhibits a slightly higher T2 value than SN. As a result, pure T1-weighted imaging, such as T1 MPRAGE (Figure 10C), fails to differentiate these two structures, whereas incorporating T2 weighting (e.g., through introducing T2 preparation, Figure 10D) allows for clearer distinction and segmentation of these structures.

Figure 10.

Figure 10.

T1 (A) and T2 (B) maps using the proposed MRF acquisition. The STN (red arrow) exhibits slightly higher T2 values than the SN (black arrow), while their T1 values remain similar to surrounding white matter. A synthesized T1-MPRAGE image (C) demonstrates poor contrast between STN and SN. Incorporating T2 weighting (e.g., by adding a T2 preparation module) could improve differentiation between these two nuclei (D).

Discussion

In this study, we successfully developed a tailored MRF acquisition and reconstruction framework optimized for a high-performance, head-only 7T scanner. Leveraging the system’s superior gradient performance, we achieved whole-brain quantitative imaging at 560-μm isotropic resolution with a scan time under 4 minutes to allow it to be easily included in many neuroscientific and clinical research studies. Furthermore, this work marks the first demonstration of a whole-brain MRF scan at 360-μm isotropic resolution. These results push the spatial resolution boundaries of in-vivo human brain imaging and provide a practical pathway for ultra-high-resolution quantitative mapping. In addition, using the example of brain nucleus SN and STN, we demonstrated that the combined use of T1 and T2 can facilitate the differentiation of these brain structures. This suggests that performing multi-parametric quantitative imaging can help optimize conventional contrast-weighted sequences to maximize contrast for specific structures of interest, thereby enabling more accurate structural delineation and segmentation.

Ablation studies highlighted the necessity of several key modifications to both the acquisition and reconstruction processes in order to achieve reliable high-resolution imaging at 7T and high-performance gradient system. The two major innovations of correcting FA attenuation and gradient rewinder compensation prove important for obtaining high-quality T2 maps. While this correction was developed for 7T, retrospective analysis suggests it could also be beneficial at 3T, where moderate FA attenuation effects are still present in certain regions, such as the front lobe of brain and in body applications with severe B0 inhomogeneity. Additionally, we implemented gradient rewinder compensation to mitigate signal dephasing caused by imperfect gradient moment nulling in FISP-based sequences. This correction, implemented with a small blip gradient derived from Skope field camera measurements, effectively preserves the stimulated echo and improves the accuracy of T2 estimation. The same principle could be extended to other sequences such as TSE, SPACE, or GRASE where echo refocusing is critical. The compensation gradient could be integrated directly into the gradient crusher or rewinder to avoid incurring an increased in echo-spacing.

Despite the multitude of the corrections/mitigation approaches involved, the proposed MRF pipeline has been designed for seamless usability. B1+ and B0 mapping modules were embedded directly within the MRF sequence to minimize inter-scan motion and streamline workflow. Given that the B1+ and B0 fields are inherently smooth, ultra-high-resolution field maps offer limited practical value. Moreover, a joint estimation of B1+ and B0 fields together with tissue parameters in the MRF framework is feasible, but would require modifications to our MRF acquisition, to increase its sensitivity to these fields and enable them to be estimated accurately. This would significantly lengthen the overall acquisition. Therefore, we adopted dedicated, rapid, low-resolution B1+ and B0 estimation modules instead. Furthermore, motion correction is integrated into the acquisition, ensuring motion robustness in routine applications, including studies involving motion-prone populations. Skope-based measurements and corresponding correction strategies demonstrated good stability, indicating that a single phantom-based calibration session may remain valid over extended periods without the need for frequent recalibration. These characteristics should make the fast 560-μm MRF protocol a promising tool that can be widely deployed in neuroscience research, enabling reproducible and high-resolution brain mapping in a time-efficient manner.

The need for accurate B1+ mapping to achieve high fidelity T1 and T2 mapping was demonstrated, where we achieved this using a rapid B1+ mapping sequence that makes use of B-S pulse. Here, the use of adiabatic design rather than conventional Fermi design of B-S pulse enables short duration pulses to be employed within the SAR constraint to achieve high accuracy in B1+ mapping. This work has demonstrated the effectiveness of the proposed B1+ correction. However, in areas where B1+ is very low, such as in the lower part of the brain, the MRF’s FA train is at markedly lower flip angles, making T2 quantification more difficult even when the actual flip angle is accounted for. On the other hand, a 20-ms adiabatic tanh pulse was employed for inversion preparation with the assumption that it would be relatively insensitive to B1+ variations. However, as in extreme cases (for example in cerebellum), insufficient inversion efficiency due to B1+ may lead to incomplete Mz inversion and faster-than-expected recovery, resulting in underestimation of the T1 values. Future work would explore the use of pTx pulse design8,5557, such as through k-t pulse5860 or GRAPE pulse61 design, to provide more uniform excitation. In particular, an interleaved binomial kT-point design62 for water-selective imaging could be a good fit here. Considering that conventional pTx techniques often require complex B1+ field calibration procedures and per subject pulse calculation, adoption of plug-and-play63 or semi–plug-and-play64 pTx solutions could also help avoid adding complexity to our MRF acquisition.

Despite applying both B1+ and RF frequency response corrections, accurate T1 and T2 estimations remain challenging in regions with substantial blood flow, such as in the venous structures, due to spin history effects65. Addressing this issue may require physical modeling of flow-related signal evolution. Furthermore, inaccuracies in the estimated T1 and T2 values can affect the PD quantification, since PD is typically computed after the dictionary matching process of MRF by dividing the sum of squared magnitudes of the acquired signal evolutions over the 500 TRs at each voxel by that of the corresponding matched dictionary entries5. If the estimated T1 and T2 are incorrect, the associated matched dictionary entry will also be incorrect, leading to a mismatch in the expected signal evolution and an altered sum of squared values, thereby introducing additional bias into the PD estimation. This bias in PD can subsequently impact synthesized MRA-like images obtained via MIP. Therefore, the MRA-like images presented in this study should not be interpreted as precise vascular measurements but rather as a proof-of-concept exploration of additional information that may be extracted from the MRF datasets. On the other hand, due to the influence of the receiver coil sensitivity profile, the PD maps often exhibit peripheral-to-central signal shading, which may in turn compromise the accuracy of synthesized MRF-like images. This issue could potentially be mitigated through the application of appropriate correction techniques66. The data acquisition strategy developed in this study enables rapid and robust k-space encoding. However, at ultra-high-resolution such as 360-μm isotropic, the MRF acquisition can become more motion limited which we have dealt with, and SNR-limited rather than image-encoding speed limited. Therefore, an important future direction is to integrate the proposed acquisition approach with advanced denoising or regularization techniques for ultra-high resolution. Unrolled deep learning-based reconstruction67 represents a promising avenue to pursue. However, further development is required, particularly in addressing the challenges associated with training such large-scale reconstruction problems. Methods such as GLEAM68 and related frameworks may provide useful solutions in this context.

Conclusion

This study presents a fast and robust MRF framework tailored for a head-only high-performance NexGen 7T scanner, enabling whole-brain quantitative imaging at 560-μm isotropic resolution in under 4 minutes and achieving the first whole-brain in-vivo 360-μm multi-parametric maps. Key innovations include embedded fast B0 and B1+ mapping, gradient rewinder compensation, RF frequency response correction, and redesigned water-only RF pulse. These methods address challenges of field inhomogeneities, gradient imperfections, and low SNR at ultra-high resolution. Field estimation and motion estimation are seamlessly integrated into the MRF framework, ensuring workflow efficiency. This method not only provides a rapid, high-resolution quantitative imaging approach that enables the application of meso-scale quantitative imaging in clinical and neuroscience research, but also pushes higher the current resolution limits of multi-parametric quantitative MRI, offering new possibilities for studying quantitative changes of fine-scale structure in the human brain.

Supplementary Material

supporting information

Acknowledgement

This work was supported by: NIH research grants: R01EB020613, R01MH116173, R01EB019437, R01EB028797, R01EB016695, R01HD114719, U01EB025162, P41EB030006, U01EB026996, R03EB031175, U24NS129949, K99EB035178.

The authors would like to convey the appreciation to Dr. Paul Weavers and Dr. Cameron Cushing for their help on setting up the Skope field camera.

Reference

  • 1.Weiskopf N, Edwards LJ, Helms G, Mohammadi S, Kirilina E. Quantitative magnetic resonance imaging of brain anatomy and in vivo histology. Nat Rev Phys. 2021;3(8):570–588. doi: 10.1038/s42254-021-00326-1 [DOI] [Google Scholar]
  • 2.Wearn AR, Nurdal V, Saunders-Jennings E, et al. T2 heterogeneity: a novel marker of microstructural integrity associated with cognitive decline in people with mild cognitive impairment. Alzheimers Res Ther. 2020;12(1):105. doi: 10.1186/s13195-020-00672-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Liao C, Cao X, Iyer SS, et al. High-resolution myelin-water fraction and quantitative relaxation mapping using 3D ViSTa-MR fingerprinting. Magn Reson Med. December 2023. doi: 10.1002/mrm.29990 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Lüsebrink F, Mattern H, Yakupov R, et al. Comprehensive ultrahigh resolution whole brain in vivo MRI dataset as a human phantom. Sci Data. 2021;8(1):138. doi: 10.1038/s41597-021-00923-w [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Ma D, Gulani V, Seiberlich N, et al. Magnetic resonance fingerprinting. Nature. 2013;495(7440):187–192. doi: 10.1038/nature11971 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Yu Z, Hodono S, Dergachyova O, et al. Simultaneous 3D acquisition of 1 H MRF and 23 Na MRI. Magn Reson Med. 2022;87(5):2299–2312. doi: 10.1002/mrm.29135 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Lim SI, Widmaier MS, Wenz D, Yun J, Xin L. Time-efficient relaxation measurements by 31P MR fingerprinting in human brain at 7T. bioRxiv. May 2022. doi: 10.1101/2022.05.23.493067 [DOI] [Google Scholar]
  • 8.Cloos MA, Knoll F, Zhao T, et al. Multiparametric imaging with heterogeneous radiofrequency fields. Nat Commun. 2016;7:12445. doi: 10.1038/ncomms12445 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Buonincontri G, Schulte RF, Cosottini M, Tosetti M. Spiral MR fingerprinting at 7T with simultaneous B1 estimation. Magn Reson Imaging. 2017;41:1–6. doi: 10.1016/j.mri.2017.04.003 [DOI] [PubMed] [Google Scholar]
  • 10.Buonincontri Guido, Pedro A Gómez Matteo Cencini, Fallo Giada, Biagi Laura, Rolf F Schulte, and Michela Tosetti. Fast parameter mapping at 7T with SSFP MR Fingerprinting using 3D radial trajectories. In: ISMRM 2018.; :4260. [Google Scholar]
  • 11.Zhou Z, Li Q, Liao C, et al. Optimized three-dimensional ultrashort echo time: Magnetic resonance fingerprinting for myelin tissue fraction mapping. Hum Brain Mapp. 2023;44(6):2209–2223. doi: 10.1002/hbm.26203 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Koolstra K, Beenakker JWM, Koken P, Webb A, Börnert P. Cartesian MR fingerprinting in the eye at 7T using compressed sensing and matrix completion-based reconstructions. Magn Reson Med. 2019;81(4):2551–2565. doi: 10.1002/mrm.27594 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Zimmermann M, Abbas Z, Sommer Y, et al. QRAGE-Simultaneous multiparametric quantitative MRI of water content, T1, T2*, and magnetic susceptibility at ultrahigh field strength. Magn Reson Med. 2025;93(1):228–244. doi: 10.1002/mrm.30272 [DOI] [PubMed] [Google Scholar]
  • 14.Versteeg E, Liu H, van der Heide O, Fuderer M, van den Berg CAT, Sbrizzi A. High SNR full brain relaxometry at 7T by accelerated MR-STAT. Magn Reson Med. 2024;92(1):226–235. doi: 10.1002/mrm.30052 [DOI] [PubMed] [Google Scholar]
  • 15.Leroi L, Gras V, Boulant N, et al. Simultaneous proton density, T1, T2, and flip-angle mapping of the brain at 7 T using multiparametric 3D SSFP imaging and parallel-transmission universal pulses. Magn Reson Med. 2020;84(6):3286–3299. doi: 10.1002/mrm.28391 [DOI] [PubMed] [Google Scholar]
  • 16.Gras V, Pracht ED, Mauconduit F, Le Bihan D, Stöcker T, Boulant N. Robust nonadiabatic T2 preparation using universal parallel-transmit kT -point pulses for 3D FLAIR imaging at 7 T. Magn Reson Med. 2019;81(5):3202–3208. doi: 10.1002/mrm.27645 [DOI] [PubMed] [Google Scholar]
  • 17.Gras V, Poser BA, Wu X, Tomi-Tricot R, Boulant N. Optimizing BOLD sensitivity in the 7T Human Connectome Project resting-state fMRI protocol using plug-and-play parallel transmission. Neuroimage. 2019;195:1–10. doi: 10.1016/j.neuroimage.2019.03.040 [DOI] [PubMed] [Google Scholar]
  • 18.Metere R, Kober T, Möller HE, Schäfer A. Simultaneous quantitative MRI mapping of T1, T2* and magnetic susceptibility with multi-echo MP2RAGE. PLoS One. 2017;12(1):e0169265. doi: 10.1371/journal.pone.0169265 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Caan MWA, Bazin PL, Marques JP, de Hollander G, Dumoulin SO, van der Zwaag W. MP2RAGEME: T1, T2 *, and QSM mapping in one sequence at 7 tesla. Hum Brain Mapp. 2019;40(6):1786–1798. doi: 10.1002/hbm.24490 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Jun Y, Fujita S, Chen Y, et al. Efficient mesoscale multiparametric quantitative MRI using 3D-QALAS at 7T with self-supervised learning. In: ISMRM Annual Meeting. Concord, CA: ISMRM; 2025. doi: 10.58530/2025/0815 [DOI] [Google Scholar]
  • 21.Feinberg DA, Beckett AJS, Vu AT, et al. Next-generation MRI scanner designed for ultra-high-resolution human brain imaging at 7 Tesla. Nat Methods. 2023;20(12):2048–2057. doi: 10.1038/s41592-023-02068-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Davids M, Dietz P, Ruyters G, et al. Peripheral nerve stimulation informed design of a high-performance asymmetric head gradient coil. Magn Reson Med. 2023;90(2):784–801. doi: 10.1002/mrm.29668 [DOI] [PubMed] [Google Scholar]
  • 23.Feinberg DA, Ma SJ, Walker E, et al. Reassessment of peripheral nerve stimulation thresholds for the Impulse model-optimized asymmetric head gradient coil. Magn Reson Med. 2025;94(3):1326–1338. doi: 10.1002/mrm.30523 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Park S, Beckett A, Häkkinen S, et al. Higher spatial resolution and sensitivity in whole brain functional MRI at 7T using 3D EPI accelerated with variable density CAIPI sampling and temporal random walk. Magn Reson Med. 2025;94(2):678–693. doi: 10.1002/mrm.30512 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Cao X, Liao C, Iyer SS, et al. Optimized multi-axis spiral projection MR fingerprinting with subspace reconstruction for rapid whole-brain high-isotropic-resolution quantitative imaging. Magnetic Resonance in Medicine. 2022;88(1):133–150. doi: 10.1002/mrm.29194 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Cao X, Ye H, Liao C, Li Q, He H, Zhong J. Fast 3D brain MR fingerprinting based on multi-axis spiral projection trajectory. Magn Reson Med. 2019;82(1):289–301. doi: 10.1002/mrm.27726 [DOI] [PubMed] [Google Scholar]
  • 27.Yablonski M, Zhou Z, Cao X, et al. Fast and reliable quantitative measures of white matter development with magnetic resonance fingerprinting. Imaging Neuroscience. 2025;3. doi: 10.1162/imag_a_00470 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Barmet C, De Zanche N, Pruessmann KP. Spatiotemporal magnetic field monitoring for MR. Magn Reson Med. 2008;60(1):187–197. doi: 10.1002/mrm.21603 [DOI] [PubMed] [Google Scholar]
  • 29.Jiang Y, Ma D, Seiberlich N, Gulani V, Griswold MA. MR fingerprinting using fast imaging with steady state precession (FISP) with spiral readout. Magn Reson Med. 2015;74(6):1621–1631. doi: 10.1002/mrm.25559 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Cao X, Liao C, Zhu Z, et al. Three-dimensional high-isotropic-resolution MR fingerprinting optimized for 0.55 T. Magn Reson Med. January 2025. doi: 10.1002/mrm.30420 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Tesiram YA, Bendall MR. Universal equations for linear adiabatic pulses and characterization of partial adiabaticity. J Magn Reson. 2002;156(1):26–40. doi: 10.1006/jmre.2002.2531 [DOI] [PubMed] [Google Scholar]
  • 32.Khalighi MM, Rutt BK, Kerr AB. Adiabatic RF pulse design for Bloch-Siegert B1+ mapping. Magn Reson Med. 2013;70(3):829–835. doi: 10.1002/mrm.24507 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Chung S, Kim D, Breton E, Axel L. Rapid B1+ mapping using a preconditioning RF pulse with TurboFLASH readout. Magn Reson Med. 2010;64(2):439–446. doi: 10.1002/mrm.22423 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Pauly J, Le Roux P, Nishimura D, Macovski A. Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm [NMR imaging]. IEEE Trans Med Imaging. 1991;10(1):53–65. doi: 10.1109/42.75611 [DOI] [PubMed] [Google Scholar]
  • 35.Norbeck O, Sprenger T, Avventi E, et al. Optimizing 3D EPI for rapid T1 -weighted imaging. Magn Reson Med. 2020;84(3):1441–1455. doi: 10.1002/mrm.28222 [DOI] [PubMed] [Google Scholar]
  • 36.Hu S, Chen Y, Zong X, Lin W, Griswold M, Ma D. Improving motion robustness of 3D MR fingerprinting with a fat navigator. Magn Reson Med. 2023;90(5):1802–1817. doi: 10.1002/mrm.29761 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Kim DH, Adalsteinsson E, Spielman DM. Simple analytic variable density spiral design. Magn Reson Med. 2003;50(1):214–219. doi: 10.1002/mrm.10493 [DOI] [PubMed] [Google Scholar]
  • 38.Tamir JI, Uecker M, Chen W, et al. T2 shuffling: Sharp, multicontrast, volumetric fast spin-echo imaging. Magn Reson Med. 2017;77(1):180–195. doi: 10.1002/mrm.26102 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Zhang T, Pauly JM, Levesque IR. Accelerating parameter mapping with a locally low rank constraint. Magn Reson Med. 2015;73(2):655–661. doi: 10.1002/mrm.25161 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Cao X, Wang K, Liao C, et al. Efficient T2 mapping with blip-up/down EPI and gSlider-SMS (T2 -BUDA-gSlider). Magn Reson Med. 2021;86(4):2064–2075. doi: 10.1002/mrm.28872 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Wang N, Liao C, Cao X, et al. Spherical echo-planar time-resolved imaging (sEPTI) for rapid 3D quantitative T 2 * and susceptibility imaging. Magn Reson Med. 2025;93(1):121–137. doi: 10.1002/mrm.30255 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Iyer SS, Ong F, Cao X, et al. Polynomial Preconditioners for Regularized Linear Inverse Problems. SIAM J Imaging Sci. 2024;17(1):116–146. doi: 10.1137/22M1530355 [DOI] [Google Scholar]
  • 43.Schauman SS, Iyer SS, Sandino CM, et al. Deep learning initialized compressed sensing (Deli-CS) in volumetric spatio-temporal subspace reconstruction. MAGMA. February 2025. doi: 10.1007/s10334-024-01222-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Robison RK, Li Z, Wang D, Ooi MB, Pipe JG. Correction of B0 eddy current effects in spiral MRI. Magnetic Resonance in Medicine. 2019;81(4):2501–2513. doi: 10.1002/mrm.27583 [DOI] [PubMed] [Google Scholar]
  • 45.Brodsky EK, Klaers JL, Samsonov AA, Kijowski R, Block WF. Rapid measurement and correction of phase errors from B0 eddy currents: Impact on image quality for non-cartesian imaging. Magnetic Resonance in Medicine. 2013;69(2):509–515. doi: 10.1002/mrm.24264 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Cox JS RW;. H. AFNI: Software for analysis and visualization of functional magnetic resonance neuroimages. Computers and Biomedical Research. 1996;29(29):162–173. [DOI] [PubMed] [Google Scholar]
  • 47.Cox RW, Hyde JS. Software tools for analysis and visualization of fMRI data. NMR in Biomedicine. 1997;10(4–5):171–178. doi: 10.1002/(SICI)1099-1492(199706/08)10:4/5<171::AID-NBM453>3.0.CO;2-L [DOI] [PubMed] [Google Scholar]
  • 48.Saad ZS, Glen DR, Chen G, Beauchamp MS, Desai R, Cox RW. Alignment using Local Pearson Correlation. Neuroimage. 2009;44(3):839–848. doi: 10.1016/j.neuroimage.2008.09.037.A [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Ostenson J, Robison RK, Zwart NR, Welch EB. Multi-frequency interpolation in spiral magnetic resonance fingerprinting for correction of off-resonance blurring. Magn Reson Imaging. 2017;41:63–72. doi: 10.1016/j.mri.2017.07.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Ma D, Coppo S, Chen Y, et al. Slice profile and B1 corrections in 2D magnetic resonance fingerprinting. Magn Reson Med. 2017;78(5):1781–1789. doi: 10.1002/mrm.26580 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Yang M, Jiang Y, Ma D, Mehta BB, Griswold MA. Game of Learning Bloch Equation Simulations for MR Fingerprinting. 2020.
  • 52.Shajan Gunamony DF. 8-channel transmit 64-channel receive compact head coil Next Gen 7T scanner head gradient insert. In: ISMRM 2022, London.; :1447. [Google Scholar]
  • 53.Seginer A, Schmidt R. Phase-based fast 3D high-resolution quantitative T2 MRI in 7 T human brain imaging. Sci Rep. 2022;12(1):14088. doi: 10.1038/s41598-022-17607-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Schmidt J High resolution quantitative T2 mapping human brain 7 T using multi-echo spin-echo sequence dictionary-based modeling. Imaging Neuroscience. 2025. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Setsompop K, Wald LL, Alagappan V, Gagoski BA, Adalsteinsson E. Magnitude least squares optimization for parallel radio frequency excitation design demonstrated at 7 tesla with eight channels. Magnetic Resonance in Medicine. 2008;59(4):908–915. doi: 10.1002/mrm.21513 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Setsompop K, Alagappan V, Gagoski BA, et al. Broadband slab selection with B1+ mitigation at 7T via parallel spectral-spatial excitation. Magnetic Resonance in Medicine. 2009;61(2):493–500. doi: 10.1002/mrm.21834 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Li Q, Liao C, Ye H, et al. Squeezed trajectory design for peak RF and integrated RF power reduction in parallel transmission MRI. IEEE Trans Med Imaging. 2018;37(8):1809–1821. doi: 10.1109/TMI.2018.2828112 [DOI] [PubMed] [Google Scholar]
  • 58.Cloos MA, Boulant N, Luong M, et al. k T -points: Short three-dimensional tailored RF pulses for flip-angle homogenization over an extended volume. Magnetic Resonance in Medicine. 2012;67(1):72–80. doi: 10.1002/mrm.22978 [DOI] [PubMed] [Google Scholar]
  • 59.Hao S, Fessler JA, Noll DC, Nielsen JF. Joint design of excitation k-space trajectory and RF pulse for small-tip 3D tailored excitation in MRI. IEEE Transactions on Medical Imaging. 2016;35(2):468–479. doi: 10.1109/TMI.2015.2478880 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Malik SJ, Keihaninejad S, Hammers A, Hajnal JV. Tailored excitation in 3D with spiral nonselective (SPINS) RF pulses. Magnetic Resonance in Medicine. 2012;67(5):1303–1315. doi: 10.1002/mrm.23118 [DOI] [PubMed] [Google Scholar]
  • 61.Khaneja N, Reiss T, Kehlet C, Schulte-Herbrüggen T, Glaser SJ. Optimal control of coupled spin dynamics: design of NMR pulse sequences by gradient ascent algorithms. J Magn Reson. 2005;172(2):296–305. doi: 10.1016/j.jmr.2004.11.004 [DOI] [PubMed] [Google Scholar]
  • 62.Löwen D, Pracht ED, Stirnberg R, Liebig P, Stöcker T. Interleaved binomial kT -points for water-selective imaging at 7T. Magn Reson Med. 2022;88(6):2564–2572. doi: 10.1002/mrm.29376 [DOI] [PubMed] [Google Scholar]
  • 63.Gras V, Vignaud A, Amadon A, Le Bihan D, Boulant N. Universal pulses: A new concept for calibration-free parallel transmission. Magn Reson Med. 2017;77(2):635–643. doi: 10.1002/mrm.26148 [DOI] [PubMed] [Google Scholar]
  • 64.Herrler J, Liebig P, Gumbrecht R, et al. Fast online-customized (FOCUS) parallel transmission pulses: A combination of universal pulses and individual optimization. Magn Reson Med. 2021;85(6):3140–3153. doi: 10.1002/mrm.28643 [DOI] [PubMed] [Google Scholar]
  • 65.Cao Xiaozhi, Liao Congyu, Wang Zhixing, Li Qing, Ye Huihui, Chen Ying, He Hongjian, Zhong Jianhui. A model-based velocity mapping of blood flow using MR fingerprinting. In: Proc Intl Soc Mag Reson Med.; 2017. [Google Scholar]
  • 66.Volz S, Nöth U, Jurcoane A, Ziemann U, Hattingen E, Deichmann R. Quantitative proton density mapping: correcting the receiver sensitivity bias via pseudo proton densities. Neuroimage. 2012;63(1):540–552. doi: 10.1016/j.neuroimage.2012.06.076 [DOI] [PubMed] [Google Scholar]
  • 67.Huang P, Zhang C, Zhang X, Li X, Dong L, Ying L. Self-supervised deep unrolled reconstruction using Regularization by denoising. IEEE Trans Med Imaging. 2024;43(3):1203–1213. doi: 10.1109/TMI.2023.3332614 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 68.Ozturkler B, Sahiner A, Ergen T, et al. GLEAM: Greedy LEarning for large-scale accelerated MRI reconstruction. arXiv [eessIV]. July 2022. http://arxiv.org/abs/2207.08393. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

supporting information

RESOURCES