Skip to main content
PLOS One logoLink to PLOS One
. 2026 May 6;21(5):e0348631. doi: 10.1371/journal.pone.0348631

From movement to METs: A validation of ActTrust® for energy expenditure estimation and physical activity classification in young adults

Elias dos Santos Batista 1,2,3, Stephania Ruth Basilio Silva Gomes 4, Ayrton Bruno de Morais Ferreira 1, Lucas G S França 5,6, John Fontenele Araújo 3,4,7, Arnaldo Luis Mortatti 1, Mario André Leocadio-Miguel 8,*
Editor: Julio Alejandro Henriques Castro da Costa9
PMCID: PMC13148694  PMID: 42090415

Abstract

Estimating physical activity (PA) levels is a challenging and expensive task. An alternative could be the use of actigraphy devices to estimate PA. This has been previously done to a number of devices, including ActiGraph® GT3X + . In this study, we validated ActTrust® against the widely used GT3X+ and compared activity counts to metabolic equivalents (METs) derived from indirect calorimetry during treadmill walking and running. Fifty-six young adults (34 men, 22 women) participated in controlled effort exercises including light, moderate, vigorous, and very vigorous activity intensities. We developed a linear model to estimate energy expenditure (EE) from movement count of combinations of devices placed at hip or wrist. We then estimated cut-off points for each intensity range. Our results showed correlations between treadmill speed and both METs (r = 0.95, p < 0.05) and movement counts from both GT3X+ and ActTrust devices placed either on the hip (r = 0.94, p < 0.05; r = 0.93, p < 0.05) or on the wrist (r = 0.88, p < 0.05; r = 0.88, p < 0.05), respectively. Our proposed model performed well with balanced accuracies above 0.77 for all intensity ranges and over 0.9 for light and moderate activity. This is the first study to model estimate and validate PA intensity thresholds on ActTrust® devices. Our findings support the use of ActTrust® devices as simple, cost-effective tool for 24-hour assessments of EE.

Introduction

Physical activity (PA) is recognised for providing several health benefits in humans, mainly preventing and controlling chronic non-communicable diseases [1]. Furthermore, the practice of PA is associated with improved subjective and mental well-being and physical self-perception [2]. The World Health Organization (WHO) recommends 150–300 minutes of moderate-intensity aerobic PA, or at least 75–150 minutes of vigorous-intensity aerobic PA, or an equivalent combination of PA of both intensities during the week for adults. Also important is the WHO’s global action plan on physical activity, which aims to decrease the prevalence of sedentarism in adults and adolescents by 15 per cent by 2030 [3].

One way to define the intensity of PA is through the metabolic equivalent (MET) of task. This mechanism expresses energy expenditure (EE) and indicates the oxygen needs for different activities, which at rest is equivalent to 3.5 ml O2/ kg/min [4,5]. The gold standard methodological approach to assessing energy expenditure is the doubly labelled water (DLW) technique [6]. However, there are limitations mainly related to this method’s high cost, complexity and scalability. Alternatively, accelerometry emerges as a practical, scalable and economically viable method to estimate energy expenditure (EE) and quantify PA based on movement and body acceleration, generating numerical variables of frequency, duration and intensity of the actions performed [7,8]. The challenges inherent in using accelerometry include the diversity of designs in accelerometer models, the disparities in signal processing between the different sensors and the multiplicity of options regarding possible placements [9–11]. Furthermore, it is possible to verify a variation in the algorithms that calculate the counts, both of the activity and other extracted parameters on different devices [12].

The ActiGraph® GT3X+ accelerometer has the most significant number of validation studies and published thresholds and has been widely used to validate devices with the same purpose [13]. The ActTrust® (Condor Instruments, São Paulo, Brazil) is a device that has a triaxial accelerometer and is specific for inferring sleep from locomotor activity [14]. It offers several advantages over other actigraphy devices: (1) significantly lower cost, making it accessible for large-scale studies in resource-limited settings; (2) longer battery life enabling extended monitoring periods; (3) integrated light and temperature sensors for comprehensive physiological assessment. However, despite these practical advantages, the ActTrust® has not been validated against gold-standard measures of EE, which limits its use in research and clinical contexts.

The validation of accelerometers through indirect calorimetry is crucial for developing mathematical equations capable of predicting energy expenditure using these devices and establishing cutoff points for classifying the intensity of PA derived from accelerometer data [15]. Based on this context, the two main objectives of this study are: (i) to define equations for estimating EE using ActTrust in young adults, and (ii) to establish cutoff points for the ActTrust movement variable (light, moderate, vigorous and very vigorous PA).

Materials and methods

Participants

Fifty-six healthy young adults were recruited for this study by online advertising. We established musculoskeletal or cardiovascular diseases that could hinder PA as excluding criteria. Participants were also excluded if they had any other contraindications to exercise or were taking medication that affected their metabolic rate. All participants completed the Physical Activity Readiness Questionnaire (PAR-Q) [16–18], and no one was excluded from the study for answering yes to one or more questions. The statistics of the characteristics of the participants are presented in Table 1.

Table 1. Demographic and physical characteristics of the participants. Mean (st. dev.).

Sex Sample size Age (y) Height (cm) Weight (kg) Body mass index
(kg/m²)
Male 34 28.3 (4.6) 172 (6.2) 78.2 (11.9) 23.1 (1.6)
Female 22 26.8 (5.0) 160 (7.3) 59.9 (9.4) 26.7 (2.5)

The study was approved by the Human Ethics Committee of the Federal University of Rio Grande do Norte, under protocol number 3.672.214 (CAAE: 15857419.5.0000.5537). It was performed according to the declaration of Helsinki and complied with the Ethical Standards in Sport and Exercise Science Research. Participants provided written consent to participate in the study. Recruitment took place between March 14, 2022 and October 24, 2022.

The sample size of 56 participants used in this study is consistent with previous validation research comparing accelerometer-based activity monitors against indirect calorimetry. Similar studies validating energy expenditure estimates from devices such as the ActiGraph GT3X+ have used samples ranging from 31 to 50 participants [8,19]. Given the similarity in the study design, controlled laboratory setting and repeated measures across multiple activity intensities and device placements, our sample size provides sufficient statistical power to detect meaningful associations and develop robust predictive models.

Measuring activity

The ActiGraph GT3X+ (ActiGraph Corporation, headquartered in Pensacola, FL, Firmware version 4.1.0) is a lightweight device that employs triaxial accelerometers and offers a dynamic range ranging from −6 g to +6 g. For the current study, a sampling rate of 30 Hz was chosen, as this range is expected to capture the majority of accelerations arising from human movement sufficiently [20]. After digitisation, the signal passed through a digital filter that restricts the accelerometer’s frequency range to 0.25–2.5 Hz. We used the vector magnitude (VM) data (vector representing the summed value of the three axes). Data from the GT3X+ device were retrieved and processed using ActiLife version 6.13.4, software provided by ActiGraph Corporation based in Pensacola, FL.

The ActTrust® device (Condor Instruments, São Paulo, SP, Brazil) is an actigraphy tool that features a rechargeable battery, a digital tri-axial accelerometer, a skin temperature sensor and a light sensor. It samples acceleration data at 25 Hz, ranging from 0.03 to 4 g. First, this data undergoes processing through a band-pass filter (0.5–2.7 Hz), after which the norm is calculated from the three axes. The resultant value is then integrated over an epoch (1–60 seconds) and stored in memory. This equipment is designed for activity and sleep monitoring and processes activity using parameters such as PIM (proportional integration mode), ZCM (zero crossing mode) and TAT (time above threshold) [21]. The Act Studio software (Condor Instruments, São Paulo, Brazil) downloads, visualises, and exports the collected data.

Importantly, this study used the highest temporal resolution available for both actigraphy devices, according to their specific firmware. Each 1-second epoch was calculated by aggregating filtered and rectified raw acceleration data into a single 1-second bin. This was consistent with validation approaches for wrist and hip-worn accelerometers across the same range of walking speeds [8,19].

Measuring oxygen uptake

Oxygen uptake was measured ‘breath-by-breath’ continuously during each condition using indirect calorimetry (Quark CPET, Cosmed®, Italy). Before testing each subject, we calibrated the metabolic cart with a known gas mixture (16% O2 and 5% CO2) and volume. One test had to be repeated seven days later due to an error in the treadmill’s security system (Super ALT, Inbramed®, Porto Alegre, Brazil). Occasional errant breath episodes (e.g., due to coughing, swallowing or talking) were excluded from the data set when exceeding three standard deviations of the mean, the latter being defined as the average of two following and two preceding sampling intervals [22].

Experimental procedure

The entire data collection took place between March 21, 2022, and November 1, 2022. All GT3X+ and ActTrust devices were initialiased via computer interfaces to collect data in 1-second epochs [19]. The GT3X + was initialised to collect the VM data, and the ActTrust to collect activity in the PIM mode. Each participant wore two pairs of GT3X+ and ActTrust devices simultaneously, securely placed on the hip on the side of the participant’s dominant lower limb through an elastic belt and the other on the non-dominant wrist.

Participants refrained from exercise and fasted for at least 4 hours before the testing session. The protocol consisted of 5 conditions (of 10-min duration each), using a progressive incremental design, interspersed with 5-minute rest periods: (i) resting; (ii) treadmill (Quasar Med 4.0, h p cosmos, Nussdorf-Traunstein, Germany) walking at 3 km·h−1; (iii) treadmill walking at 5 km·h−1; (iv) treadmill walking or running at 7 km·h−1; and (v) treadmill running at 9 km·h−1. Prior to initiating the protocol, devices were synchronised using a manual time-matching approach [23], allowing precise temporal alignment between the accelerometers and breath-by-breath metabolic data.

Data analysis

To ensure that data reflected true physiological steady states, we adopted the protocol described in previous studies [8,19]. Thus, we calculated the mean activity counts from the four central minutes of the Vector Magnitude (VM) for GT3X+ and the counts (proportional integral mode – PIM) for ActTrust to estimate activity levels. METs (Metabolic Equivalents) were individually calculated (VO²) similarly.

First, we assessed the effect of the different devices and placement with an ANOVA according to the model Treadmill speed~Activity1/2:Deviceplacement. Secondly, we then modelled the association between METs and activity for each device using a general linear model defined in equation 1. A square root transformation was applied to both METs and activity counts due to the non-linear relationship. Model assumptions were verified through examination of residual plots and Q-Q plots for normality – included in supplementary material S3 Fig in S1 File.

MET1/2~Activity1/2*Deviceplacement (1)

We evaluated two devices according to their placement, hence our model can be written as in equation 2.

MET1/2=β0†+β1†Activity1/2 (2)

with β0†=β0+β2ACTThip+β3ACTTwrist+β4GT3X+wrist and β1†=β1+β5ACTThip+β6ACTTwrist+β7GT3X+wrist. ACTThip, ACTTwrist, and GT3X+wrist are binary variables set to adjust the model values to each configuration. GT3X+hip was defined as the reference level.

Analyses were conducted using R v. 4.4.1 (R Foundation for Statistical Computing, Vienna, Austria) [24] and auxiliary packages tidyverse v. 2.0.0 [25], caret v. 6.0.94 [26], MuMIn v. 1.48.4 [27], MetBrewer v. 0.2.0 [28], cowplot v. 1.1.3 [29], pROC v. 1.18.5 [30], broom v. 1.0.6 [31], lmtest v. 0.9.40 [32], car v. 3.1.2 [33], effectSize v. 1.0.2 [34], and msm v. 1.8.2 [35]. The significance level was set at α = 0.05. The data and scripts used in the analyses featured in this article are available at https://github.com/circadia-bio/ACTT_validation_study [36].

We estimated metabolic rate in METs to derive the cut-points for moderate, vigorous and very vigorous exercise intensity levels according to the intervals (3.0–5.99, 6.0–8.99 and above 9.0 METs).

We employed performance metrics for model diagnostics and computed R2 and confusion matrix. Finally, we evaluated the ability of the new cut points from the linear regression model to accurately classify the physical activity intensity levels derived from EE objective measurements, according to the different MET bands with sensitivity, specificity, balanced accuracy, and area-under-the-curve (AUC) bench marks in a similar fashion to previous works in literature [8].

Results

The sample comprised young adults (34 men, 22 women) aged 18–35. The participants’ characteristics are displayed in Table 1. Energy expenditure values obtained with indirect calorimetry reflected each condition, increasing proportionally according to the treadmill speed, see Fig 1.

Fig 1. Scores for each treadmill speed band.

Fig 1

a) Metabolic equivalents (VO2) obtained from all participants for each speed band. b) Activity counts obtained for each speed band for every combination of placement and device, i.e., GT3X+ (hip), ACTT (hip), ACTT (wrist), GT3X+ (wrist). The scores across the different devices show a similar profiles increasing towards higher speeds, which is compatible to the variation exhibited by the ergo-spirometry metabolic equivalents measures. Point shape denotes sex (square: male, n = 34; circle: female, n = 22).

Activity counts and treadmill velocity

We performed a two-way ANOVA to understand the outputs of each device (GT3X+ and ActTrust) according to the placement and condition (treadmill velocity). Movement counts from the GT3X+ devices were mostly distinct (F(19, 1087) = 1347, p < 0.001, η2=0.96, 95% CI [0.96, 1.00]) with significantly lower activity magnitudes than those from the ActTrust devices at 3, 5, 7 and 9 km·h−1 (p < 0.05).

In contrast, placement did not affect the measures for GT3X+ regardless of hip or wrist acquisitions for 3 km·h−1 and 5 km·h−1. Movement counts of both devices (GT3X+ and ActTrust) placed on the hip or wrist increased with the increment of the treadmill speed (see Fig 1). Statistical details of the ANOVA analysis and post-hoc Tukey’s Honestly Significant Difference (HSD) [37,38] tests were included in the supplementary table S1 in S1 File.

Definition of equations to estimate energy expenditure

Our main objective was to analyse the relationship between motion measurements obtained with the ActTrust® in comparison to the GT3X + ® and metabolic equivalent based on gas exchange in young adults. First, we explored how treadmill speed is associated with METs and movement counts. Specifically, speed was highly correlated to both METs (r = 0.95, p < 0.05) and movement counts from both GT3X+ and ActTrust devices placed either on the hip (r = 0.94, p < 0.05; r = 0.93, p < 0.05) or on the wrist (r = 0.88, p < 0.05; r = 0.88, p < 0.05), respectively. Our second step was to develop a linear regression model to estimate METs based on the movement counts recorded by each device at each placement (see Fig 2c). The resulting equations are crucial for estimating physical activity in METs from the movement counts of the GT3X+ and ActTrust devices placed on the hip and wrist (see Table 2).

Fig 2. Cross device score analyses.

Fig 2

a) Pearson correlation between the activity counts magnitudes measured at each device and placement combination. b) Model of Activity count and METs following the relationship Treadmill speed1/2~Activity1/2×Deviceplacement. c) Confusion matrix relating predicted intensity ranges from model in b) and metabolic equivalents (VO2) obtained from simultaneous ergo-spirometry measurements, with a high true positive rate for the classes [0,3), [3,6) and [9,∞).

Table 2. Cut-points and regression equations with 95% confidence intervals per device and placement.

Cut-point, counts/min (95% CI) MET1/2=β0+β1×Activity1/2
Device 3 METs 6 METs 9 METs β0 (95% CI) β1 (95% CI)
GT3X+ (hip) 1132 (1041, 1223) 4853 (4688, 5018) 9468 (9135, 9800) 1.062 (1.019, 1.105) 0.0199 (0.0193, 0.0206)
ACTT (hip) 5057 (4624, 5491) 23339 (22522, 24157) 46410 (44764, 48055) 1.107 (1.066, 1.148) 0.0088 (0.0085, 0.0091)
ACTT (wrist) 3761 (3357, 4165) 22368 (21503, 23232) 47203 (45409, 48997) 1.233 (1.195, 1.271) 0.0081 (0.0079, 0.0084)
GT3X+ (wrist) 1698 (1525, 1871) 9503 (9143, 9862) 19787 (19047, 20527) 1.207 (1.168, 1.245) 0.0127 (0.0123, 0.0132)

GT3X + ® and ActTrust® movement count cut points to classify physical activity intensity

Finally, we derived cut-points based on MET ranges [0,3) or light, [3,6) or moderate, [6,9) or vigorous, [9,∞) or very vigorous physical activity intensity ranges based on the movement counts from both devices and placement. We then assessed the ability of the new cut points from our linear regression model to accurately classify the physical activity intensity levels objectively derived from EE measurement (METs). We estimated the sensitivity (recall), specificity, balanced accuracy, area-under-the-curve (AUC), and confusion matrix for each intensity zone (for metrics see Table 3). The obtained confusion matrices show a consistent high accuracy for the [0,3); [3,6); and [9, ∞) classes, featuring true positive rates above 75%, both when aggregating all the devices and placements (see Fig 2b) and for specific combinations of device and placement (see Fig 3). Sensitivity and balanced accuracy are reduced for vigorous and very vigorous activity but remained well above 0.6, see Table 3.

Table 3. Sensitivity, specificity, balanced accuracy, and area under the curve for the [0,3), [3,6), [6,9), and [9,∞) intensity ranges.

Metric Class: [0,3) Class: [3,6) Class: [6,9) Class: [9,∞)
Sensitivity 0.8810 0.9190 0.6630 0.6441
Specificity 0.9762 0.9182 0.8932 0.9556
Balanced Accuracy 0.9286 0.9186 0.7781 0.7998
AUC 0.9820 0.6307 0.7813 0.9550

Fig 3. Confusion matrices for each combination of device and placement, i.e., a) GT3X+ (hip), b) GT3X+ (wrist), c) ACTT (hip), ACTT (wrist).

Fig 3

The individual devices confusion matrices reproduce a similar profile to the one observed for the entire model, with a higher true positive rate for the classes [0,3), [3,6) and [9,∞).

Discussion

This study compared the ActTrust® and ActiGraph GT3X + ® activity counts during resting, treadmill walking and running. The main findings are as follows. Firstly, we established the suitability of our indirect calorimetry setup by measuring the metabolic equivalent as a function of the conditions (treadmill speed, resting). The speed of the treadmill was highly and significantly correlated with the estimated METs. Secondly, we demonstrated that movement counts increased as speed increased, and both ActTrust® and GT3X + ® could differentiate among the different conditions, independently of the placement, either on the wrist or the hip. As a next step, we then modelled the association between METs and activity for each device X placement using a general linear model. In addition, we established cut points for movement counts of GT3X+ and ActTrust devices to classify the intensity of physical activity. These points were derived for the categories of moderate, intense and very intense physical activity (see Table 2), considering both the movement counts of the devices and the placement. Finally, we evaluated the ability of the new cut points from the linear regression models to accurately classify the intensity levels of physical activity objectively derived from EE measurement (METs). For this analysis, we calculated the balanced accuracy, sensitivity (recall), and specificity for each cut point and each linear model (device X placement).

Our results demonstrated that the movement count of both devices (GT3X+ and ActTrust), placed on both the hip and the wrist, increased with the increase in the velocity of the treadmill, as did the energy expenditure (METs) obtained from indirect calorimetry. Our results are consistent with similar validation studies [39,40]. If we compare our results with studies which used different devices [8,19], while our model’s sensitivity and accuracy are reduced for vigorous and very vigorous activity, it still achieves accuracies above 0.90 and higher AUC scores than previous studies. Precisely, our model achieved area under the ROC curve of 0.98, 0.63, 0.78 and 0.95 for light, moderate, vigorous and very vigorous physical activity, respectively. As a comparison, the artificial neural network model developed by Santos-Lozano et al., resulted in AUC of 0.8, 0.6, 0.7 and 0.6 using a very similar protocol [8]. Therefore, we found that the model we proposed is accurate with a simpler method for the estimation of energy expenditure (EE) than the equations previously developed. Furthermore, our approach is fully explainable unlike neural network models.

The cut-points derived in the present study for the GT3X+ placed on the hip were lower than most published GT3X reference values across all thresholds. Compared to Sasaki et al. [19], our cut-points were 58%, 21%, and 2% lower at the MET 3, 6, and 9 thresholds, respectively, and 65%, 43%, and 18% lower than Santos-Lozano et al. [8] respective cut-off points for adults. Importantly, however, there is substantial variability in cut-points between these two quite similar studies, as they differ by 16% at MET 3, 39% at MET 6 and 20% at MET 9. The overall variability suggests that sample-level characteristics are the main drivers of the observed discrepancies rather than device or methodological artefacts.

Nevertheless, this is the first study to develop movement-based cut-off points for wrist and hip-worn ActTrust® devices. These cut-offs reflect moderate, intense and very intense physical activity categories. The study successfully classified physical activity intensity across the spectrum, demonstrating excellent accuracy. However, it is not possible to compare these values with other studies. For the discrimination between intensity categories based on METs, we propose thresholds to be used for the hip and wrist-worn ActTrust® devices. This allows researchers to examine how much walking people perform and distinguish it from running using these accelerometers.

Beyond estimating the intensity of physical activity, the value of validating ActTrust® lies in its ability to capture the entire 24-hour rest-activity cycle within a single device. Therefore, establishing PA cut-points for ActTrust® is a necessary first step toward leveraging its broader monitoring capabilities, which extend beyond exercise to the assessment of sleep and circadian rhythms. Actigraphy devices, such as the ActTrust® accelerometers have been used in sleep and circadian research for decades [41–43]. Actually, the sleep and wake states have been classically inferred by algorithms based on activity counts, provided frequency, duration and intensity of movement [44]. Although sleep studies are designed to collect 24-hour data, including rest and physical activity throughout the day, the simultaneous analysis of sleep/circadian data, physical activity, and sedentary behaviour is limited [13]. This gap is notable given strong evidence linking regular exercise, reduced sedentary behaviour, and healthy sleep to improved longevity and disease prevention [45,46].

Despite this clear dependence between physical activity and sleep, it is essential to move one step further and take advantage of the complete 24-hour rest-activity cycle data obtained through actigraphy. Precise health guidelines can only be determined by combining measurements of sleep, sedentary behaviour, and physical activity [13]. That said, expanding the availability of cut-off points for physical activity based on multiple actigraphy devices increases the potential to improve recommendations, guidelines and, ultimately, health and longevity. Therefore, studies like ours and the introduction of more afordable actigraphy devices, such as the ActTrust®, have the potential to stimulate future research by providing tools to extract physical activity and sleep information, even retrospectively. In fact, one of the main advantages of using accelerometry over self-reported (retrospective) physical activity is that it is not prone to recognition or memory biases, as recall measures are [47]. The validation of devices that were exclusively dedicated to inferring sleep-wake cycles as tools to identify and classify physical activity according to the metabolic equivalent opens up the possibility of using such data not only prospectively, but also retrospectively. Studies such as ours are crucial to developing personalised programmes that maximise long-term adherence to physical activity and the long-term benefits of exercise. We acknowledge that adults should aim for at least 150–300 minutes of moderate-intensity aerobic physical activity, or 75–150 minutes of vigorous-intensity aerobic physical activity, or an equivalent combination throughout the week, to achieve substantial health benefits [48]. Likewise, adults should usually aim for 7 or more hours of regular sleep each night to promote optimal health [49]. However, only validated accelerometer devices can provide information on these two essential proxies of health simultaneously (sleep and physical activity).

It is important to acknowledge the limitations in the findings of this study. Although a significantly large number of participants (N = 56) was considered, the participants were young healthy adults, with no functional limitations, which limits generalisability. As a result, the findings may not be applicable to individuals with health conditions and other age groups without further validation. In addition, the study was carried out in a laboratory setting and focused on specific physical activities, walking and running, which may also limit the universality of our findings. Although these activities are common among populations around the world, it is crucial to assess physical activities in real life and home environments. Moreover, the fixed progressive protocol may have introduced order effects such as those linked to warm-up or fatigue. While this approach was chosen for safety, and mirrors clinical exercise testing protocols, future studies could employ counterbalanced designs or include longer rest periods to eliminate carryover effects.

Further work is needed to define appropriate cut-off points for other groups, such as children and older people, as well as different device placements, as our findings are limited to devices worn on the wrist and hip. Furthermore, the exercise protocol followed a fixed incremental order, which precludes ruling out potential order effects such as progressive fatigue or warm-up adaptation; future studies should consider counterbalanced or randomised condition sequences to address this limitation.

Conclusion

In this study we have for the first time validated measures of the ActTrust®. When paired with a simple linear model, we can estimate energy expenditure and classify the intensity of physical activity. Our model distinguished between light, moderate, vigorous, and very vigorous activity intensity ranges with benchmark metrics comparable to or exceeding those reported in the literature, whilst adopting a much simpler and explainable model than black-box approaches such as neural networks, facilitating interpretation and implementation by practitioners. Nevertheless, we did not directly compare performance against such alternative modelling approaches in this study.

These results demonstrated the feasibility of using ActTrust® as a cost effective and scalable tool to estimate PA. It also has potential for potential for clinical use following validation in patient populations with diverse health conditions and functional capacities. Our results also provide hints on how to harmonise outcomes from different devices whilst estimating PA and analysing activity measures. Future work should extend our analyses to free-living conditions, comparison with commercial wearable devices other than GT3X + ® and more naturalistic activity patterns.

Supporting information

S1 Table. ANOVA analysis and post-hoc Tukey’s Honestly Significant Difference 316 (HSD) table.

(CSV)

pone.0348631.s001.csv (22.5KB, csv)
S2 Table. Area under the curve (AUC) for one-vs-rest physical activity intensity classification stratified by sex.

AUC values were computed for each intensity class using the model prediction score against the true class derived from indirect calorimetry for male (n = 34) and female (n = 22) participants. Intensity classes are defined as light [0,3), moderate [3,6), vigorous [6,9), and very vigorous [9,∞) METs.

(CSV)

pone.0348631.s002.csv (106B, csv)
S1 Figure. Confusion matrices for physical activity intensity classification stratified by sex.

Row-normalised confusion matrices showing the proportion of observations assigned to each predicted intensity class relative to the true class derived from indirect calorimetry for (A) male (n = 34) and (B) female (n = 22) participants. Intensity classes are defined as light [0,3), moderate [3,6), vigorous [6,9), and very vigorous [9,∞) METs.

(PDF)

pone.0348631.s003.pdf (5.8KB, pdf)
S2 Figure. Bland-Altman plots comparing model-predicted and measured metabolic equivalents (METs) for each device and placement combination.

Each panel plots the difference between predicted and measured METs against their mean for each device and placement with GT3X + hip (A), GT3X+ wrist (B), ActTrust hip (C), and ActTrust wrist (D). Red dashed lines indicate the mean bias and blue dotted lines indicate the 95% limits of agreement. There is no evidence of systematic proportional bias across the range of measured values.

(PDF)

pone.0348631.s004.pdf (66.1KB, pdf)
S3 Figure. Regression diagnostics for the general model.

A) Normal Q-Q plot of model residuals with points following the reference blue line, featuring minor deviations at the tails. B) Residuals versus fitted values with no systematic trend supporting the assumption of homoscedasticity.

(PDF)

pone.0348631.s005.pdf (122.5KB, pdf)

Data Availability

The data and source code used in the analyses featured in this article are available at https://github.com/circadia-bio/ACTT_validation_study.

Funding Statement

The author(s) received no specific funding for this work.

References

  • 1.Anderson E, Durstine JL. Physical activity, exercise, and chronic diseases: A brief review. Sports Med Health Sci. 2019;1(1):3–10. doi: 10.1016/j.smhs.2019.08.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Buecker S, Simacek T, Ingwersen B, Terwiel S, Simonsmeier BA. Physical activity and subjective well-being in healthy individuals: a meta-analytic review. Health Psychol Rev. 2021;15(4):574–92. doi: 10.1080/17437199.2020.1760728 [DOI] [PubMed] [Google Scholar]
  • 3.Bull FC, Al-Ansari SS, Biddle S, Borodulin K, Buman MP, Cardon G, et al. World Health Organization 2020 guidelines on physical activity and sedentary behaviour. Br J Sports Med. 2020;54(24):1451–62. doi: 10.1136/bjsports-2020-102955 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.de Jong A. The metabolic equivalent: reevaluating what we know about the MET. ACSM’s Health & Fitness Journal. 2010;14(4):43–6. [Google Scholar]
  • 5.Franklin BA, Brinks J, Berra K, Lavie CJ, Gordon NF, Sperling LS. Using Metabolic Equivalents in Clinical Practice. Am J Cardiol. 2018;121(3):382–7. doi: 10.1016/j.amjcard.2017.10.033 [DOI] [PubMed] [Google Scholar]
  • 6.Speakman J. Doubly Labelled Water: Theory and Practice. Springer Science & Business Media.
  • 7.Migueles JH, Cadenas-Sanchez C, Ekelund U, Delisle Nyström C, Mora-Gonzalez J, Löf M, et al. Accelerometer Data Collection and Processing Criteria to Assess Physical Activity and Other Outcomes: A Systematic Review and Practical Considerations. Sports Med. 2017;47(9):1821–45. doi: 10.1007/s40279-017-0716-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Santos-Lozano A, Santín-Medeiros F, Cardon G, Torres-Luque G, Bailón R, Bergmeir C, et al. Actigraph GT3X: validation and determination of physical activity intensity cut points. Int J Sports Med. 2013;34(11):975–82. doi: 10.1055/s-0033-1337945 [DOI] [PubMed] [Google Scholar]
  • 9.van Hees VT, Thaler-Kall K, Wolf KH, Brønd JC, Bonomi A, Schulze M, et al. Challenges and opportunities for harmonizing research methodology: Raw accelerometry. Methods of information in medicine. 2016;55(06):525–32. [DOI] [PubMed] [Google Scholar]
  • 10.Rosenberger ME, Haskell WL, Albinali F, Mota S, Nawyn J, Intille S. Estimating activity and sedentary behavior from an accelerometer on the hip or wrist. Med Sci Sports Exerc. 2013;45(5):964–75. doi: 10.1249/MSS.0b013e31827f0d9c [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Redenius N, Kim Y, Byun W. Concurrent validity of the Fitbit for assessing sedentary behavior and moderate-to-vigorous physical activity. BMC Med Res Methodol. 2019;19(1):29. doi: 10.1186/s12874-019-0668-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Neishabouri A, Nguyen J, Samuelsson J, Guthrie T, Biggs M, Wyatt J, et al. Quantification of acceleration as activity counts in ActiGraph wearable. Sci Rep. 2022;12(1):11958. doi: 10.1038/s41598-022-16003-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Rosenberger ME, Buman MP, Haskell WL, McConnell MV, Carstensen LL. Twenty-four Hours of Sleep, Sedentary Behavior, and Physical Activity with Nine Wearable Devices. Med Sci Sports Exerc. 2016;48(3):457–65. doi: 10.1249/MSS.0000000000000778 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Bellone GJ, Plano SA, Cardinali DP, Chada DP, Vigo DE, Golombek DA. Comparative analysis of actigraphy performance in healthy young subjects. Sleep Sci. 2016;9(4):272–9. doi: 10.1016/j.slsci.2016.05.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Chen KY, Bassett DR Jr. The technology of accelerometry-based activity monitors: current and future. Med Sci Sports Exerc. 2005;37(11 Suppl):S490-500. doi: 10.1249/01.mss.0000185571.49104.82 [DOI] [PubMed] [Google Scholar]
  • 16.Chisholm D, Collis M, Kulak L, Davenport W, Gruber N. Physical activity readiness. BC Med J. 1975;17(2):375–8. [Google Scholar]
  • 17.Warburton DE, Gledhill N, Jamnik VK, Bredin SS, McKenzie DC, Stone J. Evidence-based risk assessment and recommendations for physical activity clearance: Consensus Document 2011. Applied Physiology, Nutrition, and Metabolism. 2011;36(S1):S266–98. [DOI] [PubMed] [Google Scholar]
  • 18.Schwartz J, Oh P, Takito MY, Saunders B, Dolan E, Franchini E, et al. Translation, Cultural Adaptation, and Reproducibility of the Physical Activity Readiness Questionnaire for Everyone (PAR-Q+): The Brazilian Portuguese Version. Front Cardiovasc Med. 2021;8:712696. doi: 10.3389/fcvm.2021.712696 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Sasaki JE, John D, Freedson PS. Validation and comparison of ActiGraph activity monitors. J Sci Med Sport. 2011;14(5):411–6. doi: 10.1016/j.jsams.2011.04.003 [DOI] [PubMed] [Google Scholar]
  • 20.Matlary RED, Holme PA, Glosli H, Rueegg CS, Grydeland M. Comparison of free-living physical activity measurements between ActiGraph GT3X-BT and Fitbit Charge 3 in young people with haemophilia. Haemophilia. 2022;28(6):e172–80. doi: 10.1111/hae.14624 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Ancoli-Israel S, Cole R, Alessi C, Chambers M, Moorcroft W, Pollak CP. The role of actigraphy in the study of sleep and circadian rhythms. Sleep. 2003;26(3):342–92. doi: 10.1093/sleep/26.3.342 [DOI] [PubMed] [Google Scholar]
  • 22.Lamarra N, Whipp BJ, Ward SA, Wasserman K. Effect of interbreath fluctuations on characterizing exercise gas exchange kinetics. J Appl Physiol (1985). 1987;62(5):2003–12. doi: 10.1152/jappl.1987.62.5.2003 [DOI] [PubMed] [Google Scholar]
  • 23.Crouter SE, Kuffel E, Haas JD, Frongillo EA, Bassett DR. Refined two-regression model for the ActiGraph accelerometer. Medicine & Science in Sports & Exercise. 2023;42(5):1029–37. doi: 10.1249/MSS.0b013e3181c37458 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.R Core Team. R: A Language and Environment for Statistical Computing. 2024.
  • 25.Wickham H, Averick M, Bryan J, Chang W, McGowan L, François R, et al. Welcome to the Tidyverse. JOSS. 2019;4(43):1686. doi: 10.21105/joss.01686 [DOI] [Google Scholar]
  • 26.Kuhn M. Building Predictive Models in R Using the caret Package. J Stat Soft. 2008;28(5). doi: 10.18637/jss.v028.i05 [DOI] [Google Scholar]
  • 27.Bartoń K. MuMIn: Multi-Model Inference. 2024.
  • 28.Mills BR. MetBrewer: Color Palettes Inspired by Works at the Metropolitan Museum of Art. https://CRAN.R-project.org/packagRioGrandedoNorte=MetBrewer 2022.
  • 29.Wilke CO. cowplot: Streamlined Plot Theme and Plot Annotations for ’ggplot2’. https://CRAN.R-project.org/package=cowplot 2024.
  • 30.Robin X, Turck N, Hainard A, Tiberti N, Lisacek F, Sanchez J-C, et al. pROC: an open-source package for R and S+ to analyze and compare ROC curves. BMC Bioinformatics. 2011;12:77. doi: 10.1186/1471-2105-12-77 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Robinson D, Hayes A, Couch S. broom: Convert statistical objects into tidy tibbles. 2024.
  • 32.Zeileis A, Hothorn T. Diagnostic checking in regression relationships. 2002. https://CRAN.R-project.org/doc/Rnews/
  • 33.Fox J, Weisberg S. An R companion to applied regression. 3rd ed. Thousand Oaks CA: Sage. 2019. [Google Scholar]
  • 34.Ben-Shachar M, Lüdecke D, Makowski D. effectsize: Estimation of Effect Size Indices and Standardized Parameters. JOSS. 2020;5(56):2815. doi: 10.21105/joss.02815 [DOI] [Google Scholar]
  • 35.Jackson CH. Multi-State Models for Panel Data: ThemsmPackage forR. J Stat Soft. 2011;38(8). doi: 10.18637/jss.v038.i08 [DOI] [Google Scholar]
  • 36.dos Santos Batista E, Basilio Gomes SR, Bruno de Morais Ferreira A, França LGS, Fontenele Araújo J, Mortatti AL. From movement to METs: A validation of ActTrust® for energy expenditure estimation and physical activity classification in young adults. 2025. [DOI] [PubMed]
  • 37.Miller RG. Simultaneous Statistical Inference. New York, NY: Springer New York. 1981. [Google Scholar]
  • 38.Yandell BS. Practical Data Analysis for Designed Experiments. 1st ed. Routledge. 2017. [Google Scholar]
  • 39.Davoudi A, Wanigatunga AA, Kheirkhahan M, Corbett DB, Mendoza T, Battula M, et al. Accuracy of Samsung Gear S Smartwatch for Activity Recognition: Validation Study. JMIR Mhealth Uhealth. 2019;7(2):e11270. doi: 10.2196/11270 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Skotte J, Korshøj M, Kristiansen J, Hanisch C, Holtermann A. Detection of physical activity types using triaxial accelerometers. J Phys Act Health. 2014;11(1):76–84. doi: 10.1123/jpah.2011-0347 [DOI] [PubMed] [Google Scholar]
  • 41.Yuan H, Hill EA, Kyle SD, Doherty A. A systematic review of the performance of actigraphy in measuring sleep stages. J Sleep Res. 2024;33(5):e14143. doi: 10.1111/jsr.14143 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Sadeh A. The role and validity of actigraphy in sleep medicine: an update. Sleep Med Rev. 2011;15(4):259–67. doi: 10.1016/j.smrv.2010.10.001 [DOI] [PubMed] [Google Scholar]
  • 43.Evangelista LA, Eckeli AL, Stelzer FG, Carvalho VD, Soares HF. Validation of a national actigraph in children and adolescents. Sleep. 2025;138:107776. doi: 10.1016/j.sleep.2025.107776 [DOI] [Google Scholar]
  • 44.Fekedulegn D, Andrew ME, Shi M, Violanti JM, Knox S, Innes KE. Actigraphy-Based Assessment of Sleep Parameters. Ann Work Expo Health. 2020;64(4):350–67. doi: 10.1093/annweh/wxaa007 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Morris CJ, Yang JN, Scheer FAJL. The impact of the circadian timing system on cardiovascular and metabolic function. Prog Brain Res. 2012;199:337–58. doi: 10.1016/B978-0-444-59427-3.00019-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Duncan MJ, Murphy L, Oftedal S, Fenwick MJ, Vincent GE, Fenton S. The associations between physical activity, sedentary behaviour, and sleep with mortality and incident cardiovascular disease, cancer, diabetes and mental health in adults: a systematic review and meta-analysis of prospective cohort studies. J Act Sedentary Sleep Behav. 2023;2(1):19. doi: 10.1186/s44167-023-00026-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Burchartz A, Anedda B, Auerswald T, Giurgiu M, Hill H, Ketelhut S, et al. Assessing physical behavior through accelerometry – State of the science, best practices and future directions. Psychology of Sport and Exercise. 2020;49:101703. doi: 10.1016/j.psychsport.2020.101703 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Burtscher J, Millet GP, Burtscher M. Pushing the Limits of Strength Training. Am J Prev Med. 2023;64(1):145–6. doi: 10.1016/j.amepre.2022.07.018 [DOI] [PubMed] [Google Scholar]
  • 49.Consensus Conference Panel, Watson NF, Badr MS, Belenky G, Bliwise DL, Buxton OM, et al. Recommended Amount of Sleep for a Healthy Adult: A Joint Consensus Statement of the American Academy of Sleep Medicine and Sleep Research Society. J Clin Sleep Med. 2015;11(6):591–2. doi: 10.5664/jcsm.4758 [DOI] [PMC free article] [PubMed] [Google Scholar]

Decision Letter 0

Julio Alejandro Henriques Castro da Costa

9 Sep 2025

-->PONE-D-25-31357-->-->From Movement to METs: A Validation of ActTrust® for Energy Expenditure Estimation and Physical Activity Classification in Young Adults-->-->PLOS ONE

Dear Dr. Leocadio-Miguel,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Oct 24 2025 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:-->

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Julio Alejandro Henriques Castro da Costa

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1.Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please note that PLOS One has specific guidelines on code sharing for submissions in which author-generated code underpins the findings in the manuscript. In these cases, we expect all author-generated code to be made available without restrictions upon publication of the work. Please review our guidelines at https://journals.plos.org/plosone/s/materials-and-software-sharing#loc-sharing-code and ensure that your code is shared in a way that follows best practice and facilitates reproducibility and reuse.

3. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match.

When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section.

4. Please note that your Data Availability Statement is currently missing the repository name and/or the DOI/accession number of each dataset OR a direct link to access each database. If your manuscript is accepted for publication, you will be asked to provide these details on a very short timeline. We therefore suggest that you provide this information now, though we will not hold up the peer review process if you are unable.

5. Please amend either the abstract on the online submission form (via Edit Submission) or the abstract in the manuscript so that they are identical.

6. If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

-->Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. -->

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

Reviewer #4: Partly

**********

-->2. Has the statistical analysis been performed appropriately and rigorously? -->

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

Reviewer #4: No

**********

-->3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.-->

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

Reviewer #4: Yes

**********

-->4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.-->

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

Reviewer #4: Yes

**********

-->5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)-->

Reviewer #1: This was quite an interesting study and had a clear and concise presentation of methods and results. The authors have presented the findings in a logical manner, which makes it easy for readers to follow the narrative. The findings are novel and contribute to the technological development for improving physical activity among the population. However, I have a few comments that the authors may consider to further strengthen the study.

1. Was the gender difference considered for validation? This could be relevant since there is an expected difference between males and females in terms of body composition and energy expenditure, which may act as confounding.

2. While correlation matrices were reported, additional analysis such as Bland-Altman plots could reveal further details about the systematic bias.

Addressing these aspects may help improve the robustness and translational value of the work.

Reviewer #2: Dear researcher,

I received with interest your manuscript submitted to PLOS ONE entitled "From Movement to METs: A Validation of ActTrust® for Energy Expenditure Estimation and Physical Activity Classification in Young Adults." It is a relevant, timely, and methodologically sound study in many respects. The validation of accessible and low-cost devices, such as ActTrust®, is crucial for the expansion of physical activity and health research, especially in resource-limited settings. Congratulations on your initiative and the technical rigor employed.

However, as your advisor and reviewer, I feel compelled to point out areas that can and should be improved to increase the article's quality, clarity, scientific rigor, and chances of acceptance. Let's get into the details:

1. Text Structure and Clarity

Positive Points:

The abstract is clear, straightforward, and follows the IMRaD structure (Introduction, Methods, Results, and Discussion) appropriately.

The introduction contextualizes the problem well: the importance of physical activity, the role of METs, the limitations of DLW, and the need for accelerometer validation.

The objective is well defined: to validate ActTrust® and establish cutoff points for activity intensities.

Points for improvement:

Redundancy in the text: There is repetition of sentences between the abstract, introduction, and discussion. For example, the phrase "physical activity is recognized for providing health benefits" appears three times. Avoid this. Each section should have a unique function.

Lack of flow in some paragraphs: Some sections are too dense or poorly connected. Example:

"Accelerometers are not only used to estimate exercise or energy expenditure; they are extremely useful to assess the sleep-wake cycle..."

- This transition is abrupt. Use connectives and transitional paragraphs.

Suggestion: Revise the text with a focus on clarity, conciseness, and logical progression. Use shorter sentences and avoid unnecessary jargon.

2. Methodology

Strong Points:

Sample size (N=56): Adequately justified based on similar studies. Good statistical power.

Control of experimental conditions: Standardized protocol (fasting, resting, fixed speeds, use of an elastic belt).

Use of indirect calorimetry (Quark CPET): Well-applied gold standard.

Preprocessing of respiration data: Exclusion of erratic episodes with a robust criterion (±3 standard deviations).

Data and code availability: Excellent open practice (GitHub). This strengthens reproducibility.

Points for improvement:

Poor justification for choosing 1 second as the epoch:

You state that you used 1 second based on [8], but you don't discuss why this epoch is suitable for all intensities, especially light walking (3 km/h), where movements are smoother.

Problem: Very short epochs increase noise; very long ones lose temporal resolution. Suggestion: Include a brief theoretical justification or cite studies that validated 1s for low-intensity activities.

Lack of detail on synchronization between devices and calorimetry:

How did you synchronize the timing between the metabolism (Quark CPET) and the accelerometers?

Without precise synchronization, there is a risk of temporal misalignment, especially with sudden changes in intensity.

Suggestion: Include a paragraph in "Experimental Procedure" explaining how the synchronization was achieved (e.g., manual timing, triggering, software).

Lack of randomization of the order of conditions:

The protocol always follows: rest → 3 → 5 → 7 → 9 km/h.

This may introduce an order effect (fatigue, learning, warm-up).

Suggestion: State whether the order was fixed and discuss how this may have affected the results (limitation).

3. Statistical Analysis

Strong Points:

Use of two-way ANOVA to compare devices and positions: appropriate.

Generalized linear model with interaction between device and position: well-formulated.

Presentation of robust metrics: sensitivity, specificity, balanced accuracy, AUC.

Points for improvement:

Transformation of the dependent variable (MET^1/2):

You used MET to the square root as the dependent variable.

This is unusual and not justified in the text.

Transformations should be explained: was it to normalize residuals? To increase linearity?

Suggestion: Justify the transformation or test whether the model with linear MET performs as well (or better). If not necessary, remove it.

Regression model with multiple interactions:

The model (Eq. 2) includes complex interaction terms, but there is no test of regression assumptions (normality of residuals, homoscedasticity, multicollinearity).

Suggestion: Include a paragraph in "Data Analysis" about validating the assumptions. Add residual plots in the Supplemental Information.

Low AUC for vigorous activity (0.63):

The AUC for the [3,6] class is 0.63—below the acceptable level (0.7).

This indicates that the model does not classify moderate activity well, despite its high balanced accuracy.

Suggestion: Discuss this in the Discussion as an important limitation, not just a technical finding.

4. Results and Interpretation

Positive Points:

Clear and well-presented figures.

Tables with equations and cutoff points are extremely useful for future researchers.

Confusion matrices are well interpreted.

Points for Improvement:

Claim of "high accuracy" without context:

They say the model has "balanced accuracies above 0.77," which is good, but don't compare it to previous studies.

Suggestion: Compare your cutoff points with those of GT3X+ in the literature (e.g., Santos-Lozano et al., 2013). This provides context.

Claim that the model is "simpler and more explainable":

True, but there is no direct comparison with complex models (e.g., neural networks).

- Suggestion: Rephrase: "Our linear model is simpler and more explainable than black-box models, although we did not directly test alternative models in this study."

5. Discussion

Positive Points:

Good discussion on the importance of integrating sleep and physical activity.

Well-recognized limitations: young sample, controlled environment, few activities.

Areas for Improvement:

Lack of comparison with devices other than the GT3X+:

ActTrust® is new, but could be compared with other low-cost devices (e.g., Fitbit, Garmin).

Suggestion: Briefly mention how ActTrust® positions itself in this ecosystem.

Conclusion Too Optimistic:

They say the device can be used in "clinical practice," but there is no clinical validation.

Suggestion: Soften: "has potential for clinical use, after validation in clinical populations and natural settings."

Reviewer #3: The manuscript by Batista et al. is about validating the results of the ActTrust accelerometer to the ActiGrapgh GT3X+ accelerometer and metabolic equivalents in a group of young, healthy participants. These devices were placed on the hip and wrist of participants and studied at rest and various activity levels on treadmill. The potential impact of this paper is that it would show that this newer technology has similar outcomes to current devices.

Overall, there is a lack of motivation for the experiments performed. Why not continue using the ActiGraph GT3X+ device? What benefits does the ActTrust device have? In line 27-29 it is described that the ActTrust device is specific for “inferring sleep from locomotor activity”, but then there is no verification for its use in sleep which seems like it’s main benefit over previous technology. Additionally, more description could be provided in results to give more information to the reader.

Specific Comments:

Experimental procedures (Line 94-96): Were participants wearing both devices simultaneously? Does location of the sensors on the extremities affect results (if one device is more superior or lateral to the other)?

Table 1: It says 2*Sample size in the titles. Typo?

Fig 1: It would be helpful to differentiate between female and male participants. Maybe changing the shape or color of the points would allow readers to distinguish between males and females

Fig 1B: There is a cluster of maybe 10-15 dots in both devices (though ACTT more distinct than GT3X+) but only on the wrist at 7 km/h. Is there an explanation for why this group is separate? Is it the same participants for both devices? Are they all/majority male or female? Why do we not see this grouping in the hip or metabolic equivalents graphs? Why is it only at this moderate rate (7 km/h)?

Table 2: what are the movement counts? 3, 6, 9? Intensity?

For the results, it would be more thorough and helpful to explain a bit more about each sub figure. Instead of referring to Fig 2, what is Fig 2a telling us, Fig 2b etc.

Fig 2 legend missing “d)”

Fig 2C seems like it would fit better in figure 3

Table 3. Sensitivity in the 2 higher classes are quite a bit lower. AUC is quite a bit lower in class [3,6) than the other classes. Explanations?

A substantial part of the discussion refers to use of these devices and sleep, but there was no sufficient explanation on why these experiments did not validate this device on longer time course/including sleep. It would be beneficial to either include data of the devices and sleep, or limit discussion to the exercise model tested.

The limitations in the last paragraph of the discussion (lines 246-256) are great and could be expanded. In particular, why these results may not be applicable in unhealthy or aged populations.

Reviewer #4: Overall Assessment:

This manuscript presents a well-designed validation study comparing the ActTrust® accelerometer against the widely validated ActiGraph GT3X+ for estimating energy expenditure (EE) and classifying physical activity (IA) intensity in young adults. The study is methodologically sound, clearly written, and addresses a relevant gap in the literature. The use of indirect calorimetry as a criterion measure strengthens the validity of the findings. The authors provide novel cut-points for ActTrust® devices and demonstrate their utility in classifying PA intensity with good accuracy. I recommend major revisions before acceptance, primarily to improve clarity, statistical reporting, and contextualization of results.

MAJOR COMMENTS

1. Sample Characteristics and Generalizability:

- The sample consisted of 56 young, healthy adults. While the sample size is adequate for a validation study, the lack of diversity in age, health status, and functional capacity limits the generalizability of the cut-points. The authors should explicitly acknowledge this limitation and recommend caution when applying these thresholds to other populations (e.g., older adults, clinical populations).

2. Statistical Model and Reporting:

- The use of a square-root transformation for both METs and activity counts is justified but should be more clearly explained in the Methods section. Additionally, the model equations in Table 2 are presented without confidence intervals or measures of uncertainty (e.g., standard errors). Providing these would enhance the reproducibility and utility of the models.

3. Comparison with Existing Literature:

- The authors briefly mention that their model outperforms previous studies in certain metrics (e.g., AUC for moderate activity) but do not provide a detailed comparative discussion. A more thorough comparison with existing cut-points from similar devices (e.g., ActiGraph, Fitbit) would help contextualize the novelty and practical significance of the findings.

4. Confusion Matrix Interpretation:

- The confusion matrices (Fig 3) show reduced sensitivity for vigorous and very vigorous activities. The authors should discuss potential reasons for this (e.g., device placement, biomechanical differences between walking and running) and suggest ways to improve classification in these intensity ranges.

---

MINOR COMMENTS

1. Abstract:

- The abstract should include key numerical results (e.g., correlation coefficients, accuracy metrics) to better summarize the findings.

2. Methods:

- Clarify the rationale for selecting the 4 central minutes of each condition for analysis. Was this to avoid initialization effects or ensure steady-state metabolism?

- Specify the software and packages used for statistical analysis more clearly (e.g., version numbers, functions used).

3. Results:

- The ANOVA results are mentioned but not fully reported. Consider including effect sizes (e.g., partial eta-squared) and post-hoc test results in the main text or supplementary materials.

4. Figures and Tables:

- Ensure all figures and tables are referenced in the text in numerical order.

- Fig 1 and Fig 3 are referenced but not included in the provided manuscript text. Please verify that all figures are properly captioned and accessible.

5. Data and Code Availability:

- The GitHub repository is provided, which is excellent for reproducibility. Ensure that all data and scripts are well-documented and accessible.

6. Ethics and Compliance:

- The study received ethical approval and participants provided informed consent -no ethical concerns.

- The data availability statement is clear and compliant with PLOS ONE policies.

7. Conclusion:

- This is a valuable contribution to the field of accelerometry-based physical activity monitoring. The validation of ActTrust® devices provides researchers and clinicians with a cost-effective alternative for estimating EE and classifying PA intensity. With the suggested revisions, this manuscript will be suitable for publication in PLOS ONE.

**********

-->6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.-->

Reviewer #1: No

Reviewer #2: Yes: João Carlos Alves Bueno

Reviewer #3: No

Reviewer #4: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Attachment

Submitted filename: Letter .pdf

pone.0348631.s006.pdf (79KB, pdf)
PLoS One. 2026 May 6;21(5):e0348631. doi: 10.1371/journal.pone.0348631.r002

Author response to Decision Letter 1


24 Mar 2026

Responses to Reviewer’s Comments:

Response to Reviewer #1

We thank Reviewer #1 for the positive assessment and valuable suggestions.

Comment 1.1: Gender differences in validation

Reviewer: Was the gender difference considered for validation? This could be relevant since there is an expected difference between males and females in terms of body composition and energy expenditure, which may act as confounding.

Response: This is an excellent point. While our primary model pooled males and females together (adjusting for individual differences through the within-subject design), we agree that examining sex-specific validation is important given known differences in body composition and energy expenditure patterns. We have repeated part of our analyses and found similar model performance for both male and female individuals.

Changes Made: We have included confusion matrices and AUC indices separating male and female individuals in supplementary materials.

S1 Figure: Confusion matrices for physical activity intensity classification stratified by sex. Row-normalised confusion matrices showing the proportion of observations assigned to each predicted intensity class relative to the true class derived from indirect calorimetry for (A) male and female (n=22) participants. Intensity classes are defined as light [0,3), moderate [3,6), vigorous [6,9), and very vigorous [9,∞) METs.

S2 Table: Area under the curve (AUC) for one-vs-rest physical activity intensity classification stratified by sex. AUC values were computed for each intensity class using the model prediction score against the true class derived from indirect calorimetry for male (n=34) and female (n=22) participants. Intensity classes are defined as light [0,3), moderate [3,6), vigorous [6,9), and very vigorous [9,∞) METs.

Class Male Female

[0,3) 0.988 0.974

[3,6) 0.635 0.622

[6,9) 0.765 0.808

[9,∞) 0.961 0.943

Comment 1.2: Bland-Altman plots

Reviewer: While correlation matrices were reported, additional analysis such as Bland-Altman plots could reveal further details about the systematic bias.

Response: We thank the reviewer for the suggestion. Additional analyses, included in the supplementary material, demonstrate that there is no evidence of systematic proportional bias across the measured values.

Changes Made: We have created Supplementary Figure S2 showing Bland-Altman plots for each device-placement combination, alongside estimated bias and standard deviation.

S2 Figure: Bland-Altman plots comparing model-predicted and measured metabolic equivalents (METs) for each device and placement combination. Each panel plots the difference between predicted and measured METs against their mean for each device and placement with GT3X+ hip (A), GT3X+ wrist (B), ActTrust hip (C), and ActTrust wrist (D). Red dashed lines indicate the mean bias and blue dotted lines indicate the 95% limits of agreement. There is no evidence of systematic proportional bias across the range of measured values.

Response to Reviewer #2

We thank Dr. Bueno for the comprehensive and constructive review. We address each point below.

Comment 2.1: Text redundancy

Reviewer: There is repetition of sentences between the abstract, introduction, and discussion. Each section should have a unique function.

Response: We appreciate this observation and have reviewed the manuscript accordingly.

Changes made: We removed lines 1-2 abstract and 216-218 (discussion section).

Comment 2.2: Paragraph flow and transitions

Reviewer: Some sections are too dense or poorly connected with abrupt transitions.

Response: We have improved paragraph connections and added transitional sentences throughout.

Comment 2.3: Epoch justification

Reviewer: Poor justification for choosing 1 second as the epoch, especially for light walking (3 km/h).

Response: We followed epoch size used in literature.

Changes Made: We have clarified that in the manuscript. Lines 87-91.

Comment 2.4: Synchronization details

Reviewer: Lack of detail on synchronization between devices and calorimetry.

Response: We performed synchronisation based on timestamps.

Changes made: We detailed the procedure in lines 114-117.

Comment 2.5: Fixed protocol order

Reviewer: The protocol always follows rest → 3 → 5 → 7 → 9 km/h, which may introduce order effects.

Response: We acknowledge this limitation. The progressive protocol was chosen for safety and to reflect clinical practice.

Changes Made: We edited experimental procedure (lines 110-111). Moreover, we added to Limitations: " Moreover, the fixed progressive protocol may have introduced order effects such as those linked to warm-up or fatigue. While this approach was chosen for safety, and mirrors clinical exercise testing protocols, future studies could employ counterbalanced designs or include longer rest periods to eliminate carryover effects." (lines 314-317).

Comment 2.6: MET transformation justification

Reviewer: Transformation of dependent variable (MET^1/2) is unusual and not justified.

Response: This is a common procedure to linearise data used in many areas of signal processing though not very common in actigraphy. We have clarified that in the manuscript.

Changes Made: We added to the text: "A square root transformation was applied to both METs and activity counts due to the non-linear relationship." (lines 128-129)

Comment 2.7: Regression assumptions

Reviewer: No test of regression assumptions (normality, homoscedasticity, multicollinearity).

Response: We conducted these diagnostics but did not include them. Thank you for spotting this issue.

Changes Made: We added to the text: "Model assumptions were verified through examination of residual plots and Q-Q plots for normality – included in supplementary material S3 Figure." (lines 128-131 – data analysis section).

S3 Figure: Regression diagnostics for the general model. A) Normal Q-Q plot of model residuals with points following the reference blue line, featuring minor deviations at the tails. B) Residuals versus fitted values with no systematic trend supporting the assumption of homoscedasticity.

Comment 2.8: Low AUC for moderate activity

Reviewer: AUC for [3,6] class is 0.63—below acceptable level (0.7). Should be discussed as important limitation.

Response: Our outcomes are similar to those found in Santos-Lozano A et., 2013. Although we recognise this could be higher, it reflects the heterogeneity and complexity of human movement. We have acknowledged that issue in the discussion.

Changes Made: Added to Discussion: “if we compare our results with studies which used different devices (8,19), while our model’s sensitivity and accuracy are reduced for vigorous and very vigorous activity, it still achieves accuracies above 0.90 and higher AUC scores than precious studies. Precisely, our model achieved area under the ROC curve of 0.98, 0.63, 0.78 and 0.95 for light, moderate, vigorous and very vigorous physical activity, respectively. As a comparison, the artificial neural network model developed by Santos-Lozano et al., resulted in AUC of 0.8, 0.6, 0.7 and 0.6 using a very similar protocol (8). (lines 229-236).

Comment 2.9: Context for accuracy claims

Reviewer: Claim of "high accuracy" without comparison to previous studies.

Response: We have added explicit comparisons with established literature.

Changes Made: Added to Discussion: “if we compare our results with studies which used different devices (8,19), while our model’s sensitivity and accuracy are reduced for vigorous and very vigorous activity, it still achieves accuracies above 0.90 and higher AUC scores than precious studies. Precisely, our model achieved area under the ROC curve of 0.98, 0.63, 0.78 and 0.95 for light, moderate, vigorous and very vigorous physical activity, respectively. As a comparison, the artificial neural network model developed by Santos-Lozano et al., resulted in AUC of 0.8, 0.6, 0.7 and 0.6 using a very similar protocol (8). (lines 229-236).

Comment 2.10: Model simplicity claim

Reviewer: Claim about simplicity lacks direct comparison with complex models.

Response: We clarified our text to facilitate the understanding of our claims. We have added direct comparison with the literature.

Changes Made: “The cut-points derived in the present study for the GT3X+ placed on the hip were lower than most published GT3X reference values across all thresholds. Compared to Sasaki et al. [19] our cut-points were 58%, 21%, and 2% lower at the MET 3, 6, and 9 thresholds, respectively, and 65%, 43%, and 18% lower than Santos-Lozano et al. [8] respective cut-off points for adults. Importantly, however, there is substantial variability in cut-points between these two quite similar studies, as they differ by 16% at MET 3, 39% at MET 6 and 20% at MET 9. The overall variability suggests that sample-level characteristics are the main drivers of the observed discrepancies rather than device or methodological artefacts. Nevertheless, this is the first study to develop movement-based cut-off points for wrist and hip-worn ActTrust® devices. These cut-offs reflect moderate, intense and very intense physical activity categories. The study successfully classified physical activity intensity across the spectrum, demonstrating excellent accuracy. However, it is not possible to compare these values with other studies. For the discrimination between intensity categories based on METs, we propose thresholds to be used for the hip and wrist-worn ActTrust® devices. This allows researchers to examine how much walking people perform and distinguish it from running using these accelerometers.” (Lines 241-258).

Comment 2.11: Comparison with other devices

Reviewer: Lack of comparison with devices other than GT3X+ (e.g., Fitbit, Garmin).

Response: This is a good suggestion for a future study. We will consider that in the future.

Comment 2.12: Overly optimistic conclusion

Reviewer: Claim about "clinical practice" without clinical validation.

Response: We have appropriately softened this claim.

Changes Made: Modified Conclusion: "These results demonstrated the feasibility of using ActTrust® as a cost effective and scalable tool to estimate PA. It also has potential for potential for clinical use following validation in patient populations with diverse health conditions and functional capacities. Our results also provide hints on how to harmonise outcomes from different devices whilst estimating PA and analysing activity measures. Future work should extend our analyses to free-living conditions, comparison with commercial wearable devices other than GT3X+® and more naturalistic activity patterns." (lines 308-314).

Response to Reviewer #3

We thank Reviewer #3 for highlighting important gaps in our manuscript, particularly regarding the motivation for the study and device benefits.

Comment 3.1: Lack of motivation for experiments

Reviewer: Why not continue using ActiGraph GT3X+? What benefits does the ActTrust device have?

Response: This is a crucial point that of our study. We have now clearly articulated the advantages of ActTrust®.

Changes Made: Expanded Introduction: "The ActiGraph® GT3X+ accelerometer has the most significant number of validation studies and published thresholds and has been widely used to validate devices with the same purpose [13]. The ActTrust® (Condor Instruments, São Paulo, Brazil) is a device that has a triaxial accelerometer and is specific for inferring sleep from locomotor activity [14]. It offers several advantages over other actigraphy devices: (1) significantly lower cost, making it accessible for large-scale studies in resource-limited settings; (2) longer battery life enabling extended monitoring periods; (3) integrated light and temperature sensors for comprehensive physiological assessment. However, despite these practical advantages, the ActTrust® has not been validated against gold-standard measures of EE, which limits its use in research and clinical contexts.” (lines 25-41).

Comment 3.2: No sleep validation despite device purpose

Reviewer: The ActTrust device is described as specific for "inferring sleep from locomotor activity," but there is no verification for its use in sleep.

Response: We understand the confusion. ActTrust® has been validated for sleep assessment in previous studies [43], but never for physical activity intensity classification. Our study fills this gap.

Changes made: “Beyond estimating the intensity of physical activity, the value of validating ActTrust® lies in its ability to capture the entire 24-hour rest-activity cycle within a single device. Therefore, establishing PA cut-points for ActTrust® is a necessary first step toward leveraging its broader monitoring capabilities, which extend beyond exercise to the assessment of sleep and circadian rhythms. Actigraphy devices, such as the ActTrust accelerometers have been used in sleep and circadian research for decades [41-43]”. (lines 259-265).

Comment 3.3: Simultaneous device placement

Reviewer: Were participants wearing both devices simultaneously? Does location of sensors affect results?

Response: Yes, participants wore both devices simultaneously (added to line 105). Resulting in four possible combinations of placement and device. As expected the location of sensors does yield different measures in both device types but those seem mostly systematic changes as seen in Fig 2b.

Comment 3.4: Table 1 typo

Reviewer: Table 1: It says 2Sample size in the titles. Typo?*

Response: Thank you for catching this formatting error.

Changes Made: Corrected Table 1 header to remove "2*" and properly format the "Sample size" column heading.

Comment 3.5: Gender differentiation in Figure 1

Reviewer: It would be helpful to differentiate between female and male participants in Figure 1.

Response: We included an additional analysis in the supplementary materials S1 Figure and S2 Table – see response to Reviewer 1.

Comment 3.6: Cluster in Figure 1B at 7 km/h

Reviewer: There is a cluster of 10-15 dots at 7 km/h on wrist for both devices. Is there an explanation?

Response: The explanation of that feature is a matter for a future study. This is probably a issue reported in other studies. These have not added the dots but their error bars increase in that intensity range.

Comment 3.7: Table 2 clarity

Reviewer: What are the movement counts? 3, 6, 9? Intensity?

Response: We should have been clearer about what these columns represent.

Changes Made: Revised Table 2 caption to "Model equations and cut-off points for each device and placement combination. Movement counts correspond to thresholds (in counts) for transitions between intensity categories: 3 METs (light to moderate transition), 6 METs (moderate to vigorous), and 9 METs (vigorous to very vigorous)."

Comment 3.8: More thorough results description

Reviewer: For the results, explain more about each sub-figure instead of just referring to Fig 2.

Response: We agree that more detailed description would help readers.

Comment 3.9: Figure 2 legend missing "d)"

Reviewer: Fig 2 legend missing "d)"

Response: Figure only has 3 panels (a, b, and c)

Comment 3.10: Figure 2C placement

Reviewer: Fig 2C seems like it would fit better in figure 3.

Response: Thank you for the suggestion. We believe that Fig 2C belongs in Figure 2.

Comment 3.11: Table 3 lower sensitivity explanation

Reviewer: Sensitivity in the 2 higher classes are quite a bit lower. AUC is lower in class [3,6). Explanations?

Response: These are important patterns that deserve discussion. We decided to directly compare our results (AUC) with previous literature. Our outcomes are similar to those found in Santos-Lozano A et., 2013. (lines 229-250).

Comment 3.12: Sleep discussion without sleep data

Reviewer: A substantial part of the discussion refers to sleep, but there was no explanation why these experiments did not validate this device on sleep.

Response: We clarify that ActTrust® is already validated for sleep (previous studies), but not for physical activity. Our goal was to complete the validation picture.

Changes made: “Beyond estimating the intensity of physical activity, the value of validating ActTrust® lies in its ability to capture the entire 24-hour rest-activity cycle w

Attachment

Submitted filename: Response to reviewers.pdf

pone.0348631.s007.pdf (985.4KB, pdf)

Decision Letter 1

Julio Alejandro Henriques Castro da Costa

12 Apr 2026

-->PONE-D-25-31357R1-->-->From Movement to METs: A Validation of ActTrust® for Energy Expenditure Estimation and Physical Activity Classification in Young Adults-->-->PLOS One

Dear Dr. Leocadio-Miguel,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by May 27 2026 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:-->

  • A letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

-->

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

As the corresponding author, your ORCID iD is verified in the submission system and will appear in the published article. PLOS supports the use of ORCID, and we encourage all coauthors to register for an ORCID iD and use it as well. Please encourage your coauthors to verify their ORCID iD within the submission system before final acceptance, as unverified ORCID iDs will not appear in the published article. Only the individual author can complete the verification step; PLOS staff cannot verify ORCID iDs on behalf of authors.

We look forward to receiving your revised manuscript.

Kind regards,

Julio Alejandro Henriques Castro da Costa

Academic Editor

PLOS One

Journal Requirements:

If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise.

Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

-->Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.-->

Reviewer #3: (No Response)

**********

-->2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. -->

Reviewer #3: Yes

**********

-->3. Has the statistical analysis been performed appropriately and rigorously? -->

Reviewer #3: Yes

**********

-->4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.-->

Reviewer #3: Yes

**********

-->5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.-->

Reviewer #3: Yes

**********

-->6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)-->

Reviewer #3: Thank you to the authors for taking the time to address my comments on the original manuscript. I appreciate the work that was done to modify the manscript and figures. A minor note:

I appreciate the new supplemental figures highlighting sex differences. I still think it would be helpful to the reader to distinguish sex on the graphs in figure 1. For instance squares for men, and circles for women. It could also give insight into the clustering in 7km/h that I mentioned in another comment on my initial review.

**********

-->7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.-->

Reviewer #3: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

To ensure your figures meet our technical requirements, please review our figure guidelines: https://journals.plos.org/plosone/s/figures

You may also use PLOS’s free figure tool, NAAS, to help you prepare publication quality figures: https://journals.plos.org/plosone/s/figures#loc-tools-for-figure-preparation.

NAAS will assess whether your figures meet our technical requirements by comparing each figure against our figure specifications.

-->

PLoS One. 2026 May 6;21(5):e0348631. doi: 10.1371/journal.pone.0348631.r004

Author response to Decision Letter 2


15 Apr 2026

We thank Reviewer #3 for their continued engagement with our manuscript and for confirming that the statistical analysis, data availability, and presentation now meet the standards required for publication in PLOS ONE.

In response to the reviewer's remaining comment regarding the distinction of biological sex in Figure 1, we have updated the figure accordingly. Point shape now encodes biological sex throughout both panels of Figure 1, with squares representing male participants and circles representing female participants, as suggested. This addition is reflected in the updated figure legend, which now reads: "Point shape denotes biological sex (square: male, n = 34; circle: female, n = 22)." We hope this provides the visual clarity the reviewer had in mind and offers the insight into the clustering observed at 7 km·h−1that was noted in the initial review.

We believe this revision fully addresses the comment.

We thank the reviewer once more for their constructive and thorough engagement throughout the review process.

Kind regards,

Mario

Attachment

Submitted filename: Responses to reviewers.pdf

pone.0348631.s008.pdf (53.3KB, pdf)

Decision Letter 2

Julio Alejandro Henriques Castro da Costa

19 Apr 2026

From Movement to METs: A Validation of ActTrust® for Energy Expenditure Estimation and Physical Activity Classification in Young Adults

PONE-D-25-31357R2

Dear Dr. Leocadio-Miguel,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager® and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact billing support.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Julio Alejandro Henriques Castro da Costa

Academic Editor

PLOS One

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

-->Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.-->

Reviewer #3: All comments have been addressed

**********

-->2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. -->

Reviewer #3: Yes

**********

-->3. Has the statistical analysis been performed appropriately and rigorously? -->

Reviewer #3: Yes

**********

-->4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.-->

Reviewer #3: Yes

**********

-->5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.-->

Reviewer #3: Yes

**********

-->6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)-->

Reviewer #3: (No Response)

**********

-->7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.-->

Reviewer #3: No

**********

Acceptance letter

Julio Alejandro Henriques Castro da Costa

PONE-D-25-31357R2

PLOS One

Dear Dr. Leocadio-Miguel,

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS One. Congratulations! Your manuscript is now being handed over to our production team.

At this stage, our production department will prepare your paper for publication. This includes ensuring the following:

* All references, tables, and figures are properly cited

* All relevant supporting information is included in the manuscript submission,

* There are no issues that prevent the paper from being properly typeset

You will receive further instructions from the production team, including instructions on how to review your proof when it is ready. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few days to review your paper and let you know the next and final steps.

Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

You will receive an invoice from PLOS for your publication fee after your manuscript has reached the completed accept phase. If you receive an email requesting payment before acceptance or for any other service, this may be a phishing scheme. Learn how to identify phishing emails and protect your accounts at https://explore.plos.org/phishing.

If we can help with anything else, please email us at customercare@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Julio Alejandro Henriques Castro da Costa

Academic Editor

PLOS One

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Table. ANOVA analysis and post-hoc Tukey’s Honestly Significant Difference 316 (HSD) table.

    (CSV)

    pone.0348631.s001.csv (22.5KB, csv)
    S2 Table. Area under the curve (AUC) for one-vs-rest physical activity intensity classification stratified by sex.

    AUC values were computed for each intensity class using the model prediction score against the true class derived from indirect calorimetry for male (n = 34) and female (n = 22) participants. Intensity classes are defined as light [0,3), moderate [3,6), vigorous [6,9), and very vigorous [9,∞) METs.

    (CSV)

    pone.0348631.s002.csv (106B, csv)
    S1 Figure. Confusion matrices for physical activity intensity classification stratified by sex.

    Row-normalised confusion matrices showing the proportion of observations assigned to each predicted intensity class relative to the true class derived from indirect calorimetry for (A) male (n = 34) and (B) female (n = 22) participants. Intensity classes are defined as light [0,3), moderate [3,6), vigorous [6,9), and very vigorous [9,∞) METs.

    (PDF)

    pone.0348631.s003.pdf (5.8KB, pdf)
    S2 Figure. Bland-Altman plots comparing model-predicted and measured metabolic equivalents (METs) for each device and placement combination.

    Each panel plots the difference between predicted and measured METs against their mean for each device and placement with GT3X + hip (A), GT3X+ wrist (B), ActTrust hip (C), and ActTrust wrist (D). Red dashed lines indicate the mean bias and blue dotted lines indicate the 95% limits of agreement. There is no evidence of systematic proportional bias across the range of measured values.

    (PDF)

    pone.0348631.s004.pdf (66.1KB, pdf)
    S3 Figure. Regression diagnostics for the general model.

    A) Normal Q-Q plot of model residuals with points following the reference blue line, featuring minor deviations at the tails. B) Residuals versus fitted values with no systematic trend supporting the assumption of homoscedasticity.

    (PDF)

    pone.0348631.s005.pdf (122.5KB, pdf)
    Attachment

    Submitted filename: Letter .pdf

    pone.0348631.s006.pdf (79KB, pdf)
    Attachment

    Submitted filename: Response to reviewers.pdf

    pone.0348631.s007.pdf (985.4KB, pdf)
    Attachment

    Submitted filename: Responses to reviewers.pdf

    pone.0348631.s008.pdf (53.3KB, pdf)

    Data Availability Statement

    The data and source code used in the analyses featured in this article are available at https://github.com/circadia-bio/ACTT_validation_study.


    Articles from PLOS One are provided here courtesy of PLOS

    RESOURCES