ABSTRACT
Comprehensive two‐dimensional liquid chromatography (LC × LC) is an effective strategy to increase peak capacity and improve separation performance for complex samples, offering a valuable alternative to conventional one‐dimensional liquid chromatography (1D‐LC). However, LC × LC method development and optimization remain challenging due to the interconnection of numerous parameters within and between dimensions. Both thermodynamic factors (e.g., stationary phase, mobile phase, temperature) and kinetic factors (e.g., gradient profile, sampling rate, column geometry, flow rates) strongly influence the analytical performance. To achieve effective method optimization, two‐dimensional liquid chromatography (2D‐LC) instrumentation‐related parameters must first be characterized, as they directly influence both method development and overall performance. In this work, we propose a practical guide (protocol) for comprehensive 2D‐LC system characterization. For clarity and ease of implementation, all calculations associated with this protocol can be performed using the Characterization module of the 2D‐LC Smart Calculator software, freely accessible online. The protocol usability was validated through inter‐laboratory testing. The impact of instrumental features on LC × LC performance was then assessed by in silico evaluation of the theoretical LC × LC performance of the characterized systems. The results clearly demonstrate the benefits of the proposed protocol, emphasizing the importance of system‐specific characterization for reliable LC × LC implementation. In addition, practical recommendations are provided to help users optimize their instruments when certain parameters are not adapted.
Keywords: 2D‐LC, bidimensional liquid chromatography, characterization, LC × LC, system dispersion
1. Introduction
Comprehensive two‐dimensional liquid chromatography (LC × LC) has gained increasing attention across application fields that require the analysis of complex samples, including pharmaceutical [1, 2, 3, 4, 5], environmental [6], food [7], and biological [8] analysis. By coupling two orthogonal separation mechanisms, LC × LC provides a resolving power that exceeds conventional one‐dimensional LC [9], thus enabling the resolution of analytical problems that were previously unattainable.
However, effective LC × LC method development remains challenging due to the strong interdependence of numerous thermodynamic and kinetic parameters that govern performance across both dimensions [10]. Although LC × LC optimization is inherently complex, several guidelines have been proposed to support the selection of appropriate operating conditions [11, 12, 13, 14]. Nevertheless, a critical and often overlooked step preceding method optimization, is the thorough characterization of the LC × LC system itself. Instrument‐related factors (such as system dispersion, dwell volume and system pressure) can substantially influence the achievable resolution and may limit the efficiency of the technique if not properly understood [9, 15]. To date, a practical guideline specific for two‐dimensional liquid chromatography (2D‐LC) system characterization has not yet been established.
In this work, we present a straightforward and broadly applicable protocol for 2D‐LC system characterization. The procedure is evaluated through inter‐laboratory testing to confirm its applicability across different instrument platforms. The LC × LC performance of the characterized systems is then examined in silico to illustrate how instrumental characteristics influence the separation capabilities of a comprehensive LC × LC method. These findings highlight the need for system‐specific assessment—and instrument reorganization when necessary—for the reliable and efficient implementation of LC × LC. In addition, practical guidance is provided to support users in reorganizing their systems when key instrumental parameters are not sufficiently optimized.
2. General Aspects
This protocol enables the determination of several system‐related parameters that influence 2D‐LC method optimization, performance, and data visualization. These parameters are briefly described below, and their detailed impact on chromatographic performance is further discussed in Section 4.
2.1. Extra‐Column Variance and Extra‐Column Volume
When an analyte is injected into an LC system, it migrates through the system and column as a discrete band. The observed peak width reflects dispersive processes occurring both within the column (column dispersion) and within the system components (extra‐column dispersion) [15]. The total observed peak variance (σ 2 tot) can be expressed as:
| (1) |
where σ 2 col represents the column variance and σ 2 ext corresponds to the extra‐column variance.
σ 2 ext arises from the tubing and fittings comprised between the injector and the detector. To experimentally measure σ 2 ext, the column is replaced by a zero‐dead‐volume union, thus eliminating σ 2 col from Equation (1) and directly allowing to measure σ 2 ext.
High σ 2 ext values decrease the separation performance and this effect is even more pronounced in LC × LC. Indeed, in typical LC × LC analyses, the 2D columns are short (commonly 3–5 cm), because on‐line coupling requires high flow rates to achieve sufficiently short 2D analysis times and thus maintain an adequate sampling rate for 1D peaks [15]. Furthermore, the column internal diameters are low (typically 2.1 mm) in order to take full advantage of the possibilities offered by ultra‐high performance liquid chromatography (UHPLC). Such small internal volumes in addition to very short gradient times in 2D drastically reduce σ 2 col, which makes the contribution of σ 2 ext (Equation 1) much more significant in 2D than in 1D where, on the other hand, gradient times are much longer. As a result, all sources of extra‐column dispersion must be minimized to preserve an acceptable balance between σ 2 ext and σ 2 col.
Despite this requirement, commercial LC × LC instruments are often designed with flexibility and user convenience as priorities rather than minimization of extra‐column dispersion. Typical configurations (with separate pump, injector, thermostat, and detector modules connected by relatively long tubing) can therefore compromise performance under typical LC × LC conditions.
2.2. Dwell Volume
The dwell volume (Vd ) corresponds to the volume of the fluidic path between the point where the mobile phases are mixed in the LC pump and the column inlet [16]. To experimentally determine Vd , mobile phase B is generally spiked with a tracer compound that serves as an indicator of solvent composition over time. After replacing the column with a zero‐dead‐volume union, a gradient from predominantly A to predominantly B is applied, and the signal of the tracer is monitored. As discussed later, the Vd in 1D (1 Vd ) and 2D (2 Vd ) have a less pronounced impact on the chromatographic performance, compared to σ 2 ext. However, Vd can still affect the overall quality of LC × LC separations [16], especially in 2D as it reduces the available gradient time.
2.3. Time Offset
The time offset (t offset) corresponds to the delay between the start of the data acquisition and the actual injection valve switch. Although this parameter does not influence the practical performance of the method, it affects the visualization and alignment of chromatographic data. Because the time offset is instrument‐ and software‐dependent, its accurate determination is essential for correct data representation.
In 2D‐LC systems, two detectors are typically used; therefore, the 1D time offset is defined as 1 t offset and the 2D time offset as 2 t offset. The latter is particularly critical in full LC × LC, where modulation is typically performed using two identical sample loops. Fractions from 1D are continuously collected in these two loops, which are alternately filled and sequentially injected into 2D to ensure uninterrupted and comprehensive transfer of the 1D effluent. As a result, the 2D detector records a series of 2D chromatograms corresponding to each modulation cycle. These individual chromatograms must then be separated and recombined appropriately during data processing. As illustrated in Figure 1, an incorrect 2 t offset characterization leads to a misalignment of the 2D chromatograms, which distorts the visualization and ultimately compromises the accuracy of the contour plot.
FIGURE 1.

Effect of correct (a) and incorrect (b) determination of the second‐dimension time offset (i.e., the delay between 2D acquisition start and the actual 2D injection valve switching) on data alignment and cut reconstruction in LC × LC. The dashed vertical lines indicate the valve‐switching events.
2.4. Extra‐Column Pressure
The extra‐column pressure (P ext) corresponds to the pressure generated by the instrument itself rather than the column. P ext in the first dimension (1 P ext) is generally not critical, as flow rates are typically low. The main potential limitation is an insufficiently high system pressure to ensure stable operation at very low flow rates (5–20 µL/min). However, when using a 1D column, this situation is unlikely and, if it occurs, it can be readily mitigated by adding a restriction capillary. Consequently, 1 P ext is not systematically monitored. In contrast, the P ext in 2D (2Pext) is a critical parameter. Depending on its magnitude, the maximum achievable 2D flow rate—at a given pressure limit dictated by the column—may be constrained. Such limitations can directly impact the peak capacity and overall separation performance, as further increases in 2D flow rate may no longer be possible if 2 P ext is too high.
3. Experimental
3.1. Chemicals and Reagents
Ultrapure water obtained from a Milli‐Q purification system (Millipore, Bedford, MA, USA) is used for the preparation of all mobile phases and aqueous solutions. LC–MS grade acetonitrile (ACN), and acetone (certified analytical reagent grade) are purchased from Thermo Fisher Scientific (Reinach, Switzerland). Methyl 4‐(hydroxymethyl)benzoate (98%) and lidocaine hydrochloride monohydrate are obtained from Sigma‐Aldrich (Buchs, Switzerland).
3.2. Samples and Mobile Phases
A mobile phase consisting of 40% ACN and 60% H2O (v/v) is prepared. For a total volume of 500 mL, 200 mL of ACN are combined with 300 mL of ultrapure water.
For dwell‐volume determination, mobile phase A consists of 250 mL of ultrapure water, while mobile phase B is prepared by adding 25 µL of acetone to 250 mL of ultrapure water to obtain a 0.1% (v/v) aqueous acetone solution.
Methylparaben and lidocaine stock solutions (1 mg/mL): A mass of 1.0 mg of methylparaben is accurately weighed into a 1.0 mL vial. Around 1 mL of the solvent mixture described in Section 3.2.1.1 (40% ACN and 60% H2O (v/v)) is then added, and the solution is vortexed until complete dissolution is achieved. This results in a methylparaben stock solution at a concentration of 1 mg/mL in 40:60 (v/v) ACN:H2O.
The same procedure is applied for lidocaine, resulting in a lidocaine stock solution at a concentration of 1 mg/mL in 40:60 (v/v) ACN:H2O.
Methylparaben working solution (0.01 mg/mL): Around 10.0 µL of the methylparaben 1 mg/mL stock solution is transferred into a 1.0 mL vial. A volume of 990 µL of the same 40:60 (v/v) ACN:H2O mixture is added, and the solution is vortexed. A final methylparaben concentration of 0.01 mg/mL in 40:60 (v/v) ACN‐H2O is thereby obtained.
Lidocaine working solution (0.001 mg/mL): 1.0 µL of the lidocaine 1 mg/mL stock solution is transferred into a 1.0 mL vial. A volume of 999 µL of the 40:60 (v/v) ACN:H2O mixture is then added, and the solution is vortexed. A final lidocaine concentration of 0.001 mg/mL in 40:60 (v/v) acetonitrile–water is thereby obtained.
3.3. Instrumentation and Software
The experiments are performed on a 1290 Infinity II series 2D‐LC instrument from Agilent Technologies (Waldbronn, Germany) including a 1D‐1290 binary pump, a 2D‐1290 high‐pressure binary pump (engineered to operate at high pressure), a 1290 thermostatted autosampler with a flow‐through needle injector equipped with a 20 µL sample loop, a 1D‐Multicolumn Thermostat (MCT) column compartment, a 2D‐Thermostatted Column Compartment (TCC), a 1D‐Variable Wavelength Detector (VWD), and a 2D‐diode‐array UV absorbance detector (DAD). The interface connecting the two dimensions is a 5‐position /10‐port Active Solvent Modulation (ASM) valve connected via four 1.9 µL transfer capillaries (170 × 0.12 mm) to two 2‐position/14‐port parking deck valves, mounted with two sample loops of 40 µL each. The pressure release kit (PRK) is installed to protect the 1D‐VWD flow cell from pressure spikes arising from the 2D‐LC valve switching. The second‐dimension separation is hyphenated either with a DAD operating at 80 Hz and a response time of 0.031 s, or with an Agilent 6530 Q‐TOF mass spectrometer equipped with a Dual Jet Stream ESI source. The MS is operated in positive ion mode with an extended dynamic range and an m/z range of 100–3200 (full scan). The following source parameters are applied: Gas temperature 325°C, gas flow 10 L/min, nebulizer pressure 20 psi, sheath gas temperature 400°C, and sheath gas flow 12 L/min. The capillary voltage (V cap) was set to 3500 V, nozzle voltage to 0 V, fragmentor to 200 V, skimmer to 65 V, and Oct 1 RF V pp to 750 V. The instrument described here is referred as “Agilent (1)” in the manuscript. The other instruments used in this work and their specific modules used are described in Table S1.
Data acquisition was performed using Agilent MassHunter WorkStation Data Acquisition software. Data was processed with Agilent MassHunter Qualitative Analysis 12.0.
All calculations associated with this protocol can be done via the Characterization module of the 2D‐LC Smart Calculator software [17] developed by Heinisch et al.
The predicted peak capacities, sample dilution factors, and run numbers are obtained using the Optimization module of the 2D‐LC Smart Calculator software [17].
3.4. Instrument Configurations and Data Collection
3.4.1. Determination of 1σ2 ext, 1 V ext, and 1 T offset
A solution of methylparaben at 0.01 mg/mL is used for the determination of 1 σ 2 ext, 1 V ext, and 1 t offset. The instrument is configured to perform UV detection after the 1D column oven, and the detector outlet is directed to the waste to avoid connection to the 2D system as only 1D parameters are measured during this step. The conditions listed in Table 1 are then applied at a flow rate of 400 µL/min and repeated at flow rates of 200, 100, and 25 µL/min. A restriction capillary should preferably be used instead of the zero‐dead volume union to ensure sufficient backpressure for proper pump operation. Importantly, this restrictor must have a very low internal volume to minimize its contribution to 1 σ 2 ext, 1 V ext. Therefore, a very short capillary with a narrow internal diameter should be used, typically around 25 µm. Retention times (tr ) and peak widths at half height (w 50%) of the methylparaben peaks are measured (as illustrated in Figure 2) and recorded for each flow rate (as reported in Table S2).
TABLE 1.
Operating conditions for the 1D UV‐based measurement of 1D extra column variance, extra‐column volume and time‐offset evaluation.
| Parameter | Condition |
|---|---|
| Column | Zero‐dead‐volume union |
| Mobile phase | 40% ACN / 60% H2O (v/v) |
| Flow rate | 400 µL/min (to be varied at 200, 100, and 25 µL/min) |
| Injection volume | 0.5 µL |
| Analyte | Methylparaben, 0.01 mg/mL in 40% ACN / 60% H2O |
| Detection | UV 260 nm, acquisition rate = 80 Hz |
| Column temperature | Ambient |
| Total analysis time | 5 min |
FIGURE 2.

Retention time and w50% measurements for methylparaben at four different flow rates (25 µL/min, orange; 100 µL/min, yellow; 200 µL/min, light blue; 400 µL/min, dark blue) in the first dimension of the LC × LC setup for 1σ2 ext, 1Vext, and 1toffset determination.
3.4.2. Determination of 2σ2 ext, 2 V ext, 2 P ext, and 2 T offset
Either 0.01 mg/mL methylparaben (for UV detection in the 2D), or 0.001 mg/mL lidocaine (for MS detection in the 2D), is used. The system is configured as such that the effluent from the 1D directly enters the 2D immediately after the 1D column oven. When UV detection is used in 1D, the 1D UV detector is positioned between the column outlet and the 2D interface. The chromatographic conditions listed in Table 2 are then applied with a 2D flow rate of 1000 µL/min. This procedure is repeated at 2D flow rates of 1500, 2000, and 2500 µL/min. For each flow rate, the 2D extra‐column pressure (2 P ext), 2D tr and w 50% are measured for the most suitable peak (i.e., no peak saturation and correct signal to noise ratio). These measurements can be performed using either UV (as illustrated in Figure 4) or MS detection in 2D, depending on the configuration of the analytical setup.
TABLE 2.
Operating conditions for the 2D measurement of 2σ2 ext, 2Vext, 2Pext, 2time‐offset evaluation.
| Parameter | Condition/Setting |
|---|---|
| First dimension | |
| Column | Zero‐dead‐volume union or short tubing that generates between 10 and 100 bar at 200 µL/min (e.g.: tubing with 25 cm length and 0.063 mm diameter) |
| Mobile phase | 40% ACN / 60% H2O (v/v) |
| Flow rate | 25 µL/min |
| Injection volume | 0.5 µL of prepared standard solution |
| Analyte | Methylparaben (UV) or lidocaine (MS), in 40% ACN / 60% H2O |
| Detection | UV 80 Hz acquisition rate or no detection |
| Temperature | Ambient |
| Interface/Modulation | |
| Interface mode | Comprehensive |
| Sampling time | 0.3 min |
| Second dimension | |
| Column | Zero‐dead‐volume union |
| Mobile phase | 40% ACN / 60% H2O (v/v) |
| Flow rate | 1000 µL/min (to be varied at 1500, 2000, and 2500 µL/min) |
| Detection (UV or MS) |
UV: 260 nm, 80 Hz acquisition rate MS: >10 spectra/s, EIC = 235.1805 m/z (lidocaine) |
| Temperature | Ambient |
FIGURE 4.

Retention time and w50% measurements for methylparaben using UV detection at four different flow rates (1000 µL/min, brown; 1500 µL/min, yellow; 2000 µL/min, light green; 2500 µL/min, dark green) in the second dimension of the LC × LC setup for 2σ2 ext, 2Vext, and 2toffset determination. The black trace corresponds to the peak in the first dimension.
3.4.3. Determination of 1 Vd and 2 Vd
For 1 Vd determination, a zero‐dead‐volume union is used in 1D. A UV detector is placed immediately after the union, and the detector outlet is directed to the waste to avoid any connection with the 2D system during this step. Mobile phases consisting of (A) 100% H2O and (B) 99.9% H2O / 0.1% acetone (v/v) are employed. The gradient shown in Figure S1 is applied at a flow rate of 200 µL/min (a higher flow rate can be used if the final plateau in the chromatogram is not clearly visible). Detection is performed at 254 nm.
For 2 Vd determination, the 1D pumping system must remain operational; therefore, the 1D column is replaced with a zero‐dead‐volume union, and a low flow rate (0.1 mL/min of 40% ACN / 60% H2O) is employed in 1D. The instrument is configured for 2D‐LC operation with either 2D UV detection at 254 nm or 2D MS detection. The 2D column is replaced with a zero‐dead‐volume union. In the second dimension, a mobile phase consisting of (A) 100% H2O and (B) 99.9% H2O / 0.1% acetone is used; when MS detection is implemented, acetone needs to be replaced with an ionizable modifier (e.g. lidocaine). The gradient shown in Figure S1 is applied in the second dimension at a flow rate of 200 µL/min. The second dimension is flushed with the gradient without any injection. It is essential to ensure that the 2D gradient starts immediately and independently of any injection event. Depending on the 2D‐LC systems and software, this is achieved through different strategies, as listed in Table S4.
4. Results
4.1. Experimental Determination of the Relevant 2D‐LC System Parameters
4.1.1. Determination of First Dimension Parameters
To determine σ 2 ext, the peak width at half height (w 50%) measurement was deliberately used instead of statistical moments. Indeed, implementing moment calculations would be less user‐friendly as it introduces additional manual data handling steps, thereby increasing the risk of operator‐dependent errors. As shown in Table S2, the w50% values of methylparaben peaks at different 1D flow rates (F) allowed determination of 1 σ 2 ext. Indeed, under the assumption of a gaussian peak, the variance can be determined at half the peak height by:
| (2) |
As illustrated in Figure 3a , 1 σ 2 ext values measured at different flow rates were used to establish a linear regression between 1 σ 2 ext and flow rate, allowing 1 σ 2 ext to be extrapolated to any desired first‐dimension flow rate relevant to a given application.
FIGURE 3.

Plots of (a) 1σ2 e x t as a function of flow rate and (b) retention time as a function of 1/flow for methylparaben, with corresponding linear regressions. These plots were constructed using the data generated in Figure 2.
The retention times (tr ) of the methylparaben peaks recorded at the various flow rates as shown in Table S2, are used to obtain the slope of the linear relationship between tr and the reciprocal F (1/F). As illustrated in Figure 3b, this slope directly provides the 1 V ext, while the intercept of the plot corresponds to the 1 t offset.
4.1.2. Determination of Second Dimension Parameters
The w 50% values of the methylparaben (for UV analysis) or lidocaine (for MS analysis) peak measured in 2D at different 2D flow rates allow determination of the evolution of 2 σ 2 ext as a function of flow rate (according to Equation 2) as shown in Table S3. The corresponding 2 σ 2 ext can then be established and extrapolated to any desired 2D flow rate as illustrated by the grey trace in Figure 5a,c.
FIGURE 5.

Plots of 2σ2 e x t (grey) and 2Pext (blue) as a function of mobile phase flow rate for future extrapolation of parameters at any flow rate (a, c) and retention time as a function of 1/flow for 2toffset and 2Vext calculation (b, d) for methylparaben with UV detection (a, b) and lidocaine with MS detection (c, d). Linear regression fits are shown in all plots.
2 P ext can also be monitored for all flow rates to plot 2 P ext as a function of the 2D flow (blue trace in Figure 5a,c) to deduce the pressure that will be provided by the system when working at high flow rate for the LC × LC experiments.
The tr recorded at the different flow rates are first corrected using a modulo function based on the sampling time (0.3 min in this case; Table 2). Because the retention times measured for the compound detected in the second dimension at different flow rates do not necessarily correspond to the peak eluting during the first modulation cycle, a correction is required to reference all measurements to a given modulation event. Accordingly, each tr value is divided by the sampling time, and the remainder values of this division are then used to construct a linear relationship with the reciprocal of the flow rate (1/F), as shown in Figure 5b,d. The slope of this plot directly yields 2 V ext. The intercept of the plot corresponds to the 2 t offset.
4.1.3. Determination of Dwell Volumes
1 Vd is determined based on the 1D experiments described in Section 3.5.3 and 2 Vd is determined based on the 2D experiments described in Section 3.5.4. The t 50 values in 1D and in 2D are determined from the initial and final plateau of the gradient performed in 1D and in 2D (to calculate 1 Vd and 2 Vd respectively), according to the procedure illustrated in Figure S1. Based on the measured value of t 50, the dwell time (td ) can be obtained using the following equation:
| (3) |
The corresponding dwell volume (Vd ) is then calculated from the flow rate using:
| (4) |
4.2. Validation of the Automated Protocol Across Multiple Laboratories
To improve ease of use and ensure high‐quality guidance, all protocol steps, experimental procedures, and data processing described above can be supported by the Characterization module of the 2D‐LC Smart Calculator software [17].
The usability of this protocol was evaluated through inter‐laboratory testing. The study included four Agilent 1290 Infinity 2D‐LC systems and one Shimadzu Nexera‐e system. Instrument modules and configurations are summarized in Table S1, and tubing dimensions are provided in Table 3. The 2D‐LC system‐characterization procedure—supported by the Characterization module of the 2D‐LC Smart Calculator software [17]—was successfully applied by all participating laboratories, regardless of the instrument model or configuration, demonstrating that the protocol is broadly applicable. The measured 2D‐LC systems parameters are reported in Table 4 for each instrument.
TABLE 3.
Main characteristics and tubing configurations of the different 2D‐LC systems used in this work (four different Agilent 1290 Infinity II 2D‐LC Solution and one Shimadzu Nexera‐e).
| System | 1D mixer volume | 2D mixer volume | Interface | 1D tube column to UV | 2D tube column to detector | 1D UV flow cell volume | 2D UV flow cell volume |
|---|---|---|---|---|---|---|---|
| Agilent (1) | 35 µL | 35 µL |
5‐position /10‐port ASM valve (G4243A) connected via four 1.9 µL transfer capillaries (170 × 0.12 mm) to two 2‐position/14‐port parking deck valves (MHC), mounted with two sample loops of 40 µL each (420 × 0.35 mm) |
28 cm × 100 µm | 34 cm × 100 µm | 2 µL | 1 µL |
| Agilent (2) | 35 µL | 35 µL |
2‐position /4‐port duo valve (5067‐4244) with two sample loops of 40 µL each (815 × 0.25 mm) |
15 cm × 127 µm | 15 cm × 100 µm (+ divert valve when MS was used) | 1 µL | 1 µL |
| Agilent (3) | 35 µL | 35 µL |
2‐position /4‐port duo valve (5067‐4244) with two sample loops of 20 µL each (637 × 0.2 mm) |
25 cm × 100 µm | 32 cm × 100 µm | 13 µL | 1 µL |
| Agilent (4) | No external mixer | 35 µL |
5‐position/10 port ASM valve (G4243A) with two sample loops of 60 µL each (831 × 0.35 mm) |
20 cm × 127 µm | 15 cm × 127 µm | 2 µL | 1 µL |
| Shimadzu (1) | 20 µL | 20 µL |
Two high speed/high pressure 2‐position, 6‐port valves (FCV‐0206H3) with microelectric actuator, with two sample loops of 10 µL each (210 × 0.25 mm) |
55 cm × 100 µm | 55 cm × 100 µm | 8 µL | 8 µL |
TABLE 4.
Characterization of various 2D‐LC instruments (four different Agilent 1290 Infinity II 2D‐LC Solution and one Shimadzu Nexera‐e). All the values were experimentally obtained using the protocol previously described to characterize 2D‐LC instruments. 1D values for 0.2 mL/min and 2D values for 2.6 mL/min.
| Time offset | Dwell volume | Extra column variance | Extra column volume | Extra column pressure | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| System | 1toffset (s) (UV) | 2toffset (s) (UV) | 2toffset (s) (MS) | 1VD (µL) | 2VD (µL) | 1σ2 ext [µL2] (UV) | 2σ2 ext [µL2] (UV) | 2σ2 ext [µL2] (MS) | 1Vext [µL] (UV) | 2Vext [µL] (UV) | 2Vext [µL] (MS) | 2Pext (bar) (UV) | 2Pext (bar) (MS) |
| Agilent (1) | 0.18 | 1.11 | 1.18 | 136 | 86 | 6 | 5 | 29 | 19 | 8 | 6 | 212 | 244 |
| Agilent (2) | 0.30 | 2.30 | 2.30 | 180 | 90 | 5 | 7 | 134 | 15 | 14 | 17 | 122 | 112 |
| Agilent (3) | 0.54 | 0.31 | — | 808 | 122 | 15 | 12 | — | 29 | 17 | — | 138 | — |
| Agilent (4) | −2.34 | 0.64 | — | 838 | 92 | 6 | 33 | — | 25 | 15 | — | 227 | — |
| Shimadzu (1) | −0.84 | 1.06 | 1.16 | 212 | 94 | 143 | 53 | 406 | 63 | 9 | 49 | 94 | 45 |
To illustrate the impact of those parameters on global performance, the Optimization module of the 2D‐LC Smart Calculator [17] software was used. First, the values obtained with the system Agilent 1 and UV detection (see values in Table 4) were used in the Calculator to simultaneously optimize three parameters: (i) Peak capacity, which reflects the separation power and should be maximized; (ii) dilution factor, which represents sample dilution induced by the 2D process and should be minimized to ensure acceptable sensitivity; and (iii) the number of second‐dimension runs, which is directly related to total analysis time and is ideally reduced. This in silico optimization enables the determination of optimal values for 1D‐column length, sampling time and 1D‐flow‐rate. Those in silico optimized conditions for this 2D‐LC analysis are listed in Figure S2 and summarized in Table S5. Once the conditions were optimized for the Agilent (1) instrument, the parameters of the 2D‐LC system were studied independently of each other with values ranging from the lowest to the highest, in accordance with the parameter values obtained with the different instruments (Table 4).
The influence of these parameters on the instrument performance is illustrated in Figure 6. As the run number varies only with the total analysis time and therefore changes in 1 Vd , and this to a very limited extent, the present result analysis focuses solely on the peak capacity and the dilution factor, which are used as performance indicators (Figure 6).
FIGURE 6.

Impact of different parameters 2Pext (grey trace), 1Vd (dark blue trace), 2Vd (light blue trace), 1σ2 e x t (yellow trace), 2σ2 e x t with UV detection (orange trace) and 2σ2 e x t with MS detection (brown trace) on (a) the effective peak capacity and (b) dilution factor in an LC × LC setup under representative operating conditions (Table S5). All the parameters are normalized for graphical representation.
4.3. Interlaboratory Comparison and Discussion
4.3.1. Time Offset
As shown in Table 4, time offset values need to be characterized as they are highly instrument‐ and software‐dependent. Depending on the situation, the time offset can be positive (the sample is injected after the acquisition starts) or negative (the sample is injected before the acquisition starts). Although these values are important for data visualization and alignment (as illustrated in Figure 1), they do not affect chromatographic performance.
4.3.2. System Pressure
As shown in Table 4, 2 P ext values are instrument‐dependent and can vary between 45 to 244 bar (for a fixed 2D flow rate of 2600 µL /min), depending on the instrument and especially the tubing geometry (length and internal diameter). To assess the specific impact of 2 P ext, a 2D upper pressure limit of 1100 bar was considered, and the highest flow rate that can be used at this pressure limit was considered to calculate the corresponding peak capacity, and dilution factor, thereby highlighting the effect of 2 P ext on performance (Figure 6, grey trace). Indeed, as 2 P ext directly limits the maximum flow rate that can be applied, the allowable 2D flow rate varies between instruments. Because peak capacity depends on flow rate, high 2 P ext values restrict the attainable flow rate and consequently reduce peak capacity. This is illustrated in Figure 6a (grey trace), showing a decrease in peak capacity when instruments with high 2 P ext are used, whereas impact on dilution factor is negligible (Figure 6b, grey trace). As 2D separations typically operate at the highest possible flow rates, it is essential to anticipate the pressure generated at any given flow rate to avoid exceeding the system's pressure capability. The tube diameter is the key factor for external pressure, but it also has a significant impact on solute dispersion. It is therefore essential to find the right balance between external pressure and extra‐column variance, as discussed later.
4.3.3. Dwell Volumes
The 1 Vd varies substantially between instruments from 136 to 838 µL. With modern binary 1D pumps such as the Agilent 1290 or the Shimadzu Nexera LC‐40B X3, 1 Vd values are typically around 200 µL, or even lower when equipped with the smallest available mixers. Larger 1 Vd may be encountered with quaternary pump designs, such as the Agilent 1260 series. However, even if the dwell volume can impact retention and selectivity [16], the main impact of a high 1 Vd is a substantial increase in analysis time. Indeed, the main practical consequence of a large 1 Vd is due to the low 1D flow rates typically used in LC × LC: A large 1 Vd increases the initial delay of the 1D separation, resulting in unnecessary loss of analysis time and increase in run number. However, the impact of 1 Vd on overall 2D‐LC performance remains limited. As illustrated in Figure 6a,b (dark blue trace), neither the effective peak capacity nor the dilution factor is significantly affected by changes in 1 Vd .
In contrast, a large 2 Vd has a significant impact on effective peak capacity. To reduce 2 Vd , a strategy is to select the smallest available mixer (typically 20 or 35 µL depending on the provider) and reduce the internal diameter of the tubing between the pumps and the injector. However, as shown in Table 4, all instruments exhibit a 2 Vd of approximately 90 µL (excluding the injection loop). The main difference therefore arises from the loop volume, which must be adapted to the injected volume. Indeed, when 2 Vd is increased in silico to mimic loop volumes of 50, 100, and 150 µL, a non‐negligible loss in peak capacity is observed, reaching up to −33% (Figure 6a, light blue trace). This can be explained by the systematic delay induced by the increase of 2 Vd at the beginning of each 2D gradient. This reduces the effective gradient time available within each modulation cycle and therefore the overall peak capacity while having a very slight impact on the dilution factor (Figure 6b, light blue trace).
4.3.4. Extra Column Variances
Typical Agilent 1200 systems exhibit extra‐column variances in the range of 10–20 µL2 [15]. As shown in Table 4, this trend is confirmed for most of the Agilent instruments with UV detection. Higher values are explained by tubing with a 127 µm internal diameter instead of 100 µm tubing downstream of the 2D‐LC column oven. Furthermore, similar σ 2 ext values were obtained in D1 at 200 µL/min and in D2 at 2600 µL/min. Although this may appear unexpected, it can be explained by the highly optimized configuration in D2, which features very low volumes compared with the configuration used in D1. In addition, both 1 σ 2 ext and 2 σ 2 ext are consistently higher for the Shimadzu instrument compared with the Agilent systems. This difference may arise from the intrinsically larger internal volumes of the Shimadzu platform, from the longer lengths of tubing required to interconnect the modules, or from the higher‐volume UV flow cell used for this system.
The influence of V ext is typically negligible for large‐volume columns—where σ 2 col dominates σ 2 tot—or under low flow rate conditions [15]. Consequently, 1D, in a 2D‐LC system, is usually less affected by extra‐column effects than 2D (as shown in Figure 6, yellow trace). For this reason, low importance is usually given to reduce 1 σ 2 ext (as illustrated in Table 4 with 1 σ 2 ext values ranging from 5 to 143 µL2). However, as illustrated in Figure 6b (yellow trace), an increase in σ 2ext leads to a higher dilution factor, since broader peaks in the first‐dimension result in greater dilution than narrower ones.
Nevertheless, the impact of 2 σ 2 ext is more important; 2D employs small‐volume columns with steep gradients, resulting in peaks that are intrinsically narrow and highly sensitive to dispersion [15, 18]. As illustrated in Table 4, the 2 σ 2 ext is highly dependent on the instrumentation. However, as shown in Figure 6 (orange trace), increasing 2 σ 2 ext from 5 to 50 µL2 reduces the peak capacity and increases the dilution factor by a factor greater than two, making 2 σ 2 ext the most critical factor in LC × LC.
To reduce 1 σ 2 ext and 2 σ 2 ext, several technical solutions can be implemented. The easiest strategy is to use tubing with the shortest possible length and, above all, the smallest internal diameter—particularly downstream of the second‐dimension column—since extra‐column dispersion scales with the fourth power of the internal diameter [18]. This approach is strongly recommended, even after installation of the 2D‐LC system by the instrument provider, as factory installations are not always optimal (e.g., module placement or tubing selection). This approach may, however, involve compromises, as excessively narrow tubing can lead to an undesired increase in pressure. The use of tubing and fittings from the same manufacturer as the instrument is also advised to avoid mismatched connections that introduce additional dead volume.
The contribution of the detector to 1 σ 2 ext and 2 σ 2 ext can be lowered by working with a detector flow cell with lower volumes (1 µL or less), although this may potentially impact the background noise and limits of detection. Furthermore, MS detection should be used with high cautions, as the MS interface can introduce substantial additional volumes (connection tubing, nebulizer probe tubing) that significantly increases 2 σ 2 ext [19, 20, 21], as illustrated in Figure 6. If MS‐compatible mobile phases are used in both dimensions, the use of a divert valve prior to the 2D detector should be avoided, as it greatly increases 2 V ext (c.f. Table 4).
Coiled loops and coiled tubing are preferred over straight configurations for the sample loops, as straight loops represent the worst‐case scenario in terms of band broadening during loop filling and emptying [22]. Furthermore, a “First‐In–Last‐Out” (countercurrent) filling loop strategy is recommended for second‐dimension injections, since greater dispersion occurs when the loop is filled in the same direction in which it is emptied [23].
The importance of system characterization and instrumental optimization prior to 2D‐LC analysis is further illustrated in Figure 7. Indeed, one might assume that comparable performance could still be achieved across instruments, even if they are not optimally configured in practice, provided that the method conditions are specifically optimized for each system. To evaluate this hypothesis, an instrument‐specific in silico optimization was performed using the Optimization module of the 2D‐LC Calculator. Starting from the initial conditions reported in Table S5, the highest flow rates compatible with a maximum overall second‐dimension pressure of 1100 bar were applied for each instrument. Instrument‐specific experimentally determined values of P ext, σ 2 ext, and Vd (with UV detection) were incorporated into the simulations. As illustrated in Figure 7, the overall performance of the instrument is clearly governed by 2 σ 2 ext. After simulation of the maximum effective peak capacities and the corresponding dilution factors obtained via Pareto optimization for each 2D‐LC instrument (Figure 7b), the conclusion is straightforward: Instruments exhibiting high 2 σ 2 ext values (reflected by larger peak widths at 50% height in Figure 7a) show a markedly reduced maximum peak capacity and a significantly increased dilution factor compared to systems with low 2 σ 2 ext, even after instrument‐specific optimization. These results highlight the importance of this protocol, as thorough instrument characterization is essential to ensure practical optimization prior to method development. Even under optimal operating conditions, LC × LC performance remains strongly constrained by the instrumental setup itself.
FIGURE 7.

(a) Superimposed 2D UV chromatograms used for the determination of 2σ2 ext for the five instruments included in the study. The methylparaben peak intensity was normalized across all datasets, and retention times were corrected using a modulo function to align all signals within the same modulation cycle. (b) Optimized effective peak capacities (grey) and dilution factor (black) obtained after instrument‐specific in silico optimization for each specific instrument.
5. Conclusions
This work provides a practical and broadly applicable protocol for 2D‐LC system characterization, which can be supported either by the Characterization module of the Excel tool 2D‐LC Smart Calculator, which can be downloaded for free [17], or carried out independently by following the procedure detailed in this article.
Through inter‐laboratory evaluation across multiple platforms and configurations, the usability of the described procedure was confirmed, and the collected data demonstrated that instrument‐related parameters (particularly extra‐column variance, dwell volumes, and second‐dimension pressure limitations) vary substantially between systems and can markedly influence achievable dilution factors, peak capacities, run number, and overall separation performance in LC × LC.
Importantly, this characterization is intended to be performed once for a given instrument configuration and does not need to be repeated for each new LC × LC method, unless modifications to the system configuration are made. Moreover, the protocol relies on inexpensive and readily available compounds (e.g., lidocaine, methylparaben, and acetone) and can be completed within approximately two days for a full system characterization. By enabling users to quantify these parameters and anticipate their impact on method performance, the protocol provides a solid foundation for informed method development and instrument optimization. Adoption of this characterization procedure and application of the practical recommendations to optimize instruments will facilitate the development of comprehensive LC × LC methods, thereby supporting a wider and more effective use of 2D‐LC in complex analytical applications.
Author Contributions
Megane K. Aebischer: writing – original draft, investigation, visualization, conceptualization. Marie Pardon: writing – review and editing, investigation. Clémence Gadot: writing – review and editing, investigation. Katia Arena: writing – review and editing, investigation. Niklas Carstensen: writing – review and editing, investigation. Michael Laemmerhofer: writing – review and editing. Paola Dugo: writing – review and editing. Francesco Cacciola: writing – review and editing. Luigi Mondello: writing – review and editing. Morgan Sarrut: writing – review and editing. Deirdre Cabooter: writing – review and editing. Sabine Heinisch: writing – original draft, conceptualization. Davy Guillarme: writing – original draft, supervision, conceptualization.
Conflicts of Interest
The authors declare no conflicts of interest.
Supporting information
Supporting File: jssc70434‐sup‐0001‐SuppMat.docx.
Acknowledgments
The authors gratefully acknowledge the continuous support of Shimadzu Corporation to the University of Messina.
Open access publishing facilitated by Universite de Geneve, as part of the Wiley ‐ Universite de Geneve agreement via the Consortium Of Swiss Academic Libraries.
Aebischer M. K., Pardon M., Gadot C., et al. “A Protocol for Characterizing Comprehensive Two‐Dimensional Liquid Chromatography Systems.” Journal of Separation Science 49, no. 5 (2026): e70434. 10.1002/jssc.70434
This paper is included in the Special Collection “Protocols Papers” edited by Sebastiaan Eeltink and Michael Lämmerhofer.
Data Availability Statement
The free and open‐access 2D‐LC Smart Calculator (Excel‐based tool) is available at: https://farma‐unites.unige.ch/en/guillarme‐lab/tools/2d‐lc‐smart‐calculator
References
- 1. Li Q., Lynen F., Wang J., Li H., Xu G., and Sandra P., “Comprehensive Hydrophilic Interaction and Ion‐Pair Reversed‐Phase Liquid Chromatography for Analysis of Di‐ to Deca‐Oligonucleotides,” Journal of Chromatography A 1255 (2012): 237–243, 10.1016/j.chroma.2011.11.062. [DOI] [PubMed] [Google Scholar]
- 2. Aebischer M. K., Abdurahman Y., Heinisch S., and Guillarme D., “Investigating the Total Breakthrough Behavior of Oligonucleotides in Ion‐Pairing Chromatography and Comprehensive HILIC × IP‐RPLC,” Journal of Chromatography A 1765 (2026): 466486, 10.1016/j.chroma.2025.466486. [DOI] [PubMed] [Google Scholar]
- 3. Stoll D. R., Talus E. S., Harmes D. C., and Zhang K., “Evaluation of Detection Sensitivity in Comprehensive Two‐Dimensional Liquid Chromatography Separations of an Active Pharmaceutical Ingredient and Its Degradants,” Analytical and Bioanalytical Chemistry 407, no. 1 (2015): 265–277, 10.1007/s00216-014-8036-9. [DOI] [PubMed] [Google Scholar]
- 4. Sandra K. and Sandra P., “The Opportunities of 2D‐LC in The Analysis of Monoclonal Antibodies,” Bioanalysis 7, no. 22 (2015): 2843–2847, 10.4155/bio.15.210. [DOI] [PubMed] [Google Scholar]
- 5. Sandra K., Piens K., Bracke D., Sandra P., and Vervecken W., “Detailed Glycosylation Analysis of Therapeutic Enzymes Utilizing Comprehensive 2D‐LC–MS,” LCGC International (2025): 26–30, 10.56530/lcgc.int.nf5273s4. [DOI] [Google Scholar]
- 6. Chapel S., Pardon M., and Cabooter D., “Systematic Approach to Online Comprehensive 2D‐LC Method Development for Organic Micropollutant Profiling in Wastewater,” Journal of Chromatography A 1749 (2025): 465861, 10.1016/j.chroma.2025.465861. [DOI] [PubMed] [Google Scholar]
- 7. Caño‐Carrillo I., Gilbert‐López B., Montero L., Martínez‐Piernas A. B., García‐Reyes J. F., and Molina‐Díaz A., “Comprehensive and Heart‐Cutting Multidimensional Liquid Chromatography–Mass Spectrometry and Its Applications in Food Analysis,” Mass Spectrometry Reviews 43, no. 5 (2024): 936–976. [DOI] [PubMed] [Google Scholar]
- 8. Hand R. A., Bassindale T., Turner N., and Morgan G., “Application of Comprehensive 2D Chromatography in the Anti‐Doping Field: Sample Identification and Quantification,” Journal of Chromatography B 1178 (2021): 122584, 10.1016/j.jchromb.2021.122584. [DOI] [PubMed] [Google Scholar]
- 9. Stoll D. R. and Carr P. W., Multi‐Dimensional Liquid Chromatography: Principles, Practice, and Applications [Internet], 1st ed. (CRC Press, 2022), https://www.taylorfrancis.com/books/9781003090557. [Google Scholar]
- 10. Sarrut M., D'Attoma A., and Heinisch S., “Optimization of Conditions in On‐Line Comprehensive Two‐Dimensional Reversed Phase Liquid Chromatography. Experimental Comparison With One‐Dimensional Reversed Phase Liquid Chromatography for the Separation of Peptides,” Journal of Chromatography A 1421 (2015): 48–59, 10.1016/j.chroma.2015.08.052. [DOI] [PubMed] [Google Scholar]
- 11. Aebischer M. K., Chapel S., Guillarme D., and Heinisch S., “Theoretical and Practical Guidelines for Solvent Dilution Between the Two Dimensions in Online Comprehensive Two‐Dimensional Liquid Chromatography,” Journal of Chromatography A 1718 (2024): 464725, 10.1016/j.chroma.2024.464725. [DOI] [PubMed] [Google Scholar]
- 12. Pirok B. W. J., Gargano A. F. G., and Schoenmakers P. J., “Optimizing Separations in Online Comprehensive Two‐Dimensional Liquid Chromatography,” Journal of Separation Science 41, no. 1 (2018): 68–98, 10.1002/jssc.201700863. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Tirapelle M., Duanmu F., Chia D. N., Besenhard M. O., Mazzei L., and Sorensen E., “Method Development of Comprehensive Two‐Dimensional Liquid Chromatography: A New Metric for Preliminary Column Selection,” Journal of Chromatography A 1741 (2025): 465593, 10.1016/j.chroma.2024.465593. [DOI] [PubMed] [Google Scholar]
- 14. Schoenmakers P. J., Vivó‐Truyols G., and Decrop W. M. C., “A Protocol for Designing Comprehensive Two‐Dimensional Liquid Chromatography Separation Systems,” Journal of Chromatography A 1120, no. 1–2 (2006): 282–290, 10.1016/j.chroma.2005.11.039. [DOI] [PubMed] [Google Scholar]
- 15. Desmet G. and Broeckhoven K., “Extra‐Column Band Broadening Effects in Contemporary Liquid Chromatography: Causes and Solutions,” TrAC Trends in Analytical Chemistry 119 (2019): 115619, 10.1016/j.trac.2019.115619. [DOI] [Google Scholar]
- 16. Stoll D. R., “The Gradient Delay Volume, Part I: Theory,” LCGC International 1, no. 1 (2024): 6–10. [Google Scholar]
- 17. Guillarme D., Aebischer M., Buff M., and Heinisch S.. “2D‐LC Smart Calculator.” ISPSO, published 2025, accessed April 30, 2026, https://farma‐unites.unige.ch/en/guillarme‐lab/tools/2d‐lc‐smart‐calculator. [Google Scholar]
- 18. Vanderlinden K., Broeckhoven K., Vanderheyden Y., and Desmet G., “Effect of Pre‐ and Post‐Column Band Broadening on the Performance of High‐Speed Chromatography Columns Under Isocratic and Gradient Conditions,” Journal of Chromatography A 1442 (2016): 73–82, 10.1016/j.chroma.2016.03.016. [DOI] [PubMed] [Google Scholar]
- 19. Grata E., Guillarme D., Glauser G., et al., “Metabolite Profiling of Plant Extracts by Ultra‐High‐Pressure Liquid Chromatography at Elevated Temperature Coupled to Time‐of‐Flight Mass Spectrometry,” Journal of Chromatography A 1216, no. 30 (2009): 5660–5668, 10.1016/j.chroma.2009.05.069. [DOI] [PubMed] [Google Scholar]
- 20. Rodriguez‐Aller M., Gurny R., Veuthey J. L., and Guillarme D., “Coupling Ultra High‐Pressure Liquid Chromatography With Mass Spectrometry: Constraints and Possible Applications,” Journal of Chromatography A 1292 (2013): 2–18, 10.1016/j.chroma.2012.09.061. [DOI] [PubMed] [Google Scholar]
- 21. Spaggiari D., Fekete S., Eugster P. J., et al., “Contribution of Various Types of Liquid Chromatography–Mass Spectrometry Instruments to Band Broadening in Fast Analysis,” Journal of Chromatography A 1310 (2013): 45–55, 10.1016/j.chroma.2013.08.001. [DOI] [PubMed] [Google Scholar]
- 22. Moussa A., Lauer T., Stoll D., Desmet G., and Broeckhoven K., “Modelling of Analyte Profiles and Band Broadening Generated by Interface Loops Used in Multi‐Dimensional Liquid Chromatography,” Journal of Chromatography A 1659 (2021): 462578, 10.1016/j.chroma.2021.462578. [DOI] [PubMed] [Google Scholar]
- 23. Van Den Hurk R. S., Bos T. S., Stoll D. R., and Pirok B. W. J., “Significance of Radial Dispersion to Effective Modulation in Two‐Dimensional Liquid Chromatography,” Journal of Chromatography A 1763 (2025): 466456, 10.1016/j.chroma.2025.466456. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supporting File: jssc70434‐sup‐0001‐SuppMat.docx.
Data Availability Statement
The free and open‐access 2D‐LC Smart Calculator (Excel‐based tool) is available at: https://farma‐unites.unige.ch/en/guillarme‐lab/tools/2d‐lc‐smart‐calculator