Abstract
Background
Accurate pre-operative assessment of axillary lymph node status is key for planning surgeries in early and newly diagnosed breast cancer. Ultrasound (US) and magnetic resonance imaging (MRI) are widely used, but their performance is inconsistent across studies. This meta-analysis evaluates and compares the predictive performance of MRI and US for pre-operative lymph node metastasis in early and newly diagnosed breast cancer.
Methods
We searched PubMed/MEDLINE, Embase, Scopus, and Web of Science (inception-31 December 2025) for studies on diagnostic performance in preoperative axillary staging using ultrasound and/or MRI. Two independent reviewers extracted or reconstructed true positive, true negative, false positive, false negative, and applied QUADAS-2. Sensitivity and specificity were pooled with random-effects GLMM, with direct comparisons and prevalence-based predictive values.
Results
Twenty-nine studies (n = 8,544) were included. Pooled sensitivity/specificity were 61.1%/92.0% for MRI and 59.8%/90.3% for ultrasound, yielding LR + 7.64 vs. 6.19 and DOR 18.1 vs. 13.9. HSROC AUCs were 0.863 (MRI) and 0.843 (ultrasound). In paired cohorts (k = 12), sensitivity did not differ, and MRI showed a non-significant specificity advantage.
Conclusion
MRI and ultrasound showed comparable, moderate sensitivity and high specificity for preoperative axillary metastasis. MRI tended toward higher specificity and positive predictive value, whereas sensitivity did not differ in paired studies. Imaging can guide triage, but histopathology remains necessary for definitive staging.
Supplementary Information
The online version contains supplementary material available at 10.1007/s10238-026-02087-6.
Keywords: Early breast cancer, Axillary lymph node metastasis, Magnetic resonance imaging, Ultrasound, Preoperative axillary staging, Meta-analysis
Introduction
Breast cancer is the most diagnosed malignancy among women worldwide and is a leading cause of cancer-related mortality even with continued advances in screening, molecular classification, and treatment strategies [1]. Recent global statistics show that over 2.3 million women are diagnosed with breast cancer each year, with more than 600,000 related mortalities [2]. Earlier detection and having a multidisciplinary approach have improved survival, but disease prognosis still varies widely because of different tumor biologies and disease progression at the presentation [3]. Axillary lymph node (ALN) status is a strong predictor of recurrence and overall survival in the breast cancer, therefore accurate assessment of ALN is important for determining stage, choosing best treatment strategy regarding surgical interventions, systemic treatment, and radiotherapy [4, 5]. Axillary lymph node dissection (ALND) has historically been the gold standard for evaluating nodal metastasis. Also, it is clearly associated with higher morbidity such as nerve injuries, restricted mobility of shoulder, more pain and infection, and lymphedema [6]. To minimize these complications, sentinel lymph node biopsy (SLNB) has become the current standard of care for clinically node-negative early breast cancer and povides accurate staging while significantly reducing surgical morbidity [7]. On the other hand, SLNB is still an invasive procedure and may not be necessary for all patients, especially for patients whose nodal involvement probability is low. Furthermore, a proportion of patients still undergo unnecessary nodal surgery due to limitations in preoperative clinical evaluation [6, 8].
These issues have led to the development of non-invasive imaging tools capable of effectively predicting lymph node status before surgery; these tools present the potential to stratify patients and guide surgical decisions before an incision is made. Among various modalities, ultrasonography (US) and magnetic resonance imaging (MRI) are the most widely employed tools to evaluate nodal involvement in a preoperative setting [9]. Ultrasound is frequently used owing to its ready availability, low cost, absence of radiation, and ability to assess morphological features of axillary lymph nodes such as shape, cortical thickness, and hilum appearance [10]. MRI, on the other hand, provides superior soft-tissue contrast and offers thorough visualization of axillary anatomy. Functional sequences, such as diffusion-weighted imaging, may further improve the detection of metastatic nodes [11].
Despite the documented utility of US and MRI, their diagnostic performance varies substantially across studies. Reported sensitivities of ultrasound range from 26% to more than 80%, reflecting differences in operator skill, criteria for positive nodes, and patient populations [12]. While MRI generally demonstrates higher sensitivity than US, its specificity is variable, and access, cost, and standardization challenges remain significant. A better understanding of their predictive capabilities may inform clinical pathways that minimize invasive procedures and improve patient outcomes [13].
The application of preoperative ALN status assessment has increased to lessen the axillary surgery extension in early breast cancer. Still, the performance of US and MRI widely varies in the literature. Prior meta-analyses have generally evaluated each modality in isolation, applied inconsistent positivity thresholds, or included few paired cohorts, making it difficult to quantify the clinical trade-off between false positives and false negatives. Moreover, most syntheses stop at sensitivity and specificity and do not translate accuracy into predictive values across plausible pre-test risks, which limits bedside interpretability. To address these gaps, we performed an updated diagnostic performance meta-analysis to compare ultrasound and MRI using a consistent modeling framework, incorporating head-to-head within-study contrasts when available, and estimating prevalence-dependent predictive values to support patient-specific decision-making. We further provide head-to-head comparisons, scenario-based PPV/NPV estimates, and an assessment of small-study effects to clarify the clinical implications and robustness of the pooled diagnostic accuracy estimates.
Methods
We undertook this systematic evidence review based on best practices and reported the findings as per a checklist provided by PRISMA (the Preferred Reporting Items for Systematic Reviews and Meta-Analyses) [14].
Eligibility criteria
We included original, human diagnostic accuracy studies that evaluated preoperative axillary nodal status using ultrasound and/or MRI in patients with early or newly diagnosed breast cancer with either node positive or negative status, no metastasis, no history of breast cancer surgery, and when imaging is done in the preoperative window. The reference standard had to be histopathology (sentinel lymph node biopsy and/or axillary dissection), and studies needed to report sufficient data to derive a 2 × 2 table of true positive (TP), false positive (FP), false negative (FN), true negative (TN), or equivalent accuracy metrics. We excluded studies restricted to recurrent or metastatic disease, studies enrolling male breast cancer patients, and reports without an appropriate histopathologic reference or without extractable accuracy data.
Search strategy
We searched databases of published studies PubMed/MEDLINE, Scopus, Web of Science, and Embase from inception to 31 December 2025, and screened references to select studies that met our inclusion criteria at first. Database-specific strategies comprised of vocabulary and keywords for breast cancer, axillary lymph nodes, ultrasound/sonography, MRI, and preoperative assessment. The full search strings and results by database are provided in Table S1.
Inclusion and exclusion criteria
Eligible studies were those that used histopathologic examination, either sentinel lymph node biopsy or axillary lymph node dissection as the reference standard, and reported sufficient data to calculate diagnostic performance measures such as true/false positives and negatives, sensitivity, specificity, or predictive values. Prospective and retrospective clinical studies, including cohort, case-control, and cross-sectional diagnostic accuracy designs, were considered. Only original human research published in English with full-text availability was included.
We included studies enrolling women with newly diagnosed primary invasive breast cancer undergoing preoperative axillary imaging (ultrasound and/or MRI) prior to definitive surgery, with histopathology (SLNB and/or ALND) as the reference standard. Early breast cancer was operationally defined as non-metastatic (Stage I-III) disease when staging information was reported. Studies were excluded if they focused on recurrent breast cancer, in situ-only carcinoma, or patients primarily treated with neoadjuvant systemic therapy unless diagnostic accuracy data were extractable for pre-treatment imaging.
Study selection process
Records were de-duplicated and screened in two stages (title/abstract, then full text) by two independent reviewers. Disagreements were resolved by discussion or, when consensus was not reached, by adjudication by a third reviewer. Reasons for full-text exclusion were documented. A PRISMA flow diagram of study selection is shown in Fig. 1.
Fig. 1.
Flow diagram indicating the selection of articles
Data extraction
Two reviewers independently extracted study characteristics (design, setting, sample size), participant details, index-test, and the reference standard. For each modality within a study, we captured or derived TP/FP/FN/TN. When multiple thresholds were reported, we used the authors’ primary definition. When contingency tables were not explicitly presented, counts were reconstructed from reported sensitivity/specificity and complementary metrics, then verified by a second reviewer. Discrepancies at the title/abstract and full-text stages were resolved by discussion; if consensus was not reached, a third senior reviewer adjudicated, with final decisions and exclusion reasons documented. Inter-reviewer agreement was assessed using Cohen’s kappa (κ) at both stages and interpreted using Landis-Koch benchmarks.
Risk of bias in studies (Quality assessment)
Risk of bias and applicability concerns were assessed using QUADAS-2 across four domains (patient selection, index test, reference standard, and flow/timing) and the first three domains for applicability [15]. Each domain was rated as low, high, or unclear risk, guided by QUADAS-2 signaling questions. Assessments were performed independently by two reviewers, with disagreements resolved by consensus. Results were summarized using traffic-light plot (Fig. 2).
Fig. 2.
Risk of Bias Assessment
Statistical analysis
Diagnostic accuracy for each imaging modality was summarized using 2 × 2 contingency tables (true positive [TP], false positive [FP], false negative [FN], true negative [TN]) against histopathology as the reference standard, as prespecified in the eligibility criteria. Where studies did not explicitly report a full 2 × 2 table, cell counts were algebraically reconstructed from the reported sample size and available accuracy indices (e.g., sensitivity, specificity, accuracy, PPV/NPV), and internal consistency was checked against the published metrics (allowing for rounding), with verification by a second reviewer. Reconstructed per-study counts are provided in Table S2.
For each study and modality, sensitivity (TP/[TP + FN]) and specificity (TN/[TN + FP]) were calculated or extracted from reported data, with 95% confidence intervals derived using the Wilson method. Primary meta-analytic pooling of sensitivity and specificity was performed using a generalized linear mixed-effects model (GLMM) for proportions on the logit scale, producing random-effects pooled estimates with 95% confidence intervals. From the pooled sensitivity and specificity, we derived positive and negative likelihood ratios (LR + and LR−) and the diagnostic odds ratio (DOR) for each modality. Between-study heterogeneity was quantified using I² (reported alongside pooled estimates).
To improve clinical interpretability, pooled sensitivity and specificity were translated into PPV and NPV at prespecified pre-test probabilities (10%, 20%, and 30%) using standard Bayesian relationships; PPV/NPV curves across prevalence are shown in Figure S1 (MRI) and Figure S2 (ultrasound). For direct (within-cohort) comparisons, studies reporting both MRI and ultrasound were analyzed using within-study log-odds ratios for sensitivity and specificity (MRI vs. ultrasound), which were pooled using random-effects meta-analysis and presented as pooled odds ratios (Table 4).
Table 4.
Head-to-head comparison (MRI vs US), pooled odds ratios
| Outcome | k | Pooled OR (95CI) | I2 | tau2 | P overall |
|---|---|---|---|---|---|
| Sensitivity (MRI vs US) | 12 | 0.96 (0.61-1.51) | 0.8% | 0.3555 | 0.842 |
| Specificity (MRI vs US) | 12 | 1.27 (0.74-2.18) | 0.8% | 0.3809 | 0.351 |
Random-effects meta-analysis of within-study log-odds ratios comparing MRI vs ultrasound. Outcomes shown for sensitivity and specificity
We assessed small-study effects/publication bias using Deeks’ funnel-plot asymmetry test, implemented as a regression of ln(DOR) on 1/√(effective sample size) weighted by effective sample size. Diagnostic trade-offs were additionally summarized using HSROC curves (Rutter-Gatsonis) from the bivariate random-effects model (Reitsma), with AUC reported (Fig. 5). Study-level sensitivity and specificity are displayed as modality-specific plots.
Fig. 6.
HSROC (Rutter-Gatsonis) from the bivariate random-effects model (Reitsma). On the left, MRI, group x studies: 26 - sensitivity = 57.5% - specificity = 90.9% - AUC = 0.863. On the right, US, studies: 17 - sensitivity = 59.4% - specificity = 88.9% - AUC = 0.843
Results
Characteristics of included studies
Twenty-nine studies, published between 2004 and 2025, were selected for this study (total sample size = 8544). Of the 29 included studies, 11 (37.9%) were prospective and 18 (62.1%) were retrospective. Regarding imaging modality, 5 studies (17.2%) used ultrasound alone, 11 studies (37.9%) used MRI alone, and 13 studies (44.8%) employed a combined MRI + ultrasound approach, indicating substantial variability in diagnostic strategies. The studies originated from countries across Asia, Europe, and North America, demonstrating broad international interest in improving non-invasive preoperative assessment of axillary lymph node status in early breast cancer (Table 1).
Table 1.
Summary of articles included in the present study
| Study | Country | Design | Sample size | Modality | Age, Mean |
|---|---|---|---|---|---|
| Abe H (2013), [17] | USA | Retrospective | 50 | MRI, US | 59.9 |
| Ahn HS (2019), [18] | South Korea | Retrospective | 74 | MRI | 51.5 |
| Aktaş A (2022), [19] | Turkey | Retrospective | 336 | MRI, US | 50.1 |
| Arslan G (2016), [20] | Turkey | Prospective | 35 | MRI | 52.86 |
| Assing MA (2017), [16] | USA | Retrospective | 271 | MRI, US | 58 |
| Atallah D (2020), [21] | Lebanon | Retrospective | 169 | MRI | 48 |
| Barco I (2016), [22] | Spain | Prospective | 1,505 | MRI, US | 58.5 |
| Chayakulkheeree J (2019), [23] | Thailand | Retrospective | 131 | MRI | 51.33 |
| de Felice C (2014), [24] | Italy | Retrospective | 26 | MRI | 56 |
| Di Micco R (2025), [25] | Italy | Prospective | 246 | MRI | 56.3 |
| Ergul N (2015), [26] | Turkey | Prospective | 24 | MRI | 47 |
| Fernández A G (2011), [27] | Spain | Prospective | 507 | MRI, US | 58 |
| Hieken TJ (2013), [28] | USA | Retrospective | 988 | MRI, US | 61.1 |
| Jung NY (2015), [29] | South Korea | Retrospective | 105 | MRI | 53.3 |
| Kim SH (2017), [30] | South Korea | Retrospective | 147 | MRI | 49.2 |
| Motomura K (2015), [31] | Japan | Prospective | 70 | MRI | 55 |
| Panda SK (2019), [32] | India | Prospective | 40 | MRI, US | 55 |
| Riedel F (2021), [33] | Germany | Retrospective | 611 | US | NA |
| Rukanskiene D (2020), [34] | Lithuania | Prospective | 227 | US | 59 |
| Schumacher K (2021), [35] | USA | Retrospective | 349 | MRI, US | 63 |
| Stadnik TW (2006), [36] | Belgium | Retrospective | 10 | MRI | 56 |
| Upadhyaya VS (2019), [37] | Singapore | Retrospective | 605 | US | 56.6 |
| Valente SA (2012), [38] | USA | Retrospective | 244 | MRI, US | 58 |
| van Nijnatten TJA (2016)1, [39] | Netherlands | Retrospective | 377 | MRI, US | 58 |
| Wang J (2021), [40] | China | Prospective | 85 | US | 42.36 |
| Wang LJ (2025), [41] | Taiwan | Prospective | 247 | MRI, US | 52.38 |
| Xu Y (2018), [42] | China | Prospective | 92 | US | 51 |
| Yu Y (2021), [43] | China | Retrospective | 803 | MRI, US | 48.7 |
| Zhang H (2019), [44] | China | Retrospective | 170 | MRI, US | NA |
1Reported numbers for this study are calculated from the average of two MRI readers metrics
Risk of bias assessment
Risk of bias and applicability were evaluated using QUADAS-2 across patient selection, index test, reference standard, and flow/timing (Fig. 2). Overall study quality was mixed across the 29 included studies. Patient selection was the most frequent source of concern: only 4/29 studies were judged low risk, whereas 19/29 had some concerns and 6/29 were high risk, mainly because enrolment and exclusions were not fully described. Flow and timing showed the weakest profile (2/29 low risk; 17/29 some concerns; 10/29 high risk), suggesting that participant flow and the interval between imaging and histopathology were often insufficiently reported. By contrast, the index test and reference standard domains were more robust, with a larger proportion of studies rated low risk. Applicability concerns were generally low for the index test (22/29 low) and reference standard (23/29 low), while patient selection applicability was more variable (14/29 low).
Pooled performance
Both MRI and ultrasound showed moderate sensitivity and high specificity for preoperative axillary nodal assessment (Table 2; Figs. 3 and 4). In the primary random-effects GLMM, MRI achieved a pooled sensitivity of 61.1% (95% CI 48.3%−72.5%) and a pooled specificity of 92.0% (95% CI 88.3%−94.6%), corresponding to LR + 7.64, LR − 0.42, and a DOR of 18.1. Ultrasound yielded a pooled sensitivity of 59.8% (95% CI 49.0%−69.8%) and a pooled specificity of 90.3% (95% CI 84.6%−94.1%), with LR + 6.19, LR − 0.44, and a DOR of 13.9. Conventional inconsistency was minimal (I²=0.9% for both sensitivity and specificity). However, τ² (logit scale) and the 95% prediction intervals indicated that performance may vary across settings, particularly for sensitivity (MRI sensitivity PI 11.3%−95.1%, τ²=1.5769; MRI specificity PI 61.9%−98.8%, τ²=0.9525; ultrasound sensitivity PI 20.5%−89.6%, τ²=0.7504; ultrasound specificity PI 52.6%−98.7%, τ²=1.1070). Table 3
Table 2.
Study-level pooled diagnostic performance (sensitivity and specificity) for MRI and ultrasound (GLMM)
| Modality | Metric | Pooled | 95% CI | PI | I2 | τ² | LR+ | LR- | DOR |
|---|---|---|---|---|---|---|---|---|---|
| MRI | Sensitivity | 61.1% | 48.3% − 72.5% | 11.3% −95.1% | 0.9% | 1.5769 | 7.64 | 0.42 | 18.1 |
| MRI | Specificity | 92.0% | 88.3% − 94.6% | 61.9% − 98.8% | 0.9% | 0.9525 | |||
| US | Sensitivity | 59.8% | 49.0% − 69.8% | 20.5% − 89.6% | 0.9% | 0.7504 | 6.19 | 0.44 | 13.9 |
| US | Specificity | 90.3% | 84.6% − 94.1% | 52.6% − 98.7% | 0.9% | 1.1070 |
Pooled sensitivity and specificity (95% CIs) estimated via a GLMM meta-analysis of proportions on the logit scale (primary approach). LR+, LR-,PI, and DOR are derived from the pooled Se/Sp
Fig. 3.
Study-level sensitivity and specificity (MRI)
Fig. 4.
Study-level sensitivity and specificity (Ultrasound)
Table 3.
PPV and NPV at prespecified pre-test probabilities
| Prevalence | MRI PPV | MRI NPV | Ultrasound PPV | Ultrasound NPV |
|---|---|---|---|---|
| 10% | 45.9% | 95.5% | 40.7% | 95.3% |
| 20% | 65.6% | 90.4% | 60.7% | 90.0% |
| 30% | 76.6% | 84.7% | 72.6% | 84.0% |
PPV and NPV computed from pooled sensitivity and specificity (random-effects, logit scale). Values shown at 10%, 20%, and 30% pre-test probabilities
When translated to common clinical prevalence, MRI showed better PPVs compared to ultrasound at all given pre-test probabilities; on the other hand, NPVs were similar. For instance, at a 20% prevalence, MRI had a PPV of 65.6%, while ultrasound’s PPV was 60.7%. Regarding NPV at a 20% prevalence, the values were 90.4% and 90.0%, respectively. These patterns emphasize MRI’s superior rule-in performance without sacrificing the ability to rule out nodal disease in lower-risk settings (Table 4, Figures S1-S2).
Head-to-head comparison
Direct within-study contrasts reinforced these modality-level findings. Pooled odds ratios showed no difference in terms of sensitivity between MRI and ultrasound (OR 0.96; 95% CI, 0.61–1.51; p = 0.842; I²=0.8%). Regarding specificity, MRI had a better specificity, but the difference was non-significant (OR 1.27; 95% CI, 0.74–2.18; p = 0. 351; I²=0.8%). Thus, when both tests are applied to the same populations, MRI tends to be more specific, an attribute that reduces false positives and may spare unnecessary axillary interventions (Table 4, Figures S3-S4).
Publication bias
Potential reporting bias was assessed using Deeks’ funnel-plot asymmetry test. For MRI, the regression slope was statistically significant (p = 0.001), indicating asymmetry compatible with small-study effects. In contrast, the ultrasound funnel plot showed no evidence of asymmetry (p = 0.176). Because the test regresses ln(DOR) on the inverse square root of the effective sample size, asymmetry may reflect inflation of diagnostic odds ratios in smaller studies. This pattern suggests that pooled MRI estimates may be more vulnerable to selective publication or to systematic differences between smaller and larger cohorts. However, Deeks’ test is also influenced by some factors, including variation in MRI sequences, diagnostic thresholds for nodal positivity, and verification pathways. Accordingly, the finding should be interpreted as a cautionary signal rather than definitive proof of publication bias. Overall, the bias assessment supports greater confidence in the stability of the ultrasound summary estimates, while emphasizing the need to prioritize larger, methodologically well-described MRI studies when translating these results into clinical decision-making. Fig. 5
Fig. 5.
Deeks funnel (left: MRI, right: Ultrasound). MRI slope p-value = 0.001, Ultrasound slope p-value = 0.176. Non-significant slope suggests no small-study effects. Regression of ln (DOR) on 1/sqrt (ESS) weighted by ESS
Influence diagnostics were examined to identify studies with disproportionate impact on Deeks’ regression slope (Tables S3-S4). For MRI, the largest Cook’s distance was observed for Hieken et al. (2013), followed by Barco et al. (2016), indicating that a small number of heavily weighted cohorts contributed meaningfully to the observed funnel asymmetry. Several smaller MRI studies contained zero cells, but their Cook’s distances were low, suggesting limited leverage on the slope after continuity correction. For ultrasound, the most influential observations were again Hieken et al. (2013) and Barco et al. (2016), with moderate influence from Upadhyaya et al. (2019) and Schumacher et al. (2021). Overall, these diagnostics support interpreting Deeks’ findings as driven by a small subset of influential studies rather than uniformly across the evidence base.
Sensitivity analysis
A sensitivity analysis was performed by excluding the smallest 25% of studies based on effective sample size (Table 5). After exclusion, MRI estimates were derived from 19 studies (vs. 26 in the primary analysis) and showed a lower pooled sensitivity (52.6% vs. 61.1%), while specificity remained essentially unchanged (91.9% vs. 92.0%). In contrast, ultrasound results were stable: sensitivity was nearly identical after exclusion (59.6% vs. 59.8%), with a small reduction in specificity (88.8% vs. 90.3%). These findings suggest that MRI sensitivity may be more sensitive to small-study contributions, whereas ultrasound pooled performance is comparatively robust under this restriction.
Table 5.
Sensitivity analysis excluding the smallest 25% of studies (by ESS)
| Modality | Studies: Primary After Exclusion |
Sensitivity: Primary → After Exclusion | Specificity: Primary → After Exclusion |
|---|---|---|---|
| MRI | 26 → 19 | 61.1% → 52.6% | 92.0% → 91.9% |
| Ultrasound | 17→ 13 | 59.8% → 59.6% | 90.3% → 88.8% |
Exclusion based on ESS derived from 2 × 2 tables (continuity correction applied if needed); pooled sensitivity and specificity re-estimated using the same GLMM approach
HSROC curves
HSROC analysis was used to summarize the joint performance of sensitivity and specificity across studies while accounting for threshold effects (Fig. 6). For MRI, the Rutter-Gatsonis HSROC derived from the bivariate random-effects model yielded a summary sensitivity of 57.5% and specificity of 90.9% across 26 studies, with an area under the curve (AUC) of 0.863. For ultrasound, the corresponding summary point was a sensitivity of 59.4% and specificity of 88.9% across 17 studies, with an AUC of 0.843. Both AUC values exceed 0.80, supporting good discriminative ability for preoperative nodal assessment. The modest separation between AUCs suggests slightly better overall discrimination for MRI, largely attributable to higher specificity rather than superior sensitivity. The HSROC curves also highlight an important trade-off: neither modality simultaneously achieves high sensitivity and high specificity, which aligns with the moderate pooled sensitivities reported in the primary synthesis. From a clinical perspective, this implies that imaging alone cannot replace histopathology for definitive axillary staging, particularly when missed metastasis would change management. Nevertheless, the high-specificity region of the MRI curve supports its role as a confirmatory test when a node is classified as suspicious, as it may reduce false-positive driven escalation of axillary surgery. Ultrasound, conversely, maintains slightly higher sensitivity and remains well suited for first-line screening and triage given its accessibility and real-time assessment. Taken together, the HSROC findings support a complementary pathway in which ultrasound is used broadly to identify patients who may require further assessment, and MRI is applied selectively to increase rule-in confidence and refine preoperative planning.
Discussion
In this diagnostic accuracy meta-analysis of 29 studies (n = 8,544), MRI and ultrasound showed broadly comparable performance for preoperative detection of axillary nodal metastasis. Sensitivity was moderate for both tests, whereas specificity was consistently high. Under the primary random-effects GLMM, pooled sensitivity was 61.1% for MRI and 59.8% for ultrasound, while pooled specificity was 92.0% and 90.3%, respectively, translating into stronger rule-in performance for MRI (LR + 7.64; DOR 18.1) than for ultrasound (LR + 6.19; DOR 13.9). Statistical inconsistency was low (I²=0.9%), but the estimated between-study variance (τ², logit scale) and wide 95% prediction intervals, most notably for sensitivity, suggest that real-world accuracy could differ meaningfully by institution and protocol. When mapped to clinically plausible pre-test probabilities (10%−30%), PPV increased as prevalence rose (MRI: 45.9% to 76.6%; ultrasound: 40.7% to 72.6), whereas NPV remained relatively stable and high across scenarios (Table 3). In direct head-to-head comparisons (k = 12), MRI and ultrasound did not differ in sensitivity (OR 0.96, 95% CI 0.61–1.51), and MRI showed a non-significant tendency toward higher specificity (OR 1.27, 95% CI 0.74–2.18).
The pooled sensitivity estimates in our study, 61.1% for MRI and 59.8% for US, are largely consistent with Literature. A systematic review by Alvarez et al. found MRI sensitivity ranging from 60 to 80%, mirroring our pooled MRI sensitivity of 61.1% [45].
Importantly, our study further clarifies the specificity advantage of MRI. Earlier meta-analyses have reported MRI specificities between 85% and 93%, but without a consistent head-to-head comparison with US [46, 47]. Our pooled MRI specificity of 92.0% not only confirms the strong rule-in value of MRI but also demonstrates a non-significant superiority compared to US (90.3%) in direct comparison. This aligns with the findings from Baxter et al., who showed MRI offered improved specificity when high-resolution sequences or diffusion-weighted imaging were implemented [48]. Our meta-analysis strengthens this observation through larger combined data and standardized statistical modeling.
The diagnostic odds ratio (DOR), a comprehensive measure integrating sensitivity and specificity, also favored MRI (18.1 vs. 13.9 for US), consistent with prior work. A meta-analysis by Li et al. found MRI DOR values around 20–25, and US around 10–15, closely matching our results and reinforcing the higher discriminatory performance of MRI [49]. These comparative findings support the use of MRI, particularly when higher diagnostic certainty is required, such as in cases with ambiguous US findings or in patients for whom accurate pre-operative mapping is critical.
A particularly valuable contribution of this meta-analysis is the predictive value modeling across clinically relevant pre-test probabilities. Prior studies rarely explored PPV/NPV across varying prevalence. Our analysis shows that MRI consistently produces higher PPVs at all examined nodal metastasis prevalence (10%, 20%, and 30%). For instance, at 30% prevalence, the PPV for MRI is 76.6%, compared to 72.6% for US. These differences have meaningful clinical implications: when MRI identifies suspicious nodes, the likelihood of true metastasis is substantially higher. Conversely, both modalities achieve similarly high NPVs, above 87% across all prevalence levels, indicating strong utility for ruling out nodal disease. This finding aligns with an earlier study by Kim et al., who highlighted the high negative predictive value of both modalities for excluding significant metastatic burden [50].
Because PPV and NPV are prevalence-dependent, we modeled predictive values across clinically plausible pre-test probabilities using Bayes’ rule, rather than reporting a single “fixed” PPV/NPV that may not generalize across settings [Bours, 2021 #56]. We selected 10%, 20%, and 30% to represent low-, intermediate-, and higher-risk pathways encountered in practice. For example, in patients with small tumors (≤ 2 cm) and negative preoperative axillary ultrasound, the SOUND trial still reported a 13.7% rate of nodal involvement, supporting a 10–15% scenario [Gentilini, 2023 #57]. Conversely, broader clinically node-negative populations often have nodal positivity closer to one quarter overall (i.e., ~ 74% sentinel-node negative), making 20–30% a reasonable bracket for symptomatic presentations or cohorts enriched with higher-risk features [Yu, 2022 #58]. In our study, this prevalence-based framing clarified the clinical trade-off: MRI’s higher specificity translated into consistently higher PPV than ultrasound across all modeled prevalences, while NPVs remained very similar; at 20% prevalence, PPV was 65.6% vs. 60.7% and NPV 90.4% vs. 90.0% (MRI vs. US).
The direct head-to-head comparison incorporated in our meta-analysis is in line with priori explained findings. While many prior meta-analyses evaluated MRI or US independently, relatively few studies directly contrasted both modalities in paired cohorts. Our inclusion of twelve such studies provides clearer insight: MRI and US do not differ significantly in sensitivity, but MRI is tend to be more specific. This nuanced comparative perspective was lacking in earlier reviews and is essential for evidence-based decision-making.
Integrating imaging performance with explicit clinical decision pathways is essential for translating diagnostic accuracy into safe axillary de-escalation. In our meta-analysis, ultrasound and MRI demonstrated only moderate sensitivity for detecting nodal metastasis (59.8% and 61.1%, respectively), whereas specificity was high, particularly for MRI (90.3% for ultrasound and 92.0% for MRI), supporting imaging as a pragmatic “rule-in” rather than a definitive “rule-out” strategy in isolation. This performance profile becomes more actionable when embedded in structured preoperative algorithms that define thresholds for escalation, confirmatory biopsy, or selective omission of surgical staging. For example, Chan and colleagues showed that adding MRI meaningfully altered management: MRI led to cancellation of breast-conserving surgery/IORT in a subset of patients, including cases where axillary nodal metastases were detected only on MRI, while also introducing measurable system burdens such as longer waiting time [51]. These decision-level effects underscore that the clinical value of imaging is not limited to summary accuracy metrics but also includes how it redirects treatment intensity. Importantly, Argentiero and co-authors emphasize that diagnostic strategies should not follow a single template and argue for integrating biological profiling into multistep work-ups, recognizing microenvironmental and molecular heterogeneity as key drivers of disease behavior [52]. Future research in axillary staging should therefore prioritize validated multimodal prediction frameworks that combine imaging with clinicopathologic and biologic variables to improve false-negative control while supporting individualized de-escalation [53].
Despite the strengths of our analysis, including advanced GLMM modeling, predictive value assessment, and publication bias evaluation, certain limitations must be acknowledged. Variations in imaging techniques, operator expertise, nodal size thresholds, and histopathological reference standards may introduce heterogeneity across studies. Some MRI studies used conventional sequences only, whereas others incorporated diffusion-weighted imaging or dedicated axillary protocols, contributing to variability. Additionally, MRI availability, higher cost, and longer acquisition times remain barriers in resource-limited settings. We couldn’t comprehensively evaluate the impact of imaging protocols heterogeneity because the key technical and interpretive details were not consistently reported and it is another main limitation. For MRI, reporting was similarly variable with respect to field strength, coil setup, axillary coverage, sequence selection (conventional sequences vs. diffusion-weighted imaging and/or dedicated axillary protocols), and reader blinding to histopathology. This incomplete and inconsistent reporting limited descriptive synthesis of protocol factors and precluded robust subgroup analyses or meta-regression to explain residual variability. For ultrasound, studies varied in criteria for nodal positivity (e.g., cortical thickness thresholds, hilar effacement, morphology, and Doppler features), operator experience, and whether suspicious nodes underwent confirmatory needle sampling.
Conclusion
MRI and ultrasound each play distinct roles in the pre-operative assessment of axillary lymph node status in early breast cancer. Ultrasound offers slightly higher sensitivity and can be effectively used as an initial screening tool. MRI demonstrates non-significant higher specificity, greater diagnostic odds ratio, and superior predictive value, making it a valuable confirmatory modality when precise nodal evaluation is required. The combined use of both imaging strategies may improve diagnostic accuracy, guide personalized surgical planning, and reduce unnecessary axillary procedures. Future standardized, prospective trials are warranted to refine decision-making algorithms and further optimize axillary staging.
Supplementary Information
Below is the link to the electronic supplementary material.
Acknowledgements
None.
Abbreviations
- US
Ultrasound
- MRI
Magnetic resonance imaging
- ALN
Axillary lymph node
- ALND
Axillary lymph node dissection
- AUC
Area under the curve
- CI
Confidence interval
- DOR
Diagnostic odds ratio
- ESS
Effective sample size
- FN
False negative
- FP
False positive
- TN
True negative
- TP
True positive
- GLMM
Generalized linear mixed-effects model
- HSROC
Hierarchical summary receiver operating characteristic
- I²
I-squared heterogeneity statistic
- IORT
Intraoperative radiotherapy
- Κ
Cohen’s kappa coefficient
- LR+
Positive likelihood ratio
- LR-
Negative likelihood ratio
- MEDLINE
Medical Literature Analysis and Retrieval System Online
- PPV
Positive predictive value
- NPV
Negative predictive value
- OR
Odds ratio
- PI
Prediction interval
- PRISMA
Preferred Reporting Items for Systematic Reviews and Meta-Analyses
- QUADAS-2
Quality Assessment of Diagnostic Accuracy Studies-2
- SLNB
Sentinel lymph node biopsy
- SOUND trial
Sentinel Node vs. Observation After Axillary UltraSound trial
- τ²
Between-study variance (tau-squared)
Author contributions
Conception and Design: HJT, RB, FS, FNR, DY, SMAnalysis and interpretation of data: FNR, SS, ZA, DYData collection: RB, FS, FNR, SS, DYRisk of Bias Assessment: DY, FNR, HTJParticipating in drafting or revising the article: HJT, RB, FS, FNR, SS, DY, ZASupervision, Project Administration, Review & Editing: DY and ZA.
Funding
None.
Data availability
The dataset generated or analyzed in this study is available from the corresponding author upon a reasonable request.
Declarations
Competing interests
The authors declare no competing interests.
Consent for publication
Not applicable.
Clinical trial registration number
Not applicable.
Footnotes
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Luo C, Li N, Lu B, Cai J, Lu M, Zhang Y, et al. Global and regional trends in incidence and mortality of female breast cancer and associated factors at National level in 2000 to 2019. Chin Med J (Engl). 2022;135(1):42–51. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Arnold M, Morgan E, Rumgay H, Mafra A, Singh D, Laversanne M, et al. Current and future burden of breast cancer: global statistics for 2020 and 2040. Breast. 2022;66:15–23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Nolan E, Lindeman GJ, Visvader JE. Deciphering breast cancer: from biology to the clinic. Cell. 2023;186(8):1708–28. [DOI] [PubMed] [Google Scholar]
- 4.Loibl S, André F, Bachelot T, Barrios CH, Bergh J, Burstein HJ, et al. Early breast cancer: ESMO clinical practice guideline for diagnosis, treatment and follow-up. Ann Oncol. 2024;35(2):159–82. [DOI] [PubMed] [Google Scholar]
- 5.Cianfrocca M, Goldstein LJ. Prognostic and predictive factors in early-stage breast cancer. Oncologist. 2004;9(6):606–16. [DOI] [PubMed] [Google Scholar]
- 6.Choi HY, Park M, Seo M, Song E, Shin SY, Sohn YM. Preoperative axillary lymph node evaluation in breast cancer: current issues and literature review. Ultrasound Q. 2017;33(1):6–14. [DOI] [PubMed] [Google Scholar]
- 7.Park KU, Somerfield MR, Anne N, Brackstone M, Conlin AK, Couto HL, et al. Sentinel lymph node biopsy in early-stage breast cancer: ASCO guideline update. J Clin Oncol. 2025;43(14):1720–41. [DOI] [PubMed] [Google Scholar]
- 8.Wu T, Long Q, Zeng L, Zhu J, Gao H, Deng Y, et al. Axillary lymph node metastasis in breast cancer: from historical axillary surgery to updated advances in the preoperative diagnosis and axillary management. BMC Surg. 2025;25(1):81. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Rahbar H, McDonald ES, Lee JM, Partridge SC, Lee CI. How can advanced imaging be used to mitigate potential breast cancer overdiagnosis? Acad Radiol. 2016;23(6):768–73. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Di Paola V, Mazzotta G, Pignatelli V, Bufi E, D’Angelo A, Conti M et al. Beyond N staging in breast cancer: importance of MRI and Ultrasound-based imaging. Cancers (Basel). 2022;14(17):4270. 10.3390/cancers14174270 [DOI] [PMC free article] [PubMed]
- 11.Sood R, Rositch AF, Shakoor D, Ambinder E, Pool KL, Pollack E, et al. Ultrasound for breast cancer detection globally: A systematic review and Meta-Analysis. J Glob Oncol. 2019;5:1–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Redman A, Lowes S, Leaver A. Imaging techniques in breast cancer. Surg-Oxf Int Ed. 2016;34(1):8–18. [Google Scholar]
- 13.Aristokli N, Polycarpou I, Themistocleous SC, Sophocleous D, Mamais I. Comparison of the diagnostic performance of magnetic resonance imaging (MRI), ultrasound and mammography for detection of breast cancer based on tumor type, breast density and patient’s history: a review. Radiography Lond. 2022;28(3):848–56. [DOI] [PubMed] [Google Scholar]
- 14.Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD, et al. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ. 2021;372:n71. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Whiting PF, Rutjes AW, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, et al. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. 2011;155(8):529–36. [DOI] [PubMed] [Google Scholar]
- 16.Assing MA, Patel BK, Karamsadkar N, Weinfurtner J, Usmani O, Kiluk JV, et al. A comparison of the diagnostic accuracy of magnetic resonance imaging to axillary ultrasound in the detection of axillary nodal metastases in newly diagnosed breast cancer. Breast J. 2017;23(6):647–55. [DOI] [PubMed] [Google Scholar]
- 17.Abe H, Schacht D, Kulkarni K, Shimauchi A, Yamaguchi K, Sennett CA, et al. Accuracy of axillary lymph node staging in breast cancer patients: an Observer-Performance study comparison of MRI and ultrasound. Acad Radiol. 2013;20(11):1399–404. [DOI] [PubMed] [Google Scholar]
- 18.Ahn HS, Jang M, Kim SM, La Yun B, Lee SH. Usefulness of preoperative breast magnetic resonance imaging with a dedicated axillary sequence for the detection of axillary lymph node metastasis in patients with early ductal breast cancer. Radiol Med (Torino). 2019;124(12):1220–8. [DOI] [PubMed] [Google Scholar]
- 19.Aktaş A, Gürleyik MG, Aydın Aksu S, Aker F, Güngör S. Diagnostic value of axillary ultrasound, MRI, and (18)F-FDG-PET/ CT in determining axillary lymph node status in breast cancer patients. Eur J Breast Health. 2022;18(1):37–47. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Arslan G, Altintoprak KM, Yirgin IK, Atasoy MM, Celik L. Diagnostic accuracy of metastatic axillary lymph nodes in breast MRI. SpringerPlus. 2016;5(1):735. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Atallah D, Moubarak M, Arab W, El Kassis N, Chahine G, Salem C. MRI-based predictive factors of axillary lymph node status in breast cancer. Breast J. 2020;26(11):2177–82. [DOI] [PubMed] [Google Scholar]
- 22.Barco I, Chabrera C, García-Fernández A, Fraile M, González S, Canales L, et al. Role of axillary ultrasound, magnetic resonance imaging, and ultrasound-guided fine-needle aspiration biopsy in the preoperative triage of breast cancer patients. Clin Transl Oncol. 2017;19(6):704–10. [DOI] [PubMed] [Google Scholar]
- 23.Chayakulkheeree J, Pungrassami D, Prueksadee J. Performance of breast magnetic resonance imaging in axillary nodal staging in newly diagnosed breast cancer patients. Pol J Radiol. 2019;84:e413–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.de Felice C, Cipolla V, Stagnitti A, Porfiri LM, Guerrieri D, Musella A, et al. Diagnostic accuracy of 1.5 Tesla breast magnetic resonance imaging in the pre-operative assessment of axillary lymph nodes. Eur J Gynaecol Oncol. 2015;36(4):447–51. [PubMed] [Google Scholar]
- 25.Di Micco R, Botteri E, Canevari C, Gallivanone F, Antunovic L, Scifo P et al. [< sup>18 F]FDG PET/MRI vs Sentinel node biopsy for axillary staging of early breast cancer patients. A prospective single-arm trial. Eur J Cancer. 2025;224:115519. 10.1016/j.ejca.2025.115519 [DOI] [PubMed]
- 26.Ergul N, Kadioglu H, Yildiz S, Yucel SB, Gucin Z, Erdogan EB, et al. Assessment of multifocality and axillary nodal involvement in early-stage breast cancer patients using 18F-FDG PET/CT compared to contrast-enhanced and diffusion-weighted magnetic resonance imaging and sentinel node biopsy. Acta Radiol. 2015;56(8):917–23. [DOI] [PubMed] [Google Scholar]
- 27.García Fernández A, Fraile M, Giménez N, Reñe A, Torras M, Canales L, et al. Use of axillary ultrasound, ultrasound-fine needle aspiration biopsy and magnetic resonance imaging in the preoperative triage of breast cancer patients considered for sentinel node biopsy. Ultrasound Med Biol. 2011;37(1):16–22. [DOI] [PubMed] [Google Scholar]
- 28.Hieken TJ, Trull BC, Boughey JC, Jones KN, Reynolds CA, Shah SS, et al. Preoperative axillary imaging with percutaneous lymph node biopsy is valuable in the contemporary management of patients with breast cancer. Surgery. 2013;154(4):831–40. [DOI] [PubMed] [Google Scholar]
- 29.Jung NY, Kim SH, Kim SH, Seo YY, Oh JK, Choi HS, et al. Effectiveness of breast MRI and 18F-FDG PET/CT for the preoperative staging of invasive lobular carcinoma versus ductal carcinoma. J Breast Cancer. 2015;18(1):63–72. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Kim SH, Shin HJ, Shin KC, Chae EY, Choi WJ, Cha JH, et al. Diagnostic performance of fused diffusion-weighted imaging using T1-weighted imaging for axillary nodal staging in patients with early breast cancer. Clin Breast Cancer. 2017;17(2):154–63. [DOI] [PubMed] [Google Scholar]
- 31.Motomura K, Izumi T, Tateishi S, Tamaki Y, Ito Y, Horinouchi T, et al. Superparamagnetic iron oxide-enhanced MRI at 3 T for accurate axillary staging in breast cancer. Br J Surg. 2016;103(1):60–9. [DOI] [PubMed] [Google Scholar]
- 32.Panda SK, Goel A, Nayak V, Shaik Basha S, Pande PK, Kumar K. Can preoperative ultrasonography and MRI replace Sentinel lymph node biopsy in management of axilla in early breast cancer, a prospective study from a tertiary cancer center. Indian J Surg Oncol. 2019;10(3):483–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Riedel F, Schaefgen B, Sinn H-P, Feisst M, Hennigs A, Hug S, et al. Diagnostic accuracy of axillary staging by ultrasound in early breast cancer patients. Eur J Radiol. 2021;135:109468. 10.1016/j.ejrad.2020.109468. [DOI] [PubMed] [Google Scholar]
- 34.Rukanskienė D, Veikutis V, Jonaitienė E, Basevičiūtė M, Kunigiškis D, Paukštaitienė R, et al. Preoperative axillary ultrasound versus sentinel lymph node biopsy in patients with early breast cancer. Medicina (Kaunas). 2020;56(3):127. 10.3390/medicina56030127. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Schumacher K, Inciardi M, O’Neil M, Wagner JL, Shah I, Amin AL, et al. Is axillary imaging for invasive lobular carcinoma accurate in determining clinical node staging? Breast Cancer Res Treat. 2021;185(3):567–72. [DOI] [PubMed] [Google Scholar]
- 36.Stadnik TW, Everaert H, Makkat S, Sacré R, Lamote J, Bourgain C. Breast imaging. Preoperative breast cancer staging: comparison of USPIO-enhanced MR imaging and 18F-fluorodeoxyglucose (FDC) positron emission tomography (PET) imaging for axillary lymph node staging–initial findings. Eur Radiol. 2006;16(10):2153–60. [DOI] [PubMed] [Google Scholar]
- 37.Upadhyaya VS, Lim GH, Chan EYK, Fook-Chong SMC, Leong LCH. Evaluating the preoperative breast cancer characteristics affecting the accuracy of axillary ultrasound staging. Breast J. 2020;26(2):162–7. [DOI] [PubMed] [Google Scholar]
- 38.Valente SA, Levine GM, Silverstein MJ, Rayhanabad JA, Weng-Grumley JG, Ji L, et al. Accuracy of predicting axillary lymph node positivity by physical examination, mammography, ultrasonography, and magnetic resonance imaging. Ann Surg Oncol. 2012;19(6):1825–30. [DOI] [PubMed] [Google Scholar]
- 39.van Nijnatten TJA, Ploumen EH, Schipper RJ, Goorts B, Andriessen EH, Vanwetswinkel S, et al. Routine use of standard breast MRI compared to axillary ultrasound for differentiating between no, limited and advanced axillary nodal disease in newly diagnosed breast cancer patients. Eur J Radiol. 2016;85(12):2288–94. [DOI] [PubMed] [Google Scholar]
- 40.Wang J, Ben Z, Gao S, Lyu S, Wei X. The role of ultrasound elastography and virtual touch tissue imaging in the personalized prediction of lymph node metastasis of breast cancer. Gland Surg. 2021;10(4):1460–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Wang L-J, Yu J-C, Hong Z-J. The predict value of lymph node status pre-operation by ultrasound, mammography and MRI in early breast cancer. J Formos Med Assoc. 2025;S0929-6646(25)00457-7. 10.1016/j.jfma.2025.08.038 [DOI] [PubMed]
- 42.Xu Y, Bai X, Chen Y, Jiang L, Hu B, Hu B, et al. Application of real-time elastography ultrasound in the diagnosis of axillary lymph node metastasis in breast cancer patients. Sci Rep. 2018;8(1):10234. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Yu Y, He Z, Ouyang J, Tan Y, Chen Y, Gu Y, et al. Magnetic resonance imaging radiomics predicts preoperative axillary lymph node metastasis to support surgical decisions and is associated with tumor microenvironment in invasive breast cancer: a machine learning, multicenter study. eBioMedicine. 2021;69:103460. 10.1016/j.ebiom.2021.103460. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Zhang H, Guo LL, Tao WJ, Zhang JD, Bai GJ. Comparison of the clinical application value of mo-targeted X-ray, color doppler ultrasound and MRI in preoperative comprehensive evaluation of breast cancer. Saudi J Biol Sci. 2019;26(8):1973–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Alvarez S, Añorbe E, Alcorta P, López F, Alonso I, Cortés J. Role of sonography in the diagnosis of axillary lymph node metastases in breast cancer: a systematic review. AJR Am J Roentgenol. 2006;186(5):1342–8. [DOI] [PubMed] [Google Scholar]
- 46.Abdullah KA, Marziali S, Nanaa M, Escudero Sánchez L, Payne NR, Gilbert FJ. Deep learning-based breast cancer diagnosis in breast MRI: systematic review and meta-analysis. Eur Radiol. 2025;35(8):4474–89. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Liang X, Yu J, Wen B, Xie J, Cai Q, Yang Q. MRI and FDG-PET/CT based assessment of axillary lymph node metastasis in early breast cancer: a meta-analysis. Clin Radiol. 2017;72(4):295–301. [DOI] [PubMed] [Google Scholar]
- 48.Baxter GC, Graves MJ, Gilbert FJ, Patterson AJ. A meta-analysis of the diagnostic performance of diffusion MRI for breast lesion characterization. Radiology. 2019;291(3):632–41. [DOI] [PubMed] [Google Scholar]
- 49.Li Z, Gao Y, Gong H, Feng W, Ma Q, Li J, et al. Different imaging modalities for the diagnosis of axillary lymph node metastases in breast cancer: A systematic review and network Meta-Analysis of diagnostic test accuracy. J Magn Reson Imaging. 2023;57(5):1392–403. [DOI] [PubMed] [Google Scholar]
- 50.Kim WH, Kim HJ, Lee SM, Cho SH, Shin KM, Lee SY, et al. Preoperative axillary nodal staging with ultrasound and magnetic resonance imaging: predictive values of quantitative and semantic features. Br J Radiol. 2018;91(1092):20180507. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Chan WY, Cheah WK, Ramli Hamid MT, Md Shah MN, Fadzli F, Kaur S, et al. Impact of preoperative magnetic resonance imaging on surgery and eligibility for intraoperative radiotherapy in early breast cancer. PLoS One. 2022;17(10):e0274385. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Argentiero A, Solimando AG, Brunetti O, Calabrese A, Pantano F, Iuliani M, et al. Skeletal metastases of unknown primary: biological landscape and clinical overview. Cancers (Basel). 2019;11(9):1270. 10.3390/cancers11091270. PMID: 31470608; PMCID: PMC6770264. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Kim S-Y, Choi Y, Kim YS, Ha SM, Lee SH, Han W, et al. Use of imaging prediction model for omission of axillary surgery in early-stage breast cancer patients. Breast Cancer Res Treat. 2023;199(3):489–99. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The dataset generated or analyzed in this study is available from the corresponding author upon a reasonable request.