Abstract
This study investigated the effect of surface polishing on roughness, mechanical performance, and biological properties of a liquid crystal display (LCD)‑printed denture base polymer. Disc‑shaped samples of three-dimensional (3D) printing polymer were fabricated and allocated to polished (P) or not polished (NP) groups for comparative analysis. Polishing was performed manually in a standardized manner using sequential silicon carbide sandpapers, while the NP samples were sanded to reproduce the texture generated by the printing lines. Surface roughness was measured by contact and non‑contact profilometry, and topography analyzed by scanning electron microscopy (SEM). Mechanical properties were assessed by compressive stress testing and wettability by water contact angle. Biological assays included C. albicans biofilm proliferation (colony-forming units, CFU), metabolism (Alamar blue assay) and coverage (Live/Dead). Cytotoxicity tests with L‑929 fibroblasts under direct and indirect contact were also performed. Protein adsorption was evaluated using Alexa Fluor 488‑labeled Bovine Serum Albumin (BSA). Data were analyzed with Student’s t‑test (α = 0.05). Polishing significantly reduced roughness to the clinically acceptable threshold, with SEM and laser microscopy confirming reduction of irregularities. Mechanical testing showed similar Young’s modulus values between groups, while polished samples exhibited higher contact angles. Polished surfaces reduced C. albicans biofilm metabolism and coverage on the samples, although CFU counts did not differ significantly. Cytotoxicity tests demonstrated cell metabolism comparable to the control group (100% viable cells) over time. Protein adsorption was equivalent between groups. In conclusion, polishing reduced surface roughness and C. albicans metabolism and coverage on LCD‑printed denture base, while maintaining mechanical properties and cytocompatibility.
Graphical abstract
Supplementary Information
The online version contains supplementary material available at 10.1038/s41598-026-45942-y.
Keywords: Polymers, Candida albicans, Surface properties, Printing, Three-dimensional
Subject terms: Biotechnology, Health care, Materials science, Medical research
Introduction
Denture stomatitis (DS) is an inflammatory condition affecting up to 65% of denture wearers1. It is more frequent among hospitalized elderly individuals, smokers, and patients with systemic conditions2–4. Although its etiology is multifactorial, Candida albicans plays a central role by adhering to denture surfaces, forming biofilms, and expressing virulence factors that facilitate tissue invasion5. Because the denture base acts as a reservoir for Candida, its surface characteristics are decisive in either preventing or aggravating DS1.
Prevention of DS have focused primarily on the denture, since Candida spp. are more frequently isolated from the internal surface of dentures than from the corresponding mucosa1,6. In clinical practice, however, the intaglio surface of conventional dentures is rarely polished due to concerns about reduced retention. In fact, this is mainly determined by the border and posterior palatal seals, which stabilize the saliva film and create a suction effect. Surface tension at the periphery also contributes, but the key factors are accurate base adaptation and an effective border seal7.
Consistently, a recent study by Neeraja et al.8 evaluated the influence of surface roughness on the retention of complete dentures fabricated with different materials and techniques. The authors reported an inverse correlation between roughness and retention, with smoother denture base surfaces yielding higher retention values. This observation has led to proposals for polishing the inner denture surface as a strategy to minimize biofilm accumulation.
The correlation between surface roughness and Candida adhesion to heat-polymerized polymethyl methacrylate (PMMA) resins remains controversial9,10. While Izumida et al.11 found no significant difference in biofilm formation between surfaces of different roughness, other studies observed higher fungal adhesion on rough substrates12,13. Yamauchi et al.14 also showed that topographical irregularities increased C. albicans adhesion and proliferation, reinforcing the need for smooth intaglio surfaces to reduce microbial colonization on denture bases6,15,16.
Beyond roughness, other physicochemical properties of denture base materials can influence microbial adhesion and biofilm formation17. Surface charge may modulate binding forces between microorganisms and the substrate, while also affecting protein adsorption and salivary pellicle development. Wettability and surface free energy are also relevant. For example, microorganisms often show affinity for hydrophobic surfaces. Moreover, roughness can amplify wetting properties, enhancing the intrinsic character of the material18,19.
3D printing has become an effective method for producing denture bases, reducing material waste and fabrication costs compared with conventional techniques20–22. Among these technologies, liquid crystal display (LCD) printing is widely used in clinical practice because it is considered a faster and lower-cost system23. However, due to the layer-by-layer printing process and lower accuracy than digital light processing (DLP) and stereolithography (SLA), a more pronounced stair-step phenomenon has been reported, contributing to increased roughness of 3D-printed objects24.
These irregularities may promote microbial accumulation if not properly finished, demonstrating the importance of establishing finishing protocols that smooth the surface to clinically acceptable levels. Quirynen et al.25 reported that roughnessabove Ra = 0.2 μm significantly increase microbial adhesion, while reductions below this threshold do not provide additional benefit. This limit has since been used as a reference for polishing procedures on 3D-printed polymers, although studies mainly addressed surface roughness without evaluating microbiological outcomes26.
Despite the widespread adoption of 3D printing in prosthodontics, the existing literature has not fully addressed the fact that the printing process itself generates surface texture, even though digital impressions aim to capture oral anatomy accurately. In addition, finishing and polishing procedures for additively manufactured denture bases are not consistently standardized for clinical application, leaving uncertainty about how best to achieve clinically acceptable surface quality.
To date, no studies have evaluated the effect of polishing protocols on LCD‑printed denture bases with respect to C. albicans adhesion and related properties. Therefore, this study aimed to investigate whether surface polishing influenced surface roughness as well as the mechanical and biological performance of an LCD‑printed denture base polymer. The null hypothesis was that polishing would not significantly affect surface roughness and topography, compressive stress, wettability, C. albicans adhesion and biofilm formation or cytocompatibility.
Materials and methods
Preparation of samples
The sample size for each test was defined according to previously published in vitro studies with similar experimental design27–30. Circular samples (14 mm x 1.2 mm) were virtually designed and sliced (Autodesk Meshmixer v.3.5.474, Autodesk, Inc; Anycubic Photon Workshop V2.1.24), followed by the printing (Anycubic Photon Mono SE printer, Shenzhen Anycubic Technology Co., Ltd.) with a commercially available polymer for denture base (Cosmos Denture, Yller Biomaterials SA). The samples were printed at a 90‑degree print orientation with a layer thickness of 0.05 mm, based on Shim et al.31. After printing, the cleaning process was performed with isopropyl alcohol under agitation (Form Wash, Formlabs), followed by post‑curing in a UV light-curing unit (Form Cure, Formlabs), strictly following the manufacturer’s instructions (see Supplementary Information for details).
Polishing procedure
After preparation, samples were divided into polished (P) and not polished (NP) groups. For the NP group, a preliminary evaluation determined the average roughness of LCD‑printed samples. This value was reproduced in a standardized manner using 80‑grit silicon carbide sandpaper, ensuring all NP samples presented comparable roughness representative of the inherent texture generated by the printing lines, allowing consistent comparison with the polished group. Polishing of the P group was performed using a sequential silicon carbide sandpaper procedure (200-, 400-, 600-, 800-, 1000-, and 1200-grit), adapted from de Foggi et al.9, with the aim of achieving the clinically acceptable threshold for surface roughness (0.2 μm)25.
All polishing steps were carried out manually under continuous running water to minimize heat generation and debris accumulation. Each sanding stage was timed with a digital timer (30–60 s) and executed with uniform circular movements under light and constant pressure, controlled through previous operator training. A single trained operator conducted all procedures to reduce variability and to ensure reproducibility across all samples. Sandpaper sheets were replaced regularly to maintain abrasive efficiency. After polishing, samples were cleaned in an ultrasonic bath for 10 min and stored in distilled water at 37 °C for 48 h before testing.
Surface roughness (Ra)
Subsequently, the samples were evaluated by both contact (SJ 400, Mitutoyo Corp.) and non-contact (VK-X250 3D Laser Scanning Microscope, Keyence) profilometers. Contact measurements (n = 9) were conducted using 5 μm radius diamond tip at speed of 0.5 mm/s, with resolution of 0.01 μm, 0.8 mm interval, and 2.4 mm transverse length. Three readings were performed per sample across predetermined, representative areas. For non-contact analysis (n = 2), Differential Interference Contrast (DIC) optical images were acquired at 10× and 20× magnifications. Then, Ra was quantified using the MultiFile Analyzer software based on three representative images of distinct locations per sample to determine mean values.
Surface topography
The surface topography was evaluated by Scanning Electronic Microscopy (SEM) analysis. Samples from each group (n = 1) were dehydrated, metallized with carbon, and positioned in the microscope (JEOL JSM, 6610LV) to obtain the images at final magnifications of 100×, 250×, and 500×.
Water contact angle (WCA)
The wettability of the samples was evaluated using a drop-shape analyzer (Krüss DSA25 system). A single droplet of deionized water (0.40 µL) was deposited on each sample (n = 9). After 10 s of contact, images were captured and analyzed. The WCA was calculated with the DSA4 software32.
Compressive stress
The compressive test was performed using an ElectroForce 3200 mechanical tester (TA Instruments) equipped with a 450 N load cell and operated at a strain rate of 1%, simulating localized compressive loads on denture bases during mastication. Circular samples (P group: n = 6; NP group: n = 7) were prepared with 3 mm diameter and 1 mm thickness. The linear elastic region was defined as the portion of the stress–strain curve showing proportionality between these two parameters, confirmed by linear regression (R² ≥ 0.99). After testing, the interval between 15% and 18% strain met this criterion (R² = 0.99), and the slope of this region was subsequently used to calculate the Young’s modulus (MPa)33,34. When evident inconsistencies were observed in stress–strain curves, the data were excluded.
Adhesion and biofilm formation of C. albicans
This test consisted of cultivating Candida albicans (ATCC 90028) biofilms on the polymer samples, subsequently assessed by cell proliferation (colony-forming units, CFU/mL) and metabolism (AlamarBlue assay) (n = 9, each test)35.
Candida albicans was initially cultured on Sabouraud Dextrose Agar (SDA) for 48 h at 37 °C. A standardized inoculum was prepared by incubating five colonies in 10 mL of Yeast Nitrogen Base (YNB) for 16 h at 37 °C, followed by a 1:20 dilution in YNB for an additional 8 h. The resulting suspension equivalent to 1 × 106 CFU/mL was centrifuged, washed in Phosphate-Buffered Saline (PBS), and resuspended in RPMI-1640 medium. The samples were UV-disinfected (20 min/side) and placed in 24-well plates for a 90-min adhesion phase with 750 µL of the fungal suspension36. Non-adherent cells were removed by PBS washing, followed by a 48-h biofilm maturation period in 1500 µL of RPMI. Mature biofilms were harvested by timing the scraping to 1 min under standardized conditions, ensuring consistent detachment across all samples. Subsequently, serial dilutions (10− 1 to 10− 4) were plated onto SDA and incubated for 48 h at 37 °C to determine viable cell counts (CFU/mL).
C. albicans cell metabolism was analyzed using the alamarBlue® reagent (Invitrogen™). After 48 h of mature biofilm formation on the samples, 10% (v/v) of alamarBlue® solution was added to the plate and incubated for 4 hours37. Subsequently, fluorescence intensity was measured using Fluoroskan Ascent plate reader (Thermo Fisher Scientific) at 560 nm and 590 nm.
Live-stained C. albicans biofilm coverage and CLSM analysis
C. albicans biofilm coverage was assessed using the LIVE/DEAD™ FungaLight™ Yeast kit (Invitrogen™). A working solution was prepared by mixing SYTO-9 and propidium iodide (PI) at a 1:1 ratio, subsequently diluted 1:1000 in PBS. Samples (n = 3) were incubated with the staining solution for 30 min at 37 °C in the dark. Images were acquired using a confocal laser scanning microscope (CLSM 800 with Airyscan, Zeiss) at 10× magnification.
Fluorescence was detected using excitation/emission wavelengths of 488/500–550 nm for SYTO-9 and 561/580–650 nm for PI. Quantification was performed using the ZEN software by defining a single region of interest (ROI), which was consistently applied to three representative images per sample. Even though both channels were monitored, the green channel (SYTO-9) was selected for quantification and figure representation to assess viable biofilm coverage. Data were expressed as mean fluorescence intensity (arbitrary units, a.u.). The red channel (PI) was utilized exclusively to confirm cell viability across the samples.
Cytotoxicity tests
Cytotoxicity was evaluated via cell metabolism through indirect and direct exposure tests (n = 3, each test) following ISO 10993-5:2009 and ISO 10993-12:2021 standards38–40. L-929 mouse fibroblasts (NCTC clone 929 [L cell, L-929, derivative of strain L], CCL-1) were maintained in Dulbecco’s Modified Eagle medium (DMEM, Thermo Fisher Scientific) supplemented with 10% fetal bovine serum (FBS) and 1% penicillin-streptomycin. For all assays, cells (passage 13) were seeded at 5 × 103 cells/well in 96-well plates for 24 h at 37 °C.
For the indirect assay, extracts were prepared by incubating the UV-disinfected samples in 3 mL of DMEM for 24 h at 37 °C. The control group consisted of 3 mL of medium stored under the same conditions. Cells were exposed to the extracts or medium for 24 h, with daily reapplication over 1, 3, and 7 days (exhaustive method). In the direct assay, the discs were punched (6 mm diameter x 1.2 mm thickness) and placed in 96-well plates, and cells were seeded directly onto the samples following the same cell density as previously described, with medium renewal every 48 h.
Metabolic activity was quantified at days 1, 3, and 7 by adding 10% (v/v) of PrestoBlue® and subsequent 90 min of incubation. Fluorescence was measured (Excitation/Emission: 560/590 nm) using CLARIOstar plate reader (BMG LABTECH). Results were normalized to the control group (100% viable cells) and interpreted according to ISO thresholds39,40. Cell morphology and proliferation were qualitatively monitored via inverted microscopy (Nikon Eclipse Ti-S/L100, Nikon Instruments) and representative images were acquired at a selected time point.
Qualitative and quantitative BSA adsorption
BSA adsorption was evaluated by incubating samples in 1 mL of PBS for 24 h at 37 °C, followed by an incubation in 1 mL of Alexa Fluor 488-labeled Bovine Serum Albumin (BSA) (0.01 mg/mL in PBS) for 24 h at 37 °C. Unbound protein was removed through five consecutive PBS rinses (1 mL each). Samples (n = 3) were imaged via fluorescence microscopy at 20× magnification32. Protein adsorption on the surface was quantified from three representative images using ImageJ software (NIH) by calculating the fluorescence intensity (arbitrary units, a.u.) within defined ROI.
Statistical analysis
Data were subjected to descriptive analysis and assessed for normality and homogeneity of variances using the Shapiro-Wilk and Levene’s tests, respectively. Outliers, if any, were identified and excluded using the Grubbs’ test (extreme studentized deviate method) implemented in the GraphPad Prism outlier calculator. Despite the relatively small sample size, the data presented normal distribution and equal variance, supporting the use of a parametric test. Accordingly, the Student’s t‑test was applied for group comparisons at a significance level of 5% (α = 0.05). The number of statistical comparisons was limited to those directly addressing the study objectives, thereby minimizing the risk of type I error inflation. All analyses were performed in GraphPad Prism version 10.3.1 for Windows (GraphPad Software).
Results
The surface roughness of the polished (P) and not polished (NP) groups was evaluated using both contact and non-contact profilometers, and the data sets were analyzed independently. The Ra values obtained were 0.28 ± 0.03 μm and 5.81 ± 0.40 μm (Fig. 1a), and 0.86 ± 0.03 μm and 5.25 ± 0.56 μm (Fig. 1b), respectively. A statistically significant difference was observed between the groups, with the polishing protocol reducing Ra by approximately twenty-fold (p < 0.05). This distinction was confirmed qualitatively through optical microscopy and SEM images, which demonstrated pronounced irregularities in the NP group compared to the smoother and more homogeneous topography of the P samples (Fig. 1c and d).
Fig. 1.
a Surface roughness (Ra) values measured using a contact profilometer and b a non-contact profilometer. c Representative optical microscopy images of surface topography. d Representative SEM micrographs of the samples. Note. P: polished group; NP: not polished group. Optical microscopy images were obtained at 10 × (1) and 20 × (2) magnifications (scale bars: 200 μm and 100 μm). SEM images were acquired at 100 × (1), 250 × (2), and 500 × (3) magnifications (scale bars: 500 μm, 300 μm, and 100 μm). ****p < 0.001; **p = 0.008.
The stress–strain curves of the P group exhibited variable deformation, while NP group showed more consistent behavior within the group. The mean percentage strain at maximum stress was similar between the groups: 14.96% for P and 14.40% for NP (Fig. 2a). One sample from each group displayed anomalous stress–strain curves compared to the other samples and were excluded from the analysis, without formal outlier testing, as detailed in the Supplementary Information. For the calculation of Young’s modulus, a linear region of the stress–strain curve between 15% and 18% strain was selected, as this interval provided the most stable slope with R2 = 0.99. From this region, Young’s modulus was calculated as 319.80 ± 58.92 MPa for the P group and 325.30 ± 31.22 MPa for the NP group, with no statistically significant difference (p = 0.845) (Fig. 2b). For WCA analysis, the droplet profile is shown in Fig. 2c. Polished samples exhibited significantly higher water contact angle values (θ = 93.37° ± 6.30°) compared with NP samples (θ = 74.52° ± 5.24°) (p < 0.05) (Fig. 2d).
Fig. 2.
a Representative compressive stress–strain curves of polished (P) and not polished (NP) samples. b Young’s modulus values (MPa) calculated from the linear region between 15% and 18% strain (R2 = 0.99). c Representative droplet profiles used for water contact angle (WCA) measurement. d WCA values (θ = °) for both groups. Note. P: polished group; NP: not polished group. ****p < 0.001; ns: non-significant.
For the analysis of C. albicans proliferation, colony-forming units (CFU/mL) were quantified. Both groups showed growth of approximately 6 logs, with no statistically significant difference (p > 0.05) (Fig. 3a). Cell metabolism assessed by the AlamarBlue assay resulted in 10,262 ± 354 a.u. (P) and 11,870 ± 639 a.u. (NP), indicating a significant reduction in metabolic activity after polishing (p < 0.05) (Fig. 3b). For C. albicans biofilm coverage on the samples, quantitative and qualitative analyses were performed using three representative confocal images per group to obtain fluorescence intensity values. CLSM images revealed a denser green-stained biofilm on NP surfaces compared with P (Fig. 3c). Fluorescence intensity measured over a standardized area of 405,388.2 µm2 confirmed this statistically significant difference, with mean values of 56,200 ± 6,200 a.u. for P and 73,700 ± 11,800 a.u. for NP (p < 0.05) (Fig. 3d).
Fig. 3.
a Log10 CFU/mL values of C. albicans biofilm. b Metabolic activity assessed by AlamarBlue assay (relative fluorescence units, RFU). c Representative CLSM images of C. albicans biofilm coverage (live cells in green) at 10× magnification (scale bar: 100 μm). d Quantification of live biofilm coverage based on fluorescence intensity. Note. P: polished group; NP: not polished group. ****p < 0.001; **p = 0.001; ns: non-significant.
Cytotoxicity assessed by direct contact showed RFU values of 5,845 a.u. (P) and 5,727 a.u. (NP) on day 1, increasing on day 3 and reaching 12,982 a.u. (P) and 10,267 a.u. (NP) on day 7. Both groups exhibited approximately a two-fold increase in fluorescence from day 1 to day 7 (Fig. 4a). In the indirect contact test, results were expressed as percentage of cell metabolism normalized to the control group (cells incubated in culture medium). Metabolic activity in both experimental groups was comparable to the control, consistently around 100% over 7 days (Fig. 4b). In the qualitative analysis of the indirect contact test, representative images showed L-929 cells at a sub-confluent stage 24 h after seeding. After 72 h of extract exposure, cells reached high confluency with monolayer formation and displayed elongated fibroblast-like morphology (Fig. 4c).
Fig. 4.
a Relative fluorescence units (RFU) from the direct cell metabolism test at 1, 3, and 7 days. b Percentage of cell metabolism from indirect cell metabolism test, normalized to the control group, both assessed by PrestoBlue assay. c Representative images of L-929 cell morphology and proliferation 24 h after seeding (top) and monolayer formation after 72 h of extract exposure in polished (P) and not polished (NP) groups (bottom) (scale bar: 50 μm). Note. P: polished group; NP: not polished group.
Protein adsorption analysis showed a uniform distribution of BSA in the polished group, whereas the not polished group exhibited localized regions of higher accumulation (Fig. 5a). Autofluorescence background from the samples was checked and no signal was detected. Quantitatively, after background subtraction, fluorescence intensity measured over a standardized area of 133,519.6 μm² from three representative images per group resulted in mean values of 23.19 ± 2.50 a.u. (P) and 24.63 ± 0.40 a.u. (NP), with no statistically significant difference between groups (p > 0.05) (Fig. 5b).
Fig. 5.
a Representative fluorescence microscopy images of Alexa Fluor 488-labeled BSA on the sample surfaces, acquired at 20× magnification (scale bar: 50 μm). b Quantification of fluorescence intensity based on representative images. Note. BC: background control; P: polished group; NP: not polished group. ns: non-significant.
Discussion
This study evaluated the effects of surface polishing on an LCD-printed denture base polymer, focusing on surface roughness and its association with mechanical and biological responses. The null hypothesis was partially rejected, as polishing resulted in a significant reduction in roughness, while mechanical properties remained unaffected. In addition, polished surfaces demonstrated lower Candida albicans metabolic activity and live coverage, and cytotoxicity and protein adsorption were not significantly altered, suggesting this procedure did not adversely affect the biocompatibility of the material.
The surface roughness of denture base favors microbial adhesion, which under predisposing conditions such as systemic disease or poor hygiene may lead to denture stomatitis1,12. Because Candida albicans can act as an opportunistic pathogen, reducing the surface irregularities of prosthetic bases can represent an important preventive strategy. Various studies have emphasized the benefits of minimizing surface roughness through polishing protocols, as smoother polymers surfaces accumulate fewer microorganisms and facilitate routine cleaning9,41,42.
The roughness values of the not polished samples (NP) varied from previous reports comparing LCD-printed samples with conventional, DLP-printed and milled samples43–47. This suggests that surface properties are influenced by intrinsic material characteristics and processing factors such as printer technology, printing orientation, layer thickness, and post-processing procedures31. To ensure comparability, the present study employed a 90° printing orientation, which has been reported to reduce roughness and Candida albicans adhesion, representing the most favorable clinical condition, while also improving accuracy31,48,49.
In the context of 3D printing methods, LCD printing technology have been associated to disadvantages related to low resolution and accuracy47,49. However, the studies are controversial as it has been reported that DLP printers generally produce smoother surfaces, with Ra values around 1 µm44. Conversely, Viotto et al.46 reported lower roughness in LCD samples than in DLP, despite methodological differences, including a polishing standardization step prior to coffee immersion.
Both contact and non-contact profilometers yielded different absolute values but consistently demonstrated a significant reduction in surface roughness. The contact profilometer physically traces the surface profile, whereas the non-contact laser microscope captures the topography optically, allowing comparison with previous studies9,29,35. Importantly, the proposed polishing protocol achieved the clinically acceptable threshold associated with reduced microbial adhesion25,50, highlighting its practical applicability as a procedure that can be further replicated chairside or in the laboratory.
Improvements in polymers smoothness after polishing have been consistently documented41,51,52, and Kraemer Fernandez et al.53 showed that polished DLP‑printed samples had better surface characteristics than coated samples. Similar reductions in Ra have also been reported following mechanical polishing of 3D‑printed polymers54,55, although Quezada et al.55 found lower values with manual than mechanized polished samples. Nevertheless, discussion is still limited due to the lack of studies on polishing effects in LCD‑printed polymers, indicating the need for further investigation in this area.
Because denture base materials must resist localized compressive loads in areas of mucosal contact during mastication, compressive testing was performed to evaluate their strength and durability, critical properties for long-span prostheses34. Mechanical testing showed no significant differences between polished and not polished groups, consistent with previous findings on acrylic resin, where controlled polishing did not induce structural or chemical degradation55. The polishing procedure likely affected the superficial layer without compromising the integrity of the internal structure.
Rougher surfaces generally favor Candida colonization, as irregularities create retention niches that protect cells from shear forces during cleaning15,42. Although this study did not test different materials, several studies have shown that 3D-printed polymers exhibit higher microbial adhesion than PMMA and milled CAD-CAM samples, along with increased expression of virulence factors such as hyphal transition5,56,57. Consistent with these observations, Ryu et al.57 demonstrated that high roughness and consequent microbial adhesion of this material, particularly across different printing orientations, require meticulous polishing and regular cleansing.
In this study, polished samples did not significantly reduce CFU counts but altered C. albicans metabolism and live coverage on the samples’ surface. Such discrepancies may be influenced by methodological differences, since CFU assays rely on colony counting, which have been suggested to potentially underestimate the number of cells per mL58. The observed reduction could also be related to the decreased contact surface after polishing, which may limit the number of adhered cells. Conversely, rougher surfaces provide a larger contact area that favors cell adhesion13,59.
Other surface properties, such as wettability and surface free energy, may also influence microbial adhesion15,60. In this study, polished samples showed reduction in C. albicans metabolism and slightly higher water contact angle values compared to not polished surfaces. These differences, however, were modest and limited and did not establish a clear link with previous reports53,61 that associated rough and hydrophobic surfaces with enhanced microbial adhesion, as well as with complementary literature suggesting that hydrophobic materials can promote C. albicans biofilm formation through non-covalent interactions between cell wall proteins and the substrate17,42.
According to the Wenzel model18, surface roughness can modify wettability depending on the intrinsic character of the material. In this study, it was suggested that polishing might reveal the hydrophobic nature of the polymer. However, this measurement alone cannot fully characterize hydrophobicity, as increased surface contact in rough samples may facilitate drop spreading. Moreover, the literature remains inconclusive regarding the association between roughness and wettability, as well as their impact on biological outcomes9,62.
Physicochemical factors, including surface charge and polymer composition, can influence protein adsorption and thereby affect cell and microorganism adhesion63,64. Early adsorption of salivary proteins is known to mediate the initial stages of biofilm formation65,66. Daeschel et al.67 reported that protein adsorption to solid surfaces can be modulated by temperature, pH, protein characteristics, and surface properties. In the results, no significant differences in protein adsorption were observed between groups. This may be related to nanoscale protein organization influencing its adsorption independently of surface topography, an aspect not assessed in this study and deserving further investigation.
Since the denture base is the primary interface with the oral environment and in prolonged contact with the mucosa, such modifications could affect its biocompatibility, particularly as 3D printing polymers are mainly composed of methacrylate or acrylate monomers, that present potential toxicity68–70. The methods applied demonstrated sustained cytocompatibility, with metabolic activity comparable to the control group (100% viable cells) over time, although the assays may not detect subtle physiological changes. In line, Dai et al.71 evaluated the polishing of DLP‑printed samples and observed that polished samples did not exhibit additional cytotoxicity.
In clinical routine, the internal surface of dentures is usually left unpolished under the assumption this procedure could reduce retention. However, this property depends primarily on border and posterior palatal seals rather than the microtexture of the support tissue7. Although retention was not directly assessed in the present study, Neeraja et al.8 reported that smoother surfaces yielded higher retention values, suggesting that increased roughness may disrupt adaptation and compromise adhesion and cohesive forces. Since the intaglio surface is the primary site of denture stomatitis, this procedure is relevant for future studies evaluating clinical outcomes related to 3D-printed complete dentures.
This study has some limitations, including the use of only one polymer and printing technology, the absence of a mechanical polishing device, the use of flat samples that do not reproduce the complex geometry of complete dentures, and reliance on a single-specie biofilm and a single cell line in the cytotoxicity test without assessment of inflammatory or oxidative markers. Further investigations incorporating multispecies biofilm models and additional biological evaluations would help to strengthen the evidence base and broaden the applicability of the proposed protocol.
Conclusion
The results indicated that surface polishing reduced surface roughness and was associated with lower Candida albicans biofilm metabolism and coverage on LCD‑printed denture base samples, while mechanical properties and cytocompatibility were not significantly affected.
Supplementary Information
Below is the link to the electronic supplementary material.
Acknowledgements
The authors thank to São Paulo Research Foundation (FAPESP) for supporting this study through grants nº 2021/09624-3, nº 2021/09555-1, and n° 2024/01648-9.
Author contributions
ACF: Conceptualization, methodology, investigation, data analysis, writing, review and editing; JSO, LS: Methodology, data analysis; LVA: Resources, technical support (3D printing); CM, MBB, JHJ: Supervision, data interpretation, writing (review and editing).
Funding
This study was funded by the São Paulo Research Foundation (FAPESP), grants nº 2021/09624-3, nº 2021/09555-1, and n° 2024/01648-9.
Data availability
Data are provided within the manuscript or supplementary information files.
Declarations
Conflict of interests
The authors declare no competing interests.
Ethical approval
The study was conducted entirely in vitro and did not involve human participants or animal subjects; therefore, ethical approval was not required.
Footnotes
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Gendreau, L. & Loewy, Z. G. Epidemiology and etiology of denture stomatitis. J. Prosthodont.20 (4), 251–260. (2011). 10.1111/j.1532-849X.2011.00698.x [DOI] [PubMed] [Google Scholar]
- 2.Adam, R. Z. & Kimmie-Dhansay, F. Prevalence of denture-related stomatitis in edentulous patients at a tertiary dental teaching hospital. Front. Oral Health. 2, 772679. (2021). 10.3389/froh.2021.772679 PubMed PMID: 35048072; PubMed Central PMCID: PMC8757833. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Sardari, F., Khalili, P., Hakimi, H., Mahmoudaghaei, S. & Abedi, P. The prevalence of denture stomatitis in cigarette and hookah smokers and opium addicts: findings from Rafsanjan Cohort Study. BMC Oral Health. 21 (1), 455. (2021). 10.1186/s12903-021-01807-6 PubMed PMID: 34535104; PubMed Central PMCID: PMC8449426. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Martorano-Fernandes, L. et al. Oral candidiasis and denture stomatitis in diabetic patients: systematic review and meta-analysis. Braz Oral Res.34, e113. (2020). 10.1590/1807-3107bor-2020 PubMed PMID: 32965459. [DOI] [PubMed] [Google Scholar]
- 5.Vetsch, G. et al. Candida albicans colonization on CAD-CAM denture resin surface. J. Dent.157, 105756. (2025). 10.1016/j.jdent.2025.105756 [DOI] [PubMed] [Google Scholar]
- 6.Davenport, J. C. The oral distribution of Candida in denture stomatitis. Br. Dent. J.129 (4), 151–156. (1970). 10.1038/sj.bdj.4802540 PubMed PMID: 5272473. [DOI] [PubMed] [Google Scholar]
- 7.Darvell, B. W. & Clark, R. K. F. The physical mechanisms of complete denture retention. Br. Dent. J.189 (5), 248–252. (2000). 10.1038/sj.bdj.4800734 [DOI] [PubMed] [Google Scholar]
- 8.Neeraja, B. et al. Surface roughness on retention of complete dentures using conventional and digital techniques: an in vitro study. Bioinformation21 (08), 2849–2852. (2025). 10.6026/973206300212849 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.De Foggi, C. C. et al. Effect of surface roughness on the hydrophobicity of a denture-base acrylic resin and Candida albicans colonization. J. Invest. Clin. Dent.7 (2), 141–148. (2016). 10.1111/jicd.12125 [DOI] [PubMed] [Google Scholar]
- 10.Mayahara, M. et al. Effects of surface roughness and dimorphism on the adhesion of Candida albicans to the surface of resins: scanning electron microscope analyses of mode and number of adhesions. J. Invest. Clin. Dent.5 (4), 307–312. (2014). 10.1111/jicd.12055 [DOI] [PubMed] [Google Scholar]
- 11.Izumida, F. E. et al. Surface roughness and Candida albicans biofilm formation on a reline resin after long-term chemical disinfection and toothbrushing. J. Prosthet. Dent.112 (6), 1523–1529. (2014). 10.1016/j.prosdent.2014.06.001 [DOI] [PubMed] [Google Scholar]
- 12.Verran, J. & Maryan, C. J. Retention of Candida albicans on acrylic resin and silicone of different surface topography. J. Prosthet. Dent.77 (5), 535–539. (1997). 10.1016/S0022-3913(97)70148-3 [DOI] [PubMed] [Google Scholar]
- 13.Radford, D. R., Sweet, S. P., Challacombe, S. J. & Walter, J. D. Adherence of Candida albicans to denture-base materials with different surface finishes. J. Dent.26 (7), 577–583. (1998). 10.1016/S0300-5712(97)00034-1 [DOI] [PubMed] [Google Scholar]
- 14.Yamauchi, M., Yamamoto, K., Wakabayashi, M. & Kawano, J. In vitro adherence of microorganisms to denture base resin with different surface texture. Dent. Mater. J.9 (1), 19–24121. (1990). 10.4012/dmj.9.19 [DOI] [PubMed] [Google Scholar]
- 15.Pereira-Cenci, T., Cury, A. A. D. B., Cenci, M. S. & Rodrigues-Garcia, R. C. M. In vitro Candida colonization on acrylic resins and denture liners: influence of surface free energy, roughness, saliva, and adhering bacteria. Int. J. Prosthodont.20 (3), 308–310 (2007). PubMed PMID: 17580465. [PubMed] [Google Scholar]
- 16.Kulak, Y., Arikan, A. & Kazazoglu, E. Existence of Candida albicans and microorganisms in denture stomatitis patients. J. Oral Rehabil. 24 (10), 788–790. (1997). 10.1046/j.1365-2842.1997.00550.x PubMed PMID: 9372471. [DOI] [PubMed] [Google Scholar]
- 17.Goswami, R. R., Pohare, S. D., Raut, J. S. & Mohan Karuppayil, S. Cell surface hydrophobicity as a virulence factor in Candida albicans. Biosci. Biotech. Res. Asia. 14 (4), 1503–1511. (2017). 10.13005/bbra/2598 [Google Scholar]
- 18.Wenzel, R. N. Resistance of solid surfaces to wetting by water. Ind. Eng. Chem.28 (8), 988–994. (1936). 10.1021/ie50320a024 [Google Scholar]
- 19.Gad, M. M., Abualsaud, R. & Khan, S. Q. Hydrophobicity of denture base resins: a systematic review and meta-analysis. J. Int. Soc. Prev. Community Dentistry. 12 (2), 139–159. (2022). 10.4103/jispcd.JISPCD_213_21 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Alghazzawi, T. F. Advancements in CAD/CAM technology: options for practical implementation. J. Prosthodontic Res.60 (2), 72–84. (2016). 10.1016/j.jpor.2016.01.003 [DOI] [PubMed] [Google Scholar]
- 21.Stansbury, J. W. & Idacavage, M. J. 3D printing with polymers: challenges among expanding options and opportunities. Dent. Mater.32 (1), 54–64. (2016). 10.1016/j.dental.2015.09.018 [DOI] [PubMed] [Google Scholar]
- 22.Faty, M. A., Sabet, M. E. & Thabet, Y. G. A Comparison of denture base retention and adaptation between CAD/CAM and conventional fabrication techniques. Int. J. Prosthodont.36 (4), 469–478. (2023). 10.11607/ijp.7193 PubMed PMID: 37699188. [DOI] [PubMed] [Google Scholar]
- 23.Viotto, H. E. et al. Mechanical and chemical properties of a 3-dimension–printed denture base resin obtained by varying digital light processing and liquid crystal display printing angles. JADA Foundational Sci.5, 100064. (2026). 10.1016/j.jfscie.2025.100064 [Google Scholar]
- 24.Anderson, L. N., Alsahafi, T., Clark, W. A., Felton, D. & Sulaiman, T. A. Evaluation of surface roughness of differently manufactured denture base materials. J. Prosthet. Dent.133 (6), 1586–1591. (2025). 10.1016/j.prosdent.2023.08.028 [DOI] [PubMed] [Google Scholar]
- 25.Quirynen, M. et al. The influence of surface free energy and surface roughness on early plaque formation: an in vivo study in man. J. Clin. Periodontology. 17 (3), 138–144. (1990). 10.1111/j.1600-051X.1990.tb01077.x [DOI] [PubMed] [Google Scholar]
- 26.Al-Dulaijan, Y. A. et al. Comparative evaluation of surface roughness and hardness of 3D printed resins. Materials15 (19), 6822. (2022). 10.3390/ma15196822 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Panariello, B. H. D. et al. Effects of short-term immersion and brushing with different denture cleansers on the roughness, hardness, and color of two types of acrylic resin. Am. J. Dent.28 (3), 150–156. (2015). PubMed PMID: 26201226. [PubMed] [Google Scholar]
- 28.Zoccolotti, J. O. et al. Properties of an acrylic resin after immersion in antiseptic soaps: low-cost, easy-access procedure for the prevention of denture stomatitis. PLoS ONE13 8 e0203187. (2018). 10.1371/journal.pone.0203187 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Camargo, R. et al. Surface changes induced by brushing increase Candida albicans biofilms on 3D-printed denture base resin. JoF11 (9), 668. (2025). 10.3390/jof11090668 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.De Oliveira, J. S. et al. Effect of prolonged brushing with different agents on surface characteristics of conventional and 3D-printed denture resins. Sci. Rep.15 (1), 37305. (2025). 10.1038/s41598-025-21212-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Shim, J. S. et al. Surface characteristics, and microbial adhesion of 3D-printed resins with various printing orientations. J. Prosthet. Dent.124 (4), 468–475. (2020). 10.1016/j.prosdent.2019.05.034 [DOI] [PubMed] [Google Scholar]
- 32.Feliciano, A. J. et al. Complementary supramolecular functionalization enhances antifouling surfaces: a ureidopyrimidinone-functionalized phosphorylcholine polymer. ACS Biomater. Sci. Eng.9 (8), 4619–4631. (2023). 10.1021/acsbiomaterials.3c00425 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Contessi Negrini, N., Angelova Volponi, A., Sharpe, P. T. & Celiz, A. D. Tunable cross-linking and adhesion of gelatin hydrogels via bioorthogonal click chemistry. ACS Biomater. Sci. Eng.7 (9), 4330–4346. (2021). 10.1021/acsbiomaterials.1c00136 PubMed PMID: 34086456. [DOI] [PubMed] [Google Scholar]
- 34.Pantea, M. et al. Compressive and flexural strength of 3D-printed and conventional resins designated for interim fixed dental prostheses: an in vitro comparison. Materials15 (9), 3075. (2022). 10.3390/ma15093075 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.de Oliveira, J. S. et al. Cryptocarya moschata fractions decrease planktonic cells and biofilms of Candida albicans and Streptococcus mutans. Biofouling40 (10), 831–846. (2024). 10.1080/08927014.2024.2418466 PubMed PMID: 39444328. [DOI] [PubMed] [Google Scholar]
- 36.Sheridan, P. J., Koka, S., Ewoldsen, N. O., Lefebvre, C. A. & Lavin, M. T. Cytotoxicity of denture base resins. Int. J. Prosthodont.10 (1), 73–77 (1997). PubMed PMID: 9484073. [PubMed] [Google Scholar]
- 37.Hamid, R., Rotshteyn, Y., Rabadi, L., Parikh, R. & Bullock, P. Comparison of alamar blue and MTT assays for high through-put screening. Toxicol. Vitro. 18 (5), 703–710. (2004). 10.1016/j.tiv2004.03.012 PubMed PMID: 15251189. [DOI] [PubMed] [Google Scholar]
- 38.Wennberg, A., Hasselgren, G. & Tronstad, L. A method for toxicity screening of biomaterials using cells cultured on millipore filters. J. Biomed. Mater. Res.13 (1), 109–120. (1979). 10.1002/jbm.820130112 PubMed PMID: 429379. [DOI] [PubMed] [Google Scholar]
- 39.ISO 10993. Biological Evaluation of Medical Devices - Part 5: Tests for in Vitro Cytotoxicity.
- 40.ISO 10993. Biological Evaluation of Medical Devices - Part 12: Sample Preparation and Reference Materials.
- 41.Chatzivasileiou, K., Emmanouil, I., Kotsiomiti, E. & Pissiotis, A. Polishing of denture base acrylic resin with chairside polishing kits: an SEM and surface roughness study. Int. J. Prosthodont.26 (1), 79–81. (2013). 10.11607/ijp.3157 PubMed PMID: 23342338. [DOI] [PubMed] [Google Scholar]
- 42.Freitas, R. F. C. P. D. et al. Physical, mechanical, and anti-biofilm formation properties of CAD‐CAM milled or 3D printed denture base resins: in vitro analysis. J. Prosthodont.32 (S1), 38–44. (2023). 10.1111/jopr.13554 [DOI] [PubMed] [Google Scholar]
- 43.Verran, J. et al. The effect of dentifrice abrasion on denture topography and the subsequent retention of microorganisms on abraded surfaces. J. Prosthet. Dent.112 (6), 1513–1522. (2014). 10.1016/j.prosdent.2014.05.009 [DOI] [PubMed] [Google Scholar]
- 44.Li, P. et al. Effect of additive manufacturing method and build angle on surface characteristics and Candida albicans adhesion to 3D printed denture base polymers. J. Dent.116, 103889. (2022). 10.1016/j.jdent.2021.103889 PubMed PMID: 34800586. [DOI] [PubMed] [Google Scholar]
- 45.Kim, D. et al. Effects of post-curing time on the mechanical and color properties of three-dimensional printed crown and bridge materials. Polym. (Basel). 12 (11), 2762. (2020). 10.3390/polym12112762 PubMed PMID: 33238528; PubMed Central PMCID: PMC7700600. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Viotto, H. E. C. et al. Influence of 3D printing parameters on surface roughness and stainability of a denture base resin: An in vitro study. J. Prosthet. Dent.134 (5), 1965e1–1965e7. (2025). 10.1016/j.prosdent.2025.07.025 [DOI] [PubMed] [Google Scholar]
- 47.Katheng, A. et al. Comparison of mechanical properties of different 3D printing technologies. Sci. Rep.15 (1), 18998. (2025). 10.1038/s41598-025-03632-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Kim, H., Son, K., Kim, S. Y. & Lee, K. B. Comparison of the trueness of complete dentures fabricated using liquid crystal display 3D printing according to build angle and natural light exposure. JFB16 (8), 277. (2025). 10.3390/jfb16080277 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Tatlicioglu, B. D., Eroglu, E., Kiran, K. & Ucak, H. Comparison of DLP and LCD 3D printing technologies: effects on denture base accuracy, weight, residual material, resin consumption, and production time. J. Dent.165, 106270. (2026). 10.1016/j.jdent.2025.106270 PubMed PMID: 41319802. [DOI] [PubMed] [Google Scholar]
- 50.Alfouzan, A. F. et al. Efficacy of denture cleansers on microbial adherence and surface topography of conventional and CAD/CAM-processed denture base resins. Polymers15 (2), 460. (2023). 10.3390/polym15020460 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Guimarães, G. G., Sarte, M. F., Barreto, J. O., de Freitas-Pontes, K. M. & Regis, R. R. Effect of a continuous mechanical polishing protocol and toothbrushing on the surface roughness of acrylic resin teeth. Am. J. Dent.33 (4), 187–190 (2020). PubMed PMID: 32794392. [PubMed] [Google Scholar]
- 52.Sofou, A., Emmanouil, J., Peutzfeldt, A. & Owall, B. The effect of different polishing techniques on the surface roughness of acrylic resin materials. Eur. J. Prosthodont. Restor. Dent.9 (3–4), 117–122 (2001). PubMed PMID: 12192947. [PubMed] [Google Scholar]
- 53.Kraemer Fernandez, P., Unkovskiy, A., Benkendorff, V., Klink, A. & Spintzyk, S. Surface characteristics of milled and 3D printed denture base materials following polishing and coating: an in-vitro study. Mater. (Basel). 13 (15), 3305. (2020). 10.3390/ma13153305 PubMed PMID: 32722240; PubMed Central PMCID: PMC7435723. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Çakmak, G. et al. Surface roughness and color stability of 3D-printed denture base materials after simulated brushing and thermocycling. Mater. (Basel). 15 (18), 6441. (2022). 10.3390/ma15186441 PubMed PMID: 36143757; PubMed Central PMCID: PMC9503686. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Quezada, M. M., Salgado, H., Correia, A., Fernandes, C. & Fonseca, P. Investigation of the effect of the same polishing protocol on the surface roughness of denture base acrylic resins. Biomedicines10 (8), 1971. (2022). 10.3390/biomedicines10081971 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Meirowitz, A. et al. Effect of denture base fabrication technique on Candida albicans adhesion in vitro. Materials14 (1), 221. (2021). 10.3390/ma14010221 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Ryu, H. S. et al. The effects of fabrication methods and build orientation on Candida albicans adhesion on 3D-printed and conventional denture resin: an in vitro comparative study. BMC Oral Health. 25 (1), 1788. (2025). 10.1186/s12903-025-07266-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Soares, A. B. et al. Quantification methods of Candida albicans are independent irrespective of fungal morphology. Microb. Cell.11, 265–277. (2024). 10.15698/mic2024.07.831 PubMed PMID: 39081907; PubMed Central PMCID: PMC11287054. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Da Silva, W. J. et al. Surface roughness influences Candida albicans biofilm formation on denture materials. Rev. Odonto Ciênc (Online). 31 (2), 54. (2016). 10.15448/1980-6523.2016.2.15324 [Google Scholar]
- 60.Lee, M. J., Kim, M. J., Oh, S. H. & Kwon, J. S. Novel dental poly (Methyl Methacrylate) containing phytoncide for antifungal effect and inhibition of oral multispecies biofilm. Mater. (Basel). 13 (2), 371. (2020). 10.3390/ma13020371 PubMed PMID: 31941105; PubMed Central PMCID: PMC7014161. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Yan, S., Zhou, J. L., Zhang, R. J. & Tan, F. B. Evaluation of the influence of different build angles on the surface characteristics, accuracy, and dimensional stability of the complete denture base printed by digital light processing. Heliyon10 (1), e24095. (2024). 10.1016/j.heliyon.2024.e24095 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Murat, S., Alp, G., Alatalı, C. & Uzun, M. In vitro evaluation of adhesion of Candida albicans on CAD/CAM PMMA-based polymers. J. Prosthodont.28 (2), e873–e879. (2019). 10.1111/jopr.12942 PubMed PMID: 29962017. [DOI] [PubMed] [Google Scholar]
- 63.Sturz, C. R. C., Faber, F. J., Scheer, M., Rothamel, D. & Neugebauer, J. Effects of various chair-side surface treatment methods on dental restorative materials with respect to contact angles and surface roughness. Dent. Mater. J.34 (6), 796–813. (2015). 10.4012/dmj.2014-098 PubMed PMID: 26632228. [DOI] [PubMed] [Google Scholar]
- 64.Nassar, U., Meyer, A. E., Ogle, R. E. & Baier, R. E. The effect of restorative and prosthetic materials on dental plaque. Periodontol 2000. 8, 114–124. (1995). 10.1111/j.1600-0757.1995.tb 00049.x PubMed PMID: 9567950. [DOI] [PubMed] [Google Scholar]
- 65.Hu, X., Wang, T., Li, F. & Mao, X. Surface modifications of biomaterials in different applied fields. RSC Adv.13 (30), 20495–20511. (2023). 10.1039/d3ra02248j PubMed PMID: 37435384; PubMed Central PMCID: PMC10331796. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Hoshiba, T., Yoshikawa, C. & Sakakibara, K. Characterization of initial cell adhesion on charged polymer substrates in serum-containing and serum-free media. Langmuir34 (13), 4043–4051. (2018). 10.1021/acs.langmuir.8b00233 PubMed PMID: 29544251. [DOI] [PubMed] [Google Scholar]
- 67.Daeschel, M. A. & McGuire, J. Interrelationships between protein surface adsorption and bacterial adhesion. Biotechnol. Genet. Eng. Rev.15 (1), 413–438. (1998). 10.1080/02648725.1998.10647964 [DOI] [PubMed] [Google Scholar]
- 68.Penzenstadler, M. et al. In vitro analysis of composition profiles of resins for 3D printing of dentures. J. Dent.154, 105565. (2025). 10.1016/j.jdent.2025.105565 [DOI] [PubMed] [Google Scholar]
- 69.Artopoulou I. I., Pani, E. & Eliades, G. Comparative laboratory study of 3D printed denture base resin materials. J. Prosthet. Dent.134 (5), 1964e1–1964e9. (2025). 10.1016/j.prosdent.2025.08.007 [DOI] [PubMed] [Google Scholar]
- 70.Goiato, M., Freitas, E., Dos Santos, D., De Medeiros, R. & Sonego, M. Acrylic resin cytotoxicity for denture base: literature review. Adv. Clin. Exp. Med.24 (4), 679–686. (2015). 10.17219/acem/33009 [DOI] [PubMed] [Google Scholar]
- 71.Dai, J. et al. Post-processing of DLP-printed denture base polymer: impact of a protective coating on the surface characteristics, flexural properties, cytotoxicity, and microbial adhesion. Dent. Mater.38 (12), 2062–2072. (2022). 10.1016/j.dental.2022.11.008 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
Data are provided within the manuscript or supplementary information files.