. 2026 Jan 27;131(5):748–758. doi: 10.1007/s11547-026-02173-5

Endovascular treatment of traumatic thoracic aortic injuries in patients with normal anatomy and anatomical variants: safety, efficacy and long-term follow-up

Federico Fontana 1,2, Filippo Piacentino 1,2, Federica Grimoldi 1,2, Andrea Coppola 1,3,✉, Edoardo Macchi 1, Maria Cristina Cervarolo 4, Marco Franchin 4, Anna Maria Ierardi 5, Chiara Floridi 6, Gianpaolo Carrafiello 5,7, Andrea Giovagnoni 6, Massimo Venturini 1,2
PMCID: PMC13201368  PMID: 41591721

Abstract

Purpose

To evaluate the safety and the efficacy of thoracic endovascular aortic repair (TEVAR) in patients with traumatic thoracic aortic injuries (TTAIs), analyzing the mid-long-term outcomes.

Materials and methods

In this retrospective and monocentric study, 20 patients (46 ± 17.0 years old; mean ± SD) underwent TEVAR for TTAI from February 2012 to April 2023. All patients were subjected to computed tomography angiography (CTA) before discharge; afterward, the follow-up was set up by CTA or magnetic resonance imaging (MRI). Technical success, clinical success, safety, and mid-long-term follow-up were evaluated.

Results

Both technical and clinical success were 100%. No procedure-related death was reported. Safety was 95%. In four (20%) patients the left subclavian artery (LSA) was covered; in one out of these four (25%), revascularization was necessary through carotid-subclavian bypass. In a patient with an anatomic variant of aberrant right subclavian artery (ARSA) a stent placement was required. The follow-up’s median duration was 17 ± 79.5 months (median ± IQR) and in three cases there were minimal complications: a stable type IA endoleak (EL) < 1 cm, a minimal irregularity of device’s links, and a millimeter bird beak sign.

Conclusions

TEVAR for TTAI was found to be safe (3 minimal complications) and effective both in the short and mid-long term. Patients’ adherence to follow-up and contrast-induced kidney damage remains a challenge, but the use of MRI may be a valid alternative, avoiding ionizing radiation and the use of iodinated contrast media.

Keywords: TEVAR, Traumatic thoracic aortic injury, Endovascular, Long-term follow-up, Anatomical variants, MRI, Left subclavian artery

Introduction

Traumatic thoracic aortic injuries (TTAIs) are rare conditions (annual incidence of 1,5–2%) [1, 2], but associated with high mortality [3–5]: about 77–85% of victims do not reach the hospital and mortality is 95% if TTAIs are not treated [1, 4, 6, 7]. The most frequent cause of aortic traumas are motor vehicle collisions (about 80%) [3, 6, 8–11], followed by precipitations and crushing traumas [5, 8–10]. Males are more affected than females, with a mean age from 35 to 40 years old [3, 10]. The most involved aortic segment in traumatic injury is the isthmus region (zone 3) [3, 5, 6, 9, 10, 12].

TTAIs can be distinguished into four categories, according to Azizzadeh et al. [13] classification: grade I, intimal tear; grade II, intramural hematoma; grade III, pseudoaneurysm; grade IV, rupture. This classification is also widely used to set the correct treatment course [3].

Clinical presentation is not diriment to assess correct diagnosis: hemodynamic instability is found in less than half of patients, who are often paucisymptomatic (chest pain, interscapular pain, or abdominal pain is usually referred) or asymptomatic.

So, it is crucial to suspect an aortic lesion in every high kinetic energy trauma (especially in thoracic ones); moreover, is decisive to comprehend the event’s dynamic particularly in high energy trauma [5, 6, 8–10, 14].

The first-level exam may be chest radiography [9, 10], but it remains poorly sensible and specific [3], causing an elevated number of false negatives [5, 15]. The diagnostic gold standard is computed tomography angiography (CTA) [3, 5, 9, 12], capable of recognizing, in a short time and with very high sensitivity and specificity, aortic lesions and their site; allowing a rapid treatment’choice [3, 9, 12].

Nowadays TTAIs main treatment is represented by thoracic endovascular aortic repair (TEVAR) [3, 15], a method born in the early 90 s [16, 17] and indicated by Society of Vascular Surgery guidelines since 2011 [7] (confirmed by the latest 2017 guidelines) [3, 15]. From the moment of its introduction in 1994 (Dake et al.) [16] to today, the use of TEVAR is significantly increased, at the expense of classical open repair, which is associated with high complications and mortality [4, 18].

The long-term effectiveness of TEVAR is yet to be defined [4, 5, 9].

Several issues emerged, including the size of prostheses to be placed (usually designed for senile and tendentially dilated aortas) [4], the hemodynamic changes caused by the device [19], and the possible growth of the aorta (especially in younger patients) [20].

In the literature are available numerous studies reporting the safety and efficacy of TEVAR in acute and chronic aortic disease, while to our knowledge TEVAR for TTAIs is less reported. In addition, the follow-up is poor. This retrospective, single-center study aims to define the safety and efficacy of TEVAR in TTAIs, also evaluating its mid-long-term follow-up and comparing our findings with the current literature.

Material and methods

Population

This retrospective, single-center study enrolled a cohort of consecutive patients who underwent TEVAR for TTAIs from February 2012 to April 2023.

Patients with an acute thoracic trauma, underwent clinical evaluation and whole-body CTA; once the aortic lesion was diagnosed, lesions type (II, III, IV grade) [13] and any anatomical variant, a TEVAR procedure was scheduled following a multidisciplinary team evaluation of the case, including emergency physicians, anesthesiologists, general surgeons, interventional radiologsts, and vascular surgeons.

The endoprosthesis was chosen based on the CTA findings and diagnostic angiography to ensure complete coverage of the aortic lesion and gain an adequate distal and proximal landing.

TEVAR procedure

All procedures were performed in the operating room by a team consisting of interventional radiologists, vascular surgeons, and anesthesiologists with a portable C-arm angiography system (Vision RFD; Ziehm, Nuremberg, Germany; BV 300 C-Arm; Philips Medical System, Best, the Netherlands). Patients underwent general anesthesia, with cardiorespiratory monitoring and antibiotic prophylaxis (cefazolin 2 gr or vancomycin 1 gr).

Peripheral vascular accesses (right and left common femoral arteries) were surgically isolated, according to patient anatomy and procedure necessity; in some patients, a humeral access was surgically isolated or percutaneous gained. From the femoral access, through a 5 French (F) introducer, a 0.035’’ hydrophilic guide was advanced in the ascending aorta and then exchanged with a stiff guide wire (Lunderquist Extra-Stiff, Cook Medical, Bloomington, Indiana, USA). A 22 F introducer (Dryseal, W. L. Gore & Associates, Inc., Newark, Delaware, United States) was brought into the abdominal aorta. A co-axial pigtail-shaped angiographic catheter (Supertorque MB, Cordis, Hialeah, Florida, United States) was advanced to the aortic valve. An aortography was performed to confirm the CTA findings using from 24 to 36 °C of contrast media (Visipaque 320 mg I/ml, GE HealthCare, Chicago, Illinois; about 200 ml to patient) and was used to guide the future step of the procedure. The stent grafts were chosen by oversizing approximately 10–20% relative to the native thoracic aortic diameters, and with an appropriate length to cover the aortic lesion completely. The prostheses used were the Gore TAG Conformable (C-TAG; W. L. Gore & Associates, Inc., Newark, Delaware, United States) and the Medtronic Valiant Captivia (Medtronic, Dublin, Ireland). Once the aortic lesion is identified and the endograft chosen, the stent-graft was advanced on the stiff guidewire and then released to cover the lesion.

In cases where a proximal landing zone with a length of less than 1.5–2 cm was present, the left subclavian artery ostium had to be covered and obliterated. After the endoprosthesis was placed, a second aortography was performed to assess the correct stent-graft positioning and exclude complications.

Outcomes and Follow-up

Technical success and clinical success were evaluated. Technical success was based on the correct placement and patency of the aortic endoprostheses at the end of the procedure. Clinical success was based on patient survival and correct positioning and patency of the device at 30 days.

Complications were reported as defined by the CIRSE Quality Assurance Document and Standards for Classification of Complications [21].

All patients underwent postoperative CTA before discharge and were directed to follow up by CTA control at 1, 6, and 12 months and annually or biannually thereafter; alternatively, MRI was used in case of contraindications to CTA (kidney failure or iodinated contrast medium allergy). During follow-up, aortic size, prostheses characteristics, and possible presence of complications were evaluated. Patients were contacted by telephone to ascertain the continuation of follow-up and their health status.

Statistics analysis

In this observational study, only descriptive statistics were produced. Numbers and percentages are presented for categorical variables, and mean and standard deviation (SD) are presented for normally distributed variables. Kolmogorov–Smirnov test was used to assess normality.

Results

Population

Twenty patients were recruited, including 15 men and 5 women, with a mean age of 46 ± 17.0 years old (mean ± SD; range: 19–79 years old), undergoing TEVAR for TTAIs. The general characteristics of the study population are shown in Table 1 and Table 2. Ten (50%) patients had no pathologies at the time of the trauma; considering the remaining 10 patients, seven (35%) had a metabolic syndrome, including hypertension and diabetes mellitus. The most frequent cause of injury was motor vehicle collisions (15 cases, 75%), followed by precipitation (4 cases, 20%), and crushing (one case, 5%). Six (30%) patients presented with unstable hemodynamics; in the 14 hemodynamically stable subjects, five (25%) arrived intubated, 2 (10%) were completely asymptomatic, and the remaining were clinically paucisymptomatic.

Table 1.

Summary table

No (%) Mean ± SD Range
Age 46 ± 17.0 19–79
Sex
 M 15 75
 F 5 25
Comorbidity
 CV 7 35
 Not CV 2 10
Etiology
 Motor vehicle collision 15 75
 Precipitation 4 20
 Crushing 1 5
Clinical presentation
 Thoracic-abdominal pain 7 35
 Unstable hemodynamics 6 30
 OTI 5 25
 Asymptomatic 2 10
 Paraplegia 1 5
TAI grade
 II 1 5
 III 16 80
 IV 3 15
Anatomic variants
 Bovine trunk 3 15
 ARSA 1 5
 Right-sided aortic arch 1 5
Associated lesions
 Thoracic 17 85
 Limb fractures 14 70
 Abdominal 9 45
 Head 7 35
Access
 R CIA 1 5
 R EIA 1 5
 Bilateral CFA 1 5
 L CFA 6 30
 R CFA 11 55
 HA 3 15
Aortic diameter (mm) 23.8 ± 2.7 20–31
Endoprostheses
 Gore C-TAG 19 95
 Medtronic Valiant 1 5
LSA coverage 4 20
Technical success 20 100
Peri-procedural complications 3 15
Clinical success 20 100
Follow-up complications 3 15
Follow-up (months) 41.35 ± 44.23 1–127
 1 20 100
 12 10 50
 36 8 40
 > 48 5 25

SD Standard Deviation, CV cardiovascular (pathologies related to metabolic syndrome), OTI oro-tracheal intubation, ARSA aberrant right subclavian artery, R right, L Left, CIA common iliac artery, EIA external iliac artery, CFA common femoral artery, HA humeral artery, LSA left subclavian artery

Table 2.

Summary table of the presented cases

Age, Sex Etiology TAI grade Anatomic variants LSA coverage Complications Follow-up (months)
#01 28, F MVC III Bovine trunk No / 36
#02 44, M MVC IV / No / 1
#03 63, M MVC III ARSA Yes (Periscope technique) ARSA stent stenosis 108
#04 56, M Precipitation III Right-sided aortic arch No / 107
#05 55, F MVC III / No Left upper extremity ischemia 15
#06 19, M MVC III / No / 102
#07 27, M MVC II Bovine trunk No / 94
#08 42, M MVC III / No / 74
#09 78, F MVC IV / No / 4
#10 47, M MVC III / No / 79
#11 20, F MVC III / No / 4
#12 79, M Precipitation III / Yes (Coils) Misplacement Amplatzer in CCA 1
#13 58, F Precipitation IV / No / 1
#14 54, M Precipitation III / No / 127
#15 55, M MVC III / Yes (Amplatzer Vascular Plug) / 31
#16 21, M MVC III / No / 19
#17 42, M MVC III / No / 12
#18 46, M Crushing III / Yes (Surgical) / 10
#19 36, M MVC III Bovine trunk No / 1
#20 51, M MVC III / No / 1

ARSA aberrant right subclavian artery, CCA common carotid artery, LSA left subclavian artery, MVC motor vehicle collision

The mean aortic diameter was 23.8 ± 2.7 mm. All patients presented with the aortic lesion in zone 3 (Figs. 1, 2): out of them, sixteen (80%) subjects had a grade III lesion, 3 (15%) grade IV, and 1 (5%) grade II. Anatomical variants were identified in five (25%) cases (Fig. 3, 4). Also, all patients had other injuries in association with the aortic trauma: in only 2 (10%) cases thoracic trauma occurred isolated.

Fig. 1.

Fig. 1

Grade IV aortic isthmic lesion treated with Gore C-TAG 26 × 100 mm. a b Preoperative axial and para-coronal CTA and d aortography; c postoperative para-coronal CTA and e f aortography: correct positioning and patency of endoprostheses, no complications; LSA dulling from retrograde revascularization

Fig. 2.

Fig. 2

Grade III aortic isthmic lesion treated with Gore C-TAG 34 × 150 mm with coverage and exclusion of LSA (Amplatzer Vascular Plug 12 mm, arrow). a Preoperative CTA; b postoperative CTA and c volume rendering reconstruction; d e 31-months follow-up MRI: no device-related complications

Fig. 3.

Fig. 3

Grade III aortic isthmic lesion in ARSA anatomical variant treated with Gore C-TAG 36 × 150 mm and stent (Viabahn 13 × 100 mm) in ARSA, placed according to periscope technique. a b Preoperative CTA and c volume rendering (VR) reconstruction; d e postoperative CTA and f VR: correct placement and patency of aortic endoprotestheses; g 108-months follow-up MRI: no prostheses- and stent- related complications

Fig. 4.

Fig. 4

Grade III aortic isthmic lesion in Right-sided aortic arch anatomic variant treated with Gore C-TAG 31 × 150 mm. a Preoperative CTA and b aortography; c postoperative CTA and d aortography and f volume rendering: correct positioning and patency of endoprostheses, no complications

Procedure

Peripheral vascular accesses were surgically isolated: eleven (55%) right common femoral artery, 6 (30%) left common femoral artery, 1 (5%) femoral artery bilaterally, 1 (5%) right external iliac artery, and 1 (5%) right common iliac artery; in 3 (15%) cases a humeral artery access also had to be used (1 right and 2 left).

The prostheses used were the Gore C-TAG in 19 (95%) patients, and the Medtronic Valiant in one (5%) patient only. In all patients, only one prosthetic module was placed. The prostheses placed had a mean diameter of 26.8 ± 4.6 mm (mean ± SD; range: 21–36 mm) and a length of 111.5 ± 18.4 mm (mean ± SD; range: 100–150 mm).

The ostium of LSA has been excluded in four (20%) cases: in one case with coils (Concerto Helix, Medtronic, Dublin, Ireland), in two cases with Amplatzer vascular plug (Abbott, Chicago, Illinois, United States) (12 e 16 mm) (Fig. 2), and by surgical ligation in the last case.

In a patient with ARSA, the deployment through humeral access of an 13 mm × 100 mm covered stent (Viabahn, W. L. Gore & Associates, Inc., Newark, Delaware, United States) with the periscope technique (from the ARSA origin to the distal margin of the aortic graft) was required (Fig. 3).

Outcomes

Technical success was achieved in all cases (100%, Table 1): no patient died in the operating theater; all patients were stable and with properly placed and patent grafts.

We reported 3 (15%) periprocedural complications: in one case left upper extremity ischemia occurred, 15 min after the procedure, which required a Fogarty’s thrombectomy of the humeral-ulnar-radial axis with complete resolution (CIRSE Classification System for Complications Grade I). In a patient with ARSA, stent stenosis was found on XI postoperative day, which was treated with angioplasty (CIRSE Grade III). In the third case, there was a misplacement of a 16 mm Amplatzer in LACC, which was promptly removed by a 12 mm Hooker-type catheter; subsequently, the ostium of LSA was correctly obliterated and revascularization of the subclavian axis was not necessary. A brain CT after the procedure was negative and the patient reported no sequelae (CIRSE Grade I). Thus, the safety of the procedure was 95%.

All patients were alive and the endoprostheses were patent and correctly placed in all cases at 30 days after the procedure. Clinical success was achieved in 100% of the cases, Table 1 and 2). One patient died not from complications related to the procedure but from cardiovascular arrest 41 days after the procedure.

Follow-up

Follow-up was performed by CTA (17 patients, 85%) or MRI (3 patients, 15%). The follow-up median duration was 17 ± 79.5 months (median ± IQR; range: 1–127 months), and in nine (45%) cases it was also performed by phone call.

During follow-up, the following emerged: (1) a type IA EL, which, however, appears to be stable (< 1 cm, at 43 months); (2) minimal prosthetic mesh irregularity at the distal third of the device (at 36 months); and (3) a millimeter bird beak sign (at 1 month).

At the last follow-up (including the telephone survey) all but one patient, reported well-being and confirmed that they had returned to activities of daily living. Only one patient did not return to daily living due to trauma-related complications (medullary transection and consequent paraplegia).

Discussion

Endovascular management of aortic and arterial lesions, less burdened by morbidity and mortality than surgery, is generally considered the first therapeutic option [3, 5, 7, 11, 12, 15, 17, 19, 22–26]. From the earliest studies available in the literature about TTAIs, it became apparent that TEVAR avoids some of the typical disadvantages of the open approach, such as single-lung ventilation, aortic clamping, and cardiopulmonary bypass; it also reduces surgical time and blood loss [5]. TEVAR is generally associated with fewer complications without excluding their possible occurrence [4], including death, paraplegia, renal failure, transfusions, reinterventions for bleeding, cardiac complications, pneumonia, and length of recovery [3, 10]. In the literature, there are no randomized clinical trials available that directly compare open repair with endovascular technique, given the clear superiority of TEVAR in terms of reduction of mortality and complications; consequently, it is necessary to compare the two approaches through systematic reviews or retrospective trials [3, 4, 7, 27–30] (Table 3). United States Food and Drug Administration approved the use of TEVAR for TTAIs in 2012 [3], an indication confirmed in 2015 by EAST through a meta-analysis [7], and is currently the SVS-recommended approach in patients with favorable anatomy (2017 guidelines, GRADE IC) [15].

Table 3.

TEVAR versus Open Repair (OR) for TTAIs: literature review

Author Journal Year of publication Study type No. of patients Principal results
Buz et al.[29] 2008 Retrospective 35 (OR)/39 (TEVAR) More complications and mortality in the OR group
Eur J Cardiothorac Surg TEVAR free of peri-procedural complications
TEVAR is better in terms of early outcomes
Lee et al. [4] Journal of Vascular Surgery 2011 Systemic Review 1476 (OR)/699 (TEVAR)/3574 (nonsurgical treatment) TEVAR is associated with less mortality and less incidence of peri-procedural complications compared to OR
Cheng et al. [30] JAMA Network Open 2019 Retrospective 154 (OR)/133 (TEVAR) Better long-term outcome of TEVAR compared with OR
Elkbuli et al. [27] Journal of Surgical Research 2020 Retrospective 103 (OR)/172 (TEVAR) TEVAR is much superior to OR in terms of mortality

To date, endovascular repair represents the best option for TTAIs [27, 31–33], as confirmed by our study. Open Repair is indicated in cases ineligible for endovascular treatment (mainly due to unfavorable anatomy, including the presence of iliofemoral axis disease or diameters of these vessels less than 7 mm) or in cases where TEVAR fails, requiring open conversion (rate of about 4%) [3, 10].

Preoperative planning based on CTA is crucial for choosing the correct endoprosthesis. The most important and challenging measure is the proximal landing zone: there may be anatomical variants, or the length may be under 20 mm. In these cases, LSA coverage may be indicated [3, 5, 31, 34–36]; it is reported that in trauma LSA coverage occurs in more than half of the cases [37], whereas, in our case series it was needed in only 20% of patients. There are risks related to LSA coverage [3, 37]: in 2022, Grigorian et al. [37] showed that these complications would appear to be greater in patients performing a TEVAR with LSA coverage for TTAIs, compared with those who did not undergo trauma; in our patients who underwent LSA coverage, no one reported clinical complications. If the right vertebral artery is hypoplastic or atretic, if there is an anatomic alteration of the polygon of Willis, or in the case of aortocoronary bypass with internal mammary artery, should consider LSA revascularization through traditional surgery techniques (such as transposition of LSA or carotid-subclavian bypass) or by endovascular techniques (such as chimney technique or fenestrated graft) [4, 15, 36, 37].

Preoperative planning is also important to identify the presence of any anatomic variants; these are not uncommon conditions: a review published by our group [26] found that anatomic variants were present in 15–34% of cases; in particular, aberrant right subclavian artery was described in 0.4–1.2% and right-sided aortic arch in 0.1–0.2%. In these cases, the application of TEVAR in emergencies remains controversial. In our cohort, there were five patients with anatomic variants: one (5%) aberrant right subclavian artery, one (5%) right-sided aortic arch, and three (15%) bovine trunk; all of them were treated with TEVAR, and all prostheses are still correctly positioned and patent at follow-up.

The most frequently encountered complication following TEVAR is type I endoleak [3]. Other complications include stroke and spinal cord ischemia [4], access site complications (such as infection and bleeding), endoprostheses collapse, other types of endoleak, and recurrent laryngeal nerve damage [3, 5]. Overall, morbidity ranges between 3 and 36%, while mortality ranges between 0 and 20% [5], and both increase with delayed surgical correction [3]. In our case series, no major complications were reported, and the procedure-related mortality was 0%.

The aim of follow-up in patients undergoing TEVAR for TTAIs is the early diagnosis of lesion progression and complications [3, 38]. However, currently, there is no clear guidance available about the most appropriate diagnostic test, timing, and duration with which to set up follow-up. In the absence of imaging abnormalities, some experts recommend repeating the examination 12–36 months later, while others suggest performing the examination every 2–5 years [3, 4]. Additional problems are the absence of long-term outcome data in the literature and the poor compliance of patients, who are often lost to follow-up. Patients are usually young and otherwise healthy, so they may consider it unnecessary to undergo diagnostic investigations. Another important aspect to be considered is that patients are afferent from a wide territory and follow-ups are hardly performed in the center where they undergo the procedure. Typically, CTA is used, but some authors prefer performing a chest X-ray associated with MRI [4, 12]. MRI is a viable alternative to CTA both in planning [39] and in patient follow-up [3, 5]: it has sensitivity and specificity in the diagnosis of aortic lesions comparable to CTA, and it has some advantages, such as better anatomical definition (including ulcers and intramural hematomas) and assessment of left ventricular function, without exposing the patient to contrast medium and radiation [3] (which could be an encouragement for young patients); cardiac evaluation is particularly important in long-term patient evaluation, as TEVAR for TTAIs has been shown to cause cardiac and aortic remodeling, that increased morbidity especially if not recognized and treated [19]. One of the problems with MRI lies in the long imaging period, which may not be tolerated by claustrophobic patients; it may also be contraindicated in patients with metal prostheses (a common condition in polytraumatized patients), clips, or pacemakers [3]. Our experience shows that MRI is a good alternative to CTA during follow-up in patients treated with TEVAR for TTAI.

The main limitations of this study lie in its design, as it is a retrospective, single-center, in the limited sample size, and in the poor follow-up.

Conclusions

In our experience, TEVAR for TTAI has been confirmed as a safe and effective treatment. It has achieved a level of safety and efficacy that makes it the gold standard in the treatment of TTAI having excellent both short- and mid-long-term results. TEVAR is also a valid option in patients with anatomic variants. Furthermore, our experience has highlighted the feasibility of subjecting patients to MRI during follow-up to avoid the risks associated with repeated CTA throughout their lifetime, but further studies with a larger cohort of patients and a longer follow-up are required.

Author Contributions

F.F. Conceptualization; Material preparation; Data collection and analysis Formal Analysis; Writing – Review & Editing. F.P. Conceptualization, Material preparation, Data collection and analysis; Formal Analysis; Writing – Review & Editing. F.G. Material preparation, Data collection and analysis, Writing—first draft. A.C. Material preparation, Data collection and analysis; Methodology; Software and Statistics; Formal Analysis; Writing – Review & Editing. E.M. Methodology; Software and Statistics; Formal Analysis; Writing – Review & Editing. M.C.C. Validation; Resources; Data Curation; Writing – Review & Editing. M.F. Validation; Resources; Data Curation; Writing – Review & Editing. A.M.I. Validation; Resources; Data Curation; Writing – Review & Editing. C.F. Validation; Resources; Data Curation; Writing – Review & Editing. G.C. Validation; Resources; Data Curation; Writing – Review & Editing; Supervision. A.G.Validation; Resources; Data Curation; Writing – Review & Editing; Supervision. M.V. Conceptualization; Writing – Review & Editing Supervision; Project Administration.

Funding

Open access funding provided by Università degli Studi dell'Insubria within the CRUI-CARE Agreement. The authors declare that no funds, grants, or other support were received during the preparation of this manuscript.

Declarations

Competing interest

Andrea Giovagnoni and Chiara Floridi are editors for La Radiologia Medica. The authors have no other relevant financial or non-financial interests to disclose.

Ethics approval

This is an observational study. The Insubria Research Ethics Committee has confirmed that no ethical approval is required.

Consent to participate

Informed consent was obtained from all individual participants included in the study or from the relatives in case of death.

Consent to publish

No identifying information of any participant is shown in the article.

Footnotes

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Steenburg SD, Ravenel JG (2008) Acute traumatic thoracic aortic injuries: experience with 64-MDCT. Am J Roentgenol 191:1564–1569. 10.2214/AJR.07.3349 [DOI] [PubMed] [Google Scholar]
  • 2.Ungar TC, Wolf SJ, Haukoos JS et al (2005) A clinical decision rule to exclude thoracic aortic imaging in patients with blunt chest trauma following motor vehicle collisions. Ann Emerg Med 46:110. 10.1016/j.annemergmed.2005.06.406 [DOI] [PubMed] [Google Scholar]
  • 3.Dahal R, Acharya Y, Tyroch AH, Mukherjee D (2022) Blunt thoracic aortic injury and contemporary management strategy. Angiology 73:497–507. 10.1177/00033197211052131 [DOI] [PubMed] [Google Scholar]
  • 4.Lee WA, Matsumura JS, Mitchell RS et al (2011) Endovascular repair of traumatic thoracic aortic injury: clinical practice guidelines of the Society for Vascular Surgery. J Vasc Surg 53:187–192. 10.1016/j.jvs.2010.08.027 [DOI] [PubMed] [Google Scholar]
  • 5.Steenburg SD, Ravenel JG, Ikonomidis JS et al (2008) Acute traumatic aortic injury: imaging evaluation and management. Radiology 248:748–762. 10.1148/radiol.2483071416 [DOI] [PubMed] [Google Scholar]
  • 6.Smith RS, Chang FC (1986) Traumatic rupture of the aorta: still a lethal injury. Am J Surg 152:660–663. 10.1016/0002-9610(86)90444-7 [DOI] [PubMed] [Google Scholar]
  • 7.Fox N, Schwartz D, Salazar JH et al (2015) Evaluation and management of blunt traumatic aortic injury: a practice management guideline from the Eastern Association for the Surgery of Trauma. J Trauma Nurs 22:99–110. 10.1097/JTN.0000000000000118 [DOI] [PubMed] [Google Scholar]
  • 8.Parmley LF, Mattingly TW, Manion WCMD, Jahnke EJJ (1958) Nonpenetrating traumatic injury of the aorta. Circulation 17:1086–1101 [DOI] [PubMed] [Google Scholar]
  • 9.Dionigi R, Cabitza P, Carcano G et al (2017) Chirurgia-Basi teoriche e Chirurgia generale-Chirurgia specialistica, 4th edn. Berlin [Google Scholar]
  • 10.Sidaw A, Perler B (2018) Rutherford’s vascular surgery and endovascular therapy, 9th edn. Elsevier [Google Scholar]
  • 11.Watanabe K, Fukuda I, Asari Y (2013) Management of traumatic aortic rupture. Surg Today 43:1339–1346. 10.1007/s00595-012-0471-7 [DOI] [PubMed] [Google Scholar]
  • 12.Kapoor H, Lee JT, Orr NT et al (2020) Minimal aortic injury: mechanisms, imaging manifestations, natural history, and management. Radiographics 40:1834–1847. 10.1148/rg.2020200066 [DOI] [PubMed] [Google Scholar]
  • 13.Azizzadeh A, Keyhani K, Miller CC et al (2009) Blunt traumatic aortic injury: initial experience with endovascular repair. J Vasc Surg 49:1403–1408. 10.1016/j.jvs.2009.02.234 [DOI] [PubMed] [Google Scholar]
  • 14.Bade-Boon J, Mathew JK, Fitzgerald MC, Mitra B (2018) Do patients with blunt thoracic aortic injury present to hospital with unstable vital signs? A systematic review and meta-analysis. Emerg Med J 35:231–237. 10.1136/emermed-2017-206688 [DOI] [PubMed] [Google Scholar]
  • 15.Riambau V, Böckler D, Brunkwall J et al (2017) Management of descending thoracic aorta diseases. Eur J Vasc Endovasc Surg 53:4–52. 10.1016/j.ejvs.2016.06.005 [DOI] [PubMed] [Google Scholar]
  • 16.Dake MD, Miller DC, Semba CP et al (1994) Transluminal placement of endovascular stent-grafts for the treatment of descending thoracic aortic aneurysms. N Engl J Med 331:1729–1734. 10.1056/NEJM199412293312601 [DOI] [PubMed] [Google Scholar]
  • 17.Kato N, Dake MD, Miller DC et al (1997) Traumatic thoracic aortic aneurysm: treatment with endovascular stent-grafts. Radiology 205:657–662. 10.1148/radiology.205.3.9393517 [DOI] [PubMed] [Google Scholar]
  • 18.Fabian TC, Richardson JD, Croce MA et al (1997) Prospective study of blunt aortic injury: multicenter trial of the American Association for the surgery of trauma. J Trauma 42:374–380 [DOI] [PubMed] [Google Scholar]
  • 19.Mandigers TJ, Bissacco D, Domanin M et al (2022) Cardiac and aortic modifications after endovascular repair for blunt thoracic aortic injury: a systematic review. Eur J Vasc Endovasc Surg 64:176–187. 10.1016/j.ejvs.2022.05.004 [DOI] [PubMed] [Google Scholar]
  • 20.Hartley MC, Langan EM, Cull DL et al (2009) Evaluation of the diameter of the proximal descending thoracic aorta with age: implications for thoracic aortic stent grafting. Ann Vasc Surg 23:639–644. 10.1016/j.avsg.2009.05.001 [DOI] [PubMed] [Google Scholar]
  • 21.Filippiadis DK, Binkert C, Pellerin O et al (2017) Cirse quality assurance document and standards for classification of complications: the Cirse classification system. Cardiovasc Intervent Radiol 40:1141–1146. 10.1007/s00270-017-1703-4 [DOI] [PubMed] [Google Scholar]
  • 22.Venturini M, Piacentino F, Coppola A et al (2021) Visceral artery aneurysms embolization and other interventional options: state of the art and new perspectives. J Clin Med 10:2520. 10.3390/jcm10112520 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Mosquera VX, Marini M, Lopez-Perez JM et al (2011) Role of conservative management in traumatic aortic injury: comparison of long-term results of conservative, surgical, and endovascular treatment. J Thorac Cardiovasc Surg 142:614–621. 10.1016/j.jtcvs.2010.10.044 [DOI] [PubMed] [Google Scholar]
  • 24.Arbabi CN, DuBose J, Starnes BW et al (2022) Outcomes of thoracic endovascular aortic repair in patients with concomitant blunt thoracic aortic injury and traumatic brain injury from the Aortic Trauma Foundation global registry. J Vasc Surg 75:930–938. 10.1016/j.jvs.2021.09.028 [DOI] [PubMed] [Google Scholar]
  • 25.Benk J, Siepe M, Berger T et al (2022) Early and mid-term outcomes of thoracic endovascular aortic repair to treat aortic rupture in patients with aneurysms, dissections and trauma. Interact Cardiovasc Thorac Surg 35:ivac042. 10.1093/icvts/ivac042 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Piffaretti G, Carrafiello G, Ierardi AM et al (2015) Thoracic endovascular aortic repair for blunt thoracic aortic injuries in complex aortic arch vessels anatomies. Ann Vasc Surg 29:1320.e11–15. 10.1016/j.avsg.2015.03.052 [DOI] [PubMed] [Google Scholar]
  • 27.Elkbuli A, Dowd B, Spano PJ et al (2020) Thoracic endovascular aortic repair versus open repair: analysis of the National Trauma Data Bank. J Surg Res 245:179–182. 10.1016/j.jss.2019.07.076 [DOI] [PubMed] [Google Scholar]
  • 28.Moainie SL, Neschis DG, Gammie JS et al (2008) Endovascular stenting for traumatic aortic injury: an emerging new standard of care. Ann Thorac Surg 85:1625–1630. 10.1016/j.athoracsur.2008.01.094 [DOI] [PubMed] [Google Scholar]
  • 29.Buz S, Zipfel B, Mulahasanovic S et al (2008) Conventional surgical repair and endovascular treatment of acute traumatic aortic rupture. Eur J Cardiothorac Surg 33:143–149. 10.1016/j.ejcts.2007.10.017 [DOI] [PubMed] [Google Scholar]
  • 30.Cheng Y-T, Cheng C-T, Wang S-Y et al (2019) Long-term outcomes of endovascular and open repair for traumatic thoracic aortic injury. JAMA Netw Open 2:e187861. 10.1001/jamanetworkopen.2018.7861 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Fontana F, Macchi E, Piacentino F et al (2018) The evaluation of aortic diameter changes during long-term follow-up after endovascular treatment of acute blunt traumatic thoracic aortic injuries. Vasc Endovascular Surg 52:335–343. 10.1177/1538574418765385 [DOI] [PubMed] [Google Scholar]
  • 32.Harky A, Bleetman D, Chan JSK et al (2020) A systematic review and meta analysis of endovascular versus open surgical repair for traumatic ruptured thoracic aorta. J Vasc Surg 71:270–282. 10.1016/j.jvs.2019.05.011 [DOI] [PubMed] [Google Scholar]
  • 33.Salsano A, Salsano G, Spinella G et al (2021) Endovascular versus open surgical repair for ruptured descending aortic pathologies: a systematic review and meta-analysis of observational studies. Cardiovasc Intervent Radiol 44:1709–1719. 10.1007/s00270-021-02893-3 [DOI] [PubMed] [Google Scholar]
  • 34.Rousseau H, Chabbert V, Maracher MA et al (2009) The importance of imaging assessment before endovascular repair of thoracic aorta. Eur J Vasc Endovasc Surg 38:408–421. 10.1016/j.ejvs.2009.06.017 [DOI] [PubMed] [Google Scholar]
  • 35.Criado FJ, Barnatan MF, Rizk Y et al (2002) Technical strategies to expand stent-graft applicability in the aortic arch and proximal descending thoracic aorta. J Endovasc Ther Off J Int Soc Endovasc Spec 9:32–38 [PubMed] [Google Scholar]
  • 36.Ishimaru S Endografting of the Aortic Arch. J Endovasc Ther [DOI] [PubMed]
  • 37.Grigorian A, Lewis M, Wlodarczyk JR et al (2022) Left subclavian artery coverage during endovascular repair of thoracic aorta injury in trauma and non-trauma patients. Eur J Trauma Emerg Surg 48:4425–4429. 10.1007/s00068-022-02027-5 [DOI] [PubMed] [Google Scholar]
  • 38.Zierler RE, Jordan WD, Lal BK et al (2018) The society for vascular surgery practice guidelines on follow-up after vascular surgery arterial procedures. J Vasc Surg 68:256–284. 10.1016/j.jvs.2018.04.018 [DOI] [PubMed] [Google Scholar]
  • 39.Piacentino F, Fontana F, Micieli C et al (2018) Nonenhanced MRI planning for endovascular repair of abdominal aortic aneurysms: comparison with contrast-enhanced CT angiography. Vasc Endovascular Surg 52:39–45. 10.1177/1538574417740508 [DOI] [PubMed] [Google Scholar]