Abstract
Objective
This study seeks to assess the impact of quantum iterative reconstruction (QIR) on objective measures of contrast-to-noise ratio (CNR), lumen diameter, and stent strut thickness using photon-counting detector computed tomography (PCD-CT) in ultra-high resolution mode (UHR) in a coronary phantom.
Materials and methods
A three-dimensional printed phantom (four parallel vessels; 3 vessels with varying 3.5-mm stent types) was applied. Demineralized water, mixed with iodinated contrast, was circulated through the phantom by a dynamic pump. Images were acquired on a commercially available PCD-CT system using retrospective helical acquisition in UHR mode at both 120 and 140 kVp. Images were reconstructed at four QIR-levels (1‒4) using three sharp vascular kernels (Bv56u, Bv68u, and Bv76u). Quantitative evaluations of lumen diameter (mm) and stent strut thickness (mm) were assessed by attenuation profiles using an in-house MATLAB script.
Results
Increasing QIR level from 1 to 4 improved CNR from 1.5 (95% confidence interval [CI] 1.31‒1.70) (QIR-level 1) to 3.33 (95% CI: 2.89‒3.77) (QIR-level 4), p < 0.001. There was no change (QIR-level 1 to 4) in lumen diameter, range 2.95 mm (95% CI: 2.89‒3.02) to 2.97 mm (95% CI: 2.91‒3.03) (p = 0.890) or strut thickness, range 0.62 mm (95% CI: 0.59‒0.65) to 0.61 mm (95% CI: 0.58‒0.65 ((p = 0.870), regardless of contrast concentration or reconstruction kernel.
Conclusion
Increasing QIR-level in UHR mode PCD-CT in a stented coronary phantom improved CNR without affecting objective measures of lumen diameter or stent strut thickness under controlled phantom conditions.
Relevance statement
There has been debate regarding the negative impact of iterative reconstruction on spatial resolution, potentially increasing blooming artifacts caused by coronary stents, which is of special concern when imaging small stent lumens. Currently, no studies have assessed the effect of QIR on the objective assessment of coronary stents.
Key Points
Some earlier implementations of iterative reconstruction negatively impacted spatial resolution, raising concerns for small-vessel and stent imaging.
No prior studies examined the effect of QIR on coronary stent assessment.
The effect of QIR was evaluated at four levels (1‒4) in a stented coronary phantom.
Increasing QIR to the highest level significantly improved luminal CNR.
Objective measures of stent dimensions remained unchanged across all QIR levels.
Graphical Abstract
Keywords: Artifacts, Computed tomography angiography, Coronary arteries, Phantoms (imaging), Stents
Background
Current American Heart Association guidelines recommend limiting routine use of coronary computed tomography angiography (CTA) to assessment of coronary stents ≥ 3 mm in diameter due to partial volume and beam-hardening artifacts negatively affecting diagnostic confidence for evaluating small caliber stents [1]. With recent advancements in scanner technology by photon-counting detector computed tomography (PCD-CT), spatial resolution has been markedly improved compared to conventional energy-integrating detector CT scanners (150‒200 µm versus 400‒450 µm, respectively) [2, 3]. PCD-CT allows for reconstruction in both standard-resolution mode with spectral capabilities (slice thickness 0.4 mm) and ultra-high resolution (UHR) mode with increased spatial resolution (slice thickness 0.2 mm) at the expense of spectral information. For coronary stent assessment, UHR mode has been shown to provide improved lumen visualization compared to standard-resolution with spectral reconstructions [4]. In the context of coronary CTA, UHR acquisition translates to improved visualization of vessel lumen in the presence of calcifications and metallic stents [5, 6], primarily by reducing blooming artifacts attributable to partial volume effects [7].
However, improved spatial resolution and high sharpness kernels leads to higher image noise [8]. Iterative reconstruction algorithms can improve the signal-to-noise ratio in CTA, but there has been concerns regarding a negative impact on spatial resolution [9], which is undesirable when imaging small stent lumens due to a potential increase in stent blooming. The PCD-CT system (NAETOM Alpha, Siemens Healthineers), employs “Quantum Iterative Reconstruction” (QIR) with levels ranging from 1 to 4. QIR is a model-based iterative reconstruction algorithm for PCD-CT tailored to handle the increased complexity of energy-resolved spectral data. It has previously been demonstrated that QIR level 4 provides the optimal image sharpness for non-stented CTA images [6], however, this evaluation was performed using medium-smooth resolution vascular kernels and virtual monoenergetic reconstructions of standard-resolution PCD-CT, not optimized for stent visualization. Similarly, the effect of increasing QIR-level was tested in abdominal imaging with virtual monoenergetic imaging reconstructed at 2 mm with the Br36 kernel, showing improved image quality and reduced noise without a negative impact on image texture or attenuation values [10]. One study evaluated the effect of QIR-level in low-dose UHR imaging of lung parenchyma, and found a slight decrease in spatial resolution with increasing QIR-level [11]. The authors reported QIR-level 3 to provide optimal subjective image sharpness when reconstructing images at a high sharpness kernel (Bl64).
No previous studies have evaluated the effect of reconstruction at different QIR-levels in UHR PCD-CT for assessment of high-attenuation coronary stents. Therefore, the aim of the present study was to explore whether different QIR levels affected objective measures of contrast-to-noise ratio (CNR), lumen diameters and stent strut thickness in UHR scan mode at varying reconstruction kernels, contrast concentrations, and tube voltages.
Methods
Ethical considerations
Ethical approval and informed consent were not required for this study, as it was a phantom experiment and did not involve human participants or patient data.
Phantom and experimental setup
A three-dimensional printed vessel phantom containing four parallel vessels with an internal diameter of 3.5 mm and a wall thickness of 1 mm was prepared (Fig. 1). Three commercially available coronary stents (Evermine50, Meril Life Sciences Pvt. Ltd., Vapi, India; Xience Pro, Abbott Vascular; and Synergy, Boston Scientific) with a diameter of 3.5 mm were deployed in independent vessels with one vessel being left non-stented serving as a reference vessel. Stent characteristics are summarized in Table 1.
Fig. 1.

Three dimensional-printed vessel phantom. Computer-generated image of the 3D-printed phantom containing four parallel vessels, each with an internal diameter of 3.5 mm and a wall thickness of 1 mm. 3D, Three-dimensional
Table 1.
Characteristics of the coronary stents deployed in the vessel phantom
| Model | Manufacturer | Diameter at 16 atm (mm) | Length (mm) | Strut thickness (µm) | Material | Drug eluting |
|---|---|---|---|---|---|---|
| Evermine50 | Meril Life Sciences | 3.5/3.81 | 13 | 50 | Cobalt- chromium | Yes |
| Xience Pro | Abbott Vascular | 3.5/3.70 | 23 | 81 | Cobalt-chromium | Yes |
| Synergy | Boston Scientific | 3.5/3.79 | 24 | 79 | Platinum-chromium | Yes |
These three coronary stents (3.5 mm nominal diameter) from different manufacturers were expanded at 16 atm in the three-dimensional-printed phantom. Two stents were made of cobalt-chromium alloy and one of platinum-chromium. The fourth, empty phantom vessel served as a reference vessel
The phantom was connected to a pump capable of mimicking physiologic conditions through adjustable heartrate and stroke volume (SuperPump, ViVitro Labs). The pump was set at a frequency of 60 beats per min and a stroke volume of 75 mL, resulting in a cardiac output of 4.5 L/min. Slight movement along the x and y axes was introduced by the pulsatile flow generated by the pump, mimicking coronary movement. The pump circulated demineralized water deprived of air bubbles, mixed with 350 mg/mL iodinated contrast (Omnipaque, GE Healthcare). Contrast concentrations were adjusted by stepwise dilution to achieve target luminal attenuations of approximately 300, 400 and 500 HU (contrast concentrations 1, 2, and 3), as verified on reconstructed images. Three contrast concentrations were included to assess potential interactions between contrast attenuation and the effect of QIR-level. To achieve realistic attenuation properties, the phantom was placed in a cylindrical tank (diameter 32 cm) filled with water. The tank was placed with the phantom at the isocenter of the scanner.
CT acquisition and reconstruction parameters
All images were acquired on a first-generation dual-source PCD-CT system (NAEOTOM Alpha, Siemens Healtineers), software version VA50. Image data were acquired by retrospectively electrocardiographically-gated helical, UHR mode (collimation 120 × 0.2 mm), using an onboard synthetic electrocardiogram-generator to provide gating signal at 60 beats. Tube current was fixed at 99 mAs for all acquisitions. For each of the three concentrations of contrast media, three repeated scans were performed to account for variations in movement introduced by the pulsatile flow in the vessel phantom. Acquisitions were performed at both 120 and 140 kVp, for a total of 18 image datasets. Axial images were reconstructed in the 75% phase with a slice thickness of 0.2 mm and a reconstruction increment of 0.15 mm. Images were reconstructed using a 768 × 768 matrix and a reconstruction field of view of 160 mm, resulting in an in-plane pixel size of 0.21 × 0.21 mm2, and thus near-isotropic resolution. No additional z-filtering or postprocessing beyond standard reconstruction and QIR was applied. The 768 × 768 matrix was selected to enable small in-plane pixel spacing and to fully exploit the increased spatial resolution afforded by UHR-acquisition.
Reconstructions were made using three sharp body-vascular kernels (Bv56u, Bv68u, and Bv76u). For each kernel, images were reconstructed at QIR levels 1–4 to span the full clinically available range of reconstruction strengths. QIR-level 0 (i.e., QIR disabled) was not included, as UHR imaging without iterative reconstruction is unlikely to be used clinically due to excessive noise. Three sharp body vascular kernels were chosen (Bv56u, Bv68u, and Bv76u) as these kernels represent the range most commonly applied for coronary stent imaging. All acquisition and reconstruction parameters are listed in Table 2.
Table 2.
Acquisition and reconstruction parameters
| Rotation time (ms) | 250 |
| Temporal resolution (ms) | 66 |
| Collimation (mm) | 24 |
| Acquisition mode | Retrospectively gated helical |
| Pixel size (mm) | 0.21 × 0.21 |
| Helical pitch | 0.20–0.25 |
| Tube voltage (kVp) | 120/140 |
| Tube current (mAs) | 99 (fixed) |
| Slice thickness (mm) | 0.2 |
| Slice increment (mm) | 0.15 |
| Image matrix | 768 × 768 |
| Kernel | Bv56u, Bv68u, Bv76u |
| Iterative reconstruction | QIR 1, 2, 3, and 4 |
Quantitative analysis
For quantitative evaluations of stent dimensions and CNR an in-house MATLAB (MathWorks Inc.) script was used. For each stent, attenuation intensity profiles were created by drawing a line through two opposing stent struts in a cross section of each stent (Fig. 2). Optimal placement was aided by an interactive plot to ensure crossing at maximum intensity values. The profile line was drawn across the same stent struts in all scans and automatically copied to the same exact location across all reconstructions per acquisition, thus eliminating reader-induced variability. One hundred sampled attenuation values were measured for each line, using bicubic interpolation, and physical distances calculated from spatial reference information embedded in the Digital Imaging and Communications in Medicine‒DICOM metadata.
Fig. 2.

Measurement of stent attenuation profiles. a Cross-sectional stent image with a measurement line drawn through two opposing struts. b Corresponding attenuation profile along the line, generated using an in-house MATLAB script. Lumen diameter was defined as the inner distance at FWHM of lumen attenuation (blue arrow), and strut thickness as the average width at FWHM of the two strut peaks (red arrows). Stent diameter was defined by the distance between the two attenuation peaks. FWHM, Full width at half maximum
Inner stent diameter (lumen diameter) was defined as the inner distance between the stent attenuation profile at full width at half maximum, of lumen HU. The term stent diameter refers to the outer stent diameter, defined as the distance from the peak HU-values from each opposing stent strut. Strut thickness was defined as the average of the two HU-profile widths at full width at half maximum.
Image noise was measured in the phantom by placing three region of interest (ROI) with an area of 300 mm2 in the water tank and was defined as the average standard deviation of attenuation in the ROIs. Circular ROIs were drawn in the lumen of the stented segments in the phantom vessels on all performed scans, and the size, shape and position of ROIs and attenuation line were automatically kept constant between the series. CNR was calculated as:
Statistical analysis
Continuous outcome variables were described using means and compared using Wald test. Normality of distribution for each outcome was assessed using visual inspection of histograms and Q–Q plots. A log transformation was applied to CNR values to meet the assumptions of normality required, and model estimates were exponentiated to obtain geometric means and corresponding 95% confidence intervals (CI). A linear mixed-effects regression model was employed to evaluate the effect of QIR on each outcome variable (CNR, lumen diameter, strut thickness), with QIR, reconstruction kernel, contrast concentration, and tube voltage treated as a fixed effects. A random intercept was included to account for repeated measurements within scan repeats.
Interaction terms between QIR and covariates were added individually to assess potential effect modification. Given the exploratory nature of the study and the strong a priori hypotheses regarding the direction of QIR effects, no formal correction for multiple comparisons was applied. Results were interpreted with emphasis on effect sizes and consistency across reconstruction settings. A two-tailed p-value < 0.05 was considered statistically significant. All statistical analyses were conducted using STATA 18BE (StataCorp).
Results
A total of 216 reconstructions were analyzed across three stent types, three contrast concentrations, two tube voltages (120 and 140 kVp), three reconstruction kernels, and four QIR levels (1–4).
Image noise and attenuation
Mean image noise in the background water phantom was 10.99 ± 0.60 HU (mean ± standard deviation) across all scans. Mean luminal HU for the three contrast concentrations were 339 ± 5, 389 ± 5, and 547 ± 4, respectively.
CNR
CNR increased with increasing QIR level from 1.5 (95% CI: 1.31–1.70) to 3.33 (95% CI: 2.89–3.77) for QIR-level 1 and 4, respectively, p < 0.001 (Fig. 3a). CNR decreased with higher sharpness kernels (Bv56u, Bv68u and Bv76u): 3.85 (95% CI: 3.50–4.19), 1.93 (95% CI: 1.75–2.10) and 1.37 (95% CI: 1.25–1.49), respectively, p < 0.001 (Fig. 3b).
Fig. 3.
Effect of reconstruction settins on CNR. Measurements represent mean measurements across all three stents. a CNR increased progressively with higher QIR levels, from 1.5 (95% CI: 1.31–1.70) at QIR-1 to 3.33 (95% CI: 2.89–3.77) at QIR-4 (p-value for trend < 0.001). b CNR decreased with sharper kernels, from 3.85 (95% CI: 3.50–4.19) at Bv56u to 1.37 (95% CI: 1.25–1.49) at Bv76u. CNR, Contrast-to-noise ratio; QIR, Quantum iterative reconstruction
There was a significant interaction between QIR-level and kernel, with the increase in CNR by employing QIR-level 4 versus 1 being more pronounced at lower sharpness kernels: +2.98 (95% CI + 2.34; + 3.63) and +1.99 (95% CI + 1.66; + 2.31), for kernels Bv56u and Bv76u, respectively, p = 0.007 (Fig. 4). The effect of QIR-level 4 versus 1 on CNR did not change significantly with increasing contrast concentration (+ 1.38 [95% CI: 0.79–1.98] and +2.40 [95% CI: 1.42–3.39], p = 0.083 at contrast concentrations 1 and 3, respectively) indicating no interaction between QIR strength and contrast attenuation (Fig. 5). There was no impact of tube voltage on CNR when pooling across all QIR levels (2.60 [95% CI: 2.32–2.87] and 2.56 [95% CI: 2.26–2.85], p = 0.846), for 120 and 140 kVp, respectively.
Fig. 4.

Interaction between QIR-level and reconstruction kernel on CNR. CNR plotted across QIR levels 1–4 for kernels Bv56u, Bv68u, and Bv76u. A significant interaction was observed (p = 0.007), with the CNR difference between Bv56u and Bv76u being greater at QIR-4 than at QIR-1 (-2.98 versus -1.99). CNR, Contrast-to-noise ratio; QIR, Quantum iterative reconstruction
Fig. 5.

Interaction between QIR-level and contrast concentration on CNR. CNR plotted across QIR levels 1–4 for three contrast concentrations. The positive effect of QIR level 4 versus 1 did not change significantly with increasing contrast concentration, from 1.38 (95% CI: 0.79–1.98) at contrast concentration 1 to 2.40 (95% CI: 1.42–3.39) at contrast concentration 3 (p-value for interaction = 0.083). CNR, Contrast-to-noise ratio; QIR, Quantum iterative reconstruction
Quantitative measurement of strut thickness and lumen diameter
Lumen diameter and strut size did not differ significantly across QIR levels. Lumen diameter ranged from 2.95 mm (95% CI: 2.89–3.02 mm) to 2.97 mm (95% CI: 2.91–3.03 mm), p = 0.890 (Fig. 6a), and strut thickness from 0.62 mm (95% CI: 0.59–0.65 mm) to 0.61 mm (95% CI: 0.58–0.65 mm), p = 0.870 (Fig. 6b). Higher sharpness kernels (Bv76u versus Bv56u) increased the luminal diameter (mean 3.07 mm [95% CI: 3.00‒3.14] versus 2.83 [95% CI: 2.77‒2.90], p < 0.001) and decreased strut thickness (mean 0.52 mm [95% CI: 0.49‒0.55] versus 0.73 [95% CI: 0.70‒0.75], p < 0.001), respectively (Fig. 7). There was no interaction between QIR-level and kernel, contrast concentration or kVp for measures of lumen diameter or strut thickness (Fig. 5). Similarly, no significant differences were observed across kVp levels: mean luminal diameter was 2.96 mm (95% CI: 2.92–3.01) at 120 kVp versus 2.98 mm (95% CI: 2.94–3.03) at 140 kVp (p = 0.540), and mean strut thickness was 0.62 mm (95% CI: 0.60–0.64) at 120 kVp versus 0.60 mm (95% CI: 0.58–0.63) at 140 kVp (p = 0.408). Measured stent strut thickness values were substantially larger than manufacturer-specified nominal strut thicknesses (50–80 µm) across all reconstruction settings, consistent with expected blooming effects. Compared with in vitro reference values, CT-based measurements showed a mean underestimation of the outer stent diameter of 2.3%. At the sharpest reconstruction kernel (Bv76u), the mean visible lumen diameter corresponded to 82.0% of the in vitro reference outer stent diameter.
Fig. 6.
Effect of QIR-level on stent dimensions. Measurements are shown pooled across stents to illustrate relative reconstruction-dependent trends; individual stent sizes differed slightly but demonstrated consistent behavior across QIR levels and kernels. a Strut thickness and (b) lumen diameter plotted across QIR levels 1–4. No significant differences were observed: lumen diameter ranged from 2.95 mm (95% CI: 2.89–3.02) to 2.97 mm (95% CI: 2.91–3.03) (p = 0.890), and strut thickness from 0.62 mm (95% CI: 0.59–0.65) to 0.61 mm (95% CI: 0.58–0.65) (p = 0.870). QIR, Quantum iterative reconstruction
Fig. 7.
Effect of reconstruction kernel on stent dimensions across QIR levels. a Lumen diameter increased significantly with sharper kernels, from 2.83 mm (95% CI: 2.77–2.90) at Bv56u to 3.07 mm (95% CI: 3.00–3.14) at Bv76u (p < 0.001). b Strut thickness decreased significantly with kernel sharpness, from 0.73 mm (95% CI: 0.70–0.75) at Bv56u to 0.52 mm (95% CI: 0.49–0.55) at Bv76u (p < 0.001). No significant interaction with QIR-level was observed. Abb, QIR, Quantum iterative reconstruction
Discussion
This phantom study evaluated the impact of varying levels of QIR on CNR and quantitative assessment of coronary stent and lumen dimensions, using UHR PCD-CT angiography across three high sharpness kernels, three contrast concentrations, and two tube voltages. It was observed that increasing QIR levels significantly improved CNR without compromising relative measurements of stent strut thickness or lumen diameter, whilst sharper reconstruction kernels substantially influenced stent visualization by reducing measures of stent strut thickness and increasing visible luminal diameter. Trends in apparent stent diameter were consistent across the three investigated stents, supporting pooled analysis of reconstruction-dependent effects.
In the present study, we selected the UHR mode to evaluate the impact of different QIR levels, as the thinner slice thicknesses by the UHR mode are associated with markedly increased image noise. Similarly, we included only high-sharpness reconstruction kernels, known to be most effective for stent assessment [8, 12], despite their tendency to amplify image noise at higher levels.
Our results are in line with a previous study by Vecsey-Nagy et al, demonstrating that increasing QIR in UHR PCD-CT does not affect objective or subjective image sharpness, but reduces image noise, while stenosis measurements are unaffected in non-stented patients [13]. A similar conclusion has been made in PCD-CT angiographies of the lower leg [14].
Our results align with previous work demonstrating that sharper kernels improve stent lumen visualization by reducing blooming artifacts, as seen by a significant reduction in measured strut thickness and increased lumen diameter at higher sharpness kernels [8, 12, 15]. However, the differences demonstrated are in the order of one to two pixels, the clinical relevance of which will have to be weighed against CNR reduction in higher kernels. Preferably, image reading should be performed using different kernels to leverage both high spatial and contrast resolution. Based on the present findings, an intermediate-to-high sharpness reconstruction kernel such as Bv68u appears to provide the balanced trade-off for ultra-high-resolution coronary stent imaging.
In a retrospective in vivo study, Geering et al evaluated the effect of various reconstruction parameters on stent visualization, and found the best overall subjective image quality at 0.2 mm/Bv72 reconstructions [15]. Of note, they find a 20‒30% underestimation of the inner stent lumen compared to in-vitro conditions when reconstructing at Bv72 and Bv56, which is comparable to our findings in the current phantom setup.
Automated tube voltage selection works by selecting the combination of tube voltage and current that produces the prescribed image quality at the lowest radiation dose, potentially reducing the risk of stochastic damage associated with CT. The fact that lumen diameters and CNR were unaffected by the tube voltages investigated (120 and 140 kVp), suggests that automated tube voltage selection may be used in UHR-scanning of stents within the limited kVp-range allowed for the UHR scanmode, with no concerns regarding measurement accuracy, although extrapolation to in-vivo conditions should be made with caution. Due to the very high attenuation of strut material, this also supports the hypothesis that blooming artifacts are, primarily, caused by partial volume effects [7], rather than beam hardening. Definitive conclusions regarding underlying mechanisms for stent blooming are, however, beyond the scope of this phantom study, and should be investigated in a separate study.
Some important limitations of this study should be emphasized. First, only 3.5-mm stents were evaluated. Although reconstruction-related blooming effects are expected to be independent of nominal stent diameter for a given stent design, the relative impact on visible lumen area will be greater in smaller stents. Therefore, extrapolation of the present findings to smaller stent diameters should be made with caution. Of note, a recent prospective clinical study of 79 stented patients referred for ICA, sensitivity and specificity for identifying in-stent restenosis > 50% was 95% and 84%, respectively [16]. A subgroup analysis did not identify significant differences in sensitivity or specificity based on stent size. Second, as nominal coronary stent strut thicknesses (50‒80 µm) are substantially below the spatial resolution of clinical PCD-CT (200 µm), absolute validation of strut dimensions is not feasible without a physical ground truth (e.g., micro-CT), and our findings therefore reflect relative reconstruction-dependent differences. Third, given the nature of this phantom study, factors inherent to in vivo coronary imaging, including cardiac motion, vessel tortuosity, calcified plaque, and heterogeneous surrounding tissues, were not represented in the present model and may influence both image quality and apparent stent dimensions. For a complete and rigorous characterization of spatial resolution metrics, line-pair measurements, edge-based modulation transfer function, or more appropriately, task transfer function [17] could have been employed, but would require a different phantom setup outside the scope of our study. Furthermore, our measurement method, although simplistic, provides a clinically relevant and easy-to-interpret result for reading radiologists. Full width at half maximum‒FWHM provides an indirect quantification of blooming artifacts, having been used for that purpose in simulated coronary calcifications [18]. However, the inclusion of full-width, tenth-maximum measurements could have been informative in the context of accuracy for the detection of neointimal hyperplasia with relatively low density.
In the present study, only a single maximum allowable tube current was used. This highlights one of the challenges of using UHR, namely that the available tube effect is halved and scan times are increased compared to standard-mode PCD-CT. This also accounts for the relatively low CNR values achieved. The rule of thumb that CNR should be at least five, for reliable object detection [19], suggests that vessel to background tissue delineation could be impaired at the settings used in our study. Notably, a CNR above five was only achievable when applying QIR-level 4 to images reconstructed at Bv56u. Other studies have suggested using UHR mode for acquisition [20], but reconstruct at standard resolution (0.4 mm) to recover CNR, while taking advantage of the increase in resolution conferred by the small focal spot. This was, however, beyond the scope of our investigation. Most importantly, phantom studies are, by their nature, of limited generalizability, and our results should be tested in an in vivo setting. Such a study can also shed light on the effect of QIR and kernels on visibility and quantification of in-stent lesions.
In addition, we only performed a quantitative analysis, which did not measure reader confidence in the assessment of stent lumens; however, CNR is often used as a surrogate marker of image quality. As suggested by the Rose model [19], improving CNR increases the sensitivity of the human eye. Likewise, noise structure measurements (e.g., noise-power spectra) were not analyzed. Optimizing image quality may not only apply to the assessment of stents, but also to the evaluation of the degree of stenosis and high-risk plaque features in non-stented patients.
In conclusion, our results suggest that increasing QIR-level in UHR mode PCD-CT of coronary arteries increases CNR without affecting objective measures of lumen diameter or stent strut thickness under the investigated phantom conditions.
Acknowledgements
We thank Rikke Beese Dalby, Head of the Department of Radiology, University Hospital of Southern Denmark, Esbjerg, for providing scanner access. We also thank Allan Erik Rohold, Chief of Cardiology, and Kristian Korsgaard Thomsen, Chief of Cardiology Research, Department of Cardiology, University Hospital of Southern Denmark, Esbjerg, for their financial and institutional support. No large language models (LLMs) were used in the preparation of data, data analysis, or interpretation of results. An LLM was used solely for stylistic improvements during final manuscript preparation. The authors reviewed all content and take full responsibility for the final manuscript.
Abbreviations
- CI
Confidence interval
- CNR
Contrast-to-noise ratio
- CT
Computed tomography
- CTA
Computed tomography angiography
- PCD-CT
Photon-counting detector computed tomography
- QIR
Quantum iterative reconstruction
- ROI
Region of interest
- UHR
Ultra-high resolution
Author contributions
The initiative for the study was taken by STS and MWK. The study design and experimental setup were a collaboration between STS, MWK, HS, LG, AD, HPA, MBH, NPRS, and KTM. The study phantom was created by HS, AD, HPA, and MBH. Experiments were carried out by STS, MWK, and LG. Stent implantation in the phantom was performed by LOJ. Data analysis was planned and performed by STS. The manuscript was prepared by STS, LG, NPRS, KTM, and LOJ.
Funding
The study was funded by the Department of Cardiology, University Hospital of Southern Denmark. Open access funding provided by University of Southern Denmark.
Data availability
The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.
Declarations
Ethics approval and consent to participate
Not applicable. As this phantom study involved no patient data or biological material, ethical approval and informed consent were not required.
Consent for publication
As this phantom study involved no patient data or biological material, ethical approval and informed consent were not required.
Competing interests
The authors declare that they have no competing interests.
Footnotes
Publisher’s Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Gulati M, Levy PD, Mukherjee D et al (2021) 2021 AHA/ACC/ASE/CHEST/SAEM/SCCT/SCMR guideline for the evaluation and diagnosis of chest pain: executive summary: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation 144:e336–e367. 10.1161/CIR.0000000000001030 [DOI] [PubMed] [Google Scholar]
- 2.Flohr T, Schmidt B (2023) Technical basics and clinical benefits of photon-counting CT. Invest Radiol 58:441–450. 10.1097/rli.0000000000000980 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Willemink MJ, Persson M, Pourmorteza A et al (2018) Photon-counting CT: technical principles and clinical prospects. Radiology 289:293–312. 10.1148/radiol.2018172656 [DOI] [PubMed] [Google Scholar]
- 4.Hagar MT, Emrich T, Vecsey-Nagy M et al (2025) Spectral ultrahigh-resolution photon-counting CT for coronary stent imaging: evaluation in a dynamic phantom. Eur Radiol Exp 9:115. 10.1186/s41747-025-00654-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Hagar MT, Soschynski M, Saffar R et al (2024) Ultra-high-resolution photon-counting detector CT in evaluating coronary stent patency: a comparison to invasive coronary angiography. Eur Radiol 34:4273–4283. 10.1007/s00330-023-10516-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Sartoretti T, McDermott M, Mergen V et al (2023) Photon-counting detector coronary CT angiography: impact of virtual monoenergetic imaging and iterative reconstruction on image quality. Br J Radiol 96:20220466. 10.1259/bjr.20220466 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Pack JD, Xu M, Wang G et al (2022) Cardiac CT blooming artifacts: clinical significance, root causes and potential solutions. Vis Comput Ind Biomed Art 5:1. 10.1186/s42492-022-00125-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Michael AE, Schoenbeck D, Becker-Assmann J et al (2023) Coronary stent imaging in photon counting computed tomography: optimization of reconstruction kernels in a phantom. Eur J Radiol 166:110983. 10.1016/j.ejrad.2023.110983 [DOI] [PubMed] [Google Scholar]
- 9.Toia P, La Grutta L, Sollami G et al (2020) Technical development in cardiac CT: current standards and future improvements—a narrative review. Cardiovasc Diagn Ther 10:2018–2035. 10.21037/cdt-20-527 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Sartoretti T, Landsmann A, Nakhostin D et al (2022) Quantum iterative reconstruction for abdominal photon-counting detector CT improves image quality. Radiology 303:339–348. 10.1148/radiol.211931 [DOI] [PubMed] [Google Scholar]
- 11.Sartoretti T, Racine D, Mergen V et al (2022) Quantum iterative reconstruction for low-dose ultra-high-resolution photon-counting detector CT of the lung. Diagnostics (Basel) 12:522. 10.3390/diagnostics12020522 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Decker JA, O’Doherty J, Schoepf UJ et al (2023) Stent imaging on a clinical dual-source photon-counting detector CT system: impact of luminal attenuation and sharp kernels on lumen visibility. Eur Radiol 33:2469–2477. 10.1007/s00330-022-09283-4 [DOI] [PubMed] [Google Scholar]
- 13.Vecsey-Nagy M, Varga-Szemes A, Schoepf UJ et al (2024) Ultra-high-resolution coronary CT angiography on photon-counting detector CT: bi-centre study on the impact of quantum iterative reconstruction on image quality and accuracy of stenosis measurements. Eur J Radiol 176:111517. 10.1016/j.ejrad.2024.111517 [DOI] [PubMed] [Google Scholar]
- 14.Graafen D, Bart W, Halfmann MC et al (2024) In vitro and in vivo optimized reconstruction for low-keV virtual monoenergetic photon-counting detector CT angiography of lower legs. Eur Radiol Exp 8:89. 10.1186/s41747-024-00481-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Geering L, Sartoretti T, Mergen V et al (2023) First in-vivo coronary stent imaging with clinical ultra high resolution photon-counting CT. J Cardiovasc Comput Tomogr 17:233–235. 10.1016/j.jcct.2023.02.009 [DOI] [PubMed] [Google Scholar]
- 16.Sharma SP, van der Bie J, Ziedses des Plantes AC et al (2025) Diagnostic accuracy of ultra-high-resolution photon-counting detector computed tomography for detecting coronary in-stent restenosis. JACC Cardiovasc Imaging 18:1405–1407. 10.1016/j.jcmg.2025.07.009 [DOI] [PubMed] [Google Scholar]
- 17.Samei E, Bakalyar D, Boedeker KL et al (2019) Performance evaluation of computed tomography systems: summary of AAPM task group 233. Med Phys 46:e735–e756. 10.1002/mp.13763 [DOI] [PubMed] [Google Scholar]
- 18.Park D, Park EA, Jeong B et al (2024) Quantitative analysis of blooming artifact caused by calcification based on X-ray energy difference using computed tomography. Sci Rep 14:61187. 10.1038/s41598-024-61187-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Rose A (1948) The sensitivity performance of the human eye on an absolute scale. J Opt Soc Am 38:196–208. 10.1364/josa.38.000196 [DOI] [PubMed] [Google Scholar]
- 20.Fix Martinez M, Klein L, Maier J et al (2023) Potential radiation dose reduction in clinical photon-counting CT by the small pixel effect: ultra-high resolution (UHR) acquisitions reconstructed to standard resolution. Eur Radiol 34:4484–4491l. 10.1007/s00330-023-10499-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.