Abstract
Percutaneous coronary intervention (PCI) is a common minimally-invasive procedure for treating coronary artery stenosis. However, 10% of patients with non-complex lesions experience restenosis within five years of initial PCI. Wall shear stress (WSS) is a physiological marker that provides additional predictive value for restenosis. The ability to estimate WSS during PCI could identify patients at high risk for restenosis. In order to assess the accuracy of intravascular ultrasound WSS imaging in coronary geometries, a high-frequency linear array was used to image three unique coronary phantom geometries before stenting and with partially- and fully-expanded stents. Acquired 2D WSS images were registered and compared with 3D in silico results. Finally, 3D images were acquired in a single geometry using a newly-developed intravascular ultrasound matrix array. Ultrasound-derived flow velocity maps demonstrated a mean absolute percentage error of 13.04 ± 4.82% relative to simulations, with a mean correlation of 81.91 ± 14.59%. Segments with partially-expanded stents exhibited decreased mean WSS (0.0800 ± 0.0233 Pa vs. 0.1328 ± 0.0265 Pa, p = 0.0479) and increased WSS spatial variance (0.0038 ± 0.0011 Pa2 vs. 0.00069 ± 0.000090 Pa2, p = 0.0546) compared to segments with fully expanded stents. Accurate WSS imaging during PCI could stratify restenosis risk and inform long-term coronary modeling (i.e. digital twin system).
Supplementary Information
The online version contains supplementary material available at 10.1038/s41598-026-47719-9.
Keywords: Wall shear stress, Intravascular ultrasound, Stent malapposition, Blood flow velocity, Coronary artery, Percutaneous coronary intervention optimization
Subject terms: Cardiology, Computational biology and bioinformatics, Diseases, Medical research
Introduction
In 2019, coronary artery disease (CAD) affected 124 million individuals, was responsible for 9 million deaths worldwide, and was the leading contributor to disability-adjusted life years (DALYs) with 164 million years1,2. Atherosclerosis is the primary cause of CAD and is characterized by plaque accumulation within the coronary arteries. While atherosclerosis is highly multifactorial, it is generally understood to initiate with damage to the endothelium from oxidative stress, which leads to deposition of low-density lipoproteins and in turn promotes a positive feedback cycle of leukocyte recruitment, foam cell formation, necrosis and apoptosis, and smooth muscle proliferation3,4.
Symptomatic CAD-patients often experience chest pain due to ischemia. Percutaneous coronary intervention (PCI) is the preferred treatment for relieving symptoms of ischemia, with approximately 3 million procedures performed annually5,6. PCI is preferred to coronary artery bypass graft (CABG) due to its minimally invasive nature, decreased recovery times and mortality rates, and reduced risk for post-procedural stroke7,8. The introduction of drug-eluting stents (DES) further improved treatment of CAD, reducing restenosis rates to approximately 2–10% for non-complex lesions in the first year post-PCI9. While improvements from bare metal stents to DES have greatly improved patient outcomes, it remains that approximately 10% of patients with non-complex lesions still experience in-stent restenosis or stent thrombosis within 5 years of their initial PCI10–14. For complex lesions, the risk for restenosis increases to 13–20% over a 5-year period15. Additionally, 25%−33% of patients experience recurrent angina by the 1-year follow-up and/or a major adverse cardiac event (MACE) such as death, nonfatal myocardial infarction (MI), or target lesion revascularization by the 2-year follow-up16.
Current consensus guidelines recommend the use of intracoronary imaging—optical coherence tomography (OCT) or intravascular ultrasound (IVUS)—in addition to angiography to improve patient outcomes and reduce in-stent restenosis rates10,17–19. Despite these recommendations, angiography alone is used in 83.8% of all PCI cases20. Furthermore, angiography forms 2D projection images of 3D anatomy which, limits its ability to assess stent apposition, minimal luminal area (MLA), and plaque composition21. Evaluation of these criteria using angiography alone may result in stent failure due to underexpansion, edge dissection, or geographical miss, which are all associated with elevated risk for MACE and restenosis12,21.
Current intracoronary imaging techniques only provide geometric assessment in coronary arteries. Considering imaging modalities, IVUS can characterize the plaque and stent apposition with a spatial resolution of ~ 40 μm at 40 MHz and a penetration depth of ~ 6 mm22. When PCI is guided by IVUS, long-term cardiac death and revascularization rates are significantly reduced compared to angiography-guided PCI in complex lesions12. However, current IVUS devices lack the ability to image hemodynamics because they are side viewing, and thus the ultrasound beam is orthogonal to the primary direction of blood flow.
OCT provides higher spatial resolution (~ 10 μm), thus OCT can assess the plaque and geometric apposition, however, it lacks the ability to image local hemodynamics23,24. In addition to geometric apposition assessment, the ability to image hemodynamics could reveal underlying factors that may lead to restenosis in an individual patient to stratify risk of restenosis. For example, neoatherosclerosis or neointimal hyperplasia are hemodynamic-dependent causes of restenosis25. Hemodynamic characterization could influence lesion preparation, choice of stent type and size, deployment strategy, and post-implantation assessment of stent expansion and apposition, and risk of restenosis after stent deployment26.
Wall shear stress (WSS) is an established hemodynamic marker that refers to the tangential component of the normal stress acting along the boundary in an incompressible fluid27,28. WSS is correlated also an indicator of likelihood of restenosis29–32. Elevated WSS is associated with expansive vascular remodeling and increased plaque vulnerability33–37. In the PROSPECT trial, locally decreased WSS was found to provide incremental risk stratification for non-stented lesions in high-risk patients compared to plaque burden, MLA, and morphology alone31. Therefore, understanding the impact of stents on WSS at the target lesion is important for characterizing risk of restenosis post-PCI.
Post-PCI, stents have been shown to cause a variety of changes to the local hemodynamics. The oscillatory pattern of the struts along the vessel wall generates regions of high WSS at the struts and low WSS in the regions between the struts, disrupting the local hemodynamics38. Furthermore, underexpansion of stents is associated with high WSS39, while the stent footprint alone results in low local WSS40. Finally, malapposed struts induce significant changes such as transitions from laminar to disturbed flow and the generation of recirculation zones41. Thus, the ability to image these complex WSS interactions could allow interventionalists to further optimize stent placement during PCI and improve risk stratification for patients post-PCI.
In order to estimate and spatially map WSS, both anatomical and physiological measurements are required. The current standard for local WSS estimation is to use anatomical imaging modalities including computed tomography angiography (CTA) and IVUS or OCT with wire-based measurements (e.g., FFR) as inputs to computational fluid dynamic (CFD) modeling42–44. CFD in patient-specific geometries can be highly accurate if patient-specific measurements are acquired to provide the required boundary conditions, however, it is computationally expensive45,46. Several studies have utilized CFD to investigate the hemodynamic impact of stent footprints in idealized or animal artery models47–50. Recent investigations have embedded stents in patient-specific geometries and then simulated hemodynamics based on OCT51–54. These studies have shown that the presence of the stent struts leads to lower WSS and disturbed flow patterns40,55. For example, in a longitudinal study of 12 patients, a negative correlation was shown between WSS and plaque progression55. Related studies reported similar results in the absence of clear malapposition56,57. Thus, although CFD lacks the ability to estimate WSS in real-time for PCI optimization, it is the best current method for hemodynamic assessment and may offer incremental information to inform long-term clinical decision-making, including as part of a digital-twin system post-PCI.
As an alternative to CFD, imaging modalities could be augmented to map WSS and guide clinical decision-making. Currently, WSS in coronary arteries can be estimated with imaging using phase contrast MRI (PC-MRI) or ultrasound. PC-MRI is used to estimate WSS in vivo in larger vessels like carotid arteries and the aorta, however, it is challenging to accurately estimate WSS in small, mobile arteries like coronary arteries due to the spatial and temporal resolution requirements58–61. Alternatively, ultrasound (US) has been used to estimate WSS in vivo in larger arteries like the femoral, aorta, and carotid arteries62,63. Furthermore, development of ultrasound techniques like multi-angle vector flow imaging and ultrafast plane-wave imaging has provided 2D and 3D blood flow velocity mapping at even higher frame rates in the carotid bifurcation, brachial, and femoral arteries64–66.
These prior studies demonstrating US-based WSS estimation have utilized large, non-invasive arrays that lack the spatial resolution and penetration depth needed to accurately map WSS in coronary arteries. Given that intra-coronary imaging such as IVUS is recommended for PCI guidance, if forward-viewing IVUS (FV-IVUS) could simultaneously image anatomy and blood flow velocity, then WSS imaging—an additional marker with incremental value for risk stratification—could be provided to the interventional cardiologist during the PCI procedure. This would require a FV-IVUS matrix array, which we have been developing and evaluating in our lab67,68. If WSS variation in stented arteries can be accurately estimated with FV-IVUS, it may be possible to optimize stenting and identify high-risk patients for early follow-up (e.g., stress testing or angiography in 3–6 months) or adjunct therapy like cilostazol, per prior guidelines69,70. Finally, the anatomical and physiological data might also serve as inputs to a coronary health digital twin system to predict long-term patient outcomes71,72.
The goal of this work is to demonstrate that US imaging alone can accurately map WSS in coronary artery geometries pre- and post-stenting. We present the first study of US-based WSS imaging in stented coronary geometries. WSS maps were formed (1) in physical tissue-mimicking phantoms with US and (2) in silico in the same geometries via CFD. In addition to these 2D US images acquired with a commercial high-frequency, we also present as an initial proof-of-concept the first real-time 3D US-derived images of WSS using our recently-developed FV-IVUS matrix array68.
Results
CFD and US imaging in phantom geometries before stenting
The geometries used in both physical and in silico phantoms can be seen in the B-mode (grayscale) part of the US images prior to stenting (Fig. 1). While a uniformly low WSS is observed in the case of a straight geometry (Fig. 1a), a 55% stenosis (Fig. 1b) and a patient-specific stenosis (Fig. 1c) result in increased velocity and WSS near the stenosis. The 55% stenosis (Fig. 1b), which is more severe than the patient-specific stenosis in Fig. 1c, has the highest peak WSS (0.6423 Pa).
Fig. 1.
B-mode ultrasound images with vector Doppler and wall shear stress overlays of phantoms prior to stenting for a straight vessel phantom, b stenotic phantom with 55% stenosis, and c patient specific LAD vessel phantom. The mean WSS values for the middle segment are: A 0.0943 ± 0.0042 Pa (no stenosis), B 0.2750 ± 0.1489 Pa (55% stenosis), and C 0.1536 ± 0.0304 Pa (patient-specific stenosis).
CFD after stenting
The in silico velocity magnitude maps for the stented vessels are shown in Fig. 2 for a single cross-sectional plane located along the length of each vessel. For each case, the first image indicates the location of the cross-sectional plane in contact with the stent struts, and the second image shows a magnified version of the isolated velocity map. Maximum velocity values for Fig. 2 varied between 4.78 cm/s and 12.32 cm/s and were generally located near the center lumen of each vessel. The maximum velocity values were: (A) 12.32 cm/s for a 55% stenosis with a partially-expanded stent, (B) 5.28 cm/s for a 55% stenosis with a fully-expanded stent, (C) 9.19 cm/s for a patient-specific vessel geometry with a partially expanded stent, and (D) 4.78 cm/s for a patient-specific vessel geometry with a fully-expanded stent.
Fig. 2.
Wall shear stress (WSS) surface streamlines with normalized WSS vector glyphs for four stent deployment scenarios. The color corresponds to the magnitude of the WSS, and sparse white arrows in the lower panels indicate the direction of the WSS vector at each point on the surface for: A 55% stenosis, partially expanded stent; B 55% stenosis, fully expanded stent; C patient-specific geometry, partially expanded stent; and D patient-specific geometry, fully expanded stent. Multiple surface views and corresponding cross-sectional views are shown to highlight regions of WSS disturbance, flow separation, and reattachment.
The in silico 3D WSS maps for the stented region of each vessel and a cross-sectional view of the interior lumen and stent struts are shown in Fig. 3. Peak WSS values range from 1.849 to 5.338 Pa, where the maximum values for each case were: (A) 5.338 Pa for a 55% stenosis with a partially expanded stent, (B) 3.209 Pa for a 55% stenosis with a fully-expanded stent, (C) 3.521 Pa for a patient-specific vessel geometry with a partially expanded stent, and (D) 1.849 Pa for a patient-specific vessel geometry with a fully-expanded stent. At locations of stent malapposition, the WSS on the inner surface was lower than in the surrounding lumen, often exhibiting a color pattern corresponding to the shape of the local partially expanded strut (Fig. 3C, proximal end). However, when large portions of the stent are severely malapposed towards the center of the vessel, locations of high WSS are also seen, likely due to local flow acceleration induced by the intertwining free-hanging struts (Fig. 3A and D).
Fig. 3.
3D WSS contours in the region of interest (left) and cross-sectional views of the interior lumen (right) for A 55% stenosis with a partially expanded stent, B 55% stenosis with a fully-expanded stent, C Patient-specific vessel geometry with a partially expanded stent, and D Patient-specific vessel geometry with a fully expanded stent. The direction of flow is from left to right.
US imaging after stenting
The velocity maps for all stented geometries are shown in Supplementary Fig. S2. The US-derived velocity maps are shown in the first column, followed by the corresponding 2D slice from the CFD-derived velocity maps in the second column for comparison. To the right of these two maps, the range of values is shown for both US and CFD for each segment (proximal, middle, and distal). Similarly, the WSS maps for all stented geometries are shown in Fig. 4.
Fig. 4.
CFD and ultrasound-derived WSS maps for A a partially expanded stent in a 55% stenosis over 4 mm, and B the same geometry after the stent is fully expanded. In the bottom row, the ultrasound and CFD derived WSS maps are shown for C a partially expanded stent in a patient-specific geometry, and D the same geometry after the stent is fully expanded. The mean and standard deviation of values for each segment are shown on the right side. While the comparison here shows the difference in the WSS magnitude, the direction of the WSS vectors (tangent to the surface) estimated from US imaging was always consistent with the WSS from the CFD computations.
Straight geometry
For a straight geometry with a fully expanded stent (not shown), the peak US and CFD velocities were 5.325 cm/s and 5.284 cm/s, respectively. Thus, US overestimated the peak velocity by 0.77% compared to CFD. When the velocity was analyzed by region, US underestimated the velocity compared to CFD by 11.83%, 0.25%, and 13.78% in the proximal, middle, and distal regions, respectively. The MAPE between US and CFD for this straight geometry case was 9.36%. The Spearman correlation coefficient between US-derived and CFD-derived velocities was 93.97%.
55% stenosis
Considering velocity for a 55% stenosis with a partially expanded stent (see Supplementary Fig. S2 online), the peak US and CFD velocities were 10.119 cm/s and 9.601 cm/s, respectively. Thus, US overestimated the peak velocity by 5.12% compared to CFD. US underestimated the velocity by 6.41% in the proximal region. For the middle and distal regions, US overestimated the velocity by 6.71% and 5.75%, respectively. The MAPE between US and CFD for this case was 6.29%. The Spearman correlation coefficient was 92.47%.
Considering velocity for a 55% stenosis with a fully expanded stent, the peak US and CFD velocities were 6.787 cm/s and 7.207 cm/s, respectively (Supplementary Fig. S2). Thus, US underestimated the peak velocity by 5.83% compared to CFD. In the proximal and middle regions, US underestimated the velocity compared to CFD by 37.06% and 11.77%, respectively. In the distal region, US overestimated the velocity by 4.49%. The MAPE between US and CFD for this case was 17.77%. The Spearman correlation coefficient for US and CFD-derived velocity estimations was 83.13%.
Considering the WSS for a partially expanded stent (Fig. 4A), US underestimated WSS compared to CFD by 20.38% in the proximal region and overestimated WSS by 41.30% and 129.32% in the middle and distal regions, respectively. The MAPE between US-derived and CFD-derived WSS for this case was 63.63%. The spatial variances for US-derived and CFD-derived WSS were 0.0030 Pa2 and 0.0029 Pa2, respectively.
Considering the WSS for a fully expanded stent (Fig. 4B), US underestimated the WSS compared to CFD by 54.40% and 31.28% in the proximal and middle regions, respectively. For the distal region, US overestimated the mean WSS by 3.96%. The MAPE between US-derived and CFD-derived WSS for this case was 29.88%. The spatial variances for US-derived and CFD-derived WSS were 0.000764 Pa2 and 0.0020 Pa2, respectively.
Patient-specific geometry
Considering velocity for a patient-specific geometry with a partially-expanded stent (Supplementary Fig. S2), the peak US and CFD velocities were 7.749 cm/s and 9.186 cm/s, respectively. Thus, US underestimated the peak velocity by 15.64% compared to CFD. US overestimated the mean velocity compared to CFD by 18.13%, 12.07%, and 17.30% in the proximal, middle, and distal regions, respectively. Thus, the MAPE between US and CFD estimations for this case was 15.83%. The Spearman correlation coefficient was 82.43%.
Considering velocity for a patient-specific geometry with a fully expanded stent (Supplementary Fig. S2), the peak US and CFD velocities were 4.780 cm/s and 4.782 cm/s, respectively. Thus, US underestimated the peak velocity by 0.04% compared to CFD. US overestimated the velocity compared to CFD in the proximal and middle regions by 24.44% and 2.39%, respectively. In the distal third, US underestimated the mean velocity by 20.30%. Thus, the MAPE between US and CFD velocity estimations is 15.60%. The Spearman correlation coefficient for this case was 57.56%.
Considering WSS for a patient-specific geometry with a partially-expanded stent (Fig. 4C), US underestimated the WSS compared to CFD by 43.10%, 30.33%, and 57.80% in the proximal, middle, and distal regions, respectively. The MAPE between US-derived and CFD-derived WSS was 43.74%. The spatial variance for US-derived and CFD-derived WSS in this case were 0.0039 Pa2 and 0.0072 Pa2, respectively.
Considering WSS for a patient-specific geometry with a fully expanded stent (Fig. 4D), US overestimated the mean WSS compared to CFD by 34.8% in the proximal region and underestimated the mean WSS in both the middle and distal regions by 44.43% and 52.97%, respectively. The MAPE between US-derived and CFD-derived WSS was 44.07%. The spatial variance for US-derived and CFD-derived WSS in this case were 0.00079 Pa2 and 0.0011 Pa2, respectively.
Across all cases, the fully-expanded stent conditions show a consistent mean velocity through each region. Segments containing a partially expanded stent exhibited locally decreased mean WSS (0.0800 ± 0.0233 Pa vs. 0.1328 ± 0.0265 Pa, p = 0.0479) and increased mean spatial variance (0.0038 ± 0.0011 Pa2 vs. 0.00069 ± 0.000090 Pa2, p = 0.0546) compared to the fully expanded segments in the same acquisitions. The proximal region for all partially expanded cases showed a decrease in WSS at the wall associated with the malapposition compared to its middle and distal thirds. The proximal region for all partially expanded cases showed an increase in WSS at the wall opposite the malapposition. The straight geometry with a fully expanded stent (not shown) had the highest Spearman correlation of 93.97% between US and CFD velocity maps. For all velocity maps, the MAPE was 13.04 ± 4.82% with a mean Spearman correlation of 81.91 ± 14.59%.
3D ultrasound-derived velocity and WSS estimation
The flow velocity and WSS are shown for two phantom geometries (straight and simple 50% stenosis) in Fig. 5. For the straight geometry in Fig. 5A, the peak velocity remained relatively constant through the middle of the artery and decreased towards the wall. For the stenosis in Fig. 5B, the peak velocity was located at and immediately downstream from the stenosis.
Fig. 5.
Real-time 3D ultrasound-derived velocity and WSS estimation in a straight geometry and b stenotic geometry. Peak velocity and WSS in (a) were 1.4 cm/s and 0.12 Pa, respectively. Peak velocity and WSS in (b) were 3.5 cm/s and 0.22 Pa, respectively.
Discussion
Accurate velocity profiles are important for estimating WSS; the US-derived velocity profiles and WSS estimates for the non-stented cases appear to be similar to WSS spatial distributions previously reported in literature73,74. The US-derived velocity maps for partially-expanded stents exhibited a decrease in WSS at the site of the malapposition, which is consistent with clinical studies investigating increased rate of restenosis at those locations75–77.
There are several potential sources of variation between US-derived and CFD-derived velocity and WSS maps. First, the CFD inlet boundary conditions are calibrated only on the flow rate and not on any spatially-dependent data and have an impact on the CFD solution that is only partially mitigated by the flow extensions. Thus, this improper boundary condition at the inlet would negatively impact downstream results78. This error could have a significant contribution to the variation observed between the US and CFD-derived data, particularly for the patient-specific case in which the inlet cross section is not circular. In addition, since imaging was only performed in a 2D plane, the US flow profile does not capture out-of-plane variation in flow velocities, which affects the comparison between CFD (based on 3D velocities) and 2D US reconstructions78.
Another potential source of error is the error in locating the wall of the physical phantom in the 2D US imaging data and also in the 3D segmentation of the micro-CT data. This is a well-known challenge, as small discrepancies in wall segmentation can significantly affect measurements of lumen diameter, flow velocity gradients, and therefore WSS. Furthermore, the stent generates metal artifacts in both US-based and micro-CT-based segmentations, however, this artifact is more severe in the reconstructed micro-CT volume.
Another consideration is that the geometrical complexity of the scenarios investigated here proves challenging to the numerical solvers. In this case, a consolidated commercial package (Ansys Fluent) was utilized, however, the accuracy of the solution is not fully guaranteed for these complex, stented geometries. In the future, more sophisticated mathematical tools may be needed to handle this complexity, for example, the Virtual Element Method (VEM)79,80 or a mesh-free approach such as physics-informed neural networks (PINNs)81,82. The extension of these methodologies to problems of this complexity is not trivial and will be pursued in follow-up studies.
In the present study, simulations are performed under the rigid-wall assumption and compared with phantoms having walls that are stiffer than in vivo arteries. In patient-specific computational hemodynamics, several factors—both physical and numerical—affect the overall accuracy of the results. On the physical side, geometric reconstruction and boundary conditions are considered more critical to the accuracy of results than fluid-structure interaction (FSI)83,84.
While incorporating artery stiffness via FSI would improve physical fidelity, it significantly increases the complexity of the simulations, particularly in the presence of a stent, which locally alters the mechanical properties of the vessel wall85. Conversely, Chiastra et al. compared the WSS computed for a rigid vs. compliant model in a simplified setting, concluding that the rigid assumption is adequate86. Therefore, given the geometric complexity in the presented cases, discrepancies between simulations and measurements are more likely attributable to geometric and numerical factors rather than to the rigid-wall assumption alone.
Finally, registration of the 2D US plane to the 3D CFD domain may introduce variability if there is registration error. The magnitude of the geometric mismatch is approximately 0.25 to 0.5 mm. Given the magnitude of this error, it is estimated that the produced variation in the WSS is approximately 4.17 ± 3.66%. Specifically, slight angular rotations in the plane near the transducer during registration will produce increasingly larger errors moving towards the distal end of the plane. In the partially expanded cases, the struts provide anatomical landmarks that aid registration, while the fully expanded cases have greater radial symmetry and lack unique features, which increases the difficulty of accurate registration.
In the future, this can be addressed by using a two-dimensional matrix array to acquire the flow map parallel to the array at the inlet, then applying this map to the inlet boundary conditions for the CFD simulation. For segmentation errors, methods for increasing contrast between the lumen and phantom to reduce metal artifacts are currently being explored to automate the 3D segmentation of the phantom and improve the accuracy of wall identification. Finally, additional fiducial markers placed in the phantom during phantom development could aid in registration.
The non-stented straight vessel had a nearly constant flow profile with respect to depth (Fig. 1A), as expected with steady flow. Thus, the WSS values were also approximately constant with respect to depth. The fully expanded straight case (not shown) had the highest correlation between velocity trends (93.97%). This was expected, as this was the simplest stented case tested. The discrepancies observed in the proximal and distal regions were most likely due to a combination of the sources of error previously discussed. The decrease in velocity near the distal end may be due to decreased SNR farther away from the transducer. The decrease in the US-derived velocity map at the proximal end for the fully expanded case may be due to the influence of the transducer restricting flow at the outlet. Despite these discrepancies, the velocity maps agree quantitatively, with both velocity and WSS map exhibiting similar trends.
For the 55% stenosis with a fully expanded stent (Supplementary Fig. S2, Fig. 4B), trends between velocity and WSS values are quite similar, with the exception of the proximal region for the velocity map. Again, this may be caused by the proximity between the face of the transducer and the vessel outlet. The 55% stenosis with a partially expanded stent (Supplementary Fig. S2, Fig. 4A) also had strong agreement in velocity (92.47%) and WSS trends. However, the CFD WSS contains locally-elevated WSS at certain locations. One possible reason is that the US WSS estimation technique lacked sufficient resolution or sensitivity to detect these elevated points around stent struts, leading to an underestimation of peak values. Alternatively, the peaks may have been smoothed during the US-based WSS estimation process, causing a reduction in the observed maximum WSS.
The patient-specific geometry exhibited similar trends prior to stenting (Fig. 1C) to that of the non-stented 55% stenosis geometry (Fig. 1B). For the partially expanded and fully expanded stent cases with the patient-specific geometry, the WSS trends agree between US and CFD (Fig. 4C-D). The velocity trends and values were also similar for the patient-specific partially expanded case (82.43%). The patient-specific geometry with a fully expanded stent had the lowest agreement, with a correlation coefficient of 57.56%. The spatial shift in the peak velocities in part D of Supplementary Fig. S2 could be the result of the limited field of view in micro-CT images. The 13 mm FOV may not have captured enough of the vessel tortuosity, since the patient-specific geometry contained over 20 mm of reconstructed CT-angiography data.
The trends in the 3D US-derived velocity map (Fig. 5) were as expected, with the highest velocities at the lowest MLA and immediately proximal to the MLA. Congruently, the peak WSS in Fig. 5 was located at the lowest MLA.
There are a few limitations to this work. First, continuous (non-pulsatile) flow was used in the experiments, as locally low TA-WSS has been demonstrated as an indicator that is co-localized with plaque formation and rate of progression75–77. Regardless, coronary flow is physiologically pulsatile, and the steady-flow experimental configuration represents a simplification that does not capture transient WSS dynamics present in vivo. Similarly, the physiological motion of the heart was not modeled in the in vitro flow phantom platform. These displacements in the vessel from cardiac motion can contribute to erroneous WSS estimations87. Thus, future work will investigate WSS estimation in stented coronary geometries with pulsatile flow and cardiac motion, which we have previously investigated without stenting88. While we have previously demonstrated cardiac motion compensation in order to estimate WSS in 2D88, motion compensation strategies for blood flow imaging89–91 have not been implemented in 3D.
Another limitation was the use of degassed water with microbubbles instead of blood or a blood-mimicking solution, which does not fully replicate the non-Newtonian rheological behavior of blood or the interaction between red blood cells and the vessel wall. Finally, 2D US was used to estimate the WSS but fails to capture the entire coronary environment. Utilizing our forward-viewing matrix array allows for 3D WSS estimation (Fig. 5), which is preferable because all velocity components would be used to estimate WSS, yielding a more accurate estimation27. This would also allow 3D volumetric flow measurements, which may represent an additional useful marker in this application92–94. Overall, these limitations are inherent to the in vitro flow phantom platform used in this study, which cannot fully replicate in vivo pulsatile flow, cardiac motion, or blood rheology.
To our knowledge, this is the first demonstration of using US to accurately map WSS in stented coronary artery geometries, with CFD providing validation of the accuracy of US-derived WSS. The agreement in trends between US-derived and CFD-derived velocity and WSS maps further supports the feasibility of FV-IVUS to guide PCI with the appropriate FV-IVUS array. We also presented the first real-time 3D US-derived WSS estimation in coronary geometries in phantoms by applying the described methods to the acquired 2D slices of the 3D data set (Fig. 5). Accurately estimating WSS during PCI would allow for real-time stent optimization based on local hemodynamics. Stents are currently optimized with intracoronary anatomical imaging and wire-based physiological measurements to improve patient outcomes10,19. However, FV-IVUS could provide real-time 3D imaging of anatomy and physiology to optimize stent placement, which could reduce rates of restenosis and MACE. Furthermore, FV-IVUS could improve risk stratification post-PCI to aid in refining treatment plans31. For example, patients with significantly lower WSS around the stent struts may be considered for earlier follow-up compared to standard practice.
In addition, in the future, we plan to use data assimilation techniques95,96 to combine US data and CFD simulations. Such an approach would utilize mathematically sound techniques to rigorously convert the available (but potentially limited) US data in the imaging field of view to accurate boundary information to be used in the patient-specific model.
Methods
Development of coronary artery phantoms
To evaluate the feasibility of US-based wall shear stress (WSS) estimation in stented vessels, four sets of experiments were conducted with vessel-mimicking phantoms having different geometries. A total of 11 physical phantoms were created across all 4 experiments. The first three experiments each included three phantoms of identical geometry but varying stent conditions: no stent, partially expanded stent, and fully expanded stent. US imaging-derived velocity and WSS estimates were then compared with CFD. The first phantom contained a straight 3.2 mm-diameter vessel geometry, simulating an adult proximal left anterior descending (LAD) coronary artery97. The second phantom geometry contained a stenotic vessel geometry with an eccentric stenosis of 55% in diameter over a distance of 4 mm. A reduction in diameter > 50% corresponds to a 75% reduction in the cross-sectional area, which is an established threshold for significant ischemia98–100. However, only ~ 30% of stenoses greater than 50% cause clinically significant ischemia, complicating treatment decisions101,102. The third phantom geometry contained a de-identified patient-specific geometry reconstructed from CT angiography of the LAD coronary artery. This isolated stenotic segment from de-identified patient-specific CTA had a slight curvature. For these three geometries (straight, 55% stenosis, and patient-specific), physical phantoms were developed (described below), and then 2D ultrasound images were acquired for comparison with CFD in the same geometries.
Finally, the goal of the fourth experiment was to demonstrate initial proof-of-concept with the newly-developed FV-IVUS array transducer for WSS imaging. This experimental setup was only used for 3D US imaging and utilized two physical phantoms: a straight 3 mm-diameter geometry and a 3 mm-diameter geometry with a 50% diameter concentric stenosis. Each phantom was imaged with the FV-IVUS matrix array, and 3D WSS maps were generated by estimating 2D WSS maps on a slice-by-slice basis.
All geometries were 3D printed with a resolution of 25 μm using a FormLabs Form 3 + resin printer. Silicone rubber molds (Mold Star 15 Slow, Smooth-On, Inc., Macungie, PA) were cast around the prints, then filled with carnauba wax (PremiumCraft, Minneapolis, MN), a water-soluble wax. Once solidified, a gelatin-based tissue-mimicking material was poured around the wax to form the phantom. The tissue-mimicking material consisted of 7.5% gelatin, 6% graphite for acoustic scattering (3–6 μm, Graphite Powder, Besucce, China), 5% n-propanol, and 81.5% water by volume103. An additional 3 ml of glutaraldehyde were added to 7 ml of water and then added to the phantom mixture to mimic vessel-wall acoustic properties104. After congelation, the wax was dissolved and the luminal geometry was confirmed by acquiring micro-computed tomography (micro-CT) scans (µ-CT 50, Scanco Medical AG., Zürich, Switzerland) for comparison with the original CT-angiography data.
For each geometry (straight, stenotic, and patient-specific) in the first three experiments, three phantoms were created: native, partially expanded stent, and fully expanded stent. In stented cases, a stent and balloon catheter (Medtronic Resolute Integrity RX Zotarolimus-Eluting Coronary Stent System, Medtronic, Minneapolis, MN) were positioned across the stenosis and deployed following standard techniques18,105. For the partially expanded cases, US imaging was used to confirm a malapposition of approximately 25% (0.8 mm). This ensured a malapposition greater than the 0.4 mm threshold, which requires stent adjustment or repeat revascularization due to increased risk for MACE106.
Ultrasound measurements
All phantoms in the first three experiments were submerged in degassed water for imaging with a high frequency linear array (MS400, FUJIFILM VisualSonics, Inc. Toronto, Canada) positioned 1 mm in front of the vessel outlet and aligned coaxially with the lumen (Fig. 6). The transducer was connected to a research US system (Verasonics Vantage 256, Kirkland, Washington, USA). A microbubble solution (2.5 × 106 microbubbles per ml) was continuously infused using a syringe pump (PHD 2000, Harvard Apparatus, Holliston, Massachusetts), approximating an in vivo bolus injection of Definity® microbubbles in a 75-kg person, assuming 3 L of plasma within 5 L of whole blood107. A continuous flow rate of 20 mL/min was used, approximating average LAD flow108,109. Continuous flow was utilized because time-averaged wall shear stress (TAWSS) has been shown to co-locate with locally low WSS, with both promoting plaque formation and rate of progression75–77,110–112. Computational studies confirm that TAWSS is consistent with WSS over one heartbeat96.
Fig. 6.
Diagram of the experimental setup. Vessel-mimicking phantom with a stent submerged in degassed-water. Microbubble solution is infused at 20 ml/min with a syringe pump. A commercial high-frequency linear array (MS400, VisualSonics) is positioned 1 mm from the outlet. The yellow arrow indicates US beam direction, and the red arrows indicate the flow direction. Created with BioRender.com.
US imaging was performed with three unfocused plane steered plane wave events (−8°, 0°, and + 8°) transmitted at 23.44 MHz with 2-cycle pulses. The post-compounding frame rate was 10,000 fps. 1100 frames of in-phase and quadrature (IQ) data were recorded for each acquisition, and acquisitions were performed in duplicate.
The tissue signal was first removed by applying singular value decomposition (SVD) filtering to the beamformed RF data (Fig. 7)113. If we denote by
the matrix of the images, we apply SVD to decompose
as:
![]() |
1 |
Fig. 7.
Post-processing workflow from US acquisition to velocity and wall shear stress estimation.
where
and
are orthogonal and
is diagonal. The entries of
are the singular values. The largest singular value was discarded to remove stationary echoes. Transverse oscillation was introduced in the beamformed RF data for lateral velocity estimation114–116. The filtered data was then IQ demodulated and used to estimate the flow velocity through the lumen in both axial and lateral directions. The lateral velocities were estimated with a fourth-order autocorrelation estimator, and the axial velocities were estimated with a first-order autocorrelation estimator114–116. An ensemble size of 1100 frames was used. Velocities were smoothed with MATLAB’s (MATLAB, Version 23.2, R2023b) “smooth” function using the loess method and 10% data span. Loess applies a weighted local regression with tri-cubic weights and minimizes weighted least squares error within the span, which aids in noise reduction and local adaptivity.
Wall shear stress estimation
Smoothed axial and lateral velocities were used to estimate the WSS with a pseudo-spectral approach previously developed by our group27,88. The microbubble solution was assumed to be Newtonian and non-compressible, and thus constant in viscosity. The steady Navier-Stokes equations for an incompressible fluid arises from the principles of momentum and mass conservation (e.g84., :
![]() |
2 |
where
is the velocity,
is the pressure,
is the density,
denotes the divergence operator, and the deviatoric stress tensor
is:
![]() |
3 |
where µ is the viscosity. The steady version of the Navier-Stokes equations was implemented since this is the case considered in the experiments. The tangential component of the deviatoric stress tensor is the WSS defined as:
![]() |
4 |
where
is the normal unit vector outward to the boundary. The boundary was identified as the first zero-velocity value outside of the velocity data. A 2nd degree polynomial was fit locally to the edge pixels with weighted least squares regression. This is a more robust approach to minimize the effects of clutter and noise at the wall. A Fourier sine series interpolation was used at each depth for each lateral and axial velocity profiles. The lateral velocities were smoothed with a 3% moving average prior to differentiating the function due to higher sensitivity to noise. Lateral velocity components were included because this improves WSS estimation accuracy, and a Fourier approximation is used because it is more robust to noise than polynomial fitting27.
Segmentation, meshing, and CFD set-up
Each phantom from the first three experiments was scanned with a micro-CT scanner after staining for 24 h with a 30% solution of Optiray 320 (Guerbet LLC, Princeton, NJ) to enhance lumen-phantom contrast and reduce strut artifacts. Acquisition parameters were 90 kVp, 145 µA, 750 µs integration, and 3 averaging events. Each scan yielded 271 slices at a voxel size of 49 μm.
After reconstruction, phantoms were semi-automatically segmented with Dragonfly 3D World (Comet Technologies, Wünnewil-Flamatt, Switzerland). Multi-ROI labeling differentiated between the lumen, stent, and phantom wall. The stent struts were automatically segmented with intensity thresholding (lower bound: µ ≈ 12738 cm⁻¹; upper: max intensity). The lumen-phantom interface was semi-automatically segmented with a local-Otsu filter and partial manual segmentation. The local-Otsu filter binarizes the pixels within the scope of the selection operator (0.200 mm Ø – 2 mm Ø) to either foreground or background by dynamically thresholding the local intensity histogram. Manual segmentation was used where metal artifacts interfered with the local-Otsu filter.
Finally, the stent was subtracted from the stented lumen using a Boolean operation in Meshlab117, and resulting degenerate/self-intersecting triangles were repaired with MeshLab117 and Autodesk NetFabb118. The result was re-meshed isotropically with MeshLab to obtain a surface reticulation amenable to volumetric meshing. Flow extensions, 3× the reference diameter, were added with VMTK (Vascular Modeling ToolKit)119 to morph the inlet to a circular profile, due to the lack of precise inflow velocity data. This allowed application of a Poiseuille profile at the inlet, alleviating the impact of partially arbitrary boundary conditions while maintaining the experimental flow rate of 20 ml/min. At the outflow, the traction-free condition was prescribed, i.e.
, as this has minimal impact on the numerical solution120. A final repair and re-meshing were performed before volumetric meshing with first-order tetrahedra in ICEM CFD (ANSYS ICEM 21, Ansys Inc, Canonsburg, PA, USA), tuning element size to resolve stent struts, yielding 1.0–1.4.0.4 million elements. CFD simulations were performed with a finite volume-based discretization implemented in the Ansys Fluent solver (ANSYS 21, Ansys Inc, Canonsburg, PA, USA), using a pressure-based coupled algorithm with a second-order upwind scheme for the spatial discretization, with residual criteria set to
. Each simulation required approximately 10–20 min to complete on a high-end workstation (12 core, 24 processor Intel i9-9920X @ 3.50 GHz, 128 GB Total Memory). Visualization of all CFD results were done with ParaView (Kitware Inc., Clifton Park, NY, USA).
Registration and comparison between ultrasound and CFD maps
To register the 3D CFD data with the 2D ultrasound data, the coherent point drift (CPD) and iterative closest point (ICP) algorithms were used in succession in the open-source software CloudCompare (Telecom ParisTech & Électricité de France, France). Point clouds were downsampled by 30% to improve computational efficiency. CFD data was considered the stationary source, and US data was considered the moving source. The US data was translated and rotated to minimize the distance between points within the point clouds until either 600 iterations were achieved or the convergence criteria of 10− 5 was reached. Both CPD and ICP were chosen for registration because they are robust against noise121,122. Post registration, the 3D CFD-derived WSS and velocity fields were projected onto the aligned 2D plane using nearest-neighbor interpolation in CloudCompare. The planes were then exported and compared in MATLAB 2024b (Natick, Massachusetts, United States).
2D US-derived velocity and WSS maps were compared with CFD in all cases. The mean absolute percentage error (MAPE), and Spearman correlation were calculated for each region in each velocity map.
Supplementary Information
Below is the link to the electronic supplementary material.
Acknowledgements
The authors thank Amauri Assef, D.Sc. for assistance with phantom fabrication and Verasonics, and John Oshinski, Ph.D. and William Nicholson, M.D., for helpful discussions, and thank Dr. Laxminarayanan Krishnan for technical expertise and assistance help with micro-CT imaging. This work is supported by R01EB031101 from the U.S. National Institutes of Health. Some of the work was performed at the Georgia Tech Institute for Matter and Systems, a member of the National Nanotechnology Coordinated Infrastructure (NNCI), which is supported by the National Science Foundation (ECCS-2025462). The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health or the National Science Foundation.
Author contributions
TCS and BDL conceived and designed the study. TCS collected, processed, and segmented all US and CT data. SSR designed and fabricated FV-IVUS 2D array and collected all 3D US data. AV generated meshes for CFD simulations. IS and JMT performed all CFD simulations. TCS performed analysis. TCS, BDL, AV and IS wrote the manuscript. BDL and AV secured funding and supervised the research. All authors reviewed and edited the manuscript.
Funding
The research leading to these results received funding from the U.S. National Institutes of Health under grant R01EB031101. Some of the work was performed at the Georgia Tech Institutes for Matter and Systems, a member of the National Nanotechnology Coordinated Infrastructure (NNCI), which is supported by the National Science Foundation (ECCS-2025462). The funding sources had no role in the design and conduct of the study; collection, management, analysis, and interpretation of the data; preparation, review, or approval of the manuscript; and decision to submit the manuscript for publication.
Data availability
The data acquired is available upon reasonable request to the corresponding author.
Declarations
Competing interests
The authors declare no competing interests.
Footnotes
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Khan, M. A. et al. Global epidemiology of ischemic heart disease: Results from the Global Burden of Disease Study. Cureus12(7), e9349 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Ralapanawa, U. & Sivakanesan, R. Epidemiology and the magnitude of coronary artery disease and acute coronary syndrome: A narrative review. J. Epidemiol. Glob. Health11(2), 169–177 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Bentzon, J. F. et al. Mechanisms of plaque formation and rupture. Circ. Res.114 (12), 1852–1866 (2014). [DOI] [PubMed] [Google Scholar]
- 4.Shahjehan, R. D., Sharma, S. & Bhutta, B. S. Coronary Artery Disease. In StatPearls (Treasure Island, FL, 2025). [PubMed]
- 5.Gerber, Y. et al. Coronary revascularization in the community. A population-based study, 1990 to 2004. J. Am. Coll. Cardiol.50 (13), 1223–1229 (2007). [DOI] [PubMed] [Google Scholar]
- 6.Jennings, S. et al. Trends in percutaneous coronary intervention and angiography in Ireland, 2004–2011: Implications for Ireland and Europe. Int. J. Cardiol. Heart Vessel. 4, 35–39 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Head, S. J. et al. Current practice of state-of-the-art surgical coronary revascularization. Circulation136(14), 1331–1345 (2017). [DOI] [PubMed] [Google Scholar]
- 8.Liu, H. et al. A systematic review and meta-analysis of 35,409 patients undergoing PCI versus CABG for unprotected left main coronary artery diseases. Rev. Cardiovasc. Med.25(8), 282 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Bajeu, I. T. et al. Intrastent restenosis: A comprehensive review.. Int. J. Mol. Sci.10.3390/ijms25031715 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Choi, K. H. et al. Prognostic Impact of Operator Experience and IVUS Guidance on Long-Term Clinical Outcomes After Complex PCI. JACC Cardiovasc. Interv. 16 (14), 1746–1758 (2023). [DOI] [PubMed] [Google Scholar]
- 11.Dangas, G. D. et al. In-stent restenosis in the drug-eluting stent era.. J. Am. Coll. Cardiol.56(23), 1897–907 (2010). [DOI] [PubMed] [Google Scholar]
- 12.Groenland, F. T. W. et al. Intravascular ultrasound-guided versus coronary angiography-guided percutaneous coronary intervention in patients with acute myocardial infarction: A systematic review and meta-analysis.. Int. J. Cardiol.353, 35–42 (2022). [DOI] [PubMed] [Google Scholar]
- 13.Mauri, L. et al. Long-term clinical outcomes after drug-eluting and bare-metal stenting in Massachusetts. Circulation118 (18), 1817–1827 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Zahn, R. et al. Incidence and predictors of target vessel revascularization and clinical event rates of the sirolimus-eluting coronary stent (results from the prospective multicenter German Cypher Stent Registry). Am. J. Cardiol.95(11), 1302–8 (2005). [DOI] [PubMed] [Google Scholar]
- 15.Xu, N. et al. Five-year outcomes of biodegradable versus second-generation durable polymer drug-eluting stents used in complex percutaneous coronary intervention. Chin. Med. J. (Engl.)136(3), 322–330 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Ding, D. et al. Immediate post-procedural functional assessment of percutaneous coronary intervention: Current evidence and future directions. Eur. Heart J.42(27), 2695–2707 (2021). [DOI] [PubMed] [Google Scholar]
- 17.Byrne, R. A. et al. 2023 ESC Guidelines for the management of acute coronary syndromes. Eur. Heart J.12 (38), 3720–3826 (2023). [DOI] [PubMed] [Google Scholar]
- 18.Lawton, J. S. et al. 2021 ACC/AHA/SCAI guideline for coronary artery revascularization: A report of the American College of Cardiology/American Heart Association Joint Committee on clinical practice guidelines. Circulation145(3), e18–e114 (2022). [DOI] [PubMed] [Google Scholar]
- 19.Park, H. et al. Optimal stenting technique for complex coronary lesions: intracoronary imaging-guided pre-dilation, stent sizing, and post-dilation. JACC Cardiovasc. Interv13(12), 1403–1413 (2020). [DOI] [PubMed] [Google Scholar]
- 20.Jones, D. A. et al. Angiography alone versus angiography plus optical coherence tomography to guide percutaneous coronary intervention: outcomes from the Pan-London PCI cohort. JACC Cardiovasc. Interv11(14), 1313–1321 (2018). [DOI] [PubMed] [Google Scholar]
- 21.Mintz, G. S. Clinical utility of intravascular imaging and physiology in coronary artery disease. J. Am. Coll. Cardiol.64 (2), 207–222 (2014). [DOI] [PubMed] [Google Scholar]
- 22.Maehara, A. et al. IVUS-guided versus OCT-guided coronary stent implantation: A critical appraisal. JACC Cardiovasc. Imaging10(12), 1487–1503 (2017). [DOI] [PubMed] [Google Scholar]
- 23.De Bruyne, B. et al. Fractional flow reserve-guided PCI for stable coronary artery disease. N Engl. J. Med.371 (13), 1208–1217 (2014). [DOI] [PubMed] [Google Scholar]
- 24.van Zandvoort, L. J. C. et al. Improving PCI outcomes using postprocedural physiology and intravascular imaging. JACC Cardiovasc. Interv14(22), 2415–2430 (2021). [DOI] [PubMed] [Google Scholar]
- 25.Ng, J. et al. Local hemodynamic forces after stenting implications on restenosis and thrombosis. Arterioscler. Thromb. Vascular Biology37(12), 2231–2242 (2017). [DOI] [PubMed] [Google Scholar]
- 26.Truesdell, A. G. et al. Intravascular imaging during percutaneous coronary intervention: JACC state-of-the-art review. J. Am. Coll. Cardiol.81(6), 590–605 (2023). [DOI] [PubMed] [Google Scholar]
- 27.Tempestti, M. A pseudo-spectral method for wall shear stress estimation from Doppler ultrasound imaging in coronary arteries. Cardiovasc. Eng. Technol.15(6), 647–666 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Katritsis, D. et al. Wall shear stress: theoretical considerations and methods of measurement. Prog Cardiovasc. Dis.49 (5), 307–329 (2007). [DOI] [PubMed] [Google Scholar]
- 29.Candreva, A. et al. Impact of endothelial shear stress on coronary atherosclerotic plaque progression and composition: A meta-analysis and systematic review. Int. J. Cardiol.407, 132061 (2024). [DOI] [PubMed] [Google Scholar]
- 30.Papafaklis, M. I. et al. Effect of the local hemodynamic environment on the de novo development and progression of eccentric coronary atherosclerosis in humans: insights from PREDICTION. Atherosclerosis240 (1), 205–211 (2015). [DOI] [PubMed] [Google Scholar]
- 31.Stone, P. H. et al. Role of low endothelial shear stress and plaque characteristics in the prediction of nonculprit major adverse cardiac events: The PROSPECT study. JACC Cardiovasc. Imaging11(3), 462–471 (2018). [DOI] [PubMed] [Google Scholar]
- 32.Chiastra, C. et al. Coronary artery stenting affects wall shear stress topological skeleton.. J. Biomech. Eng.10.1115/1.4053503 (2022). [DOI] [PubMed] [Google Scholar]
- 33.Candreva, A. et al. Influence of intracoronary hemodynamic forces on atherosclerotic plaque phenotypes. Int. J. Cardiol.399, 131668 (2024). [DOI] [PubMed] [Google Scholar]
- 34.Cecchi, E. et al. Role of hemodynamic shear stress in cardiovascular disease. Atherosclerosis214 (2), 249–256 (2011). [DOI] [PubMed] [Google Scholar]
- 35.Dolan, J. M., Kolega, J. & Meng, H. High wall shear stress and spatial gradients in vascular pathology: a review. Ann. Biomed. Eng.41 (7), 1411–1427 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Eshtehardi, P. et al. High wall shear stress and high-risk plaque: an emerging concept. Int. J. Cardiovasc. Imaging. 33 (7), 1089–1099 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Salmasi, M. Y. et al. High wall shear stress can predict wall degradation in ascending aortic aneurysms: An integrated biomechanics study. Front. Bioeng. Biotechnol.9, 750656 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Elliott, M. C., Blair, J. & Menary, R. Highlighting hemodynamic risks for bioresorbable stents in coronary arteries. Fluids8(9), 241 (2023). [Google Scholar]
- 39.Kumar, S. et al. Stent underexpansion is associated with high wall shear stress: a biomechanical analysis of the shear stent study. Int. J. Cardiovasc. Imaging. 39 (7), 1375–1382 (2023). [DOI] [PubMed] [Google Scholar]
- 40.Shah, I. et al. Impact of the stent footprint on endothelial wall shear stress in patient-specific coronary arteries: A computational analysis from the SHEAR-STENT trial. Comput. Methods Programs Biomed.266, 108762 (2025). [DOI] [PubMed] [Google Scholar]
- 41.Wu, W. et al. Hemodynamic microenvironment of coronary stent strut malapposition. Comput. Biol. Med.184, 109378 (2025). [DOI] [PubMed] [Google Scholar]
- 42.Carrizo, S. et al. Functional assessment of coronary artery disease by intravascular ultrasound and computational fluid dynamics simulation. Rev. Port Cardiol.33 (10), 645e1–645e4 (2014). [DOI] [PubMed] [Google Scholar]
- 43.Sun, Z. & Xu, L. Computational fluid dynamics in coronary artery disease. Comput. Med. Imaging Graph.38(8), 651–663 (2014). [DOI] [PubMed] [Google Scholar]
- 44.Zhang, J. M. et al. Perspective on CFD studies of coronary artery disease lesions and hemodynamics: a review. Int. J. Numer. Method Biomed. Eng.30 (6), 659–680 (2014). [DOI] [PubMed] [Google Scholar]
- 45.Candreva, A. et al. Current and future applications of computational fluid dynamics in coronary artery disease. Rev. Cardiovasc. Med.23(11), 377 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Morris, P. D. et al. Virtual fractional flow reserve from coronary angiography: Modeling the significance of coronary lesions: Results from the VIRTU-1 (VIRTUal Fractional Flow Reserve From Coronary Angiography) study.. JACC Cardiovasc. Interv.6(2), 149–57 (2013). [DOI] [PubMed] [Google Scholar]
- 47.Balossino, R. et al. Effects of different stent designs on local hemodynamics in stented arteries. J. Biomech.41 (5), 1053–1061 (2008). [DOI] [PubMed] [Google Scholar]
- 48.Beier, S. et al. Hemodynamics in idealized stented coronary arteries: Important stent design considerations. Ann. Biomed. Eng.44(2), 315–329 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Duraiswamy, N., Schoephoerster, R. T. & Moore, J. E. Jr. Comparison of near-wall hemodynamic parameters in stented artery models. J. Biomech. Eng.131 (6), 061006 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.LaDisa, J. F. Jr. et al. Stent design properties and deployment ratio influence indexes of wall shear stress: a three-dimensional computational fluid dynamics investigation within a normal artery. J. Appl. Physiol. (1985). 97 (1), 424–430 (2004). discussion 416. [DOI] [PubMed] [Google Scholar]
- 51.Beyene, S. et al. Comparison of endothelial shear stress between ultrathin strut bioresorbable polymer drug-eluting stent vs durable-polymer drug-eluting stent post-stent implantation: An optical coherence tomography substudy from BIOFLOW II. Cardiovasc. Revasc Med.61, 26–34 (2024). [DOI] [PubMed] [Google Scholar]
- 52.Chiastra, C. et al. Patient-specific modeling of stented coronary arteries reconstructed from optical coherence tomography: Towards a widespread clinical use of fluid dynamics analyses. J. Cardiovasc. Transl Res.11(2), 156–172 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Migliori, S. et al. Application of an OCT-based 3D reconstruction framework to the hemodynamic assessment of an ulcerated coronary artery plaque. Med. Eng. Phys.78, 74–81 (2020). [DOI] [PubMed] [Google Scholar]
- 54.Wu, W. et al. Three dimensional reconstruction of coronary artery stents from optical coherence tomography: experimental validation and clinical feasibility. Sci. Rep.11 (1), 12252 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Bourantas, C. V. et al. Effect of the endothelial shear stress patterns on neointimal proliferation following drug-eluting bioresorbable vascular scaffold implantation: an optical coherence tomography study. JACC Cardiovasc. Interv. 7 (3), 315–324 (2014). [DOI] [PubMed] [Google Scholar]
- 56.Beyene, S. et al. Comparison of endothelial shear stress between ultrathin strut bioresorbable polymer drug-eluting stent vs durable-polymer drug-eluting stent post-stent implantation: An optical coherence tomography substudy from BIOFLOW II. Cardiovasc. Revascularization Med.61, 26–34 (2024). [DOI] [PubMed] [Google Scholar]
- 57.Tenekecioglu, E. et al. Early strut protrusion and late neointima thickness in the Absorb bioresorbable scaffold: A serial wall shear stress analysis up to five years. EuroIntervention15(4), e370–e379 (2019). [DOI] [PubMed] [Google Scholar]
- 58.Cheng, C. P., Parker, D. & Taylor, C. A. Quantification of wall shear stress in large blood vessels using Lagrangian interpolation functions with cine phase-contrast magnetic resonance imaging. Ann. Biomed. Eng.30 (8), 1020–1032 (2002). [DOI] [PubMed] [Google Scholar]
- 59.Fonken, J. et al. The impact of a limited field-of-view on computed hemodynamics in abdominal aortic aneurysms: evaluating the feasibility of completing ultrasound segmentations with parametric geometries. Ann. Biomed. Eng.51(6), 1296–1309 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Johnson, K., Sharma, P. & Oshinski, J. Coronary artery flow measurement using navigator echo gated phase contrast magnetic resonance velocity mapping at 3.0 T. J. Biomech.41 (3), 595–602 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Pantos, I. et al. In vivo wall shear stress measurements using phase-contrast MRI. Expert Rev. Cardiovasc. Ther.5 (5), 927–938 (2007). [DOI] [PubMed] [Google Scholar]
- 62.Wang, I. C. et al. Wall shear stress mapping for human femoral artery based on ultrafast ultrasound vector Doppler estimations. Med. Phys.48 (11), 6755–6764 (2021). [DOI] [PubMed] [Google Scholar]
- 63.Xiaoyong, T. Y., Wei, C., Juan, H., Feng, C. & Zhuo, Q. L. Evaluation efficacy wall shear stress carotid artery stenting. Heliyon 10(11) (2024). [DOI] [PMC free article] [PubMed]
- 64.Aizawa, K. et al. Brachial artery vasodilatory response and wall shear rate determined by multigate Doppler in a healthy young cohort. J. Appl. Physiol. (1985). 124 (1), 150–159 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Huang, Y. H. et al. Estimation of mouse carotid arterial wall shear stress using high-frequency ultrasound imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control70(6), 474–485 (2023). [DOI] [PubMed] [Google Scholar]
- 66.Yiu, B. Y. & Yu, A. C. Least-squares multi-angle Doppler estimators for plane-wave vector flow imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control63(11), 1733–1744 (2016). [DOI] [PubMed] [Google Scholar]
- 67.Lindsey, B. D. et al. 3-D intravascular characterization of blood flow velocity fields with a forward-viewing 2-D array. Ultrasound Med. Biol.46(9), 2560–2571 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Rojas, S. S. et al. High-frequency, 2-mm-diameter forward-viewing 2-D array for 3-D Intracoronary blood flow imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control71(8), 1051–1061 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Douglas, J. S. Jr. et al. Coronary stent restenosis in patients treated with cilostazol. Circulation112 (18), 2826–2832 (2005). [DOI] [PubMed] [Google Scholar]
- 70.Holmes, D. R. Jr. et al. ACCF/AHA clopidogrel clinical alert: approaches to the FDA boxed warning: a report of the American College of Cardiology Foundation Task Force on clinical expert consensus documents and the American Heart Association endorsed by the Society for Cardiovascular Angiography and Interventions and the Society of Thoracic Surgeons. J. Am. Coll. Cardiol.56 (4), 321–341 (2010). [DOI] [PubMed] [Google Scholar]
- 71.Dziopa, K. et al. Digital twins: reimagining the future of cardiovascular risk prediction and personalised care. Hellenic J. Cardiol.81, 4–8 (2025). [DOI] [PubMed] [Google Scholar]
- 72.Sel, K. et al. Building digital twins for cardiovascular health: From principles to clinical impact. J. Am. Heart Assoc.13(19), e031981 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Akhtar, S. et al. CFD analysis on blood flow inside a symmetric stenosed artery: Physiology of a coronary artery disease. Sci. Prog. 106 (2), 368504231180092 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Timofeeva, M. et al. Numerical simulation of the blood flow through the coronary artery stenosis: Effects of varying eccentricity. Comput. Biol. Med.146, 105672 (2022). [DOI] [PubMed] [Google Scholar]
- 75.Eslami, P. et al. Effect of wall elasticity on hemodynamics and wall shear stress in patient-specific simulations in the coronary arteries. J. Biomech. Eng.142(2), 0245031–02450310 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76.Samady, H. et al. Coronary artery wall shear stress is associated with progression and transformation of atherosclerotic plaque and arterial remodeling in patients with coronary artery disease. Circulation124(7), 779–788 (2011). [DOI] [PubMed] [Google Scholar]
- 77.Suo, J., Oshinski, J. N. & Giddens, D. P. Blood flow patterns in the proximal human coronary arteries: Relationship to atherosclerotic plaque occurrence. Mol. Cell. Biomech.5(1), 9–18 (2008). [PubMed] [Google Scholar]
- 78.Veneziani, A. & Vergara, C. Flow rate defective boundary conditions in haemodynamics simulations. Int. J. Numer. Methods Fluids. 47 (8–9), 803–816 (2005). [Google Scholar]
- 79.Da Veiga, L. B. et al. The virtual element method. Acta Numerica. 32, 123–202 (2023). [Google Scholar]
- 80.da Veiga, L. B., Mora, D. & Vacca, G. The stokes complex for virtual elements with application to navier-stokes flows. J. Sci. Comput.81(2), 990–1018 (2019). [Google Scholar]
- 81.Jin, X. W. et al. NSFnets (Navier-Stokes flow nets): Physics-informed neural networks for the incompressible Navier-Stokes equations.. J. Comput. Phys.10.1016/j.jcp.2020.109951 (2021). [Google Scholar]
- 82.Oldenburg, J. et al. Geometry aware physics informed neural network surrogate for solving Navier-Stokes equation (GAPINN).. Adv. Model. Simul. Eng. Sci.10.1186/s40323-022-00221-z (2022). [Google Scholar]
- 83.Taylor, C. A. & Figueroa, C. A. Patient-specific modeling of cardiovascular mechanics. Annu. Rev. Biomed. Eng.11, 109–134 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84.Formaggia, L., Quarteroni, A. & Veneziani, A. Cardiovascular Mathematics: Modeling and simulation of the circulatory system Vol. 1 (Springer Science & Business Media, 2010). [Google Scholar]
- 85.Formaggia, L., Lamponi, D. & Quarteroni, A. One-dimensional models for blood flow in arteries. J. Eng. Math.47 (3–4), 251–276 (2003). [Google Scholar]
- 86.Chiastra, C. et al. On the necessity of modelling fluid-structure interaction for stented coronary arteries. J. Mech. Behav. Biomed. Mater.34, 217–230 (2014). [DOI] [PubMed] [Google Scholar]
- 87.Zeng, D. et al. Effects of cardiac motion on right coronary artery hemodynamics. Ann. Biomed. Eng.31 (4), 420–429 (2003). [DOI] [PubMed] [Google Scholar]
- 88.Kim, S. et al. Dynamic coronary blood flow velocity and wall shear stress estimation using ultrasound in an ex vivo porcine heart. Cardiovasc. Eng. Technol.15(1), 65–76 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Tierney, J. et al. Adaptive clutter demodulation for non-contrast ultrasound perfusion imaging. IEEE Trans. Med. Imaging36(9), 1979–1991 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Tierney, J. et al. Independent component-based spatiotemporal clutter filtering for slow flow ultrasound. IEEE Trans. Med. Imaging39(5), 1472–1482 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 91.Ozgun, K. A. & Byram, B. C. Multidimensional clutter filtering of aperture domain data for improved blood flow sensitivity. IEEE Trans. Ultrason. Ferroelectr. Freq. Control68(8), 2645–2656 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 92.Kripfgans, O. D. et al. Three-dimensional US for quantification of volumetric blood flow: multisite multisystem results from within the quantitative imaging biomarkers alliance. Radiology296(3), 662–670 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 93.Pinter, S. Z. et al. Color flow ultrasound spatial sampling beam density for partial volume-corrected three-dimensional volume flow (3DVF): Theory, simulation, and experiment. Ultrasound Med. Biol.50(8), 1122–1133 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94.Welsh, A. W. et al. Three-dimensional US fractional moving blood volume: Validation of renal perfusion quantification. Radiology293(2), 460–468 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 95.Lefieux, A. et al. Semi-automatic reconstruction of patient-specific stented coronaries based on data assimilation and computer aided design. Cardiovasc. Eng. Technol.13(4), 517–534 (2022). [DOI] [PubMed] [Google Scholar]
- 96.Nannini, G. et al. An automated and time-efficient framework for simulation of coronary blood flow under steady and pulsatile conditions. Comput. Methods Programs Biomed.257, 108415 (2024). [DOI] [PubMed] [Google Scholar]
- 97.Dodge, J. T. Jr. et al. Lumen diameter of normal human coronary arteries. Influence of age, sex, anatomic variation, and left ventricular hypertrophy or dilation. Circulation86 (1), 232–246 (1992). [DOI] [PubMed] [Google Scholar]
- 98.Adjedj, J. et al. Visual and quantitative assessment of coronary stenoses at angiography versus fractional flow reserve: The impact of risk factors.. Circ. Cardiovasc. Imaging10.1161/CIRCIMAGING.117.006243 (2017). [DOI] [PubMed] [Google Scholar]
- 99.Harris, P. J. et al. The prognostic significance of 50% coronary stenosis in medically treated patients with coronary artery disease. Circulation62 (2), 240–248 (1980). [DOI] [PubMed] [Google Scholar]
- 100.Yadalam, P. K. et al. Advanced machine learning for estimating vascular occlusion percentage in patients with ischemic heart disease and periodontitis. Int. J. Cardiol. Cardiovasc. Risk Prev.21, 200291 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 101.Rosenthal, R. L. The 50% coronary stenosis. Am. J. Cardiol.115 (8), 1162–1165 (2015). [DOI] [PubMed] [Google Scholar]
- 102.Tonino, P. A. et al. Angiographic versus functional severity of coronary artery stenoses in the FAME study fractional flow reserve versus angiography in multivessel evaluation. J. Am. Coll. Cardiol.55 (25), 2816–2821 (2010). [DOI] [PubMed] [Google Scholar]
- 103.Madsen, E. L. et al. Tissue mimicking materials for ultrasound phantoms. Med. Phys.5 (5), 391–394 (1978). [DOI] [PubMed] [Google Scholar]
- 104.Telichko, A. V., Dahl, J. J. & Herickhoff, C. D. Cylindrical transducer array for intravascular shear wave elasticity imaging: Preliminary development. IEEE Trans. Ultrason. Ferroelectr. Freq. Control69(3), 1077–1087 (2022). [DOI] [PubMed] [Google Scholar]
- 105.Neumann, F. J. et al. 2018 ESC/EACTS Guidelines on myocardial revascularization. Eur. Heart J.40 (2), 87–165 (2019). [DOI] [PubMed] [Google Scholar]
- 106.Ng, J. C. K. et al. Stent malapposition generates stent thrombosis: Insights from a thrombosis model. Int. J. Cardiol.353, 43–45 (2022). [DOI] [PubMed] [Google Scholar]
- 107.Lindsey, B. D. et al. Acoustic characterization of contrast-to-tissue ratio and axial resolution for dual-frequency contrast-specific acoustic angiography imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 61 (10), 1668–1687 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 108.de Bruyne, B. et al. Simultaneous coronary pressure and flow velocity measurements in humans. Feasibility, reproducibility, and hemodynamic dependence of coronary flow velocity reserve, hyperemic flow versus pressure slope index, and fractional flow reserve. Circulation94 (8), 1842–1849 (1996). [DOI] [PubMed] [Google Scholar]
- 109.Hozumi, T. et al. Noninvasive assessment of coronary flow velocity and coronary flow velocity reserve in the left anterior descending coronary artery by Doppler echocardiography: comparison with invasive technique. J. Am. Coll. Cardiol.32 (5), 1251–1259 (1998). [DOI] [PubMed] [Google Scholar]
- 110.Kumar, A. et al. High coronary shear stress in patients with coronary artery disease predicts myocardial infarction. J. Am. Coll. Cardiol.72(16), 1926–1935 (2018). [DOI] [PubMed] [Google Scholar]
- 111.Kumar, A. et al. Low coronary wall shear stress is associated with severe endothelial dysfunction in patients with nonobstructive coronary artery disease. JACC: Cardiovasc. Interventions. 11 (20), 2072–2080 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 112.Tufaro, V. et al. Can fast wall shear stress computation predict adverse cardiac events in patients with intermediate non-flow limiting stenoses? Atherosclerosis401, 119099 (2025). [DOI] [PubMed] [Google Scholar]
- 113.Demene, C. et al. Spatiotemporal clutter filtering of Ultrafast ultrasound data highly increases Doppler and fUltrasound sensitivity. IEEE Trans. Med. Imaging34(11), 2271–2285 (2015). [DOI] [PubMed] [Google Scholar]
- 114.Jensen, J., Stuart, M. B. & Jensen, J. A. High frame rate vector velocity estimation using plane waves and transverse oscillation. In IEEE International Ultrasonics Symposium (IUS), 1–4 (2015).
- 115.Jing, B. et al. A transverse velocity spectral estimation method for ultrafast ultrasound Doppler imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control70(12), 1749–1760 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 116.Pihl, M. J. & Jensen, J. A. A transverse oscillation approach for estimation of three-dimensional velocity vectors, part I: concept and simulation study. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 61 (10), 1599–1607 (2014). [DOI] [PubMed] [Google Scholar]
- 117.Cignoni, P. et al. Meshlab: an open-source mesh processing tool. In Eurographics Italian chapter conference (Salerno, 2008).
- 118.Autodesk About Autodesk® Netfabb® (2022).
- 119.Izzo, R. et al. The vascular modeling toolkit: a python library for the analysis of tubular structures in medical images. J. Open. Source Softw.3 (25), 745 (2018). [Google Scholar]
- 120.Heywood, J. G., Rannacher, R. & Turek, S. Artificial boundaries and flux and pressure conditions for the incompressible Navier–Stokes equations. Int. J. Numer. Methods Fluids. 22 (5), 325–352 (1996). [Google Scholar]
- 121.Myronenko, A. & Song, X. Point set registration: Coherent point drift. IEEE Trans. Pattern Anal. Mach. Intell.32(12), 2262–2275 (2010). [DOI] [PubMed] [Google Scholar]
- 122.Zhang, J., Yao, Y. & Deng, B. Fast and robust iterative closest point. IEEE Trans. Pattern Anal. Mach. Intell.44(7), 3450–3466 (2022). [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The data acquired is available upon reasonable request to the corresponding author.