Abstract
Introduction
In vivo data on volumetric laser efficiency and clinical outcomes of the p-Tm: YAG laser from large multicenter cohorts remain limited. The present study aimed to evaluate the real-world in vivo volumetric ablation efficiency of the p-Tm: YAG laser during flexible ureterorenoscopy (FURS), with secondary assessment of clinical outcomes and safety in a multicenter cohort.
Methods
We conducted a retrospective multicenter cohort study including adult patients (≥ 18 years) who underwent p-Tm: YAG–based FURS for ureteral and/or renal calculi at three tertiary referral centers between 2023 and 2025. Preoperative stone volume was quantified using manual three-dimensional CT segmentation. Laser-on time (LOT), total delivered energy, and laser settings were prospectively recorded. Volumetric ablation speed (mm³/s) and energy consumption (J/mm³) were calculated using standardized definitions based on segmented pre- and postoperative stone volumes in patients with postoperative non-contrast CT (NCCT). Stone-free rate (SFR) was assessed 1–3 months postoperatively using NCCT and classified as Grade A (no residual fragments), Grade B (residual fragments ≤ 2 mm), and Grade C (residual fragments ≤ 4 mm). In patients without postoperative NCCT, SFR was assessed by intraoperative endoscopic evaluation and analyzed separately.
Results
Of 222 screened patients, 167 met the inclusion criteria. Median preoperative stone volume was 987 mm³ (IQR 329–3,124), and 47% of patients presented with multiple/complex stones. Median LOT was 21.4 min, and median total delivered energy was 12 kJ. In the 108 patients with paired volumetric data and postoperative NCCT, median ablation speed was 0.55 mm³/s (IQR 0.30–0.95), and median energy consumption was 13 J/mm³ (IQR 8–23). In this NCCT subgroup, SFR were 47%, 59%, and 66% for Grade A, B, and C, respectively. Among the 59 patients assessed endoscopically only, complete stone clearance was visually assessed in 20 patients (34%). Overall, 45 patients (27%) required additional stone-related treatment. Complications were uncommon, with 4.8% Clavien–Dindo grade I–II events and 0.6% grade III events; no grade IV–V complications were observed.
Conclusion
In this multicenter real-world cohort, p-Tm: YAG laser lithotripsy showed measurable volumetric efficiency and was associated with acceptable stone clearance and a favorable safety profile during FURS. These findings support the clinical feasibility of this technology in routine endourological practice; however, prospective comparative studies are warranted to further define its relative performance compared with established laser platforms.
Supplementary Information
The online version contains supplementary material available at 10.1007/s00345-026-06467-1.
Keywords: Pulsed thulium YAG, Ureterorenoscopy, Laser, Lithotripsy, Volume, Multicenter, Stone composition
Introduction
Urinary stone disease affects 5–12% of the global population, with a steadily increasing incidence attributed to lifestyle, metabolic, and dietary factors [1]. Flexible ureterorenoscopy (FURS) with endoscopic laser lithotripsy has become a cornerstone treatment for ureteral and renal calculi, offering high stone-free rates (SFR) with low morbidity [2]. According to current international guidelines, Holmium: yttrium–aluminum–garnet (Ho: YAG) and thulium fiber laser (TFL) represent the standard energy sources for endoscopic laser lithotripsy due to their effectiveness across a wide range of stone compositions and anatomical locations [1].
The Ho: YAG laser has historically been considered the gold standard for endoscopic laser lithotripsy; however, its high peak power and short pulse duration may contribute to increased retropulsion and reduced dusting efficiency [3]. In contrast, the TFL has demonstrated improved dusting capability and ablation efficiency due to its lower peak power, higher pulse frequencies, and more favorable pulse shape characteristics [4].
The pulsed thulium: yttrium–aluminium–garnet (p-Tm: YAG) laser is a newer solid-state platform introduced to combine advantageous features of both Ho: YAG and TFL technologies [5]. Operating at a wavelength of approximately 2013 nm and capable of delivering peak powers up to 3.7 kW, the p-Tm: YAG allows a wide range of pulse durations, frequencies, and energy settings, enabling both dusting and fragmentation strategies [6]. Preclinical studies suggest that p-Tm: YAG achieves ablation efficiency comparable to TFL while producing less retropulsion than Ho: YAG [7–9]. Early clinical experiences further support its potential, reporting SFR of reaching up to 95% in FURS with high safety profile [10].
However, real-world multicenter evidence remains limited, particularly regarding objective volumetric efficiency parameters such as ablation speed and energy consumption measured in vivo across heterogeneous stone burdens and anatomical scenarios [11]. Therefore, the present multicenter retrospective study was designed to primarily evaluate the real-world in vivo volumetric laser efficiency of the p-Tm: YAG laser during FURS. Secondary objectives included the assessment of SFR outcomes and perioperative safety in order to contextualize laser performance in routine clinical practice.
Methods
Study design and participants
This retrospective multicenter cohort study was conducted across three tertiary referral endourology centers: Hôpital Tenon (Paris, France), Tan Tock Seng Hospital (Singapore), and Hospital Clínic de Barcelona (Barcelona, Spain). The study included consecutive patients treated with pulsed thulium: YAG (p-Tm: YAG, Thulio®, Dornier MedTech, Wessling, Germany) laser lithotripsy and was conducted in accordance with national and institutional research regulations.
Data processing was declared to the Commission Nationale de l’Informatique et des Libertés (CNIL) under the MR-004 framework (approvals 2216615 v0 and 2230198 v0). Ethical approval was granted by the Comité d’Éthique et de la Recherche de l’Association Française d’Urologie (CERU approvals CERU_2020/003 and CERU_2023018B). Institutional review board approval was obtained at each participating site, and informed consent was waived or obtained in accordance with local ethics policies.
Eligible participants were adult patients (≥ 18 years) who underwent FURS with p-Tm: YAG laser lithotripsy for renal and/or ureteral calculi between January 2023 and September 2025. Patients were identified through prospectively maintained institutional surgical databases at each center. All participating institutions applied consistent preoperative evaluation protocols and standardized inclusion and exclusion criteria throughout the enrollment period to ensure cohort uniformity.
A total of 222 consecutive eligible patients were identified. Patients were excluded only when complete intraoperative laser data (e.g., total delivered energy and laser-on time), were unavailable, as these variables were required for volumetric efficiency analyses. Fifty-five patients were excluded due to incomplete laser parameter documentation, resulting in a final study cohort of 167 patients included in the analysis. The patient selection process is summarized in (Fig. 1).
Fig. 1.
Flow diagram of patient selection. Selection of patients included in the multicenter retrospective cohort study
Surgical procedure
Flexible ureterorenoscopy was performed under general anesthesia using ureteroscopes with an outer diameter ranging from 6.3 to 9.5 Fr, including both reusable and single-use platforms.
When required, a ureteral access sheath (9.5–12 Fr) was inserted to facilitate repeated access and optimize irrigation. In selected cases, a flexible suction access sheath (FANS) was used to provide active outflow and assist with intrarenal pressure control.
Laser lithotripsy was performed using 200–272 μm laser fibers in dusting or fragmentation settings according to standard clinical practice. Maximum laser power was limited to 20 W for intrarenal lithotripsy and 10 W for ureteral stones.
Irrigation was delivered using 0.9% saline solution at room temperature. In standard procedures, irrigation pressure was maintained at approximately 40 cmH₂O. In suction-assisted procedures, controlled irrigation inflow was combined with regulated negative-pressure outflow to maintain adequate visibility while minimizing intrarenal pressure elevation.
Stone lithotripsy was performed using the “painting technique” [12], with laser pulses delivered in long-pulse mode to optimize dusting efficiency and laser–stone interaction. At the conclusion of the procedure, residual fragments were retrieved when necessary, using nitinol baskets, and a ureteral stent was placed according to predefined clinical indications and surgeon judgment.
Data collection and outcomes
Demographic data including age, sex, body mass index, ASA score, were collected for all patients. Stone characteristics included anatomical location, number, maximum diameter, total stone volume, radiological density (Hounsfield units), and postoperative stone composition when available. Stone location was categorized as renal only, ureter only, or complex stone burden. Complex stones were defined as stones involving both the kidney and ureter or presenting as multiple/complex renal stone burden (e.g., partial or complete staghorn stones).
Preoperative stone volume was quantified using manual three-dimensional volumetric segmentation of non-contrast CT (NCCT) images using the dedicated 3D Slicer software [13]. All segmentations were conducted by a single trained investigator (C.G.G.) to ensure methodological consistency across centers. Segmentation was conducted using a standardized threshold-based approach with manual contour refinement when necessary. CT-based volumetric segmentation has been previously validated for accurate stone volume assessment in both experimental and clinical settings [14, 15]. Data was extracted at each center using a standardized data collection template. In patients presenting with complex stones, defined as stone burdens involving more than one intrarenal space, total stone burden was calculated as the sum of all segmented stone volumes per patient.
Intraoperative parameters, including pulse energy (J), pulse frequency (Hz), total power (W), total delivered energy (kJ), and laser-on time (LOT), were recorded, along with operative time and postoperative complications. Complications were classified according to the Clavien–Dindo classification system [16]. Laser settings were selected at the discretion of the operating surgeon, reflecting routine clinical practice; therefore, modulation-specific analyses were not performed.
Outcomes
The primary outcome of the study was volumetric laser efficiency, assessed using standardized efficiency metrics as previously defined [17]. These metrics were calculated using pre- and postoperative segmented stone volumes. Ablation speed (mm³/s) was defined as ablated stone volume divided by laser-on time, and energy consumption (J/mm³) was defined as total delivered energy divided by ablated stone volume. These outcomes were calculated among patients who had full data on preoperative and postoperative segmentation data.
Secondary outcomes included SFR, perioperative complications, and the need for additional stone-related interventions.
Postoperative SFR was assessed 1–3 months after surgery according to institutional follow-up protocols. When available, NCCT with 1–2 mm slice thickness was used for radiological assessment. In patients without postoperative NCCT (n = 59), SFR was determined based on intraoperative endoscopic assessment. To account for potential verification bias, SFR outcomes were analyzed separately according to the assessment modality. For clarity, SFR was described in accordance with prior literature [18]. In patients with postoperative NCCT imaging: Grade A (Zero fragment rate), no visible residual fragments; Grade B (relative stone-free), residual fragments ≤ 2 mm; and Grade C (relative stone-free), residual fragments ≤ 4 mm. Patients with residual fragments > 4 mm were classified as not stone-free.
Additional stone-related treatment was defined as any secondary procedure performed for residual stones during follow-up, including both planned staged procedures and unplanned retreatment. The dataset did not allow reliable differentiation between planned and unplanned secondary procedures, and this distinction could therefore not be analyzed.
Statistical analysis
All analyses were conducted in accordance with established statistical recommendations [19]. Continuous variables were summarized as medians and interquartile ranges (IQRs), and categorical variables as frequencies and percentages.
Group comparisons were performed using the Wilcoxon rank-sum test for continuous variables, and Fisher’s exact or Pearson’s Chi-square tests for categorical variables, as appropriate.
Violin plots were used to plot the changes in preoperative and postoperative stone volume. Additionally, box plots were used to represent the ablation speed calculated as the volume ablated for each second (mm3/s) according to stone type and composition. Given the lack of this data for the majority of patients, these results were only considered as exploratory. Stone-free rates and Zero fragment rate (ZFR) in accordance with the guidelines (Grade A, B and C) were indicated using proportions.
Predefined subgroup analyses were conducted according to anatomical stone location and imaging modality, and descriptive comparisons were used to assess potential variability across clinical scenarios. Given the moderate sample size and the large number of technical covariates, multivariable adjustment was considered potentially unstable and exploratory; therefore, findings are interpreted descriptively rather than causally.
Statistical significance was considered at p < 0.05. For all statistical analyses, R software environment for statistical computing and graphics was used (version 4.5.1; http://www.r-project.org/).
Results
Baseline characteristics
A total of 167 patients met the inclusion criteria, of whom 70 (42%) were female. The median age was 55 years (IQR 40–68), and the median BMI was 26.0 kg/m² (IQR 22.9–30.8). A history of urolithiasis was present in 17 patients (12%), and anatomical abnormalities were noted in 30 patients (18%).
Stone distribution was heterogeneous: 42 patients (25%) had isolated renal stones, 46 (28%) had ureteral stones only, and 79 (47%) presented with multiple anatomical locations. The median number of stones was 3 (IQR 1–3). Median stone density was 1579 HU (IQR 1148–1818), and the median preoperative stone volume was 987 mm³ (IQR 329–3124). Full baseline characteristics are summarized in Table 1.
Table 1.
Baseline characteristics of 167 patients treated with p-Tm: YAG lithotripsy across three tertiary centers (2023–2025)
| Characteristics | No. of patients (%) |
|---|---|
| No. of patients | 167 (100) |
| Age, years | |
| Median (IQR) | 55 (40–68) |
| Sex | |
| Female | 70(42) |
| BMI, Kg/m2 | |
| Median (IQR) | 26 (23–31) |
| ASA score | |
| 0 | 89(72) |
| 1 | 31(25) |
| 2 | 3(2.4) |
| Unknown | 44 |
| History of prior urolithiasis | 17(12) |
| Anatomic anomaly | 30(18) |
| Side | |
| Right | 65(45) |
| Left | 77(53) |
| Bilateral | 4(2.7) |
| Unknown | 21 |
| Stone location | |
| Kidney only | 42(25) |
| Ureter only | 46(28) |
| Multiple locations | 79(47) |
| Number of stones | |
| Median (IQR) | 3 (1–3) |
| Maximum stone density, Hounsfield | |
| Median (IQR) | 1579 (1148–1818) |
| Preoperative stone max diameter, mm | |
| Median (IQR) | 13 (9–20) |
| Preoperative stone volume, mm3 | |
| Median (IQR) | 987 (329–3124) |
IQR interquartile range, ASA American society of anesthesiologists
Laser and performance efficiency outcomes
The overall median laser-on time (LOT) was 21.4 min (IQR 11–40). Median pulse energy, frequency, and power were 0.8 J (IQR 0.6–0.8), 15 Hz (IQR 10–20), and 10 W (IQR 6–15), respectively (Table 2). Laser settings were comparable across centers.
Table 2.
Perioperative data and surgical outcomes
| Characteristics | No. patients (%) [N = 167 (100)] |
|---|---|
| Values (Median [IQR]) | |
| Laser on time, min | 21.39 (11–40) |
| Laser settings | |
| Energy (J) | 0.8 (0.6–0.8) |
| Pulse rate (Hz) | 15 (10–20) |
| Power (W) | 10 (6–15) |
| Pulse mode* |
Fragmentation: 126 (75%); Popcorning: 3 (1.8%); Captive fragmentation: 11 (6.6%); Combined approach: 1 (1.8%); Unknown: 26 (16%) |
| Total delivered energy (kJ) | 12 (5–30) |
| Energy consumption, J/mm3 | 13 (8–23) |
| Unknown | 59 |
| Ablation speed (mm 3/s) | 0.55 (0.30–0.95) |
| Unknown | 59 |
| Postoperative stenting | 163 (98) |
| Postoperative volume, mm3 | |
| Median (IQR) | 0 (0–777) |
| Unknown | 59 |
| UAS | |
| No | 85 (51) |
| Traditional UAS | 48 (29) |
| FANS | 22 (13) |
| Missing | 12 (7) |
| Complication grading | |
| Clavien 1–2 | 8 (5) |
| Clavien 3–4 | 1 (1) |
| Further stone surgery | 45(27) |
J Joule, Hz Hertz, W Watt, UAS ureteral access sheath, FANS Flexible aspirating navigable access sheath
* The variable “pulse mode” refers to the lithotripsy strategy selected and recorded intraoperatively on the laser platform (fragmentation, popcorning, captive fragmentation, or combined approach). In cases labeled as “Unknown,” the surgical strategy was not explicitly recorded in the operative report; however, detailed laser parameters (energy, frequency, power, total energy, and laser-on time) were available for all patients and were used for efficiency analyses
When lithotripsy was performed exclusively in the ureter, the median LOT was 15 min (IQR 6–23), with median pulse energy 0.8 J (IQR 0.60–0.80), frequency 10 Hz (IQR 5–20), and power 8 W (IQR 5–16). For kidney-only cases, the median LOT was 19 min (IQR 9–27), with median pulse energy 0.8 J (IQR 0.6–1), frequency 10 Hz (IQR 10–20), and power 10 W (IQR 6–14). In complex or multiple-location cases, the median LOT was 36 min (IQR 19–60), with pulse energy 0.8 J (IQR 0.6–1.0), frequency 15 Hz (IQR 10–20), and power 12 W (IQR 8–15) (Table 3).
Table 3.
Laser settings and lasering efficiency stratified by stone location (kidney, ureter, multiple/complex)
| Characteristics | Stone location [values (median [IQR])] | ||
|---|---|---|---|
| Kidney (n = 42) | Ureter (n = 46) | Multiple locations (n = 79) | |
| Laser on time, min | 19 (9–27) | 15 (6–23) | 36 (19–60) |
| Laser settings | |||
| Energy (J) | 0.8 (0.6–1.0) | 0.8 (0.6–0.8) | 0.8 (0.60–1.0) |
| Pulse rate (Hz) | 10 (10–20) | 10 (5–20) | 15 (10–20) |
| Power (W) | 10 (6–14) | 8 (5–16) | 12 (8–15) |
| Pulse mode |
Dusting (long pulse): 31 (74%) Popcorning: 1 (2%) Unknown: 10 (24%) |
Dusting (long pulse): 42 (91%) Captive Fragmentation: 2 (4.3%) Unknown: 2 (4.3%) |
Dusting (long pulse): 53 (67%) Popcorning: 2 (2.5%) Captive Fragmentation: 9 (11%) Combined approach: 1 (1.3%) Unknown: 14 (18%) |
| Total delivered energy (KJ) | 10 (3–21) | 6 (2.7–20) | 20 (11–40) |
| Unknown | 12 | 20 | 25 |
| J/mm 3 | 17 (8–26) | 16 (10–20) | 12 (7–18) |
| Unknown | 12 | 20 | 25 |
| Ablation speed (mm 3/s) | 0.52 (0.30–0.79) | 0.30 (0.21–0.58) | 0.66 (0.43–1.09) |
| Unknown | 12 | 20 | 25 |
The distribution of laser settings according to anatomical location is summarized in (Table 3). Dusting or long-pulse modes predominated in the ureter (91%), renal (74%), and multiple–location cases (67%).
The overall median total energy delivered was 12 kJ (IQR 5–30).
Among patients with known postoperative NCCT (n = 108), laser efficiency analysis showed a median energy consumption of 13 J/mm³ (IQR 8–23) and a median ablation speed of 0.55 mm³/s (IQR 0.30–0.95).
Stone composition data were available for 53 patients with known postoperative NCCT, allowing exploratory composition-stratified analysis (Fig. 2). Median ablation speeds were 0.75 mm³/s (IQR 0.45–1.09) for calcium oxalate stones (n = 42), 0.29 mm³/s (IQR 0.28–0.30) for brushite stones (n = 6), and 1.25 mm³/s (IQR 0.58–1.93) for carbapatite stones (n = 12).
Fig. 2.
Ablation speed according to stone composition. Violin and box plots illustrating ablation speed (mm³/s) across stone compositions (calcium oxalate, brushite, carbapatite, cystine, uric acid/struvite)
Postoperative stone-free rates and complications
Median postoperative stone volume was 0 mm³ (IQR: 0–777), corresponding to a median volume reduction of 781 mm³ (IQR: 236–2176) (Fig. 3).
Fig. 3.
Preoperative versus postoperative segmented stone volumes Violin plots comparing stone volume (cm³) before and after flexible ureteroscopy with p-Tm: YAG laser. Median postoperative volume was 0 cm³; reduction was statistically significant (p < 0.001, Wilcoxon paired test)
Radiological stone-free rate outcomes (NCCT subgroup)
Postoperative radiological SFR outcomes were assessed in the subgroup of patients who underwent postoperative NCCT (n = 108), which represented the principal efficacy population for SFR analysis. Grade A, Grade B, and Grade C SFR outcomes were 47% (51/108), 59% (64/108), and 66% (71/108), respectively.
Stone-free outcomes varied according to stone location and complexity. In patients with isolated ureteral stones, Grade A SFR was achieved in 100% (26/26) of cases. In patients with isolated renal stones (n = 30), Grade A, B, and C SFR were 59%, 72%, and 79%, respectively. In contrast, multiple-location (n = 54) cases demonstrated lower success rates, with Grade A, B, and C SFR of 29%, 38%, and 46%, respectively. Stone burden was strongly associated with treatment success: patients with a preoperative stone volume ≤ 1,000 mm³ (n = 87) achieved Grade A, B, and C SFR of 75%, 87%, and 92%, respectively, whereas among patients with a stone burden ≥ 10,000 mm³ (n = 17), only 10% (n = 2) achieved Grade C SFR.
Endoscopic assessment (no postoperative NCCT)
Patients without postoperative NCCT follow-up had baseline characteristics comparable to those of patients who underwent postoperative CT imaging (sTable 1). In 59 patients (35%), postoperative NCCT was not available, and SFR was therefore assessed by intraoperative endoscopic evaluation only. In this subgroup, complete stone clearance was visually assessed in 20 patients (34%), while residual fragments were observed in 39 patients (66%).
Additional treatments and complications
Overall, 45 patients (27%) required additional stone-related treatment during follow-up. In particular, 35% of patients in the NCCT missing subgroup underwent further stone treatment compared to 22% of the NCCT available subgroup (p = 0.063).
Postoperative morbidity was low: Clavien–Dindo grade I–II complications occurred in eight patients (4.8%), and one patient (0.6%) experienced a grade III complication. No grade IV–V adverse events were observed.
Discussion
The present multicenter cohort study evaluated the real-world clinical performance and volumetric laser efficiency of the p-Tm: YAG laser during FURS across three tertiary referral endourology centers. Using standardized pre- and postoperative volumetric segmentation through dedicated software, we assessed ablation speed (mm3/s) and energy consumption (J/mm³) in a cohort representative of contemporary endourological practice. This study was designed as a non-comparative performance evaluation to provide complementary clinical evidence regarding the efficiency and safety profile of this emerging laser technology. By applying CT-based volumetric segmentation in a multicenter clinical setting, the present study extends this methodology beyond bench testing into routine practice. This approach enables efficiency metrics to be interpreted alongside SFR outcomes and safety, representing a key methodological strength.
In this cohort of 167 patients, p-Tm: YAG demonstrated measurable lithotripsy efficiency. Median pulse energy was 0.8 J across anatomical locations, with median pulse frequency of 10 Hz for ureteral and renal stones and 15 Hz for complex located stones. Corresponding median power ranged from 8 W for ureteral stones to 12 W for complex cases. The overall median ablation speed was 0.55 mm3/s, with lower efficiency observed in ureteral stones (0.3 mm3/s) and higher values in renal (0.52 mm3/s) and multiple–location stones (0.66 mm3/s). These values reflect the efficiency of stone ablation in routine clinical practice despite procedural variability, heterogeneous stone composition, and complex anatomical conditions.
Exploratory analyses suggested variability in ablation speed according to stone composition. Calcium oxalate stones demonstrated expected ablation behavior, whereas brushite exhibited lower efficiency consistent with its known resistance to endoscopic laser lithotripsy [20]. Differences observed for other compositions should be interpreted cautiously given limited sample sizes and the absence of multivariable adjustment.
The ablation efficiency observed in this clinical cohort is consistent with findings reported in experimental and translational studies. In vitro investigations by Quarà et al. reported p-Tm: YAG ablation rates ranging from 0.38 to 0.45 mm³/s depending on stone composition and laser settings. Specifically, ablation rates of 0.38 mm³/s at 0.5 J–20 Hz and 0.42 mm³/s at 1 J–10 Hz were observed for hard stones, while softer stones demonstrated slightly higher ablation rates under similar conditions [21].
Our study reported a median energy consumption of 13 J/mm3. In vitro, Moretto et at. demonstrated that energy consumption per unit volume is strongly influenced by both pulse energy and stone composition. At 0.5 J laser setting, the energy required to ablate calcium oxalate monohydrate stones was 30 J/mm³, compared with 76 J/mm³ for cystine, and 23 J/mm³ for uric acid stones. Increasing pulse energy improved ablation efficiency for each stone composition, reducing energy consumption to 10–16 J/mm³ at 0.8 J, and to 7 J/mm³, 8 J/mm³, and to 4–8 J/mm³ respectively at 1 J [22]. The median energy consumption of 13 J/mm³ observed in our cohort falls within this experimentally predicted range, despite the increased complexity of in vivo lithotripsy, which includes fragment displacement, endoscopic maneuvering, and heterogeneous stone composition.
Clinical series evaluating p-Tm: YAG reported comparable efficiency metrics. Panthier et al. reported ablation efficiency of 14.8 J/mm3 and ablation speed of 0.75 mm³/s, with SFR of 95% in a cohort primarily composed of renal stones. Consistently, we reported variations of energy efficiency according to anatomical location. In our cohort, the lower ablation rates were recorded for kidney and ureteral stones − 17 and 16 J/mm3, respectively- while multiple located lithiasis reported the most efficient energy dissipation with an ablation rate of 12 J/mm3. The discrepancy between isolated kidney stones and complex stone burden likely represents the variations in fragment mobility, and irrigation dynamics and stone composition. Indeed, as reflected by the similar ablation rate between these categories, complex stones frequently consist of infection-related compositions, such as struvite, which demonstrate higher susceptibility to laser energy and may account for the observed differences in lithotripsy efficiency [23].
Considering the in vivo series using the same p-Tm: YAG, Kutchukian et al. reported a median ablation speed of 0.69 mm³/s and median energy consumption 14 J/mm³, with a SFR and ZFR of 79 and 50%, respectively, for a median stone volume of 1239 mm³ [24]. Similarly, in our cohort, we observed a ZFR of 47% and a SFR of 66% for renal stones with comparable stone burden. These findings support the ability of the p-Tm: YAG to achieve effective stone clearance during FURS, with complete clearance obtained in nearly half of patients undergoing FURS. This rate may be underestimated, as some patients without postoperative NCCT demonstrate complete endoscopic clearance with visual confirmation of residual dust only [25].
Complex lithiasis demonstrated lower success rates, with Grade A, B and C SFR of 29%, 38%, and 46%, respectively. These findings suggest that stones involving multiple intrarenal locations or involving both the kidney and the ureter might present increased procedural complexity and are more likely to require staged or second-look procedures to achieve complete clearance.
The clinical importance of obtaining the ZFR has been already highlighted by Werth et al., who demonstrated that patients with residual fragments between 1 and 4 mm experienced significantly higher rates of stone-related events at six months, including complication and retreatment [26]. Our findings are consistent with this observation, with overall complications rates of 5% and high-grade complications in only 1% of patients. Although the retreatment rate in our cohort was 27%, this result has not been considered among our main outcomes. Indeed, in many cases, secondary procedures were planned as part of staged management strategies in patients with large or complex stone burdens rather than reflecting procedural failure.
The outcomes of p-Tm: YAG have been provided also against the other laser systems. In vitro comparison between p-Tm: YAG and Ho: YAG under standardized laboratory conditions have reported a significantly more efficient lithotripsy capabilities with improved ablation rate of p-Tm: YAG, in particular when increased pulse durations are applied [8, 27]. Similarly, comparable dusting performance between p-Tm: YAG and TFL systems under standardized conditions have been investigated. In vitro, no statistically significant difference in ablated stone mass was observed between p-Tm: YAG and TFL when identical pulse energy and frequency settings were applied [28]. In vivo, a clinical trial comparing TFL and p-Tm: YAG, patients treated with TFL lithotripsy presented a similar SFR, while ZFR was higher in the TFL group [29]. These results suggest how thulium-based technologies might be considered complementary rather than alternative in FURS. Evidence on the laser shape dynamic and energy interaction with different stone composition might clarify which system may be preferred in clinical practice for the treatment of different urological lithiasis conditions [30]. However, comparative clinical studies remain limited, and consistent prospective multicentric evidence of superiority among the expanding platforms available is lacking [31]. Accordingly, these findings should not be interpreted as evidence of superiority or equivalence to other laser platforms, but rather as real-world efficiency and safety data contributing to the evolving literature on thulium-based lithotripsy systems.
Some limitations should be acknowledged. First, the retrospective design introduces potential documentation variability and residual confounding. Second, procedural heterogeneity across centers reflects real-world practice but limits causal inference. Third, stone composition was available only in a subset of patients, and composition-specific analyses should therefore be considered exploratory. In addition, postoperative imaging modality and timing were not fully standardized. In 59 patients, postoperative NCCT was not available; therefore, radiological stone-free outcomes were primarily interpreted within the subgroup that underwent postoperative NCCT, which represented the principal efficacy population for stone-free assessment. Although baseline characteristics were comparable between patients with and without postoperative NCCT, suggesting that differences in follow-up were mainly related to institutional practice rather than patient characteristics, some degree of verification bias cannot be completely excluded. The rate of stone-related reintervention was 27%, which may appear relatively high; however, this figure included both unplanned retreatment and planned staged procedures in patients with large or complex stone burden. Because the database did not systematically distinguish between planned staged procedures and unplanned retreatment, this outcome should be interpreted with caution. Due to the small sample size, further substratification other than location in assessing laser efficiency and clinical outcomes for the use of UAS or FANS was not possible. Finally, the absence of a comparator arm precludes any conclusion regarding the relative performance of the p-Tm: YAG laser compared with established laser platforms.
Conclusion
This multicenter real-world study provides volumetric performance data on the use of p-Tm: YAG during FURS across different clinical scenarios. In this cohort, the platform showed measurable ablation efficiency and was associated with acceptable safety and clinically useful stone clearance outcomes. These findings support the clinical feasibility of this technology in routine practice; however, due to the retrospective design and absence of a comparator arm, these results should be interpreted as descriptive real-world data rather than evidence of superiority or equivalence. Prospective comparative studies are warranted to further define its relative effectiveness and optimal clinical role.
Supplementary Information
Below is the link to the electronic supplementary material.
Acknowledgements
Carlos Gonzalez Gonzalez is the recipient of the 2025–2026 Olympus International Scholarship, awarded by the Endourological Society in partnership with Olympus, supporting his clinical fellowship and research activities.
Abbreviations
- ASA
American society of anesthesiologists
- BMI
Body mass index
- CM3
Cubic centimeters
- CT
Computed tomography
- FANS
Flexible active negative suction
- Fr
French gauge
- FURS/fURS
Flexible ureterenoscopy
- HU
Hounsfield units
- IQR
Interquartile range
- J
Joules
- J/mm3
Energy consumption per ablated stone volume
- kJ
Kilojoules
- LOT
Laser-on time
- mm
Millimeter
- mm3
Cubic millimeter
- mm3/s
Ablation speed
- NCCT
Non-contrast computed tomography
- PCNL
Percutaneous nephrolithotomy
- PEARLS
Progressive endourological association for research and leading solutions
- p-Tm:YAG
Pulsed thulium:yttrium–aluminium–garnet laser
- RIRS
Retrograde intrarenal surgery
- SER
Stone-free rate
- TEL
Thulium fiber laser
- UAS
Ureteral access sheath
- W
Watt
- ZER
Zero-fragment rate
Author contributions
Study conception and design: Carlos González González, Jia-Lun Kwok, Laurent Berth, Alba Sierra, Frédéric Panthier. Data acquisition: Carlos González González, Jia-Lun Kwok, Alba Sierra, Federico Zorzi, Nicola Nannola, Hubert Werth, Stefano Moretto, Juan Manuel López Martínez. Data analysis and interpretation: Carlos González González, Pietro Scilipoti. Drafting of the manuscript: Carlos González González, Pietro Scilipoti, Federico Zorzi. Critical revision of the manuscript for important intellectual content: Frédéric Panthier, Pietro Scilipoti, Steeve Doizi, Olivier Traxer, Mariela Corrales, Marie Chicaud, Alba Sierra, Laurent Berth. Supervision: Frédéric Panthier.
Funding
Open access funding provided by Arts et Metiers Institute of Technology. This study did not receive any specific funding. Carlos Gonzalez Gonzalez received a personal educational and research scholarship (Olympus International Scholarship 2025–2026) from the Endourological Society and Olympus, which did not influence the study design, data collection, analysis, interpretation, or manuscript preparation.
Data availability
The data that support the findings of this study are available from the corresponding author, Carlos Gonzalez Gonzalez, upon reasonable request.
Declarations
Conflict of interest
The authors declare that they have no conflict of interest but: Olivier Traxer has declared as consultant for Karl Storz, Coloplast, IPG photonics, Quanta System and Rocamed. Steeve Doizi has declared as consultant for Boston Scientific Corporation and Coloplast. Frederic Panthier has declared as consultant of Dornier MedTech.
Ethical approval and consent to participate
The study was conducted in accordance with the Declaration of Helsinki and approved by the Comité d’Éthique et de la Recherche de l’Association Française d’Urologie (CERU; approvals CERU_2020/003 and CERU_2023018B). Data processing was declared to the Commission Nationale de l’Informatique et des Libertés (CNIL) under the MR-004 regulatory framework (approvals 2216615 v0 and 2230198 v0). Due to the retrospective nature of the study, informed consent was waived or obtained in accordance with the ethical regulations of each participating institution.
Footnotes
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Skolarikos A et al (2025) European association of urology guidelines on the diagnosis and treatment of urolithiasis. Eur Urol 88:64–75 [DOI] [PubMed] [Google Scholar]
- 2.Enikeev D et al (2023) Thulium fiber laser in endourology: current clinical evidence. Curr Opin Urol 33:95–107 [DOI] [PubMed] [Google Scholar]
- 3.Ventimiglia E et al (2021) Laser lithotripsy: the importance of peak power and pulse modulation. European Urology Focus 7:22–25 [DOI] [PubMed] [Google Scholar]
- 4.Traxer O, Sierra A, Corrales M (2022) Which is the best laser for lithotripsy? Thulium fiber laser. Eur Urol Open Sci 44:15–17 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Ventimiglia E et al (2023) What to expect from the novel pulsed Thulium:YAG laser? A systematic review of endourological applications. World J. Urol. 41:3301–3308 [DOI] [PubMed] [Google Scholar]
- 6.Panthier F et al (2023) Initial clinical experience with the pulsed solid-state Thulium YAG laser from Dornier during RIRS: first 25 cases. World J. Urol. 41:2119–2125 [DOI] [PubMed] [Google Scholar]
- 7.Hernandez-Gaytan CA et al (2025) Pulsed Thulium:YAG laser for lithotripsy. Eur Urol Focus 11:725–729 [DOI] [PubMed] [Google Scholar]
- 8.Kraft L, Petzold R, Suarez-Ibarrola R, Miernik A (2022) In vitro fragmentation performance of a novel, pulsed Thulium solid-state laser compared to a Thulium fibre laser and standard Ho:YAG laser. Lasers Med Sci 37:2071–2078 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Kwok J-L et al (2023) Pulsed Thulium:YAG laser - What is the lithotripsy ablation efficiency for stone dust from human urinary stones? Results from an in vitro PEARLS study. World J Urol 41:3723–3730 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Kallidonis P et al (2025) Pulsed Thulium: YAG laser for the management of Urolithiasis: a systematic review from the EAU section of endourology. World J Urol 43:118 [DOI] [PubMed] [Google Scholar]
- 11.Kwok J-L et al (2023) Pulsed thulium:YAG laser—ready to dust all urinary stone composition types? Results from a PEARLS analysis. World J Urol 41:2823–2831 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Doizi S, Keller EX, De Coninck V, Traxer O (2018) Dusting technique for lithotripsy: what does it mean? Nat Rev Urol 15:653–654 [DOI] [PubMed] [Google Scholar]
- 13.Fedorov A et al (2012) 3D Slicer as an image computing platform for the Quantitative Imaging Network. Magn Reson Imaging 30:1323–1341 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Panthier F, Doizi S, Berthe L, Traxer O (2019) Développement d’un outil tri-dimensionnel gratuit pour la planification opératoire des chirurgies lithiasiques. Prog Urol 29:750–751 [Google Scholar]
- 15.Panthier F et al (2021) Evaluation of a free 3D software for kidney stones’ surgical planning: ‘kidney stone calculator’ a pilot study. World J Urol 39:3607–3614 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Dindo D, Demartines N, Clavien P-A (2004) Classification of surgical complications: a new proposal with evaluation in a cohort of 6336 patients and results of a survey. Ann Surg 240:205–213 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Kwok J-L et al (2024) Laser ablation efficiency, laser ablation speed, and laser energy consumption during lithotripsy: what are they and how are they defined? A systematic review and proposal for a standardized terminology. Eur Urol Focus 10:599–611 [DOI] [PubMed] [Google Scholar]
- 18.Panthier F et al (2024) Rethinking stone-free rates and surgical outcomes in endourology: a point of view from PEARLS members. Eur Urol 86:198–199 [DOI] [PubMed] [Google Scholar]
- 19.Assel M et al (2019) Guidelines for reporting of statistics for clinical research in urology. Eur Urol 75:358–367 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Keller EX, De Coninck V, Doizi S, Daudon M, Traxer O (2021) What is the exact definition of stone dust? An in vitro evaluation. World J Urol 39:187–194 [DOI] [PubMed] [Google Scholar]
- 21.Quarà A et al (2025) In vitro ablation rates of Ho:YAG, p-Tm:YAG and TFL lasers. World J Urol 43:685 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Moretto S et al (2025) In vitro comparative analysis of ablation volume and energy requirements for 1 mm3 stone ablation using ho: YAG, p-Tm: YAG, and TFL lasers. Urolithiasis 53:194 [DOI] [PubMed] [Google Scholar]
- 23.Johnson J et al (2024) Comparative analyses and ablation efficiency of thulium fiber laser by stone composition. J Urol 211:445–454 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Kutchukian S et al (2024) Innovative use of the new pulsed-thulium: YAG laser for ureteroscopic lithotripsy: can the kidney stone calculator predict lithotripsy duration? Urolithiasis 53:14 [DOI] [PubMed] [Google Scholar]
- 25.Moretto S et al (2025) Stone dust in endourology: a systematic review of its definition, management, and clinical impact. BJU Int 136:185–197 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Werth H et al (2025) Post ureteroscopy stone status, comparing stone free rate and zero fragment rate – does the difference matter ? World J Urol 43:670 [DOI] [PubMed] [Google Scholar]
- 27.Petzold R, Miernik A, Suarez-Ibarrola R (2021) Vitro dusting performance of a new solid state Thulium laser compared to Holmium laser lithotripsy. J Endourol 35:221–225 [DOI] [PubMed] [Google Scholar]
- 28.Kraft L et al (2022) Dusting efficiency of a novel pulsed Thulium:Yttrium Aluminum Garnet laser vs a Thulium fiber laser. J Endourol 36:259–265 [DOI] [PubMed] [Google Scholar]
- 29.Panthier F et al (2024) Thulium fiber laser versus pulsed Thulium:YAG for laser lithotripsy during flexible ureteroscopy. Lasers Med Sci 39:294 [DOI] [PubMed] [Google Scholar]
- 30.Panthier F et al (2025) Laser applications in endourology: endocorporeal laser lithotripsy during ureteroscopy. French J Urol 35:102967 [DOI] [PubMed] [Google Scholar]
- 31.Yu S et al (2024) Flexible ureteroscopic treatment of kidney stones: how do the new laser systems change our concepts? Asian J Urol 11:156–168 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The data that support the findings of this study are available from the corresponding author, Carlos Gonzalez Gonzalez, upon reasonable request.