Skip to main content
Transactions of the American Ophthalmological Society logoLink to Transactions of the American Ophthalmological Society
. 1996;94:135–145. doi: 10.1016/s0002-9394(14)70162-x

The use of computerized videokeratography as an aid in fitting rigid gas permeable contact lenses.

P C Donshik 1, D S Reisner 1, A E Luistro 1
PMCID: PMC1312092  PMID: 8981693

Abstract

PURPOSE: To assess the efficacy of computerized corneal topography in (1) fitting cosmetic rigid gas permeable contact lenses on normal corneas and (2) patients who have keratoconus. METHODS: In study 1, 26 patients were fitted with cosmetic contact lenses. In 13 patients, the first rigid gas permeable lens was chosen on the basis of corneal topography using the EyeSys videokeratoscope and contact lens software version 2.11w. Thirteen control patients were fitted with rigid gas permeable contact lenses by means of diagnostic contact lenses. The number of parameter changes required in each group to reach the best fit were compared. In study 2, the records of 28 patients (41 eyes) with keratoconus who had been successfully fit with rigid gas permeable lenses were reviewed. Topographic analysis using the EyeSys videokeratoscope was conducted. The following 8 data points were analyzed: maximum dioptric curvature with respect to the central visual axis, the power at the visual axis (VK), the average keratometry readings within the 3-mm zone, the average flat K at 3-mm central and 5-mm paracentral regions, and the K reading at 1.5, 2.0, and 2.5 mm superior to the visual axis. RESULTS: Study 1: In 10 eyes (38%) fitted with cosmetic contact lenses utilizing corneal topography and the EyeSys contact lens program, the first lens chosen by the computer program was correct with no additional changes in lens parameters required. In the other 16 eyes, 21 parameter changes were required. In the control group, 14 eyes (53%) required no additional change in parameters from the original contact lens that was fitted using diagnostic lenses. In the other 12 eyes, 12 parameter changes were required. Study 2: In the keratoconus group, of the 9 measurements analyzed, the final base curve of patients' rigid gas permeable lenses was correlated most closely with the average flat corneal curvature at the 3-mm central zone. CONCLUSION: Study 1: In fitting cosmetic rigid gas permeable contact lenses, the videokeratoscope with the EyeSys software 2.11w contact lens program proved to be inferior to diagnostic lenses in selecting the final correct contact lens parameters. However, it may be useful in selecting an initial diagnostic lens with which to begin rigid gas permeable fitting. Study 2: In patients with irregular corneas such as keratoconus, corneal topography is helpful in determining the initial base curve of the contact lens. In patients with keratoconus, it appears to be the average flat corneal curvature at the 3.0-mm zone.

Full text

PDF
135

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Bogan S. J., Waring G. O., 3rd, Ibrahim O., Drews C., Curtis L. Classification of normal corneal topography based on computer-assisted videokeratography. Arch Ophthalmol. 1990 Jul;108(7):945–949. doi: 10.1001/archopht.1990.01070090047037. [DOI] [PubMed] [Google Scholar]
  2. Dingeldein S. A., Klyce S. D. The topography of normal corneas. Arch Ophthalmol. 1989 Apr;107(4):512–518. doi: 10.1001/archopht.1989.01070010526024. [DOI] [PubMed] [Google Scholar]
  3. Lopatynsky M., Cohen E. J., Leavitt K. G., Laibson P. R. Corneal topography for rigid gas permeable lens fitting after penetrating keratoplasty. CLAO J. 1993 Jan;19(1):41–44. doi: 10.1097/00140068-199301000-00007. [DOI] [PubMed] [Google Scholar]
  4. Lopatynsky M., Cohen E. J., Leavitt K. G., Laibson P. R. Corneal topography for rigid gas permeable lens fitting after penetrating keratoplasty. CLAO J. 1993 Jan;19(1):41–44. doi: 10.1097/00140068-199301000-00007. [DOI] [PubMed] [Google Scholar]
  5. Maguire L. J., Bourne W. M. Corneal topography of early keratoconus. Am J Ophthalmol. 1989 Aug 15;108(2):107–112. doi: 10.1016/0002-9394(89)90001-9. [DOI] [PubMed] [Google Scholar]
  6. McMahon T. T., Robin J. B., Scarpulla K. M., Putz J. L. The spectrum of topography found in keratoconus. CLAO J. 1991 Jul;17(3):198–204. [PubMed] [Google Scholar]
  7. Rabinowitz Y. S., Garbus J. J., Garbus C., McDonnell P. J. Contact lens selection for keratoconus using a computer-assisted videophotokeratoscope. CLAO J. 1991 Apr;17(2):88–93. [PubMed] [Google Scholar]
  8. Rowsey J. J., Waring G. O., 3rd, Monlux R. D., Balyeat H. D., Stevens S. X., Culbertson W., Barron B., Nelson D., Asbell P., Smith R. Corneal topography as a predictor of refractive change in the prospective evaluation of radial keratotomy (PERK) study. Ophthalmic Surg. 1991 Jul;22(7):370–380. [PubMed] [Google Scholar]
  9. Strelow S., Cohen E. J., Leavitt K. G., Laibson P. R. Corneal topography for selective suture removal after penetrating keratoplasty. Am J Ophthalmol. 1991 Dec 15;112(6):657–665. doi: 10.1016/s0002-9394(14)77271-x. [DOI] [PubMed] [Google Scholar]
  10. Strelow S., Cohen E. J., Leavitt K. G., Laibson P. R. Corneal topography for selective suture removal after penetrating keratoplasty. Am J Ophthalmol. 1991 Dec 15;112(6):657–665. doi: 10.1016/s0002-9394(14)77271-x. [DOI] [PubMed] [Google Scholar]
  11. Szczotka L. B., Capretta D. M., Lass J. H. Clinical evaluation of a computerized topography software method for fitting rigid gas permeable contact lenses. CLAO J. 1994 Oct;20(4):231–236. doi: 10.1097/00140068-199410000-00007. [DOI] [PubMed] [Google Scholar]
  12. Wasserman D., Itzkowitz J., Kamenar T., Asbell P. A. Corneal topographic data: its use in fitting aspheric contact lenses. CLAO J. 1992 Apr;18(2):83–85. [PubMed] [Google Scholar]
  13. Wilson S. E., Klyce S. D. Advances in the analysis of corneal topography. Surv Ophthalmol. 1991 Jan-Feb;35(4):269–277. doi: 10.1016/0039-6257(91)90047-j. [DOI] [PubMed] [Google Scholar]

Articles from Transactions of the American Ophthalmological Society are provided here courtesy of American Ophthalmological Society